General information Sirona Dental Systems GmbH

Operator Manual ProFeel
General information
+
Dear customer, Thank you for purchasing your ProFeel
ment center from Sirona.
The technical documentation supplied is also part of the product. You should always keep this documentation within reach.
This Operator Manual describe your treatment center with maximum possible equipment.
For care of your treatment center, please observe the instructions "Care and Maintenance by the Practice Team" (Chapter 12).
To safeguard your warranty claims, please complete the attached “Installation Report/Warranty Passport” together with the service engineer immediately after the installation of your treatment center.
Separate Operating Instructions and the corresponding instructions for care have been attached to all dental instruments, the SIROLUX FANTASTIC, the SIVISION 3, the SIROCAM 3, the ProSmile Handy, the working stool and the tray.
Prior to start-up, you must read all Operating Instructions to familiarize yourself with the treatment center.
To prevent any personal injury or material damage, pay special attention to any notes printed in bold type or marked in one of the ways indicated below:
+
dental treat-
i
NOTE
CAUTION
WARNING
Additional information, explanation or supplement.
The information provided under this keyword directly concerns the functioning of the product and/or warns against possible operating errors.
This warning symbol warns against possible hazards for the product, the patient or the user.
In case you get stuck despite having thoroughly studied the Operating Instructions, please contact your dental depot.
+
Your ProFeel
Team
60 29 560 D 3474
Sirona Dental Systems GmbH Contents Operator Manual ProFeel
+

Contents

1 Explanation of the symbols used .............................................................................. 5
2 Warning and safety information ................................................................................ 8
2.1 General warning and safety information ........................................................................................... 8
3 Technical description ................................................................................................. 14
4 Controls and functional elements ............................................................................. 15
4.1 System overview of ProFeel+ with OTP dentist element .................................................................. 15
4.2 System overview of ProFeel
5 Putting the system into operation ............................................................................. 17
6 Patient chair................................................................................................................. 18
6.1 Articulated headrest / Flat headrest.................................................................................................. 18
6.2 MultiMotion ....................................................................................................................................... 19
6.3 Patient positioning ............................................................................................................................ 22
6.4 Chair programs – factory settings..................................................................................................... 23
6.5 Chair programs – reprogramming..................................................................................................... 24
6.6 Chair programs – reprogramming with the foot switch ..................................................................... 25
6.7 Emergency stop function for chair movement .................................................................................. 26
6.8 Patient chair, options ........................................................................................................................ 27
+
with dentist element swivel arm......................................................... 16
bеЦдблЬ
7 Dentist element and instrument functions ............................................................... 28
7.1 Locking the dentist element.............................................................................................................. 28
7.2 Swivel arm ........................................................................................................................................ 29
7.3 Starting an handpiece....................................................................................................................... 30
7.4 Adjusting the driving air pressure ..................................................................................................... 31
7.5 Instrument spray ON/OFF – chip blower .......................................................................................... 32
7.6 Handpieces – adjusting the spray intensity ...................................................................................... 33
7.7 SIROLUX FANTASTIC..................................................................................................................... 34
8 Dentist element options.............................................................................................. 35
8.1 Tumbler holder.................................................................................................................................. 35
8.2 Silicone handles, right/left................................................................................................................. 35
8.3 Tray................................................................................................................................................... 36
8.4 Silicone mat ...................................................................................................................................... 36
8.5 SPRAYVIT 3-way syringe................................................................................................................. 37
8.6 3-way syringe.................................................................................................................................... 38
8.7 Satelec Mini L.E.D. curing light......................................................................................................... 39
8.8 SL/EL 1 electric motor ...................................................................................................................... 44
8.9 SIROSONIC L endodontic scaler - normal operation ....................................................................... 46
8.10 SIROSONIC L endodontic scale - Endo mode ................................................................................. 47
8.11 ProSmile Handy................................................................................................................................ 48
60 29 560 D 3474 D 3474.201.03.06.02
3
Contents Sirona Dental Systems GmbH
Operator Manual ProFeel
9 Assistant element ....................................................................................................... 49
9.1 Suction handpiece/saliva ejector ..................................................................................................... 49
9.2 Assistant element options ................................................................................................................ 50
9.3 Satelec Mini L.E.D. curing light ........................................................................................................ 50
10 Water unit..................................................................................................................... 51
10.1 Safety............................................................................................................................................... 51
10.2 Cuspidor flushing time - factory settings (dentist element control panel)......................................... 52
10.3 Cuspidor flushing time - reprogramming (dentist element control panel) ........................................ 52
10.4 Adjusting the water amount for the cuspidor.................................................................................... 52
10.5 Operation of the tumbler filling fountain ........................................................................................... 53
10.6 Cuspidor flushing time - factory settings (assistant element control panel) ..................................... 54
10.7 Cuspidor flushing time - reprogramming (assistant element control panel) ..................................... 55
10.8 Combination of rinsing position S with tumbler filling fountain
and cuspidor flushing controlled by dentist element ........................................................................ 56
10.9 Combination of rinsing position S with tumbler filling fountain
and cuspidor flushing controlled by foot switch................................................................................ 57
10.10 Dürr wet suction............................................................................................................................... 58
10.11 Amalgam separator.......................................................................................................................... 59
10.12 Disinfection system.......................................................................................................................... 60
+
11 Maintenance ................................................................................................................ 61
11.1 Care and cleaning by the practice team .......................................................................................... 61
11.2 Inspection and maintenance ............................................................................................................ 61
11.3 Safety checks................................................................................................................................... 61
11.4 Maintenance Manual........................................................................................................................ 62
12 Care and cleaning by the practice team ................................................................... 63
12.1 Care and cleaning agents ................................................................................................................ 63
12.2 Patient chair ..................................................................................................................................... 63
12.3 Cleaning the filter, replacing the main fuse...................................................................................... 65
12.4 Dentist element ................................................................................................................................ 66
12.5 Assistant element............................................................................................................................. 69
12.6 Water unit......................................................................................................................................... 70
12.7 Disinfection system – sanitation....................................................................................................... 73
60 29 560 D 3474
Sirona Dental Systems GmbH 1 Explanation of the symbols used Operator Manual ProFeel
+

1 Explanation of the symbols used

Icon Explanation Icon Explanation
ON Patient chair up
OFF Patient chair down
ATTENTION Patient chair upright
Fuse Patient chair inclined
T 10A H
Clean with water Factory settings
Warning of danger Activate chair program S,
rinsing position (Last memory function)
Emergency stop function Activate chair program 0,
entry/exit position
bеЦдблЬ
60 29 560 D 3474 D 3474.201.03.06.02
Wrong Activate chair program 1,
upper jaw treatment position
No/not OK Activate chair program 2,
lower jaw treatment position
Right Individual programming,
rinsing position S
Maximum load Individual programming,
entry/exit position 0
Wrong Individual programming,
upper jaw treatment 1
5
1 Explanation of the symbols used Sirona Dental Systems GmbH
+
1
2
0
40
60
80
ENDO
1
2
0
40
6
0
80
ENDO
1
20
40
6
0
80
ENDO
Operator Manual ProFeel
Icon Explanation Icon Explanation
Short acoustic signal Individual programming,
lower jaw treatment 2
+
>10s
<10m
Long acoustic signal Can be sterilized in the autoclave
135°C 275°F
at 135 °C/275 °F
Time in seconds Can be thermodisinfected in the
thermodisinfector at 93 °C/197 °F
Time in minutes 3-way syringe – water
Stop chair movement 3-way syringe – air
Brake applied 3-way syringe – spray
Brake released Can be set only by the service
engineer
Speed increase Tumbler filler
Speed reduction Cuspidor flushing
Observe operating instructions
Press and hold key
of the manufacturer
Spray cooling Release key
Changing the intensity
Press key briefly
of spray air/spray water
SIROLUX FANTASTIC
Maximum brightness
60 29 560 D 3474
Sirona Dental Systems GmbH 1 Explanation of the symbols used Operator Manual ProFeel
Icon Explanation Icon Explanation
+/–
+
Increase/reduction of a variable SIROLUX FANTASTIC
Reduced brightness
Approximately Heat dissipation
Tap water Amalgam recycling
After consulting hours CW rotation of electric micromo-
tor
Or CCW rotation of electric micro-
motor
bеЦдблЬ
Individual overwriting of factory settings
60 29 560 D 3474 D 3474.201.03.06.02
7

2 Warning and safety information Sirona Dental Systems GmbH

2.1 General warning and safety information Operator Manual ProFeel

2 Warning and safety information
2.1 General warning and safety information
Intended use This dental treatment center is intended for prophylactic
treatment as well as diagnosis, therapy and dental treat­ment of humans by properly trained personnel.
This unit is not intended for operation in areas subject
to explosion hazards.
On-site installation The on-site installation must have been performed
according to our requirements. The details are described in the document “Installationsvoraussetzungen”.
Maintenance and repair As manufacturers of dental medical equipment and in
the interest of the operational safety of your system, we stress the importance of having maintenance and repair of your treatment center performed only by our­selves or by agencies expressly authorized by us. Fur­thermore components influencing the safety of the sys­tem should always be replaced with original spare parts upon failure.
When having such work done, we suggest that you request performed, including information about any modifica- tions of the rated parameters or of the operating ranges (if applicable), as well as the date, name of organization and signature.
a certificate stating the type and extent of work
+
Modifications to the system Modifications of this system which could impair the
safety of the system owner, patients or other persons are prohibited by law!
For reasons of product safety, only original Sirona accessories approved for this product, or accessories from third parties approved by Sirona, may be used. The user is responsible for dangers resulting from the use of non-approved accessories.
If any devices not approved by Sirona are connected, they must comply with the applicable standards:
IEC 60950 for information technology equipment (e.g. PCs), and IEC 60601-1 for medical electrical equipment.
Combination with other equipment Any person who assembles or modifies a medical elec-
trical system complying with the standard IEC 60601-1-1 (safety requirements for medical electrical equipment) by combining it with other equipment (e.g., by connect­ing it with a PC) is responsible for ensuring that the requirements of this regulation are met to their full extent for the safety of the patients, operators and environment.
In case of doubt, contact the manufacturer of the system components.
60 29 560 D 3474
Sirona Dental Systems GmbH 2 Warning and safety information Operator Manual ProFeel
+
2.1 General warning and safety information
Hygiene and disinfection Highly immunosuppressed patients should not come in
contact with water from the treatment center. The use of sterile solutions is recommended.
The national requirements and recommendations for hygiene and disinfection must be observed (e.g. RKI, ADA, CDC etc).
Electromagnetic compatibility (EMC) The ProFeel
IEC 60601-1-2:2001.
Medical electrical devices are subject to special precau­tionary measures regarding EMC. It must be installed and operated in compliance with the specifications pro­vided in the document “Installationsvoraussetzungen”.
Portable and mobile RF communication systems may interfere with medical electrical equipment. Therefore, the use of mobile wireless phones in practice or hospital environments must be prohibited.
+
complies with the requirements of
With video systems:
Electromagnetic disturbances in the environment of the ProFeel+ may result in reduced image quality. In such cases, it is advisable to repeat the exposure before sav­ing the image.
In case of line voltage drops, image data stored in the video camera may be lost. Deposit the camera and repeat the exposure.
Image artifacts may possibly occur as a result of radio-frequency interference caused by the transmitters of e.g. radio services or radio amateurs.
Influence on cardiac pacemakers The workstation includes strong magnets which can
affect cardiac pacemakers of patients, users and techni­cal personnel.
Therefore, please pay attention to the safety information on page 18.
bеЦдблЬ
Treatment of highly immunosuppressed patients Highly immunosuppressed patients should not come in
contact with water from the treatment center. The use of sterile solutions is recommended.
Quality of water/air supply Air and water supply must meet the requirements speci-
fied in the installation instructions. Use only clean water.
Water quality compliance To ensure compliance with the medical and national
legal requirements for water from treatment centers, Sirona recommends equipping the treatment center with a disinfection system. As owner of the treatment center, you are responsible for the water quality and may have to take alternative measures to ensure its compliance if you operate the treatment center without a disinfection system.
Please contact your specialized dealer or your relevant dental association for the respective national require­ments and measures.
60 29 560 D 3474 D 3474.201.03.06.02
9
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General warning and safety information Operator Manual ProFeel
Identification of warning and safety information To avoid personal injury and material damage, you must
also observe the warning and safety information pro­vided in the present operating instructions. They are highlighted by the caption NOTE, CAUTION or
WARNING.
i
Symbols used Observe accompanying documents
(on rating plate of chair)
Ventilation slots Under no circumstances may the ventilation slots on the
unit be covered, since otherwise the air circulation will be obstructed.
Do not spray disinfectants or other similar products into the ventilation slots.
Vacuum system The suction removal of aluminum and other metal oxides
from blasting devices via the automatic separator and the amalgam separator integrated in the treatment cen­ter is prohibited!
This would cause extreme wear and clogging of the suc­tion and water paths.
A separate vacuum system must be used in connection with metal oxide blasting devices.
+
Treatment centers equipped with a central wet suction system are generally suitable for suction removal of the above material. However, make sure to observe the instructions provided by the manufacturer of your vac­uum system.
No restrictions apply when using salt blasting devices in connection with Sirona treatment centers. However, in such cases, make sure that the system is subsequently flushed with an adequate amount of water.
Cuspidor Do not use any foaming domestic cleaning agents!
Patient chair Please note that, according to EN ISO 6875, the dental
chair has a maximum load capacity of 135 kg (tested with four-fold safety factor according to IEC 60601-1).
Electric micromotors in dentist element Intermittent operation: 1 min. ON – 7 min. OFF with a
motor current of 1.5 A.
Drive motors for chair and backrest 6 % duty time, cycle time 250 s.
Maintenance of the treatment center Despite the outstanding quality of your treatment center
and regular care by the practice team, it is required in the interest of operational safety that preventive mainte­nance be performed at predetermined intervals.
In order to guarantee the operational safety and reliability of your treatment center and to avoid damage due to nat­ural wear, you as the system owner must have your system checked through regularly by an authorized servic e en gi­neer from your dental depot. Furthermore, safety checks must be performed.
60 29 560 D 3474
Sirona Dental Systems GmbH 2 Warning and safety information Operator Manual ProFeel
+
Please contact your dental depot to obtain a mainte­nance offer.
For details please refer to chapter “Maintenance”.
2.1 General warning and safety information
Dismantling and reassembly When dismantling and reassembling the system, pro-
ceed according to the installation instructions for new installation in order to guarantee its functioning and sta­bility.
Disposal It applies generally that the national regulations have to
be complied with when disposing of this product. Please observe the regulations applying in your country.
Within the European Economic Community the directive 2002/96/EEC (WEEE) for electrical and electronic devices requires environmentally compatible recycling / disposal.
Your product is marked with the adjacent symbol. With the goal of environmentally compatible recycling / dis­posal, your product must not be disposed of with the domestic refuse.
The black bar under the "refuse bin" symbol means that it has been put onto the market after 13.08.2005. (See DIN EN 50419:2005)
Please note that this product is subject to the directive 2002/96/EEC (WEEE) and laws applicable in your coun­try and must be sent for environmentally compatible recycling / disposal.
Please contact your dealer if your product should be finally disposed of.
bеЦдблЬ
60 29 560 D 3474 D 3474.201.03.06.02
Maximum load capacity of the treatment chair
Please observe the maximum load capacity of the chair of 135 kg according to EN ISO 6875 (tested with a four-fold safety factor according to IEC 60601-1).
11
2 Warning and safety information Sirona Dental Systems GmbH
STOP
STOP
STOP
STOP
STOP
STOP
STOP
2.1 General warning and safety information Operator Manual ProFeel
Safety stop for patient chair
Headrest
Backrest
Toeboard
Hand/foot placement
right
wrong
+
CAUTION
Make sure that no obstacles (e.g. window wings, draw­ers, devices, …) extend into the movement range.
Safety stop for assistant element
Cuspidor EMERGENCY STOP
Caution!
Danger of injury: The patient's legs may get
caught or be pinched.
• Prevent collision of the patient or seat upholstery with the support arm of the cuspidor during move­ment of the chair.
In case of a collision, a safety switch (switching strip underneath the cuspidor) cuts off the chair movement.
As soon as the pressure is removed from the switching strip (e.g. by swiveling away the cuspidor), the chair can be moved again.
60 29 560 D 3474
Sirona Dental Systems GmbH 2 Warning and safety information
!
!
Operator Manual ProFeel
+
Position of the dentist element
CAUTION
When moving the chair (via program or manually), make sure that the patient does not under any circumstances come into contact with the support arm or the dentist el­ement.
Patient mirror
Caution!
Fire hazard: There could be a risk of fire due to
the burning glass effect.
• Do not expose mirror to direct sunlight!
2.1 General warning and safety information
bеЦдблЬ
CAUTION
For the patient: Patients who are on a restricted sodium diet or suffer from severe respiratory illness should not undergo treat­ment with the powder jet instrument. Patients wearing contact lenses should remove them be­fore treatment or wear protective goggles.
For more safety information, please refer to the Operat­ing Instructions supplied with the ProSmile Handy.
60 29 560 D 3474 D 3474.201.03.06.02
13

3 Technical description Sirona Dental Systems GmbH

Operator Manual ProFeel
3 Technical description
+
Model designation ProFeel
Power supply 100 V/115 V/127 V/230 V AC 50/60 Hz
Nominal current 11.5 A at 100 V AC
10 A at 115 V AC 9 A at 127 V AC 5 A at 230 V AC
Main unit fuse T 10A H, 250 V AC, Order No. 10 77 460
Operating conditions Ambient temperature: 10 °C – 40 °C (50 °F – 104 °F)
Relative humidity: 30 % – 75 % Air pressure: 700 hPa – 1060 hPa
Transport and storage conditions Temperature: -40 °C – +70 °C (-40 °F – +158 °F)
Relative humidity: 10 % – 95 % Air pressure: 500 hPa – 1060 hPa
Protection class Class I equipment
Degree of protection against electric shock Type B applied parts,
except SIROCAM 3/SIROCAM C: Type BF applied parts
Degree of protection against ingress of water Ordinary equipment (not protected)
The foot switch has an IPX1 degree of protection against liquids (drip-proof).
Operating mode Continuous operation with intermittent loading corresponding to
the dental mode of working.
+
Permanently connected unit.
Year of manufacture On rating plate of chair
Tests/approvals This dental treatment center complies with the requirements of
IEC 60601-1: 1990 + A1: 1993 + A2: 1995 and of IEC 60601-1-2: 2001.
DVGW: This system complies with the technical rules and require­ments on safety and hygiene for connection to the drinking water supply (if the appropriate DVGW marking is affixed to it).
This product bears the CE marking in accordance with the provi­sions of the Council Directive 93/42/EEC of June 14, 1993 con­cerning medical devices (MDD).
0123
60 29 560 D 3474

Sirona Dental Systems GmbH 4 Controls and functional elements

P1
P
2
1
2
S
P
0
1 0 0
1
2 0
40
6 0
8 0
E N D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
Operator Manual ProFeel
+

4.1 System overview of ProFeel+ with OTP dentist element

4 Controls and functional elements
4.1 System overview of ProFeel
5.
3.
4.
1.
6.
+
with OTP dentist element
Sensor
2.
bеЦдблЬ
or
Possibilities for assigning the instrument positions
Position 1 Position 2 Position 3 Position 4 Position 5 On right
SPRAYVIT Electric motor
or
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
or
SIROSONIC L Mini L.E.D.
Camera
or
Mini L.E.D.
60 29 560 D 3474 D 3474.201.03.06.02
side of dentist element
7.
1. Patient chair
2. Dentist element
3. Water unit
4. Assistant element
5. SIROLUX
6. Instrument foot switch with program positions
7. Instrument foot switch
i
NOTE
For operation of the SIVISION 3 video system, please re­fer to Operating Instructions 59 67 893.
15
4 Controls and functional elements Sirona Dental Systems GmbH
P
1
P2
1
2
S
P
0
1 0 0
1
2 0
4 0
6
0
8 0
E
N D O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
4.2 System overview of ProFeel
+
with dentist element swivel arm Operator Manual ProFeel
+
4.2 System overview of ProFeel
5.
3.
4.
1.
+
with dentist element swivel arm
2.
6.
Possibilities for assigning the instrument positions
Position 1 Position 2 Position 3 Position 4 Position 5
SPRAYVIT Electric motor
or
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
ProSmile
Handy
or
Airscaler
or
Camera
or
Mini L.E.D.
SIROSONIC L
7.
1. Patient chair
2. Dentist element
3. Water unit
4. Assistant element
5. SIROLUX with sensor
6. Instrument foot switch with program positions
7. Instrument foot switch
60 29 560 D 3474

Sirona Dental Systems GmbH 5 Putting the system into operation

1
2
S
P
0
100
1
20
40
60
8
0
ENDO
Operator Manual ProFeel
+
5 Putting the system into operation
Initial startup
The disinfection system adds a disinfectant to the water (1:100) to prevent the formation of microorgan­isms in the water system.
Prior to initial startup of your treatment center, sanita- tion must be performed.
If the service engineer skipped the sanitation procedure after installation of your treatment center on the basis of an agreement with you or sanitation has not been per­formed for more than 2 weeks, please perform sanitation as described in Chapter 12.6 Disinfection system – sanitation.
Sanitation takes approx. 24 hours.
Main switch
Switch ON the main switch ( l ).
; The ready-for-operation LED lights up.
CAUTION
For safety reasons, always switch the treatment center OFF (O) after consulting hours. This cuts off the air and water supply as well as the line voltage.
bеЦдблЬ
60 29 560 D 3474 D 3474.201.03.06.02
17

6 Patient chair Sirona Dental Systems GmbH

6.1 Articulated headrest / Flat headrest Operator Manual ProFeel

6 Patient chair
6.1 Articulated headrest / Flat headrest
Adjusting the headrest
AB
2
1
4
3
A Articulated headrest (Option)
The headrest must be adjusted manually.
To do this, take the load off the backrest and pull out or push in the headrest using the handle (1).
Tilt the headrest backward using the locking button (2).
B Flat headrest
The headrest must be adjusted manually.
To do this, take the load off the backrest and pull out or push in the headrest using the handle (3).
CAUTION
A strong magnet is integrated in the underside of the head cushion (4) which could affect a nearby cardiac pacemaker.
+
Therefore, please avoid bringing the magnet in close proximity of a cardiac pacemaker of a patient, user or technical personnel.
In addition, direct contact of the head cushion with magnetic cards can lead to a data loss.
60 29 560 D 3474
Sirona Dental Systems GmbH 6 Patient chair
A
Operator Manual ProFeel
+

6.2 MultiMotion

6.2 MultiMotion
Adjustment to patient's height
The headrest can be adjusted to the patient's height by pulling out or pushing in the headrest extension. This adjustment is usually required at the beginning of treat­ment!
i
NOTE
Before positioning the patient on the headrest, make sure that the head support has been optimally adjusted to the patient's height. This simplifies all subsequent work with the MultiMotion considerably, since readjust­ment to the patient's height can thus be omitted when changing over from mandibular to maxillary treatment.
The MultiMotion enables you to adjust the patient's head in a way that optimally supports viewing of areas of the mouth which are difficult to access.
bеЦдблЬ
Adjustment of hyperextension
Mandibular position
The mandibular position can be set by pulling the oper­ating handle (A). The anatomical movement of the arched extension keeps the patient's head in the sup­port.
¾ Pull the headrest out of the guide by pulling
handle A.
i
NOTE
You can reduce the adjustment noise by pressing release A.
60 29 560 D 3474 D 3474.201.03.06.02
19
6 Patient chair Sirona Dental Systems GmbH
A
B
B
6.2 MultiMotion Operator Manual ProFeel
Maxillary position
¾ Take the load off the headrest by lifting it slightly.
¾ Press release A on the handle.
¾ Let the headrest slide down slowly.
¾ Let go of the release when the headrest reaches the
desired position.
+
Rotation and tilt
The MultiMotion enables rotation of the patient's head about the longitudinal axis of his body as well as lateral tilting of his head.
The MultiMotion is set to the desired position as follows:
1. Press and hold one or both of the two side control
elements B. The rotation and tilt adjustment locks of the headrest are now released.
2. Set the headrest to the desired position.
3. Let go of the control elements B.
; The supporting surface is fixed in place.
Make sure that the headrest is indeed fixed in place after you let go of the control elements!
60 29 560 D 3474
Sirona Dental Systems GmbH 6 Patient chair
C
E
D
F
Operator Manual ProFeel
+
6.2 MultiMotion
Removing the MultiMotion
For certain treatments (e.g. of children) it may be expe­dient to remove the MultiMotion completely in order to obtain better access to the patient. The patient's head will then be supported on the contact surface of the headrest extension F.
To remove the MultiMotion, proceed as follows:
1. Press release button C.
2. Pull the complete headrest out of the headrest ex-
tension.
3. Cover the opening of the headrest extension with
cover cap E.
4. If a children's head pad D is available, place it on
the headrest extension. The head pad is held mag­netically.
i
NOTE
The children's head pad (D) is not included in the scope of supply, however, is available from your dental dealer.
bеЦдблЬ
WARNING
The children's head pad (D) contains a strong mag­net on its bottom side. This magnet could affect any cardiac pacemaker located nearby.
Therefore, do not allow patients, users and technical personnel with a cardiac pacemaker to be located near this magnet.
Furthermore, direct contact of the head pad with magnetic cards can lead to data loss.
i
NOTE
Deposit the removed MultiMotion in a safe place where it cannot fall onto the floor.
Inserting the MultiMotion
To insert the MultiMotion, proceed as follows:
¾ Check the guide of the removed headrest for con-
tamination. Remove any contamination.
¾ Remove the cover cap E.
Reinsert the MultiMotion in the guide from above until it audibly locks in place.
Pull on the headrest again to make sure that it is locked securely in place.
60 29 560 D 3474 D 3474.201.03.06.02
21
6 Patient chair Sirona Dental Systems GmbH
1
2
S
P
0
10 0
1
2 0
4 0
6
0
80
E N D O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
S
P
100
1
20
40
60
80
ENDO
S
1
0
2
P
S
1
0
2
S
1
0
2

6.3 Patient positioning Operator Manual ProFeel

6.3 Patient positioning
Positioning with the control panel
1.
2.
You can start the motorized chair movements manually with the keys on the control panel of the dentist element.
; During operation, a short acoustic signal sounds.
Overview:
1. Backrest inclined backward
+
3.
4.
2. Backrest upright
3. Patient chair down
4. Patient chair up
Positioning with the 4-way foot control
1.
3.
2.
4.
You can start the motorized chair movements manually with the 4-way foot control.
; During operation, a short acoustic signal sounds.
Overview:
1. Patient chair up
2. Backrest upright
3. Backrest inclined backward
4. Patient chair down
Positioning with the control panel on the
1.
3.
2.
4.
assistant element
You can start the motorized chair movements manually with the keys on the control panel of the assistant ele­ment.
; During operation, a short acoustic signal sounds.
Overview:
1. Backrest inclined backward
2. Backrest upright
3. Patient chair down
4. Patient chair up
60 29 560 D 3474
Sirona Dental Systems GmbH 6 Patient chair
S
S
P
0
0
P
1
P
1
2
P
2
Operator Manual ProFeel
+
6.4 Chair programs – factory settings
6.4 Chair programs – factory settings
Chair programs
1 + 2
S 0 1 2
3
< 3 s
1. 2.
There are 4 factory-set chair positions. These settings can be adjusted individually (see section “Chair pro­grams – reprogramming” on page 24).
Program S
Reserved for mouth rinsing position.
Activation:
1. Press key P.
2. Press key S within 3 seconds.
The chair moves to the mouth rinsing position.
In combination with program key S it is possible to acti­vate the cuspidor flushing function and the fountain for filling the tumbler. See Chapters 10.6 and 10.7.
bеЦдблЬ
< 3 s
1. 2.
< 3 s
1. 2.
3. When you press key P and then key S again, the
treatment center returns to its previous position (memory position).
Program 0
Reserved for entry/exit position.
Activation:
1. Press key P.
2. Press key 0 within 3 seconds.
The chair moves into the entry/exit position.
Program 1
Reserved for upper jaw treatment position.
Activation:
1. Press key P.
2. Press key 1 within 3 seconds.
The chair moves to the lower jaw treatment position.
Program 2
Reserved for lower jaw treatment position.
60 29 560 D 3474 D 3474.201.03.06.02
< 3 s
1. 2.
Activation:
1. Press key P.
2. Press key 2 within 3 seconds.
The chair moves into the lower jaw treatment position.
23
6 Patient chair Sirona Dental Systems GmbH
1
2
S
P
0
1
0 0
1
2 0
4 0
6 0
8 0
E N D O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
1
2
S
P
0
100
1
20
40
60
80
ENDO
P
S
1
0
2
1
2
S
P
0
100
1
20
40
60
80
ENDO
1
2
S
P
0
100
1
20
40
60
80
ENDO
P
S
1
0
2
1
2
S
P
0
100
1
20
40
60
80
ENDO
S 0 1 2
S 0 1 2
S 0 1 2
6.5 Chair programs – reprogramming Operator Manual ProFeel
Activation with foot switch
Instead of activating the programs with the program keys, you also have the option of activating them with the 4-way foot control plate of the foot switch.
Actuate the 4-way foot control plate in the desired direc­tion.
to the front = program S to the back = program 0 to the left = program 1 to the right = program 2
The chair moves into the selected position.
6.5 Chair programs – reprogramming
i
NOTE
The factory program is overwritten on reprogramming.
+
1.
2.
4.
3.
Each chair program (S, 0, 1, 2) can be programmed indi­vidually.
Reprogramming of the four chair programs is always done in the same way:
1. Move the chair into the desired position by pressing
the corresponding keys on the control panel.
2. Press program key
P briefly.
3. Press the desired key (S, 0, 1, 2) within 3 seconds
and hold it down.
4. After 6 short audible signals and 1 long one, release
the key.
The new position is saved.
60 29 560 D 3474