The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may result in
differences between this operating manual and your meter. Also, we cannot
guarantee that there are absolutely no errors in this manual. Therefore, we
are sure you will understand that we cannot accept any legal claims resulting
from the data, figures or descriptions.
The Oxi 3210 compact precision dissolved oxygen (D.O.) meter
enables you to perform D.O. measurements quickly and reliably.
The Oxi 3210 provides the maximum degree of operating comfort,
reliability and measuring certainty for all applications.
The proven OxiCal calibration procedure supports you in your work
with the D.O. meter.
ba75795e0203/2009
1Keypad
2Display
3Socket field
5
OverviewOxi 3210
ENTER
M
Oxi 3210
F1
F2
AR
CAL
STO
RCL
F1
CAL
STO
RCL
1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <ENTER>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<F1>/[Menü]: Opens the menu for measurement settings
<F1
_>/[Menü]: Opens the menu for system settings
<On/Off>:Switches the meter on or off
<M>:Selects the measured parameter
<CAL>:
<CAL
_>:
<STO>:
Calls up the calibration procedure
Displays the calibration data
Saves a measured value manually
<STO_>
<RCL>:
Displays the manually stored measured values
<RCL_>
<▲>: Increments values, scrolls
6
<▼>: Decrements values, scrolls
ba75795e0203/2009
Oxi 3210Overview
ENTER
AR
HOLD AR
25.0
7.92
°C
3
2
4
5
6
7
1
O
2
mg/l
Menü
22.09.2008
08:00
<ENTER>:
<ENTER
<AR>Freezes the measured value (HOLD function)
_>:
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
Switches the AutoRead measurement on or off
1.2Display
1Status information
2Measured value (with unit)
3Measured parameter
4Sensor symbol (calibration evaluation, calibration
interval)
5Measured temperature (with unit)
6Status line
7Softkeys and date + time
Function display
indicators
ErrorAn error occurred during calibration
ARStability control (AutoRead) is active
HOLDMeasured value is frozen (<AR> key)
7
ba75795e0203/2009
OverviewOxi 3210
2
1
1.3Socket field
Connectors:
1D.O. sensor
2Service interface
Caution
Only connect D.O. sensors to the meter that cannot return any
voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary D.O. sensors fulfill these conditions.
8
ba75795e0203/2009
Oxi 3210Safety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the
vicinity of the meter.
Target groupThe meter was developed for work in the field and in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructionsSafety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:
Further notes
Warning
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the meter or
the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.
ba75795e0203/2009
9
SafetyOxi 3210
2.1Authorized use
The authorized use of the meter consists exclusively of the
measurement of dissolved oxygen and temperature of liquid media in
a field and laboratory environment.
The technical specifications as given in chapter 7 T
ECHNICALDATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.
2.2General safety instructions
This meter is constructed and tested in compliance with the IEC 1010
safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICALDATA.
If the meter was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the meter. In this event, wait until the temperature of the
meter reaches room temperature before putting the meter back into
operation.
Caution
The meter is only allowed to be opened by authorized personnel.
10
ba75795e0203/2009
Oxi 3210Safety
Safe operationIf safe operation is no longer possible, the meter must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
Caution
In addition to the safety instructions mentioned here, also follow
the safety instructions of the sensors used.
The operating manuals of the sensors are available on the
supplied CD and on the Internet under www.WTW.com.
ba75795e0203/2009
11
SafetyOxi 3210
12
ba75795e0203/2009
Oxi 3210Commissioning
2
1
3Commissioning
3.1Scope of delivery
z D.O. meter Oxi 3210
z 4 batteries 1.5 V Mignon type AA
z Short instructions
z CD-ROM with detailed operating manual
3.2Initial commissioning
Perform the following activities:
z Insert the supplied batteries
z Switch on the meter
z Set the date and time
3.2.1Inserting the batteries
1Unscrew the two screws (1) on the underside of the meter.
2Open the battery compartment (2) on the underside of the
meter.
ba75795e0203/2009
3Place four batteries (type Mignon AA) in the battery
compartment.
13
CommissioningOxi 3210
Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.
Caution
Make sure that the poles of the batteries are positioned correctly.
The
±signs on the batteries must correspond to the ± signs in the
battery compartment.
4Close the battery compartment (2) and tighten the screws (1).
3.2.2Switching on the meter
1Press the <On/Off> key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test
is being performed.
Subsequently, the meter switches to the measuring mode
(measured value display).
Note
The meter has an energy saving feature to avoid unnecessary battery
depletion.
The energy saving feature switches off the measuring instrument if no
key is pressed during the adjusted interval. (How to set the switch-off
interval, see section 4.3.1).
3.2.3Setting the date and time
2See section 4.2.4
14
ba75795e0203/2009
Oxi 3210Operation
O
2
7.92
mg/l
25.1
°C
22.09.2008
08:00
Menu
4Operation
4.1Switching on the meter
Switching onPress the <On/Off> key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.
Switching offPress the <On/Off> key.
Automatic switch-off
function
Display illuminationThe meter automatically switches off the display illumination if no key
The instrument has an automatic switch-off function in order to save the
batteries (see section 4.3.1). The automatic switch-off function
switches off the measuring instrument if no key is pressed for an
adjustable period.
is pressed for 30 seconds. The illumination is switched on with the next
keystroke again.
You can also generally switch the display illumination on or off (see
section 4.3.1).
ba75795e0203/2009
15
OperationOxi 3210
4.2General operating principles
This section contains basic information on the operation of the
Oxi 3210.
Operating elements,
display
Operating modes,
navigation
An overview of the operating elements and the display is given in
section 1.1 and section 1.2.
An overview of the operating modes and navigation of the Oxi 3210 is
given in section 4.2.1 and section 4.2.2.
4.2.1Operating modes
The meter has the following operating modes:
z Measuring
The measurement data of the connected sensor is shown in the
measured value display
z Calibration
The course of a calibration with calibration information, functions
and settings is displayed
z Data storage
The meter stores measurement data manually
z Setting
The system menu or a sensor menu with submenus, settings and
functions is displayed
16
ba75795e0203/2009
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.