As our policy is one of continuous improvement, we reserve the right to modify designs
without prior notice.The content of this publication may not be copied either whole or in
part without the consent of Arjo Hospital Equipment AB.
Your Maxi Air Transfer™ is a part of a series of quality products designed for patient transfer especially
for hospital and extended care facilities.
Please contact us if you have any questions about the
operation or maintenance of your ArjoHuntleigh
equipment.
Please read this Instructions for Use
thoroughly!
Please read this Instructions for Use (IFU) in its
entirety before using your Maxi Air Transfer. Infor-
mation in this IFU is crucial to the proper operation
and maintenance of the equipment. It will help to
protect your product, and make sure that the equipment performs to your satisfaction. The information
in this IFU is important for your safety and must be
read and understood to help prevent possible injury.
Unauthorized modifications on any ArjoHuntleigh
equipment can affect its safety. ArjoHuntleigh will
not be held responsible for any accidents, incidents
or lack of performance that occur as a result of any
unauthorized modification to its products.
Definitions in this IFU:
WARNING
Means:
Safety warning. Failure to understand and obey this
warning can result in injury to you or to others.
•
CAUTION
Means:
Failure to follow these instructions can cause damage to
all or parts of the system.
•
NOTE
Means:
This is important information for the correct use of this
system or equipment.
Service and Support
A service routine must be performed on your Maxi
Air Transfer every year by qualified personnel to
ensure the safety and operating procedures of your
product. See section, Care and Preventive Mainte-
nance.
If you require further information, please contact
ArjoHuntleigh for comprehensive support and service programs to maximize the long-term safety , reliability and value of the product.
Contact your local ArjoHuntleigh representative for
replacement parts. The telephone numbers appear on
the last page of this IFU.
Means:
The name and address of the manufacturer .
Means:
Read the Instructions for use
4
Intended Use
This equipment should only be used for the
purpose specified in this Instructions for Use (IFU). Any other use is prohibited.
The Maxi Air Transfer is intended for lateral transfer
or repositioning of hospital and care facility patients.
The equipment must be used under the supervision of
trained caregivers with adequate knowledge of the
care environment, its common practices and procedures, and in accordance with the guidelines in this
Instructions for Use (IFU).
The equipment must only be used for the purposes
stated above, and removable parts must be mounted
according to the recommendations given in this IFU.
The system
Maxi Air Transfer together with the Maxi Air Trans-
fer Flites™ (patient specific mattress) is an air
assisted lateral patient transfer system for use in lateral transfers and repositioning.
Contraindications
This equipment can be unsuitable for patients with thoracic, cervical or lumbar fractures. The equipment can
be unsuitable for patients whose body shape is such that
they do not comfortably fit between the boundaries of
the mattress.
Always make a clinical assessment to make sure that
the patient is suitable for the equipment.
Installation & Service requirements,
Maxi Air Transfer
The expected life of the Maxi Air Transfer is unless
otherwise stated five (5) years, if subject to preventive maintenance being carried out in accordance
with the instruction for Care and Preventive Mainte-nance found in this IFU.
Expected life, Maxi Air Transfer Flites
The expected life of the Maxi Air Transfer Flites is
twenty (20) transfers.
Patient assessment
It is recommended that facilities establish regular
assessment routines. Caregivers should assess each
patient according to the following criteria prior to
use:
•This equipment is intended for patients who are
dependent, unable or not required to
participate in the transfer activity.
•The equipment is intended for adult patients.
•The patient safe working load (SWL) of the Maxi Air Transfer Flites is 1200 lbs. / 544 kg.
•The weight, girth and body shape of the patient
should be assessed in terms of the potential risk
for the caregivers who are to perform the patient
transfer using this equipment.
•Maxi Air Tr ansfer and the Maxi Air Transfer Flites are intended for use in hospital and
extended care facilities.
If the patient does not meet these criteria an alternative equipment/system shall be used.
5
Safety Instructions
WARNING
Do NOT use the is equipment for any type
of lifting of patients.
WARNING
Do NOT use the equipment if the power
cord is damaged. Damaged power cord
may cause serious injury to the patient and
the caregivers.
WARNING
To avoid explosion or fire, never use the
equipment in oxygen rich environment, in
presence of heat source or flammable
anaesthetic gases.
WARNING
This equipment does not contain any
HEPA-filter. This may cause problems for
patients sensitive to contaminated air.
WARNING
T o avoid bodily injury to the patient, use the
Maxi Air Transfer in an environment where
the temperature is below 34 °C (93 °F).
Temperatures above 34 °C (93 °F) can
cause the Maxi Air Transfer Flites surface
to reach a temperature of 44°C (111°F).
6
Preparations
Head end
Foot end
Left
Right
Actions before first use (when delivered)
1Visually inspect the Maxi Air Transfer and the
Maxi Air Tr ansfer Flites for damage.
2Check that all parts are supplied. Compare with
section Parts Designation in this IFU.
If any part is missing or damaged - do NOT use
the product!
3Read the IFU.
4Disinfect the Maxi Air Transfer.
See section Disinfection Instructions.
5Perform a functionality test.
See section Care and Preventive Maintenance.
6Find a good ventilated dry area for storing the
Maxi Air Tr ansfer and the Maxi Air Transfer
Flites.
7Choose a designated area where the IFU should
be kept, accessible at all times.
8Make sure to have a rescue plan ready in case of a
patient emergency.
9If you have any questions, please contact your
local ArjoHuntleigh representative for support
and service. The contact information (Arjo-
Huntleigh addresses) is listed at the end of this
IFU.
Below is the direction of the
Maxi Air Transfer Flites:
Actions before every new patient
1Select a suitable Maxi Air Transfer Flites size
according to sections Correct placement of the patient and Product Specifications.
2Check that all parts of the Maxi Air Transfer and
the Maxi Air Transfer Flites are in place.
Compare with the section Parts Designation in
this IFU.
3Carry out a thorough inspection for damage.
4If any part is missing or damaged - do NOT use
the product!
5Enter the patient name and date on the Maxi Air
Transfer Flites label. See section Labels on the
Maxi Air Transfer Flites.
6
WARNING
To prevent cross-contamination, always
follow the disinfection instructions in
this Instructions for Use.
Make sure the Maxi Air Transfer is disinfected
according to section Disinfection Instructions.
7If you have any questions, please contact your
local ArjoHuntleigh representative for support
and service. The contact information (Arjo-
Huntleigh addresses) is listed at the end of this
IFU.
Actions before every transfer
(with same patient)
1Check that all parts are in place. Compare with the
Parts Designation in this IFU.
2Carry out a thorough inspection for damage.
3If any part is missing or damaged - do NOT use
the product!
7
Product Specifications
L (Green handles)
XL (Blue handles)
XXL (Orange handles)
Maxi Air Transfer
Dimensions: 12,5 x 7 x 7”(317,5 x 178 x 178 mm)
Weight: 11 lbs. (5 kg)
Material: Fire Retardant ABS, NBR
Air Hose Length: 78” (1980 mm)
Pow e r C o r d L e n g t h : 180” ( 4570 mm)
Maxi Air Transfer Flites
Widths: L 34” (860 mm), XL 39” (990 mm) and XXL 50” (1270 mm)
Length: 78” (1980 mm)
Material: Non-woven top, nylon back
8
Intentionally left blank
9
Applied part:Type B.
protection against electric shock in accordance
with IEC 60601-1
Maxi Air Transfer and Maxi Air Transfer Flites connection
1Connect the air hose from the Maxi Air Transfer, to either
the left or right foot end connection point of the Maxi Air Transfer Flites. (See Fig. 1)
2Attach the snap lock button on the Maxi Air T ransfer Flites
to the air hose snap lock button. (See Fig. 2)
3Use the velcro lip to secure the air hose in place. (See Fig. 3)
Operation and controls
1Plug in the power cord from the Maxi Air Transfer to an
electrical outlet.
2The LED will light up to indicate that the mains is turned on.
3T o inflate the Maxi Air Transfer Flites, press the push button
on the Maxi Air Transfer.
4To deflate the Maxi Air Transfer Flites, press the push but-
ton again.
Mains disconnection device
If for any reason, the Maxi Air Transfer does not respond to the
push button, stop inflating by pulling out the power cord from the
electrical outlet.
Do not place anything in front of the electrical outlet. Make sure
the outlet is easily accessible.
Contact qualified personnel if a malfunction occurs.
12
Correct placement of the p atient
Fig. 1
Fig. 2
Fig. 3
WARNING
Always make sure the patient has the correct size of the Maxi Air Transfer Flites. If the Maxi Air Transfer Flites is too narrow, the Maxi Ai r T ransfer Flites can tip over during transfer caus-
ing serious injury to the patient and the caregivers.
WARNING
To avoid bodily injury make sure the patient is correctly positioned on the Maxi Air Transfer Flites. If the patient is incorrectly positioned, the Maxi Air Transfer Flites can tip over during
the transfer.
Placement of the patient
•Make sure the patient is placed in the middle of the Maxi Air Transfer Flites. (See Fig. 1)
•Make sure the patient’s body does NOT reach outside the boundaries of the Maxi Air Transfer Flites.
If the feet reach outside the foot end, make sure there is enough space for the transfer. (See Fig. 2)
•The patient’s head should be placed approximately 6” (150 mm) from the top of the M axi Air Transfer Flites.
Make sure the patient rests conveniently against the pillow during the transfer. (See Fig. 3)
13
Correct placement of the Maxi Air Transfer
Fig. 1
WARNING
Before every transfer , always make sure the Maxi Air Transfer is:
•secured from movement and from falling
•not in the way of the electrical outlet in use
•not placed on the floor
Incorrectly placed Maxi Air Transfer can cause hazards e.g. interrupted inflation. This can
cause serious injury to the patient and the caregivers.
WARNING
To avoid tripping hazards, make sure that the power cord on the floor is rolled up before the
transfer.
Placement of the Maxi Air Transfer
• Hang the Maxi Air T ransfer on the receiving surface by the
attachment hooks. Make sure it is secured from movement
and from falling down. (See Fig. 1)
• Make sure the Maxi Air T ransfer is not placed in front of the
electrical outlet that is being used.
• Make sure the power cord on the floor is rolled up.
14
Transfer
Fig. 1
Fig. 2
Transfer between surfaces (21 steps)
Sending surface/side = surface the patient lays on
Receiving surface/side = surface the patient is transferred to
1Apply the brakes on the sending surface.
2Make sure the patient is in a horizontal position.
3
WARNING
To avoid the patient from falling or caregiver
from being injured, ensure that there are two or
more caregivers present during the transfer.
Position one caregiver on each side of the patient.
(See Fig. 1)
4Place the Maxi Air T ransfer Flites with the labels upwards,
underneath the patient. Follow local routines.
5Make sure the patient is correctly placed on the Maxi Air
Transfer Flites.
6
WARNING
To prevent patient from rolling off the Maxi Air Transfer Flites, the safety belts must be used
during the entire movement.
Attach the safety belts loosely. If the belts are attached too
tight, they can cause discomfort during the inflation.
Tighten them after the inflation.
(See Fig. 2)
7Adjust the receiving surface to an ergonomic height.
8Place the receiving surface as close as possible to the send-
ing surface.
continue with the steps on the next page
15
9Adjust the sending surface height slightly higher than the
Fig. 3
Receiving
Sending
surface
surface
receiving surface. (See Fig. 3)
10
WARNING
T o avoid bodily injury during the transfer , always
make sure to:
•apply the brakes on both the sending and
receiving surface
•raise and lock the bed rail on the receiving
surface
Apply the brakes, raise and lock the outer bed rail on the
receiving surface.
NOTE
If the receiving surface does not have any bed rail,
the caregiver on the receiving side is responsible to
make sure the patient does not reach outside the
boundaries of the receiving surface.
11 Hang the Maxi Air Transfer on the receiving surface.
12 Plug the power cord into an electrical outlet.
13
WARNING
To avoid sudden deflation make sure the air
hose is properly connected before transfer.
Sudden deflation can cause bodily injuries to
both the patient and caregivers.
Receiving side: connect the air hose to the Maxi Air T rans-
fer Flites.
14 Sending side: firmly hold the transfer handles during the
inflation of the Maxi Air Transfer Flites. Receiving side: turn on the Maxi Air Transfer and inflate
the Maxi Air Transfer Flites.
continue with the steps on the next page
16
15
Fig. 4
Fig. 5
Fig. 6
WARNING
To avoid bodily injury, make sure the Maxi Air Transfer Flites is completely inflated before
starting the transfer. A partly inflated Maxi Air Transfer Flites can cause the patient to hit the
underlying surface, causing bodily injury.
The pull force will increase with a partial inflated
mattress and this can cause bodily injury to the
caregiver as well.
Sending side: feel underneath the Maxi Air T ransfer Flites
to make sure that the patient can’t be felt through it.
(See Fig. 4)
If any part of the patient can be felt, the Maxi Air Transfer
Flites is partly inflated. To correct this problem:
• abort the inflation,
• reposition the patient and
• inflate again.
16 Sending side: tighten the safety belts and push the Maxi Air
Transfer Flites firmly towards the receiving surface.
(See Fig. 5)
NOTE
Use the transfer handles that are closest to the
widest parts of the patient’s body. (See Fig. 6)
continue with the steps on the next page
17
17 Receiving side: meet up the air mattress when it is halfway
Fig. 7
Fig. 8
Fig. 9
over the receiving surface. (See Fig. 7)
18
WARNING
To avoid bodily injury, make sure that the Maxi Air Transfer Flites is centred on the receiving
surface, before the Maxi Air Transfer Flites is
deflated. If not centred, the patient can fall
down.
Receiving side: make sure the Maxi Air Transfer Flites is
centred. (See Fig. 8)
19 Sending side: turn off the Maxi Air T ransfer and detach the
hose from the Maxi Air Transfer Flites.
Receiving side: firmly hold the handles until the Maxi Air
Transfer Flites is completely deflated. (See Fig. 9)
20 After the transfer is completed:
• unbuckle the safety belts,
• lock all side rails (if possible),
• remove the Maxi Air Transfer Flites and
• remove and disinfect the Maxi Air Transfer, see section
Disinfection Instructions.
18
Intentionally left blank
19
Disinfection Instructions
WARNING
Do NOT wash the Maxi Air Transfer Flites!
The Maxi Air Transfer Flites is a p atient specific product and is not intended for use
between patients.
If the Maxi Air Transfer Flites is washed, the
paper coating will disappear, revealing a
DO NOT REUSE sign.
WARNING
T o prevent cross-cont amination, always follow the disinfection instructions in this
Instructions for Use.
CAUTION
Do not lower down the Maxi Air Transfer in
disinfectant solution. This could damage
the electrical components and cause internal corrosion.
For the best results, only use
ArjoHuntleigh branded disinfectant.
If you have any questions regarding disinfecting the
equipment or wish to order disinfection fluid (see
section Parts and Accessories), contact your local
ArjoHuntleigh representative.
Accessories for disinfecting the
Maxi Air Transfer
•Protective gloves
•Protective glasses
•Spray bottle
•Disinfectant fluid
WARNING
Always use protective gloves and protective eye wear when dealing with concentrated disinfectant. Avoid getting it on your
skin or in your eyes.
If contact occurs, rinse with large amounts
of water. If skin or eyes become irritated,
consult a physician.
WARNING
To avoid eye or skin irritation, never disinfect in the presence of a patient.
CAUTION
To avoid damage on the equipment only
use ArjoHuntleigh branded disinfectant s.
•Cleaning brush
•Wash cloth
Disposable paper towels
The following procedure should be carried
out between patients.
20
Disinfection of the Maxi Air Transfer
Fig. 1
A
Fig. 2
(16 steps)
1Use protective gloves and glasses. (See Fig. 1)
2Prepare the solution for submerging the air hose.
Make sure it is mixed according to the instructions
on the disinfectant bottle label.
3Remove the air hose from the Maxi Air Transfer
by turning the hose socket (A) counter clock-wise
and pull the air hose out. (See Fig. 2)
Air hose:
4Rinse the outside of the air hose with warm water
for 10 sec.
5Use a spray bottle containing disinfection fluid
mixed according to the instructions on the disinfection bottle label. Spray the outside of air hose
with disinfectant.
6Brush the surface of the air hose with the cleaning
brush until all visible dirt is removed.
7Roll up the air hose and lower it down into the disin-
fectant solution.
8Leave the air hose in the solution for a minimum
of five minutes and a maximum of ten minutes.
9Rinse the air hose with cold water until there are
no visible traces of the disinfectant.
10 Wipe the air hose dry with disposable towels.
Maxi Air Transfer:
11 Use a damp wash cloth (warm water) to wipe off all
surfaces on the Maxi Air Transfer.
12 Use a spray bottle containing disinfection fluid
mixed according to the instructions on the disinfection bottle label. Spray the Maxi Air Transfer
with the disinfectant and allow a disinfection time
according to the instructions on the disinfectant
bottle label.
13 Make sure all surfaces are disinfected and rub
them with the wash cloth.
14 Use a new damp wash cloth (cold water) to wipe off
all surfaces on the Maxi Air Transfer until there
are no visible traces of the disinfectant.
15 Wipe the Maxi Air Transfer dry with disposable
towels.
16 Insert the air hose socket to the Maxi Air T ransfer
socket inlet. Make sure it clicks into place.
21
Care and Preventive Maintenance
The Maxi Air T ransfer and the Maxi Air Transfer Flites are subject to wear and tear, and the following actions must be
performed to make sure that the product remains within its original manufacturing specification.
WARNING
To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow
the recommended maintenance schedule. In some cases due to heavy use of the product and
exposure to aggressive environment more frequent inspections should be carried out. Local
regulations and standards may be more stringent than the recommended maintenance
schedule.
WARNING
T o avoid injury to both p atient and caregiver , never modify the equipment or use incomp atible
parts.
Preventive maintenance schedule for Maxi Air Transfer and the Maxi Air Transfer Flites.
CAREGIVER OBLIGA TIONS Action/Check
DisinfectX
Visually check all exposed partsX
Perform functionality testX
Check/change the air filterX
Between
PATIENTS
Every
TRANSFER
Every
WEEK
Every
MONTH
WARNING
To avoid injury and/or unsafe product, the maintenance activities must be carried out at the
correct frequency by qualified personnel using correct tools, parts and knowledge of procedure. Qualified personnel must have documented training in maintenance of this device.
QUALIFIED PERSONNEL Action/Check
Every
YEAR
Change the air filterX
Perform functionality testX
NOTE
All Caregiver Obligations are to be checked when performing the Qualified Personnel Service.
22
Caregiver Obligations
Caregiver obligations shall be carried out by personnel with
sufficient Maxi Air Transfer and the Maxi Air Transfer Flites
knowledge following the instructions in this IFU.
Between Patients
Disinfect the Maxi Air Transfer. Make sure the Maxi Air
Transfer is disinfected between patients according to section
Disinfection Instructions.
Every transfer
•Visually check all exposed parts for;
-damages,
-tears and
-unhygienic signs.
Check especially the Maxi Air Trans fer Flites since it has been
in contact with both the patient and the caregivers.
Every week
•Perform functionality test. Check the;
-on/off function of the Maxi Air Transfer,
-air hose connection at the Maxi Air Transfer and Maxi Air Transfer Flites and
-air hose for damage.
Every month
Check/change the air filter. Check the air filter for dirt.
If needed, replace it. See section Check/change the air filter.
Yearly
Maxi Air Transfer has to be serviced according to the section
Care and Preventive Maintenance schedule by qualified per-
sonnel.
23
Check/change the air filter
Fig. 1
A
Fig. 2
C
B
Fig. 3
D
Check/change the air filter (6 Steps)
1Fold in the attachment hooks (A). (See Fig. 1)
2Pull up the rubber ring (B). (See Fig. 2)
3Pull down the cover (C). (See Fig. 2)
4Check the air filter (D) for dirt. (See Fig. 3)
5If the air filter is dirty, replace it. Do not use the Maxi Air
Transfer until the air filter has been changed.
6Attach the cover (C) and the rubber ring (B). (See Fig. 2)
If the product does not work as intended, immediately contact
your local ArjoHuntleigh representative for support.
24
Troubleshooting
Troubleshooting for Maxi Air Transfer and the Maxi Air Transfer Flites
PROBLEM ACTION
•Check power supply.
•Check power cord connections on the Maxi Air
No air flow.
Transfer and the electrical outlet.
•Check air hose connections between the Maxi Air Transfer and the Maxi Air Transfer Flites.
•Check the air hose for damage.
Low pressure in the Maxi Air Transfer Flites.
NOTE
If the problem cannot be solved with the written troubleshooting actions, please contact qualified
personnel.
•Check the Maxi Air Transfer Flites for damage.
•Make sure the air filter is clean.
25
Technical Specifications
Technical Data
Maxi Air Transfer
Weight 11 lbs. (5 kg)
Expected life5 years from manufacturing date
Protection classNot protected against ingress of water
Pollution degree2
Operating force: Push button 30.0 N
Maximum duty cycleFor continuous use
Sound Level74 db (A)
MRI classificationMR unsafe
Motor power 120 V 1100 W (USA)
230 V 1200 W (Rest of the world)
Maxi Air Transfer Flites
Safe working load (SWL)
(Maximum patient weight)
Weight 2 lbs. (1 kg)
Expected life 20 transfers
Shelf life Estimated shelf life (2 years) starts from the
Medical equipmenttype
MRI classificationMR Safe
Radiolucency classificationRadiolucent
Latex contentLatex free
1200 Ibs. (544 kg)
manufacturing date stated on LOT label on
the outer package.
WARNING
To avoid electric shock, make sure that the equipment is connected to:
• Continuously powered supply mains with protective earth.
• Separate fuse and ground fault circuit interrupter (GFCI)
• A mains disconnection device
• Equipotential bonding point
All installations must be in accordance with local codes and regulations.
26
Allowed Combinations
ProductsCompatible withModel
Maxi Air Transfer ArjoHuntleighMaxi Air Transfer Flites L, XL, XXL
Maxi Air Transfer
Flites L, XL, XXL
No other combinations are allowed.
Environmental Conditions
Operating
Temperature+10°C to + 34°C (+50 to +93°F)
Humidity10-70% at +20°C (+68°F)
ArjoHuntleighMaxi Air Transfer
HovertechAIR200G
(only compatible with Maxi Air Transfer Flites size L)
Manufactured before 20120101
HovertechAIR400G
Manufactured before 20120101
HovertechHT AIR1200
Manufactured before 20120101
AirPalPA-1200
Manufactured before 2011
Atmospheric pressure700 hPa to 1060 hPa
Transport and Storage
Temperature-40°C to +80°C (-40°F to +176°F) Maxi Air T ransfer
-17°C to +43°C (+0°F to +110°F) Maxi Air Transfer Flites
Humidity10-70% at +20°C (+68°F)
Atmospheric pressure500 hPa to 1060 hPa
27
Recycling
The device should be recycled according to national regulations.
PackageWood and corrugated cardboard
recyclable
Maxi Air TransferElectric, metal and plastic parts shall be separated and
recycled according to (WEEE) and according to markings on the unit.
Disposal (Environmental Protection)
Maxi Air TransferThe device is marked with the WEEE symbol to indicate
that it is electronic equipment covered by the Directive
2002/96/EC on waste electrical and electronic equipment.
In European countries the crossed out wheeled-bin
WEEE symbol reminds you that all the electrical and
electronic products, batteries and accumulators must be
taken to separate collection at end of their working life.
This requirement applies in the European Union. Do not
dispose these products as unsorted municipal waste.
You can return your device and accessories to ArjoHuntleigh.
You can also contact your local authorities for advice on
disposal.
Maxi Air Transf er FlitesThe hospitals should follow local routine for disposal of
contaminated material.
28
Intentionally left blank
29
Labels on Maxi Air Transfer Flites
TEX
REF
T ype B, Applied part: protection against electrical
shock in accordance with
IEC 60601-1.
Follow Instructions for Use
Read the Instruction for
Use before use
LOT number
MR-Safe
Foot end of mattress
Manufacturing date
The shelf life (2 years) of
the Maxi Air Transfer Flites
starts from the manufacturing date on the LOT label
on the outer package
LALAEX
Latex free product
Model number
Do notwash
Do not reuse
Appears if the Maxi Air Transfer Flites is washed
The name and address of
the manufacturer.
30
Maxi Air Transfer Flites
MAS014000-WW
LOT
Patient Name:
Date:
L
Data label
31
Labels on Maxi Air Transfer
A
B
C
Fig. 1
A1
A2
D
IP2XProtection class: The
product must only be used
in a dry location.
Class IInsulation class
V ~ HzVoltage and AC frequency
WMotor power
Separate electrical and
electronic components for
recycling in accordance
with the European Directive 2002/96/EC (WEEE)
Follow Instructions for
Use
The equipment has been EMC-tested and fulfils all
requirements according to IEC-60601-1-2.
Identification label
The identification label is found under the top lid.
It states the product number, the serial number and the
manufacturing date.
Position of Identification label
Remove the attachment hooks, handle and lid
(See Fig. 1)
1Turn the screws (A) to an unlocked (A1) position.
2Remove the attachment hooks and handle (B)
by pulling the handle upward and side to side at
the same time.
3Remove the lid (C) to find the Identification
label (D).
XXXXXXXXXXSerial Number
(located inside)
Push button
MR unsafe
Classified by Underwriters
Laboratories Inc. with
respect to electrical
shocks, fire and mechanical, hazards and other
specified hazards only in
accordance with the
Approvals and List of
Standards. See Approv-
als and List of Standards
Attach the attachment hooks, handle and lid
(See Fig. 1)
1Attach the lid (C).
2Push down the attachment hooks (B) and handle.
Make sure they click into place.
3Turn the screws (A) to a locked (A2) position.
4ML5Certificate control number
for UL.
The name and address of
the manufacturer.
32
Product no
XXX XXXX-
Serial no
XX
XXXXXXXXXX
Made by Arjo in Eslov, Sweden
YYMMDD
Identification Label
Push button
Maxi Air Transfer
Product Name
MR unsafe
Company Logo
Maxi Air Transfer, Model: MASXXXXXX-US
ArjoHuntleigh AB,Verkstadsvägen 5, 241 38 Eslöv (SWEDEN). Made in Sweden
Data label
33
Approvals and List of Standards
CERTIFICATESTANDARDS
UL EN/IEC 60601-1: 2005, Medical electrical equipment.
General requirements for safety.
UL CAN/CSA-C22.2 No. 601.1, Medical equipment Certified for Canada
34
Electromagnetic comp atibility
Medical electrical equipment needs special precautions regarding EMC (Electromagnetic compatibility) and needs to
be used according to the EMC information below . The Maxi Air Transfer is intended for use in the electromagnetic
environment specified below . The customer or the user of the Maxi Air Transfer should make sure that it is used in
such an environment.
WARNING
Stacking or placing other electrical equipment next to this device is not recommended, it can
interfere with the equipment’s operation and safety. Portable and mobile radio-frequency (RF)
communications equipment can interfere with this equipment operation and safety.
WARNING
The equipment may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take action, such as reorienting, relocating the equipment or
shielding the location.
Guidance and manufacturer’s declaration - electromagnetic emissions
Voltage fluctuations/
flicker emissions
IEC-61000-3-3
Group 1The Maxi Air Transfer uses RF energy only for its internal func-
tion. Therefor, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment. RF emissions.
Class B
Class B
Complies
The Maxi Air Transfer is suitable for use in all establishments
other than domestic and those directly connected to the public low
voltage power supply network that supplies buildings used for
domestic purposes.
35
Guidance and Manufacturer’s declaration - Electromagnetic Immunity
The Maxi Air T ransfer is intended for use in the electromagnetic environment specified below. The customer or
the user of the Maxi Air Transfer should ensure that it is used in such an environment.
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment
Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical fast
Transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage variations on power
supply input lines
IEC 61000-4-11
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input-output
± 1 kV line(s) to
line(s)
± 2 kV line(s) to PE
< 5% UT
(>95% dip in UT)
For 0,5 cycle
40% UT
(60% dip in UT)
70% UT
(30% dip in UT)
< 5% UT
(>95% dip in UT) for
five seconds
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input-output
± 1 kV line(s) to line(s)
± 2 kV line(s) to PE
< 5% UT
(>95% dip in UT)
For 0,5 cycle
40% UT
(60% dip in UT)
70% UT
(30% dip in UT)
< 5% UT
(>95% dip in UT) for
five seconds
Floors should be wood, concrete
or ceramic tile. If floors are covered with synthetic material, the
relative humidity should be at
least 30%
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
If the user of the Maxi Air Trans-fer requires continued operation
during mains interruptions, it is
recommended that the Maxi Air Transfer be powered from an
interruptible power supply or battery.
Power Frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE
UT is the AC mains voltage prior to application of the test level.
3A/m3A/mPower frequency magnetic fields
36
should be at levels characteristic
of a typical commercial or hospital environment.
Guidance and Manufacturer’s declaration - Electromagnetic Immunity
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity TestIEC 60601 Test LevelCompliance LevelElectromagnetic Environment Guid-
ance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz to 80
MHz
3 Vrms 80 kHz to 2,5
GHz
3V
3 Vm
Portable and mobile RF communications equipment should be used no
closer to any part of the Maxi Air Transfer including cables, than the
recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d=1,2√P
d=1,2√P 80 MHz to 800 MHz
d=2,3√P 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the recommended separation distance in
meters (m)
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a) should be
less than the compliance level in
each frequency range. b) Interference may occur in the vicinity of
equipment marked with following
symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
37
Parts and Accessories
Power cord
US 8563082
Cover
86 62963
Air hose
8662971
For disinfectant, contact
your local representative.
Air filter
6305180
Maxi Air Transfer Flites
L, XL, XXL
38
AUSTRALIA
ArjoHuntleigh Pty Ltd
78, Forsyth street
O’Connor
AU-6163 Western Australia
Tel: +61 89337 4111
Free: +1 800 072 040
Fax: + 61 89337 9077
C. Psimitis Co Ltd
Dimitriou Andr. 59
GR-16121 KAISARIANI ATTIKIS
Τηλ: 21 0724 36 68
Φάξ: 21 0721 55 53
ESPAÑA
ArjoHuntleigh Ibérica S.L.
Ctra. de Rubí, 88 1ª planta - A1
08173 Sant Cugat del Vallés
ES- BARCELONA 08173
Tel: +34 93 583 11 20
Fax: +34 93 583 11 22
E-mail: info.es@arjohuntleigh.com
FAR EAST
ARJO Far East Limited
Unit 3A, 4/F., block B Hoi Luen
Industrial Centre
55 Hoi Yuen Road,
Kwun Tong, Kowloon
HONG KONG
Tel: +852 2508 9553
Fax: +852 2508 1416
ArjoHuntleigh UK
ArjoHuntleigh House
Houghton Hall Park
Houghton Regis
UK-DUNSTABLE LU5 5XF
Tel: +44 (0) 1582 745 700
Fax: +44 (0) 1582 745 745
E-mail:
sales.admin@ArjoHuntleigh.com
USA
ArjoHuntleigh Inc.
2349 W Lake Street Suite 250
Addison, IL 60101
Tel: +1 630 307 2756
Free: +1 800 323 1245 Institutional
Free: +1 800 868 0441 Home Care
Fax: +1 630 307 6195
E-mail: us.info@ArjoHuntleigh.com
GETINGE GROUP is a leading global provider of products and systems that
contribute to quality enhancement and cost efficiency within healthcare and life
sciences. We operate under the three brands of ArjoHuntleigh, GETINGE and
MAQUET. ArjoHuntleigh focuses on patient mobility and wound management
solutions. GETINGE provides solutions for infection control within healthcare and
contamination prevention within life sciences. MAQUET specializes in solutions,
therapies and products for surgical interventions and intensive care.
www.ArjoHuntleigh.com
ArjoHuntleigh AB
Verkstadsvägen 5
241 38 Eslöv
SWEDEN
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.