Flintec FRD-57 User Manual

Remote Display
Flintec GmbH Bemannsbruch 9 74909 Meckesheim Germany
9 In case of any questions… Tel. : +49 6226 92400
9 You can visit us
http://www.flintec.net
FLINTEC
TYP FRD-57
Remote Display FRD-57
Datum
27.09.04
Code
FRD-57
Seite
1 von 3
Connection FRD- 57
The connection should be conducted with an 2-wire shielded cable. The shield should be connected at the FT-Indicator .
Fig.1 shows the terminal block, Jumper and Dip switch at the remote indicator.
Fig. 1
Spannungsversorgung 10VAC
20mA current loop
RS485
RS232
A
B
+Tx
-Tx
Tx
SGND
SHLD
FLINTEC
JP1
JP2
1 2 3 4
S1
Fig. 2
Remote Display FRD-57
Baudrate 1200 2400 9600 19200
Dip SW (on)
Data E,7,1 N,8,1 E,8,1
Dip SW (on)
ST2
Settings of the Dip Switch ( S1 )
xxxxxxx 1 2 1,2
xxxxxxx 3 4 3,4
Jumper Settings ( JP1 , JP 2 )
RS-232 20mA cur.Loop RS-485
Datum
27.09.04
xxxxxxx
Code
FRD-57
Seite
2 von 3
Sample for Settings Indikator FT-01 / FT-02
.
Connection via RS-232 Interface
Connect the Tx and GND at the FT-Indicator with Tx and SGND at the remote Indicator. The shield should conneted via the screwed cable gland (single-edge) Please consider the max. cable length with RS-232 respectively TTY Interface is 15 meters and with RS-485 max. 1000 meters should not exceed.
The Jumper JP1 and JP2 at the remote indicator must be open .
The Dip switch S1 adjust as following :
1 = on ( 2400 Baud )
2 = off
3 = on ( non parity, 8 data bits, 1 Stop bit ) 4 = off
Settings at Indicator FT-01 / FT-02 Setup 2
Setup 2 Parameter Value Description
FLINTEC
2.t 02 cont. sending of weight value
2.7 0 control mode misprint off
2.d 08 8 Data- bits / nonparity
Remote Display FRD-57
Datum
27.09.04
Code
FRD-57
Seite
3 von 3