Druck PACE5000, PACE6000 Operating Manual

0 (0)

GE

Sensing & Inspection Technologies

Pressure Automated Calibration Equipment

User manual K0443

PACE5000

PACE6000

g

© The General Electric Company. All rights reserved

Introduction

This technical manual provides operating instructions for the PACE Pneumatic Pressure Controllers.

The features shown and described in this manual may not be available on some models.

Safety

The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. Do not use this equipment for any other purpose than that stated.

This publication contains operating and safety instructions that must be followed to ensure safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage.

Use suitably qualified * technicians and good engineering practice for all procedures in this publication.

Pressure

Do not apply pressures greater than the maximum working pressure to the equipment.

Toxic Materials

There are no known toxic materials used in construction of this equipment.

Maintenance

The equipment must be maintained using the procedures in this publication. Further manufacturer’s procedures should be carried out by authorized service agents or the manufacturer’s service departments.

Technical Advice

For technical advice contact the manufacturer.

* A qualified technician must have the necessary technical knowledge, documentation, special test equipment and tools to carry out the required work on this equipment.

K0443 Issue No. 3

i

Abbreviations

The following abbreviations are used in this manual; the abbreviations are the same in the singular and plural.

abs

Absolute

a.c.

Alternating current

ALT

Altitude

BSP

British pipe thread

CAS

Calculated airspeed

CSK

Countersunk

d.c.

Direct current

DPI

Digital Pressure Instrument

e.g.

For example

etc.

And so on

Fig.

Figure

ft

Foot

g

Gauge

Hg

Mercury

HTS

High tensile steel

Hz

Hertz

IAS

Indicated airspeed

i.e.

That is

IEC

International Electrotechnical Commission

IEEE 488

Institute of Electrical and Electronic Engineers standard 488 data

in

Inch

kg

Kilogram

kts/kn

knot

LCD

Liquid crystal display

m

Metre

mA

Milliampere

max

Maximum

mbar

Millibar

min

Minute or minimum

mm

Millimetre

mV

Millivolts

MWP

Maximum working pressure

No.

Number

NPT

National Pipe Thread

PACE

Pressure Automated Calibration Equipment

Para.

Paragraph

PDCR

Pressure transducer

PED

Pressure Equipment Directive

psi

Pounds per square inch

PTX

Pressue transmitter

K0443 Issue No. 3

ii

ROC

Rate of Climb

RS232

Serial communications data standard

SCPI

Standard Commands for Programmable Instruments

UUT

Unit under test

V

Volts

VFC

Volts-free contact

+ve

Positive

-ve

Negative

°C

Degrees Celsius

Associated publications

K0447 PACE 5000/6000 User Guide and Safety Instructions

K0450 PACE Series Calibration Manual

K0476 Pressure Control Module User Guide and Safety Instructions

K0472 Remote Communications Manual

K0469 Heritage Communications Manual - Instrument Emulation

Symbols

The equipment contains the following symbols to identify hazards.

This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.

This symbol, on the instrument, indicates that the user should refer to the user manual.

This symbol, on the instrument, indicates do not throw-away in domestic bin, hazardous material, dispose correctly in accordance with local regulations.

 

 

 

 

K0443 Issue No. 3

iii

Pressure units and conversion factors

 

Pressure units

Factor (hPa)

Pressure units

Factor (hPa)

 

 

 

 

 

 

 

 

mbar

1.0

cmH2O @ 20°C

0.978903642

 

 

 

 

 

 

 

 

bar

1000.0

mH2O @ 20°C

97.8903642

 

 

 

 

 

 

 

 

Pa (N/m2)

0.01

kg/m2

0.0980665

 

 

hPa

1.0

kg/cm2

980.665

 

 

kPa

10.0

torr

1.333223684

 

 

 

 

 

 

 

 

MPa

10000.0

atm

1013.25

 

 

 

 

 

 

 

 

mmHg @ 0°C

1.333223874

psi

68.94757293

 

 

 

 

 

 

 

 

cmHg @ 0°C

13.33223874

lb/ft2

0.4788025898

 

 

mHg @ 0°C

1333.223874

inH2O @ 4°C

2.4908891

 

 

 

 

 

 

 

 

inHg @ 0°C

33.86388640341

inH2O @ 20°C

2.486413

 

 

 

 

 

 

 

 

mmH2O @ 4°C

0.0980665

inH20 @ 60°F

2.487641558

 

 

 

 

 

 

 

 

cmH2O @ 4°C

0.980665

ftH2O @ 4°C

29.8906692

 

 

mH2O @ 4°C

98.0665

ftH2O @ 20°C

29.836983

 

 

mmH2O @ 20°C

0.097890364

ftH20 @ 60°F

29.8516987

 

 

 

 

 

 

 

 

 

 

 

 

 

Unit Conversion

To convert FROM pressure VALUE 1 in pressure UNITS 1

TO pressure VALUE 2 in pressure UNITS 2, calculate as follows: VALUE 2 = VALUE 1 x FACTOR 1

FACTOR 2

Note:

The PACE instrument contains selectable pressure units and user defined units. Use the conversion factors to calculate a user defined unit from the table above. Refer to the data sheets SDS0001 or SDS0008 for the list of selectable pressure units.

K0443 Issue No. 3

iv

 

CONTENTS

 

Section

Title

Page

1

Description......................................................................................................................

1-1

1.1

Introduction ....................................................................................................................

1-1

2

Installation.......................................................................................................................

2-1

2.1

Packaging........................................................................................................................

2-1

2.2

Packaging for Storage and Transportation.....................................................

2-1

2.3

Preparation for Use.....................................................................................................

2-1

2.4

Pneumatic connections ............................................................................................

2-2

2.5

Rack-mount option .....................................................................................................

2-7

2.6

Electrical connections................................................................................................

2-8

3

OPERATION......................................................................................................................

3-1

3.1

Preparation .....................................................................................................................

3-1

3.2

Power-up Sequence ...................................................................................................

3-2

3.3

Measure Mode...............................................................................................................

3-3

3.4

Control Mode..................................................................................................................

3-5

3.5

Operation and Example Procedures...................................................................

3-8

3.6

Global Set-up .................................................................................................................

3-12

3.7

Barometric Reference Option.................................................................................

3-13

3.8

Supervisor Set-up ........................................................................................................

3-14

3.9

Instrument Status ........................................................................................................

3-15

K0443 Issue No. 3

v

4

MAINTENANCE...............................................................................................................

4-1

4.1

Introduction ....................................................................................................................

4-1

4.2

Visual inspection ..........................................................................................................

4-1

4.3

Cleaning ...........................................................................................................................

4-1

4.4

Test .....................................................................................................................................

4-1

4.5

Calibration.......................................................................................................................

4-1

4.6

Replacement parts......................................................................................................

4-2

4.7

Fuse replacement........................................................................................................

4-2

4.8

Filter replacement .......................................................................................................

4-4

4.9

Pressure module replacement ..............................................................................

4-5

5

TESTING AND FAULT FINDING.................................................................................

5-1

5.1

Introduction ....................................................................................................................

5-1

5.2

Standard Serviceability Test ...................................................................................

5-1

5.3

Fault Finding...................................................................................................................

5-2

5.4

Approved Service Agents .........................................................................................

5-3

6

REFERENCE AND SPECIFICATION ..........................................................................

6-1

6.1

Installation notes..........................................................................................................

6-1

6.2

Operational Requirements ......................................................................................

6-4

6.3

Icons.................................................................................................................................. .

6-6

6.4

Measure Set-up.............................................................................................................

6-9

6.5

Control Set-up................................................................................................................

6-10

6.6

Global Set-up .................................................................................................................

6-11

6.7

Supervisor Set-up ........................................................................................................

6-12

6.8

Options..............................................................................................................................

6-15

K0443 Issue No. 3

vi

6.8

Calibration.......................................................................................................................

6-28

6.9

Communications - Instrument Emulation.......................................................

6-29

6.10

Specification...................................................................................................................

6-29

6.11

Return Goods/Material Procedure.......................................................................

6-29

6.12

Packaging Procedure.................................................................................................

6-30

K0443 Issue No. 3

vii

intentionally left blank

K0443 Issue No. 3

viii

PACE Pressure Controller User Manual

1 Description

1.1Introduction

The PACE5000 single-channel and PACE6000 single/dual-channel, Pressure Automated Calibration Equipment measures and controls pneumatic pressures and displays, on a touch-screen, the pressure measurement and controller status. The touch-screen enables selections and settings in both measure and control modes. The instrument can be operated remotely through communication interfaces.

Figure 1-1 PACE5000 General view

Figure 1-2 PACE6000 General view

The rear of the instrument houses all the electrical and pneumatic output and input connections. The electrical connections provide an a.c. power supply, serial and parallel communication interfaces, d.c. output and logic input and output. The system pneumatic controller module contains a positive and negative pressure supply port, an output port, vent port and reference port.

K0443 Issue No. 3

1 - 1

1 Description

The instrument can be mounted in a standard 19 inch rack system (rack-mount option).

Figure 1-3 PACE5000 Rear view

Figure 1-4 PACE6000 Rear view

Options available for the PACE5000 and PACE6000 are detailed in datasheet SDS0001 (PACE5000) and SDS00008 (PACE6000).

Further information and notes on applications are available in section 6 of this manual and on the GE Sensing & Inspection web site at www.gesensinginspection.com.

K0443 Issue No. 3

1 - 2

PACE Pressure Controller User Manual

2 Installation

2.1Packaging

On receipt of the PACE5000 or PACE6000 check the contents of the packaging against the following list:

Packaging List

i)PACE5000 or PACE6000 Pressure Controller.

ii)Cable, power supply.

iii)User guide and CD (UD-0001) containing the full documentation suite.

iv)Pneumatic Control Module blanking plate (keep this plate for future use).

CAUTIONS:

1.AFTER REMOVING A CONTROL MODULE, USE A BLANKING PLATE TO KEEP THE FLOW OF COOLING AIR.

2.AFTER UNPACKING AN INSTRUMENT THAT HAS BEEN IN COLD CONDITIONS ALLOW TIME TO STABILISE AND ANY CONDENSATION TO EVAPORATE.

2.2Packaging for Storage or Transportation

To store the instrument or to return the instrument* for calibration or repair carry out the following procedures:

1.Pack the instrument as detailed in section 6.8, Reference and Specification.

2.Return the instrument for calibration or repair complete the return goods procedure detailed in section 6.8, Reference and Specification.

*This procedure also applies to the pressure control module as a separate item.

2.3Preparation for Use

The instrument can be used as a:

Free-standing instrument positioned on a horizontal surface.

Rack-mounted in a standard 19 inch rack using the rack-mount accessory option kit (refer to section 2.5).

For free-standing instruments, use the two front feet on the base to elevate the instrument to provide a better viewing angle.

Note: The cooling air outlet on the underside of the instrument must not be obstructed. Allow a free flow of air around the instrument, especially at high ambient temperatures.

K0443 Issue No. 3

2 - 1

2 Installation

2.4 Pneumatic connections WARNINGS:

TURN OFF THE SOURCE PRESSURE AND CAREFULLY VENT THE PRESSURE LINES BEFORE DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.

ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.

BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE. REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT.

Connection

 

adaptor

Input

supply +

G 1/8

 

supply -

G 1/8

Output

 

G 1/8

 

Vent

G 1/8

 

Reference

G 1/8

Refer to the data sheets SDS0001 or SDS0008 for a complete range of adaptors.

Pressure supply (Figure 2-1)

1.The pressure supply must be clean, dry gas, nitrogen or air and at the correct pressure, refer to the specification (Section 6).

2.Ensure the user systems can be isolated and vented.

3.Connect pressure and vacuum supplies to the SUPPLY + and SUPPLY - connection ports.

4.Connect the Unit Under Test (UUT) to the required outlet connection port.

Example instrument

1 BSP connector 2 Bonded seal

Note: For instruments with NPT connections, use adequate pressure sealing.

Figure 2-1, Pneumatic Connections

K0443 Issue No. 3

2 - 2

PACE Pressure Controller User Manual

Installation

The instrument requires a positive pressure supply, instruments operating in an absolute range or negative pressure range require a vacuum supply. A vacuum supply should be used for a fast response for instruments operating near atmospheric pressure. For dual channel operation two independent pressure and vacuum supplies can be used.

Important notes

When using two pressure modules, make sure that the modules are compatible, refer to Section 4.9. All connections must comply with the Pressure Equipment Directive (PED).

When connecting the output ports of two pressure modules together make sure both are either:

below 70 bar. OR

between 100 to 210 bar.

To comply with the Pressure Equipment Directive (PED) do not mix these categories.

Supply equipment

Pneumatic supplies should have isolation valves and, where necessary, conditioning equipment. The positive pressure supply should be regulated to between 110% of the full-scale pressure range and MWP stated on the control module.

To protect the control module, for ranges above 100 bar, from over-pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply pressure to below the MWP.

On instruments without a negative supply, the positive pressure discharges from the system to atmosphere through the negative supply port. Fit the diffuser to the negative port to diffuse airflow.

During system pressure vent operations, the pressure discharges from the system to atmosphere through the vent port. Fit a diffuser to the vent port to diffuse airflow.

K0443 Issue No. 3

2 - 3

2 Installation

Pneumatic Connection Examples (Figures 2-2 and 2-4)

These examples show a single channel connection detail using supply equipment described above.

SUPPLY

 

 

OUTLET VENT REF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

a

Figure 2-2, Pneumatic Connections without vacuum supply

Key to Figure 2-2

1

Pressure source

2

Conditioner

3

Filter

 

 

4

Regulate to between 110% full-scale and MWP

 

 

 

 

5

Diffuser *

6

Unit under test

7

Optional reservoir †

8

Protection device

9

Optional differential connection

a

atmosphere

Notes: Refer to Section 6 Reference and Specification for details of other system components.

*High pressure gas exhaust - depending on pressure range.

Optimum controller transient response and minimum time to set-point may be degraded if either the pneumatic supply or vacuum system has restricted flow. Installing a reservoir volume, which has larger capacity than the load volume, located in close proximity to the controller supply ports can improve the controller response.

To protect the control module, for ranges above 100 bar, from over-pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply pressure to below the MWP.

Optional differential connection kit.

K0443 Issue No. 3

2 - 4

PACE Pressure Controller User Manual

SUPPLY

OUTLET VENT REF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

12

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

11

 

 

 

 

 

 

 

 

 

13

a

 

 

 

 

 

 

 

 

 

*

 

Figure 2-3, Pneumatic Connections with vacuum supply

 

Key to Figure 2-3

1

Pressure source

2

Conditioner

3

Filter

4

Regulate to between 110% full-scale and MWP

 

 

5

Diffuser*

6

Unit under test

7

Optional reservoir †

8

Protection device

9

Optional differential connection

10

Oil mist trap

11

Vacuum source

12

Normally open electrical release valve

13

Check valve ‡

a

atmosphere

 

 

Notes:Refer to Section 6 Reference and Specification for details of other system components.

*High pressure gas exhaust - depending on pressure range.

Optimum controller transient response and minimum time to set-point may be degraded if either the pneumatic supply or vacuum system has restricted flow. Installing a reservoir volume, which has larger capacity than the load volume, located in close proximity to the controller supply ports can improve the controller response.

Optional vacuum system kit.

To protect the control module, for ranges above 100 bar, from over-pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply pressure to below the MWP.

Optional differential connection kit.

K0443 Issue No. 3

2 - 5

2 Installation

SUPPLY

OUTLET VENT REF

 

 

 

 

 

 

 

 

11

 

 

 

13

a

 

*

12

Figure 2-4, Pneumatic Connections with negative gauge pressure generator Key to Figure 2-4

1

Pressure source

2

Conditioner

3

Filter

 

 

4

Regulate to between 110% full-scale and MWP

 

 

 

 

5

Diffuser *

6

Unit under test

7

Optional reservoir †

8

Protection device

9

Optional differential connection

a

atmosphere

10

Vacuum generator

11

Source pressure (regulated compressed air supply)

12

Exhaust to atmosphere

13

Check valve

 

 

 

 

Notes:Refer to Section 6 Reference and Specification for details of other system components.

*High pressure gas exhaust - depending on pressure range.

Optimum controller transient response and minimum time to set-point may be degraded if either the pneumatic supply or vacuum system has restricted flow. Installing a reservoir volume, which has larger capacity than the load volume, located in close proximity to the controller supply ports can improve the controller response.

Optional negative gauge generator kit.

To protect the control module, for ranges above 100 bar, from over-pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply pressure to below the MWP.

Optional differential connection kit.

K0443 Issue No. 3

2 - 6

PACE Pressure Controller User Manual

2.5 Rack-mount option (Figure 2-5) General

There must be enough space at the rear of the instrument for all the cables and pipes. The length of the cables and pipes must allow for the removal and fitment of the instrument. The cooling air of the instrument must not be obstructed. Allow a free flow of air through the equipment rack and around the instrument, especially at high ambient temperatures.

Example instrument

Figure 2-5 Rack-mounting

Procedure

• Locate and secure the back brackets

to the rear of the equipment rack.

• Unscrew and remove the four countersunk screws from each of the instrument side panels.

• Locate the two brackets on each side of the instrument and secure with the four countersunk screws.

• Support the instrument and connect the cables and pipes.

K0443 Issue No. 3

2 - 7

2 Installation

Refer to the electrical connections below before finally fitting the instrument into the equipment rack.

Temporarily locate and screw in the two spigots* to each side of the equipment rack.

Locate and slide the instrument into the rack, locating the instrument on the spigots*.

Secure the instrument in the equipment rack with two of the screws and washers (supplied).

Remove the two spigots* and replace with the remaining two screws and washers (supplied).

2.6 Electrical connections WARNINGS

1.THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY PROTECTIVE SAFETY GROUND.

2.ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE REAR PANEL.

General

The instrument must be connected to the correct electrical power supply as stated on a label next to the power connector. Also see section 6 Reference and Specification.

Make sure that the power supply is off before connecting the power cable.

Requirements for rack-mounted instruments

Install an isolator in the power supply circuit. The power supply connector and switch on the rear panel of the instrument will not be accessible.

Set the power supply isolator to OFF. Connect the power supply and set the power supply switch to ON before sliding the instrument into the rack.

Set the power supply isolator to ON .

Check that the front panel display shows the power-up sequence.

Connecting (Figure 2-6)

To connect the power supply to the instrument proceed as follows:

Insert the IEC power connector (1) into the power supply connector at the rear of the instrument.

Set the ON/OFF switch to ON.

Check that the front panel display shows the power-up sequence.

K0443 Issue No. 3

2 - 8

PACE Pressure Controller User Manual

 

 

 

 

 

 

 

1

IEC power connector 2

Fuse carrier

3

Fuse

Figure 2-6 Electrical Connections

Pressure Control Module Input and Output Connectors

24V DC Output @ 100mA maximum

Using a 4-way connector:

pin “+”

=

 

pin “-”

=

This facility can energise external equipment.

+24 Vdc

0 Vdc

An integral self-resetting fuse protects this output.

Logic (switch) Input

 

Using a 4-way connector:

Input

 

Output

This facility can be used to trigger the instrument from a pressure switch contact during the Pressure Switch Task (see Section 3.4). Connections are not polarised and can be connected either way. Integral opto-isolators protect this input circuit.

K0443 Issue No. 3

2 - 9

Druck PACE5000, PACE6000 Operating Manual

2 Installation

Communication Connections

Connect the appropriate connectors into the rear panel communications ports and, if appropriate, secure with the captive screws.

Note: The RS232 and IEEE 488 interfaces are both enabled at power-up. Set the required parameters in Supervisor Setup/communications menu, see Section 3.8.

 

 

 

Figure 2-7, Communication Connectors

1

RS232

2

Canbus (option) 3

IEEE488

4

USB B

5

USB A

6

Ethernet (option)

 

 

 

RS232 Interface

When using the RS232 interface, a cable must be connected directly from the instrument to a suitable port on the computer in a ‘point to point’ link. The pin connections for the 9-pin D-type, RS232 connector and the relationship between the instrument and the RS232 control signals, together with device interconnection interface is shown in Table 2-1. The instrument is configured as Data Circuit Terminating Equipment (DCE).

K0443 Issue No. 3

2 - 10

PACE Pressure Controller User Manual

 

Instrument

 

 

Control Line

 

Computer/Printer

 

 

 

 

 

 

 

 

 

 

 

Instrument

Connector

 

Signal

RS232

 

Connector Type

 

 

 

 

 

 

Function

9-way

 

Direction

Terminology

 

 

 

 

 

D-type

 

 

 

 

9-way

25-way

 

 

Pin No.

 

 

 

 

 

 

 

 

 

 

D-type

D-type

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pin No.

Pin No.

 

 

 

 

 

 

 

 

 

 

 

RxD (I/P)

 

3

 

 

TxD

 

3

2

 

 

 

 

 

 

 

 

 

 

 

TxD (O/P)

 

2

 

 

RxD

 

2

3

 

 

 

 

 

 

 

 

 

 

 

GND

 

5

 

 

GND

 

5

7

 

 

 

 

 

 

 

 

 

 

 

CTS (I/P)

 

7

 

 

RTS

 

7

4

 

 

 

 

 

 

 

 

 

 

 

RTS (O/P)

 

8

 

 

CTS

 

8

5

 

 

 

 

 

 

 

 

 

 

 

Pulled

 

 

 

 

RLSD

 

 

 

 

high

 

1

 

 

 

1

8

 

 

 

 

(DCD)

 

 

internally

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Not

 

4

 

 

DTR

 

4

20

 

connected

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pulled

 

 

 

 

DSR

 

 

 

 

high

 

6

 

 

 

6

6

 

 

 

 

DCE Ready

 

 

internally

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Equipment

Connector

 

 

Cable

 

-

1

 

chassis

shell

 

 

Screen

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 2-1, RS232 Connections

 

 

Handshaking connections

 

 

 

 

 

 

 

Software handshaking use:

TXD, RXD and GND.

 

 

Hardware handshaking use:

TXD, RXD, GND, CTS, RTS and DTR.

 

 

K0443 Issue No. 3

2 - 11

2 Installation

IEEE 488 Interface

This interface complies with IEEE 488 standard. The IEEE 488 parallel interface connects a computer/controller to one or more PACE instrument and possibly other instruments. Up to 30 instruments can be connected through a high-speed data bus to the computer/controller.

Note: The length of each IEEE 488 cable must be less than 3 metres to comply with the EMC requirements, see Section 6 Reference and Specification.

Single Unit Installation (Figure 2-8)

Connect an IEEE 488 connector/cable assembly to the rear panel IEEE 488 connector of the instrument.

Connect the other end of the connector/cable assembly to the IEEE 488 connector on the controller/computer.

Change the IEEE 488 communication parameters as described in the Supervisor set-up (refer to Section 6.7).

Multiple Unit Installation (Figure 2-8)

To install multiple units use stacking plugs to link from first instrument to the second instrument. Proceed as follows.

Connect a pair of IEEE 488 stacking connectors to the rear panel IEEE 488 connector of the instrument

1Connector to rear panel of first instrument.

2Connector from controller/computer.

3Connector to rear panel of second instrument.

Connect the other end of one of the connectors to the IEEE 488 connector on the controller/ computer and the other connector into the next instrument

1

2

3

Repeat this procedure for all the instruments in the system.

Use the Supervisor set-up (communications) menu on each instrument to set-up the required communication parameters (refer to Section 3.8).

K0443 Issue No. 3

2 - 12

PACE Pressure Controller User Manual

Figure 2-8 - IEEE 488 Connection

K0443 Issue No. 3

2 - 13

2 Installation

intentionally left blank

K0443 Issue No. 3

2 - 14

Loading...
+ 60 hidden pages