3.2. Key features and Specifications .................................................................................................................. 6
3.4.1. Accessories supplied with the instrument .................................................................................................................. 8
4.2.1. Power Supply and Grounding .................................................................................................................................... 9
4.2.3. Digital I/O Connection .............................................................................................................................................. 11
4.2.4. Serial Interface RS232 ............................................................................................................................................. 11
4.2.5. Communication Interface ......................................................................................................................................... 11
5.1. Installation of the xFace Software.............................................................................................................. 13
5.2. Connection to the xFace Software............................................................................................................. 13
5.3. Setup and Calibration ................................................................................................................................ 14
5.3.2. RS-485 Serial Interface Setup ................................................................................................................................. 15
5.3.3. Scale Build and Calibration ...................................................................................................................................... 16
5.4. A/D Status (Performance Test) .................................................................................................................. 19
5.5. Setup of Digital I/O ..................................................................................................................................... 19
5.6. Bus Interface Setup ................................................................................................................................... 20
5.7. Bus Addressing via Setup Switch .............................................................................................................. 21
5.8. Back up Settings and Calibration Data ...................................................................................................... 21
6. FAD-40 / FAD-40MB WITH RS-485 INTERFACE ....................................................... 22
6.1. Front View of FAD-40 (MB) ....................................................................................................................... 22
6.7. Modbus Data Structure (for FAD-40MB and FAD-40EN only) .................................................................. 28
7. FAD-40PB WITH PROFIBUS INTERFACE ................................................................ 31
7.1. Front View of FAD-40PB ........................................................................................................................... 31
7.6. Profibus and Profinet Data Structure (for FAD-40Px only) ........................................................................ 34
8. FAD-40PN WITH PROFINET INTERFACE ................................................................. 38
8.1. Front View of FAD-40PN ........................................................................................................................... 38
8.6. Profibus and Profinet Data Structure (for FAD-40Px only) ........................................................................ 41
9. FAD-40EN WITH ETHERNET INTERFACE ................................................................ 42
9.1. Front View of FAD-40EN ........................................................................................................................... 42
9.6. Modbus Data Structure .............................................................................................................................. 44
10. FAD-40CO WITH CANOPEN INTERFACE ................................................................. 45
10.1. Front View of FAD-40CO ....................................................................................................................... 45
13. INDEX .......................................................................................................................... 54
RIGHTS AND LIABILITIES
All rights reserved.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any
means, mechanical, photocopying, recording, or otherwise, without the prior written permission of Flintec
GmbH
No patent liability is assumed with respect to the use of the information contained herein. While every
precaution has been taken in the preparation of this book, FLINTEC assumes no responsibility for errors or
omissions. Neither is any liability assumed for damages resulting from the use of the information contained
herein.
The information herein is believed to be both accurate and reliable. FLINTEC, however, would be obliged to be
informed if any errors occur. FLINTEC cannot accept any liability for direct or indirect damages resulting from
the use of this manual.
FLINTEC reserves the right to revise this manual and alter its content without notification at any time.
Neither FLINTEC nor its affiliates shall be liable to the purchaser of this product or third parties for damages,
losses, costs, or expenses incurred by purchaser or third parties as a result of: accident, misuse, or abuse of
this product or unauthorized modifications, repairs, or alterations to this product, or failure to strictly comply with
FLINTEC operating and maintenance instructions.
FLINTEC shall not be liable against any damages or problems arising from the use of any options or any
consumable products other than those designated as Original FLINTEC Products.
NOTICE: The contents of this manual are subject to change without notice.
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 2 of 56
1. SAFETY INSTRUCTIONS
CAUTION READ this manual BEFORE operating or servicing this equipment. FOLLOW
these instructions carefully. SAVE this manual for future reference. DO NOT allow untrained
personnel to operate, clean, inspect, maintain, service, or tamper with this equipment.
ALWAYS DISCONNECT this equipment from the power source before cleaning or performing
maintenance. CALL FLINTEC ENGINEERING for parts, information, and service.
WARNING ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS EQUIPMENT.
EXERCISE CARE WHEN MAKING CHECKS, TESTS AND ADJUSTMENTS THAT MUST BE
MADE WITH POWER ON. FAILING TO OBSERVE THESE PRECAUTIONS CAN RESULT
IN BODILY HARM.
WARNING FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD CONNECT TO
PROPERLY GROUNDED OUTLET ONLY. DO NOT REMOVE THE GROUND PRONG.
WARNING DISCONNECT ALL POWER TO THIS UNIT BEFORE REMOVING THE FUSE
OR SERVICING.
WARNING BEFORE CONNECTING/DISCONNECTING ANY INTERNAL ELECTRONIC
COMPONENTS OR INTERCONNECTING WIRING BETWEEN ELECTRONIC EQUIPMENT
ALWAYS REMOVE POWER AND WAIT AT LEAST THIRTY (30) SECONDS BEFORE ANY
CONNECTIONS OR DISCONNECTIONS ARE MADE. FAILURE TO OBSERVE THESE
PRECAUTIONS COULD RESULT IN DAMAGE TO OR DESTRUCTION OF THE
EQUIPMENT OR BODILY HARM.
CAUTION OBSERVE PRECAUTIONS FOR HANDLING ELECTROSTATIC SENSITIVE
DEVICES.
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 3 of 56
2. DECLARATION OF CONFORMITY
EU-Konformitätserklärung
EU-Declaration of Conformity
Monat/Jahr: month/year:
04/2016
Hersteller: Manufacturer:
Flintec GmbH
Anschrift: Address:
Bemannsbruch 9
D-74909 Meckesheim
Deutschland / Germany
Produktbezeichnung: Product name:
FAD-40 A/D Converter Series
Das bezeichnete Produkt stimmt mit folgenden Vorschriften der Europäischen Richtlinien überein:
This product confirms with the following regulations of the Directives of the European Community
Richtlinie 2014/30/EU des Europäischen
Parlaments und des Rates vom 26. Februar 2014
zur Angleichung der Rechtsvorschriften der
Mitgliedstaaten über die elektromagnetische
Verträglichkeit und zur Aufhebung der Richtlinie
2004/108/EG.
Directive 2014/30/EU of the European Parliament and of
the Council of 26th February 2014 on the approximation of
the laws of the Member States relating to electromagnetic
compatibility and repealing Directive 2004/108/EC.
Richtlinie 2014/35/EU Niederspannungs-Richtlinie
Directive 2014/35/EU Low Voltage Directive
Die Absicherung aller produktspezifischen
Qualitätsmerkmale erfolgt auf Basis eines
zertifizierten Qualitätsmanagement-Systems nach
DIN ISO 9001.
All product-related features are assured by a quality
system in accordance with ISO 9001.
Diese Erklärung bescheinigt die Übereinstimmung mit
den genannten Richtlinien, beinhaltet jedoch keine
Zusicherung von Eigenschaften.
This declaration certifies the conformity with the listed
directives, but it is no promise of characteristics.
Folgende Normen werden zum Nachweis der Übereinstimmung mit den Richtlinien eingehalten:
As a proof of conformity with the directives following standards are fulfilled:
EN 61326-1
Elektrische Mess-, Steuer-, Regel- und Laborgeräte - EMV-Anforderungen - Teil 1: Allgemeine
Anforderungen (IEC 61326-1:2013)
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General
requirements (IEC 61326-1:2012)
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 4 of 56
3. INTRODUCTION
FAD-40 A/D Converter
only with RS485 interface
FAD-40PB A/D Converter with
Profibus DP interface
FAD-40EN A/D Converter with
Ethernet TCP/IP interface
FAD-40MB A/D Converter with
Modbus RTU interface
FAD-40PN A/D Converter with
Profinet interface
FAD-40CO A/D Converter with
CANopen interface
3.1. Overview
The type FAD-40 A/D Converter Series consists of powerful and economic state-of-the-art instruments for static
and dynamic weighing applications plus force and torque measurements including digital I/O control.
Each instrument of the series converts the analogue low level signal from a load cell or a strain gauge sensor to
a digital high-resolution and high-accuracy signal and transmits the digital data to an external PLC or PC
system. As a special feature the instruments can switch between unipolar and bipolar input range without
affecting the external resolution. Digital I/O ports give further possibilities to extend the bus system with I/O
control in the cabinet and fast response setpoint output from the weighing instrument.
The type FAD-40 A/D Converter Series comprises various instruments for different industrial bus systems:
Figure 3.1 – FAD-40 A/D Converter series instruments
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 5 of 56
3.2. Key features and Specifications
Key Features
FAD-40
FAD-40MB
FAD-40PB
FAD-40PN
FAD-40EN
FAD-40CO
Bipolar signal input in Force and Count modes
Yes
Yes
Yes
Yes
Yes
Yes
500 to 100 000 counts external resolution at calibrated data
Yes
Yes
Yes
Yes
Yes
Yes
Unipolar 1 million and bipolar 2 million counts is adjusted and filtered
to external resolution in count mode
Yes
Yes
Yes
Yes
Yes
Yes
High internal resolution up to 8 million counts
Yes
Yes
Yes
Yes
Yes
Yes
Up to 800 conversions per second
Yes
Yes
Yes
Yes
Yes
Yes
2 digital inputs/outputs + 1 additional digital outp. (configurable)
Yes
Yes
Yes
Yes
Yes
Yes
Standard Serial interface RS-232C
No
No
Yes
Yes
Yes
Yes
Serial interface RS-485A
Yes
Yes
No
No
No
No
Modbus RTU interface
No
Yes
No
No
No
No
Profibus DP-V1 interface
No
No
Yes
No
No
No
Profinet interface
No
No
No
Yes
No
No
Ethernet (Modbus TCP/IP) interface
No
No
No
No
Yes
No
CANopen interface
No
No
No
No
No
Yes
BSI data interface
Yes
Yes
No
No
No
No
Bus address selection by setup switch (from 0 to 7)
Yes
Yes
Yes
No
No
Yes
Setup by Flintec xFace PC software
Yes
Yes
Yes
Yes
Yes
Yes
Auto-zero tracking and auto-zero at power-up
Yes
Yes
Yes
Yes
Yes
Yes
Motion detection
Yes
Yes
Yes
Yes
Yes
Yes
Zeroing and Taring by bus commands
Yes
Yes
Yes
Yes
Yes
Yes
Adaptive digital filter for fast and stable reading
Yes
Yes
Yes
Yes
Yes
Yes
Up to 6 load cells (350 Ω) or 18 load cells (1100 Ω)
Yes
Yes
Yes
Yes
Yes
Yes
Electronic calibration (eCal) without test weights
Yes
Yes
Yes
Yes
Yes
Yes
Zero and Span calibrations over bus interface
Yes
Yes
Yes
Yes
Yes
Yes
Zero adjustment
Yes
Yes
Yes
Yes
Yes
Yes
Span adjustment with test weights
Yes
Yes
Yes
Yes
Yes
Yes
Span adjustment with temporary zeroing for unloaded tanks
Yes
Yes
Yes
Yes
Yes
Yes
Power-on testing with setup switch
Yes
Yes
Yes
Yes
Yes
Yes
24 V DC power supply (11 to 28 V DC input voltage range)
Yes
Yes
Yes
Yes
Yes
Yes
A/D Converter:
A/D converter type:
24 bit Delta-Sigma radiometric with integral analog and digital filters
Analog input range:
0 mV to +18 mV ( unipolar ) ; - 18 mV to +18 mV ( bipolar )
Linearity:
< 0.0015 % FS
Temperature coefficient:
< 2 ppm/°C
Min. input per vsi
0.1 μV/d (non approved)
Conversion rate:
Up to 800 measurement values per second
Internal resolution:
Up to 8 million counts
External resolution
Up to 100 000 counts (weight value, force, torque) respective
1 million raw counts (unipolar) respective 2 million counts (bipolar)
Calibration and Weighing Functions:
Calibration:
Electronic calibration without test weights (eCal) or calibration by test weights
Digital filter:
10 step programmable adaptive filter
Weighing functions:
Tare, zero, auto zero tracking, motion detection, auto-zero at power-up, save
tare at power-off, increased resolution
Load cells:
Excitation:
5 V DC at 58...1200 Ω, max. 100 mA
Number of load cells:
Up to 6 load cells à 350 Ω or 18 load cells à 1100 Ω in parallel
Connection:
4- or 6-wire technique, cable length 250 m/mm² for 6-wire connection
Specifications
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 6 of 56
Communication and Setup:
Serial interface
RS232C with 9600 baud (8,N,1);
for FAD-40 and FAD-40MB: RS485 with up to 57600 baud (8N1, 7E1, 7O1)
Other interfaces
Depends on instrument type
Response time
< 4 ms (delay after each read or write command)
Setup and calibration
By PC software via RS232C respective RS485, backup data stored on PC
Digital Inputs + Outputs
2x configurable I/O
Selectively configured as input (10...26 V DC) or
open collector output (24 V DC, max. 100 mA)
1x additional output
Open collector output (24 V DC, max. 100 mA)
Input functions
Zero, tare, clear tare or control input to communication interface
Output functions
Setpoint output with / without hysteresis, tolerance band output or control
output from communication interface
Power supply:
DC power supply
10 to 28 VDC, < 200 mA, not galvanically isolated
Environment and Enclosure:
Operation temperature:
Between -10 °C and +40 °C at maximum 85% RH max, non-condensing
Enclosure and protection
Polyamide, for DIN-rail mounting, IP20
Instrument with RS485 interface: Type FAD-40
Communication:
Serial interface RS485
1200 to 57600 baud (8N1, 7E1, 7O1), bus capability up to 31 units
Communication mode
Requested
Dimensions & weight
99 x 22.5 x 114.5 mm (L x W x H), weighs appr. 110 g
Instrument with Modbus RTU interface: Type FAD-40MB
Communication:
Serial interface RS485
1200 to 57600 baud (8N1, 7E1, 7O1), bus capability up to 31 units
Communication mode
Requested or Modbus RTU
Address range
1...31
Dimensions & weight
99 x 22.5 x 114.5 mm (L x W x H), weighs appr. 110 g
Instrument with Profibus DP interface: Type FAD-40PB
Communication:
Profibus DP-V0 + DP-V1
9,6 kbit/s to 12 Mbit/s (automatic), galvanically isolated interface
Address range
1...126
Dimensions & weight
99 x 45 x 114.5 mm (L x W x H), weighs appr. 150 g
99 x 45 x 114.5 mm (L x W x H), weighs appr. 150 g
22,5 mm 45 mm
99 mm
114,5 mm
View A View B Side view
View A: FAD-40, FAD-40MB
View B: FAD-40PB, FAD-40PN, FAD-40EN, FAD-40CO
The following accessories are supplied together with the instrument. If any part
is missing, please contact your supplier.
4-pin and 5 mm pitch plug, light gray
1 1 2 2 2 2 7-pin and 3.81 mm pitch plug for load cell cable, black
1 1 1 1 1 1 6-pin and 3.81 mm pitch plug for Digital I/O and RS-485, black
1 1 3-pin and 3.81 mm pitch plug for Digital I/O, green
1 1 1 1 Installation CD (xFace software, technical documentation)
1 1 1 1 1
1
3.3. Housing
FAD-40 instruments come within a polyamide housing sealed to IP20. They are prepared for mounting on NS
37/7 or NS 35/15 standard DIN rails (see drawings).
Figure 3.2 – Dimensions
3.4. Accessories
3.4.1. Accessories supplied with the instrument
Table 3.1 – Accessories supplied with instrument
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 8 of 56
4. INSTALLATION
FAD-40 and
FAD-40MB
(front view)
FAD-40EN,
FAD-40PB,
FAD-40PN and
FAD-40CO
(front view)
24V 0V
24V 0V
TX RX GND Shield
Warning: Do not forget to connect the Shield pin to the reference ground.
PRECAUTION: Please read this manual carefully before you install the instrument. If you apply all
recommendations in this chapter you will increase the reliability and long term performance of your system.
4.1. Recommendations
4.1.1. Control Cabinet Design
Warning: Please follow the following warnings for designing the control cabinet which will increase the
reliability of your system.
The control cabinet should be designed therefor the A/D converters can operate safely. The panel should be
placed in a clean area, without getting direct sun light if possible, with a temperature between -10 ºC and +40
ºC, humidity not exceeding 85% non-condensing. All external cables should be installed safely to avoid
mechanical damages.
FAD-40 instruments are very low level signal measuring instruments. To avoid electrical noise, the instruments
should be separated from equipment that produces electrical noise. Preferably use a metal cabinet against
radio frequency interference, to protect against electromagnetic disturbance the cabinet shall be connected to
ground. Keep the load cell cable trays separated from others, if possible. If there is noise-generating equipment
such as heavy load switches, motor control equipment, inductive loads etc., please be careful against the EMC
interference in the cabinet. If possible protect FAD-40 instruments by a Faraday cage or install them in a
separate section or install them far away from this kind of equipment. Install parallel reverse diodes to the DC
inductive loads like relays, solenoids etc. to minimize voltage peaks on the DC power lines.
4.1.2. Cabling
All cables coming to the control cabinet shall be shielded. Please use separate cable trays for these low signal
level cables. Distance from load cell cables, interface cables and DC power supply cables to power line cables
shall be 50 cm at minimum.
4.1.3. Mechanical Installation
After designing the control panel and installing DIN rails according to the recommendation in this chapter, place
the FAD-40 instruments on the DIN rail Be sure that the mechanical installation of the instruments is done
properly.
4.2. Electrical Connections
Warning: Please always remember that FAD-40 instruments are very low voltage measuring instruments. Your
control cabinet design and proper installation increases the reliability and the performance of the instrument.
Please do not forget that the instrument must be powered off before inserting or removing any peripheral
connector. All required electrical connections should be done as described below.
4.2.1. Power Supply and Grounding
The power supply voltage of the instrument shall be between 12 V DC and 28 V DC. The current consumption
of the power supply will be calculated by multiplying 0.2 A and the quantity of instruments. The pin configuration
of the 24 V DC power supply connector located at the bottom front of the instrument is shown below in
figure 4.1.
Figure 4.1 – The pin layout of 24 V DC connectors
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 9 of 56
The quality of the instrument’s ground will determine the accuracy and the safety of your measuring system. A
Load Cell
Connector
4 wire LC connection
6 wire LC connection
WARNING: Connect the load cell cable shield to the reference ground or the
shield pin of the load cell connector.
Junction box 6 Wire-cableFT-10 Indicator
Bridges
4W-Load cell
Shield
Shield
+ Exc
+ Sig
- Sig
- Exc
Shield
proper ground connection is needed to minimize extraneous electrical noise effects on the measurement. A
poor ground can result in an unsafe and unstable operation. It is important that the instrument should not share
power lines with noise-generating equipment such as heavy load switching, motor control equipment, inductive
loads, etc. If the condition of the power line in the plant is poor, prepare a special power line and grounding.
Before interfering the instrument, turn off the power and wait at least for 30 seconds.
4.2.2. Load Cell Connection
To avoid damages, the load cell wiring should be made carefully before energizing the instrument. Load cell
connection details are shown below in figure 4.2. In 4-wire installations the sense and excitation pins with the
same polarity should be short circuited at the connector side. If you have a junction box in your system, use a 6
wire cable between FAD-40 and the junction box, and short circuit these pins at junction box for better
performance.
Figure 4.2 – Load cell connection
When you use a junction box and one or more load cells with 4 wires you have to short-circuit the excitation
(Exc) and sense wires (Sen) the same polarity as show on the figure 4.21.
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 10 of 56
4.2.3. Digital I/O Connection
FAD-40EN,
FAD-40PB,
FAD-40PN and
FAD-40CO
(Front view)
24V 0V
TX RX GND Shield
Please refer to the corresponding chapter:
FAD-40 (RS485)
see chapter 6.2
FAD-40MB (Modbus RTU)
see chapter 6.2
FAD-40PB (Profibus)
see chapter 7.2
FAD-40PN (Profinet)
see chapter 8.2
FAD-40EN (Ethernet and Modbus TCP)
see chapter 9.2
FAD-40CO (CANopen)
see chapter 10.2
FAD-40 instruments have the digital I/O connectors on the instrument’s front. I/O1 is always an output, I/O2 and
I/O3 can be configured for input or output. The I/O connection diagram is shown in Figure 4.3
Figure 4.3 – Digital I/O connection
4.2.4. Serial Interface RS232
You can use the serial interface RS232 with the xFace software for configuration and adjustment.
Figure 4.4 – The pin layout of RS232 connector
4.2.5. Communication Interface
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 11 of 56
4.3. Commissioning
PRECAUTION: Please read this manual carefully before energizing the instrument. Perform the commissioning
operation according the procedure given in this chapter. Only trained persons are allowed for cleaning,
commissioning, checking and servicing of the instrument. The interference of untrained person may cause
some unwanted damages or injuries.
Before energizing the instrument, please make the required mechanical and electrical installations. After power
on, you have to setup your FAD-40 instrument before you can start to use the bus interface.
Install the xFace software onto your PC as described in chapter 5 Setup. The xFace software is used for setup,
calibration and testing of FAD-40 instruments.
After you have successfully checked the performance of the instrument with xFace, you can begin to use the
instrument in your application.
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 12 of 56
5. SETUP
Typ FAD-30
Typ FAD-40: Select one of the FAD-40
models and press OK button.
Typ MCS-08
Figure 5.1 – xFace type selector window
Protocol: Select the setup port of the instrument. FAD-40 and FAD-40MB
instruments can be setup over RS-485. Other models can be setup over RS232C.
Com Port: Select the communication port of the PC
Address: Select the RS-485 address, if RS-485 is selected
Baud rate: Select the RS-485 baud rate, if RS-485 is selected
Setting: Select the RS-485 communication port setting, if RS-485 is selected
Figure 5.2 – xFace Connection Settings
PRECAUTION: Please read this manual carefully before energizing the instrument. Perform the commissioning
according the procedure given in chapter 4.3. Only trained person are allowed for commissioning, checking,
cleaning and servicing of the instrument. The interference of untrained person may cause some unwanted
damages or injures.
FAD-40 series instruments are setup and calibrated by the xFace software supplied with the instrument.
The instruments shall be setup in the sequence described below before you can use the bus interface.
Install the xFace software onto your PC
Connect your PC with the instrument over the serial interface (RS232 respective RS485)
Setup and calibrate the instrument
Check the performance of the instrument
5.1. Installation of the xFace Software
Please follow following steps to install the xFace software:
Close all applications on your PC
Insert the CD that contains the xFace software into the CD-ROM drive
Double click “Setup.exe” to start the installation. The setup Wizard is displayed.
Follow the menus in the setup wizard step by step.
After finishing the installation, the Setup Wizard will inform you about the success of the software
installation. Click the OK button.
After closing the Setup Wizard you can start to use the xFace software.
5.2. Connection to the xFace Software
The connection between FAD-40 and the xFace software is done via the RS485 serial port or the RS-232C
service port for all models. Alternatively Ethernet (for FAD-40EN only) can be used for this connection. You can
purchase a suitable PC connection cable as an accessory from Flintec (Refer to chapter 3.4.2).
After running the xFace software select the FAD-40 model you use (see figure 5.1.). Select the PC’s
communication port within the Connection settings menu in the tools tab (see figure 5.2) and click the connect
icon. After the communication between FAD-40 and your PC has successfully started the traffic light of the
connect icon turns from red to green.
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 13 of 56
5.3. Setup and Calibration
Visual Weight Display: Displays the weight, count or force value
Toolbox: Contains shortcuts of some special commands
A/D Converter Parameters: This block allows the user to setup the A/D Converter’s parameter
related to the operation mode. Refer to chapter 5.3.1
RS-232C: RS-232C serial port communication settings (not for FAD-40 and FAD-40MB)
RS-485: RS-485 port communication settings. (FAD-40 and FAD-40MB only)
Description: This block provides some clear-text explanations
Read from A/D Converter: Click this button to read the parameter settings from the instrument
Write to A/D Converter: Click this button to save the parameter settings to the instrument
Default: Click this button to load the factory default settings
Clear: Empties the parameter settings
Setting
0 1 2 3 4 5 6 7 8
9
Values/s
1600
800
400
200
100
100
100
50
50
50
Settling time [ms]
80
140
250
180
300
420
380
620
720
1800
WARNING: Read this manual carefully before you setup and calibrate the instrument. This will increase the
performance of your weighing system by applying proper setup and calibration.
Figure 5.3 – FAD-40 Setup Parameters
For entering the parameter values enter the setup tab, adjust the parameter settings and then write this data to
the A/D Converter. After changing parameter settings the instrument always requires a re-calibration.
5.3.1. Scale Parameters
In the setup menu the scale parameters of FAD-40 can be viewed, changed or saved to the instrument. These
parameter settings are:
Increased External Resolution
For service purposes this parameter enables the 10 times higher resolution of the weight value than the defined
interval in weighing and force mode. Increased resolution must be disabled for use in normal operation (Not
available in Count Mode). Default setting: ‘Disable’.
Digital Filter
Environmental noise like bounding forces, air flow, vibration, motor control instruments etc. may disturb the load
cell signal. The selection of proper filter settings determines how quickly the system will react to the load cell
signal.**Load cell signal digital filtering is done according to this parameter settings. The settings can be
changed from 0 (fastest settling) to 9 (slowest settling). Default setting: ‘7’.
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 14 of 56
Motion Detection
Disable
± 0.3e
± 0.5e (default setting)
± 1e
± 2e
Disable
± 60
± 100 (default setting)
± 200
± 400
Disable (default setting)
± 2%
± 10%
Disable
± 2% (default setting)
± 20%
± 40%
Disable (default setting)
± 0,5e
± 1e
± 3e
Disable:
Port will be disabled. Select “Disable” if this port is not used.
BSI
(default setting for FAD40)):
Communicates in BSI data format as a slave.
Refer to chapter 5.9 for details.
Modbus RTU
(default setting for FAD-40MB)):
Modbus RTU communication (for FAD-40MB only).
Refer to chapter 5.11 for details
This parameter defines the motion detection window which determines a stable weighing. If motion detection is
not required, this parameter can be disabled. The available motion detection window values are:
Weighing and Force Mode:
Count Mode:
Power On Zero
This parameter enables automatic zeroing after powering on the instrument. The automatic zeroing is only
done if the total zero drift from the calibrated zero signal is in the defined zeroing window. This zeroing window
will be defined in the percentage of the scale capacity.
To avoid unwanted zeroing at power on this parameter should be disabled or carefully set up in applications
like silo weighing, tank weighing and automatic weighing applications. If the weight is not within the power on
zeroing range, the instrument is powered on without zeroing (Not available in Count Mode).
Zeroing by Key (Zeroing Range)
Zeroing of the scale is performed if the difference between load cell signal value and unloaded load cell signal
value at the calibration is in the selected percentage of the scale capacity. Zeroing can be done by zeroing
command when the scale is stable. Zeroing can be done by zeroing command when the scale is stable.
The available zeroing ranges in the percentage of the scale capacity are:
Auto Zero Tracking
AZT automatically re-adjusts the scale to zero for compensating defined small deviations around the center of
zero. AZT only works within the defined zeroing range and stops working if this range is left.To avoid unwanted
zeroing this parameter should be disabled or carefully set up in applications like silo weighing, tank weighing
and automatic weighing applications (Not available in Count Mode). The available AZT window values are;
Tare
If this function is enabled, the weight is tared when the tare command is received. Additional conditions for
taring are positive gross weight and no motion. Multi-taring is possible with FAD-40 series (Not available in
Count Mode). Default setting: ‘Enable’.
Save Tare at Power Off
If this function is enabled, the tare value is stored at power off and the instrument starts up in Net mode at
power on (Not available in Count Mode). Default setting: ‘Enable’.
5.3.2. RS-485 Serial Interface Setup
Data Format
This parameter defines the serial communication port data format for RS-485. The available data formats are:
NOTE: xFace automatically connects to the instrument whatever the data format is. After xFace is
disconnected, the instrument returns to operate with the last saved data format.
WARNING: Disable this interface if it is not used to increase the interfacing performance.
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 15 of 56
Checksum
1200
2400
4800
9600 (default setting)
19200
38400
57600
Mode: Here the user selects the operation mode and the polarity.
Write to A/D Converter: Click this button to save the operation mode and polarity to A/D
Converter.
Calibration block: This block allows the user to calibrate with test weights.
eCal Calibration: This block allows the user to calibrate without test weights.
Calibration Coefficients: This function allows the user to restore a calibration if the calibration
coefficients have been noted before.
The checksum can be enabled or disabled within BSI data format. The checksum calculation can be found in
the related data format description. Default setting: ‘Enable’.
Carriage Return (CR)
Carriage Return can be enabled or disabled within Continuous output data format. Default setting: ‘Enable’.
Line Feed (LF)
Line Feed can be enabled or disabled within Continuous output data format. Default setting: ‘Enable’.
The RS-485 setup is done by the xFace software as described in this chapter. Select the Setup tab after xFace
has established a connection with your instrument. The Setup tab (see figure 5.3) includes all RS-485 settings.
Baud Rate
One of the following baudrate will be selected for the RS-485 communication port.
Data Length and Parity
The data length and parity can be selected as 8 None 1 (default), 7 Odd 1 or 7 Even 1.
Address of A/D Converter
The address range is 01 (default) to 31. If you enter 0, the instrument will operate without address data.
Modbus RTU Setup (for FAD-40MB only)
RS-485 Data Format: ‘Modbus RTU’,
RS-485 Data Length & Parity: ‘8 none 1’,
RS-485 Address: ‘01’ to ‘31’
5.3.3. Scale Build and Calibration
Scale build, operation mode selection and scale calibration is performed in the calibration tab of xFace as
shown in figure 5.4. Please follow the procedure in the sequence as described in the following chapters.
Figure 5.4 – Calibration
FAD-40 Technical Manual, Rev.2.0; July 2016 Page 16 of 56
Loading...
+ 39 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.