Inficon Stripe CDG045Dhs Operating Manual

Capacitance Diaphragm Gauge
Stripe™ CDG045Dhs
tina84e1-a (2019-07) 1
Operating Manual
Incl. EU Declaration of Conformity
Model:
PN: SN:
V W

Product Identification

In all communications with INFICON, please specify the infor­mation given on the product nameplate. For convenient refer­ence copy that information into the space provided below.

Validity

This document applies to products of the Stripe CDG045Dhs series.
Part numbers of standard products are indicated below. OEM products have other part numbers and different parameter settings (e.g. factory setting of setpoint) as defined in the corresponding ordering information.
2 tina84e1-a (2019-07)
The part number (PN) can be taken from the product nameplate. If not indicated otherwise in the legends, the illustrations in this
document correspond to gauges with DN 16 ISO-KF vacuum connection. They apply to gauges with other vacuum connec­tions by analogy.
We reserve the right to make technical changes without prior notice.
tina84e1-a (2019-07) 3

Intended Use

The temperature compensated Capacitance Diaphragm Gauges of the Stripe CDG045Dhs series are intended for absolute pressure measurement of gases in their respective pressure ranges ( 2).
Ideally the measurement values can be read out digitally via EtherCAT interface or analog. The gauges can also be operated in connection with an INFICON Vacuum Gauge Controller (VGC series) or another appropriate controller.

Functional Principle

A ceramic diaphragm is deflected by pressure. The deflection is measured capacitively and converted into a digital or into an analog linear output signal by the digital electronics. The digital output signal can only be read out via the EthreCAT interface.
The output signal is independent of the gas type. Very accurate pressure measurement is achieved by heating the
sensor to a constant temperature of 45°C which results in a com­pensation of changes in the ambient conditions and a reduced deposition of process products and by-products in process appli­cations.

Trademarks

Stripe™ INFICON Gmb H
®
VCR
Swagelok Marketing Co.

Patents

EP 1070239 B1, 1040333 B1 US Patents 6528008, 6591687, 7107855, 7140085
4 tina84e1-a (2019-07)

Scope of Delivery

1× gauge Stripe CDG045Dhs 1× pin for adjusting settings via buttons 1× Calibration Test Report 1× Operating Manual German 1× Operating Manual English
tina84e1-a (2019-07) 5
Contents
Product Identification 2 Validity 2 Intended Use 4 Functional Principle 4 Trademarks 4 Patents 4 Scope of Delivery 5
1 Safety 8
1.1 Symbols Used 8
1.2 Personnel Qualifications 8
1.3 General Safety Instructions 9
1.4 Liability and W arranty 9
2 Technical Data 10 3 Installation 16
3.1 Vacuum Connection 16
3.2 Power Connection 19
3.2.1 D-sub, 15-pin Connector 20
3.2.2 EtherCAT Connector 21
3.2.3 Mini USB Type B Connector (Diagnostic Port) 22
4 Operation 23
4.1 Status Indication 23
4.2 Zeroing the Gauge 24
4.2.1 <ZERO> Adjustment 24
4.3 Switching Functions SP1, SP2, ATM, Status 28
4.4 Activating the Factory Setting (Factory Reset) 34
4.7 Diagnostic Port (USB Interface) 35
4.8 EtherCAT Operation 35
5 Deinstallation 37 6 Maintenance, Repair 39
6 tina84e1-a (2019-07)
7 Returning the Product 39 8 Disposal 40 Further Information 41 EU Declaration of Conformity 43
For cross-references within this document, the symbol ( XY) is used, for cross-references to further documents, listed under "Further Information", the symbol ( [Z]).
tina84e1-a (2019-07) 7
DANGER
WARNING
Caution
Skilled personnel

1 Safety

1.1 Symbols Used

Information on preventing any kind of physical injury.
Information on preventing extensive equipment and environ­mental damage.
Information on correct handling or use. Disregard can lead to malfunctions or minor equipment damage.
Notice
<…> Labeling

1.2 Personnel Qualifications

All work described in this document may only be carried out by persons who have suitable technical training and the neces­sary experience or who have been instructed by the end-user of the product.
8 tina84e1-a (2019-07)

1.3 General Safety Instructions

Adhere to the applicable regulations and take the necessary precautions for the process media used.
Consider possible reactions with the product materials.
Adhere to the applicable regulations and take the necessary precautions for all work you are going to do and consider the safety instructions in this document.
Before beginning to work, find out whether any vacuum com­ponents are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contamin­ated parts.
Communicate the safety instructions to all other users.

1.4 Liability and Warranty

INFICON assumes no liability and the warranty becomes null and void if the end-user or third parties
disregard the information in this document
use the product in a non-conforming manner
make any kind of interventions (modifications, alterations etc.)
on the product
use the product with accessories not listed in the product documentation.
The end-user assumes the responsibility in conjunction with the process media used.
Gauge failures due to contamination are not covered by the warranty.
tina84e1-a (2019-07) 9
Measurement range
"Validity"
1)
0.01 … 0.05 F.S.
0.2% of reading
0.1 … 1100 F.S.
0.15% of reading
1 … 1100 F.S.
0.0025% F.S./ °C
Temperature effect on span
0.01% of reading / °C
Resolution
0.003% F.S.
Gas type dependence
none
(measurement signal)
Measurement range
0 … +10 V
(limited to +10.5 V)
Relationship voltage-pressure
linear
Output impedance
0 (short-circuit proof)
Loaded impedance
>10 k
Measurement rate
1 kHz
(analog output signal)
interface)

2 Technical Data

Precision
Accuracy
Temperature effect on zero
0.01 … 0.02 F.S.
0.05 … 0.5 F.S.
Output signal analog
Voltage range –5 … +10.5 V
Step response time 3)
Delay time (digital output signal, EtherCAT
1)
ISO/WD 15725-1
2)
Non-linearity, hysteresis, repeatability at 25 °C ambient operating
temperature without temperature effects after operation of 2 h.
3
)
Increase 10 … 90 % F.S.R.
4
)
Depending on pressure range, conductance vacuum connection and the
default filter setting.
10 tina84e1-a (2019-07)
2)
0.0100% F.S./ °C
0.0050% F.S./ °C
2 … 20 ms 4)
2 ms
Voltage
5 V
contact ( 24)
Pulse
>1 s … <5 s
status
Setpoints
Setting range SP1 & SP2
1 … 99% F.S.
pressure
Hysteresis 5)
1% F.S.
High Trip Point
floating (NO)
closed
LED lit solid
open
LED off
Status relay
mon (pin 5)
warning
warming up, error
Cable
USB cable type A/mini B
Identification
Resistance R
Ident
13.2 k referenced to sup­ply common
Remote Zero Adjust digital input for zero adjust-
ment with external switching
External switching contact
30 V (dc) / <5 mA (dc)
Switching functions setpoints SP1& SP2, AT M,
Setting range ATM Factor 0.5 … 1.1 from ATM
Switching characteristics 5) Low Trip Point (default)
Relay contact 30 V (dc) / 0.5 A (dc)
Relay contact 30 V (dc) / 0.5 A (dc)
connected to supply com-
closed measurement mode
open no supply voltage
Diagnostic port
5
)
The hysteresis and the switching characteristics can be programmed via
the serial interface or the diagnostic port.
tina84e1-a (2019-07) 11
Mini USB type B, 5-pin
EtherCAT interface
3CC9-xxx-23G0
[9], [10]
Data rate
100 Mbps
Note address
explicit device identification
Physical layer
100Base-Tx (IEEE 802.3)
input and output
or higher)
Cable length
100 m
For further information on the EtherCAT interface [5], [6]
DANGER
has to be fused 6).
±15 V (dc)
at operating temperature
9 W
Fuse to be connected 6)
1.25 AT
voltage and overload.
Specification, data format, communication protocol 3CC9-xxx-2380
[7], [8]
EtherCAT connector 2×RJ45, 8-pin, socket
Cable 8-pin, shielded, Ethernet
Patch Cable (CAT5e quality
Supply
The gauge may onl y be connected to power sup­plies, instruments or control devices that conform to the requirements of a grounded protective extra­low voltage (PELV). The connection to the gauge
Supply voltage
at the gauge
Power consumption while being heated
+14 … +30 V (dc) or
≤14 W
The gauge is protected against reverse polarity of the supply
6)
INFICON controllers fulfill this requirement.
12 tina84e1-a (2019-07)
Electrical connection
15-pin D-sub, male
Sensor cable
15-pin plus shielding
15 m (0.25 mm²/conductor)
75 m (0.25 mm²/conductor)
135 m (0.25 mm²/conductor)
(R
cable
1.0 ).
Grounding concept
"Power Connection"
stainless steel AISI 316L
Internal volume
4.2 cm3
0.01 … 0.5 F.S.
1.3 bar | 130 kPa
Bursting pressure (absolute)
6 bar | 600 kPa
Admissible temperatures
Bakeout
110 °C at the flange
at +40°C
3000 m NN
Degree of protection
IP 30
tal
Cable length
Supply voltage 15 V
8 m (0.14 mm²/conductor)
Supply voltage 24 V 43 m (0.14 mm²/conductor)
Supply voltage 30 V 88 m (0.14 mm²/conductor)
For longer cables, larger conductor cross-sections are required
Materials exposed to vacuum ceramics (Al2O3 99.5%),
Admissible pressure (absolute)
1000 / 1100 F.S. 1 … 500 F.S.
Storage Operation
4 bar | 400 kPa
2.6 bar | 260 kPa
–20 °C … +85 °C +10 °C … +40 °C
Relative humidity 80% at temperatures
+31 °C, decreasing to 50%
Use indoors only, altitude up to
Mounting orientation
0.01 … 0.05 F.S.
0.1 … 1100 F.S.
standing upright or horizon­tal standing upright to horizon-
tina84e1-a (2019-07) 13
15.7
31.5
110.4
3.8
DN 16 ISO-KF
½¨ tube
ø82
ø85
DN 16 CF-R
8VCR
female
Weight
837 … 897 g

Dimensions [mm]

14 tina84e1-a (2019-07)
1 2 3 4 5 6 7 8 9 10
0
0.9×F.S.
0.5×F.S.
0.6×F.S.
1.0×F.S.
0.7×F.S.
0.8×F.S.
0.1×F.S.
0.2×F.S.
0.3×F.S.
0.4×F.S.
1.1×F.S.
Pressure p
Measurement U
out
[V]
0.0×F.S.
Torr
mbar 7)
Pa 7
1.3332…
133.3224…

Analog Measurement Signal vs. Pressure

c 1.00
p = (U
/ 10 V) × p (F.S.)
out
Conversion Torr Pascal
1013.25 / 760 =
101325 / 760 =
Example: Gauge with 10 Torr F.S.
Measurement signal U
= 6 V
out
p = (6 V / 10 V) × 10 Torr
= 0.6 × 10 Torr = 6 Torr
7)
Source: NPL (National Physical Laboratory)
Guide to the Measurement of Pressure and Vacuum, ISBN 0904457x / 1998
tina84e1-a (2019-07) 15
WARNING
impacts.
DANGER
suited to overpressure.
DANGER
Use O-rings provided with an outer centering ring.

3 Installation

Fragile components The ceramic sensor may be damaged by impacts. Do not drop the product and prevent shocks and

3.1 Vacuum Connection

Overpressure in the vacuum system >1 bar Injury caused by released parts and harm caused
by escaping process gases can result if clamps are opened while the vacuum system is pressurized.
Do not open any clamps while the vacuum system is pressurized. Use the type clamps which are
Overpressure in the vacuum system >2.5 bar KF flange connections with elastomer seals (e.g.
O-rings) cannot withstand such pressures. Process media can thus leak and possibly damage your health.
16 tina84e1-a (2019-07)
DANGER
measures to fulfill this requirement.
Caution
damages.
Caution
tools when working in this area.
Protective ground Products that are not correctly connected to ground
can be extremely hazardous in the event of a fault. Electrically connect the gauge to the grounded
vacuum chamber. This connection must conform to the requirements of a protective connection ac­cording to EN 61010:
CF and VCR flanges fulfill this requirement.
For gauges with a KF flange, use a conductive
metallic clamping ring.
For gauges with a ½" tube, take appropriate
Vacuum component Dirt and damages impair the function of the vac-
uum component. When handling vacuum components, take appro-
priate measures to ensure cleanliness and prevent
Dirt sensitive area Touching the product or parts thereof with bare
hands increases the desorption rate. Always wear clean, lint-free gloves and use clean
Mount the gauge so that no vibrations occur.
tina84e1-a (2019-07) 17

Mounting orientation

0.01 … 0.05 F.S.
Seal with
centering ring
Clamp
Protective lid
standing upright or horizontal acc. to the "Calibration Test Report"
0.1 … 1100 F.S.
If adjustment should be possible after the gauge has been in­stalled, be sure to install it so that the buttons can be accessed with a pin.
Remove the protective lid and connect the product to the vac­uum system.
18 tina84e1-a (2019-07)
standing upright to horizontal
Keep the protective lid.
DANGER
has to be fused 8).

3.2 Power Connection

Make sure the vacuum connection is properly made
( 16).
tina84e1-a (2019-07) 19
The gauge may onl y be connected to power sup­plies, instruments or control devices that conform to the requirements of a grounded protective extra­low voltage (PELV) The connection to the gauge
Ground loops, differences of potential, or EMC problems
may affect the measurement signal. For optimum signal quality, please do observe the following notes:
Use an overall metal braided shielded cable. The connector must have a metal case.
Connect the cable shield to ground at one side via the connector case. Make sure the connector case has direct contact to the cable's shield on its whole ci r­cumference. Do not connect the other side of the shield.
Connect the supply common with protective ground directly at the power.
Use differential measurement input (signal common and supply common conducted separately).
Potential difference between supply common and housing ≤18 V (overvoltage protecti on).
8)
INFICON controllers fulfill this requirement.
c
ase
14…
30 V
Remote Zero Adjust
1
M
4
3
1 8 9
10
Ident
13 14
2
12
7
11
5
15
680 n
F
18 V
10
6
SP1
SP2
Status
Pin 1, 4
Relay SP1, closing contact
15
8
9
1
Pin 2
Signal output (measuring signal) or thresholds SP1/2
Pin 3
Status
Pin 5
Supply common
Pin 7, 11
Supply (+14…+30 V)
Pin 8, 9
Relay SP2, closing contact
Pin 10
Gauge identification or Remote Zero Adjust
Pin 12
Signal common
Pin 15
Housing (Chassis Ground)
Case
Connector case
Pin 6, 13, 14: n.c.

3.2.1 D-sub, 15-pin Connector

If no sensor cable is available, make one according to the following diagram (cable length and conductor cross-sections 13).
Electrical connection
20 tina84e1-a (2019-07)
D-sub, 15-pin
female
soldering side
8 1
Pin 1
TD+
Transmission data +
Pin 2
TD-
Transmission data -
Pin 3
RD+
Receive data +
Pin 6
RD-
Receive data -
Sensor cable
FCC68 cable <IN> port
FCC68 cable <OUT> port

3.2.2 EtherCAT Connector

EtherCAT is a communications interface. It is powered
via the sensor cable.
If no EtherCAT cables are available, make them according to the following diagram. Connect the EtherCAT cables.
FCC68, 8-pin, soldering side
Pin 4, 5, 7 and 8: n.c.
tina84e1-a (2019-07) 21
5
1
Pin 1
VBUS (5 V)
Pin 2
D-
Pin 3
D+
Pin 4
ID
Pin 5
GND
Mini USB type B

3.2.3 Mini USB Type B Connector (Diagnostic Port)

A standard USB cable can be used. If no cable is available, make one according to the following diagram. Connect the cable.
Mini USB typ B, soldering side
22 tina84e1-a (2019-07)
LED
LED status
Meaning
<STATUS>
off
no supply voltage
lit solid green
measurement mode
long blinks
warming up
lit solid red
error
<SP1>
lit green green
setpoint 1 closed
blinking green
waiting for setpoint 1 input
off
setpoint 1 open
<SP2>
lit solid green
setpoint 2 closed
blinki ng green
waiting for setpoint 2 input
off
setpoint 2 open

4 Operation

Put the gauge into operation. If you are using an INFICON con­troller, define the measurement range (
A warm-up time of at least ½ hour should be allowed; for precise pressure measurements a warm-up time of at least 2 hours is required.

4.1 Status Indication

[1], [2], [3]).
blinking green
EtherCAT LEDs [5], [6].
tina84e1-a (2019-07) 23
short blinks
warning, over/underrange
F.S.
U / 90°
1000 Torr/mbar
2 mV
100 Torr/mbar
10 mV
10 Torr/mbar
50 mV
1 Torr/mbar
300 mV
0.1 Torr/mbar
1.8 V
0.01 … 0.05 Torr/mbar
0 V *)

4.2 Zeroing the Gauge

The gauge is factory calibrated while "standing upright" or in horizontal position ( "Calibration Test Report").
Perform a zero adjustment, when the gauge is operated
for the first time.
Due to long time operation or contamination, a zero drift could occur and zero adjustment may become necessary.
For adjusting the zero, operate the gauge under the same con­stant ambient conditions and in the same mounting orientation as normally.
The output signal (measuring signal) is depending on the moun­ting orientation. The signal difference between the vertical and horizontal mounting orientation is:
*) The gauge is factory calibrated while "standing upright" or in
horizontal position ("Calibration Test Report").
If the gauge is operated via a controller, the zero of the
whole measuring s ystem has to be adjusted on the controller: first, adjust the zero of the gauge and then, the zero of the controller.
4.2.1 <ZERO> Adjustment
The zero can be adjusted via
the <ZERO> button on the gauge,
the diagnostic port ( [4]),
the EtherCAT interface ( [5], [6]),
24 tina84e1-a (2019-07)
the digital input "Remote Zero": Apply the supply
F.S.
zero adjustment
0.01 Torr
voltage to pin 10 (pulse 11),
an INFICON Vacuum Gauge Controller (VGC series).
While the gauge is being heated and/or under atmos-
pheric pressure, the zeroing function is locked in order for operating errors to be prevented.
Evacuate the gauge to a pressure according to the table
below:
Recommended final pressure for
1100 mbar 1000 Torr 500 Torr/mbar 200 Torr/mbar 100 Torr/mbar 50 Torr/mbar 20 Torr/mbar 10 Torr/mbar 5 Torr/mbar 2 Torr/mbar 1 Torr/mbar
0.5 Torr/mbar
0.25 Torr/mbar
0.1 Torr/mbar
0.05 Torr/mbar
0.02 Torr/mbar
<5×10 <3×10-2 Torr <1×10-2 Torr <5×10-3 Torr <3×10-3 Torr <1×10-3 Torr <5×10-4 Torr <3×10-4 Torr <1×10-4 Torr <5×10-5 Torr <3×10-5 Torr <1×10-5 Torr <5×10-6 Torr <3×10-6 Torr <1×10-6 Torr <5×10-7 Torr
-2
Torr
<7×100 Pa <7×100 Pa
<4×100 Pa <2×10-0 Pa <7×10-1 Pa <4×10-1 Pa <2×10-1 Pa <7×10-2 Pa <4×10-2 Pa <2×10-2 Pa <7×10-3 Pa <4×10-3 Pa <2×10-3 Pa <7×10-4 Pa <4×10-4 Pa <2×10-4 Pa <7×10-5 Pa
<7×10-2 mbar
<4×10-2 mbar <2×10-2 mbar <7×10-3 mbar <4×10-3 mbar <2×10-3 mbar <7×10-4 mbar <4×10-4 mbar <2×10-4 mbar <7×10-5 mbar <4×10-5 mbar <2×10-5 mbar <7×10-6 mbar <4×10-6 mbar <2×10-6 mbar
If the final pressure is too high for zero adjustment (>25% of the F.S.), the zero cannot be reached and the <STATUS> LED blinks green. If this is the case, activate the factory setting and adjust the zero again ( 34).
Operate the gauge for at least 2 hours (until the signal is
stable).
tina84e1-a (2019-07) 25
Briefly press the <ZERO> button with a pin (max.
max. ø1.1 mm
Press the button briefly
ø1.1 mm). or …
… apply supply voltage to Remote Zero at pin 10 (pulse 11).
The zero adjustment runs automatically. The <STATUS> indicator flashes until the adjustment (duration ≈8 s) is completed.
After zero adjustment, the gauge automatically returns
to the measurement mode.
The <STATUS> LED blinks green if
the signal output is negative (< -20 mV) when the final pres­sure has been attai ned
the zero adjustment has failed.
4.2.2 <ZERO> Adjustment with Ramp Function
The ramp function allows to adjust the pressure value at a known reference pressure within the measurement range of the gauge.
26 tina84e1-a (2019-07)
It also permits to adjust an offset of the characteristic curve in
max. ø1.1 mm
Keep the button depressed
order to
compensate for the offset of the measuring system or
obtain a slightly positive zero for a 0 … 10 V AD converter.
The offset should not exceed 5% of the F.S. (+500 mV). At a higher positive offset, the upper limit of the measurement range is exceeded.
Zero adjustment using the ramp function can be
performed via
the <ZERO> button on the gauge,
the EtherCAT interface ( [5], [6]),
the diagnostic port ( [4]),
Recommended procedure for adjusting the offset of a
measuring system: → Notice 24.
Operate the gauge for at least 2 hours (until the signal is
stable).
Push the <ZERO> button with a pin (max. ø1.1 mm) and
keep it depressed. The <STATUS> LED starts blinking. After 5 s, the zero adjustment value, starting at the current output value, keeps continually changing (ramp) until the button is released or until the setting limit (max. 25% F.S.) is reached.
tina84e1-a (2019-07) 27
Change of direction (inverse ramp): Release the button. Press and keep it depressed again within 3 … 5 s (the flashing frequency of the <STATUS> indicator changes briefly).
Fine adjustment: Release the button. Briefly press it again within 0 … 3 s. The zero adjustment value chang­es by one unit.
If the <ZERO> button is released for m ore than 5 s, the
gauge returns to the measurement mode.
The <STATUS> LED blinks green if the signal output is negative (< -20 mV).

4.3 Switching Functions SP1, SP2, ATM, Status

The gauge provides two adjustable and programmable switching functions, which are independent of each other. A switching contact is provided for each switching function:
Vacuum pressure set points SP1 & SP2 (default)
Atmospheric pressure set point (ATM set point)
Gauge status

Switching characteristics and hysteresis of the set points

Low Trip Point (default)
High Trip Point
The switching characteristics and the hysteresis of each
set point can only be programmed via the diagnostic port ( [4]) and the E therCAT interface ( [5], [6]).
28 tina84e1-a (2019-07)
Mea
surement sign
al
(pressure p)
Time t
OffOn
Hysteresis (1% F.S.)
Off
Threshold value
Reset value
Setpoint
Measurement signal (pressure p)
Time t
OffOn
Hysteresis (1% F.S.)
Off
Threshold value
Setpoint
Reset value

Low Trip Point (default)

If the pressure in the vacuum system is lower than the setpoint, the corresponding LED (<SP1> or <SP2>) is lit soli d and the corresponding switching contact is closed.
tina84e1-a (2019-07) 29

High Trip Point

If the pressure in the vacuum system is higher than the setpoint, the corresponding LED (<SP1> or <SP2>) is lit soli d and the corresponding switching contact is closed.
DANGER
malfunction will cause.
4.3.1 Adjusting the Vacuum Pressure Setpoints (SP1 & SP2)
The two setpoints can be set to any pressure within the meas­urement range of the gauge.
The thresholds of the setpoints SP1 & SP2 can be
adjusted via
the buttons on the gauge
the diagnostic port ( [4])
the EtherCAT interface ( [5], [6]).
The current threshold setting
is output at the measurement signal output instead of the
pressure signal and can be measured with a voltmeter after the <SP> button is pressed
can be read via the diagnostic port ( [4]) and the
EtherCAT interface ( [5], [6]).

Adjusting Setpoint <1>

Push the <SP> button with a pin (max. ø1.1 mm). The
30 tina84e1-a (2019-07)
Malfunction If processes are controlled via the signal output,
keep in mind that by pushing an <SP> button the measurement signal is suppressed and the cor­responding threshold value is output instead. This can cause malfunctions.
Push the <SP> button only if you are sure that no
gauge changes to the switching function mode and outputs the current threshold value at the measurement value output for about 10 s (LED <1> blinks).
max. ø1.1 mm
Press the button
max. ø1.1 mm
Keep the button depressed
For changing the threshold value, push the <ZERO> button
and keep it depressed. The threshold keeps changing from the current value (ramp) until the button is released or until the limit of the setting range is reached.
Change of direction (inverse ramp): Release the button. Press and keep it depressed again within 3 … 5 s (the flashing frequency of the <STATUS> indicator changes briefly).
Fine adjustment: Release the button. Briefly press it again within 0 … 3 s. The threshold value changes by one unit.
tina84e1-a (2019-07) 31
If the <ZERO> button is released for more than 5 s, the
Low Trip Point Hysteresis
High Trip Point Hysteresis
Setpoint Mode
gauge returns the measurement mode.
The factory setting of the reset value is 1% F.S. above
the Low Trip Point and 1% F.S. below the High Trip Point (hysteresis).
If after programming of the hysteresis the button <SP> is
pushed, the factory setting of the hysteresis (1%) is reactivated.

Programming setpoint SP1

Programmable parameters: ( [4], [5], [6])
Low Trip Point Low Trip Enable
High Trip Point

Adjusting setpoint SP2

The adjustment procedure is the same as for setpoint SP1.
4.3.2 Programming the ATM Setpoint
The vacuum pressure setpoints SP1 & SP2 can be programmed to atmospheric pressure setpoint (ATM setpoint) vi a the diag­nostic port ( [4]) or the EtherCAT interface ( [5], [6]).
Switching characteristics Low / High Trip Point and hysteresis 28.
The threshold value of the ATM setpoint is defined as a factor of the current atmospheric pressure. The switching contact switch­es when the pressure in the vacuum system has reached the defined threshold value.
32 tina84e1-a (2019-07)
High Trip Enable
Example: Setpoint of the High Trip P oint with threshold 110% of
M
easurement signal
(pressure p)
1078 mbar
980 mbar
Atmospheric pres/ sure (ambience)
ATM setpoint 110% (= factor 1.1)
Hysteresis (1% F.S.)
OffOnOff
Threshold value
Reset value
Low Trip Point Hysteresis
High Trip Point Hysteresis
Setpoint Mode
the atmospheric pressure (= factor 1.1) Hysteresis: 1%
The current programmed ATM threshold setting
can be read / written via the diagnostic port ( [4])
can be read / written via the EtherCAT interface ( [5],
[6]).

Programming ATM setpoint

Programmable parameters: ( [4], [5], [6])
High Trip Enable
tina84e1-a (2019-07) 33
Factor of ATM Low Trip Enable
Switching characteristics of the setpoints 28. Diagnostic port [4].
4.3.3 Programming Gauge Status
The vacuum pressure setpoints SP1 & SP2 can be programmed to gauge status via the diagnostic port ( [4]) or the EtherCAT interface ( [5], [6]).
The status of the gauge is output as a digital signal.

4.4 Activating the Factory Setting (Factory Reset)

All user defined parameters (e.g. zero, filter) are restored to their default values.
Loading of the default parameters is irreversible.
Loading the default parameters:
Put the gauge out of operation. Keep the <ZERO> button depressed for at least 5 s while
the gauge is being put into operation (Power ON).
34 tina84e1-a (2019-07)
Mini USB type B
Caution
nostic port.

4.5 Diagnostic Port (USB Interface)

The diagnostic port <DIAG> permits to output the pressure reading and all status information and to enter all settings at the same time ( [4]). A standard USB cable type A/mini B can be used.
Required software: T-Gauge. Please contact your local INFICON service center.

4.6 EtherCAT Operation

Data transmission errors Any attempt to simultaneously operate the gauge
via the EtherCAT interface and the diagnostic port causes data transmission errors.
Therefore the gauge must not be operated simul­taneously via the EtherCAT interface and the diag-
tina84e1-a (2019-07) 35
For operating the gauge via EtherCAT, prior installation of the
2
4
068
A
E
2
4
068
A
E
2
4
068
A
E
device specific ESI file is required on the bus master side. This file can be downloaded from our website (www.inficon.com).
Explicit Device Address Setting (default 00
)
hex
During device initialization, the device address switches are read by the device firmware. This device address is supported to the master as Explicit Device Identification.
The explicit device address is set in hexa­decimal form (00 … FFF <x100>, <x10> and <x1> switches.
) via the
hex
Example: Device address = 0xDDD (dec 3549): 0x100 * 0xD (dec 3328) + 0x10 * 0xD (dec 208) + 0x1 * 0xD (dec 13)

Status LED

LEDs on the gauge inform on the gauge status and the current EtherCAT status ( [5], [6]).
36 tina84e1-a (2019-07)
WARNING
impacts.
DANGER
when handling contaminated parts.
Caution
damages.

5 Deinstallation

Fragile components The ceramic sensor may be damaged by impacts. Do not drop the product and prevent shocks and
Contaminated parts Contaminated parts can be detrimental to health
and environment. Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant regulations and take the necessary precautions
Vacuum component Dirt and damages impair the function of the vac-
uum component. When handling vacuum components, take appro-
priate measures to ensure cleanliness and prevent
tina84e1-a (2019-07) 37
Caution
tools when working in this area.
Dirt sensitive area Touching the product or parts thereof with bare
hands increases the desorption rate. Always wear clean, lint-free gloves and use clean
Vent the vacuum system. Put the gauge out of operation. Unfasten the lock screws and disconnect the sensor cable. Remove the gauge from the vacuum system and install the
protective lid.
38 tina84e1-a (2019-07)
WARNING
declaration of contamination *).

6 Maintenance, Repair

Under clean operating conditions, the product requires no main­tenance.
Gauge failures due to contamination are not covered by
the warranty. We recommend checking the zero at regular intervals ( 24).
INFICON assumes no liability and the warranty becomes null and void if any repair work is carried out by the end-user or third parties.

7 Returning the Product

*)
Form under www.inficon.com
Products that are not clearly declared as "free of harmful sub­stances" are decontaminated at the expense of the customer.
Products not accompanied by a duly completed declaration of contamination are returned to the sender at his own expense.
tina84e1-a (2019-07) 39
Forwarding contaminated products Contaminated products (e.g. radioactive, toxic,
caustic or microbiological hazard) can be detrimen­tal to health and environment.
Products returned to INFICON should preferably be free of harmful substances. Adhere to the forward­ing regulations of all involved countries and for­warding companies and enclose a duly completed
DANGER
when handling contaminated parts.
WARNING
relevant local regulations.

8 Disposal

Contaminated parts Contaminated parts can be detrimental to health
and environment. Before beginning to work, find out whether any
parts are contaminated. Adhere to the relevant regulations and take the necessary precautions
Substances detrimental to the environment Products or parts thereof (mechanical and electric
components, operating fluids etc.) can be detrimen­tal to the environment.
Dispose of such substances in accordance with the

Separating the components

After disassembling the product, separate its com ponents ac­cording to the following criteria:
Contaminated components Contaminated components (radioactive, toxic, caustic or bio-
logical hazard etc.) must be decontaminated in accordance with the relevant national regulations, separated according to their materials, and disposed of.
Other components Such components must be separated according to their ma-
terials and recycled.
40 tina84e1-a (2019-07)

Further Information

[1] Operating Manual
Single-Channel Controller VGC401 tinb01e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[2] Operating Manual
Two- & Three-Channel Measurement and Control Unit VGC402, VGC403 tinb07e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[3] Operating Manual
One-, Two- & Three-Channel Measurement and Control Unit VGC501, VGC502, V GC503 tina96e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[4] Communication Protocol
Diagnostic Port
via T-Gauge tira84e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[5] Communication Protocol
EtherCAT CDG045Dhs
(ETG.5003.2080 S (R) V1.0.0)
tira68e1 (nur englisch) INFICON AG, LI–9496 Balzers, Liechtenstein
[6] Communication Protocol
EtherCAT CDG045Dhs
(ETG.5003.2080 S (R) V1.3.0)
tirb45e1 (nur englisch) INFICON AG, LI–9496 Balzers, Liechtenstein
[7] ETG.5003.1 S (R) V1.0.0: Semiconductor Device profile
– Part 1: Common Device Profile (CDP)
[8] ETG.5003.2080 S (R) V1.0.0: Semiconductor Device
profile – Part 2080: Specific Device Profile (SDP): Vacuum Pressure Gauge
[9] ETG.5003.1 S (R) V1.1.0: Semiconductor Device profile
– Part 1: Common Device Profile (CDP)
tina84e1-a (2019-07) 41
[10] ETG.5003.2080 S (R) V1.3.0: Semiconductor Device
profile – Part 2080: Specific Device Profile (SDP): Vacuum Pressure Gauge
42 tina84e1-a (2019-07)
5 July 2019
5 July 2019

EU Declaration of Conformity

We, INFICON, hereby declare that the equipment mentioned below complies with the provisions of the following directives:
2014/30/EU, OJ L 96/79, 29.3.2014
(EMC Directive; Directive relating to electromagnetic compatibility)
2011/65/EU, OJ L 174/88, 1.7.2011
(RoHS Directive; Directive on the restriction of the use of certain hazard­ous substances in electrical and electronic equipment)

Product

Capacitance Diaphragm Gauge
Stripe™ CDG045Dhs

Standards

Harmonized and international/national standards and specifi­cations:
EN 61000-6-2:2005
EN 61000-6-3:2007 + A1:2011
EN 61010-1:2010 (Safety requirements for electrical equipment for
measurement, control and laboratory use)
EN 61326:2013; Group 1, Cl ass B
cal equipment for measurement, control and laboratory use)

Manufacturer / Signatures

INFICON AG, Alte Landstraße 6, LI-9496 Balzers
(EMC: generic immunity standard)
(EMC: generic emission standard)
(EMC requirements for electri-
tina84e1-a (2019-07) 43
ti na84e1-a
LI–9496 Balzers Liechtenstein Tel +423 / 388 3111 Fax +423 / 388 3700
Original: German tina84d1-a (2019-07) reachus@inficon.com
www.inficon.com
Loading...