Fresenius Injectomat 2000 MC Service Manual

Page 1
TECHNICAL MANUAL
Injectomat 2000 MC
Page 2
Page 3
TABLE OF CHANGES
The information given in this document only concern devices of Injectomat 2000 MC / Pilot Delta
Technical reference N° ...............................................NT 1023
Revision date: ............................................................02/07/02
Applicable from serial N°.............................................18091461
Date Revision no. Pages concerned Changes
Reference Injectomat 2000 MC / Pilot
02/07/02 B0 55 & 61
Delta RS232 HE13 power supply board,
Equivalence table
Page 4
Page 5
TABLE OF CONTENT
1 OVERVIEW.......................................................................................... 5
1.1. Block diagram .............................................................................................................. 5
1.2. Precautions before use................................................................................................6
1.3. Overall product specifications...................................................................................... 6
1.3.1. Biological specifications ............................................................................................ 6
1.3.2. Mechanical specifications.......................................................................................... 6
1.3.3. Dimensions................................................................................................................ 6
1.3.4. Electrical specifications.............................................................................................6
1.3.5. Electronic specifications............................................................................................ 6
1.3.6. Injectomat 2000 MC / Pilot Delta Operator's guide................................................... 6
2 ELECTRONIC BOARD........................................................................7
2.1. MOTOR POWER SUPPLY AND CONTROL BOARD................................................. 7
2.1.1. Functional description ............................................................................................... 7
2.1.2. Description of connectors........................................................................................ 14
2.1.3. Electrical layout....................................................................................................... 16
2.1.4. Installation layout..................................................................................................... 16
2.2. CPU BOARD.............................................................................................................. 16
2.2.1. Functional description ............................................................................................. 16
2.2.2. Description of connectors........................................................................................ 19
2.2.3. Electrical layout....................................................................................................... 21
2.2.4. Installation layout..................................................................................................... 21
2.3. DISPLAY BOARD ...................................................................................................... 21
2.3.1. Overview................................................................................................................. 21
2.3.2. Functional description ............................................................................................. 21
2.3.3. Description of connectors........................................................................................ 23
2.3.4. Power consumption................................................................................................. 24
2.3.5. Electrical layout....................................................................................................... 24
2.3.6. Implantation layout..................................................................................................24
3 Configurations..................................................................................24
3.1. Configuration of the pressure functionalities.............................................................. 24
3.2. Configurations of the various parameters.................................................................. 24
3.3. CHECKING THE PILOT............................................................................................. 29
3.3.1. The After Sale Service test ..................................................................................... 29
3.3.2. Running time tESt.1............................................................................................ 30
3.3.3. Lights test tESt.2................................................................................................30
3.3.4. Key board test tESt.3 .......................................................................................... 30
3.3.5. Battery voltage display tESt.4 ............................................................................. 31
3.3.6. Last 10 alarms codes tESt.5 .............................................................................. 31
3.3.7. Total running time tESt.6.................................................................................... 32
3.3.8. TTL Serial link test: tESt.7................................................................................... 32
3.3.9. Serial link test: tESt.8.......................................................................................... 32
3.3.10. Strength on the plunger display: tESt.9............................................................ 33
3.3.11. Software version tESt.A.....................................................................................33
3.3.12. Analog input display tESt.B ............................................................................... 33
3.3.13. Driving block position display tESt.C ................................................................. 34
3.3.14. Calibration values display tESt.E.......................................................................34
3.3.15. Syringe type display tESt.F ............................................................................... 34
3.3.16. Displays of the last 10 events before the last blocking error tESt.J.................. 34
3.3.17. Drug library tESt.L............................................................................................. 34
4 REPLACING SUB-ASSEMBLIES.....................................................34
NT 1023 Rev.B0 page : 1
Page 6
4.1. Mounting the flexible circuit........................................................................................ 35
4.2. Wiring the components on the flexible circuit............................................................. 35
4.2.1. Mounting the potentiometer.....................................................................................36
4.2.2. Wiring the potentiometer.........................................................................................37
4.2.3. Plug holder connector ............................................................................................. 37
4.2.4. Lubricating the mechanical parts............................................................................. 37
5 MAINTENANCE.................................................................................39
5.1. Recommendations .....................................................................................................39
5.2. Cleaning and disinfection...........................................................................................39
5.3. Storage.......................................................................................................................39
5.4. Routine inspections.................................................................................................... 39
5.4.1. Checking before use ...............................................................................................40
5.4.2. Checking protocol (Technical manual).................................................................... 40
5.5. Technical check procedure PILOT anaesthesia 2 .....................................................42
5.6. Checking disengagement system ..............................................................................43
5.7. Checking force sensor ............................................................................................... 44
5.8. Checking back-pressure ............................................................................................44
5.9. Checking registered syringe list /syringe list label...................................................... 45
5.10. Checking Mains/Battery operation ........................................................................... 45
5.11. Checking linearity..................................................................................................... 45
5.11.1. Equipment used .................................................................................................... 45
5.11.2. Operating mode..................................................................................................... 45
5.12. Checking end of infusion..........................................................................................46
5.12.1. End of infusion pre-alarm......................................................................................46
5.12.2. End of infusion alarm............................................................................................. 46
5.13. Checking the Led's and keyboard............................................................................ 47
5.14. Battery autonomy test...............................................................................................47
5.15. Continuity test........................................................................................................... 47
5.16. Trouble Shooting...................................................................................................... 48
5.17. Error message Er(-)0, Er01, Er(-)2, Er03, CFPc...................................................... 49
5.18. Flow rate control protocol: flow rate measurement with computer...........................49
5.18.1. Equipment used: ................................................................................................... 50
5.18.2. installation..............................................................................................................50
5.18.3. Operating mode..................................................................................................... 51
5.18.4. Installation drawing................................................................................................ 51
5.19. Flow rate control: flow rate control measurement with scales.................................. 51
5.19.1. Equipment used: ................................................................................................... 51
5.19.2. Installation .............................................................................................................52
5.19.3. Operating mode..................................................................................................... 52
5.20. Flow rate control: flow rate measurement using a test tube..................................... 53
5.20.1. Equipment used .................................................................................................... 53
5.20.2. Installation .............................................................................................................53
5.20.3. Operating mode..................................................................................................... 53
6 ANNEX 1: ILLUSTRATED PARTS LIST........................................... 54
6.1. Subassembly traceability table................................................................................... 54
6.1.1. Introduction.............................................................................................................. 54
6.1.2. Replacement parts table ......................................................................................... 54
6.2. Exploded views and related parts lists .......................................................................55
6.2.1. Mechanical part list.................................................................................................. 55
6.2.2. Electronical parts list ............................................................................................... 60
6.2.3. Operator’s guide of Injectomat 2000 MC / Pilot Delta(NU)...................................... 61
Page : 2 NT 1023 Rev.B0
Page 7
6.2.4. Labels......................................................................................................................61
7 ANNEX 2: ELECTRONIC LAYOUT................................................... 62
7.1. Rear door wiring.................................................... Fehler! Textmarke nicht definiert.
7.2. Power supply and control board.................................................................................62
7.2.1. Electronic layout...................................................................................................... 62
7.2.2. Installation layout..................................................................................................... 62
7.3. CPU board.................................................................................................................. 62
7.3.1. Electronic layout...................................................................................................... 62
7.3.2. Installation layout..................................................................................................... 62
7.4. Display board ............................................................................................................. 62
7.4.1. Electronic layout...................................................................................................... 62
7.4.2. Installation layout..................................................................................................... 62
8 ADDENDA AND INFORMATION BULLETINS.................................... 1
9 USEFUL ADDRESSES........................................................................4
NT 1023 Rev.B0 page : 3
Page 8
Page : 4 NT 1023 Rev.B0
Page 9
1 OVERVIEW
keeper
1.1. Block diagram
Ext
12/15V
15W
230V
EPROM
512K x 8
UART
RAM
8K x 8
interface
Power supply
Battery
EEPROM
16K
Bus
BUS
SPI
ON / OFF
CPU
Time
DC-DC
Converter
Watch
dog
opto
Interface
ADC
Motor driver
Step
by step
motor
Motor rotation sensor
Syringe barrel sensor
Antisiphon sensor
Occlusion strength sensor
Displacement sensor
Nurse call (option)
RS232
Keyboard
Master
Buzzer
LCD driver
LED driver
LCD
Display
LED
Display
Disengagement Switch
Flange Switch
NT 1023 Rev.B0 page : 5
Page 10
1.2. Precautions before use
Please consult the user guide
1.3. Overall product specifications
1.3.1. Biological specifications
Infusion liquid only comes into contact with the syringe and associated disposable.
1.3.2. Mechanical specifications
Device operation is based on a "lead screew/nut" principle. The mechanism pushes the piston of a sy­ringe, of a given diameter, in a linear manner.
1.3.3. Dimensions
q H x L x D 120 x 330 x 155 mm. q CarrWeight 2.2 kg approximately.
1.3.4. Electrical specifications
q Power supply 230V - 50-60 Hz.(Check on the pump the identification label). q Maximum consumption 23 VA. q Fuse F2 100 mAT 250V IEC 127 q Battery 6V - 1.2Ah./1.3 Ah q External power supply 12 - 15 DC - 15W
1.3.5. Electronic specifications
The Injectomat 2000 MC / Pilot Delta syringe pump is fitted with 3 circuit boards whose features vary in line with product specifications and options.
q Motor power supply and control board. q CPU board. q Keyboard display board.
1.3.6. Injectomat 2000 MC / Pilot Delta Operator's guide
Operator's Guide can be obtained from our After Sales Service (see chapter 10.Useful addresses).
Page : 6 NT 1023 Rev.B0
Page 11
2 ELECTRONIC BOARD
2.1. MOTOR POWER SUPPLY AND CONTROL BOARD
2.1.1. Functional description
In order to simplify wiring the motor power supply control board concentrates 6 functional modules, which may be described separately:
q power supply module, q motor control module, q analog output module, q disengage and anti-siphon opto switch module, q microswitch input module, q optional nurse call and RS232 interface module.
2.1.1.1. Power supply module The power supply module consists of a cut-out power unit. It supplies all electronic components and
charges a 1,1 -1,2 or 1,3 Ah backup battery from a mains voltage input or a 12/15 volt DC power source. It generates the + 5V and Vbat voltage required by the electronic components. Finally it comprises an ON/OFF switch controlling the various power supplies.
2.1.1.1.1. Mains power supply
The mains power supply generates a DC voltage ranging from 10 and 16 volts for a maximum current of 1,2 A.
J1 mains input connector: 1 phase
2 neutral Transformer: TR1(see Electrical chart), 15VA output voltage: 9 Vac Fuse protection: F2 Principal characteristics chap1.1 Primary filtering 4.7 nf 4000 V HR capacitor, type DS1510 VDE Secondary filtering C10 Chemical Capacitor
Output voltage measured on TP3 for mains voltage: 230V measured (±± 10%)
Min Max Unit power off: 14 16 V dc 7 ohm charged on J4: 10 16 V dc primary current charged: 80 mA ac Maximum voltage 16 volts limited by the diode D 41 Minimum voltage 10 volts limited by the mains voltage - 10% and U1(MAX 652)
voltage higher than 10 volts.
NT 1023 Rev.B0 page : 7
Page 12
2.1.1.1.2. External 12-15V AC/DC power supply
The DC power supply input is designed to provide the syringe pump with a constant, external power source, such as a 12V battery.
Maximum input voltage ± 15 volts protection against polarity inversion by the PR2 WO4 diode bridge. Minimum input voltage ± 11 volts 1.2 A limited by input voltage MAX 625 and loss, diode bridge on the PR2. Limitation ± 16 volts maximum for through D 41.
2.1.1.1.3. Cut-out charger/controller
The controller is powered either from the mains or from an external DC power source. It generates a maximum voltage of 6.9V VBC, as required to charge the 1.1/1.2 Ah gelified lead battery, connected to J4, and power the electronic components.
J4 Connector: 1 battery +
2 battery -
VBC power comes directly from the battery if neither the mains nor the external power source are con­nected. Otherwise power, from an external source, supplies the electronic components and charges the battery via diode D8 and the delayed protection fuse, F1, 1.6A.
Maximum fuse resistance 0.5 Ohms Controller operation is indicated by two signals responsible for reporting operation using an external
power source either mains power or the external DC power unit. LDSECT 10 mA drives a diode which checks that the SECT diode is on, using a TTL signal, with + 5V
pull up collector open, mains presence active at 0. Cut-out controller: U1(MAX625) output voltage 7.05V 1.3 A min Output voltage measured on J4.1 for 230 V mains:
Min. Max. Unit Power OFF 3 mA charge on J4. : 6.7V 7V V DC 8 ohm charge on J4: 6.5V 7V V DC
On J4 the voltage must never exceed 7V, the maximum voltage of the charged battery. The 6.5 minimum voltage is higher than the battery pre-alarm threshold.
2.1.1.1.4. ON/OFF control
The VBAT and + 5V control system is implemented using the following circuits: U2 4011, U4 4528 and an G6AK 234P flip-flop relay. This system is powered, at all times, by the VBC voltage.
2.1.1.1.4.1 System:
3 inputs: TON ON key dry contact/GND TOFF OFF key dry contact/GND CDALIM active TTL signal with voltage cut-out
2 ouputs: VBAT Battery power/mains power. OFF TTL signal collector, + 5V PULL-UP open, OFF key pressed down, active at 0
Page : 8 NT 1023 Rev.B0
Page 13
2.1.1.1.4.2 Operation:
Press TON briefly to turn power on. Press TOFF continuously (5s<t<7s) to turn power off during a technical Failure. Press OFF 3 seconds to power OFF via CDALIM micro signal. The device can set ON or OFF via an external. Master module using the CD ON or CD OFF signal.
2.1.1.1.5. VBAT and + 5V power supply
VBAT voltage corresponds to mains voltage taken directly from the power unit/charger. Voltage is not controlled. It powers the display system and the motor, both of which are heavy duty energy consumers.
This voltage is available on TP1 and J2
Min Max
VBAT 6,5V 7V
The + 5V ± 5% is generated, using VBAT voltage, by the NS 2931 V3 controller low drop-out 0.6V for a 100 mA output current.
It is thus possible to make the best possible use of the battery. This voltage is available on TP2. The 5V rise time must be greater than 100 ms to allow for the RESET function on the CPU board.
2.1.1.2. Motor control module The control module of step by step motor is equipped with a gear reduction of 89.286. which makes the
double threaded screw 2 mm turn.
- One motor step is equivalent to 0.8233 µm of linear displacement of the driving bloc.
- One motor turn is equivalent to 22.4 µm of linear displacement of the driving bloc.
2.1.1.2.1. Motor control
Injectomat 2000 MC / Pilot Delta motor control is implemented by a stepper motor driver, dual pole con­trol module for a motor - UBB 5N model - (11.5 Ohm coil) built using an ST L293E IC7 motor control cir­cuit.
It features two functional modules: The motor control electronic parts, built around the L293E U13 circuit, optimises consumption and opti-
mal motor torque according to the pump flow rate.
2.1.1.2.1.1 Input signals
These signals are generated by the CPU board microprocessor and available on J02. They drive the U15 ULN2803 circuit.
2.1.1.2.1.2 Output signals
These signals are connected to the motor coils via J5 connector.
Phase A A Motor phase control J2.5 Phase B B Motor phase control J2.6 Phase C C Motor phase control J2.7 Phase D D Motor phase control J2.8 I Motor current reduction control J2.9 B00ST Booster activation and regulation current J2.10
NT 1023 Rev.B0 page : 9
Page 14
2.1.1.2.1.3 Booster module
BOOST = 0 The motor is powered via VBAT voltage. BOOST = 1 Booster function activated, the motor is powered at 12 Volts. L2 reactor, D18 diode, C15 capacitor and T8 transistor assembly allows a voltage of 12 V +/- 2 volts to
be obtained from VBAT. This voltage is available on TP5. The uncoupling frequency, 100 kHz, is generated by U9 oscillator, the booster is activated when the
boost line is at 1.
2.1.1.2.1.4 Soft-start module
The soft-start module, which is designed around the T9 transistor and C21 capacitor allows the pick up current of the uncoupling elevator to be limited when BOOST goes to 1.
2.1.1.2.1.5 Current regulation module
BOOST = 1 and I = 1 regulation motor current module is activated. The regulated current in the motor is 240 mA +/- 10% per motor phase. The signals amplitude and current image are available in TP6 and TP7. The current image of each coil is compared to a fixed level, which is implemented from the Divider bridge
using R31, R41/R37 and R42. The control is achieved through the U11 toggle, via a divider bridge by inhibition or confirmation of the
L293E control H bridge driving the CE1 and CE2 inputs. The calibrating frequency of 25 kHz is supplied by the U10 toggle.
2.1.1.2.1.6 Reduction current module
When the Boost signal sets at 0, the I line is used as current reducer.
I = 1 The H divider bridge is controlled by the A, B,C and D lines. I = 0 The H divider bridge is inhibited, the coils are not forwarded.
According to its rotation frequency (step/second), the motor is driven in one of the 3 control modes.
Mode Frequency motor (step/second) Control description Phase 1 from 0 to 32.3 Current reduction one ON supply Phase 2 from 32.3 to 90 No current reduction and regulation, one ON supply Phase 3 from 90 to 850 Current reduction, booster ON two supply
Page : 10 NT 1023 Rev.B0
Page 15
2.1.1.3. Analog output module The analog output module is built around a 10 bit, 5 channel analog/digital converter (MC 145053 U17)
with an SPI bus. The following SPI CLK, SI, SO, CSADC bus signals are available on connector J2: In addition the component generates an end of conversion signal (EOC). The CDANA signal, which is active when set to 1, controls the transistor T14 IRFD 9120 which digitally drives VREF. This voltage supplies the sensors and serves as a reference value for the ADC convertor. All the test points are concentrated on connector J9. Measurement of VREF on J9.7 VREF Pulsated signal of 5Volt ± 0.25V.
Convertor input:
ANO VBAT battery voltage measurement. AN1 not in use AN2 internal occlusion gauge bridge AN3 NU AN4 Absolute potentiometric position sensor
2.1.1.3.1. Battery voltage measurement
VBAT voltage is measured using a peak detection circuit comprising D19, R59, R60 and C23 in order to overcome the lower voltage created by the motor's pulsing demand for current. The voltage is available on J9.3.
For VBAT = 6.5V, V(J9.3)= 4V ± 10% motor running at 150 ml/h
2.1.1.3.2. Gauge bridge interfaces
The only Injectomat 2000 MC / Pilot Delta includes a force sensor fixed on the pusher.
2.1.1.3.2.1 Force sensor characteristics
Sensor technical Complete bridge with 4 gauges Impedance 350 Ohms ± 15% or 1 KOhms ± 15% Measurement range 0 to 150 N Surcharge 250 N Zero
< ± 10mV Sensibility 8.5 to 12 mV at 150 N zero derivation
2.1.1.3.2.2 Operation
85 µV /year
The force sensor generates a differential voltage proportional to the force sensor applied on the driving bloc. This force is amplified by a gain of 200 +/- 20% via an amplifier built around U18 TLC 251. The po­tentiometer P1 allows the offset to be compensated and for any other offset to be reset from the begin­ning. The sensor measurement chain transfer function can be defined by calibrating the sensor with two known forces.
AN3 J9.4 Pulsated amplitude signal sets at 0.6 V +/- 0.05 V for no force applied on the driving bloc
2.1.1.3.2.3 Force sensor connector
J8.1 VREF Gauge bridge power supply (+) J8.2 S(-) Out put Gauge bridge (-) J8.3 S(+) Out put Gauge bridge (+) J8.4 GND Gauge bridge power supply (-)
NT 1023 Rev.B0 page : 11
Page 16
2.1.1.3.3. Driving bloc position sensor
The absolute position of the driving bloc is obtained by a potentiometric sensor driven by a movement of the driving bloc.
The transfer function of the sensor can be characterised by calibration in two known positions. The sensor is powered by a pulsated voltage. The output is filtered by R62 and C22 linked directly to the
input AN4 (J2) Potentiometer connector:
J3.1 VREF J3.2 Center point J9.2 J3.3 GND
2.1.1.4. Opto switch module The opto switch module comprise 2 optical switches:
q A motor rotation detection switch q A syringe position head detector switch
2.1.1.4.1. Motor rotation opto
The opto switch is mounted on a disk which is pierced with a hole and assembled on the motor. It is used to check motor rotation, the opto diode is controlled in pulse mode to save energy. The optical switch is connected on J5.
Control T11 transistor Current limitation (R51) at 8mA Output T10 transistor TTL level J2.14 Control signal CDOPT1 activate at 1 J2.11 Output signal SOPT1 activate at 1 J5.7 Diode anode J5.8 Diode cathode J5.10 Transistor transmitter J5.9 Transistor collector TdON max 100 µsec TdOFF max 200 µsec
An anti-rebound device made of U20 flip-flop, reshapes the SOPT1 signal. The CDOPT1 and SOPT1 are emitted by the CPU board, and available on J2 connector
2.1.1.4.2. Anti-siphon opto switch
The opto switch is mounted on the plunger holder, it is used to check the presence or not of the syringe head.
 It is connected on J8 connector.
Control T12 Transistor Current limitation (R52) 8mA Output T13 Transistor TTL level Control signal CDOPT2 activate at 1 J2.15 Output signal SOPT2 activate at 1 J2.12 Anode diode J8.6 Cathode diode J8.5 Transistor transmitter common ground J8.10 Transistor collector J8.7 SOPT2 0V Anti - siphon present SOPT2 5V Anti - siphon missing
Page : 12 NT 1023 Rev.B0
Page 17
J8 is the ribbon cable linking the occlusion, disengagement and position sensors located on the syringe pump driver. The CDOPT2 and SOPT2 signals are emitted by the CPU board and available on J2 connector. The opto diode is powered by pulsated voltage in order to save energy.
2.1.1.5. Micro switch module
2.1.1.5.1. Disengagement micro - switch
The microswitch is mounted on the driving bloc ribbon cable. The centre point of the microswitch is con­nected to the ground.
The signals are available on J2 connector.
J8.8 DEB/ON NU J8.9 DEB/OFF OV engaged / 5V disengaged J8.10 GRD
2.1.1.5.2. Nurse call option
This is implemented by a monostable inverter relay RL2, whose two contacts and common point are available on J6 connector, the relay is driven by the BUZ signal which also drives the buzzer on the dis­play board.
J6.6 common point J6.7 contact normally open cut out power 24V/ 1A J6.8 contact normally shut
2.1.1.5.3. RS 232 option
The RS232 option interface RXD1 and TXD1 signals, in compliance with the V24 standard, signals come from the UART 2691 serial link external controller on the CPU board. It is implemented using a U19 LT 1180 CS circuit, associated with +/- 12V voltage generator, C31, C32, C33, C34 capacitors. This circuit is operational only if the pins 2 and 5 of J6 are short-circuited.
J6.1 output transmits data TX1 J6.2 +5V (DSR) J6.3 input receives data RX1 J6.4 GND J6.5 confirmation (DTR) J6.17 CTS J6.18 RTS
2.1.1.5.4. Configuration link and Master plug
The syringe pump Injectomat 2000 MC / Pilot Delta may be fitted to master module connected to the SUB 15 points plug located on the pump rear panel.
The module master communication link is done by the RX2 and TX2 signals.
J6.13 +VBAT Master power J6.14 RX2 Receive data J6.15 TX2 Transmit data J6.16 GND Master ground J6.9 CD-ON Syringe pump ON via master J6.10 CD-OFF Syringe pump OFF via master J6.12 I-SECT Master led main signal J6.11 I-OPTOM Motor rotation opto master control signal J6.19 BUZ Pilot buzzer command signal
NT 1023 Rev.B0 page : 13
Page 18
2.1.2. Description of connectors
2.1.2.1. J1 mains connector
Pin description
1 Neutral 2 Phase
2.1.2.2. J2 board / CPU connection
Pin Description
1 + 5V controlled power supply 2 GND power supply 3 + VBAT power supply 4 GND power supply 5 phase A motor control 6 phase B motor control 7 phase C motor control 8 phase D motor control 9 I signal motor control 10 BOOST signal booster command 11 sopt1 opto rotation module out put 12 sopt2 opto anti-siphon module out put 13 not in use 14 cdopt1 opto rotation control module 15 cdopt2 opto anti-siphon module control 16 OFF signal off key pressed ON/OFF 17 SECT mains power on signal power supply 18 CDALIM power cut signal 19 LDSECT mains LED control 20 CTS clear to send 21 DEB/OFF disengage signal active, set to 0 22 RTS request to send 23 OCC/OFF occlusion signal active, set to 0 24 BUZ nurse call relay control 25 EOC end of conversion ADC 26 CSADC selection bus SPI ADC 27 CLK clock bus SPI ADC 28 SI data IN bus SPI ADC 29 SO data out bus SPI ADC 30 CDANA analog sensor power control 31 RX2 receive data TTL line 2 32 TX2 transmit data TTL line 2 33 TXD1 transmit data TTL line 1 34 RXD1 receive data TTL line 1 35 TOFF OFF key 36 TON ON key 37 + VBAT power supply 38 GND power supply 39 + 5V power supply 40 GND power supply
Page : 14 NT 1023 Rev.B0
Page 19
2.1.2.3. J3 potentiometric sensor connector
Pin Description
1 VREF 2 centre point 3 GND
2.1.2.4. J4 internal battery connector
Pin Description
1 2
2.1.2.5. J5 motor connector
Pin Description
1 + VBAT 2 + VBAT 3 PHASE D 4 PHASE C 5 PHASE B 6 PHASE A 7 opto rotation anode diode/ + 5V 8 opto rotation cathode diode 9 opto rotation collector transistor 10 opto rotation transmitter transistor / GND
battery + battery -
2.1.2.6. J6 rear panel connector The connector on the rear panel concentrates signals from the external gauge bridge, the optional RS232
series link, the nurse call relay output and the configuration series link.
Pin Description
1 TX1 transmit data line 1 2 + 5V power supply 3 RX1 receive data line 1 4 GND power supply 5 DTR interface confirm 6 APP-INF COM nurse call relay common point 7 APP-INF NO nurse call relay normally open 8 APP-INF NF nurse call relay normally closed 9 CD ON external ON 10 CD OFF external OFF 11 I-OPTON motor control out put 12 I-SECT main led 13 + V BAT external power plug 14 RX 2 receive data line 2 15 T X 2 receive data line 2 16 GND power supply 17 CTS clear to send 18 RTS Request to send 19 BUZ buzzer external control
2.1.2.7. J7 external DC power supply connector
Pin Description
1 External power +/­2 External power -/ +
2.1.2.8. J8 pump ribbon cable connector The pump ribbon cable connector concentrates all the signals from the sensors located in the plunger:
disengage microswitch, gauge bridge and anti-siphon opto switch.
NT 1023 Rev.B0 page : 15
Page 20
Pin Description
1 + VREF + internal gauge bridge 2 E1 internal gauge bridge/occlusion input on 3 E2 internal gauge bridge/occlusion input off 4 GND internal gauge bridge 5 C DOPT2 anti-siphon cathode diode 6 + 5V opto anti-siphon anode diode / + 5V 7 S OPT 2 opto anti-siphon collector transistor 8 DEB / ON disengage microswitch on 9 DEB / OFF disengage microswitch off 10 GND
Important: Disassemble the ribbon cable holder on the supply board before extracting the me-
chanical assembly from the lower housing.
2.1.2.9. J9 Test Points
Pin Description
1 GND 2 out put position sensor 3 out put low battery control 4 out put force sensor amplifier 5 N.U. 6 out put optical switch motor control 7 force and position sensor voltage Ref. 8 out put optical switch syringe led detection
2.1.3. Electrical layout
(Refer to Annex 2)
2.1.4. Installation layout
(Refer to Annex 2)
2.2. CPU BOARD
Overview
The CPU board is fitted to Injectomat 2000 MC / Pilot Delta version, around a 80C32 microproces-
sor used in open mode. It concentrates all the peripheral devices directly connected to the 80C32 bus. It is connected to the power supply board by a 40 contacts ribbon cable and to the display board by fixed connectors. It forms a single unit, with the display board, which is fixed to the front panel.
The CPU board uses CMOS technology in order to minimise power consumption. Current used: 5 Volts 80 mA maximum.
2.2.1. Functional description
The CPU board comprises six functional units:
q Ram rom decoding processor q Reset WATCH DOG q Parallel port extensions
display/keyboard interface, motor interface, sensor interface.
Page : 16 NT 1023 Rev.B0
Page 21
q SPI BUS q Asynchronous serial link q Optical sensor
2.2.1.1. RAM ROM decoding processor Decoding is carried out by an IC3 80C32 circuit, running at 12 MHz, clocked by Q1. It is used in open
mode, with the EA*/VP line connected to GND. Address/data de-multiplexing is carried out by a 74HC573 U3. On this BUS are implemented:
32 Ko static RAM U6 27C040 512 Ko U4
2.2.1.2. Reset watch-dog The RESET WATCH-DOG module comprises two TL7705 U10 and U11 circuits. Operation: the U10 circuit generates RESET signals, active at 1, for the processor and the UART; RST*
active at 0 generates RESET signals for the other peripheral devices. The signals are activate in two cases:
q when the system is powered up, q as soon as the WATCH-DOG circuit is triggered. It remains active until the power is turned off.
2.2.1.2.1. Reset at power-up
The TL 7705 circuit guarantees the minimum duration of the reset lines, in the active state, once the + 5V voltage has exceeded the circuit operating threshold (4.75V). It returns to the active state, if the + 5V volt­age drops below the threshold or if the RESTIN* (U11.2) is at 0. The line is driven by the WATCH-DOG module.
The duration of the reset, in the active state, at power-up is set by the C10 capacitor 220nF 100 ms.
2.2.1.2.2. Watch-dog
The WATCHDOG circuit comprises the following elements: U10, U11, C12, D2, D1, R4, C11. When powered up the capacitor C12 is charged by the U10 circuit, via diode D1.
The charge is maintained at a threshold of over 1.5 V during operation. The software writes, every 1 ms. This writing generates a 5V/1µs impulse on the U7.10 output, which re-
charges the capacitor C12 via a high pass peak detector circuit made up C8, D3, D2. The C12 capacitor discharges in resistor R4.
When the software stops, the capacitor C12 completely discharges. The U11 RESTIN* line falls to 0 and the RESET signals are activated, stopping all syringe-pump control operations in the inactive state. Fault signals, the blinking FAIL diode and the continuous BUZZER are stuck in an active state.
WATCH-DOG trigger time is less than 400 ms.
2.2.1.3. Keyboard/display interfacing
2.2.1.3.1. Display registers
The display system is made of LED’s and of a 2 lines of 20 characters LCD graphic screen. The U7 cir­cuit allows to address the matrix of the LED’s, the U8 circuit allows the writing and reading in the display controller.
The LED’s matrix are DIG0 to DIG7 and SEG0 to SEG7. The FAIL LED shows the device is failed, the command is inverted to be active by default at RESET. It is
active at 1 on the display board. The FAIL diode is out of the matrix to be able to light on when the micro­possor does not work.
NT 1023 Rev.B0 page : 17
Page 22
2.2.1.3.2. Buzzer
The BUZZER command is inverted and controls the transistor, T3, which is mounted as a common trans­mitter. Working in parallel, the transistor collector drives the BUZZER on the display board and the nurse call relay on the motor control power supply board. After starting the pump, the BUZZER is activated for added safety.
BUZZ signal: J5 pin 6 and J3 pin 24 50 mA 6.75 Volts maximum.
2.2.1.3.3. Keyboard register
The keyboard is based on a 6 x 3 matrix, with 2 separate keys - TON and TOFF - with 1 common point (GND). They turn power ON and OFF respectively and are connected to the display board. TON and TOFF signals only transits via the CPU board.
The columns of the keyboard are driven by the same signals as the columns in the display matrix, thus facilitating simultaneous keyboard and display monitoring. Register U15 reads the status of the three key­board lines, LIG1, LIG2 and LIG3 in order to check whether a key has been activated.
2.2.1.3.4. Motor control register
The U12 motor control register generates 4 motor phase signals, A,B,C and D, the I current control sig­nal, the motor rotation opto control CDOPT1 signal and the BOOST BOOSTER control signal.
2.2.1.3.5. Sensor status register
The U16 sensor status register reads the microswitch digital sensors and the syringe pump opto switch.
2.2.1.4. SPI bus EEPROM The SPI bus is synchronous series communication bus using various peripheral circuits. The SPI bus is
driven by the 80C32 ports. The Injectomat 2000 MC / Pilot Delta syringe pump has 2 peripheral devices on the SPI bus:
q The EEPROM 16 Ko 24C128 U12 located on the CPU board. q The MC 145053 analog/digital converter located on the motor board.
This bus has 2 communication lines: 80C32 (see electrical diagrams)
Micoprocessor Ports
CLK clock generated by the microprocessor P 1.1 SI peripheral to processor data (input) P 1.2
2.2.1.5. Asynchronous serial links The Injectomat 2000 MC / Pilot Delta syringe pump is fitted with two asynchronous serial links.
RS232 option line 1 TTL configuration serial link line 2
2.2.1.5.1. RS232 serial link
The RS232 serial link is implemented using the U8 SCC2691 circuit, which controls asynchronous com­munication, and RS232 interface circuit which is located on the motor power supply board.
The circuit is clocked by quartz Q2 at 3.6864 MHz. It includes a programmable baud rate generator. It generates interrupts on the ITRS232 line connected to the processor at INTO.
The SCC2691 is on the microprocessor bus. It drives the RXD1 receive data signals on J3 (pin 34) and TXD1 transmit data signals on J3 (pin 35).
The RS232 option is reserved for dialogue with the host computer responsible for monitoring or control­ling the system.
Page : 18 NT 1023 Rev.B0
Page 23
2.2.1.5.2. TTL serial link
The TTL serial link is driven by the serial link controller inside the microprocessor. It uses one of the in­ternal timers to generate its baud rate, from the basis of the processor 12 MHz clock. The serial link drives the TXD2 transmit data and RXD2 receive data lines. The input and output lines are buffered by the U9 74HC14 buffer trigger inverter circuit.
The lines are available on J3.
TX2 J3 pin 33 RX2 J3 pin 34
This serial communication line is reserved for Injectomat 2000 MC / Pilot Delta software configuration and, when appropriate, for connecting an external MASTER module.
2.2.1.6. Opto switch sensors Injectomat 2000 MC / Pilot Delta syringe pump opto switch sensors are for piston pressure and syringe
body presence and motor rotation. The syringe body opto switch interface is located on the CPU board. It is implemented using transistors
T1, T2 and T3. Resistor R9 limits current in the diode to 8 mA.
2.2.1.6.1. Opto switch body syringe measurement.
q Note: The CI opto and the obturator are specific to the Injectomat 2000 MC / Pilot Delta equipped
with "flange detector" and are not compatible with previous versions.
q 2 opto devices allow to detect 2 syringe sizes: 60 cc and 20 cc.
J2.1 opto diode +5V anode J2.2 opto cathode common point and transistor transmitter J2.3 opto 1 transistor collector J2.4 opto 2 transistor collector
Control signal: CODPT3 activate at 1 driven by microprocessor line T0. Output signal: SOPT4 activate at 1 register U16 D1 address $ C000 Output signal: SOPT3 activate at 1 register U16 D0 address $ C000
.
SOPT3 SOPT4 Syringe clamp detection alarm, high position 1 0 60 cc detection 1 1 20 cc detection 0 1 Syringe clamp detection alarm, low position 0 0
The motor rotation and piston presence opto switch interfaces are located on the POWER SUPPLY BOARD.
2.2.1.6.2. Motor rotation opto switch
Control signal: CDOPT1 activate at 1 Output signal: SOPT1 activate at 1
2.2.1.6.3. Piston presence opto switch (anti-siphon)
Control signal: CDOPT2 activate at 1 Output signal: SOPT2 activate at 1
2.2.2. Description of connectors
2.2.2.1. J1 not used This connector is not used for the moment but it will be used for a future evolution of the software.
NT 1023 Rev.B0 page : 19
Page 24
2.2.2.2. J2 Opto switch - syringe body connector
Pin Description
1 ground 2 flange switch 3 opto +5 V diode anode 4 cathode diode transmitter transistor opto 1 and opto 2 common points 5 collector transistor opto 1 6 collector transistor opto 2
2.2.2.3. J3 Power supply/CPU ribbon cable  A 40 channel ribbon cable is soldered directly to J3, linking the power supply and the CPU.
Pin Description
1 + 5V controlled power supply 2 GND " 3 + VBAT " 4 GND " 5 A phase motor control 6 B phase " 7 C phase " 8 D phase " 9 I signal " 10 BOOST signal " 11 SOPT1 opto rotation module output 12 SOPT2 opto anti-siphon module output 13 APINF nurse call independent from buzzer signal 14 CDOPT1 opto rotation module control 15 CDOPT2 opto anti-siphon module control 16 OFF ON/OFF key depressed signal 17 SECT mains power supply on signal 18 CDALIM power cut signal 19 LDSECT mains LED control 20 CTS Clear to send line 2 21 DEB/OFF disengage active at 0 signal 22 RTS Request to send line 2 23 OCC/OFF occlusion active at 0 signal 24 BUZ nurse call relay control 25 EOC end of conversion ADC 26 CSADC selection SPI ADC bus 27 CLK clock SPI ADC bus 28 SI data INSPI ADC bus 29 SO data out SPI ADC bus 30 CDANA analogue sensor power supply control 31 RX2 receive TTL data line 2 32 TX2 transmit TTL data line 2 33 TXD1 transmit TTL data line 1 34 RXD1 receive TTL data line 1 35 TON ON key 36 TOFF OFF key 37 + VBAT power supply 38 GND 39 + 5V 40 GND
2.2.2.4. J4 Display board interconnection
Pin Description
1 SEG1 display matrix line 1 2 SEG2 display matrix line 2 3 SEG3 display matrix line 3 4 SEG4 display matrix line 4
Page : 20 NT 1023 Rev.B0
Page 25
5 SEG5 display matrix line 5 6 SEG6 display matrix line 6 7 SEG7 display matrix line 7 8 SEG8 display matrix line 8 9 COL1 display matrix column 1 10 COL2 display matrix column 2 11 COL3 display matrix column 3 12 diode FAIL control 13 RDCRT current reduction control 14 LIG1 keyboard interface line 1 15 LIG2 keyboard interface line 2 16 LIG3 keyboard interface line 3 17 LDSECT mains LED lighting control 18 + 5V power supply 19 VBAT power supply 20 GND power supply
2.2.2.5. J5: Display/CPU connection
Pin Description
1 TON ON key 2 TOFF OFF key 3 SI SPI bus 4 CLK SPI bus 5 CSLCD bus 7 VBAT power supply 8 GND power supply
2.2.3. Electrical layout
(Refer to Annex 2)
2.2.4. Installation layout
(Refer to Annex 2)
2.3. DISPLAY BOARD
2.3.1. Overview
The display board is mounted directly beneath the front plate of the syringe pump. It brings together all the facilities for operator/device dialogue: Keyboard, buzzer and display.
It is connected to the CPU by rigid connectors, forming a sandwich, with the former, held in place by struts.
The soft keyboard is connected to the display board.
2.3.2. Functional description
The display board comprises four modules:
q The electroluminescent display, q The keyboard interface, q The liquid crystal display, q The buzzer.
2.3.2.1. Electroluminescent display The electroluminescent display is made of eighteen LED's and five 7 segment display units, with the dec-
imal point, except two of them which are marked (*) in the table below.
NT 1023 Rev.B0 page : 21
Page 26
The diodes and display units are driven in a multiplexed, 8 segments x 8 digit matrix. The LED's and dis­play units are mounted with a common cathode.
The 8 segments are driven by signals SEG0 to SEG7 and the 8 digit by signals DIG0 to DIG7. The 2 LED’s, "MAIN PRESENCE" and "FAIL" are controlled independently of the matrix.
2.3.2.1.1. LED’s table
The following table lists the various diodes used in different models.
Ref Name Type Seg Dig
LD1 Mains On yellow * * LD2 Body Alarm red 1 0 LD3 Piston Alarm red 2 0 LD4 ml green 5 1 LD5 Battery green 7 1 LD6 Fail red * * LD7 Alarm red 4 to7 0 LD8 Worm screw 3 green 2 1 LD9 Prealarm orange 1 to 4 2 LD10 Occlusion Alarm red 0 0 LD11 Worm screw 2 green 3 1 LD12 Infusion End Alarm orange 0 2 LD13 Worm screw 1 green 4 1 LD14 NU LD15 Connection to PC green 0 1 LD16 disengagement alarm red 3 0 LD17 ml/h green 5 1 LD18 Battery Alarm red 5 2 LD19 Validation Demand green 1 1
2.3.2.1.2. Seven segment display units
Ref Name Type Seg Dig
U3 hundreds green 0 to 7 6 U4 tens green 0 to 7 5 U5 units green 0 to 7 4 U6 tenths orange 0 to 7 3 U2 thousands green 0 to 7 7
2.3.2.2. Keyboard interface
The keyboard is an 18 key matrix keyboard. The keys are arranged in 3 rows of 6, with two separate keys with a common point (GND), TON and TOFF, and power on and off switches, connected to J2.
Ref Digit Ligne
SILENT 2 2 STOP 1 2 VALIDATION 0 2 PURGE 3 2 BOLUS 2 0 DECAL 4 0 SELINC 3 0 SELDEC 2 1 ENTER 3 1 HISTO 4 1
Description of the connector
J2. 1 DIG 5 J2. 2 DIG 4
Page : 22 NT 1023 Rev.B0
Page 27
J2. 3 DIG 3 J2. 4 DIG 2 J2. 5 DIG 1 J2. 6 DIG 0 J2. 7 LINE 0 J2. 8 LINE 1 J2. 9 LINE 2 J2. 10 TON J2. 11 TOFF J2. 12 GND
2.3.2.3. LCD display unit The LCD display unit is "chip on glass" type, the controller is fixed on the glass. It has 2 lines of 20 char-
acters. The BUS gestion is multiplexed with the command of SEG0 and SEG7. It has a double powered retrolighting. The power supply of the LED pair, one by one, allows an optimum light for a minimum consumption.
2.3.2.4. The buzzer The buzzer is an auto-exit buzzer supplied by VBAT. It is driven by the BUZZ signal, available on connector J3, pin 6, which is generated by the CPU board. It
is mounted in parallel with the optional nurse call circuit, located on the motor power supply board.
2.3.2.5. Time keeper Real time device DS1307
2.3.3. Description of connectors
2.3.3.1. J2 Display board / CPU board connection
Pin Description
1 SEG0 display matrix and command LCD display line 1 2 SEG1 display matrix and command LCD display line 2 3 SEG2 display matrix and command LCD display line 3 4 SEG3 display matrix and command LCD display line 4 5 SEG4 display matrix and command LCD display line 5 6 SEG5 display matrix and command LCD display line 6 7 SEG6 display matrix and command LCD display line 7 8 SEG7 display matrix and command LCD display line 8 9 COL1 display matrix and keyboard column 1 10 COL2 display matrix and keyboard column 2 11 COL3 display matrix and keyboard column 3 12 FAIL diode FAIL control Fail 13 RDCRT display control writing command 14 LIG1 keyboard interface line 1 15 LIG2 keyboard interface line 2 16 LIG3 keyboard interface line 3 17 LDSECT LED mains lighting control LED sector 18 + 5V power supply 19 VBAT power supply 20 GND power supply
2.3.3.2. J3 CPU board connection
Pin Description
1 TON ON key 2 TOFF OFF key 3 SI bus SPI
NT 1023 Rev.B0 page : 23
Page 28
4 CLK bus SPI
OPT
OK
OK
OK
OK
OK
OK
OK
OK
OK
5 CSLCD bus SPI 6 BUZZ BUZZER control 7 VBAT power supply 8 GND power supply
2.3.4. Power consumption
Main supply (without battery) Battery supply (without main supply)
measurement made on the battery
0 ml/h 1200 ml/h 0 ml/h 1200 ml/h
31 mA / 230v ~ 47 mA / 230v ~
< 125 mA / 6,0 VDC 700 mA / 6,0 VDC
2.3.5. Electrical layout
(Refer to Annex 2)
2.3.6. Implantation layout
(Refer to Annex 2)
3 Configurations
Fresenius Vial recommends the presence of its qualified personnel or of a trained member of the Technical Department of your es­tablishment to help you implement the configuration procedures you wish to choose.
Note: press STOP to cancel modification at any time - Press to leave configuration mode at any time.
3.1. Configuration of the pressure functionalities
Moving to the pressure configuration mode: press +
Press Configuration mode Valid. Choice
Pres 1 : variable threshold mode.
Pres 2 :pressure limit by syringe capacity.
: Pres appears on display.
Pres 1 900 mmHg
50/60 ml syringes
Pres 2 50 cc 800 mmHg
20 ml syringes
Pres 2 20 cc 600 mmHg
Note: at any time, it is possible to press START key to store programmed values and resume Pres 2.
Press ( ) to select
preset pressure limit at a value sets between 100 and 900 mmHg at power on
or use the last programmed value (___)
from 100 to 900 mmHg
from 100 to 900 mmHg
Valid.
Pres 3 : pressure drop warning level.
Pres 3 100 mmHg
from 50 to 1100 mmHg (disable this warning 0 mmHg)
3.2. Configurations of the various parameters
Moving to configuration of the various parameters mode, press simultaneously + : Par appears on display.
Press Configuration mode Valid. Display
Par 1 : preset parameters at power on.
Par1
(parameters: flow rate, rate mode unit, drug name, drug concentration, patient weight)
Press ( ) to select
use last parameters preset parameters:
flow rate: 00.0 ml/h
Page : 24 NT 1023 Rev.B0
Valid.
Page 29
rate mode proposed :ml/h
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
OK
patient weight, V/T values : preset at last values before selection
Par 2 : syringe selection type.
Par 3: maximum flow rate selectable via the
keyboard.
Par2 SEL4
50 ml syringe
Par3 50 cc 800 ml/h
20 ml syringe
Par3 20 cc 550 ml/h
Par 4 : selectable syringes/syringe list restric-
tion.
Par 5 : prime compulsory.
Par 6 : rapid infusion start up.
BD Plastipak 50 cc
Par 5
Par 6
Press Configuration mode Valid. Display
Par 7 : KVO rate.
KVO = 1.0 ml/h KVO = 2.0 ml/h
SEL3 only one syringe per capacity with
automatic confirm
SEL4 multisyringe mode
from 0.1 to 1200 ml/h
from 0.1 to 600 ml/h
selectable syringe on display not selectable on display
Selection for any syringe
priming compulsory no priming compulsory
running rapid infusion start up no rapid infusion start up
Press ( ) to select
From 0.1 to 5 ml/h
Valid.
Par 9 : RS 232 baud rate.
Par A : empty syringe mode (finish on occlu-
sion).
Par B : frequency of maintenance and date of
next maintenance
Par C : drug library (see page 9)
Par D : syringe flanges detection.
Par E : pre-programmed bolus dose rate and
volume preset value at power on
Par 9
4800, 9600, 19200 Bauds
19200
Par A
activated disabled
Par B 200 h 12/05/2000
from 1 hour to 9 999 hours of continuous use (preset at 3 000 hours)
Note : select 0000h to desable
Par B 200 h 12/05/2000
Select date and shift to next item by pressing OK key.
Note : select 00/00/0000 to desable date of next maintenance mode
Note: at any time, it is possible to press START key to store programmed parameters and resume Par B.
ADRENALIN
Drug library configuration
200ng/ml
Par D
activated disabled
Par E
use the last rate and volume settings used a preset fixed value for rate
volume = 0.1 ml
Preset
fixed
value
50 ml syringe
Par E 50cc 800.0 ml/h
20 ml syringe
From 50.0 to 1200.0 ml/h
From 50.0 to 600.0 ml/h
Par E 20cc 600.0 ml/h
Par F : bolus rate value at power on
Par F
use the last set rate setting
Page : 28 NT 1023 Rev.B0
Page 30
used a preset fixed value
OK
OK
OK
OK
OK
OK
OK
Preset
fixed
value
Par G : rate mode selection
Par J : mains disconnection warning inhibition.
Par P : name of the service
Important: measurements of the friction forces on the plunger stoppers do not show stable characteristic. For this reason, we recommend not to use Star mode (Par6).
50 ml syringe
Par F 50cc 800.0 ml/h
20 ml syringe
Par F 20cc 600.0 ml/h
Ml/h
Par J
Par P ICU 15210 ____
From 50.0 to 1200.0 ml/h
From 50.0 to 600.0 ml/h
Activate or disable rate mode with . Shift to the next item by pressing OK OK
key
activated inhibited
inhibited mains disconnection warning active
Select the service name
3.3. CHECKING THE PILOT
 N.B.: the tests outlined below do not include the occlusion tests, the flow rate tests, the electrical safety tests,
etc.
3.3.1. The After Sale Service test
The ASS test is activated by pressing simultaneously on the keys:
and
until following display:
TESt.1
The validation LED flashes.If The device display, for example:
START
is not pressed, within 3 seconds, the device returns to normal running.
Test1Test1
0h0h
The keys , and
START
allows to scroll in the following list:
 "TEst.1TEst.1" = displays running time with zero reset if necessary and service date modification  "TEst.2TEst.2" = tests all indicator lights (LED's, 7-segment display unit AND LCD screen)  "TEst.3TEst.3" = tests keyboard.  "TEst.4TEst.4" = displays battery voltage
NT 1023 Rev.B0 page : 29
Page 31
"TEst.5TEst.5" = displays code of last 10 alarms.
STOP
STOP
 "TEst.6TEst.6" = displays total running time.  "TEst.7TEst.7" = TTL serial link test.  "TEst.8TEst.8" = RS232 serial link test.  "TEst.9TEst.9" = displays force on plunger.  "TEst.ATEst.A" = software version, check sum, loading date and language.  "TEst.BTEst.B" = displays ADC analog inputs.  "TEst.CTEst.C" = displays driving block position.  "TEst.ETEst.E" = displays calibration values.  "TEst.FTEst.F" = displays calibration syringe type.  "TEstTEst.JJ" = displays of the last 10 events before the last blocking error.  "TEst.LTEst.L" = drugs library.
3.3.2. Running time tESt.1tESt.1
This test allows to display, first, the running time in hours, days and months.
Press on
START
gives the maintenance date. This date may actualised with the key board.
3.3.3. Lights test tESt.2tESt.2
This test allows to check the lighting of LED’s of the front panel, of the 7 segments display and of the LCD screen. First, all the items light at the same time and then light one after another.
This test can be stopped at any time by pressing
3.3.4. Key board test tESt.3tESt.3
This test allows to test the correct functioning of every key. The message “Test 3” is permanently displayed. When pressing on a key, its name displayed on the screen
OFF Switch off
SIAL Alarm silent / Information keys STOP Stop infusion
START
START Validation (note: pressing longer than 2 seconds on this key, drives back
to tests choice)
BOLUS
BOLU Bolus manual control
DECAL Moving the traveller
INC Increase
DEC Decrease
OK
OK Enter, validation
OPT
OPT Options menu
Page : 30 NT 1023 Rev.B0
Page 32
1 2 3
6 : DEC +
1 : INC +++ 2 : INC ++
4 5 6
3 : INC + 4 : DEC +++ 5 : DEC ++
Nota: ON, can’t be tested If several keys are pressed simultaneously, the device displays “Err” and gives an alternative "BIP".
The key works normally: the message “OFF”is displayed as soon as pressed, the device is switched off if pressed more than one second.
3.3.5. Battery voltage display tESt.4tESt.4
This test displays the battery voltage over 5 digits. The voltage is calculated in accordance with corresponding ana­log input value and calibrating values. The values used are those for escape from an alarm or pre-alarm (6.3 V) or starting an alarm (5.7 V).
The display is in volts and tenths of volts. The display is continuously updated according to the voltage changes. The battery and mains LED's are also updated. The battery LED begins to flash if the voltage displayed is below the
calibrated pre-alarm threshold and stops flashing if the voltage is above the pre-alarm output threshold. Press the CONFIRM key to select another test.
3.3.6. Last 10 alarms codes tESt.5tESt.5
This test displays the codes of the last 10 events on the display units. Three types of events are memorised :
q ALARM q ERROR q SWITCH OFF: two cases
Normal SWITCH OFF by pressing the OFF key. Abnormal SWITCH OFF due to misfonctioning.
When an alarm goes off, anAA” is displayed followed by a number identifying the alarm:
Alarm Description
10 battery 11 syringe clamp 12 end of infusion 13 volume limit 14 disengagement 15 plunger head 16 occlusion 25 Flange 26 Main line disconnection
NT 1023 Rev.B0 page : 31
Page 33
In case of error, an “E” is displayed followed by error number:
Error Description
01 rotation control 03 communication
32 / 52 / 72 / 82 segment advance check
44 CPU / UART frequency control 50 ADC access self-test 52 advance check during take-up
14 / 60 checking coherence of syringe parameters (incoherence of the syringe diameter in relation
to the number of motor steps for 0.1 ml calculated when the syringe is confirmed)
70 incorrect motor frequency (motor step period too big or too small, calculated from the sy-
ringe diameter and the selected flow rate) 72 advance check over the whole length 80 important electromagnetic interference’s or bad key board.
16 / 55 Time keeper
18 Max infused (1,7l) reached 28 Language file error
24 / 34 / 44 / 56 Software error
66 LCD error
The error codes: 10 (internal Ram self-test(+)
20 (external Ram self-test(+) 30 (EEPROM check-sum self-test) and 40 (EEPROM access)
These errors cannot be stored in the EEPROM; the running of the device is too pertubated to allow it to write in the EEPROM. When normally stop, the “OFF” message is displayed. For abnormal stop “OFF” + flashing "F" (Fail) are displayed on the LCD screen. The events are numbered from 0 to 9. 0 is the last event, 9 is the eldest one.
The keys and allows to scroll the events one way or the other Example:
Test 5 8Test 5 8
E 01E 01
Means: event N° 8 was a type 01 error (rotation control)
3.3.7. Total running time tESt.6tESt.6
Use this test to display the total running time of the Pilot. Unlike “Test 1" which resets the time at zero when the service date is modified, it is not possible to modify this time manually.
Example:
Test 6Test 6
28 H28 H
3.3.8. TTL Serial link test: tESt.7tESt.7
This test allows to verify the TTL serial link TTL (80C32), by placing a plug on which the Rx and Tx lines are " short­circuited "(pin 2 and 3).
If link is correct: LTOKLTOK is displayed If link is not correct or if plug is not connected: LTER LTER is displayed
3.3.9. Serial link test: tESt.8tESt.8
Page : 32 NT 1023 Rev.B0
Page 34
This test allows to verify the RS 232 serial link (2691), by placing a plug on which the Rx and Tx, RTS and CTS, DSR and +5 V lines are "short-circuited" (pin 2 and 3), (7 and 8),(4 and 6)
LROKLROK = RS 232 correct link. LRERLRER = break between Tx and Rx NORCNORC = - break between RTS and CTS,
- break between DSR and + 5 V
- no plug
3.3.10. Strength on the plunger display: tESt.9tESt.9
This test displays the strength applied on the plunger. The strength is calculated according to the value of the corre­sponding analogue input and the calibrating values. The display is in grams. The value is continually updated ac­cording to changes in the strength value.
3.3.11. Software version tESt.AtESt.A
This test allows to display the software number version, the check sum and the loading date.
OK
Press on
allow to display the language.
Example:
rst
1
screen
Test A V01.1a
OCFO 01/12/2000
nd
2
screen
Test A Français
V01.0 09/09/1998
Press on for other language screen
3.3.12. Analog input display tESt.BtESt.B
This test allows to read the hexadecimal value of the 5 analogic inputs and of the 3 converter test inputs. The chan­nel number is displayed above this value.
Example
Test B M
200
The keys and allows to pass from one channel to an other. The analogue inputs are divided as follows:
0 battery voltage 1 N.U. 2 force sensor 3 NU 4 potentiometer displacement
NT 1023 Rev.B0 page : 33
Page 35
L converter zero test, between 0000 and 0004 if correct M converter mid-scale test, between 01FB and 204 if correct H converter full-scale test, between 03B and 3FF if correct
3.3.13. Driving block position display tESt.CtESt.C
This test displays the position of the driving block. The position is calculated in relation to the value of the corre­sponding analogue input and calibrating values. The display is in mm and tens of mm. The value is continually up­dated as the driving block moves. The value displayed is ± 0,1 mm.
3.3.14. Calibration values display tESt.EtESt.E
This test allows to show the calibration values stored in the EEPROM. Every value is displayed on 3 digits. The first line gives the shorted name of the value.
Example:
Test E LOWTest E LOW
082082
The keys and allows to pass from one channel to an other.
bat1 alarm and pre-alarm battery voltage: 6.3 V bat2 pre-alarm battery voltage: 5.9 V bat3 alarm battery voltage: 5.7 V HIGH displacement potentiometer with large 115.0 mm spacer LOW displacement potentiometer with small 20.0 mm spacer 0G force meter with 0 kg 5Kg force meter with 5 kg
3.3.15. Syringe type display tESt.FtESt.F
This test displays the type of syringe fitted to the Pilot. The type is defined with the indications given by the optos of the syringe clamp.
The type of syringe is displayed with its capacity: 20cc (20/25cc), 50cc (50/60cc). The capacity may not be dis­played if the parameters in EEPROM indicate that certain types of syringes are not included. When the syringe clamp is in the higher and the lower position 4 dashes are displayed.
The display is constantly updated in relation to changes in the syringe clamp system.
3.3.16. Displays of the last 10 events before the last blocking error tESt.JtESt.J
This test allows to display the 10 last events before the last blocking error.
3.3.17. Drug library tESt.LtESt.L
This test allows to read the name, the author and the date of the protocol library registered in the device.
4 REPLACING SUB-ASSEMBLIES
q Important: Disassemble the flexible circuit holder on the supply board before extracting the mechanical as-
sembly from the lower box.
q Important: Device operation must be completely checked after all intervention inside the device.
Page : 34 NT 1023 Rev.B0
Page 36
4.1. Mounting the flexible circuit
diam.12 tube
q Important: Handle the flat cable with great care when mounting this sub-assembly: damage to the flat cable
will result in complete disassembly of the mechanical block.
Fitting the flat cable
1. Take the flat cable and, using the "flat cable insertion" tool, wind the flat cable on the opposite side to the 10­point connector in the tool slot, keeping the flat cable tightened, wind it onto the equipment.
hood
2. Take the black hood of the flat cable insertion equipment and position it on the wound flat cable part.
3. Insert the driving block cover and the input bearing on the diam. 12 tube, in the proper direction for mounting the various components.
Important: The input bearing flange must be placed on the side external to the driving block cover.
q Visually check that the "input bearing + driving block cover" are properly mounted on the tube.
4. Insert the "flat cable insertion" tool in the diam. 12 tube on the input bearing side.
q The perpendicular flat cable parts are placed in the slits found at the ends of the tube.
5. Remove the black hood and the flat cable insertion equipment.
6. Correctly position the flat cable in the two slits visible at each end of the tube.
Important: The flat cable must not be twisted inside the tube.
7. Place the flexible circuit guards at both ends of the tube, passing the flat cable between the two holding lugs
8. Place the diam. 12 tube on the driving block making match the indexing finger of the driving block with the hole of the tube (opposite side of the 10 point connector)
flat cable
9. Match up the flexible circuit hole with the driving block holder centring tube.
10. Secure the clamping collar onto the driving block using two 2.5 x10 screws and two washers.
4.2. Wiring the components on the flexible circuit
q Important: it is essential that the component mounting direction be respected.
1. Position the strut (1.5mm) between the 4 pin photo switch and weld the syringe head detection photo switch flattening it against the flat cable.
Important: use a silicone between the optical switch and the flat ribbon cable. The silicon should not be over the soldering area.
NT 1023 Rev.B0 page : 35
Page 37
Important: verify there is no resistor continuity between the optical switch axis stop and the flat ribbon cable.
Photo switch
2. Weld the disengaging switch flattening it against the flat cable.(at roughly 1.5 mm).
3. Cut the part of the flat cable corresponding to the back pressure microswitch of version A2 (see cutting zone in the diagram below).
Disengaging switch
Diam 12 tube
Syringe head detection
Photo switch anode long lug
Weld the disengaging microswitch without flattening it against the flat cable (at roughly 1.5 mm).
4. Weld the force sensor wires
Important: when disassembling a force sensor, take care not to damage the welding pellets.
green wire
lower marking
5. Pass the flat cable in the force sensor oblong.
Important: When disassembling the flexible circuit from the driving block holder, take care not to detach or dam­age the holder protection square. The purpose of this square is to hold the flexible circuit correctly in place and avoid short-circuits with the holder.
6. Mount the force sensor on the driving block holder using a TF HC M4x10 screw with weak lactate.
Important: avoid all contact between the force sensor and its holder.
7. Mount the contact plate on the force sensor using a TF M4x10 screw with weak loctite.
Important: Before calibrating the force sensor, adjust the threshold voltage (0.6V + 0.05V) using an oscilloscope between pin 1 (earth) of J09 and pin 4 of J09 (square pulse).
Important: The device must be in the calibration mode (EtA9), obtained by simultaneously pressing the SILENCE ALARM key, the bolus key and the "ON" key.
8. Check that the amplitude of the square pulse increases when a manual force is exerted on the force sensor. When the force is removed from the sensor, the signal must return to the initial position.
Otherwise, check that the sensor is correctly mounted (sensor/holder friction).
4.2.1. Mounting the potentiometer
1. Disassemble the reducer flask end shield.
Page : 36 NT 1023 Rev.B0
Page 38
2. Mount the potentiometer on the flask (take the nut on a thread).
3. Position the flask in equipment T 300 869 and lock it in place using the knurled screw.
4. Position the potentiometer in the equipment and bring it up against the end shield.
5. Tighten the potentiometer.
6. Extract the flask from equipment T 300 869.
7. Mount the pinion on the potentiometer (match up the indexing half flat).
q The pinion large diameter must be flattened against the potentiometer.
8. With the potentiometer facing you, turn the pinion in an anticlockwise direction until it blocks, then turn it 1/4 of a turn in the opposite direction.
9. Mount the moving mechanical assembly on the reducer frame.
10. Insert the flask on the guides and rack.
q Check the position of the input bearing which must be on the driving block side. q Important: Take care not to damage the flexible circuit when mounting (folding).
11. Secure the end shield using the three M3x3 TC screws.
12. Secure the input bearing using the two M3x3 TC screws.
4.2.2. Wiring the potentiometer
1. Weld the 3 wires perpendicular to the lugs by placing them in the holes.
Important: Do not fold the potentiometer lugs
2. Mount the pinion and the potentiometer matching the indexing half flat with the
3
21
brown red
black
small diameter of the pinion towards the potentiometer.
1 2 3
3. Mount the pinion hocking diameter 4 Truarc ring using a pair of flat mose fliers bearing down on the potenti­ometer shaft half flat.
4. With the pinion facing you, turn the potentiometer anticlockwise until it blocks, then turn it 1/4 of a turn in the opposite direction.
5. Disengage the moving assembly and slide it until it blocks against the reduction gear flange.
6. See EtA6EtA6 calibrating test for the calibration of the potentiometer (section 3.2.2.).
4.2.3. Plug holder connector
If the total pins on the plug holder connector does not correspond to the total pins on the power supply board con­nector, the connecting method should be done as follows:
plug holder
straight
Power supply board
4.2.4. Lubricating the mechanical parts
NT 1023 Rev.B0 page : 37
Page 39
NOTE: Use silicon grease exclusively. Parts to be lubricated:
The two mechanical block guide rods The screw rod The top of the rack at guide level The full length of the rack teeth The disengaging spring housing in the 1/2 nut The inside of the mechanical block.
Page : 38 NT 1023 Rev.B0
Page 40
5 MAINTENANCE
5.1. Recommendations
The qualified technicians in your establishment or our After-Sales Service should be notified of any abnormal op­eration of the device.
For further information concerning troubleshooting or usage procedure, please contact our After-Sales Service or our Commercial Department. (see Useful Addresses, chapter 10).
If the device has to be returned to our After-Sales Service, it must be packed very carefully, if possible in its original packaging before being sent.
FRESENIUS VIAL is not liable for loss or damage to the device during transport to our After-Sales Service.
5.2. Cleaning and disinfection
The syringe pump forms a part of the patient's immediate environment. It is advisable to clean and disinfect the device's external surfaces on a daily basis, in order to protect patient and staff.
q Disconnect the power cable from the wall socket before commencing cleaning. q Do not place in an AUTOCLAVE, nor IMMERSE the device, and do not allow liquids to enter either the de-
vice's casing, or it's power supply cover.
q Use a cloth soaked in DETERGENT-DISINFECTANT, previously diluted with water if required, to destroy
micro-organisms
- Avoid abrasive scrubbing which could scratch the casing.
- Neither rinse, nor wipe surfaces.
q If the device is located in a high contamination risk unit, it is advisable to leave it in the room during aerial
disinfection, after having disinfected it in using a moist cloth.
q Do not use:
- TRICHLOROETHYLENE-DICHLOROETHYLENE,
- AMMONIA,
- AMMONIUM CHLORIDE,
- CHLORINE and AROMATIC HYDROCARBON,
- ETHYLENE DICHLORIDE-METHYLENE CHLORIDE,
- CETONE,
- BASED CLEANING PRODUCTS.
q These aggressive agents could damage the plastic parts and lead to apparatus malfunctions. q Take care with ALCOHOL BASED SPRAYS (20-40% alcohol); They lead to tarnishing of, and small cracks
in, the plastic, and do not provide the requisite cleansing action prior to disinfection.
q Please contact the appropriate service, handling cleaning and disinfection products within your es-
tablishment, for further details.
5.3. Storage
In the case of prolonged storage time, disconnect the battery through the battery door located below the PILOT. This operation should be made by a fully competent technician.
Storage place should be dry and temperate.
q Temperature between 0 and 40°. q Maximum relative humidity 85%, no condensation.
After storage, a full recharge of the battery is recommended before putting the syringe pump into use, in order to avoid any risk caused by micro power cuts in the mains supply and to insure maximum autonomy.
5.4. Routine inspections
NT 1023 Rev.B0 page : 39
Page 41
In order to ensure an optimal functioning of the device, regular servicing inspections and tests are recommended every 24 months. A servicing procedure should include the points listed below. These service checks are not cov­ered by any contract or agreement with FRESENIUS VIAL and are under the responsibility of user’s technical staff. For further information, please contact our After-Sale-Service.
q NOTE: Incorrect servicing inspections may cause wrong functioning of the PILOT.
5.4.1. Checking before use
(see Operator’s guide)
5.4.2. Checking protocol (Technical manual)
In order to ensure that the device is functioning optimally, regular servicing inspections and tests are recommended every 24 months.
A servicing check procedure should include the points listed below. For further information, please contact our After-Sales Service. The following servicing tests must be performed in accordance with the instructions defined in the Technical Man-
ual.
q Note: Failure to comply with these maintenance procedures could damage the device and lead to functional
failures.
5.4.2.1. Checking the syringe clasp detection arm
You have the tool test T301521-B.
1. Start the device by pressing the ON key.
2. Place a syringe plunger head, (take off the plunger from the syringe body) under the anti-siphon arm to stop the syringe plunger head detection alarm.
3. Pull the syringe clasp to the higher position and press the flange detection switch.
q Ensure the syringe head alarm is present.
3. Place the tool reference T301385B with the clasp on the biggest diameter, then move the tool to check the 4 diameters
q For the 2 biggest diameters, the Injectomat 2000 MC / Pilot Delta should display 50 ml capacity. q For the 2 smallest diameters, the Injectomat 2000 MC / Pilot Delta should display 20 ml capacity.
4. Take off the test tool, then place the syringe clasp in lower position and press on the flange detection switch
q Ensure the syringe clasp alarm is present.
You do not have the tool test T301521-B.
1. Start the device by pressing the ON key.
2. Place the syringe clamp in open position and press the flange detection switch.
q Ensure the syringe clasp alarm is present.
3. Place and clamp with the syringe clasp, every syringe type of the list registered in the PILOT
q Ensure right detection for each syringe capacity
4. Take off the syringe then place the syringe clasp in low position and press on the flange detection switch
q Ensure the syringe clasp alarm is present.
5.4.2.2. Checking the flange detection system
1. Start the device by pressing the ON key.
2. Place a syringe in the Injectomat 2000 MC / Pilot Delta and clamp it with the clasp.
Ensure right detection of the syringe.
Page : 40 NT 1023 Rev.B0
Page 42
3. Place a syringe on the Injectomat 2000 MC / Pilot Delta with flange OUT of the provided slot. Put on the syringe clasp.
Ensure the syringe clasp alarm is present.
5.4.2.3. Checking anti-siphon arm
q NOTE: this check should be done for every type of syringe registered on the PILOT
1 Start the device by pressing the ON key. 2 Place and clamp a syringe in the Injectomat 2000 MC / Pilot Delta with the clasp and pull the drive into max-
imum forward position to be sure the drive is not in contact with the head of the syringe.
3 When anti-siphon arm is in low position:
q Ensure plunger head presence detection alarm is present.
4 When anti-siphon arm is in high position:
q Ensure plunger head presence detection alarm is present.
5 For one of the registered syringes, correctly placed (clasp and drive in good position)/
q Ensure plunger head presence detection alarm is absent.
6 Pull the anti-siphon arm to check it is correctly fixed. NOTE: If anti-siphon arm is dismounted, use stop-screw-glue when remounting.
NT 1023 Rev.B0 page : 41
Page 43
5.5. Technical check procedure PILOT anaesthesia 2
For preventive maintenance, periodical check is recommended. Use the technical manual for good running of the procedure.
NOTE: Make a full battery charge (16 hours) before starting check procedure.
Page 44
5.6. Checking disengagement system
1.Start the device.
NT 1023 Rev.B0 page : 43
Page 45
2.Place a 50/60 ml syringe in the PILOT
3.Activate the disengaging lever.
q Ensure an acoustic and visual alarm is present.
4.Maintain the disengagement lever in high position and move the driving block
q Check easy driving block displacement (manual displacement).
5.Release the lever
q Ensure there is no alarm. q Check driving block locking.
5.7. Checking force sensor
1. Proceed to test 9 (see section 3.3: checking the PILOT)
2. Use the disengagement system to pull the pusher backward so no force is applied to the sensor.
q Displayed value should be between 0 and 100 g
3. Press and release the force sensor
q Displayed value should be between 0 and 100 g
4. Apply a known force (about 5 Kg) on the pusher.
q Displayed value should be 5 Kg +/- 250g.
5.8. Checking back-pressure
NOTE: The Injectomat 2000 MC / Pilot Delta initiates the sensor after pusher engagement. When testing back­pressure, the infusion should be started with no pressure on the line.
1. Start the device by pressing the ON key.
2. Place a manometer (or any other pressure measuring instrument) at the syringe outlet.
3. Select BD 60 ml Syringe.
4. Select the Medium Limit Pressure by pressing the LIMIT PRESSURE key.
q M = (medium limit pressure) = 500 mmHg +/- 75 or 0.65 bar ± 0.1 bar.
4. Select maximum flow rate
5. Start the infusion (press START/CONFIRM).
q Ensure there is no acoustic and visual alarm (back-pressure Led off), q Check infusion indicators are flashing. q Check visual and acoustic alarm for a 500 mmHg +/- 75 or 0.65 bar ± 0.1 bar pressure.
NOTE: If the pressure value measured according to the selection made falls outside the reference values, refer to the "EtA9EtA9 " calibration test (see Chap.3.2.3.).
NOTE: Before calibrating the force sensor, check the voltage between point J09.1 (earth) and J09.4 using an oscil­loscope. Voltage value = 0.6 V + 0.05 V (square pulse).
Page : 44 NT 1023 Rev.B0
Page 46
5.9. Checking registered syringe list /syringe list label.
This test allows to check if the list of syringes registered in the Injectomat 2000 MC / Pilot Delta is in accordance with the list printed on the label.
Proceed to HH test (section 3.3: CHECKING THE PILOT)
5.10. Checking Mains/Battery operation
1. Connect the device to a mains supply and check the presence of the mains Led (permanent yellow LED on).
2. Disconnect the device from the mains.
3. Connect the device battery lugs to a stabilised supply set at 6.3 V.
Important: Respect the +/-" polarities.
4. Place the device in the normal operating mode.
5. Select a syringe (from the syringes proposed by the PILOT) and validate.
6. Select a flow rate and validate.
7. Set the voltage on the stabilised supply between 5.8 V. and 6 V.
q The battery discharge pre-alarm is activated
NOTE: The acoustic alarm can be temporarily silenced (2 minutes) by pressing the SILENCE ALARM key.
8. Turn down the voltage of the stabilised supply.
Check the battery discharge prealarm is activated between 5.6 and 5.8 Volts.
NOTE: If the values indicated above are not respected, refer to the "EtA4EtA4" test (Section: 3.2.1).
5.11. Checking linearity
Carry out this test to check the displacement of the driving block for a 50/60 ml B-D Plastipak syringe at a flow rate of 50 ml/h. Any non-correspondence between the value measured and the table value indicates a mechanical and electronic failure which could cause flow rate errors.
A checking software ISCTRL allow to carry out this test automatically (call FRESENIUS VIAL After-Sale- Service for further information)
5.11.1. Equipment used
q Stop clock q Electronic calliper q 50 ml B-D Plastipak syringe
5.11.2. Operating mode
1. Switch the device on by pressing on the ON key.
2. Place a 50/60 ml B-D Plastipak syringe on the device.
over view
3. Measure the distance X (in millimetres) as shown on the diagram -> X
1
4. Select a 50 ml B-D Plastipak syringe.
5. Select a flow rate of 50 ml/h.
NT 1023 Rev.B0 page : 45
Page 47
6. Start the infusion by pressing on the CONFIRMATION key and start the stop clock at the same time.
7. Stop the infusion after 50 min, then measure X again -> X
8. The displacement X= X1 - X
in mm, must comply with the values indicated below.
2,
2
Table of "distance X" displacement measurements
NOTE: For precise measurement avoid any movement of the pusher during measurement.
Time Displacement in mm, for a flow rate selection of 50 m
minutes seconds minimum Average Maximum
50 00 74.96 75.72 76.47 50 30 75.71 76.47 77.24 51 00 76.46 77.23 78.00 51 30 77.21 77.99 78.77 52 00 77.96 78.74 79.53 52 30 78.71 79.50 80.30 53 00 79.46 80.26 81.06 53 30 80.21 81.02 81.83 54 00 80.95 81.77 82.59 54 30 81.70 82.53 83.35 55 00 82.45 83.29 84.12 55 30 83.20 84.04 84.88 56 00 83.95 84.80 85.65 56 30 84.70 85.56 86.41 57 00 85.45 86.32 87.18 57 30 86.20 87.07 87.94 58 00 86.95 87.83 88.71 58 30 87.70 88.59 89.47 59 00 88.45 89.34 90.24 59 30 89.20 90.10 91.00 60 00 89.95 90.86 91.77 60 30 90.70 91.62 92.53 61 00 91.45 92.37 93.30 61 30 92.20 93.13 94.06 62 00 92.95 93.89 94.83
q Important: if the distance measured does not correspond to the value indicated on the table refer to
"EtA6EtA6" Calibration mode.
5.12. Checking end of infusion
5.12.1. End of infusion pre-alarm
1. Start the device.
2. Select a 60 ml B-D PLASTIPAK syringe, 20 ml full.
3. Select a flow rate of 120 ml/hr.
q For normal flow rates, the end of infusion pre-alarm is activated 5 minutes before end of infusion. q For higher flow rates (50 ml/hr), the pre-alarm is activated when the volume remaining to be infused equals
10% of total syringe capacity.
4. Ensure the end of infusion pre-alarm is present.
5. Press the SILENCE ALARM key to silence the acoustic alarm and check the end of infusion alarm (see in
5.4.2.7.2.).
5.12.2. End of infusion alarm
1. At the end of infusion:
Page : 46 NT 1023 Rev.B0
Page 48
q Check the acoustic and visual end of infusion alarm.
over view
2. Measure the " hard height ": X, with 18.6 < X < 19.5 for a 50 ml B- D PLASTIPAK syringe.
q Important:
For accurate checking of the hard height, do not move the driving block when measuring.
3. If X value is out the tolerated ones see " Eta6", section 3.2.,CALIBRATION.
5.13. Checking the Led's and keyboard
Refer to tSt2tSt2 and tSt3tSt3 (see section 3.2.: CALIBRATION)
5.14. Battery autonomy test
1. Medium battery autonomy is 7 hours (minimum 5) when the device operates with a 50 ml B-D PLASTIPAK sy­ringe at a flow rate of 5 ml/hr (without Master module connection type)
2. The battery discharge pre-alarm warns the user the remaining autonomy is roughly 60 minutes (at 5 ml/h) be­fore infusion will completely stop (total battery discharge alarm).
3. Battery recharging time is 16 hours minimum (100% of its capacity).
5.15. Continuity test
Using an multimeter connected ohmmeter, check the presence of an infinite electrical resistance between:
q phase and metal tube, q neutral and metal tube.
NT 1023 Rev.B0 page : 47
Page 49
5.16. Trouble Shooting
Problem Causes
Perfusion end detected too early (about 10 ml).
Syringe used does not fit to selected one.
No occlusion prealarm and
alarm at perfusion end.
Major variation in flow rate or displacement control.
Occlusion alarm when
switching on.
Occlusion alarm when work­ing.
Disengagement alarm when switching on.
Detection syringe piston un­suitable alarm.
Syringe barrel holder unsuit­able alarm.
Syringe flange detection un­suitable alarm.
Syringe used does not fit to selected one.
Syringe used does not fit to selected one.
Bad calibration of strength sensor.
strength sensor out of order.
Cut in flexible circuit.
Pressure limit selected too low.
Bad calibration of strength sensor.
Disengagement microswitch out of order.
Opto sensor and/or obturator of syringe piston out of or-
der.
Opto sensor and/or obturator of syringe holder out of or­der.
Flange detection switch and/or connectic out of order.
Bad display : segments or
LED’s.
Driving transistors and/or connectic of display board out of order.
Alarm without error code. Wrong power supply (6,9V) See MAX 652
Bad CPU board.
In case of drop….. Check mechanic set and centring of diameter 12 tube
See exploded view.
Page : 48 NT 1023 Rev.B0
Page 50
5.17. Error message Er(-)0, Er01, Er(-)2, Er03, CFPc
Error code Error code definition Cause ER (-) 0
Device self test on start up was not successful
Er10 Er20 Er30 Er40
Electronic control + soft­ware anomaly
Internal RAM anomaly External RAM anomaly EPROM anomaly Eeprom anomaly
Check Sum: faulty RAM, EPROM, EEPROM
When rewriting EEPROM on switching off the device the Check Sum is re­written in the memory to save the pa­rameters.
r If the Hard cutoff circuit time is shorter than the Soft circuit time, the device is switched off before EEPROM
is fully written: Check Sum not conform Er50 Er60 Er70
Er80
ER0 1 Problem detected in motor con­trol mechanism or in the motor itself. ER(-) 2* Plunger advance checking has detected a error greater than the acceptable one (+)-) from 1 to 7: info: A.S.S. Er32
Er52 Er72
ADC anomaly Syringe parameters
anomaly Motor frequency anomaly
Keyboard anomaly
Motor supply failure Motor rotation photo switch + associ-
Mechanical advance checking anomaly
Short distance anomaly Slack adjustment anomaly All length anomaly
r Er(-)0 or CFPc: When the device is
in CFPc reconfiguration is compulsory:
faulty WATCH DOG
Faulty keyboard or short circuit in the
keyboard or excessive electromagnetic
interference.
ated circuits brake.
Potentiometer or ADC Often connector
calibration or position potentiometer
unscrewed
Er82 ER(-)4
Possible misfonctioning
Er14 Er24 Er34
CFPc The device self-test on configu­ration was not satisfactory
Flow rate anomaly Motor and flow rate calcu-
lation parameters anoma­lies
Motor period calculation anomaly Motor rotation direction anomaly
Flow rate / period calcula­tion anomaly Configuration anomaly Erroneous parameters.
Functioning or configuration wrong pa-
rameters
5.18. Flow rate control protocol: flow rate measurement with computer
NT 1023 Rev.B0 page : 49
Page 51
The test procedure outlined below can be carried out with a 50 ml or 20 ml syringe. The operating mode described below, reflects the flow rate Measurement software used by Fresenius Vial according to the Pr EN 60-601-2-24 Standard for Infusion pumps. It is up to the user to adapt this procedure to the software he uses.
5.18.1. Equipment used:
q Scales coupled to a microcomputer
scales sensitivity (in compliance with IEC 601.1 standard, 2nd part):
flow rate value (x) scale sensivity
x 5 ml/h
5 ml/h < x 30 ml/h
1/10000e
1/1000e
x > 30 ml/h 1/100e
q Multi scales acquisition program q Test tube or beaker with 1 ml graduating q Liquid: distilled water +/ oil) q Luer Lock type plastic syringe (50 or 20 ml) q Catheter extension with Luer Lock end piece (length 100 cm, inside diameter 2.5 mm). q Needle:
flow rate value (x) nodale type
x < 30 ml/h G 26
x 30 ml/h G 18 ou G 21
5.18.2. installation
1. The equipment should be installed according to the installation drawings shown in below.
q Important: Make sure that the horizontal installation plane is respected.
2. Fill the syringe with 50 ml of distilled water. Prime if necessary to eliminate any air bubbles.
3. Secure the female Luer Lock end piece of the catheter extension onto the syringe and the male Luer Lock end piece onto the needle.
4. Install the syringe onto the device.
5. Fill the test tube with water ensuring that the needle is dipped in the liquid (1 cm) and add several drops of oil in order to create a greasy film on the surface of the liquid. In this way the user will avoid any measurement error due to evaporation of the liquid.
6. Place the test tube in the centre of the scales platform.
7. Place the needle inside the test tube.
q Important: The infusion line (needle/catheter extension) must not be in contact or rest on the Scales/test
tube assembly at any time.
8. Start the device by pressing on the ON key (Injectomat 2000 MC / Pilot Delta in mains supply mode) then prime the infusion line using the PRIME/BOLUS key.
q Important: check that there are no air bubbles.
Page : 50 NT 1023 Rev.B0
Page 52
5.18.3. Operating mode
q Remark: the software works following the operating mode described in the Pr EN 60-601-2-24 Standard for
infusion pumps.
1. Start the acquisition program for the scales.
2. Enter the data necessary to carry out the program without validating the flow rate.
3. Adjust the scales to the specified flow rate.
4. Confirm the flow rate on the microcomputer so that the automatic setting of the scales can take place.
5. Start infusion when 00.00 appears on the scales display screen.
6. When the specified time is over, note the error percentage displayed on the screen.
5.18.4. Installation drawing
catheter extension support stand
RS 232 cord
anti-vibration measurement table
5.19. Flow rate control: flow rate control measurement with scales
q In compliance with IEC 601.2 standard, 2nd part - 62D standard project for infusion pumps.
5.19.1. Equipment used:
q Stop clock q Scales - scales sensitivity:
flow rate value (x) scale sensivity
x 5 ml/h
5 ml/h < x 30 ml/h
x > 30 ml/h 1/100e
q Test tube or beaker with 1 ml graduating q Liquid: distilled water +/ oil)
1/10000e
1/1000e
q Luer Lock type plastic syringe (50 or 20 ml) q Catheter extension with Luer Lock end piece (length 100 cm, inside diameter 2.5 mm) q Needle:
flow rate value (x) needle type
x < 30 ml/h G 26 x 30 ml/h
G 18 ou G 21
NT 1023 Rev.B0 page : 51
Page 53
5.19.2. Installation
1. The equipment should be installed according to the installation drawings shown in 5.18.3.
Remark: Make sure that the horizontal installation plane is respected.
2. Fill the syringe with 50 ml of distilled water. Prime if necessary to eliminate any air bubbles.
3. Secure the female Luer Lock end piece of the catheter extension onto the syringe and the male Luer Lock end piece onto the needle.
4. Install the syringe on the device following the instructions described in the Operator's Guide (See Operator's Guide PILOT anaesthesia 2, Chap. 3: positioning the syringe).
5. Fill the test tube with water ensuring that the needle is dipped in the liquid (1 cm) and add several drops of oil in order to create a greasy film on the surface of the liquid. In this way the user will avoid any measurement error due to evaporation of the liquid.
6. Place the test tube in the centre of the scales platform.
7. Place the needle inside the test tube.
Important: The infusion line (needle/catheter extension) must not be in contact or rest on the Scales/test tube as­sembly at any time.
8. Start the device by pressing on the ON key (Injectomat 2000 MC / Pilot Delta in mains supply mode) then drain the perfusion line using the PRIME/BOLUS key.
Important: check that there are no air bubbles.
5.19.3. Operating mode
1. Select the flow rate
q Important: for low flow rates (< 5 ml/hr) validate and wait for the infusion to stabilise for 1 hour. For higher
flow rates, 10 to 30 minutes are sufficient for this stabilisation.
2. Set the scales at 00.00 g
3. Start infusion by pressing on the START/CONFIRM key and set off the stop clock at the same time (if neces­sary make a note of the stop clock start value).
4. The test lasts for 1 hour. When over, press on the STOP key to stop the infusion.
5. Note the value in grams of "infused" liquid.
6. Calculate the difference between the design value and the real value.
Remark: 1 gram = 1 ml.
7. The error percentage can be calculated from this difference.
Formula:
Measured value - Design value
X 100 = percentage
D designed value
Page : 52 NT 1023 Rev.B0
Page 54
5.20. Flow rate control: flow rate measurement using a test tube
5.20.1. Equipment used
q Stop clock q Test tube or beaker with 1 ml 0.5 ml graduating q Liquid: distilled water +/ oil) q Luer Lock type plastic syringe (50 or 20 ml) q Catheter extension with Luer Lock end piece (length 150 cm, inside diameter 2.5 mm). q Needle
flow rate value (x) needle type
x < 30 ml/h G 26 x 30 ml/h
5.20.2. Installation
1. The equipment should be installed according to the installation drawings shown in 5.18.3.
q Remark: Make sure that the horizontal installation plane is respected.
G 18 ou G 21
2. Fill the syringe with 50 ml of distilled water. Prime if necessary to eliminate any air bubbles.
3. Secure the female Luer Lock end piece of the catheter extension onto the syringe and the male Luer Lock end piece onto the needle.
4. Install the syringe on the device.
5. Fill the test tube with water ensuring that the needle is dipped in the liquid (1 cm) and add several drops of oil in order to create a greasy film on the surface of the liquid. In this way the user will avoid any measurement error due to evaporation of the liquid.
6. Place the needle inside the test tube.
5.20.3. Operating mode
1. Select the flow rate
q Important: for low flow rates (5 ml/h) validate and wait for the infusion to stabilise for 1 hour. If possible, use
a smaller test tube guaranteeing greater precision in ml reading. For higher flow rates, 10 to 30 minutes are sufficient for this stabilisation.
2. Start infusion by pressing on the START/CONFIRM key and set off the stop clock at the same time (if neces­sary make a note of the stop clock start value).The length of the test is determined by the time necessary for a 50 ml infusion in the test tube.
3. Calculate the difference between the design value and the real value.
Real flow rate =
50 ml
Time in hours
4. The error percentage is calculated from this difference (Measurement error +/-1%),
Measured value - Design value
X 100 = percentage
D designed value
NT 1023 Rev.B0 page : 53
Page 55
6 ANNEX 1: ILLUSTRATED PARTS LIST
6.1. Subassembly traceability table
6.1.1. Introduction
This chapter allows the technician to find which component has been changed on the product, also to order the right part when it necessary for the pump maintenance.
6.1.2. Replacement parts table
The table below list the main modification made to improve the product. The Injectomat 2000 MC / Pilot Delta serial number should be used when looking for components.
Equivalence table for
Injectomat 2000 MC / Pilot Delta
Serial N° From:
16685686
To : 18091461 Display board CPU board Power supply and control board
Eprom Upper case IEC kit Upper case export kit Lower case IEC kit Lower case DIN kit Lower case NL kit Lower case IS kit Wired motor Battery connector Potentiometer wired
167549 167566
167568 167543
167557 167537 (Injectomat)
167991-V1.1a 167991-V1.2a
199250 199250
199251 199251
199200 199200
199201 199201
199202 199202
199203 199203
167965 167965
167961 166901
167963 167963
From:
18091461
167536 (Pilot Delta)
Page : 54 NT 1023 Rev.B0
Page 56
6.2. Exploded views and related parts lists
6.2.1. Mechanical part list
Upper and lower cases
NT 1023 Rev.B0 page : 55
Page 57
6.2.1.1. Upper case Reference Repère Quantity Designation
167744 99 1 Diam.17.5 Pilot Delta 167681 99 1 Diam.17.5 Injectomat 167476 100 1 Syringe clamp Pilot Delta 167516 100 1 Syringe clamp Injectomat 167458 113 1 Syringe clamp shaft 167361 102 1 Diam 5 retaining ring 167310 103 1 Syringe clamp compression spring 167452 104 1 Injected PC opto support 168343 111 1 PILOTE ANES window 167679 112 1 Injectomat 2000 MC German front panel 168159 112 1 Pilot Delta French front panel 167627 1 Injected Diam.6 block 199560 1 Female hybrid M3x12 spacer 167067 1 CPU board protector film 167632 1 Buzzer foam 167042 122 2 Opto 167178 123 1 Flange pilot opto IC 167462 124 1 Pilot 20/60 cc shutter 167377 125 1 Flange pilot switch joint 167365 126 1 ALPS SKHCAF switch 167376 127 1 Flange pilot switch support 199618 128 1 TCB 2.2 x 8 Eco-Syn screw 199518 129 1 Flexible stiffened washer 167632 1 Buzzer bell foam
6.2.1.2. Lower case
Reference Repère Quantity Designation
170685 105 1 9 points Sub d 170687 106 1 15 points Sub d 177201 115 1 16 VA transformer 167942 117 1 Main socket support 167093 107 1 Buzzer bell 167285 108 1 Socket support 167200 109 1 3 points female socket 167432 110 1 Injected buzzer adjustment button 167059 1 Guide rail (Pilot Delta) 167514 1 Guide rail (Injectomat) 167299 1 Buzzer adjustment washer 167297 1 Buzzer flexible washer 167249 119 6 Adhesive black block 174019 120 1 6V 1.2/1.3 Ah battery 166901 121 1 Wired Battery sockets HE 13 167122 1 flexible Circuit retainer, Injected 167355 1 Main socket fiber joint
Page : 56 NT 1023 Rev.B0
Page 58
Mechanical plunger unit PILOT anaesthesia 2
NT 1023 Rev.B0 page : 57
Page 59
6.2.1.3. Plunger / mechanism unit
Ref.: Diag. Ref. Number Component
167360 200/207 2 Diam. 3.5 mm retaining ring 167465 201 1 Half nut spring cap 167469 202 1 Half nut spring 167460 203 1 Disengagement came 167471 204 1 Upper machined half nut 167472 206 1 Lower machined half nut 167475 208 1 Tube retainer 167281 209 1 Mechanical block + injected pad 167275 210 1 Flexible circuit clip 167271 220 1 Flexible circuit 167241 221 1 Complete disengagement shaft 167292 222 1 Diam.12 tube(version 20/60 ml) 167403 223 1 Injected centring ring 177203 1 O ring (to be placed in the centring ring) 177204 1 Inox plate (To be placed between the centring ring and the flask) 167291 224 1 Anti-siphon arm 167521 225 1 Pusher housing (injectomat) 167382 225 Pusher housing (Pilot Delta) 167259 226 1 Pusher housing clip 167442 227 1 Injected M 0.5 rack 167683 228 1 Diam. 36 protective sticker 167384 229 1 Injected contact plate 167132 230 1 Force sensor 173408 232 1 OMRON microswitch 162311 249 1 RP I 131 type photo inter 167288 253 1 C pusher support 167497 254 1 Pusher protection film 168231 256 1 Disengagement finger 167298 258 2 Flexible circuit protector 167272 259 1 Pusher/tube fastener 190714 260 1 Retaining ring 167385 261 1 Anti-siphon came 167264 1 Anti-siphon spring 167260 270 1 Disengagement lever 167245 271 1 Disengagement lever spring 167056 272 1 Pusher cover (Pilot Delta) 167519 272 1 Pusher cover (Injectomat) 167464 205 1 Half nut spring shaft 167487 257 1 Disengagement shaft bearing
Page : 58 NT 1023 Rev.B0
Page 60
Motor reduction bloc
Mechanical kits
NT 1023 Rev.B0 page : 59
Page 61
6.2.1.4. Motor reduction block
Reference Diag. Ref. Number Component
176443 294 1 Injected M5 gear 167963 296 1 HE13 wired potentiometer 167128 300 1 Motor rotation photo switch 168401 301 1 Support opto 167111 302 1 Motor rotation blade 167082 303 1 Pilot 2 Ph motor 167153 304 1 Pilot motor gear 167101 305 1 R100 reduction gear mainframe
6.2.1.5. Mechanical kits
Reference Diag. Ref. Number Component
170416 1 Main cord 167969 105/6/8/9 1 HE13 wired socket support C/D (Pilot Delta) 199188 105/6/8/9 1 HE13 wired socket support C/D (Injectomat) 167659 122/123 1 HE13 wired opto printed circuit 199136 298 1 pilot anest 2 mechanical kit
6.2.2. Electronical parts list
6.2.2.1. VA Power supply board
Reference Diag. Ref. Number Component
167536 116 1 Injectomat 2000 MC / Pilot Delta RS232 HE13 power
supply board (Pilot Delta)
167537 116 1 Injectomat 2000 MC / Pilot Delta RS232 HE13 power
supply board (Injectomat)
6.2.2.2. led display board
Reference Diag. Ref. Number Component
167566 114 1 PILOT Anaesthesia 2/B 9 LEDS display board
6.2.2.3. CPU board
Reference Diag. Ref. Number Component
167543 178 1 CPU board W. flange det. ANT + HE13
Warning
A Injectomat 2000 MC / Pilot Delta CPU Board should never be used on Pilot C, and vice versa. A CPU board is configured for one device type only. This configuration include the calibration modes depends to the components characteristics, the syringe list specific to the product code and the pump serial number located on the pump iden­tifical label stacked on the lower housing.
Page : 60 NT 1023 Rev.B0
Page 62
6.2.3. Operator’s guide of Injectomat 2000 MC / Pilot Delta(NU)
It is possible to get this Operator’s guide, references:169540 (Injectomat 2000 MC in German), 169545 (Injectomat 2000 MC in English), 169610 (Pilot Delta in French) on simple request to our commercial service (see useful ad­dresses chapter 9).
6.2.4. Labels
6.2.4.1. Door Label
FR DIN IS BNF IT SP S B P 167892 167892 167892 167892 167892 167892 167892 167892 167892
6.2.4.2. Danger selection label
FR DIN IS BNF IT SP S B P 167741 167979 167776 167811 167778 167858 167886 167692 167692
6.2.4.3. Battery door label
FR DIN IS BNF IT SP S B P 167196 167196 167196 167196 167196 167196 167196 167196 167196
6.2.4.4. Main DANGER label (inside)
FR DIN IS BNF IT SP S B P 171091 171091 171091 171091 171091 171091 171091 171091 171091
6.2.4.5. Mains label(outside)
FR DIN IS NL IT SP S B P 168197 168197 168197 168197 168197 168197 168197 168197 168197
NT 1023 Rev.B0 page : 61
Page 63
7 ANNEX 2: ELECTRONIC LAYOUT
7.1. Power supply and control board
7.1.1. Electronic layout
Designation DESIGN. REF
Motor power supply and control board ( 5/5 X A3) D1A0DB000
7.1.2. Installation layout
Designation DESIGN. REF
Motor power supply and control board Pilote Delta (1/1 x A3) Motor power supply and control board Injectomat 2000MC (1/1 x A3)
7.2. CPU board
7.2.1. Electronic layout
Designation DESIGN. REF
CPU board (3/3 x A3) D1A0EY00
7.2.2. Installation layout
Designation DESIGN. REF
CPU board PILOTE AILETTE RS232 (C) (1/2 x A3 only) D3A0EY00
D3A0DB004 D3A0DB005
7.3. Display board
7.3.1. Electronic layout
Designation DESIGN. REF
Display board Pilote anesthesie 2 (3/3 x A3) D197GT000
7.3.2. Installation layout
Designation DESIGN. REF
Display board Pilote anesthesie 2 (2/2x A3) D397GT000
Page : 62 NT 1023 Rev.B0
Page 64
NT 1023 Rev.B0 page : 63
Page 65
8 ADDENDA AND INFORMATION BULLETINS
NT 0818 Rev.A1
Page 66
NT 1023 Rev.B0
Page 67
NT 0818 Rev.A1
Page 68
9 USEFUL ADDRESSES
All requests for information or documentation (technical file, tubing catalogue or commercial documenta­tion should be addressed to:
CUSTOMER SERVICE INTERNATIONAL
Fresenius Vial
Le Grand Chemin, 38590 Brezins FRANCE
Tel.: 33 (0)4 76 67 10 81or 10 54 Fax: 33 (0)4 76 65 52 22
AFTER-SALES SERVICES
INTERNATIONAL Fresenius Vial
Le Grand Chemin, 38590 Brezins FRANCE
BELGIUM FRESENIUS NV/SA Belgique
DIVISION VIAL MEDICAL
Tel.: 33 (0)4 76 67 10 76 Fax: 33 (0)4 76 65 56 66
Tel.: 32/3 880 73 07
Fax: 32/3 880 50 07 Molenberglei 7 2627 Schelle BELGIQUE
GERMANY
Changes may be made at regular intervals, for inclusion in subsequent editions.
This technical manual may not be reproduced in whole or in part without the written consent of Fresenius Vial S.A.
FRESENIUS MCM
AM-NEUNEN BERG 8 63749 ALZENAU GERMANY
It is possible that this document contains errors or typing mistakes.
COPYRIGHT © 1998, Fresenius Vial S.A
Tel.: 49/60 23 97 22-0
Fax: 49/60 23 43 06
Fresenius Vial S.A. - head office: Le Grand Chemin - 38590 Brezins (FRANCE)
With directory and supervisor board - capital 90128000 F - SIREN Grenoble B 408 720 282
NT 1023 Rev.B0
Loading...