Sirona C8+ User manual

`UHI=`UH=qìêå
lйЙк~нбеЦ=fелнкмЕнбзел===
bеЦдблЬ
Sirona Dental Systems GmbH Contents Operating Instructions C8
, C8+ Tur n

Contents

1 Explanation of the symbols used ......................................................................... 6
2 Warning and safety information ........................................................................... 9
2.1 General warning and safety information .................................................................................... 9
2.2 Dentist element - rear and side delivery of instruments............................................................. 14
2.3 Satelec Mini L.E.D. curing light.................................................................................................. 15
2.4 SIROSONIC L............................................................................................................................ 17
2.5 SIROLUX E/SIROLUX FANTASTIC.......................................................................................... 17
2.6 Fresh water bottle...................................................................................................................... 17
3 Maintenance............................................................................................................ 18
3.1 Care and cleaning by the practice team .................................................................................... 18
3.2 Inspection and maintenance...................................................................................................... 18
3.3 Safety checks ............................................................................................................................ 19
4 Technical description ............................................................................................ 20
5 System versions..................................................................................................... 21
6 Putting the system into operation ........................................................................ 25
6.1 System ON/OFF switch ............................................................................................................. 25
7 Patient chair............................................................................................................ 26
7.1 Patient positioning ..................................................................................................................... 26
7.2 Adjusting the armrest................................................................................................................. 27
7.3 Flat headrest / articulated headrest ........................................................................................... 27
7.4 MultiMotion headrest ................................................................................................................. 28
7.5 Changing the patient position by turning (C8
7.6 Chair programs – factory settings.............................................................................................. 32
7.7 Chair programs – reprogramming.............................................................................................. 33
7.8 Foot switch – reprogramming .................................................................................................... 34
7.9 Emergency stop function for chair movement ........................................................................... 35
7.10 Patient chair, options ................................................................................................................. 36
+
Turn only).......................................................... 31
ЗЙмнлЕЬÑê~å´~áë bеЦдблЬÉëé~¥çä
59 42 896 D 3417 D 3417.201.09.02.02
3
Contents Sirona Dental Systems GmbH
Operating Instructions C8
, C8+ Tur n
8 Dentist element....................................................................................................... 37
8.1 Locking the dentist element ...................................................................................................... 37
8.2 Adjusting the instrument position .............................................................................................. 38
8.3 Swivel arm................................................................................................................................. 38
8.4 Dentist element C8
8.5 Dentist element C8
8.6 Assistant element C8
8.7 Starting an instrument ............................................................................................................... 43
8.8 Adjusting the driving air pressure.............................................................................................. 44
8.9 Instrument spray ON/OFF – chip blower................................................................................... 45
8.10 Instruments – adjusting the spray intensity – not C8
8.11 Instruments – adjusting the spray intensity – C8
+
Cart .......................................................................................................... 39
+
Side Delivery............................................................................................ 39
+
Rear Delivery........................................................................................ 41
+
Rear Delivery........................................ 46
+
Rear Delivery.............................................. 47
9 Dentist element options......................................................................................... 48
9.1 X-ray image viewer ................................................................................................................... 48
9.2 Tumbler holder.......................................................................................................................... 48
9.3 Silicone handles, right/left ......................................................................................................... 49
9.4 Tray........................................................................................................................................... 49
9.5 Silicone mat............................................................................................................................... 49
9.6 SPRAYVIT 3-way syringe ......................................................................................................... 50
9.7 3-way syringe ............................................................................................................................ 51
9.8 Satelec Mini L.E.D. curing light ................................................................................................. 52
9.9 SL/EL 1 electric motor............................................................................................................... 53
9.10 SIROSONIC L scaler ................................................................................................................ 55
9.11 SIROSONIC L endodontic scaler.............................................................................................. 56
9.12 Purge function – not for C8
9.13 Purge function - C8
+
+
Rear Delivery........................................................................................... 58
Rear Delivery............................................................................... 57
10 Assistant element................................................................................................... 59
10.1 Suction handpiece/saliva ejector .............................................................................................. 59
11 Assistant element options..................................................................................... 60
11.1 3-way syringes .......................................................................................................................... 60
11.2 Microswitch for vacuum system control .................................................................................... 60
11.3 Air jet pump............................................................................................................................... 60
11.4 Satelec Mini L.E.D. curing light ................................................................................................. 61
11.5 Control panel, assistant element............................................................................................... 61
4 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH Contents Operating Instructions C8
, C8+ Tur n
12 Water unit................................................................................................................ 62
12.1 Cuspidor flushing time – factory settings ................................................................................... 62
12.2 Cuspidor flushing time – reprogramming................................................................................... 63
12.3 Tumbler filling – factory settings ................................................................................................ 64
12.4 Tumbler filling – reprogramming ................................................................................................ 64
12.5 Combination of rinsing position S
with tumbler filling and flushing controlled by dentist element ................................................... 65
12.6 Control panel – assistant element – cuspidor flushing .............................................................. 67
12.7 Control panel – assistant element – cuspidor flushing – reprogramming.................................. 67
12.8 Control panel – assistant element – tumbler filling .................................................................... 68
12.9 Control panel – assistant element –
combination of mouth rinsing position S with tumbler filling and cuspidor flushing ................... 69
12.10 Fresh water bottle...................................................................................................................... 70
12.11 Tap water/fresh water bottle selector – C8
12.12 Changeover between tap water and fresh water bottle - C8+ Side Delivery.............................. 72
12.13 Changeover between tap water and fresh water bottle - C8
12.14 Hydrocolloid connection............................................................................................................. 74
12.15 Dürr wet suction system ............................................................................................................ 75
12.16 Amalgam separator ................................................................................................................... 76
12.17 Disinfection system.................................................................................................................... 77
12.18 Conversion of water unit right-handed Ù left-handed version .................................................. 78
+ .................................................................................................. 71
+
Rear Delivery ............................. 73
ЗЙмнлЕЬÑê~å´~áë bеЦдблЬÉëé~¥çä
13 SIROLUX E operation light – optional .................................................................. 81
13.1 ON/OFF – brightness reduction................................................................................................. 81
13.2 Focusing .................................................................................................................................... 82
14 SIROLUX FANTASTIC operation light – optional ................................................ 83
14.1 ON/OFF – brightness reduction................................................................................................. 83
14.2 Focusing .................................................................................................................................... 84
15 Care and cleaning by the practice team............................................................... 85
15.1 Care and cleaning agents.......................................................................................................... 85
15.2 Patient chair............................................................................................................................... 86
15.3 Dentist element.......................................................................................................................... 89
15.4 Assistant element ...................................................................................................................... 94
15.5 Water unit .................................................................................................................................. 96
15.6 Disinfection system – terminating sanitation.............................................................................. 104
15.7 SIROLUX E operation light ........................................................................................................ 107
15.8 SIROLUX FANTASTIC operation light ...................................................................................... 109
16 Options.................................................................................................................... 111
16.1 Working stool ............................................................................................................................. 111
16.2 Working stool C ......................................................................................................................... 112
59 42 896 D 3417 D 3417.201.09.02.02
5

1 Explanation of the symbols used Sirona Dental Systems GmbH

Operating Instructions C8
, C8+ Tur n
1 Explanation of the symbols used
Symbol Explanation Symbol Explanation
ON Patient chair up
OFF Patient chair down
CAUTION Patient chair upright
Fuse Patient chair inclined
T 10A H
Clean with water Factory settings
Warning of danger Activate chair program S, rinsing
position (Last memory function)
Emergency stop function Activate chair program 0,
entry/exit position
Wrong Activate chair program 1,
upper jaw treatment position
no/not OK Activate chair program 2,
lower jaw treatment position
Right Individual programming,
rinsing position S
Maximum load Individual programming,
entry/exit position 0
Wrong Individual programming,
upper jaw treatment 1
Individual programming, lower jaw treatment 2
59 42 896 D 3417
6 D 3417.201.09.02.02
Sirona Dental Systems GmbH 1 Explanation of the symbols used
+
1
2
0
40
60
80
ENDO
1
20
40
6
0
80
ENDO
1
20
40
6
0
80
ENDO
Operating Instructions C8
, C8+ Tur n
Symbol Explanation Symbol Explanation
>10s
<10m
Short acoustic signal Can be sterilized in the autoclave
135°C 275°F
at 135°C/275°F
Long acoustic signal Can be thermodisinfected in the
thermodisinfector at 93 °C/197 °F
Time in seconds 3-way syringe – water
Time in minutes 3-way syringe – air
Stop chair movement 3-way syringe – spray
Brake applied Can be set only by the service
engineer
bеЦдблЬ
Brake released Tumbler filler
Speed increase Cuspidor flushing
Speed reduction Press and hold key
Observe operating instructions of
Release key
the manufacturer
Spray cooling Press key briefly
Changing the intensity of spray air/spray water
SIROLUX E/ SIROLUX FANTASTIC
Maximum brightness
Increase/reduction of a variable SIROLUX E/
+ / –
SIROLUX FANTASTIC
Reduced brightness
59 42 896 D 3417 D 3417.201.09.02.02
7
1 Explanation of the symbols used Sirona Dental Systems GmbH
Operating Instructions C8
, C8+ Tur n
Symbol Explanation Symbol Explanation
Heat dissipation or
Approximately Amalgam recycling
Tap water CW rotation of electric micromo-
tor
After consulting hours CCW rotation of electric micro-
motor
Reprogramming
8 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information Operating Instructions C8

, C8+ Turn 2.1 General warning and safety information

2 Warning and safety information

2.1 General warning and safety information
Intended use
General safety information
Combination with other equipment
This dental treatment center is intended for diagnosis, therapy and den­tal treatment of humans by properly trained personnel.
This unit is not intended for operation in areas subject to explosion haz­ards.
The unit may only be operated by properly trained personnel.
The “on-site installation” must have been performed according to our requirements. As manufacturers of dental medical equipment and in the interest of the operational safety of your system, we stress the importance of having maintenance and repair of your treatment center performed only by ourselves or by agencies expressly authorized by us. Furthermore, safety-critical system components must always be replaced with origi- nal spare parts upon failure.
When having such work done, we suggest that you request stating the type and extent of work performed, including information about any modifications of the rated parameters or of the operating ranges (if applicable), as well as the date, name of organization and sig­nature. Modifications to this system which might affect the safety of the system owner, patients or other persons are prohibited by law!
For reasons of product safety, this product may be operated only with original Sirona accessories or third-party accessories expressly approved by Sirona. The user is responsible for any damage resulting from the use of non-approved accessories. If any equipment not approved by Sirona is connected, it must comply with the applicable standards:
EN 60950 for information technology equipment (e.g. PCs), and IEC 60601-1 for medical electrical equipment.
Any person who assembles or modifies a medical electrical system complying with the standard IEC 60601-1-1 (safety requirements for medical electrical equipment) by combining it with other equipment (e.g., by connecting it with a PC) is responsible for ensuring that the requirements of this regulation are met to their full extent for the safety of the patients, the operators and the environment.
In case of doubt, contact the manufacturer of the system compo­nents.
a certificate
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
9
2 Warning and safety information Sirona Dental Systems GmbH
2.1 General warning and safety information Operating Instructions C8
, C8+ Tur n
Electromagnetic compatibility (EMC) The C8+, C8+ Turn complies with the requirements of
IEC 60601-1-2: 2001.
Medical electrical equipment is subject to special precautionary mea­sures regarding EMC. It must be installed and operated in compliance with the specifications provided in the document “Installation Require­ments” for the respective treatment center.
Portable and mobile RF communications equipment may interfere with medical electrical equipment. Therefore, the use of mobile wireless phones in practice or hospital environments must be prohibited.
With video systems:
Electromagnetic disturbances in the environment of the treatment cen­ter may result in reduced image quality. In such cases, it is advisable to repeat the exposure before saving the image.
When a line drop occurs, the image data stored in the video camera may be lost. Deposit the camera and repeat the exposure.
Image artifacts may possibly occur as a result of radio-frequency interference caused by the transmitters of e.g. radio services or radio amateurs.
Influencing of cardiac pacemakers
Identification of warning and safety informa­tion
Symbols used
Ventilation slots
Water/air supply quality
Water quality compliance
The treatment center contains strong magnets. These magnets can affect cardiac pacemakers implanted in patients, users and technical personnel.
Therefore, please observe the safety information on page 25.
To avoid personal injury and material damage please also observe the warning and safety information provided in the present operating instructions. All such information is highlighted by the signal words NOTE, CAUTION or WARNING.
Observe accompanying documents (on rating plate of chair)
Under no circumstances may the ventilation slots on the unit be cov­ered, since otherwise the air circulation will be obstructed.
Do not spray disinfectants or other similar products into the ventilation slots.
Air and water supply must meet the requirements specified in the instal­lation instructions.
Use only clean water and, wherever this is not possible, use the fresh water bottle (option).
To ensure compliance with the medical and national legal requirements for water from treatment centers, Sirona recommends equipping the treatment center with a disinfection system. As owner of the treatment center, you are responsible for the water quality and may have to take alternative measures to ensure its compliance if you operate the treat­ment center without a disinfection system.
Please contact your specialized dealer or your relevant dental associa­tion for the respective national requirements and measures.
DVGW approval
Treatment centers with a disinfection system have a DVGW approval. This approval of the treatment center expires if a media block is installed in the base of the chair (see page 36 and page 74).
10 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information Operating Instructions C8
, C8+ Turn 2.1 General warning and safety information
Vacuum system
Cuspidor
Electric micromotors in dentist element
Drive motors for chair and backrest
Maintenance of the treatment center
The suction removal of aluminum and other metal oxides from blasting devices via the automatic separator and the amalgam separator inte­grated in the treatment center is prohibited!
This would cause extreme wear and clogging of the suction and water paths.
A separate vacuum system must be used in connection with metal oxide blasting devices.
Treatment centers equipped with a central wet suction system are gen­erally suitable for suction removal of the above material. However, make sure to observe the instructions provided by the manufacturer of your vacuum system.
No restrictions apply when using salt blasting devices in connection with Sirona treatment centers. However, in such cases, make sure that the system is subsequently flushed with an adequate amount of water.
Do not use any foaming domestic cleaning agents!
Intermittent operation: 1 min. ON – 7 min. OFF with a motor current of
1.5 A.
6% duty time, cycle time 250s.
Despite the outstanding quality of your treatment center and regular care by the practice team, it is required in the interest of operational safety that preventive maintenance be performed at scheduled inter­vals.
bеЦдблЬ
In order to ensure the operational safety and reliability of your treatment center and to avoid damage due to natural wear, you as the system owner must have your system checked regularly by an authorized ser­vice engineer from your dental depot. Furthermore, safety checks must be performed.
Please contact your dental depot to obtain a maintenance offer.
For details please refer to chapter 3 “Maintenance”.
Dismantling and reassembly
Disposal Please observe the disposal regulations applicable in your country.
When dismantling and reassembling the system, proceed according to the installation instructions for new installation in order to guarantee its functioning and stability.
Within the European Economic Area, this product is subject to Direc­tive 2002/96/EC as well as the corresponding national laws. This directive requires environmentally sound recycling/disposal of the product.
The product must not be disposed of as domestic refuse!
Please contact your dealer if final disposal of your product is required.
59 42 896 D 3417 D 3417.201.09.02.02
11
2 Warning and safety information Sirona Dental Systems GmbH
STOP
STOP
STOP
STOP
2.1 General warning and safety information Operating Instructions C8
, C8+ Tur n
Maximum load capacity of the treatment chair
Please observe the maximum load capac­ity of the patient chair of 135kg according to EN ISO 6875 (tested with a four-fold safety factor according to EN 60601-1).
Safety stop for patient chair
1.
Headrest
Backrest
Toeboard
right
2.
Hand/foot placement
wrong
ATTENTION
Make sure that no obstacles (e.g. window wings, drawers, devices, …) extend into the movement range.
12 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information
STOP
STOP
Operating Instructions C8
, C8+ Turn 2.1 General warning and safety information
Safety stop for assistant element
Safety stop for cuspidor
C8+ Turn
+
“Long cuspidor”
C8
Position of the dentist element
bеЦдблЬ
ATTENTION
When moving the chair (via program or manually), make sure that the patient does not under any circumstances come into contact with the support arm or the dentist element.
59 42 896 D 3417 D 3417.201.09.02.02
13
2 Warning and safety information Sirona Dental Systems GmbH

2.2 Dentist element - rear and side delivery of instruments Operating Instructions C8

, C8+ Tur n
2.2 Dentist element - rear and side delivery of instruments
Side delivery of instruments
max.
2 kg
4.4 lbs
To ensure its mechanical stability, install the dentist element only on cabinets which have been approved by Sirona.
Rear instrument delivery
To ensure its mechanical stability, install the dentist element only on cabinets which have been approved by Sirona.
ATTENTION
Never place any objects on this dentist el­ement.
Position of dentist element: rear and side delivery of instru­ments
ATTENTION
When moving the chair (via program or manually), make sure that the patient does not under any circumstances come into contact with the support arm or the dentist element.
59 42 896 D 3417
14 D 3417.201.09.02.02
Sirona Dental Systems GmbH 2 Warning and safety information Operating Instructions C8

, C8+ Turn 2.3 Satelec Mini L.E.D. curing light

2.3 Satelec Mini L.E.D. curing light
ATTENTION
Always use the glare shield! The light beams emitted by this instrument may be hazardous and may never be di­rected to the eyes of any person, even if this person wears goggles with protective lenses. The light may be directed exclu­sively to the treatment area in the oral cav­ity.
bеЦдблЬ
right
L.E.D. RADIATION
DO NOT STARE INTO BEAM OR VIEW
DIRECTLY WITH
OPTICAL INSTRUMENTS
CLASS 2M LASER PRODUCT
P < 1 mW, λ = 420 nm – 480 nm
IEC 60825-1 Ed. 1.2
!
wrong
L.E.D. beam exit aperture
!
ATTENTION
Do not stare into the beam path with the glass rod removed (Class 2M laser prod­uct). Viewing the LED light exit aperture with certain optical instruments (e.g. a magni­fying lens) within a distance of 100 mm may constitute an eye hazard.
Optical specification
Wavelength λ: = 420 nm – 480 nm
L.E.D. laser power P: < 1 mW
ATTENTION
Heavy line voltage drops may lead to a shutdown of the Mini L.E.D. The fault is signaled by a double beep. Check the se­lected operating mode and restart the Mini L.E.D.
59 42 896 D 3417 D 3417.201.09.02.02
15
2 Warning and safety information Sirona Dental Systems GmbH
2.3 Satelec Mini L.E.D. curing light Operating Instructions C8
, C8+ Tur n
Additional safety information
Any condensation occurring in the hand­piece of the Mini L.E.D. may cause impairments (e.g. fogging of the L.E.D.). When moving the handpiece from a cool environment to a warm room, always wait for it to reach room temperature before putting it into operation.
Curing lights must not be used on persons who are suffering from or have in the past been afflicted by photobiological reactions (including solar urticaria and erythropoetic porphyria). Nor should they be used on persons currently being treated with any medicine which increases one’s sensitiv­ity to light (including methoxsalene and chlorotetracycline).
Any persons who have in the past suf­fered from retinal or eye lens disorders or have undergone eye surgery, especially for gray cataracts, must consult their med­ical eye specialist before using the light or undergoing treatment with the Mini L.E.D. Caution is advisable even if the patient gives his or her consent, since the light intensity can cause accidents. It is espe­cially advisable to always wear appropri­ate protective goggles (UV filter).
16 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 2 Warning and safety information
LEMON
LEMON
0
P
E ND
O
1 0
0
1
20
4 0
6
0
8 0
Operating Instructions C8

, C8+ Tur n 2.4 SIROSONIC L

2.4 SIROSONIC L
wrong
!

2.5 SIROLUX E/SIROLUX FANTASTIC

right
wrong
right
!
bеЦдблЬ
ATTENTION
Never operate without cover: Watch out for overheating of lamp compo­nents and let them cool down.
SIROLUX FANTASTIC
ATTENTION
Pay attention to elevated temperature of the lamp components; and let these cool down.

2.6 Fresh water bottle

LEMON
wrong
Fresh water bottle
ATTENTION
Use only Sirona fresh water bot­tles!
LEMON
59 42 896 D 3417 D 3417.201.09.02.02
17

3 Maintenance Sirona Dental Systems GmbH

3.1 Care and cleaning by the practice team Operating Instructions C8

, C8+ Tur n
3 Maintenance
Despite the outstanding quality of your treatment center, it is necessary to per­form inspections and preventive mainte­nance as well as safety tests at scheduled intervals in order to ensure its operational safety.
These events are documented in the Maintenance Manual.
Care and cleaning is done by the practice team.
For details please refer to the following sections.
3.1 Care and cleaning by the practice team
The practice team is responsible for regu­lar cleaning and care of the treatment center, with the aim of minimizing the risk of contamination for patients and users.
The work to be performed is specified in the chapter "Care and cleaning by the practice team".
The surfaces (apart from upholstery) can be sprayed and wiped with surface disin­fectants.

3.2 Inspection and maintenance

In order to ensure the operational safety and reliability of your treatment center and to avoid damage due to natural wear, annual inspection and maintenance must be performed on your treatment center. This is done by an authorized service engineer from your dental depot.
The steps to be performed as well as the parts which must be replaced are speci­fied in the document "Maintenance Cer- tificate".
An overview of the inspection and mainte­nance work performed is additionally recorded by the service engineer in the "Installation Report/Warranty Pass-
port". This document is part of the "Main­tenance Manual".
18 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 3 Maintenance Operating Instructions C8

, C8+ Tur n 3.3 Safety checks

3.3 Safety checks
Medical equipment is designed in such a way that the first occurrence of a fault does not create a hazard to the safety of patients, users or other persons. Hence it is important to detect such faults before a second fault occurs, which might then lead to safety hazards.
For that reason it is essential to perform safety tests every 2 years which aim par­ticularly at detecting electrical faults (e.g. isolation defects). This is done by an authorized service engineer from your dental depot, most practically together with the work to be performed according to section 3.2.
The check includes a visual inspection as well as measurements of the protective ground wire connections and the equiva­lent leakage currents.
The inspections and measurements to be performed are specified in the "Mainte- nance Manual". The measured values must be documented there by the service engineer.
Safety tests must also be performed and documented during initial startup, after extensions/upgrades (conversion) of your treatment center and after repair
work which might affect the electrical safety of the system.
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
i
NOTE
The treatment center must not be operat­ed if it has failed to pass the safety tests!
19

4 Technical description Sirona Dental Systems GmbH

3.3 Safety checks Operating Instructions C8
, C8+ Tur n
4 Technical description
Model designation
Power supply
Nominal current
Main unit fuse
Operating conditions
Transport and storage conditions
Protection class
Degree of protection against electric shock
Degree of protection against ingress of water
Mode of operation
+
C8
, C8+ Tu r n ,
100 V/115 V/127 V/230 V AC 50/60 Hz
11.5 A at 100 V AC 10 A at 115 V AC 9 A at 127 V AC 5 A at 230 V AC
T 10 A H, 250 V AC, Order No. 10 77 460
Ambient temperature: 10°C – 40°C (50°F – 104°F) Relative humidity: 30% – 75% Air pressure: 700hPa – 1060hPa
Temperature: -40°C – +70°C (-40°F – +158°F) relative humidity: 10% – 95% Air pressure: 500hPa – 1060hPa
Class I device
Type B applied parts, except SIROCAM 3/SIROCAM C: Type BF applied parts
Ordinary equipment (not protected) The foot switch has an IPX1 degree of protection against liquids (drip-proof).
Continuous operation with intermittent loading corresponding to the dental mode of working.
Permanently connected unit.
Year of manufacture
(on rating plate of chair)
Tests/approvals This dental treatment center complies with the requirements of
IEC 60601-1: 1988 + A1: 1991 + A2: 1995 and of IEC 60601-1-2: 2001.
DVGW: This system complies with the technical rules and require­ments on safety and hygiene for connection to the drinking water supply (if the appropriate DVGW marking is affixed to it).
This product bears the CE mark in accordance with the provisions of Council Directive 93/42/EEC of June 14, 1993 concerning medi­cal devices.
0123
20 D 3417.201.09.02.02
59 42 896 D 3417

Sirona Dental Systems GmbH 5 System versions

P 1
P 2
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
2
0
4 0
6 0
80
E N
D O
P 1
P2
P
2
0
1
S
C
8
+
1
2
S
P
0
100
1
2
0
4 0
6 0
8
0
E N
D O
Operating Instructions C8
, C8+ Tur n
5 System versions
C8+ OTP
1. Patient chair/Patient chair Turn
2. OTP dentist element
5.
2.
4.
3.
1.
3. Water unit
4. Assistant element/Assistant element
mounted at the middle of the chair
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
1
1
bеЦдблЬ
6.
7.
C8+ with swivel arm
1. Patient chair/Patient chair Turn
2. Dentist element with swivel arm
5.
2.
4.
3.
1.
3. Water unit
4. Assistant element/Assistant element
mounted at the middle of the chair
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
1
1
6.
7.
1. Alternative equipment, not illustrated
59 42 896 D 3417 D 3417.201.09.02.02
21
5 System versions Sirona Dental Systems GmbH
1
0
0
1
2
0
4
0
60
80
P
1
2
S
0
E
N
D
O
P 1
P 2
Operating Instructions C8
, C8+ Tur n
C8+ Cart
1. Patient chair/Patient chair Turn
2. Dentist element
5.
3. Water unit
1
4. Assistant element/Assistant element
4.
3.
2.
1.
mounted at the middle of the chair
5. SIROLUX
1
6. Instrument foot switch with program
positions
7. Instrument foot switch
6.
7.
5.
2.
C8+ Side Delivery
1. Patient chair /
Patient chair, Turn
2. Dentist element
1
3. Water unit
1.
4. Assistant element
3.
4.
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
7.
6.
1. Alternative equipment, not illustrated
22 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 5 System versions
0
P
1
2
S
E N D
O
1 0 0
1
2
0
4 0
6 0
8
0
P
2
0
1
S
C 8
+
1
2
S
P
0
10
0
1
2 0
4
0
6 0
8 0
E N D
O
Operating Instructions C8
, C8+ Tur n
3.
2.
1.
C8+ Rear Delivery
1. Patient chair /
Patient chair, Turn
2. Dentist element
1
3. Instrument holder
4. Instrument foot switch with program
bеЦдблЬ
positions
5. Instrument foot switch
4.
5.
C8+ Turn
1. Patient chair Turn
5.
2.
4.
3.
1.
2. OTP dentist element/Dentist element
with swivel arm Cart dentist element Cabinet dentist element
2
1
1
3. Water unit
4. Assistant element/Assistant element
mounted at the middle of the chair
1
5. SIROLUX
6. Instrument foot switch with program
positions
7. Instrument foot switch
6.
P 1
P2
7.
8.
8. Swivel arm for right-handedÙleft-
handed conversion
1. Alternative equipment, not illustrated
2. Alternative equipment, not illustrated
59 42 896 D 3417 D 3417.201.09.02.02
23
5 System versions Sirona Dental Systems GmbH
Operating Instructions C8
, C8+ Tur n
C8+ without dentist element
1. Patient chair/Patient chair Turn
2. Water unit
4.
3.
1.
2.
3. Assistant element/Assistant element
mounted at the middle of the chair
4. SIROLUX
i
NOTE
1
1
For operation of the SIVISION 3 video system, please refer to Operating In­structions 59 72 257.
24 D 3417.201.09.02.02
59 42 896 D 3417

Sirona Dental Systems GmbH 6 Putting the system into operation

1
2
S
P
0
100
1
20
40
60
80
ENDO
Operating Instructions C8

, C8+ Tur n 6.1 System ON/OFF switch

6 Putting the system into operation
6.1 System ON/OFF switch
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
25

7 Patient chair Sirona Dental Systems GmbH

0
P
1
2
S
E N D O
100
1
2
0
4 0
6 0
8
0
P
2
0
1
S
C8
+
1
2
S
P
0
1 0
0
1
2
0
4 0
60
8 0
E N D
O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
END
O
S
1
0
2
P
2
0
1
S
C8
+
P
S
1
0
2
S
1
0
2

7.1 Patient positioning Operating Instructions C8

, C8+ Tur n
7 Patient chair
7.1 Patient positioning
1. 2.
Option
3.
4.
Option
Option
0
S
2
1
26 D 3417.201.09.02.02
P
59 42 896 D 3417
Sirona Dental Systems GmbH 7 Patient chair Operating Instructions C8

, C8+ Tur n 7.2 Adjusting the armrest

7.2 Adjusting the armrest
A Folding armrest (optional)
C
B
A

7.3 Flat headrest / articulated headrest

B Wide backrest
C Narrow backrest
ATTENTION
Risk of collision! The following versions must not be retrofit­ted with an armrest:
+
with fixed water unit in combination
• C8 with long cuspidor
A Flat headrest
bеЦдблЬ
B Articulated headrest (optional)
A
1
ATTENTION
The head pad (A) contains a strong mag-
B
net (1) on its bottom side. This magnet could affect any cardiac pacemaker locat­ed nearby.
Therefore, do not allow patients, users and technical personnel with a cardiac pacemaker to be located near this mag­net.
Furthermore, direct contact of the head pad with magnetic cards can lead to data loss.
59 42 896 D 3417 D 3417.201.09.02.02
27
7 Patient chair Sirona Dental Systems GmbH
A

7.4 MultiMotion headrest Operating Instructions C8

, C8+ Tur n
7.4 MultiMotion headrest
Adjustment to patient's height
The headrest can be adjusted to the patient's height by pulling out or pushing in the headrest extension. This adjustment is usually required at the beginning of treat­ment!
i
NOTE
Before positioning the patient on the headrest, make sure that the head support has been optimally adjusted to the patient's height. This simplifies all subsequent work with the MultiMotion considerably, since readjust­ment to the patient's height can thus be omitted when changing over from mandibular to maxillary treatment.
The MultiMotion enables you to adjust the patient's head in a way that optimally supports viewing of areas of the mouth which are difficult to access.
Adjustment of hyperextension
Mandibular position
The mandibular position can be set by pulling the oper­ating handle (A). The anatomical movement of the arched extension keeps the patient's head in the sup­port.
¾ Pull the headrest out of the guide by pulling
handle A.
i
NOTE
You can reduce the adjustment noise by pressing release A.
28 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 7 Patient chair
A
B
B
Operating Instructions C8
, C8+ Tur n 7.4 MultiMotion headrest
Maxillary position
¾ Take the load off the headrest by lifting it slightly.
¾ Press release A on the handle.
¾ Let the headrest slide down slowly.
¾ Let go of the release when the headrest reaches the
desired position.
bеЦдблЬ
Rotation and tilt
The MultiMotion enables rotation of the patient's head about the longitudinal axis of his body as well as lateral tilting of his head.
The MultiMotion is set to the desired position as follows:
1. Press and hold one or both of the two side control
elements B. The rotation and tilt adjustment locks of the headrest are now released.
2. Set the headrest to the desired position.
3. Let go of the control elements B.
; The supporting surface is fixed in place.
Make sure that the headrest is indeed fixed in place after you let go of the control elements!
59 42 896 D 3417 D 3417.201.09.02.02
29
7 Patient chair Sirona Dental Systems GmbH
C
E
D
F
7.4 MultiMotion headrest Operating Instructions C8
, C8+ Tur n
Removing the MultiMotion
For certain treatments (e.g. of children) it may be expe­dient to remove the MultiMotion completely in order to obtain better access to the patient. The patient's head will then be supported on the contact surface of the headrest extension F.
To remove the MultiMotion, proceed as follows:
1. Press release button C.
2. Pull the complete headrest out of the headrest ex-
tension.
3. Cover the opening of the headrest extension with
cover cap E.
4. If a children's head pad D is available, place it on
the headrest extension. The head pad is held mag­netically.
i
NOTE
The children's head pad (D) is not included in the scope of supply, however, is available from your dental dealer.
WARNING
The children's head pad (D) contains a strong mag­net on its bottom side. This magnet could affect any cardiac pacemaker located nearby.
Therefore, do not allow patients, users and technical personnel with a cardiac pacemaker to be located near this magnet.
Furthermore, direct contact of the head pad with magnetic cards can lead to data loss.
i
NOTE
Deposit the removed MultiMotion in a safe place where it cannot fall onto the floor.
Inserting the MultiMotion
To insert the MultiMotion, proceed as follows:
¾ Check the guide of the removed headrest for con-
tamination. Remove any contamination.
¾ Remove the cover cap E.
Reinsert the MultiMotion in the guide from above until it audibly locks in place.
Pull on the headrest again to make sure that it is locked securely in place.
30 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 7 Patient chair Operating Instructions C8

, C8+ Turn 7.5 Changing the patient position by turning (C8+ Turn only)

7.5 Changing the patient position by turning (C8
A
+
Turn only)
Turning the patient chair
The patient chair can be turned max. 17° to either side.
Risk of collision! When turning the chair, make sure that it does not collide with any other compo­nents of the treatment center, persons, furnishings or equipment. Turn the chair only if all treatment instru­ments are in place.
1. Step on foot pedal A to release the
2. Turn the patient chair to the desired
3. The chair locks in place as soon as
ATTENTION
locking brake.
position.
you release foot pedal A.
bеЦдблЬ
17°
17°
59 42 896 D 3417 D 3417.201.09.02.02
31
7 Patient chair Sirona Dental Systems GmbH
S
S 001 2
1
2
S
P
P
0
P
P
P
S
1
0
2
1 2
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
2
0
4 0
6 0
80
END
O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
ENDO
P
1
P
2
S
0
1
2
P
2
0
1
S
C8
+
7.6 Chair programs – factory settings Operating Instructions C8
, C8+ Tur n
7.6 Chair programs – factory settings
Last Memory function
S
1.S
2.S
< 3 s < 3 s
1. 2. 1. 2.
Option
Option
+
without dentist element:
C8
0
S
2
alternatively
1
P
< 3 s < 3 s
1. 2. 1.
2.
P1
P
2
Option
32 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 7 Patient chair
0
P
1
2
S
E N D O
1 0
0
1
20
4
0
6 0
8 0
P
2
0
1
S
C8
+
1
2
S
P
0
1 00
1
2 0
4
0
6 0
8
0
E
N D O
P
2
0
1
S
C8
+
1
2
S
P
0
10
1
20
40
60
80
E
N
DO
P
S
1
0
2
1
2
S
P
0
100
1
20
40
60
80
E
ND
O
1
2
S
P
0
100
1
20
40
60
80
E
N
D
O
P
S
1
0
2
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
EN
D
O
S 0 1 2
S 0 1 2
S 0 1 2
P
2
0
1
S
C8
+
Operating Instructions C8
, C8+ Turn 7.7 Chair programs – reprogramming
7.7 Chair programs – reprogramming
Option
Option
+
C8
without dentist element:
alternatively
0
S
2
1
P
bеЦдблЬ
1.
2.
4.
> 3 s
3.
59 42 896 D 3417 D 3417.201.09.02.02
33
7 Patient chair Sirona Dental Systems GmbH
7.8 Foot switch – reprogramming Operating Instructions C8
, C8+ Tur n
7.8 Foot switch – reprogramming
> 3 s
P1
P
2
Option
34 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 7 Patient chair
0
P
1
2
S
E N D
O
1 0
0
1
2 0
4
0
6 0
8 0
P
S
1
0
2
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
2 0
4 0
6
0
8 0
E N
D O
P
2
0
1
S
1
2
S
P
0
ENDO
P1
P2
P1
P2
P1
P2
P1
P2
P1
P2
P
2
0
1
S
C8
+
Operating Instructions C8

, C8+ Turn 7.9 Emergency stop function for chair movement

7.9 Emergency stop function for chair movement
2.
1.
Option
bеЦдблЬ
Option
Option
0
S
2
3.
3.
1
P
or
alternatively
59 42 896 D 3417 D 3417.201.09.02.02
35
7 Patient chair Sirona Dental Systems GmbH

7.10 Patient chair, options Operating Instructions C8

, C8+ Tur n
7.10 Patient chair, options
Patient chair media block
Additional equipment which is approved for dental use (e.g. by EMS) and the hydrocolloid coolant supply (see page 74) may be connected to a media block on the base of the chair (observe manufacturers’ specifications for the additional equip­ment).
water
Air
T 6.3 AH
Line voltage
Technical data and requirements
The media block is equipped with quick couplings for air and water and an outlet with a separate fuse (6.3A) for non-heat­ing equipment.
Sirona does not assume any liability for possible damage to the connected equip­ment!
Pressure Flow rate
min. 2.5 bar
water
Air 4,4 ± 0.5 bar max. 70 Nl/min
max. 6.0 bar (corresponding to input pressure on treatment center)
max. 1000 ml/min
DVGW approval:
The DVGW approval of the treatment center expires if a media block is installed.
36 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 8 Dentist element Operating Instructions C8

, C8+ Turn 8.1 Locking the dentist element

8 Dentist element

8.1 Locking the dentist element
Brake applied
Maximum load
Brake released
bеЦдблЬ
ATTENTION
A = You must not place any objects on this dentist element!
Brake applied
A
A
Maximum load
Brake released
Option
59 42 896 D 3417 D 3417.201.09.02.02
37
8 Dentist element Sirona Dental Systems GmbH
Click

8.2 Adjusting the instrument position Operating Instructions C8

, C8+ Tur n
8.2 Adjusting the instrument position

8.3 Swivel arm

1.
OK
3.
2.
ATTENTION
IMPORTANT: Swivel arms that are locked in place must not be pushed backward! For resetting, pull out the instru-
NO
ment hose again!
38 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 8 Dentist element
1
0
0
1
20
40
6
0
80
P
1
2
S
0
E
N
D
O
10
0
1
2
0
40
6
0
8
0
P
1
2
S
0
E
N
D
O
2
k
g
2
k
g
m
a
x
.
m
a
x
.
Operating Instructions C8

, C8+ Turn 8.4 Dentist element C8+ Cart

8.4 Dentist element C8
8.5 Dentist element C8
+
Cart
+
Side Delivery
4.
Maximum load capacity
2kg max.
ATTENTION
Risk of toppling over! Move cart slowly and only on even floors without any obsta­cles!
bеЦдблЬ
ATTENTION
Distribute the additional loads on the den­tist element evenly!
The stiffness of the support arm is set with a friction brake (B)
1.
B
1. Loosen the positioning of the dentist
element by turning the fixing screw (B) counterclockwise.
2. Position the dentist element at the de-
sired height.
3. Hold the dentist element in this posi-
tion with one hand.
3.
2.
4. Lock the position of the dentist ele-
ment by turning the locking screw clockwise.
i
NOTE
The maximum load of the dentist element is 2 kg (4.4 lbs)
59 42 896 D 3417 D 3417.201.09.02.02
39
8 Dentist element Sirona Dental Systems GmbH
1
0
0
1
2
0
4
0
6
0
8
0
P
1
2
S
0
E
N
D
O
1
00
1
20
40
6
0
8
0
P
1
2
S
0
E
N
D
O
1
00
1
20
40
6
0
8
0
P
1
2
S
0
E
N
D
O
8.5 Dentist element C8
Side Delivery Operating Instructions C8+, C8+ Tur n
Parking the dentist element
i
NOTE
Avoid collisions with the cabinet when parking the dentist element.
1.
Hold the dentist element so that there is no risk of crushing your hands.
2.
OK
40 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 8 Dentist element
0
P
1
2
S
ENDO
1 0
0
1
20
4
0
6
0
8 0
0
P
1
2
S
E
N D
O
1
0 0
1
2 0
4 0
6 0
8 0
Operating Instructions C8

, C8+ Turn 8.6 Assistant element C8+ Rear Delivery

8.6 Assistant element C8
+
Rear Delivery
i
NOTE
Never place any objects on this dentist el­ement! It is prohibited to apply pressure to the components.
Setting the height of the instru­ment holder
bеЦдблЬ
ATTENTION
When releasing brake A, hold the support arm securely to prevent it from moving up­ward too quickly.
A
To disengage the brake, move lever A up or down.
59 42 896 D 3417 D 3417.201.09.02.02
41
8 Dentist element Sirona Dental Systems GmbH
1
1
2
0
4
0
6
0
8
0
1
0
0
ENDO
2
P
0
S
1
1
2
0
4
0
6
0
8
0
1
0
0
ENDO
2
P
0
S
8.6 Assistant element C8
Rear Delivery Operating Instructions C8+, C8+ Tur n
Parking the dentist element
i
NOTE
Avoid collisions with the cabinet when parking the dentist element. Hold the dentist element so that there is no risk of crushing your hands.
1.
2.
OK
42 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 8 Dentist element
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
20
40
6
0
8
0
E N D
O
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
20
40
6
0
8
0
E N D
O
0
P
1
2
S
ENDO
1 0 0
1
2
0
4 0
60
80
P
2
0
1
S
C8
+
1
2
S
P
0
1 00
1
2 0
40
6
0
8
0
E N D
O
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
20
40
6
0
8
0
E N D
O
P1
P2
0
P
1
2
S
E
N D O
1 0
0
1
2 0
4 0
6
0
8 0
Operating Instructions C8

, C8+ Tur n 8.7 Starting an instrument

8.7 Starting an instrument
A = Unused holders should be sealed
A
1.
with a ball!
bеЦдблЬ
ATTENTION
Do not operate the instrument without spray. Always place the instrument back in the correct holder.
i
NOTE
Also observe the operating instructions for the treatment instruments and the video system.
2.
or
NO
ATTENTION
Use only treatment instruments approved by Sirona!
OK
A
OK
A
A
NO
59 42 896 D 3417 D 3417.201.09.02.02
43
8 Dentist element Sirona Dental Systems GmbH
_
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
2 0
4 0
6
0
8 0
E N
D
O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
2 0
4 0
6
0
8 0
E N
D
O
!
!
+
P1
P2
0
P
1
2
S
E N
D O
100
1
2 0
4 0
60
80

8.8 Adjusting the driving air pressure Operating Instructions C8

, C8+ Tur n
8.8 Adjusting the driving air pressure
Only by a service engineer
A
i
NOTE
The manometer (A) indicates the current driving air pressure.
alternatively
or
59 42 896 D 3417
44 D 3417.201.09.02.02
Sirona Dental Systems GmbH 8 Dentist element
P
2
0
1
S
C8
+
1
2
S
P
0
10
0
1
2 0
4 0
6 0
80
E
N D O
P
2
0
1
S
C8
+
1
2
S
P
0
10
0
1
2 0
4 0
6 0
8 0
E
N D O
0
P
1
2
S
E
N D O
1 0
0
1
20
40
60
80
P1P2P1
P2
Operating Instructions C8
, C8+ Turn 8.9 Instrument spray ON/OFF – chip blower
8.9 Instrument spray ON/OFF – chip blower
bеЦдблЬ
A
A
B
B
A = Spray ON/OFF
B = Chip blower
alternatively
or
i
NOTE
For the adjacent foot switch, "Spray ON" is indicated by a green signal in the upper left corner.
59 42 896 D 3417 D 3417.201.09.02.02
45
8 Dentist element Sirona Dental Systems GmbH
P
1
P
2
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
4 0
6 0
8
0
E N
D O
_
+
P1
P2
!
!
!
+
+
-
P
2
0
1
S
C
8
+
1
2
S
P
0
1
0 0
1
2 0
4 0
6
0
80
EN
DO
+
8.10 Instruments – adjusting the spray intensity – not C8
Rear Delivery Operating Instructions C8+, C8+ Tur n
8.10 Instruments – adjusting the spray intensity – not C8
+
Rear Delivery
2.
1.
3.
4.
alternatively
or
5.
A
B
1.
A = Spray WATER
1.
2.
B = Spray AIR
C8+Cart / C8+ Side Delivery
The control choke may be adjusted only by a service engineer.
46 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 8 Dentist element
0
P
1
2
S
EN
D O
10 0
1
2
0
4 0
6 0
8
0
_
+
!
!
+
Operating Instructions C8
, C8+ Turn 8.11 Instruments – adjusting the spray intensity – C8+ Rear Delivery
8.11 Instruments – adjusting the spray intensity – C8
1. 2.
3.
4.
5.
+
Rear Delivery
Spray WATER (A)
1. Insert the burr.
2. Preselect spray on the instrument foot
switch.
3. Remove the instrument.
4. Actuate the instrument foot switch.
5. Adjust the spray water using the rota-
ry knob on the instrument holder or rack, located on the lower part of the dentist element front panel.
bеЦдблЬ
A
B
Spray AIR (B)
i
NOTE
The control choke may be set only by a service engineer.
59 42 896 D 3417 D 3417.201.09.02.02
47

9 Dentist element options Sirona Dental Systems GmbH

9.1 X-ray image viewer Operating Instructions C8

, C8+ Tur n
9 Dentist element options
9.1 X-ray image viewer
i
NOTE
Not available with C8+ Side Delivery and C8+ Rear Delivery versions.

9.2 Tumbler holder

1. x
2. x
i
NOTE
Not available with C8+ Side Delivery and C8+ Rear Delivery versions.
48 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 9 Dentist element options Operating Instructions C8

, C8+ Turn 9.3 Silicone handles, right/left

9.3 Silicone handles, right/left

9.4 Tray

A
i
NOTE
Not available with C8+Rear Delivery ver­sion.
i
NOTE
Not available with C8+Rear Delivery ver­sion.
B
A = Silicone mat, optional
Option
bеЦдблЬ

9.5 Silicone mat

Maximum load
i
NOTE
Not available with C8+Rear Delivery ver­sion.
59 42 896 D 3417 D 3417.201.09.02.02
49
9 Dentist element options Sirona Dental Systems GmbH
P
2
0
1
S
C8
+
1
2
S
P
0
1 0
0
1
2
0
4
0
60
8
0
E N
D O
P
2
0
1
S
C8
+
1
2
S
P
0
1 0
0
1
2
0
4
0
60
8
0
E N
D O
0
P
1
2
S
E
N D O
1 0
0
1
20
40
6 0
80
Option

9.6 SPRAYVIT 3-way syringe Operating Instructions C8

, C8+ Tur n
9.6 SPRAYVIT 3-way syringe
3 min
275°F135°C
C
A
B
Right button
D
Click
Left button
Check
Both buttons
Right button
Left button
50 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 9 Dentist element options
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
2 0
4
0
6 0
8 0
E N D
O
P
2
0
1
S
C8
+
1
2
S
P
0
1 00
1
2 0
4
0
6 0
8 0
E N D
O
0
P
1
2
S
ENDO
1 0
0
1
2 0
4 0
6 0
8 0
Operating Instructions C8

, C8+ Tur n 9.7 3-way syringe

9.7 3-way syringe
bеЦдблЬ
3 min
135°C 275°F
A
B
Left button
Right button
Both buttons
Check
59 42 896 D 3417 D 3417.201.09.02.02
51
9 Dentist element options Sirona Dental Systems GmbH
P
2
0
1
S
C 8
+
1
2
S
P
0
1
0 0
1
2 0
40
6
0
8 0
E N
D O
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
20
4 0
6
0
8 0
E N
D O
P
2
0
1
S
C
8
+
1
2
S
P
0
1 0
0
1
2 0
40
6
0
80
E N
D O
0
P
1
2
S
E N D O
1
0 0
1
2 0
4 0
6 0
8
0

9.8 Satelec Mini L.E.D. curing light Operating Instructions C8

, C8+ Tur n
9.8 Satelec Mini L.E.D. curing light
Technical data
Model: Mini L.E.D.
Weight of handpiece without hose: 105 g
Dimensions: dia. 23 x 240 mm
Handpiece power supply: 5 V DC/2 A
Optical specification
Wavelength: 420 nm – 480 nm
Light power dia. 7.5 mm (standard ver­sion): 1100 mW/cm
Light power dia. 5.5 mm (optional, order from Satelec): 2000 mW/cm
Power: 450 mW – 500 mW
Laser power: < 1 mW
2
2
Mode 1
0 510
right!
Sec.
0 512,5
Mode 2
Sec.
0 510 15 20
wrong!
DO NOT STARE INTO THE BEAM
OR VIEW IT DIRECTLY WITH
P < 1 mW, λ = 420 nm – 480 nm
iKbKaK=ДЙ~г Йсбн=~йЙкнмкЙ
right!
LED RADIATION
OPTICAL INSTRUMENTS
CLASS 2M LASER
IEC 60825-1 Ed. 1.2
Mode 3
Mode 1,2,3
Sec.
wrong!
ATTENTION
Also observe the safety information on the Mini L.E.D. on pp. 15 ff.
ATTENTION
Use the glare shield! Do not stare into the light exit aperture of the glass rod!
ATTENTION
Do not stare into the beam path with the glass rod removed (Class 2M laser prod­uct). Viewing the LED light exit aperture with certain optical instruments (e.g. a magni­fying lens) within a distance of 100 mm may constitute an eye hazard.
52 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 9 Dentist element options
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
2 0
4 0
6 0
80
E
N D
O
0
P
1
2
S
EN DO
1 00
1
2 0
4 0
6 0
8 0
1
0
0
2
0
4 0
60
80
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
2 0
4 0
6 0
80
E
N D
O
1
2
_
+
+
+
-
-
100
1
20
40
60
80
100
1
20
40
60
80
100
1
20
40
60
80
= 70= 60
1
2
S
P
0
1 0
0
1
2 0
4 0
6 0
8 0
E N D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
100
1
20
40
60
80
100
1
20
40
60
80
P
1
P2
1 …... 60
60
2
1
2
S
P
0
1 0
0
1
2 0
40
6 0
8 0
E N D
O
P
r
o
S
m
i
l
e
P
r
o
S
m
i
l
e
s
i
r
o
n
a
.
0
P
1
2
S
E
N D O
1 0 0
1
2 0
40
6
0
8 0
60
Operating Instructions C8

, C8+ Tur n 9.9 SL/EL 1 electric motor

9.9 SL/EL 1 electric motor
Use only Sirona electric motors!
Speed in % LED/display
120406080100
1 - 4 X
5 - 14 X X
A
2.
1.
A
15 - 24 X
25 - 34 X X
35 - 44 X
45 - 54 X X
55 - 64 X
65 - 74 X X
75 - 84 X
85 - 94 X X
95 - 100 X
A = Preselected maximum speed
Speed +
Speed -
i
NOTE
The set value is saved when the instru­ment is placed back in its holder.
bеЦдблЬ
Motor speed check with the Sirona
1.
unit foot switch
alternatively
or
59 42 896 D 3417 D 3417.201.09.02.02
2.
2.
53
9 Dentist element options Sirona Dental Systems GmbH
1
2
S
P
0
1 0 0
1
2 0
4 0
6 0
8 0
E
N D O
1
2
S
P
0
1 0
0
1
2 0
4 0
6 0
8 0
E N D
O
S
P
100
1
20
40
60
8
0
ENDO
M
M
ENDO
0
P
1
2
S
E
ND O
1 0 0
1
2 0
4 0
60
8 0
M
M
9.9 SL/EL 1 electric motor Operating Instructions C8
, C8+ Tur n
2.
B
1.
Reversing the direction of motor rotation
B = Direction of rotation
Motor CCW
Motor CW
i
NOTE
CCW rotation is not saved when the in­strument is placed back in its holder.
54 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 9 Dentist element options
C8
+
1
2
S
P
0
1 0 0
1
2 0
4
0
6 0
80
E
N D O
1
2
S
P
0
1 0 0
1
2 0
4
0
6 0
8
0
EN
D O
_
+
_
_
+
+
1
2
S
P
0
100
1
20
40
6
0
80
ENDO
60
100
1
20
40
60
80
P
1
P2
60
100
1
20
40
60
80
+
-
-
+
100
1
20
40
60
80
100
1
20
40
60
80
= 70= 60
0
P
1
2
S
ENDO
1
0
0
1
2
0
4
0
6 0
8
0
_
_
+
+
Operating Instructions C8

, C8+ Tur n 9.10 SIROSONIC L scaler

9.10 SIROSONIC L scaler
wrong
!
right
!
Power in % LED/display
120406080100
1 - 4 X
5 - 14 X X
15 - 24 X
25 - 34 X X
35 - 44 X
45 - 54 X X
55 - 64 X
65 - 74 X X
B
A
75 - 84 X
85 - 94 X X
95 - 100 X
A = Water +/-
B = Preselected maximum intensity
bеЦдблЬ
A
Intensity -
Intensity +
A
B
alternatively
or
ATTENTION
The set value is saved when the instru­ment is placed back in its holder.
59 42 896 D 3417 D 3417.201.09.02.02
55
9 Dentist element options Sirona Dental Systems GmbH
0
P
1
2
S
ENDO
1
00
1
20
40
60
80
_
_
+
+
1
1
2
S
P
0
1 0
0
1
2 0
4 0
6 0
8 0
E N DO
C8
+
1
2
S
P
0
1 0
0
1
2 0
4 0
6 0
8 0
E ND O
_
+
_
+
1
2
S
P
0
100
1
20
4
0
60
80
ENDO
100
1
20
40
60
80
+
_
1
2
S
P
0
100
1
20
4
0
60
80
ENDO
ENDO
+
+
-
-
60
100
1
20
40
60
80
100
1
20
40
60
80
100
1
20
40
60
80
P
1
P2
60
= 60 = 70

9.11 SIROSONIC L endodontic scaler Operating Instructions C8

, C8+ Tur n
9.11 SIROSONIC L endodontic scaler
In endodontics mode, the maximum inten­sity of the ultrasonic handpiece is limited
B
A
to ensure safe operation.
Power LED/display
120406080100
Step 1 X
Step 2 X
Step 3 X
Step 4 X
Step 5 X
Step 6 X
A = Water +/-
A
A
B
B = Preselected maximum intensity
Intensity +
Intensity -
alternatively
or
ATTENTION
The set value is saved when the instru­ment is placed back in its holder.
56 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 9 Dentist element options
+
1
2
S
P
+
+
+
+
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
2 0
4
0
6 0
8
0
E N
D O
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
2 0
4
0
60
8
0
E N
D O
2
3
4
5
30 s60 s 30 s 30 s
P1
P 2
O
F
F
P
U
R
G
E
OFF OFF
PURGE PURGE
Operating Instructions C8
, C8+ Turn 9.12 Purge function – not for C8+ Rear Delivery
9.12 Purge function – not for C8
1.
A
2.
5.
3.
+
Rear Delivery
Application
The purge function serves for purging the water paths.
1. Remove the instruments intended for
purging and set them to maximum water flow.
2. Pull off the handpieces.
3. If an SL motor is installed, attach the
purge adapter (A).
4. Put the instruments back in place.
5. Remove the instrument from holder 2
and hold it above the cuspidor.
6. Set the switch to PURGE.
7. The instrument will automatically be
purged for 60 seconds.
i
NOTE
To interrupt the purge function, actuate the foot switch and hold it down.
bеЦдблЬ
6.
6.
PURGE
OFF
11.
or
8. Place the instrument back in its hold-
er and remove the next instrument.
9. After a delay of 5 seconds, the instru-
ment will automatically be purged for 30 seconds.
10. Repeat steps 8. through 9. for all in-
struments connected.
11. Once you have finished purging all in-
struments, set the switch to OFF.
i
NOTE
All handpieces can also be removed si­multaneously. The purge function is then
performed automatically.
PURGE
OFF
11.
PURGE
OFF
59 42 896 D 3417 D 3417.201.09.02.02
57
9 Dentist element options Sirona Dental Systems GmbH
+
+
+
+
0
P
1
2
S
E N
D O
1
00
1
2
0
4 0
6 0
8
0
+
-
0
P
1
2
S
E N
D O
1
0 0
1
2
0
4 0
6 0
80
2
3
4
5
30 s60 s 30 s 30 s
O
F
P
U
R
G
E
O
F
F
9.13 Purge function - C8
Rear Delivery Operating Instructions C8+, C8+ Tur n
9.13 Purge function - C8
1.
2.
A
3.
5.
+
Rear Delivery
P
U
R
G
E
O
F
F
Application
The purge function serves for purging the water paths.
1. Remove the instruments intended for
purging and set them to maximum water flow.
2. Pull off the handpieces.
3. If an SL motor is installed, attach the
purge adapter (A).
4. Put the instruments back in place.
5. Remove the instrument from holder 2
and hold it above the cuspidor.
6. Set the switch to PURGE.
7. The instrument will automatically be
purged for 60 seconds.
i
NOTE
To interrupt the purge function, actuate the foot switch and hold it down.
8. Place the instrument back in its hold-
er and remove the next instrument.
9. After a delay of 5 seconds, the instru-
ment will automatically be purged for 30 seconds.
10. Repeat steps 8. through 9. for all in-
struments connected.
11. Once you have finished purging all in-
struments, set the switch to OFF.
i
NOTE
All handpieces can also be removed si­multaneously. The purge function is then
performed automatically.
6.
PURGE
11.
PURGE
OFF
F
59 42 896 D 3417
58 D 3417.201.09.02.02
Sirona Dental Systems GmbH 10 Assistant element Operating Instructions C8

, C8+ Turn 10.1 Suction handpiece/saliva ejector

10 Assistant element

10.1 Suction handpiece/saliva ejector
A
A = Suction handpiece
B
B = Saliva ejector
bеЦдблЬ
wrong
59 42 896 D 3417 D 3417.201.09.02.02
59
11 Assistant element options Sirona Dental Systems GmbH

11.1 3-way syringes Operating Instructions C8

, C8+ Tur n

11 Assistant element options

11.1 3-way syringes

11.2 Microswitch for vacuum system control

11.3 Air jet pump

60 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 11 Assistant element options Operating Instructions C8

, C8+ Turn 11.4 Satelec Mini L.E.D. curing light

11.4 Satelec Mini L.E.D. curing light
Operation: see Chapter 9.8
bеЦдблЬ

11.5 Control panel, assistant element

59 42 896 D 3417 D 3417.201.09.02.02
61
12 Water unit Sirona Dental Systems GmbH
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
2
0
4
0
6 0
8
0
E
N D
O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
ENDO
0
P
1
2
S
EN
DO
1
0 0
1
2
0
4
0
6 0
8 0
P
2
0
1
S
C8
+
12.1 Cuspidor flushing time – factory settings Operating Instructions C8
, C8+ Tur n

12 Water unit

12.1 Cuspidor flushing time – factory settings
30 s
Option
max. 1 kg
62 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 12 Water unit
P
2
0
1
S
C8
+
Operating Instructions C8
, C8+ Turn 12.2 Cuspidor flushing time – reprogramming
12.2 Cuspidor flushing time – reprogramming
10 s 10 m
bеЦдблЬ
1.
2s
10 s 10 m
3.
2.
Option
alternatively
4.
5.
5.
+
C8
C8+ Turn
59 42 896 D 3417 D 3417.201.09.02.02
63
12 Water unit Sirona Dental Systems GmbH
P
2
0
1
S
C8
+
1
2
S
P
0
1 0
0
1
2
0
4
0
60
8 0
E N
D O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
ENDO
0
P
1
2
S
E
N
D O
100
1
2 0
4 0
60
80
P
2
0
1
S
C8
+
+
100
1
20
40
60
80
ENDO
+
100
1
20
40
60
80
ENDO
P
2
0
1
S
C8
+
12.3 Tumbler filling – factory settings Operating Instructions C8
, C8+ Tur n
12.3 Tumbler filling – factory settings
4 s
Option
max. 1 kg
alternatively
12.4 Tumbler filling – reprogramming
> 3 s < 15 s
1.
2s
2.
> 3 s < 15 s
3.
Option
alternatively
64 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 12 Water unit
S
P
++
P
2
0
1
S
C8
+
1
2
S
P
0
1 00
1
2
0
4 0
6 0
80
END
O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
END
O
0
P
1
2
S
EN
D
O
1
0
0
1
2 0
4 0
60
8
0
S
P
+
S
S
P
+
S
S
P
S
P
2
0
1
S
C8
+
Operating Instructions C8
, C8+ Turn 12.5 Combination of rinsing position S with tumbler filling and flushing controlled by dentist element
12.5 Combination of rinsing position S
with tumbler filling and flushing controlled by dentist element
1. 2.
< 3 s
Option
< 3 s
A
bеЦдблЬ
Option
B
C
< 3 s
59 42 896 D 3417 D 3417.201.09.02.02
< 3 s
D
65
12 Water unit Sirona Dental Systems GmbH
++
P1
P2
S
+
P1
P2
S
+
S
P1
P2
S
S
P1
P2
S
12.5 Combination of rinsing position S with tumbler filling and flushing controlled by dentist element Operating Instructions C8
, C8+ Tur n
Combination of rinsing position S with tumbler filling and flushing controlled by foot switch
Only by a service engineer
max. 1 kg
P1
P
2
Option
66 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 12 Water unit Operating Instructions C8
, C8+ Turn 12.6 Control panel – assistant element – cuspidor flushing
12.6 Control panel – assistant element – cuspidor flushing
30 s
max. 1 kg
12.7 Control panel – assistant element – cuspidor flushing – reprogramming
10 s 10 m
2.1.
bеЦдблЬ
2s
10 s 10 m
3.
4.
5.
+
C8
C8+ Turn
59 42 896 D 3417 D 3417.201.09.02.02
67
12 Water unit Sirona Dental Systems GmbH
P
S
1
0
2
S
0
S
0
12.8 Control panel – assistant element – tumbler filling Operating Instructions C8
, C8+ Tur n
12.8 Control panel – assistant element – tumbler filling
Operation
4 s
Reprogramming
> 3 s < 15 s
1.
2s
2.
> 3 s < 15 s
3.
68 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 12 Water unit
S
P
++
P
S
1
0
2
S
P
+
S
S
P
+
S
S
P
S
Operating Instructions C8
, C8+ Turn 12.9 Control panel – assistant element – combination of mouth rinsing position S with tumbler filling
12.9 Control panel – assistant element –
combination of mouth rinsing position S with tumbler filling and cuspidor flushing
Only by a service engineer
< 3 s
Option
0
S
2
1
P
bеЦдблЬ
< 3 s
< 3 s
< 3 s
59 42 896 D 3417 D 3417.201.09.02.02
69
12 Water unit Sirona Dental Systems GmbH
LEMON
!
!
!
!
!
+
100
1
2
0
4 0
60
80

12.10 Fresh water bottle Operating Instructions C8

, C8+ Tur n
12.10 Fresh water bottle
ATTENTION
Use only Sirona fresh water bot­tles!
Fill with Sirona: Dentosept P
3.
H2O
4.
or
5.
A
100 : 1
2.1.
wrong
LEMON
ATTENTION
Replace deformed, scratched or discol­ored bottles immediately!
Operating temperature: +10 °C – +45 °C.
Do not continue using the bottle after its maximum period of use has expired!
Example above: Replace bottle by No­vember 2008 at the latest
!
59 42 896 D 3417
70 D 3417.201.09.02.02
Sirona Dental Systems GmbH 12 Water unit
P 1
P
2
P 1
P
2
+
Operating Instructions C8
, C8+ Turn 12.11 Tap water/fresh water bottle selector – C8
12.11 Tap water/fresh water bottle selector – C8
A
B
S
+
bеЦдблЬ
You can change over between tap water and the fresh water bottle with switch S.
A Tap wa t e r
B Fresh water bottle
59 42 896 D 3417 D 3417.201.09.02.02
alternatively
or
71
12 Water unit Sirona Dental Systems GmbH
1
0
0
1
2
0
40
60
8
0
P
1
2
S
0
END
O
PURGE
OFF
12.12 Changeover between tap water and fresh water bottle - C8
+
Side Delivery Operating Instructions C8+, C8+ Tur n
12.12 Changeover between tap water and fresh water bottle - C8
You can change over between tap water and the fresh water bottle with switch S.
PURGE
OFF
S
B
PURGE
S
OFF
PURGE
OFF
A
A Tap water
B fresh water bottle
+
Side Delivery
72 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 12 Water unit
0
P
1
2
S
E N
D
O
100
1
2
0
4 0
6 0
8
0
Operating Instructions C8

, C8+ Turn 12.13 Changeover between tap water and fresh water bottle - C8+ Rear Delivery

12.13 Changeover between tap water and fresh water bottle - C8
You can change over between tap water and the fresh water bottle with
switch S.
A Tap wa te r
PU
RG
E
O
FF
B Fresh water bottle
+
Rear Delivery
bеЦдблЬ
B
PURGE
S
OFF
A
PURGE
OFF
59 42 896 D 3417 D 3417.201.09.02.02
73
12 Water unit Sirona Dental Systems GmbH

12.14 Hydrocolloid connection Operating Instructions C8

, C8+ Tur n
12.14 Hydrocolloid connection
Extra equipment
If you want to work with a hydrocolloid coolant supply, you can obtain an adapter, Order No.: 33 15 814, free of charge with the enclosed voucher.
Please state your address and the serial number of the chair (see “Installation Report/Warranty Passport”).
OPEN
CLOSE
74 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 12 Water unit Operating Instructions C8

, C8+ Turn 12.15 Dürr wet suction system

12.15 Dürr wet suction system
Daily after consulting hours 1liter
H2O
bеЦдблЬ
C8+ Turn
10 s
simultaneous-
59 42 896 D 3417 D 3417.201.09.02.02
75
12 Water unit Sirona Dental Systems GmbH
2005
1324567 81091112 13 14 15 1716 18 19 20 21 22 2423 25 26 27 28 29 3130
2004
1324567 81091112 13 14
15 1716 18 19 20 21
22 2423 25 26 27 28 29 3130
2003
1324567
81091112 13 14 15 1716 18 19 20 21 22 2423 25 26 27 28 29 3130
2002
1324567
81091112 13 14 15 1716 18 19 20 21 22 2423 25 26 27 28 29 3130
....
100 %
95 %
0 %
0 %
AM
TEST
SAN
AMALG
DESINF
AM
TEST
SAN
AMALG
DESINF
AM
TEST
SAN
AMALG
DESINF
AM
TEST
SAN
AMALG
DESINF
AM
TEST
!
!
!

12.16 Amalgam separator Operating Instructions C8

, C8+ Tur n
12.16 Amalgam separator
AM
A
TEST
A
M A
L G
D
E S
IN F
SAN
A = ! Please replace
amalgam container !
B
AM TEST
A
M A
L G
D
E S
IN F
SAN
AM TEST
A
M A
L G
D
E S
I N F
SAN
B = !! Replace
amalgam container !!
Error
AMALGAM test
AM
TEST
at least 1x annually
AM TEST
A
M A
L G
D
E S
IN F
SAN
OK
!
59 42 896 D 3417
76 D 3417.201.09.02.02
Sirona Dental Systems GmbH 12 Water unit Operating Instructions C8

, C8+ Tur n 12.17 Disinfection system

12.17 Disinfection system
bеЦдблЬ
A
A = ! If the DESINF indicator flashes,
refill 1 liter of DENTOSEPT P!
Refill with
Sirona: Dentosept P
59 42 896 D 3417 D 3417.201.09.02.02
77
12 Water unit Sirona Dental Systems GmbH
12.18 Conversion of water unit right-handed Ù left-handed version Operating Instructions C8
, C8+ Tur n
12.18 Conversion of water unit right-handed Ù left-handed version
Conversion only possible on a C8+ Turn
ATTENTION
Danger of falling objects! Prior to con­version, remove all objects loosely depos­ited on any components of the treatment center.
i
NOTE
Please observe the minimum clear­ances specified in the document “Di­mension Drawing, Technical Data”.
78 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 12 Water unit Operating Instructions C8
, C8+ Turn 12.18 Conversion of water unit right-handed Ù left-handed version
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
79
12 Water unit Sirona Dental Systems GmbH
12.18 Conversion of water unit right-handed Ù left-handed version Operating Instructions C8
, C8+ Tur n
80 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 13 SIROLUX E operation light – optional
0
P
1
2
S
E N D
O
10
0
1
2 0
40
6
0
8 0
P
2
0
1
S
C8
+
1
2
S
P
0
1
0 0
1
2 0
4 0
6
0
8 0
ENDO
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
ENDO
P
2
0
1
S
C8
+
0
P
1
2
S
E N D
O
1
0 0
1
20
4
0
6
0
80
P
2
0
1
S
C
8
+
1
2
S
P
0
1
0 0
1
2 0
40
6
0
80
E N
D O
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
8
0
ENDO
Operating Instructions C8
, C8+ Turn 13.1 ON/OFF – brightness reduction
13 SIROLUX E operation light – optional
13.1 ON/OFF – brightness reduction
Option
alternatively
bеЦдблЬ
wrong right
Composite setting
Light intensity: 22000 lux
Light intensity: 8000 lux
i
NOTE
Use this setting to prevent premature cur­ing of composite fillings.
ATTENTION
Do not use without transparent cover!
59 42 896 D 3417 D 3417.201.09.02.02
81
13 SIROLUX E operation light – optional Sirona Dental Systems GmbH
700 mm 28 ’’
180 mm 7 ’’
75 mm
3 ’’
Click

13.2 Focusing Operating Instructions C8

, C8+ Tur n
13.2 Focusing
ATTENTION
When focusing, make sure you turn the cover cap only slightly and just up to its stop. If the cover cap is turned counter­clockwise past the stop, it will spring out of its catch and the halogen lamp can fall down.
82 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 14 SIROLUX FANTASTIC operation light – optional
0
P
1
2
S
E N D O
1 0
0
1
2
0
4 0
6 0
80
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
2
0
4 0
6 0
80
E N DO
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
ENDO
P
2
0
1
S
C8
+
0
P
1
2
S
EN
DO
1
0 0
1
2 0
4 0
6
0
80
P
2
0
1
S
C8
+
1
2
S
P
0
1 0 0
1
20
4 0
6 0
80
ENDO
P
2
0
1
S
C8
+
1
2
S
P
0
100
1
20
40
60
80
ENDO
Operating Instructions C8
, C8+ Turn 14.1 ON/OFF – brightness reduction
14 SIROLUX FANTASTIC operation light – optional
14.1 ON/OFF – brightness reduction
Option
alternatively
bеЦдблЬ
Composite setting
Light intensity: 24000 lux
Light intensity: 8000 lux
i
NOTE
Use this setting to prevent premature cur­ing of composite fillings.
59 42 896 D 3417 D 3417.201.09.02.02
83
14 SIROLUX FANTASTIC operation light – optional Sirona Dental Systems GmbH
700 mm 28 ’’
200 mm 8 ’’
100 mm
4 ’’
Click

14.2 Focusing Operating Instructions C8

, C8+ Tur n
14.2 Focusing
ATTENTION
When focusing, make sure you turn the cover cap only slightly and just up to its stop. If the cover cap is turned counter­clockwise past the stop, it will spring out of its catch and the halogen lamp can fall down.
84 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team Operating Instructions C8

, C8+ Turn 15.1 Care and cleaning agents

15 Care and cleaning by the practice team

15.1 Care and cleaning agents
ATTENTION
Use only cleaning and care agents which are approved by Sirona!
You will find a list of all agents approved at the time of delivery in the documenta­tion folder supplied with your treatment center.
A continuously updated list of approved agents can be downloaded from the Sirona Internet homepage under REF 59 70 905: Ö www.sirona.com
If you do not have Internet access, you can use one of the following two ways to order the list:
Order from your local dental depot
Order from Sirona Tel: ++49 (0) 62 51/16-16 16 FAX: ++49 (0) 62 51 / 16-18 18
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
85
15 Care and cleaning by the practice team Sirona Dental Systems GmbH

15.2 Patient chair Operating Instructions C8

, C8+ Tur n
15.2 Patient chair
Upholstery/foot switch/chair
Wipe disinfection
Care and cleaning of the uphol­stery:
Care and cleaning of the synthetic leather upholstery must be performed weekly. The cleaning agent, fluid or cloths also perform care simultaneously.
ATTENTION
P1
P 2
HO
2
Disinfecting the upholstery
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
ATTENTION
Use only disinfectants which are approved by Sirona (see page 85, section 15.1)!
Do not use any other disinfectants!
Pedal (foot switch)
Clean the bottom of the base plate using a damp cloth.
86 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team
G
Operating Instructions C8
, C8+ Tur n 15.2 Patient chair
MultiMotion disinfecting the surfaces
The surfaces can be spray and wipe disin­fected with surface disinfectants.
bеЦдблЬ
Intensive cleaning and disinfection of the head support
The contact surface of the headrest can be removed to facilitate intensive cleaning and disinfecting if neces- sary.
The surfaces can be spray and wipe disin­fected with surface disinfectants.
¾ Press one side of head pad G inward from the
outside (to unlock it) and remove it in a upward direction.
i
NOTE
Fast and frequent removal of the head pad from the front or rear can damage the snap-in noses.
¾ Clean and disinfect the head pad and headrest.
¾ Hook on the head pad again laterally.
ATTENTION
Damaged pads, e.g. with scratched surfaces, must be replaced.
For reorders:
Contact surface: REF 61 82 047
59 42 896 D 3417 D 3417.201.09.02.02
87
15 Care and cleaning by the practice team Sirona Dental Systems GmbH
15.2 Patient chair Operating Instructions C8
, C8+ Tur n
Cleaning the filter, replacing the main fuse
ATTENTION
Be sure to switch the treatment center
A
OFF before replacing the fuses!
A = T 10A H
88 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team
1
2
S
P
0
0
0
1
2
0
40
60
8
0
E
N
D
O
Operating Instructions C8

, C8+ Tur n 15.3 Dentist element

15.3 Dentist element
Instrument hoses/instrument rack
ATTENTION
1.
2.
3.
4.
2.
3 min
135°C 275°F
5.
i
For care and cleaning of the motors, the straight and contra-angle handpieces, the highspeed handpiece and the SIROSONIC L scaler, please refer to the
7.
corresponding operating instructions.
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
ATTENTION
SPRAYVIT – relieve pres­sure!
NOTE
bеЦдблЬ
Wipe disinfection
6.
5x
5x
93°C 197°F
8.
A = Unused holders should be sealed
7.
with a ball!
A
6.
5x
8.
5x
59 42 896 D 3417 D 3417.201.09.02.02
89
15 Care and cleaning by the practice team Sirona Dental Systems GmbH
1
2
S
P
0
100
1
20
40
60
80
ENDO
0
P
1
2
S
E N
D O
1
0 0
1
2
0
40
6
0
8 0
15.3 Dentist element Operating Instructions C8
, C8+ Tur n
Instrument hoses / Instrument holder - C8
+
Rear Delivery
ATTENTION
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
ATTENTION
SPRAYVIT – relieve pressure!
i
NOTE
For care and cleaning of the motors, the straight and contra-angle handpieces, the highspeed handpiece and the SIROSONIC L scaler, please refer to the corresponding operating instructions.
A
Wipe disinfection accord­ing to section 15.1.
A = Unused holders should be sealed with a ball!
90 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team
HO
2
Operating Instructions C8
, C8+ Tur n 15.3 Dentist element
SPRAYVIT 3-way syringe
3 min
bеЦдблЬ
135°C 275°F
3 min
1.
2.
3.
3-way syringe
Spray disinfection/wipe disinfection
59 42 896 D 3417 D 3417.201.09.02.02
91
15 Care and cleaning by the practice team Sirona Dental Systems GmbH
15.3 Dentist element Operating Instructions C8
, C8+ Tur n
Satelec Mini L.E.D.
ATTENTION
135°C 275°F
3 min
L.E.D. light is switched off before using disinfectants. Do not use any abrasive cleaners.
Prior to each use, check the light guides to make sure that no composite residues have adhered to them. If this is the case, remove these residues immediately and check to make sure that the surface of the insert has not been damaged. If you find any damage, replace the damaged light guide.
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
Always make sure that the Mini
92 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team
P
2
0
1
S
C8
+
1
2
S
P
0
1
0 0
1
2 0
4 0
6
0
80
E
N D O
P
1
P 2
P
2
0
1
S
C8
+
1
2
S
P
0
1 0
0
1
2 0
4 0
6 0
80
E N D O
P
2
0
1
S
C8
+
1
2
S
P
0
1 0
0
1
20
4 0
6 0
80
E N D O
P
2
0
1
S
C
8
+
1
2
S
P
0
1
0 0
1
20
4 0
6 0
8 0
E
N D O
0
P
1
2
S
E N
D O
100
1
2
0
4
0
60
8
0
Operating Instructions C8
, C8+ Tur n 15.3 Dentist element
Surfaces
ATTENTION
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
135°C 275°F
3 min
Wipe disinfection
bеЦдблЬ
Option
Option
59 42 896 D 3417 D 3417.201.09.02.02
93
15 Care and cleaning by the practice team Sirona Dental Systems GmbH
i
135°C 275°F
3 min
f

15.4 Assistant element Operating Instructions C8

, C8+ Tur n
15.4 Assistant element
Suction handpiece/saliva ejector
No amalgam!
Amalgam recycling
According to manufacturer’s speci
Wipe disinfection
Every 2 hours: 2 glasses
Daily after work: 1 Liter (0.25 Gallons)
ATTENTION
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
94 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team
i
135°C 275°F
3 min
Operating Instructions C8
, C8+ Tur n 15.4 Assistant element
Assistant element – suction hand­piece/saliva ejector (USA)
No amalgam!
According to manufacturer’s specifications
Amalgam recycling
Wipe disinfection
Every 2 hours: 2 glasses
bеЦдблЬ
59 42 896 D 3417 D 3417.201.09.02.02
Daily after work: 1 Liter (0.25 Gallons)
ATTENTION
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
95
15 Care and cleaning by the practice team Sirona Dental Systems GmbH
Soap
0,75 l
Desinf.
Soap

15.5 Water unit Operating Instructions C8

, C8+ Tur n
15.5 Water unit
Water unit
ATTENTION
No amalgam!
Use only cleaning agents which are approved by Sirona (see page 85, section 15.1)!
Amalgam recycling
Spray disinfection/wipe disinfection
ATTENTION
Do NOT use household de­tergents or acidic agents!
C
+
C8
1. 3. 4.2.
C8+ Turn
1 x
weekly
Daily before the lunch
break and after consulting hours
The vacuum system is subjected daily to septic secretions, saliva and blood. For hygienic reasons the system must there­fore be disinfected with a detergent/disin­fectant approved by Sirona.
Air jet pump
60 s
96 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team
P1
P2
1
2
S
P
0
1 0
0
1
20
4
0
6 0
8
0
ENDO
P1
P2
10
0
1
20
4 0
60
8 0
Operating Instructions C8
, C8+ Tur n 15.5 Water unit
Fresh water bottle
i
1.
NOTE
Perform sanitation very 4 weeks!
Sanitation (1)
2.
Fill with
Sirona: Dentosept P
alternatively
or
bеЦдблЬ
3.
20 s
5.1 5.2 5.3 5.4
5.
Tumbler filling 2x
6.
4.
Switch OFF the system
20 s
for 24 hours
59 42 896 D 3417 D 3417.201.09.02.02
97
15 Care and cleaning by the practice team Sirona Dental Systems GmbH
1
0 0
1
2
0
40
60
8 0
P1
P2
1
2
S
P
0
1
00
1
2
0
40
60
80
ENDO
P1
P2
15.5 Water unit Operating Instructions C8
, C8+ Tur n
Fresh water bottle
1.
H2O
Sanitation (2)
Fill with water
2.
alternatively
or
4.
20 s
3.
5.
20 s
6.1 6.2 6.3 6.4
6.
Tumbler filling 2x
98 D 3417.201.09.02.02
59 42 896 D 3417
Sirona Dental Systems GmbH 15 Care and cleaning by the practice team Operating Instructions C8
, C8+ Tur n 15.5 Water unit
Amalgam separator
C = Change amalgam container! D = Change amalgam container
at least 1x annually
bеЦдблЬ
C
D
E = Signal during removal!
2.
C8+ Turn
1.
F
E
F = Clean siphon once a month
G = Do not reuse!
3.
4.
59 42 896 D 3417 D 3417.201.09.02.02
Amalgam recycling!
Replacement: Amalgam rotor Order No. 14 34 138
G
5.
6.
99
15 Care and cleaning by the practice team Sirona Dental Systems GmbH
15.5 Water unit Operating Instructions C8
, C8+ Tur n
Disinfection system – microbio­logical water test
Must be performed after a working pause of more than one week. Use a “Total Count Tester” for this purpose. The first three testers will be supplied free of charge if you return the attached voucher. Please state your address and the serial number of the chair (see “Installation Report/Warranty Passport”). Additional testers can be ordered under Order No. 58 53 775.
30 s
ATTENTION
The “Total Count Tester” has a maximum durability of 1 year after its date of delivery.
The cardboard disk contains a dehy­drated nutrient medium. It is activated by the sample and serves as culture medium for a number of bacteria. The number of germs provides information on the hygienic condition of the water.
35°C 24 h
Performing the test
Allow water to run out of the
SPRAYVIT syringe for about 1 minute.
Then take a water sample from this
syringe. Fill in water up to the upper mark on the container. To neutralize the disin­fecting agent of the water sample, add approx. 1.5 g of fixing salt (sodium thiosulfate). Do not touch the inside of the contain­er or the part of the tester to be im­mersed.
Immerse the tester in the filled con-
tainer for 30 seconds. The cardboard disk with the nutrient medium will now absorb 1ml of the water sample.
Remove the tester and shake off ex-
cessive water. Empty the container completely.
Place the tester into the empty con-
tainer for incubation: Keep either for two days at room temperature or for 24 hours at a temperature of 35 °C.
Then count all germs found on the
surface of the tester.
ATTENTION
If the number of germs significantly ex­ceeds 100, then sanitation is required (see next page).
59 42 896 D 3417
100 D 3417.201.09.02.02