Mindray BeneHeart R3, BeneHeart R3A Service Manual

BeneHeart R3/BeneHeart R3A
Electrocardiograph
Service Manual
Intellectual Property Statement
Mindray) owns the intellectual property rights to this product and this manual. This manual
may refer to information protected by copyrights or patents and does not convey any license
under the patent rights of Mindray, nor the rights of others. Mindray does not assume any
liability arising out of any infringements of patents or other rights of third parties.
, , and are the registered trademarks or trademarks owned
by Mindray in China and other countries.
Revision History
This manual has a revision number. This revision number changes whenever the manual is
updated due to software or technical specification change. Contents of this manual are
subject to change without prior notice.
Version number 2.0 Release time: October 2014
© 2013-2014 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
I
Preface
Manual Purpose
This manual provides detailed information about the assembling, dissembling, testing and
troubleshooting of the equipment to support effective troubleshooting and repair. It is not
intended to be a comprehensive, in-depth explanation of the product architecture or
technical implementation. Observance of the manual is a prerequisite for proper equipment
maintenance and prevents equipment damage and personnel injury.
Intended Audience
This manual is for biomedical engineers, authorized technicians or service representatives
responsible for troubleshooting, repairing and maintaining the equipments
Passwords
A password may be required to access different modes. The passwords are listed below:
Maintain Mode: 888888 Demo Mode: 2088
II
Contents
1 General .............................................................................................................................. 1-1
1.1 Revision History ............................................................................................................. 1-1
1.2 Safety Information .......................................................................................................... 1-1
1.2.1 Warnings ............................................................................................................. 1-2
1.2.2 Cautions ............................................................................................................. 1-2
1.2.3 Notes .................................................................................................................. 1-3
1.3 Equipment Symbols ........................................................................................................ 1-3
2 Theory of Operation ........................................................................................................ 2-1
2.1 Introduction ..................................................................................................................... 2-1
2.2 Connectors for External Equipment ................................................................................ 2-2
2.3 Main Unit ........................................................................................................................ 2-4
2.4 Top housing assembly ..................................................................................................... 2-5
2.4.1 Main Control Board ........................................................................................... 2-5
2.4.2 Keypad Board..................................................................................................... 2-6
2.4.3 AC/DC Power Board .......................................................................................... 2-6
2.4.4 Recorder Drive Board ........................................................................................ 2-6
2.4.5 Parameter Board ................................................................................................. 2-6
3 Unpacking and Installation ............................................................................................. 3-1
3.1 Unpacking the Equipment ............................................................................................... 3-1
3.2 Preparation for Installation .............................................................................................. 3-1
3.3 Turning On the Equipment .............................................................................................. 3-2
4 Hardware and Software Upgrade ................................................................................... 4-1
4.1 Hardware Upgrade .......................................................................................................... 4-1
4.1.1 Upgrading the Equipment to have Wi-Fi Function ............................................ 4-1
4.2 Software Upgrade............................................................................................................ 4-2
4.2.1 Installing Software Upgrade Tool ...................................................................... 4-3
4.2.2 Software Upgrade Procedure ............................................................................. 4-5
5 Testing and Maintenance ................................................................................................. 5-1
5.1 Introduction ..................................................................................................................... 5-1
5.1.1 Test Equipment ................................................................................................... 5-1
5.1.2 Test Report ......................................................................................................... 5-1
5.1.3 Preventative Maintenance .................................................................................. 5-1
5.1.4 Recommended Frequency .................................................................................. 5-2
5.2 Visual Inspection ............................................................................................................. 5-2
5.3 Power-on Test .................................................................................................................. 5-3
1
5.4 ECG Performance Test .................................................................................................... 5-3
5.5 Recorder check ................................................................................................................ 5-3
5.6 Battery Check .................................................................................................................. 5-4
5.7 Maintenance Mode .......................................................................................................... 5-4
6 Troubleshooting ................................................................................................................ 6-1
6.1 Overview ......................................................................................................................... 6-1
6.2 Parts Replacement ........................................................................................................... 6-1
6.3 Quickly Identifying Failure ............................................................................................. 6-1
6.4 General Problems and System Messages ........................................................................ 6-1
6.5 Troubleshooting Guide .................................................................................................... 6-2
6.5.1 Power On/Off Failures ....................................................................................... 6-2
6.5.2 Recorder Failures ............................................................................................... 6-3
6.5.3 Wrong system time ............................................................................................. 6-4
6.5.4 Wi-Fi Related Failures ....................................................................................... 6-5
7 Disassembly and Repair .................................................................................................. 7-1
7.1 Tools Required ................................................................................................................ 7-1
7.2 Preparations for Disassembly .......................................................................................... 7-1
7.3 Disassembling the Main Unit .......................................................................................... 7-2
7.3.1 Removing Bottom housing assembly ................................................................. 7-2
7.3.2 Remove the Power Board ................................................................................... 7-3
7.3.3 Removing the Main Control Board .................................................................... 7-3
7.3.4 Upgrading to Have Wi-Fi Function and Repair the Wi-Fi Module .................... 7-4
7.3.5 Removing the Display ........................................................................................ 7-6
7.3.6 Removing the Parameter Board ......................................................................... 7-6
7.3.7 Removing the Keypad Board ............................................................................. 7-7
7.3.8 Removing the Recorder Drive Board ................................................................. 7-8
7.3.9 Removing the Thermal Print Head ..................................................................... 7-9
8 Parts .................................................................................................................................. 8-1
8.1 Introduction ..................................................................................................................... 8-1
8.2 Main Unit ........................................................................................................................ 8-2
8.2.1 Exploded View ................................................................................................... 8-2
8.2.2 Parts List ............................................................................................................ 8-3
8.3 Top Housing Assembly (115-017634-00 or 115-017793-00) .......................................... 8-5
8.3.1 Exploded View ................................................................................................... 8-5
8.3.2 Parts List ............................................................................................................ 8-5
8.4 Recorder door assembly (115-017633-00) ...................................................................... 8-6
x
8.4.1 E
8.4.2 Parts List ............................................................................................................ 8-6
8.5 Bottom housing assembly (115-017792-00 or 115-018387-00) ...................................... 8-7
8.5.1 Exploded View ................................................................................................... 8-7
8.5.2 Parts List ............................................................................................................ 8-7
ploded View ................................................................................................... 8-6
2
9 Replacement Parts, Software Upgrade Package and Accessories ................................ 9-1
9.1 BeneHeart R3/R3A BOM (V1.0) for Repair, Upgrade and Accessories ......................... 9-1
9.2 Material Change .............................................................................................................. 9-2
A Electrical Safety Inspection ........................................................................................... A-1
A.1 Power Cord Plug ........................................................................................................... A-1
A.2 Device Enclosure and Accessories ................................................................................ A-2
A.3 Device Labeling ............................................................................................................ A-2
A.4 Protective Earth Resistance ........................................................................................... A-2
A.5 Earth Leakage Test ........................................................................................................ A-4
A.6 Patient Leakage Current ................................................................................................ A-7
A.7 Mains on Applied Part Leakage .................................................................................... A-9
A.8 Patient Auxiliary Current .............................................................................................. A-9
A.9 Scheduled Electrical Safety Inspection ....................................................................... A-14
A.10 Electrical Safety Inspection after Repair ................................................................... A-16
3
FOR YOUR NOTES
4
1 General
1.1 Revision History
Manual Version
1.0 First release April 2013
2.0
Description of Changes
Add information on Wi-Fi upgrade, repairing and
troubleshooting
Update Chapter 9 replacement parts, software upgrade
package and accessories list.
Change Time
October 2014
SN Since
1.2 Safety Information
WARNING
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
death or serious injury.
CAUTION
Indicates a potential hazard or unsafe practice that, if not avoided, could result in
minor personal injury or product/property damage.
NOTE
Provides application tips or other useful information to ensure that you get the
most from your product.
1-1
1.2.1 Warnings
WARNING
All installation operations, expansions, changes, modifications and repairs of this
product are conducted by authorized personnel.
Disconnect the equipment from external power source and remove the battery
before disassembling the equipment.
When you disassemble/reassemble the equipment, a patient leakage current test
must be performed before it is put in use.
The equipment must be connected to a properly installed power outlet with
protective earth contacts only. If the installation does not provide for a protective earth conductor, disconnect it from the power line and operate it on battery power, if possible.
Dispose of the package material, observing the applicable waste control regulations
and keeping it out of children’s reach.
1.2.2 Cautions
CAUTION
Make sure that no electromagnetic radiation interferes with the performance of the
equipment when preparing to carry out performance tests. Mobile phone, X-ray equipment or MRI devices are a possible source of interference as they may emit higher levels of electromagnetic radiation.
Before connecting the equipment to the power line, check that the voltage and
frequency ratings of the power line are the same as those indicated on the equipment’s label or in this manual.
Protect the equipment from damage caused by drop, impact, strong vibration or
other mechanical force during servicing.
1-2
1.2.3 Notes NOTE
Refer to Operation Manual for detailed operation and other information.
1.3 Equipment Symbols
Caution (Attention, consult
accompanying documents)
Battery indicator
DEFIBRILLATION-PROOF
TYPE CF APPLIED PART
Network connector
DATE OF
MANUAFACTURE
Dispose of in accordance to
your country’s requirements
Authorized representative in
the European Community
ON/OFF for part of
equipment
Alternating current (AC)
Equipotentiality
USB connector
Serial number
Environment-friendly
Used Period per Chinese
Standard SJ/T11363-2006
China Metrology
Certification
Manufacturer
The product bears CE mark indicating its conformity with the provisions of
the Council Directive 93/42/EEC concerning medical devices and fulfils
the essential requirements of Annex I of this directive.
NOTE
Some symbols may not appear on your equipment.
1-3
FOR YOUR NOTES
1-4
2 Theory of Operation
2.1 Introduction
BeneHeart R3/BeneHeart R3A Electrocardiograph (hereafter referred to as “the equipment”
or “the system”) is intended to acquire, analyze, display, store, and record the patient’s
electrocardiographic information for clinical diagnosis and study.
The equipment provides the following functions:
Display 12-lead ECG waveforms in real time; print manual, auto, and rhythm
measurements.
Perform resting ECG interpretation, output measurements and diagnosis in auto
measurement mode.
Save ECG data in internal memory and export the data through a USB memory. Enables program upgrade over the network.
Internet
Mobile Viewer
PC / ECG Management
B
u
i
l
t
ECG Server
B
USB
u
ilt
-
i
n
W
i
-
F
i
-
in
W
i
-
F
i
L A N
2-1
The above figure shows the system consists of the electrocardiograph and its peripheral
devices. The electrocardiograph:
Acquires, processes and reviews 12-lead resting ECG data; Prints ECG reports; Runs on AC power source or a battery. Exports data through a USB memory. Connects LAN or Wi-Fi to send ECG data. Connects a barcode reader to enter patient information.
2.2 Connectors for External Equipment
The connectors for external devices are located on the left and rear of the equipment.
Side View
1. Patient cable connector: connects the patient cable for ECG acquisition
2. Recorder door latch: opens the recorder door
1. Network connector: connects the equipment to the network for software upgrade
2. USB connector: connects USB disc for data transfer
1
1
2
2 2
2-2
Back View
1. AC power input
2. Equipotential Grounding Terminal: When the equipment and other devices are to be
used together, their equipotential grounding terminals should be connected together to
eliminate the potential difference between them.
1
2
Bottom View
1. Battery compartment
1
2-3
2.3 Main Unit
The equipment consists of the following parts:
Top housing assembly: consists of main control board, keyboard assembly, display,
power switch, power board (AC/DC), recorder, and parameter board.
Lower housing: includes a battery compartment.
The following figure shows the main unit architecture:
Thermal head
module
W7
Power board
W1
USB RJ45
Battery
W5
Keypad board
W2
Main control board
W6
LCD
Recorder drive
board
W3
Parameter BoardW4
Wi-Fi
(reserved)
2-4
2.4 Top housing assembly
Recorder drive board
Power board
Bosses for screwing
top and bottom
housings together
2.4.1 Main Control Board
The main control board provides the following functions:
Drives display and control back light Stores data Processes printing
Keypad board
Bosses for screwing
top and bottom
housings together
Main control board Parameter board
Implements serial port communication Drives the buzzer Provides USB and wired network connection Processes data Implements power supply conversion Charges the battery Implements power management Implements network communication
2-5
2.4.2 Keypad Board
The keypad board drives the keys. The power on/off switch, AC indicator and battery
indicator are also located on the keypad board.
2.4.3 AC/DC Power Board
The power board transforms AC power to DC power.
2.4.4 Recorder Drive Board
The recorder drive board drives the recorder module.
2.4.5 Parameter Board
Parameter board is responsible for ECG data acquisition.
2-6
3 Unpacking and Installation
3.1 Unpacking the Equipment
Open the package and take out the packing list. Check that all the articles included in the
packing list are available and the quantity and specification are correct.
All the optional parts purchased by the customer shall also be checked. Notify the supplier if provided components are not correct as compared to the packing
list.
In case of damage during transportation, keep the packing material and notify the
supplier immediately.
Keep the packing material till the equipment is accepted.
3.2 Preparation for Installation
1. Ensure that the site meets all safety, environmental and power requirements
2. Check that required power sockets are available.
3. Check that a network connector is available if the equipment needs to be connected to
the wired network.
Environmental Requirements
To avoid explosion hazard, do not use the equipment in the presence of flammable
anaesthetics, vapours or liquids. The environment where the equipment will be used should
be reasonably free from vibration, dust and corrosive substances. If these conditions are not
met, the accuracy of the system may be affected and damage may occur.
Enviornment
Check and make sure that there is no electromagnetic interference source around the
equipment, especially large medical electrical equipment such as radiological
equipment and magnetic resonance imaging equipment etc. Switch off these devices
when necessary.
Keep the examination room warm (no less than 18) to avoid muscle action voltages
in ECG signal caused by cold.
3-1
Power supply
Check that power supply specification is met and the power cord is securely connected
if mains power is used. Use only power socket that is properly grounded.
AC mains Input voltage Input power Frequency Battery
Battery Type
Run time
Shutdown delay
100-240V (±10%)
60 VA
50 Hz/60 Hz (±3 Hz)
When powered by a new fully-charged battery and at ambient temperature 25 ±5 ,
500 auto measurement reports, or 2 hours of continuous recording, or 6
hours of measurement without recording
With equipment power off and at ambient temperature 25 ℃±5 ℃,
3 h to 90% capacity
3.5 h to 100% capacity
at least 5 minutes (after a low battery message first occurs)
3.3 Turning On the Equipment
Once the equipment has been installed and checked, you can get ready for measurement and
recording:
1. Connect the equipment with AC mains. If you run the equipment on battery power,
ensure that the battery is sufficiently charged.
2. Press the power on/off switch.
3-2
4 Hardware and Software Upgrade
4.1 Hardware Upgrade
4.1.1 Upgrading the Equipment to have Wi-Fi Function
For the equipment not configured with the Wi-Fi function, there is no Wi-Fi module socket
on the main board. To upgrade the equipment, use a main board with Wi-Fi module socket
to replace the original main board.
The Glasgow 12-lead resting ECG algorithm and the PDF software package are included in
the Wi-Fi upgrade package.
The Wi-Fi upgrade package (PN: 115-029558-00) consists of the following items:
Main board PCBA, with Wi-Fi module socket Wi-Fi module kit System software package Glasgow 12-lead resting ECG algorithm PDF software package
Refer to 7.3.4 Upgrading to Have Wi-Fi Function and Repair the Wi-Fi Module for the
upgrade procedure.
4-1
4.2 Software Upgrade
You can upgrade the system by installing and running the Mindray Patient Monitor
Software Upgrade Tool (PN: 110-000493-00) on a PC with Windows operating system.
Connect the equipment to be upgraded and the PC running the upgrade tool to the same
network, or directly connect the equipment and the PC via a crossover network cable. Then
configure the IP address of the PC. You can upgrade the following software:
No.
1
2
3 110-002545-00
4 110-001201-00 TR8A recorder drive chip writing software
5
6 115-018007-00 Glasgow upgrade pack (FRU, English)
7 115-017996-00 PDF format upgrade pack (FRU)
Software type
System softwar
package
Module software
Software function
upgrade package
Note: No specific sequence is required for the upgrading of above software. For detailed
information, please refer to 4.2.2 Software Upgrade Procedure.
PN Description
110-002535-00
/
110-002525-0 Power management program
115-017992-00 Glasgow upgrade pack (FRU, Chinese)
System software package (Simplified
Chinese)
System software package (Traditional
Chinese)
M501 ECG module STM32 chip writing
software
4-2
4.2.1 Installing Software Upgrade Tool
1. Find the installation program and double click it to start
installation.
2. Select language.
3. Click [Ok] and the following screen is displayed. Click [Next] to go to the next step.
4. Enter User Name, Company name, and Serial Number "26582640". Then click [Next].
4-3
5. Specify the destination folder for installing this program. Then select [Next].
6. Select Program Folder. Then select [Next].
7. Click [Finish] to complete installation.
4-4
4.2.2 Software Upgrade Procedure
1. Connect the equipment to be upgraded and the PC running the upgrade tool to the
same network, or directly connect the equipment and the PC via a crossover network
cable.
2. Set the IP address of the PC to "77.77.1.XX" and subnet mask to "255.255.255.0".
3. Run Mindray Patient Monitor Software Upgrade Tool on the PC and set Machine to
[BeneHeart R3/R3A].
4. On the Mindray Patient Monitor Software Upgrade Tool screen, select [Select
Package] and then the packages you want to upgrade. Then select [Start].
5. Turn on the equipment to be upgraded. Press and hold "F1" key and "F4" key for 2 to 3
seconds. Then the equipment enters upgrade mode and starts software upgrade
automatically. Corresponding prompt messages are displayed to be upgraded on both
the equipment and PC.
When software upgrade is finished, restart the equipment and check if the software is
correctly upgraded.
For details of software upgrade, please refer to help and instructions for use of Mindray
Patient Monitor Software Upgrade Tool.
CAUTION
Disconnect the equipment from the patient and make sure the important data are saved
before upgrade.
Do not shut down or power off the equipment when upgrading the system software.
Otherwise, it may cause the equipment to break down.
Software upgrade should be performed by qualified service personnel only. Crossover network cable is recommended when a PC is connected for software upgrade.
NOTE
Make sure the version of the upgrade package is what you desired. To obtain the latest
upgrade package, please contact Mindray Customer Service Department.
4-5
5 Testing and Maintenance
5.1 Introduction
To ensure the equipment always functions normally, qualified service personnel should
perform regular inspection, maintenance and test. This chapter provides a checklist of the
testing procedures for the equipment with recommended test equipments and frequency.
The service personnel should perform the testing and maintenance procedures as required
and use appropriate test equipments.
The testing procedures provided in this chapter are intended to verify that the equipment
meets the performance specifications. If the equipment or a module fails to perform as
specified in any test, repairs or replacements must be done to correct the problem. If you
have any question, contact our Customer Service Department.
CAUTION
All tests should be performed by qualified service personnel only. Service personnel should acquaint themselves with the test tools and make sure that
test tools and cables are applicable.
5.1.1 Test Equipment
See the following sections.
5.1.2 Test Report
Upon the completion of the tests, the table of preventative maintenance test reports and the
table of maintenance test reports in this chapter should be kept properly.
5.1.3 Preventative Maintenance
Below are preventative maintenance tests which need to be performed on the equipment.
See the following sections for detailed maintenance procedures.
5-1
5.1.4 Recommended Frequency
Check/Maintenance Item Frequency Visual inspection
Power-on test
ECG test
Electric safety tests
Recorder check
Battery check
Performance Test
Module
Calibration
Refer to A
Electrical Safety
Inspection
Functional test
Performance test
1. When first installed or reinstalled.
1. When first installed or reinstalled.
2. Following any repairs or replacement of any main
unit parts.
1. If the user suspects that the measurement is incorrect.
2. Following any repairs or replacement of relevant
module.
3. At least once every two years.
1. Following any repair or replacement
2. After the equipment drops.
3. At least once every two years.
Following any repair or replacement of the recorder.
1. When first installed.
2. Whenever a battery is replaced.
Once per year, or if the battery run time reduced
significantly.
5.2 Visual Inspection
Inspect the equipment for obvious signs of damage. The test is passed if the equipment has
no obvious signs of damage. Follow these guidelines when inspecting the equipment:
Carefully inspect the case, display screen, keys, and knob for obvious signs of damage. Inspect the power cord, wall mount, and accessories for obvious signs of damage. Inspect all external connections for loose connectors, bent pins or frayed cables. Inspect all connectors on the equipment for loose connectors or bent pins. Make sure that safety labels and data plates on the equipment are clearly legible.
5-2
5.3 Power-on Test
This test is to verify that the equipment can power up correctly. The test is passed if the
equipment starts up by following this procedure:
1. Install the battery, and connect the equipment to AC power. The AC mains indicator
and battery indicator are illuminated.
2. Press the Power On/Off switch to turn on the equipment.
Then the equipment performs self-test as soon as the equipment is powered on.
5.4 ECG Performance Test
Tool required:
Patient simulator, eg. Medsim300B or equivalent equipment
Test Method:
1. Connect the patient cable to the equipment under test and the lead wires to the patient
simulator.
2. Set the patient simulator as follows: ECG sinus rhythm, HR = 80 bpm, with the
amplitude as 1 mV (default setting).
3. Check that the displayed ECG waveforms are clear and complete without obvious
noise.
4. Press the [ECG] key to start automatic ECG measurement. Print out the waveforms
and diagnosis.
5. Check that the displayed ECG waveforms are clear and complete with the amplitude of
lead II waveform about 10 mm (assumed the Gain is set to 10mm/mV), and diagnosis
indicating sinus rhythm and HR 80 bpm.
6. Disconnect each of the leads in turn and observe the corresponding lead off message
displayed on the screen.
5.5 Recorder check
Tool required:
None
1. Print ECG waveforms. The recorder should print correctly and the printout should be
clear.
2. Simulate some recorder problems, such as out of paper, paper jam, and etc., and then
the equipment should present the message "Check Recorder". After the problem is
removed, the recorder should be able to work correctly.
5-3
5.6 Battery Check
Tool required:
None
Functional Test
1. If the equipment is installed with a battery, remove the battery first.
2. Verify that the equipment works correctly when running powered form an AC source.
3. Verify that battery is installed. If not, install the battery.
4. Remove the AC power cord and verify that the equipment still works correctly.
Performance Test
Perform the test by referring to the chapter 10 Battery in the Operator’s Manual and verify
the operating time of the battery meets the product specification.
5.7 Maintenance Mode
The equipment includes maintenance mode for you to check the equipment conveniently.
To enter the maintenance mode, select [Menu][System Setup][Maintain Mode]→enter required password.
The password for the maintenance mode is 332888 when the equipment left the factory. You
cannot change this password. If you have problem to enter the maintenance mode, please
contact your service personnel. You can perform the following tests in the maintenance
mode:
Display test Sound test Keypad test Lead wire check Recorder test Internal communication test USB test Internal storage Test
Select a test item you want to perform, and follow the on-screen prompt.
5-4
In the maintenance mode, you can also
View device information Check work log Set the frequency of AC filter Format the internal memory Input the equipment's serial number Import hospital and department name
To exit the maintenance mode, turn off the equipment.
5-5
Maintenance and Test Report
(See the above sections for detailed test procedures and contents)
Customer name Customer address Servicing person Servicing company Equipment under test
(EUT) Model of EUT SN of EUT Hardware version Software version Test equipment Model/No. Expiration Date for
Calibration
Test Contents Test Record
Visual Inspection
The housing, display screen, keys, power cord, and accessories
have no obvious signs of damage.
The external connecting cables are not frayed and the
connector pins are not loose and bent.
The external connectors are not loose or their pins are not bent.
The safety labels and data plate are clearly legible.
Power-on Test
The power-on self test is passed. The power indicator works
5-6
Test Result
(Pass/Fail)
correctly and the equipment starts up properly.
Performance test
ECG performance test and calibration
The ECG waveforms displayed and printed out should be clear
and complete without obvious noise. The amplitude of lead II
waveform is about 10 mm and diagnosis indicate sinus rhythm
and HR 80 bpm.
ECG Lead Off prompt behaves correctly.
Recorder check
The recorder can print ECG waves correctly and the printout is
clear.
Simulate recorder problems, such as out of pater and paper
jam, the message "Check Recorder" shall be presented.
Battery Check
The equipment can operates correctly from battery power
when an AC power failure accidentally occurs.
The equipment can operate independently on a single battery.
Test conclusion:
Tested by: Test date:
5-7
6 Troubleshooting
6.1 Overview
In this chapter, equipment problems are listed along with possible causes and recommended
corrective actions. Refer to the tables to check the equipment, identify and eliminate the
troubles.
The troubles we list here are frequently arisen difficulties and the actions we recommend
can correct most problems, but not all of them. For more information on troubleshooting,
contact our Customer Service Department.
6.2 Parts Replacement
The equipment's PCBs, major parts and components are replaceable. Once you isolate a
PCB you suspect defective, follow the instructions in Chapter 7 Disassembly and Repairto
replace the PCB with a known good one and check that the trouble disappears or the
equipment passes all performance tests. Defective PCB can be sent to us for repair. If the
trouble remains, exchange the replacement PCB with the original suspicious PCB and
continue troubleshooting as directed in this chapter.
To obtain information on replacement parts or order them, refer to 9
Upgrade Package and Accessories.
Replacement Parts, Software
6.3 Quickly Identifying Failure
Some troubleshooting tasks may require you to identify the hardware version and status of
your equipment. To check equipment information,
Select [Menu][System Setup][Maintain Mode]Enter the required password to
enter the maintenance mode. In the maintenance mode, you can perform related tests to
primarily identify the failure.
6.4 General Problems and System Messages
Refer to Chapter 9 Troubleshooting of the operator’s manual.
6-1
6.5 Troubleshooting Guide
6.5.1 Power On/Off Failures
6-2
6.5.2 Recorder Failures
Recorder not work
1. Check platen in position.
2. Check paper correctly loaded.
3. Close battery door.
Message Please
N
End
check recorder
presented?
Replace
print head
Recording paper
incorrectly loaded
Message “Recorder
unavailable! presented?
N
Message Please
Y
check recorder
presented?
N
Y
Recorder runs by pressing
Manual key?
Y
N
Message
Paper type error
presented?
Y
Y
N
Cable between
recorder and main
board loose or
failed?
N
Ok after recorder drive
board replaced
N
Replace main board
Perform keypad test in
maintenance mode.
Replace keypad board if
failed.
Y
Reconnect or replace cable
EndY
Set Paper Typecorrectly
6-3
6.5.3 Wrong system time
Displayed system time is
wrong
Message
RT clock error
is presented?
N
N
No cell button or cell button failed
N
Replace main control board
Y
Y OKReset system time
Y Replace button cell
Replace main
control board
6-4
6.5.4 Wi-Fi Related Failures
Symptoms Possible Cause Action
Wi-Fi function does not
work. No Wi-Fi icon shows
at the top of the screen.
The equipment is frequently
off line or disconnects from
the network.
Unable to connect the
network. The network
disconnected symbol
shows at the top of the
screen.
Wi-Fi module is not properly
installed.
Wi-Fi module fails Replace the Wi-Fi module.
Main board fails Replace the main board.
The Wi-Fi antenna is
detached from the equipment
housing, or detached from
the Wi-Fi module, or the
antenna is damaged.
Wi-Fi module failure. Replace the Wi-Fi module.
Wi-Fi signal is unstable Check that the hospital Wi-Fi
Wrong network parameter
settings, for example, IP
address is not correctly set or
the password is not correct.
The Wi-Fi signal is weak. Check that hospital Wi-Fi
Reinstall the Wi-Fi module
properly.
1. Visually check if the Wi-Fi
antenna is damaged. If yes,
replace the antenna.
2. If the antenna is not
damaged, check that the
Wi-Fi antenna is properly
connected with the module.
Secure the antenna properly.
signal is stable.
Check if there is IP conflict.
Re-set the IP address
correctly and enter correct
password.
signal is stable and strong.
The Wi-Fi antenna is
detached from the equipment
housing, or detached from
the Wi-Fi module.
Main board fails Replace the main board.
Secure the antenna properly.
6-5
FOR YOUR NOTES
6-6
7 Disassembly and Repair
7.1 Tools Required
To disassemble and replace the parts and components, the following tools may be required:
Philips screwdrivers Tweezers Sharp nose pliers Clamp
7.2 Preparations for Disassembly
Before disassembling the equipment, finish the following preparations:
Stop measuring, turn off the equipment, and disconnect all the accessories and
peripheral devices.
Disconnect the AC power source and remove the battery.
WARNING
Before disassembling the equipment, be sure to eliminate the static charges first.
When disassembling the parts labeled with static-sensitive symbols, make sure you are wearing electrostatic discharge protection such as antistatic wristband or gloves to avoid damaging the equipment.
Properly connect and route the cables and wires when reassembling the equipment to
avoid short circuit.
Select appropriate screws to assemble the equipment. If unfit screws are tightened by
force, the equipment may be damaged and the screws or part may fall off during use, causing unpredictable equipment damage or human injury.
Follow correct sequence to disassembly the equipment. Otherwise, the equipment
may be damaged permanently.
Be sure to disconnect all the cables before disassembling any parts. Be sure not to
damage any cables or connectors.
Be sure to place removed screws and disassembled parts properly, preventing them
from being lost or contaminated.
Place the screws and parts from the same module together to facilitate reassembling. To reassemble the equipment, first assemble the assemblies, and then the main unit.
Carefully route the cables.
7-1
7.3 Disassembling the Main Unit
NOTE
To disassemble the equipment, place the equipment on a work bench free from
foreign material, avoiding damaging the screen.
All the operations should be performed by qualified service personnel only.
When disassembling the PCBAs and parts labeled with static-sensitive symbols, make sure you are wearing electrostatic discharge protection such as antistatic wristband or gloves to avoid damaging the equipment.
7.3.1 Removing Bottom housing assembly
1. Lay the equipment on the work bench. Unscrew the M3×6 screw to open the battery
compartment door. Then remove the battery.
2. Unscrew the four M4×10 screws and remove the bottom housing assembly.
7-2
7.3.2 Remove the Power Board
Unscrew the four M3×6 screws. Disconnect the AC input receptacle and the cable.
Disconnect the connection cable between the main control board and the power board. Then
take out the power board and the insulator.
AC input receptacle
and cable
7.3.3 Removing the Main Control Board
1. Release the latch locking the flexible PCB connecting the display to the main control
board by pushing the latch from Position 1 to Position 2 as shown in the picture below.
Disconnect the flexible PCB.
Position 1
Position 2
NOTE
Make sure to disconnect the flexible PCB between the main control board and the
display before removing the main control board.
7-3
2. Unscrew the four M3×6 screws. Disconnect the recorder cable, the connecting cable
between the main control board and the parameter board, the connecting cable between
the main control board and the keypad board, and the connecting cable between the
main control board and the power board.
3. Remove the main control board.
7.3.4 Upgrading to Have Wi-Fi Function and Repair the Wi-Fi
Module
Follow this procedure to upgrade the equipment to have the Wi-Fi function:
1. Remove the main control board as instructed by 7.3.3 Removing the Main Control
Board.
2. Install the Wi-Fi module on the main control board FRU that has a Wi-Fi module
connector. Connect the Wi-Fi antenna.
3. Install the main control board FRU on the electrocardiography. Secure the main
control board FRU with four M3×6 screws.
4. Stick the antenna as indicated by the following picture. Secure the antenna cable with a
tape.
7-4
Stick the Wi-Fi
p
antenna here
Stick the Wi-Fi
antenna here
Secure the Wi-Fi
cable under the
insulating plate
-
To repair the Wi-Fi module,
1. Remove the Wi-Fi antenna cable from the Wi-Fi module.
2. Remove the Wi-Fi module and replace with a new one.
Place the Wi-Fi cable below the plastic
frame and secure the cable with a ta
e.
Wi-Fi
7-5
7.3.5 Removing the Display
Before removing the display, remove the main control board as described in 7.3.3
Removing the Main Control Board.
Then you can remove the display. Follow this procedure:
1. Unscrew the four PT3×6 tapping screws to remove the displayfixing kit.
2. Remove the TFT display.
7.3.6 Removing the Parameter Board
1. Unscrew the four M3×6 tapping screws.
2. Disconnect the cable between the main control board and the parameter board.
7-6
7.3.7 Removing the Keypad Board
Before removing the keypad board, remove the parameter board as described in Section
7.3.6 Removing the Parameter Board.
Then follow this procedure to remove the keypad board:
1. Unscrew the four PT3×6 tapping screws.
2. Disconnect the cable between the main control board and the keypad board. Then
remove the keypad board.
7-7
7.3.8 Removing the Recorder Drive Board
1. Unscrew the two M3×6 screws. Disconnect the recorder cable
NOTE
Make sure to the flexible PCB which is located under the TR8A recorder and
connects the thermal print head before disassembling the recorder drive board.
7-8
2. Release the latch locking the flexible PCB connecting thermal print head to the
recorder drive board by pushing the latch from Position 1 to Position 2 as shown in the
picture below. Disconnect the flexible PCB and remove the recorder drive board.
Position 1
Position 2
7.3.9 Removing the Thermal Print Head
Before removing the thermal print head, remove the recorder drive board as described in
Section 7.3.8 Removing the Recorder Drive Board
Then follow this procedure to remove the thermal print head:
1. Unscrew the two M3×6 screws, and the take out the thermal print head.
NOTE
Open the recorder door before removing and installing the thermal print head.
7-9
FOR YOUR NOTES
7-10
8 Parts
8.1 Introduction
This chapter contains the exploded views and parts lists of the main unit. It helps the
engineer to identify the parts during disassembling the equipment and replacing the parts.
This manual is based on the maximum configuration. Your equipment may not have same
parts and the quantity of the screws or stacking sleeves etc. may be different with those
included in the parts lists.
NOTE
The part number listed in the Parts List is only for checking the FRU part number
which is also included in the Parts List. Please provide the FRU parts number if you want to purchase the spare parts.
8-1
8.2 Main Unit
8.2.1 Exploded View
8-2
8.2.2 Parts List
SN PN Description FRU part number Remarks
1 021-000108-00 Display 021-000108-00 /
2 051-001353-01 Main control board
(512MB)
051-001882-00 Main control board
(512MB, Wi-Fi)
3 / Keypad (BeneHeart
R3/Chinese)
/ Keypad (BeneHeart
R3A/Chinese)
/ Keypad (BeneHeart
R3/symbol)
/ Keypad (BeneHeart
R3A/symbol)
115-017993-00 512M/Glasgow/PDF,
115-017994-00 512M/PDF, FRU
115-017995-00 512M/Glasgow, FRU
115-029555-00 512M/Glasgow/PDF,
115-029556-00 512M/Glasgow,
115-029557-00 512M/PDF, Wi-Fi,
/
/
/
/
FRU
FRU
Wi-Fi, FRU
FRU
/
/
/
/
4 047-010217-00 Display lens 047-010217-00 /
5 9211-20-87221 AC input receptacle / /
6 115-017634-00 BeneHeart R3 top
housing assembly
115-017793-00 BeneHeart R3A top
housing assembly
7 051-001354-01 Keypad board 051-001354-01 /
8 009-003326-00 Cable between main
board and keypad board
9 022-000125-00 Power board 022-000125-00 /
047-010840-00
10
11 009-003327-00 Cable between the main
12 115-018611-00 Parameter Board (FRU) 115-018611-00 /
Power board insulator
(M501)
control board and power
board
115-017634-00 /
115-017793-00
009-003326-00 /
047-010840-00 /
009-003327-00 /
/
8-3
SN PN Description FRU part number Remarks
13 009-003325-00 Cable between the main
control board and
parameter board.
14 115-017792-00 Bottom housing assembly 115-017792-00 /
009-003325-00 /
115-018387-00 Bottom housing assembly
(BeneHeart R3A)
15 022-000122-00 Li-ion battery, 11.1V,
2500 mAh, LI13S001A
16 043-003114-00 Battery door 043-003114-00 /
043-003451-00 Battery door (M501 for
bidding)
17 043-002861-00 Z-fold paper jam
protector
18 115-017633-00 Recorder door assembly 115-017633-00 /
19 043-003112-00 Recorder link 043-003112-00 /
20 043-003110-00 Recorder button 043-003110-00 /
21 DA6H-20-22947 Battery block spring
22 024-000134-00 Thermal print head, 8
dot/mm, 72 mm, 60
mm/sec, FPC lead
115-018387-00 /
022-000122-00 /
043-003451-00 /
043-002861-00 /
DA6H-20-22947 /
024-000134-00 /
23 051-000487-00 TR8A recorder drive
board
24 009-003231-00 Recorder cable 009-003231-00 /
051-000487-00 /
8-4
8.3 Top Housing Assembly (115-017634-00 or
115-017793-00)
8.3.1 Exploded View
8.3.2 Parts List
SN PN Description FRU part number Remarks
1 / Display lens
2 / Upper housing /
3 / Display dust-proof strip, short /
4 / Display dust-proof strip, long /
5 / AC input receptacle and cable /
115-017634-00 or
115-017793-00
/
8-5
8.4 Recorder door assembly (115-017633-00)
8.4.1 Exploded View
8.4.2 Parts List
SN PN Description FRU part number Remarks
1 043-002861-00
2 / Recorder door latch
3 / Reset spring /
4 / Platen /
5 / Overlay with instructions /
6 / Recorder door /
Z-fold paper jam protector
/ /
/
115-017633-00
8-6
8.5 Bottom housing assembly (115-017792-00 or
115-018387-00)
8.5.1 Exploded View
8.5.2 Parts List
SN PN Description
1 / Rubber foot
2 / Recorder instruction labelling /
3 / Lower housing /
4 / Device labelling (Chinese) /
FRU part number
115-017792-00 or
115-018387-00
Remarks
/
8-7
FOR YOUR NOTES
8-8
9 Replacement Parts, Software Upgrade
Package and Accessories
9.1 BeneHeart R3/R3A BOM (V1.0) for Repair, Upgrade
and Accessories
SN PartNumber PartDescription
1. 051-001353-01 PCBA,Low-End 3-CHA ECG Mainboard(ALL)
2. 115-017993-00 Main board(512M/Glasgow/PDF)
3. 115-017994-00 Main board(512M/PDF)
4. 115-017995-00 Main board(512M/Glasgow)
5. 022-000125-00 POWER SUPPLY BOARD 15V 40W
6. 047-010840-00 Main Board Insulation(M501)
7. 115-018611-00 cardiofax signal acquisition board PCBA
8. 051-000487-00 Recorder driver board PCBA
9. 051-001354-01 PCBA,3-Channel ECG KeyBoard
10. 021-000108-00 LCD panel TFT 5" 800*480
11. 024-000134-00 printer 8dot/mm 72mm 60mm/secFPC
12. 115-017634-00 top cover assembly
13. 115-017793-00 top cover assembly
14. 047-010217-00 lens(M501)
15. 049-000544-00 silica gel key(Beneheart R3/icon)
16. 049-000575-00 silica gel key(Beneheart R3A/icon)
17. 115-017633-00 Recorder bin gate assembly
18. 043-002861-00 fan fold paper platen
19. 043-003112-00 recorder link
20. 043-003110-00 recorder button
21. DA6H-20-22947 battery block pole spring
22. 115-017792-00 bottom cover assembly
23. 115-018387-00 bottom cover assembly(beneHeart R3A)
24. 043-003114-00 battery door
25. 043-003451-00 battery door(M501 bid inviting)
9-1
SN PartNumber PartDescription
26. 009-003231-00 recorder connecting cable
27. 009-003325-00 Cable for to parameter
28. 009-003326-00 Cable for to keypad
29. 009-003327-00 Cable for MB to power
30. 0651-20-76879 AC power socket & connecting cable
31. 046-004749-00 Physician's Guide, Glassgow
32. 046-004817-00 Glasgow Physician's Guide(for ECG-EN)
33. 046-004650-00 BeneHeart R3 Operator's Manual (Chinese)
34. 046-004651-00 BeneHeart R3 Operator's Manual (English)
35. 046-004737-00 BeneHeart R3 Operator's Manual (German)
36. 046-004738-00 BeneHeart R3 Operator's Manual (French)
37. 046-004739-00 BeneHeart R3 Operator's Manual (Czech)
38. 046-004740-00 BeneHeart R3 Operator's Manual (Italian)
39. 046-004741-00 BeneHeart R3 Operator's Manual (Spanish)
40. 046-004742-00 BeneHeart R3 Operator's Manual (PTBL)
41. 046-004743-00 BeneHeart R3 Operator's Manual (Brazil)
42. 046-004744-00 BeneHeart R3 Operator's Manual (Russian)
43. 046-004745-00 BeneHeart R3 Operator's Manual (Turkish)
44. 046-004746-00 BeneHeart R3 Operator's Manual (Polish)
45. 046-004747-00 BeneHeart R3 Operator's Manual (Romania)
46. 046-004748-00 BeneHeart R3 Operator's Manual (HU)
47. 046-004653-00 BeneHeart R3 Quick Guide (English)
48. 115-018007-00 Glasgow upgrade Glasgow
49. 115-017996-00 PDF upgrade package
50. 043-004232-00 yoke holder
51. 043-000386-00 Bin,overmolded,value stand
52. 043-004231-00 handle
53. 034-000302-00 casters.3''
54. 043-000389-00 Base cover,value stand
9.2 Material Change
Update the service manual as per the latest BOM.
9-2
A Electrical Safety Inspection
The following electrical safety tests are recommended as part of a comprehensive
preventive maintenance program. They are a proven means of detecting abnormalities that,
if undetected, could prove dangerous to either the patient or the operator. Additional tests
may be required according to local regulations.
All tests can be performed using commercially available safety analyzer test equipment.
These procedures assume the use of a 601PROXL International Safety Analyzer or
equivalent safety analyzer. Other popular testers complying with IEC 60601-1 used in
Europe, such as Fluke, Metron, or Gerb, may require modifications to the procedure. Please
follow the instructions of the analyzer manufacturer.
The consistent use of a safety analyzer as a routine step in closing a repair or upgrade is
emphasized as a mandatory step if an approved agency status is to be maintained. The
safety analyzer also proves to be an excellent troubleshooting tool to detect abnormalities of
line voltage and grounding, as well as total current loads.
A.1 Power Cord Plug
Test Item Acceptance Criteria
The power plug pins No broken or bent pin. No discolored pins.
The plug body No physical damage to the plug body.
The power plug
The strain relief
The power plug No loose connections.
The power cord
No physical damage to the strain relief. No plug
warmth for device in use.
No physical damage to the cord. No deterioration to
the cord.
For devices with detachable power cords, inspect
the connection at the device.
For devices with non-detachable power cords,
inspect the strain relief at the device.
A-1
A.2 Device Enclosure and Accessories
A.2.1 Visual Inspection
Test Item Acceptance Criteria
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
The enclosure and accessories
No residue of fluid spillage (e.g., water, coffee, chemicals,
etc.).
No loose or missing parts (e.g., knobs, dials, terminals, etc.).
A.2.2 Contextual Inspection
Test Item Acceptance Criteria
No unusual noises (e.g., a rattle inside the case).
The enclosure and accessories
No unusual smells (e.g., burning or smoky smells,
particularly from ventilation holes).
No taped notes that may suggest device deficiencies or
operator concerns.
A.3 Device Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and
legible.
Main unit label Integrated warning labels
A.4 Protective Earth Resistance
Protective Earth Resistance is measured using the RED test lead attached to the DUT
Protective Earth terminal or enclosure. Select the test current by pressing SOFT KEY 3 to
toggle 25AMP. The front panel outlet power is turned off for this test.
The following conditions apply: L1 and L2 Open.
A-2
Preparation
1. First select the test current that will be used for performing the Protective Earth
Resistance test by pressing AMPERES (SOFT KEY 3).
2. Connect the test lead(s) between the RED input jack and the GREEN input jack.
3. Press CAL LEADS. The 601PRO will measure the lead resistance, and if less than
0.150 Ohms, it will store the reading and subtract it from all earth resistance readings
taken at the calibrated current.
If the calibration fails, the previously stored readings will be used until a passing calibration
has occurred.
WARNING
During Earth Resistance testing, the DUT must be plugged into the 601PRO
front outlet. If the DUT fails Earth Resistance, discontinue tests and label the device defective.
To Perform the Test
1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO
front panel outlet.
2. Attach the 601PRO RED input lead to the device’s Protective Earth terminal or an
exposed metal area.
3. Press shortcut key 3. The Protective Earth Resistance test is displayed.
4. Press SOFT KEY 3 to select a test current (25AMP). The selected test current is
displayed in the upper right corner of the display.
5. Press START TEST to start the test. The test current is applied while resistance and
current readings are taken. This takes approximately 5 seconds.
6. Press the print data key at any time to generate a printout of the latest measurement(s).
A-3
NOTE
When "Over" is displayed for Ohms, this signifies that a valid measurement was
not obtained because either an open connection was detected or that the measurement was not within range. Readings greater than 9.999 Ohms will be displayed as Over.
In Case of Failure
Once it reaches the limitation, stop using and inform the Customer Service Engineer for
analysis and disposal.
LIMITS
ALL COUNTRIES R = 0.2 Maximum
A.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
Leakage current is measured the following ways:
Earth Leakage Current, leakage current measured through DUT outlet Earth Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth), leakage
current measured through DUT outlet Earth
There is no need to attach a test lead; the 601PRO automatically connects the measuring
device internally.
To Perform the Test
1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO
front panel outlet, and turn on the device.
2. Attach the device's applied parts to the 601PRO applied part terminals if applicable.
3. Press shortcut key 4.The Earth Leakage test appears on the display, and the test begins
immediately:
A-4
SOFT KEY 1 toggles the DUT outlet Polarity from Normal to Off to Reverse. SOFT KEY 2 toggles the DUT outlet from Earth to No Earth. SOFT KEY 3 toggles the DUT outlet from L2 to No L2. SOFT KEY 4 toggles the AP to Earth to No AP to Earth.
4. Press the print data key at any time to generate a printout of the latest measurement.
In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities.
Notify the user or owner to correct any deviations. As a work around, check the other
outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the system problem. Remove
unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
For UL60601-1,
300 μA in Normal Condition 1000 μA in Single Fault Condition
For IEC60601-1,
500 μA in Normal Condition 1000 μA in Single Fault Condition
A-5
A.6 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All
measurements have a true RMS only response.
Preparation
Perform a calibration from the Mains on Applied Part menu.
The following outlet conditions apply when performing this test:
Normal Polarity, Earth Open, Outlet ON Normal Polarity, Outlet ON Normal Polarity, L2 Open, Outlet ON Reversed Polarity, Outlet ON Reversed Polarity, Earth Open, Outlet ON Reversed Polarity, L2 Open, Outlet ON
WARNING
If all of the applied parts correspond to the instrument type, the applied parts
will be tied together and one reading will be taken. If any of the applied parts differ from the instrument type, all applied parts will be tested individually, based on the type of applied part. This applies to Auto and Step modes only.
To Perform the Test
1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO
front panel outlet, and turn on the device.
2. Attach the applied parts to the 601PRO's applied part terminals.
3. Press shortcut key 6. The Patient Leakage test is displayed, and the test begins
immediately.
4. Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part
leakage current.
5. Modify the configuration of the front panel outlet by pressing the appropriate SOFT
KEY on the 601PRO.
6. Press the print data key at any time to generate a printout of the latest measurement.
A-6
In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities.
Notify the user or owner to correct any deviations. As a work around, check the other
outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the system problem. Remove
unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
For CF applied parts
10μA in Normal Condition 50μA in Single Fault Condition
A.7 Mains on Applied Part Leakage
The Mains on Applied Part test applies a test voltage, which is 110% of the mains voltage,
through a limiting resistance, to selected applied part terminals. Current measurements are
then taken between the selected applied part and earth. Measurements are taken with the
test voltage (110% of mains) to applied parts in the normal and reverse polarity conditions
as indicated on the display.
The following outlet conditions apply when performing the Mains on Applied Part test.
Normal Polarity; Reversed Polarity
A-7
Preparation
To perform a calibration from the Mains on Applied Part test, press CAL (SOFT KEY 2).
1. Disconnect ALL patient leads, test leads, and DUT outlet connections.
2. Press CAL to begin calibration, as shown:
If the calibration fails, the previously stored readings will be used until a passing
calibration has occurred. Also, the esc/stop key has no effect during calibration.
3. When the calibration is finished, the Mains on Applied Part test will reappear.
WARNING
A 2-beep-per-second signal indicates high voltage present at the applied part
terminals while a calibration is being performed.
High voltage is present at applied part terminals while measurements are being
taken.
To Perform the Test
1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601
2. Attach the applied parts to the 601PRO applied part terminals.
3. Attach the red terminal lead to a conductive part on the DUT enclosure.
4. Press shortcut key 7. The Mains on Applied Part test is displayed.
5. Select the desired outlet configuration and applied part to test using the appropriate
SOFT KEYS:
6. Press START TEST (SOFT KEY 1) to begin the test.
7. Press the print data key to generate a printout of the latest measurement.
A-8
NOTE
If all of the applied parts correspond to the instrument type, the applied parts
will be tied together and one reading will be taken. If any of the applied parts differ from the instrument type, all applied parts will be tested individually, based on the type of applied part. This applies to Auto and Step modes only.
In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities.
Notify the user or owner to correct any deviations. As a work around, check the other
outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the system problem. Remove
unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
For CF applied parts: 50 μA
A.8 Patient Auxiliary Current
Patient Auxiliary currents are measured between any selected ECG jack and the remaining
selected ECG jacks. All measurements may have a true RMS only response.
Preparation
1. From the MAIN MENU, or with the outlet unpowered, plug the DUT into the 601PRO
front panel outlet, and turn on the device.
2. Attach the patient leads to the 601PRO ECG jacks.
3. Define the Lead Types from the View Settings Option (refer to: Lead Type Definitions
in Section 5 of this chapter).
4. Press shortcut key 8. The Patient Auxiliary Current test is displayed, and the test
begins immediately. Display values are continuously updated until another test is
selected.
A-9
5. Press SOFT KEYS 1-4 to select leakage tests
6. Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part
leakage current:
7. Modify the configuration of the front panel outlet by pressing the appropriate SOFT
KEY on the 601PRO:
8. Press the print data key at any time to generate a printout of the latest measurement.
In Case of Failure
Check any broken of the enclosure. Replace any defective part. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded and is free of other wiring abnormalities.
Notify the user or owner to correct any deviations. As a work around, check the other
outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console. If the leakage current measurement tests fail on a new unit and if situation can not be
corrected, submit a Safety Failure Report to document the system problem. Remove
unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for analysis and
disposal.
LIMITS
For CF applied parts,
10μA in Normal Condition 50μA in Single Fault Condition
A-10
A.9 Scheduled Electrical Safety Inspection
For scheduled electrical safety inspection, test items 1, 2, 3, 4, 5, 6, 7, and 8 included in the
ELECTRICAL SAFETY INSPECTION FORM shall be performed.
ELECTRICAL SAFETY INSPECTION FORM
Location: Technician: Equipment: Control Number: Manufacturer: Model: SN: Measurement equipment /SN: Date of Calibration: INSPECTION AND TESTING Pass/Fail Limit
1 Power Cord Plug
2 Device Enclosure and Accessories
3 Device Labeling
4 Protective Earth Resistance Max 0.2
Normal condition
(NC)
5
Earth
Leakage
Single Fault
condition (SFC)
____μA
____μA
Max:
NC: 300μA(refer to
UL60601-1) *
NC: 500μA(refer to
IEC60601-1) *
SFC: 1000μA
Normal condition
Patient
6
Leakage
Current
(NC)
Single Fault
condition (SFC)
7 Mains on Applied Part Leakage
Normal condition
Patient
8
Auxiliary
Current
(NC)
Single Fault condition
(SFC)
BF____μA
CF____μA
BF____μA
CF____μA
BF____μA
CF____μA
BF____μA
CF____μA
BF____μA
CF____μA
Max:
CF applied part:
NC:10μA, SFC: 50μA
Max:
CF applied part: 50μA
Max:
CF applied part:
NC:10μA, SFC: 50μA
Note: The equipment sold to the United States shall comply with the requirement of
UL60601-1; others shall comply with the requirement of IEC60601-1.
Name/ Signature: __________________________
Date:_____________________________
A-11
A.10 Electrical Safety Inspection after Repair
The following table specifies test items to be performed after the equipment is repaired.
Repair with main unit not disassembled Repair with
main unit disassemble d
When neither power supply
PCBA nor patient
electrically-connected PCBA is
repaired or replaced
When power supply PCBA is
repaired or replaced
When patient
electrically-connected PCBA is
repaired or replaced
When both power supply
PCBA and patient
electrically-connected PCBA are
repaired or replaced
Test items: 1, 2, 3
Test items: 1, 2, 3, 4
Test items: 1, 2, 3, 4, 5
Test items: 1, 2, 3, 4, 6, 7, 8
Test items: 1, 2, 3, 4, 5, 6, 7, 8
A-12
ELECTRICAL SAFETY INSPECTION FORM
Location: Technician: Equipment: Control Number: Manufacturer: Model: SN: Measurement equipment /SN: Date of Calibration: INSPECTION AND TESTING Pass/Fail Limit
1 Power Cord Plug
2 Device Enclosure and Accessories
3 Device Labeling
4 Protective Earth Resistance Max 0.2
Normal
condition(NC)
5
Earth
Leakage
Single Fault
condition(SFC)
____μA
____μA
Max:
NC: 300μA(refer to
UL60601-1) *
NC: 500μA(refer to
IEC60601-1) *
SFC: 1000μA
Normal
Patient
6
Leakage
Current
condition(NC)
Single Fault
condition(SFC)
7 Mains on Applied Part Leakage
Normal condition(NC)
Patient
8
Auxiliary
Current
Single Fault
condition(SFC)
BF____μA
CF____μA
BF____μA
CF____μA
BF____μA
CF____μA
BF____μA
CF____μA
BF____μA
CF____μA
Max:
CF applied part:
NC:10μA, SFC: 50μA
Max:
CF applied part: 50μA
Max:
CF applied part:
NC:10μA, SFC: 50μA
Note: The equipment sold to the United States shall comply with the requirement of
UL60601-1; others shall comply with the requirement of IEC60601-1.
Name/ Signature: __________________________
Date:_____________________________
A-13
FOR YOUR NOTES
A-14
P/N: 046-005000-00 (2.0)
Loading...