Sirona CEREC MC, CEREC MC X Operating Instructions Manual

New as of:
Operating Instructions (valid for USA)
10.2017
Sirona Dental CAD/CAM System CEREC MC / MC X
English (US)
=
This product is covered by one or more of the following US patents:
• US6454629
• US6394880
• US6702649
• US7522764
• US7163443
63 98 650 D 3439
2 D 3439.201.07.10.23 10.2017
Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
3
Dentsply Sirona Table of contents Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Table of contents
1
Dear Customer, ....................................................................................................... 7
1.1 Contact information ...................................................................................... 7
2
General data............................................................................................................ 8
2.1 Identification of the danger levels................................................................. 8
2.2 Formats and symbols used .......................................................................... 9
2.3 Note PC / Acquisition Unit............................................................................ 9
3
General description ................................................................................................. 10
3.1 Certification .................................................................................................. 10
3.2 Intended use................................................................................................. 11
3.3 Further use of Sirona Dental CAD/CAM system .......................................... 12
4
Safety ...................................................................................................................... 13
4.1 Basic safety information ............................................................................... 13
4.1.1 Prerequisites ................................................................................... 13
4.1.2 Maintenance and repair................................................................... 13
4.1.3 Modifications to the product ............................................................ 13
4.1.4 Accessories ..................................................................................... 14
4.2 Opening the processing chamber door during the machining process ........ 14
4.3 Wireless phone interference with equipment ............................................... 14
4.4 Disturbance of data transmission................................................................. 14
5
Installation and startup ............................................................................................ 15
5.1 Transport and unpacking.............................................................................. 15
5.2 Disposal of packaging materials................................................................... 15
5.3 Installation site.............................................................................................. 16
5.4 Commissioning............................................................................................. 16
5.4.1 Functional elements ........................................................................ 17
5.4.2 Standard accessories...................................................................... 19
5.4.2.1 Instruments....................................................................... 19
5.4.2.2 Calibration pins................................................................. 19
5.4.2.3 Torque wrench.................................................................. 19
5.4.3 Display description .......................................................................... 21
5.4.4 Illumination of the processing chamber........................................... 21
5.4.5 Using the processing chamber screen............................................ 22
5.4.6 Connecting the bar code reader...................................................... 22
5.4.7 Installation ....................................................................................... 23
5.4.7.1 Connecting to the PC via LAN.......................................... 23
63 98 650 D 3439
4 D 3439.201.07.10.23 10.2017
Table of contents Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.7.2 Connecting the unit to the power supply ........................... 23
5.4.7.3 Installing the unit ............................................................... 23
5.4.7.4 Connecting the suction device (optional) .......................... 24
5.4.7.5 Connecting to the PC via WLAN (option).......................... 27
5.4.7.6 Operating several milling and grinding units over one access point
28
5.4.7.7 Connecting to the PC via the wireless interface (optional) 28
5.4.8 Filling the water tank ........................................................................ 30
5.4.8.1 Water tank MC / MC X ...................................................... 30
5.4.8.2 Water tank CEREC MC XL Basic...................................... 31
5.4.9 Switching the unit ON and OFF ....................................................... 32
5.5 Repacking ..................................................................................................... 33
5.6 Scope of supply............................................................................................. 33
5.7 Storage.......................................................................................................... 34
6
Operation.................................................................................................................. 35
6.1 Configuration (CEREC MC/MC X) ................................................................ 35
6.2 Calibrating the unit ........................................................................................ 36
6.3 Replacing the set screw................................................................................ 39
6.4 Machining process ........................................................................................ 40
6.4.1 Process types................................................................................... 40
6.4.1.1 Grinding............................................................................. 40
6.4.1.2 Milling ................................................................................ 41
6.4.1.3 Permitted instrument combinations................................... 42
6.4.2 Preparations..................................................................................... 42
6.4.3 Starting the machining processes .................................................... 42
6.4.4 Ending the machining processes ..................................................... 43
6.4.5 Information on the seal of approval.................................................. 43
6.5 Entering the bar code.................................................................................... 44
6.6 Manual block clamp ...................................................................................... 45
7
Service ..................................................................................................................... 46
7.1 Using the cleaning hose and the wet cleaning process ................................ 47
7.2 Changing filter bags and HEPA filters........................................................... 48
7.3 Changing the water....................................................................................... 50
7.3.1 General information.......................................................................... 50
7.3.2 Changing the water .......................................................................... 51
7.3.2.1 Water tank MC / MC X ...................................................... 51
7.3.2.2 Water tank CEREC MC XL Basic...................................... 53
7.4 Instruments ................................................................................................... 54
7.4.1 Overview of materials/instruments ................................................... 54
7.4.1.1 CEREC MC ....................................................................... 54
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
5
Dentsply Sirona Table of contents Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.4.1.2 CEREC MC X................................................................... 54
7.4.2 Changing instruments ..................................................................... 54
7.5 Care, cleaning agents, and disinfectants ..................................................... 56
7.6 Cleaning surfaces......................................................................................... 56
7.6.1 Disinfecting...................................................................................... 56
7.6.2 Protection against medicaments ..................................................... 56
7.6.3 Cleaning .......................................................................................... 56
7.7 Replacing the main fuses............................................................................. 57
7.8 Changing the filter ........................................................................................ 58
7.8.1 Water tank MC / MC X .................................................................... 58
7.8.2 Water tank CEREC MC XL Basic.................................................... 59
7.9 Removing water from the unit ...................................................................... 60
7.10 Using the tank cap opener ........................................................................... 61
8
Technical description............................................................................................... 62
8.1 System requirements ................................................................................... 62
8.1.1 CEREC MC ..................................................................................... 62
8.1.2 CEREC MC X.................................................................................. 62
8.2 Grinding and milling unit............................................................................... 62
8.2.1 General technical description.......................................................... 62
8.2.2 Technical data................................................................................. 63
8.2.3 Controller board............................................................................... 63
9
Consumable ............................................................................................................ 64
10
Disposal................................................................................................................... 66
Index........................................................................................................................ 67
63 98 650 D 3439
6 D 3439.201.07.10.23 10.2017
Table of contents Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
7
Dentsply Sirona 1Dear Customer, Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
1
Dear Customer,
Thank you for your purchase of this CEREC MC/MC X® unit from Dentsply Sirona.
This device enables you to produce dental restorations, e.g. from ceramic material with a natural appearance (CEramic REConstruction).
Improper use and handling can create hazards and cause damage. Please therefore read and follow these operating instructions carefully. Always keep them within easy reach.
Also pay attention to the safety instructions to prevent personal injury and material damage.
Your CEREC MC/MC X team
1.1
Contact information
Customer service center For technical questions, use the contact form on the internet at the
following address: http://srvcontact.sirona.com
Manufacturer's address
Sirona Dental Systems GmbH Fabrikstrasse 31 64625 Bensheim Germany
Tel.: +49 (0) 6251/16-0 Fax: +49 (0) 6251/16-2591 e-Mail: contact@dentsplysirona.com www.dentsplysirona.com
63 98 650 D 3439
8 D 3439.201.07.10.23 10.2017
2General data Dentsply Sirona
2.1Identification of the danger levels Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
2
General data
Please read this document completely and follow the instructions exactly. You should always keep it within reach.
Original language of the present document: German.
2.1
Identification of the danger levels
To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions. Such information is highlighted as follows:
Tip: Information for simplifying work.
DANGER
An imminent danger that could result in serious bodily injury or death.
WARNING
A possibly dangerous situation that could result in serious bodily injury or death.
CAUTION
A possibly dangerous situation that could result in slight bodily injury.
NOTICE
A possibly harmful situation which could lead to damage of the product or an object in its environment.
IMPORTANT
Application instructions and other important information.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
9
Dentsply Sirona 2General data Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
2.2
Formats and symbols used
The formats and symbols used in this document have the following meaning:
2.3
Note PC / Acquisition Unit
When a PC is described in this document, this refers to a PC for the acquisition unit (if present). The PC is represented symbolically.
Please observe our recommendations for PC configuration (see System requirements [ → 62]).
Prerequisite
1. First action step
2. Second action step or
Alternative action
Result
➢ Individual action step
Prompts you to do something.
See “Formats and symbols used [ → 9]”
Identifies a reference to another text passage and specifies its page number.
● List Designates a list.
“Command / menu item” Indicates commands, menu items or
quotations.
63 98 650 D 3439
10 D 3439.201.07.10.23 10.2017
3General description Dentsply Sirona
3.1Certification Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3
General description
3.1
Certification
CE mark
This product bears the CE mark in accordance with the provisions of Council Directive 2006/42/EC (machinery directive). As such, the following standards apply: DIN EN ISO 12100:2011-03, DIN EN 61010­1:2011-07 and DIN EN 61326-1:2013-07.
Examples of CE mark for connected products:
● EN 60601-1:2006 based on IEC 60601-1:2005
● EN 60950-1:2006 based on IEC 60950-1:2005
● UL 60950 second edition 2010
GOST mark
EAC certification
Mark of conformity of the Eurasian Economic Community
CAUTION
CE mark for connected products
Further products which are connected to this unit must also bear the CE mark. These products must be tested according to the applicable standards.
24
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
11
Dentsply Sirona 3General description Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3.2
Intended use
The Sirona Dental CAD/CAM System is intended for use in partially or fully edentulous mandibles and maxillae in support of single or multiple­unit cement retained restorations. For the SSO 3.5 L and SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible. The system consists of three major parts: TiBase, inCoris mesostructure, and CAD/CAM software. Specifically, the inCoris mesostructure and TiBase components make up a two-piece abutment which is used in conjunction with endosseous dental implants to restore the function and aesthetics in the oral cavity. The inCoris mesostructure may also be used in conjunction with the Camlog Titanium base CAD/ CAM (types K2244.xxxx) (K083496) in the Camlog Implant System. The CAD/CAM software is intended to design and fabricate the inCoris mesostructure. The inCoris mesostructure and TiBase two-piece abutment is compatible with the following implant systems:
● Nobel Biocare Replace (K020646)
● Nobel Biocare Branemark (K022562)
● Friadent Xive (K013867)
● Biomet 3i Osseotite (K980549)
● Astra Tech Osseospeed (K091239)
● Zimmer Tapered Screw-Vent (K061410)
● Straumann SynOcta (K061176)
● Straumann Bone Level (K053088, K062129, K060958)
● Biomet 3i Certain (K014235, K061629)
● Nobel Biocare Active (K071370)
CAUTION
Small diameter implants and large angled abutments in the anterior region of the mouth due to possible failure of the implant system.
CAUTION
Federal Law (USA) restricts the sale of this device to or on the order of a physician, dentist, or licensed practitioner.
63 98 650 D 3439
12 D 3439.201.07.10.23 10.2017
3General description Dentsply Sirona
3.3Further use of Sirona Dental CAD/CAM system Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3.3
Further use of Sirona Dental CAD/CAM system
The Sirona Dental CAD/CAM System is also:
● an optical impression system for computer assisted design and manufacturing (CAD/CAM) according to 21 CFR 872.3661. The system records the topographical characteristics of teeth, dental impressions, or stone models for use in the computer-assisted design and manufacturing of dental restorative prosthetic devices.
● an endosseous dental implant accessory according to 21 CFR
872.3980. The system is used to produce a part that the user can manually incorporate together with other 3rd party components into a dental surgery guide, a temporary accessory used with endosseous dental implants with tissue contact for less than 1 hour (exempt).
Such devices are exempt from the premarket notification procedures.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
13
Dentsply Sirona 4Safety Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
4
Safety
4.1
Basic safety information
4.1.1 Prerequisites
4.1.2 Maintenance and repair
As manufacturers of dental instruments and laboratory equipment, we can assume responsibility for the safety properties of the unit only if the following points are observed:
● The maintenance and repair of this unit may be performed only by Dentsply Sirona or by agencies authorized by Dentsply Sirona.
● Components which have failed and influence the safety of the unit must be replaced with original (OEM) spare parts.
Please request a certificate whenever you have such work performed. It should include:
● The type and scope of work.
● Any changes made in the rated parameters or working range.
● Date, name of company and signature.
4.1.3 Modifications to the product
Modifications to this product which may affect the safety of the operator, patients or third parties are prohibited by law!
NOTICE
Important information on building installation
The building installation must be performed by a qualified expert in compliance with the national regulations. DIN VDE 0100-710 applies in Germany.
NOTICE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion hazards.
NOTICE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
63 98 650 D 3439
14 D 3439.201.07.10.23 10.2017
4Safety Dentsply Sirona
4.2Opening the processing chamber door during the machining process Operating Instructions (valid for USA) Sirona Dental CAD/CAM
4.1.4 Accessories
In order to ensure product safety, this device may be operated only with original Dentsply Sirona accessories or third-party accessories expressly approved by Dentsply Sirona. In particular, only the power cable also supplied or the corresponding original spare part may be used with the unit. The user is responsible for any damage resulting from the use of non-approved accessories.
4.2
Opening the processing chamber door during the machining process
4.3
Wireless phone interference with equipment
The use of mobile wireless phones in practice or hospital environments must be prohibited to ensure safe operation of the unit.
4.4
Disturbance of data transmission
Note on wireless communication
Fragment Worldwide
Data communication between the acquisition unit and the CEREC MC/ MC X milling and grinding unit should preferably be established via the wireless interface CEREC Radio Device or WLAN.
As for all wireless connections (e.g. cell phones), heavy utilization of the available radio channels or shielding caused by building installations (e.g. metal-shielded X-ray enclosures) may impair the quality of the connection. This may become noticeable through a reduction in range and/or a slower data transmission rate. In extreme cases, it will be impossible to establish a wireless connection at all.
Dentsply Sirona has selected the best possible configuration for data communication via the wireless interface (CEREC Radio Device) or WLAN, which generally ensures perfect functioning of this connection. However, in individual cases unrestricted wireless data communication may be impossible for the reasons mentioned above and/or due to local circumstances. In such cases, a cable LAN connection should be selected to ensure uninterrupted operation. If the only LAN interface on the rear of the CEREC AC is occupied by another plug, remove this wireless interface connection, and instead connect the LAN cable with the CEREC MC/MC X milling and grinding unit.
CAUTION
Instruments that continue to run
When the processing chamber door is opened during the machining process, the instruments could continue to run for a short time.
➢ Be careful not to touch the instruments with your hand or any other
object during this time.
➢ Avoid opening the processing chamber door while the milling and
grinding unit is in operation.
➢ Before you open the processing chamber door, end any actions that
are running by pressing the
"Stop"
button on the milling and
grinding unit or in the application software.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
15
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5
Installation and startup
5.1
Transport and unpacking
All products from Dentsply Sirona are carefully checked prior to shipment. Please perform an incoming inspection immediately after delivery.
1. Check the delivery note to ensure that the consignment is complete.
2. Check whether the product shows any visible signs of damage.
If return shipment is required, please use the original packaging for shipment.
The unit must be drained prior to shipment (if it has been operated). Removing water from the unit
Transport without packaging
5.2
Disposal of packaging materials
The packaging must be disposed of in compliance with the relevant national regulations. Please observe the regulations applicable in your country.
NOTICE
Damage during transport
If the product was damaged during transport, please contact your carrying agent.
CAUTION
Damage to the unit or risk of injury during transport without packaging
There is a danger of the unit falling down if it is grasped by its plastic housing.
➢ The unit should always be carried by two persons. ➢ Do not grasp the unit by its plastic housing. ➢ Always grasp the unit by its chassis next to its feet.
63 98 650 D 3439
16 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.3Installation site Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.3
Installation site
The milling and grinding unit requires a level Approx. footprint: 700 x 420 mm (W x D). The height of the milling and grinding unit is:
● with the processing chamber door closed: 425mm
● with the processing chamber door, open: 570mm
Install the milling and grinding unit in such a way that it is not difficult to operate the main switch.
Make sure that the ventilation slots underneath and at the back of the unit remain unobstructed. The distance between the back of the unit and the wall must at least be 10 cm.
Note that the unit weighs 43 kg!
The unit must not be installed at sites with a high level of humidity or dust!
5.4
Commissioning
CAUTION
Install out of the reach of patients!
Do not install or operate the milling and grinding unit in the vicinity of the patient (place it at least 1.5 m away from the patient).
NOTICE
Installation in a cabinet
If the unit is installed in a cabinet, you must provide for adequate heat exchange.
The ambient temperature surrounding the unit must be between 5°C and 40°C.
NOTICE
Important information on initial startup
Observe the software installation instructions!
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
17
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.1 Functional elements
Unit overview
Overview of the milling and grinding unit
A Processing chamber D ON/OFF switch B Processing chamber
door catch
EDrawer
C Display F Water tank
A
B
C D E F
E
63 98 650 D 3439
18 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Ports on the back side
Connections
Machining room
Machining room
A Main switch
I = ON, 0 = OFF
E Communications interface for
suction B Fuse cover F Bar code reader connection C Power connection G Connection for suction D LAN port
Ethernet
A Instrument set D Workpiece spindle B Motor mount E Suction connection C Ceramic block
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
19
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.2 Standard accessories
5.4.2.1
Instruments
The following instruments are available for milling and grinding purposes. When replacing instruments, ensure the permitted instrument combinations are used (see "Permitted instrument combinations [ → 42]").
5.4.2.2
Calibration pins
The calibration pins are used when calibrating the instrument sets (see "Calibrating the unit [ → 36]“).
5.4.2.3
Torque wrench
To insert or replace the instruments or calibration pins, use the following torque wrench.
63 98 650 D 3439
20 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
* All-in-One
Instrument REF Usage Torque wrench Clamping format of
the force transmission
Step Bur 12 S 6240167 Grinding Triangular
Step Bur 12 6260025 Grinding
Cyl. Pointed Bur 12 S 6240159 Grinding
Step Bur 20 6259597 Grinding
Cyl. Pointed Bur 20 6259589 Grinding
Shaper 25 6299395 Milling
(dry)
Finisher 10 6299387 Milling
Calibration pin (AiO*) 6241132 Calibration
Shaper 25 RZ 6433440 Milling
(wet)
Square
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
21
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.3 Display description
These operating instructions describe how to operate the unit by executing and confirming commands via your PC.
You can also confirm commands such as
"Start" , "Stop" , "Cancel"
or
"OK"
directly on the display of your milling and grinding unit.
Possible commands are then shown above the corresponding button on the display. In the example shown, button 1, (A) would confirm the command
"Start"
and button 4, (D) would confirm the command
"Stop"
.
Display
5.4.4 Illumination of the processing chamber
The lighting of the processing chamber depends on the machining operation involved:
A Button 1 E Command B Button 2 F Display C Button 3 G ON/OFF switch D Button 4
Machining operation Lighting color Milling and grinding White Operation completed Green Error or
"Stop"
button pressed Red
63 98 650 D 3439
22 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.5 Using the processing chamber screen
1. Remove the screen from the packaging.
2. Wet the underside of the sieve with water before insertion and press it firmly against the floor of the processing chamber.
5.4.6 Connecting the bar code reader
The bar code reader is optional for the CEREC MC/MC X unit.
Connecting the bar code reader
➢ Plug the bar code reader into the serial interface to the rear of the
milling and grinding unit and secure with screws.
CAUTION
Risk of injury on instruments
Be careful not to brush against the instruments with your hand.
NOTICE
Risk of blockage in the cooling circuit
If chips enter into the cooling circuit of the machine, there is a risk that the cooling circuit will become blocked.
➢ The sieve is suitable for all restoration and material types. It is
absolutely essential that no chips enter into the cooling circuit.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
23
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.7 Installation
5.4.7.1
Connecting to the PC via LAN
An Ethernet connection is located to the rear of the unit, which can be used to connect the PC to the milling and grinding unit. Use a network cable to do this (LAN connection).
Using a network cable
Connect the PC to the LAN connection of the unit. If problems arise when connecting via a network cable, please read the
separate instructions "Operating the MC XL via LAN".
5.4.7.2
Connecting the unit to the power supply
➢ Connect the unit to the power supply using the supplied power cable.
5.4.7.3
Installing the unit
You must connect the unit to the PC before putting it into operation. This is described in the section entitled "Connecting to the PC via LAN" [ → 23] or "Connecting to the PC via WLAN (option)" [ → 27].
Searching for unit automatically
The unit is connected to the PC via a LAN cable or via WLAN.
1. Click the
"Configuration"
button in the system menu.
2. Click on the
"Devices"
button.
3. Click on the
"Scan for New Devices"
button.
All units connected to the PC are recognized. In the case of new
units, you will be prompted to enter a name.
4. Enter a name for the new unit.
NOTICE
Grounded power outlet
The unit must be connected to a grounded power outlet.
63 98 650 D 3439
24 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Search for unit manually
The unit is connected to the PC via a LAN cable or via WLAN.
1. Click the
"Configuration"
button in the system menu.
2. Click on the
"Devices"
button.
3. Click on the
"Add Device (Manual)"
button.
4. Set the network.
5. Enter the network address which appears on the
"IP address:"
display once the milling and grinding unit has been switched on.
6. Click on the
"Ok"
button.
The software attempts to contact the device.
If the connection fails, check the connection. If necessary, ask a qualified technician.
Unit removal
If you no longer require a unit (e.g. a unit is replaced), you can remove
it.
The unit is operation.
1. Click the
"Configuration"
button in the system menu.
2. Click on the
"Devices"
button.
3. Click on the unit that you wish to uninstall.
4. Click on the
"Delete Device"
button.
You will be asked if you would like to remove the unit.
5. Click on the
"YES"
button.
The device is removed.
5.4.7.4
Connecting the suction device (optional)
5.4.7.4.1 Connecting the suction tube
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
25
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
1. Connect one end of the suction tube to the available connection point on the suction device (A).
2. Connect the other end of the tube to the rear side of the milling and grinding unit (B).
Notes on the suction tube:
The suction tube is supplied at a length of approx. 2.0 m (CEREC MC/MC X/MC XL Premium Package). When connecting the suction device to the unit, please ensure that no sharp bends occur over the full length of the suction tube.
Reduce the length of the tube according to your requirements and your installation location. Note that suction power drops along the length of the tube. You obtain decent suction power if the suction device is placed immediately below the machine and you have a tube length of 1.2 m or less.
5.4.7.4.2 Connecting the power cord
1. Insert the power cable into the relevant socket on the suction system.
2. Plug the other end into an appropriate power socket with a protective ground terminal.
Note on the power cable:
The suction system must only be operated with a power cable with a plug system designed for the relevant country.
Check the voltage specification on the rating plate. The system must conform to the country-specific supply voltage.
NOTICE
Grounded power outlet
The unit must be connected to a grounded power outlet.
63 98 650 D 3439
26 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.7.4.3 Connecting the interface cable (for automatic mode)
1. Plug the 15-pole connector into the socket (A) of the processing machine.
2. Plug the 9-pole connector into the socket (B) of the suction device.
5.4.7.4.4 Automatic mode
The interface cable is connected.
➢ Set the on/off switch to the position Auto.
Note on the automatic mode:
The processing machine monitors the correct socket (interface cable and suction tube) and the operation of the suction system during the running processes.
5.4.7.4.5 Setting the suction power
Use the control dial (A) to set the suction power.
➢ Recommendation: Set the suction power to the minimum (min). To
increase the service life of the filter bags, you can increase the suction power to the maximum suction power (max).
NOTICE
We recommended replacing the filter bag for the CEREC suction device after approx. 120 restoration units. This number may differ depending on the amount of zirconium oxide materials milled and suctioned.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
27
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.7.5
Connecting to the PC via WLAN (option)
Making the connection
Connect access point
➢ Connect the LAN port A of the milling and grinding unit and the
access point, using the network cable (10m, Order No.: 61 51 521). The access point is pre-configured at the factory for this
application.
Positioning the access point 1. As a test, place the access point near the milling and grinding unit at
head level or higher.
2. Perform a communication test as described in the separate instructions (see "Operating MC XL via WLAN in infrastructure mode", chapter "Final work, analyzing connection quality"). If applicable, follow the instructions on changing channels.
3. After you have found the optimum setting, take the acquisition unit and place it in the position in which it will be operated that is farthest away from the access point.
4. From this position, repeat the communication test you conducted earlier. If the results are satisfactory, leave the access point permanently in this position.
5. If the results are not satisfactory, position the access point outside of the room in which the milling and grinding unit is located and repeat the communication test.
If the connection quality is still not adequate, WLAN
communication cannot be easily achieved under the local conditions. In this case, ask your network administrator for assistance.
NOTICE
LAN connection
Operation via a cable LAN connection is possible at any time.
63 98 650 D 3439
28 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.7.6
Operating several milling and grinding units over one access point
To operate several MC/MC X or MC XL milling and grinding units over one access point, you need the following additional components:
● 1x LAN switch (e.g. Netgear ProSave 5 Port Gigabit Switch, Model GS105)
● 1x LAN network cable (10m, Sirona Order No.: 61 51 521).
LAN switch, connection example with 2 milling and grinding units
1. Connect the LAN port A of the milling and grinding units with the LAN switch using the included 10m LAN network cable.
2. Connect the access point with the LAN switch using the additional 10 m LAN network cable.
Now, all milling and grinding units connected to the LAN switch can
be operated via WLAN.
5.4.7.7
Connecting to the PC via the wireless interface (optional)
The CEREC acquisition unit has one of the following wireless modules installed:
● Höft & Wessel HW 8614/F2
– Installation kit: 62 79 694
– Retrofit kit: 62 79 702
or
● CEREC Radio Device
– Installation kit: 65 42 521
– Retrofit kit: 65 43 073
IMPORTANT
Acquisition units and milling and grinding units can only work with one wireless module system or the other. This means that any additional acquisition or milling and grinding units that are acquired by existing customers, which need to be connected wirelessly to existing systems, will require the existing systems to be retrofitted in order to work with the new CEREC Radio Device.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
29
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.7.7.1 Höft & Wessel HW 8614/F2
1. Connect the HW 8614/F2 wireless module to the LAN port of the milling and grinding unit using the LAN crossover cable (1 m).
2. Connect the plug-in power supply included in the scope of delivery with the wireless module HW 8614/F2 and plug it into the power supply.
3. If necessary, secure the wireless module in the selected operating position using the preassembled Velcro ® tape. In doing so, ensure that the rod antenna is vertically positioned.
4. Pair the wireless module of the milling and grinding unit as described in the installation instructions included with the wireless module (REF 62 80 064).
You can pair multiple milling and grinding units with a CEREC acquisition unit. If more than 2 milling and grinding units are operated at once, the limited data bandwidth may cause processing time delays.
5.4.7.7.2 CEREC Radio Device
Creating a network
The supplied network devices have not as yet been part of a network. To enable several networks from various operators to be created next to one another or in the same area, new network devices must first form a network. This is described in the following section.
To create a new network with several new network devices, perform the following steps:
1. Place all network devices in the same area.
2. Switch all network devices on within one minute.
3. After switching on the last network device, wait approx. one minute.
4. Make sure that all network devices are connected, as indicated by the green LED; see section “LED displays” in the “CEREC Radio Device” operating instructions (REF 65 45 177).
All network devices now belong to a single network, which can be operated as an independent network, and can communicate with one another. In the event of problems, see section “'Network creation failures” in the “CEREC Radio Device” operating instructions (REF 65 45 177).
Extending the network - adding a new network device
New network devices can be added to an existing network. To prevent random network devices from becoming part of the network, the user must perform the following steps:
1. Place the new network devices next to a network device that is part of the network.
2. Switch on the new network devices.
3. Switch the existing network device off and on again within one minute.
After one minute the new network devices will become part of the
network.
63 98 650 D 3439
30 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.4.8 Filling the water tank
5.4.8.1
Water tank MC / MC X
Water tank
The water tank is drained, see "Removing water from the unit".
1. Pull out the water tank at the front of the unit.
2. Turn the tank cap counter-clockwise and take it off.
IMPORTANT
A network with CEREC Radio Devices can include up to three network devices. Larger networks are not possible.
NOTICE
Using the tank cap opener
If you find the tank cap, tank drain or filter insert hard to open by hand, use the tank cap opener (see "Using the tank cap opener").
A Filter insert C Tank B Tank cap D Tank drain
A
B
C
D
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
31
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3. Add approx. 75 ml of DENTATEC to the tank.
4. Fill the tank with water until the filter insert is completely immersed (up to the bottom edge of the cover thread, approx. 3 liters).
5. Wait for a short time until the filter insert is completely soaked; then add an appropriate amount of water.
6. Close the water tank by tightening the tank cap clockwise by hand. Do not use the tank cap opener for this.
7. Push the water tank back into the housing.
8. Switch the unit on (see Switching the unit ON and OFF [ → 32]).
9. Switch the pump on (press the
"Pump"
button) to fill the water circuit.
10. Fill the water tank up again until the filter insert is completely immersed (up to the bottom edge of the cap thread).
5.4.8.2
Water tank CEREC MC XL Basic
Water tank
The water tank is drained, see "Removing water from the unit".
1. Pull out the water tank at the front of the unit.
2. Turn the tank cap counter-clockwise and take it off.
NOTICE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit. ➢ Do not spill DENTATEC.
A Filter insert C Tank B Tank cap D Tank drain
A
D
B
C
63 98 650 D 3439
32 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.4Commissioning Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
3. Add approx. 75 ml of DENTATEC to the tank.
4. Fill the tank with water until the filter insert is completely immersed (up to the bottom edge of the cover thread, approx. 3 liters).
5. Wait for a short time until the filter insert is completely soaked; then add an appropriate amount of water.
6. Close the water tank by tightening the tank cap clockwise by hand. Do not use the tank cap opener for this.
7. Push the water tank back into the housing.
8. Switch the unit on (see Switching the unit ON and OFF [ → 32]).
9. Switch the pump on (press the
"Pump"
button) to fill the water circuit.
10. Fill the water tank up again until the filter insert is completely immersed (up to the bottom edge of the cap thread).
5.4.9 Switching the unit ON and OFF
Switch on the unit
The milling and grinding unit is connected to the power supply.
1. The main switch on the rear side of the unit is set to position I (ON).
2. Press the ON/OFF button on the front panel.
The unit switches on and the display lights up.
NOTICE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit. ➢ Do not spill DENTATEC.
NOTICE
Do not put the unit into operation at low temperatures!
If you move the unit to the operating site from a cold environment, condensation may form and result in a short circuit.
Within the machine, grease depots are included for lubricating components that can cause error messages at low temperatures.
Install the unit at room temperature.
➢ Wait until the unit has reached room temperature and is absolutely
dry (for at least one hour)
The unit is dry and can be put into operation.
NOTICE
Do not adjust the line voltage
The unit automatically adjusts to the line voltage.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
33
Dentsply Sirona 5Installation and startup Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Power-up display
When the milling and grinding unit is switched on, the display shows a picture of the milling and grinding unit trying to contact the PC.
You can start or stop the water pump by pressing the
"Pump"
button (C). This enables you to drain the water circuit without connecting to the PC (e.g. prior to transport) or fill the water circuit during startup.
You can call up the IP address by pressing the
"Config"
button (D). You can configure the milling and grinding unit in the network with this address.
Switching the unit off
The unit has finished the machining operation.
➢ Briefly press the ON/OFF button on the front panel. When you let go of the button, the unit switches off.
5.5
Repacking
The water tank is empty. The main switch on the back side of the unit is set to the 0 (OFF)
position.
1. Disconnect the power cable and the connecting cable from the back
side of the unit and stow them away.
2. Stow away the calibration tools in the drawer.
3. Check the unit for completeness according to the scope of supply!
4. Pack the unit securely.
5.6
Scope of supply
The detailed scope of supply is specified in the document "Checklist CEREC MC/MC X".
NOTICE
Repack only drained units!
Drain the unit! See "Removing water from the unit“.
63 98 650 D 3439
34 D 3439.201.07.10.23 10.2017
5Installation and startup Dentsply Sirona
5.7Storage Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5.7
Storage
Store the unit in a closed and dry room at a temperature of -10°C to 50°C for a maximum period of 12 months.
NOTICE
Repack only drained units!
Drain the unit! See "Removing water from the unit“.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
35
Dentsply Sirona 6Operation Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6
Operation
6.1
Configuration (CEREC MC/MC X)
In the
"Devices"
area of the
"CEREC SW"
software, various settings can
be subsequently modified.
1. Click the
"Configuration"
button in the system menu.
2. Click on the
"Devices"
button.
3. Click on the unit that you wish to configure.
Bar code reader
If a bar code reader is used, e. g. for in inCoris ZI, the corresponding box must be activated. The bar code reader will then always be used to read a bar code.
Extraction Unit
If a suction device is connected, and the check mark is set, the dry milling process is automatically started when processing zirconium oxide. To deactivate dry milling, you can remove the check mark once again.
CAUTION
Risk of injury on calibration pins/instruments
If you reach into the processing chamber (e.g.: when inserting/removing a ceramic block, changing instruments, or inserting/removing a calibration phantom), you may injure your hand on the calibration pins/ instruments.
Be careful not to brush against the calibration pins/instruments with your hand.
Always insert your hand in the processing chamber underneath the calibration pins/instruments.
63 98 650 D 3439
36 D 3439.201.07.10.23 10.2017
6Operation Dentsply Sirona
6.2Calibrating the unit Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.2
Calibrating the unit
Unit calibrated ex works
The unit is calibrated at the factory. No additional calibration is required during initial startup. Proceed as described below when performing a subsequent calibration.
Preparing a calibration
1. Take the calibration pins and calibration phantom out of the drawer of the unit.
2. In the software, navigate to the system menu, and click on the
"Configuration"
button.
3. Click on the
"Devices"
button.
4. Click on the unit that you wish to calibrate.
5. Click on the step
"Calibrate"
.
The milling and grinding unit then moves into position to insert the
calibration tools. A dialog box prompts you to insert the calibration pins and the calibration phantom and to close the processing chamber door again.
6. Click on the
"Start"
button.
NOTICE
Use only the supplied calibration tools
Use only the supplied calibration pins and the corresponding calibration phantom when calibrating the unit.
NOTICE
Incorrect milling or grinding result
If the unit is not calibrated, the milling or grinding result may be faulty.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
37
Dentsply Sirona 6Operation Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Inserting the calibration pins and phantom
1. Press the catch of the processing chamber door and open the door.
2. Loosen the instruments with the torque wrench and remove them.
Calibration phantom
3. To insert the calibration phantom into the block fixing, grasp it by its narrow surfaces B.
4. Clamp the calibration phantom with the ball pressure screw. Use the block clamp tool for this purpose.
5. Insert the calibration pins in the motor mount by hand. Tighten the corresponding chuck with the torque wrench until a clicking sound can be heard.
6. Close the processing chamber door.
Performing a calibration
➢ Confirm your selection in the
"Calibrate milling unit"
window with the
"Start"
button.
The automatic calibration begins and takes approx. 14 minutes.
Wait until the calibration has been completed.
NOTICE
Grasp the calibration phantom correctly
Grasping the calibration phantom by its wide surfaces may cause calibration errors.
➢ Always grasp the calibration phantom by its clamping shank (A)
when removing it from the storage box.
➢ Always grasp the calibration phantom by its narrow surfaces B when
inserting it into the block fixing.
A
B
B
63 98 650 D 3439
38 D 3439.201.07.10.23 10.2017
6Operation Dentsply Sirona
6.2Calibrating the unit Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Inserting instruments
1. Open the processing chamber door following calibration.
2. Loosen the calibration pins with the torque wrench and remove them.
3. Loosen the ball pressure screw.
4. Remove the calibration phantom by grasping it by its narrow surfaces (B).
5. Insert the instruments in the motor mount by hand. Tighten the corresponding chuck with the torque wrench until a clicking sound can be heard.
6. Close the processing chamber door. The dialog box for selecting the instruments then appears.
7. Select the inserted instruments and confirm by clicking the
"Start"
button in the window. The motor mounts move to their starting positions.
The
"Calibration succeeded"
dialog box appears.
Exiting the calibration
1. Click on the
"OK"
button.
2. Click on the step
"Exit Configuration"
.
NOTICE
Store the calibration tools in a safe place
Store the calibration pins and the calibration body in a safe place (e.g. in a storage box in the unit drawer).
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
39
Dentsply Sirona 6Operation Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.3
Replacing the set screw
1. If a block is inserted in the block fixing, remove it.
2. In the software, navigate to the system menu, and click on the
"Configuration"
button.
3. Click on the
"Devices"
button.
4. Click on the unit whose set screw you wish to replace.
5. Click on the step
"Change set screw"
.
The milling and grinding unit moves into position (12 o'clock
position) to insert the set screw.
A dialog box prompts you to replace the set screw and close the
processing chamber door again.
6. Remove the worn set screw using the block clamp tool.
7. Take a new set screw out of the drawer of the unit, insert it, and secure it in place using the block clamp tool.
8. Click on the
"Ok"
button.
NOTICE
Not to be confused
Do not confuse the set screw with the ball pressure screw!
NOTICE
Wear to the set screw
Replace the set screw every 1000 clamping procedures or if the pin has broken off.
63 98 650 D 3439
40 D 3439.201.07.10.23 10.2017
6Operation Dentsply Sirona
6.4Machining process Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.4
Machining process
6.4.1 Process types
Various process types are available for milling and grinding purposes. These vary in terms of the type of materials to be processed and the instruments to be used.
6.4.1.1
Grinding
For grinding purposes, use the following instruments as well as the appropriate torque wrench:
Instrument REF Torque wrench Clamping format of the
force transmission
Step Bur 12 S 6240167 Triangular
Step Bur 12 6260025
Cyl. Pointed Bur 12 S 6240159
Step Bur 20 6259597
Cyl. Pointed Bur 20 6259589
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
41
Dentsply Sirona 6Operation Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.4.1.2
Milling
The milling option is available from the following serial numbers onwards:
Other machines must be equipped with the milling starter kit for closed motors (REF: 64 51 079) (only for wet milling).
Activating the milling option
1. Select
"Configuration" /"Settings" /"Milling"
.
2. Check
"Activate"
.
Instruments and torque wrenches
For milling purposes, use the following instruments as well as the appropriate torque wrenches:
Machine type Serial number (wet milling) Serial number (dry milling also with
retrofit) MC X 231001 236001 MC 202001 202501
IMPORTANT
The milling process is supported in the CEREC software from version
4.3 onwards.
Instrument REF Torque wrench Clamping format of the
force transmission
Finisher 10 (dry and wet) 6299387 Triangular
Shaper 25: Milling (dry) 6299395 Triangular
Shaper 25 RZ: Milling (wet) 6433440 Square
63 98 650 D 3439
42 D 3439.201.07.10.23 10.2017
6Operation Dentsply Sirona
6.4Machining process Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.4.1.3
Permitted instrument combinations
Depending on the materials to be processed and the process type used, various instrument combinations are permitted.
For an updated table of approved instrument combinations, please visit the
"www.sirona.com"
. To access the online portal for technical
documentation, follow the
"SERVICE"
/
"Technical Documentation"
menu items in the navigation bar. The portal can also be accessed directly via the following address http://www.sirona.com/manuals. Select your product there and then open the
"Bur table"
document.
6.4.2 Preparations
Download or design a restoration (see operator's manual).When the
"Milling"
option is activated, in the
"Select Material"
material selection step, you can choose between the
"Grinding"
and
"Milling"
manufacturing processes for plastic and zirconium oxide
materials.
You are in the
"MANUFACTURE"
phase and have selected the milling and grinding unit, tested the settings, and positioned the restoration in the block.
➢ Click on the
"Start Milling"
step.
The milling and grinding unit moves into position to insert the set
screw.
6.4.3 Starting the machining processes
The instrument sets are equipped with the required instrument
combinations for the machining process.
1. Depending on the configuration, you will be prompted to enter the bar code (see also "Entering the bar code").
2. Press the catch of the processing chamber door and open the door.
3. Place the selected ceramic block in the block fixing.
4. Clamp the ceramic block with the ball pressure screw. Use the block clamp tool for this purpose.
5. Close the processing chamber door and confirm the procedure by clicking
"Start"
.
The estimated time required for the machining process will then
appear in a message window.
NOTICE
Error message during touch process!
Always be sure to insert the ceramic block that you selected for the restoration. Otherwise an error message will be displayed during the touch process.
NOTICE
Ending the machining processes
You can abort the machining process at any time by pressing the
"Stop"
button.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
43
Dentsply Sirona 6Operation Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.4.4 Ending the machining processes
1. When the machining process has been completed, open the processing chamber door.
2. Remove the restoration.
3. Loosen the ball pressure screw.
4. Remove the remainder of the ceramic block.
5. Close the processing chamber door.
6.4.5 Information on the seal of approval
Proper selection and processing of the material are decisive for the long­term clinical success of the restoration, especially in the case of zirconia. However, different types of zirconia require individually matched machine parameters. This is the reason why you can and must select different types of zirconia in the software. These machine parameters are coordinated between material partners in complex development processes. In addition to the desired fit and surface quality, they also guarantee a maximum degree of material and equipment safety. Consistently high quality of the grinding or milling result and the fit can be guaranteed and damage to the processing machines can be excluded only if certified materials are used.
Fragment: Everything excluding FONA
WARNING
Risk of injury on the remainder of the ceramic block
The remaining portion of the ceramic block may have sharp edges (e.g. A) that could injury you if it is not removed carefully.
Always grasp the remainder of the ceramic block by its metal holder.
CAUTION
Do not use defective milling or grinding results!
Machining results must be judged by the user (dentist or dental technician) and must not be used if defects are detected!
NOTICE
If you have not used the milling and grinding unit for a rather long time, we recommend you should switch it off and then open the processing chamber door so that the processing chamber can dry out.
NOTICE
Block without seal of approval
If a block is found without a seal of approval during the machining process (milling or grinding), the following message appears: "No quality label was recognized on the block. The grinding and milling processes as well as the instruments are specially verified for certified materials. Certified materials can be identified by the engraved "inLab" lettering on the block. The use of zirconium oxide materials without quality label can lead to inferior results as well as increased wear on the device and instruments. Do you still wish to start the manufacturing process?"
63 98 650 D 3439
44 D 3439.201.07.10.23 10.2017
6Operation Dentsply Sirona
6.5Entering the bar code Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.5
Entering the bar code
Barcode Reader active
If you have activated the option
"Barcode Reader"
in the system configuration dialog (e.g. for CEREC Zirconia), you must read-in both bar codes with the bar code reader. To do this, hold the bar code reader tilted to a slight angle and move it over both of the bar codes on the block continuously and evenly.
If the reading process fails, you can read-in the bar code once again by pressing
"Retry"
(button 1 on the unit display). Alternatively, you also can enter the substitute code (8-digit character string, e.g. *1234XYZ) on the PC manually.
No bar code reader available
➢ Enter the substitute code (8-digit character string, e.g. *1234XYZ) on
the PC manually.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
45
Dentsply Sirona 6Operation Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6.6
Manual block clamp
1. Insert the block (A).
2. Clamp the ceramic block securely with the ball pressure screw (B). Use the block clamp tool for this purpose.
The block is pressed laterally against the contact surface of the
block fixing and simultaneously pulled in axially. The plate of the block holder thus rests on the block fixing.
NOTICE
Wear of the ball pressure screw
The high clamping forces cause wear of the ball pressure screw.
➢ Replace the ball pressure screw every 500 clamping operations.
NOTICE
Fasten the block tightly
If the block is not tightened sufficiently, this may result in falsification of the machining result and fracturing of its ceramic material.
➢ Tighten the block securely with the block clamp tool. ➢ Check to make sure that the block is seated very firmly.
63 98 650 D 3439
46 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7
Service
NOTICE
Observe country-specific Regulations!
Some countries have legal regulations which require regular safety inspections of electrical devices or systems by the operator.
NOTICE
Perform maintenance regularly!
Have maintenance performed on your unit annually by trained technical personnel / a service engineer.
NOTICE
Observe error messages
You must observe error messages shown on the display on in the software. If the error message does not disappear even after you have performed the prompted action, contact your service engineer.
NOTICE
Machine care
Interval: Once a month
➢ Change the filter (see Changing the filter) ➢ Clean the clamping cones of the instruments according to the
cleaning set instructions (REF 61 77 161).
➢ If the jets of water do not strike the instruments, carefully remove
any foreign particles from the water nozzles with a probe.
NOTICE
Processing chamber wet cleaning process
Interval (if dry milling is predominantly used): Once a week or in the case of heavy soiling.
➢ Clean the processing chamber.
NOTICE
Do not confuse the block screw with the ball pressure screw
When operating a CEREC 3 grinding unit and a CEREC MC/MC X in the same room, be careful not to confuse the block screw of the CEREC 3 with the ball pressure screw of the CEREC MC/MC X.
NOTICE
Using the tank cap opener
If you find the tank cap, tank drain or filter insert hard to open by hand, use the tank cap opener (see "Using the tank cap opener").
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
47
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.1
Using the cleaning hose and the wet cleaning process
The processing chambers of devices used for dry milling of zirconium oxide should be cleaned regularly in order to prevent deposits of zirconium oxide dust in the chamber. A second cleaning hose with a nozzle can be used before the wet cleaning process in order to support the user when sucking zirconium oxide dust out of the device's processing chamber. This hose is available separately as a spare part (REF 65 89 795). We recommend carrying out this cleaning procedure (or the wet milling of a restoration) at least once per week if the device is used for dry milling of zirconium oxide.
Proceed as follows to use the cleaning hose and to execute the recommended wet cleaning process:
1. Remove the normal hose which is connected to the milling/grinding unit.
2. Connect the cleaning hose to the upper opening of the suction device.
3. Activate the manual operation for the suction device by setting the switch at the top of the device from "Auto" to "On".
4. Suck up the dust from the chamber as required.
NOTICE
Wear of the ball pressure screw
The high clamping forces cause wear of the ball pressure screw.
➢ Replace the ball pressure screw every 500 clamping operations.
63 98 650 D 3439
48 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.2Changing filter bags and HEPA filters Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
5. Sucking up as much dust as possible from the workpiece axis is recommended. You must also always remove the dust from the bottom of the impact pane of the block axis (see arrow).
6. Once the chamber has been cleaned using the suction device you should complete the wet cleaning process lasting 30 seconds as programmed in the software 2 to 3 times as required in order to remove any accumulated material. This function is in the
"Device
configuration …"
area of the CEREC SW / CEREC Premium SW
software if you click on the
"Service"
button and then on
"Start a
cleaning process."
. Alternatively, you can also wet grind a normal
restoration with the same effect.
7. You can remove any zirconium oxide deposits in the lower part of the workpiece axis with a nylon brush between the wet cleaning processes.
7.2
Changing filter bags and HEPA filters
Changing filter bags
If there is a significant fall in suction power, the filter bag could be full and will need replacing.
1. Loosen the two knurled nuts on the maintenance cover.
2. Remove the lid.
3. Remove the filter bag from the nozzle and put a new filter bag on.
4. Put the maintenance cover on and screw it down with the two knurled nuts.
NOTICE
We recommended replacing the filter bag for the CEREC suction device after approx. 120 restoration units. This number may differ depending on the amount of zirconium oxide materials milled and suctioned.
NOTICE
Do not jam the filter bag
Make sure that the cap is sealed properly and the filter bag is not jammed.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
49
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Changing HEPA filters
Replace the HEPA filter if suction power is still weak after replacing the filter bag. The HEPA filter is located behind the filter bag.
1. Loosen the two knurled nuts on the maintenance cover.
2. Remove the lid.
3. Remove the filter bag from the nozzle.
4. Then release and remove the two Phillips screws on the perforated sheet on the inside of the suction system.
5. Take the perforated sheet out.
NOTICE
We recommend replacing the HEPA filter for the CEREC suction device once the filter bag has been replaced around 3 to 4 times. This number may differ depending on the amount of zirconium oxide materials milled and suctioned.
63 98 650 D 3439
50 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.3Changing the water Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
6. Remove the dusty HEPA filter and insert a new HEPA filter.
7. Put the perforated sheet back on and screw it down with the two Phillips screws.
8. Put the filter bag back on again.
9. Put the maintenance cover on and screw it down with the two knurled nuts.
7.3
Changing the water
7.3.1 General information
When the water is due to be changed, a message window appears on your monitor to remind you that it is time to change the water.
Preventing odors
All coolant additives contain a biodegradable preservative. Despite this, however, odors may still develop under unfavorable conditions.
Observe the following:
● Change the water at least once a week.
● With ambient temperatures above 25°C, change the water every 2 to 3 days to prevent foul odors.
● Drain the tank if you do not intend to operate the unit for more than one week.
● Clean the tank if the odors recur.
● Add the coolant additive DENTATEC and fill the tank up to the brim with water. Let it stand for at least 24 hours and then rinse it out thoroughly with water once again.
NOTICE
Do not jam the filter bag
Make sure that the cap is sealed properly and the filter bag is not jammed.
NOTICE
Damage to the pump and drives!
Too high a proportion of ceramic in the cooling water will damage the pump and the drives.
Change the water regularly!
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
51
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.3.2 Changing the water
7.3.2.1
Water tank MC / MC X
Water tank
NOTICE
Damage to surfaces!
In the undiluted state, the coolant additive DENTATEC disintegrates plastic surfaces and can cause discoloration.
➢ Do not place DENTATEC on the unit. ➢ Do not spill DENTATEC.
NOTICE
Approved coolant additive
Use only DENTATEC as a coolant additive.
NOTICE
Change the filter insert regularly!
Replace the filter insert every third water change or after around 75 restorations.
A Filter insert C Tank B Tank cap D Tank drain
A
B
C
D
63 98 650 D 3439
52 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.3Changing the water Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
To change the water, proceed as follows:
The unit is switched on.No machining process is running.
1. Pull out the water tank at the front of the unit.
2. Open the drain opening.
3. Drain the water tank.
4. Turn the tank cap counter-clockwise and take it off. If you find the tank cap hard to open by hand, use the tank cap opener (see "Opening the tank cap").
5. Remove the filter insert from the tank and clean the filter thoroughly under running water.
6. Rinse the water tank.
7. Insert the cleaned filter with handle into the tank and press it firmly onto the base in the floor of the tank.
8. Close the drain opening.
9. Add approx. 75 ml of DENTATEC to the tank.
10. Fill the tank with water until the filter insert is completely immersed (up to the bottom edge of the cover thread, approx. 3 liters).
11. Wait for a short time until the filter insert is completely soaked; then add an appropriate amount of water.
12. Close the water tank by tightening the tank cap clockwise by hand. Do not use the tank cap opener for this.
13. Push the water tank back into the housing.
NOTICE
Foaming not permissible!
If any cleaning agents are used, this will create foam, which is not permitted.
Do not use any cleaning agents.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
53
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.3.2.2
Water tank CEREC MC XL Basic
Water tank
To change the water, proceed as follows:
The unit is switched on.No machining process is running.
1. Pull out the water tank at the front of the unit.
2. Open the drain opening.
3. Drain the water tank.
4. Turn the tank cap counter-clockwise and take it off. If you find the tank cap hard to open by hand, use the tank cap opener (see "Opening the tank cap").
5. Unscrew the side cap.
6. Remove the filter insert from the tank and clean the filter thoroughly under running water.
7. Rinse the water tank.
8. Insert the cleaned filter with its cap into the unit and screw it tight.
9. Close the drain opening.
10. Add approx. 75 ml of DENTATEC to the tank.
A Filter insert C Tank B Tank cap D Tank drain
NOTICE
Foaming not permissible!
If any cleaning agents are used, this will create foam, which is not permitted.
Do not use any cleaning agents.
A
D
B
C
63 98 650 D 3439
54 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.4Instruments Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
11. Fill the tank with water until the filter insert is completely immersed (up to the bottom edge of the cover thread, approx. 3 liters).
12. Wait for a short time until the filter insert is completely soaked; then add an appropriate amount of water.
13. Close the water tank by tightening the tank cap clockwise by hand. Do not use the tank cap opener for this.
14. Push the water tank back into the housing.
7.4
Instruments
7.4.1 Overview of materials/instruments
7.4.1.1
CEREC MC
For an updated table of approved instrument combinations, please visit the
"www.sirona.com"
. To access the online portal for technical
documentation, follow the
"SERVICE"
/
"Technical Documentation"
menu items in the navigation bar. The portal can also be accessed directly via the following address http://www.sirona.com/manuals. Select your product there and then open the
"Bur table"
document.
7.4.1.2
CEREC MC X
For an updated table of approved instrument combinations, please visit the
"www.sirona.com"
. To access the online portal for technical
documentation, follow the
"SERVICE"
/
"Technical Documentation"
menu items in the navigation bar. The portal can also be accessed directly via the following address http://www.sirona.com/manuals. Select your product there and then open the
"Bur table"
document.
7.4.2 Changing instruments
The torque wrench from the draw of the milling and grinding unit is
ready-to-hand.
1. In the software, navigate to the system menu, and click on the
"Configuration"
button.
2. Click on the
"Devices"
button.
3. Click on the unit whose instruments you wish to replace.
NOTICE
Regular replacement of the instruments
Change the instruments as soon as the system prompts you to do so.
- Milling instruments (wet grinding): Typically after no more than around 25 restorations. The type of material, block size and processing options may affect the service life.
- Shaper 25 (dry milling): After 50 restorations.
- Shaper 25 RZ (wet milling): After 30 restorations.
- Finisher 10 (dry and wet milling): After 50 restorations.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
55
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
4. Click on the
"Start"
button.
The motors travel to the change position for the instruments.
The dialog box for changing the instruments opens.
5. Press the catch of the processing chamber door and open the door.
6. Loosen the worn-out or defective instrument with the torque wrench and pull it out manually.
7. NOTICE! Do not grease the instrument! Insert the new instrument into the motor mount by hand. Tighten the corresponding chuck with the torque wrench until a clicking sound can be heard.
8. Close the processing chamber door.
9. Select the instrument(s) you have inserted on the PC monitor and click
"Start"
. You can also select the instruments on the milling and grinding unit (up/down arrow) and confirm with
"Start"
.
CAUTION
Risk of injury on instruments
If you put your hand in the processing chamber, you could injure it on the instruments.
Be careful not to brush against the instruments with your hand.
Apply the torque wrench as shown.
NOTICE
Incorrect milling or grinding results
Interchanging instruments leads to faulty milling or grinding results.
NOTICE
Cleaning cooling water nozzles
The cooling water nozzles in the processing chamber always must be kept free of lime and processing dust deposits. The corresponding cooling water jet must always strike the instrument accurately!
The cooling water nozzles are dirty.
➢ Clean the nozzles with a cleaning wire and a syringe.
NOTICE
Use only suitable instruments!
Do not use any instruments from CEREC 2 or CEREC 3 units.
63 98 650 D 3439
56 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.5Care, cleaning agents, and disinfectants Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Changing a defective instrument
If an instrument breaks during the machining phase, the corresponding motor travels to the change position. A dialog box which marks the side with the broken instrument with a red cross then opens.
The instrument is broken.
1. Change the defective instrument as described above.
2. Select the instrument which you have inserted.
3. Press the
"Start"
button.
7.5
Care, cleaning agents, and disinfectants
A continuously updated list of approved media can be downloaded from the internet via the online portal for technical documents. You can reach this portal at the address: www.dentsplysirona.com/manuals . Click on the menu item
"General documents"
and then open the
"Care,
cleaning and disinfection agents"
document
If you do not have internet access, please contact your dental depot to order the list (REF 59 70 905).
7.6
Cleaning surfaces
7.6.1 Disinfecting
Wipe surfaces down with a surface disinfectant (wiping disinfectant).
Observe the manufacturer’s instructions regarding restrictions for use.
7.6.2 Protection against medicaments
Due to their high concentrations and the substances they contain, many medicaments can dissolve, etch, bleach or discolor surfaces.
7.6.3 Cleaning
Remove dirt, grime and disinfectant residue regularly using mild, commercially available cleaning agents.
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents approved by Dentsply Sirona!
NOTICE
Do not allow liquids to run into the ventilation slots!
NOTICE
Damage to the surface
Clean the surface immediately with a moist cloth and a cleaning agent.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
57
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.7
Replacing the main fuses
Fuse holder
The power plug must be disconnected.
1. Use a screwdriver to carefully pry off the cover of the fuses on the back side of the unit.
2. Pull out the fuse holder.
3. Replace the defective fuses.
4. Reinsert the fuse holder.
5. Close the cover.
WARNING
Electric shock
Disconnect the power plug at the unit end before replacing the fuses.
NOTICE
Fuse type
Use only fuses of the same type in the fuse holder!
A Cover C Fuse holder BCC DFuse
Fuses: T5H250V Order No. 20 33,111
63 98 650 D 3439
58 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.8Changing the filter Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.8
Changing the filter
7.8.1 Water tank MC / MC X
Water tank
The tank is drained, see "Removing water from the unit".
1. Pull out the water tank at the front of the unit.
2. Turn the tank cap counter-clockwise and take it off. If you find the tank cap hard to open by hand, use the tank cap opener.
3. Take the filter insert out of the tank.
NOTICE
Change the filter regularly!
Clean the filter regularly and change it immediately when damaged. Otherwise, change it every 3 months.
If a message appears stating that the water pressure is too low, you must clean the filter or, if it is damaged, replace it immediately.
CAUTION
Filter
Use only filters approved by Dentsply Sirona!
A Filter insert C Tank B Tank cap D Tank drain
A
B
C
D
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
59
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
4. Rinse the water tank.
5. Insert a new filter with handle into the tank and press it firmly onto the base in the floor of the tank.
6. Fill the tank, see "Changing the water“ [ → 50].
7. Close the water tank by tightening the tank cap clockwise by hand. Do not use the tank cap opener for this.
8. Push the water tank back into the housing.
7.8.2 Water tank CEREC MC XL Basic
Water tank
The tank is drained, see "Removing water from the unit".
1. Pull out the water tank at the front of the unit.
2. Unscrew the cover on the side and take it out of the tank along with the filter insert.
3. Rinse the water tank.
4. Insert a new filter with cover into the tank and screw it tight.
5. Fill the tank, see "Changing the water“ [ → 50].
NOTICE
Cleaning the filter
Clean the filter approx. every 12 to 15 units under running water, but at least with every water change.
A Filter insert C Tank B Tank cap D Tank drain
Filter insert: Order No. 61 29 519
A
D
B
C
63 98 650 D 3439
60 D 3439.201.07.10.23 10.2017
7Service Dentsply Sirona
7.9Removing water from the unit Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.9
Removing water from the unit
You must remove the water from the unit if you will not be using it for a longer period of time or wish to transport it.
No machining process is running.
1. Turn the device off.
2. Pull out the water tank at the front of the unit.
3. Drain the water out of the water tank through the drain opening and reinsert the water tank in the unit.
4. Switch the unit on.
5. Press the
"Pump"
key to switch the pump on.
The water pump then starts pumping the water out of the unit.
Let the pump run until no more water escapes from the nozzles.
6. Press the
"Pump"
key to switch the pump off.
7. Pull out the water tank and empty it.
8. Push it back into the housing.
NOTICE
Cleaning the filter
Clean the filter approx. every 12 to 15 units under running water, but at least with every water change.
NOTICE
Pump button active at power-up
The
"Pump"
button appears on the display when the milling and grinding unit is switched on. You can start or stop the water pump by pressing this button.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
61
Dentsply Sirona 7Service Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
7.10
Using the tank cap opener
Opening the tank cap
The water tank has been pulled out and drained.
➢ Place the tank cap opener on the tank cap as shown, and take off the
tank cap by unscrewing it counter-clockwise.
Opening the tank drain
The water tank has been pulled out.
➢ Place the tank cap opener on the filter drain as shown, and take off
the filter drain by unscrewing it counter-clockwise.
NOTICE
Risk of damage to the tank
Use the tank cap opener only for opening the tank cap and tank drain.
Do not use the tank cap opener for closing the tank cap. To close the tank cap and tank drain, it is sufficient to tighten them clockwise by hand.
63 98 650 D 3439
62 D 3439.201.07.10.23 10.2017
8Technical description Dentsply Sirona
8.1System requirements Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
8
Technical description
8.1
System requirements
8.1.1 CEREC MC
The CEREC SW software must only be installed on CEREC acquisition units or on the 64-bit inLab PC (optional).
For all software generations 4.x and higher, CEREC AC Bluecam PC hardware version must be LN or higher (Windows 7, 64 bit). No hardware limitations for CEREC AC / AF / AI with Omnicam.
If necessary, upgrade your operating system.
The software version must be CEREC SW 4.2.0 or higher.
8.1.2 CEREC MC X
The CEREC SW / CEREC Premium SW / CEREC Premium CAM SW software must only be installed on CEREC acquisition units or on the 64­bit inLab PC (optional).
The software version must be either CEREC SW 4.2.0 or inLab SW 4.2.0 or higher.
For all software generations 4.x and higher, CEREC AC Bluecam PC hardware version must be LN or higher (Windows 7, 64 bit). No hardware limitations for CEREC AC / AF / AI with Omnicam.
8.2
Grinding and milling unit
8.2.1 General technical description
Grinding instruments
Milling instruments
● Digital feed control with force monitoring for extremely sensitive processing
● Process-controlled tool drives
● Processing repeatability: +/- 25 μm
● Step Bur 12 S
● Cyl. Pointed Bur 12 S
● Step Bur 12
● Step Bur 20
● Cyl. Pointed Bur 20
● Shaper 25 RZ (wet milling)
● Shaper 25 (dry milling)
● Finisher 10 (wet and dry milling)
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
63
Dentsply Sirona 8Technical description Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
8.2.2 Technical data
8.2.3 Controller board
● 3x 2-axis stepping motor controller with microstepping
● 2 DC motor controllers with integrated speed and current control and force monitoring
● Ethernet, RJ45 interface 10 Mbit/sec
Type designation Milling and grinding unit CEREC
MC / MC X Rated line voltage 100 V - 230 V AC Rated power frequency 50/60 Hz Nominal current 1.5 - 3.5 A Nominal power output 320VA Permissible line voltage fluctuations ±10% of nominal voltage Type of protection against electric
shock
Class 1 equipment
Degree of protection against ingress of water
Ordinary device (without
protection against ingress of
water) Overvoltage category II Ambient conditions For indoor use
Pollution degree 2
Air pressure: 700 hPa – 1060
hPa Temperature range 5°C to 40°C Humidity range 80% rel. up to 31 °C
decreasing to
50% rel. up to 40 °C Operating mode Continuous operation Dimensions (WxHxD) in mm 700 x 425 x 420 Approx. weight 43 kg
63 98 650 D 3439
64 D 3439.201.07.10.23 10.2017
9Consumable Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
9
Consumable
Grinding instruments
Milling instruments**
Left-hand side REF Right-hand side Step Bur 12 S (6x) 62 40 167 62 40 159 Cyl. Pointed Bur 12 S (6x)
Step Bur 12 (6x) 62 60 025
Cylinder Bur 12 EF (6x)* 65 35 186 65 35 178 Cylinder Pointed Bur 12 EF (6x)*
* Only with 4-motor milling/grinding units. Step Bur 20 (6x) 62 59 597 62 59 589 Cyl. Pointed Bur 20 (6x)
Left-hand side REF Right-hand side Shaper 25 (3x) – dry milling*** 62 99 395 62 99 387 Finisher 10 (3x) – dry and wet milling
Shaper 25 RZ (3x) – wet milling 64 33 440
* Milling possible: only with milling/grinding machines from the following serial numbers: inLab MC XL 129001, CEREC MC XL 129001, CEREC MC XL Premium Package 302001, CEREC MC X 231001 or replacement of the left motor on 2-motor machines or of the left motor in the motor set 1 on 4-motor machines.
*** Requires CEREC milling/grinding unit with dry milling configuration and CEREC suction.
REF Description Illustration 58 09 640 DENTATEC 1000 ml
63 87 067 Filters MC/MC X (1x) 64 29 950 Filters MC/MCX (pack of 6)
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
65
Dentsply Sirona 9Consumable Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
61 29 519 Filter unit (1x) 61 29 402 Filter unit (pack of 6)
65 78 095 Filter bag (double pack), CEREC
63 85 277 HEPA filter
65 89 795 Cleaning hose
64 79 849 Torque wrench, spare
64 79 856 Replacement torque wrench (for Shaper 25 RZ only)
62 58 987 Ball pressure screw set, spare (for manual block fixing)
62 35 126 Setscrew (for MC and MC X only)
63 5 614 Cleaning set (for manual block fixing)
62 99 403 Processing chamber screen, spare
REF Description Illustration
63 98 650 D 3439
66 D 3439.201.07.10.23 10.2017
10Disposal Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
10
Disposal
In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU). These regulations require the environmentally friendly recycling/disposal of old electrical and electronic devices. Such items must not be disposed of as domestic refuse. This has been expressed using the icon of the “crossed out trash can”.
Disposal procedure
We feel responsible for our products from the first idea to their disposal. For this reason, we give you an option to return our old electronic and electrical devices.
If you wish to dispose of your devices, please proceed as follows:
In Germany
To initiate return of the electrical device, please send a disposal request to enretec GmbH. You have the following options here:
● Use the “Returning an electrical device” button under the “eom” menu item on the enretec GmbH homepage (www.enretec.de).
● Alternatively, you can also contact enretec GmbH directly.
enretec GmbH Kanalstraße 17 16727 Velten
Tel.: +49 3304 3919-500 E-mail: eom@enretec.de
In accordance with the national disposal regulations regarding old electrical and electronic devices (ElektroG), as the manufacturer, we assume the costs for disposing of the electrical and electronic devices in question. Disassembly, transport and packaging costs shall be borne by the owner/operator.
Prior to disassembly/disposal of the product, it must be fully prepared (cleaned/disinfected/sterilized).
If your unit is not permanently installed, it will be collected from the practice. If it is permanently installed, it will be picked up curbside at your address by appointment.
Other countries
For country-specific information on disposal, contact your local dental dealers.
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
67
Dentsply Sirona Index Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
Index
B
Bar code reader, 18
Building installation, 13
C
Calibration tools
Calibration phantom, 36
Calibration pins, 36
Storage, 38
Care, cleaning, and disinfecting agents, 56
CE mark, 10
Changing, 62
Changing a defective instrument, 56
Connection
Ethernet, 23
LAN, 23
WLAN, 27
Connection for suction, 18
Connections, 18
Cooling water nozzles, 55
Customer service center, 7
D
Dimensions, 63
Disinfectant, 56
Disposal of old electrical and electronic devices, 66
E
enretec GmbH, 66
Ethernet
LAN port, 18
F
Filter
mode, 58
Order No., 59
footprint, 16
Fuse, 18
Fuse type, 57
Order No., 57
replacement, 57
H
Humidity range, 63
I
Installation site, 16
installing unit
manually, 24
Instruments, 55
Changing, 54
M
Main switch, 18
Maintenance, 13
Regulations, 46
Manufacturer's address, 7
Milling and grinding unit
Display, 21
Machining room, 18
Overview, 17
N
Nominal current, 63
O
Operating mode, 63
63 98 650 D 3439
68 D 3439.201.07.10.23 10.2017
Index Dentsply Sirona
Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
P
Packaging, 15
Packing, 33
Power connection, 18
Product safety, 14
Protection class, 63
R
Rated line voltage, 63
Repair, 13
S
Safety instructions, 8
Scope of supply, 33
T
Temperature range, 63
Transport, 15
Type designation, 63
U
Unit installation
automatically, 23
Unit removal
removal, 24
Unpacking, 15
W
Water, 63
Water tank
Changing the water, 52, 53
Filling, 30, 31
Odors, 50
Overview, 30, 31, 51, 53, 58, 59
Removing water from the unit, 60
water change, 50
Weight, 63
63 98 650 D 3439 D 3439.201.07.10.23 10.2017
69
Dentsply Sirona Index Operating Instructions (valid for USA) Sirona Dental CAD/CAM System
We reserve the right to make any alterations which may be required due to technical improvements.
© Sirona Dental Systems GmbH Sprache: englisch (US) Printed in Germany D 3439.201.07.10.23 10.2017 Ä.-Nr.: 124 428
Sirona Dental Systems GmbH Contact in the USA:
Fabrikstraße 31 64625 Bensheim Germany
Sirona Dental, Inc. 4835 Sirona Drive Charlotte, NC 28273 USA
Order No
63 98 650 D 3439
www.dentsplysirona.com