The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may result in differences between this operating manual and your instrument.
Also, we cannot guarantee that there are absolutely no errors in this
manual. Therefore, we are sure you will understand that we cannot accept any legal claims resulting from the data, figures or descriptions.
The Multi 1970i portable multiparameter measuring instrument enables
you to carry out pH measurements, dissolved oxygen (D. O.) measurements and conductivity measurements quickly and reliably.
The Multi 1970i provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven MultiCal
®
and OxiCal® calibration procedures and the procedures to determine/set up the cell constant support you in your work
with the meter. The special AutoRead function enables precise measurements.
1
ON / OFF
M
AUTO READ
RUN / ENTER
CAL
2
O
7
92
m
g/
l
1 7 6
°
C
T
P
R
E
L
1
A
R
n
g
RCL
STO
Multi 1970i
3
4
5
1Keypad
2Display
3Integrated, exchangeable sensor quiver
4Jack field
5Carrying and positioning handle
3
OverviewMulti 1970i
Note
If you need further information or application notes, you can obtain the
following material from WTW:
z Application reports
z Primers
z Safety datasheets.
You will find information on available literature in the WTW catalog or
via the Internet.
Select the measuring mode, increase values,
scroll
<▲ >
Select the measuring mode, decrease values,
scroll
<▼ >
RCL
STO
Display/transmit measured values
<RCL>
Save a measured value
<STO>
1.4Jack field
Warning
Only connect sensors to the measuring instrument that cannot return
any voltages or currents that are not allowed (> SELV and > current circuit with current limiting).
Almost all sensors - in particular WTW sensors - fulfill these conditions.
5
OverviewMulti 1970i
Maximum number
of sensors to be con-
nected
Connectors:
The following sensors can be connected to the Multi 1970i (maximum
configuration):
z One pH depth armature or
one pH electrode (combination electrode
or pH electrode + reference electrode). Please make sure that only
one pH sensor is connected at the same time.
z One D. O. sensor
z One conductivity measuring cell
z Option: One external temperature sensor
Note
The pH depth armature is affected by other sensors if it is connected to
them galvanically. For this reason, the pH depth armature cannot be
operated in the same test sample together with another sensor on the
instrument.
5
1
7
8
6
9
2
4
3
Sensor / InstrumentSocket / Posi-
tion
pH depth armature1
pH electrode or pH combination electrode2
pH combination electrode with temperature sensor2 and 4
Reference electrode3
Temperature sensor, external3 and 4
Oxygen sensor5 or 6
Conductivity measuring cell5 or 6
Printer or PC (serial interface, RS232)7
Plug-in power supply unit8
Watertight valve for internal pressure equalization9
6
Multi 1970iOverview
Note
The sensor sockets, 1 and 2 must not be assigned at the same time.
On the two sockets, 5 and 6, only different sensor types (Oxi and Cond)
may be connected at the same time.
7
OverviewMulti 1970i
8
Multi 1970iSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the measuring instrument. Consequently, all responsible personnel must read this
operating manual before working with the measuring system. The operating manual must always be available within the vicinity of the measuring system.
Target groupThe measuring instrument was developed for work in the field and in
the laboratory.
Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary safety precautions to
take when handling chemicals.
Safety instructionsThe individual chapters of this operating manual use the following safe-
ty instruction to indicate various types of danger:
Warning
indicates instructions that must be followed precisely in order to avoid
the possibility of slight injuries or damage to the instrument or the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
2.1Authorized use
The authorized use of the measuring instrument consists exclusively of
the:
z pH and ORP measurement
z measurement of the oxygen content and
z measurement of the conductivity, salinity and temperature
in the field and laboratory.
The technical specifications as given in chapter 7 T
must be observed. Only the operation and running of the measuring instrument according to the instructions given in this operating manual is
authorized. Any other use is considered to be unauthorized.
ECHNICALDATA
9
SafetyMulti 1970i
2.2General safety instructions
This instrument is built and inspected according to the relevant guidelines and norms for electronic measuring instruments (see chapter
7T
ECHNICALDATA).
It left the factory in a safe and secure technical condition.
Function and operating
safety
Safe operationIf safe operation is no longer possible, the instrument must be taken out
The smooth functioning and operational safety of the measuring instrument can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are
followed during operation.
The smooth functioning and operational safety of the measuring instrument can only be guaranteed under the environmental conditions that
are specified in chapter 7 T
ECHNICALDATA.
If the instrument was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the
instrument reaches room temperature before putting the instrument
back into operation.
of service and secured against inadvertent operation!
Safe operation is no longer possible if the measuring instrument:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
Obligations of the pur-
chaser
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the instrument.
The purchaser of the measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
10
Multi 1970iCommissioning
3Commissioning
3.1Scope of delivery
z Multi 1970i portable multiparameter measuring instrument with inte-
grated rechargeable battery
z Carrying and positioning handle
z Carrying strap
z 2 sensor quivers (pH and Oxi-LF type)
z Plug-in power supply unit
z Operating manual
3.2Power supply
Mains operation and
charging the battery
Charging time of the
battery
You can either operate the measuring instrument with the integrated rechargeable battery or with the plug-in power supply. The plug-in power
supply supplies the measuring instrument with low voltage (12 V DC).
At the same time, the rechargeable battery is charged.
approx. 16 hours. The battery is charged even when the instrument is
switched off. The LoBat display indicator appears when the battery is
nearly empty and has to be charged as soon as possible.
Warning
The line voltage at the operating site must lie within the input voltage
range of the original plug-in power supply (see chapter 7 T
TA).
ECHNICALDA-
Warning
Use original plug-in power supplies only (see
chapter 7 T
ECHNICALDATA).
11
CommissioningMulti 1970i
Connecting the plug-in
power supply unit
2
1
3
Setting the date and
time
1Connect the plug (1) to the socket (2) of the measuring instru-
ment.
2Connect the original WTW plug-in power supply (3) to an easily
accessible mains socket.
3.3Initial commissioning
Perform the following activities:
z For mains operation and charging the battery: Connect the plug-in
power supply (see section 3.2 P
z Set the date and time.
1Press the <M> key and hold it down.
2Press the <ON/OFF> key.
The display test appears briefly on the display.
3Press the <RUN/ENTER> key repeatedly until the date flashes
on the display (Day.Month display indicator).
OWERSUPPLY).
12
Multi 1970iCommissioning
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
O
Sal
Time
Day.Month
Year
LoBat
S
I
Baud
No.
Ident
0 9 0 4
Tref25
Tref20
nLF
Lin
5t2
Auto
Store
Cal
Oxi
TEC
ARng
4Set the date of the current day with <▲> <▼>.
5Confirm with <RUN/ENTER>.
The date (month) flashes in the display.
6Set the current month with <▲> <▼>.
7Confirm with <RUN/ENTER>.
The year appears on the display.
8Set the current year with <▲> <▼>.
9Confirm with <RUN/ENTER>.
The hours flash on the display.
10Set the current time with <▲> <▼>.
11Confirm with <RUN/ENTER>.
The minutes flash on the display.
12Set the current time with <▲> <▼>.
13Confirm with <RUN/ENTER>.
The instrument switches to the measuring mode.
13
CommissioningMulti 1970i
3.4Sensor quiver
To store the sensors during field operation and to keep the sensor element moist, the quiver tip contains a sponge rubber insert that can be
moistened with deionized water.
Note
For further details on proper storage, refer to the operating manual of
the sensor.
Moistening the
quiver insert
1Press the quiver out of the holder from the back side of the in-
strument and pull it out completely.
2Pull off the quiver tip and moisten the sponge rubber with
deionized water.
Warning
Do not store pH electrodes in the quiver for more than 10 hours. To
store them for a longer period of time, use the watering cap filled with
potassium chloride (3 mol/l) of the electrode.
14
Multi 1970iOperation
4Operation
4.1Operating structure
Active and inactive REL
socket
In addition to the pH/ORP sensor, the sensor on the REL1 or REL2
socket can be actuated (switched "active"). The other socket is not actuated ("inactive"). This switching over inside the instrument is made
with the aid of a relay. The active REL socket is shown on the display.
The measured variable on the display can be selected as follows using
the <M> key:
z Press <M> shortly
z Press <M> for a longer
period of time:
Several measuring modes are available within a measured variable.
Switch over between the measured variables with the <▲> or <▼> key.
The options are summarized in the following diagram:
:Switch between the pH/ORP sensor and
the sensor at the active REL socket
Change the active Rel socket
15
OperationMulti 1970i
Note
When a sensor is connected to the active REL socket, the measuring
instrument recognizes the sensor or the measuring cell and automatically switches to the measuring mode that was last active. As soon as
the sensor is disconnected from the active REL socket, the instrument
switches to the pH (mV) measuring mode again.
4.2Switching on the measuring instrument
1Press the <ON/OFF> key.
The display test appears briefly on the display.
After this, the measuring instrument automatically switches to
the measuring mode.
Measuring mode when
switching on
Sensors on the socketMeasuring mode
pHREL
(active)
REL
(inactive)
:
pH or ORP measurement - depending on the last
selected setting
:
::
:
::
:::
Note
The measuring instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the
measuring instrument off if no key has been pressed for an hour.
Last selected measuring mode.
Only the last active
REL socket is taken into account
16
The energy saving feature is not active
z if the power is supplied by the plug-in power supply,
z if the AutoStore function is active,
z if the communication cable and a PC with a running communication
program are connected,
z if the printer cable is connected (for external printers).
Multi 1970iOperation
4.3pH value / ORP voltage
4.3.1General information
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect the pH depth armature or the pH electrode to the mea-
suring instrument.
If necessary, press the <M> key repeatedly until the pH (pH
measurement) or U (measurement of the ORP voltage) display
appears.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3Calibrate or check the measuring instrument with the elec-
trode.
4Using <▲ > <▼>, toggle between the pH or mV measuring
modes.
Note
Incorrect calibration of pH electrodes leads to incorrect measured values. Calibrate regularly before measuring. You can only connect electrodes of the NTC30 type or without temperature sensor.
Warning
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
17
OperationMulti 1970i
Temperature measure-
ment in pH measure-
ments
You can perform pH measurements with or without a temperature sensor as well as with the temperature sensor of an oxygen sensor or a
conductivity measuring cell. The measuring instrument recognizes
which sensors are connected and switches automatically to the correct
mode for the temperature measurement.
The following cases are distinguishable.
Temperature sensorDisplayMode
pHCond or
Oxi
yes-TPAutomatic with
pH temperature sensor
yesyesTP
--Manual
-yesTP
flashes
The temperature value of the
second sensor (Cond or Oxi)
in the same sample is taken
over for the pH measurement*
* If you do not wish that, you can:
– either disconnect the 2nd sensor and use the manual temperature
input or
– use an electrode with a temperature sensor.
If a temperature sensor is connected, it is indicated on the display by
TP.
Note
When calibrating without a temperature sensor (no TP display indicator
displayed), enter the current temperature of the respective buffer solution manually using the <▲> <▼> keys while keeping the <RUN/EN-
TER> key depressed.
18
Multi 1970iOperation
4.3.2Measuring the pH value
1Perform the preparatory activities according to section 4.3.1.
2Immerse the pH electrode in the test sample.
3Press the <▲> <▼> keys repeatedly until pH appears on the
status display. The pH value appears on the display.
AutoRead AR
(Drift control) and hold
function
m
%
1/
°
mg/l
S/
cm
cm
%
C
REL1
AR
pH/mV
Sal
TP
K
/
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
4When measuring without a connected temperature sensor:
The AutoRead function (drift control) checks the stability of the measurement signal. The stability has a considerable impact on the reproducibility of the measured values. With the aid of the hold function the
measured value display is frozen.
6
1
Baud
No.
Ident
Tref25Tref20
Options:
z Determine the current temperature using a thermometer
z TP display indicator not displayed, socket for the second
99
8
2 4 8
Auto
Store
Lin
Oxi
Cal
TEC
ARng
and, while keeping the <RUN/ENTER> key depressed, enter this temperature value with <▲> <▼>.
sensor is free:
Connect the second sensor (Oxi or Cond) and immerse it in
the same sample. TP flashes, the temperature is automatically measured using the second sensor.
With identical measurement conditions, the following criterion is valid
for the AutoRead function:
ReproducibilityResponse time
Better than 0.02> 30 seconds
For D. O. measurements, use the AutoRead function and hold function
like this:
1Call up the pH measuring mode with <▲> <▼>.
19
OperationMulti 1970i
2Activate the AutoRead function with <AUTO READ>.
The current measured value is frozen (hold function).
3Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4If necessary, start the next AutoRead measurement with
<RUN/ENTER>.
5To terminate the AutoRead function: Press the
<AUTO READ> key.
Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
4.3.3Measuring the ORP voltage
In conjunction with an ORP electrode, e.g. SenTix ORP, the measuring
instrument can measure the ORP voltage (U) of a solution.
1Perform the preparatory activities according to section 4.3.1.
2Submerse the ORP electrode in the sample.
3Press the <▲> <▼> key until the U status display appears. The
ORP voltage (mV) of the test sample appears on the display.
4Wait for a stable measured value.
pH/
Store
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
K
/
REL2
Sal
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
Note
ORP electrodes are not calibrated. However, you can check ORP electrodes using a test solution.
1
Baud
Ident
Tref25
No.
6
8
Tref20
Lin
Oxi
23
2 4 8
Auto
Cal
20
Multi 1970iOperation
4.3.4pH calibration
Why calibrate?pH electrodes age. This changes the asymmetry (zero point) and slope
of the pH electrode. As a result, an inexact measured value is displayed. Calibration determines the current values of the asymmetry
and slope of the electrode and stores them in the measuring instrument. Thus, you should calibrate at regular intervals.
When to calibrate?z After connecting another electrode
z When the sensor symbol flashes, i.e. after the calibration interval
has expired
Calibration pointsCalibration can be made with one or two buffer solutions (single-point
or two-point calibration). The measuring instrument determines the following values and calculates the calibration lines as follows:
Determined valuesValues of the calibration lines
1-pointASYz Asymmetry = ASY
z Slope = Nernst slope (59.16 mV/
pH at 25 °C)
2-pointASY
SLO
z Asymmetry = ASY
z Slope = SLO
AutoCal TECis specially matched to the WTW technical buffer solutions as a fully au-
tomatic two-point calibration. The buffer solutions are automatically
recognized by the measuring instrument. Depending on the instrument
setting (see section 4.9 C
ONFIGURATION), the instrument displays the
relevant buffer nominal value or the current electrode voltage in mV.
The calibration can be terminated after the first buffer solution. This corresponds to a single-point calibration. To do this, the instrument uses
the Nernst slope (-59.2 mV/pH at 25 °C) and determines the asymmetry of the electrode.
AutoReadThe calibration procedure automatically activates the AutoRead func-
tion.
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>.
Displaying the calibra-
tion data
Printing the
calibration protocol
You can view the data of the last calibration on the display. The proceeding is described on page 50.
The calibration protocol contains the calibration data of the current calibration. You can transmit the calibration protocol to a printer via the serial interface (see O
INTERFACE, page 53).
UTPUTTINGTHECALIBRATIONPROTOCOLONTHE
21
OperationMulti 1970i
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.8.2
before calibrating. After a valid calibration, the record is printed.
Calibration evaluationAfter calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The
worst evaluation appears on the display.
DisplayAsymmetry
[mV]
-15 ... +15-60.5 ... -58
-20 ... +20-58 ... -57
-25 ... +25-61 ... -60.5
-30 ... +30-62 ... -61
Clean the electrode according
to the electrode operating
manual
Slope
[mV/pH]
or
-57 ... -56
or
-56 ... -50
Preparatory activities
E3
Perform error elimination according to chapter 6 W
DOIF...
HATTO
< -30 or
> 30
< -62 or
> -50
1Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status display pH (pH measurement) or U (measurement of the ORP
voltage) appears.
2Keep the buffer solutions ready.
3Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a
temperature sensor (the TP display indicator is missing from
the display).
23
OperationMulti 1970i
AutoCal TECFor this procedure, use any two WTW technical buffer solutions (pH
values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).
Note
The calibration for pH 10.01 is optimized for the WTW technical buffer
solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions can lead
to an erroneous calibration. The correct buffer solutions are given in the
WTW catalog or in the Internet.
Note
The buffer solutions are automatically recognized by the measuring instrument. Depending on the instrument setting (see
section 4.9 C
er nominal value or the current electrode voltage in mV.
Note
Skip the steps 2 and 7 if you use a pH electrode with temperature sensor or the temperature sensor of a conductivity measuring cell or a D.
O. sensor.
ONFIGURATION), the instrument displays the relevant buff-
Starting the calibration
1Press the <CAL> key. The Ct1 display and the function display
AutoCal TEC appears. The sensor symbol displays the evalu-
ation of the last calibration (or no sensor symbol in the delivery
state or after the measurement parameters have been reset).
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
K
/
Sal
TP
REL2
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
2If required, enter the temperature of the first buffer solution with
3Immerse the pH electrode in the first buffer solution.
4Press the <RUN/ENTER> key.
C
Lin
8
Oxi
T1
2 4 8
Store
Auto
TEC
ARng
Cal
1
Baud
No.
Ident
Tref25Tref20
<▲> <▼> while keeping the <RUN/ENTER> key depressed.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
24
Multi 1970iOperation
mVS/
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH
cm
K
/
REL2
Sal
TP
RCL
/
pH
3
TEC
ARng
mV
%
m
1/
°
mg/l
S/
cm
%
C
REL1
AR
cm
/
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
5When the measured value is stable, Ct2 appears.
1
Baud
No.
Ident
Tref25Tref20
Lin
69
8
2 4 8
Auto
Oxi
Cal
K
Sal
TP
REL2
RCL
pH
O
Sal
1
Time
Day.Month
Year
LoBatnLF
700
Baud
No.
Ident
8
Lin
Oxi
2 4 8
Auto
Cal
Tref25Tref20
TEC
ARng
m
%
1/
°
mg/l
S/
cm
%
C
REL1
AR
pH/mV
cm
Sal
K
/
REL2
TP
RCL
pH
S
O
Sal
Time
Day.Month
Year
LoBatnLF
Note
At this point, the AutoCal TEC calibration can be terminated with <M>.
This corresponds to a single-point calibration. To do this, the instru-
ment uses the Nernst slope (-59.2 mV/pH at 25 °C) and determines the
asymmetry of the electrode.
6Thoroughly rinse the electrode with distilled water.
7If required, enter the temperature of the second buffer solution
8Immerse the pH electrode into the second buffer solution.
9Press the <RUN/ENTER> key.
C
Lin
8
Oxi
T2
2 4 8
Store
Auto
TEC
ARng
Cal
1
Baud
No.
Ident
Tref25Tref20
with <▲> <▼> while keeping the <RUN/ENTER> key depressed.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value appears on the display. Example:
25
Loading...
+ 60 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.