These Operating Instructions in pdfformat are for information only.
They are not a replacement for the
Operating Instructions supplied with
the machine/device and options.
Hemodialysis system
Technical Manual
Edition: 5/03.09
Part no.: M40 618 1
Software 4.5/5.3 and higher
CareFreseniusMedical
Important information on the Technical Manual
How to use the Technical Manual
IdentificationThe document can be identified by the following information on the title page and
on the labels, if any:
– Edition of the technical document
– Part number of the technical document
Page identificationThe page identification 1-3, for example, refers to: chapter 1, page 3.
Editorial informationThe editorial information 1/01.05, for example, refers to the 1
2005.
ChangesChanges to the Technical Manual will be released as new editions or supple-
ments. In general: This manual is subject to change without notice.
Significance of theExplanation of the Caution and Note symbols used:
safety precautions
Caution
Advises the operator against certain procedures or actions that could cause
damage to the equipment or may have adverse effects on operators and
patients.
Note
☞
Informs the operator that if the steps are not followed as described, a specific
function will be executed incorrectly or will not be executed at all, or will not
produce the desired effect.
st
edition, January
Fresenius Medical Care40085/03.09 (TM)0-1
Important information on the system
Technician’s qualification
PurposeThis Technical Manual is intended for service technicians and is to be used for
first studies (to acquire a basic knowledge) and for reference purposes (for TSC,
Maintenance and repair). The Technical Manual, however, does not replace the
training courses offered by the manufacturer.
RequirementsKnowledge of the current Operating Instructions for the respective system.
Background experience in mechanics, electrical and medical engineering.
Precautions for working on the system
Authorized personsAssembly, extensions, adjustments, modifications or repairs may only be carried
out by the manufacturer or persons authorized by him.
Test equipment andThe activities described in this technical document require the availability of the
accessoriesnecessary technical test equipment and accessories.
SpecificationsFor the specifications of the respective system, refer to the current Operating
Instructions. Observe the information on the specifications.
PrecautionsBefore turning power on, repair any visible damage.
Prior to opening the system and when working on the open system, the following
precautions have to be observed:
– Protect the components against ingress of fluids.
– Do not touch live parts (e.g. connectors of the power cable or heater).
– Disconnect and connect all jacks, connectors and components only when the
system is turned off.
ESD precautionsWhen repairing the system and replacing spare parts, observe the applicable
ESD precautions.
Hygienic measuresThe system and the consumables are generally considered to be contaminated
and must therefore be sufficiently disinfected by the responsible organization as
specified by the manufacturer.
0-2Fresenius Medical Care40085/03.09 (TM)
Addresses
Please address any inquiries to:
ManufacturerFresenius Medical Care AG & Co. KGaA
D-61346 Bad Homburg
+49 (0)6172-609-0
www.fmc-ag.com
ServiceFresenius Medical Care
Central EuropeDeutschland GmbH
1.1.3Description of the T1 test incl. error messages
● Prerequisites for starting and running the test
Error messageDescription
Power failurePower failure while the test is in progress
Dialines not connThe dialysate lines are not in the interlock shunt.
Shunt Cover openThe interlock shunt is open.
Connect Conc.Line
Wrong conc. supplyThe concentrate connector is in the rinse chamber, or concen-
trate is not connected at all. The error message depends on the
central delivery system preselected in the setup menu.
Blood Sensed by ODThe optical detector senses blood in the system.
Flow alarmLine to or from the dialyzer kinked, malfunctions in the hydraul-
ics.
Water alarmWater supply interrupted.
XXX not calibratedA valid calibration value is missing in the NOVRAM.
● Overview of the individual test steps
Bypass test .....................................................................................................................1-8
Optical detector test .......................................................................................................1-10
Blood system test ...........................................................................................................1-12
Venous pressure system test .........................................................................................1-16
Air detector test ..............................................................................................................1-18
Display test .....................................................................................................................1-22
Arterial pressure system test ..........................................................................................1-24
Battery test .....................................................................................................................1-26
Blood leak test ................................................................................................................1-28
Temperature test ............................................................................................................1-30
Negative pressure holding test .......................................................................................1-32
Positive pressure holding test ........................................................................................1-34
UF function test ..............................................................................................................1-39
Conductivity test .............................................................................................................1-42
Diasafe/HDF filter test ....................................................................................................1-44
Online plus / Diasafe plus filter / HPU test .....................................................................1-48
Fresenius Medical Care40085/03.09 (TM)1-7
● Bypass test
H_REL_W
EM_H_OFF
V24_EN
V24B_EN
CI
V26
V24B
V24
V24
V24B
V26
X631/A20
X631/A21
X632/B27
X632/B28
X632/A4
X632/A5
X632/A6
X632/B22
X632/C26
X632/C25
X632/A6
X632/A9
X632/A10
X639/A12
X639/A17
C22
A18
X634R/
X634R/
V26
X634R/
X634R/
A23
V24
V24B
V26
X634L/C25
X634L/C12
X634L/A25
HOT_RINSE
C24
X639/
A20
X632/A26
Test description:
Check of the following functions:
– Heater relay
– Bypass (electric)
– Check of the temperature range changeover
Illustration:
LP 632
CPU 2
LP 639
(4008E/H)
Power
Logic
LP 647
(4008S/B)
LP 631
CPU 1
LP 635
(4008 E)
Display
LP 924
(4008 H)
LP 630
Mother
board
LP 649
(4008 B)
board
LP 922
(4008 S)
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
LP 633
Input
board
DATA BUS
LP 634
Output
board
1-8Fresenius Medical Care40085/03.09 (TM)
Error description:
Error messageDescription
F 01 BypassThe heater relay is switched off.
– Acknowledgement (H_REL_W, X639/A12) → X632/A10, 0 V are
missing.
F 02 BypassThe heater relay cannot be switched off by CPU2.
– Acknowledgement (H_REL_W, X639/Y12) → X632/A10, 12 V are
missing.
– Control line (EM_H_OFF, X632/A9) → X639/A17, 12 V are missing.
F 03 BypassThe temperature measurement range is set to hot rinse.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are miss-
ing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are
missing.
F 04 BypassThe extended bypass cannot be correctly switched by CPU2 (V24 = off,
V26 = on, V24B = off).
– Acknowledgement (V24, X637/C1) → X632/A4, 24 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 0 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 24 V are miss-
ing.
F 05 BypassThe extended bypass cannot be correctly switched off by CPU2 (V24 =
on, V26 = off, V24B = on).
– Acknowledgement (V24, X637/C1) → X632/A4, 0 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 24 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 0 V are missing.
F06 BypassCPU1 fails to set the temperature control to hot rinse.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 12 V are miss-
ing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 12 V
are missing.
F 07 BypassThe extended bypass cannot be correctly switched by CPU1 (V24 = off,
V26 = on, V24B = off).
– Acknowledgement (V24, X637/C1) → X632/A4, 24 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 0 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 24 V are miss-
ing.
F08 BypassCPU1 fails to reset the temperature control to dialysis.
– Control line (HOTRINSE, X634R/C24) → X639/A20, 0 V are miss-
ing.
– Acknowledgement (HOTRINSE, X634R/C24) → X632/A26, 0 V are
missing.
F09 BypassThe extended bypass cannot be correctly switched off by CPU1 (V24 =
on, V26 = off, V24B = on).
– Acknowledgement (V24, X637/C1) → X632/A4, 0 V are missing.
– Acknowledgement (V26, X637/C2) → X632/A6, 24 V are missing.
– Acknowledgement (V24B, X637/C23) → X632/A5, 0 V are missing.
F95 BypassSystem error
Fresenius Medical Care40085/03.09 (TM)1-9
● Optical detector test
Test description:
Attenuation of the optical detector.
Check of the acknowledgement of the optical detector.
Illustration:
LP 632
CPU 2
B27
A20
LP 631
CPU 1
LP 635
(4008 E)
LP 924
(4008 H)
X632/B28
X631/A21
Display
board
X632/C16
X632/C15
X632/A30
LP 649
(4008 B)
LP 922
(4008 S)
LP 630
Mother
board
LDSA
ODSA
OD_OUT
X633L/C7
LP 633
Input
boardboard
X633L/C8
OD_IN
DATA BUS
X351/5
Pven
X351/7
X351/10
LP 634
Output
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
1-10Fresenius Medical Care40085/03.09 (TM)
Error description:
Error messageDescription
F01 opt. DetectorCPU1 interprets the optical detector in a different way than does CPU2.
– Acknowledgement (OD_OUT, X633L/C7) → X632/A30 and the
digital input of P.C.B. LP 633 measure different levels.
F02 opt. DetectorCPU2 fails to recognize blood in the system.
– Acknowledgement (OD_OUT, X633L/C7) → X632/A30, 0 V are
missing.
– Detuning (ODSA, X632/C15) → X351/7 not 12V.
F03 opt. DetectorCPU1 fails to recognize blood in the system.
– Acknowledgement (OD_OUT, X633L/C7) → digital input on P.C.B.
LP 633.
– Detuning (ODSA, X632/C15) → X351/7 not 12V.
F04 opt. DetectorCPU2 recognizes that the optical detector senses opaque fluid (re-
quired because of the test in the cleaning program).
– Acknowledgement X632/A30 not 12V.
– AD28 defective.
F96 opt. DetectorSystem error.
Fresenius Medical Care40085/03.09 (TM)1-11
● Blood system test
Test description:
Check of the following functions:
– Blood alarm acknowledgement
– Blood pump switch-off
Illustration:
X632/B27
X631/A20
LP 632
CPU 2
LP 631
CPU 1
X632/C10
X632/C14
X632/C21
X632/B11
X632/A11
X632/A15
X632/B28
X631/A21
LP 630
Mother
board
CLP_CTL
LDA2
BL_AL
BPSB_VEN
BPSB_ART
SN_ART
X633L/
C13
DATA BUS
LP 633
Input
board
A13
X633L/
X634L/
C15
X634L/
A14
A15
LP 634
(4008E/H)
Output
board
LP 647
(4008S/B)
BPST_ART
BPSST_A
X634L/
B14
B15
BPSST_V
X634L/
C14
LP 635
(4008 E)
Display
LP 924
(4008 H)
LP 649
(4008 B)
board
LP 922
(4008 S)
Testgenerierung/Generation of Test
Rückmeldung/Acknowledgement
X348a/2
X348a/6
X348a/3
X348a/1
X348/V6
X348/V3
X348/V1
Pven
X351/8
X351/6
1-12Fresenius Medical Care40085/03.09 (TM)
Error description:
Error messageDescription
F09 BloodsystemAcknowledgement that CPU2 recognizes that the arterial blood pump
is inactive (BP not running).
– Acknowledgement (BPSB_ART, X348a/6) → X632/A11, 12 V miss-
ing.
– Control line (BPSST_ART, X634L/B14) → X348a/1, 12 V missing or
(BPST_ART, X634L/A14) → X348a/3, 12 V are missing.
F10 BloodsystemAcknowledgement that CPU1 recognizes that the arterial blood pump
is inactive (BP not running).
– Acknowledgement (BPSB_ART, X348a/6) → X633L/A11, 12 V are
missing.
– Control line (BPSTT_ART, X634L/B14) → X348a/1, 12 V missing or
(BPST_ART, X634L/A14) → X348a/3, 12 V missing.
– Level is raised during the T1 test.
F11 BloodsystemThe arterial blood pump cannot be stopped by CPU1.
CPU2 recognizes that the arterial blood pump remains active.
– Control line (BPSST_ART, X634L/B14) → X348a/1, 0 V missing,
as well as (BPST_ART, X634L/A14) → X348a/3, 0 V missing.
– Acknowledgement (BPSB_ART, X348a/6) → X632/A11, 0 V are
missing.
– The level is raised during the T1 test, or the up/down key on the air
detector is blocked and the level is constantly raised.
F12 BloodsystemThe arterial blood pump cannot be stopped by CPU1.
CPU1 recognizes that the arterial blood pump remains active.
– Control line (BPSST_A, X634L/B14) → X348a/1, 0 V missing,
as well as (BPST_ART, X634L/A14) → X348a/3, 0 V missing.
– Acknowledgement (BPSB_ART, X348a/6) → X633L/A11, 0 V are
missing.
F13 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
Acknowledgement that CPU2 detects that the pump is inactive (pump
is not running).
– Acknowledgement (BPSB_VEN, X348V/6) → X632/ B11, 12V
missing
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 12V missing or
(BPST_VEN, X634L/A15) → X348V/3, 12V missing
– Transistor T9 on P.C.B. LP 754 defective
– IC5 on P.C.B. LP 632 defective
– In 4008 HDF an HDF treatment was performed, followed by a
cleaning program with the substituate pump running, then the T1
test has been re-started.
The substituate pump must be switched off because otherwise the
test step will fail to be passed (problem was corrected with SW 3.20
in 4008 H/S systems: the substituate pump will be switched off
automatically on starting a cleaning program).
Fresenius Medical Care40085/03.09 (TM)1-13
F14 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
Acknowledgement that CPU1 detects that the pump is inactive (pump
is not running).
– Acknowledgement (BPSB_VEN,X348V/6) → X633L/A13, 12V
missing
– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 12V or
(BPST_VEN, X634L/A15) → X348V/3 not 12V
– IC16 on P.C.B. LP 633 defective
– P.C.B. LP 633 recognizes Single-Needle pump although it is not
connected.
F15 Bloodsystem
F16 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump.
CPU2 detects that the pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1, 0V missing as
ing
– Transistor T9 on P.C.B. LP 754 defective
– IC5 on P.C.B. LP 632 defective
– During the test the lines are inserted on the corresponding pump
using the Start/Stop key.
– P.C.B. LP 633 recognizes Single-Needle pump although it is not
connected.
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
CPU1 fails to stop the corresponding blood pump.
CPU1 detects that the pump remains active.
– Control line (BPSST_VEN, X634L/B15) → X348V/1 not 0V as well
as (BPST_VEN, X634L/A15) → X348V/3 not 0V
– Acknowledgement (BPSB_VEN, X348V/6) → X633L/A13 not 0V
– IC16 on P.C.B. LP 633 defective
– P.C.B. LP 633 recognizes Single-Needle pump although it is not
connected.
F17 Bloodsystem
Applicable for SW 4.91/2.91 and higher if SN, ONLINE-HDF or
4008 HDF pump is connected (= ADKS active)
Although the recognition of the venous blood pump (ADKS) is not
acknowledged, the 24-V supply voltage of the pump can be switched
off.
– Acknowledgement line (ADKS, X348V/7) → X633L/A10 not 12V
– Acknowledgement (BPSB_VEN, X348V/6) → X633L/A13 not 12V
– Acknowledgement (BPSB_VEN, X348V/6) → X632/B11 not 12V
– Online-HDF has already been switched on during the T1 test.
– IC16 on P.C.B. LP 633 defective.
1-14Fresenius Medical Care40085/03.09 (TM)
F18 Bloodsystem
Applicable for SW 5.00/4.10 and higher, check of the BPUS signal
(CPU, P.C.B. LP 632)
At the beginning of the test step a maximum of 40s may pass until
rotation has stopped.If the blood pump is being activated, the rotation
stop alarm must have been cleared.
– Acknowledgement line (BPUS, X348A/8) → X632/A13 not 0V
– Acknowledgement line (BPUS, X348A/8) → X632/A13 not 12V
– Blood pump speed is set to “0”: preset speed during the T1 test.
F19 Bloodsystem
Applicable for SW 5.00/4.10 and higher, check of the BPUS signal
(CPU, P.C.B. LP 631 via LP633)
At the beginning of the test step a maximum of 40s may pass until
rotation has stopped. If the blood pump is being activated, the rotation
stop alarm must have been cleared.
1. Acknowledgement line (BPUS, X348A/8) → X633L/A12 not 0V
2. Acknowledgement line (BPUS, X348A/8) → X633L/A12 not 12V
F20 BloodsystemCheck of the actual arterial BP rate.
The actual rate of the arterial BP is not zero. The actual rate of the
arterial BP does not increase.
If SN is installed: The actual rate of the venous BP is not zero. The
actual rate of the venous BP does not increase.
– Acknowledgement line (BPR_ART, X348A/10) → X633L/B3 not 0V
or acknowledgement line (BPR_ART, X348A/10) → X632/A14 not
0V
– Acknowledgement line (BPR_ART, X348A/10) → X633L/B3 no
increase or acknowledgement line (BPR_ART, X348A/10) → X632/
A14 no increase
If SN is installed:
– Acknowledgement line (BPR_VHDF, X348V/10) → X633L/B4 not
0V
– Acknowledgement line (BPR_VHDF, X348V/10) → X633L/B4 no
increase
F95 BloodsystemSystem error.
Fresenius Medical Care40085/03.09 (TM)1-15
● Venous pressure system test
X
X
Test description:
Verification of the lower limit by checking the venous zero point.
The upper limit is tested by detuning the venous pressure unit in positive direction.
(The venous line clamp is closed during the test.)
Illustration:
LP 632
X632/C16
LDSA
CPU 2
PV_DET
P_VEN
Input
LP 633
LP 634
Output
632/B27
X632/C18
X632/C17
X632/B28
boardboard
X634R/C18
VENT_V
631/A20
LP 631
CPU 1
X631/A21
LP 630
X633L/B5
DATA BUS
Mother
board
LP 635
(4008 E)
LP 649
X351/1
(4008 B)
X351/2
P
VEN
Display
board
LP 924
(4008 H)
LP 922
(4008 S)
X351/4
X351/10
Testgenerierung/Generationof Test
Rueckmeldung/Acknowledgement
1-16Fresenius Medical Care40085/03.09 (TM)
Error description
Error messageDescription
F01 VenousCPU1 (input board) shows a venous zero point deviation of more than
±12 mmHg (60 s).
– Control (VENT_VALVE, X634R/C18) → X351/1 of the vent valve in
the LD is defective.
– Acknowledgement (P_VEN, X351/4) → X633L/B5 that the voltage
value is outside the zero point tolerance.
– P-venous has not been calibrated.
F02 VenousCPU2 shows a venous zero point deviation of more than ±12 mmHg
(60 s).
– Control (VENT_VALVE, X634R/C18) → X351/1 of the vent valve in
the LD is defective.
– Acknowledgement (P_VEN, X351/4) → X632/C17, the voltage
value is outside the zero point tolerance.
– P-venous has not been calibrated.
F03 VenousWith detuning in positive direction, the achieved change in the venous
display is less than 100 mmHg (7 s).
– The test detuning is defective (PV_DET, X632/C18) → X351/2.
– Acknowledgement (P_VEN, X351/4) → X633L/B5, the change in
voltage is too low.
– P-venous has not been calibrated.
F04 VenousThe deviation in the measured value between CPU1 and CPU2 is
higher than ±12 mmHg (if Pven > 100 mmHg).
– Acknowledgement (P_VEN, X351/4) → X633L/B5 and X632/C17
measure different voltage values.
– P-venous has not been calibrated.
F95 VenousSystem error.
Fresenius Medical Care40085/03.09 (TM)1-17
● Air detector test
Test description:
– Test of the air detector by checking the alarm state.
– Switch-off of the venous line clamp in the air detector module.
Illustration:
1-18Fresenius Medical Care40085/03.09 (TM)
Error description:
Error messageDescription
F01 AirdetectorCPU1 interprets the air detector signal in a different way than does
CPU2.
– Acknowledgements (LDA1, X351/14) → X632/C13 and X633L/C10
recognize different signal levels.
F02 AirdetectorThe air detector alarm is not recognized by CPU2.
– Acknowledgement (LDA1, X351/14) → X632/C13, 0 V are missing.
– Transmission weakening (LDSA, X632/C16) → X351/10, 12 V are
– Acknowledgement (LDA2, X351/6) → X633L/C13, 24 V are miss-
ing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
F05 AirdetectorThe blood alarm signal has not been cleared (indicates an alarm).
– Acknowledgement (BL_AL, X634L/C15) → X632/C21, 12 V are
missing.
If the HDF option is used, this signal is not tested (special function).
F06 AirdetectorClosing of the air detector clamp via the CPU2 control line was not
possible.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) → X632/C14, 0 V are missing.
F07 AirdetectorOpening of the air detector clamp via the CPU2 control line was not
possible.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.
F08 AirdetectorClosing of the air detector clamp via the CPU1 control line was not
possible, or CPU2 acknowledgement is incorrect.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) → X632/C14, 0 V are missing.
Fresenius Medical Care40085/03.09 (TM)1-19
F09 AirdetectorClosing of the air detector clamp via the CPU1 control line was not
possible, or CPU1 acknowledgement is incorrect.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 0 V are missing.
– Acknowledgement (LDA2, X351/6) → X633L/C13, 0 V are missing.
F10 AirdetectorThe blood alarm message is missing.
– Acknowledgement (BL_AL, X634R/C15) → X632/C21, 0 V are
missing.
If the HDF option is used, this signal is not tested (special function).
closed).
– Acknowledgement (LDA2, X351/6) → X632/C14, 24 V are missing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
closed).
– Acknowledgement (LDA2, X351/6) → X633L/C13, 24 V are miss-
ing.
– Clamp control (CLP_CTL, X634L/C14) → X351/8, 12 V are missing.
– Clamp control (CLP_CTL, X632/C10) → X351/8, 12 V are missing.
F13 AirdetectorThe blood alarm signal has not been cleared (indicates alarm).
– Acknowledgement (BL_AL, X634L/C15) → X632/C21, 12 V are
missing.
If the HDF option is used, this signal is not tested (special function).
F14 AirdetectorRaise level key on the air detector is constantly active.
– Acknowledgement (LEVEL_UP, X351/3) → X632/C11 not 0V.
F15 AirdetectorAcknowledgement of the supply voltage for the ultrasonic output stage
not between 6.5 and 13.5 V after 3 seconds.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7)
not 12V.
– Relay on AD28 failed to drop.
F16 AirdetectorAcknowledgement of the supply voltage for the ultrasound output stage
not >14.5V after 3 seconds.
– Adapter board AD28 not connected.
– Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7)
not 16V/24V.
– Relay on AD28 is not controlled.
– No 10-Hz signal at ALARM_REST (X351/12)
1-20Fresenius Medical Care40085/03.09 (TM)
F17 AirdetectorAcknowledgement of the supply voltage for the ultrasound output stage
not between 6.5 and 13.5 V after 3 seconds.
– Adapter board AD28 not connected
– Acknowledgement (X351/11 → X633L/25A jumper to X633L/B7)
not 12V
– Relay on AD28 failed to drop
F95 AirdetectorSystem error.
Fresenius Medical Care40085/03.09 (TM)1-21
● Display test
Test description:
Check of all displays and indicators on the monitor front
– Display test
– Status LED
– Alarm LED
– Seven-segment display, all dark
– Seven-segment display, all 8888
– Bar graph
– CPU1/CPU2 alarm tone
This display test must be monitored by the user!
Illustration:
X632/ B27
X631/ A20
LP 632
CPU 2
LP 631
CPU 1
LP 635
(4008 E)
LP 924
(4008 H)
X632/ B28
X631/ A21
Display
board
X632/C29
LP 649
(4008 B)
LP 922
(4008 S)
LP 630
Mother
board
CPU2_AL
LP 633
Input
boardboard
DATA BUS
Test Display
8888
8888
8888
8888
X634R/A16
LP 634
Output
+LS X634L/A13
-LS X634L/B13
Testgenerierung/Generation of Test
Rueckmeldung/Acknowledgement
1-22Fresenius Medical Care40085/03.09 (TM)
Loading...
+ 320 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.