Sirona Sinius CS, Sinius TS, Sinius Installation Requirements

New as of:
Installation Requirements
03.2017
Sinius / Sinius CS / Sinius TS
English
=
Installation requirements
63 22 668 D3561 D3561.021.01.09.02 03.2017
3
Dentsply Sirona Table of contents Installation Requirements Sinius / Sinius CS / Sinius TS
bеЦдблЬ
Table of contents
1
General information................................................................................................. 5
1.1 Notes on the installation prerequisites ......................................................... 5
1.2 Structure of the document............................................................................ 6
1.2.1 Identification of the danger levels.................................................... 6
1.2.2 Formats and symbols used ............................................................. 6
2
Safety information.................................................................................................... 7
2.1 Installation by qualified personnel ................................................................ 7
2.2 Radiotelephones .......................................................................................... 7
2.3 Modifications and extensions of the system................................................. 7
2.4 Power connection......................................................................................... 8
3
On-site installation................................................................................................... 9
3.1 Substrate, floor............................................................................................. 9
3.1.1 Load-bearing capacity of the floor................................................... 9
3.1.2 Tensile forces.................................................................................. 9
3.1.3 Stability............................................................................................ 10
3.2 Requirements of the supply media............................................................... 11
3.2.1 Requirements of the water supply................................................... 12
3.2.1.1 Condition of the water....................................................... 12
3.2.1.2 Connection to the public drinking water system ............... 13
3.2.1.3 Installation of the water supply on the building side ......... 14
3.2.2 Requirements of the air supply........................................................ 15
3.2.2.1 Condition of the air ........................................................... 15
3.2.2.2 Installation of air supply.................................................... 15
3.2.3 Requirements of the suction machine............................................. 16
3.2.3.1 Typical pressure loss in the suction system ..................... 16
3.2.4 Requirements of the waste water installation.................................. 17
3.3 Underfloor installation of supply lines........................................................... 18
3.3.1 Installation template ........................................................................ 18
3.3.2 Installation of the supply lines in the termination panel................... 20
3.4 Cleaning the air and water pipes.................................................................. 21
3.5 Underfloor installation of the PC connections .............................................. 22
4
Dimensions, technical data...................................................................................... 24
4.1 Sinius dimensions, scale bar 1:20................................................................ 24
4.1.1 Distances in the treatment room ..................................................... 24
4.1.2 Side view ......................................................................................... 25
4.1.3 Top view.......................................................................................... 26
63 22 668 D3561
4 D3561.021.01.09.02 03.2017
Table of contents Dentsply Sirona
Installation Requirements Sinius / Sinius CS / Sinius TS
4.1.4 Top view with options....................................................................... 27
4.1.5 Distances in the treatment room with Heliodent Plus....................... 28
4.1.6 Side view with HELIODENT Plus..................................................... 29
4.2 Sinius CS dimensions, scale bar 1:20........................................................... 30
4.2.1 Distances in the treatment room ...................................................... 30
4.2.2 Side view.......................................................................................... 31
4.2.3 Top view........................................................................................... 32
4.2.4 Top view with options....................................................................... 33
4.2.5 Distances in the treatment room with Heliodent Plus....................... 34
4.2.6 Side view with HELIODENT Plus..................................................... 35
4.3 Sinius TS dimensions, scale bar 1:20 ........................................................... 36
4.3.1 Distances in the treatment room ...................................................... 36
4.3.2 Side view.......................................................................................... 37
4.3.3 Top view........................................................................................... 38
4.3.4 Top view with options....................................................................... 39
4.3.5 Distances in the treatment room with Heliodent Plus....................... 40
4.3.6 Side view with HELIODENT Plus..................................................... 41
4.4 Mounting plates............................................................................................. 42
4.5 Information on planning for the practice........................................................ 44
4.6 Technical data............................................................................................... 45
4.6.1 Operating, transport and storage conditions.................................... 47
4.6.2 Weight and packaging...................................................................... 48
4.7 Standards/Approvals..................................................................................... 49
5
Electromagnetic compatibility................................................................................... 51
5.1 Accessories................................................................................................... 51
5.2 Electromagnetic emission ............................................................................. 52
5.3 Interference immunity.................................................................................... 53
5.4 Working clearances....................................................................................... 56
5.5 Foot control wireless interface ...................................................................... 57
6
Checklist................................................................................................................... 58
6.1 Installation site .............................................................................................. 58
6.2 Construction requirements ............................................................................ 59
6.3 IT hardware ................................................................................................... 60
6.4 Network ......................................................................................................... 60
6.5 Data processing ............................................................................................ 61
63 22 668 D3561 D3561.021.01.09.02 03.2017
5
Dentsply Sirona 1General information Installation Requirements Sinius / Sinius CS / Sinius TS 1.1Notes on the installation prerequisites
bеЦдблЬ
1
General information
1.1
Notes on the installation prerequisites
This document describes the installation prerequisites for all versions of the Sinius / Sinius CS / Sinius TS treatment center.
It contains the following information:
● Required information for practice planning.
● Information for the installer and operator regarding the necessary quality of the air and water supply media.
● Information for the installer, such as how to implement the connections for air, water, waste water, suction air and the power supply.
● Information on the cabling of the PCs to be connected.
● Information about electromagnetic compatibility and the prerequisites for setting up the treatment center.
● A checklist to ensure that all installation requirements have been fulfilled.
The subsequent installation of the treatment center is described in the installation instructions (REF 6322627).
You will also need the drilling template (REF 63 34 291) for the safe and secure attachment of the treatment center to the floor.
63 22 668 D3561
6 D3561.021.01.09.02 03.2017
1General information Dentsply Sirona
1.2Structure of the document Installation Requirements Sinius / Sinius CS / Sinius TS
1.2
Structure of the document
1.2.1 Identification of the danger levels
To prevent personal injury and material damage, please observe the warning and safety information provided in this document. This information is highlighted as follows:
Tip: Information for facilitating work.
1.2.2 Formats and symbols used
The formats and symbols used in this document have the following meaning:
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possible dangerous situation that could result in serious bodily injury or death.
CAUTION
A possible dangerous situation that could result in slight bodily injury.
NOTICE
A possible harmful situation which could lead to damage of the product or an object in its environment.
IMPORTANT
Application instructions and other important information.
Prerequisite
1. First action step
2. Second action step or
Alternative action
Result
➢ Individual action step
Requests you to do something.
see “Formats and symbols used [ → 6]”
Identifies a reference to another text passage and specifies its page number.
● List Identifies a list.
“Command / menu item” Identifies commands, menu items, or
quotations.
63 22 668 D3561 D3561.021.01.09.02 03.2017
7
Dentsply Sirona 2Safety information Installation Requirements Sinius / Sinius CS / Sinius TS 2.1Installation by qualified personnel
bеЦдблЬ
2
Safety information
2.1
Installation by qualified personnel
The installation of the supply connections must be carried out by qualified technicians only.
Electrical and water supply installations; reference for INST suction machines
2.2
Radiotelephones
Mobile RF communications equipment can affect electro-medical equipment. Therefore, the use of mobile wireless phones in medical office or hospital environments must be prohibited.
2.3
Modifications and extensions of the system
Modifications to this unit which might affect the safety of the system owner, patients or other persons are prohibited by law.
For reasons of product safety, this product may be operated only with original Dentsply Sirona accessories or third-party accessories expressly approved by Dentsply Sirona. The user is responsible for any damage resulting from the use of non-approved accessories.
If any units not approved by Dentsply Sirona are connected, they must comply with the applicable standards, e.g.:
● IEC 60950-1 or IEC 62368-1 for information technology equipment (e.g. PC, Ethernet switch, Ethernet hub, Ethernet converter, USB repeater, Ethernet router, USB switch, HDMI repeater/amplifier, audio sources, display port repeater/amplifier, etc. and their power supply units), as well as
● IEC 60601-1 for medical devices.
The treatment center monitor must fulfill the requirements of the IEC 60950-1 or IEC 62368-1 standard.
The loudspeaker port of the monitor may only be connected to a device that complies with the IEC 60950-1, IEC 62368-1 (e.g. a PC) or IEC 60601-1 standard. Under no circumstances may it be connected to a stereo system or similar.
If a system is created during the installation process, the requirements of IEC 60601-1, 3rd edition, must be fulfilled. The system author assumes responsibility for conformity of the system with 93/42/EEC Directive, for example.
WARNING
Professional installation
Comply with the national regulations for electrical installation (e.g. IEC 60364-1, VDE 0100-100, IEC 60364-7-710, VDE 0100-710, National Electrical Code).
Comply with the national regulations for water supply installations (e.g. EN 1717) and sewage installations (e.g. EN 12056-1).
63 22 668 D3561
8 D3561.021.01.09.02 03.2017
2Safety information Dentsply Sirona
2.4Power connection Installation Requirements Sinius / Sinius CS / Sinius TS
2.4
Power connection
WARNING
Shock hazard
It is essential that you switch off the power supply PRIOR to beginning the installation. There is a risk of electric shock. People can be injured or electrical components of the unit destroyed.
63 22 668 D3561 D3561.021.01.09.02 03.2017
9
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.1Substrate, floor
bеЦдблЬ
3
On-site installation
3.1
Substrate, floor
Unevenness The floor must be level and horizontal in accordance with DIN 18 202.
In the case of an uneven floor, the steel adapter plate must be used; see Mounting plates [ → 42].
3.1.1 Load-bearing capacity of the floor
Load capacity:
The minimum load-bearing capacity of the floor must be 0.5 N/cm
2
(corresponds to around 500 kg/m
2
).
3.1.2 Tensile forces
The floor must be able to accommodate tensile forces of > 1 kN at each of the marked screw fastening points (A).
It may be necessary to reinforce the surface beneath the floor.
When using a mounting plate (see Mounting plates [ → 42]) this requirement is mandatory.
0,5
N/cm
2
0,7
lbs/sq in.
A
63 22 668 D3561
10 D3561.021.01.09.02 03.2017
3On-site installation Dentsply Sirona
3.1Substrate, floor Installation Requirements Sinius / Sinius CS / Sinius TS
3.1.3 Stability
With a concrete floor
● Use a 12 mm dia. masonry drill to drill the holes.
● The depth of the drill hole (C) in the concrete floor must be at least 80 mm.
● The specified drilling depth (C) of at least 80 mm refers to the drilling length without screed (A) or impact sound insulation (B). The heavy duty anchor bolts included with delivery (2 units) are suitable for a clamping length ((A)+(B)) of up to 100 mm.
● Use 12 mm dia. wall plugs (but do not use wall plugs instead of heavy duty anchor bolts).
● Use 10x160 mm wood screws including washers.
With a wooden floor
● Provide and prepare appropriate supporting beams.
● Use a 7mm dia. wood drill to drill the holes.
● The depth of the drill hole must be at least 80mm.
● Use 10x80mm wood screws including washers.
ATTENTION! Do not use wall plugs!
Ø 12mmØ 12mm
20mm
7mm
160mm160
mm
80 mm min.
80 mm min.
C
B
A
100 mm
max.
100 mm
max.
WARNING
Any existent screed (A) and/or impact sound insulation (B) must be penetrated!
Ø 7mmØ 7mm
7mm
80 mm min.
80 mm min.
63 22 668 D3561 D3561.021.01.09.02 03.2017
11
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2Requirements of the supply media
bеЦдблЬ
3.2
Requirements of the supply media
Suction machine
Compressed air (free from oil) The compressor must be able to draw in air in an efficient, hygienic manner.
Cold water (drinking water quality)
K Steam trap
63 22 668 D3561
12 D3561.021.01.09.02 03.2017
3On-site installation Dentsply Sirona
3.2Requirements of the supply media Installation Requirements Sinius / Sinius CS / Sinius TS
3.2.1 Requirements of the water supply
3.2.1.1
Condition of the water
The water that supplies the treatment center must meet the following requirements to avoid operating faults.
Water pressure: 2.5bar (36psi) to 6bar (87psi)
Minimum flow rate:
3 l/min
Water hardness (overall hardness):
at > 2.4 mmol/l (= 12° dH), install a water softener.
Set dilution hardness to 1.43 mmol/l (= 8° dH). pH value: 6.5 to 8.5 (recommended) Particle size: ≤ 100 µm Water quality: The water quality must comply with the national
requirements for drinking water. Colony count: The colony count must fulfill the national regulations
for drinking water and must not exceed 500 CFU/ml
under any circumstances (CFU: Colony forming
units).
In the event of an increased colony count, the indoor
installation must be checked and the cause of
contamination eliminated. Alternatively you can also
change over to a stand-alone water supply for a brief
period.
Sampling and the colony count must be performed
by a competent laboratory.
Before installation of the treatment center, an
acceptable microbiological water quality for the
domestic water supply must be ensured and
documented in the form of a colony count.
63 22 668 D3561 D3561.021.01.09.02 03.2017
13
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2Requirements of the supply media
bеЦдблЬ
3.2.1.2
Connection to the public drinking water system
Treatment center isolated from the public drinking water supply
The treatment center, if it is equipped with a disinfection system, complies with the requirements of EN 1717 (free discharge with an isolation distance ≥ 20 mm) and the DVGW German Gas and Water Association). It is intrinsically safe according to the W540 worksheet and thus also complies with W270 and KTW (Guideline on the use of plastics in contact with drinking water). It can be connected directly to the public drinking water system.
The treatment center then has the "DVGW" label next to the rating plate.
Treatment center not isolated from the public drinking water supply
If compliance with EN 1717 is required in the country, the respective measures for the protection of public drinking water must be taken outside the treatment center.
This applies to models without a disinfection system.
The treatment center then does not have the "DVGW" label.
Please comply with the national requirements with regard to connecting treatment centers to the public drinking water system.
63 22 668 D3561
14 D3561.021.01.09.02 03.2017
3On-site installation Dentsply Sirona
3.2Requirements of the supply media Installation Requirements Sinius / Sinius CS / Sinius TS
3.2.1.3
Installation of the water supply on the building side
● Carry out the installation in accordance with the national installation requirements (e.g. EN 1717).
● Connect the treatment center to the cold water supply.
● Use fine filters with a mesh size between ≤ 80 µm (0.08 mm) and ≤ 100 µm (0.1 mm).
● In order to reduce the microorganisms in the water supply pipe, take account of the following when laying this pipe to the treatment center:
– Avoid long stub lines to the treatment center.
– Carry out the installation so that, where possible, other main
consumers (such as the sink) are fed from the same line downstream of the treatment center connection.
– Avoid laying the supply line parallel to hot water pipes.
● Recommendation: Install a corner valve in the water supply to the treatment center with 2 outlets and 2 stopcocks. The second outlet makes it easy to take a sample of the water for microbiological testing.
63 22 668 D3561 D3561.021.01.09.02 03.2017
15
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2Requirements of the supply media
bеЦдблЬ
3.2.2 Requirements of the air supply
The air supply is required to drive the turbines, cool the drill drives, for the cooling spray and to control the treatment center.
3.2.2.1
Condition of the air
3.2.2.2
Installation of air supply
● Carry out the installation in accordance with the national requirements.
● Install a steam trap.
Air pressure: 5.5bar (80psi) to 7.5bar (109psi)
Minimum flow rate: 50l/min Minimum flow rate with air
jet pump:
80 l/min
Air humidity: Dew point ≤ -20°C at atmospheric pressure Oil content:
≤ 0.5mg/m
3
Particle content: ≤ 100 particles of size 1 - 5 µm
in one m
3
of air (larger particles are filtered by the treatment center.)
Hygiene: Hygienically acceptable air
IMPORTANT! The treatment center does not require medical air.
63 22 668 D3561
16 D3561.021.01.09.02 03.2017
3On-site installation Dentsply Sirona
3.2Requirements of the supply media Installation Requirements Sinius / Sinius CS / Sinius TS
3.2.3 Requirements of the suction machine
● Vacuum in the suction machine at the connection point with the treatment center: min. 0.12 bar (1.7 psi), max. 0.18 bar (2.6 psi).
● With a vacuum of > 0.18 bar (2.6 psi) with no flow, the treatment center must be fitted with the "Vacuum limiter" retrofit kit (REF 59 68
826).
● Minimum suction power: 500 Nl/min
● For the suction line: adhere to the specifications in the installation instructions for "Suction machines".
● Suction system Type 1: High flow rate
● Wet or dry suction (depending on equipment)
● Suction handpiece diameter:
– 6 mm: Saliva ejector and surgical suction – 11 mm: Spray aspirator
● Filter mesh size 1.8 x 2.6 mm
● Sirona recommends using cannulae with a bypass in order to prevent return flow in adhered cannulas.
3.2.3.1
Typical pressure loss in the suction system
Suction flow of large suction hose
red ( ○ ): Only large suction device black ( ◊ ): Saliva ejector in parallel
63 22 668 D3561 D3561.021.01.09.02 03.2017
17
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.2Requirements of the supply media
bеЦдблЬ
3.2.4 Requirements of the waste water installation
● The quantity of waste water from the treatment center is max. 3 l/min.
– Carry out the waste water installation accordingly.
– Fit the waste water pipe in accordance with the information in the
table. The gradient of the waste water pipe must be at least 1°.
63 22 668 D3561
18 D3561.021.01.09.02 03.2017
3On-site installation Dentsply Sirona
3.3Underfloor installation of supply lines Installation Requirements Sinius / Sinius CS / Sinius TS
3.3
Underfloor installation of supply lines
3.3.1 Installation template
We recommend that you order the installation template (REF 33 15 830) from Sirona for laying the pipe ends in the installation field.
If necessary, you can also prepare the template yourself based on the diagram below. The specified dimensions of the segments are the dimensions of the core boreholes.
63 22 668 D3561 D3561.021.01.09.02 03.2017
19
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.3Underfloor installation of supply lines
bеЦдблЬ
1 Water supply
Pipe 10x1 mm, corner valve outlet 3/8"
2 Compressed air supply
Pipe 10x1 mm, corner valve outlet 3/8"
3 Suction line
DN40 HT-PP ISO 8283-3 (polypropylene, inner diameter approx. 36.5 mm)
4 Water drainage
DN40 HT-PP ISO 8283-3 (polypropylene, inner diameter approx. 36.5 mm)
5 Installation pipe (power supplies)
inner diameter approx. 40 mm)
6 Control cable to relays for the
suction machine ( ), call cable ( ),
Special function (#) 3 x 1.5 mm
2
(AWG16) (quality as in the power cable)
7 Power supply
3x1.5 mm
2
(AWG16) Circuit breaker: for 230 V AC: 16 A medium-blow for 100 - 115 V AC: 20 A medium-blow Recommendation: Type B automatic circuit breaker
8
Option: Heliodent Plus
radiation cable (10 m) (free end 300 ± 50 mm) CAUTION The Heliodent Plus radiation cable cannot be laid together with USB or HDMI cables. If required, allow for a separate installation pipe.
9 Installation pipe (IT)
inner diameter min. 50 mm (or corresponding flat conduit) for the PC connection
%
63 22 668 D3561
20 D3561.021.01.09.02 03.2017
3On-site installation Dentsply Sirona
3.3Underfloor installation of supply lines Installation Requirements Sinius / Sinius CS / Sinius TS
3.3.2 Installation of the supply lines in the termination panel
An installation template is available or was prepared.
1. Check the position of the supply lines against the installation template as per the practice blueprint. Ensure that sufficient space is provided between the lines and the walls; see "Scale 1:20". The center of the hole in the floor must be 269 mm (10 5/8") from the foot of the treatment center.
2. Lay the ends of the supply pipes, corner valves and lines as shown in the illustrations.
ª The top edge of the corner valves for air and water must not project
more than 60 mm from the top edge of the floor.
ª The suction and drainage pipes must be flush with the top edge of the
floor (a deviation of +5 mm is permissible). The inner diameter of both pipes is 36.5 mm.
ª The electric lines must project by at least 500 mm. ª The supply lines are laid.
Ø 50 mm
Ø 36,5 mm
>500mm
>20“
< 3/16“
< 5mm
< 60mm
< 2 3/8“
Ø 36,5 mm
3/8“
63 22 668 D3561 D3561.021.01.09.02 03.2017
21
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.4Cleaning the air and water pipes
bеЦдблЬ
3.4
Cleaning the air and water pipes
NOTICE
Chips and other foreign materials could be flushed and/or blown into the treatment center.
Metal chips can cause malfunctions of the pneumatic components. The filters become clogged with foreign materials.
➢ During installation, ensure that no chips or other foreign materials
enter the lines. ➢ Flush the water lines. ➢ Blow out the air lines. ➢ Ensure that no more foreign materials can enter the lines after they
have been flushed or blown out.
10l
>30 sec
Versorgung durch den Fußboden
Supply through the floor
Alimentation à travers du plancher
Alimentación a través del suelo
63 22 668 D3561
22 D3561.021.01.09.02 03.2017
3On-site installation Dentsply Sirona
3.5Underfloor installation of the PC connections Installation Requirements Sinius / Sinius CS / Sinius TS
3.5
Underfloor installation of the PC connections
Depending on the prevailing local conditions, the existing cable set can be installed in the cable duct of an underfloor installation by an installer.
Cable channel No. 9 is used for this purpose, see Installation template [ → 18].
Cable sets for PC connection with HDMI and USB cable for camera SiroCam AF / AF+
● 5 m: REF 64 46 111
● 10 m: REF 63 29 655
Running cables to the PC
Lines L343 (USB repeater), L339 (Ethernet), L406 (HDMI) and protective ground wire. For PCs without a HDMI output, the Audio line is also required.
To prevent transmission interference, ensure that the cables are not crossed.
A cable duct is laid from the treatment center to the location of the PC.Free length A of cables at the treatment center end: Length A = 600
mm
1. Pull the lines L343 (USB repeater), L339 (Ethernet), L406 (HDMI) and protective ground wire of the treatment center through the cable duct to the location of the PC C. For PCs without a HDMI output, insert the Audio line. For the USB line L343 the TYPE A connector must be on the PC side and the TYPE B connector on the chair side.
IMPORTANT
The cable set for the PC connection must not be routed in the same cable channel as any existing radiation cables for a Heliodent Plus unit model!
NOTICE
Electric lines are susceptible to breakages.
Any kinks or twists in the cables could damage their wires. You must then replace such cables.
➢ Ensure that electrical lines do not become kinked or twisted.
C
A
63 22 668 D3561 D3561.021.01.09.02 03.2017
23
Dentsply Sirona 3On-site installation Installation Requirements Sinius / Sinius CS / Sinius TS 3.5Underfloor installation of the PC connections
bеЦдблЬ
2. Save the accessory parts for final installation! ª The preparation of the connection for the underfloor installation of the
PC is completed.
IMPORTANT
Minimum requirements for PC
See document "Installation instructions and system requirements for PC configuration," (REF 61 94 075 SIVISION digital.
63 22 668 D3561
24 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.1Sinius dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4
Dimensions, technical data
4.1
Sinius dimensions, scale bar 1:20
Dentist element with travel track
4.1.1 Distances in the treatment room
A Recommended distances from cabinet or wall. B Center of the floor cut-out/installation area C Minimum distance with tray and/or Heliodent Plus D Hazard warning: The lamp installed here and the tray and
Heliodent Plus have a swivel range which exceeds the specified distances!
63 22 668 D3561 D3561.021.01.09.02 03.2017
25
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.1Sinius dimensions, scale bar 1:20
bеЦдблЬ
4.1.2 Side view
A Height and swivel range of the lamp B LEDview Plus C LEDview D Monitor on lamp support tube
63 22 668 D3561
26 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.1Sinius dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.1.3 Top view
1191
46 7/8”
456
18”
355
14”
1059
41 11/16”
max.
2180
85 3/4”
801
31 1/2”
1735
68 1/4”
30°
269
10 5/8”
1930
76”
A
B
410
16 1/8”
42°
90°
max.
200
7 7/8”
14°
48°
180°
max.
2220
87 3/8”
1950
76 3/4”
D
C
1935
76 3/8”
90°
115°
A Double-jointed headrest: Length of treatment center with
patient height of 176cm and default program 2
B Double-jointed headrest: Maximum length of the treatment
center
C Motorized headrest: Length of treatment center with patient
height of 176cm and default program 2
D Motorized headrest: Maximum length of the treatment center
63 22 668 D3561 D3561.021.01.09.02 03.2017
27
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.1Sinius dimensions, scale bar 1:20
bеЦдблЬ
4.1.4 Top view with options
A Tray B LEDview Plus C LEDview D Monitor on lamp support tube E Projection of tray arm F Projection of lamp arm
63 22 668 D3561
28 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.1Sinius dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.1.5 Distances in the treatment room with Heliodent Plus
A Recommended distances from cabinet or wall. B Center of the floor cut-out/installation area C Minimum distance with tray and/or Heliodent Plus DIMPORTANT!
The lamp installed here and the tray and Heliodent Plus have a swivel range which exceeds the specified distances!
E Heliodent Plus wall adapter
Observe radiation protection regulations!
F Support arm with Heliodent Plus tube assembly
63 22 668 D3561 D3561.021.01.09.02 03.2017
29
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.1Sinius dimensions, scale bar 1:20
bеЦдблЬ
4.1.6 Side view with Heliodent Plus
A Recommended installation height for wall adapter: 1110 mm (43
3/4“) B Cable bushing for radiation cable C The radiation cable between Sinius and the wall adapter are
supplied (12.5 mm outer diameter, max. cable length 10 meters
(393”).
CAUTION
The Heliodent Plus radiation cable cannot be laid together with
USB or HDMI cables. If required, allow for a separate installation
pipe.
63 22 668 D3561
30 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.2Sinius CS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.2
Sinius CS dimensions, scale bar 1:20
Dentist element with swivel arms
4.2.1 Distances in the treatment room
1250
49 1/4”
1300
51 3/16”
2250
88 9/16”
410
16 1/8”
A
269
10 5/8”
B
141
5 9/16”
Pos. 1
Pos. 2
D
C
A Recommended distances from cabinet or wall. B Center of the floor cut-out/installation area C Hazard warning: The lamp, dentist element / with tray
(optional) and the Heliodent Plus installed here have a swivel range which exceeds the specified distances!
D The dentist element and the lamp can be installed either in
Position 1 or Position 2. (Not possible with Heliodent Plus on the lamp support tube)
63 22 668 D3561 D3561.021.01.09.02 03.2017
31
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.2Sinius CS dimensions, scale bar 1:20
bеЦдблЬ
4.2.2 Side view
A Height and swivel range of the lamp B LEDview Plus C LEDview D Monitor on lamp support tube E Projection of the lamp arm in Position 1 and Position 2
63 22 668 D3561
32 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.2Sinius CS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.2.3 Top view
1275
50 1/4”
456 18”
355 14”
1059
41 11/16”
max.
200
7 7/8”
1935
76 3/8”
269
10 5/8”
180°
14°
48°
90°
115°
max.
2180
85 3/4”
1930
76”
A
B
max.
2220
87 3/8”
1950
76 3/4”
D
C
A Double-jointed headrest: Length of treatment center with
patient height of 176cm and default program 2
B Double-jointed headrest: Maximum length of the treatment
center
C Motorized headrest: Length of treatment center with patient
height of 176cm and default program 2
D Motorized headrest: Maximum length of the treatment center
63 22 668 D3561 D3561.021.01.09.02 03.2017
33
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.2Sinius CS dimensions, scale bar 1:20
bеЦдблЬ
4.2.4 Top view with options
A LEDview Plus B LEDview C Monitor on lamp support tube D Projection of the lamp arm in Position 1 and Position 2
63 22 668 D3561
34 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.2Sinius CS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.2.5 Distances in the treatment room with Heliodent Plus
A Recommended distances from cabinet or wall. B Center of the floor cut-out/installation area C Minimum distance with lamp and Heliodent Plus DIMPORTANT!
The lamp, dentist element / with tray (optional) and the Heliodent Plus installed here have a swivel range which exceeds the specified distances!
E Heliodent Plus wall adapter
Observe radiation protection regulations!
F Support arm with Heliodent Plus tube assembly
63 22 668 D3561 D3561.021.01.09.02 03.2017
35
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.2Sinius CS dimensions, scale bar 1:20
bеЦдблЬ
4.2.6 Side view with Heliodent Plus
A Recommended installation height for wall adapter: 1110 mm (43
3/4“) B Cable bushing for radiation cable C The radiation cable between Sinius and the wall adapter are
supplied (12.5 mm outer diameter, max. cable length 10 meters
(393”).
CAUTION
The Heliodent Plus radiation cable cannot be laid together with
USB or HDMI cables. If required, allow for a separate installation
pipe.
63 22 668 D3561
36 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.3Sinius TS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.3
Sinius TS dimensions, scale bar 1:20
Dentist element as swivel table
4.3.1 Distances in the treatment room
1250
49 1/4”
1100
43 5/16”
2440
96 1/16”
450
17 3/4”
A
C
269
10 5/8”
B
181
7 1/8”
A Recommended distances from cabinet or wall. B Center of the floor cut-out/installation area C Hazard warning: The lamp, dentist element / with tray
(optional) and the Heliodent Plus installed here have a swivel range which exceeds the specified distances!
63 22 668 D3561 D3561.021.01.09.02 03.2017
37
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.3Sinius TS dimensions, scale bar 1:20
bеЦдблЬ
4.3.2 Side view
A Height and swivel range of the lamp B LEDview Plus C LEDview D Monitor on lamp support tube
63 22 668 D3561
38 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.3Sinius TS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.3.3 Top view
160°
270°
180°
1651
65”
456 18”
355 14”
1059
41 11/16”
max.
200
7 7/8”
1935
76 3/8”
269
10 5/8”
180°
14°
48°
90°
115°
max.
2180
85 3/4”
1930
76”
A
B
max.
2220
87 3/8”
1950
76 3/4”
D
C
A Double-jointed headrest: Length of treatment center with
patient height of 176cm and default program 2
B Double-jointed headrest: Maximum length of the treatment
center
C Motorized headrest: Length of treatment center with patient
height of 176cm and default program 2
D Motorized headrest: Maximum length of the treatment center
63 22 668 D3561 D3561.021.01.09.02 03.2017
39
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.3Sinius TS dimensions, scale bar 1:20
bеЦдблЬ
4.3.4 Top view with options
A LEDview Plus B LEDview C Monitor on lamp support tube D Projection of the fixed dentist element support arm
63 22 668 D3561
40 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.3Sinius TS dimensions, scale bar 1:20 Installation Requirements Sinius / Sinius CS / Sinius TS
4.3.5 Distances in the treatment room with Heliodent Plus
A Recommended distances from cabinet or wall. B Center of the floor cut-out/installation area C Minimum distance with lamp and/or Heliodent Plus DIMPORTANT!
The lamp, dentist element / with tray (optional) and the Heliodent Plus installed here have a swivel range which exceeds the specified distances!
E Heliodent Plus wall adapter
Observe radiation protection regulations!
F Support arm with Heliodent Plus tube assembly
63 22 668 D3561 D3561.021.01.09.02 03.2017
41
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.3Sinius TS dimensions, scale bar 1:20
bеЦдблЬ
4.3.6 Side view with Heliodent Plus
A Recommended installation height for wall adapter: 1110 mm (43
3/4“) B Cable bushing for radiation cable C The radiation cable between Sinius and the wall adapter are
supplied (12.5 mm outer diameter, max. cable length 10 meters
(393”).
CAUTION
The Heliodent Plus radiation cable cannot be laid together with
USB or HDMI cables. If required, allow for a separate installation
pipe.
63 22 668 D3561
42 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.4Mounting plates Installation Requirements Sinius / Sinius CS / Sinius TS
4.4
Mounting plates
Adapter plate
If the dental treatment center is being installed as a replacement for an M1 previously installed at this location, an adapter plate is available for this purpose. You can use the existing drill holes for fastening. Two additional drill holes are required only in the area of the water unit. The treatment center is screwed onto the steel plate using M10 screws. The adapter plate also must be used if the unevenness of the floor exceeds 2 mm in the vicinity of the base plate.
Note: If a C-Line unit (e.g. C2
+
, M1+ ) was previously installed, the
fastening drill holes must be re-drilled.
Adapter plate thickness: 12 mm
Drilling with floor (7x)
REF 63 07 511 Sinius adapter plate
63 22 668 D3561 D3561.021.01.09.02 03.2017
43
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.4Mounting plates
bеЦдблЬ
Demonstration chair plate
For floors which do not permit permanent connection of the unit (e.g. demo operation at a trade show, floor heating), installation on a steel demo plate is possible.
The treatment center is screwed onto the steel plate using M10 screws.
If the demo chair plate is permanently installed and whenever it is used for medical purposes, is must be screwed firmly to points C on the floor using two screws.
Demonstration plate thickness: 10 mm
REF 63 08 717 Sinius demo chair plate
63 22 668 D3561
44 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.5Information on planning for the practice Installation Requirements Sinius / Sinius CS / Sinius TS
4.5
Information on planning for the practice
The following file(s) for practice planning are available for download in the dealer area of the Sirona website under DOWNLOADS => CAAD files:
● PDF file with print symbol - for to-scale printing on paper or adhesive film
● CAAD file(s) - for professional implementation planning with 2D/3D CAAD systems
63 22 668 D3561 D3561.021.01.09.02 03.2017
45
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.6Technical data
bеЦдблЬ
4.6
Technical data
Model designation: Sinius / Sinius CS / Sinius TS Power connection: 100 – 240 V AC ± 10%
50/60 Hz
Rated current: 2.2 A at 240 V
4.35 A at 115 V
5.0 A at 100 V also max. 6 A for external devices
Type of ground connection: TN-C-S system or TN-S system (acc. to
IEC 60364-1) Overvoltage category: 2 acc. to IEC 60664-1 Average power consumption
(for dimensioning an air conditioning system):
0.25 kW
Power consumption in the Standby mode:
3 W (without internal mini PC)
Main building fuse: Type B automatic circuit breaker
100 - 115 VAC: 20 A medium-blow
220 – 240 V AC: 16 A medium-blow
Protection class: Class Idevice Device class in accordance
with Directive 93/42/EEC:
Class IIa equipment
Degree of protection against electrical shock:
Type B applied parts
External intraoral camera SiroCam AF.
These are:
Applied part type BF
Degree of protection against ingress of water:
Ordinary equipment (without protection
against ingress of water)
The foot switch has an IPX1 degree of
protection against liquids (drip-proof).
Tests/Approvals: See "Standards/Approvals“ [ → 49]. Date of manufacture:
(on the rating plate) USB port: corresponds to USB 2.0 standard
63 22 668 D3561
46 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.6Technical data Installation Requirements Sinius / Sinius CS / Sinius TS
Foot control wireless interface
The wireless foot control can be equipped with various wireless modules. A sign is attached to the underside of the foot control indicating the wireless module used.
Wireless module up to April 2017:
Wireless module from April 2017:
Model designation: nanoLOC AVR Frequency: 2.4 GHz – 2.4835 GHz (ISM band) Transmitting power: < 2 mW (short-range device) Modulation type: MDMA Range: approx. 10 m Approval: See "Standards/Approvals“ [ → 49].
Model designation: BlueMod+S Frequency: 2.4 GHz – 2,480 GHz Transmitting power: < 2 mW (short-range device) Modulation type: GFSK Range: approx. 10 m Approval: See "Standards/Approvals“ [ → 49].
IMPORTANT
Minimum requirements for PC
See document "Installation instructions and system requirements for PC configuration," (REF 61 94 075 SIVISION digital.
63 22 668 D3561 D3561.021.01.09.02 03.2017
47
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.6Technical data
bеЦдблЬ
4.6.1 Operating, transport and storage conditions
Operating mode: Continuous operation with intermittent
loading corresponding to the dental
mode of working.
Drive motors for chair operation:
intermittent use, max. 2 minutes on and
18 minutes off
Permanently connected unit. Operation
is not permitted in mobile vehicles.
Transport and storage conditions:
Temperature: -40 °C – +70 °C
(-40 °F – 158 °F)
Relative humidity: 10% - 95%
Air pressure: 500 hPa – 1060 hPa Operating conditions: Ambient temperature: 10 °C – 40 °C (50
°F – 104 °F)
Relative humidity: 30% - 85%
no condensation
Air pressure: 700 hPa – 1060 hPa Installation location: ≤ 3000 m above sea level
This treatment center is not suitable for
operation in areas subject to explosion
hazards. Pollution degree: 2 acc. to IEC 60664-1
-40
+70
1060
500
95
10
+10
+40
1060
700
85
30
Local pressure (min./ max.):For details see page Requirements of the supply media [ → 11]
Air: 5.5 / 7.5 bar (79.8 / 108.8 psi)
Water: 2.5 / 6 bar (36.3 / 87 psi)
Suction air: p
u
min. 0.12 bar (1.7 psi),
max. 0.18 bar (2.6 psi)
≥ 500 Nl/min
Water flow: For details see page Requirements of the supply media [ → 11]
Water absorption max. 3 l/min
Residual water max. 3 l/min
63 22 668 D3561
48 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.6Technical data Installation Requirements Sinius / Sinius CS / Sinius TS
4.6.2 Weight and packaging
Dimensions of the packaging Weight (with packaging and
accessories/without packaging): Sinius dentist element, assistant element: 126 cm x 72 cm x 75 cm 63 kg / 42 kg Sinius CS dentist element: 136 cm x 86 cm x 65 cm 55 kg / 40 kg Sinius TS dentist element: 154 cm x 57 cm x 80 cm 45 kg / 29.5 kg Sinius CS/TS assistant element: 88 cm x 35 cm x 74 cm 15 kg / 10 kg Water unit: 82 cm x 56 cm x 120 cm 80 kg / 66 kg Chair: 153 cm x 66 cm x 88 cm 116 kg / 91 kg Upholstery: 80 cm x 60 cm x 36 cm 8.5 kg / 5.5 kg LEDview Plus: 40 cm x 40 cm x 19 cm 3 kg / 2 kg LEDview Plus support arm: 93 x 40 x 20 cm 8 kg / 6.5 kg LEDview: 98cmx60cmx42cm 12kg/8kg Support tube for lamp without
monitor adapter:
128 cm x 20 cm x 20 cm 4.9 kg / 3.9 kg
Support tube with monitor adapter: 128 cm x 38 cm x 20 cm 6.4 kg / 5.6 kg There are 2 types of accessory boxes: 120 cm x 80 cm x 54/78 cm 62.6 kg
63 22 668 D3561 D3561.021.01.09.02 03.2017
49
Dentsply Sirona 4Dimensions, technical data Installation Requirements Sinius / Sinius CS / Sinius TS 4.7Standards/Approvals
bеЦдблЬ
4.7
Standards/Approvals
The treatment center Sinius® complies with the following standards, among others:
● IEC 60601-1 (electrical and mechanical safety plus software reliability)
● IEC 60601-1-2 (electromagnetic compatibility)
● IEC 60601-1-6 / IEC 62366 (usability)
● IEC 62304 (software process)
● ISO 6875 (patient chair)
● ISO 7494-1 (dental treatment devices)
● ISO 7494-2 (dental treatment devices, water and air supply)
● ISO 9680 (operating light)
● ISO 11143 (amalgam separator), see also below (provided amalgam separator option is available)
● EN 1717 (connection to the drinking water system), see also below and chapter "Connection to the public drinking water supply" [ → 13]
Original language of this document: German
China Registration number: SFDA (I) 20162555131
Serial number: YZB/GER 1522-2013
Customer service: Sirona Dental Systems Trading (Shanghai) Co., Ltd Hotline: 400-821-4888
This product bears the CE marking in accordance with the provisions of Council Directive 93/42/EEC of June 14, 1993 concerning medical devices.
The treatment center complies with the requirements of the RoHS Directive 2011/65/EU.
The treatment center meets the requirements in accordance with CAN/ CSA-C22.2 No. 60601-1 and AAMI/ANSI ES 60601-1.
The amalgam separator achieves a separation efficiency of >95%. The unit thus fulfills the requirements of ISO 11143. Separating procedure type 1: Centrifuge system The amalgam separator is approved by the German institute for building technology (DIBt) and AFNOR (France).
0123
SIRONA
Dental Systems
GmbH 64625 Bensheim Fabrikstrasse 31
Z-64.1-14
63 22 668 D3561
50 D3561.021.01.09.02 03.2017
4Dimensions, technical data Dentsply Sirona
4.7Standards/Approvals Installation Requirements Sinius / Sinius CS / Sinius TS
When equipped with a disinfection system, the treatment center complies with the technical rules and requirements on safety and hygiene for connection to the public drinking water supply. The unit is certified according to the requirements of the DVGW (German Gas and Water Association). It is intrinsically safe in accordance with worksheet W540. The unit thus fulfills the requirements of EN 1717, see also the chapter entitled "Connection to the public drinking water supply" [ → 13].
This unit meets the requirements of BELGAQUA and may therefore be connected to the public drinking water supply in Belgium.
This unit meets the requirements of ATS and may therefore be connected to the public drinking water supply in Australia.
The wireless modules in the wireless foot control and in the treatment center meet the requirements of the R&TTE directive 1999/5/'EC. Standards:
● EN 60950-1
● EN 301489-1, EN 301489-17, EN 300328
The modules meet the requirements of the Federal Communications Commission (Part 15 of the FCC Rules).
FCC ID: SIFNANOLOCAVR0108 (up to April 2017) or FCC ID: RFRMS (from April 2017)
Industry Canada The modules meet the requirements of Industry Canada (RSS210).
IC: 7654A-nanoLOCAVR (up to April 2017) or ID: 4957A-MS (from April 2017)
The current approvals of the wireless foot switch are listed on the rating plate on the underside of the wireless foot switch.
Sinius
®
is a registered trademark of Sirona Dental Systems GmbH.
WaterMark
TM
ATS 5200.104
Certificate No. 21208
63 22 668 D3561 D3561.021.01.09.02 03.2017
51
Dentsply Sirona 5Electromagnetic compatibility Installation Requirements Sinius / Sinius CS / Sinius TS 5.1Accessories
bеЦдблЬ
5
Electromagnetic compatibility
Observance of the following information is necessary to ensure safe operation regarding EMC aspects.
Sinius / Sinius CS / Sinius TS complies with the requirements for electromagnetic compatibility (EMC) according to IEC 60601-1-2:2007
Sinius / Sinius CS / Sinius TS is hereinafter referred to as "UNIT".
5.1
Accessories
Connecting the PC
The required interface cables can be ordered from Sirona.
Heliodent Plus cable set
The UNIT may only be operated with accessories and spare parts approved by Sirona. Unapproved accessories and spare parts may lead to an increased emission or to a reduced immunity to interference.
The UNIT should not be operated in the immediate vicinity of other devices. If this proves to be unavoidable, the UNIT should be monitored to ensure that it is operating properly.
Designation of the interface cables Supplier HDMI cable, 10 m (L406) Sirona USB 2 cable with repeater, 10 m (L343),
connectors: Type A, Type B
Sirona
Ethernet cable, 10 m (L339) Sirona Audio cable, 10 m Sirona
2nd protective ground wire, 2.5 mm2, 10 m
Sirona
Designation of the interface cables Supplier Radiation cable, 10 m (L7) Sirona
63 22 668 D3561
52 D3561.021.01.09.02 03.2017
5Electromagnetic compatibility Dentsply Sirona
5.2Electromagnetic emission Installation Requirements Sinius / Sinius CS / Sinius TS
Accessories for EMC measurement
The EMC measurements were performed with the following PCs:
5.2
Electromagnetic emission
The UNIT is intended for operation in the electromagnetic environment specified below.
The customer or user of the UNIT should make sure that it is used in such an environment.
PC as peripheral device for checking the interfaces with:
Fujitsu Siemens Pentium 4 Fujitsu Siemens ESPRIMO
Mini PC MPC D1007 PC equipment: Processor Intel Pentium 4; 3 GHz Intel Core i3; 2.13GHz RAM 512 MB DDR 266 2 GB Graphics card NVIDIA GeForce 7300 LE Intel HD Graphics Hard disk drive ST3116085AS Hitachi HTS 545016) Keyboard PS/2 HID USB Mouse Logitech PS/2 Fujitsu OPT USB Interfaces: 1 x COM; 1 x LPT; 4 x USB 1 x LAN; 4 x USB, 1 x HDMI Software: SIDEXIS XG Video, Version 2.3
SIUCOM plus (Sirona)
SIDEXIS XG Video, Version 2.3
SIUCOM plus (Sirona) Operating system Microsoft XP Professional Version
2002 SP 3
Microsoft Windows 7 Professional
Emission measurement Conformity Electromagnetic environment - guidelines RF emissions according to CISPR 11 Group 1 The UNIT uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions according to CISPR 11 Class B The UNIT is intended for use in all facilities,
including residential areas and in any facilities connected directly to a public power supply providing electricity to buildings used for residential purposes.
Harmonics according to IEC 61000-3-2
Class A
Voltage fluctuations / flicker according to IEC 61000-3-3
coincides
63 22 668 D3561 D3561.021.01.09.02 03.2017
53
Dentsply Sirona 5Electromagnetic compatibility Installation Requirements Sinius / Sinius CS / Sinius TS 5.3Interference immunity
bеЦдблЬ
5.3
Interference immunity
The UNIT is intended for operation in the electromagnetic environment specified below.
The customer or user of the UNIT should make sure that it is used in such an environment.
Interference immunity tests
IEC 60601-1-2 test level
Compliance level Electromagnetic environment -
guidelines
Electrostatic discharge (ESD) according to IEC 61000-4-2
± 8KV contact discharge
± 15 KV air discharge
± 8 KV contact discharge
± 15 KV air discharge
Floors should be made of wood or concrete or finished with ceramic tiling. If the floor is covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/ burst according to IEC 61000-4-4
± 1kV for input and output lines
± 2 kV for power supply lines
± 1 kV for input and output lines
± 2 kV for power supply lines
The quality of the line power supply should be that of a typical commercial or hospital environment.
Surge voltages according to IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode voltage
± 1 kV differential mode
± 2 kV common mode voltage
The quality of the line power supply should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and variations of the power supply according to IEC 61000-4-11
<5% U
T
for ½ period
(>95% dip of U
T
)
40% U
T
for 5 periods
(60% dip of U
T
)
70% U
T
for 25 periods
(30% dip of U
T
)
<5% U
T
for 5sec.
(>95% dip of U
T
<5% UT for ½ period (>95% dip of U
T
)
40% U
T
for 5 periods
(60% dip of U
T
)
70% U
T
for 25 periods
(30% dip of U
T
)
<5% U
T
for 5sec.
(>95% dip of U
T
The quality of the line power supply should be that of a typical commercial or hospital environment.
If the user of the UNIT requires it to continue functioning following interruptions of the power supply, it is recommended to have the UNIT powered by an uninterruptible power supply or a battery.
Magnetic field of power frequencies (50/60 Hz) according to IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic fields
should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Note: U
T
is the AC supply voltage prior to application of the test level.
Portable and mobile radio equipment must not be used within the recommended working clearance from the UNIT and its cables, which is calculated based on the equation suitable for the relevant transmission frequency.
Recommended working clearance:
63 22 668 D3561
54 D3561.021.01.09.02 03.2017
5Electromagnetic compatibility Dentsply Sirona
5.3Interference immunity Installation Requirements Sinius / Sinius CS / Sinius TS
1. The higher frequency range applies at 80 MHz and 800 MHz.
2. The field strengths of fixed transmitters, such as base stations of radiotelephones and mobile agricultural radio broadcast services, amateur radio stations, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. A site survey is recommended to assess the electromagnetic environment due to fixed RF transmitters. If the measured field strength in the location in which the UNIT is used exceeds the applicable RF compliance level above, the UNIT should be observed to verify normal operation. If unusual performance characteristics are observed, it may be necessary to take additional measures such as reorientation or repositioning of the UNIT.
3. Over the frequency range 150kHz to 80 MHz, field strengths should be less than 3 V/m.
4. Test specification for radiated RF interference IEC 61000-4-3:
Conducted RF disturbance IEC 61000-4-6
3 V
eff
150 kHz to 80 MHz
1
3 V
eff
d= [1,2] √P
Radiated RF interference IEC 61000-4-3
4
3 V/m 80 MHz - 800 MHz
1
3 V/m 800 MHz - 2.7 GHz
1
3 V
eff
3 V
eff
d= [1,2] √P
at 80 MHz - 800 MHz
d= [2,3] √P
at 800 MHz - 2.7 GHz
with P as the power rating of the transmitter in watts (W) according to
the transmitter manufacturer's specifications and d as recommended
safety distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
2
should
be less than the compliance level
3
in
each frequency range.
Interference is possible in the vicinity of equipment bearing the following
graphic symbol.
Interference immunity tests
IEC 60601-1-2 test level
Compliance level Electromagnetic environment -
guidelines
63 22 668 D3561 D3561.021.01.09.02 03.2017
55
Dentsply Sirona 5Electromagnetic compatibility Installation Requirements Sinius / Sinius CS / Sinius TS 5.3Interference immunity
bеЦдблЬ
Test frequency (MHz)
Band (MHz) Service Modulation Maximum
Power (W)
Distance (m) Immunity test
level (V/m)
385 380 - 390 TETRA 400 Pulse modulation
18 Hz
1,8 0,3 27
450 430 - 470 GMRS 460,
FRS 460
FM ± 5 kHz deviation 1 kHz sine
20,328
710 745 780
704 - 787 LTE Band 13, 17Pulse modulation
217 Hz
0,2 0,3 9
810 870 930
800 - 960 GSM 800/900,
TETRA 800, iDEN 820, CDMA 850, LTE Band 5
Pulse modulation 18 Hz
20,328
1720 1845 1970
1700 - 1990 GSM 1800;
CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS
Pulse modulation 217 Hz
20,328
2450 2400 - 2570 Bluetooth,
WLAN, 802,11 b/g/n, RFID 2450, LTE Band 7
Pulse modulation 217 Hz
20,328
5240 5500 5785
5100 - 5800 WLAN 802.11
a/n
Pulse modulation 217 Hz
0,2 0,3 9
63 22 668 D3561
56 D3561.021.01.09.02 03.2017
5Electromagnetic compatibility Dentsply Sirona
5.4Working clearances Installation Requirements Sinius / Sinius CS / Sinius TS
5.4
Working clearances
Recommended working clearances between portable and mobile RF communication devices and the UNIT
The UNIT is intended for operation in an electromagnetic environment, where radiated RF interference is checked. The customer or the user of the UNIT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the UNIT - depending on the maximum output power of the communication device, as shown below.
The recommended safety distance d in meters (m) can be determined for transmitters, whose maximum power rating is not specified in the above
table, using the equation that belongs to the corresponding column, wherein P is the maximum power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1
The higher frequency range applies at 80 MHz and 800 MHz.
Note 2
These guidelines may not apply in all cases. The propagation of electromagnetic waves is influenced by their absorption and reflection by buildings, objects and persons.
Power rating of the transmitter [W]
Working clearance according to transmission frequency [m] 150 kHz - 80 MHz 80 MHz - 800 MHz 800 MHz - 2.5 GHz
d= [1.2] P d= [1.2] P d= [2.3] √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23
63 22 668 D3561 D3561.021.01.09.02 03.2017
57
Dentsply Sirona 5Electromagnetic compatibility Installation Requirements Sinius / Sinius CS / Sinius TS 5.5Foot control wireless interface
bеЦдблЬ
5.5
Foot control wireless interface
Insofar as the treatment center is equipped with a foot control, one wireless module each must be installed in the foot control and in the base of the chair of the treatment center. These modules transmit the foot control signals.
Wireless module in the wireless foot control and in the treatment center
The wireless foot control can be equipped with various wireless modules. A sign is attached to the underside of the foot control indicating the wireless module used.
Wireless module up to April 2017:
Wireless module from April 2017:
CAUTION
Interference with the wireless transmission
This wireless transmission may cause interference with or be disturbed by other radio services.
Model designation: nanoLOC AVR Frequency: 2.4 GHz – 2.4835 GHz (ISM band) Transmitting power: < 2 mW (short-range device) Modulation type: MDMA Range: approx. 10 m Approval: See "Standards/Approvals“ [ → 49].
Model designation: BlueMod+S Frequency: 2.4 GHz – 2,480 GHz Transmitting power: < 2 mW (short-range device) Modulation type: GFSK Range: approx. 10 m Approval: See "Standards/Approvals“ [ → 49].
63 22 668 D3561
58 D3561.021.01.09.02 03.2017
6Checklist Dentsply Sirona
6.1Installation site Installation Requirements Sinius / Sinius CS / Sinius TS
6
Checklist
6.1
Installation site
We recommend performing an inspection of the circumstances on location at least 4 weeks prior to installation. The checklist should help you when doing this.
This can help ensure a smooth procedure on the day that the Sinius is actually installed.
Installation location:
● Installation location:
● Unit location:
● Building number:
● Room name/number:
63 22 668 D3561 D3561.021.01.09.02 03.2017
59
Dentsply Sirona 6Checklist Installation Requirements Sinius / Sinius CS / Sinius TS 6.2Construction requirements
bеЦдблЬ
6.2
Construction requirements
Connections: Media (see On-site installation [ → 9])
● Water supply Pipe 10x1 mm, corner valve outlet 3/8"
● Compressed air supply line Pipe 10x1 mm, corner valve outlet 3/8"
● Suction line DN 40 HT-PP ISO 8283-3, inner diameter approx. 36.5 mm
● Water drainage DN 40 HT-PP ISO 8283-3, inner diameter approx. 36.5 mm
● Installation pipe (power supplies) inner diameter approx. 40 mm)
● Installation pipe (IT), inner diameter min. 50 mm (or corresponding flat conduit)
Connections: Electrical (see On-site installation [ → 9])
● Power cable: 3 x 1.5 mm
²
(AWG16)
● Type B automatic circuit breaker 230 V AC, 16 A medium-blow or
● Type B automatic circuit breaker 100-115 V AC, 20 A medium-blow
● Suction machine control cable and call cable: 3 x 1.5 mm
²
(AWG16)
● Wireless systems in 2.4 GHz frequency range available? (e.g. room monitoring systems, video transmitters, etc.)
☐ Yes ☐ No
● If yes, this may damage the wireless foot control. Please consult our Product Service team.
Uneven floors: Mounting plates (see Mounting plates [ → 42])
● No plate required (unevenness max. 2 mm, shimming plates can be used)
● Adapter plate, REF 63 07 511
● Demo plate (demo operation at a trade show), REF 63 08 717
● Load capacity of the floor is indicated (pressure and tensile loads).
● The treatment center can be safely anchored in the load-bearing structure (concrete/wood; NOT screed).
63 22 668 D3561
60 D3561.021.01.09.02 03.2017
6Checklist Dentsply Sirona
6.3IT hardware Installation Requirements Sinius / Sinius CS / Sinius TS
6.3
IT hardware
6.4
Network
Mode of operation
● Stand-alone solution (mini PC required in the base of the chair)
● PC in treatment room
PC system requirements: ☑ IT hardware and software requirements are described in the document "Installation notes and
system requirements for PC configuration", (REF 61 94 075) Sivision digital
IMPORTANT!
Please be aware that any deviations may cause malfunctions in camera mode. For further details, see document "Installation instructions and system requirements for PC configuration", (REF 61 94 075) Sivision digital.
Network:
● The entire network should be equipped with 100 MBit Ethernet.
● - Cat 5
- Cat 6
☐ ☐ 10Mbps ☐ 100Mbps
● Network connection for Sinius available.
● Network connection for external PC available.
IMPORTANT!
The use of routers between Sinius and the treatment center PC must be avoided.
● Network configuration plan available.
● Network jacks have been certified.
● Network certificate present.
● Network installation company.
● Remarks/Tasks:
63 22 668 D3561 D3561.021.01.09.02 03.2017
61
Dentsply Sirona 6Checklist Installation Requirements Sinius / Sinius CS / Sinius TS 6.5Data processing
bеЦдблЬ
6.5
Data processing
IP addresses/firewall:
● TCP/IP address range: _____ . _____. _____ . _____ - _____ . _____ . _____ . _____
● Subnet mask: _____ . _____ . _____ . ____
● Are addresses already defined/present? ☐ Yes ☐ No
● Is there a DHCP server (dynamic TCP/IP address assignment)? ☐ Yes ☐ No
NOTICE! A static address should be assigned for the Sinius. It must not lie in the dynamic address range!
● Sinius: _____ . _____ . _____ . ____
● Internal PC: _____ . _____ . _____ . ____
● External PC: _____ . _____. _____ . _____ - _____ . _____ . _____ . _____
● Standard gateway: _____ . _____ . _____ . ____
● Antivirus software available? ☐ Yes
Name:
☐ No
● Is a firewall installed? Software or hardware firewall?
☐ Yes ☐ SW
☐ HW
☐ No
● Remarks/Tasks:
Practice administration programs:
● Are connections to the practice administration programs, etc. installed or planned?
☐ Yes ☐ No
● If so, which system (manufacturer + name)?
● Remarks/Tasks:
63 22 668 D3561
62 D3561.021.01.09.02 03.2017
6Checklist Dentsply Sirona
6.5Data processing Installation Requirements Sinius / Sinius CS / Sinius TS
We reserve the right to make any alterations which may be required due to technical improvements.
© Sirona Dental Systems GmbH Sprache:englisch Printed in Germany D3561.021.01.09.02 03.2017 Ä.-Nr.: 123 221
Sirona Dental Systems GmbH Contact in the USA:
Fabrikstraße 31 64625 Bensheim Germany
Sirona Dental, Inc. 4835 Sirona Drive Charlotte, NC 28273 USA
Order No
63 22 668 D3561