Sirona c8+, c8+ turn Installation Requirements

Page 1
New as of:
10.2014
English
C8+ / C8+ Turn
Page 2
New as of:
10.2014
General information Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
2 D3417.021.04.11.02
Changes since the last version (09/2009):
Chapter or section, page
Does not apply to the following products: C8
+
Solo, C8+Stand-Alone, C8+Side delivery,
C8
+
Rear delivery
About this document
This document describes the installation requirements for the C8
+
/C8+Turn/C8+Turn treatment centers.
The following sets of Installation Instructions describe the subsequent installation procedure:
REF 59 42 912 (C8
+
)
REF 59 78 197 (C8
+
Tu r n )
You will also need the following drilling template in order to fasten the treatment center securely to the floor:
REF 59 66 911 (C8
+
,C8+Tu r n )
General information
Page 3
Sirona Dental Systems GmbH Table of contents Installation Requirements C8+/C8
+
Tu r n
60 28 000 D3417 D3417.021.04.11.02 3
English
1 Preparations....................................................................................................................... 5
1.1 Safety........................................................................................................... 6
1.2 Media quality................................................................................................ 7
1.3 Supply lines in the termination panel ........................................................... 8
1.4 Underfloor installation of SIVISION connections ......................................... 11
1.5 Mounting plates C8
+
(not for C8+Turn) ....................................................... 12
1.6 Demonstration chair plate C8
+
Turn ............................................................ 13
2 Dimensions, technical data ............................................................................................ 15
2.1 C8+1:20....................................................................................................... 16
2.2 C8
+
Cart 1:20............................................................................................... 21
2.3 C8
+
: Technical data ..................................................................................... 23
2.4 C8
+
Turn 1:20 .............................................................................................. 24
2.5 C8
+
Turn Cart 1:20 ...................................................................................... 31
2.6 C8
+
Turn: Technical data ............................................................................. 33
3 Electromagnetic compatibility ....................................................................................... 35
3.1 Accessories ................................................................................................. 36
3.2 Electromagnetic emission ............................................................................ 37
3.3 Immunity to interference .............................................................................. 38
3.4 Working clearances ..................................................................................... 40
Table of contents
Page 4
Page 5
C8+/C8+Turn
60 28 000 D3417 D3417.021.04.11.02 5
1 Preparations
1.1 Safety .................................................................................................................................... 6
1.2 Media quality ......................................................................................................................... 7
1.3 Supply lines in the termination panel ..................................................................................... 8
1.4 Underfloor installation of SIVISION connections ................................................................... 11
1.5 Mounting plates C8
+
(not for C8+Turn) ................................................................................. 12
1.6 Demonstration chair plate C8
+
Turn ...................................................................................... 13
Page 6
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
6 D3417.021.04.11.02
ATTENTION
It is essential that you comply with the warning and safety information contained in the Installationsvoraussetzungen. All such information is highlighted by the signal words NOTE, CAUTION or WARNING.
ATTENTION
The unit may only be operated by properly trained person­nel. On-site installation must be performed in accordance with our specifications. In our capacity as a manufacturer of medical electrical equipment, we only consider ourselves to be responsible for the safety characteristics of the unit if we perform mainte­nance and repair work on it ourselves or if this is performed by centers that we haveexpressly authorized to do so; and if defective components are replaced with original spare parts. When this work is carried out, we recommend that you ask those responsible for the work to provide a certificate indi­cating the nature and scope of the work, including details of changes to rated data or the working area where applicable. This certificate should also be signed and dated, and should show information about the relevant company.
ATTENTION
By law, any changes to this unit which may compromise the safety of operators, patients or third parties are not permis­sible. For reasons of product safety, this product may only be op­erated with original Sirona accessories or accessories from third parties approved by Sirona. The user assumes the risk of using non-approved accessories. If any devices not approved by Sirona are connected, they must comply with the applicable standards, e.g.:
– IEC 60950 for information technology equipment
(e.g. PCs) and
– IEC 60601-1 for medical equipment.
The monitor used at the treatment center must meet the re­quirements of IEC 60950. The loudspeaker port of the mon­itor may only be connected to a device that complies with IEC 60950 (e.g. a PC) or IEC 60601-1. Under no circum­stances may it be connected to a stereo system, for exam­ple. If a system is being assembled during installation, the re­quirements of IEC 60601-1 (3rd edition) must be adhered to. The person assembling the system assumes responsi­bility for ensuring that it conforms to Directive 93/42/EEC.
ATTENTION
The floor must be flat and level (DIN 18 202). A mounting plate must be used for uneven floors (see Section 1.5, "Mounting plates C8+(not for C8+Turn)" on page 12).
ATTENTION
The floor must have a minimum loading capacity of 0.5N/ cm².
ATTENTION
Wireless phone interference with medical electrical equip­ment: To ensure safe operation of medical electrical equipment, the use of mobile wireless phones in practice or hospital en­vironments is prohibited.
ATTENTION
Electromagnetic compatibility: The unit should not be op-
erated in the immediate vicinity of other devices. If this proves to be unavoidable, the unit should be monitored to ensure that it is operating properly.
1.1 Safety
Page 7
Sirona Dental Systems GmbH 1 Preparations Installation Requirements C8+/C8+Tu r n
60 28 000 D3417 D3417.021.04.11.02 7
English
,
Water quality
Lime deposits and corrosion residues in tap water can lead to the following malfunctions:
Premature clogging of the filters in the unit
Rapid clogging of the fine water paths and jets in the treatment instruments
For these reasons, the following points must be observed:
Permitted water pressure: 2.5 bar (36.25 psi) to 6 bar (87 psi)
Permitted minimum flow volume: 3 l/min
For water hardness (total hardness) of 2.2 mmol/l (= 12° dH ), install water softeners. Set the blend hardness to 1.4 mmol/l (= 8° dH).
Install a conventional fine filter; fineness: > 80 µm (0.08 mm).
Installation must be performed in compliance with the recommendations of the national installation require­ments (e.g. EN 1717/DIN 1988).
The water quality must comply with the national re­quirements for drinking water.
The connection must be made to cold water.
When laying the water pipe to the treatment center, comply with the following instructions to reduce the quantity of micro-organisms in the feed pipe:
– Avoid long stub lines to the treatment center.
– Carry out the installation so that, where possible,
other main consumers (such as the sink) are fed from the same line downstream of the treatment center connection.
– Avoid laying the supply line parallel to hot water
pipes.
Observe EN 1717 for the protection of public drinking water. Provided a disinfection system is also installed, the treatment center fulfills the requirements of EN 1717 and of the DVGW (German Technical and Scientific As­sociation for Gas and Water). It can be connected di­rectly to the public drinking water system.
Air quality
The air for driving the turbines, for cooling the drives and for the cooling spray must be free from oil, dry and hygienically faultless.
Suction pipe
Install steam trap K.
With a vacuum of pu >0.18 bar back pressure, the treat-
ment center must be retrofitted with the "Vacuum limiter" retrofit kit (REF 59 68 826).
1.2 Media quality
N
>80 µm
Air
Suction machine
oil-free
Cold water
Dilution hardness
(drinking water quality)
1 bar = 100 kPa = 14.5 psi
Page 8
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
8 D3417.021.04.11.02
1.3 Supply lines in the termination panel
5
3
6
6
6
7
Center base plate opening to floor
1
2
4
9
1 3/8“
min.
500mm
20”
7, 6
max.
60mm
2 3/8”
3 4 5
max. 5mm 3/16”
max. 5mm 3/16”
3/8”
12
10 x 1mm
10 x 1mm
3/8
x 1/32“
3/8
x 1/32“
1 3/8“
Foot end
36.5mm
50 mm
36.5mm
Page 9
Sirona Dental Systems GmbH 1 Preparations Installation Requirements C8+/C8+Tu r n
60 28 000 D3417 D3417.021.04.11.02 9
English
Observe the national regulations for electrical installa­tions (e.g. VDE 0100, VDE 0100, Part 710).
Comply with the national regulations for water supply installations (e.g. EN 1717, DIN 1988) and sewage in­stallations (e.g. EN 12056-1).
For the suction pipe, observe the instructions in the Suction Machine Installation Instructions.
For fastening the pipe ends in the installation field, we recommend using an installation template. They can be ordered from Sirona under REF 33 15 830.
If necessary, you can also prepare the template your­self based on the above sketch (not true to scale!).
Supply lines in the termination panel
Table 1: Supply lines
Item Description
1 Water inlet pipe 10x1 mm,
corner valve outlet 3/8"
2 Compressed air inlet pipe 10x1mm, corner
valve outlet 3/8"
3 Suction pipe DN40 HT-PP DIN 19560 (polypro-
pylene, inner diameter 36.5mm!)
4 Water drain DN40 HT-PP DIN 19560 (polypro-
pylene, inner diameter 36.5mm!)
5 Installation pipe, DN40 HT-PP DIN 19560
(polypropylene, 40mm!)
6 Suction machine control cable ( )and call
cables ( ) 3x1.5mm
2
7 Power cable 3x1.5mm
2
Fuse: 16A slow-blow Recommended: Type B automatic circuit breaker
8 Not applicable
9 Installation pipe (or corresponding flat duct) for
additional requirement e.g. practice network connecting cable
,
Page 10
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
10 D3417.021.04.11.02
Supply above the floor, "above-floor installation"
The supply pipe ends, corner valves and cables must be routed as shown above.
The retrofit kit for above-floor installation (REF 33 17 265) is required for connection.
ATTENTION
For cleaning, flush the air and water pipes thoroughly (metal chips!).
Supply through the floor, "underfloor installation"
1. The top edges of the corner valves for air and water
must not protrude more than 60mm above the upper surface of the finished floor.
2. The suction and drain pipes must be flush with the up­per surface of the floor (a deviation of +5mm is permis­sible). Internal diameter for both pipes: 36.5mm.
3. The electrical cables must protrude at least 500mm.
Supply lines in the termination panel
>5
00
m
m
<5m
m
<60m
m
>20“
3/8“
<3/16“
<2
3/8“
Ø 36,5 mm
10x1 mm 3/8” x 1/32”
DN 40
>5
00m
m
200m
m
250m
m
> 20’’
8’’
10’’
Ø 36,6mm
3/8“
20
Ø
36,5
m
m
Installation template
Page 11
Sirona Dental Systems GmbH 1 Preparations Installation Requirements C8+/C8+Tu r n
60 28 000 D3417 D3417.021.04.11.02 11
English
Important information for the installer
Depending on the prevailing local conditions, the existing cable set can be installed in the cable duct of an underfloor installation by an installer prior to the installation of the treatment center. In this case, please observe the following:
Proceed with extreme care when running the cables. Par­ticularly cables L15 and L38 are very sensitive, and must never be kinked or twisted. The cables must not overlap or cross one another.
RS232 (L37) and the XGA cable (L38) are not yet cut to length and terminated on the PC side. It would be impossi­ble to pull the cables through when installing them under floor level if a sub D connector were already connected.
These cables should always be pulled.
Free length A of cables at the treatment center end: Length A = 600mm
If S video cable L15 is equipped with both a female and a male connector, make sure that the female connector (socket) points to the connection box of the treatment cen­ter.
1. Bend the wire at the front end of cables L37 and L38 to form a hook.
2. Pull cables L15, L37, and L38 as well as the audio ca­ble and the protective earth conductor for the treatment center through the cable duct to the location of the SIV­ISION PC.
Save the accessory parts for final installation!
1.4 Underfloor installation of SIVISION connections
2.
1.
A
L15, L37, L38, audio Protective earth conductor
SIVISION
Page 12
1 Preparations Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
12 D3417.021.04.11.02
1.5 Mounting plates C8+(not for C8+Turn)
Mounting plate
Demonstration chair plate
REF 58 26 511
REF 47 08 074
For especially uneven floor conditions, a steel mounting plate is available to compensate for the irregularities of the floor (REF 58 26 511).
For floors which do not permit permanent connection of the unit (e.g. demo operation at a fair), installation on a steel demonstration chair plate is possible (REF 47 08 074).
Page 13
Sirona Dental Systems GmbH 1 Preparations Installation Requirements C8+/C8+Tu r n
60 28 000 D3417 D3417.021.04.11.02 13
English
1.6 Demonstration chair plate C8+Turn
Demonstration chair plate
For floors which do not permit permanent connection of the unit (e.g. demo operation at a fair), installation on a steel demonstration chair plate is possible (REF 59 46 269).
REF 59 46 269
C8
+
Turn
Page 14
Page 15
C8+/C8+Turn
60 28 000 D3417 D3417.021.04.11.02 15
2 Dimensions, technical data
2.1 C8+1:20 ................................................................................................................................. 16
2.2 C8
+
Cart 1:20 ......................................................................................................................... 21
2.3 C8
+
: Technical data................................................................................................................ 23
2.4 C8
+
Turn 1:20......................................................................................................................... 24
2.5 C8
+
Turn Cart 1:20 ................................................................................................................. 31
2.6 C8
+
Turn: Technical data........................................................................................................ 33
Page 16
2 Dimensions, technical data Sirona Dental Systems GmbH
2.1 C8+ 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
16 D3417.021.04.11.02
Recommended distances from cabinet or wall
2.1 C8+1:20
1700
67”
1300
51 3/16”
2500
98 7/16”
250
9 7/8”
!
!
250
9 7/8”
Hazard warning: The lamp installed here and the dentist
element have a swivel range which ex­ceeds the specified distances!
Center of floor opening
Page 17
Sirona Dental Systems GmbH 2 Dimensions, technical data Installation Requirements C8+/C8
+
Tu r n 2.1 C8+ 1:20
60 28 000 D3417 D3417.021.04.11.02 17
English
With long support arm for the dentist element
C8+1:20
* max. 755 29 3/4”
min. 395 15 1/2"
1212
47 3/4”
Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting
45
1 3/4”
90
3 1/2”
1950
76 3/4”
Height of the lamp with
SIVISION on the swivel
arm or lamp support
tube
1875
73 3/4”
Height of the lamp
without SIVISION or
SIVISION on the tray
45°
65°
10°
max. 1230 48 3/8”
min. 760 29 7/8”
SIROLUXLEDview
Page 18
2 Dimensions, technical data Sirona Dental Systems GmbH
2.1 C8+ 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
18 D3417.021.04.11.02
C8+1:20
1540
60 5/8”
860
33 7/8”
1210
47 1/2”
310
12 1/4”
max. 380 +15
15“ +5/8”
2070 +50***
81 1/2” +2“ ***
*** Depending on the
chair height
45
1 3/4”
78
3”
459
18”
430 +15
17“ +5/8”
180
7”
90°
90°
90°
60°
750
29 1/2”
770
30 3/8”
250
9 7/8”
Page 19
Sirona Dental Systems GmbH 2 Dimensions, technical data Installation Requirements C8+/C8
+
Tu r n 2.1 C8+ 1:20
60 28 000 D3417 D3417.021.04.11.02 19
English
With short support arm for the dentist element
C8+1:20
* max. 755 29 3/4”
min. 395 15 1/2"
1212
47 3/4”
Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting
45
1 3/4”
90
3 1/2”
1950
76 3/4”
Height of the lamp with
SIVISION on the swivel
arm or lamp support
tube
1875
73 3/4”
Height of the lamp
without SIVISION or
SIVISION on the tray
45°
65°
10°
max. 1440 56 3/4”
min. 970 38 1/8”
SIROLUXLEDview
Page 20
2 Dimensions, technical data Sirona Dental Systems GmbH
2.1 C8+ 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
20 D3417.021.04.11.02
C8+1:20
*** Depending on the chair height
max. 380 +15
15“ +5/8”
1170
46”
860
33 7/8”
1210
47 1/2”
310
12 1/4”
150
140
2070 +50***
81 1/2” +2“ ***
45
1 3/4”
459
18”
430 +15
17“ +5/8”
180
7”
90°
90°
90°
60°
750
29 1/2”
770
30 3/8”
250
9 7/8”
Page 21
Sirona Dental Systems GmbH 2 Dimensions, technical data Installation Requirements C8+/C8
+
Tu r n 2.2 C8+ Cart 1:20
60 28 000 D3417 D3417.021.04.11.02 21
English
2.2 C8+Cart 1:20
1212
47 3/4”
Height setting in 5 steps, 26 mm each
800
31 1/2”
* max. 755 29 3/4”
min. 395 15 1/2"
Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting
45
1 3/4”
930
36 5/8”
90
3 1/2”
1950
76 3/4”
Height of the lamp with
SIVISION on the swivel
arm or lamp support
tube
1875
73 3/4”
Height of the lamp
without SIVISION or
SIVISION on the tray
45°
65°
10°
SIROLUXLEDview
Page 22
2 Dimensions, technical data Sirona Dental Systems GmbH
2.2 C8+ Cart 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
22 D3417.021.04.11.02
C8+Cart 1:20
*** Depending on the chair height
max. 380 +15
15“ +5/8”
310
12 1/4”
Supply cable
2300 90 1/2”
420
16 1/2”
480
18 7/8”
430
17”
860
33 7/8”
1210
47 1/2”
2070 +50***
81 1/2” +2“ ***
45
1 3/4”
459
18”
430 +15
17“ +5/8”
180
7”
90°
90°
90°
60°
750
29 1/2”
770
30 3/8”
250
9 7/8”
Page 23
Sirona Dental Systems GmbH 2 Dimensions, technical data Installation Requirements C8+/C8
+
Tu r n 2.3 C8+: Technical data
60 28 000 D3417 D3417.021.04.11.02 23
English
2.3 C8+: Technical data
Technical data
Power supply line 100 V / 115 V / 127 V / 230 VAC 50 / 60 Hz
Nominal current 11.5 A .....................at 100 V AC
10 A .....................at 115 V AC
9 A .....................at 127 V AC
5 A .....................at 230 V AC
,QDGGLWLRQPD[ $IRUH[WHUQDOGHYLFHV
Main unit fuse T 10 A H, 250 V AC , REF 10 77 460
Transport and storage conditions Temperature: -40°C – +70° C (-40°F – +158°F)
Relative humidity: ..................... 10% – 95%
Air pressure: .........500 hPa – 1060 hPa
Operating conditions Ambient temperature: ......... 10°C – 40°C (50°F – 104°F)
Relative humidity: ............30% – 85% (no condensation)
Air pressure: .....................................700hPa – 1060hPa
Maximum load capacity of the patient chair 135kg, including accessories
Installation site
3000 m above sea level
Pollution degree 2 acc. to IEC 60664-1
Protection class Class I device
Device class in accordance with Directive 93/ 42/EEC
Class IIa equipment
Degree of protection against electric shock
Type B applied parts, except SIROCAM 3/SIROCAM C: Type BF applied parts
Degree of protection against ingress of water Ordinary equipment (not protected)
The foot switch has an IPX1 degree of protection against liquids (drip-proof).
Operating mode Continuous operation with intermittent loading corresponding to the
dental mode of working.
Permanently connected unit.
Year of manufacture ...............................................................
.........................(on rating plate of chair)
Dimensions of the packaging
Dentist element, assistant element, C8
+
Cart dentist element
Water unit
Chair Upholstery SIROLUX / LEDview
120cm x 63 cm x 110cm
97.5cm x 50.5 cm x 50.5 cm 120cmx63cmx96cm 157cmx64cmx82cm 120cmx52cmx40cm
10.5cm x 56 cm x 28cm
Weight (1 kg = 2.2 lbs) incl. / without packaging
Dentist element, assistant element, C8
+
Cart dentist element
Water unit
Chair Upholstery SIROLUX / LEDview
37.5kg / 10 kg + 3kg 17kg / 13 kg 44kg / 16.5 kg 120kg / 90 kg 13kg / 9 kg 14kg / 10 kg
On-site pressure readings (1 bar = 100 kPa = 14.5 psi)
Air min./max.
Water min./max.
550 kPa / 750 kPa (5.5 / 7.5bar) 250 kPa / 600 kPa (2.5 / 6bar)
Page 24
2 Dimensions, technical data Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
24 D3417.021.04.11.02
Recommended distances from cabinet or wall
2.4 C8+Turn 1:20
Hazard warning: The lamp installed here and the dentist
element have a swivel range which ex­ceeds the specified distances!
Center of floor opening
1600
63”
1600
63”
2500
98 7/16”
250
9 7/8”
250
9 7/8”
Page 25
Sirona Dental Systems GmbH 2 Dimensions, technical data Installation Requirements C8+/C8
+
Tu r n 2.4 C8+ Turn 1:20
60 28 000 D3417 D3417.021.04.11.02 25
English
Minimum distances for conversion from right-handed to left-handed treatment center and vice versa 2700x2650mm
C8+Turn 1:20
Center of floor opening
1050
41 3/8”
1050
41 3/8”
2750
108 1/4”
1600
63”
1234
48 1/2”
1600
63”
Cabinet on the
right or left
max. height:
900mm
250
9 7/8”
250
9 7/8”
Page 26
2 Dimensions, technical data Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
26 D3417.021.04.11.02
Minimum distances for conversion from right-handed to left-handed treatment center and vice versa in narrow spaces 3650x2100mm
C8+Turn 1:20
1050
41 3/8”
1050
41 3/8”
Center of floor opening
3650
143 3/4”
2134
84”
1250
49 1/4”
Installation area
max. 900 mm high
250
9 7/8”
250
9 7/8”
Page 27
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20 Installation Requirements C8+/C8
+
Tu r n 2.4 C8+ Turn 1:20
60 28 000 D3417 D3417.021.04.11.02 27
English
Identical working conditions for right-handed and left-handed operators: Position for right-handed operators
C8+Turn 1:20
Center of floor opening
1050
41 3/8”
1050
41 3/8”
1600
63”
1600
63”
17°
2750
108 1/4”
1234
48 1/2”
250
9 7/8”
250
9 7/8”
Page 28
2 Dimensions, technical data C8+ / C8+ Turn 1:20 Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
28 D3417.021.04.11.02
Identical working conditions for right-handed and left-handed operators: Position for left-handed operators
C8+Turn 1:20
Center of floor opening
1050
41 3/8”
1050
41 3/8”
1600
63”
1600
63”
17°
2750
108 1/4”
1234
48 1/2”
250
9 7/8”
250
9 7/8”
Page 29
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20 Installation Requirements C8+/C8
+
Tu r n 2.4 C8+ Turn 1:20
60 28 000 D3417 D3417.021.04.11.02 29
English
C8+Turn 1:20
Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting
1950
76 3/4”
Height of the lamp with
SIVISION on the swivel
arm or lamp support
tube
1875
73 3/4”
Height of the lamp
without SIVISION or
SIVISION on the tray
45°
65°
10°
max. 1230 48 3/8”
min. 760 29 7/8”
1266
49 7/8”
* max. 770 30 3/8”
min. 410 16 1/8"
45
1 3/4”
210
8 1/4”
SIROLUXLEDview
Page 30
2 Dimensions, technical data C8+ / C8+ Turn 1:20 Sirona Dental Systems GmbH
2.4 C8+ Turn 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
30 D3417.021.04.11.02
C8+Turn 1:20
*** Depending on the chair height
max. 380 +15
15“ +5/8”
1480
58 1/4”
920
36 1/4”
1270
50”
310
12 1/4”
2070 +50***
81 1/2” +2“ ***
45
1 3/4”
800 +50***
31 1/2” +2”
400
15 3/4”
430 +15
17“ +5/8”
180
7”
90°
90°
90°
60°
750
29 1/2”
770
30 3/8”
250
9 7/8”
Page 31
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20 Installation Requirements C8+/C8
+
Tu r n 2.5 C8+ Turn Cart 1:20
60 28 000 D3417 D3417.021.04.11.02 31
English
2.5 C8+Turn Cart 1:20
Height setting in 5 steps, 26 mm each
800
31 1/2”
Standard upholstery
* max.: backrest and chair at highest setting
min.: backrest and chair at lowest setting
930
36 5/8”
1950
76 3/4”
Height of the lamp with
SIVISION on the swivel
arm or lamp support
tube
1875
73 3/4”
Height of the lamp
without SIVISION or
SIVISION on the tray
65°
10°
1266
49 7/8”
* max. 770 30 3/8”
min. 410 16 1/8"
45
1 3/4”
210
8 1/4”
45°
SIROLUXLEDview
Page 32
2 Dimensions, technical data C8+ / C8+ Turn 1:20 Sirona Dental Systems GmbH
2.5 C8+ Turn Cart 1:20 Installation Requirements C8
+
/C8+Tu r n
60 28 000 D3417
32 D3417.021.04.11.02
C8+Turn Cart 1:20
*** Depending on the chair height
max. 380 +15
15“ +5/8”
920
36 1/4”
1270
50”
420
16 1/2”
310
12 1/4”
480
18 7/8”
430
17”
Supply cable
2300 90 1/2”
2070 +50***
81 1/2” +2“ ***
45
1 3/4”
400
15 3/4”
430 +15
17“ +5/8”
180
7”
90°
90°
90°
60°
750
29 1/2”
770
30 3/8”
250
9 7/8”
Page 33
Sirona Dental Systems GmbH 2 Dimensions, technical data C8+ / C8+ Turn 1:20 Installation Requirements C8+/C8
+
Tu r n 2.6 C8+ Turn: Technical data
60 28 000 D3417 D3417.021.04.11.02 33
English
2.6 C8+Turn: Technical data
Technical data
Power supply line 100 V / 115 V / 127 V / 230 VAC 50 / 60 Hz
Nominal current 11.5 A ...........................................................at 100 V AC
10 A ..............................................................at 115 V AC
9 A ................................................................at 127V AC
5 A ................................................................at 230 V AC
,QDGGLWLRQPD[ $IRUH[WHUQDOGHYLFHV
Main unit fuse T 10 A H, 250 V AC , REF 10 77 460
Transport and storage conditions Temperature: -40°C – +70° C (-40°F – +158°F)
Relative humidity: ..................... 10% – 95%
Air pressure: .........500 hPa – 1060 hPa
Operating conditions Ambient temperature: ......... 10°C – 40°C (50°F – 104°F)
Relative humidity: ............30% – 85% (no condensation)
Air pressure: .....................................700hPa – 1060hPa
Maximum load capacity of the patient chair 135kg, including accessories Installation site
ื PDERYHVHDOHYHO
Pollution degree 2 acc. to IEC 60664-1 Protection class Class I device Device class in accordance with Directive
93/42/EEC
Class IIa equipment
Degree of protection against electric shock
Type B applied parts, except SIROCAM 3/SIROCAM C: Type BF applied parts
Degree of protection against ingress of water Ordinary equipment (not protected)
The foot switch has an IPX1 degree of protection against liquids (drip-proof).
Operating mode Continuous operation with intermittent loading corresponding to the
dental mode of working.
Permanently connected unit.
Year of manufacture ...............................................................
.........................(on rating plate of chair)
Dimensions of the packaging
Dentist element, assistant element, C8
+
Turn Cart dentist element
Water unit
Chair Connection box Upholstery SIROLUX / LEDview
120cm x 63 cm x 110cm
97.5cm x 50.5 cm x 50.5 cm 120cmx63cmx96cm 157cmx64cmx82cm 72cm x 53.2 cm x 44cm 120cmx52cmx40cm
10.5cm x 56 cm x 28cm
Weight (1 kg = 2.2 lbs) incl. / without packaging
Dentist element, assistant element, C8
+
Turn Cart dentist element
Water unit
Chair Connection box Upholstery SIROLUX / LEDview
37kg / 10 kg + 3kg kg 17kg / 13 kg 58 kg / 36kg 177kg / 140 kg 31kg / 25 kg 13kg / 9 kg 14kg / 10 kg
On-site pressure readings (1 bar = 100 kPa = 14.5 psi)
Air min./max.
Water min./max.
550 kPa / 750 kPa (5.5 / 7.5bar) 250 kPa / 600 kPa (2.5 / 6bar)
Page 34
Page 35
C8+/C8+Turn
60 28 000 D3417 D3417.021.04.11.02 35
3 Electromagnetic compatibility
3.1 Accessories ........................................................................................................................... 36
3.2 Electromagnetic emission ......................................................................................................37
3.3 Immunity to interference ........................................................................................................ 38
3.4 Working clearances ............................................................................................................... 40
NOTE
i
The C8+/C8+Turn fulfills all requirements for electromagnetic compatibility (EMC) compliant with IEC 60601-1-2.
The C8
+
/C8+Turn is referred to as "UNIT" in the sections that follow.
Observance of the following information is necessary to ensure safe operation regarding EMC aspects.
Page 36
3 Electromagnetic compatibility Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
36 D3417.021.04.11.02
The UNIT may only be operated with accessories and spare parts approved by Sirona. Unapproved accesso­ries and spare parts may lead to an increased emission or to a reduced immunity to interference.
The UNIT should not be operated in the immediate vi­cinity of other devices. If this proves to be unavoidable, the UNIT should be monitored to check and make sure that it is used properly.
The EMC measurements were performed with the following PCs:
3.1 Accessories
Making the PC connection
Designations of interface cables for PCs
Supplier
XGA cable, 10 m (L38) Sirona
S video cable, 10 m (L15) Sirona
RS232 cable, 10 m (L37) Sirona
Audio cable, 10m Sirona
2nd protective ground wire, 2.5 mm
2
,
10m
Sirona
PC as peripheral device for checking the interfaces with:
Siemens Fujitsu, Pentium IV,
3.0 GHz
Extension of the PC
Graphics card NVIDIA GeForce 7300 LE
Frame grabber card PicPort Color frame grabber
card (Leutron) REF 46 93 961
Page 37
Sirona Dental Systems GmbH 3 Electromagnetic compatibility Installation Requirements C8+/C8+Tu r n
60 28 000 D3417 D3417.021.04.11.02 37
English
The UNIT is intended for operation in the electromagnetic
environment specified below.
The customer or user of the UNIT should make sure that it
is used in such an environment.
3.2 Electromagnetic emission
Emission measurement Conformity Electromagnetic environment guidelines
RF emissions according to CISPR11Group 1 The UNIT uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions according to CISPR11Class B The UNIT is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public power supply providing electricity to buildings used for resi­dential purposes.Harmonics
according to IEC 61000-3-2
Class A
Voltage fluctuations/Flicker accord­ing to IEC 61000-3-3
Complies
Page 38
3 Electromagnetic compatibility Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
38 D3417.021.04.11.02
The UNIT is intended for operation in the electromagnetic environment specified below.
The customer or user of the UNIT should make sure that it is used in such an environment.
3.3 Immunity to interference
Interference immu­nity tests
IEC 60601-1-2 test level Compliance level Electromagnetic environment
guidelines
Electrostatic dis­charge (ESD) accord­ing to IEC 61000-4-2
± 6kV contact discharge
± 8 kV air discharge
± 6kV contact discharge
± 8kV air discharge
Floors should be made of wood or concrete or finished with ceramic til­ing. If floors are covered with syn­thetic material, the relative humidity should be at least 30%.
Electrical fast tran­sient/burst according to IEC
61000-4-4
± 1kV for input and output lines
± 2kV for power cables
± 1kV for input and output lines
± 2kV for power cables
The quality of the line power supply should be that of a typical commercial or hospital environment.
Surge voltages according to IEC
61000-4-5
± 1kV differential mode
± 2kV common mode
± 1kV differential mode
± 2kV common mode
The quality of the line power supply should be that of a typical commercial or hospital environment.
Voltage dips, short
interruptions and vari­ations of the power supply according to IEC
61000-4-11
<5%U
T
for ½ period (> 95
% dip of U
T
)
40% U
T
for 5 periods (60%
dip of U
T
)
70% U
T
for 25 periods (30%
dip of U
T
)
<5%U
T
for 5 sec. (> 95
% dip of U
T
)
<5%U
T
for ½ period (> 95
% dip of U
T
)
40% U
T
for 5 periods (60%
dip of U
T
)
70% U
T
for 25 periods
(30% dip of U
T
)
<5%U
T
for 5 sec. (> 95
% dip of U
T
)
The quality of the line power supply should be that of a typical commercial or hospital environment.
If the user of the UNIT requires it to continue functioning following inter­ruptions of the power supply, it is rec­ommended to have the UNIT powered by an uninterrupted power supply or a battery.
Magnetic field of power frequencies (50/60 Hz) according to IEC
61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a typical location in a typical commer­cial or hospital environment.
Remarks: U
T
is the AC supply voltage prior to application of the test level.
Page 39
Sirona Dental Systems GmbH 3 Electromagnetic compatibility Installation Requirements C8+/C8+Tu r n
60 28 000 D3417 D3417.021.04.11.02 39
English
Portable and mobile radio equipment must not be used within the recom­mended working clearance from the UNIT and its cables, which is calcu­lated based on the equation suitable for the relevant transmission fre­quency.
Recommended working clearance:
Conducted RF inter­ference
IEC 61000-4-6
3V
eff
150 kHz to 80 MHz
a
3V
eff
Radiated HF interfer­ence
IEC 61000-4-3
3V/m 80MHz to 800 MHz
a
3V/m 800MHz to 2.5 GHz
a
3V
eff
3V
eff
where P is the nominal transmitter output in watts (W) specified by the transmitter manufacturer and d is the recommended working clearance in meters (m).
Field strengths from fixed RF trans­mitters, as determined by an electro­magnetic site survey
b
, should be less
than the compliance level
c
in each
frequency range.
Interference is possible in the vicinity of equipment bearing the following graphic symbol.
a. The higher frequency range applies at 80MHz and 800 MHz.
b. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM/FM radio and TV broadcasts, cannot be predicted theoretically with accuracy. An investiga­tion of the location is recommended to determine the electromagnetic environment resulting from stationary RF trans­mitters. If the measured field strength in the location in which the UNIT is used exceeds the applicable RF compliance level above, the UNIT should be observed to verify normal operation. If unusual performance characteristics are observed, it may be necessary to take additional measures such as reorientation or repositioning of the UNIT.
c. Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
Interference immu­nity tests
IEC 60601-1-2 test level Compliance level Electromagnetic environment
guidelines
d 12P=
d 12P=
at 80MHz to 800 MHz
d 23P=
at 800MHz to 2.5 GHz
Page 40
3 Electromagnetic compatibility Sirona Dental Systems GmbH
Installation Requirements C8+/C8+Tu r n
60 28 000 D3417
40 D3417.021.04.11.02
Recommended working clearances between portable and mobile RF communication devices and the UNIT
The UNIT is intended for operation in an electromagnetic environment, where radiated RF interference is checked. The customer or the user of the
UNIT can help prevent
electromagnetic interference by duly observing the mini­mum distances between portable and/or mobile HF com­munication devices (transmitters) and the
UNIT. These
values may vary according to the output power of the rele­vant communication device as specified below.
For transmitters whose maximum nominal output is not specified in the above table, the recommended working clearance d in meters (m) can be determined using the equation in the corresponding column, where P is the max­imum nominal output of the transmitter in watts (W) speci­fied by the transmitter manufacturer.
Remark 1
The higher frequency range applies at 80 MHz and 800 MHz.
Remark 2
These guidelines may not be applicable in all cases. The propagation of electromagnetic waves is influenced by their absorption and reflection by buildings, objects and persons.
3.4 Working clearances
Nominal transmitter output [W]
Working clearance according to transmission frequency [m]
150kHz to 80 MHz 80MHz to 800 MHz 800MHz to 2.5GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
d 12P= d 12P= d 23P=
Page 41
Page 42
We reserve the right to make any alterations which may be required due to technical improvements.
© Sirona Dental Systems GmbH 2001-2014 Language: englisch Printed in Germany D3417.021.04.11.02 10.2014 Change No.: 119 074 Imprimé en Allemagne
Sirona Dental Systems GmbH
Fabrikstrasse 31 64625 Bensheim Germany www.sirona.com
Order no.
60 28 000 D3417