INFICON VGC403 User Manual

Operating Manual
parach
parach
Incl. EC Declaration of Conformity

VGC402

Two-Channel Measurement and Control Unit

VGC403

Three-Channel Measurement and Control Unit
tinb07e1-e 2011-07 1
Contents
1 Introduction
1.1 Validity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.1 Part number . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 Firmware version . . . . . . . . . . . . . . . . . . . . . . 4
1.1.3 Type label . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Intended use. . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2.1 Liability and warranty . . . . . . . . . . . . . . . . . . . 4
1.3 Product versions . . . . . . . . . . . . . . . . . . . . . 4
1.4 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4.1 Personnel qualifications . . . . . . . . . . . . . . . . . 5
1.4.2 Illustration of residual dangers . . . . . . . . . . . . 5
1.4.3 General safety instructions . . . . . . . . . . . . . . 5
1.4.4 Disconnecting device . . . . . . . . . . . . . . . . . . . 6
2 Technical data
2.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.1 Mechanical data. . . . . . . . . . . . . . . . . . . . . . . 7
2.1.2 Ambience . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.3 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2 Mains connection. . . . . . . . . . . . . . . . . . . . . 7
3.3.6 CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3.7 RS232C . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 Operation
4.1 Front panel . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1.2 Control buttons. . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Switching on and off . . . . . . . . . . . . . . . . . 16
4.2.1 Switching on. . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2 Switching off. . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.3 Waiting time . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3 Operating modes . . . . . . . . . . . . . . . . . . . . 16
4.4 Measurement mode . . . . . . . . . . . . . . . . . . 17
4.4.1 Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.4.2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.4.3 Control button functions . . . . . . . . . . . . . . . . 17
4.5 Parameter mode. . . . . . . . . . . . . . . . . . . . . 19
4.5.1 Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.5.2 Parameter groups . . . . . . . . . . . . . . . . . . . . 19
4.5.3 Basic operation . . . . . . . . . . . . . . . . . . . . . . 20
2.3 Channels. . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3.1 Sensor connections . . . . . . . . . . . . . . . . . . . . 8
2.3.2 Sensor supply . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3.3 Measuring technique . . . . . . . . . . . . . . . . . . . 8
2.4 Switching functions. . . . . . . . . . . . . . . . . . . 8
2.4.1 Switching function relay . . . . . . . . . . . . . . . . . 9
2.4.2 Error signal relay . . . . . . . . . . . . . . . . . . . . . . 9
2.5 Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.5.1 Analog output. . . . . . . . . . . . . . . . . . . . . . . . . 9
2.5.2 Recorder output . . . . . . . . . . . . . . . . . . . . . . . 9
2.5.3 Computer interface . . . . . . . . . . . . . . . . . . . . 9
2.6 Scope of delivery . . . . . . . . . . . . . . . . . . . . . 9
3Installation
3.1 Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2 Mechanical installation . . . . . . . . . . . . . . . 10
3.2.1 Desktop unit . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.2.2 Control panel mounted. . . . . . . . . . . . . . . . . 10
3.2.3 Mounting the unit in a rack. . . . . . . . . . . . . . 11
3.3 Connecting . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3.1 Back side of the device . . . . . . . . . . . . . . . . 11
3.3.2 Mains connection . . . . . . . . . . . . . . . . . . . . . 12
3.3.3 Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.4 SENSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.3.5 RELAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5 Parameter
5.1 Switching function parameters (PArA SP)21
5.1.1 Fundamental terms . . . . . . . . . . . . . . . . . . . 21
5.1.2 Configuring switching functions . . . . . . . . . . 22
5.1.3 Adjustment range. . . . . . . . . . . . . . . . . . . . . 22
5.2 Sensor parameters (PArA SEn). . . . . . . . . 23
5.2.1 Measurement filter (FiLt) . . . . . . . . . . . . . . . 23
5.2.2 Gas type (GAS) . . . . . . . . . . . . . . . . . . . . . . 24
5.2.3 Measuring range (FS) . . . . . . . . . . . . . . . . . 24
5.2.4 Offset (oFS) . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.2.5 Degas function (dEGAS) . . . . . . . . . . . . . . . 25
5.2.6 Sensor activation (S-on) . . . . . . . . . . . . . . . 25
5.2.7 Switch-on threshold (t-on) . . . . . . . . . . . . . . 25
5.2.8 Sensor deactivation (S-oFF) . . . . . . . . . . . . 25
5.2.9 Switch-off threshold (t-off) . . . . . . . . . . . . . . 26
5.2.10 Emission (EMi) . . . . . . . . . . . . . . . . . . . . . . . 26
5.2.11 Filament selection (FiL) . . . . . . . . . . . . . . . . 26
5.2.12 Pirani range extension (PrE) . . . . . . . . . . . . 26
5.3 General parameters (PArA GEn) . . . . . . . . 27
5.3.1 Unit of measurement (unit) . . . . . . . . . . . . . 27
5.3.2 Baud rate (bAud) . . . . . . . . . . . . . . . . . . . . . 27
5.3.3 Display format (diGit) . . . . . . . . . . . . . . . . . . 27
5.3.4 Default parameters (dEF). . . . . . . . . . . . . . . 27
5.3.5 Recorder output (Ao) . . . . . . . . . . . . . . . . . . 27
5.3.6 Error signal relay (Err-r) . . . . . . . . . . . . . . . . 29
5.4 Test parameters (PArA tESt) . . . . . . . . . . . 29
5.4.1 Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.4.2 Firmware version (Pnr) . . . . . . . . . . . . . . . . 29
5.4.3 Watchdog control (dt-C). . . . . . . . . . . . . . . . 29
5.4.4 Torr lock (tr-L) . . . . . . . . . . . . . . . . . . . . . . . 30
5.4.5 Parameter setup lock (LoC) . . . . . . . . . . . . . 30
5.4.6 RAM test (rA-t). . . . . . . . . . . . . . . . . . . . . . . 30
5.4.7 EPROM test (EP-t). . . . . . . . . . . . . . . . . . . . 30
5.4.8 EEPROM test (EE-t) . . . . . . . . . . . . . . . . . . 30
5.4.9 Display test (di-t) . . . . . . . . . . . . . . . . . . . . . 30
5.4.10 A/D converter signal (Ad-S). . . . . . . . . . . . . 30
5.4.11 A/D converter ID (Ad-i) . . . . . . . . . . . . . . . . 31
5.4.12 I/O test (io-t). . . . . . . . . . . . . . . . . . . . . . . . . 31
5.4.13 RS232C test (rS-t) . . . . . . . . . . . . . . . . . . . . 31
6 Computer interface
6.1 Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.1.1 Connection. . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.1.2 Nomenclature . . . . . . . . . . . . . . . . . . . . . . . 32
6.2 Communication . . . . . . . . . . . . . . . . . . . . . 32
6.2.1 Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.2.2 Sending (Host --> Unit) . . . . . . . . . . . . . . . . 32
6.2.3 Receiving (Unit --> Host) . . . . . . . . . . . . . . . 33
6.2.4 Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.2.5 Number formats . . . . . . . . . . . . . . . . . . . . . . 33
6.2.6 Continuous transmission of measurements. 33
6.3.31 TEE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.32 TEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.33 TID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3.34 TIO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3.35 TKB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3.36 TLC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.3.37 TRA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.3.38 TRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.3.39 UNI. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.3.40 WDT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7 Maintenance and service
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 45
7.1.1 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.2 Program transfer mode . . . . . . . . . . . . . . . 45
7.2.1 Preparations and selection . . . . . . . . . . . . . 45
7.2.2 Program transfer . . . . . . . . . . . . . . . . . . . . . 45
7.2.3 Restarting . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.3.1 Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.3.2 CAO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.3.3 CAF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.3.4 Calibrating the unit . . . . . . . . . . . . . . . . . . . . 46
6.3 Mnemonics . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.3.2 AOM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.3.3 BAU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.3.4 COM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.3.5 COR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3.6 DCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3.7 DGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3.8 ERA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3.9 ERR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3.10 EUM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.3.11 FIL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.3.12 FSR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.3.13 FUM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3.14 GAS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3.15 HVC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3.16 ITR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3.17 LOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.3.18 OFC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.3.19 OFD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.3.20 PNR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.3.21 PR1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.3.22 PRE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.3.23 PRX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.3.24 RES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.3.25 SAV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.3.26 SC1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3.27 SP1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3.28 SPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3.29 TAD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.30 TDI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8 Troubleshooting
8.1 Fault indication . . . . . . . . . . . . . . . . . . . . . 48
8.2 Error messages . . . . . . . . . . . . . . . . . . . . . 48
8.3 Technical support . . . . . . . . . . . . . . . . . . . 48
9 Storage and disposal
9.1 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Appendix
Conversion tables . . . . . . . . . . . . . . . . . . . 50
Weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Linear measures . . . . . . . . . . . . . . . . . . . . . 50
Temperature . . . . . . . . . . . . . . . . . . . . . . . . 50
Default parameters . . . . . . . . . . . . . . . . . . 51
Literature . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
ETL Certification . . . . . . . . . . . . . . . . . . . . 55
EC Declaration of Conformity. . . . . . . . . . 55
tinb07e1-e 2011-07 Vacuum Gauge Controller 3

1 Introduction

Model:
PN: SN: V Hz W

1.1 Validity

1.1.1 Part number

This document applies to the following products:
Part number Product
398-020 VGC402
398-021 VGC403
The part number can be found on the type label which is attached to one side of the unit.

1.1.2 Firmware version

This Operating Manual is based on the firmware version 302-534-D.
If the unit does not work as described, please check if it is equipped with this firmware version. See Chapter 5.4.2 Firmware version (Pnr),  29.

1.1.3 Type label

There is a type label attached to one side of the unit. In all communication with INFICON, please state the infor­mation on the type label. For this purpose you may want to copy the information into the space provided below:

1.2 Intended use

The VGC402 and VGC403 Vacuum Gauge Controller is a display and control unit for vacuum gauges made by INFICON.
It is used together with vacuum gauges of the PSG, PCG, PEG, MPG, CDG, BPG, BCG and HPG series and is used for total pressure measurements. The vacuum gauges must be operated in accordance with their respective operating manuals.
In the following, the VGC402 or VGC403 Vacuum Gauge Controller will be referred to as «Vacuum Gauge Control­ler».

1.2.1 Liability and warranty

INFICON assumes no liability and the warranty becomes null and void if the end user or third parties
Disregard the information in this document
Use the product in a non-conforming manner
Make any kind of alterations (modifications, repair work, etc.) to the product
Use the product with accessories not listed in the cor­responding product documentation
We reserve the right to make technical changes without prior notice. The figures are non-committal.

1.3 Product versions

The Vacuum Gauge Controller is available in two differ­ent versions: VGC402 and VGC403. The two products differ from each other with regard to:
Number of channels
Fig. 1-1 Type label (example)
Number of switching functions
Power consumption
Weight
See Chapter 2 Technical data,  7.
This Operating Manual describes both the VGC402 and the VGC403.

1.4 Safety

1.4.1 Personnel qualifications

All work described in this document may only be carried out by persons who have suitable technical training and the necessary experience or who have been instructed by the end user of the product.

1.4.3 General safety instructions

For all work you are going to do, adhere to the applicable safety regulations.
Also observe all safety notes given in this document and forward the information to all other users of the product.
In particular, pay attention to the following safety notes:

1.4.2 Illustration of residual dangers

This Operating Manual illustrates safety notes concern­ing residual dangers as follows:
DANGER
DANGER indicates an imminently hazardous situation which, if not avoided, will result in death or severe injury.
WARNING
WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or severe injury.
CAUTION
CAUTION indicates a potentially hazardous situation which, if not avoided, may result in moderate or minor injury or in property dam­age.
DANGER
Mains power.
The Vacuum Gauge Controller contains parts which are connected to the mains supply.
Make sure that no objects enter through the louvers of the unit. Keep the unit dry. Do not open the unit.
NOTE: A note such as this one indicates particularly impor­tant, but not safety-relevant information.
Fig. 1-2 Do not insert objects through louvers and keep unit dry
WARNING
Improper use.
Improper use can damage the Vacuum Gauge Controller.
Use the Vacuum Gauge Controller only as in­tended by the manufacturer. See Chapter 1.2 Intended use,  4.
tinb07e1-e 2011-07 Vacuum Gauge Controller 5
WARNING
A
Improper installation and operation data.
Improper installation and operation data may damage the Vacuum Gauge Controller.
Strictly adhere to the stipulated installation and operation data.

1.4.4 Disconnecting device

The Vacuum Gauge Controller is equipped with a discon­necting device according to EN 61010-1.
The disconnecting device is located at the back of the Vacuum Gauge Controller. See Fig. 1-3,  6.
The disconnecting device must be readily identifiable and easily reached by the user.
In order to disconnect the Vacuum Gauge Controller from mains, you must unplug the mains cable.
Fig. 1-3 Back side of the VGC403
A Disconnecting device

2 Technical data

106.3
91.4
3.5
122.5
128.5
103
110
285
204.5
2.5

2.1 General data

2.1.1 Mechanical data

Dimensions Width: 106.3 mm
Height: 128.5 mm (3 HE) Depth: 207 mm
See Fig. 2-1,  7
Weight VGC402:
1.04 kg
VGC403:
1.16 kg
Use Desktop unit
Control panel mounted Mounting the unit in a rack

2.1.2 Ambience

Temperature Storage: -20… +60 °C
Operation: +5…+50 °C
Relative humidity Max. 80 % (bis 31 °C),
decreasing to max. 50 % (above 40 °C)
Use Indoors only
Altitude max. 2000 m NN
Pollution degree II
Protection type IP20

2.1.3 Operation

Manually Via 4 control buttons on the
front panel
Remote control Via RS232C interface
Fig. 2-1 Dimensions (in mm)

2.2 Mains connection

Voltage 90…250 VAC
Frequency 50…60 Hz
Power consumption VGC402:
Max. 45 W
VGC403: Max. 65 W
Overvoltage category II
Protection class 1
Connection European appliance con-
nector IEC 320 C14
tinb07e1-e 2011-07 Vacuum Gauge Controller 7

2.3 Channels

2.3.3 Measuring technique

2.3.1 Sensor connections

Number of channels VGC402: 2
VGC403: 3
Sensor connections per channel
Compatible sensors Pirani:
RJ45 (FCC 68) D-Sub, 15 pins, female (connected in parallel)
PSG400, PSG400-S, PSG100-S, PSG101-S, PSG500, PSG500-S, PSG502-S, PSG510-S, PSG512-S, PSG550, PSG552, PSG554
Pirani / Capacitance: PCG400, PCG400-S, PCG550, PCG552, PCG554
Cold cathode: PEG100
Cold cathode / Pirani: MPG400, MPG401
Capacitance: CDG025, CDG025D, CDG045, CDG045D, CDG100, CDG100D, CDG160D
Hot ionization / Pirani: BPG400, BPG402 HPG400
Hot ionization / Pirani / Capacitance: BCG450
Measuring ranges Sensor dependent
Error of measurement Gain error:
0.005 % FS
Offset error: 0.01 % FS
-1
Measuring rate 50 s
Display rate 10 s
Filter time constant Slow:
Unit of measurement mbar, Pa, Torr, Micron
Possible adjustments Linear sensors (CDG):
A/D converter Resolution > 16 bit
NOTE: The measurements of the BPG/BCG/HPG/CDGxxxD are transferred digitally.
-1
Approx. 1.0 s (f
Normal (nor): Approx. 0.3 s (f
Fast: Approx. 0.06 s (f
Zero-adjust
Logarithmic sensors (PSG, PCG, PEG, MPG, BPG, BCG, HPG): Fixed correction factors for
, Ar, H2, or a variable cor-
N
2
rection factor in the range
0.10…9.99
= 0.16 Hz)
g
= 0.53 Hz)
g
= 2.65 Hz)
g

2.3.2 Sensor supply

2.4 Switching functions

Voltage +24 VDC ±5 %
Current 500 mA
(750 mA short-time)
Fuse 900 mA via PTC element
Self-resetting after switching the unit off or unplugging the sensor.
The supply meets the requirements of a ground protective extra low voltage (SELV).
Number of switching functions
Assignment Can be configured any way
Delay time Filter time constant depen-
Adjustment range Sensor dependent
Hysteresis Linear sensors (CDG): 1
VGC402: 4 VGC403: 6
dent
% FS
Logarithmic sensors (PSG, PCG, PEG, MPG, BPG, BCG, HPG): 10 % of measurement

2.4.1 Switching function relay

2.5.2 Recorder output

Contact type Change-over contact, float-
ing
Load (ohmic) Max. 60 VDC, 0.5 A
Max. 30 VAC, 1 A
Lifetime Mechanical:
Connection D-sub, 25 pins, female.
7
cycles
10
Electrical:
5
cycles at maximum load
10
See Fig. 3-8,  13.

2.4.2 Error signal relay

Number 1
Delay time 20 ms
Contact type Change-over contact, float-
ing
Load (ohmic) Max. 60 VDC, 0.5 A
Max. 30 VAC, 1 A
Lifetime Mechanical:
7
cycles
10
Electrical:
5
cycles at maximum load
10
Number 1
Voltage range 0…10 VDC
Resolution 1 mV
Accuracy ± 20 mV
Internal resistance 3300 Ω
Relation between voltage and pressure
Connection D-Sub, 9 pins, male.
Programmable
See Fig. 3-9,  14.

2.5.3 Computer interface

Default RS232C
Protocol ACK/NAK
ASCII with 3-character mne­monics. Bidirectional data flow.
Signals Only TXD and RXD used
Baud rate 9600, 19200, 38400
Connection D-Sub, 9 pins, female.
See Fig. 3-10,  14.
Connection D-sub, 25 pins, female.
See Fig. 3-8,  13.

2.5 Outputs

2.5.1 Analog output

Number 1 per channel
Voltage range 0…10 VDC
Deviation from displayed value
Internal resistance 47 Ω
Relation between voltage and pressure
Connection D-Sub, 9 pins, male.
± 50 mV
Sensor dependent
See Fig. 3-9,  14.

2.6 Scope of delivery

Designation Number
Vacuum Gauge Control­ler
Mains cable 1
Rubber strip Rubber feet
Collar screws Plastic sleeves
CD-ROM manual 1
EC Declaration of Confor­mity
Installation manual 1
1
1 2
4 4
1
tinb07e1-e 2011-07 Vacuum Gauge Controller 9

3 Installation

105
91.4
122.5
112
M3 (ø3.5)

3.1 Unpacking

1 Visually inspect the transport packaging for signs
of external damage
2 Unpack the Vacuum Gauge Controller and put the
packaging material aside
NOTE: Keep the packaging material for later use. The Vacu­um Gauge Controller must be stored and transported in the original packaging material only.
3 Examine the Vacuum Gauge Controller for com-
pleteness
4 Visually inspect the Vacuum Gauge Controller for
signs of damage
WARNING
Damaged product.
Putting a damaged product into operation can be extremely dangerous.
Never attempt to put a damaged product into operation. Secure the damaged product from unintended operation. Send a damage report to the haulage company or the insurer.
1 Turn the Vacuum Gauge Controller upside down
as shown in Fig. 3-1,  10
2 Push the supplied rubber strip onto the lower edge
of the front panel
3 Stick the supplied rubber feet to the bottom of the
casing
Fig. 3-1 Using the product as a desk-top unit
4 Turn the Vacuum Gauge Controller back to normal
orientation and place it on the required location

3.2 Mechanical installation

The Vacuum Gauge Controller can be used as follows: As a desk-top unit, mounted in a control panel, or mounted in a 19" rack. In each of these cases you must pay attention to the following safety note:
CAUTION
Too high ambient temperature.
Exceeding the maximum permitted ambient temperature may damage the unit.
Make sure that the maximum permitted ambi­ent temperature is not exceeded and that the air can flow freely through the louvers. Do not expose the unit to direct sunlight.

3.2.1 Desktop unit

In order to use the Vacuum Gauge Controller as a desk­top unit, proceed as follows:

3.2.2 Control panel mounted

In order to mount the unit in a control panel, the following cutout is required:
Fig. 3-2 Control panel cutout (in mm)
1 Insert the Vacuum Gauge Controller into the cutout
2 Fasten the unit with four M3 screws
NOTE:
A
B
C
D
EFGH
I
J
K
In order to reduce the strain on the front panel it is recommended to support the bottom of the unit.

3.3 Connecting

3.3.1 Back side of the device

3.2.3 Mounting the unit in a rack

The Vacuum Gauge Controller is designed for installation into a rack chassis adapter according to DIN 41 494 (19", 3 HE). For this purpose, 4 collar screws and 4 plastic sleeves are supplied with the unit.
WARNING
Lower protection class of the rack.
If the product is installed in a rack, it is likely to lower the protection class of the rack (pro­tection from foreign bodies and water) e.g. according to the EN 60204-1 regulations for switching cabinets.
Take appropriate measures to restore the re­quired protection class of the rack.
Fig. 3-4, 11 shows the back side of the VGC403. The connection for channel 3 (Pos. C) is not available in the VGC402.
Fig. 3-4 Back side of the VGC403
A Sensor connection, channel 1 B Sensor connection, channel 2 C Sensor connection, channel 3 D CONTROL connection E Switch for program transfer mode F RS232C connection G Mains connection / disconnecting device HMains switch I Ground screw J RELAY connection K Screw for internal protective conductor. Do not loosen this screw!
Fig. 3-3 Mounting the unit in a rack
WARNING
NOTE: In order to reduce the strain on the front panel it is recommended to equip the rack chassis adapter with a guide rail.
NOTE:
Screw for internal protective conductor.
The internal protective conductor is connect­ed to the casing with a screw (Pos. K).
Do not turn or loosen this screw.
For safe and easy installation of heavy rack chassis adapters, it is recommended to equip the rack frame with slide rails.
1 Fasten the rack chassis adapter in the rack
2 Insert the Vacuum Gauge Controller into the rack
chassis adapter
The configuration of the available connections is described in the following sections.
3 Fasten the Vacuum Gauge Controller with the sup-
plied collar screws and plastic sleeves to the rack chassis adapter
tinb07e1-e 2011-07 Vacuum Gauge Controller 11

3.3.2 Mains connection

1
8
1
8
9
15

3.3.3 Ground

The mains connection (Fig. 3-4, 11, Pos. G) is designed for a mains cable which contains a European appliance connector on the device side.
A mains cable is supplied with the unit. If the plug is not compatible with your wall socket, you have to get a suit­able mains cable:
Three-conductor cable with protective ground
2
Conductor cross-section 3 × 1.5 mm
Fig. 3-5 Three-conductor cable with protective ground (example)
or larger
The ground screw (Fig. 3-4, 11, Pos. I) can be used to connect the Vacuum Gauge Controller with the protective ground of the pumping station.
1 If required: Connect the protective ground of the
pumping station with the ground screw. Use a pro­tective conductor.

3.3.4 SENSOR

The SENSOR connection is used to connect the sensors.
For each channel, there are two connections available which are connected in parallel: An 8-pin RJ45 appliance socket and a 15-pin D-Sub appliance socket. See Fig. 3-4, 11, Pos. A…C.
Pin assignment
WARNING
Mains power.
Improperly grounded devices can be ex­tremely dangerous in the event of a fault.
Use three-wire mains or extension cables with protective ground only. Plug the mains cable into wall sockets with protective ground only.
WARNING
No mains fuse.
The Vacuum Gauge Controller is not equipped with a mains fuse.
The wall socket must be protected with a fuse (max. 10 A).
1 Connect the European appliance connector of the
mains cord with the mains connection of the unit
Fig. 3-6 SENSOR appliance socket (RJ45)
1 +24 VDC 2PGND 3 Signal 4 Ident
Fig. 3-7 SENSOR appliance socket (D-Sub, 15-pin)
1 EMI-Status 2 Signal 3 Status 4 HV_EMI 5PGND 6n.c. 7 Degas 8 Supply
5 Signal-GND 6 Status 7HV_L 8 HV_EMI
9n.c. 10 Ident 11 Supply_CDG 12 Signal-GND 13 RXD 14 TXD 15 Chassis
2 Connect the plug of the mains cable with the wall
socket
NOTE: If the unit is installed in a switching cabinet, the mains power can be supplied via a switchable central power distributor.
CAUTION
1
13
14
25
Improper sensor.
Sensors which are not designed for use with the Vacuum Gauge Controller may damage the unit.
Operate the Vacuum Gauge Controller with proper sensors only. See Chapter 2.3.1 Sen­sor connections,  8.
CAUTION
Multiple connection.
Only one sensor may be connected to each of the channels. Otherwise the connected sen­sors will be damaged.
Never connect more than one sensor per channel.
Connecting
1 Channel 1: Connect the sensor with to the CH1-A
or CH1-B connection. Use a shielded 1:1 cable.
2 Channel 2: Connect the sensor with to the CH2-A
or CH2-B connection. Use a shielded 1:1 cable.
3 Channel 3: Connect the sensor with to the CH3-A
or CH3-B connection. Use a shielded 1:1 cable.

3.3.5 RELAY

The switching functions and the error monitoring system influence the state of several relays inside of the Vacuum Gauge Controller. The RELAY connection (Fig. 3-4, 11, Pos. J) allows to utilize the relay contacts for switching purposes. The relay contacts are potential-free (floating).
Pin assignment
Fig. 3-8 RELAY appliance socket (D-Sub, 25-pin)
1GND 2n.c. 3 Error break contact (NC) 4 SP 1 break contact (NC) 5 SP 1 common contact
(COM) 6 SP 1 make contact (NO) 7GND 8 SP 2 break contact (NC) 9 SP 2 common contact
(COM)
10 SP 2 make contact (NO) 11 SP 3 break contact (NC) 12 SP 3 common contact
(COM)
13 SP 3 make contact (NO)
n.c. not connected COM common contact NC break contact (normally closed) NO make contact (normally open)
14 Error make contact (NO) 15 Error common contact
(COM) 16 SP 4 break contact (NC) 17 SP 4 common contact
(COM) 18 SP 4 make contact (NO) 19 SP 5 break contact (NC) 20 SP 5 common contact
(COM) 21 SP 5 make contact (NO) 22 SP 6 break contact (NC) 23 SP 6 common contact
(COM) 24 SP 6 make contact (NO) 25 +24 VDC, 200 mA. Meets
the requirements of a ground
protective extra low voltage
(SELV)
NOTE: Pin 25 is used for supplying relays with a higher breaking capacity. The supply contact is protected at 200 mA with a PTC element. The element is self-reset­ting when switching the unit off or unplugging the RELAY connector.
WARNING
Hazardous voltage.
Voltages above 60 VDC or 30 VAC pose a shock hazard.
The RELAY connection may be used for switching voltages of max. 60 VDC or 30 VAC only. These voltages must meet the require­ments of a ground protective extra low volt­age (SELV).
1 Connect the peripheral components with the
RELAY connection. Use a shielded connection cable.
tinb07e1-e 2011-07 Vacuum Gauge Controller 13

3.3.6 CONTROL

1
5
6 9
1
5
6
9

3.3.7 RS232C

The CONTROL connection (Fig. 3-4, 11, Pos. D) con­tains the following signal pins:
Analog outputs for the signals of the individual chan­nels
Recorder output. This is a programmable analog out­put which can be assigned to one of the three chan­nels.
HV-EMI. Used to switch the high-vacuum circuit of the PEG sensor on and off. The signal levels are: On = +24 V. Off = 0 V. See Reference [7].
Pin assignment
Fig. 3-9 CONTROL appliance plug (D-Sub, 9-pin)
1 Analog output 1 2 Analog output 3 3GND 4 HV-EMI 3 5 HV-EMI 1
6 Analog output 2 7 Recorder output 8GND 9 HV-EMI 2
1 Connect the peripheral components with the CON-
TROL connection. Use a shielded connection cable.
The RS232C serial interface (Fig. 3-4, 11, Pos. F) allows remote control of the unit via a computer or a ter­minal. See Chapter 6 Computer interface,  32.
In addition, the interface may be used for firmware updates. See Chapter 7.2 Program transfer mode,  45.
Pin assignment
Fig. 3-10 RS232C appliance socket (D-Sub, 9-pin)
1 n.c. / SUP 2TXD 3RXD 4n.c. 5GND
6DSR 7n.c. 8CTS 9GND
1 Connect the serial interface of the computer with
the RS232C connection. Use a shielded cable.
NOTE: Use a serial extension cable with a 9-pin plug and a 9-pin socket. The cable must not contain any crossed wires.
NOTE: The analog outputs (pins 1, 2, 6) differ from the dis­played values by no more than ±50 mV.

4Operation

parach
ABC
D
E
F
G

4.1 Front panel

Fig. 4-1, 15 shows the front panel of the VGC403. The VGC402 is not equipped with the switching points SP5 and SP6 (Pos. A) and the display for the channel 3 (Pos. F).
Pressure unit
The top right corner (Pos. C) of the display indicates the pressure unit: mbar, Torr, Pa, or Micron.
Channels
There is a separate display area for each of the available channels (Pos. D, E, F). From the left to the right, this area displays the following information:
Display Significance
1, 2, 3 Channel number
FAIL (flashing) Error
DEG (illuminated) Degas function is acti-
vated
HV (illuminated) High-vacuum circuit is
activated
+18
-8.8.8.8.8
Measurement or status
message
OFS (illuminated) Offset correction is acti-
vated
Fig. 4-1 Front panel of the VGC403
A Switching function indicator B Parameter mode C Pressure unit D Display area for channel 1 E Display area for channel 2 F Display area for channel 3 G Control buttons

4.1.1 Display

Switching functions
The top left corner (Pos. A) of the display indicates the switching function states. An illuminated triangle above a number indicates that the pressure is above the lower threshold value. An illuminated triangle below a number indicates that the pressure is below the upper threshold value. See Fig. 5-1,  21.
Parameter mode
The PARA indicator (Pos. B) is illuminated when the unit is set to the parameter mode.
COR (illuminated) Gas type correction is
activated

4.1.2 Control buttons

CH
The CH button is used to select a channel. This may be necessary e.g. if you want to switch a particular sensor on or off, or if you want to modify the sensor parameters. The number of the currently selected channel is flashing for a few seconds.
PARA
The PARA button is used to select the parameter mode. The PARA indicator (Pos. B) is illuminated and you can modify various parameters. See Chapter 4.5 Parameter mode, 19.
Arrow buttons (DOWN/UP)
The arrow buttons are required for entering data in the parameter mode. Pressing one of these buttons will decrease or increase the currently displayed value. In the following, these buttons will be referred to as DOWN and UP, respectively.
tinb07e1-e 2011-07 Vacuum Gauge Controller 15

4.2 Switching on and off

4.3 Operating modes

4.2.1 Switching on

1 Switch the mains switch on. See Fig. 3-4,  11,
Pos. H.
After switching on, the Vacuum Gauge Controller will per­form the following actions:
•Self test
Identify all sensors
Restore the previously set parameters
Activate measurement mode
Adapt parameters (if a sensor type has changed meanwhile)

4.2.2 Switching off

1 Switch the mains switch off. See Fig. 3-4,  11,
Pos. H.

4.2.3 Waiting time

CAUTION
Delay time.
The Vacuum Gauge Controller can be set to one of the following operating modes:
Measurement mode
The measurement mode is the standard operating mode. It displays the pressure readings of the sensors. In case of an error, a status message is displayed instead. See Chapter 4.4 Measurement mode,  17.
Parameter mode
The parameter mode gives you access to various param­eters. You can check the parameter settings or modify them using the arrow buttons. This allows you to config­ure the Vacuum Gauge Controller. See Chapter 4.5 Parameter mode,  19.
Program transfer mode
The program transfer mode is used to transfer the latest version of the firmware to the Vacuum Gauge Controller. See Chapter 7.2 Program transfer mode,  45.
After switching off, the Vacuum Gauge Con­troller requires approximately 10 seconds to initialize again.
Wait for at least 10 seconds before you switch the Vacuum Gauge Controller on again.
NOTE: If the Vacuum Gauge Controller has been installed in a control panel or a rack, it can also be switched on and off via the central power distributor.

4.4 Measurement mode

parach
parach
parach

4.4.3 Control button functions

4.4.1 Selection

The Vacuum Gauge Controller automatically selects the measurement mode after it has been switched on.
When the unit is set to the parameter mode, it will auto­matically return to the measurement mode if no button is pressed for 10 seconds.

4.4.2 Description

The measurement mode is the standard operating mode. It displays the pressure readings of the sensors. A status message is displayed if the pressure exceeds the permis­sible range. See Tab. 4-1,  17.
Display Pressure
Er Hi Significantly above the permissible
range
The FAIL indicator flashes The error signal relay switches
Reading In the permissible range
4.4.3.1 Selecting a channel
1 Press the CH button
Fig. 4-2 Pressing the CH button
The unit selects the next channel. The number of the selected channel is flashing for a few seconds.
4.4.3.2 Selecting parameter mode
1 Press the PARA button
Er Lo Significantly below the permissible
range
The FAIL indicator flashes The error signal relay switches
Er x Error message of BPG400/HPG
x = Error code (High-Byte)
Er xx Error message of BCG, BPG402
xxH = Error code
noSEn See Tab. 4-2, 18
noid See Tab. 4-2, 18
oFF See Tab. 5-14, 30
Hot See Chapter 5.2.6, 25
SELF See Chapter 5.2.8, 25
CH 1 See Chapter 5.2.8, 25
CH 2 See Chapter 5.2.8, 25
CH 3 See Chapter 5.2.8, 25
LoC See Chapter 5.4.5, 30
Tab. 4-1 Display when in measurement mode
Channels which are not connected to a sensor display noSEn. This status message disappears after approxi­mately two minutes.
Fig. 4-3 Pressing the PARA button
The unit changes to the parameter mode. See Chapter 4.5 Parameter mode, 19. It will automatically return to the measurement mode if no button is pressed for 10 seconds.
4.4.3.3 Switching high-vacuum circuit on
The high-vacuum circuit of the following sensors can be switched on manually: PEG.
For this purpose the sensor control must be set to HAnd. See Chapter 5.2.6 Sensor activation (S-on),  25.
1 Press the CH button to select the required channel
2 Keep the UP button pressed for approximately 1
second
Fig. 4-4 Press the UP button for 1 second
The sensor on the selected channel is switched on. The HV indicator is illuminated. The display shows the pres­sure reading or a status message. See Tab. 4-1,  17.
tinb07e1-e 2011-07 Vacuum Gauge Controller 17
4.4.3.4 Switching degas function on
parach
parach
4.4.3.7 Identifying a sensor
The degas function of the following sensors can be switched on manually: BPG, BCG.
1 Press the CH button to select the required channel
2 Keep the UP button pressed for approximately 1
second. See Fig. 4-4, 17.
The degas function of the sensor on the selected channel is switched on. The DEG indicator is illuminated.
4.4.3.5 Switching high-vacuum circuit off
The high-vacuum circuit of the following sensors can be switched off manually: PEG.
For this purpose the sensor control must be set to HAnd. See Chapter 5.2.8 Sensor deactivation (S-oFF),  25.
1 Press the CH button to select the required channel
2 Keep the DOWN button pressed for approximately
1 second
1 Keep the UP and DOWN buttons pressed for
approximately 1 second
Fig. 4-6 Press the UP and DOWN buttons for 1 second
The display area of the individual sensor shows the con­nected sensors. See Tab. 4-2,  18.
Display Significance
PSG Pirani gauge (PSG)
PCG Combined gauge
Pirani / Capacitance (PCG)
PEG Cold cathode gauge (PEG)
MPG Combined gauge
Cold cathode / Pirani (MPG)
Fig. 4-5 Press the DOWN button for 1 second
The sensor on the selected channel is switched off. The HV indicator is dark. The display shows the status oFF.
4.4.3.6 Switching degas function off
The degas function of the following sensors can be switched off manually: BPG, BCG.
1 Press the CH button to select the required channel
2 Keep the DOWN button pressed for approximately
1 second. See Fig. 4-5,  18.
The degas function of the sensor on the selected channel is switched off. The DEG indicator is dark.
CDG Capacitive analog gauge, linear
(CDG025, CDG045, CDG100)
CDG d Capacitive digital gauge, linear
(CDG025D, CDG045D, CDG100D, CDG160D)
bPG Combined gauge
Hot ionization / Pirani (BPG400)
bPG2 Combined gauge
Hot ionization / Pirani (BPG402)
bCG Combined gauge
Hot ionization / Pirani / Capacitance
HPG Combined gauge
Hot ionization / Pirani (HPG)
noSEn No sensor found
noid No sensor identification found
Tab. 4-2 Sensor identification
NOTE: In the case of BPG/BCG/HPG sensors, the software version number of the sensor is also shown (e.g.
1.20).

4.5 Parameter mode

4.5.1 Selection

Pressing the PARA button switches from the measure­ment mode to the parameter mode. The PARA indicator is illuminated.
When the unit is set to the parameter mode, it will auto­matically return to the measurement mode if no button is pressed for 10 seconds. The PARA indicator is dark.

4.5.2 Parameter groups

The parameter mode gives you access to various param­eters. You can check the parameter settings or modify them using the arrow buttons. This allows you to config­ure the Vacuum Gauge Controller.
Tab. 4-3, 19 shows all available parameters.
Parameter group
PArA
SP
PArA
SEn
Parameter
SP1-L SP1-H SP2-L SP2-H SP3-L SP3-H SP4-L SP4-H SP5-L SP5-H SP6-L SP6-H
FiLt GAS FS oFS dEGAS S-on S-oFF EMi FiL PrE
Parameter group
PArA
tESt
Tab. 4-3 Parameter groups and their parameters
The available parameters are subdivided into the follow­ing parameter groups:
Switching function parameters (PArA SP)
These parameters are used to assign pressure depen­dent switching functions to the channels. The switching points 5 and 6 are only available in the VGC403. See Chapter 5.1 Switching function parameters (PArA SP), 21.
Sensor parameters (PArA SEn)
These parameters concern the sensor on the currently selected channel only. There is an individual set of parameters for each channel. See Chapter 5.2 Sensor parameters (PArA SEn),  23.
General parameters (PArA GEn)
These parameters are used for general configuration of the unit. The parameters affect all channels. See Chapter 5.3 General parameters (PArA GEn),  27.
Test parameters (PArA tESt)
This parameter group is used to check individual system functions. The parameter group is not required during normal operation. For this reason it must be accessed in a special way. See Chapter 5.4 Test parameters (PArA tESt), 29.
Parameter
Pnr dt-C tr-L LoC rA-t EP-t EE-t di-t Ad-S Ad-i CALib io-t rS-t
PArA
GEn
tinb07e1-e 2011-07 Vacuum Gauge Controller 19
unit bAud diGit dEF Ao Err-r

4.5.3 Basic operation

Starting at the measurement menu, you can select and modify a specific parameter as follows:
1 Press the CH button to select the required channel.
(This is only necessary if you want to modify a sen­sor parameter.)
2 Press the PARA button
The parameter menu is selected
The PARA indicator is illuminated
3 Use the arrow buttons to select the required
parameter group
The name of the parameter group is displayed
4 Press the PARA button to select the required
parameter
The name and the value of the parameter are displayed
5 Use the arrow keys (and the CH button, if neces-
sary) to modify the parameter value
The value of the parameter is changed
6 Repeat the steps 4 and 5 to change further param-
eters of the same parameter group
The unit returns to the measurement mode after the last parameter of a parameter group has been accessed. Parameter modifications are effective immediately, and they are saved in the EEPROM automatically.

5 Parameter

p
t
SP-H
SP-L
NO
COM
NC
NO
COM
NC
NO
COM
NC
5.1 Switching function
parameters (PArA SP)
This parameter group allows you to configure the switch­ing functions. The following switching function parame­ters are available in the VGC402:
SP1-L
SP1-H
SP2-L
SP2-H
SP3-L
SP3-H
SP4-L
SP4-H
In addition, the VGC403 is equipped with the following switching function parameters:
SP5-L
SP5-H
SP6-L
SP6-H
upper threshold value. This means that the common contact of the relay is connected to the make contact.
Hysteresis
In the pressure range between the two threshold values, the previous relay state is maintained. The relay does not switch in this range, and the relay state depends on the pressure curve history. See Fig. 5-1,  21.

5.1.1 Fundamental terms

Switching functions
The VGC402 is equipped with four relays which switch in dependance of the measured pressure. The relay con­tacts are potential-free and can be used for switching via the RELAY connection. See Chapter 3.3.5 RELAY,  13. In this context we speak of the switching functions 1 … 4.
The VGC403 contains a total of six switching function relays, i.e. the switching functions 1 … 6 are available in this unit.
Fig. 5-1 Behaviour of a switching function when the pressure
changes
pPressure tTime NO make contact (normally open) COM common contact NC break contact (normally closed)
The region between the threshold values generates a hysteresis (lag) between activating and deactivating of the relay. The hysteresis prevents the switching function from rapidly switching on and off when the pressure is close to one of the threshold values.
Threshold values
The switching behavior of the individual relays is deter­mined by two parameters each: The lower threshold value and the upper threshold value of the switching func­tion.
Lower threshold value SP-L: The lower threshold value is responsible for activating the assigned switching function. The relay switches on as soon as the pressure falls below the lower threshold value. This means that the common contact of the relay is connected to the make contact.
Upper threshold value SP-H: The upper threshold value is responsible for deacti­vating the assigned switching function. The relay switches off as soon as the pressure rises above the
tinb07e1-e 2011-07 Vacuum Gauge Controller 21

5.1.2 Configuring switching functions

5.1.3 Adjustment range

Prerequisite: The parameter group SP-P is selected
1 Press the PARA button to select the required
parameter
The name and the value of the parameter are displayed
2 Use the CH button to assign the switching function
to a channel
The switching functions can be assigned to the channels any way
The two threshold values of the switching func­tion are always assigned to the same channel
3 Use the arrow buttons to modify the threshold
value
The value of the parameter is changed
4 Repeat the steps 1 to 3 to change further parame-
ters of the same parameter group
Adjustment range of the lower threshold value
The lower threshold value of a switching function can be set in the following pressure range:
Sensor
SP-L min. [mbar]
PSG 2 × 10
PCG 2 × 10
PEG 1 × 10
MPG 5 × 10
-3
*) 5 × 102
-3
*) 1.5 × 103
-9
1 × 10-2
-9
5 × 10
SP-L max. [mbar]
2
CDG/CDG d FS/1000 FS
bPG/bPG2 1 × 10
HPG 1 × 10
bCG 1 × 10
Tab. 5-1 Adjustment range of the lower threshold values
*) 2 × 10-4 mbar if
PrE
is activated (see Chapter 5.2.12, 26)
-8
5 × 102
-6
1 × 103
-8
1.5 × 10
3
Adjustment range of the upper threshold value
The upper threshold value of a switching function can be set in the following pressure ranges:
Sensor
SP-H min. [mbar]
PSG 1.1 SP-L 5 × 10
PCG 1.1 SP-L 1.5 × 10
PEG 1.1 SP-L 1 × 10
MPG 1.1 SP-L 5 × 10
SP-H max.
[mbar]
2
3
-2
2
CDG/CDG d SP-L + 0.01 FS FS
bPG/bPG2 1.1 SP-L 5 × 10
HPG 1.1 SP-L 1 × 10
bCG 1.1 SP-L 1.5 × 10
Tab. 5-2 Adjustment range of the upper threshold values
2
3
3
This means that the hysteresis amounts to 10 % of the lower threshold value (logarithmic sensors) or to 1 % of the full-scale range (linear sensors) at least. If another sensor type is connected to a channel, the respective threshold values will be adjusted automatically if neces­sary.

5.2 Sensor parameters (PArA SEn)

p
t
p
t
p
t
There is an individual set of sensor parameters for each channel. Select the required channel before you change to the parameter menu and modify the sensor parame­ters.
The number of available parameters depends on the sen­sor type which is connected to the selected channel. See Tab. 5-3, 23.
Sensor
FiLt
GAS
FS
oFS
dEGAS
S-on
S-oFF
EMi
FiL
PSG 99 9
PCG 99 9
PEG 99 99
MPG 99
CDG 999
CDG d 999
FASt
Fast. The Vacuum Gauge Controller responds quickly to signal changes. This makes it rather sensitive to signal noise.
PrE
Fig. 5-2 Measurement with filter set to FASt (example)
nor
Normal. This is the default setting. It offers a good com­promise between the response time and the sensitivity to noise.
bPG 99
bPG2 9999
bCG 999
HPG 9
Tab. 5-3 Available sensor parameters

5.2.1 Measurement filter (FiLt)

The measurement filter improves measurements when the signal is noisy or disturbed. The filter affects the read­ings on the display, the RS232C output, the recorder out­put, and the switching functions. The analog outputs, however, are not affected.
The filter can be set to one of the following values:
Fig. 5-3 Measurement with filter set to nor (example)
SLo
Slow. The Vacuum Gauge Controller responds slowly to signal changes. This makes it less sensitive to signal noise. This setting is recommended for precise compari­son measurements.
Fig. 5-4 Measurement with filter set to SLo (example)
tinb07e1-e 2011-07 Vacuum Gauge Controller 23

5.2.2 Gas type (GAS)

Sensors are normally calibrated for a measurement in nitrogen or in air. The GAS parameter is used to configure the channel to other gas types.
Display Significance
n2 Nitrogen or air. No correction of any
kind is required.
Ar Argon. The pressure reading is
determined utilizing a correction fac­tor for argon. COR is illuminated.
H2 Hydrogen. The pressure reading is
determined utilizing a correction fac­tor for hydrogen. COR is illumi­nated.
Cor Other gases. The pressure reading
is determined utilizing a variable correction factor. COR is illumi­nated.
Tab. 5-4 GAS parameter settings
Cor
If you want to perform pressure measurements in a gas type without a fixed correction factor, you may multiply the pressure reading with a variable correction factor. To this end proceed as follows:
1 Set the GAS parameter to Cor
2 Press the PARA button
The correction factor is displayed
3 Use the arrow buttons the modify the correction
factor
The value of the parameter is changed
The COR indicator is illuminated
You can adjust the correction factor of a sensor in the range 0.109.99. The setting 1.00 returns the uncor­rected pressure reading.
NOTE: The gas type correction is effective only for the fol­lowing pressures: p < 10 sor), p < 10 mbar (PCG sensor), p < 1 mbar (BCG sensor).
-2
mbar (BPG/HPG/MPG sen-

5.2.3 Measuring range (FS)

Linear sensors (CDG) require specification of the full­scale range. You can set this value using the cursor but­tons. The following values are available:
0.01 mbar
0.01 Torr, 0.02 Torr, 0.05 Torr
0.10 mbar, 0.25 mbar, 0.50 mbar
0.10 Torr, 0.25 Torr, 0.50 Torr
1 mbar, 2 mbar, 5 mbar
1 Torr, 2 Torr, 5 Torr
10 mbar, 20 mbar, 50 mbar
10 Torr, 20 Torr, 50 Torr
100 mbar, 200 mbar, 500 mbar
100 Torr, 200 Torr, 500 Torr
1000 mbar, 1100 mbar
1000 Torr
2 bar, 5 bar, 10 bar, 50 bar

5.2.4 Offset (oFS)

When the offset correction is activated, a previously specified offset value will be subtracted from each pres­sure reading. This allows to conveniently measure the relative pressure with respect to a reference pressure.
The offset correction affects the readings on the display, the RS232C output, the recorder output, and the switch­ing functions. The analog outputs, however, are not affected.
Establishing an offset value and activating the offset correction
1 Select the oFS parameter
2 Keep the UP button pressed for approximately 2
seconds
The current pressure reading becomes the new offset value
The offset correction is activated
The OFS indicator is illuminated
Deactivating the offset correction
1 Select the oFS parameter
2 Press the DOWN button
The offset correction is deactivated
The display shows oFF
The OFS indicator is dark
Activating the offset correction
1 Select the oFS parameter
2 Press the UP button
The offset correction is activated
The OFS indicator is illuminated
NOTE: Always deactivate the offset correction before ad­justing the offset of a sensor.
Adjusting the zero point of a digital CDG
1 Select the oFS parameter
2 Keep the DOWN button pressed for approximately
2 seconds
The zero point of the sensor is adjusted
The OFS indicator is illuminated after >2 s and as long as the button is being pressed
NOTE: First adjust the sensor, then the Vacuum Gauge Con­troller.
NOTE: After adjusting the zero point, a zero value is dis­played. Due to the measuring resolution of the CDG (noise, drift), a zero with plus/minus several digits are displayed.

5.2.5 Degas function (dEGAS)

Ionization sensors with a hot cathode are sensitive with regard to depositions on the electrodes. These deposi­tions can cause signal fluctuations.
The dEGAS function is to bakeout and thereby clean the electrode system of the sensor. See Reference [11].
Hot
Hot start. The sensor automatically switches on when the unit is switched on. After a power failure the measure­ment will be resumed automatically.
CH 1
By channel 1. The subsequent parameter t-on is used to specify the switch-on threshold. The sensor is switched on when the pressure on channel 1 falls below the switch­on threshold.
CH 2
By channel 2. The subsequent parameter t-on is used to specify the switch-on threshold. The sensor is switched on when the pressure on channel 2 falls below the switch­on threshold.
CH 3
By channel 3. This setting is only available if the unit is equipped with three channels. The subsequent parame­ter t-on is used to specify the switch-on threshold. The sensor is switched on when the pressure on channel 3 falls below the switch-on threshold.

5.2.7 Switch-on threshold (t-on)

BPG402 gauge: The dEGAS function only affects the currently active fila­ment. See Reference [15].
Activating the degas function
1 Select the dEGAS parameter
2 Press the UP button
The degas function is activated
The display shows on
The DEG indicator is illuminated
The cleaning process takes approximately 3 minutes. Then the degas function switches off automatically. You may also deactivate this function manually.
Deactivating the degas function
1 Select the dEGAS parameter
2 Press the DOWN button
The degas function is deactivated
The display shows oFF
The DEG indicator is dark

5.2.6 Sensor activation (S-on)

This parameter determines how the sensor is switched on. The sensor activation can be set to one of the follow­ing values:
HAnd
Manually. The sensor can be switched on by pressing the UP button. See Chapter 4.4.3.3 Switching high-vacuum circuit on,  17.
This parameter is only available if the sensor activation parameter is set to CH 1, CH 2 or CH 3. See Chapter 5.2.6 Sensor activation (S-on),  25.
The t-on parameter is used to specify a switch-on thresh­old. The sensor is switched on when the pressure on the respective channel falls below the switch-on threshold.

5.2.8 Sensor deactivation (S-oFF)

This parameter determines how the sensor is switched off. The sensor deactivation can be set to one of the fol­lowing values:
HAnd
Manually. The sensor can be switched off by pressing the DOWN button. See Chapter 4.4.3.5 Switching high-vac­uum circuit off,  18.
SELF
Self control. The subsequent parameter t-off is used to specify the switch-off threshold. The sensor is switched off when the pressure at the sensor exceeds the switch­off threshold.
CH 1
By channel 1. The subsequent parameter t-off is used to specify the switch-off threshold. The sensor is switched off when the pressure on channel 1 exceeds the switch­off threshold.
CH 2
By channel 2. The subsequent parameter t-off is used to specify the switch-off threshold. The sensor is switched
tinb07e1-e 2011-07 Vacuum Gauge Controller 25
off when the pressure on channel 2 exceeds the switch­off threshold.
Display Significance
CH 3
By channel 3. This setting is only available if the unit is equipped with three channels. The subsequent parame­ter t-off is used to specify the switch-off threshold. The sensor is switched off when the pressure on channel 3 exceeds the switch-off threshold.

5.2.9 Switch-off threshold (t-off)

This parameter is only available if the sensor deactivation parameter is set to CH 1, CH 2 or CH 3. See Chapter 5.2.8 Sensor deactivation (S-oFF),  25.
The t-off parameter is used to specify a switch-off thresh­old. The sensor is switched off when the pressure on the respective channel exceeds the switch-off threshold.

5.2.10 Emission (EMi)

This parameter defines the rules for switching the emis­sion on.
Display Significance
on Range extension:
Display down to 5 × 10
Setpoint adjustment range down to 2 × 10
Tab. 5-7 PrE parameter values
-4
mbar
-5
mbar
Auto Emission is switched on and off by
the sensor electronics
MAn Emission is switched on and off
manually
Tab. 5-5 EMi parameter values

5.2.11 Filament selection (FiL)

This parameter defines the rules for selecting the active filament.
Display Significance
Auto The sensor alternately selects one
of the two filaments
Fil 1 Filament 1 ist active
Fil 2 Filament 2 ist active
Tab. 5-6 FiL parameter values

5.2.12 Pirani range extension (PrE)

Extend the display and the setpoint adjustment range.
Display Significance
oFF Normal operation

5.3 General parameters (PArA GEn)

These parameters are used for general configuration of the unit. The parameters affect all channels.

5.3.1 Unit of measurement (unit)

Unit of measurement for pressure values. The unit affects displayed pressure readings, threshold values, etc.
NOTE: The diGit parameter has no effect on CDG sensors.
NOTE: When PrE is enabled, the display of PSG and PCG gauges in the pressure range p < 10 duced by one digit.
-4
mbar is re-

5.3.4 Default parameters (dEF)

Reset all parameters to the default values (factory set­tings). Please note that this action cannot be undone.
Display Significance
bAr Pressure unit mbar or bar
torr Pressure unit Torr
PASC Pressure unit Pascal
uC Pressure unit Micron
Tab. 5-8 unit parameter values
The unit of measurement is indicated on the display. See Fig. 4-1, 15, Pos. C. For information on common pres­sure units refer to Section «Conversion tables»,  50.
NOTE: The pressure unit «Torr» can be locked. In this case torr is not available for selection. See Chapter 5.4.4 Torr lock (tr-L), 30.

5.3.2 Baud rate (bAud)

Transfer rate of the RS232C interface.
Display Significance
Proceed as follows to reset the parameters:
1 Select the dEF parameter
2 Press both the UP and the DOWN button at the
same time
The display shows SEt
All parameters are reset to the default values

5.3.5 Recorder output (Ao)

The recorder output is a programmable analog output. The recorder output voltage is a function of the pressure on the sensor. The relation between the pressure and the voltage is called the characteristic curve of the output. It can be selected.
Modifying parameter
1 Select the Ao parameter
2 Use the CH button to assign the recorder output to
a channel
3 Use the arrow buttons to select the characteristic
curve of the output
The value of the parameter is changed
9600 9600 Baud
19200 19200 Baud
38400 38400 Baud
Tab. 5-9 bAud parameter values

5.3.3 Display format (diGit)

Number of digits shown in the display.
Display Significance
2 Two digits
-1
e.g. 2.5
3 Three digits
e.g. 2.47
Tab. 5-10 diGit parameter values
tinb07e1-e 2011-07 Vacuum Gauge Controller 27
or 370
-1
or 373
Characteristic curves
Fundamentally we have to distinguish between logarith­mic and linear characteristic curves. A logarithmic char­acteristic curve is useful if the pressure range covers sev­eral orders of magnitude in the measurement. In this case it is appropriate to take the logarithm of the pressure and then scale the result in a suitable manner.
A linear characteristic curve is useful if the pressure range covers only a few orders of magnitude in the mea­surement. In this case the recorder output voltage is pro­portional to the pressure value. You can specify which pressure value will result in the maximum output voltage.
The available characteristic curves will be described in the following. In each case it is shown how to calculate the pressure p (in mbar) from the recorder output voltage U (in volts).
LoG
Logarithmic representation of the entire measuring range.
LoG +0
Logarithmic representation of a part of the measurement range (2.5 V/decade).
Sensor Pressure (in mbar)
PSG p = 10^[U/(10/7) - 4]
PCG p = 10^[U/(10/7) - 4]
PEG p = 10^[U/(10/7) - 9]
MPG p = 10^[U/(10/12) - 9]
CDG/CDG d p = 10^[U/(10/4) - 4] * FS
bPG/bPG2 p = 10^[U/(10/12) - 9]
bCG p = 10^[U/(10/12) - 9]
HPG p = 10^[U/(10/9) - 6]
LoG A
Logarithmic representation of the entire measuring range (compatible with VGC012, VGC023, VGC032).
Sensor Pressure (in mbar)
PSG p = 10^[U/(10/6) - 3]
PCG p = 10^[U/(10/7) - 4]
PEG p = 10^[U/(9/7) - 9 - 7/9]
MPG p = 10^[U/(10/11) - 8]
CDG/CDG d p = 10^[U/(10/4) - 4] * FS
bPG p = 10^[(U - 7.75)/0.75]
bPG2 p = 10^[U - 8]
Sensor Pressure (in mbar)
All types p = 10^[U/(10/4) - 4]
LoG +3
Logarithmic representation of a part of the measurement range (2.5 V/decade).
Sensor Pressure (in mbar)
All types p = 10^[U/(10/4) - 1]
LoGC1
Logarithmic representation matched to the following sen­sor combination:
PSG on channel 1
PEG on channel 2
Sensor Pressure (in mbar)
PSG + PEG p = 10^[U/(10/12) - 9]
LoGC2
Logarithmic representation matched to the following sen­sor combination:
CDG on channel 1
CDG on channel 2
This characteristic curve is only useful if the sensors have different measuring ranges. The total measuring range of the sensor combination is represented logarithmically in the range 0…10 V.
bCG p = 10^[(U - 7.75)/0.75]
HPG p = 10^[U/(10/9) - 6]
LoGC3
Logarithmic representation matched to the following sen­sor combination:
LoG -6
Logarithmic representation of a part of the measurement range (2.5 V/decade).
CDG on channel 1
CDG on channel 2
CDG on channel 3
This characteristic curve is only useful if the sensors have
Sensor Pressure (in mbar)
different measuring ranges. The total measuring range of the sensor combination is represented logarithmically in
All types p = 10^[U/(10/4) - 10]
LoG -3
Logarithmic representation of a part of the measurement range (2.5 V/decade).
Sensor Pressure (in mbar)
All types p = 10^[U/(10/4) - 7]
the range 0…10 V.
NOTE: The three sensors must be sorted with regard to their measuring range (FS). The sort order may be increas­ing or decreasing.
Lin n
Linear representation. U = 10 V is equivalent of p = 10
n
mbar. The exponent n may be any integer value in the range -10… +3.
Sensor Pressure (in mbar)
All types p = U/10 * 10
iM221
Logarithmic representation of the IM221 controller (1 V/decade) U = 8 V is equivalent of p = 10
n
-2
mbar.

5.4 Test parameters (PArA tESt)

This parameter group is intended for test and service pur­poses. It is used to examine additional system data, to set basic system parameters, and to check individual system functions.
The parameter group tESt is not required for normal oper­ation. For this reason it is not accessible normally.
Controller Pressure (in mbar)
IM221 p = 10^[U - 10]
LoGC4
Logarithmic representation of 12 decades (0.83 V / decade) matched to the following sensor combination:
PCG on channel 1
BPG402 on channel 2
Sensor Pressure (in mbar)
PCG+BPG402 p = 10^[U/(10/12) - 9]
U = 10 V is equivalent of p = 1000 mbar. The switching point between the sensors is 10
PM411
Nonlinear characteristic curve of the output as with the PM411 board.
-2
mbar.

5.3.6 Error signal relay (Err-r)

The Err-r parameter is used to specify what kind of error will trigger the error signal relay.
Display Significance
ALL All errors
no SE Device errors

5.4.1 Selection

When switching on
The parameter group tESt becomes available if you switch on the Vacuum Gauge Controller as follows:
1 Press the PARA button and keep it pressed
2 Switch the mains switch on. See Fig. 3-4, 11,
Pos. H.
The Vacuum Gauge Controller is switched on
The parameter group tESt is selected
The PARA indicator is illuminated
During normal operation
During normal operation it is also possible to activate the parameter group tESt from the measurement mode:
1 Press the PARA button
The parameter menu is selected
The PARA indicator is illuminated
2 Keep the UP and DOWN buttons pressed for
approximately 5 seconds
The firmware version is displayed
The parameter group tESt is selected
NOTE: When the parameter group tESt is selected, the Vac­uum Gauge Controller will not automatically return to the measurement mode. In order to return to the mea­surement mode, press the PARA button repeatedly until all test parameters have been run through.
CH 1 Sensor 1 and device errors
CH 2 Sensor 2 and device errors
CH 3 Sensor 3 and device errors
Tab. 5-11 Err-r parameter values

5.4.2 Firmware version (Pnr)

Displays the firmware version number. The last character represents the modification index.
Example: 302-534-D

5.4.3 Watchdog control (dt-C)

Behavior of the system monitoring system (watchdog control) in the event of an error.
Display Significance
Auto An error message from the watch-
dog control is acknowledged auto­matically after 2 seconds
tinb07e1-e 2011-07 Vacuum Gauge Controller 29
Display Significance
oFF An error message from the watch-
dog control must be acknowledged by the user
Tab. 5-12 dt-C parameter values

5.4.7 EPROM test (EP-t)

Test the program memory. Press the UP button to start the test.
Display Significance
run Test is running

5.4.4 Torr lock (tr-L)

This parameter affects the general parameter unit. When the lock is activated, the unit of measurement «Torr» can­not be selected. See Chapter 5.3.1 Unit of measurement (unit), 27.
Display Significance
oFF Unit of measurement «Torr» can be
selected
on Unit of measurement «Torr» cannot
be selected
Tab. 5-13 tr-L parameter values

5.4.5 Parameter setup lock (LoC)

This parameter affects the parameter mode. When the lock is activated, the user can inspect but not modify parameter values.
Display Significance
oFF Parameters can be inspected and
modified
on Parameters can be inspected only
Tab. 5-14 LoC parameter values
PASS Test completed without errors
Err Test completed and errors detected.
A four-digit checksum is displayed.
Tab. 5-16 EPROM test
Please contact your local INFICON service center if the test fails repeatedly.

5.4.8 EEPROM test (EE-t)

Test the parameter memory. Press the UP button to start the test.
Display Significance
run Test is running
PASS Test completed without errors
Err Test completed and errors detected
Tab. 5-17 EEPROM test
Please contact your local INFICON service center if the test fails repeatedly.

5.4.9 Display test (di-t)

Test the display. In this test all segments of the display are illuminated simultaneously at first. Then the individual segments of the display are activated one after the other.

5.4.6 RAM test (rA-t)

Test the main memory. Press the UP button to start the test.
Display Significance
run Test is running
PASS Test completed without errors
Err Test completed and errors detected
Tab. 5-15 RAM test
Please contact your local INFICON service center if the test fails repeatedly.
Press the UP button to start the test.

5.4.10 A/D converter signal (Ad-S)

Display the A/D converter output signal (in volts) for each of the channels. When applying a reference voltage to the input signal pin of the SENSOR connection, this allows you to check the A/D converters of the respective chan­nel. See Chapter 3.3.4 SENSOR,  12.
NOTE: When the signal input is not connected, a quickly fluctuating value is displayed because of the high measurement sensitivity of the unit.
Press the UP button to start the test.

5.4.11 A/D converter ID (Ad-i)

For each channel, display a signal (in volts) which is caused by a resistor inside the connected sensor. This signal is used for identification of the connected sensors.
Press the UP button to start the test.

5.4.12 I/O test (io-t)

Test all relays. In this test the relays are switched on and off one after the other, and the relay states are shown on the display. A circuit indicator or an ohmmeter may be used to verify the relay states on the RELAY connection. See Chapter 3.3.5 RELAY,  13.
Press the UP button to start the test.
CAUTION
Relay test.
In this test the relays switch irrespective of the actual pressure. This may cause unintended switching of devices.
The test starts as soon as the rS-t parameter has been selected. You may e.g. use a terminal program and input characters via the keyboard. Each of the input characters is returned to the terminal from the Vacuum Gauge Con­troller. The data transfer between the two units is visible on the terminal screen only.
Press the PARA button to quit the test and to return to the measurement mode.
Unplug the RELAY connection before per­forming a relay test.
Display Significance
off All relays switched off
r1-H r1-L
r2-H r2-L
r3-H r3-L
r4-H r4-L
r5-H r5-L
r6-H r6-L
r7-H r7-L
Switching function 1 relay on Switching function 1 relay off
Switching function 2 relay on Switching function 2 relay off
Switching function 3 relay on Switching function 3 relay off
Switching function 4 relay on Switching function 4 relay off
Switching function 5 relay on Switching function 5 relay off
Switching function 6 relay on Switching function 6 relay off
Error signal relay on Error signal relay off
Tab. 5-18 Relay test

5.4.13 RS232C test (rS-t)

Test the RS232C interface. In this test, the Vacuum Gauge Controller echoes each character received from from the serial interface back to the interface.
tinb07e1-e 2011-07 Vacuum Gauge Controller 31

6 Computer interface

6.1 Basics

6.1.1 Connection

The Vacuum Gauge Controller is able to communicate with a computer via a serial interface (RS232C). The con­nection socket and the required connection cable are described in Chapter 3.3.7 RS232C,  14.

6.1.2 Nomenclature

The following terms and symbolic styles will be used in the description of the computer interface:
Term Significance
Host Computer or terminal
Sending (S) Data transfer from the Host to the
Vacuum Gauge Controller
Receiving (R) Data transfer from the Vacuum
Gauge Controller to the Host
ASCII American Standard Code for Infor-
mation Interchange
Tab. 6-1 Terms
Square brackets [...]
Square brackets identify optional parameters. The items enclosed by the brackets may appear, but they are not essential. The brackets are not actually used in the com­mand.
Term Value Significance
<CR> 0Dh Carriage return. Carriage
return.
<NAK> 15h Negative acknowledge.
Negative acknowledge.
Tab. 6-2 Control characters

6.2 Communication

6.2.1 Protocol

The following protocol is used in the communication:
8 data bits
No parity bit
•1 stop bit
The baud rate can be selected. See Chapter 5.3.2 Baud rate (bAud), 27. Hardware handshake is not used.
Messages are transferred as ASCII strings. Blanks (spaces) in the string are ignored. The information is exchanged bidirectionally, i.e. data and control com­mands can be exchanged in both directions.
The input buffer of the Host must have a capacity of at least 75 bytes.

6.2.2 Sending (Host --> Unit)

Angle brackets <...>
Abbreviations enclosed by angle brackes identify control characters. The entire expession including the brackets is replaced by a numerical value. See Tab. 6-2,  32.
Term Value Significance
<EXT> 03h End of text (Ctrl-C). Reset
the interface. Delete the input buffer.
<ENQ> 05h Enquiry (Ctrl-E). Request
data transmission.
<ACK> 06h Acknowledge. Positive
acknowledge.
<LF> 0Ah Line feed. Line feed.
Messages of the Host are composed of mnemonics and parameters. Mnemonics are command abbreviations and always consist of three ASCII characters. See Chapter 6.3 Mnemonics, 34. The control characters <CR> or <CR><LF> signal the end of the message.
The Vacuum Gauge Controller checks every message it receives. Afterwards it sends a positive or a negative acknowledgement to the Host.
In a symbolic representation this process can be illus­trated as follows:
S: Mnemonic [parameters]<CR>[<LF>]
S: <ENQ>
R: <ACK><CR><LF> or <NAK><CR><LF>

6.2.3 Receiving (Unit --> Host)

The Host may request data from the unit. To this end the Host first sends a message which describes what kind of data is requested. The Vacuum Gauge Controller then stores the requested data in the output buffer of the inter­face.
Afterwards the Host sends the control character <ENQ> to the unit. This prompts the unit to send to contents of the output buffer to the Host.
In a symbolic representation this process can be illus­trated as follows:
S: Mnemonic [parameters]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: Data<CR><LF>
If the Vacuum Gauge Controller receives a message which cannot be interpreted (syntax error) it stores the respective error status in the output buffer. See Chapter 6.3.9 ERR,  36.
R: 0,9.0000E-01,2.2000E+00<CR><LF>
Setting the filter
S: FIL,1,2,1<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: 1,2,1<CR><LF>
Behavior in case of a syntax error
S: FOL,1,2,1<CR>[<LF>]
R: <NAK><CR><LF>
S: <ENQ>
R: 0001<CR><LF>

6.2.5 Number formats

The following data is always stored in the exponential for­mat in the Vacuum Gauge Controller:
Pressure values
Offset values
Threshold values

6.2.4 Examples

Inquiring the sensor identification
S: TID<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: PSG,CDG,noSen<CR><LF>
Inquiring the sensor status
S: HVC<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: 0,0,0<CR><LF>
Inquiring parameters of the switching function 1
S: SP1<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: 0,2.0000E-01,5.0000E+00<CR><LF>
Inquiring parameters of the switching function 2
S: SP2,0,9E-1,2.2E0<CR>[<LF>]
R: <ACK><CR><LF>
Output
The above data is always output in the exponential for­mat. A five-digit mantissa and a two-digit exponent are used. Both parts of the number may contain a sign.
Symbolic representation: ±a.aaaaE±aa Example: 1.2500E-01
In the case of logarithmic sensors (PSG, PCG, PEG, MPG, BPG, BCG, HPG) the last two digits of the man­tissa are always zero. Linear sensors (CDG) use all digits of the mantissa.
Input
The above data may be input either in the exponential for­mat or in the fixed point format. The input data is automat­ically converted to the exponential format by the unit.
Example: 1.25E-1 and 0.125 are both valid input data.

6.2.6 Continuous transmission of measurements

After the unit has been switched on, it starts to continu­ously send measurements to the serial interface. By default one set of measurements is sent every second.
The continuous measurement transmission stops when the Host sends a character to the serial interface. The transmission can be resumed with the COM command. See Chapter 6.3.4 COM,  35.
tinb07e1-e 2011-07 Vacuum Gauge Controller 33

6.3 Mnemonics

Mnemonic Significance

6.3.1 Overview

Mnemonic Significance
AOM Analog output mode. Characteristic
curve of the recorder output.
BAU Baud rate. Transfer rate of the
RS232C interface.
COM Continuous mode. Continuous
transmission of measurements to the serial interface.
COR Correction factor.
DCD Display control digits. Number of
digits shown in the display.
DGS Degas.
ERA Error relay allocation.
ERR Error status.
EUM Switch the emmision.
FIL Filter. Measurement filter.
FUM Select the filament.
FSR Full scale range. Full scale range of
linear sensors (CDG).
GAS Gas type correction.
HVC High vacuum circuit on/off. Switch
the high vacuum circuit of sensors on/off.
PRE Pirani range extension.
PRX Pressure sensors. Pressure read-
ings of all sensors.
RES Reset. Reset the serial interface.
SAV Save parameters to EEPROM.
SC1 Sensor 1 control.
SC2 Sensor 2 control.
SC3 Sensor 3 control.
SP1 Setpoint 1. Switching function 1.
……
SP6 Setpoint 6. Switching function 6.
SPS Setpoint status. Switching function
status.
TAD Test A/D converter. Test the A/D
converter.
TDI Test display. Test the display.
TEE Test EEPROM. Test the EEPROM.
TEP Test EPROM. Test the EPROM.
TID Transmitter identification. Sensor
identification.
TIO Test I/O. Test the relays.
TKB Test keyboard. Test the keyboard.
TLC Torr lock.
ITR Read a data string from the
BPG/BCG/HPG/CDGxxxD sensor.
LOC Parameter setup lock.
OFC Offset correction. Offset correction
function for linear sensors (CDG).
OFD Offset display. Offset correction val-
ues for linear sensors (CDG).
PNR Program number. Firmware version
number.
PR1 Pressure sensor 1. Pressure read-
ing of sensor 1.
PR2 Pressure sensor 2. Pressure read-
ing of sensor 2.
PR3 Pressure sensor 3. Pressure read-
ing of sensor 3.
TRA Test RAM. Test the RAM.
TRS Test RS232C interface. Test the
RS232C interface.
UNI Unit of measurement.
WDT Watchdog control.
Tab. 6-3 Mnemonics

6.3.2 AOM

Analog output mode. Characteristic curve of the recorder output. See Chapter 5.3.5 Recorder output (Ao),  27.
S: AOM[,a,b]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
Parameter Significance
a Transfer rate
0 = 9600 baud (default) 1 = 19200 Baud 2 = 38400 Baud
R: a,b<CR><LF>
Parameter Significance
a Channel number
0 = Channel 1 1 = Channel 2 2 = Channel 3
b Characteristic curve
0 = Logarithmic LoG 1 = Logarithmic LoG A
2 = Logarithmic LoG -6 3 = Logarithmic LoG -3 4 = Logarithmic LoG +0 5 = Logarithmic LoG +3
6 = Logarithmic LoGC1 7 = Logarithmic LoGC2 8 = Logarithmic LoGC3
9 = Linear Lin -10 10 = Linear Lin -9 11 = Linear Lin -8 12 = Linear Lin -7 13 = Linear Lin -6 14 = Linear Lin -5 15 = Linear Lin -4 16 = Linear Lin -3 17 = Linear Lin -2 18 = Linear Lin -1 19 = Linear Lin +0 20 = Linear Lin +1 21 = Linear Lin +2 22 = Linear Lin +3
23 = iM221 24 = Logarithmic LoGC4 25 = PM411

6.3.3 BAU

Baud rate. Transfer rate of the RS232C interface. See Chapter 5.3.2 Baud rate (bAud),  27.
S: BAU[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
NOTE: The acknowledgement of the BAU command will al­ready be sent with the changed transfer rate.

6.3.4 COM

Continuous mode. Continuous transmission of measure­ments to the serial interface.
S: COM,a<CR>[<LF>]
R: <ACK><CR><LF>
The acknowledgement is immediately followed by the continuous measurement transmission. The measure­ments are always output in the exponential format.
R: b,±c.ccccE±cc,d,±e.eeeeE±ee,f,±g.ggggE±gg <CR><LF>
Parameter Significance
aPeriod
0 = 100 milliseconds 1 = 1 second (default) 2 = 1 minute
b Status of channel 1
0 = Measurement data ok 1 = Underrange 2 = Overrange 3 = Sensor error 4 = Sensor switched off 5 = No sensor 6 = Identification error 7 = BPG/BCG/HPG error
±c.ccccE±cc Reading of sensor 1 in current unit
of measurement
d Status of channel 2
(see above)
±e.eeeeE±ee Reading of sensor 2 in current unit
of measurement
f Status of channel 3
(see above)
±g.ggggE±gg Reading of sensor 3 in current unit
of measurement
R: a<CR><LF>
tinb07e1-e 2011-07 Vacuum Gauge Controller 35

6.3.5 COR

Correction factor. See Chapter 5.2.2 Gas type (GAS), 24.
S: COR[,a.aa,b.bb,c.cc]<CR>[<LF>]
R: <ACK><CR><LF>
Parameter Significance
a Sensor 1
0 = Degassing off (default) 1 = Degassing on
S: <ENQ>
R: a.aa,b.bb,c.cc<CR><LF>
Parameter Significance
a.aa Correction factor of channel 1
0.10…9.99 (default: 1.00)
b.bb Correction factor of channel 2
(see above)
c.cc Correction factor of channel 3
(see above)
NOTE: The correction factor is only used when the gas type is set to «Other gas». See Chapter 6.3.14 GAS, 38.

6.3.6 DCD

Display control digits. Number of digits shown in the dis­play. See Chapter 5.3.3 Display format (diGit),  27.
S: DCD[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
b Sensor 2
(see above)
NOTE: The degas function is switched off automatically af­ter 3 minutes. It may be also be stopped prematurely.

6.3.8 ERA

Error relay allocation. See Chapter 5.3.6 Error signal relay (Err-r),  29.
S: ERA[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
Parameter Significance
a Error relay allocation
0 = All errors 1 = Device errors 2 = Sensor 1 and device errors 3 = Sensor 2 and device errors 4 = Sensor 3 and device errors
R: a<CR><LF>
Parameter Significance
a Number of digits
2 = 2 digits (default) 3 = 3 digits
NOTE: When PrE is enabled, the display of PSG and PCG gauges in the pressure range p < 10 duced by one digit.
-4
mbar is re-

6.3.7 DGS

Degas. See Chapter 5.2.5 Degas function (dEGAS), 25.
S: DGS[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>

6.3.9 ERR

Error status.
S: ERR<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: aaaa<CR><LF>
Parameter Significance
aaaa Error status
0000 = No error 1000 = Device error (FAIL illum.) 0100 = Hardware not installed 0010 = Parameter invalid 0001 = Syntax error
NOTE: The error status is a binary number. It may be com-
bined by the logical operator OR. Example: 1001 = Device error and syntax error.

6.3.10 EUM

Switch the emmision. See Chapter 5.2.10 Emission (EMi), 26.
S: EUM[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Emission for channel 1
0 = Manually 1 = Automatic (default)
b Emission for channel 2
(see above)
c Emission for channel 3
(see above)

6.3.11 FIL

Filter. Measurement filter. See Chapter 5.2.1 Measure­ment filter (FiLt),  23.
S: FIL[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Filter for channel 1
0 = Fast 1 = Medium (default) 2 = Slow

6.3.12 FSR

Full scale range. Full scale range of linear sensors (CDG). See Chapter 5.2.3 Measuring range (FS),  24.
S: FSR[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Full scale range of sensor 1
0 = 0.01 mbar 1 = 0.01 Torr 2= 0.02 Torr 3 = 0.05 Torr
4 = 0.10 mbar 5 = 0.10 Torr 6 = 0.25 mbar 7 = 0.25 Torr 8 = 0.50 mbar 9 = 0.50 Torr
10 = 1 mbar 11 = 1 Torr 12 = 2 mbar 13 = 2 Torr 14 = 5 mbar 15 = 5 Torr
16 = 10 mbar 17 = 10 Torr 18 = 20 mbar 19 = 20 Torr 20 = 50 mbar 21 = 50 Torr
22 = 100 mbar 23 = 100 Torr 24 = 200 mbar 25 = 200 Torr 26 = 500 mbar 27 = 500 Torr
b Filter for channel 2
(see above)
c Filter for channel 3
(see above)
b Full scale range of sensor 2
c Full scale range of sensor 3
tinb07e1-e 2011-07 Vacuum Gauge Controller 37
28 = 1000 mbar 29 = 1100 mbar 30 = 1000 Torr
31 = 2 bar 32 = 5 bar 33 = 10 bar 34 = 50 bar
(see above)
(see above)

6.3.13 FUM

6.3.15 HVC

Select the filament. See Chapter 5.2.11 Filament selec­tion (FiL),  26.
S: FUM[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Filament for channel 1
0 = Automatic (default) 1 = Filament 1 2 = Filament 2
b Filament for channel 2
(see above)
c Filament for channel 3
(see above)

6.3.14 GAS

Gas type correction. See Chapter 5.2.2 Gas type (GAS), 24.
High vacuum circuit on/off. Switch the high vacuum circuit of sensors on/off. See Chapter 4.4.3 Control button func­tions, 17.
S: HVC[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Sensor 1
0 = Off 1 = On
b Sensor 2
(see above)
b Sensor 3
(see above)
NOTE: In order to switch a sensor on/off, the sensor control must be set to «Hand». See Chapter 6.3.26 SC1, 41.
S: GAS[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Gas type for channel 1
0 = Nitrogen/air (default) 1 = Argon 2 = Hydrogen 3 = Other gas
b Gas type for channel 2
(see above)
c Gas type for channel 3
(see above)
NOTE: When «Other gas» is selected, the gas type depen­dence of the measurements will be corrected by a variable correction factor. See Chapter 6.3.5 COR, 36.

6.3.16 ITR

Read a data string from the BPG/BCG/HPG/CDGxxxD sensor.
The measurements of the BPG/BCG/HPG/CDGxxxD are transferred digitally. A data string consists of 8 bytes (hexadecimal numbers) which are separated from each other by a comma. See Reference [11]. The data strings of the sensors are separated from each other by double space characters.
S: ITR<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: aa,aa,aa,aa,aa,aa,aa,aa bb,bb,bb,bb,bb,bb, bb,bb cc,cc,cc,cc,cc,cc,cc<CR><LF>
Parameter Significance
aa,aa,aa,aa,aa,aa,aa,aa Data string of sensor 1
bb,bb,bb,bb,bb,bb,bb,bb Data string of sensor 2
cc,cc,cc,cc,cc,cc,cc,cc Data string of sensor 3

6.3.17 LOC

Parameter setup lock. See Chapter 5.4.5 Parameter setup lock (LoC),  30.
S: LOC[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
Parameter Significance
Parameter Significance
±b.bbbbE±bb Offset value of sensor 2 in current
unit of measurement (see above)
±c.ccccE±cc Offset value of sensor 3 in current
unit of measurement (see above)
a Parameter setup lock
0 = Off (default) 1 = On

6.3.18 OFC

Offset correction. Offset correction function for linear sen­sors (CDG). See Chapter 5.2.4 Offset (oFS),  24.
S: OFC[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Offset correction of channel 1
0 = Off (default) 1 = On 2 = Determine the offset value and activate offset correction function 3 = Adjust the zero point of a CDGxxxD

6.3.20 PNR

Program number. Firmware version number. See Chapter 5.4.2 Firmware version (Pnr),  29.
S: PNR<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
Parameter Significance
a Firmware version
Example: 302-534-D

6.3.21 PR1

Pressure sensor 1. Pressure reading of sensor 1.
S: PR1<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,±b.bbbbE±bb<CR><LF>
b Offset correction of channel 2
(see above)
c Offset correction of channel 3
(see above)

6.3.19 OFD

Offset display. Offset correction values for linear sensors (CDG). See Chapter 5.2.4 Offset (oFS),  24.
S: OFD[,±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc] <CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: ±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc<CR><LF>
Parameter Significance
±a.aaaaE±aa Offset value of sensor 1 in current
unit of measurement
Default: 0.0000E+00
Parameter Significance
a Status of channel 1
0 = Measurement data ok 1 = Underrange 2 = Overrange 3 = Sensor error 4 = Sensor switched off 5 = No sensor 6 = Identification error 7 = BPG/HPG error
±b.bbbbE±bb Pressure reading of sensor 1 in the
current unit of measurement
NOTE: The commands PR2 and PR3 concern the sensors 2 and 3, respectively. The commands are analogous to the PR1 command.
tinb07e1-e 2011-07 Vacuum Gauge Controller 39

6.3.22 PRE

Pirani range extension. See Chapter 5.2.12 Pirani range extension (PrE), 26.
S: PRE[,a,b,c]<CR>[<LF>]
Parameter Significance
±f.ffffE±ff Pressure reading of sensor 3 in the
current unit of measurement
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Range extension for
Sensor 1
0 = Off (default) 1 = On
b Range extension for
sensor 2 (see above)
c Range extension for
sensor 3 (see above)

6.3.23 PRX

Pressure sensors. Pressure readings of all sensors.
S: PRX<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,±b.bbbbE±bb,c,±d.ddddE±dd,e,±f.ffffEff<CR><LF>
Parameter Significance
a Status of channel 1
0 = Measurement data ok 1 = Underrange 2 = Overrange 3 = Sensor error 4 = Sensor switched off 5 = No sensor 6 = Identification error 7 = BPG/BCG/HPG error
±b.bbbbE±bb Pressure reading of sensor 1 in the
current unit of measurement
c Status of channel 2
(see above)
±d.ddddE±dd Pressure reading of sensor 2 in the
current unit of measurement

6.3.24 RES

Reset. Reset the serial interface.
Deletes the input buffer. All queued error messages are sent to the Host. The unit returns to the measurement mode.
S: RES[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: b,c,d,...<CR><LF>
Parameter Significance
a 1 = Perform reset
b,c,d... Queued error messages
0 = No error 1 = Watchdog control has been trig­gered 2 = Task(s) not executed 3 = EPROM error 4 = RAM error 5 = EEPROM error 6 = Display error 7 = A/D converter error 8 = UART error 9 = Sensor 1 general error 10 = Sensor 1 ID error 11 = Sensor 2 general error 12 = Sensor 2 ID error 13 = Sensor 3 general error 14 = Sensor 3 ID error

6.3.25 SAV

Save parameters to EEPROM.
The command SAV,0 resets all parameters to their default values (factory settings). See Chapter 5.3.4 Default parameters (dEF),  27.
The command SAV,1 saves parameter values which have been changed via the serial interface in the EEPROM. These values will be preserved even when the unit is switched off.
S: SAV,a<CR>[<LF>]
e Status of channel 3
(see above)
R: <ACK><CR><LF>
Parameter Significance
a Save parameters
0 = Save default parameters 1 = Save user parameters

6.3.27 SP1

Setpoint 1. Switching function 1. See Chapter 5.1 Switch­ing function parameters (PArA SP),  21.
S: SP1[,a,b.bbbbE±bb,c.ccccE±cc]<CR>[<LF>]
R: <ACK><CR><LF>
NOTE: Parameters which have been changed manually (control buttons) are saved in the EEPROM automat­ically. The SAV command is not required in this case.

6.3.26 SC1

Sensor 1 control. See Chapter 5.2.6 Sensor activation (S-on), 25 and Chapter 5.2.8 Sensor deactivation (S­oFF), 25.
S: SC1[,a,b,c.ccE±cc,d.ddE±dd]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c.ccE±cc,d.ddE±dd<CR><LF>
Parameter Significance
a Sensor activation
0 = Manual (default) 1 = Hot start 2 = By channel 1 3 = By channel 2 4 = By channel 3
b Sensor deactivation
0 = Manual (default) 1 = Self control 2 = By channel 1 3 = By channel 2 4 = By channel 3
S: <ENQ>
R: a,b.bbbbE±bb,c.ccccE±cc<CR><LF>
Parameter Significance
a Switching function assignment
0 = Channel 1 1 = Channel 2 2 = Channel 3
b.bbbbE±bb Lower threshold value in the current
unit of measurement
c.ccccE±cc Upper threshold value in the current
unit of measurement
NOTE: The commands SP2…SP6 concern the switching functions 2…6, respectively. The commands are analogous to the SP1 command.

6.3.28 SPS

Setpoint status. Switching function status. See Chapter 5.1 Switching function parameters (PArA SP), 21.
S: SPS<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c,d,e,f<CR><LF>
c.ccE±cc Activation value in the current unit of
measurement
d.ddE±dd Deactivation value in the current
unit of measurement
NOTE: The commands SC2 and SC3 concern the sensors 2 and 3, respectively. The commands are analogous to the SC1 command.
tinb07e1-e 2011-07 Vacuum Gauge Controller 41
Parameter Significance
a Status of switching function 1
0 = Off 1 = On
b Status of switching function 2
(see above)
c Status of switching function 3
(see above)
d Status of switching function 4
(see above)
e Status of switching function 5
(see above)
Parameter Significance
f Status of switching function 6
(see above)

6.3.29 TAD

Parameter Significance
a Test status
0 = Off 1 = On
Test A/D converter. Test the A/D converter. See Chapter 5.4.10 A/D converter signal (Ad-S), 30 and Chapter 5.4.11 A/D converter ID (Ad-i),  31.
S: TAD<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: ±a.aaaa,±b.bbbb,±c.cccc,±d.dddd,±e.eeee,±f.ffff <CR><LF>
Parameter Significance
±a.aaaa ADC channel 1 Reading of sensor 1
in volts.
0.0000… +11.0000
±b.bbbb ADC channel 2. Reading of sensor 2
in volts.
0.0000… +11.0000
±c.cccc ADC channel 3. Reading of sensor 3
in volts.
0.0000… +11.0000
±d.dddd ADC channel 4. Identification of
sensor 1 in volts.
0.0000… +5.0000

6.3.31 TEE

Test EEPROM. Test the EEPROM. See Chapter 5.4.8 EEPROM test (EE-t),  30.
CAUTION
EEPROM life.
A large number of write operations will reduce the EEPROM life.
Do not repeat the EEPROM test more often than necessary (e.g. in program loops).
S: TEE<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: aaaa<CR><LF>
The control character <ENQ> starts the test. It takes approximately one second to complete the test.
Parameter Significance
aaaa Error status. See Chapter 6.3.9
ERR, 36.
±e.eeee ADC channel 5. Identification of
sensor 2 in volts.
0.0000… +5.0000
±f.ffff ADC channel 6. Identification of
sensor 3 in volts.
0.0000… +5.0000

6.3.30 TDI

Test display. Test the display. See Chapter 5.4.9 Display test (di-t),  30.
S: TDI[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>

6.3.32 TEP

Test EPROM. Test the EPROM. See Chapter 5.4.7 EPROM test (EP-t),  30.
S: TEP<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: aaaa,bbbb<CR><LF>
The control character <ENQ> starts the test. It takes approximately 5 seconds to complete the test.
Parameter Significance
aaaa Error status. See Chapter 6.3.9
ERR, 36.
bbbb Check sum (hexadecimal)

6.3.33 TID

Transmitter identification. Sensor identification. See Chapter 4.4.3.7 Identifying a sensor,  18.
S: TID<CR>[<LF>]
R: <ACK><CR><LF>
Parameter Significance
a Test status
0 = Off 1 = On
S: <ENQ>
R: a,b,c<CR><LF>
Parameter Significance
a Identification of sensor 1
PSG PCG PEG MPG CDG BPG BPG402 BCG HPG noSen noid
b Identification of sensor 2
(see above)
c Identification of sensor 3
(see above)
bb Relay status
00 = All relays off 01 = Switching function 1 relay on 02 = Switching function 2 relay on 04 = Switching function 3 relay on 08 = Switching function 4 relay on 10 = Switching function 5 relay on 20 = Switching function 6 relay on 40 = Error signal relay on 7F = All relays on
NOTE: The relay status is a hexadecimal number. It may be combined by the logical operator OR. Example: 24 = Switching functions relays 3 and 6 on.

6.3.35 TKB

Test keyboard. Test the keyboard.
S: TKB<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>

6.3.34 TIO

Test I/O. Test the relays. This command allows to switch a single relay or several relays at a time.
CAUTION
Relay test.
In this test the relays switch irrespective of the actual pressure. This may cause unintended switching of devices.
Unplug the RELAY connection before per­forming a relay test.
S: TIO[,a,bb]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,bb<CR><LF>
R: aaaa<CR><LF>
The control character <ENQ> starts the test. The Vac­uum Gauge Controller polls the keyboard and sends a message to the computer.
Parameter Significance
aaaa Keyboard status
0000 = No button pressed 1000 = CH pressed 0100 = PARA pressed 0010 = DOWN pressed 0001 = UP pressed
NOTE: The keyboard status is a binary number. It may be combined by the logical operator OR. Example: 0011 = DOWN and UP pressed at the same time.
tinb07e1-e 2011-07 Vacuum Gauge Controller 43

6.3.36 TLC

Torr lock. See Chapter 5.4.4 Torr lock (tr-L),  30.
S: TLC[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
Parameter Significance
a Unit of measurement
0 = mbar/bar (default) 1 = Torr 2 = Pascal 3 = Micron
Parameter Significance
aTorr lock
0 = Off (default) 1 = On

6.3.37 TRA

Test RAM. Test the RAM. See Chapter 5.4.6 RAM test (rA-t), 30.
S: TRA<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: aaaa<CR><LF>
The control character <ENQ> starts the test. It takes approximately one second to complete the test.
Parameter Significance
aaaa Error status. See Chapter 6.3.9
ERR, 36.

6.3.40 WDT

Watchdog control. See Chapter 5.4.3 Watchdog control (dt-C), 29.
S: WDT[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
Parameter Significance
a Error acknowledgement
0 = Manually 1 = Automatic (default)

6.3.38 TRS

Test RS232C interface. Test the RS232C interface. See Chapter 5.4.13 RS232C test (rS-t),  31.
S: TRS<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
The control character <ENQ> starts the test. The test can be stopped by pressing Ctrl-C.

6.3.39 UNI

Unit of measurement. See Chapter 5.3.1 Unit of mea­surement (unit),  27.
S: UNI[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>

7 Maintenance and service

7.1 Maintenance

The Vacuum Gauge Controller does not require any spe­cial maintenance work.

7.1.1 Cleaning

For cleaning the outside of the unit, a slightly moistened cloth will usually do. Do not use any aggressive or scour­ing cleaning agents.
DANGER
Mains power.
The Vacuum Gauge Controller contains parts which are connected to the mains supply.
Make sure that no objects enter through the louvers of the unit. Keep the unit dry. Do not open the unit.

7.2 Program transfer mode

If your Vacuum Gauge Controller requires an updated firmware version, e.g. for using a new sensor type, please contact your local INFICON service center. You may also visit our website www.inficon.com updates are available for download.
The user parameters set by you are generally still avail­able after the firmware update. It is however recom­mended that you make a note of the settings before updating. See Section «Default parameters»,  51.

7.2.1 Preparations and selection

where firmware

7.2.2 Program transfer

The firmware for the Vacuum Gauge Controller is deliv­ered as a self-extracting *.exe file or as a packed *.zip file.
1 Copy the *.exe or the *.zip file into an empty direc-
tory
2 Unpack the file. One of the extracted files is a
batch file *.bat.
3 By default, the program transfer is assumed to run
via the COM1 serial interface. Proceed as follows if you want to use another serial interface:
3.1 Click the batch file with the right mouse but­ton
A menu appears
3.2 From the menu, select the option «Edit»
The batch file is loaded into a text editor
3.3 Change the COM1 entry to the interface you want to use (e.g. COM2)
3.4 Save and close the modified batch file
4 Execute the batch file by double-clicking it with the
mouse
The new firmware is being transferred to the Vacuum Gauge Controller. You can monitor the individual steps on the PC screen. After approximately 1 minute the trans­fer is completed.

7.2.3 Restarting

Proceed as follows after the firmware has been trans­ferred completely:
1 Switch the Vacuum Gauge Controller off
2 Wait at least 10 seconds to make sure that the
Vacuum Gauge Controller can initialize
1 Switch the Vacuum Gauge Controller off
2 Connect the RS232C socket (Fig. 3-4, 11, 2 10,
Pos. F) with a serial interface of the PC (e.g. COM1). See Chapter 3.3.7 RS232C,  14.
3 Press the button behind the opening (Fig. 3-4,
11, Pos. E) with a pencil and switch the Vacuum Gauge Controller on
NOTE: The display remains dark. The Vacuum Gauge Con­troller is set to the program transfer mode.
tinb07e1-e 2011-07 Vacuum Gauge Controller 45
3 Switch the Vacuum Gauge Controller on again
4 Check if the current parameter settings still agree
with the previous ones. See Section «Default parameters», 51.
The Vacuum Gauge Controller is ready for operation again.

7.3 Calibration

7.3.1 Basics

The Vacuum Gauge Controller can only measure with high accuracy when it is calibrated precisely. The Vac­uum Gauge Controller is calibrated by INFICON before it is shipped. Normally there is no need to change the cali­bration data.
CAUTION
Calibration.
If you input incorrect calibration data, the Vacuum Gauge Controller cannot perform accurate measurements anymore.
NOTE: The CAO command can also be used with other pa­rameters. See Chapter 7.3.4 Calibrating the unit, 46.

7.3.3 CAF

Calibration factor. Calibration factor of the A/D converter. The command is intended for service technicians of INF­ICON only.
S: CAF[,a.aaaaE±aa,b.bbbbE±bb,c.ccccE±cc] <CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a.aaaaE±aa,b.bbbbE±bb,c.ccccE±cc<CR><LF>
The interface commands for calibrating the unit are intended for service technicians of INFICON only.
The calibration affects the A/D converters of the individ­ual channels. The measuring curve of an ideal A/D con­verter is a straight line which has a slope of one and runs through the origin, i.e.:
Factor = 1 (slope of the line)
Offset = 0 (intersection with the y axis)
The curves of real A/D converters differ slightly from these ideal values. Calibrating the unit means to deter­mine the gain factors and the offset voltages of the indi­vidual A/D converters and to store these calibration val­ues.
The interface commands CAF and CAO are used to access the calibration data of the unit.

7.3.2 CAO

Calibration offset. Calibration offset of the A/D converter. The command is intended for service technicians of INF­ICON only.
S: CAO[,±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc] <CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: ±a.aaaaE±aa,±b.bbbbE±bb,±c.ccccE±cc<CR><LF>
Parameter Significance
±a.aaaaE±aa Calibration offset of channel 1
±b.bbbbE±bb Calibration offset of channel 2
±c.ccccE±cc Calibration offset of channel 3
Parameter Significance
a.aaaaE±aa Calibration factor of channel 1
b.bbbbE±bb Calibration factor of channel 2
c.ccccE±cc Calibration factor of channel 3
NOTE: The CAF command can also be used with other pa­rameters. See Chapter 7.3.4 Calibrating the unit, 46.

7.3.4 Calibrating the unit

The Vacuum Gauge Controller can be calibrated auto­matically with the interface commands CAF and CAO. The following auxiliary tools are required:
D-Sub plug, 15-pin
Soldering equipment
High-precision voltage source for 10 volts (10.000 V)
Calibration offset
1 Unplug the sensor of the respective channel
2 Connect pin 2 (Signal) and Pin 12 (Signal-GND) of
a 15-pin D-Sub plug with a wire strap. See Chapter 3.3.4 SENSOR,  12.
3 Put the D-Sub plug with the wire strap onto the
sensor connection of the respective channel
The input of the A/D converter is short-circuited
4 Use the serial interface to send the command
CAO,a to the unit. For the parameter a, use:
0 = Channel 1
1 = Channel 2
2 = Channel 3
The calibration offset of the respective channel is deter­mined and stored in the EEPROM.
Calibration factor
5 Remove the D-Sub plug with the wire strap from
the sensor connection of the respective channel
6 Connect pin 2 (Signal) and pin 12 (Signal-GND) of
a 15-pin D-Sub plug with a high-precision voltage source. See Chapter 3.3.4 SENSOR,  12.
Pin 2 = +10.000 V
•Pin 12 = GND
7 Put the D-Sub plug with the voltage source con-
nections onto the sensor connection of the respec­tive channel
8 Use the serial interface to send the command
CAF,a to the unit. For the parameter a, use:
0 = Channel 1
1 = Channel 2
2 = Channel 3
The calibration factor of the respective channel is deter­mined and stored in the EEPROM. Now the channel is calibrated.
tinb07e1-e 2011-07 Vacuum Gauge Controller 47

8 Troubleshooting

8.1 Fault indication

A fault in the Vacuum Gauge Controller is indicated as follows:
FAIL flashes and the display shows an error message.
See Chapter 8.2 Error messages,  48.
The error signal relay opens

8.2 Error messages

Display Possible cause and corrective action
SE Sensor error. Error in the connection of
the respective sensor.
Press PARA to acknowledge. If the cause has not been removed, then
noSEn or noid will be displayed.
dt The watchdog control has been trig-
gered. Severe electrical fault or an oper­ating system error. Or: The Vacuum Gauge Controller has been switched off and on without suffi­cient delay.
Press PARA to acknowledge. The Vac­uum Gauge Controller will acknowledge automatically after 2 s if the watchdog control is set to auto.
Display Possible cause and corrective action
Er x Error message of BPG400 / HPG.
0 = No communication with the sensor x = Error code (High-Byte).
See Reference [11], Reference [12].
Er xx Error message of BCG, BPG402.
xxH = Error code.
See Reference [13], Reference [15].

8.3 Technical support

If the fault persists even after the message has been acknowledged several times and/or the sensor has been exchanged, please contact your local INFICON service center.
rA Error in the main memory (RAM).
Press PARA to acknowledge.
EP Error in the program memory (EPROM).
Press PARA to acknowledge.
EE Error in parameter memory (EEPROM).
Press PARA to acknowledge.
di Error in the display driver.
Press PARA to acknowledge.
Ad Error in the A/D converter.
Press PARA to acknowledge.
tF Task fail. Error in the operating system.
Press PARA to acknowledge.
UA Error in UART.
Press PARA to acknowledge.

9 Storage and disposal

9.1 Packaging

Please keep the original packaging. The packaging is required for storing the Vacuum Gauge Controller and for shipping it to an INFICON service center.

9.2 Storage

The Vacuum Gauge Controller may only be stored in a dry room. The following requirements must be met:
Ambient temperature -20…+60 °C
Humidity As low as possible. Prefera-
bly in an air-tight plastic bag with a desiccant.

9.3 Disposal

The product must be disposed of in accordance with the relevant local regulations for the environmentally safe disposal of systems and electronic components.
tinb07e1-e 2011-07 Vacuum Gauge Controller 49

Appendix

Conversion tables

Weights

1 È = ...Æ kg lb oz slug
kg 1 2.205 35.27 6.852 × 10
lb 0.454 1 16 3.108 × 10
oz 28.35 × 10
slug 14.59 32.17 514.8 1

Pressure

1 È = ...Æ Pa (= N/m2) mbar Torr (= mm Hg) psi (= lb/in2)
-3
62.5 × 10-3 1 1.943 × 10-3
-2
-2
-2
Pa 1 10
mbar 10
Torr 1.333 × 10
psi 6.895 × 10
2
1 0.75 1.45 × 10-2
2
1.333 1 1.934 × 10-2
3
68.95 51.71 1
0.987 × 10-5 1.45 × 10-4

Linear measures

1 È = ...Æ mm m inch ft
-3
mm 1 10
m 1000 1 39.37 3.281
inch 25.4 2.54 × 10
ft 304.8 0.3048 12 1
3.937 × 10-2 3.281 × 10-3
-2
1 8.333 × 10-2

Temperature

È = ...Æ Kelvin Celsius Fahrenheit
Kelvin 1 °C + 273.15 (°F + 459.67) × 5/9
Celsius K - 273.15 1 (°F × 5/9) - 17.78
Fahrenheit (K × 9/5) - 459.67 (°C + 17.78) × 9/5 1

Default parameters

Display Default User
-11
-11
mbar
mbar
SP -L 1 × 10
SP -H 9 × 10
FiLt nor
GAS n2
Cor 1.00
FS 1000 mbar
oFS oFF
0.0000E+00 mbar
[3] Operating Manual
Compact Pirani Gauge PSG500/-S, PSG502-S, PSG510-S, PSG512-S tina44e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[4] Operating Manual
Pirani Standard Gauge PSG550, PSG552, PSG554 tina60e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[5] Operating Manual
Pirani Capacitance Gauge PCG400, PCG400-S tina28e1 INFICON AG, LI–9496 Balzers, Liechtenstein
S-on HAnd
1.00E-03 mbar
S-oFF HAnd
1.00E-03 mbar
PrE oFF
unit bAr
bAud 9600
diGit 2
Ao LoG
Err-r ALL
dt-C Auto
tr-L oFF
LoC oFF
EMi Auto
FiL Auto
[6] Operating Manual
Pirani Capacitance Diaphragm Gauge PCG550, PCG552, PCG554 tina56e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[7] Operating Manual
Penning Gauge PEG100 tina14e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[8] Operating Manual
Capacitance Diaphragm Gauge CDG025 tina01e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[9] Operating Manual
Capacitance Diaphragm Gauge CDG045, CDG045-H tina07e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[10] Operating Manual
Capacitance Diaphragm Gauge CDG100 tina08e1 INFICON AG, LI–9496 Balzers, Liechtenstein

Literature

The operating manuals listed below can be downloaded from the website www.inficon.com
[1] Operating Manual
Pirani Standard Gauge PSG400, PSG400-S tina04e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[2] Operating Manual
Pirani Standard Gauge PSG100-S, PSG101-S tina17e1 INFICON AG, LI–9496 Balzers, Liechtenstein
tinb07e1-e 2011-07 Vacuum Gauge Controller 51
in the PDF file format.
[11] Operating Manual
Bayard-Alpert Pirani Gauge BPG400 tina03e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[12] Operating Manual
High Pressure / Pirani Gauge HPG400 tina31e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[13] Operating Manual
Triple Gauge BCG450 tina40e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[14] Operating Manual
Inverted Magnetron Pirani Gauge MPG400, MPG401 tina48e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[15] Operating Manual
Bayard-Alpert Pirani Gauge BPG402 tina46e1 INFICON AG, LI–9496 Balzers, Liechtenstein

Index

A
A/D converter calibration . . . . . . . . . . . . . . . . . . . . . . . 46
A/D converter test . . . . . . . . . . . . . . . . . . . . . . . . . 30
B
Back side. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Baud rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
, 42
, 35
[16] Operating Manual
Capacitance Diaphragm Gauge CDG025D tina49e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[17] Operating Manual
Capacitance Diaphragm Gauge CDG045D tina51e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[18] Operating Manual
Capacitance Diaphragm Gauge CDG100D tina52e1 INFICON AG, LI–9496 Balzers, Liechtenstein
[19] Operating Manual
Capacitance Diaphragm Gauge CDG160D, CDG200D tina53e1 INFICON AG, LI–9496 Balzers, Liechtenstein
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Calibration factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Calibration offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Characteristic curve of recorder output . . . . . . . . . . . . 27
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Continuous transmission of measurements . . .33
Number formats . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Receiving . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Sending . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Computer interface . . . . . . . . . . . . . . . . . . . . . . . . 14
Resetting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Connections
CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Mains connection . . . . . . . . . . . . . . . . . . . . . . . . . . 12
RELAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
RS232C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
SENSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Continuous transmission of measurements . . . . .33
Control buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
CONTROL connection . . . . . . . . . . . . . . . . . . . . . . . . 14
Control panel mounted . . . . . . . . . . . . . . . . . . . . . . . . 10
Conversion tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Correction factor . . . . . . . . . . . . . . . . . . . . . . . . . . 24
, 35
, 32
, 44
, 35
, 36
D
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . 55
Default parameters
Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Degas function . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Switching off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Degassing
see degas function
Disconnecting device . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Display
Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
, 40
, 36
, 36 , 42
E
EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . 55
EEPROM test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
EPROM test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Error messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Error signal relay . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Error status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
ETL Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
, 42 , 42
, 36
F
Factory settings. . . . . . . . . . . . . . . . . . . . . . . .27, 40, 51
Fault indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Filament selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Firmware
Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
, 37
, 29, 39
G
General parameters . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Baud rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Default parameters . . . . . . . . . . . . . . . . . . . . . . . . 27
Display format . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Error signal relay . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Recorder output . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Unit of measurement . . . . . . . . . . . . . . . . . . . . . . . 27
Ground connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
H
Hysteresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Switching the degas function on . . . . . . . . . . . . . . 18
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Mnemonics
Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Mounting the unit in a rack . . . . . . . . . . . . . . . . . . . . . 11
, 37
O
Offset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24, 39
Operating modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . 17
Parameter mode . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Program transfer mode . . . . . . . . . . . . . . . . . . . . . 45
P
Parameter groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Parameter mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Basic operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Parameter setup lock . . . . . . . . . . . . . . . . . . . . . . 30
Parameters
General parameters . . . . . . . . . . . . . . . . . . . . . . . . 27
Sensor Parameters . . . . . . . . . . . . . . . . . . . . . . . . 23
Switching function parameters . . . . . . . . . . . . . . . . 21
Test parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Part number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Pin assignment
CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
RELAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
RS232C . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
SENSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Pirani range extension . . . . . . . . . . . . . . . . . . . . . 26
Product versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
, 38
, 40
I
I/O test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31, 43
Identifying sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
L
Literature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Lower threshold value . . . . . . . . . . . . . . . . . . . . . . . . . 21
Adjustment range. . . . . . . . . . . . . . . . . . . . . . . . . . 22
M
Mains connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Measurement filter . . . . . . . . . . . . . . . . . . . . . . . . 23
Measurement mode . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Identifying a sensor . . . . . . . . . . . . . . . . . . . . . . . . 18
Selecting a channel . . . . . . . . . . . . . . . . . . . . . . . . 17
Status report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Switching a sensor off . . . . . . . . . . . . . . . . . . . . . . 18
Switching a sensor on . . . . . . . . . . . . . . . . . . . . . . 17
Switching the degas function off . . . . . . . . . . . . . . 18
, 37
R
RAM test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30, 44
Range extension . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Recorder output . . . . . . . . . . . . . . . . . . . . . . . . . . 27
RELAY connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Residual dangers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
RS232C connection . . . . . . . . . . . . . . . . . . . . . . . . . . 14
RS232C test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
, 40 , 35
, 44
S
Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Sensor
Degassing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Identifying. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Switching off. . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Switching on. . . . . . . . . . . . . . . . . . . . . . . . . . . 17
SENSOR connection. . . . . . . . . . . . . . . . . . . . . . . . . . 12
Sensor control . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Sensor identification . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Sensor parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Degas function . . . . . . . . . . . . . . . . . . . . . . . . . 25
, 38 , 38
, 41
, 36
tinb07e1-e 2011-07 Vacuum Gauge Controller 53
Filament selection . . . . . . . . . . . . . . . . . . . . . . . . . 26
Gas type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Measurement filter . . . . . . . . . . . . . . . . . . . . . . 23
Measuring range . . . . . . . . . . . . . . . . . . . . . . . 24
Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Pirani range extension . . . . . . . . . . . . . . . . . . . 26
Sensor activation . . . . . . . . . . . . . . . . . . . . . . . 25
Sensor deactivation . . . . . . . . . . . . . . . . . . . . . 25
Switching the emission. . . . . . . . . . . . . . . . . . . 37
Switching the emission on . . . . . . . . . . . . . . . . . . . 26
Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Switching function parameters . . . . . . . . . . . . . . . . . . 21
Switching functions . . . . . . . . . . . . . . . . . . . . . . . . 21
Configuring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Switching off
Degas function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Switching on
Degas function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Switching the emission on. . . . . . . . . . . . . . . . . . . . . . 26
, 38 , 37 , 37 , 39 , 40 , 41 , 41 , 38
, 41
, 38
, 38
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Watchdog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
, 44
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Test parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
A/D converter ID. . . . . . . . . . . . . . . . . . . . . . . . 31
A/D converter signal. . . . . . . . . . . . . . . . . . . . . 30
Display test. . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
EEPROM test. . . . . . . . . . . . . . . . . . . . . . . . . . 30
EPROM test . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Firmware version . . . . . . . . . . . . . . . . . . . . . . . 29
I/O test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Keyboard test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Parameter setup lock . . . . . . . . . . . . . . . . . . . . 30
RAM test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
RS232C test. . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Torr lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Watchdog control . . . . . . . . . . . . . . . . . . . . . . . 29
Threshold values . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Adjustment range . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Torr lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Type label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
, 42 , 42 , 42 , 42 , 42 , 39 , 43
, 38 , 44 , 44
, 44 , 44 , 41
, 44
U
Unit of measurement . . . . . . . . . . . . . . . . . . . . . . . 27, 44
Update. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Upper threshold value . . . . . . . . . . . . . . . . . . . . . . . . . 21
Adjustment range . . . . . . . . . . . . . . . . . . . . . . . . . . 22
W
Waiting time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

ETL Certification

3103457
ETL LISTED
The products VGC402 and VGC403 comply with the requirements of the following standards:
UL 61010-1, Issued: 2004/07/12 Ed: 2 Rev: 2005/07/22
CAN/CSA C22.2#61010-1, Issued: 2004/07/12

EC Declaration of Conformity

We, INFICON, hereby declare that the equipment men­tioned below complies with the provisions of the Directive relating to electrical equipment designed for use within certain voltage limits 2006/95/EC and the Directive relat­ing to electromagnetic compatibility 2004/108/EC.
Products
VGC402, VGC403 Vacuum Gauge Controller
Part numbers
398-020 398-021
Standards
Harmonized and international/national standards and specifications:
EN 61010-1:2001 (Safety requirements for electrical equipment for measurement and control use)
EN 61000-3-2:2006 (EMC, limits for harmonic current emissions)
EN 61000-3-3:1995 + A1:2001 + A2:2005 (EMC, limita­tion of voltage changes, voltage fluctuations and flicker)
EN 61000-6-2:2005 (EMC, generic immunity standard)
EN 61000-6-3:2007 (EMC, generic emission standard)
Signatures
INFICON AG, Balzers
November 19, 2008 November 19, 2008
Markus Truniger Product Manager
Dr. Urs Wälchli Managing Director
tinb07e1-e 2011-07 Vacuum Gauge Controller 55
+423 / 388 3111 +423 / 388 3700
LI-9496 Balzers Liechtenstein Tel Fax reachus@inficon.com
www.inficon.com
t i nb07e1- e
Original: tinb07d1-e (2011-07)
Loading...