The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may result in differences between this operating manual and your instrument.
Also, we cannot guarantee that there are absolutely no errors in this
manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
The portable Oxi 1970i measuring instrument enables you to carry out
oxygen measurements rapidly and reliably.
The Oxi 1970i provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven OxiCal
®
calibration procedure supports you in your work
with the measuring instrument. The special AutoRead function enables
precise measurements.
1
ON / OFF
M
AUTO READ
RUN / ENTER
CAL
2
O
7
92
m
g/
l
1 7 6
°
C
T
P
R
E
L
1
A
R
n
g
RCL
STO
Oxi 1970i
3
4
ba75691e0101/2007
5
1Keypad
2Display
3Integrated, exchangeable sensor quiver
4Socket field
5Carrying and positioning handle
5
OverviewOxi 1970i
Note
If you need further information or application notes, you can obtain the
following material from WTW:
z Application reports
z Primers
z Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
Select the measuring mode, increase values,
scroll
<▲ >
Select the measuring mode, decrease values,
scroll
<▼ >
RCL
STO
Display/transmit measured values
<RCL>
Save a measured value
<STO>
ba75691e0101/2007
7
OverviewOxi 1970i
1.3Socket field
SensorsYou can use the following sensors with the Oxi 1970i:
z Oxygen depth armature
z D. O. sensor CellOx 325, or
z D. O. sensor DurOx 325, or
z Self-stirring D. O. sensor StirrOx G
Connectors
2
3
4
1
5
Sensor / InstrumentSocket
Oxygen sensor1
Integrated stirrer StirrOx G2
Printer or PC (serial interface, RS232)3
Plug-in power supply unit4
Watertight valve for internal pressure equalization5
Warning
Only connect sensors to the measuring instrument that cannot return
any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Nearly all sensors - especially WTW sensors
- fulfill these conditions.
8
ba75691e0101/2007
Oxi 1970iSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the measuring instrument . Consequently, all responsible personnel must read this
operating manual before working with the measuring system. The operating manual must always be available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in
the laboratory.
We assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take
when handling chemicals.
Safety
instructions
Further notes
The individual chapters of this operating manual use safety instructions
such as the label shown below to indicate various hazards or dangers:
Warning
indicates instructions that must be followed precisely in order to avoid
the possibility of slight injuries or damage to the instrument or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of
the measurement of the dissolved oxygen (D. O.) content of liquid media in the field and laboratory.
The technical specifications as given in chapter 7 T
be observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is authorized.
Any other use is considered to be unauthorized.
ECHNICALDATA must
ba75691e0101/2007
9
SafetyOxi 1970i
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter
7T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
The smooth functioning and operational safety of the measuring instrument can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are
followed during operation.
The smooth functioning and operational safety of the measuring instrument can only be guaranteed under the environmental conditions that
are specified in chapter 7 T
ECHNICALDATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
of service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
10
Obligations
of the purchaser
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
ba75691e0101/2007
Oxi 1970iCommissioning
3Commissioning
3.1Scope of delivery
z Portable measuring instrument, Oxi 1970i with integrated recharge-
able battery
z Carrying and positioning handle
z Carrying strap
z Sensor quiver
z Plug-in power supply unit
z Operating manual
3.2Power supply
Mains operation and
charging the battery
Charging time of the
battery
You can either operate the measuring instrument with the integrated rechargeable battery or with the plug-in power supply. The plug-in power
supply supplies the measuring instrument with low voltage (12 V DC).
At the same time, the rechargeable battery is charged.
Note
The automatic stirring function of the self-stirring D. O. sensor
StirrOx G can only be used when the plug-in power supply is connected.
approx. 16 hours. The battery is charged even when the instrument is
switched off. The LoBat display indicator appears when the battery is
nearly empty and has to be charged as soon as possible.
Warning
The line voltage at the operating site must lie within the input voltage
range of the original plug-in power supply (see chapter 7 T
TA).
ECHNICALDA-
Warning
Use original plug-in power supplies only (see
chapter 7 T
ECHNICALDATA).
ba75691e0101/2007
11
CommissioningOxi 1970i
Connecting the plug-in
power supply unit
2
1
3
Setting the date and
time
1Connect the plug (1) to the socket (2) of the measuring instru-
ment.
2Connect the original WTW plug-in power supply (3) to an easily
accessible mains socket.
3.3Initial commissioning
Perform the following activities:
z For mains operation and charging the battery: Connect the plug-in
power supply (see section 3.2 P
z Set the date and time.
1Press the <M> key and hold it down.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
3Press the <RUN/ENTER> key repeatedly until the date flashes
on the display (Day.Month display indicator).
OWERSUPPLY).
12
ba75691e0101/2007
Oxi 1970iCommissioning
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
0 9 0 4
Tref25
Tref20
nLF
Lin
5t2
Auto
Store
Cal
Oxi
TEC
ARng
4Set the date of the current day with <▲> <▼>.
5Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
6Set the current month with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The year appears on the display.
8Set the current year with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The hours flash on the display.
10Set the current time with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The minutes flash on the display.
12Set the current time with <▲> <▼>.
13Confirm with <RUN/ENTER>.
The instrument switches to the measuring mode.
ba75691e0101/2007
13
CommissioningOxi 1970i
3.4Sensor quiver
To store the sensors during field operation and to keep the sensor element moist, the quiver tip contains a sponge rubber insert that can be
moistened with deionized water.
Note
For further details on proper storage, refer to the operating manual of
the sensor.
Moistening the
quiver insert
1Press the quiver out of the holder from the back side of the in-
strument and pull it out completely.
2Pull off the quiver tip and moisten the sponge rubber with
deionized water.
14
ba75691e0101/2007
Oxi 1970iOperation
4Operation
4.1Switching on the measuring instrument
1Connect the D. O. sensor to the measuring instrument.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
The relative slope for the sensor type that was just connected
subsequently appears for approx. one second.
The measuring instrument then automatically switches to the
measuring mode that was last selected. The display shows the
relevant measured value.
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour. The
energy saving feature is not active when the AutoStore function is active.
The energy saving feature is also not active
z if the power is supplied by the plug-in power supply,
z if the communication cable and a PC with a running communication
program are connected,
z if the recorder cable is connected,
z if the printer cable is connected (for external printers).
ba75691e0101/2007
15
OperationOxi 1970i
4.2Measuring
4.2.1General information
You can measure the following variables:
z D. O. concentration
z D. O. saturation
The measuring instrument is supplied with the following functions:
z AutoRange (automatic switchover of the measurement range). If a
measuring range is exceeded, AutoRange causes the measuring instrument to change automatically to the next higher measuring
range and back again. Therefore, the instrument always measures
in the measuring range with the highest possible resolution. The
function can be switched off.
z The AutoRead function (drift control) for checking the stability of the
measurement signal. This ensures the reproducibility of the measuring signal. For details of how to switch the AutoRead function on/off,
see page 19.
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect the D. O. sensor to the measuring instrument (see
section 1.3 S
OCKETFIELD).
2Calibrate or check the measuring instrument with the D. O.
sensor. How to calibrate is described in section 4.3.
3Select the measuring mode with <M>.
Note
Incorrect calibration of D. O. sensors will result in incorrect measured
values. Calibrate at regular intervals.
Note
The measuring instrument automatically recognizes the type of the D.
O. sensor (CellOx 325 or DurOx 325).
Temperature sensorThe D. O. sensor has an integrated temperature sensor that always
measures the current temperature of the test sample.
16
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
ba75691e0101/2007
Oxi 1970iOperation
4.2.2Measuring the D. O. concentration
Note
When measuring the concentration of test samples with a salt content
of more than 1 g/l, a salinity correction is required. For this, you have to
measure and input the salinity of the measured medium first. This is described in section 4.3.3 on page 27.
Switching the salt content correction on or off, see below.
To measure the D. O. concentration with and without salt content correction, proceed as follows:
1Perform the preparatory activities according to section 4.2.1.
2Immerse the D. O. sensor in the test sample.
3Press the <M> key repeatedly until the D. O. concentration in
mg/l appears on the display.
Switching on/off the salt
content correction
m
m
%
1/
°
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
pH/mV
cm
K
/
REL2
Sal
TP
RCL
Sal
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
Proceed as follows to switch on the salt content correction:
1In the D. O. concentration measuring mode, switch on the salt
pH
7
Lin
8
Oxi
92
1 7 6
Auto
Store
Cal
TEC
ARng
1
Baud
No.
Ident
Tref25Tref20
content correction with <▲ > . The SAL display indicator appears on the display. The value that was entered is taken into
consideration during the measurement (see section 4.3.3).
S
O
Sal
Time
Day.Month
Year
LoBatnLF
1
Baud
No.
Ident
Tref25Tref20
Lin
5
8
1 7 6
Oxi
42
Auto
Store
Cal
TEC
ARng
ba75691e0101/2007
2Switch the salt content correction off with <▼ >. The SAL dis-
play indicator is no longer displayed.
17
OperationOxi 1970i
4.2.3Measuring the D. O. saturation
You can measure the D. O. saturation as follows:
1Perform the preparatory activities according to section 4.2.1
2Immerse the D. O. sensor in the test sample.
3Press the <M> key repeatedly until the saturation in % appears
on the display.
/mV
%
m
1/
°
S/
C
pH
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
8
Baud
No.
Ident
8
Tref25Tref20
Lin
Oxi
92
1 7 6
Auto
Store
Cal
TEC
ARng
4.2.4Automatic selection of the measuring range, AutoRange
SensorsYou can use the sensors listed in section 1.3 on page 8. The sensor
type is automatically recognized and the resolution is adjusted if necessary (see chapter 7 T
AutoRangeTwo measuring ranges are available for each measuring mode. If a
measuring range is exceeded, AutoRange causes the measuring instrument to change automatically to the next higher measuring range
and back again. Therefore, the instrument always measures in the
measuring range with the highest possible resolution.
To switch the automatic measurement range selection on/off, see
section 4.6 C
on, ARng appears on the display.
pH
S
ONFIGURATION. When the AutoRange function is switched
O
Sal
Time
Day.Month
Year
LoBatnLF
1
Baud
No.
Ident
Tref25Tref20
7
8
Lin
Oxi
08
1 7 6
Auto
Cal
ECHNICALDATA).
mg/l
%
m
S/
1/
cm
%
°
ARng
C
REL1
AR
Store
TEC
pH/mV
cm
K
/
REL2
Sal
TP
RCL
18
ba75691e0101/2007
Oxi 1970iOperation
4.2.5AutoRead AR (drift control) and hold function
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
CriteriaWith identical measurement conditions, the following criteria are valid
for the AutoRead function:
Measuring modeReproducibilityResponse time
D. O. concentration
D. O. saturation
index
Use the AutoRead function and hold function like this with the sensors
z Oxygen depth armature
z CellOx 325
z DurOx 325
Note
The special features of the self-stirring D. O. sensor StirrOx G can be
found in section 4.2.6 on page 20.
1Select the measuring mode with <M>.
2Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
better than 0.05 mg/l> 10 seconds
better than 0.6 %> 10 seconds
ba75691e0101/2007
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
Sal
TP
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
4If necessary, start the next AutoRead measurement with
7
Lin
8
Oxi
08
1 7 6
Auto
Cal
1
Baud
No.
Ident
Tref25Tref20
<RUN/ENTER>.
Store
TEC
ARng
19
OperationOxi 1970i
5To terminate the AutoRead function: Press the <AR> key.
4.2.6Sensor with stirrer (StirrOx G)
ConnectionConnect the 2-pole stirrer plug to the stirrer output of the measuring in-
strument (designation REL1).
Note
The automatic stirring function of the self-stirring D. O. sensor
StirrOx G can only be used when the plug-in power supply is connected.
Stirrer functionThe D. O. sensor, StirrOx G, has a temperature sensor and stirrer inte-
grated. Using the key on the sensor you switch the stirrer on.
Measuring with
AutoRead
You can switch on the stirrer and start an AutoRead measurement at
the same time. To do so, proceed as follows:
1Select the measuring mode with <M>.
2Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).
3Submerse the sensor in the test sample and press the stirrer
button. The stirrer is startet. At the same time, the AutoRead
measurement starts. AR flashes until a stable measured value
is achieved. Subsequently, the stirrer switches itself off.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
7
Baud
No.
Ident
8
Tref25Tref20
Lin
Oxi
08
1 7 6
Auto
Store
Cal
TEC
ARng
20
ba75691e0101/2007
Loading...
+ 45 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.