Thermo Menu
Service Manuals
Operation Manual
Calibration Tester
Model 9600
Welch Allyn 9600 Calibration Tester
Calibration
ModelTester
9600
|
Operation Manual |
CONTENTS |
|
INTRODUCTION ................................................................................ |
4 |
WHAT THE 9600 CALIBRATION SYSTEM IS ........................................ |
4 |
WHAT THE 9600 CALIBRATION SYSTEM DOES ................................. |
4 |
MAJOR COMPONENTS OF THE 9600 .............................................. |
5 |
TEMPERATURE SELECT SWITCH ...................................................... |
5 |
POWER INDICATOR ............................................................................. |
5 |
READY INDICATOR ............................................................................... |
5 |
OVER-TEMPERATURE DETECTOR ..................................................... |
6 |
THERMOMETER HOLES ...................................................................... |
6 |
USING THE 9600 SYSTEM ................................................................ |
7 |
SETUP PROCEDURE ........................................................................... |
7 |
LEVEL 1 CALIBRATION TEST .......................................................... |
7 |
THE MODEL 9000 THERMOMETER .................................................... |
8 |
THERMISTOR THERMOMETERS ......................................................... |
9 |
CELSIUS-ONLY THERMISTOR THERMOMETERS .............................. |
9 |
LEVEL 2 CALIBRATION TEST ........................................................ |
10 |
LEVEL 2 CALIBRATION TEST ............................................................. |
10 |
TROUBLESHOOTING ..................................................................... |
11 |
MODEL 9600 CALIBRATION CHECKS ................................................ |
11 |
ENVIRONMENTAL CHECKS ............................................................... |
12 |
MODEL 9000 CALIBRATION ADJUSTMENT .................................. |
12 |
PRIOR TO CALIBRATION ADJUSTMENT ........................................... |
13 |
ESTABLISHING A BASELINE PRIOR TO MAKING |
|
CALIBRATION ADJUSTMENTS ........................................................... |
13 |
SETTING THE THERMOMETER TO THE |
|
CALIBRATION ADJUSTMENT MODE .................................................. |
14 |
SPECIFICATIONS ........................................................................... |
15 |
LIMITED WARRANTY ...................................................................... |
16 |
3
Welch Allyn 9600 Calibration Tester
The 9600 Calibration Tester is a device for testing the calibration of
Welch Allyn thermometers. The unit contains solid-state circuitry for longterm stability, with virtually no maintenance.
The 9600 System provides a convenient “reference patient” for testing the calibration of:
nThermistor-based thermometers (for example, Models 500, 600, 650, 678, 679, 767T, SureTemp® and SureTemp4®).
nInfrared-technology Model 9000 and Model 9020 Welch Allyn thermometers.
To maintain long-term stability, as well as to ensure strict control over varying environmental conditions, only the highest precision components are used in the construction of the 9600 System. For this reason, the instrument must be factory calibrated on a regular basis. The initial calibration, if performed within the warranty period, is free of charge. Please contact your Welch Allyn Customer Service representative at (800) 8542904 for further information.
The 9600 Calibration Tester provides a tightly regulated temperature source that is compatible with all Welch Allyn thermometer products. The device’s “temperature” can be taken with the knowledge and assurance that precise readings will be obtained as the unit is factory calibrated to very close tolerances. User adjustments are not required or recommended.
With the 9600 System, the difficult and time-consuming setup, precision calibration, and maintenance requirements of water baths and other general purpose equipment, such as ovens and environmental chambers, is eliminated.
4
Operation Manual
The Temperature Select Switch and the POWER and READY Indicators are located on the front of the Main Unit. Whenever the unit is plugged into a standard wall outlet, the POWER indicator will illuminate. There is no power ON/OFF switch. The READY indicator will only illuminate when the temperature stabilizes at the temperature you have selected on the Temperature Select Switch.
NOTE: For specific operational details about the components of the 9600 Calibration Tester, see the Specifications Section at the end of this manual.
When unpacking the 9600 Calibration Tester, notice that the Power Module is permanently connected to the Main Unit. This is to assure a continuous power supply to the unit, since intermittent power would cause the unit to lose temperature control and would turn off the READY Indicator.
After the 9600 warms up, the unit will draw about 7 watts of power from a standard wall outlet; therefore, it may be left on continuously. However, if the unit will not be used for long periods of time, it may be unplugged.
The 9600 Calibration Tester contains an independent temperature sensing circuit that senses temperature changes more precisely than the thermometers it tests.
Because the unit contains this independent circuitry, a failure in the tempera- ture-control circuit will not alter its “READY-sensing” ability. It is virtually impossible for the READY indicator to illuminate if the temperature is incorrect.
NOTE: Do not use the unit if the READY indicator is not illuminated. If the 9600 does not show “Ready” within a few minutes of the time allowed in the Specifications Section of this manual, and the environment is within acceptable operating conditions, the unit is malfunctioning and should be returned to the factory for repair or replacement.
5