İNFUSOMAT FMS ERROR CODE LİST
100defective RAM memory
101UMOT cannot be switched on
102UMOT still switched on despite overvoltage
103UMOT still switched on despite MOTEIN=0
104UMOT still switched on despite undervoltage
105ON/OFF key pressed longer than 14 sec
106defective air sensor (calibration value?)
107defective program memory
108defective program flow
109different number of pump head cycles
110different keyboard gaps between 80c838 and 68hc11
111different program versions between 80c535 and 68hc11
112defective program flow
113testbit!=0 out of switch-on test
116defective program memory - text
117defective program memory - text does not match with program
118reset during active operation
119defective ROM
FF01 |
dummy for test |
FF02 |
battery not present / missing battery current |
FF03 |
defective RAM memory |
FF04 |
defective program memory |
FF05 |
defective program memory |
FF06 |
calibration data error from EEPROM |
FF07 |
pump head cycle not plausible |
FF08 |
failure / inaccuracy of system clock |
FF09 |
failure 100msec system clock |
FF10 |
reset during active operation |
FF12 |
no dynamic pressure sensor signal (EDB) |
FF16 |
defective membrane keyboard (from IFMC02003 on) |