The HandyLab 200 compact precision conductivity meter enables you
to perform conductivity measurements quickly and reliably.
The HandyLab 200 provides the maximum degree of operating comfort, reliability and measuring certainty for all applications.
The proven procedures for determining or adjusting the cell constant
support your work with the conductivity meter.
1 Keypad
2Display
3 Connectors
ba77063e0103/2014
5
OverviewHandyLab 200
F1
CAL
STO
RCL
ENTER
AR
1.1Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <ENTER>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<ENTER
_>).
<F1>:
_>:
<F1
<F2>:
<On/Off>:Switches the meter on or off
<M>:Selects the measured parameter
<CAL>:
<CAL
_>:
<STO>:Saves a measured value manually
<RCL>: Displays the manually stored measured values
<>: Increments values, scrolls
<>: Decrements values, scrolls
Softkey providing situation dependent functions, e.g.:
<F1>/[Menu]: Opens the menu for measurement settings
_>/[Menu]: Opens the menu for system settings
<F1
Calls up the calibration procedure
Displays the calibration data
<ENTER>:
<ENTER
<AR>Freezes the measured value (HOLD function)
6
_>:
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
Switches the AutoRead measurement on or off
ba77063e0103/2014
HandyLab 200Overview
HOLD AR
25.0
1417
°C
3
2
4
5
6
7
1
μS/cm
[Tr25][nLF][ 0.475 1/cm]
ϰ
15.03.2014
10:00
Menu
1.2Display
1 Status information
2 Measured value (with unit)
3 Measured parameter
4 Sensor symbol (calibration evaluation, calibration interval)
Function display
indicators
5 Measured temperature (with unit)
6 Status line
7 Softkeys and date + time
ErrorAn error occurred during calibration
ARStability control (AutoRead) is active
HOLDMeasured value is frozen (<AR> key)
ba77063e0103/2014
7
OverviewHandyLab 200
1.3Connectors
1 Conductivity measuring cell
2 Service interface
Caution
Only connect measuring cells to the meter that cannot return any
voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary measuring cells fulfill these conditions.
8
ba77063e0103/2014
HandyLab 200Safety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating manual before working with the meter.
The operating manual must always be available within the vicinity of the
meter.
Target groupThe meter was developed for work in the field and in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructionsSafety instructions in this operating manual are indicated by the warn-
ing symbol (triangle) in the left column. The signal word (e.g. "Caution")
indicates the level of danger:
Further notes
Warning
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the instrument
or the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating manuals.
ba77063e0103/2014
9
SafetyHandyLab 200
2.1Authorized use
Authorized use of the meter consists exclusively of the measurement
of conductivity, resistivity, salinity, TDS (total dissolved solids) and temperature in a laboratory or field environment.
The technical specifications as given in chapter 7 T
ECHNICALDATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.
2.2General safety instructions
This meter is constructed and tested in compliance with the IEC 1010
safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICALDATA.
If the meter was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning
of the meter. In this event, wait until the temperature of the meter
reaches room temperature before putting the meter back into operation.
Caution
The meter is only allowed to be opened by authorized personnel.
10
ba77063e0103/2014
HandyLab 200Safety
Safe operationIf safe operation is no longer possible, the meter must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
has been damaged in transport
has been stored under adverse conditions for a lengthy period of
time
is visibly damaged
no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
EEC directives for protective labor legislation
National protective labor legislation
Safety regulations
Safety datasheets of the chemical manufacturers.
Caution
In addition to the safety instructions mentioned here, also follow
the safety instructions of the sensors used.
ba77063e0103/2014
11
CommissioningHandyLab 200
2
1
3Commissioning
3.1Scope of delivery
Conductivity meter HandyLab 200
4 batteries 1.5 V Mignon type AA
Short instructions
CD-ROM with
– detailed operating manual
3.2Initial commissioning
Perform the following activities:
Insert the supplied batteries
Switching on the meter
Set the date and time
3.2.1Inserting the batteries
1Unscrew the four screws (1) on the underside of the meter.
2Open the battery compartment (2) on the underside of the
meter.
12
ba77063e0103/2014
HandyLab 200Commissioning
3Place four batteries (type Mignon AA) in the battery compart-
ment.
4Close the battery compartment.
Caution
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the
battery compartment.
Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.
3.2.2Switching on the meter
± signs in the
1Press the <On/Off> key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test
is being performed.
Subsequently, the meter switches to the measuring mode
(measured value display).
Note
The meter has an energy saving feature to avoid unnecessary battery
depletion.
The energy saving feature switches off the meter if no key is pressed
during the adjusted interval. (How to set the switch-off interval, see section 4.3.1).
3.2.3Setting the date and time
1See section 4.2.4
ba77063e0103/2014
13
OperationHandyLab 200
ϰ
1417
µS/cm
25.0
°C
[Tr25][nLF][0.475 1/cm]
15.03.2014
10:00
Menu
Menu
4Operation
4.1Switching on the meter
Switching onPress the <On/Off> key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.
Switching offPress the <On/Off> key.
Automatic switch-off
function
Display illuminationThe meter automatically switches off the display illumination if no key
The instrument has an automatic switch-off function in order to save the
batteries (see section 4.3.1). The automatic switch-off function
switches off the meter if no key is pressed for an adjustable period.
is pressed for 30 seconds. The illumination is switched on with the next
keystroke again.
You can also generally switch the display illumination on or off (see
section 4.3.1).
14
ba77063e0103/2014
HandyLab 200Operation
4.2General operating principles
This section contains basic information on the operation of the
HandyLab 200.
Operating elements,
display
Operating modes,
navigation
An overview of the operating elements and the display is given in section 1.1 and section 1.2.
An overview of the operating modes and navigation of the
HandyLab 200 is given in section 4.2.1 and section 4.2.2.
4.2.1Operating modes
The instrument has the following operating modes:
Measuring
The measurement data of the connected sensor are shown in the
measured value display
Calibration
The course of a calibration with calibration information, functions
and settings is displayed
Storing in memory
The meter stores the measurement data manually
Setting
The system menu or a sensor menu with submenus, settings and
functions is displayed
ba77063e0103/2014
15
OperationHandyLab 200
General
Clock
Service information
Reset
System
15.03.2014
10:00
Back
Language:Deutsch
Illumination:On
Contrast
:48 %
Switchoff time
:30 min
General
15.03.2014
10:00
Back
4.2.2Navigation
Measured value displayIn the measured value display, you can
open the menu for calibration and measurement settings with <F1>
(short
keystroke)
Open the Storage & config menu with the sensor-independent set-
tings with <F1
_> (long keystroke (approx. 2 s).
Change the display in the measurement window by pressing <M>
(e.g. conductivity −> resistivity −> −> −> ).
Menus and dialogsThe menus for settings and dialogs in procedures contain further sub-
elements. The selection is made with the <><> keys. The current
selection is displayed with a frame.
Submenus
The name of the submenu is displayed at the upper edge of the
frame. Submenus are opened by confirming with <ENTER>. Example:
16
Settings
Settings are indicated by a colon. The current setting is displayed on
the right-hand side. The setting mode is opened with <ENTER>.
Subsequently, the setting can be changed with <><> and
<ENTER>. Example:
ba77063e0103/2014
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.