Currentness of firmwareThe process of consistently improving our products includes the
The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may
result in differences between this operating manual and your
instrument. Also, we cannot guarantee that there are absolutely no
errors in this manual. Therefore, we are sure you will understand that
we cannot accept any legal claims resulting from the data, figures or
descriptions.
continuous further development of instrument firmware. The current
Multi 3500i firmware is available on the Internet. It can easily be
downloaded on your meter using the enclosed AK 340/B cable and a
PC. For more detailed information, refer to the appendix of this
operating manual or to the Internet under http://www.WTW.com
The Multi 3500i compact precision handheld meter enables you to carry out pH measurements, ISE measurements, dissolved oxygen (DO)
measurements and conductivity measurements quickly and reliably.
The Multi 3500i handheld meter provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
®
The proven MultiCal
cedures to determine/set up the cell constant support you in your work
with the meter. The special AutoRead function enables precise measurements.
and OxiCal® calibration procedures and the pro-
3
2
M
C
A
L
STO
P
R
T
O
ESC
1
1Keypad
2Display
3Socket field
Note
If you need further information or application notes, you can obtain the
following material from WTW:
z Application reports
z Primers
z Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
3
OverviewMulti 3500i
1.2SETs of equipment
The measuring instrument is also available as part of individual SETs
of equipment.
You will find additional information on this and other accessories in the
WTW catalog or via the Internet.
16
15
14
13
12
10
11
17
18
19
9
8
7
6
5
4
2
5
4
3
1
4
Multi 3500iOverview
Set (sample configuration):
1Multi 3500i measuring instrument, carrying strap with 2 carrying
clips, armoring
2Cond/Oxi beaker with beaker clip
3pH beaker
4Stand
5Plastic beaker, 50 ml
6Storing solution for pH electrodes
750 ml pH buffer solution, STP 4
850 ml pH buffer solution, STP 7
9Calibration and control standard for conductivity measuring
cells, 50 ml
1050 ml ELY/G electrolyte solution for DO sensors
1150 ml RL/G cleaning solution for DO sensors
12Exchange membrane caps for DO sensors (3 pieces)
13SF 300 polishing strip for the maintenance of DO sensors
14Conductivity measuring cell
15pH combination electrode
16DO sensor
17Operating manual + short operating manual
18Equipment case
19Plug-in power supply unit
MultiLab pilot CD-ROM
5
OverviewMulti 3500i
1.3Keypad
PRTM
ESCSTOCAL
MENU
ENTER
Key functions
PRT
CAL
STO
ESC
M
Select the measured variable <M>:
– pH value / ORP / ion concentration
– DO concentration / DO saturation / DO partial
pressure
– Conductivity / specific resistance / salinity /
TDS
Output display contents to RS232 interface (e.g.
print)
<PRT>
Switch the measuring instrument on/off
<ON/OFF>
Calibrate the currently selected measured vari-
able
<CAL>
Store a measured value
<STO>
Switch to the next higher menu level /
cancel input
<ESC>
Set values
<▲>, <▼>
MENU
ENTER
6
Open a menu / confirm input
<MENU/ENTER>
Multi 3500iOverview
1.4Display
The graphical display can indicate up to three measuring windows at
the same time. The illumination enables to read the display even in the
darkness. You can modify the display to meet your requirements in
multiple ways.
Measured variable
Measured value (with unit)
pH
O
2
[SAL]
[nLF][Tr25]
6.949
7.29
731
mg/l
µS/cm
05.09.03 09:25
1.5Socket field
24.0
24.0
24.0
°C
°C
°C
Temperature display
Sensor symbol
Status line
5
4
123
1DO sensor or conductivity measuring cell or combined con-
ductivity / DO sensor
2pH electrode or ISE combination electrode
3pH temperature sensor
4Plug-in power supply unit
5RS232 serial interface
7
OverviewMulti 3500i
8
Multi 3500iSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the measuring instrument. Consequently, all responsible personnel must read this
operating manual before working with the measuring system. The operating manual must always be available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in
the laboratory.
Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to
take when handling chemicals.
Safety instructionsThe individual chapters of this operating manual use the following safe-
ty instruction to indicate various types of danger:
Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the instrument
or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of
the:
z pH and ORP measurement
z ISE measurement
z measurement of dissolved oxygen (DO) and
z conductivity measurement
in the field and laboratory.
The technical specifications as given in chapter 7 T
be observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is authorized. Any other use is considered to be unauthorized.
ECHNICALDATA must
9
SafetyMulti 3500i
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter
7T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
The smooth functioning and operational safety of the measuring instrument can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are
followed during operation.
The smooth functioning and operational safety of the measuring instrument can only be guaranteed under the environmental conditions that
are specified in chapter 7 T
ECHNICALDATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
of service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
Obligations of the
purchaser
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of this measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
10
Multi 3500iCommissioning
3Commissioning
3.1Scope of delivery
z Multi 3500i handheld measuring instrument with 4 rechargeable bat-
teries, 1.2 V type AA in the instrument
z Plug-in power supply with Euro plug,
exchange plugs for USA, UK, and Australia are enclosed
z Operating manual and short operating manual
z MultiLab pilot CD-ROM
3.2Power supply
Mains operation and
charging the batteries
Charging time of the
batteries
You can operate the measuring instrument either with the built-in rechargeable batteries or with the plug-in power supply. The plug-in power supply supplies the measuring instrument with low voltage (9 V DC).
At the same time, the rechargeable batteries are charged. The batteries are charged even when the instrument is switched off.
approx. 36 hours. The LoBat display indicator appears when the batteries are nearly empty and have to be charged as soon as possible.
Caution
The line voltage at the operating site must lie within the input
voltage range of the original plug-in power supply (see chapter
7T
ECHNICALDATA).
Caution
Use original plug-in power supplies only (see
chapter 7 T
Note
The batteries should not be completely discharged. If you do not operate the instrument for a longer period of time you should charge the batteries every six months.
ECHNICALDATA).
Automatic switchoffThe instrument has an automatic switchoff function in order to save the
batteries (see section 4.3.3).
Display illumination
with battery operation
During battery operation, the measuring instrument automatically
switches off the display illumination if no key has been pressed for
15 seconds. The illumination is switched on with the next keystroke
again. The display illumination can be switched off completely.
11
CommissioningMulti 3500i
4
3
2
1
1If necessary, replace the Euro plug (1) on the plug-in power
supply unit (2) by the country-specific plug suitable for your
country.
2Connect the plug (3) to the socket (4) of the measuring instru-
ment.
3Connect the plug-in power supply unit to an easily accessible
mains socket.
3.3Initial commissioning
Perform the following activities:
z For mains operation and charging the batteries: Connect the plug-in
power supply unit (see section 3.2 P
z Switch on the measuring instrument (see section 4.1)
z Set the language (see section 4.2.3)
z Set the date and time (see section 4.2.4)
Note
When you set the language, date and time according to the mentioned
sections of this operating manual you will quickly be familiar with the
simple operation of the Multi 3500i.
OWERSUPPLY).
12
Multi 3500iOperation
4Operation
4.1Switching on the measuring instrument
Switching onPress the <ON/OFF> key.
The measured value display appears.
pH
6.949
25.0
Switching offPress the <ON/OFF> key.
Automatic switchoffThe instrument has an automatic switchoff function in order to save the
batteries (see section 4.3.3). The automatic switchoff switches off the
measuring instrument if no key is pressed for an adjustable period.
The automatic switchoff is not active
z if the power is supplied by the plug-in power supply unit,
z if the Automatic data storage function is active, or with automatic
data transmission
z if the communication cable and a PC with a running communication
program are connected,
z if the printer cable is connected (for external printers).
Display illumination
with battery operation
During battery operation, the measuring instrument automatically
switches off the display illumination if no key has been pressed for
15 seconds. The illumination is switched on with the next keystroke
again.
°C
13
OperationMulti 3500i
4.2General operating principles
This section contains basic information of the operation of the
Multi 3500i.
Operating elements,
display
Operating modes,
navigation
An overview of the operating elements and the display is given in
section 1.3 and section 1.4.
An overview of the operating modes and navigation of the Multi 3500i
is given in section 4.2.1 and section 4.2.2.
4.2.1Operating modes
The instrument has the following operating modes:
z Measuring
The measuring data of one to three sensors is displayed in the
measured value display
z Calibration
The course of a calibration with calibration information, functions
and settings is displayed
z Data storage
The measuring instrument stores measuring data automatically or
manually
z Transmitting data
The measuring instrument transmits measuring data and calibration
records to the serial interface automatically or manually.
z Setting
The system menu or a sensor menu with submenus, settings and
functions is displayed
14
Multi 3500iOperation
4.2.2Navigation
Measured value displayIn the measured value display, you can
z select a measuring window with <▲><▼> and
open the relevant measuring menu by shortly
pressing
<MENU/ENTER>.
z open the system menu with the sensor-independent settings by
pressing <MENU/ENTER> for a long
time (approx. 1 s).
z change the display in the selected measuring window
(e. g. pH < − > mV) by pressing <M>.
Menus and dialogsThe menus for settings and dialogs in courses contain further sub-
menus. The selection is made with the <▲><▼> keys. The current se-
lection is displayed in reverse video.
z Submenus
The name of the submenu is displayed at the upper edge of the
frame. Submenus are opened by confirming with <MENU/ENTER>.
Example:
System
General
Interface
Clock function
Reset
z Settings
Settings are indicated by a colon. The current setting is displayed on
the right-hand side. The setting mode is opened with
<MENU/ENTER>. Subsequently, the setting can be changed with
<▲><▼> and <MENU/ENTER>.
Example:
System
Language:Deutsch
Beep
:Off
Illumination:On
Contrast:48 %
Temperature unit
Switchoff time:30 min
:°C
15
OperationMulti 3500i
z Functions
Functions are designated by the name of the function. They are immediately carried out by confirming with <MENU/ENTER>.
Example: Display the Calibration record function.
pH
Calibration type:TEC
Calibration interval
Unit for slope:mV/pH
Calibration record
:
i 2.00 4.01 7.00 10.01
:7 d
MessagesInformation or operating instructions are designated by the i symbol.
They cannot be selected.
Example:
pH
i Buffer recognitionTEC
i Immerse sensor in buffer 1
Set temperature:25 °C
Continue
Note
The principles of navigation are explained in the two following sections
by reference of examples:
z Set the language (see section 4.2.3)
z Set the date and time (see section 4.2.4).
16
Multi 3500iOperation
4.2.3Example 1 on navigation: Setting the language
1Press the <ON/OFF> key.
The measured value display appears.
The instrument is in the measuring ode.
pH
7.00
25.0
°C
2Open the system menu by pressing <MENU/ENTER> for a
long
time (approx. 1 s).
The instrument is in the setting mode.
Storage & config
Data storage
Display
System
3Select the System submenu with <▲><▼>.
The current selection is displayed in reverse video.
4Open the System
submenu with <MENU/ENTER>.
System
General
Interface
Clock function
Reset
5Select the General submenu with <▲><▼>.
The current selection is displayed in reverse video.
17
OperationMulti 3500i
6Open the General submenu with <MENU/ENTER>.
System
Language:Deutsch
Beep
:Off
Illumination:On
Contrast:48 %
Temperature unit
Switchoff time:30 min
:°C
7Open the setting mode for the Language with
<MENU/ENTER>.
System
Language:Deutsch
Beep:Off
Illumination:On
Contrast
Temperature unit:°C
Switchoff time:30 min
:48 %
8Select the required language with <▲><▼>.
9Confirm the setting with <MENU/ENTER>.
The setting becomes active the next time the system menu is
called up.
10To make further settings, switch to the next higher menu level
with <ESC>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring ode.
18
Multi 3500iOperation
4.2.4Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date
and time are indicated in the status line of the measured value display.
The indication can be switched off. When storing measured values and
calibrating, the current date and time are automatically stored as well.
The correct setting of the date and time and date format is important for
the following functions and displays:
z Current date and time
z Calibration date
z Identification of stored measured values.
Therefore, check the time at regular intervals.
Note
After a fall of the supply voltage (empty batteries), the date and time are
reset to 01.01.2003, 00:00 hours.
Setting the date, time
and date format
The data format can be switched from the display of day, month, year
(dd.mm.yy) to the display of month, day, year (mm/dd/yy or mm.dd.yy).
1In the measured value display:
Open the system menu by pressing <MENU/ENTER> for a
long
time (approx. 1 s).
The instrument is in the setting operating mode.
2Select and confirm the System / Clock function menu with
<▲><▼> and <MENU/ENTER>.
The setting menu for the date and time opens up.
3Select and confirm the Time menu with <▲><▼> and
<MENU/ENTER>.
The seconds are highlighted.
System
Time:14:53:40
Date
:30.10.03
Date format:dd.mm.yy
4Change and confirm the setting with <▲><▼> and <MENU/
ENTER>.
The minutes are highlighted.
19
OperationMulti 3500i
5Change and confirm the setting with <▲><▼> and <MENU/
ENTER>.
The hours are highlighted.
6Change and confirm the setting with <▲><▼> and <MENU/
ENTER>.
The time is set.
7If necessary, set the Date and Date format. The setting is made
similarly to that of the time.
8If necessary, select and set the Date with <▲><▼> and
<MENU/ENTER>.
9To make further settings, switch to the next higher menu level
with <ESC>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring operating mode.
20
Multi 3500iOperation
4.3System settings (system menu)
The system menu comprises the following settings:
z Data storage (see section 4.3.1)
z Display (see section 4.3.2)
z System (see section 4.3.3).
4.3.1Data storage
This menu contains all functions to display, edit and erase stored measured values and calibration records.
Note
Detailed information on the data storage functions of the Multi 3500i is
given in section 4.8.
4.3.2Display
With the aid of Display submenu, you can modify the measured value
display to meet your requirements. When doing so, you can display or
hide the following elements:
z pH/ISE measuring window
z DO measuring window
z Conductivity measuring window
z Date indication in the status line
z Time indication in the status line
Note
When several sensors or multiple sensors (e.g. ConOx) are connected,
all available measured variables are automatically displayed. If you do
not wish to have all measured values displayed, you can hide measured values of individual sensors.
21
OperationMulti 3500i
SettingsTo open the system menu in the measured value display, press the
<MENU/ENTER> key for approx. 1 s
. After completing the settings,
switch to the measured value display with <M>.
Menu itemSettingDescription
Display / TimeOn
Off
Display / DateOn
Off
Display / pHOn
Off
Display / O2On
Off
Display / CondOn
Off
Display of the time in the system
status line
Display of the date in the system
status line
Display of the pH/ISE measuring
window.
This menu item is only visible if the
electrode and an additional sensor
are connected.
Display of the DO measuring window.
This menu item is only visible if a
DO sensor is connected.
Display of the conductivity measuring window.
This menu item is only visible if a
conductivity measuring cell is connected.
22
Multi 3500iOperation
4.3.3System
OverviewThe following sensor-independent instrument features can be adjusted
in the system menu/System and its submenus:
z Menu language
z Beep on keystroke
z Display illumination
z Display contrast
z Unit of the temperature display
z Interval of the automatic switchoff
z Data interface
z Clock and date function
z Function to reset all sensor-independent system settings to the de-
fault condition
SettingsTo open the system menu in the measured value display, press the
<MENU/ENTER> key for approx. 1 s
. After completing the settings,
switch to the measured value display with <M>.
Menu itemSettingDescription
System / General /
Language
Deutsch
English
Select the menu language
(further)
System / General /
Beep
System / General /
Illumination
System / General /
Contrast
System / General /
Temperature unit
On
Off
On
Off
Switch on/off the beep on
keystroke
Switching the display illumination on/off
0 ... 100 %Changing the display con-
trast
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperatures are displayed with the selected unit.
System / General /
Switchoff time
10, 20, 30, 40,
50 min,
1 2 3 4 5 10 15
20 24 h
The automatic switchoff automatically switches the measuring instrument off if no
entry is made for a specified
period of time (switchoff interval). This saves the batteries.
23
OperationMulti 3500i
Menu itemSettingDescription
System / Interface /
Baud rate
1200, 2400,
4800, 9600,
Baud rate of the data interface
19200
System / Interface /
Output format
ASCII
CSV
Output format for data transmission
For details, see section 4.9
System / Interface /
Header
Yes
No
Option for output in csv format. "Yes" creates a header
in the table.
System / Clock function
Time
Date
Settings of time and date.
For details, see section 4.2.4
Date format
System / Reset-Resets the system settings to
the default values.
For details, see
section 4.10.2
24
Multi 3500iOperation
4.4pH value / ORP voltage
4.4.1General information
You can measure the following variables:
z pH value [ ]
z ORP [mV]
Caution
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result. The
RS232 interface is not galvanically isolated.
Temperature
measurement
For reproducible pH measurements, it is essential to measure the temperature of the test sample.
You have the following possibilities of measuring the temperature:
z Automatic measurement of the temperature by the temperature sen-
sor (NTC30 or Pt 1000) integrated in electrode.
z Measurement of the temperature by the integrated temperature sen-
sor of a simultaneously connected DO sensor or conductivity measuring cell in the test sample.
z Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is
connected and automatically switches on the temperature measurement.
25
OperationMulti 3500i
The display of the temperature indicates the active temperature measuring mode:
Temperature
sensor
Resolution
of the temp.
Mode
display
pHCond or
Oxi
yes-0.1 °CAutomatic with pH tem-
yesyes0.1 °C
perature sensor
-- 1°CManual
-yes0.1 °C,
measured
temperature
value flashes
The temperature value of
the second sensor (Cond
or Oxi) in the same test
sample is taken over for
pH measurement*
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature
input or
– use an electrode with a temperature sensor.
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect a pH or ORP electrode to the measuring instrument.
The pH/ISE measuring window is displayed.
2If necessary, select the pH or mV display with <M>.
3Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without a temperature sensor.
4Calibrate or check the measuring instrument with the elec-
trode.
26
Multi 3500iOperation
4.4.2Measuring the pH value
1Perform the preparatory activities according to section 4.6.1.
2Immerse the pH electrode in the test sample.
pH
6.949
AutoRead
(Drift control)
CriteriaWith identical measurement conditions, the following applies:
24.8
3Select the pH or mV display with <M>.
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured variable flashes until a stable measured value is available.
Measured
variable
pH valueBetter than 0.01> 30 seconds
°C
ReproducibilityResponse time
27
OperationMulti 3500i
4.4.3Measuring the ORP voltage
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
1Perform the preparatory activities according to section 4.6.1.
2Submerse the ORP electrode in the sample.
U
AutoRead
(drift control)
CriteriaWith identical measurement conditions, the following applies:
157.0
24.8
3Select the mV display with <M>.
The AutoRead function (drift control) continually checks the stability of
the measurement signal. The stability has a considerable impact on the
reproducibility of measured values. The display of the measured variable flashes until a stable measured value is available.
Measured variableReproducibilityResponse time
ORP voltagebetter than 0.6 mV> 30 seconds
mV
°C
28
Loading...
+ 77 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.