Xylem SuSix User Manual

Turbidity and Suspended Solids Transmitter
Turbidity and Suspended
Solids Transmitter
Installation and User Manual
1EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
CE CERTIFICATE OF CONFORMITY
This product complies with the requirements concerning electromagnetic compatibility (EMC) stipulated in Council directive no. 89/336/EEC of 3rd may 1989, alteret at directive no. 92/31/EEC, on the approximation of the laws of the member states relating to electromag­netic compatibility.
CE approvals
We declare that the SuSix® transmitter and ultrasonic sensors comply with the values stipulated in EN-61010-1 and EN 61326-1.
Ex approval
The SuSix® sensor is approved for mounting in explosive amospheres: Ex II G3.
2EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Table of Contents
1. Introduction 7
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Operating Principles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2. Safety, Repair and Product Identification 11
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Physical Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Product Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3. Sensor Installation 12
Motion in the liquid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Mounting / position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Mechanical installation of a submerged sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Position of submerged sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Mechanial installation of sensor in a tube (in-line) . . . . . . . . . . . . . . . . . . . . . . . . . .16
Mounting guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Installation / Removal using Insertion Fitting 155210 . . . . . . . . . . . . . . . . . . . . . . .20
Required Tools and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Preparing the Insertion Fitting before welding to tube . . . . . . . . . . . . . . . . . . . . . . . .20
Installation of the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Removal of the Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4. Sensor Calibration 23
Turbidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Suspended Solids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
Delete Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
5. Converter 29
Electrical Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Changing the Power Supply Voltage 230/115 V AC . . . . . . . . . . . . . . . . . . . . . . . . .31
Analogue Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Digital Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Converter Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Sensor and Mounting Kit / Connection Board Terminals . . . . . . . . . . . . . . . . . . . . . .36
6. System Configuration Examples 37
Magflux® with Blind Lid and SuSix® with Display Unit . . . . . . . . . . . . . . . . . . . . . . . 38
Wiring Schematic #1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
SuSix® with Blind Lid, Magflux® with Blind Lid and a Remote Display . . . . . . . . . .40
Wiring Schematic #2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
EN 5.10 SuSix Manual 1406
3
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
7. Startup 43
Initial Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Initial Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
Language Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Display Read-out, one connected unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Display Read-out, several connected units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Display Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8. SuSix® Menus 47
Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Specify Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Data Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
Graph Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Set Sensor Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Converter Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Sensor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
mA Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Solid curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
High alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Low alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Sensor error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Service Menu - Digest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Service Menu - Detailed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
Display Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Set clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
Modbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Factory settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Display SW version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
9. Mechanical Dimensions 85
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Converter and Display Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Immersion Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
Insertion Fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
10. Maintenance 89
Cleaning the Sensor - Once a Month . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Cleaning the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
4EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Functional Tests - Once a Year . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
Replacing Wipers and Gaskets - Every Two Years . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Maintenance interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Required materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
Remove the used wiper and gasket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
Install the new gasket and wiper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Appendix A. Pop-up and Error Messages 95
Appendix B. MJK-Field Link Software 99
Important Notes! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
A. Connect a PC to the SuSix® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
B. Save Log Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
C. Upgrade Display Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .101
D. Upgrade Converter Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103
E. Install/Add Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Appendix C. Front Panel Cut-out Drawing 105
Appendix D. Log Files 107
Appendix E. Digital Input/Output Connections 111
Main Menu Overview 112
Converter Setup Menu Overview 114
Service Menu Overview 122
Display Configuration Menu Overview 126
Index Index-1
EN 5.10 SuSix Manual 1406
5
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
This page intentionally left blank.
6EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
1. Introduction
Thank you for choosing the MJK SuSix® Turbidity and Suspended Solids Transmitter. We have done our utmost to design and manufacture a quality transmitter that satisfies your requirements.
The SuSix® Transmitter, consisting of a sensor, a converter and a display, is easy to install, calibrate and put into service. However, read this manual first to get familiar with the SuSix® Transmitter and all its features.
The equipment must be handled and operated as instructed by the manufacturer, MJK Automation to ensure stable operation and accurate measurements.
The SuSix® sensor is equipped with two light-emitting diodes (LEDs, transmitters) and four emitter diodes (receivers). The readout is based on several measurements utilizing all the diodes to compensate for errors like dirt adhesion, worn diodes or scratches on the lenses.
SuSix® is available in two versions – a sensor with or without a dirt wiper and a screw thread for submerged installation in tanks or a fitting for instal­lation in tubes. The principles for use are identical for both versions.
You can always contact your local representative or the MJK hot lines for advice and guidance:
Europe Tel.: +45 45 56 06 56 E-mail: mjk@mjk.com
Denmark Tel.: +45 45 56 06 56 E-mail: mjk@mjk.dk
Visit our web sites at www.mjk.com to learn more about MJK Automation, our other products and the people behind them.
SuSix® is a registered trademark of MJK Automation A/S, Denmark.
MJK Automation is a Xylem Brand
EN 5.10 SuSix Manual 1406
7
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
About this Manual
This manual is divided into a table of contents, ten chapters, five appendi­ces, four fold-out pages and one index.
1. Introduction
Contains a presentation of the MJK SuSix® Turbidity and Suspended
Solids Transmitter, the structure of this manual and the operating prin­ciples.
2. Safety, Repair and Product Identification
Provide answers to issues regarding safety, mounting, repair,
restrictions and product identification.
3. Sensor Installation
Contains installation guidelines such as correct and wrong positioning
of the sensor in the liquid, min. and max. distances to walls and other solid faces, and installation in tubes.
4. Sensor Calibration
Turbidity measurements require no calibration, but detailed calibration
procedures and flow charts for suspended solids calibration are avail­able.
5. Converter
Describes the physical specifications and installation rules for the con-
verter such as power supply, in- and outputs, sensor/converter/ display configurations, etc.
6. System Configuration Examples
Illustrate how SuSix® Turbidity and Suspended Solids Transmitters can
be interconnected with for example MJK MagFlux® flow meters in real­life scenarios.
7. Startup
Describes important checks and basic settings to get started including
the display and keyboard user interface.
8. SuSix® Menus
Contains a comprehensive description of the SuSix® menus, options
and utilities.
9. Mechanical Dimensions
Lists the SuSix® Turbidity and Suspended Solids Transmitter dimen-
sions and specifications.
8EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
10. Maintenance
Procedures for cleaning the sensor, the converter, replacing wipers and
functional test are described.
* * * * *
Appendix A. Pop-up Events / Error Messages
Lists possible pop-up messages, explains their meaning and offer
solutions to error conditions.
Appendix B. MJK-Field Link Software Upgrade
Describes in detail how to utilise the unique and intuitive MJK-Field Link
software program to upload new display and converter firmware versions.
Appendix C. Front Panel Cut-out Drawing
A 1:1 scale drawing of the front panel outline and cut-out area for
installation and mounting purposes.
Appendix D. Log Files
An example log file illustrates the format and explains the entries.
Appendix E. Digital Input/Output Connections
Illustrates how the digital in- and outputs can be interconnected.
* * * * *
Main Menu Overview
A continuous presentation of the main menu structure
Converter Setup Menu Overview
A continuous presentation of the converter setup menu structure
Service Menu Overview
A continuous presentation of the service menu structure
Display Configuration Menu Overview
A continuous presentation of the display configuration menu structure
EN 5.10 SuSix Manual 1406
9
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Operating Principles
The SuSix® transmitter is an instrument for measuring turbidity and sus­pended solids. The transmitter consists of a sensor, a converter and a display unit.
Operating principles
The SuSix® sensor uses a patented 6-channel, multi-beam, pulsed infra­red light system providing a single sensor for both turbidity and suspended solids measurement. The turbidity measurement complies with ISO stand­ard 7027.
The advanced optical system is combined with a patented progressive al­gorithm using neural logic to generate a reliable high quality measurement. All of the optical systems within the transmitter is fitted with automatic self-diagnostics that ensure accurate measurements. The optical algorithm compensates for errors due to fouling or aging of the optical array, and an additional feature is a system that compensates for gas bubbles in the sample media.
Sensor with wiper Effect of wiper
Physical fouling of the sensor head can be eliminated by an optional sen­sor wiper system (see pictures above).
10EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
2. Safety, Repair and Product Identification
Safety instructions
• Read this manual carefully.
• Pay attention to the environment on the installation site.
• Wear necessary protective equipment and follow all current safety regulations.
• SuSix® can invoke a start signal for dangerous machinery. Always ensure that connected machinery and other equipment are effectively put out of service (that is to remove the main fuses and lock main and security switches in off-position) before commencing configuration, fault finding, servicing, maintenance work, etc.
WARNING: There is a risk of lethal, electrical shock from the "Mains supply" terminals "N" and "L". Be careful not to touch these terminals while the SuSix® is being serviced.
• WARNING: Incorrect installation or use can lead to body injury and/or material damage!
Physical Mounting
The SuSix® converter/display unit must not be mounted in explosion hazardous areas!
SuSix® may only be installed and commissioned by authorized MJK per­sonnel, and all national and local regulations must be complied with.
Repair
Repair must only be made by authorized MJK personnel or by a service representative approved by MJK.
Product Identification
A delivery will usually consist of a SuSix® sensor and a SuSix® converter with or without a display unit.
Check that the item(s) delivered corresponds to the ordered item(s). The part number is printed on an identification label sticked onto the converter shipping box and on a label on the converter itself.
EN 5.10 SuSix Manual 1406
11
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
3. Sensor Installation
Important! To ensure safe and optimal functionality of the SuSix® transmitter, the instructions below must be followed closely.
• The sensor must be mounted in a location where it is possible to per-
form a representative suspended solids measurement.
• If the suspended solids transmitter is mounted in open air, protect it
against the influence of the elements using for example a rain roof.
• Fittings and mountings must be capable of carrying the transmitter’s
weight (approx. 5 kg), and vibrations from for example pumps and other equipment must be avoided.
• The system should be installed to facilitate access to the suspended
solids transmitter operator panel and sensor without risking personal injury during normal operation, maintenance, cleaning of sensor, etc.
• The suspended solids transmitter must be connected to a mains outlet
that complies with national and local regulations.
• All cables must be installed, connected and mounted to minimize me-
chanical impact and electrical influence from other installations.
• The sensor cable’s maximum tensile load is adjusted to the sensor
weight.
• The sensor is a mechanical instrument. Be careful not to damage the
sensor windows. The accuracy will deteriorate, if the optics is subject to heavy pollution, or if a steel brush is used for cleaning.
• All fuses are located on the suspended solid transmitter’s motherboard.
WARNING!
The suspended solids transmitter must not be powered until the transmitter is correctly mounted and connected!
WARNING!
Connections to the suspended solids transmitter must not be disconnected while the transmitter is powered.
12EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Motion in the liquid
The liquid has to remain in motion and mixed. It is possible to achieve this by constantly circulating the liquid, or by installing the sensor at a place with a constant flow.
Faulty measurement due to stagnant liquid
As soon as the liquid becomes stagnant, and the suspended solids start for sedimentation, the sensor does not recognize the suspended solids anymore. For calibration in a sample bucket it is necessary to stir gently the liquid, or to constantly move the sensor in order to achieve reliable measuring results. Observe that you keep the minimum distances from the bucket walls.
EN 5.10 SuSix Manual 1406
13
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Mounting / position
The flow direction should be parallel to the sensor surface. This prevents turbulences at the sensor head.
Correct installation
In case of low suspended solids concentration, whirls and turbulences can falsify the measuring result.
Incorrect installation!
14EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Mechanical installation of a submerged sensor
Preparation
Select an easy-to-get-to position for a measuring point which can produce a valid and representative measuring value.
Position of submerged sensor
The following three steps are of vital importance for getting optimal meas­uring values.
1. The sensor (the optics) must be submerged at least 5 mm in the liquid.
2. The sensor to wall distance must be at least 250 mm.
EN 5.10 SuSix Manual 1406
15
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
3. The distance from the sensor head (the optics) to solid faces (for exam-
ple bottom and walls) must be at least 30 - 80 mm (depending on the liquid’s concentration).
Mechanial installation of sensor in a tube (in-line)
Mounting guidelines
1. The optical windows row have to be arranged lengthwise to the me-
dium flow direction (1).
2. Keep the minimal insertion depth >5 mm (2).
3. Control the minimum distance to the opposite wall. Must be >30 …80
mm depending on the concentration (3).
16EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
4. Make sure that the sensor is mounted within ± 45 degrees from the horizontal plane (see below).
- 45 degrees
0
+ 45 degrees
EN 5.10 SuSix Manual 1406
17
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
5. An in-line welding stud must be welded to the tube at an angle of 90
degrees in all directions (see below).
90 degrees
18EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
6. This drawing below illustrates a SuSix® sensor installed in an in-line clamp fitting with parts and dimensions.
MJK, 206310-12 SuSix® sensor
GPA, ZFP 040-R040
PP- d40 x 1½” inner RG
1½” welding stud, 1½” outer thread RG
122 mm
EN 5.10 SuSix Manual 1406
5 mm
19
1½”
nippel
Min. Ø100 mm
90 degrees
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Installation / Removal using Insertion Fitting 155210
Warnings : Max. operating pressure is 6 bar (720 N). Max. residual pressure when changing sensor is 1 bar (120 N). The sensor guide (2) is not a safety component The inser tion fitting is not fully assembled from the factory Be careful when dealing with corroding and toxic substances Be careful when dealing with lines under pressure
Required Tools and Materials
• 5 mm hex key
• LOCTITE® 577 sealant
4
2
3
T
6
1
5
A B
7
Measure position Service position
Preparing the Insertion Fitting before welding to tube
The nuts on the ball valve housing (see A and B arrows) have not been tightened and secured with a sealant from the factory. This allows the customer to rotate, position and secure the ball valve as required on the installation site.
1. Loosen/remove the A and B nuts, and rotate the ball valve to its final
position.
20EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
2. Apply Loctite 577 around the male threads (skipping the leader), tighten the A and B nuts, and allow the sealant to cure for about 72 hours (3 days) to achieve full strenght.
3. Weld the Insertion Fitting to the tube (see also pages 17 - 19).
Installation of the Sensor
1. Set the sensor guide handle (7) in "Service" position.
2. Open the air relief valve on the underside (1) with a 5 mm hex key.
3. Carefully insert the sensor into the insertion fitting's built-in gasket (inside 6) and lead the other end into the sensor guide (2).
4. Close the air relief valve on the underside (1) with a 5 mm hex key.
5. Make sure that the residual, in-line pressure is max. 1 bar !
6. Support the sensor end piece (3) and slowly open the ball valve (4).
Note: Rebounds may occur at this point !
7. Release (pull out) the safety lock (5), and at the same time pull the sen­sor guide handle (7).
8. When the end piece (3) touches the insertion fitting body (6), the sensor guide handle (7) has reached its "Measure" position, and the safety lock (5) engages (falls into place) with audible click.
9. Make sure that the safety lock (5) has been engaged (in locked posi­tion).
Removal of the Sensor
1. Make sure that the residual, in-line pressure is max. 1 bar !
2. Release (pull out) the safety lock (5), and at the same time push the sensor guide handle (7) forwards to extract the sensor from the inser­tion fitting body (6).
Note: Rebounds may occur at this point !
3. The sensor guide handle (7) is in "Service" position, when the safety lock (5) engages with a click.
4. Close the ball valve (4).
5. Open the air relief valve (1).
6. Support the sensor end piece (3), and carefully lift the sensor out of the sensor guide (2).
7. Close the air relief valve (1).
8. Remove the sensor.
EN 5.10 SuSix Manual 1406
21
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
This page intentionally left blank.
22EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
4. Sensor Calibration
Turbidity
The SuSix® sensor is factory calibrated for turbidity measurements. No further calibration is required.
Suspended Solids
Calibration is required prior to using the SuSix® for suspended solids measurements.
Important! Suspended solids calibration must be carried out at fixed intervals.
If a distinct change in composition and/or colour of the process appears, the SuSix® must be re-calibrated immediately.
Note: Representative measuring values can only be expected if the calibration point is obtained at a sludge concentration of more than 50% of the application’s nominal range.
Calibration may be 1-point, 2-point or 3-point as required. Normally a 1-point calibration will be sufficient.
Zero-point calibration is not required!
Calibration Procedures
1. Take out the SuSix® sensor and clean it according to the directions in the section about cleaning the sensor during maintenance.
2. Extract a sample from the actual process. The necessary volume must be at least 10 litres. The sample is ana­lyzed at a laboratory, and the content of suspended solids can be determined. This value must be kept as it is required later on.
EN 5.10 SuSix Manual 1406
23
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
3. Dip the SuSix® sensor in the sample stirring continuously, or leave it in
the fluid in which the suspended solids measurement is going to take place.
Ensure sure that the extracted sample does not sediment (deposit) by stirring continuously.
Important! If the signal from the suspended solids sensor changes very slowly on the display, make sure that the signal attenuation is not set too high. Check likewise in the “Averaging” menu that the response time is not set too high.
The recommended response time is 5 seconds.
24EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Converter Setup
accordingly
Adjust samples
Adjust sample 1 Measured 10,2 g/l 2 Measured 8,11 g/l 3 Measured 5,87 g/l Back Select OK
Clear Measurement sample No Yes
Back Select Ok
Clear Measurement sample
4. Proceed to “Sensor Calibration” in the “Converter Setup” menu (see below). Note that the options "Sensor Calibration", "Cleaning" and "Solid curve" will be marked with an "NA" (not applicable), if they cannot be selected (for example if wrong units have been chosen).
Sensor Calibration Averaging Units mA output Cleaning Solid curve High Alarm Low Alarm Sensor error Status Factory settings Service menu Back Select OK
Sensor Calibration
Sensor Calibration Measurement samples Adjust samples Clear ”Measurement sample” Clear ”Adjust samples”
Back Select OK
Measurement samples
Select sample
1. point Measured
2. point Measured
3. point Measured Back Select OK
5. Select “Measurement samples”, “1. point” and close with OK.
EN 5.10 SuSix Manual 1406
25
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
accordingly
Sensor Calibration
Adjust samples
Adjust
1. point 00010,2 g/l
Back Select >
Adjust sample 1 Measured 10,2 g/l 2 Measured 8,11 g/l 3 Measured 5,87 g/l Back Select OK
Next
Adjust
2. point 00008,11 g/l
Back Select >
Adjust
3. point 00005,87 g/l
Sensor Calibration Measurement samples Adjust samples Clear Measurement sample Clear Adjust samples
Back Select OK
Measurement samples
Select sample
1. point Measured
2. point Measured
3. point Measured Back Select OK
Clear Adjust samples
Clear Adjustment sample No Yes
Back Select Ok
2. point
1. point
3. point
Next
Ok
Ok
Select point
Select point
Clear Measurement sample No Yes
Back Select Ok
Clear Measurement sample
1./2./3. point Start sample? Progress [ 0% ]
Exit Start
6. Press “Start” to start loading of the actual sample’s values.
During this process a bar graph is displayed indicating the progress
from 0-100 %.
Start point
1./2./3. point Measureing Progress [||||||||||||||||| ] 45%
Cancel
Measure point 1/2/3 Done Progress [ 100% ]
Exit Next
26EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Averaging
Quality improvement No Yes
Back Select >
accordingly
Averaging
(from 1 sek to 120 min)
Back Select >
Adjust calibration
Adjust samples
Clear Adjust samples
Clear Adjustment sample No Yes
Back Select Ok
Clear Measurement sample
Clear Measurement sample
7. After data loading is finished, a realignment of the measured value with respect to the previously measured laboratory value is required. Select ” Adjust samples” in the “Sensor Calibration” menu, select ”1. point” and close with OK.
Adjust sample 1 Measured 10,2 g/l 2 Measured 8,11 g/l 3 Measured 5,87 g/l Back Select OK
Adjust
1. point
1. point
Back Select >
Ok
Select point
No Yes
Back Select Ok
00010,2 g/l
Next
8. Select ”Adjust” in ”1. point” and enter the value from the analyzed labo­ratory sample. If a 2- or 3-point calibration is required, repeat the previous steps re­placing “1. point” with “2. point” and/or “3. point”.
EN 5.10 SuSix Manual 1406
27
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Averaging
Quality improvement No Yes
Back Select >
NTU (Turbidity) FTU (Turbidity) FNU (Turbidity) EBC (Turbidity) % (Suspended solids) mg/l (Suspended solids) g/l (Suspended solids) Kg/l (Suspended solids)
Back Select OK
accordingly
Averaging
0001 sec
(from 1 sek to 120 min)
Back Select >
Yes
Clear Adjust samples
Clear Adjustment sample No Yes
Back Select Ok
Clear ”Measurement sample”
Averaging
Units
Quality improvement No Yes
Back Select >
20 mA output
Units NTU (Turbidity) FTU (Turbidity) FNU (Turbidity) EBC (Turbidity) % (Suspended solids) mg/l (Suspended solids) g/l (Suspended solids) Kg/l (Suspended solids)
Back Select OK
4mA output
4 mA = 00.00 g/l
Back Select OK
accordingly
Use mA output Yes No Back Select OK
Averaging
0001 sec
(from 1 sek to 120 min)
Back Select >
Yes
Clear ”Adjust samples”
Delete Calibration
If you want to delete a calibration point (the measured value), you can do it from the ”Clear "Measurement sample"” menu.
Clear ”Measurement sample” No Yes
Back Select Ok
If you want to delete an alignment point (the entered value), you can do it from the ”Clear "Adjustment samples"” menu.
Clear ”Adjustment sample” No Yes
Back Select Ok
28EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
5. Converter
Electrical Mounting
Warning!
The SuSix® converter/display unit must not be mounted in explosion hazardous areas!
Loosen the four screws (positions indicated by arrows) and remove the display unit to gain access to the terminals.
Max. four MJK converters with Modbus can be connected to one display unit:
Display unitConverter #4Converter #1
+DC GND A B
Pink
Grey
Shield
Brown
display unit
display unit
+ +A B C2 N02C NO N01 PE N L
- -
-mA- -D02--D01--DI--Display-
shielded cable.
521269
C1
2 wire
4 wire
shielded cable.
Mains supply
521268
Fuse 24V AC: 500mA T
Fuse 115V AC: 125mA T
Fuse 230V AC: 63mA T
Line
EN 5.10 SuSix Manual 1406
29
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
Power Supply
The flow converter must be supplied from a properly fused mains outlet, a 24 volt AC outlet, or a 10 - 30 V DC power supply/battery.
Power Supply 230 V AC, 115 V AC or 24 V AC
Terminal Designation
PE Protective ground
N 230 / 115 / 24 V AC neutral
L 230 / 115 / 24 V AC live
Power Supply 10 - 30 V DC
Terminal Designation
PE Protective ground
DC neutral
+ DC live
The internal fuse ratings are:
Internal Fuse Ratings
Voltage Rating Order no. Dimension
230 V AC 0.063 mA T 550030 5 x 20 mm
115 V AC 0.125 mA T 550035 5 x 20 mm
24 V AC 0.5 mAT 550049 5 x 20 mm
10 - 30 V DC 1.0 AT 550051 5 x 20 mm
The technical specifications/requirements for a 10 - 30 V DC power supply/battery are:
Technical Specifications for 10 - 30 V DC Power Supply
Power consumption without display < 5 W
Power consumption with display < 8 W
Peak start current @12 V DC,1 second Approx. 1,5 A
Peak start current @24 V DC,1 second Approx. 1 A
30EN 5.10 SuSix Manual 1406
FW: 841013-005/843003-001
Loading...
+ 102 hidden pages