Thank you for your purchase of this CEREC MC XL Premium Package®
unit from Dentsply Sirona.
This device enables you to produce dental restorations, e.g. from ceramic
material with a natural appearance (CEramic REConstruction).
Improper use and handling can create hazards and cause damage.
Please therefore read and follow these operating instructions carefully.
Always keep them within easy reach.
Also pay attention to the safety instructions to prevent personal injury and
material damage.
Your
CEREC MC XL Premium Package team
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
5
2General dataDentsply Sirona
2.1Identification of the danger levelsOperating Instructions CEREC MC XL Premium Package
General data
2
Please read this document completely and follow the instructions exactly.
You should always keep it within reach.
Original language of the present document: German.
2.1
Identification of the danger levels
To prevent personal injury and material damage, please observe the
warning and safety information provided in these operating instructions.
Such information is highlighted as follows:
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily injury
or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTICE
A possibly harmful situation which could lead to damage of the product
or an object in its environment.
IMPORTANT
Application instructions and other important information.
Tip: Information for simplifying work.
63 98 536 D 3439
6D 3439.201.06.10.02 10.2017
Dentsply Sirona 2General data
Operating Instructions CEREC MC XL Premium Package2.2Formats and symbols used
2.2
2.3
Formats and symbols used
The formats and symbols used in this document have the following
meaning:
Prerequisite
1. First action step
2. Second action step
or
➢ Alternative action
Result
➢ Individual action step
See “Formats and symbols
used [ → 7]”
● ListDesignates a list.
“Command / menu item”Indicates commands, menu items or
Prompts you to do something.
Identifies a reference to another text
passage and specifies its page
number.
quotations.
Note PC / Acquisition Unit
When a PC is described in this document, this refers to a PC for the
acquisition unit (if present). The PC is represented symbolically.
Please observe our recommendations for PC configuration (see System
requirements [ → 70]).
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
7
3General descriptionDentsply Sirona
24
3.1CertificationOperating Instructions CEREC MC XL Premium Package
General description
3
3.1
Certification
CE mark
This product bears the CE mark in accordance with the provisions of
Council Directive 2006/42/EC (machinery directive). As such, the
following standards apply: DIN EN ISO 12100:2011-03, DIN EN 610101:2011-07 and DIN EN 61326-1:2013-07.
CAUTION
CE mark for connected products
Further products which are connected to this unit must also bear the CE
mark. These products must be tested according to the applicable
standards.
Examples of CE mark for connected products:
● EN 60601-1:2006 based on IEC 60601-1:2005
● EN 60950-1:2006 based on IEC 60950-1:2005
● UL 60950 second edition 2010
GOST mark
EAC certification
Mark of conformity of the Eurasian Economic Community
63 98 536 D 3439
8D 3439.201.06.10.02 10.2017
Dentsply Sirona 3General description
Operating Instructions CEREC MC XL Premium Package3.2Intended use
3.2
Intended use
This unit produces computer-aided dental restorations, e.g. from naturalappearing ceramic material.
This unit must not be used for any other purpose. If the unit is used for
any purpose other than the one mentioned above, it may be damaged.
Intended use also includes compliance with these Operating Instructions
and the relevant maintenance instructions.
CAUTION
Follow the instructions
If the instructions for operating the unit described in this document are
not observed, the intended protection of the user may be impaired.
For the USA only
CAUTION: Federal law (USA) restricts sale of this device to or on the
order of a physician, dentist, or licensed practitioner.
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
9
4SafetyDentsply Sirona
4.1Basic safety informationOperating Instructions CEREC MC XL Premium Package
Safety
4
4.1
Basic safety information
4.1.1Prerequisites
NOTICE
Important information on building installation
The building installation must be performed by a qualified expert in
compliance with the national regulations. DIN VDE 0100-710 applies in
Germany.
NOTICE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion
hazards.
NOTICE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
4.1.2Maintenance and repair
As manufacturers of dental instruments and laboratory equipment, we
can assume responsibility for the safety properties of the unit only if the
following points are observed:
● The maintenance and repair of this unit may be performed only by
Dentsply Sirona or by agencies authorized by Dentsply Sirona.
● Components which have failed and influence the safety of the unit
must be replaced with original (OEM) spare parts.
Please request a certificate whenever you have such work performed. It
should include:
● The type and scope of work.
● Any changes made in the rated parameters or working range.
● Date, name of company and signature.
4.1.3Modifications to the product
Modifications to this product which may affect the safety of the operator,
patients or third parties are prohibited by law!
63 98 536 D 3439
10D 3439.201.06.10.02 10.2017
Dentsply Sirona 4Safety
Operating Instructions CEREC MC XL Premium Package4.2Opening the processing chamber door during the machining process
4.1.4Accessories
In order to ensure product safety, this device may be operated only with
original Dentsply Sirona accessories or third-party accessories expressly
approved by Dentsply Sirona. In particular, only the power cable also
supplied or the corresponding original spare part may be used with the
unit. The user is responsible for any damage resulting from the use of
non-approved accessories.
4.2
4.3
Opening the processing chamber door during the
machining process
CAUTION
Instruments that continue to run
When the processing chamber door is opened during the machining
process, the instruments could continue to run for a short time.
➢ Be careful not to touch the instruments with your hand or any other
object during this time.
➢ Avoid opening the processing chamber door while the milling and
grinding unit is in operation.
➢ Before you open the processing chamber door, end any actions that
are running by pressing the
grinding unit touch display or in the application software.
"Stop"
button on the milling and
Wireless phone interference with equipment
The use of mobile wireless phones in practice or hospital environments
must be prohibited to ensure safe operation of the unit.
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
11
4SafetyDentsply Sirona
4.4Disturbance of data transmissionOperating Instructions CEREC MC XL Premium Package
Note on wireless communication
4.4
Disturbance of data transmission
Fragment Worldwide
Data communication between the acquisition unit and the CEREC MC XL
Premium Package milling and grinding unit should preferably be
established via the wireless interface CEREC Radio Device or WLAN.
As for all wireless connections (e.g. cell phones), heavy utilization of the
available radio channels or shielding caused by building installations (e.g.
metal-shielded X-ray enclosures) may impair the quality of the
connection. This may become noticeable through a reduction in range
and/or a slower data transmission rate. In extreme cases, it will be
impossible to establish a wireless connection at all.
Dentsply Sirona has selected the best possible configuration for data
communication via the wireless interface (CEREC Radio Device) or
WLAN, which generally ensures perfect functioning of this connection.
However, in individual cases unrestricted wireless data communication
may be impossible for the reasons mentioned above and/or due to local
circumstances. In such cases, a cable LAN connection should be
selected to ensure uninterrupted operation. If the only LAN interface on
the rear of the CEREC AC is occupied by another plug, remove this
wireless interface connection, and instead connect the LAN cable with the
CEREC MC XL Premium Package milling and grinding unit.
63 98 536 D 3439
12D 3439.201.06.10.02 10.2017
Dentsply Sirona 5Installation and startup
Operating Instructions CEREC MC XL Premium Package5.1Transport and unpacking
Installation and startup
5
5.1
Transport and unpacking
All products from Dentsply Sirona are carefully checked prior to shipment.
Please perform an incoming inspection immediately after delivery.
1. Check the delivery note to ensure that the consignment is complete.
2. Check whether the product shows any visible signs of damage.
NOTICE
Damage during transport
If the product was damaged during transport, please contact your
carrying agent.
If return shipment is required, please use the original packaging for
shipment.
The unit must be drained prior to shipment (if it has been operated).
Removing water from the unit [ → 67]
Transport without packaging
CAUTION
Damage to the unit or risk of injury during transport without packaging
There is a danger of the unit falling down if it is grasped by its plastic
housing.
bеЦдблЬ
5.2
➢ The unit should always be carried by two persons.
➢ Do not grasp the unit by its plastic housing.
➢ Always grasp the unit by its chassis next to its feet.
Disposal of packaging materials
The packaging must be disposed of in compliance with the relevant
national regulations. Please observe the regulations applicable in your
country.
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
13
5Installation and startupDentsply Sirona
5.3Installation siteOperating Instructions CEREC MC XL Premium Package
5.3
Installation site
CAUTION
Install out of the reach of patients!
Do not install or operate the milling and grinding unit in the vicinity of the
patient (place it at least 1.5 m away from the patient).
The milling and grinding unit requires a level Approx. footprint: 700 x 420
mm (W x D). The height of the milling and grinding unit is:
● with the processing chamber door closed: 425mm
● with the processing chamber door, open: 570mm
Install the milling and grinding unit in such a way that it is not difficult to
operate the main switch.
Make sure that the ventilation slots underneath and at the back of the unit
remain unobstructed. The distance between the back of the unit and the
wall must at least be 10 cm.
Note that the unit weighs 43 kg!
The unit must not be installed at sites with a high level of humidity or dust!
NOTICE
Installation in a cabinet
5.4
If the unit is installed in a cabinet, you must provide for adequate heat
exchange.
The ambient temperature surrounding the unit must be between 5°C
and 40°C.
Commissioning
NOTICE
Important information on initial startup
Observe the software installation instructions!
63 98 536 D 3439
14D 3439.201.06.10.02 10.2017
Dentsply Sirona 5Installation and startup
A
BCDEF
E
Operating Instructions CEREC MC XL Premium Package5.4Commissioning
5.4CommissioningOperating Instructions CEREC MC XL Premium Package
Ports on the back side
Connections
AMain switch
I = ON, 0 = OFF
ECommunications interface for
suction
BFuse coverFBar code reader connection
CPower connectionGConnection for suction
DLAN port
Ethernet
Processing chamber
Processing chamber
AInstrument set 1DCeramic block
BInstrument set 2EWorkpiece spindle
CMotor mountFSuction connection
63 98 536 D 3439
16D 3439.201.06.10.02 10.2017
Dentsply Sirona 5Installation and startup
Operating Instructions CEREC MC XL Premium Package5.4Commissioning
5.4.2Standard accessories
5.4.2.1
5.4.2.2
Instruments
The following instruments are available for milling and grinding purposes.
When replacing instruments, ensure the permitted instrument
combinations are used (see "Permitted instrument combinations [ → 43]").
Calibration pins
The calibration pins are used when calibrating the instrument sets (see
"Calibrating the unit [ → 38]“).
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
17
5Installation and startupDentsply Sirona
5.4CommissioningOperating Instructions CEREC MC XL Premium Package
5.4.2.3
Torque wrench
To insert or replace the instruments or calibration pins, use the following
torque wrench.
InstrumentREFUsageTorque wrenchClamping format of
the force
transmission
Step Bur 12 S62 40 167GrindingTriangular
Step Bur 1262 60 025Grinding
Cyl. Pointed Bur 12 S62 40 159Grinding
Step Bur 2062 59 597Grinding
Cyl. Pointed Bur 2062 59 589Grinding
Cylinder Bur 12 EF65 35 186Extra-fine grinding
Cylinder Pointed Bur 12 EF 65 35 178Extra-fine grinding
Shaper 2562 99 395Milling
(dry)
Finisher 1062 99 387Milling
(wet and dry)
Calibration pin (AiO*)62 41 132Calibration
Shaper 25 RZ64 33 440Milling
(wet)
Square
* All-in-One
63 98 536 D 3439
18D 3439.201.06.10.02 10.2017
Dentsply Sirona 5Installation and startup
Operating Instructions CEREC MC XL Premium Package5.4Commissioning
5.4.3Touch display description
In these Operating Instructions, operation is described in such a manner
that you can execute and confirm commands via your PC or the touch
display, such as
commands also have buttons on the touch display.
"OK"
"Cancel"
"Start" , "Stop" , "Cancel"
or
"OK"
. Other possible
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
19
5Installation and startupDentsply Sirona
5.4CommissioningOperating Instructions CEREC MC XL Premium Package
Touch display overview
Touch display overview
AUnit serial numberD
BDesignationEEdit unit settings
CSoftware version
"Pump"
Start/stop water pump
5.4.4Illumination of the processing chamber
The lighting of the processing chamber depends on the machining
operation involved:
Machining operationLighting color
Milling and grindingWhite
Operation completedGreen
Error or
"Stop"
button pressedRed
button,
63 98 536 D 3439
20D 3439.201.06.10.02 10.2017
Dentsply Sirona 5Installation and startup
Operating Instructions CEREC MC XL Premium Package5.4Commissioning
5.4.5Using the processing chamber screen
CAUTION
Risk of injury on instruments
Be careful not to brush against the instruments with your hand.
NOTICE
Risk of blockage in the cooling circuit
If chips enter into the cooling circuit of the machine, there is a risk that
the cooling circuit will become blocked.
➢ The sieve is suitable for all restoration and material types. It is
absolutely essential that no chips enter into the cooling circuit.
1. Remove the screen from the packaging.
2. Wet the underside of the sieve with water before insertion and press
it firmly against the floor of the processing chamber.
5.4.6Connecting the bar code reader
Tool holder
➢ Insert the block clamp tool on the left and the bar code reader on the
right.
Connecting the bar code reader
➢ Plug the bar code reader into the serial interface to the rear of the
milling and grinding unit and secure with screws.
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
21
5Installation and startupDentsply Sirona
5.4CommissioningOperating Instructions CEREC MC XL Premium Package
5.4.7Installation
5.4.7.1
5.4.7.2
Connecting to the PC via LAN
An Ethernet connection is located to the rear of the unit, which can be
used to connect the PC to the milling and grinding unit. Use a network
cable to do this (LAN connection).
Using a network cable
Connect the PC to the LAN connection of the unit.
If problems arise when connecting via a network cable, please read the
separate instructions "Operating the MC XL via LAN".
Connecting the unit to the power supply
5.4.7.3
NOTICE
Grounded power outlet
The unit must be connected to a grounded power outlet.
➢ Connect the unit to the power supply using the supplied power cable.
Installing the unit
You must connect the unit to the PC before putting it into operation. This
is described in the section entitled "Connecting to the PC via LAN" [ → 22]
or "Connecting to the PC via WLAN (option)" [ → 27].
63 98 536 D 3439
22D 3439.201.06.10.02 10.2017
Dentsply Sirona 5Installation and startup
Operating Instructions CEREC MC XL Premium Package5.4Commissioning
Searching for unit automatically
The unit is connected to the PC via a LAN cable or via WLAN.
1. Click the
2. Click on the
3. Click on the
All units connected to the PC are recognized. In the case of new
4. Enter a name and a short designation for the new unit.
The unit name, the designation, and the software version are
"Configuration"
"Devices"
"Scan for New Devices"
units, you will be prompted to enter a name.
then shown on the touch display.
button in the system menu.
button.
button.
Touch display overview
AUnit serial numberD
BDesignationEEdit unit settings
CSoftware version
"Pump"
Start/stop water pump
button,
bеЦдблЬ
63 98 536 D 3439
D 3439.201.06.10.02 10.2017
23
5Installation and startupDentsply Sirona
5.4CommissioningOperating Instructions CEREC MC XL Premium Package
Search for unit manually
The unit is connected to the PC via a LAN cable or via WLAN.
1. Click the
2. Click on the
3. Click on the
4. Set the network.
5. Enter the network address which appears on the
display once the milling and grinding unit has been switched on.
6. Click on the
The software attempts to contact the device.
If the connection fails, check the connection. If necessary, ask a qualified
technician.
"Configuration"
"Devices"
"Add Device (Manual)"
"Ok"
button.
button in the system menu.
button.
button.
"IP address:"
Unit removal
✔ If you no longer require a unit (e.g. a unit is replaced), you can remove
it.
✔ The unit is operation.
1. Click the
2. Click on the
3. Click on the unit that you wish to uninstall.
4. Click on the
You will be asked if you would like to remove the unit.
"Configuration"
"Devices"
button.
"Delete Device"
button in the system menu.
button.
5. Click on the
The device is removed.
"YES"
button.
63 98 536 D 3439
24D 3439.201.06.10.02 10.2017