Proper shielding of room and operator
position is essential.
Since these requirements vary from state
to state it is the assembler's/installer's
responsibility that all local radiation
safety requirements are met.
61 88 085 D3352
2D3352.103.02.01.02
Schick Technologies, Inc.
Maintenance Instructions CDRPanElite
9Checking the X-Ray Beam for panorama exposure......................................................22
9.1Preparing the Pan adjustment........................................................ ........ 22
9.2Checking the Pan sensor adjustment................................................ .. ... 23
9.3Checking the Pan diaphragm adjustment................................. ... ........... 25
9.4Checking the Pan filter adjustment.................................................... ..... 27
9.5Checking the Pan symmetry................. .. ................................................ 29
10Checking the X-Ray Beam for cephalometer exposure................................................ 31
10.1Preparing the Ceph adjustment............................................. ... ... ........... 31
10.2Checking the Ceph primary diaphragm adjustment................................ 32
10.3Checking the Ceph rotation fixpoint adjustment..................................... 35
10.4Checking the Ceph main X-ray beam direction adjustment.................... 37
11Checking the alignment of the ear plugs....................................................................... 39
11.1Preparing the ear plugs ......................................................... ................. 39
11.2Checking the ear plug alignment........................................ .................... 40
12Checking the laser............................................................................................................ 43
English
61 88 085 D3352
D3352.103.02.01.02
3
Schick Technologies, Inc.
Maintenance Instructions CDRPanElite
61 88 085 D3352
4D3352.103.02.01.02
Schick Technologies, Inc.1 General
ATTENTION
ATTENTION
ATTENTION
WARNING
ON
OFF
Maintenance Instructions CDR PanElite
1 General
To stay in compliance with the DHHS requirements the
CDR PanElite must be maintained annually.
It is the responsibility of the user to insure that the equipment is maintained with the manufacturer's recommended
Maintenance Instructions to insure compliance with the
Federal Performance Standard.
The manufacturer and the assembler/installer are relieved
from responsibility in those cases where the user fails to
have the manufacturer's recommended maintenance performed.
The actual maintenance inspection and consequent service must be accomplished by a trained serviceman. Neither the inspection nor service is part of the equipment warranty.
Technical instructions required
Operating Instructions
Service Manual
Instruments and adjustment tools required
All measuring devices must be suitable for measurements
on the mains potential.
This includes the pulse counter.
This means: A voltage of 1000 V between a connection and
ground must be possible.
1. Digital multimeter Fluke 87 III or equivalent.
Accuracy:
DC voltage ± 0.1 % of reading plus 0.02% of range
DC current ± 0.4 % of reading plus 0.1% of range.
Power Supply Adequacy
To assure that the CDR PanElite system performance is in
accordance with Schick specifications, an adequate power
supply for permanent installation is essential.
The Federal Performance Standard for diagnostic X-ray
units, code of Federal Regulations, Title 21 CFR, Subchapter J, mandates an adequate power supply.
Duty Cycle
Between exposures a minimum cool-off time is maintained
(automatic exposure blockage, see Operating Instructions
manual).
Operating Instructions
During measurements and adjustments it is necessary to
energize or de-energize the unit. For all operating steps
please refer to the Operating Instructions manual.
CAUTION with PC Boards!
All PC boards are fitted with electronic components sensi-
tive to electrostatic discharge (ESD). Electrostatic charges
are unavoidable due to friction caused by clothing, carpeting etc.
English
2. Electromechanical pulse counter, model KESSLER
ELLIS KT 203 ±1 pulse, or equivalent..
3. Adjustment set with alignment tool for X-ray beam , test
block and needle phantom delivered with the unit, (customer's property).
RADIATION!
Observe the radiation protection guide lines as outlined in
the Operating Instructions manual.
X-radiation is emitted as long as the exposure switch is depressed.
The X-ray indicator must light up on the Control Pad during
radiation. An acoustic signal should also be heard.
To prevent damage of electronic chips do not touch components.
Always handle circuit boards by their edges.
Electrical Shock Hazard!
Always turn unit OFF before connecting and disconnecting
the test leads to the test points
Life-threatening voltage on DX6.
61 88 085 D3352
D3352.103.02.01.02
5
2 Visual Check Schick Technologies, Inc.
CDR PanElite
Service program
Software
Maintenance Instructions CDR PanElite
2 Visual Check
•Look for mechanical damage that may affect radiation
safety.
•Verify that all labels are affixed and legible.
Defaced labels must be replaced.
Order labels from Schick (address, see rear)
in writing stating:
•Customer Name
•Customer Address
All Model Numbers with Serial Numbers should be legible
on the unit for identification purposes. For serial numbers
see also Installation Report.
61 88 085 D3352
6D3352.103.02.01.02
Schick Technologies, Inc.3 Indicator Lights and audible Signal
P114.1648
Thisproductcomplies with DHHS
regulations21CFR Subchapter J,
applicableatdate of manufacture.
WARNING:THISX-RAYUNIT MAY BE DANGEROUS TO PATIENT
ANDOPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING
INSTRUCTIONSARE OBSERVED.UNAUTHORIZEDUSE IS PROHIBITED.
WARNING:THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT
ANDOPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING
INSTRUCTIONSARE OBSERVED.UNAUTHORIZEDUSEIS PROHIBITED.
Remote
Radiation indicator
ON
OFF
Unit ON LED
Maintenance Instructions CDR PanElite
3 Indicator Lights and audible Signal
•Unit ON LED:
•Testing the keys and the functions on the user in-
Press the main switch to the ”
I” position to turn unit ON.
Power ON tests to take about 1 minute.
The ”Unit ON LED” in the upper right corner of the user
interface will then be lit".
terface.
– When the height is adjusted, an acoustic signal is
active.
– If the forehead support functions correctly, the max.
holding torque is still present.
– If remote is installed, test the remote keys and ver-
ify proper operation of the display.
– The exposure switch must not be defective or bro-
ken.
English
See also Operating Instructions under ”Preparing the Exposure” subchapter ”Switching the system on”.
61 88 085 D3352
D3352.103.02.01.02
7
3 Indicator Lights and audible Signal Schick Technologies, Inc.
– X-ray head must be in the initial position (If not,
press return key R).
– Make sure that CDR PanElite Service program is
ready to take an image.
For more details and possible error messages see
Operating Instructions.
– Set the
– Select
–Press and hold the exposure switch until the
exposure terminates automatically.
–The exposure ends when the rotation and radiation
switch off automatically.
–The radiation indication must light up during the
exposure period.
–Simultaneously an audible beep must sound at the
unit.
P1 exposure program using the menu keys.
68kV/8mA using submenu keys.
•Interrupt an exposure – deadman feature:
– Observe a cool-off time of max. 5 mins. between
exposures (automatic exposure blockage).
– Setting same as above.
–Press the exposure switch until X-ray lights up
and subsequently
terminate immediately. The radiation indicator
lights up.
Defective light indicators constitute a safety hazard to the
patient as well as to the operator.
The user is not permitted to use the unit, until repairs
are made!
release – the exposure must
8D3352.103.02.01.02
61 88 085 D3352
Schick Technologies, Inc.4 Power supply adequacy
CAUTION RADIATION
NOTE
i
3.
1.2.
n
K1
300 VAC
4. - 5.
K1
*
o
Maintenance Instructions CDR PanElite
4 Power supply adequacy
•To determine power supply adequacy, the line volt-
age drop
1. Be sure power is disconnected at the central distribu-
tion panel!
•Remove front cover (for details see Service Manual).
2. Select 300VAC line voltage range on multimeter .
Connect measuring leads to terminal K1, L and N.
3. Connect power and switch unit ON
Wait 1 min. for self-adjustment of the unit.
Press key R o to return X-ray tube head into the initial
position.
4. Make sure that CDR PanElite Service program is ready
to take an image.
5. Select highest exposure level e.g., 90kV/12mA.
–Press and hold the exposure switch until the meter
reading is obtained.
during exposure must be measured.
n.
Line voltage, no load:Max. permissible line volt-
age drop:
180 – 208V9V
208 – 230V8V
230 – 240V7.5V
240 – 264V7V
Record reading.
•Turn unit OFF.
•Remove meter leads and refit cover.
If the voltage drop is not within the specified range advise the customer, that an adequate power supply must be
installed. Refer to Pre-Installation Instructions.
Disconnect unit and do not release for use!
English
61 88 085 D3352
D3352.103.02.01.02
9
5 kV – verification, kV ramp during panoramic exposure Schick Technologies, Inc.
WARNING
WARNING
ATTENTION
NOTE
i
NOTE
i
1.
2.
A
B
Maintenance Instructions CDR PanElite
5 kV – verification, kV ramp during panoramic
exposure
The electronics of the X-ray tube assembly are always
connected to line voltage.
Always switch the X-ray unit off and wait until V203 is
no longer illuminated before contacting the test leads.
The test leads and measuring instruments used must
have a dielectric strength of at least 1000V!
Be sure to use a battery-powered measuring instrument with shock-hazard-protected sockets.
Use only test leads with shock protection.
The X-ray tube assembly moves during the measurement.
Therefore, be sure to use test leads of sufficient length and
place the measuring instrument in a location where it is firmly seated so that it doesn’t fall down during rotation.
Measurements:
62kV correspond to 3.1V, tolerance +/-10%
During exposure the kV is increased in the central region
depending on kV/mA selected by up to 10%. This increase
can be measured in VDC.
1. Remove cover (for details see Service Manual).
2. Remove the four screws A and remove cover plate B of