Xylem handylab pH 11 User Manual

Operating manual
AR
O
A
L
P
H
6
4
2
9
9
2
6
°
C
TP
M
handylab pH 11
ba75366d04 01/2011
pH Meter
handylab LF 12
Accuracy when
going to press
Warranty We guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our instruments are the result of continuous development. This may result in differences between this operating manual and your instrument. Al­so, we cannot guarantee that there are absolutely no errors in this man­ual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
purchase. The instrument warranty covers manufacturing faults that are discov­ered within the warranty period. The warranty does not cover compo­nents that are replaced during maintenance work, e.g. batteries.
The warranty claim extends to restoring the instrument to readiness for use but not, however, to any further claim for damages. Improper han­dling or unauthorized opening of the instrument invalidates any warran­ty claim.
Copyright
© 2011, SI Analytics GmbH
Reproduction in whole - or even in part - is prohibited without the express written permission of SI Analytics GmbH, Mainz. Printed in Germany.
2
ba75366d04 01/2011
handylab pH 11 List of contents
handylab pH 11 - List of contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Jack field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Switching on the measuring instrument . . . . . . . . . . . . . . . . . 15
4.2 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2 Measuring the pH value . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.3 Measuring the ORP voltage . . . . . . . . . . . . . . . . . . . . 18
4.3 Calibrating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.3.1 AutoCal TEC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.3.2 AutoCal DIN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.3.3 ConCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.4 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.5 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
ba75366e04 01/2011
5 Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . . . . . 35
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
3
List of contents handylab pH 11
4
ba75366e04 01/2011
handylab pH 11 Overview
AR
O
C
A
L
PH
6
4
2
9
9
2
6
°
C
TP
1
2
3
M

1 Overview

The compact precision handheld meter handylab pH 11 enables you to carry out pH measurements rapidly and reliably. The handylab pH 11 provides the maximum degree of operating comfort, re­liability and measuring certainty for all applications.
The proven calibration procedures and the special AutoRead function sup­port you in your work with the handylab pH 11.
1 Keypad
2 Sample display
3 Jack field
ba75366e04 01/2011
5
Overview handylab pH 11
CAL
RUN/ENTER
AR
M
M
CAL
AR
RUN/ENTER

1.1 Keypad

Key functions
Select the measuring mode <M>: –pH value
–ORP voltage
Switch measuring instrument on/off
<ON/OFF>
Calibration of the current measured variable
<CAL>
Activate/deactivate the AutoRead function
<AR>
Select the measuring mode Increase values, scroll
<>
Select the measuring mode Decrease values, scroll
<>
Confirm entries, start AutoRead
<RUN/ENTER>
6
ba75366e04 01/2011
handylab pH 11 Overview
%
°F
S
ConCal
Auto
Store
Time
Baud
Day.Month No.
Ident
Year
RCL
TempError
mg/L
% ppm
mol/L
ISE
8
8.
8
88
8
8
mV/pH
°C
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCal TEC
TP
1
UpH
Status display
Meas. value display
Function and temperature display
Sensor symbol
R
e
f
12
3

1.2 Display

1.3 Jack field

The meter has either a DIN socket or a BNC socket. The figure shows a meter with DIN socket (1).
ba75366e04 01/2011
1 pH electrode or single-rod measuring electrode
2 Reference electrode
3 Temperature probe
Caution
Only connect electrodes to the measuring instrument that cannot return any voltages or currents that are not allowed (> SELV and > current circuit with current limiting). Nearly all electrodes - especially SI Analytics electrodes - fulfill these condi­tions.
7
Overview handylab pH 11

1.4 Technical Data

Dimensions
and weight
Mechanical
structure
Electrical safety
Test certificates
Ambient
conditions
Length [mm] 172
Width [mm] 80
Height [mm] 37
Weight [kg] Approx. 0.3
Type of protection IP 66
Protective class III
cETLus, CE
Storage - 25 °C ... + 65 °C
Operation -10 °C ... + 55 °C
Relative atmospheric
< 90 % of annual average
humidity
pH/ORP
measuring ranges
Precision of
pH/ORP
(± 1 digit)
Measuring range Resolution
pH - 2.000 ... + 19.999
- 2.00 ... + 19.99
U [mV] - 999.9 ... + 999.9
- 1999 ... + 1999
0.001
0.01
0.1 1
T [°C] - 5.0 ... + 105.0 0.1
pH (± 2 pH units from the
± 0.005 at + 15 °C ... + 35 °C ± 0.01
calibration point)
U [mV] ± 0.3 at + 15 °C ... + 35 °C
± 1
T [°C] NTC 30:
Accuracy ± 0.1
PT 1000: Accuracy Operating temperature ± 0.5 0 °C ... 15 °C ± 0.1 15 °C ... 35 °C ± 1 35 °C ... 55 °C
8
ba75366e04 01/2011
handylab pH 11 Overview
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Correction
function
Power supply
Applicable
guidelines and
norms
Temperature input
Manually [°C] - 20 ... + 130
Batteries 4 x 1.5 V alkali-manganese batteries, Type
AA
Operational life Approx. 5000 operating hours
EMC EC guideline 2004/108/EC
EN 61326-1 EN 61000-3-2 EN 61000-3-3 FCC Class A
Instrument safety EC guideline 2006/95/EC
EN 61010-1 ANSI/UL 61010-1 CAN/CSA-C22.2 No. 61010-1
Climatic class VDI/VDE 3540
Type of protection EN 60529
ba75366e04 01/2011
9
Overview handylab pH 11
10
ba75366e04 01/2011
handylab pH 11 Safety

2Safety

This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the pH meter. Conse­quently, all responsible personnel must read this operating manual before working with the measuring system. The operating manual must always be available within the vicinity of the measuring system.
Target group The measuring instrument was developed for work in the field and in the lab-
oratory. We assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to take when handling chemicals.
Safety
instructions
Further notes
The individual chapters of this operating manual use safety instructions such as the label shown below to indicate various hazards or dangers:
Caution
indicates instructions that must be followed precisely in order to avoid the possibility of slight injuries or damage to the instrument or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.

2.1 Authorized use

The authorized use of the measuring instrument consists exclusively of the pH and ORP measurement in the field and laboratory. The technical specifications as given in section 1.4 T observed. Only the operation and running of the measuring instrument ac­cording to the instructions given in this operating manual is authorized. Any other use is considered to be unauthorized.
ECHNICAL DATA must be
ba75366e04 01/2011
11
Safety handylab pH 11

2.2 General safety instructions

This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see section 1.4 T D
ATA).
It left the factory in a safe and secure technical condition.
ECHNICAL
Function and oper-
ating safety
The smooth functioning and operational safety of the measuring instrument can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are followed during oper­ation.
The smooth functioning and operational safety of the measuring instrument can only be guaranteed under the environmental conditions that are specified in section 1.4 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm environ­ment, the formation of condensate can lead to the faulty functioning of the in­strument. In this event, wait until the temperature of the instrument reaches room temperature before putting the instrument back into operation.
Safe operation If safe operation is no longer possible, the instrument must be taken out of
service and secured against inadvertent operation! Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of time
z is visibly damaged
Obligations of the
purchaser
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
12
ba75366e04 01/2011
handylab pH 11 Commissioning

3 Commissioning

3.1 Scope of delivery

z Handheld meter, handylab pH 11
z Operating manual and short operating manual
z 4 batteries, 1.5 V Mignon type AA (in the instrument)
ba75366e04 01/2011
13
Commissioning handylab pH 11
14
ba75366e04 01/2011
handylab pH 11 Operation

4 Operation

4.1 Switching on the measuring instrument

1 Press the <ON/OFF> key.
The display test appears briefly on the display. Subsequently, the slope and asymmetry used appear for ap­prox. one second one after the other. In addition, the calibration procedure of the last calibration is shown (AutoCal TEC or AutoCal DIN or no display in the delivery state or following ini­tialization). The measuring instrument then automatically switches to the measuring mode that was last selected.
Note
The measuring instrument has an energy saving feature to avoid unneces­sary battery depletion. The energy saving feature switches the measuring in­strument off if no key has been pressed for an hour. The energy saving feature is not active when the AutoStore function is active.
ba75366e04 01/2011
15
Operation handylab pH 11

4.2 Measuring

4.2.1 General information

You can measure the following variables:
z pH value
z ORP voltage
Preparatory
activities
Temperature
sensor
Perform the following preparatory activities when you want to measure:
1 Connect the electrode to the measuring instrument.
2 Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a temperature sensor.
3 In conjunction with the electrode, check or calibrate the mea-
suring instument. How to calibrate is described in section 4.3.
4 Select the measuring mode with <M>.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values. Calibrate regularly before measuring.
You can measure with or without a temperature sensor. If a temperature sen­sor is connected, it is indicated on the display by TP.
Note
The measuring instrument automatically recognizes the type of the tempera­ture sensor used. Therefore, you can connect temperature sensors of the NTC30 or Pt1000 type.
16
The temperature measurement is absolutely essential for a reproducible pH measurement. If the measurement is made without a temperature sensor, proceed as follows:
1 Measure the current temperature of the test sample using a
thermometer.
2 Enter the temperature on the instrument: Set the temperature
value of the test sample using <> <>.
ba75366e04 01/2011
handylab pH 11 Operation
mV/pH
mol/L
8
Auto
Store
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCal TEC
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
mg/L
% ppm
ISE
8
4
2
29
9
6
°C
TP
pH

4.2.2 Measuring the pH value

1 Perform the preparatory activities according to section 4.2.1.
2 Immerse the pH electrode in the test sample.
3 Press the <M> key until pH appears on the status display. The
pH value appears on the display.
Setting the
resolution
AutoRead
(drift control)
The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the <M> key while pressing the <RUN/ENTER> key.
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values.
For identical measurement conditions, the following criteria apply:
Reproducibility Response time
Better than 0.02 > 30 seconds
1 Call up the pH measuring mode with <M> and <> <> .
2 Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3 Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This measured value is transmitted to the interface.
ba75366e04 01/2011
4 If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5 To terminate the AutoRead function: Press the <AR> key.
17
Operation handylab pH 11
/pH
ConCal
Auto
Store
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCal TEC
8
pH
1
%
°F
S
Time
Baud
Day.Month No.
Ident
Year
RCL
TempError
mg/L
% ppm
mol/L
ISE
3
1
2
2
0
mV
°C
U
TP
1
7
Note
The current AutoRead measurement can be terminated at any time (accept­ing the current value) by pressing <RUN/ENTER>.

4.2.3 Measuring the ORP voltage

In conjunction with an ORP electrode, e.g. BlueLine 31Rx, the measuring in­strument can measure the ORP voltage (U) of a solution.
1 Perform the preparatory activities according to section 4.2.1.
2 Immerse the ORP electrode in the test sample.
3 Press the <M> key until U appears on the status display.
The ORP voltage (mV) of the test sample appears on the dis­play.
4 Wait for a stable measured value.
18
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
ba75366e04 01/2011
handylab pH 11 Operation

4.3 Calibrating

Why calibrate? pH electrodes age. This changes the asymmetry (zero point) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the asymmetry and slope of the electrode and stores them in the measuring instrument. Thus, you should calibrate at regular intervals.
When to calibrate? z After connecting another electrode
z after a voltage interruption, e. g. after changing the batteries
You can select between 3 calibration procedures:
AutoCal TEC is specially adapted to the SI Analytics technical buffer solutions as a fully au-
tomatic two-point or three-point calibration. The buffer solutions are auto- matically recognized by the measuring instrument. Depending on the instrument setting (see section 4.4 C the relevant buffer nominal value or the current electrode voltage in mV. The calibration can be terminated after the first buffer solution. This corresponds to a single-point calibration. For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
ONFIGURATION), the instrument displays
AutoCal DIN is specially adapted to the permanently programmed buffer solutions in ac-
cordance with DIN19266 as a fully automatic two-point or three-point cali- bration. The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see section
4.4 C
ONFIGURATION), the instrument displays the relevant buffer nominal val-
ue or the current electrode voltage in mV. The calibration can be terminated after the first buffer solution. This corresponds to a single-point calibration. For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and de­termines the asymmetry of the electrode.
ConCal This function is a conventional two-point calibration using two buffer solu-
tions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point calibra-
tion using any buffer solution and is used as a high-speed method.
AutoRead When calibrating with AutoCal TEC and AutoCal DIN, the AutoRead function
is automatically activated. The current AutoRead measurement can be terminated at any time (accept­ing the current value) by pressing <RUN/ENTER>.
ba75366e04 01/2011
19
Operation handylab pH 11
Displaying
calibration data
Calibration
evaluation
Each time the instrument is switched on, the calibration data are shown on the display for a short time (see section 4.1 S
STRUMENT). In order to view the calibration data, switch the measuring instru-
WITCHING ON THE MEASURING IN-
ment off and switch it on again.
After the calibration, the measuring instrument automatically evaluates the current status of the electrode. The asymmetry and slope are evaluated sep­arately. The worst evaluation appears on the display.
Display Asymmetry
[mV]
Slope [mV/pH]
-15 ... +15 -60.5 ... -58
-20 ... +20 -58 ... -57
-25 ... +25 -61 ... -60.5 or
-57 ... -56
Preparatory
activities
-30 ... +30 -62 ... -61 or
-56 ... -50
Clean the electrode according to the electrode operating manual
E3
Eliminate the error according to chapter 6 W
HAT TO DO IF...
< -30 or > 30
< -62 or > -50
1 Switch on the measuring instrument with <ON/OFF>.
2 Connect the pH electrode to the measuring instrument.
3 Keep the buffer solutions ready.
4 Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without temper­ature sensor.
20
ba75366e04 01/2011
handylab pH 11 Operation
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1C
t
°C
TP
pH
AutoCal TEC

4.3.1 AutoCal TEC

For this procedure, use any two or three technical buffer solutions in ascend­ing or descending order (pH values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the SI Analytics technical buffer solution pH 10.01. Other buffer solutions can lead to an erroneous calibra­tion. The correct buffer solutions are given in the SI Analytics catalog or in the Internet.
Note
The buffer solutions are automatically recognized by the measuring instru­ment. Depending on the instrument setting (see section 4.4 C the instrument displays the relevant buffer nominal value or the current elec­trode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
ONFIGURATION),
1 Press the <CAL> key repeatedly until the Ct1 display indicator
and the function display AutoCal TEC appears. The sensor symbol displays the evaluation of the last calibration (or no sensor symbol in the delivery state or after the measurement parameter has been reset).
2 If necessary, enter the temperature of the first buffer solution
with <> <>.
3 Immerse the pH electrode in the first buffer solution.
4 Press the <RUN/ENTER> key.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. Example:
ba75366e04 01/2011
21
Operation handylab pH 11
cm
3
7
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
AutoCal DIN
Tref25
Auto
AutoCal DIN
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OO UpHUpH SalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.Month No.
Ident
Year
Time
Baud
Day.Month No.
Ident
Year
0
03
7
pHpH
C
°
C
°
AutoCal TEC
AutoCal TEC
ARAR
mV
88
44
22
TP
TP
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
2C
t
°C
TP
pH
AutoCal TEC
5 When the measured value is stable, Ct2 appears.
Note
At this point, the AutoCal TEC calibration can be terminated with <M>. This corresponds to a single-point calibration. For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
6 To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
7 If necessary, enter the temperature of the second buffer solu-
tion with <> <>.
8 Immerse the pH electrode in the second buffer solution.
9 Press the <RUN/ENTER> key.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. Example:
22
ba75366e04 01/2011
handylab pH 11 Operation
1
TDS
S
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
AutoCal DIN
Tref25
Auto
AutoCal DIN
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OO UpHUpH SalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDS
1
S
mbarmbar
Time
Baud
Day.Month No.
Ident
Year
Time
Baud
Day.Month No.
Ident
Year
88
44
08
14
4
pHpH
C
°
C
°
AutoCal TEC
AutoCal TEC
ARAR
mV
1
22
TP
TP
O
L
S
RCL
Sal
cm
AutoCal DIN
Day.Month
S
Tref20
Cal
Ident
No.
Year
Time
Baud
1
TDS
Sal
Store
AR
Tref25
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
Lin
OU
nLF
Auto
LoBat
ARng
°
F
9
4
pH
5
mV/pH
AutoCal TEC
YA
S
0
RCL
Sal
cm
AutoCal DIN
Time
Day.Month
Sal
Store
AR
8
No.
Ident
Tref20
Tref25
Year
Baud
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
TDS
6
Lin
Cal
1
S
OU
nLF
Auto
LoBat
ARng
°
F
2
pH
mV
AutoCal TEC
10 When the measured value is stable, AR disappears. The in-
strument displays the value of the slope (mV/pH) and the eval­uation of the calibration in calibration range I.
ba75366e04 01/2011
11 Press the <RUN/ENTER> key. The instrument displays the
value of the asymmetry in calibration range I.
Note
At this point, the AutoCal TEC calibration can be terminated with <M>. This corresponds to a two-point calibration. The instrument uses the slope and asymmetry of the calibration range I.
23
Operation handylab pH 11
Three-point
calibration
12 Press the <RUN/ENTER> key
to continue the three-point calibration. Ct3 appears on the display.
13 Thoroughly rinse the electrode with distilled water.
14 If necessary, enter the temperature of the third buffer solution
with <> <>.
15 Immerse the pH electrode in the third buffer solution.
16 Press the <RUN/ENTER> key.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. When the measured value is stable, AR disappears. The instrument displays the value of the slope (mV/pH) and the evaluation of the calibration in calibration range II.
17 Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibra­tion range II. The three-point calibration is completed.
18 To return to the measuring mode: Press the <M> key.
Note
You can prematurely terminate the three-point calibration by pressing <M>. The values of the two-point calibration for slope and asymmetry will then re­main stored.
24
ba75366e04 01/2011
handylab pH 11 Operation
AutoCal TEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1C
d
°C
TP
pH
AutoCal DIN

4.3.2 AutoCal DIN

For this procedure, use two or three different standard buffer solutions ac­cording to DIN 19266 in ascending or descending order (type A, C, D or F with pH values at 25 °C of: 1.679 / 4.006 / 6.865 / 9.180).
Note
The buffer solutions are automatically recognized by the measuring instru­ment. Depending on the instrument setting (see section 4.4 C the instrument displays the relevant buffer nominal value or the current elec­trode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
1 Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol dis­plays the evaluation of the last calibration (or no sensor symbol in the delivery state or after the measurement parameter has been reset).
ONFIGURATION),
ba75366e04 01/2011
2 If necessary, enter the temperature of the buffer solution with
<> <>.
3 Immerse the pH electrode in the first buffer solution.
4 Press the <RUN/ENTER> key.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. Example:
25
Operation handylab pH 11
AutoCal TEC
AutoCal TEC
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
Tref25
Auto
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OO UpHUpH SalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.Month No.
Ident
Year
Time
Baud
Day.Month No.
Ident
Year
0
17
4
pHpH
C
°
C
°
ARAR
mV
88
44
22
TP
TP
7
1
AutoCal DIN
AutoCal DIN
AutoCal TEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
2C
d
°C
TP
pH
AutoCal DIN
5 When the measured value is stable, Cd2 appears.
Note
At this point, the AutoCal DIN calibration can be terminated with <M>. This corresponds to a single-point calibration. For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
26
6 To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
7 If necessary, enter the temperature of the second buffer solu-
tion with <> <>.
8 Immerse the pH electrode in the second buffer solution.
9 Press the <RUN/ENTER> key.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. Example:
ba75366e04 01/2011
handylab pH 11 Operation
AutoCal TEC
AutoCal TEC
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
Tref25
Auto
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OO UpHUpH SalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.Month No.
Ident
Year
Time
Baud
Day.Month No.
Ident
Year
1
89
9
pHpH
C
°
C
°
ARAR
mV
88
44
22
TP
TP
2
1
AutoCal DIN
AutoCal DIN
AutoCal TEC
YA
S
0
RCL
Sal
cm
Time
Day.Month
Sal
Store
AR
8
No.
Ident
Tref20
Tref25
Year
Baud
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
TDS
6
Lin
Cal
1
S
OU
nLF
Auto
LoBat
ARng
°
F
2
pH
mV
AutoCal DIN
10 When the measured value is stable, AR disappears. The in-
strument displays the value of the slope (mV/pH) and the eval­uation of the calibration in calibration range I.
ba75366e04 01/2011
11 Press the <RUN/ENTER> key. The instrument displays the
value of the asymmetry in calibration range I.
Note
At this point, the AutoCal DIN calibration can be terminated with <M>. This corresponds to a two-point calibration. The instrument uses the slope and asymmetry of the calibration range I.
27
Operation handylab pH 11
Three-point
calibration
12 Press the <RUN/ENTER> key
to continue the three-point calibration. Cd3 appears on the display.
13 Thoroughly rinse the electrode with distilled water.
14 If necessary, enter the temperature of the third buffer solution
with <> <>.
15 Immerse the pH electrode in the third buffer solution.
16 Press the <RUN/ENTER> key.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. When the measured value is stable, AR disappears. The instrument displays the value of the slope (mV/pH) and the evaluation of the calibration in calibration range II.
17 Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibra­tion range II. The three-point calibration is completed.
18 To return to the measuring mode: Press the <M> key.
Note
You can prematurely terminate the three-point calibration by pressing <M>. The values of the two-point calibration for slope and asymmetry will then re­main stored.
28
ba75366e04 01/2011
handylab pH 11 Operation

4.3.3 ConCal

Single-point
calibration
Two-point
calibration
Use any buffer solution for this rapid method. The calibration will be the more exact the nearer the pH value of the buffer solution is to that of the test sample.
For this procedure, use two buffer solutions:
z pH 7.0 ± 0.5
z any other buffer solution
Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature sensor.
1 Press the <CAL> key repeatedly until the display ASY and the
function display ConCal appears. The sensor symbol displays the evaluation of the last calibration (or no sensor symbol in the delivery state or after the measurement parameter has been reset).
ba75366e04 01/2011
2 If necessary, enter the temperature of the first buffer solution
with <> <>.
3 Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).
4 Press the <RUN/ENTER> key.
The measured pH value appears on the display.
5 Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <> <>.
6 Press the <RUN/ENTER> key. The value of the asymmetry
(mV) and the sensor symbol appear on the display.
7 Press the <RUN/ENTER> key. SLO appears on the display.
Note At this point, the ConCal calibration can be broken off with <M>. This corre-
29
Operation handylab pH 11
AutoCal DIN
AutoCal TEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
OS
L
°C
TP
pH
ConCal
sponds to a single-point calibration. For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and the specified asymmetry of the elec­trode.
8 To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
9 If necessary, enter the temperature of the second buffer solu-
tion with <> <>.
10 Immerse the pH electrode in the second buffer solution.
11 Press the <RUN/ENTER> key. The second pH value appears
on the display.
12 Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <> <>.
13 When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display. The probe symbol shows the evaluation of the current calibra­tion.
30
14 Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
ba75366e04 01/2011
handylab pH 11 Operation
15 To return to the measuring mode: Press the <M> key.
ba75366e04 01/2011
31
Operation handylab pH 11

4.4 Configuration

You can adapt the measuring instrument to your individual requirements. To do this, the following parameters can be changed (the status on delivery is marked in bold):
Display during the pH calibration
pH resolution 0.01 or 0.001
Note
You can leave the configuration menu at any time with <M> . The parameters that have already been changed are stored.
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
Buffer nominal value, or current electrode voltage
Display during the
pH calibration
3 Press the <ON/OFF> key.
The display test appears briefly on the display. CAL diSP ap­pears on the display.
4 Select the required display during the pH calibration with
<> <>.
mV: Display of the current electrode voltage /pH: Display of the buffer nominal value.
5 Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH measuring mode.
32
ba75366e04 01/2011
handylab pH 11 Operation
Setting the
resolution
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <M> key.
The measured values are displayed with a high resolution, e.g. pH = 4.012.
3 Press the <RUN/ENTER> and <M> keys once more.
The measured values are displayed with a low resolution, e.g. pH = 4.01.
ba75366e04 01/2011
33
Operation handylab pH 11
Tref25
Tref20
AutoCal TEC
RCL
LoBat
Cal
ARng
AR
Lin
nLF
Store
Sal
°
°
TP
cm
1/
F
C
%
/
K
8
8
1
S/
cm
m
M
TDS
Sal
S
Auto
mbar
%
mg/l
O
mV/pH
cm
AutoCal DIN
Time
Baud
Day.Month No.
Ident
Year
o
n
n
I
I
pH

4.5 Reset

You can reset (initialize) the measurement parameters and the configuration parameters separately from one another.
Measurement
parameters
Resetting the
measuring
parameters
The following measured parameters (pH InI) are reset to the default condi- tion:
Measuring mode pH
Asymmetry 0 mV
Slope -59.16 mV/pH
Calibration procedures AutoCal TEC
Temperature, manual 25 °C
Resolution (pH display) 0.01 (low resolution)
Display during the pH calibration
Note
The calibration data gets lost when the measuring parameters are reset. Recalibrate after performing a reset.
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <CAL> key.
Buffer nominal value
34
3 Using <> <>, switch between no and YES.
YES: Resetting the measuring parameters no: Retaining settings.
4 Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH measuring mode.
ba75366e04 01/2011
handylab pH 11 Maintenance, cleaning, disposal
1
2

5 Maintenance, cleaning, disposal

5.1 Maintenance

The measuring instrument is almost maintenance-free. The only maintenance task is replacing the batteries. LoBat indicates that the batteries should be changed. The batteries are then largely depleted.
Replacing the
batteries
1 Open the housing after the instrument has been switched off:
– Undo the four screws on the underside of the instrument
– Pull down the lower cover (1).
ba75366e04 01/2011
2 If necessary, take the four depleted batteries (2) out of the bat-
tery compartment.
3 Place four new batteries (type Mignon AA) in the battery com-
partment.
4 Close the lower cover (1).
Caution
Make sure that the poles of the batteries are the right way round. The
± signs on the batteries must correspond to the ± signs in the battery
compartment. Only use leakproof alkaline manganese batteries.
35
Maintenance, cleaning, disposal handylab pH 11
Note
For the maintenance of the electrodes, follow the corresponding operating manual.

5.2 Cleaning

Occasionally wipe the outside of the measuring instrument with a damp, lint­free cloth. Disinfect the housing with isopropanol as required.
Caution
The housing is made of a synthetic material (ABS). Thus, avoid contact with acetone and similar detergents that contain solvents. Remove any splashes immediately.

5.3 Disposal

Packing This measuring instrument is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the instrument against damage during transport.
Batteries This note refers to the battery regulation that applies in the Federal Republic
of Germany. We would ask end-consumers in other countries to follow their local statutory provisions.
Note
This instrument contains batteries. Batteries that have been removed must only be disposed of at the recycling facility set up for this purpose or via the retail outlet. It is illegal to dispose of them in household refuse.
Measuring
instrument
Dispose of the measuring instrument as electronic waste at an appropriate collection point. It is illegal to dispose of the instrument in household refuse.
36
ba75366e04 01/2011
handylab pH 11 What to do if...

6 What to do if...

Error message 0FL
Cause Remedy
pH electrode:
– Not connected – Connect electrode
Error message E3
– Air bubble in front of the
– Remove air bubble
diaphragm
– Air in the diaphragm – Extract air or
moisten diaphragm
– Cable broken – Replace electrode
– Gel electrolyte dried out – Replace electrode
Cause Remedy
Electrode
– Diaphragm contaminated – Clean diaphragm
– Membrane contaminated – Clean membrane
– Moisture in the plug – Dry plug
– Electrolyte out of date – Replenish electrolyte or
replace electrode
– Electrode worn out – Replace electrode
ba75366e04 01/2011
– Electrode broken – Replace electrode
Measuring instrument:
– Incorrect calibration procedure – Select correct procedure
– Incorrect solution temperature
– Set up correct temperature
(without temperature sensor)
– Socket damp – Dry socket
37
What to do if... handylab pH 11
Buffer solutions
– Incorrect buffer solutions – Change calibration procedure
– Buffer solutions too old – Use only once.
Note the shelf life
– Buffer solutions depleted – Change solutions
No stable mea-
Cause Remedy
sured value
pH electrode:
– Diaphragm contaminated – Clean diaphragm
– Membrane contaminated – Clean membrane
Test sample:
LoBat display
– pH value not stable – Measure with air excluded if
necessary
– Temperature not stable – Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low – Use suitable electrode
– Temperature too high – Use suitable electrode
– Organic liquids – Use suitable electrode
Cause Remedy
– Batteries almost empty – Replace batteries
(see section 5.1 MAINTENANCE)
38
ba75366e04 01/2011
handylab pH 11 What to do if...
Obviously incor-
rect
measured values
Instrument does not react to key-
stroke
Cause Remedy
pH electrode:
– pH electrode unsuitable – Use suitable electrode
– Temperature difference
between buffer and test
– Adjust temperature of buffers
or test samples
sample too high
– Measurement procedure not
– Follow special procedure
suitable
Cause Remedy
– Operating condition
undefined or EMC load unallowed
– Processor reset:
Press the <CAL> and <ON/OFF> keys at the same time and release them again. The software version is displayed.
Display to
Cause Remedy
– Time-out of the interface – Check the instrument that is
connected
ba75366e04 01/2011
39
What to do if... handylab pH 11
40
ba75366e04 01/2011
handylab pH 11 Lists

7Lists

This chapter provides additional information and orientation aids.
Abbreviations The list of abbreviations explains the indicators and the abbreviations that ap-
pear on the display and in the manual.
Specialist terms The glossary briefly explains the meaning of the specialist terms. However,
terms that should already be familiar to the target group are not described here.
Index The index will help you to find the topics that you are looking for.
ba75366e04 01/2011
41
Lists handylab pH 11
Abbreviations
AR AutoRead (drift control)
ARng Automatic range switching
Measuring instrument measures with highest resolution
ASY Asymmetry
AutoCal DIN Automatic pH calibration with buffer solutions prepared
according to DIN 19 266
AutoCal TEC Automatic pH calibration with SI Analytics technical buf-
fer solutions according to DIN 19267
°C Temperature unit, degrees Celsius
Cal Calibration
Cd... Display indicator during calibration for pH measure-
ments. Indicates the selection of the buffer data record for buffer solutions prepared according to DIN 19 266
ConCal Conventional single-point or two-point calibration for pH
measurements
Ct... Display indicator during calibration for pH measure-
ments. Indicates the selection of the buffer data records for SI Analytics technical buffer solutions
InI Initialization
Resets individual basic functions to the status they had on delivery
LoBat Batteries almost empty(Low Battery)
mV Voltage unit
mV/pH Unit of the electrode slope (internat. mV)
OFL Display range exceeded (Overflow)
pH pH value
S Slope (internat. k)
SELV Safety Extra Low Voltage
SLO Slope setting on calibration
42
TP Temperature measurement active (Temperature Probe)
U
ASY
Asymmetry
ba75366e04 01/2011
handylab pH 11 Lists
Glossary
Adjusting To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains with­in the tolerance.
Asymmetry Designation for the offset potential of a pH electrode. It is the measur-
able potential of a symmetrical electrode, the membrane of which is immersed in a solution with the pH of the nominal electrode zero point (electrodes: pH = 7).
AutoRange Name of the automatic selection of the measuring range.
AutoRead Name for a function to check the stability of the measured value.
Calibration Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Of­ten, this expression is also used when the measuring system is adjust­ed at the same time (see adjusting).
Diaphragm The junction is a porous body in the housing wall of reference elec-
trodes or electrolyte bridges. It forms the electrical contact between two solutions and makes electrolyte exchange more difficult. The ex­pression, junction, is also used for ground or junction-less transitions.
Electrode zero point The zero point of a pH electrode is the pH value at which the electro-
motive force of the pH electrode at a specified temperature is zero. Normally, this is at 25 °C.
Electromotive force of
an electrode
The electromotive force U of the electrode is the measurable electro­motive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the electrode. Its dependency on the pH results in the electrode function which is characterized by the parameters, slope and zero point.
Measured parameter The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
Measured value The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Measuring system The measuring system comprises all the devices used for measuring,
e. g. measuring instrument and probe. In addition, there is the cable and possibly an amplifier, terminal strip and armature.
Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
MultiCal
ba75366e04 01/2011
vent.
®
Name stating that a measuring instrument provides several calibration procedures.
43
Lists handylab pH 11
Offset potential The measurable potential of a symmetrical electrode, the membrane
of which is immersed in a solution with the pH of the nominal electrode zero point. The asymmetry is part of the offset potential.
ORP voltage The ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface (e. g. a gold or platinum surface).
pH value The pH is a measure of the acidic or basic effect of an aqueous solu-
tion. It corresponds to the negative decadic logarithm of the molal hy­drogen ions activity divided by the unit of the molality. The practical pH value is the value of a pH measurement.
Potentiometry Name of a measuring technique. The signal (depending on the mea-
sured parameter) of the electrode is the electrical potential. The elec­trical current remains constant.
Reset Restoring the original condition of all settings of a measuring system.
Resolution Smallest difference between two measured values that can be dis-
played by a measuring instrument.
Slope The slope of a linear calibration function.
Standard solution The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Test sample Designation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample and original sample are identical if the test sample was not processed.
44
ba75366e04 01/2011
handylab pH 11 Index
Index
A
Asymmetry 19 Authorized use 11 AutoCal DIN 19, 25 AutoCal TEC 19, 21 AutoRead 17
C
Calibration evaluation 20 Calibration procedures 19 ConCal 19, 29
D
Default condition 34 Display 7 Displaying calibration data
20
Drift control 17
E
Energy saving feature 15
S
Safety 11 Safety precautions 11 Scope of delivery 13 Setting the resolution 33 Setting the resolution (pH) 17 Single-point calibration 19 Single-point calibration (pH)
22, 23, 26, 27, 30
Slope 19
T
Temperature sensor 16 Three-point calibration 19
AutoCal DIN 28 AutoCal TEC 24
Two-point calibration 19
ConCal 29
Two-point calibration (pH)
19, 22, 23, 26, 27, 30
I
Initialization 34
K
Keys 6
L
LoBat 38
M
Measuring ranges
pH/ORP
8
O
Operating safety 12 ORP electrode 18 ORP voltage 18
R
Replacing the batteries 35 Reset 34
45
Index handylab pH 11
46
SI Analytics GmbH Postfach 2443 D-55014 Mainz Hattenbergstr. 10 D-55122 Mainz
Telefon +49 (0) 61 31/66 5111 Telefax +49 (0) 61 31/66 5001 Email: support@
si-analytics.com
Internet: www.si-analytics.com
Loading...