Sirona CEREC MC, CEREC MC X Operating Instructions Manual

New as of:
Operating Instructions (valid for USA)
10.2017
Sirona Dental CAD/CAM System CEREC MC / MC X
English (US)
=
This product is covered by one or more of the following US patents:
• US6454629
• US6394880
• US6702649
• US7522764
• US7163443
63 98 650 D 3439
2 D 3439.201.07.10.23 10.2017
Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
3
Dentsply Sirona Table of contents Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Table of contents
1
Dear Customer, ....................................................................................................... 7
1.1 Contact information ...................................................................................... 7
2
General data............................................................................................................ 8
2.1 Identification of the danger levels................................................................. 8
2.2 Formats and symbols used .......................................................................... 9
2.3 Note PC / Acquisition Unit............................................................................ 9
3
General description ................................................................................................. 10
3.1 Certification .................................................................................................. 10
3.2 Intended use................................................................................................. 11
3.3 Further use of Sirona Dental CAD/CAM system .......................................... 12
4
Safety ...................................................................................................................... 13
4.1 Basic safety information ............................................................................... 13
4.1.1 Prerequisites ................................................................................... 13
4.1.2 Maintenance and repair................................................................... 13
4.1.3 Modifications to the product ............................................................ 13
4.1.4 Accessories ..................................................................................... 14
4.2 Opening the processing chamber door during the machining process ........ 14
4.3 Wireless phone interference with equipment ............................................... 14
4.4 Disturbance of data transmission................................................................. 14
5
Installation and startup ............................................................................................ 15
5.1 Transport and unpacking.............................................................................. 15
5.2 Disposal of packaging materials................................................................... 15
5.3 Installation site.............................................................................................. 16
5.4 Commissioning............................................................................................. 16
5.4.1 Functional elements ........................................................................ 17
5.4.2 Standard accessories...................................................................... 19
5.4.2.1 Instruments....................................................................... 19
5.4.2.2 Calibration pins................................................................. 19
5.4.2.3 Torque wrench.................................................................. 19
5.4.3 Display description .......................................................................... 21
5.4.4 Illumination of the processing chamber........................................... 21
5.4.5 Using the processing chamber screen............................................ 22
5.4.6 Connecting the bar code reader...................................................... 22
5.4.7 Installation ....................................................................................... 23
5.4.7.1 Connecting to the PC via LAN.......................................... 23
63 98 650 D 3439
4 D 3439.201.07.10.23 10.2017
Table of contents Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.7.2 Connecting the unit to the power supply ........................... 23
5.4.7.3 Installing the unit ............................................................... 23
5.4.7.4 Connecting the suction device (optional) .......................... 24
5.4.7.5 Connecting to the PC via WLAN (option).......................... 27
5.4.7.6 Operating several milling and grinding units over one access point
28
5.4.7.7 Connecting to the PC via the wireless interface (optional) 28
5.4.8 Filling the water tank ........................................................................ 30
5.4.8.1 Water tank MC / MC X ...................................................... 30
5.4.8.2 Water tank CEREC MC XL Basic...................................... 31
5.4.9 Switching the unit ON and OFF ....................................................... 32
5.5 Repacking ..................................................................................................... 33
5.6 Scope of supply............................................................................................. 33
5.7 Storage.......................................................................................................... 34
6
Operation.................................................................................................................. 35
6.1 Configuration (CEREC MC/MC X) ................................................................ 35
6.2 Calibrating the unit ........................................................................................ 36
6.3 Replacing the set screw................................................................................ 39
6.4 Machining process ........................................................................................ 40
6.4.1 Process types................................................................................... 40
6.4.1.1 Grinding............................................................................. 40
6.4.1.2 Milling ................................................................................ 41
6.4.1.3 Permitted instrument combinations................................... 42
6.4.2 Preparations..................................................................................... 42
6.4.3 Starting the machining processes .................................................... 42
6.4.4 Ending the machining processes ..................................................... 43
6.4.5 Information on the seal of approval.................................................. 43
6.5 Entering the bar code.................................................................................... 44
6.6 Manual block clamp ...................................................................................... 45
7
Service ..................................................................................................................... 46
7.1 Using the cleaning hose and the wet cleaning process ................................ 47
7.2 Changing filter bags and HEPA filters........................................................... 48
7.3 Changing the water....................................................................................... 50
7.3.1 General information.......................................................................... 50
7.3.2 Changing the water .......................................................................... 51
7.3.2.1 Water tank MC / MC X ...................................................... 51
7.3.2.2 Water tank CEREC MC XL Basic...................................... 53
7.4 Instruments ................................................................................................... 54
7.4.1 Overview of materials/instruments ................................................... 54
7.4.1.1 CEREC MC ....................................................................... 54
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
5
Dentsply Sirona Table of contents Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.4.1.2 CEREC MC X................................................................... 54
7.4.2 Changing instruments ..................................................................... 54
7.5 Care, cleaning agents, and disinfectants ..................................................... 56
7.6 Cleaning surfaces......................................................................................... 56
7.6.1 Disinfecting...................................................................................... 56
7.6.2 Protection against medicaments ..................................................... 56
7.6.3 Cleaning .......................................................................................... 56
7.7 Replacing the main fuses............................................................................. 57
7.8 Changing the filter ........................................................................................ 58
7.8.1 Water tank MC / MC X .................................................................... 58
7.8.2 Water tank CEREC MC XL Basic.................................................... 59
7.9 Removing water from the unit ...................................................................... 60
7.10 Using the tank cap opener ........................................................................... 61
8
Technical description............................................................................................... 62
8.1 System requirements ................................................................................... 62
8.1.1 CEREC MC ..................................................................................... 62
8.1.2 CEREC MC X.................................................................................. 62
8.2 Grinding and milling unit............................................................................... 62
8.2.1 General technical description.......................................................... 62
8.2.2 Technical data................................................................................. 63
8.2.3 Controller board............................................................................... 63
9
Consumable ............................................................................................................ 64
10
Disposal................................................................................................................... 66
Index........................................................................................................................ 67
63 98 650 D 3439
6 D 3439.201.07.10.23 10.2017
Table of contents Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
7
Dentsply Sirona 1Dear Customer, Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
1
Dear Customer,
Thank you for your purchase of this CEREC MC/MC X® unit from Dentsply Sirona.
This device enables you to produce dental restorations, e.g. from ceramic material with a natural appearance (CEramic REConstruction).
Improper use and handling can create hazards and cause damage. Please therefore read and follow these operating instructions carefully. Always keep them within easy reach.
Also pay attention to the safety instructions to prevent personal injury and material damage.
Your CEREC MC/MC X team
1.1
Contact information
Customer service center For technical questions, use the contact form on the internet at the
following address: http://srvcontact.sirona.com
Manufacturer's address
Sirona Dental Systems GmbH Fabrikstrasse 31 64625 Bensheim Germany
Tel.: +49 (0) 6251/16-0 Fax: +49 (0) 6251/16-2591 e-Mail: contact@dentsplysirona.com www.dentsplysirona.com
63 98 650 D 3439
8 D 3439.201.07.10.23 10.2017
2General data Dentsply Sirona
2.1Identification of the danger levels Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
2
General data
Please read this document completely and follow the instructions exactly. You should always keep it within reach.
Original language of the present document: German.
2.1
Identification of the danger levels
To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions. Such information is highlighted as follows:
Tip: Information for simplifying work.
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily injury or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTICE
A possibly harmful situation which could lead to damage of the product or an object in its environment.
IMPORTANT
Application instructions and other important information.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
9
Dentsply Sirona 2General data Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
2.2
Formats and symbols used
The formats and symbols used in this document have the following meaning:
2.3
Note PC / Acquisition Unit
When a PC is described in this document, this refers to a PC for the acquisition unit (if present). The PC is represented symbolically.
Please observe our recommendations for PC configuration (see System requirements [ → 62]).
Prerequisite
1. First action step
2. Second action step or
Alternative action
Result
➢ Individual action step
Prompts you to do something.
See “Formats and symbols used [ → 9]”
Identifies a reference to another text passage and specifies its page number.
● List Designates a list.
“Command / menu item” Indicates commands, menu items or
quotations.
63 98 650 D 3439
10 D 3439.201.07.10.23 10.2017
3General description Dentsply Sirona
3.1Certification Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3
General description
3.1
Certification
CE mark
This product bears the CE mark in accordance with the provisions of Council Directive 2006/42/EC (machinery directive). As such, the following standards apply: DIN EN ISO 12100:2011-03, DIN EN 61010­1:2011-07 and DIN EN 61326-1:2013-07.
Examples of CE mark for connected products:
● EN 60601-1:2006 based on IEC 60601-1:2005
● EN 60950-1:2006 based on IEC 60950-1:2005
● UL 60950 second edition 2010
GOST mark
EAC certification
Mark of conformity of the Eurasian Economic Community
CAUTION
CE mark for connected products
Further products which are connected to this unit must also bear the CE mark. These products must be tested according to the applicable standards.
24
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
11
Dentsply Sirona 3General description Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3.2
Intended use
The Sirona Dental CAD/CAM System is intended for use in partially or fully edentulous mandibles and maxillae in support of single or multiple­unit cement retained restorations. For the SSO 3.5 L and SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible. The system consists of three major parts: TiBase, inCoris mesostructure, and CAD/CAM software. Specifically, the inCoris mesostructure and TiBase components make up a two-piece abutment which is used in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity. The inCoris mesostructure may also be used in conjunction with the Camlog Titanium base CAD/ CAM (types K2244.xxxx) (K083496) in the Camlog Implant System. The CAD/CAM software is intended to design and fabricate the inCoris mesostructure. The inCoris mesostructure and TiBase two-piece abutment is compatible with the following implant systems:
● Nobel Biocare Replace (K020646)
● Nobel Biocare Branemark (K022562)
● Friadent Xive (K013867)
● Biomet 3i Osseotite (K980549)
● Astra Tech Osseospeed (K091239)
● Zimmer Tapered Screw-Vent (K061410)
● Straumann SynOcta (K061176)
● Straumann Bone Level (K053088, K062129, K060958)
● Biomet 3i Certain (K014235, K061629)
● Nobel Biocare Active (K071370)
CAUTION
Small diameter implants and large angled abutments in the anterior region of the mouth due to possible failure of the implant system.
CAUTION
Federal Law (USA) restricts the sale of this device to or on the order of a physician, dentist, or licensed practitioner.
63 98 650 D 3439
12 D 3439.201.07.10.23 10.2017
3General description Dentsply Sirona
3.3Further use of Sirona Dental CAD/CAM system Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3.3
Further use of Sirona Dental CAD/CAM system
The Sirona Dental CAD/CAM System is also:
● an optical impression system for computer assisted design and manufacturing (CAD/CAM) according to 21 CFR 872.3661. The system records the topographical characteristics of teeth, dental impressions, or stone models for use in the computer-assisted design and manufacturing of dental restorative prosthetic devices.
● an endosseous dental implant accessory according to 21 CFR
872.3980. The system is used to produce a part that the user can manually incorporate together with other 3rd party components into a dental surgery guide, a temporary accessory used with endosseous dental implants with tissue contact for less than 1 hour (exempt).
Such devices are exempt from the premarket notification procedures.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
13
Dentsply Sirona 4Safety Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
4
Safety
4.1
Basic safety information
4.1.1 Prerequisites
4.1.2 Maintenance and repair
As manufacturers of dental instruments and laboratory equipment, we can assume responsibility for the safety properties of the unit only if the following points are observed:
● The maintenance and repair of this unit may be performed only by Dentsply Sirona or by agencies authorized by Dentsply Sirona.
● Components which have failed and influence the safety of the unit must be replaced with original (OEM) spare parts.
Please request a certificate whenever you have such work performed. It should include:
● The type and scope of work.
● Any changes made in the rated parameters or working range.
● Date, name of company and signature.
4.1.3 Modifications to the product
Modifications to this product which may affect the safety of the operator, patients or third parties are prohibited by law!
NOTICE
Important information on building installation
The building installation must be performed by a qualified expert in compliance with the national regulations. DIN VDE 0100-710 applies in Germany.
NOTICE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion hazards.
NOTICE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
63 98 650 D 3439
14 D 3439.201.07.10.23 10.2017
4Safety Dentsply Sirona
4.2Opening the processing chamber door during the machining process Operating Instructions (valid for USA) Sirona Dental CAD/CAM
4.1.4 Accessories
In order to ensure product safety, this device may be operated only with original Dentsply Sirona accessories or third-party accessories expressly approved by Dentsply Sirona. In particular, only the power cable also supplied or the corresponding original spare part may be used with the unit. The user is responsible for any damage resulting from the use of non-approved accessories.
4.2
Opening the processing chamber door during the machining process
4.3
Wireless phone interference with equipment
The use of mobile wireless phones in practice or hospital environments must be prohibited to ensure safe operation of the unit.
4.4
Disturbance of data transmission
Note on wireless communication
Fragment Worldwide
Data communication between the acquisition unit and the CEREC MC/ MC X milling and grinding unit should preferably be established via the wireless interface CEREC Radio Device or WLAN.
As for all wireless connections (e.g. cell phones), heavy utilization of the available radio channels or shielding caused by building installations (e.g. metal-shielded X-ray enclosures) may impair the quality of the connection. This may become noticeable through a reduction in range and/or a slower data transmission rate. In extreme cases, it will be impossible to establish a wireless connection at all.
Dentsply Sirona has selected the best possible configuration for data communication via the wireless interface (CEREC Radio Device) or WLAN, which generally ensures perfect functioning of this connection. However, in individual cases unrestricted wireless data communication may be impossible for the reasons mentioned above and/or due to local circumstances. In such cases, a cable LAN connection should be selected to ensure uninterrupted operation. If the only LAN interface on the rear of the CEREC AC is occupied by another plug, remove this wireless interface connection, and instead connect the LAN cable with the CEREC MC/MC X milling and grinding unit.
CAUTION
Instruments that continue to run
When the processing chamber door is opened during the machining process, the instruments could continue to run for a short time.
➢ Be careful not to touch the instruments with your hand or any other
object during this time.
➢ Avoid opening the processing chamber door while the milling and
grinding unit is in operation.
➢ Before you open the processing chamber door, end any actions that
are running by pressing the
"Stop"
button on the milling and
grinding unit or in the application software.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
15
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5
Installation and startup
5.1
Transport and unpacking
All products from Dentsply Sirona are carefully checked prior to shipment. Please perform an incoming inspection immediately after delivery.
1. Check the delivery note to ensure that the consignment is complete.
2. Check whether the product shows any visible signs of damage.
If return shipment is required, please use the original packaging for shipment.
The unit must be drained prior to shipment (if it has been operated). Removing water from the unit
Transport without packaging
5.2
Disposal of packaging materials
The packaging must be disposed of in compliance with the relevant national regulations. Please observe the regulations applicable in your country.
NOTICE
Damage during transport
If the product was damaged during transport, please contact your carrying agent.
CAUTION
Damage to the unit or risk of injury during transport without packaging
There is a danger of the unit falling down if it is grasped by its plastic housing.
➢ The unit should always be carried by two persons. ➢ Do not grasp the unit by its plastic housing. ➢ Always grasp the unit by its chassis next to its feet.
63 98 650 D 3439
16 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.3Installation site Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.3
Installation site
The milling and grinding unit requires a level Approx. footprint: 700 x 420 mm (W x D). The height of the milling and grinding unit is:
● with the processing chamber door closed: 425mm
● with the processing chamber door, open: 570mm
Install the milling and grinding unit in such a way that it is not difficult to operate the main switch.
Make sure that the ventilation slots underneath and at the back of the unit remain unobstructed. The distance between the back of the unit and the wall must at least be 10 cm.
Note that the unit weighs 43 kg!
The unit must not be installed at sites with a high level of humidity or dust!
5.4
Commissioning
CAUTION
Install out of the reach of patients!
Do not install or operate the milling and grinding unit in the vicinity of the patient (place it at least 1.5 m away from the patient).
NOTICE
Installation in a cabinet
If the unit is installed in a cabinet, you must provide for adequate heat exchange.
The ambient temperature surrounding the unit must be between 5°C and 40°C.
NOTICE
Important information on initial startup
Observe the software installation instructions!
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
17
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.1 Functional elements
Unit overview
Overview of the milling and grinding unit
A Processing chamber D ON/OFF switch B Processing chamber
door catch
EDrawer
C Display F Water tank
A
B
C D E F
E
63 98 650 D 3439
18 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Ports on the back side
Connections
Machining room
Machining room
A Main switch
I = ON, 0 = OFF
E Communications interface for
suction B Fuse cover F Bar code reader connection C Power connection G Connection for suction D LAN port
Ethernet
A Instrument set D Workpiece spindle B Motor mount E Suction connection C Ceramic block
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
19
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.2 Standard accessories
5.4.2.1
Instruments
The following instruments are available for milling and grinding purposes. When replacing instruments, ensure the permitted instrument combinations are used (see "Permitted instrument combinations [ → 42]").
5.4.2.2
Calibration pins
The calibration pins are used when calibrating the instrument sets (see "Calibrating the unit [ → 36]“).
5.4.2.3
Torque wrench
To insert or replace the instruments or calibration pins, use the following torque wrench.
63 98 650 D 3439
20 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
* All-in-One
Instrument REF Usage Torque wrench Clamping format of
the force transmission
Step Bur 12 S 6240167 Grinding Triangular
Step Bur 12 6260025 Grinding
Cyl. Pointed Bur 12 S 6240159 Grinding
Step Bur 20 6259597 Grinding
Cyl. Pointed Bur 20 6259589 Grinding
Shaper 25 6299395 Milling
(dry)
Finisher 10 6299387 Milling
Calibration pin (AiO*) 6241132 Calibration
Shaper 25 RZ 6433440 Milling
(wet)
Square
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
21
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.3 Display description
These operating instructions describe how to operate the unit by executing and confirming commands via your PC.
You can also confirm commands such as
"Start" , "Stop" , "Cancel"
or
"OK"
directly on the display of your milling and grinding unit.
Possible commands are then shown above the corresponding button on the display. In the example shown, button 1, (A) would confirm the command
"Start"
and button 4, (D) would confirm the command
"Stop"
.
Display
5.4.4 Illumination of the processing chamber
The lighting of the processing chamber depends on the machining operation involved:
A Button 1 E Command B Button 2 F Display C Button 3 G ON/OFF switch D Button 4
Machining operation Lighting color Milling and grinding White Operation completed Green Error or
"Stop"
button pressed Red