HP (Hewlett-Packard) E4402B, E4403B, E4411B, E4408B, E4407B User Manual

...
5 (1)

Getting Started Guide

Agilent Technologies ESA Spectrum Analyzers

This manual provides documentation for the following instruments:

ESA-E Series

E4401B (9 kHz - 1.5 GHz)

E4402B (9 kHz - 3.0 GHz)

E4404B (9 kHz - 6.7 GHz)

E4405B (9 kHz - 13.2 GHz)

E4407B (9 kHz - 26.5 GHz)

and ESA-L Series

E4411B (9 kHz - 1.5 GHz) E4403B (9 kHz - 3.0 GHz) E4408B (9 kHz - 26.5 GHz)

Manufacturing Part Number: E4401-90464

Supersedes: E4401-90437

Printed in USA

November 2003

© Copyright 2001-2003 Agilent Technologies, Inc.

WARNING

CAUTION

NOTE

Notice

The information contained in this document is subject to change without notice.

Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

The following safety symbols are used throughout this manual. Familiarize yourself with the symbols and their meaning before operating this analyzer.

Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.

Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the analyzer. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.

Note calls out special information for the user’s attention. It provides operational information or additional instructions of which the user should be aware.

Additional Information

For the latest information about this analyzer, including firmware upgrades, application information, and product information, see the following URL:

http://www.agilent.com/find/esa/

2

Contents

1. Installation and Setup

Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Checking the Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 AC Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Turning on the Analyzer for the First Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Why Aren’t All the Personality Options Loaded in Memory? . . . . . . . . . . . . . . . . . . 16 Using an External Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Firmware Revision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Running Internal Alignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Printer Setup and Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Protecting Against Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Legal Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

2. Front and Rear Panel Features

Front Panel Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Front-Panel Connectors and Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Display Annotations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Rear-Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

Key Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

Front and Rear Panel Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

3. Making a Basic Measurement

Using the Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Entering Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Using Menu Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Presetting the Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Creating a User Preset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Viewing a Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

4. Viewing Catalogs and Saving Files

 

File Menu Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

53

Locating and viewing files in the catalog . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

53

Creating a directory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

55

Formatting a Floppy Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

56

Saving a File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

57

Step 1. Set up the analyzer trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

57

Step 2. Save the file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

59

Loading a file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

61

Renaming a File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

63

Copying a File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

64

Deleting a File . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

65

Using the Alpha Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

66

3

Contents

5. Options and Accessories

Ordering Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69

Option Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

50 Ohm/75 Ohm Minimum Loss Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

75 Ohm Matching Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

AC Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

AC Probe (Low Frequency) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

Broadband Preamplifiers and Power Amplifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83

Carrying Strap (Part Number E4401-60028) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84

External Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84

GPIB Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84

HP/Agilent 11970 Series Harmonic Mixers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84

HP/Agilent 11974 Series Preselected Millimeter Mixers . . . . . . . . . . . . . . . . . . . . . . 84

Agilent E1779A Battery Pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

Parallel Interface Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

RF and Transient Limiters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

RF Bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

RS-232 Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86

Static Safe Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86

6. In Case of Difficulty

 

Types of Spectrum Analyzer Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

89

Before Calling Agilent Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

90

Check the Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

90

Read the Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

92

Service Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

92

Calling Agilent Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

92

Returning an Analyzer for Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

94

4

1 Installation and Setup

5

Installation and Setup

This chapter provides the following information that you may need when you first receive your spectrum analyzer:

“Initial Inspection” on page 7

“Power Requirements” on page 10

“Turning on the Analyzer for the First Time” on page 15

“Printer Setup and Operation” on page 19

“Protecting Against Electrostatic Discharge” on page 22

“Running Internal Alignments” on page 18

“Legal Information” on page 23

6

Chapter 1

Installation and Setup

Initial Inspection

Initial Inspection

Inspect the shipping container and the cushioning material for signs of stress. Retain the shipping materials for future use, as you may wish to ship the analyzer to another location or to Agilent Technologies for service. Verify that the contents of the shipping container are complete. The following table lists the items shipped with the analyzer.

Item

Description

 

 

Accessories

 

 

 

Adapter, Type-N (m) to BNC (f)

Not shipped with Option 1DP. Two adapters are shipped with

 

Option 1DN.

 

 

Adapter, BNC (m) to F (f), 75

Shipped only with Option 1DP. Two adapters shipped with

 

Option 1DQ.

 

 

Adapter, Type-N (m) to SMA (f)

Shipped only with Option 1DN for Agilent E4402B, E4403B,

 

E4404B, E4405B, E4407B, and E4408B. Not shipped with

 

Option BAB.

 

 

Adapter, APC 3.5 (f) to APC 3.5 (f)

Shipped only with Option BAB.

 

 

Adapter, BNC (f) to SMA (m)

Shipped only with Option BAB.

 

 

Cable, BNC (m) to BNC (m), 203 mm

Shipped only with Agilent E4402B, E4403B, E4404B,

 

E4405B, E4407B, and E4408B.

 

 

Cable, SMA (m) to Type-N (m), 220 mm

Shipped only with Option 1DN for Agilent E4402B, E4403B,

 

E4404B, E4405B, E4407B, and E4408B.

 

 

IntuiLink Toolbar software, CD-ROM

Provides a set of connectivity tools that enable you to easily

 

move data from your analyzer to your PC.

 

 

Power Cable (See Table 1-3. on page 13)

Connection for power source.

 

 

Standard Documentation Set

 

 

 

Getting Started Guide

Covers unpacking and setting up the analyzer, analyzer

 

features, and how to make a basic measurement. Includes

 

information on options and accessories, and what to do if you

 

have a problem.

 

 

User’s and Programmer’s Reference

Describes analyzer features in detail, including front-panel

 

key descriptions, basic spectrum analyzer programming

 

information, and SCPI command descriptions.

 

 

Measurement Guide

Provides details on how to measure various signals, and how

 

to use catalogs and files.

 

 

Specifications Guide

Documents specifications, safety, and regulatory information.

 

 

Instrument Messages and Functional Tests

Includes instrument messages (and suggestions for

 

troubleshooting them), and manual functional tests.

 

 

Programming Conversion Guide

Describes SCPI programming command compatibility for

 

8590, ESA series analyzers.

 

 

Chapter 1

7

Installation and Setup

Initial Inspection

 

Item

Description

 

 

 

 

 

Documentation CD-ROM

Includes the documents in the standard set (listed above).

 

 

You can view and print the information as needed. See the

 

 

CD-ROM jacket for installation information.

 

 

 

 

 

 

 

 

NOTE

If you purchased one or more optional measurement personalities, the related guides for the

 

options you ordered are included.

 

Service documentation is not included in the standard documentation set. See “Options” on

 

page 69 for information on ordering.

 

 

8

Chapter 1

Installation and Setup

Initial Inspection

If There Is a Problem

If the shipping materials are damaged or the contents of the container are incomplete:

Contact the nearest Agilent Technologies office to arrange for repair or replacement (Table 6-2. on page 93). You will not need to wait for a claim settlement.

Keep the shipping materials for the carrier’s inspection.

If you must return an analyzer to Agilent Technologies, use the original (or comparable) shipping materials (see “Returning an Analyzer for Service” on page 94).

Chapter 1

9

Installation and Setup

Power Requirements

Power Requirements

The only physical installation of your Agilent spectrum analyzer is a connection to a power source.

Line voltage does not need to be selected.

WARNING Failure to ground the analyzer properly can result in personal injury. Before turning on the analyzer, you must connect its protective earth terminals to the protective conductor of the main power cable. Insert the main power cable plug into a socket outlet that has a protective earth contact only. DO NOT defeat the earth-grounding protection by using an extension cable, power cable, or autotransformer without a protective ground conductor.

If you are using an autotransformer, make sure its common terminal is connected to the protective earth contact of the power source outlet socket.

This is a Safety Class 1 Product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.

CAUTION VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4° C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.

This analyzer has autoranging line voltage input. Be sure the supply voltage is within the specified range.

NOTE

For more information regarding analyzer specifications, see the Specifications

 

guide.

 

 

10

Chapter 1

Installation and Setup

Power Requirements

Table 1-1. AC Power Requirements

Description

Specifications

 

 

Voltage

90 to 132 Vrms (47 to 440 Hz)

 

 

Voltage

195 to 250 Vrms (47 to 66 Hz)

 

 

Power Consumption, On

< 300 W

 

 

Power Consumption, Standby

< 5 W

 

 

Table 1-2. DC Power Requirements

Description

Specifications

 

 

Voltage

12 to 20 Vdc

 

 

Power Consumption

< 200 W

 

 

Power Consumption, Standby

< 5 W

 

 

Checking the Fuse

Where IEC regulations apply, use a 5 by 20 mm, rated F5A, 250 V IEC approved fuse. This fuse may be used with input line voltages of 115 V or 230 V. Its part number is 2110-0709.

Where UL/CSA regulations apply, use a 5 by 20 mm rated fast blow, 5 A, 125 V UL/CSA approved fuse (part number 2110-0756). This fuse may only be used with an input line voltage of 115 V.

The line fuse is housed in a fuse holder in the upper left hand corner of the rear panel.

To remove the fuse, first disconnect the power cord from the analyzer. Then insert the tip of a screwdriver into the slot at the middle of the fuse holder, and turn counterclockwise to extend the fuse holder.

WARNING For continued protection against fire hazard, replace the line fuse only with the same type and rating. The use of other fuses or material is prohibited.

AC Power Cord

The analyzer is equipped with a three-wire power cord, in accordance with international safety standards. This cable grounds the analyzer cabinet when connected to an appropriate power line outlet. The cable appropriate to the original shipping location is included with the analyzer.

Various AC power cables are available that are unique to specific geographic areas. You can order additional AC power cables for use in different areas. AC Power Cords, on page 13 lists the available AC power cables, illustrates the plug configurations, and identifies the geographic area in which each cable is appropriate.

Chapter 1

11

Installation and Setup

Power Requirements

NOTE

The front panel switch is a standby switch only; it is not a LINE switch

 

(power disconnecting device).

 

 

 

 

WARNING

Install the product so that the detachable power cord is readily

 

identifiable and easily reached by the operator. The detachable

 

power cord is the product disconnecting device. It disconnects the

 

mains circuits from the mains supply before other parts of the

 

product. The front panel switch is only a standby switch and is not a

 

LINE switch. Alternatively, an externally installed switch or circuit

 

breaker (which is readily identifiable and is easily reached by the

 

operator) may be used as a disconnecting device.

 

 

 

 

CAUTION

Always use the three-prong AC power cord supplied with this product. Failure

 

to ensure adequate earth grounding by not using this cord can cause product

 

damage.

 

 

 

 

CAUTION

This analyzer has autoranging line voltage input. Be sure the supply voltage

 

is within the specified range. (Refer to the specifications guide for your

 

analyzer.)

 

 

12

Chapter 1

HP (Hewlett-Packard) E4402B, E4403B, E4411B, E4408B, E4407B User Manual

Installation and Setup

Power Requirements

Table 1-3. AC Power Cords

Chapter 1

13

Installation and Setup

Power Requirements

Battery Information

The analyzer uses a lithium battery to enable the internal memory to retain data. The date when the battery was installed is on a label on the rear panel of the analyzer. See Figure 1-1.

The minimum life expectancy of the battery is 7 years at 25 ° C, or 1 year at 55 ° C. If you experience problems with the battery or the recommended time period for battery replacement has elapsed, see “Returning an Analyzer for Service” on page 94

If you wish to replace the battery yourself, you can purchase the service documentation that provides all necessary test and maintenance information.

After replacing the analyzer battery, write the date of battery replacement on the rear-panel label.

Figure 1-1. Rear-Panel Battery Information Label

You can order the service documentation for Agilent spectrum analyzers through your Agilent Sales and Service office. The documentation is described under “Options” on page 69.

NOTE

If the analyzer’s clock does not work, the problem is the battery. See

 

“Returning an Analyzer for Service” on page 94.

 

 

WARNING Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent type recommended. Discard used batteries according to the manufacturer’s instructions.

14

Chapter 1

Installation and Setup

Turning on the Analyzer for the First Time

Turning on the Analyzer for the First Time

Plug in the power cord.

WARNING If this product is to be energized via an external auto transformer for voltage reduction, make sure that its common terminal is connected to a neutral (earthed pole) of the power supply.

CAUTION The analyzer is shipped with a a transportation disk inserted in the disk drive to prevent damage to the disk drive during transportation. This transportation disk, or a blank floppy disk, should be inserted in the disk drive whenever transporting the analyzer.

NOTE

Do not connect anything else to the analyzer yet.

Choose a power on preference using the switch on the rear panel (refer to Rear-Panel Features, on page 33, item 14). The PWR ALWAYS ON setting turns the analyzer on whenever external power is applied. This mode is useful if an external power switch is used to control a rack of several instruments. Nevertheless, if you set the analyzer to standby using the front panel Standby key (see Rear-Panel Features, on page 33, item 23) and the external power is removed and restored within 20 seconds, the analyzer will remain in standby.

The PWR NORM setting assigns analyzer on/off control to the front-panel On and Standby keys (see “Front-Panel Connectors and Keys” on page 26, item 23). If the analyzer is on and the external power is removed and restored within 20 seconds, the analyzer will turn on. On the other hand, if the external power is removed and restored after 20 seconds, the analyzer will remain in standby regardless of the front-panel switch settings.

Press the | (On) key to turn the analyzer on.

Information Screen

An information screen appears during the initialization process. The information screen contains the analyzer product number and a URL for accessing product support information on the World Wide Web. See “Additional Information” on page 2.

NOTE

The information screen displays for approximately 10 seconds before the

 

initialization process is complete.

 

Record the firmware revision and serial number, and keep it for reference. If

 

you should ever need to call Agilent Technologies for service or with any

 

questions regarding your analyzer, it will be helpful to have this information

Chapter 1

15

Installation and Setup

Firmware Revision

readily available. You can also obtain the firmware revision and serial number by pressing System, More, Show System.

Allow the analyzer to warm-up for 5 minutes before making a calibrated measurement. To meet its specifications, the analyzer must meet operating temperature conditions.

If the analyzer is an Agilent Technologies E4402B, E4403B, E4404B, E4405B, E4407B, or E4408B, connect a BNC cable from the AMPTD REF OUT to INPUT 50 using an adapter. After a 5 minute warm-up, press System, Alignments, Align Now, All.

CAUTION When operating in dc coupled mode on analyzers with Option UKB ensure protection of the input mixer by limiting the input level to 0 Vdc, +30 dBm.

When operating in ac coupled mode, ensure protection of the input mixer by limiting the input level to 50 Vdc, +30 dBm.

NOTE

It is normal to hear clicking when the Auto Alignment function is on. See

 

“Running Internal Alignments” on page 18 for more information.

 

 

Why Aren’t All the Personality Options Loaded in Memory?

Many measurement personality options are available for use with this instrument. If the option is loaded in the instrument, you must also have a license key entered, to use it.

Some versions of instrument hardware my not have enough memory to accommodate all the options that you have ordered. If this is the case you will need to swap the applications in/out of memory, as needed. It may also be possible to upgrade your hardware to have more memory. Contact your local sales/service office.

Using an External Reference

If you wish to use an external 10 MHz source as the reference frequency, connect an external reference source to the 10 MHz REF IN connector on the rear panel. The signal level should be greater than –15 dBm.

NOTE

It is not necessary to connect the 10 MHz REF OUT to the 10 MHz REF IN on

 

the rear of the analyzer. Doing so results in a “Frequency Reference Error”

 

message.

 

 

1.To use an external frequency reference, connect it to the EXT REF IN connector on the rear panel (see “Rear-Panel Features” on page 33).

Firmware Revision

To view the firmware revision of your analyzer, press System, More, Show System. If you call Agilent Technologies regarding your analyzer, it is helpful to have this revision and the analyzer serial number available.

16

Chapter 1

Installation and Setup

Firmware Revision

TIP

You can get automatic electronic notification of new firmware releases and

 

other product updates/information by subscribing to the Agilent Technologies

 

Test & Measurement E-Mail Notification Service for the PSA and ESA Series

 

at http://www.agilent.com/find/notifyme

 

 

Chapter 1

17

Installation and Setup

Running Internal Alignments

Running Internal Alignments

Each time the analyzer is powered on, the internal alignment routine runs automatically.

The analyzer was shipped from the factory with the Alignments mode set to Auto,

Align All.

NOTE

When the Alignment routine runs, you will hear the attenuator settings

 

changing, which generates noise. This is not an indication of trouble.

 

 

Manually Performing an Alignment

If the analyzer is an Agilent Technologies E4402B, E4403B, E4404B, E4405B, E4407B, or E4408B, connect a BNC cable from the AMPTD REF OUT to INPUT 50 using an adapter. After a 5 minute warm-up, press System, Alignments, Align Now, All.

NOTE

It is normal to hear clicking when the Auto Alignment function is on. During

 

the interval between sweeps, portions of the analyzer’s circuitry are

 

realigned. Some of the circuitry is controlled by relays. It is the rapid

 

switching of these relays between sweeps which causes the clicking sound.

 

Under normal operation, these relays will last over 50 years.

 

To eliminate the clicking sound, turn off the automatic alignment. (See the

 

Alignments key description in your User’s guide.”) With Auto Align turned off,

 

however, the Align Now All function should be performed periodically. For

 

more information on how often to perform Align Now All when the Auto

 

Alignment function is off, refer to the appropriate “Specifications and

 

Characteristics” chapter in your specifications guide.

 

 

If Auto Align, Off is selected, refer to the Specifications guide for the conditions required to maintain calibration.

18

Chapter 1

Installation and Setup

Printer Setup and Operation

Printer Setup and Operation

A printer can be connected to your analyzer if it is equipped with an external I/O interface. Supported printers accept Hewlett-Packard Printer Control Language Level 3 (PCL3) or 5 (PCL5). Refer to the documentation or specifications supplied with your printer, or contact the manufacturer to identify your printer’s language.

Equipment

IEEE 1284 compliant printer cable.

Supported and tested printers are listed below. Note that there are many PCL3/5 printers that may work with your analyzer, however, they have not been tested.

PCL3 printers include most HP DeskJet printers.

PCL5 printers include most HP LaserJet printers.

NOTE

The following printers are not compatible with your analyzer.

 

• HP Deskjet 720C, 722C, 820C and 1600C

 

• Epson MX-80, FX-85, Stylus, and LQ-570

 

 

 

 

 

 

 

 

 

 

 

Printer Models

Language Type

Color Capable

 

 

 

 

 

HP DeskJet 310, 320, 350C, 400L

PCL3

yes

 

 

 

 

 

HP DeskJet 500C, 550C, 600, 660C, 672C,

PCL3

yes

 

680C, 682C, 690C, 693C

 

 

 

 

 

 

 

HP DeskJet 840C, 850C, 870C, 890C, 895C

PCL3

yes

 

 

 

 

 

HP DeskJet 935C, 970C, 990C

PCL3

yes

 

 

 

 

 

HP DeskJet 1120C, 1150C

PCL3

yes

 

 

 

 

 

HP Inkjet 2000C

PCL 3

yes

 

 

 

 

 

HP LaserJet III

PCL3/5

no

 

 

 

 

 

HP LaserJet 4P

PCL3/5

no

 

 

 

 

 

HP LaserJet 5L, 5M, 5N, 5P, 5SI

PCL3/5

no

 

 

 

 

 

HP LaserJet 6L, 6MP

PCL5

no

 

 

 

 

 

HP LaserJet 2100 Series,

PCL3/5

no

 

 

 

 

 

HP LaserJet 4050N

PCL3/5

yes

 

 

 

 

 

HP LaserJet 5000GN

PCL3/5

yes

 

 

 

 

 

HP Professional Series 2500CM

PCL3

yes

 

 

 

 

 

HP Professional Series 2500CM

PCL3

yes

 

 

 

 

 

 

Chapter 1

19

Installation and Setup

Printer Setup and Operation

Interconnection and Setup

1.Turn off the printer and the analyzer.

2.Connect the printer to the analyzer parallel I/O interface connector using an IEEE 1284 compliant parallel printer cable.

3.If appropriate, configure your printer using configuration menus or switches. Refer to your printer’s documentation for more specific information on configuring your printer.

4.Turn on the analyzer and printer.

5.Press Print Setup on the front panel and then press the Printer Type menu key. Printer Type accesses the following keys:

None

None disables the analyzer from attempting to print to a printer. This is

 

the appropriate setting if no printer is connected to the analyzer.

Custom

Custom allows you to access the Define Custom menu keys. The Define

 

Custom menu keys allow you to specify printer characteristics such as

 

PCL Level and printer color capability.

Auto

Auto enables the analyzer to automatically attempt to identify the

 

connected printer when the Print key is pressed or when Printer Type is

 

set to Auto.

6.Press Printer Type to access the Printer Type menu keys. Press Auto to make the analyzer attempt to identify the connected printer. When you press Auto, the analyzer will respond in one of the three following ways:

The Print Setup menu will be displayed with the Auto key selected and no new message will be displayed in the display status line. This indicates that the analyzer has successfully identified the connected printer and no further setup is required. As long as Auto remains selected in the Printer Type menu, the analyzer will attempt to identify the printer when the front panel Print key is pressed. The selected printer will be displayed by pressing System, More 1 of 3, Show System.

The Print Setup menu will be displayed with the Custom key selected and one of the following diagnostic messages will be displayed in the display status line:

Unknown printer, Define Custom to set up printer

No printer response, Define Custom to set up printer Invalid printer response, Define Custom to set up printer

This indicates that the analyzer was unable to automatically identify the connected printer, and Custom has been selected in the Printer Type menu. Press Print Setup, Define Custom to select specific printer characteristics such as the printer language (PCL3 or PCL5) and color printing capability. Once you have set these characteristics to match those of your connected printer, the printer setup process is complete. As long as Custom remains selected in the Printer Type menu, the analyzer will not attempt to automatically identify the connected printer when the front panel Print key is pressed.

The Print Setup menu will be displayed with the None key selected and the following message will appear in the display status line:

20

Chapter 1

Installation and Setup

Printer Setup and Operation

Unsupported printer, Printer Type set to None

This indicates that the analyzer has successfully identified the connected printer, but the printer is not supported by the analyzer. As long as None is selected in the Printer Type menu, the analyzer will respond to any print command by displaying the message Printer Type is None in the display status line.

7.Select the desired paper size by pressing Print Setup, More, Page Size, then choose the appropriate page size for which your printer is configured. This setting will remain unchanged with Preset or Power Cycle.

The factory default page size is Letter. The page size will be reset to Letter if System,

More, Restore Sys Defaults is executed.

Testing Printer Operation

When you have completed the analyzer’s printer setup, press Print Setup, then press Print on the front panel. If the printer is ready and the printer setup was successful, a printout of the analyzer display will be printed. If the printer is not ready, the message Printer Timeout will appear on the analyzer display. Printer Timeout will remain on the display until the printer is ready or until you press ESC to cancel the printout request.

NOTE

There may be some small discrepancies in the color mapping of the analyzer

 

display to your color printer. Due to differences in display and printer

 

technologies, the default display colors do not map exactly to the printer

 

colors. For example trace 1 is yellow on your analyzer display while it maps to

 

green on your printer.

 

 

Chapter 1

21

Installation and Setup

Protecting Against Electrostatic Discharge

Protecting Against Electrostatic Discharge

Electrostatic discharge (ESD) can damage or destroy electronic components (the possibility of unseen damage caused by ESD is present whenever components are transported, stored, or used).

Test Equipment and ESD

To help reduce ESD damage that can occur while using test equipment:

Before connecting any coaxial cable to an analyzer connector for the first time each day, momentarily short the center and outer conductors of the cable together.

Personnel should be grounded with a 1 Mresistor-isolated wrist-strap before touching the center pin of any connector and before removing any assembly from the analyzer.

Be sure that all instruments are properly earth-grounded to prevent build-up of static charge.

WARNING Do not use these first three techniques when working on circuitry with a voltage potential greater than 500 volts.

Perform work on all components or assemblies at a static-safe workstation.

Keep static-generating materials at least one meter away from all components.

Store or transport components in static-shielding containers.

Always handle printed circuit board assemblies by the edges. This reduces the possibility of ESD damage to components and prevent contamination of exposed plating.

For information on ordering static-safe accessories, see “Accessories” on page 83.

Additional Information about ESD

For more information about ESD and how to prevent ESD damage, contact the Electrostatic Discharge Association (http://www.esda.org). The ESD standards developed by this agency are sanctioned by the American National Standards Institute (ANSI).

22

Chapter 1

Installation and Setup

Legal Information

Legal Information

WARNING This is a Safety Class 1 Product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall be inserted only in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.

If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.

Warranty

This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of three years from date of shipment. During the warranty period, Agilent Technologies Company will, at its option, either repair or replace products that prove to be defective.

For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.

Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free.

LIMITATION OF WARRANTY

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.

NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

EXCLUSIVE REMEDIES

THE REMEDIES PROVIDED HERIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.

Chapter 1

23

Installation and Setup

Legal Information

24

Chapter 1

2 Front and Rear Panel Features

This chapter gives you an overview of the front and rear panels of your analyzer. For details on analyzer keys and remote programming, refer to the User’s and Programmer’s Guide. For connector specifications (including input/output levels), see the Specifications guide.

25

Front and Rear Panel Features

Front Panel Overview

2.1 Front Panel Overview

This section provides information on the analyzer’s front panel, including:

Front Panel Connectors and Keys, see below

“Display Annotations” on page 30

2.1.1 Front-Panel Connectors and Keys

1

Viewing Angle keys adjust the display so that it can be optimally viewed from

 

different angles.

2

Esc. The Esc (escape) key cancels any entry in progress. Esc will abort a print (if

 

one is in progress) and clear error messages from the status line at the bottom

 

of the display. It also clears input and tracking generator overload conditions.

3

Menu keys are the unlabeled keys next to the screen. The menu key labels are

 

the annotation on the screen next to the unlabeled keys. Most of the labeled

 

keys on the analyzer front panel (also called front-panel keys) access menus of

 

keys having related functions.

4

FREQUENCY Channel, SPAN X Scale, and AMPLITUDE Y Scale are the three large

 

keys that activate the primary analyzer functions and access menus of related

 

functions. The secondary labels on these keys (Channel, X Scale, and Y Scale)

 

are used in some measurements.

5

CONTROL functions access menus that adjust the resolution bandwidth, adjust

 

the sweep time, and control the analyzer display. They also set other analyzer

26

Chapter 2

 

Front and Rear Panel Features

 

Front Panel Overview

 

parameters needed for making measurements.

6

MEASURE accesses a menu of keys that automate some common analyzer

 

measurements. Once a measurement is running, Meas Setup accesses additional

 

menu keys for defining your measurement. Meas Control and Restart access

 

additional measurement control functions.

7

SYSTEM functions affect the state of the entire analyzer.

 

Various setup and alignment routines are accessed with the System key.

 

The green Preset key resets the analyzer to a known state.

 

The File key menu saves/loads setups, traces, states, limit-line tables,

 

screens, measurement results, and amplitude correction factors to or from

 

analyzer memory or the floppy disk drive. The Save key immediately

 

executes the Save Now function defined under File in your User’s guide.

 

The Print Setup menu keys configure hardcopy outputs. The Print key

 

immediately sends hardcopy data to the printer. See your User’s guide for

 

more details.

8

MARKER functions control the markers, read out frequencies and amplitudes

 

along the analyzer trace, automatically locate the signals of highest amplitude,

 

and access functions like Marker Noise and Band Power.

9

The Media Door on the right side of the front panel accesses the 3.5 inch disk

 

drive and the Earphone connector. The earphone connector provides a

 

connection for an earphone jack which bypasses the internal speaker.

10

The Data Control Keys, which include the step keys, knob, and numeric

 

keypad, change the numeric value of an active function such as center

 

frequency, start frequency, resolution bandwidth, and marker position.

The data controls will change the active function in a manner prescribed by that function. For example, you can change center frequency in fine steps with the knob, in discrete steps with the step keys, or to an exact (1 Hz resolution) value with the numeric keypad.

The Knob provides fine incremental changes of functions such as center frequency, reference level, and marker position. Clockwise rotation of the knob increases values. The extent of alteration is determined by the size of the measurement range. The speed at which the knob is turned affects the rate at which the values are changed.

For slow sweeps, the analyzer uses a smooth panning feature which is designed to move the trace display to the latest function value as the knob is turned. When center, stop or, start frequency or reference level is adjusted, the signal will shift right or left or up or down with the rotation of the knob before a new sweep is actually taken. An asterisk is placed in the message block (the upper right-hand corner of the analyzer display) to indicate that the data on the screen does not reflect data at the current setting.

The Numeric Keypad allows entry of exact values for many of the analyzer functions. You may include a decimal point in the number portion. If not, the decimal point is placed at the end of the number.

Numeric entries must be terminated with a units key. When a numeric entry is begun, the menu keys show the units key labels. The units keys change depending on what the active function is. For example, the units keys for

Chapter 2

27

Front and Rear Panel Features

Front Panel Overview

frequency span are GHz, MHz, kHz, and Hz, whereas the units for reference level are +dBm, dBm, mV, µ V, and µ A.

NOTE If an entry from the numeric keypad does not coincide with an allowed function value (for example, that of a 12 MHz bandwidth), the analyzer defaults to the nearest allowable value.

 

The Step Keys ( ) increase or decrease the active function value. The step

 

size depends upon the current analyzer measurement. Each press results in

 

a single step change. For those parameters with fixed values (resolution

 

bandwidth), the next value in a sequence is selected each time a step key is

 

pressed. Step size is predictable (e.g., 10% of span for center frequency) and

 

can be set for some functions (i.e., center frequency). Out-of-range values or

 

out-of-sequence values will not occur using these keys.

11

VOLUME. The VOLUME knob adjusts the volume of the internal speaker. The

 

speaker is turned on and off with the Speaker On Off key in the Det/Demod menu.

12

EXT KEYBOARD. The EXT KEYBOARD connector is a 6-pin mini-DIN connector.

 

The keyboard can be used to enter screen titles and filenames.

 

 

NOTE

To avoid damage to the analyzer, always turn off power before plugging a keyboard

 

into the analyzer.

 

 

13

PROBE POWER provides power for high-impedance ac probes or other

 

accessories. (+15 V, − 12.6 V, 150 mA maximum)

14

LO OUTPUT provides the proper local oscillator signal for use with external

 

mixers (Option AYZ).

15

IF INPUT connects to the IF OUTPUT of the external mixer (Option AYZ).

16

Return. The Return key accesses the previously selected menu. Continuing to

 

press Return accesses earlier menus. Return also terminates entry of alpha

 

numeric functions (e.g., Title).

17

AMPTD REF OUT provides an amplitude reference signal of 50 MHz at –20 dBm.

 

Agilent ESA models E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B

 

only.

18

Tab Keys are used to move around in the Limit editor, the Correction editor and

 

similar table-driven forms.

19

INPUT 50(INPUT 75for Option 1DP) is the signal input for the analyzer.

 

 

 

CAUTIONWhen operating in dc coupled mode on analyzers with Option UKB,

 

ensure protection of the input mixer by limiting the input level to

 

0 Vdc, +30 dBm.

 

When operating in ac coupled mode, ensure protection of the

 

input mixer by limiting the input level to 50 Vdc, +30 dBm.

 

 

20

The Next Window key can be used to select the active window in functions which

 

support split-screen display modes, such as Zone markers. (Refer to “Zone” in

 

the User’s guide for more information.) In such modes, pressing Zoom switches

28

Chapter 2

 

Front and Rear Panel Features

 

Front Panel Overview

 

between the split-screen and full-sized display of the active window.

21

Help. Press the Help key and then any front panel or menu key to get a short

 

description of the key function and the associated SCPI command. The next key

 

you press will remove the help window from the display.

22

RF OUT 50for Option 1DN or RF OUT 75(for Option 1DQ) is the source

 

output for the built-in tracking generator. Option 1DN or 1DQ only.

CAUTION If the tracking generator output power is too high, it may damage the device under test. Do not exceed the maximum power that the device under test can tolerate.

23

The | (On) key turns the analyzer on, while the Standby key turns most of the

 

analyzer off. An analyzer alignment is performed (if Auto Align is on) every time

 

the analyzer is turned on. After turning on the analyzer, allow 5 minutes of

 

warm-up time to ensure the analyzer will meet all specifications.

NOTE The analyzer continues to draw power even if the line power switch is in standby. The detachable power cord is the analyzer disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the analyzer. The front-panel switch is only a standby switch and is not a LINE switch (disconnecting device).

Chapter 2

29

Front and Rear Panel Features

Front Panel Overview

2.1.2 Display Annotations

Table 2-1

Screen Annotation

 

 

 

 

Item

Description

Associated Function Key

 

 

 

1a

Detector mode

Detector

2

Reference level

Ref Level

 

 

 

3

Active function block

Refer to the description of the activated

 

 

 

function.

 

 

 

 

4

Screen title

 

Change Title

 

 

 

5

Time and date display

Time/Date On Off

 

 

 

6a,b

RF attenuation

Attenuation Auto Man

7

Marker frequency

Marker or

 

 

 

Marker Count On Off

 

 

 

8

Marker amplitude

Marker

 

 

 

 

30

Chapter 2

Loading...
+ 70 hidden pages