Mindray BeneVision N22, BeneVision N17, BeneVision N19, BeneVision N15, BeneVision N12 Quick Reference Manual

BeneVision N22/N19/N17/N15/ N12
Quick Reference Guide
Note: The BeneVision N22/N19/N17/N15/N12 Quick Reference Guide is not intended as a replacement to the Operating Instructions. This document is a guideline only to be used as an aid to comprehensive Inservice Training.
Table of Contents
QUICK KEYS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
PATIENT SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Admit Patient And Enter Demographics (Manual Entry) . . . . . . . . . . . . . . . . . . . .4
Enter Demographics from the CMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Enter Demographics from the ADT Database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Discharge a Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
STANDBY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Enter Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Resume Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Realtime Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Adjust Recorder’s Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
PRINT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Realtime Waveforms (Laser printer connection required) . . . . . . . . . . . . . . . . . . .8
Adjust Printed Waveforms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
End Case Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Adjust ECG Lead . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Enable Pacemaker Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Initiate ECG Relearn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Initiate ST Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Initiate QT Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Selecting Leads for QT Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Initiate and Print a 12-lead ECG Report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Initiate A Non-Invasive Blood Pressure Measurement (NIBP) . . . . . . . . . . . . . . .11
Table of Contents (cont'd)
Stop Non-Invasive Blood Pressure Measurement (NIBP) . . . . . . . . . . . . . . . . . . .11
Set A Non-Invasive Blood Pressure Measurement Interval . . . . . . . . . . . . . . . . .12
Configure A Custom Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Factors that Affect Blood Pressure Measurement Accuracy . . . . . . . . . . . . . . . . .13
INVASIVE BLOOD PRESSURES (OPTIONAL) . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Adjust Invasive Blood Pressure Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Zero Invasive Blood Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
CONTINUOUS CARDIAC OUTPUT (OPTIONAL) . . . . . . . . . . . . . . . . . . . . . . . 16
Adjust the Displayed Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
CENTRAL VENOUS OXYGEN SATURATION (OPTIONAL) . . . . . . . . . . . . . . . 16
Print Oxygenation Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
BISPECTRAL INDEX (OPTIONAL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Adjust BIS Waveform Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
BIS Expand View (Bilateral Sensor Required) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Adjust Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
ALARMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting Parameter Alarm Properties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Changing Arrhythmia Alarm Settings(Arrhythmia Option Required) . . . . . . .18
Disable Non-Lethal Arrhythmia Alarms (Arrhythmia Option Required) . . . . . .19
Accessing Help When Technical Alarms Occur . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
RESET ALARMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
PAUSE ALARMS/PAUSE ALARM TONES . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
HEMODYNAMIC MONITORING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Initiate A Cardiac Output Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Initiate Pulmonary Artery Wedge Pressure (PAWP) . . . . . . . . . . . . . . . . . . . . . . . . .22
Table of Contents (cont'd)
Initiate Hemodynamic Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
SCREEN DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
REVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Tabular Trends . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Arrhythmia and Alarm Events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Full Disclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
OxyCRG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
VIEWING OTHER BEDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Entering the Remote View Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Adding a Bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Resetting Alarms for Remote Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
Removing a Bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Saving a Manual Event . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Displaying the Alarm Watch Tile on the Main Screen . . . . . . . . . . . . . . . . . . . . . . .28
VOLUMES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Alarm Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Key Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
Beat Tone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
DATE AND TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
PARAMETER COLORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
LOAD DEFAULT (PRESET) SETTINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
LOCK/UNLOCK TOUCHSCREEN DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
PAIR THE TM80 TELEMETRY MONITOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Pair the TM80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Unpair the TM80 Telemetry Monitor from the N Series Bedside Monitor . . . .32
Unpair the N Series Bedside Monitor from the TM80 Telemetry Monitor . . . .32
Table of Contents (cont'd)
PAIR BP10 NIBP MODULE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Pair the BP10 Module with the N Series Bedside Monitor . . . . . . . . . . . . . . . . . .33
Unpair BP10 Module from the N Series Bedside Monitor . . . . . . . . . . . . . . . . . . .33
CLEANING/DISINFECTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Cleaning the Main Unit/Module/SMR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Disinfecting the Main Unit/Module/SMR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Cleaning and Disinfecting the Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
STERILIZATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Changes made from the N Series bedside monitor effecting the BeneVision Central Monitoring System and changes made from the BeneVision Central Monitoring System effecting N Series bedside monitor are indicated with the
symbol.
BeneVision N22/N19/N17/N15/N12
QUICK KEYS
Quick Reference Guide
Main Menu
Opens the main menu.
Alarm Setup
Opens the Alarm dialog.
Audio Pause
Pauses alarm tone.
Review
Opens Review dialog for the patient’s historical data.
Patient Management
Opens the Patient Management dialog.
NIBP Start/Stop
Starts an NIBP measurement or stops the current NIBP measurement.
NIBP
More
Shows more quick keys.
Alarm Reset
Acknowledges alarms.
Alarm Pause
Pauses the current and future alarms for a preconfigured time.
Standby
Opens Standby mode.
Screen Setup
Opens the Screen Setup dialog.
Stop All
Stops all NIBP measurements.
NIBP Measure
Starts NIBP STAT measurement.
Zero IBP
Zeros all IBPs.
PAW P
Opens the PAWP screen.
Opens the NIBP Measure dialog.
C.O. Measure
Starts cardiac output procedure.
Loops
Opens the respiratory Loops screen.
1
BeneVision N22/N19/N17/N15/N12
Quick Reference Guide
Venipuncture
Opens the Venipuncture dialog.
Parameters Setup
Opens the Parameters Setup dialog.
Manual Event
Triggers a manual event.
OxyCRG
Opens the OxyCRG screen.
Privacy Mode
Enters the privacy mode.
CPB Mode
Enters the CPB mode.
Start TOF
Starts/stops TOF measurement.
Remote View
Opens the split-screen view of another patient(s).
Minitrends
Opens the Minitrends screen.
ECG Full- Screen
Opens the 12-lead ECG full screen.
Night Mode
Enters the night mode.
Intubation Mode
Enters the intubation mode.
Volume
Opens the Volume dialog.
Calculations
Opens the Calculations dialog.
Print
Initiates a real-time report.
2
Freeze
Freezes waveforms.
Load Config
Opens the Load Config dialog to load
preconfigured settings.
Record
Starts/Stops a recording.
BeneVision N22/N19/N17/N15/N12
Quick Reference Guide
Unit
Opens the Unit dialog.
ECG Lead/Gain
Opens the ECG Lead/Gain dialog.
HemoSight
Opens the HemoSight dialog.
GCS
Opens the GCS dialog.
Enters the Discharged
Patients dialog.
Discharge Patient
Opens the Discharge Patient dialog to discharge the patient and restore default settings.
Rotate Screen (for N22/ N19)
Changes the screen orientation (between portrait and landscape).
Call Help
Calls for help.
EWS
Opens the EWS screen.
SepsisSight
Opens the SepsisSight dialog.
ECG 24h Sum
Views the 24-hour ECG summary.
3
BeneVision N22/N19/N17/N15/N12
Quick Reference Guide
GESTURES
The following gestures on the touchscreen are enabled:
Swipe down the touchscreen with two fingers to access the Discharge Patient dialog box.
Swipe from left to right on the touchscreen with two fingers to change the screen from normal screen, to minitrends screen, to big numerics screen, to normal screen.
Swipe from right to left on the touchscreen with two fingers to change the screen from normal screen, to big numerics screen, to mini trends screen, to normal screen.
PATIENT SETUP
Admit Patient And Enter Demographics (Manual Entry)
The Patient Management dialog allows the clinician to admit a patient and enter patient demographics. Admitting also restores default (preconfigured) settings and clears the previous patient’s information.
After a patient is manually discharged, the monitor automatically admit a new patient. To manually admit a patient on the N Series bedside monitor:
a. The monitor prompts to admit a new patient in the following situations:
After being switched o for the selected time period*, the monitor automatically discharge the previous patient and admit a new patient at startup.
If the monitor has not detected certain patient vital signs (ECG, SpO RR, NIBP) for 30 minutes, the clinician will be prompted to admit a new patient once continuous monitoring is detected.
*Time period configurable.
b. To open the Patient Management dialog:
Select the patient information area at the top left corner of the screen.
Select the Patient Management
4
quick key.
, PR,
2
BeneVision N22/N19/N17/N15/N12
Select the Main Menu quick key, and select Patient Management
from the Patient Management column.
Select Patient Management from the discharge screen.
c. Select Patient Category and select OK to confirm.
d. Select patient field(s) and enter patient information as required using the
onscreen keyboard.
Quick Reference Guide
e. Select
f. Select OK to confirm and return to the main screen.
Note: If the patient’s barcode is scanned, the medical number and associated information is entered into the monitor.
button to accept the information.
Enter Demographics from the Central Monitoring System (CMS)
If the N Series bedside monitor is connected to the CMS, entering patient information at the CMS will enter the patient information into the N Series bedside monitor. In addition, entering demographic information into the N Series monitor will automatically enter information into the CMS.
Enter Demographics from the ADT Database
If the N Series bedside monitor is connected to a hospital’s ADT server the patient demographic information can be imported into the N Series bedside monitor. To import patient information from the Hospital’s ADT server:
a. Enter the Find Patient dialog in either of the following ways:
Select the Main Menu Patient Management column.
Select Find Patient from the Patient Management dialog.
b. Verify the ADT tab is selected.
quick key, and select Find Patient from the
c. Enter search field such as patient ID.
d. Select Search to display a list of the search results.
e. Select a patient from the patient list.
f. Select
import that patient’s information and return to patient management.
adjacent to the patient information N Series bedside monitor to
5
BeneVision N22/N19/N17/N15/N12
g. Select OK to confirm and return to the main screen.
Note: Find patient from the ADT server is only enabled if ADT Query is
supported.
Note: Loading patient information from the ADT server updates only patient information in the N Series bedside monitor. The patient’s monitoring data is unaffected.
Quick Reference Guide
Discharge a Patient
The Discharge function clears all patient information and restores default (preconfigured) settings.
Swipe down the touchscreen with two  ngers, and select Discharge to
con rm.
Select the Discharge Patient
con rm.
Select the Patient Management Patient.
Select the Main Menu Patient Management column.
Select Discharge patient from the standby screen.
And select Discharge to confirm.
quick key, and select Discharge to
quick key, and select Discharge
quick key, and select Discharge from the
STANDBY
Enter Standby
The Standby function allows the clinician to suspend monitoring on a patient. While in standby, the patient is not being monitored; however, all data and settings are preserved.
To place a patient into Standby on the N Series bedside monitor:
a. Select the Standby quick key.
Or
a. Select the Main Menu
Management column.
quick key, and select Standby from the Patient
b. Set Location from the drop down list to define where the patient is when
the monitor enters the standby mode.
6