6.5.2 Certificate of Conformity and System Validation .............................................. 141
6.6 Scope of delivery and ordering designations.......................................................... 143
Index ........................................................................................................................147
784 KFP Titrino
Overview
784 KFP Titrino
Instructions for Use
784 KFP Titrino
1
Overview
1 Overview
Front view of instrument:
1
2
3
4
5
2
784 KFP Titrino
1Exchange Unit
2Display
3Control keys and indicator lamps on the Titrino
Key < >Power switch
Key < 4 > Switching stirrer ON/OFF
Key <DOS>Dosing key. Dispensing is performed as long as
<DOS> is being pressed. Used e.g. to prepare
the Exchange Unit. The dispensing rate can be
set with potentiometer (5).
Key <STOP/FILL>- Stops procedures, e.g. titrations, conditioning.
- Filling after manual dosing with <DOS>.
Key <START>Starts procedures, e.g. titrations, conditioning.
Identical with key <START> of the separate
keypad.
Indicator lamps:
"COND."Lamp flashes when conditioning is performed
and the titration vessel is still wet. It is on if con-
ditioning is ok.
"STATISTICS"Lamp is on when the "statistics" function
(calculation of mean and standard deviation) is
on.
"SILO"Lamp is on when silo memory (for sample data)
is on.
Overview
4Setting of display contrast
5Controls the dosing rate during manual dosing with <DOS> and
subsequent filling
784 KFP Titrino
3
Overview
Rear view of instrument:
1
A
6
7
8
9
RS 232
A
Keyboard
C
2
Remote
Pol
1.784
Nr.
B
100 - 240 V
f = 50 - 60 Hz
S = 15 W
Made by Metrohm
Herisau Switzerland
12
E
11
10
4
784 KFP Titrino
6RS232 interfaces
2 separate interfaces for the connection of printer, balance, and computer
7Remote lines (input/output)
for the connection of the Remote Box, Sample Changers, robots etc.
8Connection of electrodes
• 1 measuring input for polarized electrodes, e.g. KF electrode
9Connection for separate keypad
10Connection for power cable
With power supplies where the voltage is subject to severe HF disturbances, the Titrino should be operated via an additional power filter,
e.g. Metrohm 615 model.
Overview
11Connection for stirrer
728 Magnetic Stirrer, 722 Rod Stirrer, 703 or 727 Ti Stand
Supply voltage: 10 VDC (I ≤ 200 mA)
12Rating plate
with fabrication, series and instrument number
784 KFP Titrino
5
2.1 Keypad
SILO
–9STATISTICS
MEAS/HOLD
+
→↑←
↓
STOP
START
/6*
USER METH
)3DEF
(2C-FMLA
MODE
REPORTS
PRINT
USER
CURVE
2 Manual operation
2.1 Keypad
PARAMCONFIG
SMPL
DATA
PARAMParameters.
SMPL DATA Sample data.
CONFIGConfiguration.
EP
7
8
STATISTICS ON/OFF switching of statistics calcu-
lations of consecutive determination,
see page 28.
MEAS/HOLD ON/OFF switching of measurements
RS
4
5
between titrations and hold during titrations.
SILOON/OFF switching of silo memory for
C
1
C-FMLACalculation values, see page 27.
sample data, see page 38.
DEFFormulas, data output, see page 25ff.
MN
0
ABC
.
;
–
USER METH Management of internal method mem-
ory, see page 35.
PRINTPrinting of reports, see page 33.
REPORTSResult output.
MODEMode dialogue, see page 14.
USERUser name, see page 34.
CLEARENTER
CURVESwitching result/curve display, see
page 33.
←,→Selection of special values (dialog
marked with ":").
QUIT
↑,↓Cursor key for navigation.
CLEARClears values, set special values.
6.2130.050
ENTERStores values.
STOPStops methods.
QUITQuits inquiries, waiting times, printing.
STARTStarts methods.
The third functions (inscriptions in the triangle) on the keys of the keypad are used for
formula entry, see page 25.
6
784 KFP Titrino
2.2 Principle of data input
remo
te box: OF
F
• If you press a key, you will find a group of inquiries
Place the cursor to ">store method" and press
<ENTER>.
The name of the method which is currently in the
working memory is displayed.
• Delete this name with <CLEAR>.
• Open the "text writing mode" with key <ABC>.
You can now select the desired character by means
of the cursor keys, then confirm this character. Select the next character...
When you confirmed the last character, i.e. your
name is complete, you quit the text writing mode
with <QUIT>.
Confirm now the name with <ENTER>.
• During text input you can correct typing errors with
<CLEAR>:
<CLEAR> deletes the characters one by one.
• If you wish to modify an existing name, do not de-
lete the existing name before you start the text input
mode. Proceed then as follows:
1. Press <USER METH>, place the cursor to
">store method" and press <ENTER>.
2. Open the text writing mode directly: Press key
<ABC>.
3. <CLEAR> now deletes the characters one by
one or you can add additional characters.
4. If your text is complete, leave the text writing
mode with <QUIT> and confirm the text with
<ENTER>.
8
784 KFP Titrino
2.4 Configuration, key <CONFIG>
Key <CONFIG> serves to enter device specific data.
Monitoring of instrument validation, service interval and
printout of diagnostic report.
peripheral units:
Selection of printer, balance, stirrer control and COM
for manual report output.
auxiliaries:
e.g. setting of dialog language, date, time, type of result display.
RS232 settings COM1 and 2:
RS parameters for the COM's.
common variables:
Values of common variables.
2.4 Configuration, key <CONFIG>
>monitoring
validation: OFF
time interval 365 d
time counter 0 d
service: OFF
The display texts of the Titrino are shown on the left
side. The values are the default values.
Monitoring functions
Monitoring the validation interval (ON, OFF)
Monitoring is carried out at the end of the titrations and
when the Titrino is switched on. If the monitoring
responds the message "validate instrument" appears.
The message vanishes with <CLEAR>. At the same
time the counter is reset to zero.
If "ON" has been set:
Time interval for validation (1...9999 d)
see also page 139.
Time counter (0...9999 d)
Counts the number of days since the last time the
counter was reset.
Monitoring the service interval (ON, OFF)
Monitoring is carried out after the Titrino has been
switched on. If the monitoring responds the message
"Service is due" appears. The message vanishes with
<CLEAR>.
next service YYYY-MM-DD
784 KFP Titrino
If "ON" has been set:
Date of next service (YYYY-MM-DD)
9
2.4 Configuration, key <CONFIG>
system test report: OFFSystem test report printout (ON, OFF)
With "ON" the report of the system test is printed out
after the Titrino has been switched on, see also
page 139.
>peripheral units
send to COM1: IBM
send to COM2: IBM
man.reports to COM: 1
balance: Sartorius
Settings for peripheral units
Selection of printer (Epson, Seiko, Citizen, HP, IBM) at
the Titrino COM1
"Epson", for Epson
"Seiko", e.g. for DPU-414
"Citizen", e.g. for iDP 562 RS
"HP" e.g. for Desk Jet types. Place curves always at the
beginning of a page as you cannot have them over 2
pages.
"IBM" for all printers with IBM character set Table 437
and IBM graphics, as well as for the data transmission
to a computer or a data system.
COM of Titrino for the output of manually triggered
reports (1, 2, 1&2)
Manually triggered reports (e.g. with <PRINT> ....).
Exception <PRINT><REPORTS>: These reports are
outputted on the COM as defined in the method.
Selection of balance (Sartorius, Mettler, Mettler AT,
AND, Precisa)
Sartorius:Models MP8, MC1
Mettler:Models AM, PM and balances with 011,
012, and 016 interfaces
Mettler AT: Model AT
AND:Models ER-60, 120, 180, 182, FR-200, 300
and FX-200, 300, 320
Precisa:Models with RS232C interface
10
stirrer control: OFF
remote box: OFF
keyboard: US
Automatic switching ON/OFF of the stirrer in the titration
sequence (ON, OFF)
If stirrer control is ON, the stirrer will be switched on at
the beginning and switched OFF at the end of a determination. For KFT with conditioning the stirrer will be
switched off in the inactive state.
For stirrer control the red switch on the stirrer unit must
be ON.
Connection of a remote box (ON ,OFF)
To the remote socket for PC keyboard and barcode
reader, see page 118.
If "ON" has been set:
Type of PC keyboard (US, German, French, Spanish,
Swiss.)
The PC keyboard is used as an input aid, see
page 119.
784 KFP Titrino
2.4 Configuration, key <CONFIG>
barcode: inputTarget for barcode reader (input, method, id1, id2, id3,
smpl size)
The barcode reader is used as an input aid, see
page 118.
Input:The barcode string goes to the entry field in
which the cursor is currently located.
Method:The barcode string goes to the entry field
"Methods" in the silo memory.
Id1:The barcode string goes to the entry field
"Id1". (Similar for Id2 and Id3.)
Smpl size: The barcode string goes to the entry
field "smpl size".
>auxiliaries
dialog: english
date 1999-08-15
time 08:13
run number 0
auto start OFF
start delay 0 s
General settings
Selection of dialog language (english, deutsch,
français, español, italiano, portugese, svenska)
Current date (YYYY-MM-DD)
Format: Year-month-day, entry with leading zeros.
Current time (hh-mm)
Format: Hours-minutes, entry with leading zeros.
Current run number for result output (0...9999)
The sample number is set to 0 when the instrument is
switched on and incremented on every determination.
Automatic start of titrations. (1...9999, OFF)
Number of automatic starts ("number of samples").
Used for instrument interconnections in which the
external instrument does not initiate a start. Not
advisable in connections with Sample Changers.
Start delay (0...999999 s)
Delay time after start of methods. Abort start delay time
with <QUIT>.
result display: bold
dev.label.
program 784.0010
784 KFP Titrino
Type of result display at the end of the determination
(bold, standard)
bold: The calculated results are displayed in bold
characters.
standard: Displays the whole information, e.g. results,
endpoints, messages etc.
Individual identification of devices (up to 8 ASCII
characters)
Will be printed in the result report, see page 32.
The values of all common variables are displayed. For
creating common variables see page 30.
Settings with key <CONFIG> and power ON
Proceed as follows:
1. Switch the Titrino OFF.
2. Press <CONFIG> and keep it pressed during switching the Titrino ON.
The display shows the following:
Setup
>lock
>curve
lock:
Locking keys <CONFIG>, <PARAM> and <SMPL
DATA>, and the functions "recall method", "store
method" and "delete method" of the internal method
memory in the Titrino.
curve:
Changes the appearance of the curve printout.
>lock
lock
"ON" means that the corresponding function is not
accessible anymore.
<configuration>: OFF
<parameters>: OFF
<smpl data>: OFF
recall method: OFF
store method: OFF
delete method: OFF
The corresponding key is locked.
The corresponding function in the internal method
memory of the Titrino is locked.
12
784 KFP Titrino
2.4 Configuration, key <CONFIG>
>curve
grid: ON
frame: ON
scaling: Full
width 0.5
length 0.05
curve
The settings are valid for COM1 and COM2.
If you change the printer type, the following settings are
initialized according to the printer.
Grid drawing (ON, OFF)
Frame drawing (ON, OFF)
Type of scaling (Full, Auto)
Full: The scaling goes from the greatest to the smallest
value.
auto: The scaling from tick to tick, e.g. the
smallest/greatest values lie in between the first/last tick.
Width (0.2...1.00)
1 is greatest width. If you set 1, you may loose the label
at the right margin.
Length (0.01...1.00)
Time axis:
Curve length
0.01100 cm
0.110 cm
0.5 2 cm
11 cm
784 KFP Titrino
13
2.5 Key <MODE>
MODE
2.5 Key <MODE>
;
–
mode
mode: KFT
measured quantity: Ipol
With the key <MODE> the dialogue „mode“ is
opened.
The 784 Titrino is a titrator, specially designed for Karl
Fischer water determinations. It has exclusively the
mode KFT (Karl Fischer Titration).
With <ENTER> the selection of the measured
quantity is activated. Select Ipol or Upol with <←> or
<→> and confirm it also with <enter>.
The following modes can be selected:
• KFT: Karl Fischer Titration, the determination of
water content.
The mode KFT is equipped with a set of standard
parameters. They only need few settings in order to be
ready to work.
14
784 KFP Titrino
2.6 Parameters, key <PARAM>
The key <PARAM> is used for the entry of values that
PARAM
2.6.1 Parameters for KFT
determine the modes. Values marked with "cond." are
accessible during the conditioning in the KFT mode.
"**titr." means that these values can be changed during the titration. They influence the ongoing determination. Other values can only be changed in the inactive
state.
The display texts of the Titrino are shown on the left
side. The values are the default values.
<CLEAR> sets "max.".
This parameter determines primarily the addition rate
outside the control range, see page 21.
The maximum rate depends on the Exchange Unit:
Exchange Unitmax. rate
<CLEAR> sets "min.".
This parameter determines the addition rate at the start
and the end of the titration, see also page 21.
This parameter influences the titration speed and
therefore its accuracy very strongly: A smaller
"min.volume incr." results in a slower titration.
Type of stop criteria (drift, time)
Titration stops, if EP and stop drift are reached
(1...999 µL/min)
Titration stops if there is no dosing during t(delay)
(0...999 s, INF)
<CLEAR> sets "INF".
Switch off when EP is reached and the set time after
the last dosing has elapsed.
stop time OFF s
**titr.
>titration parameters
titr.direction: –
pause 1 0 s
**titr.
If t(delay) is "INF"
Stop after a time (0...999999 s, OFF)
<CLEAR> sets "OFF".
Stop after the set time after start of titration. "OFF"
means no stop, i.e. titration for an "infinitely" long time.
Titration parameters
Direction is set automatically (+, –, auto)
auto: The direction is set automatically by the Titrino.
+:Direction of higher voltage (more "positive"),
larger currents.
– : Direction of lower voltage, smaller currents.
Pause 1 (0...999999 s)
Waiting time before start volume, e.g. for equilibration
of the electrode after start. The waiting time can be
aborted with <QUIT>.
16
784 KFP Titrino
2.6 Parameters, key <PARAM>
start V: OFF
cond.
start V 0.0 ml
cond.
factor 0
cond.
dos.rate max. ml/min
**titr.
pause 2 0 s
**titr.
Type of start volume (OFF, abs., rel.)
"OFF": start volume switched off
"abs.": absolute start volume in mL
"rel.":relative start volume to sample size.
If "abs." is set:
Absolute start volume (0...999.99 mL)
If "rel." is set:
Factor for relative start volume (0...±999999).
Calculated as: start V in mL = factor ∗ sample size
Dosing rate for start volume (0.01...150 mL/min, max.)
<CLEAR> sets "max.".
The maximum rate depends on the Exchange Unit:
Exchange Unitmax. rate
Waiting time after start volume, e.g. reaction time after
dosing of a start volume. The waiting time can be
aborted with <QUIT>.
extr.time 0 s
**titr.
I(pol) 50 µA
U(pol) 400 mV
electrode test: OFF
temperature 25.0 °C
cond.
time interval 2 s
cond.
Extraction time (0...999999 s)
During this time the titration is running. It will not be
stopped (also if EP is already reached), before the
extraction time is terminated. The extraction time can
be aborted with <QUIT>.
Polarization current (–127...127 µA)
or the
polarization potential (–1270...1270 mV, in steps of
10 mV) is inquired.
Electrode test (OFF, ON)
Test for polarized electrodes. Performed on
changeover from the inactive standby state to a
measurement. "OFF" means that the test is not
performed.
Titration temperature (–170.0...500.0 °C).
The temperature can be entered manually.
Time interval (1...999999 s)
Time interval for acquisition of a measured value into
the measuring point list.
784 KFP Titrino
17
2.6 Parameters, key <PARAM>
>stop conditions
stop V: abs.
**titr.
stop V 99.99 ml
**titr.
factor 999999
**titr.
filling rate max. ml/min
**titr.
Stop conditions for titration
If this is not "normal", i.e. after reaching the EP
Type of stop volume (abs., rel., OFF)
"abs.": absolute stop volume in mL
"rel.":relative stop volume to sample size
"OFF": stop volume switched off. Stop volume is not
monitored.
The stop volume applies for one single titration. The
conditioning volume is not monitored.
If "abs." is set:
Absolute stop volume (0...9999.99 mL)
If "rel." is set:
Factor for relative stop volume (0...±999999)
Calculated as: Stop V in mL = factor ∗ sample size
Filling rate after titration (0.01...150 mL/min, max.)
<CLEAR> sets "max.".
The maximum rate depends on the Exchange Unit:
Exchange Unitmax. rate
Automatic conditioning of titration vessel (ON, OFF)
If conditioning is "ON", the solution is constantly kept at
the endpoint.
When conditioning is performed, the volume drift can
be displayed during the conditioning:
Display of drift during conditioning (ON, OFF).
Volume drift.
Type of drift correction (auto, man., OFF)
Type of drift correction: (EP – drift ∗ time)
auto: drift value at start is valid
man.: see below
OFF : no correction
Value for manual drift correction (0.0...99.9 µL/min)
Request of identifications after start of titration (id1,
id1&2, all, OFF)
After start, sample identifications can be requested
automatically: Only id1, id1 & id2, all three id's or no
inquiries.
18
784 KFP Titrino
2.6 Parameters, key <PARAM>
req.smpl size: OFF
cond.
limit smpl size: OFF
cond.
low lim. 0.0
cond.
up lim. 999999
cond.
oven: no
cond.
Request of sample size after start of titration (value,
unit, all, OFF)
"all" the value and the unit will be requested.
Limiting value check for sample size (ON, OFF)
With "ON" the error message "sample size out." appears if the entry is outside the set limits. The limiting
values are shown in the display window.
The absolute value of the limit is checked during sample size input and during the calculation of the results.
If "ON" has been set:
Lower limit for sample size (0.0...999 999)
Upper limit for sample size (0.0...999 999)
Connected oven (COM1, COM2, no)
COM of the Titrino to which the oven is connected.
If an oven is connected via RS232 an inquiry will be
made for the oven results and these will be inserted
into the result report of the Titrino. The report output on
the oven must be switched OFF (see also page 115).
Set "no" if no oven has been connected or if you have
not connected the oven to Titrino the via RS232 interface.
activate pulse: OFF
cond.
Pulse output on I/O line L6 (L6, pin 1) of the remote
socket (first, all, cond., OFF) e.g. start of a Dosimat,
see page 126.
784 KFP Titrino
19
2.6 Parameters, key <PARAM>
<START>
(Sta
rt delay)
(Sta
rt conditions)
Calculation
sData output
(Reconditioning)
(Stirrer OFF
)
Titration sequence of KFT
(Activate pulse)
(Stirrer ON)
(Preconditioning)
(<START>
(Activate pulse)
(Start delay)
(Request ident.)
(Request smpl size)
(Extraction time)
Titration with test of
stop criterion
After the start, the activate pulse is outputted and the stirrer switched on.
The start delay time is waited off.
If conditioning is on, the sample solution is titrated until
the EP is reached. The display shows then
drift OK 2.3 µl/min
or
KFT conditioning
The vessel is now conditioned. The titration can be started
with <START>.
The sample identifications and the sample size are requested.
Pause 1 is waited off, the start volume is dispensed and
pause 2 waited off.
The titration is executed. If the extraction time is not yet
over when the endpoint has been reached, the titration will
only be terminated after the extraction time is over.
The stirrer is switched off when there is no conditioning.
Calculations are carried out.
Data are outputted.
Conditioning is carried out.
20
784 KFP Titrino
2.6 Parameters, key <PARAM>
Control parameters for KFT
The control parameters can be set according to your samples. The default parameters
are already set to get satisfactory results. Optimize the control parameters for specific
samples only.
During the titration, reagent dosing is carried out in 3 phases:
U/mV
t/s
Initial
dosing
I/uA
KFT Ipol
Continuous
dosing
EP
Control
range
V/mL
V/mL
1. Initial dosing:
Here the dosing rate
increases constantly up to
"max.rate".
2. Continuous dosing:
Dosing is performed at the
maximum rate "max.rate"
until the control range
(dynamics) is reached.
3. Control range (dynamics):
In this range, dosing is
performed in single steps.
The last dosing steps are
controlled by the
parameter "min.volume
incr.".
EP
KFT Upol
V/mL
Remarks:
• Titer determination should have the same modes and the same parameters as the
sample titrations.
• In most cases the preset default parameters provide precise and reproducible results.
• Ipol gives relatively steep curves. They are flatter with Upol.
784 KFP Titrino
21
2.6 Parameters, key <PARAM>
Mechanisms of the KFT parameters in Ipol mode
U/mV
+
Titration
(direct titration)
EP at U
Titration
(Back titration)
I(pol)
Excess H2O
Excess Iodine-
darker
solution
+
I/µA
• The position and curve characteristics of the line between the ranges of excess water
or excess iodine depends on the type of sample and the ingredients of the working
media.
• The endpoint has to be set close to the range margin, but always within the iodine
excess range. If the endpoint is set too close to the limit, an overdose of KF reagent
may be dispensed. The steeper the curve at the preset polarization current I(pol), the
more delicate it is to titrate to a steady and reproducible endpoint. For the titration of
troublesome samples, the particular conditions must be optimized by trial and error.
Always mind the color of the working medium at the end point as your guideline.
Remark: In most cases the preset default parameters provide precise and
reproducible results.
• If you select negative values for a parameter, you should choose negative values for all
other parameters too in order to have compatible parameter sets.
• In KFT Upol mode, all parameters work in a similar manner as in KFT Ipol mode.
22
784 KFP Titrino
2.6 Parameters, key <PARAM>
The determination of the free water is easily done, as far as the specifications of the
reagent manufacturer regarding the "water capacity" of the reagents are considered.
Problems may occur with specific sample matrices. The relevant literature contains many
precise analysis instructions. Here, we give you some useful hints for the sample
addition.
Sample addition
For sample addition various accessories are available, for example injection syringes,
weighing spoons etc.
It is a good idea to apply the back-weighing technique, except when you are disposing
small amounts of liquids with a micro-syringe. Fill the syringe or the weighing spoon with
your sample just before the titration (e.g. during conditioning) and tare it on a balance
with an appropriate precision, which ideally is connected to the Titrino. Sample size
request should always be switched on (see page 19). Add the sample during the request
for the sample size, and weigh the syringe or weighing spoon again to evaluate the
weight difference. Use the transfer of balance data to enter the correct sample size if
possible. The weight has to be transmitted in gram. If you transfer the sample size from
the balance, the sample size request is confirmed automatically, and the titration is
started at once. Otherwise use the keypad to enter the weight.
If the sample size request is switched off, and preconditioning is set, you are explicitly
prompted to add your sample. For six seconds, the message "add sample" is displayed
after which the titration is started by the Titrino itself. This waiting time can be extended by
pressing <MEAS/HOLD>.
Solid samples
Use the glass weighing spoon 6.2412.000 and dispose the sample through the opening
for the septum stopper.
Liquid samples
Use a disposable syringe (2...20 mL) or a micro-syringe with a long needle attached.
Puncture the septum and dispose the sample carefully. When using disposable syringes,
add the liquid sample carefully without dipping into the solvent. Draw back the last drop
of sample into the syringe before you redraw the needle. For disposing an exact, small
volume of liquid sample, use a micro-syringe. Proceed as described above, but dip the
needle into the preconditioned solvent and dispose the sample carefully. Here drawing
back of solvent or sample adhering to the needle is not recommended.
Always keep in mind to exchange the septum as soon as it shows any large punctures
impairing the tightness of the titration cell.
Pasty, viscous samples
For samples which cannot be disposed with a needle because of their high viscosity, you
can handle your samples with disposable syringes with a large volume without using a
needle. Dispose your sample through the opening of the septum stopper. Back-weigh the
syringe after addition of the sample. Be sure to wipe off any adhering excess sample
substance of the syringe before taring it.
In any case, pay attention to prevent the penetration of atmospheric moisture into the
titration vessel when you are disposing the sample. If you still have to open the titration
cell for any reason, determine a blank value and take it into account for the calculation of
the titration result (see page 132).
784 KFP Titrino
23
2.6 Parameters, key <PARAM>
In the following table we attempt to show you solutions related more to the instrument's
side:
What to do if ...
ProblemPossible causes and remedial action
Dosing at end too long and
increments too small. "Is never
finished!"
The increments at the end of the
titration are too large.
"Overshoots".
Solution becomes too brown at
the end of the titration.
Solution becomes darker with
every titration.
• Increase "min.volume incr.".
• Change stop criterion. Try to increase stop drift or
use a short stopping time as stop criterion, e.g.
• For problematic samples change solvents with
ketones or aldehydes in 2-methoxyethanol or with
amines mixture in methanol/glacial acetic, e.g., see
literature.
• Lower "max.rate". The following experiment gives you
a reference point for the optimum max.rate: Drift
display during conditioning and add sample without
starting the titration. Select a value below the
maximum drift as "max.rate".
• Optimize setup of electrode and buret tip and
improve stirring.
• The methanol fraction in the solvent is too low.
Change the solvent.
• Electrode could be coated; wipe off with acetone.
• Renew solvent.
• Electrode could be coated; wipe off with acetone.
The drift increases with every
titration.
The endpoint is reached "too
rapidly".
The titration times become
longer and longer.
• Does your sample evolve water extremely sluggishly?
Work with the KF oven.
• Are acids esterified in your sample? Change solution
more frequently. Increase buffer capacity of the
solvent.
• Does your sample contain ketones or aldehydes?
Use special reagents suitable for ketones and
aldehydes.
• Reduce "max.rate".
• With 2-component reagents the buffer capacity of the
solvent can be exhausted. Change solution.
• If the drift increases at the same time, see above.
Key <DEF> contains various inquiries for result
calculations and data output. The data of this key are
method specific and they are stored in the method
memory together with the method.
formula:
Formulas for result calculations.
The display texts of the Titrino are shown on the left
side. The values are the default values.
Input of formulas
Enter formula number (1...9)
You can calculate up to 9 results per method.
Enter a number 1...9.
Input of formula
Example:
RS1=EP1∗C01/C00
Enter formula by means of 3rd functions of keyboard.
Here you will find operands, mathematical operations
and parentheses. Operands require a number as an
identification. You can use the following operands:
EPX: EP's. X = 1...9
RSX: Results which have already been calculated with
previous formulas. X = 1...9
CXX: Calculation constants. XX = 00...45
784 KFP Titrino
Rules:
• Calculation operations are performed in the
algebraic hierarchy: ∗ and / before + and –.
• Store formula with <ENTER>.
• Calculation quantities and operands can be deleted
with <CLEAR> one by one.
• To delete a complete formula press <CLEAR>
repeatedly until only RSX remains in the display.
Confirm with <ENTER>.
If a formula is stored with <ENTER>, result text,
number of decimals, result unit and the settings of the
result limit control will be requested:
25
2.7 Result calculations
RS1 text RS1
Text for result output (up to 8 characters)
Text input see page 8.
RS1 decimal places 2
RS1 unit: %
Number of decimal places for result (0...5)
Selection of result unit (%, ppm, g/L, mg/mL, mol/L,
mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to
6 characters).
RS1 limit control: OFF
Limit control for the result (ON, OFF)
The limits are checked each time a result is calculated.
Sets line L13 of the remote socket (OFF, active, pulse)
if the result lies outside the limits.
Enter next formula, e.g. for RS2.
Meaning of the calculation variables CXX:
C00Sample size, see page 37.
C01...C19Method specific operands, see page 27. They are stored with the
method in the method memory.
C21...C23Sample specific operands, see page 37ff.
C26, 27 Mean values from silo calculations, see page 41ff.
C30...C39Common variables.
C40Initial measured value of the sample.
C41End volume.
C42Determination time.
C43Volume drift for KFT with conditioning.
C44Temperature.
C45Dispensed start volume.
26
784 KFP Titrino
Input of method specific operands C01...C19, key <C-FMLA>
C-FMLA
2.7 Result calculations
C
1
With <C-FMLA> the operands C01...C19 can be put
in. For the calculation, the operands are used, which
were introduced in the formula.
The inputs method specific and are store in method
memory.
The calculation report can be printed with the key sequence
<PRINT><←/→> (press keys repeatedly until "calc" appears in the display) <ENTER>.
Example for a calculation report:
'fm
784 KFP Titrino 02134 784.0010
user sn
date 1999-08-15 time 15:01 3
KFT Ipol KF-Blank
>calculations
Water=(EP1-C38)*C39*C01/C00/C02;2;%
Titer=C39;4;mg/ml
Blank=C38;4;ml
C00= 1.0
C01= 0.1
C02= 1
C38= 0.0
C39= 0.0
------------
Calculation report
mode und method name
formulas:
result name=formula;number of decimal
places;result unit
sample size
method specific operands
values of the common variables that are
used
784 KFP Titrino
27
Loading...
+ 123 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.