Xylem LF 1100+, 1100T+, 4100+, 5100+, 5100T+ User Manual

Instruction manual for the
SCHOTT
®
Instruments
Conductivity measuring cells type
LF 513T, 613T, 713T, 713T-250, LF 813T, 913T, 913T-ID, LF 1100+, 1100T+, 4100+, 5100+, 5100T+
®
®
Picture of exemplary electrodes
®
SI Analytics GmbH Hattenbergstrasse 10 55122 Mainz Germany Telefon +49 (0) 6131/66 5119 Telefax +49 (0) 6131/66 5001 e-mail: support@si-analytics.com
www.si-analytics.com
Putting into operation
The LF913 T-ID conductivity measuring cell is equipped with an automatic sensor detection system in the connection plug. This system stores the sensor type and the serial number as well as the calibration data. To be able to use the automatic sensor detection feature, you will need a measurement device supporting this function. For further information, please refer to the operating instructions of the measurement device. 2 pole measuring cells (LF1100+,1100T+, 4100+, 5100+, 5100T+, 513T) are delivered platinated. 4 channel and 5 channel conductivity measuring cells (LF613T, 713T 713T-250 and LF 813T, 913T, 913T-ID) are immediately ready for use and do not need to be platinumplated. Please check the cell constant prior to the first use. The interval of the next following verification will depend on the circumstances of use. The conductivity measuring cell (except LF4100+) has to immerse by a minimum of 65 mm (ring mark) in the solution to be measured. The measurement value may be read off as soon as the signal has stabilised. In the case of temperature changes, this may take up to two minutes. The detachable threaded protection cap (only PPO shaft) at the front end of the measurement cell (LF513T, 613T, 813T) will protect the glass tube with the platinum rings and determines the cell constant.
Storage and maintenance
The storage conditions for conductivity measuring cells may be within the range of a 0-50°C and 5-95% relative humidity. Any conditions leading to condensation of water at the plug ought to be avoided. Prior to any extended period of storage, the electrodes (platinum rings) of the measuring cell should be cleaned, if applicable (please refer below), and the measuring cell should be stored in its original cardboard box.
Cleaning
Thorough cleaning is particularly recommended prior to measuring low conductivities. To clean the device, please detach the sensor. Unscrew and remove the protective cap (only LF513T, 613T, 813T) of the mesuring cell. 2 pole measuring cells should only be rinsed with water, if necessary with diluted hydrochloric acid or sodium hydroxide. After mechanical treatment a platination is necessary, therefore the electrode can be sent to SI Analytics. Contamination on the sensor can be removed with the following procedure: Grease or oil, carefully clean the glass part and the ring electrodes using warm water containing a dishwashing detergent and a sponge. In the case of serious contamination, you may also gently remove it with an abrasive powder. In the presence of lime, please place the parts shortly (five minutes) in acetic acid (10Vol%). To clean the protective cap, please use a small bottle brush or a cotton bud. When using acids, leaches and solvents for cleaning, please make sure that they are compatible with the material of the shaft!
An optimum cleaning of the inner glass part of the LF 713 T, LF 713T­250 and LF 913 T-ID conductivity measuring cells is achieved with a tobacco pipe cleaner. The use of acids (exception: HF!), leaches and solvents is uncritical for these cells.
Exemplified check and setting of the cell constant
To verify the cell constant, you need test solutions, for instance the SI Analytics ampoules set LF 995. Immerse the measuring cell, except the LF 4100+, in a suitable test solution by a minimum of 65 mm, for instance in KCl 0.01 mol/l. Then keep the temperature constant and set the corresponding conductivity value on the conductometer, for instance 1.413 mS/cm at 25°C for KCl 0.01 mol/l. For other solutions or different temperatures, please take the appropriate values from the table. To minimise the measuring incertainty over the entire range of use, calibration should be performed both at low and high conductivities. The appropriate cell constant should be set as a function of the measurement range. Setting and reading off of the cell constant is done according to the operating instructions of the conductometer used.
Electrical conductivity mS/cm of KCl solutions as a function of temperature
KCl concentration mol/l
Temp.(°C) 0,001 0,01 0,1 1
18 0,127 1,225 11,19 98,24 19 0,13 1,251 11,43 100,16 20 0,133 1,278 11,67 102,09 21 0,136 1,305 11,91 104,02 22 0,138 1,332 12,15 105,94 23 0,141 1,359 12,39 107,89 24 0,144 1,386 12,64 109,84 25 0,147 1,413 12,88 111,8
Quality
Each electrode must meet the strict quality requirements of final testing. The working life depends mainly on the usage conditions. Hydrofluoric acid, sodium hydroxide and hot phosphoric acid attac glass.
Further information
Further informations can be found in the laboratory cataloque of SI Analytics GmbH.
Subject to technical amendment.
GA_LF513-613-813-913_us_T814838_090504
8
Plug pin configuration 4 pole conductivity cell: LF 613T, 713T, 713T-250 5 pole conductivity cell: LF 813T, 913T, 913T-ID
NTC 30k
8 NTC
6 not used
1 NTC
4 Current electrode 1
Screen + drain wire
2 Voltage electrode 1
5 Voltage electrode 2
3 Current electrode LF 813T, LF 913T, LF 913T-ID, 2a, 2b
7 Screen
Plug pin configuration 2 pole conductivity cell: LF 513 T
8 NTC
NTC 30k
6 not used
1 NTC
4 Current electrode
Pt 1000
2 Voltage electrode
5 Voltage electrode
3 Current electrode
7 Screen
Screen + drain wire
Plug, front view :
2
5
4
3
1
7 6
Plug head configuration Configuration 2 pole conductivity cell: LF 1100+, 1100T+, 4100+, 5100+, 5100T+
4 Pt 1000
5 Pt 1000
1 Current electrode
6 Voltage electrode
3 Voltage electrode
2 Current electrode
Screen + drain wire
SMEK
View on top of
electrode head
5
3
7
4
1
6
Loading...