Welch Allyn Tycos User manual

0 (0)
Welch Allyn Tycos User manual

Manuals

Operator Manual

Service Manual

Aneroid Sphygmomanometers

PN 95P504 Rev A

Copyright 2000

Welch Allyn Inc.

 

Aneroid Sphygmomanometers Service Manual

Revision A

 

 

 

 

Revision History

 

 

Revision

Section

Date

Author

Description

A

All

8/31/00

JDB

Introduction of Aneroid Service

 

 

 

 

Manual

Welch Allyn Inc.

Page 2 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

 

Revision A

 

 

Table of Contents

 

 

Section 1: General Information .......................................................................................................................

 

4

1.1

To Service Personnel ............................................................................................................................

 

4

1.2

Tools and Equipment ............................................................................................................................

 

5

1.3

Test Specifications ................................................................................................................................

 

6

1.4

Test Equipment Bench Layout ..............................................................................................................

 

7

Section 2: Classic Pocket Aneroid Gauges .....................................................................................................

 

8

2.1

Opening/Removing and Replacing Parts of the Pocket Gauge* ...........................................................

8

2.2

Calibration of the Pocket Gauge ...........................................................................................................

 

9

2.3

Testing ................................................................................................................................................

 

11

Section 3: Classic Hand Aneroid Gauge .......................................................................................................

 

12

3.1

Opening/Removing and Replacing Parts of the Hand Gauge*............................................................

12

3.2

Calibration of the Hand Gauge ...........................................................................................................

 

13

3.3

Testing ................................................................................................................................................

 

15

Section 4: Classic Wall and Mobile Aneroid Gauge.....................................................................................

 

16

4.1

Opening/Removing and Replacing Parts of the Wall and Mobile Gauge.* ........................................

16

7670 series Wall and Mobile Aneroid. .................................................................................................

 

16

Classic Wall and Mobile Aneroid. ........................................................................................................

 

16

4.2

Calibration of the Wall and Mobile Gauge .........................................................................................

 

17

4.3

Testing ................................................................................................................................................

 

19

Section 5: Economy Aneroid Gauge .............................................................................................................

 

20

5.1

Opening/Removing and Replacing Parts of the Economy Gauge* .....................................................

20

5.2

Calibration of the Economy Gauge .....................................................................................................

 

21

5.3

Testing ................................................................................................................................................

 

23

Section 6: Silver Ring Hand Aneroid...........................................................................................................

 

24

6.1

Opening/Removing and Replacing Parts of the Silver Ring ...............................................................

24

6.2

Calibration of the Silver Ring Hand Aneroid......................................................................................

 

25

6.3

Testing of the Silver Ring Hand Aneroid............................................................................................

 

27

Section 7: Pro Check Hand Aneroid .............................................................................................................

 

28

7.1

Opening/Removing and Replacing Parts of Pro Check Hand Aneroid ...............................................

28

7.2

Calibration of the Pro Check Hand Aneroid .......................................................................................

 

29

7.3

Testing of the Pro Check Hand Aneroid .............................................................................................

 

31

Section 8: Pictures and Replacement Parts ...................................................................................................

 

32

Section 8.1: Classic Pocket Aneroid Replacement Parts...........................................................................

32

Section 8.2: Classic Hand Aneroid Replacement Parts.............................................................................

33

Section 8.3: 767 Wall/Mobile Aneroid Replacement Parts.......................................................................

34

Section 8.4: Classic Wall/Mobile Aneroid Replacement Parts .................................................................

35

Section 8.5: Economy Aneroid Replacement Parts...................................................................................

 

36

Section 8.6: Silver Ring Hand Aneroid Replacement Parts ......................................................................

37

Section 8.7: Pro Check Hand Aneroid Replacement Parts........................................................................

38

Section 8.8: Welch Allyn Tycos Movement .............................................................................................

 

39

Section 8.9: Welch Allyn S+K Movement................................................................................................

 

40

Welch Allyn Inc.

Page 3 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

Section 1: General Information

1.1 To Service Personnel

The information contained in this manual is subject to change without notice and should not be construed as commitment by Welch Allyn Inc.

Welch Allyn Inc. assumes no responsibility for any errors that may appear in this manual. If the product and/or its operation varies significantly from any description herein, please contact the

Welch Allyn Inc. Technical Service Center 800-450-9275 (828) 684-4895

Fax (828) 687-1002

Welch Allyn Inc.

95 Old Shoals Road

Arden, North Carolina 28704

USA

This product has been designed to provide a high degree of safety and reliability. However, Welch Allyn Inc. can not guarantee against: deterioration of components due to aging, normal wear, tampering, and abuse.

Warning: If in the course performing calibration or repair on the gauge the device becomes inoperable and must be sent in to the Welch Allyn Inc. Service Center, The warranty will not be honored and you will be charge for all repairs.

Welch Allyn Inc.

Page 4 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

1.2 Tools and Equipment

The following is a list of the tools and equipment that is required to do the repair and calibration of the aneroid gauges.

Tool Number

Description

 

 

T106845A

Calibration Tool

 

T105243A

Pointer Removal Tool

 

T105243B

Pointer Removal Tool

 

T107345

Span Wrench

 

T110657

Calibration Dial Adapter, Economy

T105721

Calibration Dial, Pocket and Hand Gauge

T112663

Calibration Dial, Walls and Mobiles

T106763

Calibration Stand (Wall Aneroid)

 

T112592

Calibration Dial (Silver Ring)

 

T105258

Wrench (Economy)

 

T108913

3/8” Open End Wrench

 

T108897

Movement Removal Wrench

 

T113265

Thumbscrew Wrench

 

T110543

Pliers to Grip Lure

 

T105644

Collet

 

T55804-1

Pocket Screw Cap Assembly

 

T108212

Economy Fixture

 

T107569

Hand Fixture

 

T74555

Pocket Fixture

 

T112655

Top Plate Wrench

 

T113249

Screw Cap Removal Tool

 

T112927

5/16” Open End Wrench

 

T112765

¼” Open End Wrench

 

T34938

Pointer Block

 

T112511B

Tweezers

 

T112726

Jewelers Screw Driver

 

T112595

Calibration Tool (Silver Ring)

 

“Disytest”

Calibration Dial for Pro Check (Order From S+K)

T112833

Movement Removal Tool (S+K Movement)

T112653

Torque Wrench (S+K movement)

T112653A

Adapter for Torque wrench (S+K Movement)

Test Equipment

Description

 

Other Tools Needed but

5088-01

Bulb and Valve

Not Provided By Welch Allyn

T112854

500cc Volume

1. Small Hammer

9585TPK4

“T” Fitting

2. 5/8” Socket

9586TPK4

“Y” Fitting

3. ½” Socket

97P24

Pneumatic Tubing (4ft.)

4. ¾” Open End Wrench

200-2000IN

Netech Digital Pressure Meter

5. 9/16” - 24 Thread Tap

“or”

 

 

 

6. ¼” - 40 Thread Tap

2270-01

Setra Digital Pressure Meter.

7. 7/16” - 36 Thread Tap

T111962A

Tension Set Tool

 

T112459

Torque Screwdriver

 

T112459A

Torque bit.

 

Welch Allyn Inc.

Page 5 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

1.3 Test Specifications

The following are the test specification for the Welch Allyn Tycos Aneroid Gauge:

1.The gauge will be accurate to within +/- 3 mmHg* at the following test pressures:

a)0 mmHg (pointer inside the oval/rectangle)

b)60 mmHg

c)150 mmHg

d)200 mmHg

e)300 mmHg

*This accuracy is based on the use of a calibrated digital pressure meter. If using a calibrated aneroid or mercury column for measuring the accuracy of the aneroid in question the resulting accuracy is +/- 6 mmHg.

Welch Allyn Inc.

Page 6 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

1.4 Test Equipment Bench Layout

500cc Volume

 

Digital

 

 

Pressure

 

 

Meter

 

 

 

Gauge

Under

Test

Figure 1.5

Welch Allyn Inc.

Page 7 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

Section 2: Classic Pocket Aneroid Gauges

2.1 Opening/Removing and Replacing Parts of the

Pocket Gauge*

1.Place the pocket gauge into the Pocket Gauge Fixture (T74555).

2.Remove the bezel by unscrewing it from the case and aside. Remove the crystal and aside.

3.Remove the pointer by using the pointer removal tool (T105243A).

4.Pry out the dial by using a small flat head screwdriver.

5.Remove the locking nut using the span wrench (T107345).

6.Remove the screw cap using (T113249) and remove the capsule.

Note: the arm of the movement fits into the top of the capsule.

7.Remove the 2 screws that are holding down the gauge movement using T108897 and remove the movement.

Reassemble in reverse order.

* - See section 8.1 for exploded view drawing for the Pocket Gauge.

Welch Allyn Inc.

Page 8 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

2.2 Calibration of the Pocket Gauge

1.Attach the pneumatic tubing of test equipment to the gauge.

2.Place gauge in to fixture (T74555)

3.Remove bezel, crystal, pointer and dial.

To set the tension:

4.Tighten all screws.

5.View sector gear and ensure proper position (i.e. right edge of sector gear should align with mark on base of movement).

6.If sector gear not aligned properly, loosen the capsule nut using a 3/8” wrench and adjust the sector gear by turning the capsule until sector gear is in correct position per step 5. Tighten capsule nut.

7.Loosen sector stop lug and pull toward bottom of gauge.

8.Using your left index finger, push the sector gear all the way to the post and hold that in position.

9.Place tension setting tool (T119628) onto pinion at the 9 o’clock position.

10.Rotate tool clock wise to the 3 o’clock position.

11.Release sector gear and continue rotation to the 6 o’clock position sector gear should be moving during this step).

12.Bring lug back to correct position and lock into place.

To set the span:

13.Place Test Dial onto gauge. Place pointer onto pinion. Rotate dial until pointer is at “0” at the bottom of the gauge.

14.Apply 320 mmHg to the Gauge.

15.Release pressure to 300 mmHg to the gauge.

16.If point is less than 300, Place the calibration tool (T106854A) into the radius plate in slot above screw. Pull calibration tool toward bottom of gauge moving the sector gear upward until pointer reached “300.” Release pressure and check “0” point.

17.If point is greater than 300, Using the calibration tool (T106845A) positioned between the sector gear and radius plate on the left of the movement, rotate the tool clock wise until pointer is at “300” (this brings the radius plate down). Release pressure and check “0” point.

18.If not at “0” upon release of pressure, rotate test dial until pointer is on the “0” mark of test dial and re-apply 300 mmHg. Repeat steps 14-17 until span is

set.

19.Check for leak by pinching off the tubing from the test equipment while pressure is on the gauge. There should be no visible leak.

Welch Allyn Inc.

Page 9 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

Check and finish calibration:

20.Remove the pointer and test dial, replace the gauge’s dial.

21.Place pointer onto pinion and resting in center of oval. Tap pointer with small hammer slightly to secure.

22.Apply 320 mmHg to the gauge.

23.Release pressure to 300 mmHg and take readings here and at given points (see section 1.3).

24.If pointer is greater than 300, pull pointer until you feel slight resistance counter-clock wise, tap pointer with small hammer and recheck. Make sure pointer returns to the oval each time. Repeat until satisfied.

25.If less than 300, rotate pointer 1 complete revolution clockwise until resistance is felt. Push pointer a little farther and tap pointer with small hammer and recheck. Make sure pointer returns to the oval each time.

Repeat until satisfied.

26.Place crystal and bezel back onto gauge and check.

Please see Section 8.8 for a drawing of the Welch Allyn Tycos Movement.

Welch Allyn Inc.

Page 10 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

2.3 Testing

1.Hook up the Pocket Gauge that needs to be tested to the pneumatic tubing of the test equipment.

2.Apply 320 mmHg to the gauges.

3.Release the pressure to 300 mmHg on the digital pressure meter, or a calibrated aneroid or mercury gauge. Note: The aneroid or mercury gauge must be accurate to +/- 3 mmHg at the stated target pressures using a calibrated digital pressure meter. If using a calibrated aneroid or mercury column the accuracy must be +/- 6 mmHg.

4.Check reading of the pressure meter or gauge to that of the Pocket Gauge under test. Refer to section 1.3 for specifications.

Welch Allyn Inc.

Page 11 of 40

Welch Allyn Inc.

Aneroid Sphygmomanometers Service Manual

Revision A

 

Section 3: Classic Hand Aneroid Gauge

3.1 Opening/Removing and Replacing Parts of the Hand

Gauge*

1.Pull bulb off of valve body (48S95).

2.Place the Hand Gauge into the Hand Gauge Fixture (T107569).

3.Remove the bezel by unscrewing it from the case and aside. Remove the crystal and aside.

4.Remove the pointer by using the pointer removal tool (T105243A).

5.Pry out the dial by using a small flat head screwdriver.

6.Remove the cover of the Hand Gauge by unscrewing the two screws (9P2651) with a Phillips head screwdriver.

7.Loosen the locking nut (32P527) from the capsule and remove both from the case.

8.Remove the 9P1799 screws from the case and movement and remove the movement from the case.

9.Remove the Thumbscrew by using tool # T113265 (Thumbscrew wrench) and turning the thumbscrew counter-clock wise.

10.Remove the Thumbscrew valve body (42P252) using a 5/8” socket.

11.Remove the valve body (48S95) from the case using a 1/2” socket.

Reassemble in reverse order

* - See section 8.2 for an exploded view drawing of the Hand Gauge.

Welch Allyn Inc.

Page 12 of 40

Loading...
+ 28 hidden pages