Sirona CDR PanElite Maintenance manual

CDR PanElite
j~бенЙе~еЕЙ=fелнкмЕнбзел====
English
Schick Technologies, Inc.
Maintenance Instructions CDR PanElite
General information
ATTENTION !
61 88 085 D3352
2 D3352.103.02.01.02
Schick Technologies, Inc. Maintenance Instructions CDRPanElite
Contents
1 General................................................................................................................................ 5
2 Visual Check....................................................................................................................... 6
3 Indicator Lights and audible Signal.................................................................................. 7
4 Power supply adequacy.....................................................................................................9
5 kV – verification, kV ramp during panoramic exposure ............................................... 10
6 Tube Current Verification ................................................................................................ 14
7 Exposure Time Verification.............................................................................................17
8 Checking the X-Ray Beam...............................................................................................20
8.1 Diaphragm/system adjustment menu..................................................... 20
9 Checking the X-Ray Beam for panorama exposure......................................................22
9.1 Preparing the Pan adjustment........................................................ ........ 22
9.2 Checking the Pan sensor adjustment................................................ .. ... 23
9.3 Checking the Pan diaphragm adjustment................................. ... ........... 25
9.4 Checking the Pan filter adjustment.................................................... ..... 27
9.5 Checking the Pan symmetry................. .. ................................................ 29
10 Checking the X-Ray Beam for cephalometer exposure................................................ 31
10.1 Preparing the Ceph adjustment............................................. ... ... ........... 31
10.2 Checking the Ceph primary diaphragm adjustment................................ 32
10.3 Checking the Ceph rotation fixpoint adjustment..................................... 35
10.4 Checking the Ceph main X-ray beam direction adjustment.................... 37
11 Checking the alignment of the ear plugs....................................................................... 39
11.1 Preparing the ear plugs ......................................................... ................. 39
11.2 Checking the ear plug alignment........................................ .................... 40
12 Checking the laser............................................................................................................ 43
English
61 88 085 D3352 D3352.103.02.01.02
3
Schick Technologies, Inc.
Maintenance Instructions CDRPanElite
61 88 085 D3352
4 D3352.103.02.01.02
Schick Technologies, Inc. 1 General
ATTENTION
ATTENTION
ATTENTION
WARNING
ON
OFF
Maintenance Instructions CDR PanElite
1 General
To stay in compliance with the DHHS requirements the
CDR PanElite must be maintained annually.
It is the responsibility of the user to insure that the equip­ment is maintained with the manufacturer's recommended Maintenance Instructions to insure compliance with the Federal Performance Standard. The manufacturer and the assembler/installer are relieved from responsibility in those cases where the user fails to have the manufacturer's recommended maintenance per­formed. The actual maintenance inspection and consequent ser­vice must be accomplished by a trained serviceman. Nei­ther the inspection nor service is part of the equipment war­ranty.
Technical instructions required
Operating Instructions Service Manual
Instruments and adjustment tools required
All measuring devices must be suitable for measurements on the mains potential. This includes the pulse counter. This means: A voltage of 1000 V between a connection and ground must be possible.
1. Digital multimeter Fluke 87 III or equivalent.
Accuracy: DC voltage ± 0.1 % of reading plus 0.02% of range DC current ± 0.4 % of reading plus 0.1% of range.
Power Supply Adequacy
To assure that the CDR PanElite system performance is in accordance with Schick specifications, an adequate power supply for permanent installation is essential.
The Federal Performance Standard for diagnostic X-ray units, code of Federal Regulations, Title 21 CFR, Subchap­ter J, mandates an adequate power supply.
Duty Cycle
Between exposures a minimum cool-off time is maintained (automatic exposure blockage, see Operating Instructions manual).
Operating Instructions
During measurements and adjustments it is necessary to energize or de-energize the unit. For all operating steps please refer to the Operating Instructions manual.
CAUTION with PC Boards!
All PC boards are fitted with electronic components sensi-
tive to electrostatic discharge (ESD). Electrostatic charges
are unavoidable due to friction caused by clothing, carpet­ing etc.
English
2. Electromechanical pulse counter, model KESSLER ELLIS KT 203 ±1 pulse, or equivalent..
3. Adjustment set with alignment tool for X-ray beam , test block and needle phantom delivered with the unit, (cus­tomer's property).
RADIATION! Observe the radiation protection guide lines as outlined in the Operating Instructions manual. X-radiation is emitted as long as the exposure switch is de­pressed. The X-ray indicator must light up on the Control Pad during radiation. An acoustic signal should also be heard.
To prevent damage of electronic chips do not touch compo­nents. Always handle circuit boards by their edges.
Electrical Shock Hazard! Always turn unit OFF before connecting and disconnecting the test leads to the test points
Life-threatening voltage on DX6.
61 88 085 D3352 D3352.103.02.01.02
5
2 Visual Check Schick Technologies, Inc.
CDR PanElite
Service program Software
Maintenance Instructions CDR PanElite
2 Visual Check
Look for mechanical damage that may affect radiation safety.
Verify that all labels are affixed and legible. Defaced labels must be replaced.
Order labels from Schick (address, see rear) in writing stating:
Customer Name
Customer Address
All Model Numbers with Serial Numbers should be legible on the unit for identification purposes. For serial numbers see also Installation Report.
61 88 085 D3352
6 D3352.103.02.01.02
Schick Technologies, Inc. 3 Indicator Lights and audible Signal
P1 14.1 64 8
Thisproductcomplies with DHHS regulations21CFR Subchapter J, applicableatdate of manufacture.
Dateofmanufacture:
SironaDentalSystems GmbH Fabrikstraße31 D-64625Bensheim,Germany
Thisproductcomplies with DHHS regulations21CFR Subchapter J, applicableatdate of manufacture.
Dateofmanufacture:
SironaDentalSystems GmbH Fabrikstraße31 D-64625Bensheim,Germany
P10
11.5
64
8
S
k
V
m
A
P 10
11.5
64
8
Program
s
k
V
m
A
Prog.
S
k
V
m
A
P 10
11.5
64
8
Program
WARNING:THISX-RAYUNIT MAY BE DANGEROUS TO PATIENT ANDOPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONSARE OBSERVED.UNAUTHORIZEDUSE IS PROHIBITED.
WARNING:THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT ANDOPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONSARE OBSERVED.UNAUTHORIZEDUSEIS PROHIBITED.
Remote
Radiation indicator
ON
OFF
Unit ON LED
Maintenance Instructions CDR PanElite

3 Indicator Lights and audible Signal

Unit ON LED:
Testing the keys and the functions on the user in-
Press the main switch to the ”
I” position to turn unit ON.
Power ON tests to take about 1 minute. The ”Unit ON LED” in the upper right corner of the user
interface will then be lit".
terface.
– When the height is adjusted, an acoustic signal is
active.
– If the forehead support functions correctly, the max.
holding torque is still present.
– If remote is installed, test the remote keys and ver-
ify proper operation of the display.
– The exposure switch must not be defective or bro-
ken.
English
See also Operating Instructions under ”Preparing the Expo­sure” subchapter ”Switching the system on”.
61 88 085 D3352 D3352.103.02.01.02
7
3 Indicator Lights and audible Signal Schick Technologies, Inc.
CAUTION RADIATION
CAUTION RADIATION
ATTENTION
P1 14.1 64 8
CAUTION RADIATION ! Observe Radiation Protection Guide Lines
see Operating Instructions
Unit ON LED
Return key R
menu keys
submenu key
Radiation indicator
Maintenance Instructions CDR PanElite
Take an exposure:
– X-ray head must be in the initial position (If not,
press return key R).
– Make sure that CDR PanElite Service program is
ready to take an image. For more details and possible error messages see
Operating Instructions. – Set the – Select
Press and hold the exposure switch until the
exposure terminates automatically.
The exposure ends when the rotation and radiation
switch off automatically.
The radiation indication must light up during the
exposure period.
Simultaneously an audible beep must sound at the
unit.
P1 exposure program using the menu keys.
68kV/8mA using submenu keys.
Interrupt an exposure – deadman feature:
– Observe a cool-off time of max. 5 mins. between
exposures (automatic exposure blockage).
– Setting same as above.
Press the exposure switch until X-ray lights up
and subsequently terminate immediately. The radiation indicator lights up.
Defective light indicators constitute a safety hazard to the
patient as well as to the operator.
The user is not permitted to use the unit, until repairs are made!
release – the exposure must
8 D3352.103.02.01.02
61 88 085 D3352
Schick Technologies, Inc. 4 Power supply adequacy
CAUTION RADIATION
NOTE
i
3.
1. 2.
n
K1
300 VAC
4. - 5.
K1
*
o
Maintenance Instructions CDR PanElite
4 Power supply adequacy
To determine power supply adequacy, the line volt-
age drop
1. Be sure power is disconnected at the central distribu-
tion panel!
Remove front cover (for details see Service Manual).
2. Select 300VAC line voltage range on multimeter . Connect measuring leads to terminal K1, L and N.
3. Connect power and switch unit ON Wait 1 min. for self-adjustment of the unit. Press key R o to return X-ray tube head into the initial position.
4. Make sure that CDR PanElite Service program is ready to take an image.
5. Select highest exposure level e.g., 90kV/12mA.
Press and hold the exposure switch until the meter reading is obtained.
during exposure must be measured.
n.
Line voltage, no load: Max. permissible line volt-
age drop:
180 – 208V 9V
208 – 230V 8V
230 – 240V 7.5V
240 – 264V 7V
Record reading.
Turn unit OFF.
Remove meter leads and refit cover.
If the voltage drop is not within the specified range ad­vise the customer, that an adequate power supply must be installed. Refer to Pre-Installation Instructions.
Disconnect unit and do not release for use!
English
61 88 085 D3352 D3352.103.02.01.02
9
5 kV – verification, kV ramp during panoramic exposure Schick Technologies, Inc.
WARNING
WARNING
ATTENTION
NOTE
i
NOTE
i
1.
2.
A
B
Maintenance Instructions CDR PanElite
5 kV – verification, kV ramp during panoramic
exposure
The electronics of the X-ray tube assembly are always connected to line voltage.
Always switch the X-ray unit off and wait until V203 is no longer illuminated before contacting the test leads.
The test leads and measuring instruments used must have a dielectric strength of at least 1000V!
Be sure to use a battery-powered measuring instru­ment with shock-hazard-protected sockets.
Use only test leads with shock protection.
The X-ray tube assembly moves during the measurement. Therefore, be sure to use test leads of sufficient length and place the measuring instrument in a location where it is firm­ly seated so that it doesn’t fall down during rotation.
Measurements: 62kV correspond to 3.1V, tolerance +/-10%
During exposure the kV is increased in the central region depending on kV/mA selected by up to 10%. This increase can be measured in VDC.
1. Remove cover (for details see Service Manual).
2. Remove the four screws A and remove cover plate B of
the electronics box.
61 88 085 D3352
10 D3352.103.02.01.02
Schick Technologies, Inc. 5 kV – verification, kV ramp during panoramic exposure
CAUTION RADIATION
NOTE
i
3.
4.
5.
6. - 7.
KV– KV+
20VDC
com.
400 VDC
PC-board DX6
C
D
n
*
o
Maintenance Instructions CDR PanElite
3. Connect digital voltmeter to KV+ and KV– test points on PC board DX6 and select range 20 VDC.
4. Refit cover C and tighten screw D securely.
5. Switch unit ON
Wait 1 min. for power on of the unit. Press key R position.
6. Make sure that CDR PanElite Service program is ready to take an image.
7. Select P1 program and 62kV/8mA.
n.
o to return X-ray tube head into the initial
Measurement:
Press and hold the exposure switch until the
exposure terminates automatically.
The following values must be obtained
(see also diagram on next page)
up to 5.25 seconds:
from 6.25 to 7.85 s KV-ramp: e.g., 4 V ±0.4V
after 8.85 seconds:
3.1 V ±0.3V
3 V ±0.3V
English
61 88 085 D3352 D3352.103.02.01.02
If specified values cannot be obtained, see Service Man­ual for troubleshooting.
11
5 kV – verification, kV ramp during panoramic exposure Schick Technologies, Inc.
15
ramp
max. 4.05V
min. 3.45V
VDC
s
Exposure time
0 2
1
2
3
4
5
6
7
8
9
1 104 5 6 7 8 9 11 12 13 143
5.25s 8.85s
ramp
1.6s
14.1s Complete P1 exposure time
7.05s
max. 3.41V
min. 2.79V
6.25s 7.85s
Nominal 3.1V
Nominal 4.0V
Maintenance Instructions CDR PanElite
kV – Verification, kV–ramp during panoramic exposure
kV ramp diagram
with program P1 and 62kV/8mA set on the Control Pad and temple support fully open.
12 D3352.103.02.01.02
61 88 085 D3352
Schick Technologies, Inc. 5 kV – verification, kV ramp during panoramic exposure
8.
Maintenance Instructions CDR PanElite
Turn unit OFF.
8. Remove top cover and meter leads.
English
61 88 085 D3352 D3352.103.02.01.02
13
6 Tube Current Verification Schick Technologies, Inc.
WARNING
WARNING
WARNING
Maintenance Instructions CDR PanElite

6 Tube Current Verification

The electronics of the X-ray tube assembly are always connected to line voltage.
Always switch the X-ray unit off and wait until V203 is no longer illuminated before contacting the test leads.
The test leads and measuring instruments used must have a dielectric strength of at least 1000V!
Be sure to use a battery-powered measuring instru­ment with shock-hazard-protected sockets.
Use only test leads with shock protection.
Be sure to switch the X-ray unit off before removing the jumper for the mA measuring jack.
14 D3352.103.02.01.02
61 88 085 D3352
Schick Technologies, Inc. 6 Tube Current Verification
CAUTION RADIATION
NOTE
i
NOTE
i
C
D
1. 2.
MA+MA-
20VDC
com.
PC-board DX6
·
A
400 VDC
3.
4. - 5.
n
*
o
Maintenance Instructions CDR PanElite
1. Remove shunt jumper A from MA+/MA – test points on PC board DX6. Connect digital multimeter to VDC and select range 20 mVDC.
2. Reattach cover C and tighten screw D securely.
3. Switch unit ON
Wait 1 min. for self-adjustment of the unit. Press key R o to return X-ray tube head into the initial position.
4. Select 66kV/8mA. Make sure that CDR PanElite Ser­vice program is ready to take an image.
5. If P1 program and 66kV/8mA are selected . The unit must be ready for radiation.
n.
Measurement:
Press and hold the exposure switch until meter reading is obtained.
The multimeter shall indicate 8mV ±1.6mV. Record reading.
Readings: 1mV corresponds to a tube current of 1mA, toler-
ance +/-20%.
English
61 88 085 D3352 D3352.103.02.01.02
If specified values cannot be obtained, see Service Man­ual, chapter ”Tube Current Verification”.
15
6 Tube Current Verification Schick Technologies, Inc.
6.
If specified value is obtained switch unit OFF.
6. Remove upper cover and meter leads
Replace jumper!
Maintenance Instructions CDR PanElite
61 88 085 D3352
16 D3352.103.02.01.02
Schick Technologies, Inc. 7 Exposure Time Verification
WARNING
WARNING
C
D
400 VDC
PC board DX6
1.
to N to L
T2
of 110VAC/60Hz power receptacle
T1
Pulse counter
2.
Maintenance Instructions CDR PanElite
7 Exposure Time Verification
The electronics of the X-ray tube assembly are always connected to line voltage.
Always switch the X-ray unit off and wait until V203 is no longer illuminated before contacting the test leads.
The test leads and measuring instruments used must have a dielectric strength of at least 1000V!
Be sure to use a battery-powered measuring instru­ment with shock-hazard-protected sockets.
Use only test leads with shock protection.
1. Connect... – test point
110V power receptacle
– the common lead of the pulse counter to
110V power receptacle
– the second measuring lead of the pulse counter to
test point T1
2. Refit cover C and tighten screw D securely.
T2 of terminal on PC board DX6 to L of
English
N of the
61 88 085 D3352 D3352.103.02.01.02
17
7 Exposure Time Verification Schick Technologies, Inc.
CAUTION RADIATION
NOTE
i
6.3.
4. - 5.
n
*
o
3. Switch unit ON n.
Wait 1 min. for power on of the unit. Press key R o to return X-ray tube head to the initial position.
4. Select 66kV/8mA.
Make sure that CDR PanElite Service program is ready to take an image.
Maintenance Instructions CDR PanElite
Measurement:
Press and hold the exposure switch until the expo­sure terminates automatically.
5. Select service routine S2 (Selecting a service routine: See Service Manual):
Set radiation time to 1 sec.
Measured values at 50 Hz 50 pulses +/-10%
Measured values at 60 Hz 60 pulses +/-10%
Record average pulse count.
If specified value cannot be obtained, see Service Man­ual, chapter ”Exposure Time Verification”.
If specified value is obtained, switch unit OFF.
6. Remove upper cover and meter leads.
18 D3352.103.02.01.02
61 88 085 D3352
Loading...
+ 40 hidden pages