Dear customer,Thank you for purchasing your ProFeel
ment center from Sirona.
The technical documentation supplied is also part of the
product. You should always keep this documentation
within reach.
This Operator Manual describe your treatment center with
maximum possible equipment.
For care of your treatment center, please observe the
instructions "Care and Maintenance by the Practice Team" (Chapter 12).
To safeguard your warranty claims, please complete the
attached “Installation Report/Warranty Passport”
together with the service engineer immediately after the
installation of your treatment center.
Separate Operating Instructions and the corresponding
instructions for care have been attached to all dental
instruments, the SIROLUX FANTASTIC, the
SIVISION 3, the SIROCAM 3, the ProSmile Handy,
the working stool and the tray.
Prior to start-up, you must read all Operating Instructions
to familiarize yourself with the treatment center.
To prevent any personal injury or material damage,
pay special attention to any notes printed in bold type or
marked in one of the ways indicated below:
+
dental treat-
i
NOTE
CAUTION
WARNING
Additional information, explanation or supplement.
The information provided under this keyword directly
concerns the functioning of the product and/or warns
against possible operating errors.
This warning symbol warns against possible hazards for
the product, the patient or the user.
In case you get stuck despite having thoroughly studied
the Operating Instructions, please contact your dental
depot.
+
Your ProFeel
Team
60 29 560 D 3474
2D 3474.201.03.06.02
Sirona Dental Systems GmbH Contents
Operator Manual ProFeel
+
Contents
1Explanation of the symbols used .............................................................................. 5
2Warning and safety information ................................................................................ 8
2.1General warning and safety information ........................................................................................... 8
with dentist element swivel arm......................................................... 16
bеЦдблЬ
7Dentist element and instrument functions ............................................................... 28
7.1Locking the dentist element.............................................................................................................. 28
7.2Swivel arm ........................................................................................................................................ 29
7.3Starting an handpiece....................................................................................................................... 30
7.4Adjusting the driving air pressure ..................................................................................................... 31
8.4Silicone mat ...................................................................................................................................... 36
12.3Cleaning the filter, replacing the main fuse...................................................................................... 65
12.4Dentist element ................................................................................................................................ 66
12.7Disinfection system – sanitation....................................................................................................... 73
4D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH1 Explanation of the symbols used
Operator Manual ProFeel
+
1 Explanation of the symbols used
IconExplanationIconExplanation
ONPatient chair up
OFFPatient chair down
ATTENTIONPatient chair upright
FusePatient chair inclined
T 10A H
Clean with waterFactory settings
Warning of dangerActivate chair program S,
rinsing position
(Last memory function)
Emergency stop functionActivate chair program 0,
entry/exit position
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
WrongActivate chair program 1,
upper jaw treatment position
No/not OKActivate chair program 2,
lower jaw treatment position
RightIndividual programming,
rinsing position S
Maximum loadIndividual programming,
entry/exit position 0
WrongIndividual programming,
upper jaw treatment 1
5
1 Explanation of the symbols used Sirona Dental Systems GmbH
+
1
2
0
40
60
80
ENDO
1
2
0
40
6
0
80
ENDO
1
20
40
6
0
80
ENDO
Operator Manual ProFeel
IconExplanationIconExplanation
Short acoustic signalIndividual programming,
lower jaw treatment 2
+
>10s
<10m
Long acoustic signalCan be sterilized in the autoclave
135°C 275°F
at 135 °C/275 °F
Time in secondsCan be thermodisinfected in the
thermodisinfector at 93 °C/197 °F
Time in minutes3-way syringe – water
Stop chair movement3-way syringe – air
Brake applied3-way syringe – spray
Brake releasedCan be set only by the service
engineer
Speed increaseTumbler filler
Speed reductionCuspidor flushing
Observe operating instructions
Press and hold key
of the manufacturer
Spray coolingRelease key
Changing the intensity
Press key briefly
of spray air/spray water
SIROLUX FANTASTIC
Maximum brightness
6D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH1 Explanation of the symbols used
Operator Manual ProFeel
IconExplanationIconExplanation
+/–
+
Increase/reduction of a variableSIROLUX FANTASTIC
Reduced brightness
ApproximatelyHeat dissipation
Tap waterAmalgam recycling
After consulting hoursCW rotation of electric micromo-
tor
OrCCW rotation of electric micro-
motor
bеЦдблЬ
Individual overwriting of factory
settings
60 29 560 D 3474
D 3474.201.03.06.02
7
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General warning and safety informationOperator Manual ProFeel
2 Warning and safety information
2.1General warning and safety information
Intended useThis dental treatment center is intended for prophylactic
treatment as well as diagnosis, therapy and dental treatment of humans by properly trained personnel.
This unit is not intended for operation in areas subject
to explosion hazards.
On-site installationThe on-site installation must have been performed
according to our requirements. The details are described
in the document “Installationsvoraussetzungen”.
Maintenance and repairAs manufacturers of dental medical equipment and in
the interest of the operational safety of your system,
we stress the importance of having maintenance and repair of your treatment center performed only by ourselves or by agencies expressly authorized by us. Furthermore components influencing the safety of the system should always be replaced with original spare parts upon failure.
When having such work done, we suggest that you
request
performed, includinginformationabout anymodifica-
tions of therated parameters orof theoperating ranges
(if applicable), as well as the date, name of organization
and signature.
a certificatestatingthe type and extent of work
+
Modifications to the systemModifications of this system which could impair the
safety of the system owner, patients or other persons
are prohibited by law!
For reasons of product safety, only original Sirona
accessories approved for this product, or accessories
from third parties approved by Sirona, may be used.
The user is responsible for dangers resulting from the
use of non-approved accessories.
If any devices not approved by Sirona are connected,
they must comply with the applicable standards:
IEC 60950 for information technology equipment
(e.g. PCs), and IEC 60601-1 for medical electrical
equipment.
Combination with other equipmentAny person who assembles or modifies a medical elec-
trical system complying with the standard IEC 60601-1-1
(safety requirements for medical electrical equipment)
by combining it with other equipment (e.g., by connecting it with a PC) is responsible for ensuring that the
requirements of this regulation are met to their full extent
for the safety of the patients, operators and environment.
In case of doubt, contact the manufacturer of the system
components.
8D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH2 Warning and safety information
Operator Manual ProFeel
+
2.1 General warning and safety information
Hygiene and disinfectionHighly immunosuppressed patients should not come in
contact with water from the treatment center. The use of
sterile solutions is recommended.
The national requirements and recommendations for
hygiene and disinfection must be observed (e.g. RKI,
ADA, CDC etc).
Electromagnetic compatibility (EMC)The ProFeel
IEC 60601-1-2:2001.
Medical electrical devices are subject to special precautionary measures regarding EMC. It must be installed
and operated in compliance with the specifications provided in the document “Installationsvoraussetzungen”.
Portable and mobile RF communication systems may
interfere with medical electrical equipment. Therefore,
the use of mobile wireless phones in practice or hospital
environments must be prohibited.
+
complies with the requirements of
With video systems:
Electromagnetic disturbances in the environment of the
ProFeel+ may result in reduced image quality. In such
cases, it is advisable to repeat the exposure before saving the image.
In case of line voltage drops, image data stored in the
video camera may be lost. Deposit the camera and
repeat the exposure.
Image artifacts may possibly occur as a result of
radio-frequency interference caused by the transmitters
of e.g. radio services or radio amateurs.
Influence on cardiac pacemakersThe workstation includes strong magnets which can
affect cardiac pacemakers of patients, users and technical personnel.
Therefore, please pay attention to the safety information
on page 18.
bеЦдблЬ
Treatment of highly immunosuppressed patientsHighly immunosuppressed patients should not come in
contact with water from the treatment center. The use of
sterile solutions is recommended.
Quality of water/air supplyAir and water supply must meet the requirements speci-
fied in the installation instructions. Use only clean water.
Water quality complianceTo ensure compliance with the medical and national
legal requirements for water from treatment centers,
Sirona recommends equipping the treatment center with
a disinfection system. As owner of the treatment center,
you are responsible for the water quality and may have
to take alternative measures to ensure its compliance if
you operate the treatment center without a disinfection
system.
Please contact your specialized dealer or your relevant
dental association for the respective national requirements and measures.
60 29 560 D 3474
D 3474.201.03.06.02
9
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General warning and safety informationOperator Manual ProFeel
Identification of warning and safety informationTo avoid personal injury and material damage, you must
also observe the warning and safety information provided in the present operating instructions. They are
highlighted by the captionNOTE, CAUTION or
WARNING.
i
Symbols usedObserve accompanying documents
(on rating plate of chair)
Ventilation slotsUnder no circumstances may the ventilation slots on the
unit be covered, since otherwise the air circulation will be
obstructed.
Do not spray disinfectants or other similar products into
the ventilation slots.
Vacuum systemThe suction removal of aluminum and other metal oxides
from blasting devices via the automatic separator and
the amalgam separator integrated in the treatment center is prohibited!
This would cause extreme wear and clogging of the suction and water paths.
A separate vacuum system must be used in connection
with metal oxide blasting devices.
+
Treatment centers equipped with a central wet suction
system are generally suitable for suction removal of the
above material. However, make sure to observe the
instructions provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in
connection with Sirona treatment centers. However, in
such cases, make sure that the system is subsequently
flushed with an adequate amount of water.
CuspidorDo not use any foaming domestic cleaning agents!
Patient chairPlease note that, according to EN ISO 6875, the dental
chair has a maximum load capacity of 135 kg
(tested with four-fold safety factor according to
IEC 60601-1).
Electric micromotors in dentist elementIntermittent operation: 1 min. ON – 7 min. OFF with a
motor current of 1.5 A.
Drive motors for chair and backrest6 % duty time, cycle time 250 s.
Maintenance of the treatment centerDespite the outstanding quality of your treatment center
and regular care by the practice team, it is required in the
interest of operational safety that preventive maintenance be performed at predetermined intervals.
In order to guarantee the operational safety and reliability
of your treatment center and to avoid damage due to natural wear, you as the system owner must have your system
checked through regularly by an authorized servic e en gineer from your dental depot. Furthermore, safety checks
must be performed.
60 29 560 D 3474
10D 3474.201.03.06.02
Sirona Dental Systems GmbH2 Warning and safety information
Operator Manual ProFeel
+
Please contact your dental depot to obtain a maintenance offer.
For details please refer to chapter “Maintenance”.
2.1 General warning and safety information
Dismantling and reassemblyWhen dismantling and reassembling the system, pro-
ceed according to the installation instructions for new
installation in order to guarantee its functioning and stability.
DisposalIt applies generally that the national regulations have to
be complied with when disposing of this product. Please
observe the regulations applying in your country.
Within the European Economic Community the directive
2002/96/EEC (WEEE) for electrical and electronic
devices requires environmentally compatible recycling /
disposal.
Your product is marked with the adjacent symbol. With
the goal of environmentally compatible recycling / disposal, your product must not be disposed of with the
domestic refuse.
The black bar under the "refuse bin" symbol means that
it has been put onto the market after 13.08.2005. (See
DIN EN 50419:2005)
Please note that this product is subject to the directive
2002/96/EEC (WEEE) and laws applicable in your country and must be sent for environmentally compatible
recycling / disposal.
Please contact your dealer if your product should be
finally disposed of.
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
Maximum load capacity of the treatment chair
Please observe the maximum load capacity of the chair
of 135 kg according to EN ISO 6875 (tested with a
four-fold safety factor according to IEC 60601-1).
11
2 Warning and safety information Sirona Dental Systems GmbH
STOP
STOP
STOP
STOP
STOP
STOP
STOP
2.1 General warning and safety informationOperator Manual ProFeel
Safety stop for patient chair
•Headrest
•Backrest
•Toeboard
Hand/foot placement
right
wrong
+
CAUTION
Make sure that no obstacles (e.g. window wings, drawers, devices, …) extend into the movement range.
Safety stop for assistant element
Cuspidor EMERGENCY STOP
Caution!
Danger of injury: The patient's legs may get
caught or be pinched.
• Prevent collision of the patient or seat upholstery
with the support arm of the cuspidor during movement of the chair.
In case of a collision, a safety switch (switching strip
underneath the cuspidor) cuts off the chair movement.
As soon as the pressure is removed from the switching
strip (e.g. by swiveling away the cuspidor), the chair can
be moved again.
12D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH2 Warning and safety information
!
!
Operator Manual ProFeel
+
Position of the dentist element
CAUTION
When moving the chair (via program or manually), make
sure that the patient does not under any circumstances
come into contact with the support arm or the dentist element.
Patient mirror
Caution!
Fire hazard: There could be a risk of fire due to
the burning glass effect.
• Do not expose mirror to direct sunlight!
2.1 General warning and safety information
bеЦдблЬ
CAUTION
For the patient:
Patients who are on a restricted sodium diet or suffer
from severe respiratory illness should not undergo treatment with the powder jet instrument.
Patients wearing contact lenses should remove them before treatment or wear protective goggles.
For more safety information, please refer to the Operating Instructions supplied with the ProSmile Handy.
60 29 560 D 3474
D 3474.201.03.06.02
13
3 Technical description Sirona Dental Systems GmbH
Operator Manual ProFeel
3 Technical description
+
Model designationProFeel
Power supply100 V/115 V/127 V/230 V AC 50/60 Hz
Nominal current11.5 A at 100 V AC
10 A at 115 V AC
9 A at 127 V AC
5 A at 230 V AC
Main unit fuseT 10A H, 250 V AC, Order No. 10 77 460
Operating conditionsAmbient temperature: 10 °C – 40 °C (50 °F – 104 °F)
Degree of protection against electric shockType B applied parts,
except SIROCAM 3/SIROCAM C:
Type BF applied parts
Degree of protection against ingress of waterOrdinary equipment (not protected)
The foot switch has an IPX1 degree of protection against liquids
(drip-proof).
Operating modeContinuous operation with intermittent loading corresponding to
the dental mode of working.
+
Permanently connected unit.
Year of manufactureOn rating plate of chair
Tests/approvalsThis dental treatment center complies with the requirements of
IEC 60601-1: 1990 + A1: 1993 + A2: 1995 and of
IEC 60601-1-2: 2001.
DVGW: This system complies with the technical rules and requirements on safety and hygiene for connection to the drinking water
supply (if the appropriate DVGW marking is affixed to it).
This product bears the CE marking in accordance with the provisions of the Council Directive 93/42/EEC of June 14, 1993 concerning medical devices (MDD).
0123
14D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH4 Controls and functional elements
P1
P
2
1
2
S
P
0
1
0
0
1
2
0
40
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
Operator Manual ProFeel
+
4.1 System overview of ProFeel+ with OTP dentist element
4 Controls and functional elements
4.1System overview of ProFeel
5.
3.
4.
1.
6.
+
with OTP dentist element
Sensor
2.
bеЦдблЬ
or
Possibilities for assigning the instrument positions
Position 1 Position 2Position 3 Position 4Position 5On right
SPRAYVIT Electric motor
or
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
or
SIROSONIC L Mini L.E.D.
Camera
or
Mini L.E.D.
60 29 560 D 3474
D 3474.201.03.06.02
side of
dentist
element
7.
1. Patient chair
2. Dentist element
3. Water unit
4. Assistant element
5. SIROLUX
6. Instrument foot switch with program positions
7. Instrument foot switch
i
NOTE
For operation of the SIVISION 3 video system, please refer to Operating Instructions 59 67 893.
15
4 Controls and functional elements Sirona Dental Systems GmbH
P
1
P2
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
4.2 System overview of ProFeel
+
with dentist element swivel armOperator Manual ProFeel
+
4.2System overview of ProFeel
5.
3.
4.
1.
+
with dentist element swivel arm
2.
6.
Possibilities for assigning the instrument positions
Position 1 Position 2Position 3 Position 4Position 5
SPRAYVIT Electric motor
or
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
or
Camera
or
Mini L.E.D.
SIROSONIC L
7.
1. Patient chair
2. Dentist element
3. Water unit
4. Assistant element
5. SIROLUX with sensor
6. Instrument foot switch with program positions
7. Instrument foot switch
60 29 560 D 3474
16D 3474.201.03.06.02
Sirona Dental Systems GmbH5 Putting the system into operation
1
2
S
P
0
100
1
20
40
60
8
0
ENDO
Operator Manual ProFeel
+
5 Putting the system into operation
Initial startup
The disinfection system adds a disinfectant to the
water (1:100) to prevent the formation of microorganisms in the water system.
Prior to initial startup of your treatment center, sanita-tion must be performed.
If the service engineer skipped the sanitation procedure
after installation of your treatment center on the basis of
an agreement with you or sanitation has not been performed for more than 2 weeks, please perform sanitation
as described in Chapter 12.6 Disinfection system – sanitation.
Sanitation takes approx. 24 hours.
Main switch
Switch ON the main switch ( l ).
; The ready-for-operation LED lights up.
CAUTION
For safety reasons, always switch the treatment
center OFF (O) after consulting hours. This cuts off the
air and water supply as well as the line voltage.
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (1).
Tilt the headrest backward using the locking button (2).
BFlat headrest
The headrest must be adjusted manually.
To do this, take the load off the backrest and pull out or
push in the headrest using the handle (3).
CAUTION
A strong magnet is integrated in the underside of the
head cushion (4) which could affect a nearby cardiac
pacemaker.
+
Therefore, please avoid bringing the magnet in close
proximity of a cardiac pacemaker of a patient, user or
technical personnel.
In addition, direct contact of the head cushion with
magnetic cards can lead to a data loss.
18D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH6 Patient chair
A
Operator Manual ProFeel
+
6.2 MultiMotion
6.2MultiMotion
Adjustment to patient's height
The headrest can be adjusted to the patient's height by
pulling out or pushing in the headrest extension. This
adjustment is usually required at the beginning of treatment!
i
NOTE
Before positioning the patient on the headrest, make
sure that the head support has been optimally adjusted
to the patient's height. This simplifies all subsequent
work with the MultiMotion considerably, since readjustment to the patient's height can thus be omitted when
changing over from mandibular to maxillary treatment.
The MultiMotion enables you to adjust the patient's head
in a way that optimally supports viewing of areas of the
mouth which are difficult to access.
bеЦдблЬ
Adjustment of hyperextension
Mandibular position
The mandibular position can be set by pulling the operating handle (A). The anatomical movement of the
arched extension keeps the patient's head in the support.
¾ Pull the headrest out of the guide by pulling
handle A.
i
NOTE
You can reduce the adjustment noise by pressing
release A.
60 29 560 D 3474
D 3474.201.03.06.02
19
6 Patient chair Sirona Dental Systems GmbH
A
B
B
6.2 MultiMotionOperator Manual ProFeel
Maxillary position
¾ Take the load off the headrest by lifting it slightly.
¾ Press release A on the handle.
¾ Let the headrest slide down slowly.
¾ Let go of the release when the headrest reaches the
desired position.
+
Rotation and tilt
The MultiMotion enables rotation of the patient's head
about the longitudinal axis of his body as well as lateral
tilting of his head.
The MultiMotion is set to the desired position as follows:
1. Press and hold one or both of the two side control
elements B.
The rotation and tilt adjustment locks of the headrest
are now released.
2. Set the headrest to the desired position.
3. Let go of the control elements B.
; The supporting surface is fixed in place.
Make sure that the headrest is indeed fixed in place
after you let go of the control elements!
20D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH6 Patient chair
C
E
D
F
Operator Manual ProFeel
+
6.2 MultiMotion
Removing the MultiMotion
For certain treatments (e.g. of children) it may be expedient to remove the MultiMotion completely in order to
obtain better access to the patient. The patient's head
will then be supported on the contact surface of the
headrest extension F.
To remove the MultiMotion, proceed as follows:
1. Press release button C.
2. Pull the complete headrest out of the headrest ex-
tension.
3. Cover the opening of the headrest extension with
cover cap E.
4. If a children's head pad D is available, place it on
the headrest extension. The head pad is held magnetically.
i
NOTE
The children's head pad (D) is not included in the scope
of supply, however, is available from your dental dealer.
bеЦдблЬ
WARNING
The children's head pad (D) contains a strong magnet on its bottom side. This magnet could affect any
cardiac pacemaker located nearby.
Therefore, do not allow patients, users and technical
personnel with a cardiac pacemaker to be located
near this magnet.
Furthermore, direct contact of the head pad with
magnetic cards can lead to data loss.
i
NOTE
Deposit the removed MultiMotion in a safe place where it
cannot fall onto the floor.
Inserting the MultiMotion
To insert the MultiMotion, proceed as follows:
¾ Check the guide of the removed headrest for con-
tamination. Remove any contamination.
¾ Remove the cover cap E.
Reinsert the MultiMotion in the guide from above until it
audibly locks in place.
Pull on the headrest again to make sure that it is
locked securely in place.
60 29 560 D 3474
D 3474.201.03.06.02
21
6 Patient chair Sirona Dental Systems GmbH
1
2
S
P
0
10
0
1
2
0
4
0
6
0
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
S
P
100
1
20
40
60
80
ENDO
S
1
0
2
P
S
1
0
2
S
1
0
2
6.3 Patient positioningOperator Manual ProFeel
6.3Patient positioning
Positioning with the control panel
1.
2.
You can start the motorized chair movements manually
with the keys on the control panel of the dentist element.
; During operation, a short acoustic signal sounds.
Overview:
1. Backrest inclined backward
+
3.
4.
2. Backrest upright
3. Patient chair down
4. Patient chair up
Positioning with the 4-way foot control
1.
3.
2.
4.
You can start the motorized chair movements manually
with the 4-way foot control.
; During operation, a short acoustic signal sounds.
Overview:
1. Patient chair up
2. Backrest upright
3. Backrest inclined backward
4. Patient chair down
Positioning with the control panel on the
1.
3.
2.
4.
assistant element
You can start the motorized chair movements manually
with the keys on the control panel of the assistant element.
; During operation, a short acoustic signal sounds.
Overview:
1. Backrest inclined backward
2. Backrest upright
3. Patient chair down
4. Patient chair up
22D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH6 Patient chair
S
S
P
0
0
P
1
P
1
2
P
2
Operator Manual ProFeel
+
6.4 Chair programs – factory settings
6.4Chair programs – factory settings
Chair programs
1 + 2
S012
3
< 3 s
1.2.
There are 4 factory-set chair positions. These settings
can be adjusted individually (see section “Chair programs – reprogramming” on page 24).
Program S
Reserved for mouth rinsing position.
Activation:
1. Press key P.
2. Press key S within 3 seconds.
The chair moves to the mouth rinsing position.
In combination with program key S it is possible to activate the cuspidor flushing function and the fountain for
filling the tumbler.
See Chapters 10.6 and 10.7.
bеЦдблЬ
< 3 s
1.2.
< 3 s
1.2.
3. When you press key P and then key S again, the
treatment center returns to its previous position
(memory position).
Program 0
Reserved for entry/exit position.
Activation:
1. Press key P.
2. Press key 0 within 3 seconds.
The chair moves into the entry/exit position.
Program 1
Reserved for upper jaw treatment position.
Activation:
1. Press key P.
2. Press key 1 within 3 seconds.
The chair moves to the lower jaw treatment position.
Program 2
Reserved for lower jaw treatment position.
60 29 560 D 3474
D 3474.201.03.06.02
< 3 s
1.2.
Activation:
1. Press key P.
2. Press key 2 within 3 seconds.
The chair moves into the lower jaw treatment position.
Instead of activating the programs with the program
keys, you also have the option of activating them with the
4-way foot control plate of the foot switch.
Actuate the 4-way foot control plate in the desired direction.
to the front = program S
to the back = program 0
to the left = program 1
to the right = program 2
The chair moves into the selected position.
6.5Chair programs – reprogramming
i
NOTE
The factory program is overwritten on reprogramming.
+
1.
2.
4.
3.
Each chair program (S, 0, 1, 2) can be programmed individually.
Reprogramming of the four chair programs is always
done in the same way:
1. Move the chair into the desired position by pressing
the corresponding keys on the control panel.
2. Press program key
P briefly.
3. Press the desired key (S, 0, 1, 2) within 3 seconds
and hold it down.
4. After 6 short audible signals and 1 long one, release
the key.
The new position is saved.
24D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH6 Patient chair
Operator Manual ProFeel
+
6.6 Chair programs – reprogramming with the foot switch
6.6Chair programs – reprogramming with the foot switch
The chair programs can also be reprogrammed with the
foot switch.
1. Move the chair into the desired position by pressing
the corresponding keys.
2. Press the 4-way foot control plate in the desired di-
rection for at least 3 seconds.
1.
2.
2.
> 3 s
3.
to the front = program S
to the back = program 0
to the left = program 1
to the right = program 2
3. After 6 short acoustic signals and 1 long one, re-
lease the 4-way foot control plate.
The chair position is saved.
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
25
6 Patient chair Sirona Dental Systems GmbH
1
2
S
P
0
1
0
0
1
2
0
4
0
60
8
0
END
O
P
r
o
S
m
i
l
e
1
2
S
P
0
ENDO
P
S
1
0
2
P1
P2
P
1
P2
P1
P2
P1
P2
P1
P2
6.7 Emergency stop function for chair movementOperator Manual ProFeel
6.7Emergency stop function for chair movement
.
a
n
o
P
r
i
s
r
o
S
m
i
l
e
+
There are several ways to stop the chair movement
instantly:
1. Control panel
Each stroke of keys P, 0, 1 and 2 stops the chair
movement.
2. 4-way foot control
Each actuation of the 4-way foot control stops the
chair movement.
3. Chair foot switch
If the foot switch is actuated, all chair movements stop.
26D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH6 Patient chair
Operator Manual ProFeel
+
6.8 Patient chair, options
6.8Patient chair, options
Media block on the patient chair
Additional equipment which is approved for dental use
(e.g. by EMS) may be connected to a media block on the
base of the chair (observe manufacturers’ specifications
for the additional equipment).
Water
Air
Fuse
Line voltage
230V
Technical data and requirements•Connected equipment is treated with disinfected wa-
The media block is equipped with quick couplings for air
and water and an outlet with a separate fuse (6.3 A) for
non-heating equipment.
For treatment centers with a disinfection system, the
DVGW approval (EN1717/DIN1988) remains valid also
with an integrated media block. In this case, the additional equipment connected to the treatment center also
complies with the requirements of EN 1717/DIN 1988.
ter if a disinfection system is installed. This water
contains 0.1 ‰ to 0.2 ‰ hydrogen peroxide during
regular operation and 1.4 % of the same agent during sanitation.
bеЦдблЬ
CAUTION
Before connecting any additional equipment, check to
make sure that it is suitable for hydrogen peroxide.
If the connected equipment is not suitable for the higher
sanitation concentration, Sirona recommends disconnecting it from the treatment center during sanitation and
sanitizing it separately in accordance with the manufacturer’s specifications.
Sirona does not assume any liability for possible damage to the connected equipment!
PressureFlow rate
Water2.2 ± 0.2 barmax. 300 ml/min
Air4.4 ± 0.5 barmax. 70 Nl/min
60 29 560 D 3474
D 3474.201.03.06.02
27
7 Dentist element and instrument functions Sirona Dental Systems GmbH
1
2
S
P
0
1
0
0
1
2
0
40
60
80
ENDO
1
2
S
P
0
1
0
0
1
2
0
40
60
8
0
ENDO
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
max.
2 kg
4.4 lbs
1
2
S
P
0
100
1
2
0
4
0
60
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
max.
1 kg
2.2 lbs
1
2
S
P
0
100
1
2
0
4
0
60
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
s
i
r
o
n
a
.
7.1 Locking the dentist elementOperator Manual ProFeel
7 Dentist element and instrument
functions
7.1Locking the dentist element
OTP dentist element
1. Loosen the positioning of the dentist element by turning the fixing screw (B) counterclockwise.
2. Position the dentist element at the desired height.
3. Hold the dentist element with one hand in that posi-
tion.
4. Lock the position of the dentist element by turning
the fixing screw clockwise.
+
Option
B
i
NOTE
The maximum load of the dentist element is 2 kg (4.4 lbs)
Dentist element with swivel arm
CAUTION
A = Never place any objects on this dentist element!
1. Loosen the positioning of the dentist element by
turning the fixing screw (B) counterclockwise.
2. Position the dentist element at the desired height.
A
A
B
3. Hold the dentist element with one hand in that posi-
tion.
4. Lock the position of the dentist element by turning
the fixing screw clockwise.
Option: Tray
i
NOTE
The maximum load of the tray is 1 kg (2.2 lbs)
28D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH7 Dentist element and instrument functions
Click
Operator Manual ProFeel
+
7.2 Swivel arm
7.2Swivel arm
Operation
1.
The swivel arms of the instrument hoses snap into the
end position (1) (not with SPRAYVIT).
To unlock the swivel arm, pull it to the front (2) and then
to the back (3).
bеЦдблЬ
CAUTION
Do not press the swivel arms to the rear while they are
locked; otherwise the locking mechanism may be destroyed!
OK
The spring elements optionally built into the swivel arms
2.
3.
NO
provide optimum access to the patient.
The instrument picked up first is ready for use.
To start an instrument after picking it up, actuate the foot
switch.
60 29 560 D 3474
D 3474.201.03.06.02
Optimizing the position
2
3
The position of the instruments on the rack can be optimized by shifting the guide rollers (2) on the swivel
arms (3).
29
7 Dentist element and instrument functions Sirona Dental Systems GmbH
C8
+
1
2
S
P
0
1
0
0
1
2
0
40
6
0
80
E
N
D
O
1
2
S
P
0
1
0
0
1
2
0
40
6
0
80
E
N
D
O
P1
P
2
1
2
S
P
0
1
0
0
1
2
0
40
6
0
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
20
4
0
60
8
0
E
N
D
O
7.3 Starting an handpieceOperator Manual ProFeel
7.3Starting an handpiece
1. Remove the instrument from the instrument holder,
(OTP dentist element) or rack (dentist element with
swivel arm).
2. Actuate the foot switch.
The instrument starts running.
1.
A
or
2.
NO
OK
CAUTION
Do not operate the handpieces without spray.
Always place the instrument back in the correct holder.
i
NOTE
Also observe the operating instructions for all type of
handpieces and the video system.
CAUTION
Use only treatment instruments approved by Sirona!
+
i
NOTE
A = Unused instrument holders must be sealed with a
OK
NO
ball!
60 29 560 D 3474
30D 3474.201.03.06.02
Sirona Dental Systems GmbH7 Dentist element and instrument functions
1
2
S
P
0
1
0
0
1
20
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
ENDO
Operator Manual ProFeel
+
7.4 Adjusting the driving air pressure
7.4Adjusting the driving air pressure
CAUTION
The driving air pressure must be set by a service engineer only.
CAUTION
The driving air pressure is set according to the specification of the handpiece manufacturer.
CAUTION
The driving air is factory set for ProSmile Handy.
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
31
7 Dentist element and instrument functions Sirona Dental Systems GmbH
With the foot switch the spray can be switched ON/OFF
by pressing the left button (A) on the foot switch with the
handpiece removed.
Chip blower
The chip blower is activated by pressing the right
button (B) on the foot switch with the handpiece removed. As long as you keep the button pressed, an air
jet escapes from the nozzle of the burr instrument.
+
B
A
i
NOTE
For the foot switch pictured opposite, "Spray ON" is
shown by a green indicator in the top left corner.
A
B
60 29 560 D 3474
32D 3474.201.03.06.02
Sirona Dental Systems GmbH7 Dentist element and instrument functions
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
1
2
S
P
0
1
0
0
1
2
0
4
0
60
80
E
N
D
O
+
Operator Manual ProFeel
+
7.6 Handpieces – adjusting the spray intensity
7.6Handpieces – adjusting the spray intensity
!
1.
3.
4
P
1
P2
!
2.
P1
P
2
s
i
r
o
n
a
.
e
l
i
m
S
o
r
P
The spray intensity is preset at the factory. However,
it can be adjusted by means of the control valve at the
bottom front part of the dentist element.
1. Insert the bur.
bеЦдблЬ
2. Preselect spray on the handpiece foot switch.
3. Remove the handpiece.
4. Actuate the handpiece foot switch.
5. Adjust the spray water using the knob at the hand-
piece holder or rack, located at the bottom of the
dentist element front side.
+
5.
Spray WATER
_
+
Spray AIR
1. Open the flap on the base of the water unit.
2. Adjust the amount of air by means of the valve.
This setting is then valid for all handpieces.
2.
1.
60 29 560 D 3474
D 3474.201.03.06.02
33
7 Dentist element and instrument functions Sirona Dental Systems GmbH
1
2
S
P
0
1
0
0
1
2
0
40
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
00
1
20
40
60
80
ENDO
A
B
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
100
1
20
40
60
80
ENDO
7.7 SIROLUX FANTASTICOperator Manual ProFeel
7.7SIROLUX FANTASTIC
Switching ON/OFF
To switch ON/OFF, press the corresponding key on the
dentist element control panel.
Light intensity: 24000 lux
Proximity switch
The lamp can also be switched ON and OFF via a proximity switch.
Move your hand into the area under the black cover (A).
+
The brightness can be regulated by turning the
knob (B).
Brightness reduction
Adjust the brightness with the corresponding key on the
control panel.
Light intensity: 24000 lux
Light intensity: 8000 lux
Further information
For more information on operation, care and cleaning,
please refer to the Operating Instructions of the
SIROLUX FANTASTIC.
34D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH8 Dentist element options
1
2
S
P
0
1
0
0
1
20
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
100
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
Operator Manual ProFeel
+
8.1 Tumbler holder
8 Dentist element options
8.1Tumbler holder
Mounting positions
The tumbler holder B can be mounted in four different
positions on both dentist element versions.
B
bеЦдблЬ
8.2Silicone handles, right/left
A
A
135°C 275°F
3 min
A
The silicone handles A serve for hygienic protection of
the control panel and the gripping area on the dentist
element.
They can be removed for cleaning and sterilizing.
The silicone handles can be sterilized in the autoclave at
temperatures up to 135 °C/275 °F, 3 minutes.
A
60 29 560 D 3474
D 3474.201.03.06.02
35
8 Dentist element options Sirona Dental Systems GmbH
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
max.
1kg
2.2 lbs
max.
1kg
2.2
lbs
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
8.3 TrayOperator Manual ProFeel
8.3Tray
Installation
A tray can be mounted on the dentist element version
with swivel arm.
+
93°C197°F
B
B
1
8.4Silicone mat
Place the tray B in position and clamp it with the
lever (1).
CAUTION
Maximum load: 1 kg (2.2 lbs)
The tray can be removed for cleaning. To do so, release
the lever (1).
C
135°C 275°F
3 min
The silicone mat C serves for hygienic protection of the
storage surface.
It can be removed for cleaning and sterilizing.
The silicone mat can be sterilized in the autoclave at
temperatures up to 135 °C/275 °F, 3 minutes.
60 29 560 D 3474
36D 3474.201.03.06.02
Sirona Dental Systems GmbH8 Dentist element options
1
2
S
P
0
1
00
1
20
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
10
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
Operator Manual ProFeel
+
8.5 SPRAYVIT 3-way syringe
8.5SPRAYVIT 3-way syringe
Remove the SPRAYVIT 3-way syringe and press the
corresponding medium button(s).
•Left buttonTaste = air
•Right button = water
•Both buttons = spray
Nozzle versions:
135°C275°F
3 min
D
Click
C
A
B
•A, Standard – short
•B, Option – bent
The nozzles A, B and the handle sleeve C can be pulled
off for cleaning and sterilizing. The nozzles can be sterilized in the autoclave at temperatures up to 135 °C/
275°F, 3 minutes.
For easier cleaning turn the insert D while pulling it off.
CAUTION
After having refitted the nozzle and handle sleeve, check
wether they are firmly engaged.
bеЦдблЬ
check
check
Left button
Right button
Both buttons
60 29 560 D 3474
D 3474.201.03.06.02
37
8 Dentist element options Sirona Dental Systems GmbH
135°C275°F
1
2
S
P
0
1
0
0
1
2
0
40
60
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
60
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
8.6 3-way syringeOperator Manual ProFeel
8.63-way syringe
Remove the 3-way syringe and press the corresponding medium button(s).
•Left button = water
•Right button = air
•Both buttons = spray
The nozzle (A) can be pulled off for cleaning and ster-
ilizing after pressing the release ring (B). The nozzle
can be sterilized in the autoclave at temperatures up to
135 °C/275 °F, 3 minutes.
CAUTION
After having refitted the nozzle and handle sleeve,
check wether they are firmly engaged.
+
3 min
A
B
Check
Left button
Right button
Both buttons
38D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH8 Dentist element options
Operator Manual ProFeel
+
8.7 Satelec Mini L.E.D. curing light
8.7Satelec Mini L.E.D. curing light
Safety information
CAUTION
Always use the glare shield!
The light beams emitted by this instrument may be hazardous and may never be directed to the eyes of any person, even if this person wears goggles with protective
lenses. The light may be directed exclusively to the treatment area in the oral cavity.
right
!
DO NOT STARE INTO BEAM OR VIEW
L.E.D. RADIATION
DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 2M LASER PRODUCT
P < 1 mW, l = 420 nm - 480 nm
IEC 60825-1 Ed. 1.2
wrong
L.E.D. beam
exit aperture
!
CAUTION
Do not stare into the beam path with the glass rod removed (Class 2M laser product).
Viewing the LED light exit aperture with certain optical instruments (e.g. a magnifying lens) within a distance of
100 mm may constitute an eye hazard.
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
CAUTION
Heavy line voltage drops may lead to a shutdown of the
Mini L.E.D. The fault is signaled by a double beep. Check
the selected operating mode and restart the Mini L.E.D.
39
8 Dentist element options Sirona Dental Systems GmbH
8.7 Satelec Mini L.E.D. curing lightOperator Manual ProFeel
Any condensation occurring in the handpiece of the
Mini L.E.D. may cause impairments (e.g. fogging of the
L.E.D.). When moving the handpiece from a cool environment to a warm room, always wait for it to reach room
temperature before putting it into operation.
Curing lights must not be used on persons who are suffering from or have in the past been afflicted by photobiological reactions (including solar urticaria and erythropoetic porphyria). Nor should they be used on persons
currently being treated with any medicine which
increases one’s sensitivity to light (including methoxsalene and chlorotetracycline).
Any persons who have in the past suffered from retinal
or eye lens disorders or have undergone eye surgery,
especially for gray cataracts, must consult their medical
eye specialist before using the light or undergoing treatment with the Mini L.E.D. Caution is advisable even if the
patient gives his or her consent, since the light intensity
can cause accidents. It is especially advisable to always
wear appropriate protective goggles (UV filter).
+
Installation
1. Connect the supply cable to the treatment center.
2. Screw the handpiece on.
4
2
3
1
3. Attach the sterilized light guide to the handpiece.
Make sure to insert the light guide correctly.
4. The light guide engages with a “click”.
5. Slip the glare shield onto the light guide. It protects
your eyes against reflected light.
40D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH8 Dentist element options
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
00
1
2
0
40
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
Operator Manual ProFeel
+
8.7 Satelec Mini L.E.D. curing light
Installation options
•In the assistant element, holder 4
•Additional holder (not possible with swivel arm)
bеЦдблЬ
7
4
1
Handpiece and accessories
1. Handpiece
6
2. Light guide
3. Supply cable
3
4. Status LEDs
2
5. Glare shield
6. ON/OFF switch
7. Function button
5
Operating modes
The Mini L.E .D. has three different modes. To select the
desired menu, press the function button.
4.1 Status LED
The status LED signals the following states:
4.1
4.2
4.3 4.4
ColorMode
greenNormal operation
red, flashingOverheat protection
60 29 560 D 3474
D 3474.201.03.06.02
41
8 Dentist element options Sirona Dental Systems GmbH
8.7 Satelec Mini L.E.D. curing lightOperator Manual ProFeel
4.2 Quick curing mode
In this mode the Mini L.E.D. operates at full power
for 10 seconds.
LED 4.2 shows your selection.
0510
Sec.
In this mode, the light intensity output is approximately
as follows:
•1100 mW/cm² (± 10 %) with the standard light guide,
dia. 7.5 mm.
•2000 mW/cm² (± 10 %) with the optional booster
light guide, dia. 5.5 mm.
4.3 Pulsed curing mode
This mode is selected by pressing the function
button (7). The light operates at full power in “pulsed”
012.5
Sec.
mode; it emits radiation in 10 consecutive light pulses
of 1 s each. Between the individual pulses there is a
250 ms break. LED 4.3 shows your selection.
4.4 “Soft start” mode
+
05101520
Sec.
This mode is selected by pressing the function
button (7). The “soft start” mode provides:
•A 10-second “soft start” from 0 to 1100 mW/cm²,
or from 0 to 2000 mW/cm² with the (optional) “booster light guide”, dia. 5.5 mm.
•Full power during 10 seconds.
LED 4.4 shows your selection.
Handling
After you switch on the treatment center, the Mini L.E.D.
is by default in the quick curing mode (4.2).
If you change the operating mode of the Mini L.E.D.
while the treatment center is in operation, the new mode
remains valid even after the Mini L.E.D. is placed in its
holder.
After you have selected a mode, the Mini L.E.D. is ready
for operation.
Position the light guide as close as possible to the composite material surface you want to photopolymerize.
Make sure that the light guide never touches the material
to be cured, as this may cause damage to the light guide
and reduce its effectiveness.
•Never work without the glare shield (5).
•Never look into the light which is reflected by the
tooth surface.
42D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH8 Dentist element options
Operator Manual ProFeel
+
To start the polymerization cycle, briefly press the
ON/OFF switch (6). This is confirmed by an audible
signal.
6
In each mode, an audible signal sounds every 5 seconds.
The end of the cycle is also indicated by an audible signal. You may interrupt the cycle at any time though by
lightly pressing the ON/OFF switch.
8.7 Satelec Mini L.E.D. curing light
Technical data
Model:Mini L.E.D.
Weight of handpiece without
hose:
Dimensions:dia. 23 mm x 240 mm
Power supply of handpiece:5 V DC/2 A
Thermal safety:Overheat protection
105 g
bеЦдблЬ
Wavelength:420 nm – 480 nm
Light power dia. 7.5 mm
(standard version):
Light power dia. 5.5 mm
(optional, available from
Satelec):
Light power:450 mW – 500 mW
L.E.D. laser power:< 1 mW
1100 mW/cm
2000 mW/cm
2
2
60 29 560 D 3474
D 3474.201.03.06.02
43
8 Dentist element options Sirona Dental Systems GmbH
1
2
S
P
0
10
0
1
20
40
6
0
8
0
EN
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
P
0
1
S
1
2
S
P
0
10
0
1
2
0
4
0
6
0
8
0
E
N
DO
1
2
_
+
+
-
100
1
20
40
60
80
100
1
20
40
60
80
100
1
20
40
60
80
= 70= 60
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
60
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
100
1
20
40
60
80
P1
P
2
1 …... 60
2
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
100
1
20
40
60
80
8.8 SL/EL 1 electric motorOperator Manual ProFeel
8.8SL/EL 1 electric motor
CAUTION
Use only Sirona electric motors!
Setting the maximum speed
1. Remove the handpiece.
; The display shows the value last set as maximum
1.
A
2.
speed.
2. Modify the speed using control knob A.
; The set value is shown on the display.
Speed in %LED/display
120406080100
1 - 4•
5 - 14••
15 - 24•
25 - 34••
35 - 44•
45 - 54••
55 - 64•
65 - 74••
75 - 84•
85 - 94••
95 - 100•
+
3. Put the instrument back in place.
The set value is saved during this action.
Motor speed control with the Sirona unit foot
switch
1.
2.
1.
1. Remove the handpiece.
2. Actuate the foot switch.
60
The speed instantly
rises to the preset maximum value.
44D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH8 Dentist element options
S
P
100
1
20
40
60
80
ENDO
M
M
ENDO
1
2
S
P
0
1
0
0
1
2
0
40
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
2
1
2
S
P
0
1
0
0
1
20
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
M
M
Operator Manual ProFeel
+
8.8 SL/EL 1 electric motor
Reversing the direction of rotation of the
motor
1.
1.
1. Remove the handpiece.
; The default setting is clockwise rotation.
2. Press key B.
2.
B
; The LED lights up.
The new direction of rotation remains valid until the
bеЦдблЬ
handpiece is placed back in its holder, or until key B is
pressed again.
Table 1: Explanation of the symbols
Motor CCW
Motor CW
60 29 560 D 3474
D 3474.201.03.06.02
45
8 Dentist element options Sirona Dental Systems GmbH
1
2
S
P
0
1
0
0
1
2
0
4
0
60
8
0
E
N
D
O
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
80
E
N
D
O
_
+
_
_
+
+
1
2
S
P
0
100
1
2
0
40
60
8
0
ENDO
100
1
20
40
60
80
100
1
20
40
60
80
100
1
20
40
60
80
P
1
P
2
60
= 70
100
1
20
40
60
80
+
-
-
+
= 60
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
60
8.9 SIROSONIC L endodontic scaler - normal operationOperator Manual ProFeel
8.9SIROSONIC L endodontic scaler - normal operation
Setting the power intensity
+
right !
wrong
!
The full power of the ultrasonic oscillator is available in
normal operation.
1. Remove the handpiece.
; The display shows the last value set.
2. Adjust the intensity with control knob B.
; The set value is shown on the display.
3. Actuate the foot switch.
The intensity changes to the preset maximum value.
4. Put the handpiece back it its holder. The set value is
A
saved.
Power in %LED/display
120406080100
1 - 4•
AB
5 - 14••
15 - 24•
25 - 34••
35 - 44•
45 - 54••
55 - 64•
65 - 74••
75 - 84•
85 - 94••
95 - 100•
A = Water +/-
B = Preselected maximum intensity
Intensity +
Intensity -
46D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH8 Dentist element options
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
80
E
N
D
O
_
+
_
_
+
+
1
2
S
P
0
100
1
2
0
40
60
80
ENDO
100
1
20
40
60
80
100
1
20
40
60
80
100
1
20
40
60
80
P
1
P2
60
= 70
100
1
20
40
60
80
+
-
-
+
= 60
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
100
1
2
0
40
6
0
80
ENDO
ENDO
60
Operator Manual ProFeel
+
8.10 SIROSONIC L endodontic scale - Endo mode
8.10SIROSONIC L endodontic scale - Endo mode
The Endo mode is especially suitable for paradontal
treatment.
In the Endo mode, the maximum intensity of the ultrasonic handpiece is restricted to ensure safe operation.
bеЦдблЬ
PowerLED/display
1 20406080100
Step 1•
Step 2•
Step 3•
Step 4•
A
Step 5•
Step 6•
A
B
A = Water +/-
B = Preselected maximum intensity
Intensity +
Intensity -
CAUTION
The set value is saved when the handpiece is placed
back in its holder.
60 29 560 D 3474
D 3474.201.03.06.02
47
8 Dentist element options Sirona Dental Systems GmbH
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
1
2
S
P
0
100
1
20
40
6
0
80
ENDO
1
2
S
P
0
1
0
0
1
2
0
40
60
80
E
ND
O
ENDO
1
2
S
P
0
100
1
20
40
6
0
80
ENDO
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
1
2
S
P
0
100
1
20
40
6
0
80
ENDO
1
2
S
P
0
1
0
0
1
2
0
40
60
80
E
ND
O
ENDO
1
2
S
P
0
100
1
20
40
6
0
80
ENDO
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
1
2
S
P
0
1
0
0
1
2
0
40
60
80
E
N
D
O
ENDO
1
2
S
P
0
1
0
0
1
20
4
0
6
0
8
0
E
NDO
P
r
o
S
m
i
l
e
8.11 ProSmile HandyOperator Manual ProFeel
8.11ProSmile Handy
+
10 sec
10 sec
10 sec
r
i
o
s
n
a
.
The ProSmile Handy removes dental plaque, soft
deposits and stains on all visible tooth surfaces as well
as in fissures and interdental spaces.
For more details on the function and operation,
P
r
o
S
e
l
m
i
please refer to the Operating Instructions of the
ProSmile Handy.
The media supply is suitable for the ProSmile Handy.
Connection of the ProSmile Handy/SIROAIR
B
.
.
.
a
a
a
n
n
n
o
o
o
s
s
s
r
r
r
i
i
i
P
P
P
r
r
r
o
o
o
S
S
S
m
m
m
e
e
e
l
l
l
i
i
i
i
NOTE
The ProSmile Handy is plugged onto a Sirona turbine
hose (holder 2/3/4).
For operation in position 2
1. Remove the ProSmile Handy.
s
s
s
i
i
i
.
.
.
r
r
r
a
a
a
o
o
o
n
n
n
e
e
e
l
l
l
P
P
P
i
i
i
r
r
r
o
o
o
m
m
m
S
S
S
2. Press key B for ten seconds.
; The function is selected and saved.
For operation in positions 3 and 4
; No selection required
48D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH9 Assistant element
Operator Manual ProFeel
+
9.1 Suction handpiece/saliva ejector
9 Assistant element
9.1Suction handpiece/saliva ejector
Suction handpiece
Adjust the suction intensity with the slider.
You can angle the suction handpiece by turning it.
bеЦдблЬ
Saliva ejector
Adjust the suction intensity with the lever.
60 29 560 D 3474
D 3474.201.03.06.02
49
9 Assistant element Sirona Dental Systems GmbH
9.2 Assistant element optionsOperator Manual ProFeel
CAUTION
Avoid absolutely sucking in the oral mucosa!
wrong
9.2Assistant element options
SPRAYVIT E / 3-way syringe
+
9.3Satelec Mini L.E.D. curing light
Operation: see Chapter 8.7
50D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH10 Water unit
max.
3 kg
6.6 lbs
Operator Manual ProFeel
+
10.1 Safety
10 Water unit
10.1Safety
Weights exceeding 3 kg must not be placed on the glass
plate.
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
CAUTION
Do not support yourself on the glass plate!
Make sure that the patient does not support himself/herself on the glass plate e.g. when leaving the patient chair.
51
10 Water unit Sirona Dental Systems GmbH
1
2
S
P
0
100
1
2
0
4
0
6
0
8
0
ENDO
1
2
S
P
0
10
0
1
20
40
60
80
ENDO
+
100
1
20
40
60
80
E
N
DO
+
100
1
20
40
60
80
E
N
D
O
+
100
1
2
0
40
60
80
EN
D
O
+
10.2 Cuspidor flushing time - factory settings (dentist element control panel)Operator Manual ProFeel
10.2Cuspidor flushing time - factory settings (dentist element control
panel)
When you press the cuspidor flushing key briefly, the
cuspidor is flushed for 30 seconds (factory setting).
30 s
+
C8
When you press the key again while the cuspidor is being flushed, flushing stops.
10.3Cuspidor flushing time - reprogramming (dentist element control
panel)
+
10 s10 m
1. Press the cuspidor flushing key for approx. 2 sec-
onds until an intermittent acoustic signal sounds.
1.
2.
2. Release the key.
3. You can preselect a cuspidor flushing time
between 10 seconds and 10 minutes.
4. Press the key again briefly after the desired flushing
2 s
time has elapsed. A long acoustic signal sounds.
; The cuspidor flushing time is saved now.
3.
10 s10 m
4.
10.4Adjusting the water amount for the cuspidor
1. Open the flap on the base of the water unit.
2. Adjust the amount of water with the valve.
1.
52D 3474.201.03.06.02
2.
60 29 560 D 3474
Sirona Dental Systems GmbH10 Water unit
1
2
S
P
0
1
0
0
1
20
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
1
2
S
P
0
100
1
20
40
60
80
ENDO
P
S
1
0
2
+
100
1
2
0
40
60
80
ENDO
Operator Manual ProFeel
+
10.5 Operation of the tumbler filling fountain
10.5Operation of the tumbler filling fountain
Activation of the tumbler filling function at the
dentist element
The fountain keeps running as long as the key on the
control panel is pressed.
.
a
n
o
P
r
i
s
r
o
S
m
i
l
e
Activation of the tumbler filling function at the
assistant element
bеЦдблЬ
The fountain keeps running as long as the key on the
control panel is pressed.
Activation of tumbler filling function with the
button on the support arm; this is intended for
operation by the patient.
The tumbler filling function keeps running as long as the
key on the support arm is pressed.
CAUTION
The tumbler must not be filled with cuspidor flushing water!
This water is not disinfected by the disinfection system.
60 29 560 D 3474
D 3474.201.03.06.02
53
10 Water unit Sirona Dental Systems GmbH
P
S
1
0
2
10.6 Cuspidor flushing time - factory settings (assistant element control panel)Operator Manual ProFeel
Activation of tumbler filling function with the
button on the support arm by the patient
The tumbler filling function keeps running as long as the
key on the support arm is pressed.
i
NOTE
The fountain is removable and sterilizable.
+
10.6Cuspidor flushing time - factory settings (assistant element
control panel)
When you press the cuspidor flushing key briefly, the
cuspidor is flushed for 30 seconds (factory setting).
When you press the key again while the cuspidor is
being flushed, flushing stops.
54D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH10 Water unit
S
1
0
2
S
1
0
2
S
1
0
2
Operator Manual ProFeel
+
10.7 Cuspidor flushing time - reprogramming (assistant element control panel)
10.7Cuspidor flushing time - reprogramming (assistant element
control panel)
2 s
3.
10 s10 m
10 s10 m
2.
You can preselect a time between > 10 seconds and
< 10 minutes.
1. Press and hold down the cuspidor flushing function
bеЦдблЬ
key. After 2 seconds an intermittent acoustic signal
sounds.
1.
2. Release the key.
3. The intermittent acoustic signal continues to sound
until the end of the flushing time (10 minutes max.).
4.
When the desired flushing time has elapsed, press the
key again briefly. A long acoustic signal confirms that
programming is complete now.
5.
60 29 560 D 3474
D 3474.201.03.06.02
55
10 Water unit Sirona Dental Systems GmbH
S
P
++
1
2
S
P
0
1
00
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
1
2
S
P
0
100
1
20
40
60
80
END
O
S
P
+
S
S
P
+
S
S
P
S
10.8 Combination of rinsing position S with tumbler filling fountain and cuspidor flushing controlled by dentist elementOperator Manual ProFeel
10.8Combination of rinsing position S with tumbler filling fountain
and cuspidor flushing controlled by dentist element
With the factory setting, neither the tumbler filling nor the
cuspidor flushing function can be activated using the
Skey.
< 3 s
< 3 s
The combinations shown on the left can be programmed
by a service engineer.
Only by a service engineer
+
.
a
n
o
P
r
i
s
r
o
S
m
i
l
e
< 3 s
< 3 s
56D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH10 Water unit
S
P
1
P
2
S
++
P
1
P
2
S
+
P
1
P
2
S
+
S
P
1
P
2
S
Operator Manual ProFeel
+
10.9 Combination of rinsing position S with tumbler filling fountain and cuspidor flushing controlled by foot switch
10.9Combination of rinsing position S with tumbler filling fountain
and cuspidor flushing controlled by foot switch
max.1 kg
Only by a service engineer
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
57
10 Water unit Sirona Dental Systems GmbH
10 sec
1 L H 0
2
i
10.10 Dürr wet suctionOperator Manual ProFeel
10.10Dürr wet suction
Draw off 1 liter of cold, clear water daily after
consulting hours. Press the self-cleaning
button simultaneously.
When the amalgam container reaches a filling level of
95 %, the AMALG indicator lights up.
The filling level is also indicated by a short acoustic signal.
A) !Please replace the amalgam container!
A
When the amalgam container has reached a filling level
of 100 %, the AMALG indicator lights up and a long audible signal sounds.
B) !!Replace the container immediately!!
bеЦдблЬ
B
The AMALG indicator flashes and an intermittent acoustic signal sounds.
An error has occurred.
Please contact your service engineer.
Functional test of the safety signal of the
amalgam separator
When you press the AM TEST key, the AMALG indicator
starts flashing; an acoustic signal is emitted simultaneously.
The AMALGAM test must be performed at
AM
TEST
least 1x annually.
60 29 560 D 3474
D 3474.201.03.06.02
59
10 Water unit Sirona Dental Systems GmbH
DESINF
AM
TEST
SAN
A
MALG
AM
TEST
SAN
AMALG
DESINF
DESINF
A
M
T
E
S
T
S
A
N
AMALG
AM
TEST
SAN
AMALG
DESINF
DENTOSEPTPDENTOSEPT
P
DENTOSEPTPDENTOSEPT
P
10.12 Disinfection systemOperator Manual ProFeel
10.12Disinfection system
Function:
The disinfection system adds a disinfectant to the water
for all treatment instruments and the tumbler filling function to reduce the germ count in the water.
i
NOTE
Disinfectant is not added to the water for the cuspidor
flushing function.
The water unit has an indicator for the disinfectant filling
level.
A) !If the DESINF indicator flashes, refill 1 liter of
DENTOSEPT P!
A
+
Refilling DENTOSEPT P
1. Open the cover of the water unit.
2. Refill DENTOSEPT P.
The DESINF indicator stops flashing.
i
NOTE
Your treatment center comes with a filler neck which can
be screwed onto the DENTOSEPT P bottle.
60D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH11 Maintenance
Operator Manual ProFeel
+
11 Maintenance
Despite the outstanding quality of your treatment center,
it is necessary to perform inspections and preventive
maintenance as well as safety tests at predetermined
intervals in order to ensure its operational safety.
These events are documented in the Maintenance
Manual.
Care and cleaning is done by the practice team.
For details please refer to the following sections.
11.1Care and cleaning by the practice team
The practice team is responsible for regular cleaning
and care of the treatment center, with the aim of minimizing the risk of contamination for patients and users.
The work to be performed is specified in the chapter
“Care and cleaning by the practice team”.
The surfaces (apart from upholstery) can be sprayed
and wiped with surface disinfectants.
bеЦдблЬ
11.2Inspection and maintenance
11.3Safety checks
In order to guarantee the operational safety and reliability of your treatment center and to avoid damage due to
natural wear, annual inspection and maintenance must
be performed on your treatment center. This is done by
an authorized service engineer from your dental depot.
The steps to be performed as well as the parts which
must be replaced are specified in the document
“Maintenance Certificate”.
An overview of the inspection and maintenance work
performed is additionally recorded by the service engineer in the “Installation Report/Warranty Passport”.
This document is part of the “Maintenance Manual”.
Medical equipment is designed in such a way that the
first occurrence of a fault does not create a hazard to the
safety of patients, users or other persons. Hence it is
important to detect such faults before a second fault
occurs, which might then lead to safety hazards.
60 29 560 D 3474
D 3474.201.03.06.02
61
11 Maintenance Sirona Dental Systems GmbH
11.4 Maintenance ManualOperator Manual ProFeel
For that reason it is essential to perform safety tests
every 2 years which aim particularly at detecting electrical faults (e.g. isolation defects). This is done by an
authorized service engineer from your dental depot,
most practically together with the work to be performed
according to section 11.2.
The check includes a visual inspection as well as measurements of the protective ground wire connections
and the equivalent leakage currents.
The inspections and measurements to be performed are
specified in the “Maintenance Manual”. The measured
values must be documented there by the service engineer.
Safety tests must also be performed and documented
during initial startup, after extensions/upgrades (con-version) of your treatment center and after repair work
which might affect the electrical safety of the system.
i
NOTE
The treatment center must not be operated if it has failed
to pass the safety tests!
+
11.4Maintenance Manual
Keep this document near your treatment center.
Any inspection and maintenance work as well as all
safety tests are documented by the service engineer in
the Maintenance Manual.
We recommend that the user always keep the documentation in the chapter “Reporting of incidents to author-ities/manufacturers” up-to-date, regardless of any
legal requirements.
62D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
Operator Manual ProFeel
+
12.1 Care and cleaning agents
12 Care and cleaning by the
practice team
12.1Care and cleaning agents
CAUTION
Use only cleaning and care agents which are approved
by Sirona!
You will find a list of all agents approved at the time of
delivery in the documentation folder supplied with your
treatment center.
A continuously updated list of approved agents can be
downloaded from the Sirona Internet homepage under
REF 59 70 905: Ö www.sirona.com
If you do not have Internet access, you can use one of
the following two ways to order the list:
bеЦдблЬ
12.2Patient chair
P1
P
2
1.2.
HO
2
• Order from your local dental depot
• Order from Sirona
Tel: ++49 (0) 62 51/16-16 16
FAX: ++49 (0) 62 51/16-18 18
Upholstery/foot switch
CAUTION
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
Cleaning the upholstery
For weekly care of the imitation leather upholstery, use
only agents approved by us.
Disinfecting the upholstery
CAUTION
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
60 29 560 D 3474
D 3474.201.03.06.02
Do not use any other disinfectants!
Spray disinfection/wipe disinfection
Pedal (foot switch)
Clean the bottom of the base plate using a damp cloth.
63
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
G
12.2 Patient chairOperator Manual ProFeel
Disinfecting the surfaces
The surfaces can be spray and wipe disinfected with surface disinfectants.
Intensive cleaning and disinfection of the head
support
+
The contact surface of the headrest can be removed to
facilitate intensive cleaning and disinfecting if necessary.
The surfaces can be spray and wipe disinfected with surface disinfectants.
¾ Press one side of head pad G inward from the
outside (to unlock it) and remove it in a upward
direction.
i
NOTE
Fast and frequent removal of the head pad from the front
or rear can damage the snap-in noses.
¾ Clean and disinfect the head pad and headrest.
¾ Hook on the head pad again laterally.
CAUTION
Damaged pads, e.g. with scratched surfaces, must be
replaced.
For reorders:
Contact surface:REF 61 82 047
64D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
Operator Manual ProFeel
+
12.3 Cleaning the filter, replacing the main fuse
12.3Cleaning the filter, replacing the main fuse
CAUTION
Before replacing the fuse, switch OFF the unit.
A
The filters must be cleaned under running water.
A = T 10A H
bеЦдблЬ
60 29 560 D 3474
D 3474.201.03.06.02
65
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
1
2
S
P
0
00
1
2
0
40
60
80
EN
DO
12.4 Dentist elementOperator Manual ProFeel
12.4Dentist element
Instrument hoses/instrument rack
+
1.
3.
CAUTION
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
2.
4.
135°C275°F
5.
CAUTION
SPRAYVIT
3 min
i
NOTE
Relieve pressure!
For care and cleaning of the motors, the straight and con-
7.
tra-angle handpieces, the highspeed handpiece and the
SIROSONIC L scaler, please refer to the corresponding
operating instructions.
6.
5x
5x
Wipe disinfection
93°C 197°F
8.
7.
A
i
NOTE
A = Unused holders should be sealed with a ball!
6.
5x
8.
5x
66D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
HO
2
Operator Manual ProFeel
+
12.4 Dentist element
SPRAYVIT 3-way syringe
CAUTION
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
3 min
Chemical disinfection according to the manu-
facturer’s specifications is permissible only externally.
CAUTION
Never place or immerse parts in disinfectant solutions!
The nozzles can be steam-sterilized in systems with a
drying device and pre- or post-vacuum.
i
NOTE
Do not reprocess the nozzles in a thermodisinfector.
Only the handle sleeve and the nozzles can be sterilized.
Sterilize only in autoclaves at up to 135 °C, 3 minutes
(2.1 bar).
CAUTION
Do not exceed 140 °C during the drying phase!
Care afterwards:
Wipe out the handle sleeve joint with a Q-tip dipped in
T1 spray regularly, especially after each disinfection
and/or sterilization cycle, to prevent the dry O-rings from
sticking and being damaged.
T1 spray, Order No.: 59 01 665
bеЦдблЬ
135°C 275°F
3 min
60 29 560 D 3474
D 3474.201.03.06.02
3-way syringe
Spray disinfection/wipe disinfection
Replacing the O-rings in the nozzle
i
A
B
C
NOTE
The O-rings must be replaced every 3 months!
Remove the three O-rings to be replaced with a hook
probe.
A To slide on the backmost O-ring, insert the supplied
tool up to the stop.
B To slide on the middle O-ring, pull out the tool up to the
mark.
67
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
80
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
EN
DO
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P1
P
2
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
8
0
E
N
D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
12.4 Dentist elementOperator Manual ProFeel
C The front O-ring can be inserted manually without
using any tool.
Do not use any pointed objects.
Attachment to the handle sleeve: Observe instructions
for care (page 67)!
O-ring set, insertion tool and 10 O-rings
Order No. 41 76 751
Surfaces
CAUTION
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
Wipe disinfection
+
Option
Option
135°C 275°F
3 min
Satelec Mini L.E.D.
CAUTION
135°C 275°F
3 min
switched off before using disinfectants. Do not use any
abrasive cleaners.
Prior to each use, check the light guides to make sure
that no composite residues have adhered to them. If this
is the case, remove these residues immediately and
check to make sure that the surface of the insert has not
been damaged. If you find any damage, replace the
damaged light guide.
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
Always make sure that the Mini L.E.D. light is
68D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
i
135°C 275°F
3 min
Operator Manual ProFeel
+
12.5 Assistant element
12.5Assistant element
No amalgam!
3 min
According to manufacturer’s
specifications
Suction handpiece/saliva ejector
bеЦдблЬ
CAUTION
Amalgam recycling
Amalgam residues caught in the sieve must
not be disposed of in the cuspidor or the
sink.
Wipe disinfection
Draw off 2 tumblers of clean water
every 2 hours.
Daily after consulting hours, draw off 1 liter
(0.25 gallons) of cold, clear water to clean
the hoses.
CAUTION
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
60 29 560 D 3474
D 3474.201.03.06.02
69
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
Soap
0,75 l
(1/4 Gallon)
Desinf.
135°C 275°F
3 min
Soap
12.6 Water unitOperator Manual ProFeel
12.6Water unit
Cuspidor
CAUTION
Use only cleaning agents which are approved
by Sirona (see page 63 section 12.1)!
Spray disinfection/wipe disinfection
CAUTION
!Do NOT use household detergents or
acidic agents!
+
Daily before the lunch break and after
consulting hours
The vacuum system is subjected daily to septic secretions, saliva and blood. For hygienic reasons, the system
must therefore be disinfected with a detergent/disinfec-
A
tant approved for drains and sewage lines by Sirona.
i
NOTE
If the treatment center is equipped with an amalgam separator, a coarse trap (A) is located in the cuspidor.
70D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
C= When the acoustic signal sounds and the amalgam
indicator lights up, replace the rotor!
D= Replace the rotor in the amalgam separator at least
once a year, independently of whether the AMALG indi-
C
D
F
E
cator lights up or not.
This prevents unpleasant odors from developing.
Replacing the rotor
1. Open the flap.
2. Remove the rotor
(signal (E) during removal).
bеЦдблЬ
F
4.
2.
1.
3. Put the accumulated amalgam residues in the old
rotor for proper disposal.
4. Fit the transport cap on the full rotor until it locks in
place. This releases a bonding agent. Therefore, do
not remove the transport cap again!
5. Return the rotor in the special packaging for disposal
(send it to one of the addresses listed on the slip that
comes with the new amalgam rotor).
6. Insert the new rotor in the lower part.
3.
Do not install a used rotor!
7. Close the flap again
Amalgam recycling!
Replacement:
Amalgam rotor
Order No. 14 34 138
Siphon
5.
6.
Clean the siphon (F) in case of poor drainage of the cuspidor, otherwise once a month.
To do so, first open the flap and then unscrew the cap.
Have a collector ready to catch the contents. Screw the
cap back on and close the flap.
60 29 560 D 3474
D 3474.201.03.06.02
71
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
12.6 Water unitOperator Manual ProFeel
Disinfection system – microbiological water
test
Must be performed after a working pause of more than
one week. Use a “Total Count Tester” for this purpose.
The first three testers will be supplied free of charge if
you return the attached voucher. Please state your
address and the serial number of the chair (see “Installation Report/Warranty Passport”). Additional testers
can be ordered under Order No. 58 53 775.
+
35 °C
24 h
30 s
CAUTION
The “Total Count Tester” has a maximum durability of
1 year after its date of delivery.
The cardboard disk contains a dehydrated nutrient
medium. It is activated by the sample and serves as culture medium for a number of bacteria. The number of
germs provides information on the hygienic condition of
the water.
Performing the test
•Allow water to run out of the SPRAYVIT syringe for
about 1 minute.
•Then take a water sample from this syringe.
Fill in water up to the upper mark on the container.
To neutralize the disinfecting agent of the water
sample, add approx. 1.5 g of fixing salt (sodium thiosulfate).
Do not touch the inside of the container or the part of
the tester to be immersed.
•Immerse the tester in the filled container for 30 seconds.
The cardboard disk with the nutrient medium will
now absorb 1ml of the water sample.
•Remove the tester and shake off excessive water.
Empty the container completely.
•Place the tester into the empty container for incubation: Keep either for two days at room temperature or
for 24 hours at a temperature of 35 °C.
•Then count all germs found on the surface of the
tester.
CAUTION
If the number of germs significantly exceeds 100, then
sanitation is required (see next page).
60 29 560 D 3474
72D 3474.201.03.06.02
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
DENTOSEPT
P
D
EN
TOSEPT
P
DENTOSEPT
P
+=
=+
D
E
S
I
N
F
AM
TEST
SAN
A
M
A
L
G
D
E
S
I
N
F
AM
TEST
SAN
A
M
A
L
G
AM
TEST
SAN
AMALG
DESINF
AM
TEST
SAN
AMALG
DESINF
Operator Manual ProFeel
+
12.7 Disinfection system – sanitation
12.7Disinfection system – sanitation
Sanitation cycle
AM
TEST
AMALG
DESINF
SAN
AM
TEST
AMALG
No sanitation possible,
DESINF
SAN
Refill with Dentosept P!
o.k.!
P
TOSEPT
EN
D
Starting the sanitation
Perform sanitation very 4 weeks.
bеЦдблЬ
3.
Sanitation can be started only when the DESINF LED is
not lit. When the red DESINF LED lights up, refill
Dentosept P!
CAUTION
The sanitation procedure cannot be interrupted.
CAUTION
> 2 s
1.
3.
2.
4.
1. Press the SAN key for at least 2 seconds.
2. The SAN LED lights up and a short acoustic signal
sounds. The sanitation procedure starts.
The following steps are performed automatically:
3. The mixing tank is emptied via the tumbler outlet.
4. The SAN LED starts flashing as soon as the mixing
tank is empty.
60 29 560 D 3474
D 3474.201.03.06.02
4.
73
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
D
E
S
I
N
F
AM
TEST
SAN
A
M
A
L
G
D
E
S
I
N
F
AM
TEST
SAN
A
M
A
L
G
AM
TEST
SAN
AMALG
DESINF
AM
TEST
SAN
AMALG
DESINF
D
E
S
I
N
F
AM
TEST
SAN
A
M
A
L
G
AM
TEST
SAN
AMALG
DESINF
12.7 Disinfection system – sanitationOperator Manual ProFeel
Flushing of the tumbler outlet
DENTOSEPT P = 1.41 % H2O
2
+
approx. 10 s
100 % DENTOSEPT P
1.
100 % DENTOSEPT P
1. The mixing tank is filled with Dentosept P
(approx. 10 seconds).
The SAN LED is flashing.
2.
2. The tumbler filling fountain is flushed.
The SAN LED is flashing.
3. After the flushing procedure is complete, an audible
signal sounds (two short beeps).
The SAN LED continues flashing.
3.
OK !
74D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
P1
P
2
1
2
S
P
0
1
0
0
1
2
0
4
0
6
0
80
ENDO
P
1
P
2
Operator Manual ProFeel
+
12.7 Disinfection system – sanitation
Flushing of the instruments
1.
DENTOSEPT P = 1.41 % H2O
1. Remove the SPRAYVIT and flush it for at least
20 seconds.
2. Activate the spray function with the foot switch.
2
2.
15 s
20 s
3. Remove an instrument.
4. Flush the instrument for at least 15 seconds by actu-
ating the foot switch.
5. Repeat steps 3 and 4 for all instruments connected.
6. Switch off the treatment center.
bеЦдблЬ
3. - 5.
; The ready-for-operation LED goes out.
Switch OFF the unit for 24 hours
6.
60 29 560 D 3474
D 3474.201.03.06.02
75
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
!
1
2
S
P
0
1
0
0
1
2
0
4
0
60
80
EN
D
O
D
ESINF
AM
TEST
SAN
A
M
ALG
DESIN
F
AM
TEST
SAN
AMALG
AM
TEST
SAN
AMALG
DESINF
AM
TEST
SAN
AMALG
DESINF
12.7 Disinfection system – sanitationOperator Manual ProFeel
Terminating sanitation
2.
DENTOSEPT P = 1.41 % H2O
2
+
100 % DENTOSEPT P
1.
5.
1. Switch on the treatment center after 24 hours.
2. The ready-for-operation LED lights up.
3. The treatment center is in sanitation mode:
The SAN LED is flashing.
A permanent acoustic signal sounds.
4. Press the SAN key (approx. 3 seconds).
The following steps are performed automatically:
> 3 s
3.
4.
5. The mixing tank is emptied via the tumbler filling
fountain.
6. The SAN LED starts flashing as soon as the mixing
tank is empty.
5 s5 s
6.
6.
76D 3474.201.03.06.02
60 29 560 D 3474
Sirona Dental Systems GmbH12 Care and cleaning by the practice team
D
E
S
IN
F
AM
TEST
SAN
A
M
A
L
G
D
E
S
IN
F
AM
TEST
SAN
A
M
A
L
G
AM
TEST
SAN
AMALG
DESINF
AM
TEST
SAN
AMALG
DESINF
D
E
S
I
NF
AM
TEST
SAN
A
M
A
L
G
AM
TEST
SAN
AMALG
DESINF
Operator Manual ProFeel
+
12.7 Disinfection system – sanitation
bеЦдблЬ
7.
1 % DENTOSEPT P
7. The mixing tank is filled with water and Dentosept P.
8. The tumbler filling fountain is flushed with a mixture
8.
1 % DENTOSEPT P
9.
of 1 % Dentosept P and 99 % water.
60 29 560 D 3474
D 3474.201.03.06.02
9. After the flushing procedure is complete, a permanent acoustic signal sounds repeatedly (3 short
beeps, 10 second pause).
9.
10 s10 s
77
12 Care and cleaning by the practice team Sirona Dental Systems GmbH
P1
P
2
P1
P2
12.7 Disinfection system – sanitationOperator Manual ProFeel
Flushing of the instruments
The instruments are flushed with a mixture of
1 % Dentosept P and 99 % water.
1.
1. Flush the SPRAYVIT for approx. 20 seconds.
20 s
+
2.
15 s
2. Activate the spray function for the instruments by
means of the foot switch.
3. Remove an instrument.
4. Start the instrument flushing procedure by actuating
the foot switch (approx. 15 seconds).
5. Repeat steps 3 and 4 for all instruments connected.
; THE SAN LED goes out. The sanitation procedure is