Watson-Marlow 520Du User Manual

4.8 (6)

WATSON-MARLOW BREDEL MANUALS

m-520du-gb-05

Watson-Marlow 520Du pumps

Contents

1

Declaration of conformity

3

 

18.16

Language

51

2

Declaration of incorporation

3

 

18.17

Defaults

51

3

Warranty

4

 

18.18

Beep

52

4

When you unpack your pump

5

 

18.19

Security code

52

5

Information for returning

 

 

18.20

Exit

54

 

pumps

6

19

MemoDose

55

6

Peristaltic pumps - an overview

7

 

19.1

Changing dosing speed

57

7

Safety notes

8

 

19.2

Footswitch operation

 

8

Specifications

10

 

 

with MemoDose

57

 

8.1

Dimensions

14

20

Pin out details

58

9

Good pump installation practice

15

21

Exit

 

59

 

9.1

General recommendations

15

22

Automatic control wiring

60

 

9.2

Do’s and do not’s

15

 

22.1

Speed: analogue input 1

63

10

Connecting to a power supply

17

 

22.2

Scaling: analogue input 2

64

11

Start-up check list

20

 

22.3

Speed: analogue output

65

12

Switching on for the first time

20

 

22.4

Tachometer frequency

 

13

Switching on subsequently

22

 

 

output

65

14

Manual operation

23

 

22.5

Run / stop input

66

 

14.1

Keypad functions

 

 

22.6

Direction input

67

 

 

in manual mode

23

 

22.7

Auto / manual toggle

67

 

14.2

Keypad lock

26

 

22.8

MemoDose input

68

 

14.3

Keypad beep

26

 

22.9

Leak detection input

68

 

14.4

Manual operation and

 

 

22.10

Logic outputs

70

 

 

remote digital inputs

 

 

22.11

Supply voltages

73

 

 

and outputs

26

23

Automatic control and operation 74

15

Main menu

27

24

Network control and operation

77

 

15.1

Keypad in menu screens

27

 

24.1

RS232 command strings

79

 

15.2

Main menu entry

28

25

Troubleshooting

80

16

PIN-secure process protection

29

 

25.1

Error codes

81

17

Calibrate

30

26

Drive maintenance

82

18

Setup

 

34

27

Drive spares

82

 

18.1

Trim

35

28

The 520R, 520R2 and 520RE

 

 

18.2

Analogue

36

 

pumpheads

82

 

18.3

Display

40

 

28.1

Pumphead fitting

83

 

18.4

Pump number

41

29

520R, 520R2 and 520RE

 

 

18.5

Baud

41

 

installation

85

 

18.6

Stop bits

42

 

29.1

Opening the guard

85

 

18.7

Xon/Xoff

42

 

29.2

520R and 520R2

 

 

18.8

Flow units

43

 

 

tube loading

85

 

18.9

Run time

44

 

29.3

520RE: fitting the

 

 

18.10

Outputs

45

 

 

drain port

87

 

18.11

Remote stop

47

 

29.4

520RE Element loading

87

 

18.12

Auto-restart

48

 

29.5

520RE Element connection 90

 

18.13

Set maximum speed

49

30

520R, 520R2 and 520RE

 

 

18.14

Backlight

50

 

maintenance

91

 

18.15

ROM

50

 

 

 

 

Watson-Marlow 520Du User Manual

1

31

520R, 520R2 and 520RE

 

 

rotor settings

92

32

Pumphead spares

93

33

Flow rates

94

34

Tubing and element part

 

 

numbers

102

35

520 series pumping

 

 

accessories

105

36

Trademarks

106

37

Patient-connected applications

106

38

Publication history

106

39

Decontamination certificate

107

Watson-Marlow 520Du User Manual

2

1 Declaration of conformity

This declaration was issued for Watson-Marlow 520Du pumps on February 1, 2004. When this pump unit is used as a stand-alone pump it complies with: Machinery Directive 2006/42/EC, EMC Directive 2004/108/EC.

This pump is ETL listed: ETL control number 3050250. Cert to CAN/CSA std C22.2 No 1010-92. Conforms to UL std 61010A, April 30, 2002.

See 8 Pump specifications.

2 Declaration of incorporation

When this pump unit is to be installed into a machine or is to be assembled with other machines for installations, it must not be put into service until the relevant machinery has been declared in conformity with the Machinery Directive 2006/42/EC. See 8 Pump specifications.

Responsible person: Christopher Gadsden, Managing Director, Watson-Marlow Limited, Falmouth, Cornwall TR11 4RU, England. Telephone +44 (0) 1326 370370 Fax +44 (0) 1326 376009.

The information in this user guide is believed to be correct at the time of publication. However, Watson-Marlow Limited accepts no liability for errors or omissions. Watson-Marlow Bredel has a policy of continuous product improvement, and reserves the right to alter specifications without notice. This manual is intended for use only with the pump it was issued with. Earlier or later models may differ. The most up-to-date manuals appear on the Watson-Marlow website: http://www.watson-marlow.com

Watson-Marlow 520Du User Manual

3

3 Five year warranty

520 cased pumps, 620 cased pumps and 720 cased pumps

For any 520, 620 or 720 cased pump purchased after 1 January 2007, WatsonMarlow Limited (“Watson-Marlow”) warrants, subject to the conditions and exceptions below, through either Watson-Marlow, its subsidiaries, or its authorised distributors, to repair or replace free of charge, any part of the product which fails within five years of the day of manufacture of the product. Such failure must have occurred because of defect in material or workmanship and not as a result of operation of the product other than in normal operation as defined in this pump manual.

Watson-Marlow shall not be liable for any loss, damage, or expense directly or indirectly related to or arising out of the use of its products, including damage or injury caused to other products, machinery, buildings, or property, and Watson-Marlow shall not be liable for consequential damages, including, without limitation, lost profits, loss of time, inconvenience, loss of product being pumped, and loss of production. This warranty does not obligate Watson-Marlow to bear any costs of removal, installation, transportation, or other charges which may arise in connection with a warranty claim.

Conditions of and specific exceptions to the above warranty are:

Conditions

Products must be returned by pre-arrangement, carriage-paid, to WatsonMarlow, or a Watson-Marlow approved service centre.

All repairs or modifications must have been made by Watson-Marlow Limited, or a Watson-Marlow approved service centre or with the express permission of Watson-Marlow.

Warranties purporting to be on behalf of Watson-Marlow made by any person, including representatives of Watson-Marlow, its subsidiaries, or its distributors, which do not accord with the terms of this warranty shall not be binding upon Watson-Marlow unless expressly approved in writing by a Director or Manager of Watson-Marlow.

Exceptions

The warranty shall not apply to repairs or service necessitated by normal wear and tear or for lack of reasonable and proper maintenance.

All tubing and pumping elements as consumable items are excluded.

Products which, in the judgment of Watson-Marlow, have been abused, misused, or subjected to malicious or accidental damage or neglect are excluded.

Electrical surge as a cause of failure is excluded.

Chemical attack is excluded

All pumphead rollers are excluded.

The 620R family of pumpheads are excluded from all warranty when pumping above 2 bar while above 165rpm.

Pumpheads from the 313/314 and the Microcassette ranges and any 701 extension pumpheads are excluded and retain their one-year standard pumphead warranty. The drive they are attached to is subject to the five-year warranty as set out here.

Ancillaries such as leak detectors are excluded.

Watson-Marlow 520Du User Manual

4

4 When you unpack your pump

Unpack all parts carefully, retaining the packaging until you are sure all components are present and in good order. Check against the components supplied lists, below.

Packaging disposal

Dispose of packaging materials safely, and in accordance with regulations in your area. Pay particular attention to the expanded polystyrene shockproof shells. The outer carton is made of corrugated cardboard and can be recycled.

Inspection

Check that all components are present. Inspect components for damage in transit. If anything is missing or damaged, contact your distributor immediately.

Components supplied (520Du models)

The 520Du pump drive unit fitted with:

520R2 or other pumphead (see 8 Pump specifications) if specified as a pump

The designated mains power lead for your pump

PC-readable CDROM containing these operating instructions

Quick Start manual

Note: Some versions of this product will include components different from those listed above. Check against your purchase order.

Watson-Marlow 520Du User Manual

5

Storage

This product has an extended shelf life. However, care should be taken after storage to ensure that all parts function correctly. Users should be aware that the pump contains a battery with an unused life of seven years. Long-term storage is not recommended for peristaltic pump tubing. Please observe the storage recommendations and use-by dates which apply to tubing you may wish to bring into service after storage.

5 Information for returning pumps

Equipment which has been contaminated with, or exposed to, body fluids, toxic chemicals or any other substance hazardous to health must be decontaminated before it is returned to Watson-Marlow or its distributor.

A certificate included at the rear of these operating instructions, or signed statement, must be attached to the outside of the shipping carton. This certificate is required even if the pump is unused. See 39 Decontamination certificate.

If the pump has been used, the fluids that have been in contact with the pump and the cleaning procedure must be specified along with a statement that the equipment has been decontaminated.

Watson-Marlow 520Du User Manual

6

6 Peristaltic pumps - an overview

Peristaltic pumps are the simplest possible pump, with no valves, seals or glands to clog or corrode. The fluid contacts only the bore of a tube, eliminating the risk of the pump contaminating the fluid, or the fluid contaminating the pump. Peristaltic pumps can operate dry without risk.

How they work

A compressible tube is squeezed between a roller and a track on an arc of a circle, creating a seal at the point of contact. As the roller advances along the tube, the seal also advances. After the roller has passed, the tube returns to its original shape, creating a partial vacuum which is filled by fluid drawn from the inlet port.

Before the roller reaches the end of the track, a second roller compresses the tube at the start of the track, isolating a packet of fluid between the compression points. As the first roller leaves the track, the second continues to advance, expelling the packet of fluid through the pump’s discharge port. At the same time, a new partial vacuum is created behind the second roller into which more fluid is drawn from the inlet port.

Backflow and siphoning do not occur, and the pump effectively seals the tube when it is inactive. No valves are needed.

The principle may be demonstrated by squeezing a soft tube between thumb and finger and sliding it along: fluid is expelled from one end of the tube while more is drawn in at the other.

Animal digestive tracts function in a similar way.

Suitable applications

Peristaltic pumping is ideal for most fluids, including viscous, shear-sensitive, corrosive and abrasive fluids, and those containing suspended solids. They are especially useful for pumping operations where hygiene is important.

Peristaltic pumps operate on the positive displacement principle. They are particularly suitable for metering, dosing and dispensing applications. Pumps are easy to install, simple to operate and inexpensive to maintain.

Watson-Marlow 520Du User Manual

7

7 Safety notes

In the interests of safety, this pump and the tubing selected should only be used by competent, suitably trained personnel after they have read and understood this manual, and considered any hazard involved. If the pump is used in a manner not specified by Watson-Marlow Ltd, the protection provided by the pump may be impaired.

Any person who is involved in the installation or maintenance of this equipment should be fully competent to carry out the work. In the UK this person should also be familiar with the Health and Safety at Work Act 1974.

This symbol, used on the pump and in this manual, means: Caution, refer to accompanying documents.

This symbol, used on the pump and in this manual, means: Do not allow fingers to contact moving parts.

This symbol, used on the pump and in this manual, means: Recycle this product under the terms of the EU Waste Electrical and Electronic Equipment (WEEE) Directive.

Fundamental work with regard to lifting, transportation, installation, starting-up, maintenance and repair should be performed by qualified per-

sonnel only. The unit must be isolated from mains power while work is being carried out. The motor must be secured against accidental start-up.

There is a user-replaceable type T2,5A H 250V fuse in the fuseholder in the centre of the switchplate at the back of the pump. In some countries,

the mains power plug contains an additional replaceable fuse. There is a fuse on the interface card which self-resets after five seconds. There are no user-serviceable fuses or parts inside this pump.

Watson-Marlow 520Du User Manual

8

There are moving parts inside the pumphead. Before opening the toolunlockable pumphead guard, ensure that the following safety directions are followed.

Ensure that the pump is isolated from the mains power.

Ensure that there is no pressure in the pipeline.

If a tube failure has occurred, ensure that any fluid in the pumphead has been allowed to drain to a suitable vessel, container or drain.

Ensure that protective clothing and eye protection are worn if hazardous fluids are pumped.

Primary operator protection from rotating parts of the pump is provided by the pumphead safeguard. Note that safeguards differ, depending on the type of pumphead. See the pumphead section of this manual.

This pump must be used only for its intended purpose.

The pump must be accessible at all times to facilitate operation and maintenance. Access points must not be obstructed or blocked. Do not fit any devices to the drive unit other than those tested and approved by Watson-Marlow. Doing so could lead to injury to persons or damage to property for which no liability can be accepted.

If hazardous fluids are to be pumped, safety procedures specific to the particular fluid and application must be put in place to protect against injury to persons.

The exterior surfaces of the pump may get hot during operation. Do not take hold of the pump while it is running. Let it cool after use before handling it. The drive unit must not be run without a pumphead fitted.

This product does not comply with the ATEX directive and must not be used in explosive atmospheres.

Watson-Marlow 520Du User Manual

9

8 Pump specifications

A nameplate is fixed to the rear of the pump. It contains manufacturer and contact details, product reference number, serial number and model details.

520Du, IP31 model

This pump can be controlled from the keypad or remotely. It features:

Manual control

Speed adjustment; run and stop; direction control; “max” key for rapid priming.

Remote control

The pump can be digitally controlled with a contact closure or logic input signal to operate the pump.

Analogue control

The pump speed can be controlled through an analogue signal input in the ranges 0-10V or 4-20mA. Scaling can be controlled similarly using Analogue signal input 2.

Outputs

A 0-10V, 4-20mA or 0-1258Hz output signal provides feedback of the pump speed. There are four digital status outputs which can be configured in software for a variety of pump parameters.

MemoDose

Allows precise repeat dosing. Stores in memory a pulse count from the motor. This count is repeated each time START is pressed to provide a single-shot dose.

Calibration

Full calibration, plus default figures for a range of pumpheads and tubes. Calibration dose facility.

RS232 serial communications

Full pump control from a PC or other controller with the ability to network up to 16 pumps.

Watson-Marlow 520Du User Manual

10

IP (Ingress Protection) and NEMA definitions

 

IP

 

 

 

NEMA

 

1st Digit

 

2nd Digit

 

 

 

Protected against

 

 

 

 

 

ingress of solid objects

 

Protection against

 

 

 

with a diameter of

 

 

Indoor use to provide a

 

 

dripping water falling

 

 

more than 2.5mm.

 

 

degree of protection

3

1

vertically. No harmful

2

Tools, wires etc with a

against limited amounts

 

thickness of more than

 

effect must be

 

of falling water and dirt

 

 

produced

 

 

2.5mm are prevented

 

 

 

 

 

 

 

 

 

from approach

 

 

 

 

 

Protected against

 

 

 

 

Indoor use to provide a

 

harmful dust deposits.

 

 

 

 

degree of protection

 

Ingress of dust is not

 

 

Protection against

12

against dust, falling dirt

 

totally prevented but

 

 

water projected from a

 

and dripping, non-

 

the dust must not

 

 

nozzle against the

 

corrosive liquids

5

enter in sufficient

 

5

equipment (enclosure)

 

 

 

quantity to interfere

 

 

from any direction.

 

Indoor use to provide a

 

with satisfactory

 

 

There must be no

 

 

 

 

 

degree of protection

 

operation of the

 

 

harmful effect (water

 

 

 

 

13

against dust and

 

equipment. Complete

 

 

jet)

 

 

 

 

spraying water, oil and

 

protection against

 

 

 

 

 

 

 

 

 

non-corrosive coolants

 

contact

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Indoor or outdoor* use

 

 

 

 

 

 

to provide a degree of

 

 

 

 

Protection against

 

protection against

 

Protection against

 

 

heavy seas or powerful

 

splashing water, wind-

 

ingress of dust (dust-

 

 

water jets. Water must

 

blown dust and rain,

6

tight). Complete

 

6

not enter the

4X

hose-directed water;

 

protection against

 

 

equipment (enclosure)

 

undamaged by the

 

contact

 

 

in harmful quantities

 

formation of ice on the

 

 

 

 

(splashing over)

 

enclosure. (Resist

 

 

 

 

 

 

corrosion: 200-hour salt

 

 

 

 

 

 

spray)

 

 

 

 

 

 

 

* 520N cased pumps are rated to NEMA 4X (indoor use) only.

Watson-Marlow 520Du User Manual

11

Pump specifications

Control range (turndown ratio)

0.1-220rpm (2 200:1)

Supply voltage/frequency

100-120V/200-240V 50/60Hz 1ph

 

±10% of nominal voltage.

Maximum voltage fluctuation

A well regulated electrical mains supply

is required along with cable connections

 

 

to the best practice of noise immunity

Installation category

II

(overvoltage category)

 

Power consumption

135VA

Full load current

<0.6A at 230V; <1.25A at 115V

Eprom version

Accessible through pump software

 

IP31 to BS EN 60529. Equivalent to

 

NEMA 2, suiable for indoor use.

Enclosure rating

Protected against dripping water and

 

falling dirt. May be wiped with a damp

 

cloth, but should not be immersed

Pumphead options

520R, 501RL, 313, 314, 505L,

505BA, 505CA, 314MC, 318MC

 

Operating temperature range

5C to 40C, 41F to 104F

Storage temperature range

-40C to 70C, -40F to 158F

Maximum altitude

2,000m, 6,560ft

Humidity (non-condensing)

80% up to 31C, 88F, decreasing linearly

to 50% at 40C, 104F

 

Pollution degree

2

Noise

<70dB(A) at 1m

 

 

* Protect from prolonged UV exposure.

Note: 520 drive models are c ETL us listed. Cert to std CAN/CSA C22.2 No 101092. Conforms to std UL 61010A-1 April 30, 2002.

Note: 520 drive models have been tested in accordance with BS EN 61000-6- 2:2001 (EN 61000-4-4) Fast Transient and Burst Tests to Industrial limits - ie: Level 3 : 2kV.

Watson-Marlow 520Du User Manual

12

Standards

 

Safety of machinery—electrical equipment of machines:

 

BS EN 60204-1

 

Safety requirements for electrical equipment for measurement,

 

control and laboratory use:

 

BS EN 61010-1 incorporating A2 Category 2, Pollution degree 2

 

Degrees of protection provided by enclosures (IP code):

 

BS EN 60529 amendments 1 and 2

 

Conducted emissions:

 

BS EN 55011 A1 and A2 Class A, called by BS EN61000-6-4

 

Radiated emissions:

 

BS EN 55011 A1 and A2 Class A, called by BS EN61000-6-4

EC

Electrostatic discharge: BS EN 61000-4-2

Radiated RF immunity:

harmonised

standards

BS EN 61000-4-3 A1 and A2, called by BS EN 61000-6-2

 

Fast transient burst:

 

BS EN 61000-4-4 A1 and A2, Level 3 (2kV),

 

called by BS EN 61000-6-2

 

Surge immunity:

 

BS EN 61000-4-5 A1 and A2, called by BS EN 61000-6-2

 

Conducted RF immunity:

 

BS EN 61000-4-6, called by BS EN 61000-6-2

 

Voltage dips and interruptions:

 

BS EN 61000-4-11, called by BS EN 61000-6-2

 

Mains harmonics: BS EN 61000-3-2 A2

 

Pumps and pump units for liquids—common safety

 

requirements: BS EN 809

 

UL 61010A-1

Other

CAN/CSA-C22.2 No 61010-1

Conducted emissions FCC 47CFR, Part 15.107

standards

Radiated emissions FCC 47CFR, Part 15

 

 

NEMA 4X to NEMA 250 (indoor use) for IP66 products only

 

 

Watson-Marlow 520Du User Manual

13

8.1 Dimensions

520Du IP31 model

Unit weights

 

Drive only

+ 520R, 520R2

+ 520REL,

+ 505L

 

520REM, 520REH

IP31

9.70kg

10.60kg

10.52kg

12.18kg

21lb 6oz

23lb 5oz

23lb 3oz

26lb 14oz

 

Watson-Marlow 520Du User Manual

14

9 Good pump installation practice

9.1 General recommendations

A correctly engineered installation will promote long tube life. Site the pump on a flat, horizontal, rigid surface, free from excessive vibration, to ensure correct lubrication of the gearbox. Allow a free flow of air around the pump to ensure that heat can be dissipated. Ensure that the ambient temperature around the pump does not exceed 40C.

The STOP key on the keypad will always stop the pump. However, it is recommended that a suitable local emergency stop device is fitted into the mains supply to the pump.

Do not stack pumps more than three high. When pumps are stacked, ensure that the ambient temperature around all the pumps in the stack does not exceed 40C.

The pump may be set up so that the direction of rotor rotation is clockwise or count- er-clockwise, whichever is convenient. Please note, however, that for the 520R and 501RL pumpheads tube life will be greater if the rotor rotates clockwise; and that performance against pressure will be maximised if the rotor rotates counter-clock- wise. To achieve 4 bar and 7 bar pressures using a 520RE pump and the appropriate rotor and element, the pump must rotate counter-clockwise.

Peristaltic pumps are self-priming and self-sealing against backflow. No valves are required in inlet or discharge lines, except as described below. Valves in the process flow must be opened before the pump operates. Users are advised to fit a pressure relief device between the pump and any valve on the discharge side of the pump to protect against damage caused by accidental operation with the discharge valve closed. Users of 520RE pumps at pressures up to 4 bar and 7 bar are advised to fit a non-return valve between the pump and the discharge pipework to avoid the sudden release of pressurised fluid in the unlikely event of element failure.

9.2 Do’s and do not’s

Do not build a pump into a tight location without adequate airflow around the pump.

Do not strap the control and mains power cables together.

Watson-Marlow 520Du User Manual

15

Do keep delivery and suction tubes as short and direct as possible - though ideally not shorter than 1m - and follow the straightest route. Use bends of large radius: at least four times the tubing diameter. Ensure that connecting pipework and fittings are suitably rated to handle the predicted pipeline pressure. Avoid pipe reducers and lengths of smaller bore tubing than the pumphead section, particularly in pipelines on the suction side. Any valves in the pipeline (not usually needed with a self-prim- ing peristaltic pump) must not restrict the flow. Any valves in the flow line must be open when the pump is running.

Do use suction and delivery pipes equal to or larger than the bore of the tube in the pumphead. When pumping viscous fluids use pipe runs with a bore several times larger than the pump tube.

Do ensure that on longer tube runs at least 1m of smooth bore flexible tubing is connected to the inlet and discharge port of the pumphead to help to minimise impulse losses and pulsation in the pipeline. This is especially important with viscous fluids and when connecting to rigid pipework.

Do site the pump at or just below the level of the fluid to be pumped if possible. This will ensure flooded suction and maximum pumping efficiency.

Do keep the pumphead track and all moving parts clean and free from contamination and debris.

Do run at slow speed when pumping viscous fluids. When using the 520R pumphead, a 6.4mm or 4.8mm bore tube with a 2.4mm wall will give best results. Tube smaller than this will generate a high friction loss, so reducing the flow. Tube with a larger bore may not have sufficient strength to restitute fully. Flooded suction will enhance pumping performance in all cases, particularly for materials of a viscous nature.

Do recalibrate after changing pump tubes, fluid, or any connecting pipework. It is also recommended that the pump is recalibrated periodically to maintain accuracy.

IP31 models may be wiped with a wet cloth, but should not be hosed or immersed. The front of IP31 models is further protected against light spillages onto the pump.

When using Marprene or Bioprene continuous tubing, do re-tension the tube after the first 30 minutes of running.

Tube selection: The chemical compatibility lists published in Watson-Marlow publications are guides. If in doubt about the compatibility of a tube material and the duty fluid, request a Watson-Marlow tube sample card for immersion trials.

Watson-Marlow 520Du User Manual

16

10 Connecting this product to a power supply

Set the voltage selector to 115V for 100-120V 50/60Hz supplies or 230V for 200-240V 50/60Hz supplies. Always check the voltage selector switch before connecting the mains supply.

Make suitable connection to an earthed single-phase mains electricity supply.

We recommend using commercially available supply voltage surge suppression where there is excessive electrical noise.

Power cable: The pump is supplied fitted with a cable gland and approximately 2.8m of power cable. Recommended cable: H05RN-F3G0.75; SJTW 105C 3- 18AWG; max OD 8mm (5/16in).

This drawing shows the cable connections from below with the interface card cover removed.

Watson-Marlow 520Du User Manual

17

Watson-Marlow 520Du User Manual

If the mains power cable is inappropriate for your installation, it can be changed.Ensure that the pump is isolated from the mains power.

Remove the six screws from the interface card cover underneath the pump. Lift off the interface card cover. You may find it convenient to remove the cover completely; if so, remove the cover earth lead.

Undo the terminal block connectors. Remove the restraining clip by sliding its jaws sideways in opposite directions.

Loosen the cable gland using a 19mm wrench and remove the gland and the cable.

Thread a replacement cable through the three parts of the gland, the pump case and the restraining clip. Connect the new cable to the block connectors, following the drawing above.

Tighten the restraining clip and the gland to 2.5Nm. Check that the card cover earth link is secure. Replace the card cover, checking that the earth wire is not pinched beneath the cover lip. Take care that the sealing strip is properly seated to ensure a seal.

Watson-Marlow 520Du User Manual

18

Input line fusing: type T2,5A H 250V 20mm time-delayed fuse, located in a fuseholder in the centre of the switchplate at the rear of the pump.

Power interruption: This pump has an auto-restart feature which, when active, will restore the pump to the operating state it was in when power was lost. See 18.12 Auto-restart.

Stop / start power cycles: Do not power up/power down for more than 100 starts per hour, whether manually or by means of the auto-restart facility. We recommend remote control where a high frequency of power cycles is required.

Watson-Marlow 520Du User Manual

19

11 Start-up check list

Note: See also 29.2 Tube loading.

Ensure that proper connections are achieved between the pump tube and suction and discharge piping.

Ensure proper connection has been made to a suitable power supply.

Ensure that the recommendations in the section 9 Good pump installation practice are followed.

12 Switching the pump on for the first time

Note: This manual uses bold type to highlight the active option in menu screens: “English” in the first screen represented here. The active option appears on the pump display in inverse text.

Switch on the power supply at the rear of the pump. The pump runs a poweron test to confirm proper functioning of the memory and hardware. If a fault is found, an error message is displayed. See 25.1 Error codes.

The pump displays a language menu. Use the UP and DOWN keys to select your language. Press the ENTER key to confirm your choice.

The information which follows assumes that your choice was English. When the language is chosen this menu will not appear again and all menus will appear in the language you chose. (Language can be reset as described later. See 18.16 Language.)

The pump displays the Watson-Marlow start-up screen for four seconds, followed by the pump model identity screen for four seconds, and then the manual mode main screen.

The rotation symbol on the display indicates clockwise rotation. The speed of rotation is the pump’s maximum. Other initial start-up operational parameters are listed in the table below.

Watson-Marlow 520Du User Manual

20

First-time start-up defaults

Language

Not set

 

Analogue inputs

4-20mA

Speed

Maximum

 

User trim

None

Direction

Clockwise

 

Remote stop

Open=run

Pumphead

520R

 

Pump number

1

Tube size

9.6mm

 

Baud rate

9600

Calibration

from head and tube

 

Stop bits

2

 

 

 

table

 

Xon/Xoff

Off

 

 

 

 

 

Backlight

On

 

Output 1

Run/Stop *

Keypad lock

Off

 

Output 2

Direction †

Auto-restart

Off

 

Output 3

Auto/Man ‡

Scaling

Off

 

Output 4

General alarm

 

 

 

 

 

Pump status

Stopped

 

Run*

= High

Beeper

On

 

Clockwise rotation †

= High

Manual

rpm

 

Auto ‡

= High

screen

 

 

 

 

 

Security

Not set

 

 

 

code

 

 

 

 

 

 

 

 

 

 

 

 

Note: The settings shown above for Run, Clockwise rotation and Auto are those in force on initial start-up for the functions available on Output 1, Output 2 and Output 3 respectively. For example, a high signal on Output 2 indicates clockwise rotation. These can be exchanged later according to user requirements.

The pump is now ready to operate according to the defaults listed above.

All operating parameters may be changed by means of key-presses. See 14 Manual operation.

Watson-Marlow 520Du User Manual

21

13 Switching the pump on in subsequent power cycles (if not in auto-restart mode)

Switch on the power supply at the rear of the pump. The pump runs a poweron test to confirm proper functioning of the memory and hardware. If a fault is found, an error message is displayed. See 25.1 Error codes.

The pump displays the Watson-Marlow start-up screen for four seconds followed by the pump model identity screen for four seconds, and then the manual mode main screen.

Note: If ANY key is pressed during the display of any of the preliminary screens, the display jumps to the next screen. Quickly pressing any two keys or any key twice immediately after switch-on causes the display to jump to the manual mode main screen. Once in the manual mode main screen, keys assume their normal functions - see 14.1 Keypad functions in manual mode below. A subsequent press on START causes the pump to operate.

Start-up defaults are those in place when the pump was switched off last. Check that the pump is set to operate as you require it.

The pump is now ready to operate.

All operating parameters may be changed by means of key-presses. See 14 Manual operation.

Watson-Marlow 520Du User Manual

22

14 Manual operation

14.1 Keypad functions in manual mode

All settings and functions of the pump in manual mode are set and controlled by means of key-presses. Immediately after the start-up display sequence detailed above, the manual mode main screen will be displayed. The currently selected rotation direction is indicated on the display by a clockwise or count- er-clockwise segmented arrow. If an exclamation mark ( ! ) shows, it indicates that Auto-restart is on (see 18.12 Auto-restart). If a padlock icon ( ) shows, it indicates that Keypad lock is on (see 14.2 Keypad lock).

Note: A number of the controls listed below are shortcuts to commands which are also available through the Main menu. See 15 Main menu.

A brief single press on each key triggers a beep sound (if enabled - see 18.18 Beep) and causes the pump to function as follows:

START: starts the pump at the speed and in the direction shown on the display. The rotation symbol will become animated to confirm that the pump is operating. We recommend that the speed is reduced to a minimum (0.1 rpm) before starting the pump.

If the pump is running when START is pressed, it causes the information shown on the manual mode main screen to cycle from revolutions per minute, to flowrate in a choice of units (via a warning screen if flowrate has not been calibrated and if this is the first cycle since power-up) to rpm, flowrate and Run time. An example is shown here. (See 17 Calibrate. See 18.9 Run time.) This cycle operates in a similar way if ENTER is pressed when the pump is stopped. The default can be altered from within the Setup menu. See 18.3 Display.

Watson-Marlow 520Du User Manual

23

MAX: while pressed, MAX operates the pump at the maximum allowed speed and in the direction shown on the display. When released, the pump returns to its previous status.

Note: Priming can be achieved by pressing the MAX key until fluid flows through the pump and reaches the point of discharge, and then releasing the MAX key.

AUTO/MAN: displays a screen offering three choices: manual control, analogue control or network (RS232) control. Use the UP and DOWN keys to make a choice. Press the ENTER key to confirm your decision. When started under analogue or network control, the pump runs at the speed set by any analogue or RS232 signal applied to the pump, and in the direction shown in the display.

STOP: has no effect if the pump is not running. If the pump is running, pressing STOP stops the pump. The display will continue to show the previous speed and direction. The pump will return to this speed and direction when the START key is pressed again.

UP: increases the speed shown on the display in minimum steps of 0.1 rpm (unless the speed displayed is already the maximum allowed speed). If the pump is then started by pressing the START key, it will operate at the new speed. If the pump is running when UP in pressed, the change takes effect immediately.

Note: After a speed change (if the pump has been calibrated), a screen showing the new rpm figure and the new flowrate is displayed for four seconds before returning the user to the previously set manual mode main screen: rpm or flowrate.

DOWN: decreases the speed shown on the display in minimum steps of 0.1 rpm. If the pump is then started by pressing the START key, it operates at the new speed. The minimum speed possible is 0.1 rpm. If the pump is running when DOWN is pressed, the change takes effect immediately.

Note: After a speed change (if the pump has been calibrated), a screen showing the new rpm figure and the new flowrate is displayed for four seconds before returning the user to the previously set manual mode main screen: rpm or flowrate.

Note: You can reduce the pump speed from 0.1 rpm to 0 rpm by a further press on the DOWN key. The pump is still in the running state and the rotation symbol will continue to move. Press the UP key to return the pump to the minimum speed.

Watson-Marlow 520Du User Manual

24

DIRECTION: toggles the direction of rotation shown on the display. If the pump is then started by pressing the START key, it rotates in the new direction. If the pump is running when DIRECTION is pressed, the change takes effect immediately.

ENTER: cycles the information shown on the manual mode main screen from revolutions per minute, to flowrate in a choice of units (via a warning screen if flowrate has not been calibrated and if this is the first cycle since power-up) to rpm, flowrate and Run time. (See START, above. See 17 Calibrate. See 18.9 Run time.) This cycle operates when the pump is stopped and when it is running. The default can be altered from within the Setup menu. See 18.3 Display.

MENU: causes the main menu to be displayed, from which all aspects of pump setup can be controlled. See 15 Main menu.

Keypress combinations cause the pump to function as follows:

UP and DIRECTION on power-up: toggles the keypad beep on and off.

START on power-up: switches on the Auto-restart facility. See 18.12 Autorestart.

STOP on power-up: switches off the Auto-restart facility. See 18.12 Autorestart.

STOP and DIRECTION on power-up: allows the user to press UP and DOWN keys to toggle the sense of remote run / stop control between open=stop and open=run.

STOP and UP while the pump is stopped: turns the display backlight on.

STOP and DOWN while the pump is stopped: turns the display backlight off.

MAX and UP: sets the pump to maximum allowed speed.

MAX and DOWN: sets the pump to minimum speed.

DIRECTION and DOWN: interrupts the display to show the pump’s ROM version for four seconds.

START pressed and held for two seconds: toggles the keypad lock on and off. Only the START and STOP keys are active when keypad lock is on. The padlock icon is displayed.

STOP pressed and held for two seconds: toggles the keypad lock on and off. Only the START and STOP keys are active when keypad lock is on. The padlock icon is displayed.

STOP STOP within half a second: shortcut entry to MemoDose; when in MemoDose, shortcut return to manual mode main screen. See 19 MemoDose.

Note: The maximum allowed speed of the drive defaults to 220 rpm. It is possible to set this limit at any speed up to this value. See 18.13 Set maximum speed.

Watson-Marlow 520Du User Manual

25

14.2 Keypad lock

The keypad can be locked to prevent changes to pump speed or other settings, and make it possible only to start or stop the pump. The padlock symbol shows on the display.

While the pump is running, hold down the START key for two seconds. The padlock symbol shows and only the START and STOP keys function.

The keypad may also be locked while the pump is stopped. Hold down the STOP key for two seconds. The padlock symbol shows and only the START and STOP keys function.

To unlock the keypad while the pump is running hold down the START key for two seconds. The padlock symbol is removed. If the pump is stopped hold down the STOP key until the padlock symbol is removed.

14.3 Keypad beep

The pump keypad can operate silently or indicate a positive key-press with a beep sound.

To toggle the sound on and off, stop the pump. Turn off the mains power switch at the rear of the pump.

Depress the UP and DIRECTION keys while switching on the mains power switch at the rear of the pump.

14.4 Manual operation and remote digital inputs and outputs

The remote run / stop, direction and leak-detected inputs are operational.

The remote status outputs are all fully functional.

The STOP key acts as an overriding emergency stop. The run / stop input will not start the pump in manual mode, but once the START key has been pressed, the remote run / stop input will stop and start the pump according to its operational state.

If you invert the operation of the remote run / stop switch to operate as open=stop, you must connect pin 7 to 19, lower D-connector, to be able to start the pump from the keypad.

If STOP is pressed the remote run / stop switch will have no effect.

You cannot invert the remote direction signal.

Watson-Marlow 520DuN IP66/NEMA 4X User Manual

26

15 Main menu

15.1 Keypad functions in menu screens

In addition to their functions in other operations, the following keys have specific actions in menu screens:

STOP: In general, STOP functions as a “go back” key, taking the user up one menu level without making a change.

UP: The UP key is used in menu item selection: it moves a highlight up the menu. When a numerical entry screen is displayed, pressing UP increases the number displayed.

DOWN: The DOWN key is used in menu item selection: it moves a highlight down a menu. When a numerical entry screen is displayed, pressing DOWN decreases the number displayed.

ENTER: The ENTER key functions in a similar way to the “enter” key of a personal computer: it confirms key-presses made immediately before. In menu item selection, it triggers the action or display

selected from a menu using the UP and DOWN keys.

Note: Confirmation screens are displayed for 4 seconds. While they are displayed, a single press on any key removes them.

Watson-Marlow 520DuN IP66/NEMA 4X User Manual

27

15.2 Main menu entry

The MENU key displays the main menu. It operates at any point in the pump’s activity except where error screens are displayed, or screens where UP and DOWN keys are used to enter values.

The main menu offers five options: Calibrate, Setup, Pin out details, MemoDose and Exit. Use the UP and DOWN keys to make a choice. Press the ENTER key to confirm your decision.

Calibrate

Calibrate allows the user to calibrate the pump with default figures for a range of pumpheads and tubes, as well as to refine the flowrate figures with a calibration dose facility.

Setup

Setup allows the user to set the pump’s operating parameters under 20 headings:

Trim, Analogue, Display, Pump number, Baud, Stop bits, Xon/Xoff, Flow units, Run time, Outputs, Remote stop, Auto-restart, Set max speed, Backlight, ROM, Language, Defaults, Beep, Security and Exit.

Pin out details

Selecting Pin out details causes the pump to display a warning screen and then its preset pin and voltage details under 18 headings: Analogue input 1, Analogue input 2, Analogue output, Tacho output, Direction enable, Direction input, Run/Stop input, Leak input, Auto/Man toggle, Dose input, Output 1, Output 2, Output 3, Output 4, Supply, 0 volts, Earth and Others.

MemoDose

The MemoDose facility is used to remember the number of revolutions needed to dispense a set volume of fluid, and cause the pump to dispense that volume repeatedly.

Exit

If Exit is selected, the pump returns to its last manual state with the pump stopped.

Watson-Marlow 520DuN IP66/NEMA 4X User Manual

28

16 PIN-secure process protection

The 520DuN and 520Du feature PIN-secure process protection. This allows the pump to be configured to suit the application, and for the setup to be protected by two levels of PIN code. See 18.19 Security code setup.

Menu option

With Main

With User

Code set and

or keypress

code set

code set

keypad locked

Menu

 

 

 

Available

Available

Not available

Calibrate

Available

Available

Not available

Accept

Available

Available

Not available

 

 

 

 

Change

Main PIN

Main PIN or User

Not available

needed

PIN needed

 

 

Setup

Main PIN

Main PIN needed;

Not available

needed

User PIN invalid

 

 

MemoDose

Available

Available

Available*

 

 

 

 

Pin out details

Available

Available

Not available

Max

Available

Available

Not available

 

 

 

 

Dir

Main PIN

Main PIN or User

Not available

needed

PIN needed

 

 

Auto / Man

Available

Available

Not available

Up

Available

Available

Not available

Down

Available

Available

Not available

Max and Up

Available

Available

Not available

Max and Down

Available

Available

Not available

 

 

 

 

Start

Available

Available

Available

Stop

Available

Available

Available

 

 

 

 

Enter

Available

Available

Not available

 

 

 

 

Keypad lock

Main PIN

Main PIN or

Main PIN or

needed

User PIN needed

User PIN needed

 

Auto-restart

Available

Available

Available

 

 

 

 

* When the keypad is locked, MemoDose is available via its access shortcut: press STOP twice; it is not available through the menu structure.

The main code allows changes to Calibrate, Setup, Direction and Keypad lock.

The secondary (User) code permits Calibrate, Direction change and Keypad lock but bars Setup change.

If either code is used in conjunction with Keypad lock, all keys are disabled except STOP and START.

To activate and set a security code, see 18.19 Security code setup.

If the main code is lost or forgotten: The Setup main security code can be bypassed by entering a special key sequence; all codes can then be cancelled and reset. Contact Watson-Marlow or your distributor for details.

Watson-Marlow 520Du User Manual

29

17 Calibrate

The pump can display flowrate in a choice of units as well as speed in revolutions per minute. It must first be calibrated.

Pump calibration can be limited to users who correctly enter a three-digit security code. If a security code has been set, selecting Change from the calibration details screen and confirming with the ENTER key causes the pump to display the Security code entry sequence. See 16 PIN-secure process protection. If no security code has been set, the pump displays the first screen of the calibration sequence.

To dispense the correct amount of fluid the pump must know which pumphead is fitted and the tube size in the pumphead. The pump is pre-programmed with default flow rate figures for a range of pumpheads and tubes. You may also measure the flow from the pump and enter this value for the most accurate calibration.

Use the UP and DOWN keys to highlight Calibrate in the main menu. Press ENTER to confirm.

The pump displays its current settings: pumphead, tube size and flowrate; the default for each standard pumphead is the largest tube the pumphead can take. Example information is shown here.

Use the UP and DOWN keys to highlight Accept or Change. Press ENTER to confirm.

If Accept is selected, the default or previously set flowrate data for that pumphead and tube size are used in flowrate calculations. The pump redisplays the main menu.

If Change is selected with a security code in place, the pump starts its security code entry sequence. See 16 Security code entry. When the correct code is entered, the pump offers three options: Head and tube; Tube; and Calibration dose. If Change is selected with no security code in place, the three options are displayed immediately.

Use the UP and DOWN keys to make a selection. Press ENTER to confirm.

Watson-Marlow 520Du User Manual

30

Loading...
+ 77 hidden pages