Getinge Arjohuntleigh Parker Instructions For Use Manual

Parker
Instructions for Use
04.AL.01_11GB • 10/2017
...with people in mind
WARNING
To avoid injury, always read this Instructions for Use and accompanied documents before using the product.
Mandatory to read the Instructions for Use
Design Policy and Copyright ® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. © ArjoHuntleigh AB 2013.
As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.The content of this publication may not be copied either whole or in part without the consent of ArjoHuntleigh AB.
Content s
Foreword ................................................................................... 4
Intended Use .............................................................................5
Safety Instructions ....................................................................6
Preparations ..............................................................................7
Parts Designation ......................................................................8
Control Panel ..........................................................................10
Hand Controls .........................................................................11
Product Description/Functions ................................................12
Drain Control Knob ..............................................................12
Fill Function .........................................................................12
Auto Fill Function (option) ...................................................12
Scalding Protection .............................................................12
Temperature Display ...........................................................12
Emergency Stop ..................................................................13
Shower Handle ....................................................................13
Functions Lock ....................................................................13
Door Lock ............................................................................13
Air Spa System (option) ......................................................14
Sound & Vision® System (option) .......................................14
Disinfection Unit (option) .....................................................15
Battery Back-up ...................................................................15
Head Cushion, Back Support Cushions and
Leg Rest (accessories) ........................................................16
Bathe Patient ...........................................................................17
Bathe Patient using a Standing & Raising Aid ........................23
Bathe Patient using a Ceiling Lift ............................................26
Cleaning and Disinfection Instructions ....................................28
Care and Preventive Maintenance ..........................................42
Troubleshooting ......................................................................47
Technical Specifications .........................................................49
Labels .....................................................................................54
List of Standards and Certificates............................................56
Electromagnetic Compatibility .................................................57
Parts and Accessories ............................................................61
ArjoHuntleigh Addresses ............................................Last page
3
Foreword
Thank you for purchasing ArjoHuntleigh Equipment
Your Parker® bath 500 is part of a series of quality products designed especially for hospitals, nursing homes and other health care uses.
Please contact us if you have any questions about the operation or maintenance of your ArjoHuntleigh equipment.
Please read this Instructions for Use thoroughly!
Please read this Instructions for Use (IFU) in its entirety before using your Parker bathing system. Information in this IFU is crucial to the proper operation and maintenance of the equipment, will help protect your product, and ensure that the equipment performs to your satisfaction. Some of the information in this IFU is important for your safety and must be read and understood to help prevent possible injury .
Unauthorized modifications on any ArjoHuntleigh equipment may affect its safety. ArjoHuntleigh will not be held responsible for any accidents, incidents or lack of performance that occur as a result of any unauthorized modification to its products.
Definitions in this IFU
WARNING
Means: Safety warning. Failure to understand and obey this warning may result in injury to you or to others.
CAUTION
Means: Failure to follow these instructions may cause damage to all or parts of the system or equipment.
NOTE
Means: This is important information for the correct use of this system or equipment.
Means: The name and address of the manufacturer .
Service and Support
A service routine has to be done to your Parker bath to ensure the safety and operating procedures of your product. See section Care and Preventive Maintenance on page 42.
If you require further information, please contact ArjoHuntleigh for comprehensive support and service programs to maximize the long-term safety , reliability and value of the product.
Contact ArjoHuntleigh for replacement parts. Contact information appears on the last page of this IFU.
4
Intended Use
This equipment must be used in accordance with these safety instructions. Anyone using this equipment must also have read and understood the instructions in this Instructions for Use (IFU).
If there is anything you do not completely understand, ask your ArjoHuntleigh representative.
This bath is intended for assisted bathing and showering of adult residents in care facilities (excluding use at home).
The Parker bath must be used by appropriately trained caregivers with adequate knowledge of the care environment, its common practices and procedures, and in accordance with the guidelines in the Instructions for Use (IFU).
Any other use is prohibited. The bath should only be used for the purpose specified here.
Installation & Service Requirements
The Parker bath must be installed by appropriately trained personnel according to the Assembly and
Installation Instructions, requirements in the Assembly and Installation Instructions can only be superseded by
local code. The expected life of this equipment, unless otherwise
stated, is ten (10) years, subject to preventive maintenance being carried out in accordance with the instructions in section Care and Preventive Maintenance on page 42.
Resident Assessment
We recommend that facilities establish regular assessment routines.
Caregivers should assess each resident according to the following criteria prior to use:
The resident's weight should not exceed 210 kg/ 463 lbs (Max Patient Weight).
Active and semi-active residents (i.e. able to sit upright self supported on the side of a bed or toilet) can use the bath. Correct transfer equipment shall be used, if needed.
Residents with no capacity to support them selves at all (i.e. able to sit upright with support) can use the bath, only with a ceiling and a 2-point spreader bar with loop sling for transfer and bath.
The resident should have the ability to understand and respond to instructions, or have limited physical capacity to move, to ensure a safe position during the bathing procedure.
If a resident does not meet these criteria an alternative equipment or system shall be used.
Indication for Use
T o relieve pain and itching and as an aid in the healing process of inflamed and traumatized tissue, and it serves as a setting for removal of contaminated tissue.
5
Safety Instructions
WARNING
To avoid injury to both the patient and the caregiver, never modify the equipment or use incompatible parts.
WARNING
To avoid explosion or fire, never use the equipment in oxygen rich environment, in presence of heat source or flammable anaesthetic gases.
WARNING
To avoid electric shock, never use other electrical equipment near this device.
WARNING
To avoid injury, make sure that the patient is not left unattended at any time.
WARNING
To avoid the device in use from tipping, do not raise or lower other equipment close to it and be aware of stationary object when lowering.
6
Preparations
Site Preparations
Prior to the installation of the Parker bath the floor and walls within the required area must be finished.
Material: The floor should be made of a slip-resistant material, free of bumps.
Floor Slope: When using a floor drain, the slope towards the drain should not exceed 1:50.
Floor attachments: The floor construction must be suitable for anchoring the bolts
Actions Before First Use
(11 steps)
1 Visually check the package for damage. If the
product looks damaged due to freight, make a complaint to the transport agency.
2 The packaging should be recycled according to
local regulations.
3 Install the Parker bath by qualified personnel. See
Assembly and Installation Instructions.
4 Check that all parts of the product are supplied.
Compare with section Parts Designation on page 8. If any part is missing or damaged - do NOT use the product.
10
WARNING
To prevent scalding, make sure that the water mixer has been calibrated. Confirm that responsible organization has established calibration routines.
Make sure the Parker bath thermostatic mixer has been calibrated according to the Assembly and Installation Instructions by qualified personnel or an ArjoHuntleigh technician.
11 Make sure to have a rescue plan ready in case of a
patient emergency.
Actions Before Every Use
(3 steps)
1 Check that all parts are in place. Compare to
section Parts Designation on page 8.
2 If any part is missing or damaged - do NOT use
the product.
3
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use.
5 Unpack the hand control(s) from the plastic. 6 Read the IFU. 7 Disinfect the product according to section
Cleaning and Disinfection Instructions on page 28.
8 Perform a functionality test. See section Care and
Preventive Maintenance on page 42.
9 Choose a designated area, where the IFU should
be kept and is easily accessible at all times.
Before every use, make sure the Parker bath is cleaned and disinfected according to section
Cleaning and Disinfection Instructions on page 28.
NOTE If you have any questions, please contact your
local ArjoHuntleigh representative for support and service. The contact numbers appear on the last page of this IFU.
7
Part s Designation
Type B Applied part. Degree of Protection - Electrical shock
1 Trigger handle 2 Lock button 3 USB connection
(with Sound & Vision system option)
4 USB memory stick
(with Sound & Vision system option)
5 Disinfectant flowmeter
(with Disinfection option)
6 Sound & Vision system hand control
(with Sound & Vision system option)
7 Disinfection spray handle
(with Disinfection option) 8 Door lock handle 9 Shower stand (for GB only) 10 Filling spout
26 Compartment for disinfectant bottle
(with Disinfection option)
27 Sound & Vision system speakers
(with Sound & Vision system option)
28 Disinfection suction part
(with Disinfection option)
29 Disinfectant bottle
(with Disinfection option)
30 Lock keys for disinfectant compartment
(with Disinfection option) 31 Head cushion (accessory) 32 Back support cushions (accessory) 33 Leg rest (accessory)
11 Shower holder/shower handle 12 Emergency stop button 13 Grab handle 14 Hand control 15 Cover
(with Disinfection and/or Sound & Vision system
option) 16 Control panel 17 Door hinge 18 Sound & Vision system light
(with Sound & Vision system option) 19 Water sensory knobs (3 pcs) 20 Drain 21 Drain control knob with built-in surface overflow
outlet 22 Disinfectant/Rinsing Inlet
(with Air Spa system option) 23 Bathtub 24 Air Spa system outlets (9 pcs)
(with Air Spa system option) 25 Height adjustable feet
8
19
24
26
25
16 17
21
22
101211
9
13
8
20
18
23
2
1
MI
X
8 0
70
6 0
5 0
4 0
30
2 0
30
29
28
27
3 4
5 6
7
14 15
333231
9
Control Panel
1
89 12
2
10
11
6
7543
14 1513
1 LED for shower indication 2 Filling water and Shower temperature display
with scalding protection 3 Bath water temperature display with scalding
protection. 4 LED, lock function indicator 5 Mixing gradient 6 LED, ON/OFF disinfectant indicator (option) 7 Disinfection ON/OFF (option)
8 Fill ON/OFF (Auto fill option) 9 Shower ON/OFF 10 Air Spa system - whole tub ON/Intensity/OFF
(option)
11 Air Spa system - foot well ON/Intensity/OFF
(option) 12 Function lock ON/OFF 13 Knob for water temperature adjustment 14 LED, battery back-up indicator 15 LED, mains indicator
10
1 2
3
4
5 6
Hand Controls
1 2
4
5
8
7
6
3
Hand Control
1 Raise bathtub 2 Lower bathtub 3 Tilt bathtub to upright position 4 Tilt bathtub to reclined position 5 Air Spa system - whole tub ON/Intensity/OFF (option) 6 Air Spa system - foot well ON/Intensity/OFF (option)
Sound & Vision® System Hand Control (option)
1 LED - activity indicator 2 Sound & Vision system light - ON/OFF 3 Automatic light colour change - ON/OFF 4 MP3 player ON/PAUSE 5 Reduce (–) volume. 6 Increase (+) volume 7 MP3 track forward 8 MP3 track reverse
11
Fig. 1
Fig. 2
Fig. 3
Product Description/Functions
Drain Control Knob
The drainage plug opens and closes by turning the drain control knob in the bathtub. (See Fig. 1)
Open - turn clockwise
Close - turn counter clockwise The drain control knob is equipped with a built in surface
overflow outlet to prevent water from overflowing the tub.
Fill Function
• Press the tub filling button to start filling the footwell
or the whole tub. (See Fig. 2)
Press the tub filling button again to stop the filling.
(See Fig. 2)
Auto Fill Function (option)
A Parker bath equipped with auto fill function, prefills the footwell. (See Fig. 2)
Activate it by pressing the auto fill button.
Once the filling level is reached it stops by itself.
Press the auto fill button again and filling continues for another 60 s.
Stop filling at any time by pressing the auto fill button.
Scalding Protection
If a dangerous temperature level of 45 °C (113 °F) is reached, the electronic scalding protection shuts off the water after 10 s.
Activated scalding protection: a yellow flashing light and a beeping sound on both temperature displays.
If scalding protection turns on, get the patient out safely . S top using the bath and contact an ArjoHuntleigh service technician.
Temperature Display
The Parker bath has two temperature displays:
(See Fig. 3)
Top temperature display: shows the active filling or shower temperature. If filling or shower is not turned on, the display shows “– –. –” During disinfection the display shows the temperature of the disinfectant mixture.
The bottom temperature display: shows the current bathtub water temperature.
12
Emergency Stop
Fig. 4
T rigger
Lock Button
Fig. 5
Fig. 6
A
Press the emergency stop button, to stop all functions.
(See Fig. 4)
To release the emergency stop button, turn it clockwise.
(See Fig. 4)
If the emergency stop button was used due to an equipment error , stop using the bath and contact an ArjoHuntleigh technician.
Shower Handle
The shower handle is located in the holder or the stand.
(See Fig. 5)
Press the shower button to start the shower. The green LED will turn on when active.
Press the trigger on the shower handle to regulate the water . Hold the trigger in for constant water flow.
T o get continuous water flow without having to hold in the trigger do as follows:
Press the trigger and push the lock button on the side. The water will continue to flow without having to press the trigger.
Unlock by pressing the handle trigger and the lock button will pop out by itself.
Functions Lock
Press the function lock button on the control panel and it locks all the functions except:
scalding protection
emergency stop
For example, if the function lock is activated during a shower, the shower can only be turned off after deactivating the function lock.
The function lock LED is flashing yellow when active.
Door Lock
The Parker bath has a door handle that works as a door lock.The door handle needs to be aligned with the sticker showing an open or closed lock. (See Fig. 6)
To unlock the door: press the door handle down until aligned with the sticker showing an open lock. (A)
To lock the door: press the door handle up until the handle is aligned with the sticker showing a closed lock. (This may require some physical force). (B)
B
13
Air Spa System (Option)
MI
X
8
0
7
0
6
0
5
0
4
0
3
0
2
0
Fig. 7
A
B
A Parker bath with Air Spa system allows the patient to sit comfortably and enjoy a massage by air jets in the water .
Both the Air Spa system tub button and the Air Spa system footwell button have two adjustable intensities that can be selected. The maximum treatment time is 15 minutes.
The Air Spa system can be activated either for the whole bathtub or just in the footwell.
NOTE Operating the Air Spa system with soapy water can
cause excessive foaming.
Sound & Vision® System (option)
WARNING
To avoid electric shock, only use the USB memory stick in the connector on the bath.
The USB outlet is located behind the cover . (See Fig. 7)
Unscrew the black cap counter clockwise to get to the USB outlet (A) and plug in the Sound & Vision system memory stick (B).
T o get colored light in the tub, press the Sound & Vision system light button on the Sound & Vision system hand control.
Press the Sound & Vision system light button one more time to change light colour.
Press the Sound & Vision system light button for 2 s to turn the light off.
To start a rotation of different light colours press the Sound & Vision system automatic light colour change button.
Continue with the steps on the next page.
14
MI
X
8
0
70
6
0
50
4
0
30
2
0
Fig. 8
To start the music press the MP3 button on the Sound & Vision system hand control. Only MP3 files can be read and only files from the root (not in “folders”).
Sound & Vision system hand control LED flashes when active.
NOTE The music function will stop automatically 30 minutes
after activation. Press the MP3 player button on the Sound & Vision system hand control for continued listening.
NOTE The Sound & Vision system will stop automatically
30 minutes after activation. Press the Sound & Vision system light ON/OFF button on the Sound & Vision system hand control for continued lighting.
Disinfection Unit (Option)
The Parker bath has the a built in disinfection unit as an option.
(See Fig. 8)
There is a disinfectant bottle with an adjustable top to adjust the mixing value of the disinfectant concentration and water . It is located in the lockable compartment between the bathtub legs.
There is also a disinfection spray handle and flowmeter, located behind the cover.
T o use the disinfection unit: 1 Press the disinfection button and hold it in for 5 s to
activate it. 2 The LED will turn green when active. 3 If the disinfectant bottle is empty, the LED will start
flashing when pressing the disinfection button. A beeping
sound comes on and it is not possible to disinfect the bath. Replace with a full disinfection bottle and try again (see section
Replace Disinfectant Bottle on page 30).
15
Battery Back-up
Fig. 9
A battery will work as a back-up if there is a power failure. It is possible to help a patient out of a bathtub during that time. Everything on the bath will work under the battery back-up except for the auto fill function, Air Spa system and Sound & Vision system.
If the battery fails to work during a power failure contact your ArjoHuntleigh technician.
CAUTION Never put the patient in the bathtub when the power
indicator light on the control panel is out. Ensure that the bath is connected to a power outlet or the battery might run out of power.
Contact your local ArjoHuntleigh technician if problem occurs.
Battery charge level when the main power supply is not connected.
Y ellow LED on No mains connected
No mains connected and battery is
Y ellow LED flashing
No LED light
running low. The Battery is automatically charged when the Parker bath is connected to the mains.
Connect the Parker bath to the mains to charge the battery.
Head Cushion, Back Support Cushions and Leg Rest (accessories)
Extra cushions are available for added comfort for the patient. The back support cushions have suction cups for easy attachment to the bath. (See Fig. 9)
These accessories can be ordered from your local ArjoHuntleigh representative.
CAUTION Always fit leg rest on the opposite side of the door.
If hung on the door, damage to the door may occur.
16
Fig. 1 Fig. 2
Fig. 3
Bathe Patient
To complete a full patient bath follow these 36 steps.
WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the flow of the water away from the patient.
Fill the Footwell with Water
(Step 1-3 of 36)
1 Ensure that the bathtub is in a upright position. (See Fig. 1)
2 Close the drainage plug by rotating the drain control knob
counter clockwise. (See Fig. 2)
3 Parker bath with fill: start filling the footwell by pressing
the tub filling button. (See Fig. 3)
• Feel the water temperature with a naked hand and let the patient feel the water. Adjust the knob to a comfortable temperature.
• At desired footwell water level, stop the filling by pressing the tub filling button again.
Parker bath with Auto fill: activate prefill by pressing the auto fill button. (See Fig. 3)
• Feel the water temperature with a naked hand and let the patient feel the water. Adjust the knob to a comfortable temperature.
• The filling stops automatically when the footwell is filled.
Continue with the steps on the next page.
17
Bathe the Patient
Fig. 4
Fig. 5
(Step 4-14 of 36)
4 Open the bath door by pressing the door handle down. Make
sure no objects are in the path of opening the door. Raise the door to a fully open position.
5
WARNING
To avoid falling, make sure the patient always remains in a seated position.
Adjust the bathtub height by pressing up or down button on the hand control, in order to get the right patient seating height. (See Fig. 4)
6
WARNING
To avoid falling, make sure the patient uses the grab rails for support when entering or exiting the bath.
If needed, help the patient get seated in the Parker bath.
(See Fig. 4)
7
WARNING
To avoid falling, make sure the patient always remains in a seated position.
If needed, help the patient to place their feet into the footwell. (See Fig. 5)
8
WARNING
To avoid entrapment, make sure to keep the patients hair, arms and feet close to the body and use designated grab supports during any movement.
Close and lock the bat h do or by pulling the door handle fully up (with some physical force) until aligned with the closed lock sticker.
Continue with the steps on the next page.
18
Fig. 6
Fig. 7
9
WARNING
To avoid entrapment, make sure to keep the patients hair, arms and feet close to the body and use designated grab supports during any movement.
Recline the bathtub by pressing the tilt back button on the hand control.
10 Position the bathtub to the preferred working height by
pressing up or down buttons on the hand control. (See Fig. 6)
11 Adjust bath water depth by pressing
fill/auto fill button.
12 Adjust the water temperature by rotating the temperature
knob on the control panel.
13 Feel the water temperature with a naked hand and let the
patient feel the water as well. (See Fig. 7)
14 At a desired water level, turn the water off.
Parker bath with fill: Press filling button again when the water has reached desired level to turn it off.
Parker bath with Auto fill: The filling stops automatically after 60 s. To stop filling within this time frame, press auto fill button. If the button is pressed once again, the filling starts for another 60 s.
Shower the Patient
(Step 15-24 of 36)
15 Lift off the shower handle from the holder or stand. 16 Hold the shower handle away from the patient. 17 Press the shower button on the control panel (the LED
lights up when active).
Continue with the steps on the next page.
19
Fig. 8
Fig. 9
18
WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the flow of the water away from the patient.
Press the trigger on the shower handle directing the water stream away from the patient.
19 The top display shows the shower water temperature. Allow
a few seconds for the display to adjust the temperature. If needed, turn the knob to adjust the water temperature.
20 Also check by placing your naked hand in the water stream
that the water temperature is not to hot. (See Fig. 8)
21 Allow the patient to feel the water with his or her hand
before the water stream is directed onto the patient. If needed, adjust water temperature by turning the knob.
(See Fig. 9)
22 T o keep the trigger handle in for a longer period, activate the
lock button by pushing it in. T o deactivate, push the trigger .
23 When showering is finished, press the shower button on
the control panel.
24 Make sure the shower trigger is unlocked before placing the
shower handle back into the holder or stand.
NOTE Raise the bathtub, for optimum use of the shower
stand, and turn the shower stand away from the patient when not in use. (GB only)
Continue with the steps on the next page.
20
Air Spa System (option)
MI
X
8
0
7
0
6
0
5
0
4
0
3
0
2
0
Fig. 10
Fig. 11 Fig. 12
(Step 25-25 of 36)
25 T o activate the Air Spa system, use either the buttons on the
control panel or on the hand control.
• T o activate the Air Spa system in the whole bath, press the
Air Spa - whole bath button. The Air Spa system will
start with low intensity in the whole tub.
• If pressed a second time the Air Spa system intensity will increase.
• If pressed a third time the Air Spa system will stop.
• The above sequence apply to the Air Spa system ­footwell button as well.
Sound & Vision® System (option)
(Step 26 of 36)
26 Activate the Sound & V ision system through the USB outlet
and the Sound & Vision system hand control, which is located behind the cover. Unscrew the black cap counter clockwise to get to the USB outlet and plug in the Sound & V ision system USB memory stick. (See Fig. 10)
End the bath cycle
(Step 27-35 of 36)
27 When bathing is complete, drain the bat ht ub by opening the
drain. Turn the drain control knob clockwise.
28 With intervals, tilt the bathtub to upright position by
pressing the tilt up button on the hand control. Try to keep the patient in the water as long as possible.
29 In the mean time, keep the patient warm by showering the
upper body or covering him or her with a towel.
30 When the bathtub is in an upright position, lean the patient
forward and ask him or her to grab the handles. Shower and dry the patient’s upper body and place a towel around the patient to preserve heat. (See Fig. 11)
31 Lower the bathtub to the lowest position. (See Fig. 12)
Continue with the steps on the next page.
21
Fig. 13
Fig. 14 Fig. 15
32 Make sure that the water level is below the door opening.
(See Fig. 13)
33 Unlock and open the bath door by pulling the door handle
down and lift the door up. Make sure no objects are in the path of opening the door.
34 If necessary, gently grab the patient’s legs and dry them
before placing them on the floor. (See Fig. 14)
35
WARNING
To avoid falling, make sure the patient uses the grab rails for support when entering or exiting the bath.
Remove the patient and dry remaining wet areas of the patient’s body.
Disinfect after each Patient
(Step 36-36 of 36)
36 Make the bath ready for the next patient. The bath must be
fully disinfected after each use to prevent cross­contamination. See section Cleaning and Disinfection
Instructions on page 28. (See Fig. 15)
NOTE To preserve the door seal, leave the door slightly open
when not in use. Place the hand control on the hand control holder when not in use.
22
Fig. 1
Fig. 2
Fig. 3
Bathe Patient using a Standing & Raising Aid
Parker bath integrates with ArjoHuntleigh products according to Allowed Combinations in section Technical Specifications on page 49.
Patient Transfer from Sara 3000 to Parker Bath
(30 steps)
1 Lower the bathtub to the lowest position. 2 Start filling the footwell. (See Fig. 1)
3
WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the flow of the water away from the patient.
Check the water temperature with your naked hand.
(See Fig. 2)
4 Open the bath door . Make sure no objects are in the path of
opening the door. 5 Transfer the patient to the bathroom. 6 Place the sling around the patient and fasten it to the
standing and raising aid. 7 Activate the standing and raising aid and raise the patient. 8 Pull down the patient’s trousers or skirt. 9 Transfer the patient to the Parker bath. (See Fig. 3)
Continue with the steps on the next page.
23
Fig. 4
Fig. 5
Fig. 6
10 Lower the patient in order to place him or her on the bat htu b
seat. (See Fig. 4) 11 Detach the sling and pull the standing and raising aid away. 12 Undress the patient’s upper half. 13 Take off the patient’s trousers or skirt. 14 Let the patient feel the water temperature in the footwell.
15 If needed, gently, spin the patient around and place his or
her feet in the footwell. (See Fig. 5) 16
WARNING
To avoid entrapment, make sure to keep the patients hair, arms and feet close to the body and use designated grab supports during any movement.
Close the bath door. 17 Perform the bathing, see section Bathe the Patient on
page 18. 18 Lower the bathtub after ending the bathing cycle. 19 Position the standing and raising aid close to the Parker
bath. 20 Open the bath door . Make sure no objects are in the path of
opening the door. 21 Gently, spin the patient around and place his or her feet
outside the bath and dry them thoroughly. (See Fig. 6)
Continue with the steps on the next page.
24
22 Dress the patient’s upper body. 23 Dry the patient’s feet and legs. 24 Pull the patient’ s trousers or skirt up to his or her ankle area. 25 Put the sling on. 26 Adjust the height of the bathtub until the patient’s feet are
positioned on the standing and raising aid. 27 Attach the sling to the standing and raising aid. 28 Activate the lift and raise the patient up. 29 Dry the remaining wet areas of the patient’s body. 30 Pull the patient’s trousers or skirt all the way up.
25
Fig. 1
Fig. 2
Fig. 3
Bathe Patient using a Ceiling Lift
Parker bath integrates with ArjoHuntleigh products according to Allowed Combinations in section Technical Specifications on page 49.
Patient Transfer from a Ceiling Lift to Parker Bath
(16 steps)
1 Close door and prefill the bath with the bat htu b in an upright
position. 2 When prefilled, recline the bat ht ub and lower it to its lowest
position. (See Fig. 1) 3 Put the patient in a sling.
4
WARNING
To prevent scalding, always check the water temperature with your naked hand before directing the water on the patient. Do not use gloves as it may insulate to the extent that the water temperature can be misjudged. Point the flow of the water away from the patient.
Check the water temperature with your naked hand.
(See Fig. 2)
5
WARNING
To avoid entrapment, make sure to keep the patients hair, arms and feet close to the body and use designated grab supports during any movement.
Transfer the patient over the closed bath door to the reclined
bathtub. (See Fig. 3)
Continue with the steps on the next page.
26
Fig. 4
Fig. 5
Fig. 6
Fig. 7
6 Raise the bathtub to an ergonomic working height. Adjust
the height of the spreader bar, if needed. 7 Let there be some tension in the sling from the spreader bar
to keep the patient in place. (See Fig. 4) 8 Fill the bathtub with water to an appropriate level.
9 Make sure to feel the water temperature and adjust, if
needed. (See Fig. 5) 10 Perform the bathing (see section Bathe the Patient on
page 18). 11 Begin to drain the bathtub by opening the drain. 12 Lower the bat htub to its lowest position. 13 Lift patient slightly up from the bathtub and shower the
patient’s back. (See Fig. 6)
14 Dry the patient in the sling and transfer him or her over the
closed bath door to a preferred helping aid. Finish drying the
patient. (See Fig. 7) 15 Tilt the bathtub to upright position to drain the bath
completely. 16 Disinfect the bath to prevent cross-contamination. See
section Cleaning and Disinfection Instructions on page 28.
27
Cleaning and Disinfection Instructions
Thorough disinfection routines are necessary to prevent the build-up of bacteria on and in the equipment and to prevent cross-contamination. This is particularly vital in cases where the Parker bath is in a shared facility.
WARNING
To avoid eye and skin damage, always use protective glasses and gloves. If contact occurs rinse with plenty of water. If eyes or skin becomes irritated, contact a physician. Always read the material safety data sheet of the disinfectant.
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this
Instructions for Use.
ArjoHuntleigh Disinfectants
For optimal performance only use ArjoHuntleigh Disinfectants. If you have any questions regarding disinfecting the equipment or wish to order disinfectant, see section Parts and Accessories on page 61.
If the Parker bath is equipped with Air Spa system, the bath must be disinfected with its integrated disinfection system.
If the Parker bath does not have an integrated disinfection system, the Multi Clean disinfection system must be used.
Always disinfect the Air Spa system, even when the blower has not been used.
Accessories for Cleaning and Disinfection
Protective glasses
Protective gloves
Spray bottle with disinfectant or disinfection spray handle with disinfectant.
Spray bottle with water/shower handle with water
WARNING
To avoid eye or skin irritation, never disinfect in the presence of a patient.
CAUTION To avoid damage on the equipment only
use ArjoHuntleigh branded disinfectants.
Cloths – wet and dry
Disposable towels
Soft bristled brush
Long handled soft bristled brush
28
Microbiological Formation Prevention
In order to enhance the on-site curative actions (chlorine or thermal shock), which do not guarantee a reduction of the contamination long term, make sure to follow the instructions below .
Installation precautions
In order to limit the growth of bacterial flora, it is necessary to take action at four levels:
• Avoid stagnating water and make sure proper water circulation;
• Take precautions against the formation of scale and corrosion based on the water quality;
• Maintain a high water temperature all the way from the hot water dispenser to the different distribution pipes;
• Mix hot and cold water as close as possible to the point of use.
Use
• Make sure that the water circulates in the bath and the shower on a daily basis even if the bathtub is not used; and particular to make sure to remove any water that may be left behind in the hose.
• Let the water flow approximately 5 minutes before the first bath of the day.
• Clean and disinfect the bathtub according to the IFU before the first bath of the day and after the bath of each patient.
Do not forget to contact ArjoHuntleigh for the annual Preventive Maintenance visit.
29
Replace Disinfectant Bottle
Fig. 1
Fig. 2
MI
X
8 0
70
6 0
50
4 0
3 0
20
Arjo Clean (30)
Arjo General Purpose Disinfectant (35)
Arjo Disinfectant Cleanser IV (60)
Cen-Kleen IV (30)
Arjo Guard (15)
Fig. 3
(10 steps)
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for
Use.
1 Make sure that the Parker bath is in an upright position.
(See Fig. 1)
2 Open the compartment for disinfectant bottle. (See Fig. 2) 3 On the disinfectant bottle:
• open the bottle lock
• disconnect the suction part from the bottle.
4 Discard the old bottle and replace with a new one. 5 Open the cover and grab the disinfection spray handle. 6 Press the disinfection button for 5 s. 7 Spray disinfectant in the bath while keeping an eye on the
flowmeter.
8 Check the mix ratio on the flowmeter so that it is in
accordance with the setting range stated in this IFU.
(See Fig. 3)
Continue with the steps on the next page.
30
Fig. 4
9 If necessary, adjust the bottle cap by hand to get the right
mix ratio. (See Fig. 4)
10 Adjust water temperature so that the disinfectant mixtures
reaches approximately 20 °C (68 °F).
Clean and Disinfect with a Spray Bottle
(31 steps)
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use.
Remove visual residue (Step 1-5 of 31)
1 Raise the bathtub to convenient working position. 2 Make sure all functions are turned off. 3 Open the drain to empty any left over water. 4 Detach and place the accessories (if applicable) such as
head cushion, back support cushions and leg rest, at the bottom of the bathtub.
5 Shower and rinse the bathtub and the accessories in the
bottom of the bathtub with water to remove visible residue. Turn the accessories and rinse again.
Continue with the steps on the next page.
31
Clean
Fig. 1
Fig. 2
(Step 6-17 of 31)
6 Put on protective gloves and eye wear. 7 Mix the disinfectant according to the label instructions
stated on the disinfectant bottle.
8 Spray disinfectant in drain. Use a long handled brush to
scrub drain. 9 Close the drain. 10 Spray disinfectant using the spray bottle on all internal
bathtub surfaces and the accessories at the bottom of the
bathtub. Turn the accessories and spray on the other side.
Pay attention to parts and knobs and sensors attached to the
bathtub surface such as drain control knob/surface overflow
outlet. (See Fig. 1)
CAUTION Never use green or brown Scotch-Brite nor steel
wool on the bath as it may scratch the surface.
11 Scrub the surfaces of the bathtub, loose parts and
accessories with a soft bristle brush or cloth (to remove any
deposits e.g. skin flakes). Scrub thoroughly, especially on
handles and hand control.
(See Fig. 2)
12 Open the drain. 13 Raise the Parker bath to upright position. 14 Rinse all parts clean with plenty of water having a
temperature of approximately 25 °C (77 °F) using the
shower handle to remove disinfectant. 15 Soak a cloth in disinfectant. 16 Wipe all other contact areas such as the outside of the
bathtub (particularly behind the head), metal sensory knobs
and hand control with the cloth. 17 Wet a new cloth with water and remove all traces of
disinfectant on contact areas. The cloth should occasionally
be rinsed in running water when removing disinfectant.
Continue with the steps on the next page.
32
Disinfect
Fig. 3
Fig. 4
(Step 18-31 of 31)
18 Use a spray bottle and spray disinfectant in drain. 19 Close the drain. 20 Spray disinfectant on all internal bathtub surfaces and the
accessories at the bottom of the bathtub. Turn the
accessories and spray on the other side. Pay attention to
parts and knobs and sensors attached to the bathtub surface
such as drain control knob/surface overflow outlet.
(See Fig. 3)
21 Allow an appropriate contact time of the disinfectant
according to the instructions on the disinfectant bottle. 22 Open the drain. 23 Raise the Parker bath to upright position. 24 Push the button for shower. 25 Rinse all parts clean with water having a temperature of
approximately 25 °C (77 °F) using the shower handle to
remove the disinfectant. (See Fig. 4) 26 Soak a cloth in disinfectant. 27 Wipe all other contact areas such as the outside of the
bathtub (particularly behind the head), metal sensory knobs
and hand control with the cloth. 28 Allow an appropriate contact time of the disinfectant
according to the instructions on the disinfectant bottle. 29 Wet a new cloth with water and remove all traces of
disinfectant on contact areas. The cloth should occasionally
be rinsed in running water when removing disinfectant. 30 Leave to air dry. 31 Attach any accessories.
33
Clean and Disinfect with Disinfection Spray
MI
X
80
70
6 0
5 0
4
0
3 0
20
Arjo Clean (30)
Arjo General Purpose Disinfectant (35)
Arjo Disinfectant Cleanser IV (60)
Cen-Kleen IV (30)
Arjo Guard (15)
Fig. 1
Fig. 2
Handle
(66 steps)
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use.
Remove visual residue (Step 1-10 of 66)
1 Determining the disinfectant mixing ratio: Spray
disinfectant in the bathtub while keeping an eye on the
flowmeter. 2 Check the level of the disinfectant in the disinfectant bottle. 3 Check the mix ratio of the disinfectant on the flowmeter so
that it is in accordance with the setting stated in this IFU.
(See Fig. 1)
4 If necessary, adjust the bottle cap by hand to get the right
mix ratio. 5 Raise the bathtub to convenient working
position. (See Fig. 2) 6 Make sure all functions are turned off. 7 Open the drain and empty any left over water so that the
water sensory knobs do not detect any water. If water is
detected, the disinfection function will remain
de-activated. 8 Tilt the bathtub to an upright position. (See Fig. 2) 9 Detach and place the accessories (if applicable) such as
head cushion, back support cushions and leg rest, at the
bottom of the bathtub. 10 Shower and rinse the bathtub and the accessories in the
bottom of the bathtub with water to remove visible residue.
Turn the accessories and rinse again.
Continue with the steps on the next page.
34
Clean
MI
X
80
70
6 0
50
4
0
3 0
20
Arjo Clean (30)
Arjo General Purpose Disinfectant (35)
Arjo Disinfectant Cleanser IV (60)
Cen-Kleen IV (30)
Arjo Guard (15)
Fig. 3
Fig. 4
Fig. 5
(Step 11-38 of 66)
11 Open the cover, lift out the disinfection spray handle and
direct the disinfectant spray towards the bottom of the
bathtub. 12 Press the disinfection button for 5 s on the control
panel to turn it on. 13 Press the trigger on disinfection spray handle and spray
disinfectant in the bathtub while keeping an eye on the
flowmeter. (See Fig. 3) 14 Spray until the flowmeter shows the correct mixing value.
If the flow meter value is not correct, adjust it according to
section Cleaning and Disinfection Instructions on page 28.
(See Fig. 3)
15 Adjust water temperature so that disinfectant mixtures
reaches approximately 20 °C (68 °F). 16 Spray disinfectant in the drain. Use a long handled brush to
scrub the drain. 17 Close the drain. 18 Spray disinfectant using the disinfection spray handle on all
internal bathtub surfaces and the accessories at the bottom
of the bathtub. Turn the accessories and spray on the other
side. Pay attention to parts and knobs and sensors attached
to the bathtub surface such as drain control knob/surface
overflow outlet. (See Fig. 4) 19 Scrub the surfaces of the bathtub, loose parts and
accessories with a soft bristle brush or cloth (to remove any
deposits e.g. skin flakes). Scrub thoroughly, especially on
handles and hand control.
For Parker bath without Air Spa system continue from step
31. Additional steps 20-30 for Air Spa system only:
20 Tilt the Parker bath backwards so that all nozzles will come
in contact with disinfectant. 21 Release trigger and place the disinfection spray handle into
the disinfectant/rinsing inlet (some force may be needed).
(See Fig. 5)
22 The inlet is fitted with insert locks and can be turned either
way to be secured tightly. (See Fig. 5)
Continue with the steps on the next page.
35
Fig. 6
NOTE Disinfection should be distributed through all Air Spa
system nozzles or proper disinfection can not be guaranteed. Contact an ArjoHuntleigh technician if problem occurs.
23 Press the trigger handle and hold it until the disinfectant
flows from all Air Spa system nozzles. 24 Let it run for another 30 s. 25 Release the trigger and remove the disinfection spray
handle from the disinfectant/rinsing inlet. 26 Press the disinfection button for 5 s to turn disinfection
off. 27 Place the disinfection spray handle back in its holder. 28 Place the shower handle into the disinfectant/rinsing inlet
and turn it either way to secure it tightly. 29 Press the trigger on the shower handle and hold it until clean
water flows from all nozzles. 30 Let it run for another 30 s. 31 Open the drain. 32 Raise the Parker bath to upright position. 33 Rinse all parts clean with plenty of water having a
temperature of approximately 25 °C (77 °F) using the
shower handle to remove disinfectant. (See Fig. 6) 34 Press the shower button on the front panel to turn it off. 35 Place the shower handle back into the holder. 36 Soak a cloth in disinfectant. 37 Wipe all other contact areas such as the outside of the
bathtub (particularly behind the head), metal sensory knobs
and hand control with the cloth. 38 Wet a new cloth with water and remove all traces of
disinfectant on contact areas. The cloth should occasionally
be rinsed in running water when removing disinfectant.
Continue with the steps on the next page.
36
Disinfect
Fig. 7
Fig. 8
(Step 39-66 of 66)
39 Direct the disinfection spray handle towards the drain and
spray disinfectant in the drain. 40 Close the drain. 41 Spray disinfectant using the disinfection spray handle on all
internal bathtub surfaces and the accessories at the bottom
of the bathtub. Turn the accessories and spray on the other
side. Pay attention to parts and knobs and sensors attached
to the bathtub surface such as drain control knob/surface
overflow outlet. (See Fig. 7)
For Parker bath without Air Spa system continue from step
55. Additional steps 42-54 for Air Spa system only:
42 Tilt the Parker bath backwards so that all nozzles will come
in contact with disinfectant. 43 Release trigger and place the disinfection spray handle into
the disinfectant/rinsing inlet (some force may be needed).
(See Fig. 8)
44 The inlet is fitted with insert locks and can be turned either
way to be secured tightly. (See Fig. 8)
NOTE Disinfection should be distributed through all Air Spa
system nozzles or proper disinfection can not be guaranteed. Contact an ArjoHuntleigh technician if problem occurs.
45 Press the trigger handle and hold it until the disinfectant
flows from all Air Spa system nozzles. 46 Let it run for another 30 s. 47 Release the trigger and remove the disinfection spray
handle from the disinfectant/rinsing inlet. 48 Press the disinfection button for 5 s to turn disinfection
off. 49 Allow an appropriate contact time of the disinfectant
according to the instructions on the disinfectant bottle. 50 Place the disinfection spray handle back in its holder. 51 Place the shower handle into the disinfectant/rinsing inlet
and turn it either way to secure it tightly.
37
Continue with the steps on the next page.
Fig. 9
52 Press the trigger on the shower handle and hold it until clean
water flows from all nozzles. 53 Let it run for another 30 s. 54 Take out the shower handle from the disinfectant/rinsing
inlet. 55 Open the drain. 56 Raise the Parker bath to upright position. 57 Rinse all parts clean with water having a temperature of
approximately 25 °C (77 °F) using the shower handle to
remove the disinfectant. (See Fig. 9) 58 Soak a cloth in disinfectant. 59 Wipe all other contact areas such as the outside of the
bathtub (particularly behind the head), metal sensory knobs
and hand control with the cloth. 60 Allow an appropriate contact time of the disinfectant
according to the instructions on the disinfectant bottle. 61 Wet a new cloth with water and remove all traces of
disinfectant on contact areas. The cloth should occasionally
be rinsed in running water when removing disinfectant. 62 Leave to air dry. 63 Attach any accessories. Additional steps 64-66 for Air Spa system only: 64 Press twice on the Air Spa system button for the whole
bath, to activate the highest Air Spa system intensity. 65 Keep it running for 30 s to remove water inside the Air Spa
system jets. 66 Turn off Air Spa system by pressing the Air Spa
system - whole tub button again.
38
Clean and Disinfect Parker Bath using
Arjo Clean (30)
Arjo General Purpose Disinfectant (35)
Arjo Disinfectant Cleanser IV (60)
Cen-Kleen IV (30)
Arjo Guard (15)
Fig. 1
Fig. 2
Fig. 3
MultiClean
(33 steps)
WARNING
To prevent cross-contamination, always follow the disinfection instructions in this Instructions for Use. (See Fig. 1)
Follow steps 1- 30 after bathing each patient.
1 Check the level of the disinfectant in the disinfectant bottle. 2 Raise the bathtub to convenient working height. 3 Open the drain and make sure it is empty of water. 4 Detach and place the accessories (if applicable) such as
head cushion, back support cushions and leg rest, at the
bottom of the bathtub. 5 Shower and rinse the bath and accessories with water. 6 Lift out the spray handle from the Multi Clean disinfection
system and hold it over the bathtub. 7 Turn the ON/OFF control on the Multi Clean to the ON
position. Turn the function selection control to the
disinfection setting (yellow). 8 Adjust water temperature so that disinfectant mixtures
reaches approximately 20 °C (68 °F). 9 Apply the disinfectant by spraying all internal bath surfaces,
and the accessories in the bottom of the tub. (See Fig. 2) 10 Scrub the surfaces with a soft bristle brush or cloth to
remove any build up of biofilm which may have been
deposited.
For Parker bath without Air Spa system continue from step
16. Additional steps 11 - 15 for Air Spa system only:
11 The Parker bath must be tilted backwards to disinfect
through all nozzles. 12 Place the disinfection spray handle into the disinfectant/
rinsing inlet and turn either way to secure it (some force
may be needed). (See Fig. 3)
Continue with the steps on the next page.
39
Fig. 4
Fig. 5
NOTE Disinfection should be distributed through all Air Spa
system nozzles or proper disinfection can not be guaranteed. Contact an ArjoHuntleigh technician if problem occurs.
13 Press the trigger handle and hold it until the disinfectant
flows from all Air Spa system nozzles. 14 Let it run for another 30 s. 15 Remove the disinfection spray handle from the disinfectant/
rinsing inlet. 16 Turn the Multi Clean control to the OFF position. 17 Place the disinfection spray handle in its holder. 18 Allow an appropriate contact time of the disinfectant
according to the instructions on the disinfectant bottle,
during this time clean all surfaces. (See Fig. 4) 19 Wipe all other contact areas such as the outside of the bath
(particularly behind the head area), metal sensory knobs,
door seals and the hand control with a cloth soaked in
disinfectant. 20 Turn the function selection control on the Multi Clean to the
rinse setting (blue).
For Parker bath without Air Spa system continue from step
26. Additional steps 21-25 for Air Spa system only:
21 The Parker bath must be tilted backwards to rinse through
all nozzles. 22 Place the water spray handle into the disinfectant/rinsing
inlet and turn either way to secure it (some force may be
needed). 23 Press the trigger handle and hold it until clean water flows
from all nozzles. 24 Let it run for another 30 s. 25 Remove the shower handle from the disinfectant/rinsing
inlet. 26 Rinse the bath and the accessories. (See Fig. 5) 27 Place the spray handle back into the holder on the
Multi Clean and turn the unit off.
40
Continue with the steps on the next page.
Fig. 6
Additional steps 28-30 for Air Spa system only: 28 Press twice on the Air Spa system button for the whole
bath, to activate the highest Air Spa system intensity. 29 Keep it running for 30 s to remove residual water from
within the Air Spa system jets. 30 Turn off Air Spa system by pressing the Air Spa
system - whole tub button again.
Follow additional steps 31-33 after the last bath of the day .
31 Wipe the inner wall of the bath vigorously with a cloth
soaked in the disinfectant to remove any build up of biofilm
that may have accumulated during the day. (See Fig. 6) 32 Rinse thoroughly with water until no residual disinfectant is
visible. 33 Leave to air dry so that the bath is clean, disinfected and
ready to be used the next morning.
41
Care and Preventive Maintenance
The Parker bath is subject to wear and tear , and the following actions must be performed when specified to ensure that the product remains within its original manufacturing specification.
WARNING
To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow the recommended maintenance schedule. In some cases due to heavy use of the product and exposure to aggressive environment more frequent inspections should be carried out. Local regulations and standards may be more stringent than the recommended maintenance schedule.
NOTE Product cannot be maintained and serviced while in use with the patient.
PREVENTIVE MAINTENANCE SCHEDULE: Parker bath
CAREGIVER OBLIGATIONS Action/Check
Clean and Disinfect X Visually check all exposed parts X Check pillows/leg rest X Visually check hoses, pipes and connections X Check hand control and cable X Check door/seal X Perform functionality test X Check/clean shower heads X Exercise the thermostat mixer X Check/clean filters incoming water X Yearly checks by qualified personnel only X Check electrical installation
(to be performed by authorized electrician)
Between
Patients
Intervals according to local requirements
Every
WEEK
Every
MONTH
Every YEAR
WARNING
To avoid injury to both the patient and the caregiver, never modify the equipment or use incompatible parts.
42
Caregiver Obligations
Fig. 1
Fig. 2
Fig. 3
Caregiver obligations shall be carried out by personnel with sufficient Parker bathknowledge following the instructions in this IFU.
Between Patients
Clean and disinfect:
Before every use, make sure the Parker bath is cleaned and
disinfected according to section Cleaning and Disinfection
Instructions on page 28.
Every week
Visually check exposed parts:
V isually check all exposed parts, especially where personal contact is made by either the patient or caregiver. Make sure no cracks or sharp edges have developed that could cause the patient or user injury or that has become unhygienic and can cause cross-contamination. Any use of chlorine or alcohol can damage the bathtub and cause cracks.
Check accessories:
Check pillows/leg rest to make sure there are no cracks or tears that would allow water to get into the filling. If there is such damage, replace. (See Fig. 1)
Visually check hoses, pipes and connections:
Visually inspect for leaks of any kind by looking under the tub. (See Fig. 2)
Check hand control and cable:
Visually check the condition of the hand control and its cable. Replace as per below if cut or damaged. (See Fig. 3)
The hand control can be replaced without ArjoHuntleigh support. T o replace the hand control: Pull the cord connector out from under the back shell located behind the back seat area of the Parker bath. Reach in and plug in the new cord connector of the new hand control.
43
Continue with the steps on the next page.
Fig. 4 Fig. 5
Fig. 6
Check door/seal:
1 Open and close the door and check its supports for correct
function i.e the door should not drop by itself during closing. Check the lock for proper function. (See Fig. 4)
2 Examine the door seal for cuts, tears or distortion.
Perform functionality test (See Fig. 5):
1 Test all functions. Water, Air Spa system and tub
movement.
2 With water on: check shower hose and shower handle for
damage. Replace if damaged. The shower handle can be replaced without ArjoHuntleigh support.
3 If the shower stand on the Parker b ath for GB is damaged,
contact and ArjoHuntleigh technician.
Check the emergency stop function:
Check the emergency stop function by pushing down the button. (See Fig. 6) No filling or showering is possible. Both temperature display background lights will turn off.
Continue with the steps on the next page.
44
Fig. 7
Fig. 8
Fig. 9
A
B
C
Every month
Remove and clean the shower head (See Fig. 7):
A
1 Loosen the screw.
2 Remove the parts according to the illustration.
3 Clean carefully.
4 Re-assemble. The parts (A) must be assembled before
they can be set into the handle.
5 Tighten the screw carefully.
Exercise the thermostat mixer:
Turn the temperature control back and forth over the whole temperature range with water on. This is particularly important in areas with hard water. (See Fig. 8)
Check/clean filters for incoming water:
WARNING
Pressurised system and risk of scalding. Make sure that the valve (A) is closed before removing the cap, this is especially important for the hot water valve.
A rinsing valve with a filter is mounted on each of the two water supply hoses found under the bath. The filter in both valves should be rinsed in the order listed. (See Fig. 9)
1 Close the valves (A). The handles (red for hot water and
blue for cold) should be in a cross position.
2 Remove the cap (B) with a wrench.
3 Pull out the filter (C) and clean it carefully in water.
4 Reassemble.
5 Open the valves.
45
Yearly Checks by Qualified Personnel Only
The Parker bath must be serviced once a year in accordance with the Maintenance and Repair Manual.
T o make sure that you have a problem free and safe operation, it is recommended to have regular scheduled services with calibration of the thermostatic mixer and thermometer by a trained technician. Relocation of the bath, significant changes or modifications in local water supply may require re-calibration of the thermostatic mixing valve.
Contact your local ArjoHuntleigh service representative to sign for a service agreement.
WARNING
To avoid injury and/or unsafe product, the maintenance activities must be carried out at the correct frequency by qualified personnel using correct tools, parts and knowledge of procedures. Qualified personnel must have documented training in maintenance of this device.
NOTE All Caregiver Obligations are to be checked when
performing the Qualified Personnel Service. For details, see separate service instructions.
46
Troubleshooting
NOTE Before starting the troubleshooting check: Power supply, Water supply and the LEDs on control
panel. If the problem can not be solved with the problem solving actions below, please contact an ArjoHuntleigh authorized technician.
PROBLEM ACTION
Hand control does not work or control panel buttons do not work and the green power indicator LED is on (not the Sound & Vision system hand control).
Hand control still not working (not the Sound & Vision system hand control).
Everything but fill and Air Spa system works, the green power indicator LED is off.
The temperature displays have no backlight.
1 Check that the emergency stop is not pressed
down. If pressed down then release it by turning the button clockwise.
2 Make sure the Lock function is not active with
LED flashing. If lock function is active press the lock button for 2 s to deactivate it.
3 If the control panel works check the attachment of
the hand control. If attachment is intact, replace the hand control.
4 If attachment of the hand control is intact then
check the Air Spa system function on the control panel. If the Air Spa system buttons on the control panel work, the hand control needs to be replaced.
1 The power to the bathtub is off with the battery
LED indicator turned on. If it is running on battery, finish the bathing and get the patient out safely.
2 Contact an electrician to make sure the power
supply is working correctly before next bathing session.
Nothing works, the power indicator LED is off and temperature display is off.
Fill or shower function stops with a beeping sound. 1 The back light on the temperature display is
It is only possible to lower the lift. The lifting device of this equipment is equipped with a
1 The power is off and there is a flashing or no
battery LED indicator.
2 The battery needs to be replaced and the bat htub
will not function until the power is on. Contact qualified personnel for a battery replacement.
flashing.
2 Water is too hot and scalding protection has been
activated. There is a wrong temperature setting or water supply fault.
safety nut. If the safety nut will be activated it will only be possible to raise or lower the equipment any more. In case of this scenario stop using the equipment and contact qualified personnel for repair of the equipment.
47
PROBLEM ACTION
Air Spa system stops during operation. (Option) 1 Normal, if used more than 15 minutes.
2 If used less than 15 minutes it may be overheated
due to high ambient temperature or clogged filter. Can be restarted when cooled off.
3 Power failure.
Only the water flow to the shower is slow . Check and clean the shower head according to section
Care and Preventive Maintenance on page 42.
Water flow from shower and filling spout is slow. 1 In order to feel if any hot water is coming in, feel
the incoming hot water line to see if hot.
2 Check and clean filter for incoming water
according to section Care and Preventive
Maintenance on page 42.
Disinfection does not work. (Option, not for GB) 1 Make sure the bathtub is empty of any left over
water. If there is water in the bathtub the disinfection LED on the control panel will flash and beep.
Auto fill function will not stop filling after footwell level. (Option)
Sound & Vision system hand control is not working. (Option)
2 Make sure disinfection button is pressed in for at
least 5 s.
3 If the disinfection LED on the control panel is
flashing and beeping, it can also indicate that the level of disinfectant bottle is running low and needs to be replaced. See section Cleaning and Disinfection Instructions on page 28.
4 Make sure shower, fill or Air Spa system is not
activated and try again.
5 If power is off and the bathtub is running on
battery, the disinfection will not work.
Metal sensory plugs need cleaning because of collected build up. See section Cleaning and Disinfection Instructions on page 28.
1 The LED on the remote control should flash when
active.
2 If it is not flashing, replace the Sound & Vision
system hand control.
Sound & Vision system sound is not working, but bathtub light is working. (Option)
No Sound & Vision system light but sound is working. (Option)
1 The USB memory stick may be defective. If
possible, check the USB LED light for indication.
Solid light = power ON, Flashing light = reading. 2 Test another USB memory stick. Light is broken and it needs to be replaced by qualified
personnel.
48
Technical Specifications
A full technical description of the Parker can be found in the service documentation, please contact ArjoHuntleigh for a copy . Contact information appears on the last page of this IFU.
TECHNICAL DATA
General
Tub length in the upright position 1855 mm / 73 inches Tub length in reclined position 1960 mm / 77 1/8 inches Tub width 750 mm / 29 1/2 inches Height of door with bathtub in highest position, feet in lowest
2510 mm / 98 7/8 inches
position Maximum patient weight 210 kg / 463 lbs SWL (Safe Working Load) (patient + water) 300 kg / 661 lbs Transfer height at lowest position 565 mm / 22 1/4 inches Transfer height at highest position 685 mm / 27 inches Maximum duty cycle Lift/Hi-Lo 1 minute ON, 9 minutes OFF Maximum duty cycle Air Spa system 15 minutes ON, 15 minutes OFF Operating forces of controls (Control Panel) 4 N Operating forces of controls (Hand Control) 3 N Operating forces of controls (Sound & Vision System Hand
3 N
Control) Medical Equipment Type B
Pollution degree 2 Degree of fire protection In accordance with standard
EN/ISO 60601-1
Water
Water consumption (prefilled foot well volume) 96 l / 25.4 US gallons Filling time at 3.0 bar (43.5 PSI) dynamic pressure (prefill) 100 s Water content up to overflow level 230 l / 60.8 US gallons Water content up to auto fill level 96 l / 25.4 US gallons Emptying time from auto fill level (96 l) 3 minutes
Thermometer
Range 10-50 °C (50-122 °F) Accuracy +/-1 °C (1.8 °F) Resolution 0.1 °C (0.18 °F)
49
Motion
Raising time with full load (approximately) 16 +/- 5 s Lowering time with full load (approximately) 12 +/- 5 s
Sound Levels
Sound level at raising / lowering, no load 61 dBA Sound level at raising / lowering, maximum load 60 dBA Sound level of Air Spa system, high intensity 83 dBA Sound level tilt, no load 60 dBA
Battery
Battery Type Lead, 24 V Battery Type Number 59776
IP Classification
IP Class IP X4 IP Class hand control and Sound & Vision System Hand control IP X7
Weight
Parker bath depending on selected options (excluding packaging) 124 - 131 kg
273 - 289 lbs
Recycling
Package Wood and corrugated cardboard, recyclable The product Electric, steel and plastic parts shall be separated and recycled according
to WEEE
Battery Recyclable, Not for disposal. Environmental danger.The device contains a
lead battery that shall be disposed according to local regulations
Disposal of Parker bath
Bathtub The device is marked with the WEEE symbol to indicate that it is electronic
equipment covered by the Directive 2012/19/EU on waste electrical and electronic equipment. In European countries the crossed out wheeled-bin WEEE symbol reminds you that all the electrical and electronic products, batteries and accumulators must be taken to separate collection at end of their working life. This requirement applies in the European Union. Do not dispose these products as unsorted municipal waste. You can return your device and accessories to ArjoHuntleigh or you can contact your local authorities for advice on disposal.
50
Allowed Combinations
Parker bath integrates with the following ArjoHuntleigh products: Stedy, Sara Lite,
Sara 3000, Sara Plus, Maxi Sky 600 (with 2-point
spreader bar), Mesh sling with head support (MLA4060 XS-L, XL) and, Mesh sling without head support (MLA4531 S-L, XL)
No other combinations are allowed
SITE PREPARATION
NOTE All drainage, water and electrical installation work should be carried out by qualified personnel, in
accordance with local and national regulations.
Floor
Maximum floor load 450 kg / 0.8 m Maximum floor pressure 0.6 N/mm
Operating Conditions
2
2
Ambient temperature 10 °C - 40 °C
(50 °F -104 °F) Relative humidity 30% - 95% Rh Atmospheric pressure 800 hPa to 1060 hPa
T ransport and S torage
Ambient temperature -20 °C to 70 °C
(-4 °F to 158 °F) Relative humidity 10% - 80% at 20 °C (68 °F) Atmospheric pressure 500 hPa - 1100 hPa
51
Electrical Data: EU and North America
Voltage 230 V (EU)
120 V (North America) Frequency 50 Hz (EU)
60 Hz (North America) Power consumption (maximum) 900 VA Number of phases Single phase AC Fuses 230 V 10 A
120 V 13 A Protective earth Yes Potential equalisation Yes Power supply cord replaced by qualified personnel
Mains disconnection device
Permanently installed in wall, visible and accessible at all times Mains transient voltage 4 kV* Creepage distance 3 mm* Air clearance 1.8 mm* Direction of movement Up and down, right and left,
clockwise and counterclockwise
* If local regulations are higher, they must be met.
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point All installations must be in accordance with local codes and regulations.
52
WATER SUPPLY
Drain requirement
Flow capacity 100 l/minute
26.4 US gallons/minute
Position (floor, Wall) either
Water (Hot and Cold)
Connection 3/4” B.S.P female Upper limit static pressure 6 bar / 87 PSI Recommended dynamic pressure, at 8 °C (46 °F) and at
3 bar / 43 PSI
80 °C (176 °F) flow of 58 l/minute (15.3 US gallons/minute) Lower limit dynamic pressure at 8 °C and at 80 °C and
0.8 bar / 12 PSI
26 l/minute flow (GB) Lower limit dynamic pressure at 8 °C (46 °F) and
1 bar / 15 PSI 80 °C (176 °F) degrees and 29 l/minute (7.7 US gallons/minute) flow.(North America without disinfection)
Lower limit dynamic pressure at 8 °C (46 °F) and
2 bar / 29 PSI 80 °C (176 °F) degrees and 45 l/minute (11.9 US gallons/minute) flow. (All other markets) (North America with disinfection)
Maximum pressure difference between hot and cold water +/- 0.5 bar / +/- 7 PSI Cold water lower limit temperature 2 °C (36 °F) Cold water upper limit temperature 20 °C (68 °F) Hot water lower limit temperature 60 °C (140 °F) Hot water upper limit temperature 80 °C (176 °F)
Microbiological Demands
Fulfil WHO:s demands on drinking water and the Drinking Water Directive (DWD) in EU
53
Labels
Label explanation
Applied Part Type B, Applied part: protection
against electrical shock in accordance with IEC 60601-1.
Attention Label
Approval Label
Calibration Label
Important user information for correct use.
States approvals for the current product.
States that a calibration has to be done before first use and after every installation. See Assembly and Installations Instructions and call an ArjoHuntleigh technician.
Caution Label States a potential hazard for the
product.
Data Label States technical performance and
requirements, e.g. Input Power, Input Voltage and maximum Patient Weight etc.
Identification Label
States the product identification, serial number and year and month of manufacturing.
Product Feature Label
States the product feature name and recycling requirements.
Symbol explanation
Lifting Stroke
Label for equipotential connection point (bonding lug)
IP X4 Degree of Protection (i.e. the
product is protected against splashing water)
Air Spa system Duty Cycle
Time span of having Air Spa system turned ON and OFF.
Lift Duty Cycle T ime span of lifting the Parker
bath and rest time. Read the IFU before use
Separate electrical and electronic components for recycling in accordance with the European Directive 2012/19/EU (WEEE)
Water Supply (Water Fittings) Regulations 1999 (the Water Byelaws 2000 in Scotland)
CE marking in accordance with The Medical Devices Directive 93/42/EEC.
0086
The equipment has been EMC­tested and fulfils all requirements according to IEC 60601-1-2.
Classified by Underwriters Laboratories Inc. with respect to electrical shock, fire, mechanical, hazards and other specified hazards only in accordance with the Approvals and List of Standards.
Recyclable
Max Patient
Maximum weight of the patient
Weight Max Load Maximum weight of patient and
water Input Voltage Voltage feed Input Power Power Fuse F1
Identity of fuses Fuse F2 Fuse F3
Class 1 T ype of protection - grounded
54
Name and address of the manufacturer
Manufacturing date
8360890
8418710 A
Sound & Vision
ARJO HOSPITAL EQUIPMENT AB, ESLÖV SWEDEN
Identification Label
Caution Label
Equipotential Bonding Lug Label
Approval Label
Sound & Vision System Data Label US
Data Label EU Data Label US
Calibration Label (located on the control panel before first use)
Product Feature Label
Approval Label
Applied Part
Attention Label
Calibration Label (located on the control panel before first use)
55
List of Standards and Certificates
ST ANDARDS/ CERTIFICA TES
DESCRIPTION
EN/IEC 60601-1:2005 AMD1:2012 ANSI/AAMI ES60601-1
(2005) AMD 1 (2012)
CAN/CSA-C22.2 No.
60601.1-1:14 ANSI Z124. 1 Standards for Plastic Bathtub Units ANSI/ASME A1 12.19.7 Whirlpool Bathtub Appliances ASSE 1001
CSA B64.1.1 ANSI/ASSE 1014 Handheld Shower ANSI/ASSE 1016 Individual Thermostatic, Pressure Balancing and Combination Pressure Balancing
EN 1717 Protection against pollution of portable water in water installation and general
Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.
Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.
Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance.
Pipe Applied Atmospheric T ype Vacuum Breaker
and Thermostatic Control Valves for Individual Fixtures.
requirements of devices to prevent pollution by backflow .
WRAS Water Supply (Water Fittings) Regulations 1999 (the Water Byelaws 2000 in
Scotland) WRAS approved to the UK Water Supply Byelaws. WRAS approval to: National Health Service Model Engineering Specification D
08-TMV3. Thermostatic Mixing Valves (Healthcare Premises).
Designation Code: HP/LP-S41/T44.
56
Electromagnetic Compatibility
Medical electrical equipment needs special precautions regarding EMC (Electromagnetic compatibility) and needs to be used according to the EMC information below .
Portable and mobile RF communications equipment can affect the Parker bath.
WARNING
Stacking or placing other electrical equipment next to this device is not recommended, it can interfere with the equipment’s operation and safety. Portable and mobile radio-frequency (RF) communications equipment can interfere with this equipment operation and safety.
WARNING
The equipment may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take action, such as reorienting, relocating the equipment or shielding the location.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Parker bath is intended for use in the electromagnetic environment specified below . The customer or the user of the Parker bath should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11
RF emissions CISPR 11 Harmonic emission IEC 61000-3-2
Voltage fluctuations/flicker emission
IEC 61000-3-3
Group 1 The Parker bath uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B The Parker bath is suitable for use in all establishments,
including domestic establishments and those directly connected to the public low-voltage power supply network that
Class A
Complies
supplies buildings used for domestic purposes.
57
Guidance and manufacturer’s declaration – electromagnetic immunity
The Parker bath is intended for use in the electromagnetic environment specified below. The customer or the user of the Parker bath should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electrostatic discharge (ESD)
± 6 kV contact ± 8 kV air
± 6 kV contact ± 8 kV air
IEC 61000-4-2
Electrical fast transient/burs
IEC 61000-4-4 Surge
IEC 61000-4-5 Voltage dips,
short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
± 2 kV for power supply lines
± 1 kV for input/output lines ± 1 kV line(s) to line(s)
± 2 kV line(s) to earth <5% U
T
(>95% dip in UT) for 0.5 cycle
40% U
T
(60% dip in UT) 70% U
T
(30% dip in UT) <5% U
T
± 2 kV for power supply lines input/output lines not applicable
± 1 kV line(s) to line(s) ± 2 kV line(s) to earth
<5% U 40% U 70% U <5% U
T T T
T
(>95% dip in UT) for 5 s
(>95% dip in UT) for 5 s
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Power frequency (50/60 Hz)
Magnetic field IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE - U
is the a.c. mains voltage prior to application of the test level
T
Lift-functions, tilt-functions and temperature measurement were determined to be Essential Performance.
58
Guidance and manufacturer’s declaration – electromagnetic immunity
The Parker bath is intended for use in the electromagnetic environment specified below . The customer or the user of the Parker bath should assure that it is used in such an environment.
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC 60601 test level
3 V rms 150 kHz to
80 MHz
3 V/m 80 MHz to
2.5 GHz
Compliance level
Electromagnetic environment – guidance
10 V rms Portable and mobile RF communications equipment
should be used no closer to any part of the Parker bath, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
3 V/m Recommended separation distance
d=0.35¥P d=1.2¥P 80 MHz to 800 MHz d=2.3¥P 800 MHz to 2.5 GHz Where P is the maximum output power rating of the
transmitters in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey ,
a
should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 - At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy . To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Parker bath is used exceeds the applicable RF compliance level above, the Parker bath should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary , such as re-orienting or relocating the Parker bath.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
59
Recommended separation dist ances between port able and mobile RF communications equipment and the Parker bath
The Parker bath is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Parker bath can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Parker bath as recommended below , according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 800 MHz
d=0.35¥P
80 MHz to 800 MHz
d=1.2¥P
800 MHz to 2.5 GHz
d=2.3¥P
0.01 0.04 0.12 0.23
0.1 0.11 0.38 0.73 1 0.35 1.2 2.3 10 1.1 3.8 7.3 100 3.5 12 23 For transmitter rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where Pis the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer .
NOTE 1 - At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
60
Parts and Accessories
400.00.00.02- Grey
400.00.00.15 - Blue Back Support Cushion
For disinfectant contact local ArjoHuntleigh representative
400.00.00.09 - Grey
400.00.00.16 - Blue Head Cushion
8257376-031 Leg Rest
61
Intentionally left blank
62
AUSTRALIA ArjoHuntleigh Pty Ltd 78, Forsyth Street O’Connor AU-6163 Western Australia Tel: +61 89337 4111 Free: +1 800 072 040 Fax: + 61 89337 9077
BELGIQUE / BELGIË ArjoHuntleigh NV/SA Evenbroekveld 16 BE-9420 ERPE-MERE Tél/Tel: +32 (0) 53 60 73 80 Fax: +32 (0) 53 60 73 81 E-mail: info.belgium@arjohuntleigh.be
BRASIL Maquet do Brasil Equipamentos Médicos Ltda Rua Tenente Alberto Spicciati, 200 Barra Funda, 01140-130 SÃO PAULO, SP - BRASIL Fone: +55 (11) 2608-7400 Fax: +55 (11) 2608-7410
CANADA ArjoHuntleigh 90 Matheson Boulevard West Suite 300 CA-MISSISSAUGA, ON, L5R 3R3 Tel/Tél: +1 905 238 7880 Free: +1 800 665 4831 Institutional Free: +1 800 868 0441 Home Care Fax: +1 905 238 7881 E-mail: info.canada@arjohuntleigh.com
ýESKÁ REPUBLIKA ArjoHuntleigh s.r.o. Hlinky 118 CZ-603 00 BRNO Tel: +420 549 254 252 Fax: +420 541 213 550
DANMARK ArjoHuntleigh A/S Vassingerødvej 52 DK-3540 LYNGE Tel: +45 49 13 84 86 Fax: +45 49 13 84 87 E-mail: dk_kundeservice@arjohuntleigh.com
DEUTSCHLAND ArjoHuntleigh GmbH Peter-Sander-Strasse 10 DE-55252 MAINZ-KASTEL Tel: +49 (0) 6134 186 0 Fax: +49 (0) 6134 186 160 E-mail: info-de@arjohuntleigh.com
ESPAÑA ArjoHuntleigh Ibérica S.L. Ctra. de Rubí, 88 1ª planta - A1 08173 Sant Cugat del Vallés ES- BARCELONA 08173 Tel: +34 93 583 11 20 Fax: +34 93 583 11 22 E-mail: info.es@arjohuntleigh.com
FRANCE ArjoHuntleigh SAS 2 Avenue Alcide de Gasperi CS 70133 FR-59436 RONCQ CEDEX Tél: +33 (0) 3 20 28 13 13 Fax: +33 (0) 3 20 28 13 14 E-mail: info.france@arjohuntleigh.com
HONG KONG Getinge Group Hong Kong Ltd 1510-17, 15/F, Tower 2 Kowloon Commerce Centre 51 Kwai Cheong Road Kwai Chung HONG KONG Tel: +852 2207 6363 Fax: +852 2207 6368
IITALIA ArjoHuntleigh S.p.A. Via Giacomo Peroni 400-402 IT-00131 ROMA Tel: +39 (0) 6 87426211 Fax: +39 (0) 6 87426222 E-mail: Italy.promo@arjohuntleigh.com
MIDDLE EAST Getinge Group Middle East
Office G005 - Nucleotide Complex, Dubai Biotechnology & Research Park, P.O.Box 214742, Dubai, United Arab Emirates Tel: +971 (0)4 447 0942 E-mail: Info.ME@getinge.com
NEDERLAND ArjoHuntleigh Nederland BV Biezenwei 21 4004 MB TIEL Postbus 6116 4000 HC TIEL Tel: +31 (0) 344 64 08 00 Fax: +31 (0) 344 64 08 85 E-mail: info.nl@arjohuntleigh.com
NEW ZEALAND ArjoHuntleigh Ltd 41 Vestey Drive Mount Wellington NZ-AUCKLAND 1060 Tel: +64 (0) 9 573 5344 Free Call: 0800 000 151 Fax: +64 (0) 9 573 5384 E-mail: nz.info@ArjoHuntleigh.com
NORGE ArjoHuntleigh Norway AS Olaf Helsets vei 5 N-0694 OSLO Tel: +47 22 08 00 50 Faks: +47 22 08 00 51 E-mail: no.kundeservice@arjohuntleigh.com
ÖSTERREICH ArjoHuntleigh GmbH Dörrstrasse 85 AT-6020 INNSBRUCK Tel: +43 (0) 512 204 160 0 Fax: +43 (0) 512 204 160 75
POLSKA ArjoHuntleigh Polska Sp. z o.o. ul. Ks Piotra Wawrzyniaka 2 PL-62-052 KOMORNIKI (PoznaĔ) Tel: +48 61 662 15 50 Fax: +48 61 662 15 90 E-mail: arjo@arjohuntleigh.com
PORTUGAL ArjoHuntleigh em Portugal MAQUET Portugal, Lda. (Distribudor Exclusivo) Rua Poeta Bocage n.º 2 - 2G PT-1600-233 Lisboa Tel: +351 214 189 815 Fax: +351 214 177 413 E-mail: Portugal@arjohuntleigh.com
SUISSE / SCHWEIZ ArjoHuntleigh AG Fabrikstrasse 8 Postfach CH-4614 HÄGENDORF Tél/Tel: +41 (0) 61 337 97 77 Fax: +41 (0) 61 311 97 42
SUOMI
ArjoHuntleigh Finland c/o Getinge Finland Oy Riihitontuntie 7 C 02200 Espoo Finland Puh: +358 9 6824 1260
E-mail: Asiakaspalvelu.finland@arjohuntleigh.com
SVERIGE Arjo Sverige AB Hans Michelsensgatan 10 SE-211 20 MALMÖ Tel: +46 (0) 10 494 7760 Fax: +46 (0) 10 494 7761 E-mail: kundservice@arjohuntleigh.com
UNITED KINGDOM ArjoHuntleigh UK ArjoHuntleigh House Houghton Hall Park Houghton Regis UK-DUNSTABLE LU5 5XF Tel: +44 (0) 1582 745 700 Fax: +44 (0) 1582 745 745 E-mail: sales.admin@arjohuntleigh.com
USA ArjoHuntleigh Inc. 2349 W Lake Street Suite 250 US-Addison, IL 60101 Tel: +1 630 307 2756 Free: +1 800 323 1245 Institutional Free: +1 800 868 0441 Home Care Fax: +1 630 307 6195 E-mail: us.info@arjohuntleigh.com
ࢤࢸ࢕ࣥࢤࢢ࣮ࣝࣉ࣭ࢪࣕࣃࣥᰴᘧ఍♫ ࠛ ᮾி㒔ရᕝ༊ᮾရᕝ ࢫࣇ࢕࢔ࢱ࣮࣡ኳ⋤Ὢ) 㟁ヰ㸸
)D[㸸
www.arjohuntleigh.com
Address page - REV 20: 08/2017
ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden
www.arjohuntleigh.com
ArjoHuntleigh is a world-leading provider of integrated products and
solutions that improve the lives of patients and residents with reduced
mobility. We help healthcare facilities deliver wellness
and effective everyday care, early mobilisation, safe patient handling,
venous thromboembolism prevention, pressure injury prevention,
hygiene routines, bariatric care and diagnostics. With extensive know-
ledge and experience, we strive to improve effi ciency and ensure
a safer and dignifi ed environment for caregivers and their patients.
Loading...