WarrantyWe guarantee the instrument described for 3 years from the date of
The use of advanced technology and the high quality standard of our
instruments are the result of continuous development. This may result
in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot accept
any legal claims resulting from the data, figures or descriptions.
purchase.
The instrument warranty covers manufacturing faults that are discovered within the warranty period. The warranty does not cover components that are replaced during maintenance work, e.g. batteries.
The warranty claim extends to restoring the instrument to readiness for
use but not, however, to any further claim for damages. Improper handling or unauthorized opening of the instrument invalidates any warranty claim.
To ascertain the warranty liability, return the instrument and proof of
purchase together with the date of purchase freight paid or prepaid.
The compact precision handheld meter Multi 3400i enables you to carry out pH measurements, dissolved oxygen (D. O.) measurements and
conductivity measurements quickly and reliably.
The Multi 3400i handheld meter provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven MultiCal
®
and OxiCal® calibration procedures and the procedures to determine/set up the cell constant support you in your work
with the meter. The special AutoRead function enables precise measurements.
9
O
2
7
6
A
R
n
O
m
g/l
°
C
T
g
P
CAL
7
1
M
RCL
ST
O
AR
1Keypad
2Display
3Jack field
Note
If you need further information or application notes, you can obtain the
following material from WTW:
z Application reports
z Primers
z Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
5
OverviewMulti 3400i
1.2SETs of equipment
The measuring instrument is also available as part of individual SETs
of equipment.
You will find additional information on this and other accessories in the
WTW catalog or via the Internet.
16
15
14
13
12
10
11
17
18
19
9
8
7
6
5
4
2
5
4
3
1
Set (sample configuration):
1Multi 3400i meter, carrying strap with 2 carrying clips, armoring
2Cond/Oxi beaker with beaker clip
3pH beaker
4Stand
5Plastic beaker
6Storing solution for pH electrodes
6
Multi 3400iOverview
750 ml pH buffer solution, STP 4
850 ml pH buffer solution, STP 7
9Calibration and control standard for conductivity measuring
cells, 50 ml
1050 ml ELY/G electrolyte solution for D.O. probes
1150 ml RL/G cleaning solution for D.O. probes
12WP 90/3 exchange membrane heads for D. O. probes
(3 pieces)
13SF 300 abrasive film for D.O. probes
14Conductivity measuring cell
– TetraCon® 325-3 or
– TetraCon® 325
15pH electrode
– SenTix 41-3 or
– SenTix 41
16Oxygen sensor
– CellOx 325-3 or
– CellOx 325
17Operating manual + short operating manual
18Equipment case
19Plug-in power supply unit
7
OverviewMulti 3400i
1.3Keypad
STOM
ARRCLCAL
RUN/ENTER
Key functions
M
STO
CAL
RCL
AR
Select the measured variable <M>:
– pH value / ORP voltage
– D. O. concentration / D. O. saturation
– Conductivity / salinity
Save a measured value
<STO>
Switch measuring instrument on/off
<ON/OFF>
Calibrate the currently set measured variable
<CAL>
Display/transmit measured values
<RCL>
Activate/deactivate the AutoRead function
<AR>
Select the measuring mode, increase values,
scroll
<▲ >
Select the measuring mode, decrease values,
scroll
The following overview diagram shows which keys you have to press
to select between the different measuring modes:
Note
When a D. O. probe or a conductivity measuring cell with a temperature
probe is connected, the measuring instrument recognizes the probe or
measuring cell and automatically switches to the oxygen measuring
mode or conductivity measuring mode last selected. As soon as the
probe/measuring cell is disconnected, the instrument switches to the
pH or ORP measuring mode.
10
Multi 3400iSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the measuring instrument. Consequently, all responsible personnel must read this
operating manual carefully before working with the measuring system.
The operating manual must always be available within the vicinity of the
measuring system.
Target groupThe measuring instrument was developed for work in the field and in
the laboratory.
Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to
take when handling chemicals.
Safety instructionsThe individual chapters of this operating manual use the following safe-
ty instruction to indicate various types of danger:
Warning
indicates instructions that must be followed precisely in order to avoid
the possibility of slight injuries or damage to the instrument or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of
the:
z pH and ORP measurement
z measurement of the oxygen content and
z measurement of the conductivity, salinity and temperature
in the field and laboratory.
The technical specifications as given in chapter 7 T
be observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is authorized. Any other use is considered to be unauthorized.
ECHNICALDATA must
11
SafetyMulti 3400i
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter 7
T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
The smooth functioning and operational safety of the measuring instrument can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are
followed during operation.
The smooth functioning and operational safety of the measuring instrument can only be guaranteed under the environmental conditions that
are in specified in chapter 7 T
ECHNICALDATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
of service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
Obligations of the pur-
chaser
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
12
Multi 3400iCommissioning
3Commissioning
3.1Scope of delivery
z Multi 3400i handheld meter
z Plug-in power supply unit
z Operating manual and short operating manual
z 4 batteries, 1.5 V Mignon type AA (in the instrument)
Setting the date and
time
For details of scope of delivery of SETs, see section 1.2 SET
EQUIPMENT and WTW catalog.
3.2Initial commissioning
Perform the following activities:
z Set the date and time
z Connect the plug-in power supply (optional)
1Press the <M> key and hold it down.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
3Press the <RUN/ENTER> key repeatedly until the date ap-
pears on the display.
4Set the date of the current day with <▲> <▼>.
5Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
SOF
6Set the current month with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The year appears on the display.
8Set the current year with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The hours flash on the display.
10Set the current time with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The minutes flash on the display.
12Set the current time with <▲> <▼>.
13
CommissioningMulti 3400i
13Confirm with <RUN/ENTER>.
The instrument switches to a measuring mode.
14Switch the instrument off using <ON/OFF>.
Connecting the
plug-in power supply
You can either operate the measuring instrument with batteries or with
the plug-in power supply. The plug-in power supply supplies the measuring instrument with low voltage (12 VDC). This saves the batteries.
Warning
The line voltage at the operating site must lie within the input voltage
range of the original plug-in power supply (see chapter 7 T
TA).
ECHNICALDA-
Warning
Use original plug-in power supplies only (see
chapter 7 T
ECHNICALDATA).
2
1
3
14
1Plug the jack (1) into the socket (2) of the measuring instru-
ment.
2Connect the original WTW plug-in power supply (3) to an easily
accessible mains socket.
Multi 3400iOperation
4Operation
4.1Switching on the measuring instrument
1Press the <ON/OFF> key.
The display test appears briefly on the display.
After this, the measuring instrument automatically switches to
the measuring mode.
Measuring mode when
switching on
Connected sensorMeasuring mode
No sensor or pH/ORP electrodepH or ORP measurement (de-
pending on the last selected setting)
D. O. probe or conductivity measuring cell
2 sensors of any typeLast selected measuring mode
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour.
The energy saving feature is not active
z if the power is supplied by the plug-in power supply,
z if the AutoStore function is active,
z if the communication cable and a PC with a running communication
program are connected,
z if the printer cable is connected (for external printers).
Last selected measuring mode
AutoRange measurement range selection
There are several measuring ranges available for both oxygen and
conductivity measurements. As a rule, the measuring instrument is in
the measuring range with the highest possible resolution. If a measuring range is exceeded, AutoRange causes the measuring instrument to
change automatically to the next higher measuring range.
The AutoRange function is always active with oxygen and conductivity
measurements and cannot be switched off.
Note
The measuring ranges for oxygen and conductivity measurements can
be found in chapter 7 T
ECHNICALDATA.
15
OperationMulti 3400i
4.2pH value / ORP voltage
4.2.1General information
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3Calibrate or check the measuring instrument with the elec-
trode.
4Using <▲ > <▼>, toggle between the pH or mV measuring
modes.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values. Calibrate regularly before measuring. You can only connect electrodes of the NTC30 type or without temperature sensor.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
16
Multi 3400iOperation
Temperature measure-
ment in pH measure-
ments
You can perform pH measurements with or without a temperature sensor as well as with the temperature sensor of an oxygen sensor or a
conductivity measuring cell. The measuring instrument recognizes
which sensors are connected and switches automatically to the correct
mode for the temperature measurement.
The following cases are distinguishable.
Temperature sensorDisplayMode
pHCond or
Oxi
yes-TPAutomatic with
pH temperature sensor
yesyesTP
--Manual
-yesTP
flashes
The temperature value of the
second probe (Cond or Oxi) in
the same sample is taken
over for the pH measurement*
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature
input or
– use an electrode with a temperature sensor.
If a temperature sensor is connected, it is indicated on the display by
TP.
Note
When calibrating without a temperature sensor (no TP display indicator
displayed), enter the current temperature of the respective buffer solution manually using the <▲> <▼> keys while keeping the <RUN/EN-
TER> key depressed.
17
OperationMulti 3400i
4.2.2Measuring the pH value
1Perform the preparatory activities according to section 4.2.1.
2Immerse the pH electrode in the test sample.
3Press the <▲> <▼> keys until pH appears in the status display.
The pH value appears on the display.
AutoRead AR
(Drift control)
TDS
S
6
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
4When measuring without a connected temperature sensor:
Options:
z Determine the current temperature using a thermometer
z TP display indicator not displayed, socket for the second
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values.
pH
9
2
8
AutoCal DIN
Tref20
Lin
Cal
AutoCalTEC
and, while keeping the <RUN/ENTER> key depressed, enter this temperature value with <▲> <▼>.
probe is free:
Connect the second probe (Oxi or Cond) and immerse it in
the same sample. TP flashes, the temperature is automatically measured using the second probe.
9
4
SalOU
mV/pH
%
m
mbar
1/
°
F
8
Auto
ARng
mg/l
S/
cm
M
cm
°
%
C
Store
AR
cm
Sal
K
/
RCL
TP
18
For identical measurement conditions, the following criteria apply:
ReproducibilityResponse time
Better than 0.02> 30 seconds
1Call up the pH measuring mode with <▲> <▼>.
2Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
Multi 3400iOperation
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the <AR> key.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
4.2.3Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. SenTix ORP, the measuring
instrument can measure the ORP voltage (U) of a solution.
1Perform the preparatory activities according to section 4.2.1.
2Submerse the Redox electrode in the sample.
3Press the <▲> <▼> keys until the U status display appears.
The ORP voltage (mV) of the sample appears on the display.
4Wait for a stable measured value.
Lin
2
8
Tref20
Cal
pH
U
2
AutoCal DINAutoCalTEC
3
4
SalO
mV
%
m
mbar
1/
°
F
8
Auto
ARng
/pH
mg/l
S/
cm
M
cm
°
%
C
Store
AR
cm
Sal
K
/
RCL
TP
TDS
S
8
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
19
OperationMulti 3400i
4.2.4pH calibration
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the measuring instrument. Thus, you should calibrate at regular intervals.
When to calibrate?z After connecting another electrode
z When the sensor symbol flashes:
– after the calibration interval has expired
– after a voltage interruption, e. g. after changing the batteries
AutoCal TECis specially matched to the WTW technical buffer solutions as a fully au-
tomatic two-point calibration. The buffer solutions are automatically
recognized by the measuring instrument. Depending on the instrument
setting (see section 4.8 C
ONFIGURATION), the instrument displays the
relevant buffer nominal value or the current electrode voltage in mV.
The calibration can be terminated after the first buffer solution. This corresponds to a single-point calibration. To do this, the instrument uses
the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
AutoReadThe calibration procedure automatically activates the AutoRead func-
Displaying the calibra-
tion data
Printing the
calibration protocol
tion.
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
You can view the data of the last calibration on the display. The proceeding is described on page 48.
The calibration record contains the calibration data of the current calibration. You can transmit the calibration protocol to a printer via the serial interface (see O
INTERFACE, page 51).
UTPUTTINGTHECALIBRATIONPROTOCOLONTHE
Note
You can automatically print a calibration record after the calibration. To
do so, connect a printer to the interface according to section 4.7.2 before calibrating. After a valid calibration, the record is printed.
Calibration evaluationAfter calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The
worst evaluation appears on the display.
DisplayAsymmetry
[mV]
-15 ... +15-60.5 ... -58
-20 ... +20-58 ... -57
-25 ... +25-61 ... -60.5
-30 ... +30-62 ... -61
Clean the electrode according
to the electrode operating
manual
Slope
[mV/pH]
or
-57 ... -56
or
-56 ... -50
E3
Eliminate the error according to
chapter 6 W
HATTODOIF...
< -30 or
> 30
... -62 or
... -50
21
OperationMulti 3400i
Preparatory activities
AutoCal TECFor this procedure, use any two WTW technical buffer solutions (pH
1Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2Keep the buffer solutions ready.
3Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a
temperature sensor (the TP display indicator is missing from
the display).
values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the WTW technical buffer
solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions can lead
to an erroneous calibration. The correct buffer solutions are given in the
WTW catalog or in the Internet.
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see section
4.8 C
ONFIGURATION), the instrument displays the relevant buffer nomi-
nal value or the current electrode voltage in mV.
Starting the calibration
Note
Skip the steps 2 and 7 if you use a pH electrode with temperature sensor or the temperature sensor of a conductivity measuring cell or a D.
O. probe.
1Press the <CAL> key. The Ct1 display and the function display
AutoCal TEC appears. The sensor symbol displays the evaluation of the last calibration (or no sensor symbol in the delivery
state or after the measurement parameter has been reset).
22
Multi 3400iOperation
TDS
S
C
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
2If required, enter the temperature of the first buffer solution,
while keeping the <RUN/ENTER> key depressed, using <▲> <▼>.
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
7
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
pH
OU
Sal
/pH
mV
%
mg/l
m
S/
cm
cm
M
t
2
8
AutoCal DIN
Tref20
Cal
Lin
AutoCal TEC
3
22
0
AutoCal DIN
Tref20
Cal
Lin
AutoCal TEC
mbar
1
Sal
cm
1/
TP
°
K
%
/
°
C
F
8
4
44
Store
Auto
ARng
AR
RCL
TDSTDS
mV
SS
cm
cm
TP
°
°
C
F
88
Store
Auto
ARng
AR
RCL
7
11
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
Lin
0
0
Tref20
Cal
pHpH
OOUpHUpHSalSal
03
AutoCal DIN
AutoCal TEC
mbarmbar
mV/pHmV/pH
%%
mm
1/1/
°
Auto
ARng
F
mg/lmg/l
S/S/
°
cmcm
MM
cm
%%
C
Store
ARAR
AR
//
cm
SalSal
KK
RCL
TP
5When the measured value is stable, Ct2 appears.
Lin
t
8
Tref20
Cal
pH
OU
2
2
4
AutoCal DIN
AutoCal TEC
Sal
mV
%
m
mbar
1/
°
F
8
Auto
ARng
/pH
mg/l
S/
cm
M
cm
°
%
C
Store
AR
cm
Sal
K
/
RCL
TP
23
TDS
S
C
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
Note
At this point, the AutoCal TEC calibration can be terminated with <M>.
This corresponds to a single-point calibration. To do this, the instru-
ment uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the
asymmetry of the electrode.
OperationMulti 3400i
6Thoroughly rinse the electrode with distilled water.
7If required, enter the temperature of the first buffer solution,
while keeping the <RUN/ENTER> key depressed, using <▲> <▼>.
8Immerse the pH electrode in the second buffer solution.
9Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
TDS
mV
S
1
1
Baud
Time
Day.MonthNo.
Year
Ident
Tref25
LoBat
nLF
10When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The probe symbol
shows the evaluation of the current calibration.
TDS
22
0
Tref20
Lin
44
AutoCal DIN
Cal
AutoCal TEC
pH
OU
S
9
1
Time
Day.Month
Year
LoBat
5
Baud
No.
Ident
Tref25nLF
Tref20
Lin
S
Cal
4
L
AutoCal DIN
AutoCal TEC
cm
TP
°
°
C
F
88
Store
Auto
ARng
AR
RCL
Sal
mV/pH
%
mg/l
m
S/
cm
cm
M
mbar
Sal
cm
1/
TP
°
K
%
/
°
C
F
O
Store
Auto
ARng
AR
RCL
TDS
S
4
1
Time
Day.MonthNo.
Year
Tref25
LoBat
nLF
Baud
Ident
Lin
08
0
Tref20
Cal
pHpH
OOUpHUpHSalSal
14
AutoCal DIN
AutoCal TEC
mbarmbar
mV/pHmV/pH
%%
mm
1/1/
°
F
Auto
ARng
mg/lmg/l
S/S/
MM
cm
°
%%
C
Store
ARAR
AR
cmcm
cmcm
SalSal
TP
KK
//
RCL
24
11Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
Multi 3400iOperation
No.
Lin
0
8
Tref20
Cal
pH
OU
2
S
AutoCal DIN
AutoCal TEC
Sal
mV
%
m
mbar
1/
°
F
YA
Auto
ARng
mg/l
S/
cm
M
cm
°
%
C
Store
AR
cm
Sal
K
/
RCL
TP
TDS
S
6
1
Baud
Time
Day.Month
Year
Ident
Tref25
LoBat
nLF
12Switch to the measuring mode with <M>.
25
OperationMulti 3400i
4.3Dissolved oxygen
4.3.1General information
Note
D. O. measurements with the Multi 3400i can only be carried out using
a CellOx 325 or StirrOx G D. O. probe. The stirrer of the StirrOx G
D. O. probe has to be supplied with voltage separately using the
NT/pH Mix 540 power supply.
You can measure the following variables:
z D. O. concentration
z D. O. saturation
The measuring instrument is supplied with the following functions:
z AutoRange (automatic switchover of the measurement range),
z The AutoRead function (drift control) for checking the stability of the
measurement signal. This ensures the reproducibility of the measuring signal. For details of how to switch the AutoRead function on/off,
see page 29.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect the D. O. probe to the measuring instrument. The in-
strument automatically switches over to D. O. measurement
(O
status display).
2
If the D. O. probe is already connected, press the <M> repeat-
edly until the O
status display appears.
2
2Calibrate or check the measuring instrument with the sensor.
How to calibrate is described in section 4.3.5 from page 20.
3Use <▲> <▼> to toggle between the measuring modes, D. O.
concentration (mg/L) and D. O. saturation (%).
Note
Incorrect calibration of D. O. probes will result in incorrect
measured values. Calibrate at regular intervals.
Temperature sensorThe D. O. probe has an integrated temperature sensor that always
measures the current temperature of the test sample.
26
Loading...
+ 59 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.