HP M1705B, M1772A, M1771A, M1770A, PageWriter 300 Service Manual

...
4 (3)

PageWriter 100, 200 and 300 Series Cardiographs

M1770A, M1771A, and M1772A

Mobile Cart M1705B

Service Manual

+

+3 3DUW 1R 0

&RS\ULJKW ‹

3ULQWHG LQ 86$ 2FWREHU

+HZOHWW 3DFNDUG &RPSDQ\

 

3ULQWHG LQ 8 6 $

(GLWLRQ

 

October 22, 1998 3:00 pm DRAFT

Notice

Notice

The information in this document is subject to change without notice.

Hewlett-Packard makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

This document contains or refers to proprietary information which is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of HewlettPackard Company.

 

Responsibility of the Manufacturer

 

Hewlett-Packard only considers itself responsible for any effects on safety, reliability and

 

performance of the equipment if all the following are true:

 

• Assembly operations, extensions, re-adjustments, modifications or repairs are done by

 

persons authorized by Hewlett-Packard.

 

• The electrical installation of the relevant room complies with the IEC or national require-

 

ments.

 

• The instrument is used according to the instructions for use presented in this manual.

 

 

 

As with all electronic

equipment, radio frequency interference between this

WARNING

 

cardiograph and any existing RF transmitting or receiving equipment at the

 

 

installation site, including electrosurgical equipment, should be evaluated

 

carefully and any limitations noted before the equipment is placed in service.

 

Monitoring during electrosurgery should not be attempted and monitoring

 

electrodes should be removed from the patient to preclude the possibility of

 

burns. Radio frequency generation from electrosurgical equipment and close

 

proximity transmitters may seriously degrade cardiograph performance. Hewlett-

 

Packard assumes no liability for failures resulting from RF interference between

 

HP medical electronics and any radio frequency generating equipment at levels

 

exceeding those established by applicable standards.

 

 

 

ii

CAUTION

CAUTION

Notice

Like all electronic devices, this cardiograph is susceptible to electrostatic discharge (ESD). Electrostatic discharge typically occurs when electrostatic energy is transferred to the patient, the electrodes, or the cardiograph. ESD may result in ECG artifact that may appear as narrow spikes on the cardiograph display or on the printed report. When ESD occurs, the cardiograph’s ECG interpretation may be inconsistent with the physician’s interpretation.

ESD discharges to exposed metal on the rear of the cardiograph can occasionally cause an error message to appear on the cardiograph display. The cardiograph returns to normal operation after turning the power off, then on again.

The data transmission cable must have a suppression device attached to assure the cardiograph’s compliance with the European Radiated Emissions Standard found in CISPR 11. If your data transmission cable does not include a suppression device, compliance can be achieved by attaching one of the following suppression devices to the cable, near the cardiograph:

part number 0443164251

Fair-Rite Products Corporation

 

 

P. O. Box J

 

 

One Commercial Row

 

 

Wallkill, New York 12589

 

 

telephone: (914) 895-2055

 

 

FAX: (914)895-2629

 

 

or

 

 

Euro-Schaffner, S. A.

 

 

1 B Avenue de Suisse - BP 16

 

 

68311 Illzach Cedex, France

 

 

telephone: 33-8-931-0400

 

 

FAX: 33-8-931-0401

part number 28B2025-0A0

Steward

 

 

East 36th Street

 

 

P. O. Box 510

 

 

Chattanooga, TN 37401

 

 

telephone: (615) 867-4100

 

 

FAX: (615) 867-4102

 

 

or

 

 

Steward EMC, S.A.

 

 

Rue Fritz-Couvoisier 40

 

 

Ch-2300 La Chaux-de-Fonds

 

 

Switzerland

 

 

telephone: 41-39-282-387

 

 

Fax: 41-39-280-277

 

 

 

This is to certify that this equipment is in accordance with the Radio Interference Requirements of the EMC Directive.

iii

Notice

Warranty

Hewlett-Packard warrants this medical product against defects in materials and workmanship for a period of three years in certain geographics, or one year with onsite support.

If Hewlett-Packard receives notice of such defects during the warranty period, HewlettPackard shall, at its option, either repair or replace hardware products which prove to be defective.

Hewlett-Packard software and firmware products that are designated by Hewlett-Packard for use with a hardware product, when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects during the warranty period. Hewlett-Packard shall repair or replace software media and firmware that do not execute their programming instructions due to such defects. Hewlett-Packard does not warrant that the operation of the software, firmware, or hardware shall be uninterrupted or error free.

If Hewlett-Packard is unable, within a reasonable time, to repair or replace any product to a condition as warranted, Buyer shall be entitled to a refund of the purchase upon return of the product to Hewlett-Packard.

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from any of the following:

1.Improper or inadequate maintenance by Buyer.

2.Buyer-supplied software or interfacing.

3.Unauthorized modification or misuse.

4.Operation outside of the environmental specifications for the product.

5.Improper site preparation and maintenance.

THE WARRANTY SET FORTH ABOVE IS EXCLUSIVE AND NO OTHER WARRANTY, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

iv

Notice

Printing History

October 1994

Edition 1

May 1996

Edition 2

October 1998

Edition 3

v

Safety Summary

Safety Summary

Safety Symbols Marked on the Cardiograph

The following symbols are used on the cardiograph or the cart:

Caution - See operating instructions

Alternating current.

Meets IEC type CF leakage current requirements and is defibrillator protected. (Isolated ECG input.)

Equipotential (identifies independent protective earth conductor to the cardiograph).

Fuse.

Indicates power control for cardiograph.

Hz Indicates operating frequency in cycles per second.

The maximum weight that the cart can hold

Displays the configuration menu on the PageWriter 200/200/300pi.

Please see Chapter 2, “Performance Verification and Maintenance,” for safety requirements that apply to the cardiograph.

vi

WARNING

CAUTION

NOTE

Conventions Used in This Manual

Conventions Used in This Manual

Warning statements describe conditions or actions that can result in personal injury or loss of life.

Caution statements describe conditions or actions that can result in damage to the equipment or software.

Notes contain additional information on cardiograph usage.

TEXT

represents the labels that appear on the display.

.H\

represents keys on the key panel.

 

represents the temporary key labels that appear on the display.

Softkey

 

 

vii

Preface

Preface

This manual contains service information for the Hewlett-Packard M1770A pageWriter 300pi, PageWriter 200i, M1771A PageWriter 200, and M1772A PageWriter 100 cardiographs. The information and procedures in this manual apply to all models unless otherwise specified.

This manual is organized as follows:

Chapter 1 — Introduction. Contains a general description of the cardiographs, lists of technical specifications, and lists of accessories and options.

Chapter 2 — Performance Verification and Maintenance. Explains how to check the cardiograph’s performance using built-in self-tests, and lists maintenance procedures and safety requirements that apply to the cardiograph.

Chapter 3 — Theory of Operation. Provides an overview of how the cardiograph works and describes the operation of the major subassemblies.

Chapter 4 — Troubleshooting. Contains procedures and error codes to aid the service person in localizing faults to a replaceable subassembly.

Chapter 5 — Removal and Replacement. Contains procedures for removing and replacing each of the cardiograph’s major subassemblies.

Chapter 6 — Parts List. Lists part numbers for the cardiograph’s replaceable parts, and provides assembly drawings.

Appendix A — Connector Pin Assignments. Identifies and defines the signals assigned to the subassembly interconnections.

Index.

viii

Contents

Notice. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . .ii

Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . .ii

Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . iv

Limitation of Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . iv

Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . vi

Safety Symbols Marked on the Cardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . vi

Conventions Used in This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. .vii

Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. viii

1. Introduction

 

PageWriter 100, 200, 200i, and 300pi Series Cardiographs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-1

M1705B Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-3

Transmission and Storage, Option #A05 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-3

Inquiries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-3

Specification Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-3

Sampling Characteristics of Cardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-4

Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-8

PageWriter 100, 200, 200i, and 300pi Cardiographs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-8

Country/Region Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 1-8

Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-12

2. Performance Verification and Maintenance

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-1

Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-1

Visual Inspection and Power On Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-2

Extended Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-2

Internal Circuitry Testing of PageWriter 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-2

Internal Circuitry Testing of PageWriter 200/200i/300pi . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-4

Memory Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 2-5

How to Read the Extended Self-Test Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-11

Display Test (PageWriter 200/200i/300pi only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-12

Indicator Light Test (PageWriter 100 only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-12

Printer Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-12

Keyboard Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-13

PageWriter 200/200i/300pi Keyboard Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-13

PageWriter 100 Keyboard Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-13

ECG Simulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-14

Changing the Default Operating Language (200/200i/300pi Only) . . . . . . . . . . . . . . . . . . . . . . . . .

2-16

Resetting the Cardiograph to the Factory Default State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-17

Resetting the PageWriter 200/200i/300pi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-17

Resetting the PageWriter 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-17

Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-18

Care and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-18

Cleaning the Cardiograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-18

Cleaning the Keyboard Overlay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-19

Cleaning the Digital Array Printhead and Paper Sensor . . . . . . . . . . . . . . . . . . . . . . . . . .

2-20

ix

Cleaning the Electrodes and Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-21

Caring for the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-22

Storing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-22

Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2-23

PageWriter 100, 200/200i, and 300pi Series Performance Verification Matrix . . . . . . . . . . . . . . . .

2-24

3. Theory of Operation

 

Operational Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 3-1

ECG Data Path. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-1

Power-on and Power-off Sequences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-1

Power-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-1

Power-off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-2

Circuit Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-2

The Patient Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-3

CPU Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-4

CPU/System Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-5

System Gate Array/Real-Time Crystal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-5

ECG Front End Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-5

Interrupt Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-5

DMA for ECG Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-6

DMA for LCD Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-6

System Reset Circuitry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-6

Watch-dog Circuitry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-6

Real-Time Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-6

DRAM RAS and CAS Generation and Refresh . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-6

NVRAM Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

Keyboard Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

Display Control for PageWriter 20 and 300 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

NVRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

ECG Front End . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

DRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

ROM-1 and ROM-2 Address Space . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

Printer Gate Array/SRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-7

Motor Driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-8

Analog/Digital Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-8

Thermal Printhead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-8

System Expansion Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-8

Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-9

Voltage Selector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-9

Primary Line Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-9

Power Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-9

Rectifier/Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-10

Switch-Mode Battery Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-10

Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-10

VPRINT Boost Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-10

Switched +6V Boost Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-10

5V Linear Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-10

ECG Front End . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-11

Isolation Power Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-11

Front End Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-11

Opto-Isolators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-12

x

Integrated Front End Circuits/Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12 Defibrillator Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12 Storage and Transmission (Option #A05 0nly) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13 M177XA Expansion Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Gate Array . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 ROM-X1, ROM-X2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Flash1, Flash2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 VPROG Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 Dual UART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 18.432 MHz Oscillator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 RS232 Drivers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 Serial Port Connectors 1 and 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15

4. Troubleshooting

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-1

Maintenance Philosophy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-1

Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-1

Test Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-2

Patient Cable Test Tool (M1770-87908) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 4-2

Front End and RS-232 Port Test Tool (M1770-87909, Rev. B) . . . . . . . . . . . . . . . . . . . . . . . . . .

4-2

Testing the Instrument Signal Path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-2

Testing the Cardiograph’s RS-232 Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-2

The Error and Event Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-3

Using Extended Self-test in Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-3

How the PageWriter 100 Communicates Error Codes and Messages . . . . . . . . . . . . . . . . . . . . . . . . .

4-3

PageWriter 100 Error Code Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-3

PageWriter 100 Error Message Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-4

Troubleshooting Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-4

Testing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4-19

5. Removal and Replacement

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-1

Tool Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-1

The Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-2

Removing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-2

The Keyboard Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-4

Removing the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-4

Removing the Keyboard Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-4

Removing the Keyboard Flexible Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-6

Removing the Keyboard Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-6

Replacing the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-7

The Top Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-8

Opening and Removing the Top Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-8

Replacing the Top Cover Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-9

The Printer Door Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-10

Removing the Printer Door Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-10

Replacing the Printer Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-11

The Printer Drive Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-11

Removing the Printer Drive Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-11

Replacing the Printer Drive Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-12

xi

The Printhead Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-13

Removing the Printhead Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-13

Removing the Optical Paper Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-14

Replacing the Printhead Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-15

The Control Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-17

Removing the Control Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-17

Replacing the Control Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-17

The Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-18

Removing the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-18

Replacing the Power Supply Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-19

The Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-20

The Internal Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-21

Replacing the Power Supply Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-21

Replacing the Capacitor Board Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-22

Replacing the Printer Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-23

The Storage and Transmission Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-23

Removing the Storage and Transmission Board Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-23

6. Parts Lists

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-1

Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 6-1

Calling for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-2

Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-3

7. Connector Pin Assignments

 

CPU Assembly Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

A-1

xii

WARNING

NOTE

Introduction

This chapter introduces you to the M1770A PageWriter 200i/300pi, M1771A PageWriter 200, and M1772A PageWriter 100 cardiographs and lists their technical specifications.

These cardiographs are only to be serviced by HP qualified personnel.

Safe and effective use of medical instrumentation requires periodic inspection and preventive maintenance. Perform the preventive maintenance procedures in Chapter 2 of this manual at required intervals to ensure satisfactory instrument performance.

The cardiographs use a thermal printhead to record waveforms and label the ECG report. The paper supplied with the cardiographs is a thermal paper designed to work with this printhead and with the photo detector used to advancing the paper.

Hewlett-Packard guarantees the performance of the cardiographs only when used with Hewlett-Packard supplies, accessories, and paper that meet or exceed Hewlett-Packard specifications.

PageWriter 100, 200, 200i, and 300pi Series Cardiographs

The M1770A PageWriter 200i/300pi is Hewlett-Packard’s economical, interpretive cardiograph. The M1771A PageWriter 200 is the economical, non-interpretive cardiograph. The M1772A PageWriter 100 is the most economical, non-interpretive cardiograph. The cardiograph contains the user controls, the printer, and all the processing circuitry. All models use the same enclosure. The M1705B cart is designed for these cardiographs. Figure 1-1 shows the M1770A mounted on the cart. Option #A05 provides transmission and storage capability for the M1770A and M1771A. The StressWriter option provides capability for communication between a PageWriter 100/200/200i/300pi cardiograph and a StressWriter system.

1-1

HP M1705B, M1772A, M1771A, M1770A, PageWriter 300 Service Manual

PageWriter 100, 200, 200i, and 300pi Series Cardiographs

Figure 1-1

M1770A/M1771A PageWriter 300pi/200i/200 Cardiograph.

The PageWriter 100, 200, and 300 family of cardiographs print ECGs in Auto and Manual formats. All PageWriter cardiographs use a continuous feed, high-resolution, thermal array printer. This produces electrocardiograms on standardsize paper that can be included in patient records without cutting or mounting. ECG reports clearly show the ECG waveforms. The PageWriter 200 and 300 series cardiographs also annotate records with patient information and ECG measurements. The PageWriter 200i and 300pi also includes ECG interpretation for both adult and pediatric patients.

From a service standpoint, the PageWriter 100, 200, and 300 family of cardiographs are similar except for two major hardware differences:

The M1770A has additional firmware for ECG measurements and interpretation.

The M1771A has additional firmware for ECG measurements.

The M1772A keypanel does not have an LCD preview display and uses operational keys instead of alphanumeric keys for user input.

All other model differences are found in the cardiograph’s software.

The PageWriter 100, 200, and 300 family of cardiographs is designed for low-cost, longterm reliability. The modular design makes extensive use of VLSI and gate array technology, resulting in a minimum number of subassemblies. The modular approach means less down-time for the user, since replacing subassemblies allows faster field repairs. The internal self-test efficiently identifies faulty subassemblies, further speeding the repair process.

1-2

Inquiries

M1705B Cart

The optional M1705B Cart provides mobility for the PageWriter 100, 200, and 300 family of cardiographs. The large wheels make the cart easy to move, yet steady. Slots in the cardiograph’s feet and a thumbscrew secure the cardiograph to the cart rails. The tray directly below the cardiograph provides storage for the user’s guide. Storage for the power cable is built in. Other built-in cable retainers hold the patient data cable out of the way. Two compartmented trays provide storage for spare patient electrodes and patient cables, additional thermal paper, and consumable supplies.

Transmission and Storage, Option #A05

Transmission and storage features are available for the M1770A and M1771A with purchase of Option #A05. Up to thirty Auto ECGs can be stored in the cardiograph’s internal flash memory. Stored ECGs can be recalled later for editing, re-analyzing, printing, or transmission to another PageWriter 200/200i/300pi equipped with Option #A05, a PageWriter XLi cardiograph, a TraceMaster ECG Management System, or facsimile machine. Manual ECGs cannot be transmitted or stored.

Inquiries

 

Refer any questions or comments regarding these instruments to the nearest Hewlett-

 

Packard Sales/Service Office or to one of Hewlett-Packard’s Service Dispatch Centers.

 

Always identify the instrument by model number and serial number in all correspondence.

 

Telephone numbers for Service Dispatch Centers and Sales/Service Offices are listed in

 

Chapter 6, “Parts List.”

 

 

 

 

 

 

 

 

 

 

Specification Data

 

The following tables list the technical specifications for the cardiographs and the mobile

 

cart. Specifications are the same for all models except as noted

Table 1-1

Physical Specifications

 

 

 

Parameter

 

Specification

 

 

 

 

 

Dimensions (h × w × l)

 

4 in × 15 in × 17 in (11.5 cm × 39.2 cm × 43.7 cm)

cardiograph

 

cart

 

36 in × 17 in × 33 in (91.4 cm × 43.5 cm × 84.8 cm)

 

 

 

 

 

Weight

 

 

 

 

cardiograph

 

18.7 lbs. (8.5 kg) or less (includes battery, 200 sheets of paper, patient cable, and power cord)

cart

 

38 lbs. (17.24 kg)

 

 

 

Chemical resistance,

 

Withstands the following: isopropyl alcohol (except patient cable), mild soap and water, chlorine

cleaners

 

bleach and water (30 ml/l of water).

 

 

 

 

 

1-3

 

Specification Data

Table 1-2

Electrical Specifications

 

 

 

Parameter

 

Specification

 

 

 

Resolution

 

5 μV

ECG (internal)

 

display

 

128 row pixels by 240 column pixels

 

 

 

display sweep speed

 

23.1 mm/sec ± 1%

 

 

 

Input impedance

 

greater than 2.5 MΩ@ 10 Hz, typically greater than 100MΩ @ DC1

 

 

 

Gain accuracy

 

±5% of input signal or ±40 μV, whichever is greater2

 

 

 

Input bias

 

less than 40 nA for input leadwire, less than 500 nA for Right Leg output

 

 

leadwire1

 

 

 

Common mode rejection

 

110 dB or greater with AAMI test circuit1

 

 

 

Defibrillator recovery

 

System recovery 8 seconds after 360 joule discharge, no damage1

 

 

 

Crosstalk rejection

 

Less than 2% channel crosstalk1

 

 

 

Sample rate

 

4.096M samples per second per input leadwire (unfiltered)

 

 

 

DC offset tolerance

 

³ ±300 mVdc with less than 5% gain change1

 

 

 

Noise

 

£ 30 μV peak-to-peak RTI1

 

 

 

Standardizing voltage

 

1 mV ± 5% for 100 mSec ± 5 mSec1

 

 

 

50/60 Hz notch filter

 

50/60 Hz AC line rejection filters are always active.

 

 

 

Pacemaker pulse display

 

> 0.2 mV indication for pace pulses between 0.5 and 2 ms duration at amplitude

 

 

2 to 250 mV1

 

 

 

1Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices.

2Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices. Tested according to AAMI EC11-1991 test method 4.2.7.1.

Sampling Characteristics of Cardiograph

Physiological factors such as breathing can cause variation in amplitudes of heart beats, independent of the heart rate. In addition, sampled systems (as opposed to continuous systems) may show changes in the apparent height of the R-wave when sampling of the R- wave occurs slightly off-peak. Because the sampling rate and the heart rate are asynchronous, the time between the peak and an adjacent sample can vary from one QRS complex to the next. This results in a slight variation of displayed QRS amplitude. This effect is more pronounced with narrower signals, more commonly found in some pediatric electrocardiograms and in pacemaker pulses. The M177XA family of cardiographs minimizes this effect by:

using an integrating-type A/D converter

sampling all leadwires simultaneously

oversampling the signals

processing the oversampled signals with appropriate digital signal processing techniques

1-4

Specification Data

Table 1-3 ECG to Paper Specifications

Parameter

Specification

 

 

 

 

Frequency response of PageWriter

High Pass

Low Pass

200/200i/300pi

 

 

(-3dB bandwidth)

 

 

 

 

 

Auto ECG

0.05 Hz1

150 Hz1

0.15 Hz1

100 Hz2

 

0.5 Hz

40 Hz

Manual ECG

0.05 Hz3

150 Hz3

0.15 Hz

100 Hz2

 

0.5 Hz

40 Hz

 

 

 

 

 

Frequency response of PageWriter 100

 

 

Auto ECG1

<3 dB down at (user selectable):

0.15-150 Hz or 0.15-40 Hz

Manual ECG3

<3 dB down at (user selectable):

0.05-150 Hz or 0.05-40 Hz

 

 

 

ECG resolution on paper

 

 

voltage axis

Approximately 8 dots/mm (200 dots/in)

 

time axis

Approximately 20 dots/mm (500 dots/in) at 25mm/sec and slower

 

Approximately 10 dots/mm (250 dots/in) at 50mm/sec

 

 

 

Recorder speed

5, 10, 25, and 50 mm/sec + 1.5%4

 

 

 

ECG visibility with pace pulse

Meets or exceed AMMI EC11-1991 standard for Diagnostic Electrocardio-

 

graphic Devices

 

 

 

 

1Bold entries mean meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices. Meets frequency response standard 3.2.7.2 using methods A, D, and E with auto report filter settings at 0.05-150 Hz or 0.15-150 Hz.

2-4 dB @ 100 Hz

3Bold entries mean meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices. Meets frequency response standard 3.2.7.2 using methods A, D, and E with manual report filter settings at 0.05-150 Hz.

4Bold entries mean meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices.

1-5

Specification Data

Table 1-4

Power Supply and Battery Specifications

 

 

 

Parameter

 

Specification

 

 

 

AC line frequency

 

50 and 60 Hz (nominal)

115 V setting

 

90–132 Vac

230 V setting

 

198–264 Vac

AC input power rating

 

50 VA

 

 

 

Battery voltage

 

6 V DC, 6.5 Ah

 

 

Battery recharge time with unit off

 

(battery fully discharged)

 

 

to 90% capacity

 

7 hours typical

to full capacity

 

16 hours typical

 

 

 

Battery capacity

 

 

Auto

 

40 ECGs

continuous rhythm (Manual mode)

40 minutes

 

 

 

Battery life

 

 

AC and battery operation

 

Typically 24 months

battery only operation

 

Typically 14–16 months

 

 

 

Low battery warning

 

At least 2 Auto ECGs or at least 2 minutes of Man-

 

 

ual ECG is allowed after the Low Battery indica-

 

 

tion is given.

 

 

 

Table 1-5

Safety Specifications

 

 

 

 

 

Parameter

Specification

 

 

 

 

ECG leads source current to ground

Less than 10 μA RMS1

 

 

 

 

ECG leads sink current

 

 

120 Vac, 60 Hz1

Less than 20 μA RMS with patient cable

 

240 Vac, 50 Hz2

Less than 50 μA RMS with patient cable

 

 

 

 

Ground wire integrity

Less than 100 mΩ @ 25 A AC1

 

 

 

 

Shunting of defibrillator energy

< 10%1

 

 

 

 

Operator safety during defibrillation

< 100 μC1

 

 

 

1Meets or exceeds AAMI EC11-1991 standard for Diagnostic Electrocardiographic Devices.

2Meets or exceeds IEC 601-2-25 International Standard for Medical Electrical Equipment, Part 2 for Electrocardiographs.

1-6

Specification Data

Table 1-6

Environmental Specifications

 

 

 

 

 

Parameter

Specification

 

 

 

 

Temperature

50° to 104° F (10° to 40° C)

 

operating

 

storage

32° to 122° F (0° to 50° C)

 

 

 

 

Humidity

 

 

operating

15 to 80% RH, non-condensing

 

storage

15 to 90% RH, non-condensing

 

 

 

 

Pressure (altitude)

 

 

operating

15,000 ft. (4600 m) for 2 hours

 

storage

15,000 ft. (4600 m)

 

 

 

Table 1-7

Miscellaneous Specifications

 

 

 

 

 

Parameter

Specification

 

 

 

 

Real time clock accuracy

 

 

10 to 40° C

Less than 3 minutes deviation per month.

 

 

 

 

Printhead life

Typically 100,000 pages

 

 

 

Table 1-8

Storage and Transmission Specifications (Option #A05 only)

 

 

 

 

Parameter

Specification

 

 

 

 

Auto ECG storage capacity

30 ECGs

 

 

 

 

Transmission Protocols

 

 

DT

Data transmission standard used between HP PageWriter cardio-

 

 

graphs and HP TraceMaster ECG Management Systems

 

SCP

Standard Communications Protocol1

 

 

 

 

Modem command interfaces

 

 

Data modem

Hayes standard AT command set

 

FAX modem

EIA/TIA-578 Service Class 1

 

 

 

 

Modem protocols

 

 

Modulation

V.34

 

Error correction

V.42

 

Compression

V.42 bis

 

FAX modulation

V.17

 

 

 

1Described in the European Committee for Standardization, Standard Communications Protocol - Computer - Assisted Electrocardiography. (CEN/TC 251).

1-7

Options and Accessories

Options and Accessories

PageWriter 100, 200, 200i, and 300pi Cardiographs

These tables list the options and accessories available for the PageWriter 100, 200, and 300 family of cardiographs.

Country/Region Options

Each country/region option includes the appropriate power cord, voltage, printer, patient cable, and language. See Table 1-9, which shows the configuration of each country/region option.

1-8

Options and Accessories

Table 1-9

 

Country/Region Option Configurations

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Country/

 

 

User

Printer1

AHA/

 

Power Cord2

 

Option

Region

 

Labels

Manuals

IEC

Volts

Electrodes

 

 

 

 

 

 

 

 

 

 

ABA

North America

 

English

English

A

AHA

120

903–US

Disposable

 

 

 

 

 

 

 

 

 

 

ABB

Europe English

 

English

English

A4

IEC

220

902–Europe

None

 

 

 

 

 

 

 

 

 

 

ABC

French Canada

 

French

French

A

AHA

120

903–US

Disposable

 

 

 

 

 

 

 

 

 

 

ABD

Germany

 

German

German

A4

IEC

220

902–Europe

None

 

 

 

 

 

 

 

 

 

 

ABE

Spain

 

Spanish

Spanish

A4

IEC

220

902–Europe

None

 

 

 

 

 

 

 

 

 

 

ABF

France

 

French

French

A4

IEC

220

902–Europe

None

 

 

 

 

 

 

 

 

 

 

ABG

Australia

 

English

English

A4

AHA

240

901–Australia

Welsh

 

 

 

 

 

 

 

 

 

 

ABH

Netherlands

 

Dutch

Dutch

A4

IEC

220

902–Europe

None

 

 

 

 

 

 

 

 

 

 

ABK

Intercon

 

English

English

A

AHA

120

903–US

Welsh

 

English, Taiwan

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ABM

Latin America

 

Spanish

Spanish

A

AHA

120

903–US

Welsh

 

 

 

 

 

 

 

 

 

 

ABN

Norwegian

 

Norwe-

Norwe-

A4

IEC

230

902-Europe

None

 

 

 

gian

gian

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ABS

Sweden

 

Swedish

Swedish

A4

IEC

220

902-Europe

None

 

 

 

 

 

 

 

 

 

 

ABU

United Kingdom

 

English

English

A4

IEC

240

900–UK

None

 

 

 

 

 

 

 

 

 

 

ABX

Finland

 

Finnish

Finnish

A4

IEC

220

902-Europe

None

 

 

 

 

 

 

 

 

 

 

ABZ

Italy

 

Italian

Italian

A4

IEC

220

902–Europe

None

 

 

 

 

 

 

 

 

 

 

AB2

China

 

Chinese

Chinese

A4

IEC

220

922–China

Welsh

 

 

 

 

 

 

 

 

 

 

AB4

Singapore

 

English

English

A4

AHA

240

900–UK

Welsh

 

 

 

 

 

 

 

 

 

 

AC6

S. Korea

 

English

English

A

AHA

220

902-Europe

Welsh

 

 

 

 

 

 

 

 

 

 

ACJ

India

 

English

English

A4

IEC

240

917-SA

Welsh

 

 

 

 

 

 

 

 

 

 

AKM

China

 

Chinese

English

A4

IEC

220

922-China

Welsh

 

 

 

 

 

 

 

 

 

 

ACQ

South Africa

 

English

English

A4

IEC

220

917–SA

None

 

 

 

 

 

 

 

 

 

 

AKV

South America

 

Spanish

Spanish

A

AHA

220

902–Europe

Welsh

 

 

 

 

 

 

 

 

 

 

1In Table 1-9 an “A” in the Printer column refers to 8.5 x 11-in. paper; “A4” refers to 210 x 297-mm paper

2See Table 1-11 for HP part numbers

NOTE

Switzerland should order the desired language (German, French, Italian, or English) and

the Swiss power cord (HP 8120-2296).

 

1-9

Options and Accessories

Table 1-10

PageWriter 300pi, 200/200i and 100 Series Documentation Part Numbers

 

 

 

 

 

Document Title

Language

Part Number

 

 

 

 

 

PageWriter 100 Cardiograph User’s Guide

English

M1772-91900

 

(M1772A)

 

 

 

 

 

 

 

 

French

M1772-91901

 

 

 

 

 

 

German

M1772-91902

 

 

 

 

 

 

Dutch

M1772-91903

 

 

 

 

 

 

Spanish

M1772-91904

 

 

 

 

 

 

Italian

M1772-91905

 

 

 

 

 

 

Finnish

M1772-91910

 

 

 

 

 

 

Chinese

M1772-91908

 

 

 

 

 

PageWriter 200/200i Cardiograph Users’s

English

M1772-91900

 

Guide (M1770A, M1771A with Serial # Prefix

 

 

 

CNA, CNB, CNC, 3807A or earlier version)

 

 

 

 

 

 

 

 

French

M1770-91901

 

 

 

 

 

 

German

M1770-91902

 

 

 

 

 

 

Dutch

M1770-91903

 

 

 

 

 

 

Spanish

M1770-91904

 

 

 

 

 

 

Italian

M1770-91905

 

 

 

 

 

 

Swedish

M1770-91906

 

 

 

 

 

 

Finnish

M1770-91910

 

 

 

 

 

PageWriter 300pi Cardiograph Users’s Guide

English

M1770-91930

 

(M1770A with Serial # Prefix CND)

 

 

 

 

 

 

 

 

French

M1770-91931

 

 

 

 

 

 

German

M1770-91932

 

 

 

 

 

 

Dutch

M1770-91933

 

 

 

 

 

 

Spanish

M1770-91934

 

 

 

 

 

 

Italian

M1770-91935

 

 

 

 

 

 

Swedish

M1770-91936

 

 

 

 

 

 

Norwegian

M1770-91938

 

 

 

 

 

 

Finnish

M1770-91939

 

 

 

 

1-10

Options and Accessories

Table 1-10

PageWriter 300pi, 200/200i and 100 Series Documentation Part Numbers

 

 

 

 

 

PageWriter 300pi Predictive Instruments Quick

English

M1770-92800

 

Reference Guide (M1770A with Serial # Prefix

 

 

 

CND)

 

 

 

 

 

 

 

 

French

M1770-92801

 

 

 

 

 

 

German

M1770-92802

 

 

 

 

 

 

Dutch

M1770-92803

 

 

 

 

 

 

Spanish

M1770-92804

 

 

 

 

 

 

Italian

M1770-92805

 

 

 

 

 

 

Swedish

M1770-92806

 

 

 

 

 

 

Norwegian

M1770-92808

 

 

 

 

 

 

Finnish

M1770-92809

 

 

 

 

 

Interpretive Physician’s Guide (M1770A)

English

M1700-92908

 

 

 

 

 

 

French

M1700-92918

 

 

 

 

 

 

German

M1700-92928

 

 

 

 

 

 

Dutch

M1700-92938

 

 

 

 

 

 

Spanish

M1700-92948

 

 

 

 

 

 

Italian

M1700-92958

 

 

 

 

 

Interpretive Physician’s Guide - Addendum

English

M1700-90700

 

(M1770A)

 

 

 

 

 

 

 

 

French

M1700-90701

 

 

 

 

 

 

German

M1700-90702

 

 

 

 

 

 

Dutch

M1700-90703

 

 

 

 

 

 

Spanish

M1700-90704

 

 

 

 

 

 

Italian

M1700-90705

 

 

 

 

 

Predictive Instruments Physician’s Guide

English

M1792-93900

 

(M1770A with Serial # Prefix CND)

 

 

 

 

 

 

 

 

French

M1792-93901

 

 

 

 

 

 

German

M1792-93902

 

 

 

 

 

 

Dutch

M1792-93903

 

 

 

 

 

 

Spanish

M1792-93904

 

 

 

 

 

 

Italian

M1792-93905

 

 

 

 

1-11

Options and Accessories

Table 1-11

Power Cord Part Numbers

 

 

 

 

 

 

Power Cord Key

 

HP Part Number

 

 

 

 

 

900

 

8120-1703

 

 

 

 

 

901

 

8120-4464

 

 

 

 

 

902

 

8120-1692

 

 

 

 

 

903

 

8120-5213

 

 

 

 

 

917

 

8120-4600

 

 

 

 

 

922

 

8120-8377

 

 

 

 

Table 1-12

Options

 

 

 

 

 

Option

Description

 

 

 

 

A05

Adds storage and transmission capability

 

 

 

 

M2488A-#A70, A71, A72

StressWriter system includes PageWriter cardiograph

 

 

 

 

M2488A-#C70

StressWriter system includes interface card for PageWriter

 

 

cardiograph

 

 

 

Standard Accessories

Accessories included are based on model number and localization.

• 200 sheets of z-fold paper -English paper p/n M2481A -Metric paper p/n M2483A

PageWriter 300pi HP M1770A Cardiograph User’s Guide (PageWriter 300pi)

PageWriter 200/200i HP M1771A/M1770A Cardiograph User’s Guide (PageWriter

200/200i)

Hewlett Packard Interpretive Cardiograph Physician’s Guide (PageWriter 200i/300pi only)

PageWriter 100 HP M1772A Cardiograph User’s Guide (PageWriter 100 only)

Using the HP PageWriter 200/200i Cardiograph Operator Training Video (PageWriter 200/200i, also for use with PageWriter 300pi)

Using the HP PageWriter 100 Cardiograph Operator Training Video (PageWriter 100 only)

1-12

Performance Verification and Maintenance

Introduction

This chapter describes how to verify the cardiograph’s performance, explains preventive maintenance, presents patient safety information, and provides a performance verification checklist.

Performance Verification

Several procedures make up performance verification: visual inspection of the cardiograph exterior, execution of Extended Self-test, visual examination of an ECG recorded from an ECG patient simulator, and system safety tests using a safety analyzer.

A Performance Verification matrix which appears at the end of this section specifies the tests and inspections which must be performed following servicing of the PageWriter cardiographs. The Performance Verification test results must be recorded on

Customer Service order records

Make copies of the matrix. Fill out a copy each time the cardiograph is tested. Attach the printer test output, and simulator ECG trace to the completed matrix and file with the cardiograph’s permanent maintenance record.

NOTE

If any of the performance verification tests fail, refer to Chapter 4, Troubleshooting.

 

2-1

2Q 6WDQGE\

Performance Verification

Visual Inspection and Power On Self Test

Before beginning the inspection, press the key to put the cardiograph in Standby mode, and unplug the power cord from the wall outlet. Inspect the cardiograph for the following:

Worn or damaged power cord

Loose or missing hardware

Mechanical damage

Evidence of liquid spill

Worn printer drive gear

Worn printer roller

Corroded or damaged reusable electrodes, if present

Damaged patient cable

Dirt/paper residue on the thermal printhead

Frayed or damaged wiring

Replace any damaged or missing items, and clean the printhead and patient electrodes as necessary. Cleaning instructions are listed under “Preventive Maintenance” later in this chapter. Connect the cardiograph to AC power. Be sure that the AC indicator comes on. Turn the cardiograph On and observe that the display turns on (PageWriter 200/200i/300pi) or that the LED’s flash on (PageWriter 100).

Extended Self-test

The cardiograph’s extended self-test checks each major subassembly. The extended selftest consists of two sections: the testing of internal circuitry and the testing of printer, display, keyboard, and modem capabilities.

To begin the extended self-test, press and hold both the $XWR

and 0DQXDO keys while

turning on the cardiograph with the 2Q 6WDQGE\ key.

 

To stop or exit extended self-test, you must press 2Q 6WDQGE\

to place the cardiograph

in Standby.

 

Internal Circuitry Testing of PageWriter 100

 

These tests are performed on the internal circuitry:

 

• Memory

 

• Gate Array

 

• Front-End Interface

 

• Serial Interface

 

Each of these tests is described in detail later in this chapter.

 

When the internal circuitry test of the PageWriter 100 is performed, the LEDs show which test is being performed. The following table shows which LED is associated with each test.

2-2

Performance Verification

Table 2-1

PageWriter 100 Test LED’s

 

 

 

 

 

 

 

 

Test Number

 

Test Name

LED(s) lit on Page-

Comments/Expected Results

 

 

 

 

Writer 100

 

 

 

 

 

 

 

 

1

 

ROM-1

5 mm/mV

OK

 

 

 

 

 

 

 

2

 

ROM-2

10 mm/mV

Normally fails for 100.

 

 

 

 

 

 

 

3

 

ROM-X1

20 mm/mV

OK for A#10.

 

 

 

 

 

 

 

4

 

ROM-X2

V Leads

OK for A#10.

 

 

 

 

 

 

 

5

 

RAM

Filter

OK

 

 

 

 

 

 

 

6

 

NVRAM

Auto

OK

 

 

 

 

 

 

 

7

 

GA

I II III Lead Group

Will also flash all LEDs. OK

 

 

 

 

 

 

 

8

 

ECG FE

aVR aVL aVF Lead Group

OK

 

 

 

 

 

 

 

9

 

SIO

II aVF V2 Lead Group

OK for A#10.

 

 

 

 

 

 

 

A

 

REMOTE

V Leads

For future use - normally fails.

 

 

 

 

 

 

 

B

 

MODEM

5 mm/sec

OK for A#10, with modem or test tool

 

 

 

 

 

attached

 

 

 

 

 

 

 

C

 

FLASH1

10 mm/sec

OK for A#10.

 

 

 

 

 

 

 

D

 

FLASH2

25 mm/sec

OK for A#10.

 

 

 

 

 

 

When the tests are completed, the results are also printed out in a pass/fail format on the left-hand side of the first page of the test report.

Table 2-2

Test Results Format

 

Test Passes

Test Fails

 

 

 

Printed Result

OK

**

 

 

 

In some cases, a test failure halts the test/test printout and an error code indicating the failure is flashed on the LED’s. See Chapter 4, “Troubleshooting,” for a list of self-test failure codes and failure symptoms. The section “How the PageWriter 100 Communicates Error Codes and Messages” in Chapter 4 describes how to read the LED’s.

2-3

Performance Verification

Internal Circuitry Testing of PageWriter 200/200i/300pi

These tests are performed on the internal circuitry:

Memory

Gate Array

Front-End Interface

Serial Interface

Each test is described detail later in this chapter.

When testing the internal circuitry of the PageWriter 200/200i/300pi, the numbers 1 through 9 and the characters A through D are displayed one at a time beginning with number 1. Each number represents a particular test. If the test fails, the character X is placed after the test number, for example, 3X. The test descriptions below are listed in the order the tests are performed.

Table 2-3

PageWriter 200/200i/300pi Test Numbers

 

 

 

 

 

 

 

 

Test Number

 

Test Name

Character Displayed on Page-

Comments

 

 

 

 

Writer 200/200i/300pi

 

 

 

 

 

 

 

 

1

 

ROM-1

1

OK

 

 

 

 

 

 

 

2

 

ROM-2

2

Passes for 300pi, 200i, and 200 with

 

 

 

 

 

software revision A.05.06 or later.

 

 

 

 

 

 

 

3

 

ROM-X1

3

Passes for Options #A05 and #A10

 

 

 

 

 

only.

 

 

 

 

 

 

 

4

 

ROM-X2

4

Passes for Options #A05 and #A10

 

 

 

 

 

only.

 

 

 

 

 

 

 

5

 

RAM

5

OK

 

 

 

 

 

 

 

6

 

NVRAM

6

OK

 

 

 

 

 

 

 

7

 

GA

7

OK

 

 

 

 

 

 

 

8

 

ECG FE

8

OK

 

 

 

 

 

 

 

9

 

SIO

9

Passes for Options #A05 and #A10

 

 

 

 

 

only.

 

 

 

 

 

 

 

A

 

REMOTE

A

For future use - normally fails

 

 

 

 

 

 

 

B

 

MODEM

B

Passes for Options #A05 and #A10

 

 

 

 

 

only, with modem or test tool

 

 

 

 

 

attached

 

 

 

 

 

 

 

C

 

FLASH1

C

Passes for Options #A05 and #A10

 

 

 

 

 

only

 

 

 

 

 

 

 

D

 

FLASH2

D

For future use - normally fails

 

 

 

 

 

 

2-4

CAUTION

Table 2-4

Performance Verification

If your cardiograph is equipped with Options #A05 or #A10, do not turn it off during Test C or D. If you turn off the cardiograph while Test C or D are in process, you could lose stored ECGs, the Log of ECGs Stored, or the Log of ECGs Taken.

When the tests are completed, the results are printed out in a pass/fail format on the lefthand side of the first page of the test report.

Test Results Format

 

Test Passes

Test Fails

 

 

 

Printed Result

OK

**

 

 

 

In some cases, a test failure halts the test/test printout and an error code indicating the failure appears on the display. See Chapter 4, “Troubleshooting,” for a list of self-test failure codes and failure symptoms.

Memory Test

This test looks for failures in the following memory subsystems:

Read-Only Memory (ROM)

Random Access Memory (RAM)

Non-Volatile Random Access Memory (NVRAM)

Read-Only Memory (tests 1-4) . The cardiograph is designed to have up to four ROMs installed, two on the main control board and two on an option board. The Read-Only Memory test reads the contents and performs a 32-bit cyclic redundancy check (CRC) of each of the four ROM address spaces. The test of a ROM address space will fail if the ROM module is faulty or is not present. The four ROM address spaces are identified on the report as

ROM-1

ROM-2

ROM-X1

ROM-X2

Depending on the serial number and model of the unit you are servicing, the corresponding ROM part numbers vary. Refer to tables 2-5, 2-6, 2-7, and 2-8 for the correct ROM replacement part numbers.

2-5

Performance Verification

Table 2-5

ROM Replacement Part Numbers for PageWriter Cardiographs M1770, M1771A Serial Number

 

Prefix CNA, CNB, or CNC,3807A or earlier

 

 

 

 

 

 

Test

 

ROM Part Number

Language

 

 

 

 

 

ROM 1

U205 - Main CPU Board

 

M1770-89505

 

 

 

 

 

 

ROM 2

U206 - Main CPU Board

 

M1770-89521

English

 

 

 

M1770-89521

French

 

 

 

M1770-89522

German

 

 

 

M1770-89523

Dutch

 

 

 

M1770-89524

Spanish

 

 

 

M1770-89525

Italian

 

 

 

 

 

ROM X1

U9 - Interface Board

 

M2488-17901

 

 

 

 

 

 

ROM X2

U10 - Interface Board

 

M2488-17902

 

 

 

 

 

 

Table 2-6

ROM Replacement Part Numbers for PageWriter Cardiographs M1770, M1771A Serial Number

 

Prefix CND

 

 

 

 

 

 

 

 

Test

 

ROM Part Number

Language

 

 

 

 

 

ROM 1

U205 - Main CPU Board

 

M1770-89550

English

ROM 2

 

 

M1770-89551

French

 

 

 

M1770-89552

German

 

 

 

M1770-89553

Dutch

 

 

 

M1770-89554

Spanish

 

 

 

M1770-89555

Italian

 

 

 

M1770-89556

 

 

 

Swedish

 

 

 

M1770-89558

 

 

 

Norwegian

 

 

 

M1770-89559

 

 

 

Finnish

 

 

 

 

 

 

 

 

 

ROM X1

U9 - Interface Board

 

M2488-17901

 

 

 

 

 

 

ROM X2

U10 - Interface Board

 

M2488-17902

 

 

 

 

 

 

2-6

Loading...
+ 134 hidden pages