Installation and startup may be carried out only by personnel specifically
authorized by Sirona.
Safety
WARNING
Perilous shock hazard!
You must switch off the unit and then wait at least 1 minute, for
measurements at the tube assembly at least 4 minutes, before you start
to connect test cables or remove a cover!
For measurements in the area of the power supply terminal, the unit
must be disconnected from the junction box of the building installation
before you start to connect test cables!
DANGER
X-rays
When performing the following tests, be sure to observe the radiation
protection regulations applicable in your country (see Operating
Instructions).
DANGER
X-rays
"Radiation" is signaled by the message "X-RAY active!", a beep, and an
X-ray LED.
NOTICE
Risk of damage to boards
Please observe the usual precautionary measures for handling printed
circuit boards (ESD). Touch a ground point to discharge static electricity
before touching any boards.
IMPORTANT
It is essential that you also observe the notes about the operation of the
unit in the Operating Instructions.
4D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH2 Operation notes
Maintenance Instructions GALILEOS Comfort PLUS
Operation notes
2
Operation notes
Nominal line voltageThe volume tomograph operates in the following nominal line voltage
ranges:
● 200 – 240 V
● 50 / 60 Hz
The permissible line voltage fluctuation is ±10%.
The internal line impedance must not exceed 0.8 Ω.
Only permanent electrical connection of the unit is allowed.
IMPORTANT
The regulation "Federal Performance Standard for Diagnostic X-ray
Units, Code of Federal Regulations, Title 21 CFR, Subchapter J"
requires a corresponding power supply connection.
Cooling periodThe cooling period between two exposures is maintained by an automatic
exposure blocking function according to the pulse/pause ratio (see
Operating Instructions). The decrementing waiting time count is displayed
on the control panel.
MeasurementsAlways switch the unit off before connecting a measuring instrument.
bеЦдблЬ
For safety reasons, the power supply should be switched off at the
junction box of the building installation when performing measurements
in the vicinity of the power supply unit.
Select the correct current/voltage type and adjust the measuring range to
match the expected readings.
Perform continuity tests only on units which are switched off.
If several exposures with radiation must be taken to check a
measurement, make sure that the prescribed cool-down intervals are
observed. They are maintained by an automatic exposure blocking
function (see Operating Instructions).
The pulse/pause ratio is 1:20, i.e. a 20-second pause is maintained for
each second of radiation cycle. The pulse/pause ratio is automatically
maintained (automatic exposure blocking).
It is essential that you observe the radiation protection regulations
applicable in your country prior to radiation release.
64 05 521 D3437
D3437.103.02.02.02 03.2013
5
3 Auxiliary devices requiredSirona Dental Systems GmbH
Maintenance Instructions GALILEOS Comfort PLUS
Auxiliary devices required
3
Auxiliary devices required
CAUTION
Only use a battery-operated digital multimeter with safety sockets.
It is essential that you observe the safety and operating notes provided
in the operating instructions of the multimeter.
● Battery-operated digital multimeter of type:
–Fluke 8000 A
–Philips PM 2816 rms
–or similar
Dosimeter USA
CAUTION
It is essential that you observe the safety and operating notes provided
in the operating instructions of the dosimeter.
● Dosimeter for pulsed radiation of type:
–Mult-O-Meter 512L
–or similar
Measuring wires
CAUTION
Use exclusively fully insulated measuring wires.
Check the measuring wires for damages before use.
● Measuring wires with the following properties:
–dielectric strength > 1000V
6D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH4 Visual check
Maintenance Instructions GALILEOS Comfort PLUS
Visual check
4
Visual check
● Check for mechanical damage that could affect radiation protection.
● Check that all labels are attached and legible.
bеЦдблЬ
–Damaged/imperfect labels must be replaced.
You can order them in writing from Sirona (see back page for
address),
stating the following:
–Customer's name
–Customer's address
● Check to see whether all model numbers with serial numbers are
attached and are still legible on the device for the purpose of
identification.
For serial numbers see also Installation Report / Warranty Passport.
64 05 521 D3437
D3437.103.02.02.02 03.2013
7
5 Optical and acoustic signalsSirona Dental Systems GmbH
A
B
C
Maintenance Instructions GALILEOS Comfort PLUS
Optical and acoustic signals
5
Optical and acoustic signals
Optical and acoustic signals, Galileos
1. Turn the main switch (A) to position I.
ª The X-Ray radiation indicator C lights up briefly.
ª After approx. 2 seconds, the green LED B in the upper part of the
control panel lights up. This LED remains lit as long as the unit is
on.
ª The start screen appears on the touchscreen of the Easypad and
the initialization of the device starts running (for approx. 1
minute). The rotating element rotates briefly clockwise and
counterclockwise.
2. Check the function of the keys.
ª Press a height-adjustment key. The movement of the height-
adjustment motor is accompanied by an acoustic signal.
ª If Remote has been installed, test the buttons and the display of
the remote unit, too.
ª The release button must not be defective or damaged.
See also the Operating Instructions, in the Operation section.
64 05 521 D3437
8D3437.103.02.02.02 03.2013
Sirona Dental Systems GmbH5 Optical and acoustic signals
T
R
B
A
C
?
Exposure is performedExposure is performed
D
F
E
Maintenance Instructions GALILEOS Comfort PLUS 5.1 Checking the function of keys and buttons
5.1
Checking the function of keys and buttons
Radiation protection measures
Checking the function of keys and buttons
WARNING
The unit emits X-ray radiation.
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
✔ "Unit ON" LED display C lights up. Press the R key D to move the unit
back to the starting position. As long as no connection has been
made to SIDEXIS XG, the message is displayed in the comment line
of the control panel on the
state"
touchscreen.
1. Switch on the PC.
2. Start SIDEXIS XG.
For further information and possible error messages, see Operating
Instructions.
3. Use the -/+ keys to select a program E.
4. In the sub menu F, select a kV/mA combination with the -/+ keys.
5. Check whether the patient symbols on the touchscreen can be
selected in exactly the right position.
If problems occur during selection, adjust the touchscreen ().
6. Press release button (A) and hold it down until the end of the
exposure.
ª The exposure is released.
the comment line on the touchscreen.
During radiation, the optical radiation indicator (B) lights up on the
Easypad; during radiation, an acoustic signal is also emitted.
Radiation can be triggered multiple times during exposure.
When the rotation and radiation switch off automatically, the
exposure is complete.
"Switch SIDEXIS to ready for exposure
"Exposure is performed"
appears in
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
7. Let go of release button (A).
ª The exposure is completed.
9
5 Optical and acoustic signalsSirona Dental Systems GmbH
5.1 Checking the function of keys and buttonsMaintenance Instructions GALILEOS Comfort PLUS
8. Cancel exposure - Check deadman function
Select the same exposure parameters.
The operational readiness LED B flashes until the automatic coolingoff period of the X-ray tube assembly has expired (automatic
exposure block).
9. Press the release button A.
10. Let go of the exposure release button.
ª The exposure is immediately terminated.
The confirmation of the exposure data is displayed on the
touchscreen. The radiation time and the area dose product (the
lower two values) flash. In this way, an exposure that has been
triggered can be canceled again at any time.
CAUTION
Defective indicators and keys represent a risk to the safety of both the
patient and the operator.
The user is not permitted to operate the unit until the necessary repairs
have been made.
10D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH6 Checking the power supply connection
K1
K1
300V
AC
Maintenance Instructions GALILEOS Comfort PLUS
Checking the power supply connection
6
Checking the power supp ly connection
To check the line voltage, the
creating an X-ray. To do this, proceed as follows:
Preparing the measurement1. DANGER! After disconnecting the unit from the junction box of the
building installation, wait at least 1 minute before starting to check the
line voltage!
Disconnect the unit from the junction box of the building installation.
2. Remove the "Bottom profile" cover (see Service Manual).
line voltage drop
must be determined while
3. CAUTION! Only use fully insulated measuring wires.
Connect the measuring wires as shown in the illustration to the
connectors L and N of the power supply terminal K1.
4. On the multimeter, select the voltage measuring range "300 V
5. Re-attach the unit to the junction box of the building installation.
6. DANGER! Do not touch any live components!
Set the main switch (A) to I (see also Operating Instructions).
7. Wait for approx. 1 minute.
8. Press the R key.
ª The unit moves to its starting position.
AC
bеЦдблЬ
".
64 05 521 D3437
D3437.103.02.02.02 03.2013
11
6 Checking the power supply connectionSirona Dental Systems GmbH
Maintenance Instructions GALILEOS Comfort PLUS
Performing and analyzing a
measurement
1. Set the highest kV/mA level, e.g. 98kV /12mAs (see Operating
Instructions).
2. Make the SIDEXIS XG ready for exposure.
3. CAUTION! Activating the release button triggers X-rays.
Activate the release button and take the voltage drop reading on the
multimeter display.
ª If the measured voltage drop does
tolerance range
a suitable line voltage (according to the notes listed in the
Installation Requirements) must be installed.
(see the following table), notify the customer that
not fall within the permissible
IMPORTANT
In such a case, the unit must be switched off immediately and
disconnected from the junction box of the building installation. It must
not be placed in operation!
ª If the measured voltage drop falls within the permissible
tolerance range (see the following table), finalize the
measurement.
Permissible voltage drop:
Line voltage, with zero loadMax. permissible
line voltage drop
180-208 V9 V
208-230 V8 V
230-240 V7.5 V
240-264 V7 V
Concluding the measurement1. Switch the unit on via switch (A) (see also Operating Instructions).
2. Disconnect the unit from the junction box of the building installation.
3. DANGER! Wait at least 1 minute after disconnecting the unit from the
junction box of the building installation before removing the
measuring wires!
Remove the measuring units from the unit.
4. Re-attach the "Bottom profile" cover to the unit.
12D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH7 Checking the tube current
E
D
D
D
F
G
Maintenance Instructions GALILEOS Comfort PLUS
Checking the tube current
7
Checking the tube current
NOTICE
Damage to the measuring unit
The ring assembly and the tube assembly move during the
measurement.
Make sure that the measuring wires are sufficiently long to allow for the
ring movement and that the measuring unit is in a secure position so that
it will not fall down.
Preparing the measurement1. Switch off the unit (see Operating Instructions).
DANGER! After switching off the unit, wait at least 4 minutes (LED
V500 on the DX6 must no longer be on) before removing the cover
on the tube assembly.
2. Loosen the 3 screws (D) and remove the lid of the tube assembly
cover (E).
3. Loosen the 4 screws (F) and remove the cover plate (G).
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
13
7 Checking the tube currentSirona Dental Systems GmbH
Maintenance Instructions GALILEOS Comfort PLUS
14D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH7 Checking the tube current
E
D
Maintenance Instructions GALILEOS Comfort PLUS
4. DANGER! After switching off the unit, wait at least 4 minutes (LED
V500 on the DX6 must no longer be on) before removing jumper H
from the DX6 board.
Remove the jumper (H) from connector X302 on the DX6 board.
5. DANGER! Only use fully insulated measuring wires.
Connect the digital multimeter with the measuring wires to test points
MA- (X5-) and MA+ (X4+) at connector X302 on the DX6 board.
6. On the multimeter, select the current measuring range 20mA DC.
7. NOTICE! If the lid of the tube assembly cover is not attached, the ring
circulation is impeded and the unit can be damaged.
Temporarily install the lid of the tube assembly cover (E) on the unit
using the screw (D).
8. DANGER! Do not touch any live components!
Set the main switch (A) to I (see also Operating Instructions).
9. Wait for approx. 1 minute.
10. Press the R key.
ª The unit moves to its starting position.
Performing measurements, Galileos
Performing measurements1. Call the Service menu and the Service routine S002.5 (see Service
Manual).
2. Select kV/mA level 85 kV/ 7 mA and 4 seconds of radiation time (see
Service Manual).
3. CAUTION! Activating the release button triggers X-rays.
Release the radiation. Hold the release button pressed until the set
radiation time has expired.
IMPORTANT
1mA corresponds to a tube current of 1mA. The permissible tolerance
is ± 20 %.
Analyzing measurements➢ Read the voltage value on the display of the multimeter.
Evaluating measurements, Galileos
ª The tube current must be in the range of 7mA ± 1.4mA.
ª If the measured value does
tolerance, replace the
ª If the measured value falls within the permissible tolerance,
conclude the measurement.
not
fall within the permissible
tube assembly
(see Service Manual).
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
15
7 Checking the tube currentSirona Dental Systems GmbH
E
D
F
G
E
D
D
D
Maintenance Instructions GALILEOS Comfort PLUS
Concluding the measurement1. Switch the unit on via switch (A) (see also Operating Instructions).
2. Loosen the screw (D) and remove the lid of the tube assembly cover
(E).
3. DANGER! After switching off the unit, wait at least 4 minutes before
removing the measuring wires or reinserting the jumper!
Remove the measuring wires and bridge with the test points MA+/
MA- on the DX6 board again with the jumper (H).
4. Reattach the cover plate (G) to the tube assembly with the 4 screws
(F).
5. Reattach the lid of the tube assembly cover (E) to the unit and secure
it with the 3 screws (D).
64 05 521 D3437
16D3437.103.02.02.02 03.2013
Sirona Dental Systems GmbH8 Checking the tube voltage
A
Maintenance Instructions GALILEOS Comfort PLUS
Checking the tube voltage
8
Checking the tube voltage
Preparing the measurement1. Attach the Mult-O-Meter sensor in the middle of the X-ray detector.
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
ª The unit moves to its starting position.
Performing measurements1. Call the Service menu and the Service routine S002.5 (see Service
Manual).
2. Select kV/mA level 85 kV/ 7 mA and 4 seconds of radiation time (see
Service Manual).
3. CAUTION! Activating the release button triggers X-rays.
Release radiation. Hold the release button pressed until the set
radiation time has expired.
bеЦдблЬ
Analyzing measurements➢ Read the voltage values on the display of the Mult-O-Meter.
IMPORTANT
The measured tube voltage must correspond with the tube voltage set
of 85kV. The permissible tolerance is ± 10 %.
ª If the measured values
, replace the
range
ª If the measured values are within the permissible tolerance
range, finalize the measurement.
Concluding the measurement1. Exit the service routine.
2. Switch the unit on via switch (A) (see also Operating Instructions).
do not fall within the permissible tolerance
tube assembly
(see Service Manual).
64 05 521 D3437
D3437.103.02.02.02 03.2013
17
9 Checking the radiation timeSirona Dental Systems GmbH
A
Maintenance Instructions GALILEOS Comfort PLUS
Checking the radiation time
9
Checking the radiation time
Preparing the measurement1. Attach the Mult-O-Meter sensor in the middle of the X-ray detector.
2. Set the main switch (A) to I (see also Operating Instructions).
3. Wait for approx. 1 minute.
4. Press the R key.
ª The unit moves to its starting position.
Performing measurements1. Call the Service menu and the Service routine S002.5 (see Service
Manual).
2. Select kV/mA level 85 kV/ 7 mA and 4 seconds of radiation time (see
Service Manual).
3. CAUTION! Activating the release button triggers X-rays.
Release radiation. Hold the release button pressed until the set
radiation time has expired.
Analyzing measurements➢ Read the radiation time on the Mult-O-Meter.
ª The value displayed on the Mult-O-Meter for the radiation time
must be 4s. The permissible tolerance is ± 10 %.
ª If the measured radiation time does
tolerance, replace the
ª If the measured radiation time falls within the permissible
tolerance, finalize the measurement.
Concluding the measurement1. Exit the service routine.
2. Switch the unit on via switch (A) (see also Operating Instructions).
tube assembly
not
fall within the permissible
(see Service Manual).
18D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH10 Calculating / checking the mAs value
Maintenance Instructions GALILEOS Comfort PLUS
10
Calculating / checking the mAs value
Calculating / checking the mAs v alue
Calculate the mAs value with the following formula:
● mAs = mA measurement x time measurement
= result of Checking the tube current [ → 13] x result of Checking the
radiation time [ → 18]
= 28 mAs ± 2.8 mAs
The calculated value must be ± 10% at 28 mAs.
NOTICE
If the calculated value is not within the permitted tolerance, see the
Service Manual.
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
19
11 Checking the laser light localizersSirona Dental Systems GmbH
A
B
Maintenance Instructions GALILEOS Comfort PLUS
11
Checking the vertical laser beam (MS)
Checking the laser light localizers
Checking the laser light localizers
CAUTION
Risk of injury to eyes.
The unit contains lasers of Class 1.
Keep a distance of at least 4" (10 cm) between eye and laser. Do not
look into the laser beam.
1. Set the bite block (A) in the bite block holder, see Operating
Instructions.
2. Press the light localizer key (B) on the control panel.
ª The light localizer is switched on.
ª The vertical laser beam must be displayed in the center of the bite
block and the bite block holder. If this is not the case, adjust the
laser light localizer, see Service Manual.
3. Switch the light localizers off.
64 05 521 D3437
20D3437.103.02.02.02 03.2013
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.1 Calling the "Adjustment/Calibration" menu
Easypad
12
12.1
Adjusting and calibrating the system
Adjusting and calibrating the system
Calling the "Adjustment/Calibration" menu
Calling XG 3D and GALILEOS, ne w
Calling the "Adjustment/Calibration" menu
You can call the
"Utilities"
/
exposure "
Password prompt (see section entitled "Password protection")
The
"Select X-ray device"
only displayed if more than one unit has been set up in SIDEXIS XG.
Password protection
Password for XG 3D and GA LILEOS, new
The
"Adjustment/calibration"
password, enter the first four digits of the current system date (PC) in
reverse order. Example: On 05/30/2010, the service password is 5003.
Service mode
Service for XG 3D and GALILEOS, new
When you open the
user mode to the PC service mode logged by the PC. In PC service mode,
the control options that are available on the control panel are determined
by SIDEXIS XG and the service routine currently selected. General
control of the unit by means of the control panel (as in the user mode) is
not possible in this mode.
Easypad overview (GALILE OS)
"Adjustment/calibration"
"Constancy test..."
/
"Select X-ray component"
/
"3D"/ "Select X-ray device"
and
"Select X-ray component"
menu is password-protected. For the
"Adjustment/calibration"
menu via SIDEXIS XG:
/
"Service
/
"Adjustment/calibration"
prompts are
menu, the unit switches from
/
bеЦдблЬ
Service mode is displayed on the Easypad via the PC service image.
64 05 521 D3437
D3437.103.02.02.02 03.2013
21
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.2 Menu structureMaintenance Instructions GALILEOS Comfort PLUS
12.2
Menu structure
Menu structure
The menu is divided into four areas.
ANavigation areaStructure tree for adjustment and
calibration [ → 23]
BPreview imageShows the exposure to be taken in this
stage of the adjustment/calibration
procedure.
CMessage windowShows messages and information
about this stage of the adjustment/
calibration procedure.
DTools pictographShows which (if any) test phantom must
be used for this stage of the adjustment/
calibration procedure.
In addition to the four areas, the menu also contains the following buttons:
EImage acquisitionMakes the unit ready for exposure.
FSave valuesSaves the current adjustment/
calibration values.
GNextSwitches to the next stage of the
adjustment/calibration procedure.
HExitExits adjustment/calibration and closes
the menu.
64 05 521 D3437
22D3437.103.02.02.02 03.2013
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.2 Menu structure
12.2.1Navigation area
Navigation area
The navigation area contains a structure tree similar to the one you will
be familiar with from your Windows interface. The structure tree contains
all stages of the adjustment and calibration procedure you need to
complete in order to adjust and calibrate your system. The order in which
the elements appear in the structure tree determines the chronological
order of the procedure to adjust and calibrate the unit:
● Diaphragm
● Radiation field
● Dose measurement
● Sensor
●Iris
● Shading
● Distortion
● Geometry
In addition to the elements required for adjustment and calibration of the
unit described above, the structure tree also contains one other element:
Backup
● Calibration data
Validity of existing adjustment/calibration settings
The elements of the structure tree used for adjusting (calibrating) the
device are prefixed by symbols indicating the current status of the
corresponding adjustment or calibration operation.
The element for saving calibration data does not contain any symbols.
Green and
checked
YellowData record available,
RedInvalid data record or no
Valid data record;
adjustment/calibration is in
progress
but not yet
saved
record present
No adjustment/calibration
required
Data record must be saved
Adjustment/calibration
required
Working with the structure tree
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
You can navigate between the elements of the structure tree by clicking
on them with the mouse.
NOTICE! You must follow the prescribed sequence in order to obtain a
valid adjustment or calibration.
Click with the mouse on the small triangles in front of the elements to
collapse and expand the structure tree.
23
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.2 Menu structureMaintenance Instructions GALILEOS Comfort PLUS
12.2.2Tool pictographs
Tool pictographs
The tools pictograph shows which (if any) test phantom must be used for
this particular calibration step.
Click
"Cancel"
to quit the
"Service functions"
menu.
24D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
12.3
Adjustment and calibration via the "Service
Functions" menu
Adjustment and calibration via the "Ser vice Functions" menu
IMPORTANT
After every calibration of the unit, the reference values for the constancy
measurement must be recalculated and entered in the "Test results"
form, "Reference value" column.
12.3.1Diaphragm image
GALILEOS GAX7 and GAX5
Diaphragm image
bеЦдблЬ
Adjusting the "diaphragm open"
diaphragm setting
1. Call the
2. In the structure tree, under
"Diaphragm"
ª The
1. Set the rotary knob on the tube assembly to the "open diaphragm"
position.
2. Click on the
menu area.
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S030.5 is displayed on the control panel.
3. Take an exposure.
"Adjustment/calibration"
"3D Adjustment/calibration"
element (S030.5).
"Diaphragm"
"Diaphragm open"
menu is displayed in the action area.
button in the
menu [ → 21].
, click on the
"Image acquisition"
64 05 521 D3437
D3437.103.02.02.02 03.2013
25
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS
4. For GALILEOS Comfort and GALILEOS Comfort
Evaluate the image.
PLUS
:
A+BAdjustment OK
CAdjustment not OK
EPermissible tolerance: 30 pixels ± 5 pixels
ª The brightness distribution along the border surrounding the
image on all sides must be uniform (A).
ª The distance between the bottom edge and the lowest point in
the image should be 30 ± 5 pixels (measure with SIDEXIS scale)
(B).
If the distance between the bottom edge and the lowest point in
the image is out of tolerance (E) or the brightness distribution
along the surrounding border is not uniform (C), the diaphragm
must be adjusted mechanically.
26D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
5. If the exposure is OK (A+B), confirm this by clicking the check box
located to the right of the
ª The box will appear checked.
ª The adjustment for the "open diaphragm" diaphragm setting is
now complete.
"Diaphragm open"
button.
Adjustment of the "Upper jaw"
diaphragm setting
ª The
6. Continue the calibration procedure with the adjustment of the "upper
jaw" diaphragm setting.
✔ The
✔ The element
calibration"
✔ The
area.
1. Click on the
ª The
2. Set the rotary knob on the tube assembly to the "maxillary exposure"
position.
3. Click the
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S030.5 is displayed on the control panel.
4. Take an exposure.
"Diaphragm open"
acquisition"
menu area.
"Adjustment/calibration"
"Diaphragm"
in the structure tree (S030.5).
"Upper jaw"
button is selectable in the
"Upper jaw"
"Diaphragm"
"Upper jaw"
button is selectable in the
menu is called [ → 21].
is selected under
"3D Adjustment/
"Image
"Image acquisition"
button in the
menu is displayed in the action area.
button.
"Image acquisition"
menu area.
menu
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
27
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS
5. Evaluate the image.
AAdjustment OK
BAdjustment not OK
ª The upper edge of the lower lead diaphragm must be within
tolerance, i.e. lie inside of the auxiliary lies (A).
If the edge is out of tolerance (B), the diaphragm must be
adjusted mechanically.
28D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
6. If the exposure is OK (A), confirm this by clicking the check box
located to the right of the
ª The box will appear checked.
ª The adjustment for the "Upper jaw" diaphragm setting is now
complete.
"Upper jaw"
button.
Adjustment of the "Lower jaw"
diaphragm setting
ª The
7. Continue the calibration procedure with the adjustment of the "Lower
jaw" diaphragm setting.
✔ The
✔ The element
calibration"
✔ The
area.
1. Set the rotary knob on the tube assembly to the "lower jaw" position.
2. Click the
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S030.5 is displayed on the control panel.
3. Take an exposure.
"Lower jaw"
"Adjustment/calibration"
"Diaphragm"
in the structure tree (S030.5).
"Lower jaw"
"Lower jaw"
button is selectable.
menu is called [ → 21].
is selected under
button is selectable in the
button.
"3D Adjustment/
"Image acquisition"
menu
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
29
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS
4. Evaluate the image.
AAdjustment OK
BAdjustment not OK
ª The lower edge of the upper lead diaphragm must be within
tolerance, i.e. lie inside of the auxiliary lines (A).
If the edge is out of tolerance (B), the diaphragm must be
adjusted mechanically.
5. If the exposure is OK (A), confirm this by clicking the check box
located to the right of the
"Lower jaw"
button.
ª The box will appear checked.
ª Diaphragm adjustment is now complete.
6. Continue the calibration procedure with the radiation field
check [ → 31].
30D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
A
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
12.3.2Checking the radiation field
Checking the radiation field
IMPORTANT: The illumination must be checked once the collimator has
been adjusted.
✔ The
1. Clip the distortion phantom onto the X-ray detector cover.
2. In the structure tree, under
3. Click the
4. Press the R key to move the unit back to the starting position.
5. Press the release button. Hold down the release button and observe
6. To confirm that the lighting strips on the distortion phantom are
"Adjustment/calibration"
"Radiation field"
ª The
"Radiation field"
element (S002.6).
"Image acquisition"
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S002.6 is displayed on the control panel.
the distortion phantom. The lighting strips on the distortion phantom
(A) must not light up.
If the strips on the phantom light up at all, the system is overexposed,
and you cannot continue the adjustment. In this case, repeat the
diaphragm adjustment procedure and then check the radiation beam
field again. If the lighting strips still light up during the re-check of the
beam field, contact the SIRONA Customer Service Center (CSC) to
solve the problem.
click the check box on the left next to the text
correct?"
ª The box will appear checked.
.
menu is called [ → 21].
"3D Adjustment/calibration"
menu is displayed in the action area.
button.
, click on the
"Radiation field position
not
li t,
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
ª The beam field check is now completed.
7. Continue the calibration procedure with the dosimetry [ → 32].
Tip: Leave the distortion phantom on the unit for the next calibration
step.
31
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS
12.3.3Dosimetry
Dosimetry
A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for
dosimetry.
✔ The Adjustment/calibration menu is called [ → 21].
✔ The distortion phantom is clipped onto the cover of the X-ray detector
for protection against scratching.
1. Attach the Mult-O-Meter sensor approximately in the middle of the
distortion phantom mounted on the X-ray detector.
2. In the structure tree, under 3D Adjustment/calibration, click on the
"Dose measurement"
ª The
"Dose measurement"
element (S002.6).
menu is displayed in the action area.
3. Click the
"Image acquisition"
button.
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S002.6 is displayed on the control panel.
4. Press the R key to move the unit back to the starting position.
5. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
6. Then read off the dose from the Mult-O-Meter.
For GALILEOS Comfort
PLUS
, the value must be between 2.3 and
4.5 mGray.
If the value is outside the permissible range, check the X-ray tube
assembly.
7. To confirm that the dose is within the
and 4.5 mGray
, click the check box on the left next to the text.
permissible range between 2.3
ª The box will appear checked.
ª The dosimetry is now complete.
8. Remove the sensor from the distortion phantom and take the
phantom off the X-ray detector.
9. Continue the calibration procedure with the sensor
adjustment [ → 33].
32D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
12.3.4Sensor adjustment
Sensor adjustment
✔ The Adjustment/calibration menu is called [ → 21].
1. In the structure tree under
"Sensor"
ª The
2. Click the
ª SIDEXIS XG makes the unit ready for exposure.
element (S010.14).
"Sensor"
menu is displayed in the action area.
"Image acquisition"
"3D Adjustment/calibration"
button.
, click on the
bеЦдблЬ
ª Service routine S010.14 is displayed on the control panel.
3. Press the R key to move the unit back to the starting position.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
IMPORTANT: This process takes approx. 2-3 minutes.
ª The unit transfers the acquired images to the Reconstruction and
Control Unit (RCU). This process can take 2-3 minutes. Once the
transfer is complete, the evaluation of the sensor calibration is
displayed in the message window.
If the information in the message window indicates that
calibration is
procedure starting with Step b) until calibration is OK and/or
possible.
If you have repeated the procedure three times and still have not
attained a positive result, please contact the SIRONA Customer
Service Center (CSC).
5. If the adjustment is OK or possible, click the
ª The adjustment is saved.
ª The sensor adjustment is now complete.
6. Continue the calibration procedure with the iris adjustment [ → 34].
not OK and/or not possible
, keep repeating the
"Save values"
button.
64 05 521 D3437
D3437.103.02.02.02 03.2013
33
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS
12.3.5Iris adjustment
Iris adjustment
✔ The
1. In the structure tree under
2. Click the
"Adjustment/calibration"
"Iris"
element (S010.10).
ª The
"Iris"
menu is displayed in the action area.
"Image acquisition"
menu is called [ → 21].
"3D Adjustment/calibration"
button.
, click on the
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S010.10 is displayed on the control panel.
3. Press the R key to move the unit back to the starting position.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
IMPORTANT: This process takes approx. 2-3 minutes.
ª The unit transfers the acquired images to the Reconstruction and
Control Unit (RCU). This process can take 2-3 minutes. Once the
transfer is complete, the evaluation of the sensor calibration is
displayed in the message window.
If the information in the message window indicates that
calibration is
not OK and/or not possible
, keep repeating the
procedure starting with Step b) until calibration is OK and/or
possible.
If you have repeated the procedure three times and still have not
attained a positive result, please contact the SIRONA Customer
Service Center (CSC).
5. If the calibration is OK or possible, click the
"Save values"
button.
ª The adjustment is saved.
ª The iris adjustment is now complete.
6. Continue the calibration procedure with the shading
calibration [ → 35].
34D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
12.3.6Shading calibration
Shading calibration
✔ The
1. In the structure tree under
2. Click the
3. Press the R key to move the unit back to the starting position.
4. Press the release button. Press and hold down the button until the
"Adjustment/calibration"
"Shading"
ª The
element (S010.11 / S010.15).
"Shading"
menu is displayed in the action area.
"Image acquisition"
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S010.11 or S010.15 (extended shading
calibration) is displayed on the control panel.
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
ª The shading exposure and the evaluation of the shading
calibration is displayed.
If the information in the message window indicates that
calibration is
procedure starting with Step b) until calibration is OK and/or
possible.
If you have repeated the procedure three times and still have not
attained a positive result, please contact the SIRONA Customer
Service Center (CSC).
IMPORTANT: No foreign bodies may be visible on the shading
exposure. If this is the case, check the beam path for foreign
bodies, remove them if necessary and repeat the calibration.
not OK and/or not possible
menu is called [ → 21].
"3D Adjustment/calibration"
button.
, keep repeating the
, click on the
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
5. If the calibration is OK or possible, click the
ª The calibration is saved.
ª The shading calibration is now complete.
6. Continue the calibration procedure with the distortion
calibration [ → 36].
"Save values"
button.
35
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS
12.3.7Distortion calibration
Distortion calibration
IMPORTANT
With GALILEOS Comfort
With GALILEOS Comfort
device must be used. Distortion phantoms with a serial number <3001
may not be used for the calibration of GALILEOS Comfort
PLUS
:
Plus
, the distortion phantom delivered with the
PLUS
.
✔ The
1. In the structure tree under
"Adjustment/calibration"
"Distortion"
ª The
element (S010.12).
"Distortion"
menu is called [ → 21].
"3D Adjustment/calibration"
, click on the
menu is displayed in the action area.
2. Clip the distortion phantom onto the X-ray detector cover.
3. Click the
"Image acquisition"
button.
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S010.12 is displayed on the control panel.
4. Press the R key to move the unit back to the starting position.
5. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
IMPORTANT: This process takes approx. 2-3 minutes.
ª The unit transfers the acquired images to the Reconstruction and
Control Unit (RCU). This process can take 2-3 minutes. Once the
transfer is complete, the evaluation of the distortion calibration is
displayed in the message window.
If the information in the message window indicates that
calibration was not successful, check the calibration phantom to
make sure that it is not damaged.
If the phantom checks out OK (all balls are present and correctly
positioned), repeat the procedure starting with point d) as often
as required until the calibration is OK.
If you have repeated the procedure three times and still have not
attained a positive result, please contact the SIRONA Customer
Service Center (CSC).
36D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
6. IMPORTANT: When a new distortion calibration is saved, the
geometric calibration is set to "invalid" (red LEDs).
If the calibration is OK or possible, click on the
ª The calibration is saved.
ª The distortion calibration is now complete.
7. Remove the distortion phantom again from the X-ray detector cover.
8. Continue the calibration procedure with the geometry
calibration [ → 38].
"Save values"
button.
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
37
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS
12.3.8Geometry calibration
Geometry calibration
IMPORTANT
For an installed and configured FACESCAN:
The face scanner is automatically calibrated during the geometry
calibration of GALILEOS. In this case, make sure that the normal room
lighting is switched on during the calibration process. The room does not
have to be darkened during calibration.
✔ The
1. In the structure tree, under
"Adjustment/calibration"
"Geometry"
ª The
element (S010.13).
"Geometry"
menu is called [ → 21].
"3D Adjustment/calibration"
, click on the
menu is displayed in the action area.
2. Insert the geometric phantom in the bite block holder of the unit. Align
the phantom with a spirit level.
3. For FACESCAN: Check to make sure that the room lighting is
switched on. Switch it on if necessary.
4. Click the
"Image acquisition"
button.
ª SIDEXIS XG makes the unit ready for exposure.
ª Service routine S010.13 is displayed on the control panel.
5. Press the R key to move the unit back to the starting position.
38D3437.103.02.02.02 03.2013
64 05 521 D3437
Sirona Dental Systems GmbH12 Adjusting and calibrating the system
Maintenance Instructions GALILEOS Comfort PLUS 12.3 Adjustment and calibration via the "Service Functions" menu
6. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
IMPORTANT: This process takes approx. 2-3 minutes.
ª The unit transfers the acquired images to the Reconstruction and
Control Unit (RCU). This process can take 2-3 minutes. Once the
transfer is complete, the evaluation of the calibration is displayed
in the message window.
If the information in the message window indicates that
calibration was not successful, check the calibration phantom to
make sure that it is not damaged.
If you have repeated the procedure three times and still have not
attained a positive result, check the mechanical geometry of the
unit. Adjust the unit if necessary and then repeat the calibration.
If this still does not lead to a positive result, please contact the
SIRONA Customer Service Center (CSC).
7. If the calibration is OK or possible, click the
ª The calibration is saved.
ª Calibration of the geometry is now complete.
8. Remove the geometric phantom from the bite block holder of the unit.
"Save values"
button.
bеЦдблЬ
64 05 521 D3437
D3437.103.02.02.02 03.2013
39
12 Adjusting and calibrating the systemSirona Dental Systems GmbH
12.3 Adjustment and calibration via the "Service Functions" menuMaintenance Instructions GALILEOS Comfort PLUS