Fresenius 4008 Technical manual

4008 E / 4008 B / 4008 H / 4008 S
Caution!
These Operating Instructions in pdf­format are for information only. They are not a replacement for the Operating Instructions supplied with the machine/device and options.
Hemodialysis system
Technical Manual
Edition: 5/03.09
Part no.: M40 618 1
Software 4.5/5.3 and higher
CareFreseniusMedical
Important information on the Technical Manual
How to use the Technical Manual
Identification The document can be identified by the following information on the title page and
on the labels, if any: – Edition of the technical document – Part number of the technical document
Page identification The page identification 1-3, for example, refers to: chapter 1, page 3.
Editorial information The editorial information 1/01.05, for example, refers to the 1
2005.
Changes Changes to the Technical Manual will be released as new editions or supple-
ments. In general: This manual is subject to change without notice.
Significance of the Explanation of the Caution and Note symbols used: safety precautions
Caution
Advises the operator against certain procedures or actions that could cause damage to the equipment or may have adverse effects on operators and patients.
Note
Informs the operator that if the steps are not followed as described, a specific function will be executed incorrectly or will not be executed at all, or will not produce the desired effect.
st
edition, January
Fresenius Medical Care 4008 5/03.09 (TM) 0-1
Important information on the system
Technician’s qualification
Purpose This Technical Manual is intended for service technicians and is to be used for
first studies (to acquire a basic knowledge) and for reference purposes (for TSC, Maintenance and repair). The Technical Manual, however, does not replace the training courses offered by the manufacturer.
Requirements Knowledge of the current Operating Instructions for the respective system.
Background experience in mechanics, electrical and medical engineering.
Precautions for working on the system
Authorized persons Assembly, extensions, adjustments, modifications or repairs may only be carried
out by the manufacturer or persons authorized by him.
Test equipment and The activities described in this technical document require the availability of the accessories necessary technical test equipment and accessories.
Specifications For the specifications of the respective system, refer to the current Operating
Instructions. Observe the information on the specifications.
Precautions Before turning power on, repair any visible damage.
Prior to opening the system and when working on the open system, the following precautions have to be observed: – Protect the components against ingress of fluids. – Do not touch live parts (e.g. connectors of the power cable or heater). – Disconnect and connect all jacks, connectors and components only when the
system is turned off.
ESD precautions When repairing the system and replacing spare parts, observe the applicable
ESD precautions.
Hygienic measures The system and the consumables are generally considered to be contaminated
and must therefore be sufficiently disinfected by the responsible organization as specified by the manufacturer.
0-2 Fresenius Medical Care 4008 5/03.09 (TM)
Addresses
Please address any inquiries to:
Manufacturer Fresenius Medical Care AG & Co. KGaA
D-61346 Bad Homburg +49 (0)6172-609-0 www.fmc-ag.com
Service Fresenius Medical Care Central Europe Deutschland GmbH
Geschäftsbereich Zentraleuropa Kundendienst / Servicecenter Steinmühlstraße 24 61352 Bad Homburg Germany Phone: +49 (0)6172-609-7100 Fax: +49 (0)6172-609-7102 E-mail: ServicecenterD@fmc-ag.com
International Fresenius Medical Care service Deutschland GmbH
Service Support International Hafenstraße 9 D-97424 Schweinfurt Germany Phone: +49 (0)9721-678-333 (hotline) Fax: +49 (0)9721-678-130
Local service
Fresenius Medical Care 4008 5/03.09 (TM) 0-3
0-4 Fresenius Medical Care 4008 5/03.09 (TM)

Table of contents

Section Page
1 Description of system functions and malfunctions ................................................. 1-
1.1 Description of the T1 test ............................................................................................... 1-3
1.2 Functional description of the modules ............................................................................ 1-78
1.3 Functional description of the hydraulic unit .................................................................... 1-83
2 Technical Safety Checks / Maintenance .................................................................... 2-
2.1 Technical Safety Checks and Maintenance
for 4008 hemodialysis systems and options .................................................................. 2-3
2.2 TSC report ...................................................................................................................... 2-25
3 Adjustment instructions .............................................................................................. 3-
3.1 Overview of the DIP switches in the 4008 ...................................................................... 3-9
3.2 Calibration mode ............................................................................................................ 3-13
3.3 Hydraulics ....................................................................................................................... 3-15
3.4 Air detector ..................................................................................................................... 3-31
4 Calibration program ..................................................................................................... 4-
5 Diagnostics program.................................................................................................... 5-
5.1 General notes ................................................................................................................. 5-3
5.2 Menu structure ................................................................................................................ 5-5
5.3 Reading the analog inputs of CPU I ............................................................................... 5-7
5.4 Reading the analog inputs of CPU II .............................................................................. 5-9
5.5 Reading the digital inputs of CPU I ................................................................................ 5-10
5.6 Reading the digital inputs of CPU II ............................................................................... 5-15
5.7 Writing the analog outputs of CPU I ............................................................................... 5-19
5.8 Writing the analog outputs of CPU II .............................................................................. 5-20
5.9 Writing the digital outputs of CPU I ................................................................................ 5-21
5.10 Writing the digital outputs of CPU II ............................................................................... 5-27
5.11 Writing/Reading the digital outputs of CPU I.................................................................. 5-30
5.12 ONLINE
5.13 HPU ................................................................................................................................ 5-32
6 Setup menu ................................................................................................................... 6-
6.1 Overview Setup menu settings ....................................................................................... 6-3
6.2 Overview ......................................................................................................................... 6-7
6.3 Main menu 4008 E/B ...................................................................................................... 6-9
6.4 Main menu 4008 H/S ...................................................................................................... 6-37
7 Miscellaneous ............................................................................................................... 7-
plus
™ module ..................................................................................................... 5-31
Fresenius Medical Care 4008 5/03.09 (TM) 0-5
Section Page
8 Circuit diagrams and circuit descriptions ................................................................. 8-
8.1 Block diagram 4008 ........................................................................................................ 8-3
8.2 Block diagram of voltage supply .................................................................................... 8-4
8.3 Block diagram of screen 4008 H/S ................................................................................. 8-5
8.4 Connection layout diagram ............................................................................................. 8-6
8.5 P.C.B. LP 450-2 Level detector control (LD) .................................................................. 8-9
8.6 P.C.B. LP 493 Blood leak detector ................................................................................. 8-11
8.7 P.C.B. LP 624 Control board (BP).................................................................................. 8-12
8.8 P.C.B. LP 630 Motherboard ........................................................................................... 8-13
8.9 P.C.B. LP 631 CPU 1 ..................................................................................................... 8-14
8.10 P.C.B. LP 632 CPU 2 ..................................................................................................... 8-16
8.11 P.C.B. LP 633 Input board ............................................................................................. 8-18
8.12 P.C.B. LP 634 Output board........................................................................................... 8-20
8.13 P.C.B. LP 635 Display board.......................................................................................... 8-22
8.14 P.C.B. LP 636 External connectors................................................................................ 8-24
8.15 P.C.B. LP 950 Control board (HEP) ............................................................................... 8-25
8.16 P.C.B. LP 644-4 Display board (HEP)............................................................................ 8-26
8.17 P.C.B. LP 645 Position sensor membrane pump........................................................... 8-27
8.18 P.C.B. LP 649-2 Display board (4008 B/S) .................................................................... 8-28
8.19 P.C.B. LP 742 Interference filter..................................................................................... 8-30
8.20 P.C.B. LP 748 Display board (BP) ................................................................................. 8-31
8.21 P.C.B. LP 763 Multi interface board (COMMCO III) ...................................................... 8-32
8.22 P.C.B. LP 922 Display board (4008 S)........................................................................... 8-33
8.23 P.C.B. LP 923 Traffic light (4008 H/S) ........................................................................... 8-34
8.24 P.C.B. LP 924 Display board (4008 H) .......................................................................... 8-35
8.25 P.C.B. LP 941 Hydraulics processor .............................................................................. 8-36
8.26 Heater board (power supply unit 4008) .......................................................................... 8-37
8.27 Power board (power supply unit 4008) .......................................................................... 8-38
0-6 Fresenius Medical Care 4008 5/03.09 (TM)
Table of contents 1 Description of system functions and malfunctions
Section Page
1.1 Description of the T1 test ............................................................................................ 1-3
1.1.1 T1 test flow diagram, serial program steps .................................................................... 1-3
1.1.2 T1 test flow diagram, parallel program steps ................................................................. 1-5
1.1.3 Description of the T1 test incl. error messages .............................................................. 1-7
1.1.4 Description of system errors during the cleaning programs .......................................... 1-53
1.1.5 Error messages after turning power on .......................................................................... 1-70
1.1.6 Error messages during dialysis ...................................................................................... 1-71
1.2 Functional description of the modules ...................................................................... 1-78
1.2.1 Blood pump (arterial) ...................................................................................................... 1-78
1.2.2 Blood pump (Single-Needle), optional ........................................................................... 1-79
1.2.3 Heparin pump ................................................................................................................. 1-80
1.2.4 Air detector ..................................................................................................................... 1-82
1.3 Functional description of the hydraulic unit ............................................................. 1-83
Fig.: Flow diagram
1.3.1 Description of the hydraulic unit ..................................................................................... 1-85
1.3.2 Theory of operation of the balancing chamber .............................................................. 1-87
1.3.3 Central delivery system option ....................................................................................... 1-91
1.3.4 Program sequences during the cleaning programs ....................................................... 1-92
Fig.: Flow chart of cleaning programs – overview
.......................................................................................................... 1-83
.......................................................... 1-92
Fresenius Medical Care 4008 5/03.09 (TM) 1-1
1-2 Fresenius Medical Care 4008 5/03.09 (TM)

1.1 Description of the T1 test

1.1.1 T1 test flow diagram, serial program steps
MODULE
T1
START T1 TEST
TEST
BYPASS
TEST OK?no
yes
TEST
OPT. DETECTOR
TEST OK?no
yes
TEST
BLOOD SYSTEM
TEST OK?no
yes
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
TEST
ACCUMULATOR
TEST OK?no
yes
TEST
BLOOD LEAK DETECTOR
TEST OK?no
yes
TEST
TEMPERATURE
TEST OK?no
yes
NEG. PRESSURE
HOLDING TEST
TEST OK?no
yes
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
TEST
VENOUS PRESSURE
TEST OK?no
yes
TEST
AIR DETECTOR
TEST OK?no
yes
TEST
DISPLAY
TEST OK?no
yes
TEST
ARTERIAL PRESSURE
TEST OK?no
yes
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
POS. PRESSURE
HOLDING TEST
TEST OK?no
yes
TEST
UF-FUNCTION
TEST OK?no
yes
TEST
CONDUCTIVITY
TEST OK?no
yes
Basic hydraulics Advanced hydraulics
TEST
DIASAFE/HDF FILTER
TEST OK?no
yes
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
STORAGE
ERROR NUMBER
TEST DIASAFE PLUS /
ONLINE PLUS /
HPU TEST
STORAGE
ERROR NUMBER
1
Fresenius Medical Care 4008 5/03.09 (TM) 1-3
1
T1 TEST
UNSUCCESSFUL
no
RETURN
yes
DIALYSIS START KEY
INCORRECT
TEST STEP
TEST OK?no
yes
FURTHER INCORRECT TEST STEPS
no
RETURN
yes
ERROR
DISPLAY
NEXT INCORRECT
TEST STEP
1-4 Fresenius Medical Care 4008 5/03.09 (TM)
1.1.2 T1 test flow diagram, parallel program steps
/
MODULE
T1
START T1 TEST
TEST
OPT. DETECTOR
TEST OK
?
no
STORAGE
ERROR NUMBER
TEST
BLOOD SYSTEM
yes
TEST OK
?
1 1 1 1
STORAGE
ERROR NUMBER
TEST
VENOUS PRESSURE
TEST OK
?
STORAGE
ERROR NUMBER
TEST
AIR DETECTOR
TEST OK
?
STORAGE
ERROR NUMBER
yes
no
yes
no
yes
no
1 1
TEST
BYPASS
TEST OK
?
no
STORAGE
ERROR NUMBER
TEST
TEMPERATURE
TEST OK
?
no
STORAGE
ERROR NUMBER
NEG. PRESSURE
HOLDING TEST
TEST OK
?
no
STORAGE
ERROR NUMBER
TEST
DISPLAY
yes
TEST OK
?
STORAGE
ERROR NUMBER
yes
TEST
BLOOD LEAK DETECTOR
yes
TEST OK
?
STORAGE
ERROR NUMBER
TEST
ACCUMULATOR
yes
TEST OK
?
no
STORAGE
ERROR NUMBER
yes
no
yes
no
TEST
ARTERIAL PRESSURE
TEST OK
?
no
STORAGE
ERROR NUMBER
yes
POS. PRESSURE
HOLDING TEST
TEST OK
?
no
STORAGE
ERROR NUMBER
TEST
CONDUCTIVITY
TEST OK
?
no
STORAGE
ERROR NUMBER
TEST
DIASAFE/HDF FILTER
TEST OK
?
no
STORAGE
ERROR NUMBER
yes
yes
ERROR NUMBER
yes
1
TEST
UF-FUNCTION
TEST OK
?
no
STORAGE
Conductivity?
yes
yesno
TEST
CONDUCTIVITY
TEST OK
?
no
STORAGE
ERROR NUMBER
POS. PRESSURE
HOLDING TEST
TEST OK
?
no
STORAGE
ERROR NUMBER
TEST
UF-FUNCTION
TEST OK
?
no
STORAGE
ERROR NUMBER
yes
yes
Basic hydraulics Advanced hydraulics
TEST
TEST
DIASAFE/HDF FILTER
DIASAFE/HDF FILTER
yes yes
TEST OK
?
1 1
no
STORAGE
ERROR NUMBER
TEST DIASAFE PLUS
ONLINE PLUS /
HPU-TEST
Fresenius Medical Care 4008 5/03.09 (TM) 1-5
1
T1 TEST
UNSUCCESSFUL
no
RETURN
yes
DIALYSIS START KEY
INCORRECT
TEST STEP
TEST OK?no
yes
FURTHER INCORRECT TEST STEPS
yes
RETURN
no
ERROR
DISPLAY
NEXT INCORRECT
TEST STEP
1-6 Fresenius Medical Care 4008 5/03.09 (TM)
1.1.3 Description of the T1 test incl. error messages
Prerequisites for starting and running the test
Error message Description
Power failure Power failure while the test is in progress
Dialines not conn The dialysate lines are not in the interlock shunt.
Shunt Cover open The interlock shunt is open.
Connect Conc.Line Wrong conc. supply The concentrate connector is in the rinse chamber, or concen-
trate is not connected at all. The error message depends on the central delivery system preselected in the setup menu.
Blood Sensed by OD The optical detector senses blood in the system.
Flow alarm Line to or from the dialyzer kinked, malfunctions in the hydraul-
ics.
Water alarm Water supply interrupted.
XXX not calibrated A valid calibration value is missing in the NOVRAM.
Overview of the individual test steps
Bypass test ..................................................................................................................... 1-8
Optical detector test ....................................................................................................... 1-10
Blood system test ........................................................................................................... 1-12
Venous pressure system test ......................................................................................... 1-16
Air detector test .............................................................................................................. 1-18
Display test ..................................................................................................................... 1-22
Arterial pressure system test .......................................................................................... 1-24
Battery test ..................................................................................................................... 1-26
Blood leak test ................................................................................................................ 1-28
Temperature test ............................................................................................................ 1-30
Negative pressure holding test ....................................................................................... 1-32
Positive pressure holding test ........................................................................................ 1-34
UF function test .............................................................................................................. 1-39
Conductivity test ............................................................................................................. 1-42
Diasafe/HDF filter test .................................................................................................... 1-44
Online plus / Diasafe plus filter / HPU test ..................................................................... 1-48
Fresenius Medical Care 4008 5/03.09 (TM) 1-7
Bypass test
H_REL_W EM_H_OFF
V24_EN V24B_EN
CI
V26 V24B V24
V24
V24B
V26
X631/A20
X631/A21
X632/B27
X632/B28
X632/A4
X632/A5
X632/A6
X632/B22
X632/C26
X632/C25
X632/A6
X632/A9
X632/A10
X639/A12 X639/A17
C22
A18
X634R/
X634R/
V26
X634R/
X634R/
A23
V24
V24B
V26
X634L/C25
X634L/C12
X634L/A25
HOT_RINSE
C24
X639/ A20
X632/A26
Test description:
Check of the following functions: – Heater relay – Bypass (electric) – Check of the temperature range changeover
Illustration:
LP 632
CPU 2
LP 639
(4008E/H)
Power
Logic
LP 647
(4008S/B)
LP 631
CPU 1
LP 635
(4008 E)
Display
LP 924
(4008 H)
LP 630
Mother
board
LP 649
(4008 B)
board
LP 922
(4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
LP 633
Input
board
DATA BUS
LP 634
Output
board
1-8 Fresenius Medical Care 4008 5/03.09 (TM)
Error description:
Error message Description
F 01 Bypass The heater relay is switched off.
– Acknowledgement (H_REL_W, X639/A12) X632/A10, 0 V are
missing.
F 02 Bypass The heater relay cannot be switched off by CPU2.
– Acknowledgement (H_REL_W, X639/Y12) X632/A10, 12 V are
missing.
– Control line (EM_H_OFF, X632/A9) X639/A17, 12 V are missing.
F 03 Bypass The temperature measurement range is set to hot rinse.
– Control line (HOTRINSE, X634R/C24) X639/A20, 0 V are miss-
ing.
– Acknowledgement (HOTRINSE, X634R/C24) X632/A26, 0 V are
missing.
F 04 Bypass The extended bypass cannot be correctly switched by CPU2 (V24 = off,
V26 = on, V24B = off). – Acknowledgement (V24, X637/C1) X632/A4, 24 V are missing. – Acknowledgement (V26, X637/C2) X632/A6, 0 V are missing. – Acknowledgement (V24B, X637/C23) X632/A5, 24 V are miss-
ing.
F 05 Bypass The extended bypass cannot be correctly switched off by CPU2 (V24 =
on, V26 = off, V24B = on). – Acknowledgement (V24, X637/C1) X632/A4, 0 V are missing. – Acknowledgement (V26, X637/C2) X632/A6, 24 V are missing. – Acknowledgement (V24B, X637/C23) X632/A5, 0 V are missing.
F06 Bypass CPU1 fails to set the temperature control to hot rinse.
– Control line (HOTRINSE, X634R/C24) X639/A20, 12 V are miss-
ing.
– Acknowledgement (HOTRINSE, X634R/C24) X632/A26, 12 V
are missing.
F 07 Bypass The extended bypass cannot be correctly switched by CPU1 (V24 = off,
V26 = on, V24B = off). – Acknowledgement (V24, X637/C1) X632/A4, 24 V are missing. – Acknowledgement (V26, X637/C2) X632/A6, 0 V are missing. – Acknowledgement (V24B, X637/C23) X632/A5, 24 V are miss-
ing.
F08 Bypass CPU1 fails to reset the temperature control to dialysis.
– Control line (HOTRINSE, X634R/C24) X639/A20, 0 V are miss-
ing.
– Acknowledgement (HOTRINSE, X634R/C24) X632/A26, 0 V are
missing.
F09 Bypass The extended bypass cannot be correctly switched off by CPU1 (V24 =
on, V26 = off, V24B = on). – Acknowledgement (V24, X637/C1) X632/A4, 0 V are missing. – Acknowledgement (V26, X637/C2) X632/A6, 24 V are missing. – Acknowledgement (V24B, X637/C23) X632/A5, 0 V are missing.
F95 Bypass System error
Fresenius Medical Care 4008 5/03.09 (TM) 1-9
Optical detector test
Test description:
Attenuation of the optical detector. Check of the acknowledgement of the optical detector.
Illustration:
LP 632
CPU 2
B27
A20
LP 631
CPU 1
LP 635
(4008 E)
LP 924
(4008 H)
X632/B28
X631/A21
Display
board
X632/C16
X632/C15
X632/A30
LP 649
(4008 B)
LP 922
(4008 S)
LP 630
Mother
board
LDSA
ODSA
OD_OUT
X633L/C7
LP 633
Input
board board
X633L/C8
OD_IN
DATA BUS
X351/5
Pven
X351/7
X351/10
LP 634
Output
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
1-10 Fresenius Medical Care 4008 5/03.09 (TM)
Error description:
Error message Description
F01 opt. Detector CPU1 interprets the optical detector in a different way than does CPU2.
– Acknowledgement (OD_OUT, X633L/C7) X632/A30 and the
digital input of P.C.B. LP 633 measure different levels.
F02 opt. Detector CPU2 fails to recognize blood in the system.
– Acknowledgement (OD_OUT, X633L/C7) X632/A30, 0 V are
missing.
– Detuning (ODSA, X632/C15) X351/7 not 12V.
F03 opt. Detector CPU1 fails to recognize blood in the system.
– Acknowledgement (OD_OUT, X633L/C7) digital input on P.C.B.
LP 633.
– Detuning (ODSA, X632/C15) X351/7 not 12V.
F04 opt. Detector CPU2 recognizes that the optical detector senses opaque fluid (re-
quired because of the test in the cleaning program). – Acknowledgement X632/A30 not 12V. – AD28 defective.
F96 opt. Detector System error.
Fresenius Medical Care 4008 5/03.09 (TM) 1-11
Blood system test
Test description:
Check of the following functions: – Blood alarm acknowledgement – Blood pump switch-off
Illustration:
X632/B27
X631/A20
LP 632
CPU 2
LP 631
CPU 1
X632/C10
X632/C14
X632/C21
X632/B11
X632/A11
X632/A15
X632/B28
X631/A21
LP 630
Mother
board
CLP_CTL
LDA2
BL_AL
BPSB_VEN
BPSB_ART
SN_ART
X633L/ C13
DATA BUS
LP 633
Input
board
A13
X633L/
X634L/
C15
X634L/
A14
A15
LP 634
(4008E/H)
Output
board
LP 647
(4008S/B)
BPST_ART
BPSST_A
X634L/
B14
B15
BPSST_V
X634L/
C14
LP 635
(4008 E)
Display
LP 924
(4008 H)
LP 649
(4008 B)
board
LP 922
(4008 S)
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
X348a/2
X348a/6
X348a/3
X348a/1
X348/V6
X348/V3
X348/V1
Pven
X351/8
X351/6
1-12 Fresenius Medical Care 4008 5/03.09 (TM)
Error description:
Error message Description
F09 Bloodsystem Acknowledgement that CPU2 recognizes that the arterial blood pump
is inactive (BP not running). – Acknowledgement (BPSB_ART, X348a/6) X632/A11, 12 V miss-
ing.
– Control line (BPSST_ART, X634L/B14) X348a/1, 12 V missing or
(BPST_ART, X634L/A14) X348a/3, 12 V are missing.
F10 Bloodsystem Acknowledgement that CPU1 recognizes that the arterial blood pump
is inactive (BP not running). – Acknowledgement (BPSB_ART, X348a/6) X633L/A11, 12 V are
missing.
– Control line (BPSTT_ART, X634L/B14) X348a/1, 12 V missing or
(BPST_ART, X634L/A14) X348a/3, 12 V missing.
– Level is raised during the T1 test.
F11 Bloodsystem The arterial blood pump cannot be stopped by CPU1.
CPU2 recognizes that the arterial blood pump remains active. – Control line (BPSST_ART, X634L/B14) X348a/1, 0 V missing,
as well as (BPST_ART, X634L/A14) X348a/3, 0 V missing.
– Acknowledgement (BPSB_ART, X348a/6) X632/A11, 0 V are
missing.
– The level is raised during the T1 test, or the up/down key on the air
detector is blocked and the level is constantly raised.
F12 Bloodsystem The arterial blood pump cannot be stopped by CPU1.
CPU1 recognizes that the arterial blood pump remains active. – Control line (BPSST_A, X634L/B14) X348a/1, 0 V missing,
as well as (BPST_ART, X634L/A14) X348a/3, 0 V missing.
– Acknowledgement (BPSB_ART, X348a/6) X633L/A11, 0 V are
missing.
F13 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected (= ADKS active)
Acknowledgement that CPU2 detects that the pump is inactive (pump is not running). – Acknowledgement (BPSB_VEN, X348V/6) X632/ B11, 12V
missing
– Control line (BPSST_VEN, X634L/B15) X348V/1, 12V missing or
(BPST_VEN, X634L/A15) X348V/3, 12V missing
– Transistor T9 on P.C.B. LP 754 defective – IC5 on P.C.B. LP 632 defective – In 4008 HDF an HDF treatment was performed, followed by a
cleaning program with the substituate pump running, then the T1 test has been re-started. The substituate pump must be switched off because otherwise the test step will fail to be passed (problem was corrected with SW 3.20 in 4008 H/S systems: the substituate pump will be switched off automatically on starting a cleaning program).
Fresenius Medical Care 4008 5/03.09 (TM) 1-13
F14 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected (= ADKS active)
Acknowledgement that CPU1 detects that the pump is inactive (pump is not running). – Acknowledgement (BPSB_VEN,X348V/6) X633L/A13, 12V
missing
– Control line (BPSST_VEN, X634L/B15) X348V/1 not 12V or
(BPST_VEN, X634L/A15) X348V/3 not 12V
– IC16 on P.C.B. LP 633 defective – P.C.B. LP 633 recognizes Single-Needle pump although it is not
connected.
F15 Bloodsystem
F16 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump. CPU2 detects that the pump remains active. – Control line (BPSST_VEN, X634L/B15) X348V/1, 0V missing as
well as (BPST_VEN, X634L/A15) X348V/3 not 0V
– Acknowledgement (BPSB_VEN, X348V/6) X632/B11, 0V miss-
ing – Transistor T9 on P.C.B. LP 754 defective – IC5 on P.C.B. LP 632 defective – During the test the lines are inserted on the corresponding pump
using the Start/Stop key. – P.C.B. LP 633 recognizes Single-Needle pump although it is not
connected.
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump. CPU1 detects that the pump remains active. – Control line (BPSST_VEN, X634L/B15) X348V/1 not 0V as well
as (BPST_VEN, X634L/A15) X348V/3 not 0V – Acknowledgement (BPSB_VEN, X348V/6) X633L/A13 not 0V – IC16 on P.C.B. LP 633 defective – P.C.B. LP 633 recognizes Single-Needle pump although it is not
connected.
F17 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or 4008 HDF pump is connected (= ADKS active)
Although the recognition of the venous blood pump (ADKS) is not acknowledged, the 24-V supply voltage of the pump can be switched off. – Acknowledgement line (ADKS, X348V/7) X633L/A10 not 12V – Acknowledgement (BPSB_VEN, X348V/6) X633L/A13 not 12V – Acknowledgement (BPSB_VEN, X348V/6) X632/B11 not 12V – Online-HDF has already been switched on during the T1 test. – IC16 on P.C.B. LP 633 defective.
1-14 Fresenius Medical Care 4008 5/03.09 (TM)
F18 Bloodsystem
Applicable for SW 5.00/4.10 and higher, check of the BPUS signal
(CPU, P.C.B. LP 632)
At the beginning of the test step a maximum of 40s may pass until
rotation has stopped.If the blood pump is being activated, the rotation
stop alarm must have been cleared.
– Acknowledgement line (BPUS, X348A/8) X632/A13 not 0V
– Acknowledgement line (BPUS, X348A/8) X632/A13 not 12V
– Blood pump speed is set to “0”: preset speed during the T1 test.
F19 Bloodsystem
Applicable for SW 5.00/4.10 and higher, check of the BPUS signal
(CPU, P.C.B. LP 631 via LP633)
At the beginning of the test step a maximum of 40s may pass until
rotation has stopped. If the blood pump is being activated, the rotation
stop alarm must have been cleared.
1. Acknowledgement line (BPUS, X348A/8) X633L/A12 not 0V
2. Acknowledgement line (BPUS, X348A/8) X633L/A12 not 12V
F20 Bloodsystem Check of the actual arterial BP rate.
The actual rate of the arterial BP is not zero. The actual rate of the
arterial BP does not increase.
If SN is installed: The actual rate of the venous BP is not zero. The
actual rate of the venous BP does not increase.
– Acknowledgement line (BPR_ART, X348A/10) X633L/B3 not 0V
or acknowledgement line (BPR_ART, X348A/10) X632/A14 not 0V
– Acknowledgement line (BPR_ART, X348A/10) X633L/B3 no
increase or acknowledgement line (BPR_ART, X348A/10) X632/ A14 no increase
If SN is installed:
– Acknowledgement line (BPR_VHDF, X348V/10) X633L/B4 not
0V
– Acknowledgement line (BPR_VHDF, X348V/10) X633L/B4 no
increase
F95 Bloodsystem System error.
Fresenius Medical Care 4008 5/03.09 (TM) 1-15
Venous pressure system test
X
X
Test description:
Verification of the lower limit by checking the venous zero point. The upper limit is tested by detuning the venous pressure unit in positive direction. (The venous line clamp is closed during the test.)
Illustration:
LP 632
X632/C16
LDSA
CPU 2
PV_DET
P_VEN
Input
LP 633
LP 634
Output
632/B27
X632/C18
X632/C17
X632/B28
boardboard
X634R/C18
VENT_V
631/A20
LP 631
CPU 1
X631/A21
LP 630
X633L/B5
DATA BUS
Mother
board
LP 635
(4008 E)
LP 649
X351/1
(4008 B)
X351/2
P
VEN
Display
board
LP 924
(4008 H)
LP 922
(4008 S)
X351/4
X351/10
Testgenerierung/Generationof Test
Rueckmeldung/Acknowledgement
1-16 Fresenius Medical Care 4008 5/03.09 (TM)
Error description
Error message Description
F01 Venous CPU1 (input board) shows a venous zero point deviation of more than
±12 mmHg (60 s).
– Control (VENT_VALVE, X634R/C18) X351/1 of the vent valve in
the LD is defective.
– Acknowledgement (P_VEN, X351/4) X633L/B5 that the voltage
value is outside the zero point tolerance.
– P-venous has not been calibrated.
F02 Venous CPU2 shows a venous zero point deviation of more than ±12 mmHg
(60 s).
– Control (VENT_VALVE, X634R/C18) X351/1 of the vent valve in
the LD is defective.
– Acknowledgement (P_VEN, X351/4) X632/C17, the voltage
value is outside the zero point tolerance.
– P-venous has not been calibrated.
F03 Venous With detuning in positive direction, the achieved change in the venous
display is less than 100 mmHg (7 s).
– The test detuning is defective (PV_DET, X632/C18) X351/2.
– Acknowledgement (P_VEN, X351/4) X633L/B5, the change in
voltage is too low.
– P-venous has not been calibrated.
F04 Venous The deviation in the measured value between CPU1 and CPU2 is
higher than ±12 mmHg (if Pven > 100 mmHg).
– Acknowledgement (P_VEN, X351/4) X633L/B5 and X632/C17
measure different voltage values.
– P-venous has not been calibrated.
F95 Venous System error.
Fresenius Medical Care 4008 5/03.09 (TM) 1-17
Air detector test
Test description:
– Test of the air detector by checking the alarm state. – Switch-off of the venous line clamp in the air detector module.
Illustration:
1-18 Fresenius Medical Care 4008 5/03.09 (TM)
Error description:
Error message Description
F01 Airdetector CPU1 interprets the air detector signal in a different way than does
CPU2.
– Acknowledgements (LDA1, X351/14) X632/C13 and X633L/C10
recognize different signal levels.
F02 Airdetector The air detector alarm is not recognized by CPU2.
– Acknowledgement (LDA1, X351/14) X632/C13, 0 V are missing.
– Transmission weakening (LDSA, X632/C16) X351/10, 12 V are
missing.
F03 Airdetector Air detector clamps acknowledgement (CPU2) activated (clamp
closed).
– Acknowledgement (LDA2, X351/6) X632/C14, 24 V are missing.
– Clamp control (CLP_CTL, X634L/C14) X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) X351/8, 12 V are missing.
F04 Airdetector Air detector clamps acknowledgement (CPU1) activated (clamp
closed).
– Acknowledgement (LDA2, X351/6) X633L/C13, 24 V are miss-
ing.
– Clamp control (CLP_CTL, X634L/C14) X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) X351/8, 12 V are missing.
F05 Airdetector The blood alarm signal has not been cleared (indicates an alarm).
– Acknowledgement (BL_AL, X634L/C15) X632/C21, 12 V are
missing.
If the HDF option is used, this signal is not tested (special function).
F06 Airdetector Closing of the air detector clamp via the CPU2 control line was not
possible.
– Clamp control (CLP_CTL, X632/C10) X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) X632/C14, 0 V are missing.
F07 Airdetector Opening of the air detector clamp via the CPU2 control line was not
possible.
– Clamp control (CLP_CTL, X632/C10) X351/8, 12 V are missing.
– Acknowledgement (LDA2, X351/6) X632/C14, 24 V are missing.
F08 Airdetector Closing of the air detector clamp via the CPU1 control line was not
possible, or CPU2 acknowledgement is incorrect.
– Clamp control (CLP_CTL, X634L/C14) X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) X632/C14, 0 V are missing.
Fresenius Medical Care 4008 5/03.09 (TM) 1-19
F09 Airdetector Closing of the air detector clamp via the CPU1 control line was not
possible, or CPU1 acknowledgement is incorrect. – Clamp control (CLP_CTL, X634L/C14) X351/8, 0 V are missing. – Acknowledgement (LDA2, X351/6) X633L/C13, 0 V are missing.
F10 Airdetector The blood alarm message is missing.
– Acknowledgement (BL_AL, X634R/C15) X632/C21, 0 V are
missing.
If the HDF option is used, this signal is not tested (special function).
F11 Airdetector Air detector clamps acknowledgement (CPU2) activated (clamp
closed). – Acknowledgement (LDA2, X351/6) X632/C14, 24 V are missing. – Clamp control (CLP_CTL, X634L/C14) X351/8, 12 V are missing. – Clamp control (CLP_CTL, X632/C10) X351/8, 12 V are missing.
F12 Airdetector Air detector clamps acknowledgement (CPU1) activated (clamp
closed). – Acknowledgement (LDA2, X351/6) X633L/C13, 24 V are miss-
ing. – Clamp control (CLP_CTL, X634L/C14) X351/8, 12 V are missing. – Clamp control (CLP_CTL, X632/C10) X351/8, 12 V are missing.
F13 Airdetector The blood alarm signal has not been cleared (indicates alarm).
– Acknowledgement (BL_AL, X634L/C15) X632/C21, 12 V are
missing.
If the HDF option is used, this signal is not tested (special function).
F14 Airdetector Raise level key on the air detector is constantly active.
– Acknowledgement (LEVEL_UP, X351/3) X632/C11 not 0V.
F15 Airdetector Acknowledgement of the supply voltage for the ultrasonic output stage
not between 6.5 and 13.5 V after 3 seconds. – Adapter board AD28 not connected – Acknowledgement (X351/11 X633L/25A jumper to X633L/B7)
not 12V. – Relay on AD28 failed to drop.
F16 Airdetector Acknowledgement of the supply voltage for the ultrasound output stage
not >14.5V after 3 seconds. – Adapter board AD28 not connected. – Acknowledgement (X351/11 X633L/25A jumper to X633L/B7)
not 16V/24V. – Relay on AD28 is not controlled. – No 10-Hz signal at ALARM_REST (X351/12)
1-20 Fresenius Medical Care 4008 5/03.09 (TM)
F17 Airdetector Acknowledgement of the supply voltage for the ultrasound output stage
not between 6.5 and 13.5 V after 3 seconds.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 X633L/25A jumper to X633L/B7)
not 12V
– Relay on AD28 failed to drop
F95 Airdetector System error.
Fresenius Medical Care 4008 5/03.09 (TM) 1-21
Display test
Test description:
Check of all displays and indicators on the monitor front – Display test – Status LED – Alarm LED – Seven-segment display, all dark – Seven-segment display, all 8888 – Bar graph – CPU1/CPU2 alarm tone
This display test must be monitored by the user!
Illustration:
X632/ B27
X631/ A20
LP 632
CPU 2
LP 631
CPU 1
LP 635
(4008 E)
LP 924
(4008 H)
X632/ B28
X631/ A21
Display
board
X632/C29
LP 649
(4008 B)
LP 922
(4008 S)
LP 630
Mother
board
CPU2_AL
LP 633
Input
board board
DATA BUS
Test Display
8888
8888
8888
8888
X634R/A16
LP 634
Output
+LS X634L/A13
-LS X634L/B13
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
1-22 Fresenius Medical Care 4008 5/03.09 (TM)
Loading...
+ 320 hidden pages