Sirona ProFeel+ Maintenance manual

kЙп=лбеЕЙW=
H
mêçcÉÉä
j~бенЙе~еЕЙ=j~ем~д
NNKOMMQ
ProFeel
Model Serial number Chair
+
bеЦдблЬ
E
N
L
PE
~
MHP
AP
2 D 3474.102.01.01.02 11.2004
60 32 317 D 3474
Table of Contents
1 General information ............................................................... 5
1.1 Purpose of the Maintenance Manual ..............................5
1.2 Work to be performed .....................................................5
2 Installation Report / Warranty Passport ................................. 6
2.1 Master data of the unit ....................................................6
2.2 Inspection and maintenance ...........................................6
3 Safety checks ........................................................................ 7
3.1 Visual inspection .............................................................9
3.2 Protective ground wire test ProFeel
3.3 Measurement of equivalent leakage currents ...............11
3.3.1 Equivalent unit leakage current ............................................ 13
3.3.2 Equivalent patient leakage current ....................................... 14
3.4 Safety check (Initial test after initial start-up) ................15
3.5 Safety check (re-tests) ..................................................15
+
............................10
4 Remarks / particularities regarding the treatment center ..... 19
aÉìíëÅÜ
60 32 317 D 3474 D 3474.102.01.01.02 11.2004
3
4 D 3474.102.01.01.02 11.2004
60 32 317 D 3474
1 General information

1.1 Purpose of the Maintenance Manual

1 General information

In order to guarantee the operational safety and reliability of the system and to protect the health of patients, users and other persons, inspection and maintenance must be performed at predetermined intervals.
This includes:
Inspection and maintenance (yearly) to prevent damage due to natural wear
Safety tests (every 2 years) to ensure the technical safety of the system
This document describes the work to be performed by the service engineer. Its realization and the measurement results are documented by the ser­vice engineer. This document must be stored near the treatment center.

1.2 Work to be performed

By the service engineer: 1. Note the model and the serial number of the chair on the title page and the
relevant pages (headers) of the Maintenance Manual.
2. Complete the “Installation Report / Warranty Passport” and file it after chapter 2.
3. Perform inspection and maintenance according to the Maintenance Certificate. Document their implementation in the “Installation Report / Warranty Passport".
4. Conduct the safety tests in accordance with chapter 3. Document the results.
bеЦдблЬ
60 32 317 D 3474 D 3474.102.01.01.02 11.2004
5

2 Installation Report / Warranty Passport

2 Installation Report /
Warranty Passport

2.1 Master data of the unit

Complete the document “Installation Report / Warranty Passport” and file the “Customer Copy” after this page.
Unusual occurrences during installation can be noted down in addition on the second page of the “Dealer Copy”.

2.2 Inspection and maintenance

To avoid damage due to natural wear, an inspection must be performed every year.
The steps to be performed as well as the parts which must be replaced are specified in the document “Maintenance Certificate”. Their realization is documented there.
A separate Maintenance Certificate is produced for each maintenance event.
List the inspection and maintenance events also under the maintenance over­view in the “Installation Report / Warranty Passport”.