Metrohm 782IC User Manual

CH-9101 Herisau/Switzerland Tel. ++41 71 353 85 85 Fax ++41 71 353 89 01 Internet www.metrohm.ch E-Mail info@metrohm.ch
782 IC Column
Thermostat
782 IC Column Thermostat
8.782.1003 Instructions for Use
01.03.2000 / dö

Table of contents

Table of contents
1 Introduction............................................................................................1
1.1 Instrument description.............................................................................1
1.2 Parts and controls..................................................................................... 2
1.3 Information about the Instructions for Use..........................................4
1.3.1 Organization ..................................................................................4
1.3.2 Notation and pictograms............................................................... 5
1.4 Safety notes ...............................................................................................6
2 Installation..............................................................................................7
2.1 Setting up the instrument........................................................................ 7
2.1.1 Packaging......................................................................................7
2.1.2 Check ............................................................................................7
2.1.3 Location......................................................................................... 7
2.1.4 Arrangement of the instruments....................................................7
2.2 Mains connection......................................................................................8
2.2.1 Setting the mains voltage..............................................................8
2.2.2 Fuses.............................................................................................9
2.2.3 Mains cable and mains connection ..............................................9
2.2.4 Switching the instrument on/off..................................................... 9
2.3 IC Column Heater....................................................................................10
2.3.1 Installation in 733 IC Separation Center......................................10
2.3.2 Installation in 761 Compact IC ....................................................11
3 Operation............................................................................................... 13
4 Appendix................................................................................................. 15
4.1 Technical data..........................................................................................15
4.2 Standard equipment ...............................................................................17
4.3 Warranty and conformity.......................................................................18
4.3.1 Warranty ......................................................................................18
4.3.2 EU Declaration of conformity ......................................................19
4.3.3 Certificate of conformity and system validation .......................... 20
4.4 Index..........................................................................................................21
782 IC Column Thermostat
I
Table of contents
List of figures
Fig. 1: Front of 782 IC Column Thermostat ............................................................... 2
Fig. 2: Rear of 782 IC Column Thermostat................................................................ 3
Fig. 3: Setting the mains voltage ............................................................................... 9
II
782 IC Column Thermostat

1 Introduction

1.1 Instrument description
The 782 IC Column Thermostat consists of a control unit and the «IC Column Heater» column heating element, which is placed inside the
733 IC Separation Center or the 761 Compact IC. The separating col­umn inserted in the column heating element is brought to the preset temperature within a few minutes. The current and set temperature val­ues can be read off from the control unit at any time.

1.1 Instrument description

782 IC Column Thermostat
1
1 Introduction

1.2 Parts and controls

782 IC Column Thermostat MetrohmMetrohm
11 22 55 66
Fig. 1: Front of 782 IC Column Thermostat
11 Key "PGM"
Program key (without function)
22 Key "Down"
Key to lower the set temperature
33 Key "Up"
Key to increase the set tempera­ture
33 44
44 Key "EXIT"
Exit key (without function)
55 Current temperature
Digital display (red) of current temperature
66 Set temperature
Digital display (green) of set tem­perature
2
782 IC Column Thermostat
f = 50-60 Hz S = 12 VA
115V: 100-120V: 230V: 220-240V:
WARNING - Fire Hazard -
For continued protection replace only with the same type and rating of fuse
Fuse 0,5A(T) 0,25A(T)
1.2 Parts and controls
1010998877
Type 1.782.0010 Nr.
IC Column Heater
Made by Metrohm Herisau Switzerland
1212 1111
Fig. 2: Rear of 782 IC Column Thermostat
77 Fuse holder
Changing the fuses, see section 2.2
88 Voltage selection insert with
display of voltage
99 Mains switch
For switching the instrument on/off:
I = ON 0 = OFF
1010 Serial number
1111 Connection for IC Column
Heater
1212 Mains connection plug
Mains connection see section 2.2
782 IC Column Thermostat
3
1 Introduction

1.3 Information about the Instructions for Use

Please read through these Instructions for Use carefully before you put the 782 IC Column Thermostat into operation. The Instructions for Use contain information and warnings to which the user must pay attention in order to assure safe operation of the instrument.

1.3.1 Organization

These 8.782.1003 Instructions for Use for the 782 IC Column Ther­mostat provide a comprehensive overview of the installation, startup procedure, operation, and technical specifications of this instrument. The Instructions for Use are organized as follows:
Section 1 Introduction
General description of instruments, parts and controls and safety notes
Section 2 Installation
Mains connection, connection of IC Column Heater
Section 3 Operation
Manual operation
Section 4 Appendix
Technical data, standard equipment, options, warranty, declarations of conformity, index
To find the required information on the instruments please use either the Table of contents or the Index at the back.
4
782 IC Column Thermostat

1.3.2 Notation and pictograms

The following notations and pictograms (symbols) are used in these In­structions for Use:
1212 Part or control of 782
2626 Part or control of 732/733
3333 Part or control of 761
range Menu item, parameter or entry
<OK> Button of 761
1.3 Information about the Instructions for Use
value of 761
Hazard
This symbol draws attention to a possible danger to life or of injury if the associated directions are not followed correctly.
Warning
This symbol draws attention to possible damage to instruments or instrument parts if the associated directions are not followed cor­rectly.
Caution
This symbol marks important infor­mation. First read the associated directions before you continue.
Comment
This symbol marks additional in­formation and tips.
782 IC Column Thermostat
5
1 Introduction

1.4 Safety notes

While electrical safety in the handling of the 782 IC Column Thermostat is assured in the context of the specifications IEC 1010-1 (protection class 1, degree of protection IP40), the following points should be noted:
Mains connection
Setting the mains voltage must be effected in accordance with the instructions in section 2.2.
Opening the instrument
Inside the instrument there are no parts which must be set or adjusted by the user.
If the 782 IC Column Thermostat is connected to the power supply, the instrument must not be opened nor must parts be removed from it, otherwise there is a danger of coming into contact with compo­nents which are live. Hence, always disconnect the instrument from all voltage sources before you open it and ensure that the mains cable is disconnected from mains connection 1212 !
Protection against static charges
Electronic components are sensitive to static charging and can be destroyed by discharges. Before you touch any of the components inside the 782 IC Column Thermostat, you should earth yourself and any tools you are using by touching an earthed object (e.g. housing of the instrument or a radiator) to eliminate any static charges which exist.
6
782 IC Column Thermostat

2 Installation

2.1 Setting up the instrument

2.1.1 Packaging

The 762 IC Interface is supplied together with the separately packed accessories in special packagings containing shock-absorbing foam linings designed to provide excellent protection. The instrument itself is packed in an evacuated polyethylene bag to prevent the ingress of dust. Please store all these special packagings as only they assure transport of the instrument free from damage.

2.1.2 Check

2.1 Setting up the instrument

After receipt, immediately check whether the shipment is complete and has arrived without damage (compare with delivery note and list of accessories in section 4.2). In the case of transport damage, see instructions in section 4.4.1 "Warranty".

2.1.3 Location

Position the instrument in the laboratory at a location convenient for op­eration, free from vibrations and protected against a corrosive atmos­phere and contamination by chemicals.

2.1.4 Arrangement of the instruments

The 782 IC Column Thermostat can be piled up together with other IC instruments (e.g. 732, 733, 709).
782 IC Column Thermostat
7
2 Installation

2.2 Mains connection

Follow the instructions below for connecting to the power supply. If the instrument is operated with the mains voltage set wrongly and/or wrong mains fuse there is a danger of fire!

2.2.1 Setting the mains voltage

Before switching on the 782 IC Column Thermostat for the first time, check that the mains voltage set on the instrument (see Fig. 3) matches the local mains voltage. If this is not the case, you must reset the mains voltage on the instrument as follows:
Disconnect mains cable
1
Disconnect mains cable from mains connection plug 1212 of the 782 IC Column Thermostat.
Remove fuse holder
2
Using a screwdriver, loosen fuse holder 77 and take out com- pletely.
Change mains voltage
3
Completely remove voltage selection insert 88 by hand, rotate it through 180° and reinsert it. The required mains voltage (115 or 230 V) must now be visible from the front.
Check fuses
4
Carefully take both fuses out of fuse holder 77 and check their specifications:
100…120 V 0.5 A (slow-blow) Metrohm-No. U.600.0013 220…240 V 0.25 A (slow-blow) Metrohm-No. U.600.0010
Insert fuses
5
Change both fuses if necessary and reinsert in fuse holder 77.
Install fuse holder
6
Push fuse holder 77 back into the opening of the 782 IC Column Thermostat by hand until it clicks into place properly.
8
782 IC Column Thermostat
2.2 Mains connection
100 – 120 V220 – 240 V
77
230
115
88
77 Fuse holder 88 Voltage selection
insert with display of voltage
99
1212
Fig. 3: Setting the mains voltage

2.2.2 Fuses

Two fuses 0.5 A/slow-blow for 100120 V or 0.25 A/slow-blow for 220…240 V are installed in the fuse holder 77 of the 782 IC Column Thermostat as standard.
Ensure that the instrument is never put into operation with fuses of another type, otherwise there is danger of fire!
For checking or changing fuses, proceed as described in section 2.2.1.

2.2.3 Mains cable and mains connection

Mains cable
The instrument is supplied with one of three mains cables
99 Mains switch
1212 Mains connection
plug
6.2122.020 with plug SEV 12 (Switzerland, …)
6.2122.040 with plug CEE(7), VII (Germany, …)
6.2133.070 with plug NEMA 5-15 (USA, …)
which are three-cored and fitted with a plug with an earthing pin. If a different plug has to be fitted, the yellow/green lead (IEC standard) must be connected to protective earth (protection class 1).
Any break in the earthing inside or outside the instrument can make it a hazard!
Mains connection
Plug the mains cable into mains connection plug 1212 of the 782 IC Col- umn Thermostat (see Fig. 3).

2.2.4 Switching the instrument on/off

The 782 IC Column Thermostat is switched on and off using mains switch 99. When the instrument is switched on the two temperature dis- plays 55 and 66 light up.
782 IC Column Thermostat
9
2 Installation

2.3 IC Column Heater

2.3.1 Installation in 733 IC Separation Center

Proceed as follows to insert the 2.782.0100 IC Column Heater into the 733 IC Separation Center:
Switch off all instruments
1
Switch off 782 IC Column Thermostat and 732 IC Detector
using the mains switch.
Insert IC Column Heater
2
Unscrew the four knurled screws 3434 from the top rear panel
3535 of the 733 IC Separation Center and remove rear panel
(see Fig. 5 of 732/733 Instructions for Use).
Position the IC Column Heater in the inner compartment and
insert the cable permanently attached in opening 2929 or 3333 of the rear panel 3535.
Replace rear panel 3535 and screw to the 733 IC Separation
Center using the four knurled screws 3434.
Connect column to injector
3
Mount a 6.2744.014 compression fitting on each side of the
6.1836.000 preheating capillary (procedure see section 2.5.3 of 732/733 Instructions for Use).
Connect the long, free end of the 6.1836.000 preheating
capillary to connection "4" of the injection valve 6868 (see Fig. 16 or Fig. 17 of 732/733 Instructions for Use).
Connect the short, free end of the 6.1836.000 preheating
capillary to the inlet end of the separating column 7676 (note flow direction).
Rinse column
4
Place a beaker beneath the column outlet.
Switch on 709 IC Pump and rinse column with eluent for ca.
10 min.
Switch off 709 IC Pump.
Connect column to detector block or suppressor
5
module
without suppressor module:
Screw outlet end of separating column 7676 to the inlet capillary 8282 permanently mounted on the detector block 8181.
with suppressor module:
Screw outlet end of separating column 7676 to suppressor inlet capillary 8989 using a 6.2744.010 compression fitting.
10
782 IC Column Thermostat
2.3 IC Column Heater
Insert column in IC Column Heater
6
Remove cover of IC Column Heater.
Insert separating column with inlet and outlet capillaries into
the IC Column Heater in such a way that the outlet end of the separting column is at the connecting cable side.
Insert inlet and outlet capillaries in the slices made for them
and close IC Column Heater with the cover.
Fix IC Column Heater
7
Insert the two 6.2027.060 holders in the mounting rails 8383 (see Fig. 16 or Fig. 17 of 732/733 Instruction for Use) and fasten IC Column Heater in the holder so that the connecting cable is on the upper side.
Connect IC Column Heater
8
Plug the connecting cable permanently attached to the IC Column Heater into connection 1111 "IC Column Heater“ of the 782 IC Column Thermostat (see Fig. 2).

2.3.2 Installation in 761 Compact IC

Proceed as follows to insert the 2.782.0100 IC Column Heater into the 761 Compact IC:
Switch off all instruments
1
Switch off 782 IC Column Thermostat and 761 Compact IC using the mains switch.
Insert IC Column Heater
2
Unscrew the four knurled screws 1515 from the top rear panel 1717 of the 761 Compact IC and remove rear panel (see Fig. 2
of 761 Instructions for Use).
Position the IC Column Heater in the inner compartment and insert the cable permanently attached in one of the openings 1111 of the rear panel 1717.
Replace rear panel 1717 and screw to the 761 Compact IC using the four knurled screws 1515.
Connect column to injector
3
Mount a 6.2744.014 compression fitting on each side of the
6.1836.000 preheating capillary (procedure see section 2.5.3 of 761 Instructions for Use).
Connect the long, free end of the 6.1836.000 preheating capillary instead of the column connection capillary 2828 at the injection valve 3232 (see Fig. 14 of 761 Instructions for Use).
Connect the short, free end of the 6.1836.000 preheating
782 IC Column Thermostat
11
2 Installation
capillary to the inlet end of the separating column 8181 (note flow direction).
Rinse column
4
Place a beaker beneath the column outlet.
Open software window for manual system control.
If necessary, modify Flow rate to the value suited for the
inserted separting column and click on <Send to unit> to send this value to the 761 Compact IC.
Switch on high-pressure pump (IC pump) by clicking <On> and
rinse column with eluent for ca. 10 min.
Switch off high-pressure pump by clicking <Off>.
Connect column to detector block or suppressor
5
module
without suppressor module:
Screw outlet end of separating column 8181 to the inlet capillary 4545 permanently mounted on the detector block 4646.
with suppressor module:
Screw outlet end of separating column 8181 to suppressor inlet capillary 9696 using a 6.2744.010 compression fitting.
Insert column in IC Column Heater
6
Remove cover of IC Column Heater.
Insert separating column with inlet and outlet capillaries into
the IC Column Heater in such a way that the outlet end of the separting column is at the connecting cable side.
Insert inlet and outlet capillaries in the slices made for them
and close IC Column Heater with the cover.
Fix IC Column Heater
7
Insert the two 6.2027.060 holders in the mounting rails 2727 (see Fig. 14 of 761 Instruction for Use) and fasten IC Column Heater in the holder so that the connecting cable is on the upper side.
Connect IC Column Heater
8
Plug the connecting cable permanently attached to the IC Column Heater into connection 1111 "IC Column Heater“ of the 782 IC Column Thermostat (see Fig. 2).
12
782 IC Column Thermostat
3 Operation
Switch instrument on/off
The 782 IC Column Thermostat is switched on and off using mains switch 99 (see Fig. 2) on the rear of the instrument:
After the instrument has been switched on the two tempera­ture displays 55 and 66 light up and show that the instrument is ready for use.

3 Operation

I Instrument switched on
0 Instrument switched off
Set temperature
The desired heating temperature is set using the two keys 22 and 33. A single pressing of these keys increases or de- creases the temperature value shown in the lower display 55 in green by 0.1 °C, continuous pressing of these keys leads to a continuous increase or decrease. A temperature set once will be preserved even after switching the instrument off.
Control temperature
As soon as the instrument is switched on the IC Column Heater is heated to the set temperature. The current temperature is shown in the upper display 66 in red, the set temperature is shown in the lower display 55 in green.
782 IC Column Thermostat
13
3 Operation
14
782 IC Column Thermostat

4 Appendix

4.1 Technical data
Temperature control
Temperature range 3075 °C, to be set in steps of 0.1 °C Accuracy ± 0.5 °C Stability ± 0.2 °C
Mains connection
Mains voltage 115 V: 100...120 V ± 10 %

4.1 Technical data

230 V: 220...240 V ± 10 % Switchable with voltage selection insert in fuse holder (see section 2.2.1)
Mains frequency 50...60 Hz Power consumption 12 VA Fuse 5 mm Ø, 20 mm length
100…120 V: 0.5 A (slow-blow) 220…240 V: 0.25 A (slow-blow)
Safety specifications
Construction / Testing According to IEC 1010 / EN 61010 / UL 3101-1,
protection class 1, degree of protection IP40
Safety directions The Instructions for Use include information and
warnings to which the user must pay attention in order to assure safe operation of the instrument.
Electromagnetic compatibility (EMC)
General requirements Standards met:
IEC61326/EN61326, NAMUR
Emitted interference Standards met:
EN55011, EN55022, EN 50081-1/2
Immunity to interference Standards met:
782 IC Column Thermostat
IEC61000-4-2/EN61000-4-2 (class 4), IEC61000-4-4/EN1000-4-4 (class 4), IEC61000-4-11/EN61000-4-11, IEC61000-4-14/EN61000-4-14, IEC61000-6-2/EN61000-6-2, EN50082-1/2
15
4 Appendix
Ambient temperature
Nominal operating range +5+45°C
(at 20…80 % atmospheric humidity)
Storage, transport –40+70°C
Housing
Material of cover Polyurethane rigid foam (PUR) with fire protection
for fire class UL94VO, CFC-free
Material of base Steel, enamelled
Dimensions
Width 255 mm Height 128 mm Depth 340 mm Weight 4.4 kg (incl. accessories)
16
782 IC Column Thermostat
4.2 Standard equipment
1.59
0.25
Subject to changes ! All dimensions are given in mm.
The 2.782.0010 IC Column Thermostat includes the following parts:
Quant. Order No. Description
1 1.782.0010 Control unit "IC Column Thermostat"
1 1.782.0100 IC Column Heater
1 6.1836.000 Preheating capillary
Length = 0.8 m
1 6.2027.060 Holder
to fix the IC Column Heater in the inner compartment of the 733 IC Separation Center or 761 Compact IC

4.2 Standard equipment

Opening d = 37 mm Set of 2
1 6.2122.0X0 Mains cable
to customer's specifications: Cable socket Cable connector
Type IEC 320/C 13 Type SEV 12 (CH…) ...............................6.2122.020
Type IEC 320/C 13 Type CEE (7), VII (D)...........................6.2122.040
Type CEE (22), V Type NEMA 5-15 (USA…) ......................6.2122.070
1 6.2744.014 PEEK compression fitting
For connection of 6.1831.010 PEEK capillaries or 6.1803.020 PTFE capillar­ies. Set of 2
1 8.782.1003 Instructions for Use (English)
for 782 IC Column Thermostat
782 IC Column Thermostat
17
4 Appendix

4.3 Warranty and conformity

4.3.1 Warranty

The warranty on our products is limited to defects that are traceable to material, construction or manufacturing error which occur within 12 months from the day of delivery. In such cases the defects will be recti­fied in our workshops free of charge. Transport costs are to be paid by the customer.
For day and night operation the warranty is limited to 6 months. Glass breakage in the case of electrodes or other parts is not covered
by the warranty. Checks which we are asked to carry out during the warranty period for reasons other than material or manufacturing faults will be invoiced. For parts manufactured by third parties, insofar as these constitute an appreciable part of our instrument, the warranty stipulations of the manufacturer in question apply.
With the regard to the guarantee of accuracy, the technical specifica­tions in the instruction manual are authoritative.
With regard to defects in material, construction or design as well as the absence of guaranteed features, the purchaser has no rights or claims except those mentioned above.
If damage of the packaging is evident on receipt of a consignment or if the goods show signs of transport damage after unpacking, the carrier must be informed immediately and a written damage report demanded. Lack of an official damage report releases Metrohm from any liability to pay compensation.
If any instruments and parts have to be returned, the original packaging should be used if at all possible. This applies above all to instruments, electrodes, burette cylinders and PTFE pistons. Before embedding them in wood shavings or similar material, the parts must be packed in a dustproof package (for instruments the use of a plastic bag is im­perative). If open assemblies are enclosed in the scope of delivery that are sensitive to electromagnetic voltages (e.g. data interfaces etc.) these must be returned in the associated original protective packaging (e.g. conductive protective bag). (Exception: assemblies with built-in voltage source belong in a non-conductive protective packaging). For damage which arises as a result of non-compliance with these instruc­tions, no warranty responsibility whatsoever will be accepted by Metrohm.
18
782 IC Column Thermostat

4.3.2 EU Declaration of conformity

EU Declaration of Conformity
The METROHM AG company, Herisau, Switzerland hereby certifies, that the in­strument:
782 IC Column Thermostat
meets the requirements of EC Directives 89/336/EWG and 73/23/EWG.
4.3 Warranty and conformity
Source of the specifications:
EN 50081-1/2 Electromagnetic compatibility, basic specification
Emitted Interference
EN 50082-1/2 Electromagnetic compatibility, basic specification
Interference Immunity
EN 61010 Safety requirements for electrical laboratory measurement
and control equipment
Description of the instrument:
PC controlled chromatography data system for remote control of IC instru­ments and automatic evaluation of chromatograms
Herisau, February 18, 2000
782 IC Column Thermostat
Dr. J. Frank Ch. Buchmann
Development Manager Production and
Quality Assurance Manager
19
4 Appendix

4.3.3 Certificate of conformity and system validation

Certificate of Conformity and System Validation
This is to certify the conformity to the standard specifications for electrical appli­ances and accessories, as well as to the standard specifications for security and to system validation issued by the manufacturing company.
Name of commodity: 782 IC Column Thermostat Name of manufacturer: Metrohm Ltd., Herisau, Switzerland Principal technical information: Voltages: 100-240 V
Frequency: 50-60 Hz
This Metrohm instrument has been built and has undergone final type testing according to the standards:
IEC61326/EN61326, NAMUR, EN55011, EN55022, EN 50081-1/2, IEC61000-4-2/EN61000-4-2 (class 4), IEC61000-4-4/EN1000-4-4 (class 4), IEC61000-4-11/EN61000-4-11, IEC61000-4-14/EN61000-4-14, IEC61000-6-2/EN61000-6-2, EN50082-1/2
Electromagnetic compatibility
IEC1010, EN61010, UL3101-1 — Security specifications
It has also been certified by the Swiss Electrotechnical Association (SEV), which is member of the International Certification Body (CB/IEC).
The technical specifications are documented in the instruction manual.
Metrohm Ltd. is holder of the SQS-certificate of the quality system ISO 9001 for quality assurance in design/development, production, installation and servicing.
Herisau, February 18, 2000
Dr. J. Frank Ch. Buchmann
Development Manager Production and
Quality Assurance Manager
20
782 IC Column Thermostat

4.4 Index

4.4 Index
A
Accuracy ............................................. 15
Ambient temperature........................... 16
Appendix ............................................. 15
Arrangement of the instruments ............ 7
C
Caution .................................................. 5
Certificate of conformity and system
validation .........................................20
Check .................................................... 7
Comment............................................... 5
Compression fitting (6.2744.014)
Mount ......................................... 10,11
Ordering designation....................... 17
Conformity ........................................... 19
Connection 1111
Connect IC Column Heater ........ 11,12
Figure ................................................ 3
Control temperature ............................ 13
Current temperature ............................ 13
D
Declaration of conformity .................... 19
Degree of protection......................... 6,15
Detector block 4646
Connect separating column ............ 12
Detector block 8181
Connect separating column ............ 10
Dimensions.......................................... 16
Display 55
Figure ................................................ 2
Ready for use display........................ 9
Display 66
Figure ................................................ 2
Ready for use display........................ 9
Temperature display........................ 13
E
Earthing .............................................. 6,9
Electrical safety ..................................... 6
Electromagnetic compatibility ............. 15
EMC..................................................... 15
Emitted interference ............................ 15
EU Declaration of conformity .............. 19
F
Front ...................................................... 2
Fuse holder 77
Change fuses ....................................8
Figure ............................................. 3,9
Fuses............................................. 8,9,15
H
Hazard................................................... 5
Holder (6.2027.060)
Installation .................................. 11,12
Ordering designation ...................... 17
Housing............................................... 16
I
IC Column Heater
Connection to control unit .......... 11,12
Fix............................................... 11,12
Insert separating column............ 11,12
Installation in 733 ............................ 10
Installation in 761 ............................ 11
Immunity to interference...................... 15
Information about the Instructions for
Use.................................................... 4
Injection valve 3232
Connect preheating capillary ..... 11,12
Injection valve 6868
Connect preheating capillary .......... 10
Inlet capillary 4545
Mount .............................................. 12
Inlet capillary 8282
Mount .............................................. 10
Installation ............................................. 7
Instructions for Use (8.782.1003).......... 4
Instrument description........................... 1
Introduction ........................................... 1
K
Key 11
Figure ................................................ 2
Key 22
Figure ................................................ 2
Set temperature............................... 13
Key 33
Figure ................................................ 2
Set temperature............................... 13
Key 44
Figure ................................................ 2
Knurled screw 1515
Open rear panel .............................. 11
Knurled screw 3434
Open rear panel .............................. 10
L
Location................................................. 7
M
Mains cable...................................... 9,17
Mains connection............................. 6,15
Procedure.......................................... 8
Mains connection plug 1212
Figure ............................................. 3,9
Mains connection.............................. 9
Mains frequency.................................. 15
Mains switch 99
Figure ..............................................3,9
Switching the instrument on/off.....9,13
Mains voltage.......................................15
Mounting rail 2727
Fix IC Column Heater .......................12
Mounting rail 8383
Fix IC Column Heater .......................11
N
Notation..................................................5
O
Opening the instrument .........................6
Operation .............................................13
Organization...........................................4
P
Packaging ..............................................7
Parts and controls..................................2
Pictograms.............................................5
Power consumption .............................15
Preheating capillary (6.1836.000)
Mount ..........................................10,11
Ordering designation .......................17
Protection class..............................6,9,15
Protective earth ......................................9
R
Rear........................................................3
Rear panel 1717
Open ................................................11
Rear panel 3535
Open ................................................10
S
Safety directions...................................15
Safety notes ...........................................6
Safety specifications ............................15
Separating column 7676
Connection to injection valve ...........10
Rinse ................................................10
Separating column 8181
Connection to injection valve ......11,12
Rinse ................................................12
Serial number 1010
Figure .................................................3
Set temperature ...................................13
Setting the mains voltage....................8,9
Setting up the instrument .......................7
Stability.................................................15
Standard equipment ............................17
Static charges ........................................6
Storage.................................................16
Switch intrument on/off ........................13
Switching the instrument on/off..............9
782 IC Column Thermostat
21
4 Appendix
System validation ................................20
T
Technical data ..................................... 15
Temperature range.............................. 15
Transport .......................................... 7,16
Transport damage............................ 7,18
V
Voltage selection insert 88
Change fuses ....................................8
Figure ............................................. 3,9
W
Warning .................................................5
Warranty ..............................................18
782 IC Column Thermostat
22
Loading...