Sirona C8+ User manual

`UHI=`UH=qìêå
lйЙк~нбеЦ=fелнкмЕнбзел===
bеЦдблЬ
Sirona Dental Systems GmbH Contents Operating Instructions C8
, C8+ Tur n

Contents

1 Explanation of the symbols used ......................................................................... 6
2 Warning and safety information ........................................................................... 9
2.1 General warning and safety information .................................................................................... 9
2.2 Dentist element - rear and side delivery of instruments............................................................. 14
2.3 Satelec Mini L.E.D. curing light.................................................................................................. 15
2.4 SIROSONIC L............................................................................................................................ 17
2.5 SIROLUX E/SIROLUX FANTASTIC.......................................................................................... 17
2.6 Fresh water bottle...................................................................................................................... 17
3 Maintenance............................................................................................................ 18
3.1 Care and cleaning by the practice team .................................................................................... 18
3.2 Inspection and maintenance...................................................................................................... 18
3.3 Safety checks ............................................................................................................................ 19
4 Technical description ............................................................................................ 20
5 System versions..................................................................................................... 21
6 Putting the system into operation ........................................................................ 25
6.1 System ON/OFF switch ............................................................................................................. 25
7 Patient chair............................................................................................................ 26
7.1 Patient positioning ..................................................................................................................... 26
7.2 Adjusting the armrest................................................................................................................. 27
7.3 Flat headrest / articulated headrest ........................................................................................... 27
7.4 MultiMotion headrest ................................................................................................................. 28
7.5 Changing the patient position by turning (C8
7.6 Chair programs – factory settings.............................................................................................. 32
7.7 Chair programs – reprogramming.............................................................................................. 33
7.8 Foot switch – reprogramming .................................................................................................... 34
7.9 Emergency stop function for chair movement ........................................................................... 35
7.10 Patient chair, options ................................................................................................................. 36
+
Turn only).......................................................... 31
ЗЙмнлЕЬÑê~å´~áë bеЦдблЬÉëé~¥çä
59 42 896 D 3417 D 3417.201.09.02.02
3
Contents Sirona Dental Systems GmbH
Operating Instructions C8
, C8+ Tur n
8 Dentist element....................................................................................................... 37
8.1 Locking the dentist element ...................................................................................................... 37
8.2 Adjusting the instrument position .............................................................................................. 38
8.3 Swivel arm................................................................................................................................. 38
8.4 Dentist element C8
8.5 Dentist element C8
8.6 Assistant element C8
8.7 Starting an instrument ............................................................................................................... 43
8.8 Adjusting the driving air pressure.............................................................................................. 44
8.9 Instrument spray ON/OFF – chip blower................................................................................... 45
8.10 Instruments – adjusting the spray intensity – not C8
8.11 Instruments – adjusting the spray intensity – C8
+
Cart .......................................................................................................... 39
+
Side Delivery............................................................................................ 39
+
Rear Delivery........................................................................................ 41
+
Rear Delivery........................................ 46
+
Rear Delivery.............................................. 47
9 Dentist element options......................................................................................... 48
9.1 X-ray image viewer ................................................................................................................... 48
9.2 Tumbler holder.......................................................................................................................... 48
9.3 Silicone handles, right/left ......................................................................................................... 49
9.4 Tray........................................................................................................................................... 49
9.5 Silicone mat............................................................................................................................... 49
9.6 SPRAYVIT 3-way syringe ......................................................................................................... 50
9.7 3-way syringe ............................................................................................................................ 51
9.8 Satelec Mini L.E.D. curing light ................................................................................................. 52
9.9 SL/EL 1 electric motor............................................................................................................... 53
9.10 SIROSONIC L scaler ................................................................................................................ 55
9.11 SIROSONIC L endodontic scaler.............................................................................................. 56
9.12 Purge function – not for C8
9.13 Purge function - C8
+
+
Rear Delivery........................................................................................... 58
Rear Delivery............................................................................... 57
10 Assistant element................................................................................................... 59
10.1 Suction handpiece/saliva ejector .............................................................................................. 59
11 Assistant element options..................................................................................... 60
11.1 3-way syringes .......................................................................................................................... 60
11.2 Microswitch for vacuum system control .................................................................................... 60
11.3 Air jet pump............................................................................................................................... 60
11.4 Satelec Mini L.E.D. curing light ................................................................................................. 61
11.5 Control panel, assistant element............................................................................................... 61
4 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH Contents Operating Instructions C8
, C8+ Tur n
12 Water unit................................................................................................................ 62
12.1 Cuspidor flushing time – factory settings ................................................................................... 62
12.2 Cuspidor flushing time – reprogramming................................................................................... 63
12.3 Tumbler filling – factory settings ................................................................................................ 64
12.4 Tumbler filling – reprogramming ................................................................................................ 64
12.5 Combination of rinsing position S
with tumbler filling and flushing controlled by dentist element ................................................... 65
12.6 Control panel – assistant element – cuspidor flushing .............................................................. 67
12.7 Control panel – assistant element – cuspidor flushing – reprogramming.................................. 67
12.8 Control panel – assistant element – tumbler filling .................................................................... 68
12.9 Control panel – assistant element –
combination of mouth rinsing position S with tumbler filling and cuspidor flushing ................... 69
12.10 Fresh water bottle...................................................................................................................... 70
12.11 Tap water/fresh water bottle selector – C8
12.12 Changeover between tap water and fresh water bottle - C8+ Side Delivery.............................. 72
12.13 Changeover between tap water and fresh water bottle - C8
12.14 Hydrocolloid connection............................................................................................................. 74
12.15 Dürr wet suction system ............................................................................................................ 75
12.16 Amalgam separator ................................................................................................................... 76
12.17 Disinfection system.................................................................................................................... 77
12.18 Conversion of water unit right-handed Ù left-handed version .................................................. 78
+ .................................................................................................. 71
+
Rear Delivery ............................. 73
ЗЙмнлЕЬÑê~å´~áë bеЦдблЬÉëé~¥çä
13 SIROLUX E operation light – optional .................................................................. 81
13.1 ON/OFF – brightness reduction................................................................................................. 81
13.2 Focusing .................................................................................................................................... 82
14 SIROLUX FANTASTIC operation light – optional ................................................ 83
14.1 ON/OFF – brightness reduction................................................................................................. 83
14.2 Focusing .................................................................................................................................... 84
15 Care and cleaning by the practice team............................................................... 85
15.1 Care and cleaning agents.......................................................................................................... 85
15.2 Patient chair............................................................................................................................... 86
15.3 Dentist element.......................................................................................................................... 89
15.4 Assistant element ...................................................................................................................... 94
15.5 Water unit .................................................................................................................................. 96
15.6 Disinfection system – terminating sanitation.............................................................................. 104
15.7 SIROLUX E operation light ........................................................................................................ 107
15.8 SIROLUX FANTASTIC operation light ...................................................................................... 109
16 Options.................................................................................................................... 111
16.1 Working stool ............................................................................................................................. 111
16.2 Working stool C ......................................................................................................................... 112
59 42 896 D 3417 D 3417.201.09.02.02
5

1 Explanation of the symbols used Sirona Dental Systems GmbH

Operating Instructions C8
, C8+ Tur n
1 Explanation of the symbols used
Symbol Explanation Symbol Explanation
ON Patient chair up
OFF Patient chair down
CAUTION Patient chair upright
Fuse Patient chair inclined
T 10A H
Clean with water Factory settings
Warning of danger Activate chair program S, rinsing
position (Last memory function)
Emergency stop function Activate chair program 0,
entry/exit position
Wrong Activate chair program 1,
upper jaw treatment position
no/not OK Activate chair program 2,
lower jaw treatment position
Right Individual programming,
rinsing position S
Maximum load Individual programming,
entry/exit position 0
Wrong Individual programming,
upper jaw treatment 1
Individual programming, lower jaw treatment 2
59 42 896 D 3417
6 D 3417.201.09.02.02
Sirona Dental Systems GmbH 1 Explanation of the symbols used
+
1
2
0
40
60
80
ENDO
1
20
40
6
0
80
ENDO
1
20
40
6
0
80
ENDO
Operating Instructions C8
, C8+ Tur n
Symbol Explanation Symbol Explanation
>10s
<10m
Short acoustic signal Can be sterilized in the autoclave
135°C 275°F
at 135°C/275°F
Long acoustic signal Can be thermodisinfected in the
thermodisinfector at 93 °C/197 °F
Time in seconds 3-way syringe – water
Time in minutes 3-way syringe – air
Stop chair movement 3-way syringe – spray
Brake applied Can be set only by the service
engineer
bеЦдблЬ
Brake released Tumbler filler
Speed increase Cuspidor flushing
Speed reduction Press and hold key
Observe operating instructions of
Release key
the manufacturer
Spray cooling Press key briefly
Changing the intensity of spray air/spray water
SIROLUX E/ SIROLUX FANTASTIC
Maximum brightness
Increase/reduction of a variable SIROLUX E/
+ / –
SIROLUX FANTASTIC
Reduced brightness
59 42 896 D 3417 D 3417.201.09.02.02
7
1 Explanation of the symbols used Sirona Dental Systems GmbH
Operating Instructions C8
, C8+ Tur n
Symbol Explanation Symbol Explanation
Heat dissipation or
Approximately Amalgam recycling
Tap water CW rotation of electric micromo-
tor
After consulting hours CCW rotation of electric micro-
motor
Reprogramming
8 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information Operating Instructions C8

, C8+ Turn 2.1 General warning and safety information

2 Warning and safety information

2.1 General warning and safety information
Intended use
General safety information
Combination with other equipment
This dental treatment center is intended for diagnosis, therapy and den­tal treatment of humans by properly trained personnel.
This unit is not intended for operation in areas subject to explosion haz­ards.
The unit may only be operated by properly trained personnel.
The “on-site installation” must have been performed according to our requirements. As manufacturers of dental medical equipment and in the interest of the operational safety of your system, we stress the importance of having maintenance and repair of your treatment center performed only by ourselves or by agencies expressly authorized by us. Furthermore, safety-critical system components must always be replaced with origi- nal spare parts upon failure.
When having such work done, we suggest that you request stating the type and extent of work performed, including information about any modifications of the rated parameters or of the operating ranges (if applicable), as well as the date, name of organization and sig­nature. Modifications to this system which might affect the safety of the system owner, patients or other persons are prohibited by law!
For reasons of product safety, this product may be operated only with original Sirona accessories or third-party accessories expressly approved by Sirona. The user is responsible for any damage resulting from the use of non-approved accessories. If any equipment not approved by Sirona is connected, it must comply with the applicable standards:
EN 60950 for information technology equipment (e.g. PCs), and IEC 60601-1 for medical electrical equipment.
Any person who assembles or modifies a medical electrical system complying with the standard IEC 60601-1-1 (safety requirements for medical electrical equipment) by combining it with other equipment (e.g., by connecting it with a PC) is responsible for ensuring that the requirements of this regulation are met to their full extent for the safety of the patients, the operators and the environment.
In case of doubt, contact the manufacturer of the system compo­nents.
a certificate
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
9
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General warning and safety information Operating Instructions C8
, C8+ Tur n
Electromagnetic compatibility (EMC) The C8+, C8+ Turn complies with the requirements of
IEC 60601-1-2: 2001.
Medical electrical equipment is subject to special precautionary mea­sures regarding EMC. It must be installed and operated in compliance with the specifications provided in the document “Installation Require­ments” for the respective treatment center.
Portable and mobile RF communications equipment may interfere with medical electrical equipment. Therefore, the use of mobile wireless phones in practice or hospital environments must be prohibited.
With video systems:
Electromagnetic disturbances in the environment of the treatment cen­ter may result in reduced image quality. In such cases, it is advisable to repeat the exposure before saving the image.
When a line drop occurs, the image data stored in the video camera may be lost. Deposit the camera and repeat the exposure.
Image artifacts may possibly occur as a result of radio-frequency interference caused by the transmitters of e.g. radio services or radio amateurs.
Influencing of cardiac pacemakers
Identification of warning and safety informa­tion
Symbols used
Ventilation slots
Water/air supply quality
Water quality compliance
The treatment center contains strong magnets. These magnets can affect cardiac pacemakers implanted in patients, users and technical personnel.
Therefore, please observe the safety information on page 25.
To avoid personal injury and material damage please also observe the warning and safety information provided in the present operating instructions. All such information is highlighted by the signal words NOTE, CAUTION or WARNING.
Observe accompanying documents (on rating plate of chair)
Under no circumstances may the ventilation slots on the unit be cov­ered, since otherwise the air circulation will be obstructed.
Do not spray disinfectants or other similar products into the ventilation slots.
Air and water supply must meet the requirements specified in the instal­lation instructions.
Use only clean water and, wherever this is not possible, use the fresh water bottle (option).
To ensure compliance with the medical and national legal requirements for water from treatment centers, Sirona recommends equipping the treatment center with a disinfection system. As owner of the treatment center, you are responsible for the water quality and may have to take alternative measures to ensure its compliance if you operate the treat­ment center without a disinfection system.
Please contact your specialized dealer or your relevant dental associa­tion for the respective national requirements and measures.
DVGW approval
Treatment centers with a disinfection system have a DVGW approval. This approval of the treatment center expires if a media block is installed in the base of the chair (see page 36 and page 74).
10 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information Operating Instructions C8
, C8+ Turn 2.1 General warning and safety information
Vacuum system
Cuspidor
Electric micromotors in dentist element
Drive motors for chair and backrest
Maintenance of the treatment center
The suction removal of aluminum and other metal oxides from blasting devices via the automatic separator and the amalgam separator inte­grated in the treatment center is prohibited!
This would cause extreme wear and clogging of the suction and water paths.
A separate vacuum system must be used in connection with metal oxide blasting devices.
Treatment centers equipped with a central wet suction system are gen­erally suitable for suction removal of the above material. However, make sure to observe the instructions provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in connection with Sirona treatment centers. However, in such cases, make sure that the system is subsequently flushed with an adequate amount of water.
Do not use any foaming domestic cleaning agents!
Intermittent operation: 1 min. ON – 7 min. OFF with a motor current of
1.5 A.
6% duty time, cycle time 250s.
Despite the outstanding quality of your treatment center and regular care by the practice team, it is required in the interest of operational safety that preventive maintenance be performed at scheduled inter­vals.
bеЦдблЬ
In order to ensure the operational safety and reliability of your treatment center and to avoid damage due to natural wear, you as the system owner must have your system checked regularly by an authorized ser­vice engineer from your dental depot. Furthermore, safety checks must be performed.
Please contact your dental depot to obtain a maintenance offer.
For details please refer to chapter 3 “Maintenance”.
Dismantling and reassembly
Disposal Please observe the disposal regulations applicable in your country.
When dismantling and reassembling the system, proceed according to the installation instructions for new installation in order to guarantee its functioning and stability.
Within the European Economic Area, this product is subject to Direc­tive 2002/96/EC as well as the corresponding national laws. This directive requires environmentally sound recycling/disposal of the product.
The product must not be disposed of as domestic refuse!
Please contact your dealer if final disposal of your product is required.
59 42 896 D 3417 D 3417.201.09.02.02
11
2 Warning and safety information Sirona Dental Systems GmbH
STOP
STOP
STOP
STOP
2.1 General warning and safety information Operating Instructions C8
, C8+ Tur n
Maximum load capacity of the treatment chair
Please observe the maximum load capac­ity of the patient chair of 135kg according to EN ISO 6875 (tested with a four-fold safety factor according to EN 60601-1).
Safety stop for patient chair
1.
Headrest
Backrest
Toeboard
right
2.
Hand/foot placement
wrong
ATTENTION
Make sure that no obstacles (e.g. window wings, drawers, devices, …) extend into the movement range.
12 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information
STOP
STOP
Operating Instructions C8
, C8+ Turn 2.1 General warning and safety information
Safety stop for assistant element
Safety stop for cuspidor
C8+ Turn
+
“Long cuspidor”
C8
Position of the dentist element
bеЦдблЬ
ATTENTION
When moving the chair (via program or manually), make sure that the patient does not under any circumstances come into contact with the support arm or the dentist element.
59 42 896 D 3417 D 3417.201.09.02.02
13
2 Warning and safety information Sirona Dental Systems GmbH

2.2 Dentist element - rear and side delivery of instruments Operating Instructions C8

, C8+ Tur n
2.2 Dentist element - rear and side delivery of instruments
Side delivery of instruments
max.
2 kg
4.4 lbs
To ensure its mechanical stability, install the dentist element only on cabinets which have been approved by Sirona.
Rear instrument delivery
To ensure its mechanical stability, install the dentist element only on cabinets which have been approved by Sirona.
ATTENTION
Never place any objects on this dentist el­ement.
Position of dentist element: rear and side delivery of instru­ments
ATTENTION
When moving the chair (via program or manually), make sure that the patient does not under any circumstances come into contact with the support arm or the dentist element.
59 42 896 D 3417
14 D 3417.201.09.02.02
Sirona Dental Systems GmbH 2 Warning and safety information Operating Instructions C8

, C8+ Turn 2.3 Satelec Mini L.E.D. curing light

2.3 Satelec Mini L.E.D. curing light
ATTENTION
Always use the glare shield! The light beams emitted by this instrument may be hazardous and may never be di­rected to the eyes of any person, even if this person wears goggles with protective lenses. The light may be directed exclu­sively to the treatment area in the oral cav­ity.
bеЦдблЬ
right
L.E.D. RADIATION
DO NOT STARE INTO BEAM OR VIEW
DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 2M LASER PRODUCT
P < 1 mW, λ = 420 nm – 480 nm
IEC 60825-1 Ed. 1.2
!
wrong
L.E.D. beam exit aperture
!
ATTENTION
Do not stare into the beam path with the glass rod removed (Class 2M laser prod­uct). Viewing the LED light exit aperture with certain optical instruments (e.g. a magni­fying lens) within a distance of 100 mm may constitute an eye hazard.
Optical specification
Wavelength λ: = 420 nm – 480 nm
L.E.D. laser power P: < 1 mW
ATTENTION
Heavy line voltage drops may lead to a shutdown of the Mini L.E.D. The fault is signaled by a double beep. Check the se­lected operating mode and restart the Mini L.E.D.
59 42 896 D 3417 D 3417.201.09.02.02
15
2 Warning and safety information Sirona Dental Systems GmbH
2.3 Satelec Mini L.E.D. curing light Operating Instructions C8
, C8+ Tur n
Additional safety information
Any condensation occurring in the hand­piece of the Mini L.E.D. may cause impairments (e.g. fogging of the L.E.D.). When moving the handpiece from a cool environment to a warm room, always wait for it to reach room temperature before putting it into operation.
Curing lights must not be used on persons who are suffering from or have in the past been afflicted by photobiological reactions (including solar urticaria and erythropoetic porphyria). Nor should they be used on persons currently being treated with any medicine which increases one’s sensitiv­ity to light (including methoxsalene and chlorotetracycline).
Any persons who have in the past suf­fered from retinal or eye lens disorders or have undergone eye surgery, especially for gray cataracts, must consult their med­ical eye specialist before using the light or undergoing treatment with the Mini L.E.D. Caution is advisable even if the patient gives his or her consent, since the light intensity can cause accidents. It is espe­cially advisable to always wear appropri­ate protective goggles (UV filter).
16 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information
LEMON
LEMON
0
P
E ND
O
1 0
0
1
20
4 0
6
0
8 0
Operating Instructions C8

, C8+ Tur n 2.4 SIROSONIC L

2.4 SIROSONIC L
wrong
!

2.5 SIROLUX E/SIROLUX FANTASTIC

right
wrong
right
!
bеЦдблЬ
ATTENTION
Never operate without cover: Watch out for overheating of lamp compo­nents and let them cool down.
SIROLUX FANTASTIC
ATTENTION
Pay attention to elevated temperature of the lamp components; and let these cool down.

2.6 Fresh water bottle

LEMON
wrong
Fresh water bottle
ATTENTION
Use only Sirona fresh water bot­tles!
LEMON
59 42 896 D 3417 D 3417.201.09.02.02
17

3 Maintenance Sirona Dental Systems GmbH

3.1 Care and cleaning by the practice team Operating Instructions C8

, C8+ Tur n
3 Maintenance
Despite the outstanding quality of your treatment center, it is necessary to per­form inspections and preventive mainte­nance as well as safety tests at scheduled intervals in order to ensure its operational safety.
These events are documented in the Maintenance Manual.
Care and cleaning is done by the practice team.
For details please refer to the following sections.
3.1 Care and cleaning by the practice team
The practice team is responsible for regu­lar cleaning and care of the treatment center, with the aim of minimizing the risk of contamination for patients and users.
The work to be performed is specified in the chapter "Care and cleaning by the practice team".
The surfaces (apart from upholstery) can be sprayed and wiped with surface disin­fectants.

3.2 Inspection and maintenance

In order to ensure the operational safety and reliability of your treatment center and to avoid damage due to natural wear, annual inspection and maintenance must be performed on your treatment center. This is done by an authorized service engineer from your dental depot.
The steps to be performed as well as the parts which must be replaced are speci­fied in the document "Maintenance Cer- tificate".
An overview of the inspection and mainte­nance work performed is additionally recorded by the service engineer in the "Installation Report/Warranty Pass-
port". This document is part of the "Main­tenance Manual".
18 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 3 Maintenance Operating Instructions C8

, C8+ Tur n 3.3 Safety checks

3.3 Safety checks
Medical equipment is designed in such a way that the first occurrence of a fault does not create a hazard to the safety of patients, users or other persons. Hence it is important to detect such faults before a second fault occurs, which might then lead to safety hazards.
For that reason it is essential to perform safety tests every 2 years which aim par­ticularly at detecting electrical faults (e.g. isolation defects). This is done by an authorized service engineer from your dental depot, most practically together with the work to be performed according to section 3.2.
The check includes a visual inspection as well as measurements of the protective ground wire connections and the equiva­lent leakage currents.
The inspections and measurements to be performed are specified in the "Mainte- nance Manual". The measured values must be documented there by the service engineer.
Safety tests must also be performed and documented during initial startup, after extensions/upgrades (conversion) of your treatment center and after repair
work which might affect the electrical safety of the system.
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
i
NOTE
The treatment center must not be operat­ed if it has failed to pass the safety tests!
19

4 Technical description Sirona Dental Systems GmbH

3.3 Safety checks Operating Instructions C8
, C8+ Tur n
4 Technical description
Model designation
Power supply
Nominal current
Main unit fuse
Operating conditions
Transport and storage conditions
Protection class
Degree of protection against electric shock
Degree of protection against ingress of water
Mode of operation
+
C8
, C8+ Tu r n ,
100 V/115 V/127 V/230 V AC 50/60 Hz
11.5 A at 100 V AC 10 A at 115 V AC 9 A at 127 V AC 5 A at 230 V AC
T 10 A H, 250 V AC, Order No. 10 77 460
Ambient temperature: 10°C – 40°C (50°F – 104°F) Relative humidity: 30% – 75% Air pressure: 700hPa – 1060hPa
Temperature: -40°C – +70°C (-40°F – +158°F) relative humidity: 10% – 95% Air pressure: 500hPa – 1060hPa
Class I device
Type B applied parts, except SIROCAM 3/SIROCAM C: Type BF applied parts
Ordinary equipment (not protected) The foot switch has an IPX1 degree of protection against liquids (drip-proof).
Continuous operation with intermittent loading corresponding to the dental mode of working.
Permanently connected unit.
Year of manufacture
(on rating plate of chair)
Tests/approvals This dental treatment center complies with the requirements of
IEC 60601-1: 1988 + A1: 1991 + A2: 1995 and of IEC 60601-1-2: 2001.
DVGW: This system complies with the technical rules and require­ments on safety and hygiene for connection to the drinking water supply (if the appropriate DVGW marking is affixed to it).
This product bears the CE mark in accordance with the provisions of Council Directive 93/42/EEC of June 14, 1993 concerning medi­cal devices.
0123
20 D 3417.201.09.02.02
59 42 896 D 3417

Sirona Dental Systems GmbH 5 System versions

P 1
P 2
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
2
0
4 0
6 0
80
E N
D O
P 1
P2
P
2
0
1
S
C
8
+
1
2
S
P
0
100
1
2
0
4 0
6 0
8
0
E N
D O
Operating Instructions C8
, C8+ Tur n
5 System versions
C8+ OTP
1. Patient chair/Patient chair Turn
2. OTP dentist element
5.
2.
4.
3.
1.
3. Water unit
4. Assistant element/Assistant element
mounted at the middle of the chair
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
1
1
bеЦдблЬ
6.
7.
C8+ with swivel arm
1. Patient chair/Patient chair Turn
2. Dentist element with swivel arm
5.
2.
4.
3.
1.
3. Water unit
4. Assistant element/Assistant element
mounted at the middle of the chair
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
1
1
6.
7.
1. Alternative equipment, not illustrated
59 42 896 D 3417 D 3417.201.09.02.02
21
5 System versions Sirona Dental Systems GmbH
1
0
0
1
2
0
4
0
60
80
P
1
2
S
0
E
N
D
O
P 1
P 2
Operating Instructions C8
, C8+ Tur n
C8+ Cart
1. Patient chair/Patient chair Turn
2. Dentist element
5.
3. Water unit
1
4. Assistant element/Assistant element
4.
3.
2.
1.
mounted at the middle of the chair
5. SIROLUX
1
6. Instrument foot switch with program
positions
7. Instrument foot switch
6.
7.
5.
2.
C8+ Side Delivery
1. Patient chair /
Patient chair, Turn
2. Dentist element
1
3. Water unit
1.
4. Assistant element
3.
4.
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
7.
6.
1. Alternative equipment, not illustrated
22 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 5 System versions
0
P
1
2
S
E N D
O
1 0 0
1
2
0
4 0
6 0
8
0
P
2
0
1
S
C 8
+
1
2
S
P
0
10
0
1
2 0
4
0
6 0
8 0
E N D
O
Operating Instructions C8
, C8+ Tur n
3.
2.
1.
C8+ Rear Delivery
1. Patient chair /
Patient chair, Turn
2. Dentist element
1
3. Instrument holder
4. Instrument foot switch with program
bеЦдблЬ
positions
5. Instrument foot switch
4.
5.
C8+ Turn
1. Patient chair Turn
5.
2.
4.
3.
1.
2. OTP dentist element/Dentist element
with swivel arm Cart dentist element Cabinet dentist element
2
1
1
3. Water unit
4. Assistant element/Assistant element
mounted at the middle of the chair
1
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
6.
P 1
P2
7.
8.
8. Swivel arm for right-handedÙleft-
handed conversion
1. Alternative equipment, not illustrated
2. Alternative equipment, not illustrated
59 42 896 D 3417 D 3417.201.09.02.02
23
5 System versions Sirona Dental Systems GmbH
Operating Instructions C8
, C8+ Tur n
C8+ without dentist element
1. Patient chair/Patient chair Turn
2. Water unit
4.
3.
1.
2.
3. Assistant element/Assistant element
mounted at the middle of the chair
4. SIROLUX
i
NOTE
1
1
For operation of the SIVISION 3 video system, please refer to Operating In­structions 59 72 257.
24 D 3417.201.09.02.02
59 42 896 D 3417

Sirona Dental Systems GmbH 6 Putting the system into operation

1
2
S
P
0
100
1
20
40
60
80
ENDO
Operating Instructions C8

, C8+ Tur n 6.1 System ON/OFF switch

6 Putting the system into operation
6.1 System ON/OFF switch
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
25

7 Patient chair Sirona Dental Systems GmbH

0
P
1
2
S
E N D O
100
1
2
0
4 0
6 0
8
0
P
2
0
1
S
C8
+
1
2
S
P
0
1 0
0
1
2
0
4 0
60
8 0
E N D
O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
END
O
S
1
0
2
P
2
0
1
S
C8
+
P
S
1
0
2
S
1
0
2

7.1 Patient positioning Operating Instructions C8

, C8+ Tur n
7 Patient chair
7.1 Patient positioning
1. 2.
Option
3.
4.
Option
Option
0
S
2
1
26 D 3417.201.09.02.02
P
59 42 896 D 3417
Sirona Dental Systems GmbH 7 Patient chair Operating Instructions C8

, C8+ Tur n 7.2 Adjusting the armrest

7.2 Adjusting the armrest
A Folding armrest (optional)
C
B
A

7.3 Flat headrest / articulated headrest

B Wide backrest
C Narrow backrest
ATTENTION
Risk of collision! The following versions must not be retrofit­ted with an armrest:
+
with fixed water unit in combination
• C8 with long cuspidor
A Flat headrest
bеЦдблЬ
B Articulated headrest (optional)
A
1
ATTENTION
The head pad (A) contains a strong mag-
B
net (1) on its bottom side. This magnet could affect any cardiac pacemaker locat­ed nearby.
Therefore, do not allow patients, users and technical personnel with a cardiac pacemaker to be located near this mag­net.
Furthermore, direct contact of the head pad with magnetic cards can lead to data loss.
59 42 896 D 3417 D 3417.201.09.02.02
27
7 Patient chair Sirona Dental Systems GmbH
A

7.4 MultiMotion headrest Operating Instructions C8

, C8+ Tur n
7.4 MultiMotion headrest
Adjustment to patient's height
The headrest can be adjusted to the patient's height by pulling out or pushing in the headrest extension. This adjustment is usually required at the beginning of treat­ment!
i
NOTE
Before positioning the patient on the headrest, make sure that the head support has been optimally adjusted to the patient's height. This simplifies all subsequent work with the MultiMotion considerably, since readjust­ment to the patient's height can thus be omitted when changing over from mandibular to maxillary treatment.
The MultiMotion enables you to adjust the patient's head in a way that optimally supports viewing of areas of the mouth which are difficult to access.
Adjustment of hyperextension
Mandibular position
The mandibular position can be set by pulling the oper­ating handle (A). The anatomical movement of the arched extension keeps the patient's head in the sup­port.
¾ Pull the headrest out of the guide by pulling
handle A.
i
NOTE
You can reduce the adjustment noise by pressing release A.
28 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 7 Patient chair
A
B
B
Operating Instructions C8
, C8+ Tur n 7.4 MultiMotion headrest
Maxillary position
¾ Take the load off the headrest by lifting it slightly.
¾ Press release A on the handle.
¾ Let the headrest slide down slowly.
¾ Let go of the release when the headrest reaches the
desired position.
bеЦдблЬ
Rotation and tilt
The MultiMotion enables rotation of the patient's head about the longitudinal axis of his body as well as lateral tilting of his head.
The MultiMotion is set to the desired position as follows:
1. Press and hold one or both of the two side control
elements B. The rotation and tilt adjustment locks of the headrest are now released.
2. Set the headrest to the desired position.
3. Let go of the control elements B.
; The supporting surface is fixed in place.
Make sure that the headrest is indeed fixed in place after you let go of the control elements!
59 42 896 D 3417 D 3417.201.09.02.02
29
7 Patient chair Sirona Dental Systems GmbH
C
E
D
F
7.4 MultiMotion headrest Operating Instructions C8
, C8+ Tur n
Removing the MultiMotion
For certain treatments (e.g. of children) it may be expe­dient to remove the MultiMotion completely in order to obtain better access to the patient. The patient's head will then be supported on the contact surface of the headrest extension F.
To remove the MultiMotion, proceed as follows:
1. Press release button C.
2. Pull the complete headrest out of the headrest ex-
tension.
3. Cover the opening of the headrest extension with
cover cap E.
4. If a children's head pad D is available, place it on
the headrest extension. The head pad is held mag­netically.
i
NOTE
The children's head pad (D) is not included in the scope of supply, however, is available from your dental dealer.
WARNING
The children's head pad (D) contains a strong mag­net on its bottom side. This magnet could affect any cardiac pacemaker located nearby.
Therefore, do not allow patients, users and technical personnel with a cardiac pacemaker to be located near this magnet.
Furthermore, direct contact of the head pad with magnetic cards can lead to data loss.
i
NOTE
Deposit the removed MultiMotion in a safe place where it cannot fall onto the floor.
Inserting the MultiMotion
To insert the MultiMotion, proceed as follows:
¾ Check the guide of the removed headrest for con-
tamination. Remove any contamination.
¾ Remove the cover cap E.
Reinsert the MultiMotion in the guide from above until it audibly locks in place.
Pull on the headrest again to make sure that it is locked securely in place.
30 D 3417.201.09.02.02