WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept
any legal claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are discovered within the warranty period. The warranty does not cover components that are replaced during maintenance work, e.g. batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper handling or unauthorized opening of the instrument invalidates any warranty claim.
The compact precision handheld meter handylab pH 11 enables you to carry
out pH measurements rapidly and reliably.
The handylab pH 11 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven calibration procedures and the special AutoRead function support you in your work with the handylab pH 11.
1Keypad
2Sample display
3Jack field
ba75366e0401/2011
5
Overviewhandylab pH 11
CAL
RUN/ENTER
AR
M
M
CAL
AR
RUN/ENTER
1.1Keypad
Key functions
Select the measuring mode <M>:
–pH value
–ORP voltage
Switch measuring instrument on/off
<ON/OFF>
Calibration of the current measured variable
<CAL>
Activate/deactivate the AutoRead function
<AR>
Select the measuring mode
Increase values, scroll
<▲>
Select the measuring mode
Decrease values, scroll
<▼ >
Confirm entries, start AutoRead
<RUN/ENTER>
6
ba75366e0401/2011
handylab pH 11Overview
%
°F
S
ConCal
Auto
Store
Time
Baud
Day.MonthNo.
Ident
Year
RCL
TempError
mg/L
% ppm
mol/L
ISE
8
8.
8
88
8
8
mV/pH
°C
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCal TEC
TP
1
UpH
Status display
Meas. value display
Function and
temperature display
Sensor symbol
R
e
f
12
3
1.2Display
1.3Jack field
The meter has either a DIN socket or a BNC socket.
The figure shows a meter with DIN socket (1).
ba75366e0401/2011
1pH electrode or single-rod measuring electrode
2Reference electrode
3Temperature probe
Caution
Only connect electrodes to the measuring instrument that cannot return any
voltages or currents that are not allowed
(> SELV and > current circuit with current limiting).
Nearly all electrodes - especially SI Analytics electrodes - fulfill these conditions.
7
Overviewhandylab pH 11
1.4Technical Data
Dimensions
and weight
Mechanical
structure
Electrical safety
Test certificates
Ambient
conditions
Length [mm]172
Width [mm]80
Height [mm]37
Weight [kg]Approx. 0.3
Type of protectionIP 66
Protective classIII
cETLus, CE
Storage- 25 °C ... + 65 °C
Operation-10 °C ... + 55 °C
Relative atmospheric
< 90 % of annual average
humidity
pH/ORP
measuring ranges
Precision of
pH/ORP
(± 1 digit)
Measuring rangeResolution
pH- 2.000 ... + 19.999
- 2.00 ... + 19.99
U [mV]- 999.9 ... + 999.9
- 1999 ... + 1999
0.001
0.01
0.1
1
T [°C]- 5.0 ... + 105.00.1
pH
(± 2 pH units from the
± 0.005 at + 15 °C ... + 35 °C
± 0.01
calibration point)
U [mV]± 0.3 at + 15 °C ... + 35 °C
± 1
T [°C]NTC 30:
Accuracy ± 0.1
PT 1000:
AccuracyOperating temperature
± 0.5 0 °C ... 15 °C
± 0.1 15 °C ... 35 °C
± 1 35 °C ... 55 °C
8
ba75366e0401/2011
handylab pH 11Overview
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.
Correction
function
Power supply
Applicable
guidelines and
norms
Temperature input
Manually [°C]- 20 ... + 130
Batteries4 x 1.5 V alkali-manganese batteries, Type
AA
Operational lifeApprox. 5000 operating hours
EMCEC guideline 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Instrument safetyEC guideline 2006/95/EC
EN 61010-1
ANSI/UL 61010-1
CAN/CSA-C22.2 No. 61010-1
Climatic classVDI/VDE 3540
Type of protectionEN 60529
ba75366e0401/2011
9
Overviewhandylab pH 11
10
ba75366e0401/2011
handylab pH 11Safety
2Safety
This operating manual contains basic instructions that you must follow during
the commissioning, operation and maintenance of the pH meter. Consequently, all responsible personnel must read this operating manual before
working with the measuring system. The operating manual must always be
available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in the lab-
oratory.
We assume that, as a result of their professional training and experience, the
operators will know the necessary safety precautions to take when handling
chemicals.
Safety
instructions
Further notes
The individual chapters of this operating manual use safety instructions such
as the label shown below to indicate various hazards or dangers:
Caution
indicates instructions that must be followed precisely in order to avoid the
possibility of slight injuries or damage to the instrument or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of the
pH and ORP measurement in the field and laboratory.
The technical specifications as given in section 1.4 T
observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is authorized.
Any other use is considered to be unauthorized.
ECHNICAL DATA must be
ba75366e0401/2011
11
Safetyhandylab pH 11
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines
and norms for electronic measuring instruments (see section 1.4 T
D
ATA).
It left the factory in a safe and secure technical condition.
ECHNICAL
Function and oper-
ating safety
The smooth functioning and operational safety of the measuring instrument
can only be guaranteed if the generally applicable safety measures and the
specific safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the measuring instrument
can only be guaranteed under the environmental conditions that are specified
in section 1.4 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the instrument reaches
room temperature before putting the instrument back into operation.
Safe operationIf safe operation is no longer possible, the instrument must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of time
z is visibly damaged
Obligations of the
purchaser
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following
laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
12
ba75366e0401/2011
handylab pH 11Commissioning
3Commissioning
3.1Scope of delivery
z Handheld meter, handylab pH 11
z Operating manual and short operating manual
z 4 batteries, 1.5 V Mignon type AA (in the instrument)
ba75366e0401/2011
13
Commissioninghandylab pH 11
14
ba75366e0401/2011
handylab pH 11Operation
4Operation
4.1Switching on the measuring instrument
1Press the <ON/OFF> key.
The display test appears briefly on the display.
Subsequently, the slope and asymmetry used appear for approx. one second one after the other. In addition, the calibration
procedure of the last calibration is shown (AutoCal TEC or AutoCal DIN or no display in the delivery state or following initialization).
The measuring instrument then automatically switches to the
measuring mode that was last selected.
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the measuring instrument off if no key has been pressed for an hour. The energy saving
feature is not active when the AutoStore function is active.
ba75366e0401/2011
15
Operationhandylab pH 11
4.2Measuring
4.2.1General information
You can measure the following variables:
z pH value
z ORP voltage
Preparatory
activities
Temperature
sensor
Perform the following preparatory activities when you want to measure:
1Connect the electrode to the measuring instrument.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3In conjunction with the electrode, check or calibrate the mea-
suring instument. How to calibrate is described in section 4.3.
4Select the measuring mode with <M>.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values.
Calibrate regularly before measuring.
You can measure with or without a temperature sensor. If a temperature sensor is connected, it is indicated on the display by TP.
Note
The measuring instrument automatically recognizes the type of the temperature sensor used. Therefore, you can connect temperature sensors of the
NTC30 or Pt1000 type.
16
The temperature measurement is absolutely essential for a reproducible pH
measurement. If the measurement is made without a temperature sensor,
proceed as follows:
1Measure the current temperature of the test sample using a
thermometer.
2Enter the temperature on the instrument: Set the temperature
value of the test sample using <▲ > <▼>.
ba75366e0401/2011
handylab pH 11Operation
mV/pH
mol/L
8
Auto
Store
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
mg/L
% ppm
ISE
8
4
2
29
9
6
°C
TP
pH
4.2.2Measuring the pH value
1Perform the preparatory activities according to section 4.2.1.
2Immerse the pH electrode in the test sample.
3Press the <M> key until pH appears on the status display. The
pH value appears on the display.
Setting the
resolution
AutoRead
(drift control)
The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the <M> key
while pressing the <RUN/ENTER> key.
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
For identical measurement conditions, the following criteria apply:
ReproducibilityResponse time
Better than 0.02> 30 seconds
1Call up the pH measuring mode with <M> and <▲> <▼> .
2Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
ba75366e0401/2011
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the <AR> key.
17
Operationhandylab pH 11
/pH
ConCal
Auto
Store
LoBat
AutoCal DIN
CalError
ARng
AR
AutoCalTEC
8
pH
1
%
°F
S
Time
Baud
Day.MonthNo.
Ident
Year
RCL
TempError
mg/L
% ppm
mol/L
ISE
3
1
2
2
0
mV
°C
U
TP
1
7
Note
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.
4.2.3Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. BlueLine 31Rx, the measuring instrument can measure the ORP voltage (U) of a solution.
1Perform the preparatory activities according to section 4.2.1.
2Immerse the ORP electrode in the test sample.
3Press the <M> key until U appears on the status display.
The ORP voltage (mV) of the test sample appears on the display.
4Wait for a stable measured value.
18
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes
using a test solution.
ba75366e0401/2011
handylab pH 11Operation
4.3Calibrating
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines the current values of the asymmetry and slope of the electrode
and stores them in the measuring instrument.
Thus, you should calibrate at regular intervals.
When to calibrate?z After connecting another electrode
z after a voltage interruption, e. g. after changing the batteries
You can select between 3 calibration procedures:
AutoCal TECis specially adapted to the SI Analytics technical buffer solutions as a fully au-
tomatic two-point or three-point calibration. The buffer solutions are auto-
matically recognized by the measuring instrument. Depending on the
instrument setting (see section 4.4 C
the relevant buffer nominal value or the current electrode voltage in mV. The
calibration can be terminated after the first buffer solution. This corresponds
to a single-point calibration. For this, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
ONFIGURATION), the instrument displays
AutoCal DINis specially adapted to the permanently programmed buffer solutions in ac-
cordance with DIN19266 as a fully automatic two-point or three-point cali-bration. The buffer solutions are automatically recognized by the measuring
instrument. Depending on the instrument setting (see section
4.4 C
ONFIGURATION), the instrument displays the relevant buffer nominal val-
ue or the current electrode voltage in mV. The calibration can be terminated
after the first buffer solution. This corresponds to a single-point calibration.
For this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
ConCalThis function is a conventional two-point calibration using two buffer solu-
tions (pH 7.0 ± 0.5 and any other buffer solution) or a single-point calibra-
tion using any buffer solution and is used as a high-speed method.
AutoReadWhen calibrating with AutoCal TEC and AutoCal DIN, the AutoRead function
is automatically activated.
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.
ba75366e0401/2011
19
Operationhandylab pH 11
Displaying
calibration data
Calibration
evaluation
Each time the instrument is switched on, the calibration data are shown on
the display for a short time (see section 4.1 S
STRUMENT). In order to view the calibration data, switch the measuring instru-
WITCHINGONTHEMEASURINGIN-
ment off and switch it on again.
After the calibration, the measuring instrument automatically evaluates the
current status of the electrode. The asymmetry and slope are evaluated separately. The worst evaluation appears on the display.
DisplayAsymmetry
[mV]
Slope
[mV/pH]
-15 ... +15-60.5 ... -58
-20 ... +20-58 ... -57
-25 ... +25-61 ... -60.5
or
-57 ... -56
Preparatory
activities
-30 ... +30-62 ... -61
or
-56 ... -50
Clean the electrode according
to the electrode operating
manual
E3
Eliminate the error according to
chapter 6 W
HATTODOIF...
< -30 or
> 30
< -62 or
> -50
1Switch on the measuring instrument with <ON/OFF>.
2Connect the pH electrode to the measuring instrument.
3Keep the buffer solutions ready.
4Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without temperature sensor.
20
ba75366e0401/2011
handylab pH 11Operation
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1C
t
°C
TP
pH
AutoCal TEC
4.3.1AutoCal TEC
For this procedure, use any two or three technical buffer solutions in ascending or descending order
(pH values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the SI Analytics technical buffer
solution pH 10.01. Other buffer solutions can lead to an erroneous calibration. The correct buffer solutions are given in the SI Analytics catalog or in the
Internet.
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see section 4.4 C
the instrument displays the relevant buffer nominal value or the current electrode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
ONFIGURATION),
1Press the <CAL> key repeatedly until the Ct1 display indicator
and the function display AutoCal TEC appears. The sensor
symbol displays the evaluation of the last calibration (or no
sensor symbol in the delivery state or after the measurement
parameter has been reset).
2If necessary, enter the temperature of the first buffer solution
with <▲> <▼>.
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ba75366e0401/2011
21
Operationhandylab pH 11
cm
3
7
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
AutoCal DIN
Tref25
Auto
AutoCal DIN
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
0
03
7
pHpH
C
°
C
°
AutoCal TEC
AutoCal TEC
ARAR
mV
88
44
22
TP
TP
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
AutoCal DIN
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
2C
t
°C
TP
pH
AutoCal TEC
5When the measured value is stable, Ct2 appears.
Note
At this point, the AutoCal TEC calibration can be terminated with <M>. This
corresponds to a single-point calibration. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the
electrode.
6To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
7If necessary, enter the temperature of the second buffer solu-
tion with <▲> <▼>.
8Immerse the pH electrode in the second buffer solution.
9Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
22
ba75366e0401/2011
handylab pH 11Operation
1
TDS
S
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
AutoCal DIN
Tref25
Auto
AutoCal DIN
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDS
1
S
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
88
44
08
14
4
pHpH
C
°
C
°
AutoCal TEC
AutoCal TEC
ARAR
mV
1
22
TP
TP
O
L
S
RCL
Sal
cm
AutoCal DIN
Day.Month
S
Tref20
Cal
Ident
No.
Year
Time
Baud
1
TDS
Sal
Store
AR
Tref25
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
Lin
OU
nLF
Auto
LoBat
ARng
°
F
9
4
pH
5
mV/pH
AutoCal TEC
YA
S
0
RCL
Sal
cm
AutoCal DIN
Time
Day.Month
Sal
Store
AR
8
No.
Ident
Tref20
Tref25
Year
Baud
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
TDS
6
Lin
Cal
1
S
OU
nLF
Auto
LoBat
ARng
°
F
2
pH
mV
AutoCal TEC
10When the measured value is stable, AR disappears. The in-
strument displays the value of the slope (mV/pH) and the evaluation of the calibration in calibration range I.
ba75366e0401/2011
11Press the <RUN/ENTER> key. The instrument displays the
value of the asymmetry in calibration range I.
Note
At this point, the AutoCal TEC calibration can be terminated with <M>. This
corresponds to a two-point calibration.
The instrument uses the slope and asymmetry of the calibration
range I.
23
Operationhandylab pH 11
Three-point
calibration
12Press the <RUN/ENTER> key
to continue the three-point calibration.Ct3 appears on the display.
13Thoroughly rinse the electrode with distilled water.
14If necessary, enter the temperature of the third buffer solution
with <▲> <▼>.
15Immerse the pH electrode in the third buffer solution.
16Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display.
When the measured value is stable, AR disappears.
The instrument displays the value of the slope (mV/pH) and the
evaluation of the calibration in calibration range II.
17Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibration range II. The three-point calibration is completed.
18To return to the measuring mode: Press the <M> key.
Note
You can prematurely terminate the three-point calibration by pressing <M>.
The values of the two-point calibration for slope and asymmetry will then remain stored.
24
ba75366e0401/2011
handylab pH 11Operation
AutoCalTEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
1C
d
°C
TP
pH
AutoCal DIN
4.3.2AutoCal DIN
For this procedure, use two or three different standard buffer solutions according to DIN 19266 in ascending or descending order (type A, C, D or F
with pH values at 25 °C of: 1.679 / 4.006 / 6.865 / 9.180).
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see section 4.4 C
the instrument displays the relevant buffer nominal value or the current electrode voltage in mV.
Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.
1Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol displays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).
ONFIGURATION),
ba75366e0401/2011
2If necessary, enter the temperature of the buffer solution with
<▲> <▼>.
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
25
Operationhandylab pH 11
AutoCalTEC
AutoCalTEC
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
Tref25
Auto
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
0
17
4
pHpH
C
°
C
°
ARAR
mV
88
44
22
TP
TP
7
1
AutoCal DIN
AutoCal DIN
AutoCalTEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
ConCal
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
2C
d
°C
TP
pH
AutoCal DIN
5When the measured value is stable, Cd2 appears.
Note
At this point, the AutoCal DIN calibration can be terminated with <M>. This
corresponds to a single-point calibration. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the
electrode.
26
6To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
7If necessary, enter the temperature of the second buffer solu-
tion with <▲> <▼>.
8Immerse the pH electrode in the second buffer solution.
9Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
ba75366e0401/2011
handylab pH 11Operation
AutoCalTEC
AutoCalTEC
°
F
0
Cal
Tref20
Lin
0
Cal
Tref25
ARng
Tref25
Auto
ARng
Tref20
Lin
F
cm
1/1/
cm
AR
AR
RCL
RCL
OOUpHUpHSalSal
Store
Store
Auto
LoBat
LoBat
nLF
nLF
SalSal
%%
mg/lmg/l
mV/pHmV/pH
cmcm
°
mm
S/S/
cmcm
%%
//
KK
MM
TDSTDS
11
SS
mbarmbar
Time
Baud
Day.MonthNo.
Ident
Year
Time
Baud
Day.MonthNo.
Ident
Year
1
89
9
pHpH
C
°
C
°
ARAR
mV
88
44
22
TP
TP
2
1
AutoCal DIN
AutoCal DIN
AutoCalTEC
YA
S
0
RCL
Sal
cm
Time
Day.Month
Sal
Store
AR
8
No.
Ident
Tref20
Tref25
Year
Baud
cm
C
/
%
K
TP
°
M
mbar
1/
%
m
mg/l
cm
S/
TDS
6
Lin
Cal
1
S
OU
nLF
Auto
LoBat
ARng
°
F
2
pH
mV
AutoCal DIN
10When the measured value is stable, AR disappears. The in-
strument displays the value of the slope (mV/pH) and the evaluation of the calibration in calibration range I.
ba75366e0401/2011
11Press the <RUN/ENTER> key. The instrument displays the
value of the asymmetry in calibration range I.
Note
At this point, the AutoCal DIN calibration can be terminated with <M>. This
corresponds to a two-point calibration.
The instrument uses the slope and asymmetry of the calibration
range I.
27
Operationhandylab pH 11
Three-point
calibration
12Press the <RUN/ENTER> key
to continue the three-point calibration.Cd3 appears on the display.
13Thoroughly rinse the electrode with distilled water.
14If necessary, enter the temperature of the third buffer solution
with <▲> <▼>.
15Immerse the pH electrode in the third buffer solution.
16Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display.
When the measured value is stable, AR disappears.
The instrument displays the value of the slope (mV/pH) and the
evaluation of the calibration in calibration range II.
17Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibration range II. The three-point calibration is completed.
18To return to the measuring mode: Press the <M> key.
Note
You can prematurely terminate the three-point calibration by pressing <M>.
The values of the two-point calibration for slope and asymmetry will then remain stored.
28
ba75366e0401/2011
handylab pH 11Operation
4.3.3ConCal
Single-point
calibration
Two-point
calibration
Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the buffer
solution is to that of the test sample.
For this procedure, use two buffer solutions:
z pH 7.0 ± 0.5
z any other buffer solution
Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature sensor.
1Press the <CAL> key repeatedly until the display ASY and the
function display ConCal appears. The sensor symbol displays
the evaluation of the last calibration (or no sensor symbol in the
delivery state or after the measurement parameter has been
reset).
ba75366e0401/2011
2If necessary, enter the temperature of the first buffer solution
with <▲> <▼>.
3Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).
4Press the <RUN/ENTER> key.
The measured pH value appears on the display.
5Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <▲> <▼>.
6Press the <RUN/ENTER> key. The value of the asymmetry
(mV) and the sensor symbol appear on the display.
7Press the <RUN/ENTER> key. SLO appears on the display.
Note
At this point, the ConCal calibration can be broken off with <M>. This corre-
29
Operationhandylab pH 11
AutoCal DIN
AutoCalTEC
6
mV/pH
mol/L
8
Auto
Store
RCL
LoBat
CalError
ARng
AR
Time
Baud
Day.MonthNo.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
OS
L
°C
TP
pH
ConCal
sponds to a single-point calibration. For this, the instrument uses the
Nernst slope (-59.2 mV/pH at 25 °C) and the specified asymmetry of the electrode.
8To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
9If necessary, enter the temperature of the second buffer solu-
tion with <▲> <▼>.
10Immerse the pH electrode in the second buffer solution.
11Press the <RUN/ENTER> key. The second pH value appears
on the display.
12Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <▲> <▼>.
13When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibration.
30
14Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
ba75366e0401/2011
handylab pH 11Operation
15To return to the measuring mode: Press the <M> key.
ba75366e0401/2011
31
Operationhandylab pH 11
4.4Configuration
You can adapt the measuring instrument to your individual requirements. To
do this, the following parameters can be changed (the status on delivery is
marked in bold):
Display during the
pH calibration
pH resolution 0.01 or 0.001
Note
You can leave the configuration menu at any time with <M> . The parameters
that have already been changed are stored.
1Switch off the measuring instrument.
2Press the <M> key and hold it down.
Buffer nominal value, or current
electrode voltage
Display during the
pH calibration
3Press the <ON/OFF> key.
The display test appears briefly on the display. CAL diSP appears on the display.
4Select the required display during the pH calibration with
<▲> <▼>.
mV: Display of the current electrode voltage
/pH: Display of the buffer nominal value.
5Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH
measuring mode.
32
ba75366e0401/2011
handylab pH 11Operation
Setting the
resolution
1Press the <RUN/ENTER> key and hold it down.
2Press the <M> key.
The measured values are displayed with a high resolution, e.g.
pH = 4.012.
3Press the <RUN/ENTER> and <M> keys once more.
The measured values are displayed with a low resolution, e.g.
pH = 4.01.
ba75366e0401/2011
33
Operationhandylab pH 11
Tref25
Tref20
AutoCalTEC
RCL
LoBat
Cal
ARng
AR
Lin
nLF
Store
Sal
°
°
TP
cm
1/
F
C
%
/
K
8
8
1
S/
cm
m
M
TDS
Sal
S
Auto
mbar
%
mg/l
O
mV/pH
cm
AutoCal DIN
Time
Baud
Day.MonthNo.
Ident
Year
o
n
n
I
I
pH
4.5Reset
You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.
Measurement
parameters
Resetting the
measuring
parameters
The following measured parameters (pH InI) are reset to the default condi-
tion:
Measuring modepH
Asymmetry0 mV
Slope -59.16 mV/pH
Calibration proceduresAutoCal TEC
Temperature, manual25 °C
Resolution (pH display)0.01 (low resolution)
Display during the
pH calibration
Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.
1Press the <RUN/ENTER> key and hold it down.
2Press the <CAL> key.
Buffer nominal value
34
3Using <▲> <▼>, switch between no and YES.
YES: Resetting the measuring parameters
no: Retaining settings.
4Confirm with <RUN/ENTER>.
The measuring instrument automatically switches to the pH
measuring mode.
ba75366e0401/2011
handylab pH 11Maintenance, cleaning, disposal
1
2
5Maintenance, cleaning, disposal
5.1Maintenance
The measuring instrument is almost maintenance-free.
The only maintenance task is replacing the batteries.
LoBat indicates that the batteries should be changed. The batteries are then
largely depleted.
Replacing the
batteries
1Open the housing after the instrument has been switched off:
– Undo the four screws on the underside of the instrument
– Pull down the lower cover (1).
ba75366e0401/2011
2If necessary, take the four depleted batteries (2) out of the bat-
tery compartment.
3Place four new batteries (type Mignon AA) in the battery com-
partment.
4Close the lower cover (1).
Caution
Make sure that the poles of the batteries are the right way round.
The
±signs on the batteries must correspond to the ± signs in the battery
compartment.
Only use leakproof alkaline manganese batteries.
35
Maintenance, cleaning, disposalhandylab pH 11
Note
For the maintenance of the electrodes, follow the corresponding operating
manual.
5.2Cleaning
Occasionally wipe the outside of the measuring instrument with a damp, lintfree cloth. Disinfect the housing with isopropanol as required.
Caution
The housing is made of a synthetic material (ABS). Thus, avoid contact with
acetone and similar detergents that contain solvents. Remove any splashes
immediately.
5.3Disposal
PackingThis measuring instrument is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the
instrument against damage during transport.
BatteriesThis note refers to the battery regulation that applies in the Federal Republic
of Germany. We would ask end-consumers in other countries to follow their
local statutory provisions.
Note
This instrument contains batteries. Batteries that have been removed must
only be disposed of at the recycling facility set up for this purpose or via the
retail outlet.
It is illegal to dispose of them in household refuse.
Measuring
instrument
Dispose of the measuring instrument as electronic waste at an appropriate
collection point. It is illegal to dispose of the instrument in household refuse.
36
ba75366e0401/2011
handylab pH 11What to do if...
6What to do if...
Error message 0FL
CauseRemedy
pH electrode:
– Not connected– Connect electrode
Error message E3
– Air bubble in front of the
– Remove air bubble
diaphragm
– Air in the diaphragm– Extract air or
moisten diaphragm
– Cable broken– Replace electrode
– Gel electrolyte dried out– Replace electrode
CauseRemedy
Electrode
– Diaphragm contaminated– Clean diaphragm
– Membrane contaminated– Clean membrane
– Moisture in the plug– Dry plug
– Electrolyte out of date– Replenish electrolyte or
– pH value not stable– Measure with air excluded if
necessary
– Temperature not stable– Adjust temperature if
necessary
Electrode + test sample:
– Conductivity too low– Use suitable electrode
– Temperature too high– Use suitable electrode
– Organic liquids– Use suitable electrode
CauseRemedy
– Batteries almost empty– Replace batteries
(see section 5.1 MAINTENANCE)
38
ba75366e0401/2011
handylab pH 11What to do if...
Obviously incor-
rect
measured values
Instrument does
not react to key-
stroke
CauseRemedy
pH electrode:
– pH electrode unsuitable– Use suitable electrode
– Temperature difference
between buffer and test
– Adjust temperature of buffers
or test samples
sample too high
– Measurement procedure not
– Follow special procedure
suitable
CauseRemedy
– Operating condition
undefined or EMC load
unallowed
– Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Display to
CauseRemedy
– Time-out of the interface– Check the instrument that is
connected
ba75366e0401/2011
39
What to do if...handylab pH 11
40
ba75366e0401/2011
handylab pH 11Lists
7Lists
This chapter provides additional information and orientation aids.
AbbreviationsThe list of abbreviations explains the indicators and the abbreviations that ap-
pear on the display and in the manual.
Specialist termsThe glossary briefly explains the meaning of the specialist terms. However,
terms that should already be familiar to the target group are not described
here.
IndexThe index will help you to find the topics that you are looking for.
ba75366e0401/2011
41
Listshandylab pH 11
Abbreviations
ARAutoRead (drift control)
ARngAutomatic range switching
Measuring instrument measures with highest resolution
ASYAsymmetry
AutoCal DINAutomatic pH calibration with buffer solutions prepared
according to DIN 19 266
AutoCal TECAutomatic pH calibration with SI Analytics technical buf-
fer solutions according to DIN 19267
°CTemperature unit, degrees Celsius
CalCalibration
Cd...Display indicator during calibration for pH measure-
ments. Indicates the selection of the buffer data record
for buffer solutions prepared according to DIN 19 266
ConCalConventional single-point or two-point calibration for pH
measurements
Ct...Display indicator during calibration for pH measure-
ments. Indicates the selection of the buffer data records
for SI Analytics technical buffer solutions
InIInitialization
Resets individual basic functions to the status they had
on delivery
LoBatBatteries almost empty(Low Battery)
mVVoltage unit
mV/pHUnit of the electrode slope (internat. mV)
OFLDisplay range exceeded (Overflow)
pHpH value
SSlope (internat. k)
SELVSafety Extra Low Voltage
SLOSlope setting on calibration
42
TPTemperature measurement active (Temperature Probe)
U
ASY
Asymmetry
ba75366e0401/2011
handylab pH 11Lists
Glossary
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains within the tolerance.
AsymmetryDesignation for the offset potential of a pH electrode. It is the measur-
able potential of a symmetrical electrode, the membrane of which is
immersed in a solution with the pH of the nominal electrode zero point
(electrodes: pH = 7).
AutoRangeName of the automatic selection of the measuring range.
AutoReadName for a function to check the stability of the measured value.
CalibrationComparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
DiaphragmThe junction is a porous body in the housing wall of reference elec-
trodes or electrolyte bridges. It forms the electrical contact between
two solutions and makes electrolyte exchange more difficult. The expression, junction, is also used for ground or junction-less transitions.
Electrode zero pointThe zero point of a pH electrode is the pH value at which the electro-
motive force of the pH electrode at a specified temperature is zero.
Normally, this is at 25 °C.
Electromotive force of
an electrode
The electromotive force U of the electrode is the measurable electromotive force of an electrode in a solution. It equals the sum of all the
galvanic voltages of the electrode. Its dependency on the pH results
in the electrode function which is characterized by the parameters,
slope and zero point.
Measured parameterThe measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
Measured valueThe measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Measuring systemThe measuring system comprises all the devices used for measuring,
e. g. measuring instrument and probe. In addition, there is the cable
and possibly an amplifier, terminal strip and armature.
MolalityMolality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
MultiCal
ba75366e0401/2011
vent.
®
Name stating that a measuring instrument provides several calibration
procedures.
43
Listshandylab pH 11
Offset potentialThe measurable potential of a symmetrical electrode, the membrane
of which is immersed in a solution with the pH of the nominal electrode
zero point. The asymmetry is part of the offset potential.
ORP voltageThe ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface
(e. g. a gold or platinum surface).
pH valueThe pH is a measure of the acidic or basic effect of an aqueous solu-
tion. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH
value is the value of a pH measurement.
PotentiometryName of a measuring technique. The signal (depending on the mea-
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
ResetRestoring the original condition of all settings of a measuring system.
ResolutionSmallest difference between two measured values that can be dis-
played by a measuring instrument.
SlopeThe slope of a linear calibration function.
Standard solutionThe standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Test sampleDesignation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
44
ba75366e0401/2011
handylab pH 11Index
Index
A
Asymmetry 19
Authorized use 11
AutoCal DIN 19, 25
AutoCal TEC 19, 21
AutoRead 17