The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may result in differences between this operating manual and your instrument.
Also, we cannot guarantee that there are absolutely no errors in this
manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
The Multi 1970i portable multiparameter measuring instrument enables
you to carry out pH measurements, dissolved oxygen (D. O.) measurements and conductivity measurements quickly and reliably.
The Multi 1970i provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven MultiCal
®
and OxiCal® calibration procedures and the procedures to determine/set up the cell constant support you in your work
with the meter. The special AutoRead function enables precise measurements.
1
ON / OFF
M
AUTO READ
RUN / ENTER
CAL
2
O
7
92
m
g/
l
1 7 6
°
C
T
P
R
E
L
1
A
R
n
g
RCL
STO
Multi 1970i
3
4
5
1Keypad
2Display
3Integrated, exchangeable sensor quiver
4Jack field
5Carrying and positioning handle
3
OverviewMulti 1970i
Note
If you need further information or application notes, you can obtain the
following material from WTW:
z Application reports
z Primers
z Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
Select the measuring mode, increase values,
scroll
<▲ >
Select the measuring mode, decrease values,
scroll
<▼ >
RCL
STO
Display/transmit measured values
<RCL>
Save a measured value
<STO>
1.4Jack field
Warning
Only connect sensors to the measuring instrument that cannot return
any voltages or currents that are not allowed (> SELV and > current circuit with current limiting).
Almost all sensors - in particular WTW sensors - fulfill these conditions.
5
OverviewMulti 1970i
Maximum number
of sensors to be con-
nected
Connectors:
The following sensors can be connected to the Multi 1970i (maximum
configuration):
z One pH depth armature or
one pH electrode (combination electrode
or pH electrode + reference electrode). Please make sure that only
one pH sensor is connected at the same time.
z One D. O. sensor
z One conductivity measuring cell
z Option: One external temperature sensor
Note
The pH depth armature is affected by other sensors if it is connected to
them galvanically. For this reason, the pH depth armature cannot be
operated in the same test sample together with another sensor on the
instrument.
5
1
7
8
6
9
2
4
3
Sensor / InstrumentSocket / Posi-
tion
pH depth armature1
pH electrode or pH combination electrode2
pH combination electrode with temperature sensor2 and 4
Reference electrode3
Temperature sensor, external3 and 4
Oxygen sensor5 or 6
Conductivity measuring cell5 or 6
Printer or PC (serial interface, RS232)7
Plug-in power supply unit8
Watertight valve for internal pressure equalization9
6
Multi 1970iOverview
Note
The sensor sockets, 1 and 2 must not be assigned at the same time.
On the two sockets, 5 and 6, only different sensor types (Oxi and Cond)
may be connected at the same time.
7
OverviewMulti 1970i
8
Multi 1970iSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the measuring instrument. Consequently, all responsible personnel must read this
operating manual before working with the measuring system. The operating manual must always be available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in
the laboratory.
Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to
take when handling chemicals.
Safety instructionsThe individual chapters of this operating manual use the following safe-
ty instruction to indicate various types of danger:
Warning
indicates instructions that must be followed precisely in order to avoid
the possibility of slight injuries or damage to the instrument or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of
the:
z pH and ORP measurement
z measurement of the oxygen content and
z measurement of the conductivity, salinity and temperature
in the field and laboratory.
The technical specifications as given in chapter 7 T
must be observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is
authorized. Any other use is considered to be unauthorized.
ECHNICALDATA
9
SafetyMulti 1970i
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter
7T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
The smooth functioning and operational safety of the measuring instrument can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are
followed during operation.
The smooth functioning and operational safety of the measuring instrument can only be guaranteed under the environmental conditions that
are specified in chapter 7 T
ECHNICALDATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
of service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
Obligations of the pur-
chaser
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
10
Multi 1970iCommissioning
3Commissioning
3.1Scope of delivery
z Multi 1970i portable multiparameter measuring instrument with inte-
grated rechargeable battery
z Carrying and positioning handle
z Carrying strap
z 2 sensor quivers (pH and Oxi-LF type)
z Plug-in power supply unit
z Operating manual
3.2Power supply
Mains operation and
charging the battery
Charging time of the
battery
You can either operate the measuring instrument with the integrated rechargeable battery or with the plug-in power supply. The plug-in power
supply supplies the measuring instrument with low voltage (12 V DC).
At the same time, the rechargeable battery is charged.
approx. 16 hours. The battery is charged even when the instrument is
switched off. The LoBat display indicator appears when the battery is
nearly empty and has to be charged as soon as possible.
Warning
The line voltage at the operating site must lie within the input voltage
range of the original plug-in power supply (see chapter 7 T
TA).
ECHNICALDA-
Warning
Use original plug-in power supplies only (see
chapter 7 T
ECHNICALDATA).
11
CommissioningMulti 1970i
Connecting the plug-in
power supply unit
2
1
3
Setting the date and
time
1Connect the plug (1) to the socket (2) of the measuring instru-
ment.
2Connect the original WTW plug-in power supply (3) to an easily
accessible mains socket.
3.3Initial commissioning
Perform the following activities:
z For mains operation and charging the battery: Connect the plug-in
power supply (see section 3.2 P
z Set the date and time.
1Press the <M> key and hold it down.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
3Press the <RUN/ENTER> key repeatedly until the date flashes
on the display (Day.Month display indicator).
OWERSUPPLY).
12
Multi 1970iCommissioning
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
0 9 0 4
Tref25
Tref20
nLF
Lin
5t2
Auto
Store
Cal
Oxi
TEC
ARng
4Set the date of the current day with <▲> <▼>.
5Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
6Set the current month with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The year appears on the display.
8Set the current year with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The hours flash on the display.
10Set the current time with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The minutes flash on the display.
12Set the current time with <▲> <▼>.
13Confirm with <RUN/ENTER>.
The instrument switches to the measuring mode.
13
CommissioningMulti 1970i
3.4Sensor quiver
To store the sensors during field operation and to keep the sensor element moist, the quiver tip contains a sponge rubber insert that can be
moistened with deionized water.
Note
For further details on proper storage, refer to the operating manual of
the sensor.
Moistening the
quiver insert
1Press the quiver out of the holder from the back side of the in-
strument and pull it out completely.
2Pull off the quiver tip and moisten the sponge rubber with
deionized water.
Warning
Do not store pH electrodes in the quiver for more than 10 hours. To
store them for a longer period of time, use the watering cap filled with
potassium chloride (3 mol/l) of the electrode.
14
Multi 1970iOperation
4Operation
4.1Operating structure
Active and inactive REL
socket
In addition to the pH/ORP sensor, the sensor on the REL1 or REL2
socket can be actuated (switched "active"). The other socket is not actuated ("inactive"). This switching over inside the instrument is made
with the aid of a relay. The active REL socket is shown on the display.
The measured variable on the display can be selected as follows using
the <M> key:
z Press <M> shortly
z Press <M> for a longer
period of time:
Several measuring modes are available within a measured variable.
Switch over between the measured variables with the <▲> or <▼> key.
The options are summarized in the following diagram:
:Switch between the pH/ORP sensor and
the sensor at the active REL socket
Change the active Rel socket
15
OperationMulti 1970i
Note
When a sensor is connected to the active REL socket, the measuring
instrument recognizes the sensor or the measuring cell and automatically switches to the measuring mode that was last active. As soon as
the sensor is disconnected from the active REL socket, the instrument
switches to the pH (mV) measuring mode again.
4.2Switching on the measuring instrument
1Press the <ON/OFF> key.
The display test appears briefly on the display.
After this, the measuring instrument automatically switches to
the measuring mode.
Measuring mode when
switching on
Sensors on the socketMeasuring mode
pHREL
(active)
REL
(inactive)
:
pH or ORP measurement - depending on the last
selected setting
:
::
:
::
:::
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour.
Last selected measuring mode.
Only the last active
REL socket is taken into account
16
The energy saving feature is not active
z if the power is supplied by the plug-in power supply,
z if the AutoStore function is active,
z if the communication cable and a PC with a running communication
program are connected,
z if the printer cable is connected (for external printers).
Multi 1970iOperation
4.3pH value / ORP voltage
4.3.1General information
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect the pH depth armature or the pH electrode to the mea-
suring instrument.
If necessary, press the <M> key repeatedly until the pH (pH
measurement) or U (measurement of the ORP voltage) display
appears.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3Calibrate or check the measuring instrument with the elec-
trode.
4Using <▲ > <▼>, toggle between the pH or mV measuring
modes.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values. Calibrate regularly before measuring. You can only connect electrodes of the NTC30 type or without temperature sensor.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
17
OperationMulti 1970i
Temperature measure-
ment in pH measure-
ments
You can perform pH measurements with or without a temperature sensor as well as with the temperature sensor of an oxygen sensor or a
conductivity measuring cell. The measuring instrument recognizes
which sensors are connected and switches automatically to the correct
mode for the temperature measurement.
The following cases are distinguishable.
Temperature sensorDisplayMode
pHCond or
Oxi
yes-TPAutomatic with
pH temperature sensor
yesyesTP
--Manual
-yesTP
flashes
The temperature value of the
second sensor (Cond or Oxi)
in the same sample is taken
over for the pH measurement*
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature
input or
– use an electrode with a temperature sensor.
If a temperature sensor is connected, it is indicated on the display by
TP.
Note
When calibrating without a temperature sensor (no TP display indicator
displayed), enter the current temperature of the respective buffer solution manually using the <▲> <▼> keys while keeping the <RUN/EN-
TER> key depressed.
18
Multi 1970iOperation
4.3.2Measuring the pH value
1Perform the preparatory activities according to section 4.3.1.
2Immerse the pH electrode in the test sample.
3Press the <▲> <▼> keys repeatedly until pH appears on the
status display. The pH value appears on the display.
AutoRead AR
(Drift control) and hold
function
m
%
1/
°
mg/l
S/
cm
cm
%
C
REL1
AR
pH/mV
Sal
TP
K
/
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
4When measuring without a connected temperature sensor:
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
6
1
Baud
No.
Ident
Tref25Tref20
Options:
z Determine the current temperature using a thermometer
z TP display indicator not displayed, socket for the second
99
8
2 4 8
Auto
Store
Lin
Oxi
Cal
TEC
ARng
and, while keeping the <RUN/ENTER> key depressed, enter this temperature value with <▲> <▼>.
sensor is free:
Connect the second sensor (Oxi or Cond) and immerse it in
the same sample. TP flashes, the temperature is automatically measured using the second sensor.
With identical measurement conditions, the following criterion is valid
for the AutoRead function:
ReproducibilityResponse time
Better than 0.02> 30 seconds
For D. O. measurements, use the AutoRead function and hold function
like this:
1Call up the pH measuring mode with <▲> <▼>.
19
OperationMulti 1970i
2Activate the AutoRead function with <AUTO READ>.
The current measured value is frozen (hold function).
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the
<AUTO READ> key.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
4.3.3Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. SenTix ORP, the measuring
instrument can measure the ORP voltage (U) of a solution.
1Perform the preparatory activities according to section 4.3.1.
2Submerse the ORP electrode in the sample.
3Press the <▲> <▼> key until the U status display appears. The
ORP voltage (mV) of the test sample appears on the display.
4Wait for a stable measured value.
pH/
Store
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
1
Baud
Ident
Tref25
No.
6
8
Tref20
Lin
Oxi
23
2 4 8
Auto
Cal
20
Multi 1970iOperation
4.3.4pH calibration
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the measuring instrument. Thus, you should calibrate at regular intervals.
When to calibrate?z After connecting another electrode
z When the sensor symbol flashes, i.e. after the calibration interval
has expired
Calibration pointsCalibration can be made with one or two buffer solutions (single-point
or two-point calibration). The measuring instrument determines the following values and calculates the calibration lines as follows:
Determined valuesValues of the calibration lines
1-pointASYz Asymmetry = ASY
z Slope = Nernst slope (59.16 mV/
pH at 25 °C)
2-pointASY
SLO
z Asymmetry = ASY
z Slope = SLO
AutoCal TECis specially matched to the WTW technical buffer solutions as a fully au-
tomatic two-point calibration. The buffer solutions are automatically
recognized by the measuring instrument. Depending on the instrument
setting (see section 4.9 C
ONFIGURATION), the instrument displays the
relevant buffer nominal value or the current electrode voltage in mV.
The calibration can be terminated after the first buffer solution. This corresponds to a single-point calibration. To do this, the instrument uses
the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
AutoReadThe calibration procedure automatically activates the AutoRead func-
tion.
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
Displaying the calibra-
tion data
Printing the
calibration protocol
You can view the data of the last calibration on the display. The proceeding is described on page 50.
The calibration protocol contains the calibration data of the current calibration. You can transmit the calibration protocol to a printer via the serial interface (see O
INTERFACE, page 53).
UTPUTTINGTHECALIBRATIONPROTOCOLONTHE
21
OperationMulti 1970i
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.8.2
before calibrating. After a valid calibration, the record is printed.
Calibration evaluationAfter calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The
worst evaluation appears on the display.
DisplayAsymmetry
[mV]
-15 ... +15-60.5 ... -58
-20 ... +20-58 ... -57
-25 ... +25-61 ... -60.5
-30 ... +30-62 ... -61
Clean the electrode according
to the electrode operating
manual
Slope
[mV/pH]
or
-57 ... -56
or
-56 ... -50
Preparatory activities
E3
Perform error elimination according to chapter 6 W
DOIF...
HATTO
< -30 or
> 30
< -62 or
> -50
1Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status display pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2Keep the buffer solutions ready.
3Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a
temperature sensor (the TP display indicator is missing from
the display).
23
OperationMulti 1970i
AutoCal TECFor this procedure, use any two WTW technical buffer solutions (pH
values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the WTW technical buffer
solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions can lead
to an erroneous calibration. The correct buffer solutions are given in the
WTW catalog or in the Internet.
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see
section 4.9 C
er nominal value or the current electrode voltage in mV.
Note
Skip the steps 2 and 7 if you use a pH electrode with temperature sensor or the temperature sensor of a conductivity measuring cell or a D.
O. sensor.
ONFIGURATION), the instrument displays the relevant buff-
Starting the calibration
1Press the <CAL> key. The Ct1 display and the function display
AutoCal TEC appears. The sensor symbol displays the evalu-
ation of the last calibration (or no sensor symbol in the delivery
state or after the measurement parameters have been reset).
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
Sal
TP
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
2If required, enter the temperature of the first buffer solution with
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
C
Lin
8
Oxi
T1
2 4 8
Store
Auto
TEC
ARng
Cal
1
Baud
No.
Ident
Tref25Tref20
<▲> <▼> while keeping the <RUN/ENTER> key depressed.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
24
Multi 1970iOperation
mVS/
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
REL2
Sal
TP
RCL
/
pH
3
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
/
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
5When the measured value is stable, Ct2 appears.
1
Baud
No.
Ident
Tref25Tref20
Lin
69
8
2 4 8
Auto
Oxi
Cal
K
Sal
TP
REL2
RCL
pH
O
Sal
1
Time
Day.Month
Year
LoBatnLF
700
Baud
No.
Ident
8
Lin
Oxi
2 4 8
Auto
Cal
Tref25Tref20
TEC
ARng
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
Note
At this point, the AutoCal TEC calibration can be terminated with <M>.
This corresponds to a single-point calibration. To do this, the instru-
ment uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the
asymmetry of the electrode.
6Thoroughly rinse the electrode with distilled water.
7If required, enter the temperature of the second buffer solution
8Immerse the pH electrode into the second buffer solution.
9Press the <RUN/ENTER> key.
C
Lin
8
Oxi
T2
2 4 8
Store
Auto
TEC
ARng
Cal
1
Baud
No.
Ident
Tref25Tref20
with <▲> <▼> while keeping the <RUN/ENTER> key depressed.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
25
OperationMulti 1970i
mVS/
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
1
Time
Day.Month
Year
LoBatnLF
S
Baud
No.
Ident
Tref25Tref20
Lin
8
2 4 8
Auto
Oxi
Cal
TEC
ARng
mV
%
m
1/
°
/
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
1
Time
Day.Month
Year
LoBatnLF
401184
Baud
No.
Ident
8
Tref25Tref20
Lin
Oxi
2 4 8
Auto
TEC
ARng
Cal
10When the measured value is stable, AR disappears. The value
of the slope (mV/pH) appears on the display. The sensor symbol shows the evaluation of the current calibration.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
594
Baud
No.
Ident
Tref25Tref20
Lin
8
Oxi
248
Auto
Cal
Store
TEC
ARng
11Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
pH/
2
Store
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
59
Baud
No.
Ident
8
Tref25Tref20
Lin
248
Auto
Oxi
Cal
12Switch to the measuring mode with <M>.
26
Multi 1970iOperation
4.4Dissolved oxygen
4.4.1General information
Note
D. O. measurements with the Multi 1970i can only be carried out using
a CellOx 325 or StirrOx G D. O. sensor. The stirrer of the StirrOx G
D. O. sensor has to be supplied with voltage separately using the
NT/pH Mix 540 power supply.
You can measure the following variables:
z D. O. concentration
z Oxygen saturation
The measuring instrument is supplied with the following functions:
z AutoRange (automatic switchover of the measurement range). If a
measuring range is exceeded, AutoRange causes the measuring instrument to change automatically to the next higher measuring
range and back again. Therefore, the instrument always measures
in the measuring range with the highest possible resolution.
z The AutoRead function (drift control) for checking the stability of the
measurement signal. This ensures the reproducibility of the measuring signal. For details of how to switch the AutoRead function on/off,
see page 31.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect the D. O. sensor to the measuring instrument. If the D.
O. sensor is connected to the active REL socket, press the
<M> repeatedly if necessary until the O
status display indica-
2
tor appears. If the D. O. sensor is connected to the inactive
REL socket, press the <M> key for 2 s, until the O
status dis-
2
play indicator appears.
2Calibrate or check the measuring instrument with the sensor.
How to calibrate is described in section 4.4.5 from page 21.
3Use <▲> <▼> to toggle between the measuring modes, D. O.
concentration (mg/L) and D. O. saturation (%).
27
OperationMulti 1970i
Note
Incorrect calibration of D. O. probes will result in incorrect measured
values.
Calibrate at regular intervals.
Temperature sensorThe D. O. sensor has an integrated temperature sensor that always
measures the current temperature of the test sample.
28
Multi 1970iOperation
4.4.2Measuring the D. O. concentration
When measuring the concentration of test samples with a salt content
of more than 1 g/l, a salinity correction is required.
Note
How to enter the current salt content is described in
section 4.4.6 E
Switching the salt content correction on or off, see below.
To measure the D. O. concentration with and without salt content correction, proceed as follows:
1Perform the preparatory activities according to section 4.4.1.
2Immerse the D. O. sensor in the test sample.
3Press the <▲> <▼> key repeatedly until the D. O. concentra-
tion in mg/l appears on the display.
NTERINGTHESALTCONTENT (SALINITY) on page 35.
Switching on/off the salt
content correction
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
Proceed as follows to switch on the salt content correction:
1Perform the preparatory activities according to section 4.4.1.
2Immerse the D. O. sensor in the test sample.
3While pressing the <RUN/ENTER> key, switch on the salt con-
7
Lin
8
Oxi
92
1 7 6
Auto
Store
Cal
TEC
ARng
1
Baud
No.
Ident
Tref25Tref20
tent correction with <▲> . The SAL display indicator appears
on the display. The specified salt content is taken into consideration during the measurement.
29
OperationMulti 1970i
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
4While pressing the <RUN/ENTER> key, switch off the salt con-
4.4.3Measuring the D. O. saturation
You can measure the D. O. saturation as follows:
1Perform the preparatory activities according to section 4.4.1.
2Immerse the D. O. sensor in the test sample.
5
Lin
8
Oxi
42
1 7 6
Auto
Store
Cal
TEC
ARng
1
Baud
No.
Ident
Tref25Tref20
tent correction with <▼>. The SAL display indicator is no longer
displayed.
3Press the <▲> <▼> key repeatedly until the D. O. saturation in
% appears on the display.
/mV
%
m
1/
°
S/
C
pH
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
8
Baud
No.
Ident
8
Tref25Tref20
Lin
92
1 7 6
Auto
Oxi
Cal
Store
TEC
ARng
30
Multi 1970iOperation
4.4.4AutoRead AR (Drift control) and hold function
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
CriteriaWith identical measurement conditions, the following criteria are valid
for the AutoRead function:
Measuring modeReproducibilityResponse time
D. O. concentration
Oxygen saturation
index
For D. O. measurements, use the AutoRead function and hold function
like this:
1Call up the measuring mode with <M> and/or <▲> <▼> .
2Activate the AutoRead function with <AUTO READ>. The cur-
rent measured value is frozen (hold function).
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This measured value is transmitted to the interface.
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
1
Baud
No.
Ident
Tref25Tref20
Lin
7
8
1 7 6
Oxi
08
Auto
better than 0.05 mg/l> 10 seconds
better than 0.6 %> 10 seconds
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
ARng
C
REL1
AR
REL2
RCL
Store
Cal
TEC
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the
<AUTO READ> key.
31
OperationMulti 1970i
4.4.5D. O. calibration
Why calibrate?D. O. probes age. This changes the slope of the D. O. sensor. Calibra-
tion determines the current slope of the sensor and stores this value in
the instrument.
When to calibrate?z After connecting another D. O. sensor
z When the sensor symbol flashes (after the calibration interval has
expired).
Calibration procedureThe calibration is performed in water vapor-saturated air. Use the
®
OxiCal
-SL air calibration vessel (accessory) for the calibration.
AutoReadThe calibration procedure automatically activates the AutoRead func-
tion. The AR display indicator flashes. The calibration process is finished when AR stops flashing.
Displaying the calibra-
tion data
Printing the
calibration protocol
You can view the data of the last calibration on the display. The proceeding is described on page 50.
The calibration protocol contains the calibration data of the current calibration. You can transmit the calibration protocol to a printer via the serial interface (see page 53).
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.8.2
before calibrating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
02.03.02 14:19
Device No.: 12345678
CALIBRATION 02
Cal time: 02.03.01 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++
32
Multi 1970iOperation
Sensor evaluationAfter the calibration, the measuring instrument evaluates the current
status of the sensor against the relative slope. The evaluation appears
on the display. The relative slope has no effect on the measuring accuracy. Low values indicate that the electrolyte will soon be depleted and
the sensor will have to be regenerated.
DisplayRelative slope
S = 0.8 ... 1.25
S = 0.7 ... 0.8
S = 0.6 ... 0.7
E3
S < 0.6 or S > 1.25
Perform error elimination according to chapter 6 W
HATTODOIF...
Starting the calibrationProceed as follows to calibrate the instrument:
1Perform the preparatory activities according to section 4.4.1.
®
2Keep the OxiCal
-SL air calibration vessel ready.
Note
The sponge in the air calibration vessel must be moist (not wet). Observe the instructions in the OxiCal
®
-SL operating manual.
3Put the D. O. sensor into the air calibration vessel.
4Press the <CAL> key repeatedly until the calibration mode ap-
pears. The sensor symbol displays the evaluation of the last
calibration (or no sensor symbol in the delivery state or after the
measuring parameters have been reset).
33
OperationMulti 1970i
m
m
%
1/
°
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
pH/mV
cm
K
/
REL2
Sal
TP
RCL
Sal
TP
RCL
pH
S
O
Sal
TimeBaud
Day.Month
Year
LoBat
5Press the <RUN/ENTER> key. AutoRead is active, AR flashes.
pH
1
No.
Ident
Tref25Tref20
Lin
nLF
S
C
8
1 7 6
Oxi
AL
Auto
Store
ARng
TEC
Cal
O
Sal
TimeBaud
Day.Month
Year
LoBat
6As soon as a stable value is achieved the AR display stops
0
Lin
8
Oxi
88
1 7 6
Auto
Store
ARng
TEC
Cal
1
No.
Ident
Tref25Tref20
nLF
flashing. The calibration is finished then. The value of the relative slope appears on the display. The sensor symbol shows
the sensor evaluation (see page 33).
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
TimeBaud
Day.Month
Year
LoBat
7Switch to the measuring mode with <M>.
Note
In chapter 6 W
to take for error elimination.
1
No.
Ident
Tref25Tref20
Lin
nLF
0
88
8
1 7 6
Auto
Store
ARng
Oxi
TEC
Cal
HATTODOIF... from page 67, you will find the measures
34
Multi 1970iOperation
4.4.6Entering the salt content (salinity)
A salt content correction is required in the oxygen concentration measurement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent (the measured salinity) of the
test sample (range 0.0 - 70.0) and to switch on the salinity correction.
Note
With the Multi 1970i, you can measure the salinity. How to proceed is
described in section 4.5.3 M
EASURINGTHESALINITY on page 38.
Entering the
salt content
1Determine the salinity of the test sample (any method, see also
section 4.5.3 M
2Press the <CAL> key repeatedly until Sal appears on the dis-
play.
pH
S
O
Sal
TimeBaud
Day.Month
Year
LoBat
3Enter the salt content with <▲> <▼>.
4Switch to the measuring mode with <M>.
Note
How to switch on the salt content correction is described on page 29.
1
No.
Ident
Tref25Tref20
Lin
nLF
5
8
6 9 6
Oxi
50
Auto
Cal
EASURINGTHESALINITY on page 38).
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
ARng
C
REL1
AR
REL2
RCL
Store
TEC
35
OperationMulti 1970i
4.5Conductivity
4.5.1General information
Note
Conductivity measurements with the Multi 1970i can only be carried out
using the TetraCon 325 measuring cell.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.
The measuring instrument is supplied with the following functions:
z AutoRange (automatic switchover of the measurement range). If a
measuring range is exceeded, AutoRange causes the measuring instrument to change automatically to the next higher measuring
range and back again. Therefore, the instrument always measures
in the measuring range with the highest possible resolution.
z The AutoRead function (drift control) for checking the stability of the
measurement signal. This ensures the reproducibility of the measuring signal. For details of how to switch the AutoRead function on/off,
see page 39.
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect a conductivity measuring cell to the measuring instru-
ment. If the conductivity measuring cell is connected to the active REL socket, press the <M> key repeatedly (if necessary)
until the status display appears. If the conductivity measuring cell is connected to the inactive REL socket, press the <M>
key for 2 s until the status display appears.
2Check the selected cell constant or calibrate the measuring in-
strument with the measuring cell (see below).
3Check the selected cell constant (see below) or calibrate the
measuring instrument with the measuring cell (see
section 4.5.5 on page 40).
4Using <▲ > <▼>, toggle between the measuring modes, con-
ductivity ( in µS/cm) or salinity (SAL).
Temperature sensorThe TetraCon 325 conductivity measuring cell has a temperature sen-
sor integrated in it. The temperature sensor is shown on the display by
TP.
36
Multi 1970iOperation
Temperature
compensation
Reference temperature,
Tref
Checking the
cell constant
The nonlinear temperature compensation is set fixed and is shown on
the display by nLF.
The reference temperature (Tref) can be switched between 20 °C and
25 °C. It appears on the display as Tref20 or Tref25. To switch over the
reference temperature, see S
TURE, page 60.
1Press the <RCL> key repeatedly until CAL disp appears on the
display.
pH
S
O
Sal
TimeBaud
Day.Month
Year
LoBat
C
No.
Ident
Tref25Tref20
Lin
nLF
C
I S P
d
Oxi
AL
Auto
Store
TEC
ARng
Cal
WITCHINGOVERTHEREFERENCETEMPERA-
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
C
REL2
REL1
AR
RCL
2Press the <RUN/ENTER> repeatedly until the last calibrated
cell constant is displayed, e. g. 0.472 1/cm.
pH
S
O
Sal
C
TimeBaud
Day.Month
LoBat
No.
Year
Ident
0
Tref20
Tref25
Oxi
Lin
nLF
3To return to the measuring mode: Press the <M> key when the
correct cell constant is displayed.
4If you want to recalibrate the cell constant, proceed according
to section 4.5.5 D
INTHECONTROLSTANDARD) .
LC
C
4 7 5
Auto
Store
ARng
TEC
Cal
pH/mV
mg/l
%
m
S/
cm
1/
cm
Sal
TP
K
%
°
/
C
REL1
REL2
AR
RCL
ETERMININGTHECELLCONSTANT (CALIBRATION
37
OperationMulti 1970i
4.5.2Measuring the conductivity
You can carry out the conductivity measurements as follows:
1Perform the preparatory activities according to section 4.5.1
page 36.
2Immerse the conductivity measuring cell in the test sample.
3Press the < ▲> <▼> keys until in the status display, and the
unit µS/cm appears. The conductivity value appears on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
TimeBaud
Day.Month
Year
LoBat
Ident
Tref25
nLF
05 3
No.
0
2 4 0
Tref20
Auto
Store
Cal
Oxi
Lin
TEC
ARng
4.5.3Measuring the salinity
You can carry out the salinity measurements as follows:
1Perform the preparatory activities according to section 4.5.1
page 36.
2Immerse the conductivity measuring cell in the test sample.
3Press the <▲> <▼> keys repeatedly until the Sal status display
appears. The salinity value appears on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
TimeBaud
Day.Month
Year
LoBat
Ident
Tref25
nLF
332
No.
0
2 1 2
Tref20
Auto
Store
TEC
ARng
Cal
Oxi
Lin
38
Multi 1970iOperation
4.5.4AutoRead AR (Drift control) and hold function
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
For conductivity measurements, use the AutoRead function and hold
function like this:
1Call up the measuring mode or SAL with <M> and/or <▲ >
<▼>.
2Immerse the conductivity measuring cell in the test sample.
3Activate the AutoRead function with <AUTO READ>. The cur-
rent measured value is frozen (hold function).
4Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
pH/mV
cm
Sal
K
/
REL2
RCL
pH
S
O
Sal
C
TimeBaud
Day.Month
Year
LoBat
Ident
Tref25
nLF
05 3
No.
0
2 5 0
Tref20
Auto
Store
Cal
Oxi
Lin
TEC
ARngTPAR
5If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
6To terminate AutoRead: Press the <AUTO READ> key.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>. You can
only change to another measuring mode after completion of AutoRead.
39
OperationMulti 1970i
4.5.5Determining the cell constant (Calibration in the control
standard)
Why determine the cell
constant?
AutoReadIn calibration, the AutoRead function is automatically activated. The AR
Displaying the
calibration data
Printing the
calibration protocol
Aging slightly changes the cell constant, e. g. by coatings. As a result,
an inexact measured value is displayed. The original characteristics of
the cell can often be restored by cleaning the cell. Calibration determines the current value of the cell constant and stores this value in the
instrument.
Thus, you should calibrate at regular intervals.
In the delivery condition, the cell constant of the measuring instrument
is set to 0.475 cm
-1
(conductivity measuring cell TetraCon 325).
display indicator flashes. The calibration process is finished when AR
stops flashing.
You can view the data of the last calibration on the display. The proceeding is described on page 50.
The calibration protocol contains the calibration data of the current calibration. You can transmit the calibration protocol to a printer via the serial interface (see page 53).
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.8.2
before calibrating. After a valid calibration, the record is printed.
Sample printout:
CALIBRATION PROTOCOL
14.04.02 11:37
Device No.: 99990000
Calibration Conductivity
Cal time: 14.04.01 / 11:37
Cal interval: 180d
Cal Std.: 0.01 mol/l KCL
Calibration evaluationAfter the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the display.
-1
Display Cell constant [cm
0.450 ... 0.500 cm
]
-1
Determining the
cell constant
E3
Perform error elimination according to chapter 6 W
IF...
HATTODO
Outside the range
0.450 ... 0.500 cm
-1
You can determine the cell constant (conductivity measuring cell
TetraCon 325) as follows:
1Press the <CAL> key repeatedly until the status display
CAL appears.
2Press the <RUN/ENTER> key. The CAL display appears, as
well as
– the current, calibrated cell constant (with sensor symbol on
the display) or
– the fixed cell constant 0.475 1/cm (without sensor symbol
on the display). In this case, the measurement parameters
are initialized. See “Reset” on page 62..
/pH
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
Baud
No.
Ident
0 4 7 2
Tref20
Tref25
Lin
Oxi
AC
Auto
Cal
L
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
cm
%
C
REL1
AR
pH
O
Sal
Time
Sal
REL2
RCL
Day.Month
Year
LoBatnLF
K
/
CAL
1
Baud
No.
Ident
0
Tref20
Tref25
Lin
Oxi
4 7 5
Auto
Cal
TEC
ARng
mVS/
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
Sal
TPTP
REL2
RCL
3Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
41
OperationMulti 1970i
4Press the <RUN/ENTER> key.
The AutoRead measurement to determine the cell constant
starts. The AR display indicator flashes until a stable signal is
reached. The cell constant determined is displayed. The measuring instrument automatically stores the cell constant.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
TimeBaud
Day.Month
Year
LoBat
Ident
Tref25
nLF
LCA
No.
4 7 3
0
Tref20
Auto
Store
Oxi
Lin
Cal
TEC
ARng
5To return to the measuring mode: Press the <M> key. The de-
termined cell constant is taken over for the measurement.
Note
If error message
appears refer to chapter 6 W
E3
E3
E3E3
HATTODOIF...
42
Multi 1970iOperation
4.6Calibration intervals (Int 3, Int 4, Int 5)
For each measured variable, a time interval is stored. When it has expired, you will be reminded to calibrate. After a calibration interval has
expired, the sensor symbol of the relevant measured variable flashes.
It is still possible to measure. By calibrating the relevant sensor, the
function is reset and the interval starts anew.
The following calibration intervals are set in the factory:
Setting the calibration
interval
Measured
DesignationDefault setting
parameter
pH/ORPInt 37 days
Dissolved
Int 414 days
oxygen
ConductivityInt 5180 days
You can change each of these intervals (1 ... 999 days):
1Switch off the measuring instrument.
2Press the <M> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configuration.
4Press the <RUN/ENTER> key repeatedly, until Int 3 ... 5 to-
gether with the required measured variable (pH, O
or) ap-
2
pears. Example:
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
4nt
No.
4
1 4
Auto
Store
Oxi
Cal
TEC
ARng
Lin
d
Tref20
5Set the required interval until the next calibration with <▲>
<▼>.
6Confirm with <RUN/ENTER>.
7Switch to the measuring mode with <M>.
43
OperationMulti 1970i
4.7Saving data
The portable Multi 1970i multiparameter measuring instrument has an
internal data storage. It can store up to 500 datasets.
A complete data record consists of:
z Number of the storage location
z Date/time
z Measured values of the connected and active sensors
z Temperature values of the connected and active sensors
z Temperature measuring procedure
z ID number
You can transmit measured values (data records) to the data storage
in two ways:
z Save manually
z Switch on AutoStore (Int 1), see page 46
4.7.1Saving manually
You can transmit a measured value to the data storage as follows:
1Press the <STO> key.
The current number (location number No.) of the next free storage location appears under the current measured value on the
display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
d1
4
Tref20
Oxi
Lin
Auto
Cal
Store
TEC
969
4
ARng
2Confirm with <RUN/ENTER>.
The display switches to entering the ID number.
44
Multi 1970iOperation
Message StoFull
StoFullThis message appears when all of the 500 storage locations are occu-
StoFullStoFull
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
Tref25
nLF
d1
4
Tref20
Auto
Cal
Oxi
Lin
Store
TEC
969
2
ARng
3Using <▲> <▼>, enter the required ID number
(1 ... 999).
4Confirm with <RUN/ENTER>.
The measured values are stored. The instrument changes to
the measuring mode.
pied.
You have the following options:
Saving the current measured value.
The oldest measured value (storage location 1)
will be overwritten by this
Press
<RUN/ENTER>
Returning to the measuring mode without savingpress any key
Outputting the data storagesee
section 4.7.3
Clearing the memorysee
section 4.7.4
45
OperationMulti 1970i
4.7.2Switching on AutoStore (Int 1)
The save interval (Int 1) determines the chronological interval between
automatic save processes. After the fixed interval has expired, the current data record is transmitted to the internal storage and to the interface.
Switching on AutoStore
1Press the <RUN/ENTER> key and hold it down.
2Press the <STO> key. Int 1 appears on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
1nt
No.
Lin
d
Tref20
Auto
Cal
Oxi
F
FO
Store
ARng
TEC
3Set the required interval between the saving procedures with
The number of free memory locations appears on the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
949
No.
E
ER
Store
Auto
Cal
Oxi
TEC
ARng
Lin
F
Tref20
46
5Confirm with <RUN/ENTER>.
The prompt for the ID number appears on the display.
Multi 1970iOperation
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
949
No.
ERF
Store
TEC
1
ARng
Lin
Tref20
Auto
Cal
Oxi
6Set the required ID number with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measuring mode and start the measuring and saving procedure. Au-toStore flashes on the display.
As soon as all of the 500 storage locations are occupied, AutoStore is
terminated (Int 1 = OFF). If there are not enough storage locations
available for your measurements:
z Output and backup the data storage (see page 48) and
z clear the memory (see page 54).
Note
The AutoStore function is interrupted if you start other functions, e.g.
output the data storage. After the other function is finished, the AutoStore function is continued. By this, however, temporal gaps in the recording of the measured values will occur.
Switching off AutoStoreSwitch AutoStore off by:
z setting the save interval (Int 1) to OFF, or
z switching the measuring instrument off and then on again.
47
OperationMulti 1970i
4.7.3Outputting the data storage
You can output the contents of the data storage:
z Stored data on the display
z Calibration data on the display
z Stored data on the serial interface
z Calibration protocol on the interface
Outputting stored data
on the display
1Press the <RCL> key repeatedly until StO dISP appears on the
display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
oST
No.
SID
P
Store
Auto
Tref20
Cal
Oxi
Lin
TEC
ARng
2Press the <RUN/ENTER> key.
A measured value appears on the display.
The storage location of the data record is displayed for approx.
2 s, then the respective temperature appears.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
6
Baud
No.
Ident
8
Tref25Tref20
Lin
03
2 5 0
Auto
Store
Oxi
Cal
TEC
ARng
48
You can perform the following activities:
Display further data of the data record
Press <RUN/ENTER>
(ID number, date, time, storage location)
Toggle between two saved measured variables
Press <RUN/EN-
TER> + <M>
Advance one data record (storage location) Press <▲>
Go back one data record (storage location)Press <▼>
Multi 1970iOperation
Note
If you want to search for a certain element of the data record (e.g. date),
proceed as follows:
1Using <RUN/ENTER>, select the element (e.g. date).
2Press <▲> or <▼> repeatedly until the required element ap-
pears on the display.
After approx. 2 s the temperature of the displayed measured
value appears.
Outputting stored data
to the interface
1Press the <RCL> key repeatedly until Sto SEr appears on the
display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
oST
No.
Lin
D
Tref20
ES
r
Store
Auto
Cal
Oxi
TEC
ARng
2Press the <RUN/ENTER> key.
The complete contents of the storage are transmitted to the interface. During the data transmission the instrument increments the storage numbers. After the data transmission, the
instrument automatically switches to the last active measurement mode.
The transmitted data contains the entire contents of the storage in incrementing order of the location numbers.
Note
You can cancel the transmission with <M> or <RUN/ENTER>.
49
OperationMulti 1970i
Sample printout:
No. 1:
09.03.02 17:10
pH 10.01 25 °C
Tauto AR
Ident : 47
No. 2:
09.03.02 17:12
305 mV
Tauto
Ident : 6
No. 3:
09.03.02 17:24
7.88 mg/l 17.6° C
Tauto
Ident : 81
No. 4:
09.03.02 17:46
7.11 mg/l 17.8° C
Tauto
SAL = 17.9
Ident : 4
Outputting the calibra-
tion data
on the display
No. 5:
10.03.02 19:09
2.40 mS/cm 25.3 °C
Tauto
nLF
Tref25 C = 0.475 1/cm
Ident : 10
No. 6:
10.03.02 20:48
2.46 mS/cm 25.6 °C
Tauto
nLF
Tref25 C = 0.475 1/cm
Ident : 1
...
1Press the <RCL> key repeatedly until CAL disp appears on
the display.
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
S
O
Sal
C
TimeBaud
Day.Month
Year
LoBat
C
No.
Ident
d
Tref25Tref20
Lin
nLF
AL
I S P
Auto
Oxi
Cal
Store
TEC
ARng
50
Multi 1970iOperation
2Press the <RUN/ENTER> key.
The data of the last calibration of all measured variables appears in the following sequence:
z pH: Slope SLO and asymmetry ASY
z Oxygen: Relative slope SLO
z Cond: Cell constant C
Information concerning the calibration procedure is output as
well.
pH/
Store
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
594
Baud
No.
Ident
8
Lin
Oxi
S L O
Auto
Cal
Tref25Tref20
3Press <RUN/ENTER> to display the value of the asymmetry
(mV).
pH/
Store
TEC
Store
TEC
ARng
0
ARng
mV
%
m
1/
°
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
mg/l
S/
cm
%
C
REL1
AR
cm
Sal
K
/
REL2
pH/
cm
Sal
K
/
REL2
TP
RCL
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
4Press <RUN/ENTER> to display the relative slope of the D. O.
pH
1
Baud
No.
Ident
Tref25Tref20
Lin
sensor.
S
5
8
A S Y
Oxi
01
Auto
Cal
O
Sal
1
Time
Day.Month
Year
Tref25Tref20
LoBatnLF
Baud
No.
Ident
0
8
Lin
Oxi
88
S L
Auto
Cal
51
OperationMulti 1970i
5Press <RUN/ENTER> to display the cell constant. The dis-
played value is:
z the current, calibrated cell constant (with sensor symbol on
the display) or
z the fixed cell constant 0.475 1/cm (without sensor symbol
on the display). In this case, the measuring parameters are
pH
S
O
Sal
1
Time
Day.Month
Year
LoBatnLF
initialized (see section 4.10 R
/pH
mV
mg/l
%
m
S/
Baud
No.
Ident
0 4 7 2
Tref20
Tref25
Lin
Oxi
LCLC
C
Auto
ARng
TEC
Cal
1/
°
cm
%
C
AR
cm
/
Sal
K
REL2
RCLRCL
ESET).
pH
O
Sal
1
Time
Day.Month
Year
LoBatnLF
Baud
Ident
Tref25
mVS/
pH
mg/l
%
m
S/
ARng
1/
°
cm
%
C
REL1REL1
AR
cm
K
/
Sal
TPTP
REL2
C
No.
4 7 5
0
Auto
Tref20
Lin
Oxi
TEC
Cal
6With <M> you can switch back to the last active measuring
mode.
52
Multi 1970iOperation
Outputting the calibra-
tion protocol on the
interface
1Press the <RCL> key repeatedly until CAL SEr appears on the
display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
C
TimeBaud
Day.Month
Year
LoBat
C
No.
Ident
d
Tref25Tref20
Lin
nLF
AL
S E R
Auto
Oxi
Cal
Store
TEC
ARng
2Press the <RUN/ENTER> key.
The calibration protocol for all measured variables is transmitted to the interface. After the data transmission, the instrument
automatically switches to the last active measurement mode.
With this function, you can delete the stored data records. 500 storage
locations will then be available again.
Note
The Clear memory function only appears when there are data records
stored in the memory. Otherwise, the measuring instrument automatically switches to the last active measuring mode.
Proceed as follows to clear all data records:
1Switch off the measuring instrument.
2Press the <STO> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
TimeBaud
Day.Month
Year
LoBat
4Confirm the clearing process with <RUN/ENTER>.
Note
The calibration data remain stored and can be called up.
S
Lin
d
Oxi
To
C L R
Auto
Store
TEC
ARng
Cal
C
No.
Ident
Tref25Tref20
nLF
Pressing any other key prevents the clearing, the data records
will remain stored.
54
Multi 1970iOperation
4.8Transmitting data
You have the following possibilities of transmitting data:
z One of the following options:
– With the AutoStore function (page 46), the measured values of all
connected and active sensors are stored periodically and output
to the interface (save interval Int 1).
– With the Data transmission interval (Int 2), the measured values
of all connected and active sensors are periodically output to the
interface (see below).
– AutoStore (Int 1) covers the Data transmission interval
(Int 2).
z Press the <RUN/ENTER> key
This manually triggers a data transmission of the current measured
value to the serial interface at any time - independently of the selected intervals.
Setting the
Data transmission
interval
z With the Output data storage function (page 48), calibration data or
saved measured values are output on the interface.
4.8.1Data transmission interval (Int 2)
The interval for the data transmission (Int 2) determines the chronological interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.
The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min,
5 min, 10 min, 15 min, 30 min, 60 min):
1Press the <RUN/ENTER> key and hold it down.
2Press the <RCL> key. Int 2 appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
I
TimeBaud
Day.Month
Year
LoBat
n
No.
Ident
d
Tref25Tref20
Lin
nLF
T2
O F F
Store
Auto
Oxi
Cal
TEC
ARng
3Set the required interval between the saving procedures with
<▲> <▼>.
55
OperationMulti 1970i
4Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measuring mode.
Note
When the AutoStore function is active at the same time, the data transmission is performed according to the setting of the save interval (Int 1).
Set the save interval (Int 1) to OFF to activate the Data transmission in-terval (Int 2).
56
Multi 1970iOperation
S
4.8.2PC/external printer (RS232 interface)
Via the RS 232 interface, you can transmit the data to a PC or an external printer.
Use the AK340/B (PC) or AK325/S (ext. printer) cable to connect the
interface to the devices.
Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result.
Set up the following transmission data at the PC/printer:
The measuring instrument can be remotely controlled from a PC. This
requires the KOM pilot communication kit. It is available as an accessory. The instrument is then controlled via commands that simulate
keystrokes and request the current display contents.
3 Ground
4TxD
Note
A more detailed description is provided within the scope of delivery of
the communication kit.
57
OperationMulti 1970i
4.9Configuration
You can adapt the measuring instrument to your individual requirements. To do this, the following parameters can be changed (the status
on delivery is marked in bold):
Baud rate1200, 2400, 4800, 9600
Air pressure displayCurrent value in mbar
(no input possible)
Baudrate
Calibration
intervals
– pH: 1 ... 7 ... 999 dInt 3
1 ... 14 ... 999 dInt 4
– O
2
–: 1 ... 180 ... 999 dInt 5
Date/timeAny
Note
You can leave the configuration menu at any time with <M> . The parameters that have already been changed are stored.
1Switch off the measuring instrument.
2Press the <M> key and hold it down.
3Press the <ON/OFF> key.
The display test appears briefly on the display. The measuring
instrument then switches automatically to the setting of the
baud rate.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
Baud
No.
Ident
Tref25Tref20
Lin
nLF
8 0 0
4
Auto
Oxi
Cal
Store
TEC
ARng
58
4Select the required baud rate with <▲> <▼>.
5Confirm with <RUN/ENTER>. CAL dISP appears on the dis-
play.
Multi 1970iOperation
Display during the
pH calibration
Displaying the
air pressure
mV
%
m
1/
°
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
mg/l
S/
cm
%
C
REL1
AR
pH/
cm
Sal
K
/
REL2
pH/
cm
Sal
K
/
REL2
TP
RCL
TP
RCL
pH
S
O
Sal
TimeBaud
Day.Month
Year
LoBat
6Select the required display during the pH calibration with
7Confirm with <RUN/ENTER>. P appears on the display.
The air pressure in mbar is only displayed if a D. O. sensor is connected
to the active REL socket. Otherwise, "---" appears.
pH
C
Lin
d
Oxi
AL
I S P
Store
Auto
TEC
ARng
Cal
C
No.
Ident
Tref25Tref20
nLF
<▲> <▼>.
mV: Display of the current electrode voltage
/pH: Display of the buffer nominal value.
S
O
Sal
TimeBaud
Day.Month
Year
LoBat
C
No.
Ident
Tref25Tref20
Lin
nLF
CA
d
9 4 9
Auto
Store
Cal
Oxi
TEC
P
ARng
pH calibration interval
8Confirm with <RUN/ENTER>. Int 3 and the measured variable
pH appear on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
I
Baud
Time
Day.Month
LoBat
10Confirm with <RUN/ENTER>. Int 4 and the measured variable
No.
Year
Ident
d
Tref25
Tref20
nLF
Oxi
Lin
9Set the required interval in days (d) with <▲> <▼>.
appear on the display.
O
2
4
Auto
3nt
1
7
Store
ARng
TEC
Cal
59
OperationMulti 1970i
Dissolved oxygen
calibration interval
Conductivity
calibration interval
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
4nt
No.
4
1 4
Auto
Store
Oxi
Cal
TEC
ARng
Lin
d
Tref20
11Set the required interval in days (d) with <▲> <▼>.
12Confirm with <RUN/ENTER>. Int 5 and the measured variable
appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
5nt
No.
8 0
1
Auto
Store
Oxi
Cal
TEC
ARng
Lin
d
Tref20
Switching over the
reference temperature
13Set the required interval in days (d) with <▲> <▼>.
14Confirm with <RUN/ENTER>. The reference temperature set
for conductivity measurements appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
5t2
No.
Tref20
Lin
80d1
Auto
Store
Cal
Oxi
TEC
ARng
15Using <▲> <▼>, toggle between 25 °C (Tref25) and 20 °C
(Tref20).
16Confirm with <RUN/ENTER>.
The date flashes on the display.
60
Multi 1970iOperation
Date and time
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
0 9 0 4
Tref25
Tref20
nLF
Lin
5t2
Auto
Store
Cal
Oxi
TEC
ARng
17Set the date of the current day with <▲> <▼>.
18Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
19Set the current month with <▲> <▼>.
20Confirm with <RUN/ENTER>.
The year appears on the display.
21Set the current year with <▲> <▼>.
22Confirm with <RUN/ENTER>.
The hours flash on the display.
23Set the current time with <▲> <▼>.
24Confirm with <RUN/ENTER>.
The minutes flash on the display.
25Set the current time with <▲> <▼>.
26Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measuring mode.
61
OperationMulti 1970i
4.10Reset
You can reset (initialize) the measurement parameters and the configuration parameters separately from one another.
Measurement
parameters
pH
O
The following measurement parameters (pH/O
/ InI) can be reset to
2
the delivery status:
Measuring modepH
Asymmetry0 mV
Slope -59.16 mV
2
Measuring modeD. O. concentration
Relative slope 1.00
Salinity (value)0.0
Salinity (function)Off
Measuring mode
-1
Cell constant0.475 cm
(fixed)
Configuration
parameters
Note
The calibration data gets lost when the measuring parameters are reset. Recalibrate after performing a reset.
The following configuration parameters (InI) are reset to the delivery
status:
Baud rate4800
Display during the
Buffer nominal value
pH calibration
Interval 1
(automatically saved)OFF
Interval 2
(for data transmission)OFF
62
Multi 1970iOperation
Resetting the
measuring parameters
1Press the <RUN/ENTER> key and hold it down.
2Press the <CAL> key. The setting to reset the pH measuring
parameters appears on the display.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
09
Tref20
Lin
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
3Using <▲> <▼>, switch between no and YES.
YES: Resetting the pH measuring parameters
no: Retaining settings.
4Confirm with <RUN/ENTER>.
The measuring instrument switches to the setting to reset the
O
measuring parameters.
2
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
09
Tref20
Lin
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
5Using <▲> <▼>, switch between no and YES.
YES: Resetting the O
measuring parameters.
2
no: Retaining settings.
6Confirm with <RUN/ENTER>. The measuring instrument
switches to the setting to reset the Cond measuring parameters.
63
OperationMulti 1970i
%
m
1/
°
S/
C
pH/mV
mg/l
cm
cm
%
/
REL1
AR
K
Sal
TP
REL2
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
09
Lin
Tref20
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
7Using <▲> <▼>, switch between no and YES.
YES: Resetting the Cond measuring parameters.
no: Retaining settings.
8Confirm with <RUN/ENTER>.
The measuring instrument switches to the configuration parameters.
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
REL2
Sal
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
Ident
Tref25
nLF
No.
09
Tref20
Lin
iin
n o
Auto
Store
Cal
Oxi
TEC
ARng
Resetting the
configuration parame-
ters
9Using <▲> <▼>, switch between no and YES.
YES: Resetting the configuration parameters
no: Retaining settings.
10Confirm with <RUN/ENTER>.
The measuring instrument switches to the last active measuring mode.
64
Multi 1970iMaintenance, cleaning, disposal
5Maintenance, cleaning, disposal
5.1Maintenance
The measuring instrument is maintenance-free.
5.2Cleaning
Occasionally wipe the outside of the measuring instrument with a
damp, lint-free cloth. Disinfect the housing with isopropanol as required.
Warning
The housing components are made out of synthetic materials (polyurethane, ABS and PMMA). Thus, avoid contact with acetone and similar
detergents that contain solvents. Remove any splashes immediately.
5.3Disposal
PackingThis measuring instrument is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing protects the measuring instrument from transport damages.
Rechargeable batteryRemove the rechargeable battery from the instrument and dispose of it
at a suitable facility according to local legal requirements. It is illegal to
dispose of the rechargeable battery with household refuse. Proceed as
follows to disassemble the rechargeable battery:
1
Ni-CdCd
1
1Remove the carrying and positioning handle or the carrying
strap.
2
2Unscrew the fixing elements (1) using a hexagon key.
3Remove the instrument from the enclosure by vigorously
pressing against the socket field.
4Take out the rechargeable battery (2) and cut off the battery ca-
ble.
65
Maintenance, cleaning, disposalMulti 1970i
Measuring instrumentDispose of the measuring instrument without the rechargeable battery
as electronic waste at an appropriate collection point.
Operational lifeapprox. 600 hours with one charging
Plug-in power supply
unit
(charging device)
The following applies to all plug-in power
supplies:Connection max.
Overvoltage category II
Plug-in power supply unit
(Euro, US , UK, Australian plug)
FRIWO FW7555M/09, 15.1432
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1,5 A
73
Technical dataMulti 1970i
Serial
interface
Guidelines
and norms used
Connection of the cable AK 340/B or AK 325/S
Baud rateadjustable: 1200, 2400, 4800, 9600 Baud
TypeRS232, data output
Data bits8
Stop bits2
ParityNone
HandshakeRTS/CTS + Xon/Xoff
Cable length Max. 15m
EMCEC guideline 89/336/EEC
EN 61326-1:1998
EN 61000-3-2 A14:2000
EN 61000-3-3:1995
FCC Class A
Instrument safetyE.C. guideline 73/23/EEC
EN 61010-1 A2:1995
Manual
temperature input
Climatic classVDI/VDE 3540
IP protectionEN 60529:1991
7.2Measuring ranges, resolutions, accuracies
7.2.1pH/ORP
VariableMeasuring
range
pH- 2.00 ... + 19.990.01± 0.01 *
U [mV]- 1999 ... + 19991± 1
T [°C]- 5.0 ... + 105.00.1± 0.1
* when measuring in a range of ± 2 pH around a calibration point
VariableRangeIncrement
T
[°C]- 20 ... + 1301
manual
ResolutionAccuracy
(± 1 Digit)
74
Multi 1970iTechnical data
7.2.2Dissolved oxygen
Measuring ranges
and resolutions
AccuraciesVariableAccuracy (± 1 Digit)
Correction
functions
VariableMeasuring rangeResolution
Concentration [mg/l]0 ... 19.99
0 ... 90.0
Saturation [%]0 ... 199.9
0 ... 600
0.01
0.1
0.1
1
T [°C]0 ... 50.00.1
Concentration [mg/l]± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C
Saturation [%]± 0.5 % of measured value
when measuring in the range of ± 10 K
around the calibration temperature
T [°C]± 0.1
Temperature com-
Accuracy better than 2 % at 0 ... 40 °C
pensation
Salinity correction0 ... 70.0 SAL
Air pressure
correction
Automatic through integrated pressure sensor in the range of 500 ... 1100 mbar
75
Technical dataMulti 1970i
7.2.3Conductivity
Measuring ranges
and resolutions
AccuraciesVariableAccuracy (± 1 Digit)
VariableMeasuring rangeResolution
[µS/cm]0 ... 19991
[mS/cm]0.00 ... 19.99
0.0 ... 199.9
0 ... 500
SAL0.0 ... 70.0
0.01
0.1
1
0.1
according to the IOT
table
T [°C]− 5.0 ... + 105.00.1
[µS/cm] / [mS/cm]
Accuracy Sample temperature
± 0.5 %0 °C ... 35 °C
(nonlinear compensation)
± 0.5 % 35 °C ... 50 °C
according to EN 27 888;
extended nLF function
according to
WTW measurements
Cell constant,
calibrating
Reference
temperature
SAL
Accuracy Sample temperature
± 0.1 5 °C ... 25 °C
(range 0.0 ... 42.0)
± 0.2 25 °C ... 30 °C
T [°C] ± 0.1
C [cm-1]0.450 ... 0.500
TrefCan be set to 20 °C or 25 °C
76
Multi 1970iLists
8Lists
This chapter provides additional information and orientation aids.
AbbreviationsThe list of abbreviations explains the indicators and the abbreviations
that appear on the display and in the manual.
Specialist termsThe glossary briefly explains the meaning of the specialist terms. How-
ever, terms that should already be familiar to the target group are not
described here.
IndexThe index helps you to find the topics that you are looking for.
77
ListsMulti 1970i
Abbreviations
Conductivity value (international γ)
ARAutoRead (drift control)
ARngAutomatic range switching
Measuring instrument measures with highest resolution
ASYAsymmetry
AutoCal DINAutomatic pH calibration with buffer solutions pre-
pared according to DIN 19 266
AutoCal TECAutomatic pH calibration with WTW technical buff-
er solutions according to DIN 19267
-1
CCell constant [cm
] (internat. k)
°CTemperature unit, degrees Celsius
CalCalibration
Cd...Display indicator during calibration for pH mea-
surements. Indicates the selection of the buffer
data record for buffer solutions prepared according
to DIN 19 266
Cm...Display indicator during calibration for pH mea-
surements. Indicates the selection of buffer data
records for buffer solutions of the Merck company
ConCalConventional single-point or two-point calibration
for pH measurements
Ct...Display indicator during calibration for pH mea-
surements. Indicates the selection of the buffer
data records for WTW technical buffer solutions
E3Error message
see chapter 6 W
HATTODOIF...
InIInitialization
Resets individual basic functions to the status they
had on delivery
78
LoBatBattery nearly empty (Low Battery)
mVVoltage unit
mV/pHUnit of the electrode slope (internat. mV)
Multi 1970iLists
nLFNonlinear temperature compensation
OFLDisplay range exceeded (Overflow)
OxiCal Automatic calibration for D. O. measurements
pHpH value
SSlope (internat. k)
SALSalinity
SELVSafety Extra Low Voltage
SLOSlope setting on calibration
TCTemperature coefficient (internat. α)
TPTemperature measurement active (temperature
sensor)
25/T25Reference temperature of 25 °C
T
Ref
U
ASY
Asymmetry
79
ListsMulti 1970i
Glossary
AdjustingTo manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains within the tolerance.
AsymmetryDesignation for the offset potential of a pH electrode. It is the measur-
able potential of a symmetrical electrode, the membrane of which is
immersed in a solution with the pH of the nominal electrode zero point
(WTW electrodes: pH = 7).
AutoRangeName of the automatic selection of the measuring range.
AutoReadWTW name for a function to check the stability of the measured value.
CalibrationComparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct. Often, this expression is also used when the measuring system is adjusted at the same time (see adjusting).
Cell constant, kCharacteristic quantity of a conductivity measuring cell, depending on
the geometry.
ConductivityShort form of the expression, specific electrical conductivity. It is a
measured value of the ability of a substance to conduct an electric current. In water analysis, the electrical conductivity is a dimension for the
ionized substances in a solution.
D. O. partial pressurePressure caused by the oxygen in a gas mixture or liquid.
DiaphragmThe junction is a porous body in the housing wall of reference elec-
trodes or electrolyte bridges. It forms the electrical contact between
two solutions and makes electrolyte exchange more difficult. The expression, junction, is also used for ground or junction-less transitions.
Electrode zero pointThe zero point of a pH electrode is the pH value at which the electro-
motive force of the pH electrode at a specified temperature is zero.
Normally, this is at 25 °C.
Electromotive force of
an electrode
The electromotive force U of the electrode is the measurable electromotive force of an electrode in a solution. It equals the sum of all the
galvanic voltages of the electrode. Its dependency on the pH results
in the electrode function which is characterized by the parameters,
slope and zero point.
Measured parameterThe measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D. O. concentration.
Measured valueThe measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
80
Multi 1970iLists
Measuring systemThe measuring system comprises all the devices used for measuring,
e. g. measuring instrument and sensor. In addition, there is the cable
and possibly an amplifier, terminal strip and armature.
MolalityMolality is the quantity (in Mol) of a dissolved substance in 1000 g sol-
vent.
MultiCal
®
WTW name stating that a measuring instrument provides several calibration procedures.
Offset potentialThe measurable potential of a symmetrical electrode, the membrane
of which is immersed in a solution with the pH of the nominal electrode
zero point. The asymmetry is part of the offset potential.
OxiCal
®
WTW name for a procedure to calibrate D. O. measuring systems in
water vapor saturated air.
Oxygen saturationShort name for the relative D. O. saturation.
Note: The D. O. saturation value of air-saturated water and the D. O.
saturation value of oxygen-saturated water are different.
pH valueThe pH is a measure of the acidic or basic effect of an aqueous solu-
tion. It corresponds to the negative decadic logarithm of the molal hydrogen ions activity divided by the unit of the molality. The practical pH
value is the value of a pH measurement.
PotentiometryName of a measuring technique. The signal (depending on the mea-
sured parameter) of the electrode is the electrical potential. The electrical current remains constant.
ORP voltageThe ORP is caused by oxidizing or reducing substances dissolved in
water if these substances become effective on an electrode surface
(e. g. a gold or platinum surface).
Reference
temperature
Fixed temperature value to compare temperature-dependent measured values. For conductivity measurements, the measured value is
converted to a conductivity value at a reference temperature of 20 °C
or 25 °C.
ResetRestoring the original condition of all settings of a measuring system.
ResistanceShort name for the specific electrolytic resistance. It corresponds to
the reciprocal value of the electrical conductivity.
ResolutionSmallest difference between two measured values that can be dis-
played by a measuring instrument.
SalinityThe absolute salinity S
of the mass of dissolved salts to the mass of the solution (in g/Kg). In
practice, this dimension cannot be measured directly. Therefore, the
practical salinity is used for oceanographic monitoring. It is determined by measuring the electrical conductivity.
of seawater corresponds to the relationship
A
81
ListsMulti 1970i
Salt contentGeneral designation for the quantity of salt dissolved in water.
SampleDesignation of the sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
Setting the
temperature
compensation
Name of a function that considers the temperature influence on the
measurement and converts it accordingly. Depending on the measured parameter to be determined, the temperature compensation
functions in different ways. For conductimetric measurements, the
measured value is converted to a defined reference temperature. For
potentiometric measurements, the slope value is adjusted to the temperature of the test sample but the measured value is not converted.
SlopeThe slope of a linear calibration function.
Slope (relative)Designation used by WTW in the D. O. measuring technique. It ex-
presses the relationship of the slope value to the value of a theoretical
reference sensor of the same type of construction.
Standard solutionThe standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
TDSTotal dissolved solids
Temperature
Value of the slope of a linear temperature function.
coefficient
Temperature functionName of a mathematical function expressing the temperature behav-
ior of a test sample, a sensor or part of a sensor.
82
Multi 1970iIndex
A
Air pressure display 59
Asymmetry 21
Authorized use 9
AutoCal TEC 21, 24
AutoRead
pH calibration evaluation 23
pH value 19
Plug-in power supply 11
Precautions 9
R
Reference temperature (conductivity)
Remote control 57
Reset 62
Resetting the measuring parameters
RS232 interface 57
RS232 socket assignment
37
63
57
S
Safety 9
Salt content correction
entering the salinity
switching on 29
Save interval 46
Saving data 44
Scope of delivery 11
Setting the baud rate 58
Setting the date 12, 14, 61
Setting the reference temperature
60
Setting the time 12, 61
Single-point calibration 21
Single-point calibration (pH)
35
21, 25
Slope 21, 32
T
Temperature compensation
(conductivity)
Temperature probe
Conductivity
Temperature sensor
D. O.
pH 18
Transmitting data 55
Transmitting measured val-