Xylem MULTI-1970I User Manual

ON / OFF
Multi 1970i
M
AUTO READ
RUN / ENTER
Multi 1970i
O
CAL
7
1 7 6
92
A
R
n
g
m
°
C
R
E
L
1
STO
Operating manual
g/
l
T
P
Multi 1970i
Portable pH / oxygen / conductivity measuring instrument
ba75693e01 01/2007
Accuracy when
going to press
The use of advanced technology and the high quality standard of our instruments are the result of a continuous development. This may re­sult in differences between this operating manual and your instrument. Also, we cannot guarantee that there are absolutely no errors in this manual. Therefore, we are sure you will understand that we cannot ac­cept any legal claims resulting from the data, figures or descriptions.
Copyright
© Weilheim 2007, WTW GmbH
Reproduction in whole - or even in part - is prohibited without the ex­press written permission of WTW GmbH, Weilheim. Printed in Germany.
Multi 1970i List of contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1 General features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Jack field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10
3 Commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Sensor quiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Operating structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 Switching on the measuring instrument . . . . . . . . . . . . . 16
4.3 pH value / ORP voltage . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3.1 General information . . . . . . . . . . . . . . . . . . . . . . 17
4.3.2 Measuring the pH value . . . . . . . . . . . . . . . . . . 19
4.3.3 Measuring the ORP voltage . . . . . . . . . . . . . . . 20
4.3.4 pH calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.4 Dissolved oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.4.1 General information . . . . . . . . . . . . . . . . . . . . . . 27
4.4.2 Measuring the D. O. concentration . . . . . . . . . . 29
4.4.3 Measuring the D. O. saturation . . . . . . . . . . . . . 30
4.4.4 AutoRead AR (Drift control) and hold function . 31
4.4.5 D. O. calibration . . . . . . . . . . . . . . . . . . . . . . . . 32
4.4.6 Entering the salt content (salinity) . . . . . . . . . . . 35
4.5 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.5.1 General information . . . . . . . . . . . . . . . . . . . . . . 36
4.5.2 Measuring the conductivity . . . . . . . . . . . . . . . . 38
4.5.3 Measuring the salinity . . . . . . . . . . . . . . . . . . . . 38
4.5.4 AutoRead AR (Drift control) and hold function . 39
4.5.5 Determining the cell constant (Calibration in the
control standard) . . . . . . . . . . . . . . . . . . . . . . . . 40
4.6 Calibration intervals (Int 3, Int 4, Int 5) . . . . . . . . . . . . . . 43
4.7 Saving data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
4.7.1 Saving manually . . . . . . . . . . . . . . . . . . . . . . . . 44
4.7.2 Switching on AutoStore (Int 1) . . . . . . . . . . . . . 46
4.7.3 Outputting the data storage . . . . . . . . . . . . . . . . 48
4.7.4 Clearing the memory . . . . . . . . . . . . . . . . . . . . . 54
4.8 Transmitting data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
1
List of contents Multi 1970i
4.8.1 Data transmission interval (Int 2) . . . . . . . . . . . . 55
4.8.2 PC/external printer (RS232 interface) . . . . . . . . 57
4.8.3 Remote control . . . . . . . . . . . . . . . . . . . . . . . . . 57
4.9 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4.10 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5 Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 65
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
5.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
5.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6.1 pH system messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6.2 Oxi system messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
6.3 Conductivity system messages . . . . . . . . . . . . . . . . . . . 70
6.4 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7.2 Measuring ranges, resolutions, accuracies . . . . . . . . . . 74
7.2.1 pH/ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
7.2.2 Dissolved oxygen . . . . . . . . . . . . . . . . . . . . . . . . 75
7.2.3 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
8 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
2
Multi 1970i Overview
RUN / ENTER
Multi 1970i

1Overview

1.1 General features

The Multi 1970i portable multiparameter measuring instrument enables you to carry out pH measurements, dissolved oxygen (D. O.) measure­ments and conductivity measurements quickly and reliably. The Multi 1970i provides the maximum degree of operating comfort, re­liability and measuring certainty for all applications.
The proven MultiCal
®
and OxiCal® calibration procedures and the pro­cedures to determine/set up the cell constant support you in your work with the meter. The special AutoRead function enables precise mea­surements.
1
ON / OFF
M
AUTO READ
RUN / ENTER
CAL
2
O
7
92
m
g/
l
1 7 6
°
C
T
P
R
E
L
1
A
R
n
g
STO
Multi 1970i
3
4
5
1 Keypad
2 Display
3 Integrated, exchangeable sensor quiver
4 Jack field
5 Carrying and positioning handle
3
Overview Multi 1970i
Note
If you need further information or application notes, you can obtain the following material from WTW:
z Application reports
z Primers
z Safety datasheets.
You will find information on available literature in the WTW catalog or via the Internet.

1.2 Display

Status display indicator
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
8
88
Baud
No.
Ident
8 8 8 8
Tref25 Tref20
Lin
Auto
Oxi
Cal
Store
TEC
ARng
%
m
1/
°
S/
cm
%
C
REL1
pHmV/
mg/l
cm
K
/
REL2
AR
Sal
TP
RCL
Sensor symbol
Measured value display
Function and temperature display
4
Multi 1970i Overview
RCL
STO

1.3 Keypad

Key functions
ON / OFF
M
CAL
AUTO READ
RUN / ENTER
Calibrate the currently set measured variable
Switch measuring instrument on/off
<ON/OFF>
Select the measured variable <M>:
pH value / ORP voltage
D. O. concentration / D. O. saturation
Conductivity / salinity
<CAL>
Activate/deactivate the AutoRead function
<AUTO READ>
Confirm entries, start AutoRead, output measured values
<RUN/ENTER>
Select the measuring mode, increase values, scroll
< >
Select the measuring mode, decrease values, scroll
<>
RCL
STO
Display/transmit measured values
<RCL>
Save a measured value
<STO>

1.4 Jack field

Warning
Only connect sensors to the measuring instrument that cannot return any voltages or currents that are not allowed (> SELV and > current cir­cuit with current limiting). Almost all sensors - in particular WTW sensors - fulfill these conditions.
5
Overview Multi 1970i
Maximum number
of sensors to be con-
nected
Connectors:
The following sensors can be connected to the Multi 1970i (maximum configuration):
z One pH depth armature or
one pH electrode (combination electrode or pH electrode + reference electrode). Please make sure that only one pH sensor is connected at the same time.
z One D. O. sensor
z One conductivity measuring cell
z Option: One external temperature sensor
Note
The pH depth armature is affected by other sensors if it is connected to them galvanically. For this reason, the pH depth armature cannot be operated in the same test sample together with another sensor on the instrument.
5
1
7
8
6
9
2
4
3
Sensor / Instrument Socket / Posi-
tion
pH depth armature 1
pH electrode or pH combination electrode 2
pH combination electrode with temperature sensor 2 and 4
Reference electrode 3
Temperature sensor, external 3 and 4
Oxygen sensor 5 or 6
Conductivity measuring cell 5 or 6
Printer or PC (serial interface, RS232) 7
Plug-in power supply unit 8
Watertight valve for internal pressure equalization 9
6
Multi 1970i Overview
Note
The sensor sockets, 1 and 2 must not be assigned at the same time. On the two sockets, 5 and 6, only different sensor types (Oxi and Cond) may be connected at the same time.
7
Overview Multi 1970i
8
Multi 1970i Safety

2 Safety

This operating manual contains basic instructions that you must follow during the commissioning, operation and maintenance of the measur­ing instrument. Consequently, all responsible personnel must read this operating manual before working with the measuring system. The op­erating manual must always be available within the vicinity of the mea­suring system.
Target group The measuring instrument was developed for work in the field and in
the laboratory. Thus, we assume that, as a result of their professional training and ex­perience, the operators will know the necessary safety precautions to take when handling chemicals.
Safety instructions The individual chapters of this operating manual use the following safe-
ty instruction to indicate various types of danger:
Warning
indicates instructions that must be followed precisely in order to avoid the possibility of slight injuries or damage to the instrument or the envi­ronment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manu­als.

2.1 Authorized use

The authorized use of the measuring instrument consists exclusively of the:
z pH and ORP measurement
z measurement of the oxygen content and
z measurement of the conductivity, salinity and temperature
in the field and laboratory.
The technical specifications as given in chapter 7 T must be observed. Only the operation and running of the measuring in­strument according to the instructions given in this operating manual is authorized. Any other use is considered to be unauthorized.
ECHNICAL DATA
9
Safety Multi 1970i

2.2 General safety instructions

This instrument is built and inspected according to the relevant guide­lines and norms for electronic measuring instruments (see chapter 7T
ECHNICAL DATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
Safe operation If safe operation is no longer possible, the instrument must be taken out
The smooth functioning and operational safety of the measuring instru­ment can only be guaranteed if the generally applicable safety mea­sures and the specific safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the measuring instru­ment can only be guaranteed under the environmental conditions that are specified in chapter 7 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty func­tioning of the instrument. In this event, wait until the temperature of the instrument reaches room temperature before putting the instrument back into operation.
of service and secured against inadvertent operation! Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
Obligations of the pur-
chaser
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the fol­lowing laws and guidelines are observed when using dangerous sub­stances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
10
Multi 1970i Commissioning

3 Commissioning

3.1 Scope of delivery

z Multi 1970i portable multiparameter measuring instrument with inte-
grated rechargeable battery
z Carrying and positioning handle
z Carrying strap
z 2 sensor quivers (pH and Oxi-LF type)
z Plug-in power supply unit
z Operating manual

3.2 Power supply

Mains operation and
charging the battery
Charging time of the
battery
You can either operate the measuring instrument with the integrated re­chargeable battery or with the plug-in power supply. The plug-in power supply supplies the measuring instrument with low voltage (12 V DC). At the same time, the rechargeable battery is charged.
approx. 16 hours. The battery is charged even when the instrument is switched off. The LoBat display indicator appears when the battery is nearly empty and has to be charged as soon as possible.
Warning
The line voltage at the operating site must lie within the input voltage range of the original plug-in power supply (see chapter 7 T
TA).
ECHNICAL DA-
Warning
Use original plug-in power supplies only (see chapter 7 T
ECHNICAL DATA).
11
Commissioning Multi 1970i
Connecting the plug-in
power supply unit
2
1
3
Setting the date and
time
1 Connect the plug (1) to the socket (2) of the measuring instru-
ment.
2 Connect the original WTW plug-in power supply (3) to an easily
accessible mains socket.

3.3 Initial commissioning

Perform the following activities:
z For mains operation and charging the battery: Connect the plug-in
power supply (see section 3.2 P
z Set the date and time.
1 Press the <M> key and hold it down.
2 Press the <ON/OFF> key.
The display test appears briefly on the display.
3 Press the <RUN/ENTER> key repeatedly until the date flashes
on the display (Day.Month display indicator).
OWER SUPPLY).
12
Multi 1970i Commissioning
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
0 9 0 4
Tref25
Tref20
nLF
Lin
5t2
Auto
Store
Cal
Oxi
TEC
ARng
4 Set the date of the current day with <> <>.
5 Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
6 Set the current month with <> <>.
7 Confirm with <RUN/ENTER>.
The year appears on the display.
8 Set the current year with <> <>.
9 Confirm with <RUN/ENTER>.
The hours flash on the display.
10 Set the current time with <> <>.
11 Confirm with <RUN/ENTER>.
The minutes flash on the display.
12 Set the current time with <> <>.
13 Confirm with <RUN/ENTER>.
The instrument switches to the measuring mode.
13
Commissioning Multi 1970i

3.4 Sensor quiver

To store the sensors during field operation and to keep the sensor ele­ment moist, the quiver tip contains a sponge rubber insert that can be moistened with deionized water.
Note
For further details on proper storage, refer to the operating manual of the sensor.
Moistening the
quiver insert
1 Press the quiver out of the holder from the back side of the in-
strument and pull it out completely.
2 Pull off the quiver tip and moisten the sponge rubber with
deionized water.
Warning
Do not store pH electrodes in the quiver for more than 10 hours. To store them for a longer period of time, use the watering cap filled with potassium chloride (3 mol/l) of the electrode.
14
Multi 1970i Operation

4 Operation

4.1 Operating structure

Active and inactive REL
socket
In addition to the pH/ORP sensor, the sensor on the REL1 or REL2 socket can be actuated (switched "active"). The other socket is not ac­tuated ("inactive"). This switching over inside the instrument is made with the aid of a relay. The active REL socket is shown on the display.
The measured variable on the display can be selected as follows using the <M> key:
z Press <M> shortly
z Press <M> for a longer
period of time:
Several measuring modes are available within a measured variable. Switch over between the measured variables with the <> or <> key.
The options are summarized in the following diagram:
: Switch between the pH/ORP sensor and
the sensor at the active REL socket
Change the active Rel socket
15
Operation Multi 1970i
Note
When a sensor is connected to the active REL socket, the measuring instrument recognizes the sensor or the measuring cell and automati­cally switches to the measuring mode that was last active. As soon as the sensor is disconnected from the active REL socket, the instrument switches to the pH (mV) measuring mode again.

4.2 Switching on the measuring instrument

1 Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the measuring instrument automatically switches to the measuring mode.
Measuring mode when
switching on
Sensors on the socket Measuring mode
pH REL
(active)
REL
(inactive)
:
pH or ORP mea­surement - depend­ing on the last selected setting
:
::
:
::
:::
Note
The measuring instrument has an energy saving feature to avoid un­necessary battery depletion. The energy saving feature switches the measuring instrument off if no key has been pressed for an hour.
Last selected mea­suring mode. Only the last active REL socket is tak­en into account
16
The energy saving feature is not active
z if the power is supplied by the plug-in power supply,
z if the AutoStore function is active,
z if the communication cable and a PC with a running communication
program are connected,
z if the printer cable is connected (for external printers).
Multi 1970i Operation

4.3 pH value / ORP voltage

4.3.1 General information

Preparatory activities Perform the following preparatory activities when you want to measure:
1 Connect the pH depth armature or the pH electrode to the mea-
suring instrument. If necessary, press the <M> key repeatedly until the pH (pH measurement) or U (measurement of the ORP voltage) display appears.
2 Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a temperature sensor.
3 Calibrate or check the measuring instrument with the elec-
trode.
4 Using <> <>, toggle between the pH or mV measuring
modes.
Note
Incorrect calibration of pH electrodes leads to incorrect measured val­ues. Calibrate regularly before measuring. You can only connect elec­trodes of the NTC30 type or without temperature sensor.
Warning
When connecting an earthed PC/printer, measurements cannot be per­formed in earthed media as incorrect values would result. The RS232 interface is not galvanically isolated.
17
Operation Multi 1970i
Temperature measure-
ment in pH measure-
ments
You can perform pH measurements with or without a temperature sen­sor as well as with the temperature sensor of an oxygen sensor or a conductivity measuring cell. The measuring instrument recognizes which sensors are connected and switches automatically to the correct mode for the temperature measurement.
The following cases are distinguishable.
Temperature sensor Display Mode
pH Cond or
Oxi
yes - TP Automatic with
pH temperature sensor
yes yes TP
- - Manual
-yesTP flashes
The temperature value of the second sensor (Cond or Oxi) in the same sample is taken over for the pH measurement*
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature
input or
– use an electrode with a temperature sensor.
If a temperature sensor is connected, it is indicated on the display by TP.
Note
When calibrating without a temperature sensor (no TP display indicator displayed), enter the current temperature of the respective buffer solu­tion manually using the <> <> keys while keeping the <RUN/EN-
TER> key depressed.
18
Multi 1970i Operation

4.3.2 Measuring the pH value

1 Perform the preparatory activities according to section 4.3.1.
2 Immerse the pH electrode in the test sample.
3 Press the <> <> keys repeatedly until pH appears on the
status display. The pH value appears on the display.
AutoRead AR
(Drift control) and hold
function
m
%
1/
°
mg/l
S/
cm
cm
%
C
REL1
AR
pH/mV
Sal
TP
K
/
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
4 When measuring without a connected temperature sensor:
The AutoRead function (drift control) checks the stability of the mea­surement signal. The stability has a considerable impact on the repro­ducibility of the measured values. With the aid of the hold function the measured value display is frozen.
6
1
Baud
No.
Ident
Tref25 Tref20
Options:
z Determine the current temperature using a thermometer
z TP display indicator not displayed, socket for the second
99
8
2 4 8
Auto
Store
Lin
Oxi
Cal
TEC
ARng
and, while keeping the <RUN/ENTER> key depressed, en­ter this temperature value with <> <>.
sensor is free: Connect the second sensor (Oxi or Cond) and immerse it in the same sample. TP flashes, the temperature is automati­cally measured using the second sensor.
With identical measurement conditions, the following criterion is valid for the AutoRead function:
Reproducibility Response time
Better than 0.02 > 30 seconds
For D. O. measurements, use the AutoRead function and hold function like this:
1 Call up the pH measuring mode with <> <>.
19
Operation Multi 1970i
2 Activate the AutoRead function with <AUTO READ>.
The current measured value is frozen (hold function).
3 Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This measured value is transmitted to the interface.
4 If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5 To terminate the AutoRead function: Press the
<AUTO READ> key.
Note
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.

4.3.3 Measuring the ORP voltage

In conjunction with an ORP electrode, e.g. SenTix ORP, the measuring instrument can measure the ORP voltage (U) of a solution.
1 Perform the preparatory activities according to section 4.3.1.
2 Submerse the ORP electrode in the sample.
3 Press the <> <> key until the U status display appears. The
ORP voltage (mV) of the test sample appears on the display.
4 Wait for a stable measured value.
pH/
Store
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
Note
ORP electrodes are not calibrated. However, you can check ORP elec­trodes using a test solution.
1
Baud
Ident
Tref25
No.
6
8
Tref20
Lin
Oxi
23
2 4 8
Auto
Cal
20
Multi 1970i Operation

4.3.4 pH calibration

Why calibrate? pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is dis­played. Calibration determines the current values of the asymmetry and slope of the electrode and stores them in the measuring instru­ment. Thus, you should calibrate at regular intervals.
When to calibrate? z After connecting another electrode
z When the sensor symbol flashes, i.e. after the calibration interval
has expired
Calibration points Calibration can be made with one or two buffer solutions (single-point
or two-point calibration). The measuring instrument determines the fol­lowing values and calculates the calibration lines as follows:
Determined values Values of the calibration lines
1-point ASY z Asymmetry = ASY
z Slope = Nernst slope (59.16 mV/
pH at 25 °C)
2-point ASY
SLO
z Asymmetry = ASY
z Slope = SLO
AutoCal TEC is specially matched to the WTW technical buffer solutions as a fully au-
tomatic two-point calibration. The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see section 4.9 C
ONFIGURATION), the instrument displays the
relevant buffer nominal value or the current electrode voltage in mV. The calibration can be terminated after the first buffer solution. This cor­responds to a single-point calibration. To do this, the instrument uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymme­try of the electrode.
AutoRead The calibration procedure automatically activates the AutoRead func-
tion. The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>.
Displaying the calibra-
tion data
Printing the
calibration protocol
You can view the data of the last calibration on the display. The pro­ceeding is described on page 50.
The calibration protocol contains the calibration data of the current cal­ibration. You can transmit the calibration protocol to a printer via the se­rial interface (see O
INTERFACE, page 53).
UTPUTTING THE CALIBRATION PROTOCOL ON THE
21
Operation Multi 1970i
Note
You can automatically print a calibration protocol after the calibration. To do so, connect a printer to the interface according to section 4.8.2 before calibrating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.02 14:19 Device No.: 12345678 Calibration pH Cal time: 01.03.01 / 15:20 Cal interval: 7d AutoCal TEC Tauto Buffer 1 2.00 Buffer 2 4.01 Buffer 3 7.00 * Buffer 4 10.01 C1 184.1 mV 25.0°C C2 3.0 mV 25.0°C S1 -59.4 mV/pH ASY1 - 4 mV Probe: +++
22
Multi 1970i Operation
Calibration evaluation After calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The worst evaluation appears on the display.
Display Asymmetry
[mV]
-15 ... +15 -60.5 ... -58
-20 ... +20 -58 ... -57
-25 ... +25 -61 ... -60.5
-30 ... +30 -62 ... -61
Clean the electrode according to the electrode operating manual
Slope [mV/pH]
or
-57 ... -56
or
-56 ... -50
Preparatory activities
E3
Perform error elimination ac­cording to chapter 6 W
DO IF...
HAT TO
< -30 or > 30
< -62 or > -50
1 Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis­play pH (pH measurement) or U (measurement of the ORP voltage) appears.
2 Keep the buffer solutions ready.
3 Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a temperature sensor (the TP display indicator is missing from the display).
23
Operation Multi 1970i
AutoCal TEC For this procedure, use any two WTW technical buffer solutions (pH
values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the WTW technical buffer solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions can lead to an erroneous calibration. The correct buffer solutions are given in the WTW catalog or in the Internet.
Note
The buffer solutions are automatically recognized by the measuring in­strument. Depending on the instrument setting (see section 4.9 C er nominal value or the current electrode voltage in mV.
Note
Skip the steps 2 and 7 if you use a pH electrode with temperature sen­sor or the temperature sensor of a conductivity measuring cell or a D. O. sensor.
ONFIGURATION), the instrument displays the relevant buff-
Starting the calibration
1 Press the <CAL> key. The Ct1 display and the function display
AutoCal TEC appears. The sensor symbol displays the evalu-
ation of the last calibration (or no sensor symbol in the delivery state or after the measurement parameters have been reset).
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
Sal
TP
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
2 If required, enter the temperature of the first buffer solution with
3 Immerse the pH electrode in the first buffer solution.
4 Press the <RUN/ENTER> key.
C
Lin
8
Oxi
T1
2 4 8
Store
Auto
TEC
ARng
Cal
1
Baud
No.
Ident
Tref25 Tref20
<> <> while keeping the <RUN/ENTER> key depressed.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. Example:
24
Multi 1970i Operation
mVS/
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
REL2
Sal
TP
RCL
/
pH
3
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
/
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
5 When the measured value is stable, Ct2 appears.
1
Baud
No.
Ident
Tref25 Tref20
Lin
69
8
2 4 8
Auto
Oxi
Cal
K
Sal
TP
REL2
RCL
pH
O
Sal
1
Time
Day.Month
Year
LoBat nLF
700
Baud
No.
Ident
8
Lin
Oxi
2 4 8
Auto
Cal
Tref25 Tref20
TEC
ARng
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
Note
At this point, the AutoCal TEC calibration can be terminated with <M>. This corresponds to a single-point calibration. To do this, the instru- ment uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
6 Thoroughly rinse the electrode with distilled water.
7 If required, enter the temperature of the second buffer solution
8 Immerse the pH electrode into the second buffer solution.
9 Press the <RUN/ENTER> key.
C
Lin
8
Oxi
T2
2 4 8
Store
Auto
TEC
ARng
Cal
1
Baud
No.
Ident
Tref25 Tref20
with <> <> while keeping the <RUN/ENTER> key de­pressed.
The AR display indicator flashes. The electrode voltage (mV) or the buffer nominal value ap­pears on the display. Example:
25
Operation Multi 1970i
mVS/
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
1
Time
Day.Month
Year
LoBat nLF
S
Baud
No.
Ident
Tref25 Tref20
Lin
8
2 4 8
Auto
Oxi
Cal
TEC
ARng
mV
%
m
1/
°
/
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
1
Time
Day.Month
Year
LoBat nLF
401184
Baud
No.
Ident
8
Tref25 Tref20
Lin
Oxi
2 4 8
Auto
TEC
ARng
Cal
10 When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The sensor sym­bol shows the evaluation of the current calibration.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
594
Baud
No.
Ident
Tref25 Tref20
Lin
8
Oxi
2 4 8
Auto
Cal
Store
TEC
ARng
11 Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
pH/
2
Store
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
59
Baud
No.
Ident
8
Tref25 Tref20
Lin
2 4 8
Auto
Oxi
Cal
12 Switch to the measuring mode with <M>.
26
Multi 1970i Operation

4.4 Dissolved oxygen

4.4.1 General information

Note
D. O. measurements with the Multi 1970i can only be carried out using a CellOx 325 or StirrOx G D. O. sensor. The stirrer of the StirrOx G D. O. sensor has to be supplied with voltage separately using the NT/pH Mix 540 power supply.
You can measure the following variables:
z D. O. concentration
z Oxygen saturation
The measuring instrument is supplied with the following functions:
z AutoRange (automatic switchover of the measurement range). If a
measuring range is exceeded, AutoRange causes the measuring in­strument to change automatically to the next higher measuring range and back again. Therefore, the instrument always measures in the measuring range with the highest possible resolution.
z The AutoRead function (drift control) for checking the stability of the
measurement signal. This ensures the reproducibility of the measur­ing signal. For details of how to switch the AutoRead function on/off, see page 31.
Warning
When connecting an earthed PC/printer, measurements cannot be per­formed in earthed media as incorrect values would result. The RS232 interface is not galvanically isolated.
Preparatory activities Perform the following preparatory activities when you want to measure:
1 Connect the D. O. sensor to the measuring instrument. If the D.
O. sensor is connected to the active REL socket, press the <M> repeatedly if necessary until the O
status display indica-
2
tor appears. If the D. O. sensor is connected to the inactive REL socket, press the <M> key for 2 s, until the O
status dis-
2
play indicator appears.
2 Calibrate or check the measuring instrument with the sensor.
How to calibrate is described in section 4.4.5 from page 21.
3 Use <> <> to toggle between the measuring modes, D. O.
concentration (mg/L) and D. O. saturation (%).
27
Operation Multi 1970i
Note
Incorrect calibration of D. O. probes will result in incorrect measured values. Calibrate at regular intervals.
Temperature sensor The D. O. sensor has an integrated temperature sensor that always
measures the current temperature of the test sample.
28
Multi 1970i Operation

4.4.2 Measuring the D. O. concentration

When measuring the concentration of test samples with a salt content of more than 1 g/l, a salinity correction is required.
Note
How to enter the current salt content is described in section 4.4.6 E Switching the salt content correction on or off, see below.
To measure the D. O. concentration with and without salt content cor­rection, proceed as follows:
1 Perform the preparatory activities according to section 4.4.1.
2 Immerse the D. O. sensor in the test sample.
3 Press the <> <> key repeatedly until the D. O. concentra-
tion in mg/l appears on the display.
NTERING THE SALT CONTENT (SALINITY) on page 35.
Switching on/off the salt
content correction
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
Proceed as follows to switch on the salt content correction:
1 Perform the preparatory activities according to section 4.4.1.
2 Immerse the D. O. sensor in the test sample.
3 While pressing the <RUN/ENTER> key, switch on the salt con-
7
Lin
8
Oxi
92
1 7 6
Auto
Store
Cal
TEC
ARng
1
Baud
No.
Ident
Tref25 Tref20
tent correction with <> . The SAL display indicator appears on the display. The specified salt content is taken into consid­eration during the measurement.
29
Operation Multi 1970i
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
4 While pressing the <RUN/ENTER> key, switch off the salt con-

4.4.3 Measuring the D. O. saturation

You can measure the D. O. saturation as follows:
1 Perform the preparatory activities according to section 4.4.1.
2 Immerse the D. O. sensor in the test sample.
5
Lin
8
Oxi
42
1 7 6
Auto
Store
Cal
TEC
ARng
1
Baud
No.
Ident
Tref25 Tref20
tent correction with <>. The SAL display indicator is no longer displayed.
3 Press the <> <> key repeatedly until the D. O. saturation in
% appears on the display.
/mV
%
m
1/
°
S/
C
pH
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
8
Baud
No.
Ident
8
Tref25 Tref20
Lin
92
1 7 6
Auto
Oxi
Cal
Store
TEC
ARng
30
Multi 1970i Operation

4.4.4 AutoRead AR (Drift control) and hold function

The AutoRead function (drift control) checks the stability of the mea­surement signal. The stability has a considerable impact on the repro­ducibility of the measured values. With the aid of the hold function the measured value display is frozen.
Criteria With identical measurement conditions, the following criteria are valid
for the AutoRead function:
Measuring mode Reproducibility Response time
D. O. concentra­tion
Oxygen saturation index
For D. O. measurements, use the AutoRead function and hold function like this:
1 Call up the measuring mode with <M> and/or <> <> .
2 Activate the AutoRead function with <AUTO READ>. The cur-
rent measured value is frozen (hold function).
3 Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This mea­sured value is transmitted to the interface.
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
1
Baud
No.
Ident
Tref25 Tref20
Lin
7
8
1 7 6
Oxi
08
Auto
better than 0.05 mg/l > 10 seconds
better than 0.6 % > 10 seconds
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
ARng
C
REL1
AR
REL2
RCL
Store
Cal
TEC
4 If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5 To terminate the AutoRead function: Press the
<AUTO READ> key.
31
Operation Multi 1970i

4.4.5 D. O. calibration

Why calibrate? D. O. probes age. This changes the slope of the D. O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in the instrument.
When to calibrate? z After connecting another D. O. sensor
z When the sensor symbol flashes (after the calibration interval has
expired).
Calibration procedure The calibration is performed in water vapor-saturated air. Use the
®
OxiCal
-SL air calibration vessel (accessory) for the calibration.
AutoRead The calibration procedure automatically activates the AutoRead func-
tion. The AR display indicator flashes. The calibration process is fin­ished when AR stops flashing.
Displaying the calibra-
tion data
Printing the
calibration protocol
You can view the data of the last calibration on the display. The pro­ceeding is described on page 50.
The calibration protocol contains the calibration data of the current cal­ibration. You can transmit the calibration protocol to a printer via the se­rial interface (see page 53).
Note
You can automatically print a calibration protocol after the calibration. To do so, connect a printer to the interface according to section 4.8.2 before calibrating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.02 14:19 Device No.: 12345678 CALIBRATION 02 Cal time: 02.03.01 / 14:19 Cal interval: 14d OxiCal Tauto AR Relative Slope: 0,88 Probe: +++
32
Multi 1970i Operation
Sensor evaluation After the calibration, the measuring instrument evaluates the current
status of the sensor against the relative slope. The evaluation appears on the display. The relative slope has no effect on the measuring accu­racy. Low values indicate that the electrolyte will soon be depleted and the sensor will have to be regenerated.
Display Relative slope
S = 0.8 ... 1.25
S = 0.7 ... 0.8
S = 0.6 ... 0.7
E3
S < 0.6 or S > 1.25
Perform error elimination accord­ing to chapter 6 W
HAT TO DO IF...
Starting the calibration Proceed as follows to calibrate the instrument:
1 Perform the preparatory activities according to section 4.4.1.
®
2 Keep the OxiCal
-SL air calibration vessel ready.
Note
The sponge in the air calibration vessel must be moist (not wet). Ob­serve the instructions in the OxiCal
®
-SL operating manual.
3 Put the D. O. sensor into the air calibration vessel.
4 Press the <CAL> key repeatedly until the calibration mode ap-
pears. The sensor symbol displays the evaluation of the last calibration (or no sensor symbol in the delivery state or after the measuring parameters have been reset).
33
Operation Multi 1970i
m
m
%
1/
°
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
pH/mV
cm
K
/
REL2
Sal
TP
RCL
Sal
TP
RCL
pH
S
O
Sal
Time Baud
Day.Month
Year
LoBat
5 Press the <RUN/ENTER> key. AutoRead is active, AR flashes.
pH
1
No.
Ident
Tref25 Tref20
Lin
nLF
S
C
8
1 7 6
Oxi
AL
Auto
Store
ARng
TEC
Cal
O
Sal
Time Baud
Day.Month
Year
LoBat
6 As soon as a stable value is achieved the AR display stops
0
Lin
8
Oxi
88
1 7 6
Auto
Store
ARng
TEC
Cal
1
No.
Ident
Tref25 Tref20
nLF
flashing. The calibration is finished then. The value of the rela­tive slope appears on the display. The sensor symbol shows the sensor evaluation (see page 33).
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
Time Baud
Day.Month
Year
LoBat
7 Switch to the measuring mode with <M>.
Note
In chapter 6 W to take for error elimination.
1
No.
Ident
Tref25 Tref20
Lin
nLF
0
88
8
1 7 6
Auto
Store
ARng
Oxi
TEC
Cal
HAT TO DO IF... from page 67, you will find the measures
34
Multi 1970i Operation

4.4.6 Entering the salt content (salinity)

A salt content correction is required in the oxygen concentration mea­surement of samples with a salt content of more than 1 g/l. To do this, you have to enter the salinity equivalent (the measured salinity) of the test sample (range 0.0 - 70.0) and to switch on the salinity correction.
Note
With the Multi 1970i, you can measure the salinity. How to proceed is described in section 4.5.3 M
EASURING THE SALINITY on page 38.
Entering the
salt content
1 Determine the salinity of the test sample (any method, see also
section 4.5.3 M
2 Press the <CAL> key repeatedly until Sal appears on the dis-
play.
pH
S
O
Sal
Time Baud
Day.Month
Year
LoBat
3 Enter the salt content with <> <>.
4 Switch to the measuring mode with <M>.
Note
How to switch on the salt content correction is described on page 29.
1
No.
Ident
Tref25 Tref20
Lin
nLF
5
8
6 9 6
Oxi
50
Auto
Cal
EASURING THE SALINITY on page 38).
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
ARng
C
REL1
AR
REL2
RCL
Store
TEC
35
Operation Multi 1970i

4.5 Conductivity

4.5.1 General information

Note
Conductivity measurements with the Multi 1970i can only be carried out using the TetraCon 325 measuring cell.
Warning
When connecting an earthed PC/printer, measurements cannot be per­formed in earthed media as incorrect values would result. The RS232 interface is not galvanically isolated.
The measuring instrument is supplied with the following functions:
z AutoRange (automatic switchover of the measurement range). If a
measuring range is exceeded, AutoRange causes the measuring in­strument to change automatically to the next higher measuring range and back again. Therefore, the instrument always measures in the measuring range with the highest possible resolution.
z The AutoRead function (drift control) for checking the stability of the
measurement signal. This ensures the reproducibility of the measur­ing signal. For details of how to switch the AutoRead function on/off, see page 39.
Preparatory activities Perform the following preparatory activities when you want to measure:
1 Connect a conductivity measuring cell to the measuring instru-
ment. If the conductivity measuring cell is connected to the ac­tive REL socket, press the <M> key repeatedly (if necessary) until the status display appears. If the conductivity measur­ing cell is connected to the inactive REL socket, press the <M> key for 2 s until the status display appears.
2 Check the selected cell constant or calibrate the measuring in-
strument with the measuring cell (see below).
3 Check the selected cell constant (see below) or calibrate the
measuring instrument with the measuring cell (see section 4.5.5 on page 40).
4Using <> <>, toggle between the measuring modes, con-
ductivity ( in µS/cm) or salinity (SAL).
Temperature sensor The TetraCon 325 conductivity measuring cell has a temperature sen-
sor integrated in it. The temperature sensor is shown on the display by TP.
36
Multi 1970i Operation
Temperature
compensation
Reference temperature,
Tref
Checking the
cell constant
The nonlinear temperature compensation is set fixed and is shown on the display by nLF.
The reference temperature (Tref) can be switched between 20 °C and 25 °C. It appears on the display as Tref20 or Tref25. To switch over the reference temperature, see S
TURE, page 60.
1 Press the <RCL> key repeatedly until CAL disp appears on the
display.
pH
S
O
Sal
Time Baud
Day.Month
Year
LoBat
C
No.
Ident
Tref25 Tref20
Lin
nLF
C
I S P
d
Oxi
AL
Auto
Store
TEC
ARng
Cal
WITCHING OVER THE REFERENCE TEMPERA-
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
C
REL2
REL1
AR
RCL
2 Press the <RUN/ENTER> repeatedly until the last calibrated
cell constant is displayed, e. g. 0.472 1/cm.
pH
S
O
Sal
C
Time Baud
Day.Month
LoBat
No.
Year
Ident
0
Tref20
Tref25
Oxi
Lin
nLF
3 To return to the measuring mode: Press the <M> key when the
correct cell constant is displayed.
4 If you want to recalibrate the cell constant, proceed according
to section 4.5.5 D
IN THE CONTROL STANDARD) .
LC
C
4 7 5
Auto
Store
ARng
TEC
Cal
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
C
REL1
REL2
AR
RCL
ETERMINING THE CELL CONSTANT (CALIBRATION
37
Operation Multi 1970i

4.5.2 Measuring the conductivity

You can carry out the conductivity measurements as follows:
1 Perform the preparatory activities according to section 4.5.1
page 36.
2 Immerse the conductivity measuring cell in the test sample.
3 Press the < > <> keys until in the status display, and the
unit µS/cm appears. The conductivity value appears on the dis­play.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
Time Baud
Day.Month
Year
LoBat
Ident
Tref25
nLF
05 3
No.
0
2 4 0
Tref20
Auto
Store
Cal
Oxi
Lin
TEC
ARng

4.5.3 Measuring the salinity

You can carry out the salinity measurements as follows:
1 Perform the preparatory activities according to section 4.5.1
page 36.
2 Immerse the conductivity measuring cell in the test sample.
3 Press the <> <> keys repeatedly until the Sal status display
appears. The salinity value appears on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
Time Baud
Day.Month
Year
LoBat
Ident
Tref25
nLF
332
No.
0
2 1 2
Tref20
Auto
Store
TEC
ARng
Cal
Oxi
Lin
38
Multi 1970i Operation

4.5.4 AutoRead AR (Drift control) and hold function

The AutoRead function (drift control) checks the stability of the mea­surement signal. The stability has a considerable impact on the repro­ducibility of the measured values. With the aid of the hold function the measured value display is frozen.
For conductivity measurements, use the AutoRead function and hold function like this:
1 Call up the measuring mode or SAL with <M> and/or <>
<>.
2 Immerse the conductivity measuring cell in the test sample.
3 Activate the AutoRead function with <AUTO READ>. The cur-
rent measured value is frozen (hold function).
4 Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This measured value is transmitted to the interface.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
pH/mV
cm
Sal
K
/
REL2
RCL
pH
S
O
Sal
C
Time Baud
Day.Month
Year
LoBat
Ident
Tref25
nLF
05 3
No.
0
2 5 0
Tref20
Auto
Store
Cal
Oxi
Lin
TEC
ARngTPAR
5 If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
6 To terminate AutoRead: Press the <AUTO READ> key.
Note
The current AutoRead measurement can be terminated at any time (accepting the current value) by pressing <RUN/ENTER>. You can only change to another measuring mode after completion of AutoRead.
39
Operation Multi 1970i

4.5.5 Determining the cell constant (Calibration in the control standard)

Why determine the cell
constant?
AutoRead In calibration, the AutoRead function is automatically activated. The AR
Displaying the
calibration data
Printing the
calibration protocol
Aging slightly changes the cell constant, e. g. by coatings. As a result, an inexact measured value is displayed. The original characteristics of the cell can often be restored by cleaning the cell. Calibration deter­mines the current value of the cell constant and stores this value in the instrument. Thus, you should calibrate at regular intervals.
In the delivery condition, the cell constant of the measuring instrument is set to 0.475 cm
-1
(conductivity measuring cell TetraCon 325).
display indicator flashes. The calibration process is finished when AR stops flashing.
You can view the data of the last calibration on the display. The pro­ceeding is described on page 50.
The calibration protocol contains the calibration data of the current cal­ibration. You can transmit the calibration protocol to a printer via the se­rial interface (see page 53).
Note
You can automatically print a calibration protocol after the calibration. To do so, connect a printer to the interface according to section 4.8.2 before calibrating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
14.04.02 11:37 Device No.: 99990000 Calibration Conductivity Cal time: 14.04.01 / 11:37 Cal interval: 180d Cal Std.: 0.01 mol/l KCL
40.0 °C Conduct./Tref25: 1413µS/cm Cell Const : 0.478 1/cm Probe : +++
40
Multi 1970i Operation
Calibration evaluation After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis­play.
-1
Display Cell constant [cm
0.450 ... 0.500 cm
]
-1
Determining the
cell constant
E3
Perform error elimination ac­cording to chapter 6 W
IF...
HAT TO DO
Outside the range
0.450 ... 0.500 cm
-1
You can determine the cell constant (conductivity measuring cell TetraCon 325) as follows:
1 Press the <CAL> key repeatedly until the status display
CAL appears.
2 Press the <RUN/ENTER> key. The CAL display appears, as
well as – the current, calibrated cell constant (with sensor symbol on
the display) or
– the fixed cell constant 0.475 1/cm (without sensor symbol
on the display). In this case, the measurement parameters are initialized. See “Reset” on page 62..
/pH
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
Baud
No.
Ident
0 4 7 2
Tref20
Tref25
Lin
Oxi
AC
Auto
Cal
L
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
cm
%
C
REL1
AR
pH
O
Sal
Time
Sal
REL2
RCL
Day.Month
Year
LoBat nLF
K
/
CAL
1
Baud
No.
Ident
0
Tref20
Tref25
Lin
Oxi
4 7 5
Auto
Cal
TEC
ARng
mVS/
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
Sal
TPTP
REL2
RCL
3 Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
41
Operation Multi 1970i
4 Press the <RUN/ENTER> key.
The AutoRead measurement to determine the cell constant starts. The AR display indicator flashes until a stable signal is reached. The cell constant determined is displayed. The mea­suring instrument automatically stores the cell constant.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
Time Baud
Day.Month
Year
LoBat
Ident
Tref25
nLF
LCA
No.
4 7 3
0
Tref20
Auto
Store
Oxi
Lin
Cal
TEC
ARng
5 To return to the measuring mode: Press the <M> key. The de-
termined cell constant is taken over for the measurement.
Note
If error message
appears refer to chapter 6 W
E3
E3
E3E3
HAT TO DO IF...
42
Multi 1970i Operation
4.6 Calibration intervals (Int 3, Int 4, Int 5)
For each measured variable, a time interval is stored. When it has ex­pired, you will be reminded to calibrate. After a calibration interval has expired, the sensor symbol of the relevant measured variable flashes. It is still possible to measure. By calibrating the relevant sensor, the function is reset and the interval starts anew.
The following calibration intervals are set in the factory:
Setting the calibration
interval
Measured
Designation Default setting
parameter
pH/ORP Int 3 7 days
Dissolved
Int 4 14 days
oxygen
Conductivity Int 5 180 days
You can change each of these intervals (1 ... 999 days):
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the measuring instrument automatically switches over to configu­ration.
4 Press the <RUN/ENTER> key repeatedly, until Int 3 ... 5 to-
gether with the required measured variable (pH, O
or ) ap-
2
pears. Example:
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
4nt
No.
4
1 4
Auto
Store
Oxi
Cal
TEC
ARng
Lin
d
Tref20
5 Set the required interval until the next calibration with <>
<>.
6 Confirm with <RUN/ENTER>.
7 Switch to the measuring mode with <M>.
43
Operation Multi 1970i

4.7 Saving data

The portable Multi 1970i multiparameter measuring instrument has an internal data storage. It can store up to 500 datasets. A complete data record consists of:
z Number of the storage location
z Date/time
z Measured values of the connected and active sensors
z Temperature values of the connected and active sensors
z Temperature measuring procedure
z ID number
You can transmit measured values (data records) to the data storage in two ways:
z Save manually
z Switch on AutoStore (Int 1), see page 46

4.7.1 Saving manually

You can transmit a measured value to the data storage as follows:
1 Press the <STO> key.
The current number (location number No.) of the next free stor­age location appears under the current measured value on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
d 1
4
Tref20
Oxi
Lin
Auto
Cal
Store
TEC
969
4
ARng
2 Confirm with <RUN/ENTER>.
The display switches to entering the ID number.
44
Multi 1970i Operation
Message StoFull
StoFull This message appears when all of the 500 storage locations are occu-
StoFullStoFull
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
Tref25
nLF
d 1
4
Tref20
Auto
Cal
Oxi
Lin
Store
TEC
969
2
ARng
3 Using <> <>, enter the required ID number
(1 ... 999).
4 Confirm with <RUN/ENTER>.
The measured values are stored. The instrument changes to the measuring mode.
pied.
You have the following options:
Saving the current measured value. The oldest measured value (storage location 1) will be overwritten by this
Press
<RUN/ENTER>
Returning to the measuring mode without saving press any key
Outputting the data storage see
section 4.7.3
Clearing the memory see
section 4.7.4
45
Operation Multi 1970i

4.7.2 Switching on AutoStore (Int 1)

The save interval (Int 1) determines the chronological interval between automatic save processes. After the fixed interval has expired, the cur­rent data record is transmitted to the internal storage and to the inter­face.
Switching on AutoStore
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <STO> key. Int 1 appears on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
1nt
No.
Lin
d
Tref20
Auto
Cal
Oxi
F
FO
Store
ARng
TEC
3 Set the required interval between the saving procedures with
<> <> (selection: 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min).
4 Confirm with <RUN/ENTER>.
The number of free memory locations appears on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
949
No.
E
ER
Store
Auto
Cal
Oxi
TEC
ARng
Lin
F
Tref20
46
5 Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.
Multi 1970i Operation
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
949
No.
ERF
Store
TEC
1
ARng
Lin
Tref20
Auto
Cal
Oxi
6 Set the required ID number with <> <>.
7 Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measur­ing mode and start the measuring and saving procedure. Au- toStore flashes on the display.
As soon as all of the 500 storage locations are occupied, AutoStore is terminated (Int 1 = OFF). If there are not enough storage locations available for your measurements:
z Output and backup the data storage (see page 48) and
z clear the memory (see page 54).
Note
The AutoStore function is interrupted if you start other functions, e.g. output the data storage. After the other function is finished, the Au­toStore function is continued. By this, however, temporal gaps in the re­cording of the measured values will occur.
Switching off AutoStore Switch AutoStore off by:
z setting the save interval (Int 1) to OFF, or
z switching the measuring instrument off and then on again.
47
Operation Multi 1970i

4.7.3 Outputting the data storage

You can output the contents of the data storage:
z Stored data on the display
z Calibration data on the display
z Stored data on the serial interface
z Calibration protocol on the interface
Outputting stored data
on the display
1 Press the <RCL> key repeatedly until StO dISP appears on the
display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
oST
No.
SID
P
Store
Auto
Tref20
Cal
Oxi
Lin
TEC
ARng
2 Press the <RUN/ENTER> key.
A measured value appears on the display. The storage location of the data record is displayed for approx. 2 s, then the respective temperature appears.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
6
Baud
No.
Ident
8
Tref25 Tref20
Lin
03
2 5 0
Auto
Store
Oxi
Cal
TEC
ARng
48
You can perform the following activities:
Display further data of the data record
Press <RUN/ENTER>
(ID number, date, time, storage location)
Toggle between two saved measured vari­ables
Press <RUN/EN-
TER> + <M>
Advance one data record (storage location) Press <>
Go back one data record (storage location) Press <>
Multi 1970i Operation
Note
If you want to search for a certain element of the data record (e.g. date), proceed as follows:
1 Using <RUN/ENTER>, select the element (e.g. date).
2 Press <> or <> repeatedly until the required element ap-
pears on the display. After approx. 2 s the temperature of the displayed measured value appears.
Outputting stored data
to the interface
1 Press the <RCL> key repeatedly until Sto SEr appears on the
display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
oST
No.
Lin
D
Tref20
ES
r
Store
Auto
Cal
Oxi
TEC
ARng
2 Press the <RUN/ENTER> key.
The complete contents of the storage are transmitted to the in­terface. During the data transmission the instrument incre­ments the storage numbers. After the data transmission, the instrument automatically switches to the last active measure­ment mode.
The transmitted data contains the entire contents of the storage in in­crementing order of the location numbers.
Note You can cancel the transmission with <M> or <RUN/ENTER>.
49
Operation Multi 1970i
Sample printout:
No. 1:
09.03.02 17:10 pH 10.01 25 °C Tauto AR Ident : 47
No. 2:
09.03.02 17:12 305 mV Tauto Ident : 6
No. 3:
09.03.02 17:24
7.88 mg/l 17.6° C Tauto Ident : 81
No. 4:
09.03.02 17:46
7.11 mg/l 17.8° C Tauto SAL = 17.9 Ident : 4
Outputting the calibra-
tion data
on the display
No. 5:
10.03.02 19:09
2.40 mS/cm 25.3 °C Tauto nLF Tref25 C = 0.475 1/cm Ident : 10
No. 6:
10.03.02 20:48
2.46 mS/cm 25.6 °C Tauto nLF Tref25 C = 0.475 1/cm Ident : 1
...
1 Press the <RCL> key repeatedly until CAL disp appears on
the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
Time Baud
Day.Month
Year
LoBat
C
No.
Ident
d
Tref25 Tref20
Lin
nLF
AL
I S P
Auto
Oxi
Cal
Store
TEC
ARng
50
Multi 1970i Operation
2 Press the <RUN/ENTER> key.
The data of the last calibration of all measured variables ap­pears in the following sequence:
z pH: Slope SLO and asymmetry ASY
z Oxygen: Relative slope SLO
z Cond: Cell constant C
Information concerning the calibration procedure is output as well.
pH/
Store
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
594
Baud
No.
Ident
8
Lin
Oxi
S L O
Auto
Cal
Tref25 Tref20
3 Press <RUN/ENTER> to display the value of the asymmetry
(mV).
pH/
Store
TEC
Store
TEC
ARng
0
ARng
mV
%
m
1/
°
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
mg/l
S/
cm
%
C
REL1
AR
cm
Sal
K
/
REL2
pH/
cm
Sal
K
/
REL2
TP
RCL
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBat nLF
4 Press <RUN/ENTER> to display the relative slope of the D. O.
pH
1
Baud
No.
Ident
Tref25 Tref20
Lin
sensor.
S
5
8
A S Y
Oxi
01
Auto
Cal
O
Sal
1
Time
Day.Month
Year
Tref25 Tref20
LoBat nLF
Baud
No.
Ident
0
8
Lin
Oxi
88
S L
Auto
Cal
51
Operation Multi 1970i
5 Press <RUN/ENTER> to display the cell constant. The dis-
played value is: z the current, calibrated cell constant (with sensor symbol on
the display) or
z the fixed cell constant 0.475 1/cm (without sensor symbol
on the display). In this case, the measuring parameters are
pH
S
O
Sal
1
Time
Day.Month
Year
LoBat nLF
initialized (see section 4.10 R
/pH
mV
mg/l
%
m
S/
Baud
No.
Ident
0 4 7 2
Tref20
Tref25
Lin
Oxi
LCLC
C
Auto
ARng
TEC
Cal
1/
°
cm
%
C
AR
cm
/
Sal
K
REL2
RCL RCL
ESET).
pH
O
Sal
1
Time
Day.Month
Year
LoBat nLF
Baud
Ident
Tref25
mVS/
pH
mg/l
%
m
S/
ARng
1/
°
cm
%
C
REL1REL1
AR
cm
K
/
Sal
TPTP
REL2
C
No.
4 7 5
0
Auto
Tref20
Lin
Oxi
TEC
Cal
6 With <M> you can switch back to the last active measuring
mode.
52
Multi 1970i Operation
Outputting the calibra-
tion protocol on the
interface
1 Press the <RCL> key repeatedly until CAL SEr appears on the
display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
C
Time Baud
Day.Month
Year
LoBat
C
No.
Ident
d
Tref25 Tref20
Lin
nLF
AL
S E R
Auto
Oxi
Cal
Store
TEC
ARng
2 Press the <RUN/ENTER> key.
The calibration protocol for all measured variables is transmit­ted to the interface. After the data transmission, the instrument automatically switches to the last active measurement mode.
Sample printout:
CALIBRATION PROTOCOL
02.03.02 14:19 Device No.: 12345678
Calibration pH Cal time: 01.03.02 / 15:20 Cal interval: 7d AutoCal TEC Tauto Buffer 1 2.00 Buffer 2 4.01 Buffer 3 7.00 * Buffer 4 10.01 C1 174.1 mV 25.0°C C2 -133.3 mV 25.0°C S1 -59.4 mV/pH ASY1 - 4 mV Probe: +++
CALIBRATION 02 Cal time: 02.03.02 / 14:19 Cal interval: 14d OxiCal Tauto AR Relative Slope: 0,88 Probe: +++
Calibration Conductivity Cal time: 14.01.02 / 11:37 Cal interval: 180d Cal Std.: 0.01 mol/l KCL
40.0 °C Conduct./Tref25: 1413µS/cm Cell Const : 0.478 1/cm Probe : +++
53
Operation Multi 1970i

4.7.4 Clearing the memory

With this function, you can delete the stored data records. 500 storage locations will then be available again.
Note
The Clear memory function only appears when there are data records stored in the memory. Otherwise, the measuring instrument automati­cally switches to the last active measuring mode.
Proceed as follows to clear all data records:
1 Switch off the measuring instrument.
2 Press the <STO> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
Time Baud
Day.Month
Year
LoBat
4 Confirm the clearing process with <RUN/ENTER>.
Note
The calibration data remain stored and can be called up.
S
Lin
d
Oxi
To
C L R
Auto
Store
TEC
ARng
Cal
C
No.
Ident
Tref25 Tref20
nLF
Pressing any other key prevents the clearing, the data records will remain stored.
54
Multi 1970i Operation

4.8 Transmitting data

You have the following possibilities of transmitting data:
z One of the following options:
– With the AutoStore function (page 46), the measured values of all
connected and active sensors are stored periodically and output to the interface (save interval Int 1).
– With the Data transmission interval (Int 2), the measured values
of all connected and active sensors are periodically output to the interface (see below).
AutoStore (Int 1) covers the Data transmission interval
(Int 2).
z Press the <RUN/ENTER> key
This manually triggers a data transmission of the current measured value to the serial interface at any time - independently of the select­ed intervals.
Setting the
Data transmission
interval
z With the Output data storage function (page 48), calibration data or
saved measured values are output on the interface.

4.8.1 Data transmission interval (Int 2)

The interval for the data transmission (Int 2) determines the chronolog­ical interval between automatic data transmissions. After the selected interval expires, the current data record is transmitted to the interface.
The default setting for the interval is OFF. To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <RCL> key. Int 2 appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
I
Time Baud
Day.Month
Year
LoBat
n
No.
Ident
d
Tref25 Tref20
Lin
nLF
T2
O F F
Store
Auto
Oxi
Cal
TEC
ARng
3 Set the required interval between the saving procedures with
<> <>.
55
Operation Multi 1970i
4 Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measur­ing mode.
Note
When the AutoStore function is active at the same time, the data trans­mission is performed according to the setting of the save interval (Int 1). Set the save interval (Int 1) to OFF to activate the Data transmission in- terval (Int 2).
56
Multi 1970i Operation
S
4.8.2 PC/external printer (RS232 interface)
Via the RS 232 interface, you can transmit the data to a PC or an ex­ternal printer. Use the AK340/B (PC) or AK325/S (ext. printer) cable to connect the interface to the devices.
Warning
The RS232 interface is not galvanically isolated. When connecting an earthed PC/printer, measurements cannot be per­formed in earthed media as incorrect values would result.
Set up the following transmission data at the PC/printer:
Baud rate selectable between: 1200, 2400, 4800, 9600
Handshake RTS/CTS + Xon/Xoff
PC only:
Socket assignment
Parity none
Data bits 8
Stop bits 2
1 CT 2 RxD
2
1
4
3
RS 232
REC

4.8.3 Remote control

The measuring instrument can be remotely controlled from a PC. This requires the KOM pilot communication kit. It is available as an acces­sory. The instrument is then controlled via commands that simulate keystrokes and request the current display contents.
3 Ground 4TxD
Note
A more detailed description is provided within the scope of delivery of the communication kit.
57
Operation Multi 1970i

4.9 Configuration

You can adapt the measuring instrument to your individual require­ments. To do this, the following parameters can be changed (the status on delivery is marked in bold):
Baud rate 1200, 2400, 4800, 9600
Air pressure display Current value in mbar
(no input possible)
Baudrate
Calibration intervals
– pH: 1 ... 7 ... 999 dInt 3
1 ... 14 ... 999 dInt 4
– O
2
: 1 ... 180 ... 999 dInt 5
Date/time Any
Note
You can leave the configuration menu at any time with <M> . The pa­rameters that have already been changed are stored.
1 Switch off the measuring instrument.
2 Press the <M> key and hold it down.
3 Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring instrument then switches automatically to the setting of the baud rate.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
Baud
No.
Ident
Tref25 Tref20
Lin
nLF
8 0 0
4
Auto
Oxi
Cal
Store
TEC
ARng
58
4 Select the required baud rate with <> <>.
5 Confirm with <RUN/ENTER>. CAL dISP appears on the dis-
play.
Multi 1970i Operation
Display during the
pH calibration
Displaying the
air pressure
mV
%
m
1/
°
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
mg/l
S/
cm
%
C
REL1
AR
pH/
cm
Sal
K
/
REL2
pH/
cm
Sal
K
/
REL2
TP
RCL
TP
RCL
pH
S
O
Sal
Time Baud
Day.Month
Year
LoBat
6 Select the required display during the pH calibration with
7 Confirm with <RUN/ENTER>. P appears on the display.
The air pressure in mbar is only displayed if a D. O. sensor is connected to the active REL socket. Otherwise, "---" appears.
pH
C
Lin
d
Oxi
AL
I S P
Store
Auto
TEC
ARng
Cal
C
No.
Ident
Tref25 Tref20
nLF
<> <>.
mV: Display of the current electrode voltage /pH: Display of the buffer nominal value.
S
O
Sal
Time Baud
Day.Month
Year
LoBat
C
No.
Ident
Tref25 Tref20
Lin
nLF
CA
d
9 4 9
Auto
Store
Cal
Oxi
TEC
P
ARng
pH calibration interval
8 Confirm with <RUN/ENTER>. Int 3 and the measured variable
pH appear on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
I
Baud
Time
Day.Month
LoBat
10 Confirm with <RUN/ENTER>. Int 4 and the measured variable
No.
Year
Ident
d
Tref25
Tref20
nLF
Oxi
Lin
9 Set the required interval in days (d) with <> <>.
appear on the display.
O
2
4
Auto
3nt
1
7
Store
ARng
TEC
Cal
59
Operation Multi 1970i
Dissolved oxygen
calibration interval
Conductivity
calibration interval
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
4nt
No.
4
1 4
Auto
Store
Oxi
Cal
TEC
ARng
Lin
d
Tref20
11 Set the required interval in days (d) with <> <>.
12 Confirm with <RUN/ENTER>. Int 5 and the measured variable
appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
5nt
No.
8 0
1
Auto
Store
Oxi
Cal
TEC
ARng
Lin
d
Tref20
Switching over the
reference temperature
13 Set the required interval in days (d) with <> <>.
14 Confirm with <RUN/ENTER>. The reference temperature set
for conductivity measurements appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
5t2
No.
Tref20
Lin
8 0d 1
Auto
Store
Cal
Oxi
TEC
ARng
15 Using <> <>, toggle between 25 °C (Tref25) and 20 °C
(Tref20).
16 Confirm with <RUN/ENTER>.
The date flashes on the display.
60
Multi 1970i Operation
Date and time
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
0 9 0 4
Tref25
Tref20
nLF
Lin
5t2
Auto
Store
Cal
Oxi
TEC
ARng
17 Set the date of the current day with <> <>.
18 Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
19 Set the current month with <> <>.
20 Confirm with <RUN/ENTER>.
The year appears on the display.
21 Set the current year with <> <>.
22 Confirm with <RUN/ENTER>.
The hours flash on the display.
23 Set the current time with <> <>.
24 Confirm with <RUN/ENTER>.
The minutes flash on the display.
25 Set the current time with <> <>.
26 Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measur­ing mode.
61
Operation Multi 1970i

4.10 Reset

You can reset (initialize) the measurement parameters and the config­uration parameters separately from one another.
Measurement
parameters
pH
O
The following measurement parameters (pH/O
/ InI) can be reset to
2
the delivery status:
Measuring mode pH
Asymmetry 0 mV
Slope -59.16 mV
2
Measuring mode D. O. concentration
Relative slope 1.00
Salinity (value) 0.0
Salinity (function) Off
Measuring mode
-1
Cell constant 0.475 cm
(fixed)
Configuration
parameters
Note
The calibration data gets lost when the measuring parameters are re­set. Recalibrate after performing a reset.
The following configuration parameters (InI) are reset to the delivery status:
Baud rate 4800
Display during the
Buffer nominal value
pH calibration
Interval 1 (automatically saved) OFF
Interval 2 (for data transmission) OFF
62
Multi 1970i Operation
Resetting the
measuring parameters
1 Press the <RUN/ENTER> key and hold it down.
2 Press the <CAL> key. The setting to reset the pH measuring
parameters appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
0 9
Tref20
Lin
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
3 Using <> <>, switch between no and YES.
YES: Resetting the pH measuring parameters no: Retaining settings.
4 Confirm with <RUN/ENTER>.
The measuring instrument switches to the setting to reset the O
measuring parameters.
2
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
0 9
Tref20
Lin
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
5 Using <> <>, switch between no and YES.
YES: Resetting the O
measuring parameters.
2
no: Retaining settings.
6 Confirm with <RUN/ENTER>. The measuring instrument
switches to the setting to reset the Cond measuring parame­ters.
63
Operation Multi 1970i
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
0 9
Lin
Tref20
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
7 Using <> <>, switch between no and YES.
YES: Resetting the Cond measuring parameters. no: Retaining settings.
8 Confirm with <RUN/ENTER>.
The measuring instrument switches to the configuration pa­rameters.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
0 9
Tref20
Lin
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
Resetting the
configuration parame-
ters
9 Using <> <>, switch between no and YES.
YES: Resetting the configuration parameters no: Retaining settings.
10 Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measur­ing mode.
64
Multi 1970i Maintenance, cleaning, disposal

5 Maintenance, cleaning, disposal

5.1 Maintenance

The measuring instrument is maintenance-free.

5.2 Cleaning

Occasionally wipe the outside of the measuring instrument with a damp, lint-free cloth. Disinfect the housing with isopropanol as re­quired.
Warning
The housing components are made out of synthetic materials (polyure­thane, ABS and PMMA). Thus, avoid contact with acetone and similar detergents that contain solvents. Remove any splashes immediately.

5.3 Disposal

Packing This measuring instrument is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing pro­tects the measuring instrument from transport damages.
Rechargeable battery Remove the rechargeable battery from the instrument and dispose of it
at a suitable facility according to local legal requirements. It is illegal to dispose of the rechargeable battery with household refuse. Proceed as follows to disassemble the rechargeable battery:
1
Ni-Cd Cd
1
1 Remove the carrying and positioning handle or the carrying
strap.
2
2 Unscrew the fixing elements (1) using a hexagon key.
3 Remove the instrument from the enclosure by vigorously
pressing against the socket field.
4 Take out the rechargeable battery (2) and cut off the battery ca-
ble.
65
Maintenance, cleaning, disposal Multi 1970i
Measuring instrument Dispose of the measuring instrument without the rechargeable battery
as electronic waste at an appropriate collection point.
66
Multi 1970i What to do if...

6 What to do if...

6.1 pH system messages

Error message 0FL
0FL
0FL0FL
Cause Remedy
pH electrode:
– Not connected – Connect electrode
Error message E3
E3
E3E3
Air bubble in front of the
Remove air bubble
diaphragm
– Air in the diaphragm – Extract air or
moisten diaphragm
Cable broken Replace electrode
Gel electrolyte dried out – Replace electrode
Cause Remedy
Electrode
Diaphragm contaminated – Clean diaphragm
Membrane contaminated Clean membrane
Moisture in the plug Dry plug
Electrolyte obsolete – Replenish electrolyte or
replace electrode
Electrode obsolete – Replace electrode
Electrode broken Replace electrode
Measuring instrument:
Incorrect calibration procedure – Select correct procedure
Incorrect solution temperature
Set up correct temperature
(without temperature sensor)
– Socket damp – Dry socket
67
What to do if... Multi 1970i
Buffer solutions
Incorrect buffer solutions Change calibration procedure
Buffer solutions too old – Use only once.
Note the shelf life
– Buffer solutions depleted – Change solutions
No stable measured
Cause Remedy
value
pH electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Obviously incorrect
measured values
Test sample:
– pH value not stable – Measure with air excluded if
necessary
– Temperature not stable – Adjust temperature if
necessary
Electrode + test sample:
Conductivity too low – Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode
Cause Remedy
pH electrode:
pH electrode unsuitable – Use suitable electrode
Temperature difference
between buffer and test
– Adjust temperature of buffers
or sample
sample too high
68
– Measurement procedure not
suitable
– Follow special procedure
Multi 1970i What to do if...

6.2 Oxi system messages

Error message 0FL
Error message E3
0FL
0FL0FL
E3
E3E3
Cause Remedy
Display range exceeded
Oxygen sensor
Not connected Connect the sensor
Cable broken Replace sensor
Depleted Replace sensor
Short-circuit between gold and
lead electrode
– Clean sensor and replace it if
necessary
Cause Remedy
Invalid calibration
Oxygen sensor
Electrolyte solution depleted – Regenerate sensor
Membrane contaminated Clean membrane
Error message E7
E7
E7E7
AR flashes
continuously
Electrode system poisoned – Regenerate sensor
Obsolete Replace sensor
Broken Replace sensor
Cause Remedy
Membrane damaged
Membrane damaged Regenerate sensor
Membrane head screwed on
Screw membrane head tight
not tight enough
Cause Remedy
No stable measured value
– Membrane contaminated – Clean membrane
69
What to do if... Multi 1970i
Measured value too low
Measured value too
high
Error message OFL
OFL
OFLOFL
Cause Remedy
Insufficient flow Provide flow to the sensor
Cause Remedy
– High amount of dissolved
substances
– Air bubbles bump on the
membrane with high velocity
The carbon dioxide pressure is
Correct solubility function
using the salinity equivalent
– Avoid direct flow to the
membrane
– Measuring not possible
too high (> 1 bar)

6.3 Conductivity system messages

Cause Remedy
Error message E3
E3
E3E3
The measured value lies outside the measuring range
Measuring cell not connected Connect measuring cell
Cable broken Replace measuring cell
Cause Remedy
– Measuring cell contaminated – Clean cell and replace it if
necessary
Unsuitable calibration
Check calibration solutions
solution
70
Multi 1970i What to do if...

6.4 General errors

Display LoBat
Instrument does not
react
to keystroke
Display to
to
toto
Cause Remedy
– Battery almost empty – Charge the battery (see
section 3.2)
Cause Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Switch the instrument on while pressing the <RCL> key
Cause Remedy
Sensor symbol flashes
Message StoFull
StoFull
StoFullStoFull
You want to know which
software version is in
the instrument
– Time-out of the interface – Check the instrument
connected
Cause Remedy
– Calibration interval expired – Recalibrate the measuring
system
Cause Remedy
– All memory locations are full – Output data storage and clear
data storage
Cause Remedy
– E. g., a question by the WTW
service department
– Simultaneously press the
<CAL> and <ON/OFF> keys and release them again. The software version is displayed.
71
What to do if... Multi 1970i
72
Multi 1970i Technical data

7 Technical data

7.1 General data

Dimensions approx. 90 x 200 x 190 mm
Weight approx. 1.5 kg (without plug-in power supply)
Mechanical structure Type of protection IP 67
Electrical safety Protective class III
Test certificates CE
Ambient conditions Storage - 25 °C ... + 65 °C
Operation -10 °C ... + 55 °C
Climatic class 2
Power
supply
Rechargeable battery Nickel-cadmium(NiCad) rechargeable bat-
tery
Operational life approx. 600 hours with one charging
Plug-in power supply unit (charging device)
The following applies to all plug-in power supplies:Connection max. Overvoltage category II
Plug-in power supply unit (Euro, US , UK, Australian plug) FRIWO FW7555M/09, 15.1432 Friwo Part. No. 1883259 Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA Output: 9 V = / 1,5 A
73
Technical data Multi 1970i
Serial
interface
Guidelines
and norms used
Connection of the cable AK 340/B or AK 325/S
Baud rate adjustable: 1200, 2400, 4800, 9600 Baud
Type RS232, data output
Data bits 8
Stop bits 2
Parity None
Handshake RTS/CTS + Xon/Xoff
Cable length Max. 15m
EMC EC guideline 89/336/EEC
EN 61326-1:1998 EN 61000-3-2 A14:2000 EN 61000-3-3:1995 FCC Class A
Instrument safety E.C. guideline 73/23/EEC
EN 61010-1 A2:1995
Manual
temperature input
Climatic class VDI/VDE 3540
IP protection EN 60529:1991

7.2 Measuring ranges, resolutions, accuracies

7.2.1 pH/ORP

Variable Measuring
range
pH - 2.00 ... + 19.99 0.01 ± 0.01 *
U [mV] - 1999 ... + 1999 1 ± 1
T [°C] - 5.0 ... + 105.0 0.1 ± 0.1
* when measuring in a range of ± 2 pH around a calibration point
Variable Range Increment
T
[°C] - 20 ... + 130 1
manual
Resolution Accuracy
(± 1 Digit)
74
Multi 1970i Technical data

7.2.2 Dissolved oxygen

Measuring ranges
and resolutions
Accuracies Variable Accuracy (± 1 Digit)
Correction
functions
Variable Measuring range Resolution
Concentration [mg/l] 0 ... 19.99
0 ... 90.0
Saturation [%] 0 ... 199.9
0 ... 600
0.01
0.1
0.1 1
T [°C] 0 ... 50.0 0.1
Concentration [mg/l] ± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C
Saturation [%] ± 0.5 % of measured value
when measuring in the range of ± 10 K around the calibration temperature
T [°C] ± 0.1
Temperature com-
Accuracy better than 2 % at 0 ... 40 °C
pensation
Salinity correction 0 ... 70.0 SAL
Air pressure correction
Automatic through integrated pressure sen­sor in the range of 500 ... 1100 mbar
75
Technical data Multi 1970i

7.2.3 Conductivity

Measuring ranges
and resolutions
Accuracies Variable Accuracy (± 1 Digit)
Variable Measuring range Resolution
[µS/cm] 0 ... 1999 1
[mS/cm] 0.00 ... 19.99
0.0 ... 199.9 0 ... 500
SAL 0.0 ... 70.0
0.01
0.1 1
0.1 according to the IOT table
T [°C] 5.0 ... + 105.0 0.1
[µS/cm] / [mS/cm]
Accuracy Sample temperature ± 0.5 % 0 °C ... 35 °C
(nonlinear compensa­tion)
± 0.5 % 35 °C ... 50 °C
according to EN 27 888;
extended nLF function according to WTW measurements
Cell constant,
calibrating
Reference
temperature
SAL
Accuracy Sample temperature ± 0.1 5 °C ... 25 °C
(range 0.0 ... 42.0)
± 0.2 25 °C ... 30 °C
T [°C] ± 0.1
C [cm-1] 0.450 ... 0.500
Tref Can be set to 20 °C or 25 °C
76
Multi 1970i Lists

8Lists

This chapter provides additional information and orientation aids.
Abbreviations The list of abbreviations explains the indicators and the abbreviations
that appear on the display and in the manual.
Specialist terms The glossary briefly explains the meaning of the specialist terms. How-
ever, terms that should already be familiar to the target group are not described here.
Index The index helps you to find the topics that you are looking for.
77
Lists Multi 1970i
Abbreviations
Conductivity value (international γ)
AR AutoRead (drift control)
ARng Automatic range switching
Measuring instrument measures with highest reso­lution
ASY Asymmetry
AutoCal DIN Automatic pH calibration with buffer solutions pre-
pared according to DIN 19 266
AutoCal TEC Automatic pH calibration with WTW technical buff-
er solutions according to DIN 19267
-1
C Cell constant [cm
] (internat. k)
°C Temperature unit, degrees Celsius
Cal Calibration
Cd... Display indicator during calibration for pH mea-
surements. Indicates the selection of the buffer data record for buffer solutions prepared according to DIN 19 266
Cm... Display indicator during calibration for pH mea-
surements. Indicates the selection of buffer data records for buffer solutions of the Merck company
ConCal Conventional single-point or two-point calibration
for pH measurements
Ct... Display indicator during calibration for pH mea-
surements. Indicates the selection of the buffer data records for WTW technical buffer solutions
E3 Error message
see chapter 6 W
HAT TO DO IF...
InI Initialization
Resets individual basic functions to the status they had on delivery
78
LoBat Battery nearly empty (Low Battery)
mV Voltage unit
mV/pH Unit of the electrode slope (internat. mV)
Multi 1970i Lists
nLF Nonlinear temperature compensation
OFL Display range exceeded (Overflow)
OxiCal Automatic calibration for D. O. measurements
pH pH value
S Slope (internat. k)
SAL Salinity
SELV Safety Extra Low Voltage
SLO Slope setting on calibration
TC Temperature coefficient (internat. α)
TP Temperature measurement active (temperature
sensor)
25/T25 Reference temperature of 25 °C
T
Ref
U
ASY
Asymmetry
79
Lists Multi 1970i
Glossary
Adjusting To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or a value that is regarded as correct, or that the difference remains with­in the tolerance.
Asymmetry Designation for the offset potential of a pH electrode. It is the measur-
able potential of a symmetrical electrode, the membrane of which is immersed in a solution with the pH of the nominal electrode zero point (WTW electrodes: pH = 7).
AutoRange Name of the automatic selection of the measuring range.
AutoRead WTW name for a function to check the stability of the measured value.
Calibration Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Of­ten, this expression is also used when the measuring system is adjust­ed at the same time (see adjusting).
Cell constant, k Characteristic quantity of a conductivity measuring cell, depending on
the geometry.
Conductivity Short form of the expression, specific electrical conductivity. It is a
measured value of the ability of a substance to conduct an electric cur­rent. In water analysis, the electrical conductivity is a dimension for the ionized substances in a solution.
D. O. partial pressure Pressure caused by the oxygen in a gas mixture or liquid.
Diaphragm The junction is a porous body in the housing wall of reference elec-
trodes or electrolyte bridges. It forms the electrical contact between two solutions and makes electrolyte exchange more difficult. The ex­pression, junction, is also used for ground or junction-less transitions.
Electrode zero point The zero point of a pH electrode is the pH value at which the electro-
motive force of the pH electrode at a specified temperature is zero. Normally, this is at 25 °C.
Electromotive force of
an electrode
The electromotive force U of the electrode is the measurable electro­motive force of an electrode in a solution. It equals the sum of all the galvanic voltages of the electrode. Its dependency on the pH results in the electrode function which is characterized by the parameters, slope and zero point.
Measured parameter The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
Measured value The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
80
Multi 1970i Lists
Measuring system The measuring system comprises all the devices used for measuring,
e. g. measuring instrument and sensor. In addition, there is the cable and possibly an amplifier, terminal strip and armature.
Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
MultiCal
®
WTW name stating that a measuring instrument provides several cal­ibration procedures.
Offset potential The measurable potential of a symmetrical electrode, the membrane
of which is immersed in a solution with the pH of the nominal electrode zero point. The asymmetry is part of the offset potential.
OxiCal
®
WTW name for a procedure to calibrate D. O. measuring systems in water vapor saturated air.
Oxygen saturation Short name for the relative D. O. saturation.
Note: The D. O. saturation value of air-saturated water and the D. O. saturation value of oxygen-saturated water are different.
pH value The pH is a measure of the acidic or basic effect of an aqueous solu-
tion. It corresponds to the negative decadic logarithm of the molal hy­drogen ions activity divided by the unit of the molality. The practical pH value is the value of a pH measurement.
Potentiometry Name of a measuring technique. The signal (depending on the mea-
sured parameter) of the electrode is the electrical potential. The elec­trical current remains constant.
ORP voltage The ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface (e. g. a gold or platinum surface).
Reference
temperature
Fixed temperature value to compare temperature-dependent mea­sured values. For conductivity measurements, the measured value is converted to a conductivity value at a reference temperature of 20 °C or 25 °C.
Reset Restoring the original condition of all settings of a measuring system.
Resistance Short name for the specific electrolytic resistance. It corresponds to
the reciprocal value of the electrical conductivity.
Resolution Smallest difference between two measured values that can be dis-
played by a measuring instrument.
Salinity The absolute salinity S
of the mass of dissolved salts to the mass of the solution (in g/Kg). In practice, this dimension cannot be measured directly. Therefore, the practical salinity is used for oceanographic monitoring. It is deter­mined by measuring the electrical conductivity.
of seawater corresponds to the relationship
A
81
Lists Multi 1970i
Salt content General designation for the quantity of salt dissolved in water.
Sample Designation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample and original sample are identical if the test sample was not processed.
Setting the
temperature
compensation
Name of a function that considers the temperature influence on the measurement and converts it accordingly. Depending on the mea­sured parameter to be determined, the temperature compensation functions in different ways. For conductimetric measurements, the measured value is converted to a defined reference temperature. For potentiometric measurements, the slope value is adjusted to the tem­perature of the test sample but the measured value is not converted.
Slope The slope of a linear calibration function.
Slope (relative) Designation used by WTW in the D. O. measuring technique. It ex-
presses the relationship of the slope value to the value of a theoretical reference sensor of the same type of construction.
Standard solution The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
TDS Total dissolved solids
Temperature
Value of the slope of a linear temperature function.
coefficient
Temperature function Name of a mathematical function expressing the temperature behav-
ior of a test sample, a sensor or part of a sensor.
82
Multi 1970i Index
A
Air pressure display 59 Asymmetry 21 Authorized use 9 AutoCal TEC 21, 24 AutoRead
Conductivity Criteria 31 Oxygen 31 pH 19
39
B
Batteries
Charging Charging time 11
11
C
Calibration
Conductivity D. O. 32 pH 21
Calibration evaluation
D. O.
33
Dissolved oxygen 33
Calibration interval 43
conductivity 60 Dissolved oxygen 60 pH 59
Calibration protocol 21
conductivity 40 D. O. 32
Cell constant 37, 40 Conductivity calibration eval­uation
41
Configuration parameters 62 Connecting a printer 57
40
D
D. O.
Measuring the concentra-
tion
29
Measuring the saturation
30
Data record 44 Delivery status 62 Display 4 Drift control
Conductivity Oxygen 31 pH 19
39
E
Energy saving feature 16 Error messages 69, 70
I
Initial commissioning 12, 14 Initialization 62 Interval
Calibration (Int 3) save 46
43
J
Jack field 5
K
Keys 5
M
Mains operation 11 Measuring 36
conductivity 38 D. O. concentration 29 D. O. saturation 30 ORP voltage 20 pH value 19 Salinity 38
Measuring mode when switching on Measuring ranges
Conductivity
Measuring the conductivity
16
76
38
Measuring the salinity 38
O
Operating 15 Operating safety 10 ORP electrode 20 ORP voltage 20
P
pH calibration evaluation 23 pH value 19 Plug-in power supply 11 Precautions 9
R
Reference temperature (con­ductivity) Remote control 57 Reset 62 Resetting the measuring pa­rameters RS232 interface 57 RS232 socket assignment
37
63
57
S
Safety 9 Salt content correction
entering the salinity
switching on 29 Save interval 46 Saving data 44 Scope of delivery 11 Setting the baud rate 58 Setting the date 12, 14, 61 Setting the reference temper­ature
60
Setting the time 12, 61 Single-point calibration 21 Single-point calibration (pH)
35
21, 25
Slope 21, 32
T
Temperature compensation (conductivity) Temperature probe
Conductivity Temperature sensor
D. O.
pH 18 Transmitting data 55 Transmitting measured val-
37
36
28
83
Index Multi 1970i
ues 55 Two-point calibration 21 Two-point calibration (pH)
21, 25
84
Wissenschaftlich-Technische Werkstätten GmbH
Dr.-Karl-Slevogt-Straße 1 D-82362 Weilheim
Germany
Tel: +49 (0) 881 183-0
+49 (0) 881 183-100 Fax: +49 (0) 881 183-420 E-Mail: Info@WTW.com Internet: http://www.WTW.com
Loading...