Mindray BeneHeart D3, BeneHeart D2 Service Manual

Page 1
BeneHeart D3/BeneHeart D2
Defibrillator/Monitor
Service Manual
Page 2
Page 3

Intellectual Property Statement

intellectual property rights to this product and this manual. This manual may refer to information
protected by copyrights or patents and does not convey any license under the patent rights of Mindray,
nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents
or other rights of third parties.
, , and are the registered trademarks or
trademarks owned by Mindray in China and other countries.
Revision History
This manual has a revision number. This revision number changes whenever the manual is updated due to
software or technical specification change. Contents of this manual are subject to change without prior
notice.
Version number: 2.0
Release time: May 2021
© 2017-2021 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
I
Page 4
Preface
Manual Purpose
This manual provides detailed information about the assembling, dissembling, testing and
troubleshooting of the equipment to support effective troubleshooting and repair. It is not intended to be
a comprehensive, in-depth explanation of the product architecture or technical implementation.
Observance of the manual is a prerequisite for proper equipment maintenance and prevents equipment
damage and personnel injury.
Intended Audience
This manual is for biomedical engineers, authorized technicians or service representatives responsible for
troubleshooting, repairing and maintaining the defibrillator/ monitors
Passwords
Passwords may be required to access different modes. The passwords are listed below:
Installation mode: 888888
Service mode: 332888
Configuration mode: 315666
II
Page 5

Contents

1 Safety ............................................................................................................................................................ 1-1
1.1 Safety Information ...................................................................................................................................................................... 1-1
1.1.1 Dangers ................................................................................................................................................................................ 1-2
1.1.2 Warnings .............................................................................................................................................................................. 1-2
1.1.3 Cautions ............................................................................................................................................................................... 1-3
1.1.4 Notes ..................................................................................................................................................................................... 1-4
1.2 Equipment Symbols ................................................................................................................................................................... 1-4
2 Theory of Operation ..................................................................................................................................... 2-1
2.1 The Basics ....................................................................................................................................................................................... 2-1
2.1.1 Overview ............................................................................................................................................................................. 2-1
2.1.2 Main Functions .................................................................................................................................................................. 2-1
2.2 Components ................................................................................................................................................................................. 2-2
2.3 Main Unit ........................................................................................................................................................................................ 2-2
2.3.1 System Structure .............................................................................................................................................................. 2-3
2.3.2 System Signal Flow .......................................................................................................................................................... 2-4
2.3.3 Signal Sequence of Power-On/Off ............................................................................................................................. 2-5
2.4 Front Housing Assembly .......................................................................................................................................................... 2-5
2.5 Paddle Tray .................................................................................................................................................................................... 2-6
2.6 Rear Housing Assembly ............................................................................................................................................................ 2-6
2.6.1 Power System .................................................................................................................................................................... 2-6
2.6.2 Main Control System ....................................................................................................................................................... 2-7
2.6.3 Therapy System ................................................................................................................................................................. 2-8
2.6.4 Recorder............................................................................................................................................................................... 2-8
2.6.5 Parameter Measurement System ............................................................................................................................... 2-9
2.7 External Device Connectors .................................................................................................................................................... 2-9
3 Unpacking and Installation ......................................................................................................................... 3-1
3.1 Unpacking the Equipment ...................................................................................................................................................... 3-1
3.2 Preparation for Installation ...................................................................................................................................................... 3-2
3.2.1 Preparation for Installation Site................................................................................................................................... 3-2
3.2.2 Electrical Requirements ................................................................................................................................................. 3-3
3.3 Preparation for Power On ......................................................................................................................................................... 3-3
3.4 User Test.......................................................................................................................................................................................... 3-3
4 Testing and Maintenance ............................................................................................................................ 4-1
4.1 Introduction .................................................................................................................................................................................. 4-1
4.1.1 Test Report .......................................................................................................................................................................... 4-2
4.1.2 Recommended Frequency ........................................................................................................................................... 4-2
4.2 Visual Test ....................................................................................................................................................................................... 4-3
1
Page 6
4.3 Power On Test ............................................................................................................................................................................... 4-3
4.4 User Test.......................................................................................................................................................................................... 4-4
4.5 Password for Installation Mode.............................................................................................................................................. 4-5
4.6 Module Performance Tests ...................................................................................................................................................... 4-5
4.6.1 Manual Defibrillation Test ............................................................................................................................................. 4-5
4.6.2 Pacing Test .......................................................................................................................................................................... 4-7
4.6.3 ECG Test ................................................................................................................................................................................ 4-7
4.6.4 Resp Test .............................................................................................................................................................................. 4-8
4.6.5 SpO2 Test .............................................................................................................................................................................. 4-9
4.6.6 NIBP Tests ............................................................................................................................................................................ 4-9
4.6.7 CO2 Module Tests ........................................................................................................................................................... 4-12
4.7 Electrical Safety Tests .............................................................................................................................................................. 4-13
4.8 Recorder Check ......................................................................................................................................................................... 4-13
4.9 Factory Service .......................................................................................................................................................................... 4-13
4.9.1 Password for Service Mode ........................................................................................................................................ 4-13
4.9.2 Accessing Service Mode Menu ................................................................................................................................. 4-13
4.9.3 Impedance Measurement .......................................................................................................................................... 4-14
4.9.4 Device Information ....................................................................................................................................................... 4-15
4.9.5 Checking Failure Code ................................................................................................................................................. 4-15
4.9.6 Inputting Serial Number ............................................................................................................................................. 4-16
4.9.7 Paddle Open Circuit Display ...................................................................................................................................... 4-16
5 Hardware and Software Upgrade ............................................................................................................... 5-1
5.1 Hardware Upgrade ..................................................................................................................................................................... 5-1
5.1.1 Upgrade MPM module ................................................................................................................................................... 5-1
5.1.2 Upgrade the Therapy Module...................................................................................................................................... 5-2
5.2 Software Upgrade through a PC ........................................................................................................................................... 5-3
5.2.1 Installing Mindray Patient Monitor Software Upgrade Tool ............................................................................. 5-4
5.2.2 Software Upgrade Procedure ...................................................................................................................................... 5-6
5.3 Software Upgrade through a USB Disk ............................................................................................................................... 5-7
5.3.1 Applicable ........................................................................................................................................................................... 5-7
5.3.2 Tool ........................................................................................................................................................................................ 5-7
5.3.3 Upgrade Procedure ......................................................................................................................................................... 5-7
6 Troubleshooting ........................................................................................................................................... 6-1
6.1 Overview ........................................................................................................................................................................................ 6-1
6.2 Part Replacement ........................................................................................................................................................................ 6-1
6.3 Checking Defibrillator/Monitor Status ................................................................................................................................ 6-1
6.4 Checking Device Information ................................................................................................................................................. 6-2
6.5 Checking Technical Alarm ........................................................................................................................................................ 6-2
6.6 Troubleshooting Guide ............................................................................................................................................................. 6-2
6.6.1 Defibrillation Problems .................................................................................................................................................. 6-2
6.6.2 Pacing Problems ............................................................................................................................................................... 6-4
6.6.3 Power On/Off Problems ................................................................................................................................................. 6-4
2
Page 7
6.6.4 Display Problems .............................................................................................................................................................. 6-5
6.6.5 Alarm Problems ................................................................................................................................................................. 6-6
6.6.6 Button and Knob Problems .......................................................................................................................................... 6-6
6.6.7 Recorder Problems ........................................................................................................................................................... 6-7
6.6.8 Output Interface Problems ........................................................................................................................................... 6-8
6.6.9 Memory Card Problems ................................................................................................................................................. 6-8
6.6.10 Power Supply Problems ............................................................................................................................................... 6-8
6.6.11 Software Upgrade Problems...................................................................................................................................... 6-9
6.7 Technical Alarm Messages .................................................................................................................................................... 6-10
6.8 Error Codes ................................................................................................................................................................................. 6-15
6.8.1 Therapy Module Error Codes ..................................................................................................................................... 6-15
6.8.2 Power Module Error Codes ........................................................................................................................................ 6-17
6.8.3 Main Control Error Codes ........................................................................................................................................... 6-18
6.8.4 M51C Error Codes .......................................................................................................................................................... 6-19
7 Disassembly and Repair .............................................................................................................................. 7-1
7.1 Tools Required .............................................................................................................................................................................. 7-1
7.2 Preparations for Disassembly ................................................................................................................................................. 7-1
7.3 Disassembling the Main Unit .................................................................................................................................................. 7-2
7.3.1 Removing Hook Mount (if configured) .................................................................................................................... 7-3
7.3.2 Removing Paddle Tray .................................................................................................................................................... 7-3
7.3.3 Separating the Housing ................................................................................................................................................. 7-5
7.3.4 Discharging the Capacitor ............................................................................................................................................ 7-6
7.3.5 Disassembling the MPM Module Assembly ........................................................................................................... 7-6
7.3.6 Removing the Parameter Panel Assembly .............................................................................................................. 7-7
7.3.7 Removing the Therapy Module ................................................................................................................................... 7-7
7.3.8 Disassembling the Power Base Assembly ............................................................................................................... 7-8
7.3.9 Disassembling the AC/DC Power Supply Board .................................................................................................... 7-9
7.3.10 Disconnecting the Therapy Port Cable ............................................................................................................... 7-10
7.3.11 Removing the Main Board Assembly .................................................................................................................. 7-11
7.3.12 Checking Waterproof Strips before Reassembling ......................................................................................... 7-12
7.4 Disassembling the Front Housing Assembly ................................................................................................................. 7-14
7.4.1 Removing the Keypad Board .................................................................................................................................... 7-14
7.4.2 Removing Display Assembly ..................................................................................................................................... 7-15
7.4.3 Removing the Alarm Lamp Board and Assistant Keypad Board .................................................................. 7-16
7.4.4 Removing the Speaker ................................................................................................................................................ 7-17
7.4.5 Removing the Mode Select Knob ........................................................................................................................... 7-17
7.4.6 Removing the Encoder ................................................................................................................................................ 7-18
7.4.7 Checking Waterproof Material on the Front Housing ...................................................................................... 7-18
7.5 Removing the Recorder ......................................................................................................................................................... 7-19
7.6 Disassembling the Recorder ................................................................................................................................................ 7-19
8 Parts .............................................................................................................................................................. 8-1
8.1 Introduction .................................................................................................................................................................................. 8-1
3
Page 8
8.2 Main Unit ........................................................................................................................................................................................ 8-2
8.2.1 Exploded View.................................................................................................................................................................... 8-2
8.2.2 Parts List ............................................................................................................................................................................... 8-2
8.3 Front Housing Assembly .......................................................................................................................................................... 8-3
8.3.1 Exploded View.................................................................................................................................................................... 8-3
8.3.2 Parts List ............................................................................................................................................................................... 8-3
8.4 Rear Housing Assembly ............................................................................................................................................................ 8-4
8.4.1 Exploded View.................................................................................................................................................................... 8-4
8.4.2 Parts List ............................................................................................................................................................................... 8-5
8.5 Paddle Tray Assembly ................................................................................................................................................................ 8-6
8.5.1 Exploded View ................................................................................................................................................................... 8-6
8.5.2 Parts List ............................................................................................................................................................................... 8-6
8.6 External Paddles Assembly ......................................................................................................................................................... 8-7
8.6.1 Exploded View ..................................................................................................................................................................... 8-7
8.6.2 Parts List ............................................................................................................................................................................... 8-7
8.7 Others .............................................................................................................................................................................................. 8-8
A Electrical Safety Inspection ......................................................................................................................... A-1
4
Page 9
NOTE

1 Safety

1.1 Safety Information

DANGER
Indicates an imminent hazard that, if not avoided, will result in death, serious personal injury
or property damage.
WARNING
Indicates a potential hazard or unsafe maintenance practice that, if not avoided, could result in
death, serious personal injury, product / property damage.
CAUTION
Indicates a potential hazard or unsafe maintenance practice that, if not avoided, could result in
minor personal injury or product/property damage
Provides application tips or other useful information to ensure that you can better service your
product.
1-1
Page 10

1.1.1 Dangers

DANGER
The equipment delivers up to 360 J of electrical energy. Unless properly used as described in
these Operating Instructions, this electrical energy may cause serious injury or death. Do not
attempt to operate this defibrillator unless thoroughly familiar with these operating
instructions and the function of all controls, indicators, connectors, and accessories.
Defibrillation current can cause operator or bystander severe injury or even death. Keep
distance with the patient or metal devices connected to the patient during defibrillation.
Do not disassemble the defibrillator. It contains no operator serviceable components and
dangerous high voltages may be present. Contact authorized service personnel for repair.
To avoid explosion hazard, do not use the equipment in the presence of oxygen-rich
atmospheres, flammable anesthetics, or other flammable agents (such as gasoline). Keep the
equipment and the operating environment dry and clean.

1.1.2 Warnings

WARNING
Before putting the system into operation, the operator must verify that the equipment,
connecting cables and accessories are in correct working order and operating condition.
Make sure the synchronous input system is applied to this equipment and the input signal is
correct if necessary.
To avoid risk of electric shock, this equipment must only be connected to a supply mains with
protective earth. If the installation does not provide for a protective earth conductor,
disconnect it from the power line and operate it on smart lithium-ion batteries.
Ensure that the equipment is supplied with continuous electric power during work. Sudden
power failure leads to the loss of patient data.
Use and store the equipment in specified environmental condition. The equipment and
accessories may not meet the performance specification due to aging, stored or used outside
the specified temperature and humidity range.
This equipment is used for single patient at a time.
Medical electrical equipment which does not incorporate defibrillator protection should be
disconnected during defibrillation.
Do not defibrillate a patient who lies on the wet ground.
Do not touch the patient and live parts simultaneously.
Do not touch the patient when connecting the peripheral equipment via the I/O signal ports to
prevent patient leakage current from exceeding the requirements specified by the standard.
1-2
Page 11
Do not rely exclusively on the audible alarm system for patient monitoring. Adjustment of
alarm volume to a low level or off may result in a hazard to the patient. Remember that alarm
settings should be customized according to different patient situations and always keeping
the patient under close surveillance is the most reliable way for safe patient monitoring.
Do not perform any functional check if the equipment is connected with a patient; otherwise
the patient might be shocked.
Remain attentive to the patient during applying therapy. Delay in delivering a shock may
result in a rhythm that was analyzed as shockable converting spontaneously to non-shockable
and could result in inappropriate delivery of a shock.
For the treatment of patients with implantable pacemakers, place therapy pads or paddles
away from internal pacemaker generator if possible to help prevent damage to the pacemaker.
To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard.
Wrap and secure excess cabling to reduce risk of entanglement or strangulation by patients or
personnel.
Do not touch device connectors, recorder print head, battery connector or other live
equipment if in contact with the patient; otherwise patient injury may result.
To ensure patient safety, use only parts and accessories specified in this manual.
Package material may contaminate the environment. Properly dispose of the package material
according to applicable waste control regulations and keep it out of children’s reach.

1.1.3 Cautions

CAUTION
Use of Manual Therapy security password requires the clinician to know and remember the
password. Failure to enter correct password will prevent the delivery of manual defibrillation,
synchronized cardioversion and pacing therapy.
At the end of its service life, the equipment, as well as its accessories, must be disposed of in
compliance with the guidelines regulating the disposal of such products to avoid
contaminating the environment.
Magnetic and electrical fields are capable of interfering with the proper performance of the
equipment. For this reason make sure that all external devices operated in the vicinity of the
equipment comply with the relevant EMC requirements. Mobile phones, X-ray equipment or
MRI devices are a possible source of interference as they may emit higher levels of
electromagnetic radiation.
Before connecting the equipment to the power line, check that the voltage and frequency
ratings of the power line are the same as those indicated on the equipment’s label or in this
manual.
Always install or carry the equipment properly to avoid damage caused by drop, impact,
strong vibration or other mechanical force.
1-3
Page 12
NOTE
Dry the equipment immediately in case of rain.
Never charge and deliver shock frequently in non-clinical situations. Otherwise equipment
damage could occur.

1.1.4 Notes

The equipment use a mains plug as isolation means to the mains power supply. Do not locate
the equipment in a place difficult to operate the mains plug.
During normal use, the operator shall stand in front of the equipment.
Keep this manual in the vicinity of the equipment so that it can be obtained conveniently when
needed.
If the equipment is run on a DC power supply, a DC/AC adapter we supply should be used.
This manual describes all features and options. Your equipment may not have all of them.

1.2 Equipment Symbols

See the BeneHeart D3/ BeneHeart D2 Defibrillator/Monitor Operator’s Manual (P/N: 046-010599-00)
for information about the symbols used on this product and its packaging.
1-4
Page 13

2 Theory of Operation

2.1 The Basics

2.1.1 Overview

The BeneHeart D3/BeneHeart D2 defibrillator/monitor (hereinafter called the equipment) provides four
operating modes: Manual Defib, AED, Pacer, and Monitor. The equipment is for use in hospital and
pre-hospital settings. It adopts the most advanced biphasic defibrillation technology and can deliver up
to 360J of defibrillation energy.
The equipment has an 7.0 inch color TFT LCD display with LED Backlight.

2.1.2 Main Functions

The equipment has the following main functions:
Manual Defib Mode
In Manual Defib Mode, the operator analyzes the patient’s ECG, and, if appropriate, follows this procedure:
1 Select the Manual Defib mode, adjust the energy level if necessary
2 Charge; and
3 Deliver the shock.
Defibrillation may be performed through external paddles or multifunction electrode pads. In Manual
Defib Mode, you can also perform synchronized cardioversion.
AED Mode
In AED mode, the equipment automatically analyzes the patient’s ECG rhythm and indicates whether or
not a shockable rhythm is detected. Voice prompts provide easy-to-follow instructions and patient
information to guide you through the defibrillation process. Messages and flashing buttons are also
presented to reinforce the voice prompts.
Pacer Mode
The Pacer Mode offers non-invasive transcutaneous pacing therapy. Pace pulses are delivered through
multifunction electrode pads using a monophasic square waveform.
Monitor Mode
In Monitor Mode, the equipment is intended for monitoring, displaying, reviewing, storing and printing
multiple physiological parameters and waveforms including ECG and pulse oximetry (SpO2),
2-1
Page 14

2.2 Components

The equipment consists of a main unit, accessories and PC software.
The main unit is the core of the equipment. It provides:
Overall system control;
System power supply;
Display;
Defibrillation and pacing;
AED ;
Man-mahcine interface;
Audible and visible alarms;
Multiple parameter measurements;
External connectors and communication; and
Recording, printing and data storage.

2.3 Main Unit

The main unit is composed of the front housing assembly, rear housing assembly and the paddle tray
assembly. External paddles are rested in the paddle tray.
The front housing assembly mainly consists of LCD, keypad board, speaker, microphone, Mode Select
knob, navigation knob, alarm lamp board, front housing and front housing sheet metal, etc.
The rear housing assembly consists of CPU board, therapy module, high voltage capacitors, MPM
module, power management board, fan, measurement module panel, therapy port, recorder and
rear housing, etc.
The paddle tray is for holding the external paddles.
2-2
Page 15
B13 knob
B14 speaker
Keypad board
LCD
backlight
Indicator
Therapy
module
Paddle tray
assembly
External paddle
in-position
detection
Multi-
parameter
module
CPU and power management board
External paddle
tray
Therapy connector
50-ohm discharging
resistor
Multi-
parameter
connector panel
External paddles
Self-discharging
resistor
W8
W1
F r
o n t h o u
s i
n g
a s
s e m
b l
y
External component
W13
W14
RJ45
USB
Externally
extended IO
W2
W4
W5
W6
B15 microphone
W3
AC/DC module
C3
C4
Recorder (D3)
Battery
W11
W9
W10
R e
a r
h o u s
i
n g
a s
s e m
b l
y
indicates connection cable.
indicates hard connection socket.
High-voltage
capacitor
B12 mode
selection switch
LCD
control
C2
CPR
module
W19
M02D
D6 interface
board
C1
Inductor
Pacing board
W20
W18
W12
W17
Alarm lamp
W7
W15
CO2
watertrap
W16
Recorder (D6)
W11
Notes:
1. The D3 recorder is placed on the rear housing while the D6 recorder is placed on the front housing.
2. For D6, the multi-parameter module and the parameter connector panel are connected directly in hard connection mode; for D3, the multi-parameter module and the parameter connector panel are connected through wires.
3. The D6 needs to be connected to an external interface through the D6 interface board.

2.3.1 System Structure

2-3
Page 16
CPU and power
management board
Power management
Main
control
AM3352
MSP430
UART
Lithium
batteries
ACDC
module
Power
Therapy
module
GPIO
M51C
Module
UART+GPIO +Power
GPIO
Navigation
knob
Mode
switch
Alarm
lamp
MicrophoneSpeaker
LCD
CPR sensor
Recorder
module
M02D
module
UART+ Power
M0
Keypad board
RS232+Power +GPIO
URAT+Power
GPIO+Power
GPIO+ Power
LVDS+ Power
GPIO+ Power
UART+ GPIO+Power
LVDS+Power+GPIO
SmBUS+ GPIO+ Power

2.3.2 System Signal Flow

The system uses the CPU and power management board as the core. The CPU and power management
board is divided into the power management part and the main control part. The processor of the power
management part is MSP430, which mainly implements communication with external lithium batteries
and power management. The processor of the main control part is AM3352, which communicates with
and controls other subsystem modules (including the therapy module, M51C module, M02D module,
recorder module, CPR sensor, and keypad board on the front housing). The main control module
communicates with other subsystems through asynchronous serial ports. The keypad board on the front
housing implements the transit between man-machine interaction interfaces and signal display. The
processor is M0, which controls input and output signals of keys, knobs, module switches, and alarm
lamps.
2-4
Page 17
Keypad
board
MCU
TPS2490
Dual-line
DC/DC
PCON
ENABLE
RUN
CPU and power management board
EXTLD
Single-line
DC/DC
VCC_EN

2.3.3 Signal Sequence of Power-On/Off

The power-on/off action is triggered by the naviagation knbo on the keypad board of the front housing.
The power-on/off process is controlled by the power management processor as follows:
Power-on process: When you rotate the mode switch on the front panel of the defibrillator/monitor to a
mode other than "Off", the PCON signal value is changed from 1 to 0 to awake the MCU. The MCU
immediately outputs the ENABLE control signal to the TPS2490 enabling the VBUS. Then, the MCU enables
the RUN signal control of the dual-line DC-DC module to output 5 V (VBB) and 12 V signals.
Power-off process: When your rotate the mode switch to "Off", the PCON signal value is changed from 0 to
1, and the MCU outputs the ENABLE=0 and RUN=0 signals to shut down the VBUS and the DC-DC module.

2.4 Front Housing Assembly

The front housing assembly consists of display assembly, a keypad board, a speaker, a microphone, a
Mode Select knob, a navigation knob, an alarm lamp board, a front housing and front housing sheet metal,
etc.

Navigation Knob

You can rotate the knob clockwise or counterclockwise and then press it to confirm a selection. The knob
is connected to the keypad board.

Mode Select Knob

A 8-position encoder is used to select the operating mode (Monitor, Manual Defib, AED and Pacer) and
power-off. The unused positions are mechanically disabled.

Speaker

The speaker emits alarm tones, key-stroke tone, heart beats and PR sound. It supports the functions of
PITCH TONE and the multi-level volume. The speaker is connected to the keypad board.
2-5
Page 18
AC-DC
BAT1
BAT2
Dual-battery management
system
Charging circuit
3.3V LDO
MCU
Vbus
5V DC-DC
12V DC-DC
Multi-function port (reserved)
Speaker
LCD
M51C module
CPR Sensor
USB flash drive
Main control
3G/4G module
Keypad boards
on left and right
electrode pads
Keypad board
on front housing
3.3V DC-DC Bluetooth
Wifi
M02D module
Recorder
Therapy board
VAC
BC SMBC SMBD
AC and battery in-position and energy detection
EXTLD
AD sampling
Overvoltage
protection
Buzzer
alarm
3.3V, 5V, 12V
AD sampling at 3.3 V, 5 V, 12 V, and 18 V
3.3 V, 5 V, and 12 V overvoltage protection
18V
18V
12-17V
Dual-battery switching control
Overcurrent and power
protection

Microphone

It provides the function of voice recording.

Alarm Lamp Board

The keypad board interfaces with the alarm lamp board. The alarm lamp transmits signals to drive the
green and yellow alarm lamp. The drvie current 60mA.

2.5 Paddle Tray

The paddle tray is used to hold paddles. It has a 50 ohm test load and position detective switch inside.
When the equipment runs self tests, test current will pass through the test load.

2.6 Rear Housing Assembly

The rear housing assembly consists of the CPU and power management board, therapy module, high
voltage capacitors, M51C module, recorder, rear housing, parameter receptacle panel, and therapy port.

2.6.1 Power System

2-6
Page 19
CPU and power management board
Power management
Main control
AM3352
MSP430
UART
AC/DC board
It has AC mains as an input and outputs 18VDC.
Battery
Its rated voltage is 14.8V, 3000mAh. Or, its rated voltage is 15.1V, 5600mAh.
CPU and power management board
It manages the system power input, supplies different system power, and monitors the power status. The
power management part implements battery charging and status information management.
The priority of system power supply is AC mains, Battery. That is to say, when AC is not available, Battery is
used.

2.6.2 Main Control System

Serving as the core of the system, the main control system implements the display, storage, printing, and
review of parameters and waveforms, as well as parameter algorithm processing.
The integrated CPU and power management board consists of the MSP430 for power management and
the AM3352 for main control. The MSP430 monitors the power status and controls the power-on/off; the
AM3352 implements man-machine interaction, screen display, and AED algorithm. The main control
module part and the power management part communicate with each other through asynchronous serial
ports.
2-7
Page 20
MCU
Defibrillation charging and
discharging
Cable type
identification
ASIC
Pace detection
High-pass +
amplification
Anti-
defibrillation
Hardware
low-pass
Impedance
coupling
p-lead
Keypad
identification/lamp
8IO
High voltage portion of
defibrillation and pacing
2CH current
detection
2CH voltage
detection
2CH relay
1 wire
50-ohm self-
discharging resistor
2CH relay
MCU
Uart
AD
2CH relay
ECG input
Keypad board interface
Recorder keypad board
Power interface
Signal interface
Button & indicator interface
Recorder power module
Recorder CPU
Motor drive circuit
Power control
Motor control
FPC interface
Thermal print head

2.6.3 Therapy System

The therapy system implements the measurement of P-lead input ECG and human body impedance, as
well as the defibrillation and pacing functions. The pacing function is implemented by the optional pacing
module.

2.6.4 Recorder

The recorder implements the parameter and waveform printing function by using the universal TR6F
recorder module.
The recorder receives data from the CPU board and then sends the data to a thermal head for printing.
The recorder front panel has a key for starting/ stopping the recorder and a green indicator which is lit
when working normally. The recorder is connected to the keypad board which board provides connection
for the TR6F recorder. The block diagram and functional modules of the recorder are shown as below.
2-8
Page 21
Module Description
Power Interface Adjusts input voltage to run each module.
Recorder CPU Coordinates module communication, controls and processes module status.
Keypad board Interface
Motor Drive Circuit Receives control signals sent by the recorder CPU to drive the step motor.
Keypad and Indicator
Interface
FPC Interface
Serves as the data communication channel between the keypad board and
the recorder CPU.
Sends keypad commands to CPU and receives CPU commands to control the
indicator.
Sends print head information to CPU and receives CPU commands to control
the print head.

2.6.5 Parameter Measurement System

The M51C multi-parameter module implements the parameter measurement function. It mainly
implements 3/5-lead ECG, SpO2, NIBP, and RESP measurement functions.

2.7 External Device Connectors

1
2
3
4
5
6
7
2-9
Page 22
1. Hook
2. Battery
3. Equipotential grounding terminal: When the defibrillator/monitor and other devices are to be used
together, their equipotential grounding terminals should be connected together to eliminate the
potential difference between them.
4. External power input: It connects an AC power cord or a DC/AC adapter to run the equipment
respectively on the external AC mains or DC power supply.
5. Multifunctional connector: It connects a CPR sensor, provides ECG output and defib synchronization
input.
6. USB connector
7. Network connector: It is a standard RJ45 connector.
2-10
Page 23

3 Unpacking and Installation

This chapter provides information you need to install a defibrillator/monitor ready for use.

3.1 Unpacking the Equipment

Open the package and take out the packing list. Check that all the articles included in the packing list are
available and the quantity and specification are correct.
All the optional parts purchased by the customer shall also be checked.
Notify the supplier if provided components are not correct as compared to the packing list.
In case of damage during transportation, keep the packing material and notify the supplier
immediately.
Keep the packing material till new equipment is accepted.
The following pictures show the defibrillator/monitor and accessory packing.
Main unit packing
Accessory packing
3-1
Page 24
CAUTION

3.2 Preparation for Installation

3.2.1 Preparation for Installation Site

1. Ensure that the site meets all safety, environmental and power requirements
2. Check that required power sockets are available.
3. Check that a network connector is available if the defibrillator/monitor needs to be connected to
network.
WARNING
Only power cables provided with the system may be used. For reasons of safety, power (mains)
extension cables or adapters shall not be used.
Environmental Requirements
WARNING
To avoid explosion hazard, do not use the equipment in the presence of flammable
anaesthetics, vapours or liquids.
The environment where the defibrillator/monitor will be used should be reasonably free from
vibration, dust and corrosive substances. If these conditions are not met, the accuracy of the
system may be affected and damage may occur.
The environmental specification is as follows:
Operating environment
0 to 45℃ (at least 60 minutes of working time when the temperature
Operating Temperature
Operating humidity 10% to 95%, (non-condensing)
Operating altitude
Storage environment
Storage temperature
Storage humidity 10% to 95%, (non-condensing)
Storage altitude
reduces from room temperature to – 20)
(5 to 40 for CO
-381mmHg to + 4575mmHg (106.2kPa to 57kPa)
430mmHg to + 790 mmHg for CO
-30 to 70(-20 to 60 for CO2 module)
-381mmHg to +4575 mmHg (106.2kPa to 57kPa)
430mmHg to + 790 mmHg for CO
module)
2
module (57.3kPa to 105.3kPa)
2
module (57.3kPa to 105.3kPa)
2
3-2
Page 25

3.2.2 Electrical Requirements

Check cables and power cords. Make sure that:
1. All system cables, power cords and power plugs are not damaged, and pins are not loose. Otherwise,
remove it from use.
2. The insulation of patient cables and leadwires is not damaged, and connectors are not loose.
WARNING
Only power sockets with protective grounding can be used.
The electrical specification is as follows:
Line voltage: 100 to 240VAC (±10%)
Current: 1.8 A
Frequency: 50/60Hz (±3Hz)

3.3 Preparation for Power On

Before connecting the power cord to the defibrillator/monitor's power input, check that
The mains voltage meets the requirement.
3-wire power cord is used. The power socket should be 3-wire also. This ensures that the
defibrillator/monitor is properly grounded. Do not use 2-wire power cord or socket.
The equipotential grounding terminals should be connected together when the defibrillator/monitor
and other devices are to be used together.
The defibrillator/monitor is not placed under the infusion bag or placed where their might be liquid
spillage. This protects the defibrillator/monitor from liquid ingress.

3.4 User Test

A user test shall be performed after the defibrillator/monitor is installed. Follow this procedure:
1. Connect AC mains or install the battery.
2. Connect the external paddles. If pads are used, connect the test load.
3. Select the Main Menu button on the equipment’s front panel and select [User Test >>]. Select all test
items and press [Start] to perform user test.
NOTE
Install the battery and properly place the external paddles in the paddle tray or connect the
pads cable and 50 Ω test load. Otherwise the User Test will fail.
3-3
Page 26
See the BeneHeart D3/ BeneHeart D2 Defibrillator/Monitor Operator’s Manual (P/N: 046-010599-00)
for the detailed information on user test.
3-4
Page 27

4 Testing and Maintenance

4.1 Introduction

To ensure the equipment always functions normally, qualified service personnel should perform regular
inspection, maintenance and test. This chapter provides a checklist of the testing procedures for the
equipment with recommended test equipment and frequency. The service personnel should perform the
testing and maintenance procedures as required and use appropriate test equipment.
The testing procedures provided in this chapter are intended to verify that the equipment meets the
performance specifications. If the equipment or a module fails to perform as specified in any test, repairs
or replacement must be done to correct the problem. If the problem persists, contact our Customer
Service Department.
CAUTION
All tests should be performed by qualified service personnel only.
Care should be taken to change the settings in [Installation Mode] and [Service Mode] menus
to avoid loss of data.
Before testing, service personnel should acquaint themselves with the test tools and make sure
that test tools and cables are applicable.
When testing monitoring parameters, move the Mode Select knob to Monitor to access the
Monitor Mode.
When performing therapy function tests, move the Mode Select knob to corresponding mode.
4-1
Page 28

4.1.1 Test Report

After completing the tests, service personnel can record test results by referring the Test Report at the end
of this chapter.

4.1.2 Recommended Frequency

Test item After
repair
Visual inspection
Function
suspected 6 months
12
months
24
months
Power-on Test
User test
Function Checks Recorder check
ECG Cable Test
Manual
defibrillation
tests
Pacing test
Preventive
Maintenance
(Only Performance
Tests)
Performing Tests in
the Installation
Mode
ECG Performance test
Resp Performance test
SpO2 Performance test
Maintain NIBP Performs the NIBP
Maintain CO2 Calibrates the CO2
Version Displays the
CPR Displays the
Format Data
Card
Watchdog Test Checks if the
Modify
Password
Charge/
discharge
Energy disarming
Synchronous
defibrillation
Module calibration
leakage test and
NIBP accuracy test
module.
equipment
information
battery information
of the CPR sensor
Formats the
storage card.
equipment can be
normally restarted.
Modifies the
password for
4-2
Page 29
Test item After
repair
Performing Tests in
the Installation
Mode
Electrical safety
tests as per
IEC60601-1
entering the
installation mode.
Earth leakage
current
Patient leakage
current
Patient auxiliary
current

4.2 Visual Test

Inspect the equipment for obvious signs of damage. The test is passed if the equipment has no obvious
signs of damage. Follow these guidelines when inspecting the equipment:
Carefully inspect the housing, the display screen and the buttons for physical damage.
Inspect accessories for signs of damage.
Inspect all external connections for loose connectors, bent pins or frayed cables.
Function
suspected 6 months
12
months
24
months
Inspect all connectors on the equipment for loose connectors or bent pins.
Make sure that safety labels and data plates on the equipment are clearly legible.

4.3 Power On Test

This test is to verify that the defibrillator/ monitor can power on normally. The test is passed if the
defibrillator/ monitor starts up by following this procedure:
1. Place the external paddles on paddle tray, insert the battery in the battery compartment, and then
connect the equipment with AC mains. In this case, both the AC indicator and battery indicator shall
light.
2. Turn the Mode Select knob to Monitor. Check that the equipment passes the self test and is turned on
properly.
3. Check the display of technical alarm area, prompt area and battery status indicator on the upper right
corner of the main screen to judge whether the equipment runs normally.
If a power-on self test error happens, the service indicator is illuminated, and alarm messages are
displayed in the technical alarm area.
4-3
Page 30

4.4 User Test

Follow this procedure to perform user test:
1. If you use external paddles, place them on the paddle tray; if you use a pads cable, connect it to the
test load.
2. Insert the battery into the equipment. Connect the AC mains if no battery is available.
3. Select the Main Menu button on the equipment’s front panel. In the Main Menu, select [User Test>>].
Then a prompt “Enter user test?” pops up. Select “Yes” to enter the User Test screen.
4. Check the test items you want to perform and select [Start] to start user test
The test results indicate the condition of the system. If any item fails, the Red Cross status indicator flashes.
If you cannot pass User Test or the message “Connect paddles cable, and place paddles in paddle tray” is
shown when paddle cable is connected and paddles are placed in paddle tray, check paddles status.
Select the Monitor mode. Press and hold the [Event] hardkey, and then press the [Lead Select] hardkey on
the front panel, the following screen appears.
Observe the reading of “Lead Stat”:
0 x 382: Paddles are properly placed in paddle tray.
0 x 182: The travel switch indicating paddle status may fail, but impedance is correct.
0 x 102 :Paddles are not properly placed in paddle tray and the impedance value is not correct.
4-4
Page 31

4.5 Password for Installation Mode

Accessing installation mode is password protected. The required password is set to 888888 before the
equipment leaves the factory.

4.6 Module Performance Tests

4.6.1 Manual Defibrillation Test

Test tools:
Defibrillator/pacer analyzer
Charge/Discharge
1. Remove the batteries and connect the equipment with AC mains. Turn the Mode Select knob to
Manual Defib.
2. Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer
analyzer.
3. Enter the Configuration-Main screen. From the Record Setup menu set [Shock Event] to [On] so that
shock events can be recorded automatically if happened.
4. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as
0 or blank.
5. Select the energy level to 1J.
6. Charge/discharge the equipment to verify the energies measured by the analyzer meet the following
accuracy:
Selected Energy (J) Measured Value (J)
1 0 to 3
100 85 to 115
360 306 to 414
7 Set the energy to 100J and 360J respectively. Repeat step 6.
8. Disconnect the equipment from the AC mains. Run the equipment on fully charged battery. Move the
Mode Select knob to Manual Defib. Repeat steps 5 to 7.
9. Use multifunctional electrode pads. Repeat steps 5 to 7.
10. Verify that the equipment records the shock events automatically and correctly.
4-5
Page 32
Energy Disarming
1. Run the equipment on fully charged battery. Move the Mode Select knob to Manual Defib.
2. Connect the external paddles to the equipment and place the paddles on the defibrillator/pacer
analyzer.
3. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as
0 or blank.
4. Select the energy level to 360J.
5. Charge the equipment.
6. Verify that the charge tone is issued during charging.
7. Press the “Disarm” soft key to discharge the energy internally.
8. Verify that a prompt “Charge Removed” appears and the charge done tone stops.
9. Verify that the value measured by the analyzer is 0J or blank.
10. Enter the Configuration-Main menu, select [Manual Therapy Setup] and set [Time to Auto Disarm]
to [60s].
11. Exit “Configuration Management”. The equipment restarts automatically.
12. Set the analyzer to Energy Measurement mode. In this case, the energy value should be displayed as
0 or blank.
13. Select the energy level to 360J.
14. Charge the equipment. Count time after charging is completed.. Verify that the prompt “Shock
Removed” appears on the equipment and the energy measured by the analyzer is 0J or blank after 60
seconds.
15. Use multifunctional electrode pads. Repeat steps 3 to 14.
Synchronous Defibrillation
1. Connect the external paddles and ECG cable to the equipment. Place the paddles ECG electrodes on
the defibrillator/pacer analyzer.
2. Set the analyzer to Measurement Mode and output normal sinus rhythms, e.g. amplitude value 1mV
and HR 60bpm.
3. Enter Configuration Management. In the [Manual Therapy Setup] menu, set [Sync After Shock] to
[On].
4. Adjust the energy setting of the equipment to be 10J.
5. Press the [Sync On] soft key to start synchronous defibrillation. If Remote Sync is switched on, press
the [Sync On] soft key and select [Local] to start synchronous defibrillation
6. Select [Pads] or [Paddles] as the ECG source and begin charging.
7. When charging finishes, press and hold the “Shock” button to deliver a shock.
8. Verify that synchronous discharge succeeds and the delivery energy measured by the analyzer is
10J±2J.
9. Verify that the delay time of synchronous defibrillation measured by the analyzer is less than 60ms.
4-6
Page 33
10. Verify that the synchronous discharge mark appears on the R wave.
11. Verify that the prompt messages are correct during testing.
12. Select lead II as ECG source and perform charging. Repeat steps 7 to 11.
13. Use multifunctional electrode pads. Repeat steps 2 to 12.

4.6.2 Pacing Test

Test tools:
Defibrillator/pacer analyzer
1. Run the equipment on fully charged battery. Move the Mode Select knob to Pacer. Set [Pacer Mode]
to [Fixed].
2. Connect the pads cable to the equipment and properly place the pads on the defibrillator/pacer
analyzer.
3. Set the analyzer to Pacing Measurement mode. Use test load of 50Ω.
4. On the equipment, set [Pacer rate] to [70ppm] and [Pacer Output] to [30mA].
5. Press the [Start Pacing] soft key. Verify that the pacer rate measured by the analyzer is 70
ppm±1ppm and the pacer output measured is 30 mA±5mA.
6. Press the “Stop Pacing” soft key, and then set [Pacer rate] to [170ppm] and [Pacer Output] to
[200mA].
7. Press the [Start Pacing] soft key. Verify that the pacer rate measured by the analyzer is 170
ppm±2ppm, and the measured current is 200 mA±10mA.

4.6.3 ECG Test

Performance Test
Test tools
ECG simulator
1. Connect the simulator to the equipment’s ECG connector with ECG leadwires.
2. Set the patient simulator as follows: ECG sinus rhythm, HR=80 bpm with the amplitude as 1mV.
3. Check the ECG waves are displayed correctly without noise and the displayed HR value is within 80 ±
1 bpm.
4. Disconnect the simulator from the equipment’s ECG connector. Verify that ECG Lead Off alarm
behaves correctly.
5. On the equipment, set [Paced] to [Yes], the simulator is configured as pace signals. Verify that pace
signals are detected and pace pulse marks are displayed.
6. Connect the simulator to the equipment’s therapy module with pads.
7. Set the patient simulator as follows: ECG sinus rhythm, HR=80 bpm with the amplitude as 1mV.
4-7
Page 34
8. Check the ECG waves are displayed correctly without noise and the displayed HR value is within 80 ±
1 bpm.
9. Disconnect the simulator from the equipment’s therapy module. Verify that ECG Lead Off alarm
behaves correctly.
10. On the equipment, set [Paced] to [Yes], the simulator is configured as pace signals. Verify that pace
signals are detected and pace pulse marks are displayed.
ECG Calibration
Tool required:
Vernier caliper
1. Connect the simulator to the equipment’s ECG connector with ECG leadwires.
2. Select the ECG parameter area to enter the [ECG Setup] menu.
3. Select [Others>>] [Calibrate]. A waveform signals appear on the screen and the message [ECG
Calibrating] is displayed in the prompt information area in the lower left corner of the screen.
4. Compare the amplitude of the waveform with the wave scale. The difference should be within 5%. If
needed, you can also print out the waveform and the wave scale.
5. After ECG calibration is completed, select [Stop Calibrating].
6. Connect the simulator to the equipment’s therapy module with pads.
7. Repeat steps 3–5.

4.6.4 Resp Test

Test tools
Resp Patient simulator
1. Connect the patient simulator to the ECG connector on the module.
2. On the defibrillator/monitor, select the Resp widow to enter the Resp Setup menu. Set [Lead] to [II].
3. Configure the simulator as follows: set Lead to II, base impedance line to 1500 Ω; delta impedance to
0.5 Ω, and respiration rate to 40 rpm.
4. Check that respiration waveform is not distorted and the displayed Resp value does not exceed 40±2
rpm.
4-8
Page 35
Defibrillator/monitor
Connector for
NIBP cuff
Balloon pump
Metal

4.6.5 SpO2 Test

Test tool
Patient simulator.
1. Connect the patient simulator to the equipment’s SpO
2. Select the model and manufacturer of the SpO
2
connector.
2
module under test. Configure the parameter as SpO
96% and PR 80 bmp.
3. The displayed SpO
Mindray
Masimo
and PR values should be within the ranges listed below
2
SpO2 (%) PR (bmp)
96% ±2% 80±3
96% ±2% 80±3
MAX-A, MAX-N, MAX-P, MAX-I 96% ±2%
Nellcor
DS-100A, OXI-A/N, OXI-P/I 96% ±3%
80±3

4.6.6 NIBP Tests

Accuracy Test
The NIBP accuracy test is required at least once every two years or whenever you doubt the NIBP reading.
Tools required:
T-shape connector
2
Tubing
Balloon pump
Metal Vessel, volume 500±25 ml
Calibrated manometer for reference, accuracy not lower than 1 mmHg
To perform the accuracy test:
1. Connect the equipment as shown below.
Manometer
Tubing
vessel
2. Before inflation, the reading of the manometer should be 0. If not, disconnect the airway and
reconnect it until the readings is 0.
4-9
Page 36
monitor
onnector for
NIBP cuff
Cuff
3. Press the Main menu button on the equipment’s front panel. Select [Others>>] [Installation
Mode>>] enter the required password [Maintain NIBP] [Start Accuracy Test].
4. Compare the value of manometer with the value displayed on the equipment’s screen. The difference
should be no greater than 3 mmHg.
5. Raise the pressure in the metal vessel to 50 mmHg with the balloon pump. Repeat steps 3 and 4.
6. Raise the pressure in the metal vessel to 200 mmHg with the balloon pump. Repeat steps 3 and 4.
Note
You can replace the balloon pump and manometer with a blood pressure simulator to form a
test system.
NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once every
two years or whenever you doubt the NIBP reading.
Tools required:
An adult cuff
An air tubing
A correct sized cylinder
To perform the leakage test:
1. Connect the cuff to the equipment’s NIBP connector.
2. Wrap the cuff around the cylinder as shown below.
Defibrillator/
C
3. Press the Main menu button on the equipment’s front panel. Select [Others>>] [Installation
Mode>>] enter the required password [Maintain NIBP] [Start Accuracy Test].
After about 20 seconds, the equipment automatically deflates. This means the leakage test finishes.
When the accuracy test is completed, the result will be displayed. If the message [NIBP Pneumatic Leak]
is displayed, it indicates that the NIBP airway may have leakages. Check the tubing and connections for
leakages, and then perform a leakage test again.
Tubing
Cylinder
4-10
Page 37
onnector for NIBP
Metal
When you select the [Start Leakage Test] button, it turns to be [Stop Leakage Test]. Select [Stop Leakage
Test], leakage test stops and the button turns to be [Start Leakage Test] again.
Calibrating NIBP
Tools required:
T-shape connector
Tubing
Balloon pump
Metal Vessel, volume 500±25 ml
Calibrated manometer, accuracy higher than 1 mmHg
1. Connect the equipment as shown below
Defibrillator/monitor
C
Tubing
Manometer
cuff
Balloon pump
vessel
2. Before inflation, the reading of the manometer should be 0. If not, disconnect the airway and
reconnect it until the readings is 0.
3. Press the Main menu button on the equipment’s front panel. Select [Others>>] [Service Mode>>]
enter the required password [Calibrate NIBP].
4. Calibrate pressure. To do so, set the calibration value to 150 mmHg and adjust the pump output
pressure to 150 mmHg. After the system is stable, click the [Calibrate] button to start calibration.
5. Calibrate overpressure. To do so,
Set [Patient Cat.] to [Adu/Ped] and adjust pump output pressure to 330 mmHg. Click the
[Calibrate] button and start calibration. Or
Set [Patient Cat.] to [Neo] and adjust pump output pressure to 165 mmHg. Click the [Calibrate]
button and start calibration.
All the calibration results will be displayed in the [Calibrating NIBP] screen. If the calibration fails, please
check the connections and then perform a calibration again.
4-11
Page 38
Gas cylinder
as valve

4.6.7 CO2 Module Tests

Leakage Check
1. Access the [CO2 Setup] menu and set [Operating Mode] to [Measure]. Wait for CO
2. Block the CO
and Microstream CO
Sidestream: Check that alarm message [CO2 Filter Line Err] is displayed on the screen in 3s.
module gas inlet completely. This will cause different reactions from the Sidestream
2
modules.
2
warm-up.
2
Block the gas inlet for another 30s, if the alarm message does not disappear, the module does
not leak.
Microstream CO
module: [CO2 Purging…] is displayed in 3s. Block the gas inlet for another 30s,
2
if the alarm message [CO2 Tubing Err] appears, the module does not leak.
Module Calibration
Test tools
Gas cylinder, with 6% of CO
and balance gas N2.
2
T-shape connector
Tubing
For sidestream CO2 module, zeroing is required before calibration. Enter [CO2 Setup] menu and select
[Zero] to perform zeroing.
To calibrate the CO
module, follow this procedure:
2
1. Make sure that the CO
module has been warmed up or started up.
2
2. Connect the gas cylinder with the tubing using a T-shape connector as shown below. Check the
airway and make sure there are no leaks.
To the air
Tubing
G
Defibrillator/
monitor
3. Vent the tubing to the CO
by opening the gas valve.
2
4. Access the [Maintain CO2] menu. To do so, Press the Main Menu button on the equipment’s front
panel. Select [Others>>] [Installation Mode>>] enter the required password [Maintain
CO2].
4-12
Page 39
5. In the [Maintain CO2] menu, select a CO2 value equal to the vented CO2 concentration.
6. In the [Calibrate CO2] menu, the measured CO2 concentration is displayed. Wait till the measured
CO2 concentration becomes stable; select [Calibrate] to start CO2 calibrate.
The message [
failed, the prompt [
Calibration Completed!] is displayed after a successful calibration. If the calibration
Calibration Failed!] will be displayed. In this case, perform another calibration.

4.7 Electrical Safety Tests

See A Electrical Safety Inspection..

4.8 Recorder Check

Tools required:
None.
1. Print ECG waveforms. The recorder should print correctly and the printout should be clear.
2. Simulate some recorder problems, such as out of paper, paper jam, etc. the defibrillator/ monitor
should give corresponding prompt messages. After the problem is removed, the recorder should be
able to work correctly.
3. Switch automatic alarm recording for each parameter ON and then set each parameter’s limit outside
set alarm limits. Corresponding alarm recordings should be triggered when parameter alarms occur.

4.9 Factory Service

4.9.1 Password for Service Mode

Accessing service mode is password protected. The required password is set to 332888 before the
equipment leaves the factory.

4.9.2 Accessing Service Mode Menu

To access the factory service menu, Press the Main menu button on the equipment’s front panel. Select
[Others>>] [Maintenance>>] [Service Mode>>] enter the required passwords. The Service
Mode-Main menu is shown below.
4-13
Page 40

4.9.3 Impedance Measurement

Normally impedance measurement is unnecessary. However, you can perform impedance checking after
replacing the therapy module.
. If not pre-connected, connect the pads cable to the equipment.
1
2. Connect a test load of 300 ohms to the pads cable.
3. Start the equipment and select the Monitor mode. Press and hold the [Event] hardkey, and then
press the [Lead Select] hardkey on the front panel, the following screen appears.
4. Verify that the reading of “RT Imped” is between 3000±450.
NOTE
If 300 ohms test load is not available, you can use a 50 ohms test load to perform impedance
checking. In this case, Verify that the reading of “RT Imped” is between 500±75.
the reading of “RT Imped” is not correct, replace the therapy module.
If
4-14
Page 41

4.9.4 Device Information

Press the Main menu button on the equipment’s front panel. Select [Others>>] [Maintenance>>]
[Service Mode>>]enter the required passwords[Device Information]. In the Device Information
list, you can view the device information such as software version, system status, and etc, as shown below.
In the Device Information screen, you can select [Export] to export error codes and shock delivery data to
a USB flash memory.

4.9.5 Checking Failure Code

Press the Main Menu button on the equipment’s front panel. Select [Others>>][Maintenance>>]
[Service Mode>>]enter the required passwords[Failure Code] to check error codes. This helps the
service personnel to identify failures.
Re
fer to 6.8 Error Codes for the description of each error code.
4-15
Page 42

4.9.6 Inputting Serial Number

Press the Main Menu button on the equipment’s front panel. Select [Others>>] [Maintenance>>]
[Service Mode>>]enter the required passwords[Input Serial Number] to input the equipment’s
serial number.
After inputting the serial number, you can view it by accessing Installation Mode and select [Version].

4.9.7 Paddle Open Circuit Display

This [Paddle Open Circuit Display] switch is for testing only. In normal operation, it should be set to [Off].
Test Report
Customer name
Customer address
Servicing person
Servicing company
Equipment under test (EUT)
Model of EUT
SN of EUT
Hardware version
Software version
Test equipment Model/No. Effective date of calibration
4-16
Page 43
Test items Test records
Visual inspection
The case, display screen, buttons, knob, modules, power cord, and
accessories have no obvious signs of damage.
The external connecting cables are not frayed and the connector pins
are not loose or bent.
The external connectors are not loose or their pins are not bent.
The safety labels and data plate are clearly legible.
Power-on test
The power-on test is passed. The power indicator and alarm system work
correctly and the equipment start up properly.
Performance test
Manual Defibrillation Test
When the equipment runs on AC mains and external paddles are used,
the equipment can be properly charged and discharged; the energy
delivered meets the requirement for accuracy, and the shock
information is correctly recorded.
When the equipment runs on fully charged battery and external paddles
are used, the equipment can be properly charged and discharged; the
energy delivered meets the requirement for accuracy, and the shock
information is correctly recorded.
When the equipment runs on AC mains and multifunctional electrode
pads are used, the equipment can be properly charged and discharged;
the energy delivered meets the requirement for accuracy, and the shock
information is correctly recorded.
When the equipment runs on fully charged battery and multifunctional
electrode pads are used, the equipment can be properly charged and
discharged; the energy delivered meets the requirement for accuracy,
and the shock information is correctly recorded.
When external paddles are used, the charge tone is correctly issued
when the equipment is being charged. The prompt "Charged Removed"
is shown on the screen and the charge done tone stops when the
Disarm hotkey is pressed. The equipment does not discharge externally.
When [Time to Auto Disarm] is set to [60s], the prompt "Charged
Removed" is shown on the screen and the charge done tone stops after
60 seconds at the completion of charging. The equipment does not
discharge externally.
When pads are used, the charge tone is correctly issued when the
equipment is being charged. The prompt "Charged Removed" is shown
on the screen and the charge done tone stops when the Disarm hotkey
is pressed. The equipment does not discharge externally. When [Time to
Test results
(Pass/Fail)
4-17
Page 44
that there is no leakage.
Test items Test records
Auto Disarm] is set to [60s], the prompt "Charged Removed" is shown
on the screen and the charge done tone stops after 60 seconds at the
completion of charging. The equipment does not discharge externally.
When external paddles are used for synchronous defibrillation and ECG
source is paddles and lead II respectively, the prompt is correct and a
Sync mark appears above each R wave. The delivered energy measured
is 10J±2J and the synchronous shock is delivered within 60 ms of the
peak of the R-wave.
When pads are used for synchronous defibrillation and ECG source is
paddles and lead II respectively, the prompt is correct and a Sync mark
appears above each R wave. The delivered energy measured is 10J±2J
and the synchronous shock is delivered within 60 ms of the peak of the
R-wave.
Pacing Test
When set [Pacer rate] to [70ppm] and [Pacer Output] to [30mA], the
pacer rate measured by the analyzer is 70 ppm±1ppm and the pacer
output measured is 30 mA±5mA.
When set [Pacer rate] to [170ppm] and [Pacer Output] to [200mA], the
pacer rate measured by the analyzer is 170 ppm±2ppm, and the
measured current is 200 mA±10mA.
ECG performance test
ECG waves are displayed correctly without noise and the HR value is
within 80±1 bpm.
ECG Lead Off alarm behaves correctly.
Paced signals are detected and pace pulse marks are displayed when
[Paced] is set to [Yes]
The difference between the amplitude of the ECG calibration square
wave and that of the wave scale is not greater than 5%.
Resp test
The Resp wave is not distorted and the Resp value is within 40±2 rpm.
SpO2 test
The displayed SpO2 and PR values should be within the specified ranges.
NIBP test
The difference is within ±3 mm when 0, 50 or 200 mmHg is set for NIBP
accuracy test.
There is no leakage with NIBP, or the manual leakage test result does not
exceed 6mmHg/min.
CO2 test
Block the gas inlet of the module or watertrap. The CO2 flowrate is slower
than 10ml/min and an alarm of CO2 Filterline Err is given. It indicates
Test results
(Pass/Fail)
4-18
Page 45
Test items Test records
The displayed CO2 value is within 6±0.05.
Electrical safety tests
Refer to Appendix A Electrical Safety Inspection.
Recorder check
The recorder can print ECG waves correctly and the printout is clear.
Set the recorder to some problems such as out of paper, paper jam, etc.
the equipment gives corresponding prompt messages. After the
problem is removed, the recorder is able to work correctly.
Automatic alarm recording for each parameter functions correctly when
parameter alarms occur.
Tested by: _________________________ Date: ________________________
Test results
(Pass/Fail)
4-19
Page 46
FOR YOUR NOTES
4-20
Page 47

5 Hardware and Software Upgrade

5.1 Hardware Upgrade

5.1.1 Upgrade MPM module

You can upgrade MPM module to any of the following configuration:
ECG+Mindray SpO2 modules: 115-049471-00 MPM upgrade package
ECG+Nellcor SpO2 modules: 115-049472-00 MPM upgrade package
ECG+NIBP modules: 115-049474-00 MPM upgrade package
ECG+Mindray SpO2+NIBP modules: 115-049475-00 MPM upgrade package
ECG+Masimo SpO2+NIBP modules: 115-049476-00 MPM upgrade package
ECG+Nellcor SpO2+NIBP modules: 115-049477-00 MPM upgrade package
The upgrade procedure is as follows:
1. Remove the MPM module assembly and parameter panel assembly as described in. 7.3.5
Disassembling the MPM Module Assembly and 7.3.6 Removing the Parameter Panel Assembly.
2. Replace the old measurement module panel and MPM module with those in the MPM upgrade kit.
3. Reassemble the equipment.
If the upgrade MPM module is equipped with Masimo SpO
left corner of the front housing and a No Implied License label below the measurement module panel, as
indicated in the following pictures.
, you need to stick a Masimo label at the lower
2
Masimo label
No Implied License label
5-1
Page 48
If the upgraded MPM module is equipped with Nellcor SpO
left corner of the front housing, as indicated in the following pictures.
After upgrade the MPM module, perform the tests described in 4.6.3 ECG Tes t, 4.6.4 Resp Test, 4.6.5 SpO2
Tes t, 4.7Electrical Safety Tests, and 4.8 Recorder Check.
Nellcor label
, you need to stick a Nellcor label at the lower
2

5.1.2 Upgrade the Therapy Module

You can use 115-049460-00 pacer function upgrade kit to upgrade the therapy module so that equipment
has pacing function. After upgrading the therapy module, choose Mode label (with pacing function) with
the language you need.
Follow this procedure to upgrade the therapy module:
1. Remove therapy module as described in 7.3.7 Removing the Therapy Module. Be noted that you
need not to remove the parameter panel assembly and MPM module assembly.
2. Take off the Mode Select knob. Peel off the Mode label, see the picture below.
3. Replace the old therapy module using the one with pacing function in the upgrade kit.
4. Replace the old Mode Select knob and Mode label using new ones in the upgrade kit.
5. Reassemble the equipment.
5-2
Page 49
Mode Select knob
Mode label
After upgrade the therapy module, perform the tests described in 4.6.1 Manual Defibrillation Test, 4.6.2
Pacing Test, and 4.7 Electrical Safety Tests.

5.2 Software Upgrade through a PC

You can upgrade system software and module software using Mindray Patient Monitor Software Upgrade
Tool. You can also view software upgrade log. Mindray Patient Monitor Software Upgrade Tool can directly
run on a PC. By connecting the defibrillator/monitor to a PC via a crossover network cable, you can
upgrade the following software:
Bootstrap
System software
Language library
BMP files (including screen icons, start-up screens, standby screens)
General configurations (including passwords, company logo)
System functional configuration
FPGA program
Power module software
MPM module software and therapy module software
CO
module
2
5-3
Page 50

5.2.1 Installing Mindray Patient Monitor Software Upgrade Tool

1. Find the installation program and double click it to start installation.
2. Select installation language.
3. Click [Ok] and the following screen appears. Click [Next] to go to the next step.
4. Enter User Name, Company Name and Serial Number.
5-4
Page 51
5. Specify the destination folder for installing this program. Then select [Next].
7. Select Program Folder. Then select [Next].
8. Click [Finish] to complete installation.
5-5
Page 52
NOTE

5.2.2 Software Upgrade Procedure

1. Connect the defibrillator/monitor to be upgraded with a PC.
2. Set IP address to 77.77.1.XX and subnet mask to 255.255.255.0.
3. Run Mindray Patient Monitor Software Upgrade Tool on the PC and set Machine to BeneHeart.
4. On the Mindray Patient Monitor Software Upgrade Tool screen, select [Select Package] and select
packages you want to upgrade. Then select [Start].
5. Simultaneously hold the [Record] key and [Menu] key on the defibrillator/monitor’s front panel, and
then turn on the equipment.
After software upgrade is finished, turn off the equipment, and then disconnect AC mains and remove the
battery. If you do not disconnect the power supply after software upgrade, the status indicator will flash
and the beeper will sound.
For the details of software upgrade, refer to help and instructions for use of Mindray Patient Monitor
Software Upgrade Tool.
CAUTION
Disconnect the equipment from the patient and make sure the important data are saved
before upgrade.
Do not shut down or power off the equipment when upgrading the boot program. Otherwise,
it may cause the equipment to break down.
Program upgrade should be performed by qualified service personnel only.
Crossover network cable shall be used if a PC is connected for equipment upgrade.
After upgrading the boot program, re-upgrade the system program and other programs to
ensure compatibility.
Make sure the version of the upgrade package is the one that you desired. If you want to obtain
the latest upgrade package, contact Mindray Customer Service Department.
5-6
Page 53

5.3 Software Upgrade through a USB Disk

5.3.1 Applicable

Model SN Remark
BeneHeart D5/D6 DZ-XXXXXXXX
DI-XXXXXXXX
BeneHeart D2/D3 EZ-XXXXXXXX
EI-XXXXXXXX

5.3.2 Tool

Index Tools Qty.
1 USB disk 1

5.3.3 Upgrade Procedure

1. Prepare a USB Disk
This procedure is only applicable to the device
which software version is 04.03.00.16 and above
a. Prepare a common USB disk which the system capacity is larger than 2GB and choose FAT 32
b. Create following folders in the boot directory of the USB Disk UPGRADE_AMP/Beneheart/
c. Copy the BIOS program NewBeneHeart_Installer.pkg (Do not change this file name) to the
UPGRADE_AMP/ Beneheart
d. Copy the upgrade file (PKG or MPKG) to the UPGRADE_AMP/Beneheart.
2. Connection
a. Insert the prepared USB disk into the USB port of the machine.
3. Upgrade
a. Turn off the defibrillator, Press and hold Summary (or 12 Lead) button and Main Menu button
of D5/D6 at same time, then turn on the defibrillator, to enter upgrade mode.
b. Note: For D2/D3 machine simultaneously press and hold Menu and Record button (or NIBP
button), and turn on the defibrillator to enter the upgrade mode.
5-7
Page 54
4. Select the File
A maximum of 16 upgrade packages can be displayed in the left and right columns. Rotate the navigation
knob to select the desired upgrade package and press down the knob to confirm.
5. Completed
The upgrade completed will be showed in the screen when the upgrade is completed. Restart the device to
activate the new system software.
6. NOTE
a. Disconnect the defibrillator from the patient and make sure that important data are saved
before upgrade.
b. Do not shut down or power off the equipment when upgrading the BIOS program and FPGA
program. Otherwise, the equipment may break down.
c. Upgrade should be performed by qualified service engineer only. Indicates a potential hazard
or unsafe practice that, if not avoided, could result in minor personal injury or
product/property damage.
d. Make sure the version of the upgrade package is your desired one. If you want to obtain the
latest upgrade package, please contact Mindray Customer Service Department.
5-8
Page 55

6 Troubleshooting

6.1 Overview

In this chapter, the defibrillator/monitor problems are listed along with possible causes and recommended
corrective actions. Refer to the tables to check the defibrillator/monitor, identify and eliminate the
problems.
The problems we list here are frequently arisen difficulties and the actions we recommend can correct
most problems, but not all of them. For more information on troubleshooting, contact our Customer
Service Department.

6.2 Part Replacement

Printed circuit boards (PCBs), major parts and components in the defibrillator/monitor are replaceable.
Once you isolate a defective PCB, follow the instructions in 0
Disassembly and Repair to replace the PCB with a known good one and check that the trouble disappears
or the defibrillator/monitor passes all performance tests. If the trouble remains, replace the PCB with the
original suspicious PCB and continue troubleshooting as directed in this chapter.
To obtain information on replacement parts or order them, refer to 8 Parts.

6.3 Checking Defibrillator/Monitor Status

Some troubleshooting tasks may require you to identify the hardware version and status of your
defibrillator/ monitor. To check equipment status,
1. Select [Main Menu] [Review >>][Event Review >>]. Then you can view the information on
system start time, self check, etc.
2. You can also view the information on the defibrillator/monitor’s current status by pressing the Menu
key on the equipment’s front panel, and then selecting [Others>>]→[Maintenance >>]→[Service
Mode>>]→ enter the required password →[Device Information>>].
6-1
Page 56

6.4 Checking Device Information

Some troubleshooting may involve software compatibility. Thus it requires you to know your
defibrillator/monitor configuration and software version. For detailed information on version compatibility,
please contact our Customer Service Department.
To identify your software version, press the Menu key on the equipment’s front panel, and then select
[Others>>] [Maintenance>>][Installation Mode>>] enter the required password [Version]. In
the Version screen, you can view system software version and module software version.
You can also press the Menu key on the equipment’s front panel, and then select [Others>>]
[Maintenance>>][Service Mode>>] enter the required password [Device Information>>] to
check system software version, module software version, and device status .

6.5 Checking Technical Alarm

Before troubleshooting the defibrillator/monitor, check for technical alarm message. If an alarm message is
presented, eliminate the technical alarm first. For detailed information on technical alarm message,
possible cause and corrective action, see the BeneHeart D3/ BeneHeart D2 Defibrillator/Monitor
Operator’s Manual (P/N: 046-010599-00)l.

6.6 Troubleshooting Guide

6.6.1 Defibrillation Problems

Symptom Possible Cause Corrective Action
The equipment does
not charge by
pressing the Charge
button on the front
panel.
The equipment
cannot be charged
by pressing the
Charge button on
the external paddles.
Keypad board failure. 1. Connect external paddles. Press the
“Charge” button on the Apex paddle to start
charging. If the “Charge” button on the
paddles works, it indicates the keypad
board fails.
2. Replace the keypad board.
The Charge button fails to be pressed
down effectively due to the damaged
or dislocated silica gel keypad.
Paddles not connected properly. Reconnect the paddles.
Paddles failure. 1. Isolate the problem by connecting pads
Failure of therapy module. Replace the therapy module.
Failure of connection wire to the
therapy module.
Disassemble the keypad board to replace or
reshuffle the keypad.
to perform charging/discharging.
2. If the paddles are failed, replace them.
Replace the connection wire.
6-2
Page 57
Symptom Possible Cause Corrective Action
The equipment is
charged too slowly
A shock cannot be
delivered by
pressing the Shock
button on the
equipment’s front
panel in Manual
Defib Mode or AED
Mode.
A shock cannot be
delivered by
pressing the Shock
button on the
paddles.
The message.
“Disarming Failed” is
displayed.
The equipment can
be properly charged,
but the energy is
disarmed
automatically at the
completion of
charging or when
the equipment is
being discharged.
Defibrillation
malfunction.
Energy Select
buttons on the
equipment front
panel do not work.
Battery failure. Replace the battery or connect the
equipment with external power supply.
CPU and power management board
failure.
Therapy module failed. Replace the therapy module.
Keypad board failure. 1. Locate the problem by connecting
The Charge Button fails to be pressed
down effectively due to the failure or
dislocated silica gel keypad.
Paddles not connected properly. Reconnect the paddles.
Paddles failure. 1. Locate the problem by connecting pads
Failure of therapy module. Replace the therapy module.
Connection wire to the therapy
module broken.
Failure of therapy module. Replace the therapy module.
Too high or too low patient impedance
detected.
1. Pads/paddles are detached from the
patient.
2. Pads/paddles failure.
3. Pads cable failure.
Failure of therapy module. Replace the therapy module.
Defibrillation hardware circuit
defective.
Keypad board failure. 1. Locate the problem by connecting
The Energy Select Buttons fail to be
pressed down due to the damaged or
dislocated silica gel keypad.
Replace the CPU and power management
board.
paddles to perform charging / discharging.
2. If the keypad board is defective, replace
it.
Disassemble the keypad board to replace or
reshuffle the keypad.
to perform charging/discharging. If normal
discharge can be performed, it indicates the
paddles are defective.
2. Replace the paddles.
Replace the connection wire.
Ensure good connection between the
patient and pads/paddles. If the problem
persists, replace Pads, paddles or Pads
cable.
Replace the therapy module.
paddles to perform energy setting.
2. If the keypad board is defective, replace
it.
Disassemble the keypad board to replace or
reshuffle the keypad.
6-3
Page 58
Symptom Possible Cause Corrective Action
Energy Select
buttons on the
paddles do not work
The message
“Defibrillation
malfunction”
presented at
power-on
Paddles not connected properly. Reconnect the paddles.
Paddles failure. Replace the paddles.
Failure of therapy module. Replace the therapy module.
Connection wires to the therapy
module broken.
Configuration files not properly
upgraded.
Therapy module failure Replace the therapy module.
Replace the connection wires.
Upgrade the configuration files.

6.6.2 Pacing Problems

Symptom Possible Cause Corrective Action
Does not deliver
correct pacing
current.
Does not deliver
correct pacing rate.
Pacer Equip
Malfunction
Failure of the therapy module Replace the therapy module.
Failure of the therapy module. Replace the therapy module.
Pacer hardware failure. Replace the therapy module.

6.6.3 Power On/Off Problems

Symptom Possible Cause Corrective Action
The
defibrillator/monitor
fails to start. AC LED or
battery LED does not
light
AC mains not connected or
battery too low.
Power supply protection.
Cables defective or poorly
connected.
Power switch or keypad board
failure
AC/DC board defective Replace the AC/DC board.
CPU and power management
board failure
Check that AC mains is properly connected or
battery capacity is sufficient.
Refer to 6.6.10 Power Supply Problems.
1. Check that the cables between the power
switch and the keypad board, the keypad board
and the power management board and
between the power module and the power
management board are correctly connected.
2. Check that wires and connectors are not
defective.
Replace the power switch or keypad board.
Replace the CPU and power management
board.
6-4
Page 59

6.6.4 Display Problems

Symptom Possible Cause Corrective Action
The LCD screen is blank,
but the
defibrillator/monitor
works properly.
Secondary display does
not function.
Secondary display
displays snows or
flashing specks
Images overlapped or
distorted
The colour of images
deviates from the
standard configuration.
Images overlapped or
distorted
Connection cable defective or
poorly connected.
LCD Display failure Replace the display.
Keypad board failure Replace the keypad board.
CPU and power management
board failure.
Cable failure 1. Check that the wires between the display and
CPU and power management
board failure.
Connection cable defective or
poorly connected.
CPU and power management
board failure.
FPGA failure Update or upgrade FPGA.
Connection cable defective or
poorly connected.
Connection cable defective or
poorly connected.
Display failure Replace the display.
Keypad board failure Replace the keypad board.
CPU and power management
board failure.
FPGA failure Update or upgrade FPGA
Connection cable defective or
poorly connected.
1. Check that wires between the display and the
keypad board, and between the keypad board
and the CPU and power management board are
correctly connected.
2. Check that the cables and connectors are not
defective.
Replace the CPU and power management board.
the defibrillator/monitor are correctly connected.
2. Check that the cables and connectors are not
defective.
Replace the CPU and power management board.
1. Check that the wires between the display and
the defibrillator/ monitor are correctly connected.
2. Check that the cables and connectors are not
defective.
Replace the CPU and power management board.
1. Check that the wire between the display and
keypad board is correctly connected.
2. Check that the cables and connectors are not
defective.
1. Check that wires between the display and the
keypad board and between the keypad board
and the power management board are correctly
connected.
2. Check that the cables and connectors are not
defective.
Replace the CPU and power management board.
1. Check that the wire between the display and
keypad board is correctly connected.
2. Check that the cables and connectors are not
defective.
6-5
Page 60

6.6.5 Alarm Problems

Symptom Possible Cause Corrective Action
The alarm lamp is not
light or extinguished
but alarm sound is
issued
No alarm sound is
issued but alarm lamp
is lit
Connection cable defective or
poorly connected.
Alarm LED board failure Replace the alarm LED board.
Keypad board failure Replace the keypad board.
CPU and power management
board failure.
Audio alarm disabled Check that alarm tone volume is set to a value
Connection cable defective or
poorly connected.
FPGA audio logic ERROR Upgrade the audio logic part of the FPGA
Speaker failure Replace the speaker.
CPU and power management
board failure.
1. Check that wire between alarm LED board and
keypad board are properly connected.
2. Check that connection wires and connectors
are not defective.
Replace the CPU and power management board.
other than zero by pressing the Menu key on the
equipment’s front panel, and then selecting
[Alarm Setup >>].
1. Check that the wire between the speaker and
keypad board is properly connected.
2. Check that connection wires and connectors
are not defective.
program.
Replace the CPU and power management board.

6.6.6 Button and Knob Problems

Symptom Possible Cause Corrective Action
Buttons do not
respond.
Mode Select knob
does not respond.
Connection cable defective or
poorly connected.
Keypad board failure Replace the keypad board.
CPU and power management
board failure.
Connection cable defective or
poorly connected.
1. .Check that the wire between the keypad and
the keypad board is properly connected.
2. Check that that the wire between the keypad
board and the CPU and power management
board is properly connected.
3. Check if the connection wires and connectors
are defective.
Replace the CPU and power management board.
1. Check that wires between the knob to keypad
board, and between the keypad board and the
power management board are properly
connected.
2. Check that connecting wires and connectors
are not defective.
6-6
Page 61
Symptom Possible Cause Corrective Action
Mode Select knob
does not respond.
Navigation knob does
not respond.
Knob failure Replace Mode Select knob.
Keypad board failure Replace the keypad board.
CPU and power management
board failure.
Connection cable defective or
poorly connected.
Knob failure. Replace the navigation knob.
Keypad board failure Replace the keypad board.
CPU and power management
board failure.
Replace the CPU and power management board.
1. Check that wires between the knob to keypad
board, and between the keypad board and the
power management board are properly
connected.
2. Check that connection wires and connectors
are not defective.
Replace the CPU and power management board.

6.6.7 Recorder Problems

Symptom Possible Cause Corrective Action
No printout Connection cable defective or
poorly connected.
Recorder power supply failure Check if the power module outputs 5 V DC and
Recorder failure Replace the recorder.
Poor print quality or
paper not feeding
properly
Blank printout Paper-roll installed reversely. Reload paper-roll.
Paper roll not properly installed Stop the recorder and re-install the paper roll.
Print head dirty 1. Check the thermal print head and the paper
Recorder failure Replace recorder.
Recorder failure Replace the recorder.
1. Check the wire between the recorder and the
keypad board is connected properly.
2. Check that connection wires and connectors
are not defective.
12V DC correctly.
roller for foreign matter.
2. lean the thermal print head with an
appropriate clean solution.
6-7
Page 62

6.6.8 Output Interface Problems

Symptom Possible Cause Corrective Action
Sync input failure CPU and power management
board failure.
Interface board failure Replace the interface board.
USB Device does not
function (provided
that the peripheral
devices are good)
CPU and power management
board failure.
Replace the CPU and power management board.
Replace the CPU and power management board.

6.6.9 Memory Card Problems

Symptom Possible Cause Corrective Action
Memory card
malfunctions
Memory card failure Format the memory card. If the problem persists,
replace the CPU and power management board.

6.6.10 Power Supply Problems

Symptom Possible Cause Corrective Action
Battery failure Battery damaged. Replace battery.
Battery interface failure. 1. Check batteries are installed properly.
2. Check if the battery interface is defective.
3. If the battery interface is defective, replace the
power management board.
CPU and power management
board failure.
Batteries can not be
fully charged.
Battery cannot be
charged
No +3.3 V A output 1. Power supply failure 1. Turn off the defibrillator/monitor then restart
Battery damaged. Replace batteries.
Battery interface failure. 1. Check batteries are installed properly.
CPU and power management
board failure.
Battery failure. Replace battery and recharge the replacement
Cable defective or poorly
connected
CPU and power management
board failure.
Replace the CPU and power management board.
2. Check if the battery interface is defective.
3. If the battery interface is defective, replace the
power management board.
Replace the CPU and power management board.
battery. If the replacement battery can be
recharged, the original one fails.
1. Check batteries are installed properly.
2. Check if the battery interface is defective.
3. If the battery interface is defective, replace the
power management board.
Replace the CPU and power management board.
6-8
Page 63
management board.
NOTE
Symptom Possible Cause Corrective Action
No +3.3 V B output
Not +5.0 V output
No +12 V output
When the power module has a failure, it may cause problems to other components, e.g. the
defibrillator/monitor suddenly breaks down during start-up, as the power module may have a
power supply protection. In this case, troubleshoot the power module per the procedure
described in the table above.
Components of the main unit, SMR and parameter modules are powered by the power module.
In the event that a component malfunctions, check if the operating voltage is correct. Refer to
2 Theory of Operation for the operating voltage and measurement points of each component.
2. CPU and power management
board failure.
it.
2. If the problem remains, disconnect the AC
mains for 5 s and reconnect it, then restart the
defibrillator/ monitor.
3. If the problem still remains, replace the power

6.6.11 Software Upgrade Problems

Symptom Possible Cause Corrective Action
Boot file upgrade fails Power failure or unintended
power off during boot file
upgrade
Program upgrade fails Incorrect network connection 1. Check that the network cable is properly
Wrong upgrade package Upgrade package shall be .pkg files. Select
Incorrect IP address
configuration
Return the CPU board to factory for repair.
connected and is not too long (shorter than
50m).
2. Make sure that the network cable is of the
right type. Network cable with crossed wires
inside is used for LAN upgrade and those with
parallel wires inside for WAN.
package according to system requirement.
Configure a fixed IP address in range C as
specified for the defibrillator/monitor. We
recommend not to upgrade a program when the
defibrillator/monitor is connected to a network
with multiple PCs.
6-9
Page 64

6.7 Technical Alarm Messages

Measurement Alarm Message Cause and solution
XX XX SelfTest Err An error occurred to the XX module, or there is
XX Init Err
XX Comm Err
XX Comm Stop
XX Overrange The measured XX value is not within the
ECG ECG Lead Off The ECG electrode has become detached from
ECG YY Lead Off
(YY represents the leadwires V,
LL, LA, and RA, as per AHA
standard, or C, F, L and R as per
IEC standard.)
a problem with the communications between
the module and the host. Restart the
equipment.
specified range for XX measurement. Contact
our Customer Service Department.
the patient or the lead wire has become
disconnected from the trunk cable. Check the
connection of the electrodes and leadwires.
SpO2
Pads/Paddles off The pads/paddles have been detached from the
patient or the therapy cable is loose. Check that
the pads/paddles and therapy cable are
properly connected.
ECG Noise The ECG signal is noisy. Check for any possible
sources of signal noise form the area around the
cable and electrode, and check the patient for
excessive motion.
ECG Signal Invalid ECG amplitude is so low that ECG signal is
undetectable. Check for any possible source of
interference from the area around the cable
and electrode; check the patient’s condition.
SpO2 Sensor Off The SpO2 sensor has become detached from
SpO2 Sensor Fault
SpO2 No Sensor
SpO2 Unknow Sensor
SpO2 Sensor Incompatible
SpO2 Too Much Light There is too much light on the Sp O2 sensor.
SpO2 Low Signal The Sp O2 signal is too low or too weak. Check
the patient or the module, or there is a fault
with the Sp O
sensor has been used. Check the sensor
application site and the sensor type, and make
sure the sensor is not damaged. Reconnect the
sensor or use a new sensor.
Move the sensor to a place with lower level of
ambient light or cover the sensor to minimize
the ambient light.
sensor, or an unspecified Sp O2
2
6-10
Page 65
Measurement Alarm Message Cause and solution
SpO2
NIBP NIBP Loose Cuff The NIBP cuff is not properly connected, or
CO2
SpO2 Weak Signal the patient’s condition and change the sensor
SpO2 Weak Pulse
SpO2 Low Perf
SpO2 Interference The Sp O2 signal has been interfered. Check for
SpO2 Non-Pulsatile
SpO2 Board Fault There is a problem with the Sp O2 measurement
NIBP Air Leak
NIBP Pneumatic Leak Check the NIBP cuff and pump for leakages.
NIBP Cuff Type Wrong The cuff type applied mismatches the patient
NIBP Air Press. Err An error occurred to the air pressure. Verify that
NIBP Weak Signal The patient’s pulse is weak or the cuff is loose.
NIBP Sig. Saturated The NIBP signal is saturated due to excess
NIBP Overrange The patient’s NIBP value may be beyond the
NIBP Excessive
Motion
NIBP Equip Err An error occurred during NIBP measurement
NIBP Time Out
NIBP Measure Failed
NIBP Reset For Err An illegal reset occurred during NIBP
CO2 Sensor High
Temp
CO2 Occlusion The airway or watertrap was occluded. Check
CO2: Change
Watertrap
application site. If the error persists, replace the
sensor.
any possible sources of signal noise form the
area around the sensor, and check the patient
for excessive motion.
board. Do not use the module and contact our
Customer Service Department.
there is a leak in the airway.
category. Verify the patient category and
replace the cuff.
the equipment application site meets the
environmental requirements and check if there
is any source that affects the air pressure.
Check the patient’s condition and change the
cuff application site. If the problem persists,
change the cuff.
motion or other sources.
specified measurement range.
Check the patient’s condition and reduce the
patient motion.
and therefore the equipment cannot perform
analysis correctly. Check the patient’s condition
and NIBP connections, or replace the cuff.
measurement. Check if the airway is occluded.
Check, stop using or replace the sensor.
the airway and remove the occlusion.
Change the watertrap.
6-11
Page 66
Measurement Alarm Message Cause and solution
CO2
CPR sensor CPR Sensor Err There is a self-test error or communication
Main control system
CO2 Watertrap
Mismatch
Check the patient category, replace a matched
watertrap.
CO2 No Watertrap Check the watertrap connections.
CO2 Zero Failed Check the CO2 connections. After the sensor’s
temperature becomes stabilized, perform a zero
calibration again.
CO2 Module Error There is a problem with the CO2 module, or a
problem with the communications between the
host and the CO
module. Restart the
2
equipment.
problem
with the CPR sensor. Contact our Customer
Service Department.
CPR Sensor Low
Battery
The battery power of the CPR sensor is low.
Charge the battery by connect the CPR sensor
to the equipment.
CPR Sensor Need Service The compressions using the CPR sensor exceed
the expected numbers. Contact our Customer
Service Department.
CPR Sensor Cable Fault An error occurred to the CPR sensor cable.
Replace the CPR sensor cable.
Change CPR Sensor
Battery
The CPR sensor battery is aging. Contact our
Customer Service Department.
CPR Sensor Bat. Charge Err The CPR sensor cannot be charged. Contact our
Customer Service Department.
No Speaker Make sure that the speaker is connected.
Power Board Comm Err An error occurred to the power board, or there
is a problem with the communications between
the power board and the host. Restart the
equipment.
Keyboard Comm Err An error occurred to the keypad board, or there
is a problem with the communications between
the keypad board and the host. Restart the
equipment.
Therapy Module Comm Err An error occurred to the therapy module, or
there is a problem with the communications
between the therapy module and the host.
Restart the equipment. If the problem persists,
contact our Customer Service Department.
Main Control Selftest Err The main control voltage is abnormal. Replace
the main control board.
6-12
Page 67
Measurement Alarm Message Cause and solution
Main control system Wifi Module Fault Contact our Customer Service Department.
Machine Type Error
RT Clock Need Reset Reset system time.
RT Clock Err An error occurred to the RTC chip, or the button
cell
is depleted. Replace corresponding part.
Memory Err There is a problem with the data card. Format
the CF card. If the problem persists, contact our
Customer Service Department.
Last User Test Failed Run a successful user test.
Last Auto Test Failed Run a successful user test again.
No CMS The equipment is disconnected from the CMS.
Check the network connection.
IP Address Conflict Network IP conflicts. Check the network
settings.
Power board Power System Selftest Err An error occurred to the system power supply.
Restart the equipment.
Therapy module
Power Board Volt Err
Low Battery Change battery or connect the equipment to
the AC power source to charge the batteries.
No Battery Battery is not installed. Install the battery.
Battery Depleted! System will
shut shown imminently.
Connect to AC Mains or Replace
Batter y.
Battery Err There is a problem with the batteries. Check the
Battery Aged Replace the battery.
Battery failed charging Battery failure or power board hardware failure.
Therapy Equip selftest Err An error occurred during therapy module self
Connect the equipment to AC mains.
batteries for damage; verify that correct
batteries are used. Replace the batteries if
necessary.
Replace the battery. If the problem persists,
replace the power board.
test. Restart the equipment or replace the
therapy module low voltage board.
6-13
Page 68
Measurement Alarm Message Cause and solution
Therapy module
Monitoring module Mornitor Module Selftest Err An error occurred during MPM module
Recorder Recorder Init Err Restart the equipment.
Defib Malfunction The defibrillation function fails or both the
defibrillation and pacing functions fail. Restart
the equipment and test defibrillation function.
If the problem persists, contact our Customer
Service Department.
Pacer Malfunction! The pacing function fails. Restart the
equipment and test pacer function. If the
problem persists, contact our Customer Service
Department.
Disarming Failed There is a problem with the therapy module
disarming circuit. Replace the therapy module
low voltage board and high voltage board.
power-on self test. Replace the MPM module.
Monitor Module Reset Err MPM module reset abnormally. In this case, the
MPM module restores to default configuration.
You can ignore this problem.
Monitor Module Voltage Err The voltage of MPM module is abnormal.
Replace the MPM module.
Recordhead Overheated The recorder has been working for a prolonged
time. Clear the recording tasks and resume the
recording till the recorder’s print head cools
down.
Recorder Overcurrent Re-load the recorder paper.
Pacer Pads cable Off Check that pads cable is properly connected.
Pads Off Check that pads are properly connected.
ECG Lead Off Check that ECG leadwires are properly
connected.
Pacer Stopped Abnormally Check paddles. Check that pads well contact
with patient’s skin. Make sure pads are properly
applied, and then start pacing again.
Others Load Config Err Check if the configuration is correct, or restore
the factory configuration.
6-14
Page 69

6.8 Error Codes

6.8.1 Therapy Module Error Codes

Type Error Code Description
Power-on self-test 1 CPU (M0+) power-on self-test failed
2 RAM (M0+) power-on self-test failed
3 ROM (M0+) power-on self-test failed
4 Watchdog (M0+) power-on self-test failed
5 ADC (M0+) power-on self-test failed
6 Register (M3) power-on self-test failed
7 RAM (M3) power-on self-test failed
8 ROM (M3) power-on self-test failed
9 Watchdog (M3) power-on self-test failed
10 ASIC (M0+) power-on self-test failed
11 CPU (M3) power-on self-test failed
Reset information 45 M0+ abnormal reset
46 ASIC exception
47 M3 abnormal reset
48 M3 and M0+ communication error
Real-time self-test 49 M3 real-time ADC self-test error
50 M3 chip calculation function error
Defibrillation 51 |V1-V2| > 500 V at the onset of charging
52 V1 <= 65 V or V2 <= 65 V 1s after charging starts
53 The V1/V2 decline exceeds more than 10% of V1/V2tgt during
charging
54 |V1-V2| > 128 V during use and |V1-V2| > 108 V during self-test
in the case of charging
55 V1 >= 2400 V or V2 >= 2400 V during charging
56 Charging not completed within 25s after charging starts
57 V1 > (V1Tgt*1.2) after charging ends
58 V1 <= 50 V or V2 <= 50 V during charging retention
59 V1 > (V1Tgt*1.2) or V2 > (V1Tgt*1.2) during charging
retention
60 Overcurrent during self-discharging
61 V1 >= 40 V or V2 >= 40 V after self-discharging
62 Overvoltage protection occurred
Impedance calibration 63 Zeroing sampling value error
64 Gain calibration sampling value error
65 Slope calculation error in gain calibration
66 Zeroing failed before gain calibration
67 An error occurred when calibration information is written into
the flash memory
6-15
Page 70
Type Error Code Description
Impedance calibration 68 An error occurred when calibration information in the flash
memory is read
Pacing 71 Pacing power error
72 Pacing relay error
73 Pacing frequency error
74 Pacing current error
75 Pacing DA error
76 Pacing voltage error
77 Pacing overcurrent protection point too high
78 Pacing overcurrent protection failure
Function self-test
201 AD self-test failed in function self-test
202 Clock self-test timeout and uncompleted in function self-test
203 Clock frequency self-test failed in function self-test
210 Defibrillation charging failed in function self-test
211 Function self-test charging timeout
212 Function self-test charging retention failed
216 Discharge circuit self-test failed in function self-test
217 Discharge circuit self-test failed in function self-test
218 Discharge circuit self-test failed in function self-test
219 Discharge circuit self-test failed in function self-test
220 Discharge circuit self-test failed in function self-test
221 Discharge circuit self-test failed in function self-test
222 Discharge circuit self-test failed in function self-test
223 Discharge circuit self-test failed in function self-test
224 Discharge circuit self-test failed in function self-test
226 Self-discharging completion self-test failed in function
self-test
230 Self-test failed in minor signal impedance detection in
function self-test
231 Defibrillation relay short-circuited in minor signal impedance
detection in function self-test
232 Test load relay short-circuited in minor signal impedance
detection in function self-test
258 P-lead ECG self-test failed in function self-test
259 PFlag & RFlag self-test failed in function self-test
260 M3 5V7 power failure
261 M3 +-5 V power failure
262 M3 +18 V power failure
263 M3 3V3 power failure
264 M0 AVCC power failure
265 M0 AVSS power failure
266 M0 +2V5 power failure
6-16
Page 71
Type Error Code Description
Function self-test
267 M0 -2V5 power failure
268 M0 ASIC_VREF power failure
269 M0 DVDD power failure
270 Charging prohibited after the in-vivo cable is connected to
the human body in self-test
271 Discharging failed due to discharging circuit short-circuited
273 The capacitance of the energy-storage capacitor exceeds the
threshold in self-test
274 Defibrillation energy dual-backup check error
275 Pacing frequency dual-backup check error
276 Pacing current dual-backup check error
277 Big error in defibrillation discharge resistance

6.8.2 Power Module Error Codes

Type Error Code Description
Register failure 101 Battery discharging short circuit failure
102 Battery charging short circuit failure
103 Battery AFE discharging overcurrent failure
104 Battery AFE watchdog failure
105 Battery controller watchdog failure
106 Setting of the battery permanent failure flag bit
107 Battery overvoltage failure
108 Battery undervoltage failure
109 Battery pack overvoltage failure
110 Battery pack undervoltage failure
111 Battery Level-2 charging overcurrent failure
112 Battery Level-2 discharging overcurrent failure
113 Battery charging overcurrent failure
114 Battery discharging overcurrent failure
115 Battery charging over-temperature failure
116 Battery discharging over-temperature failure
117 Battery over-charge failure
118 Battery over-charge current failure
119 Battery over-charge voltage failure
120 Battery fast charge timeout failure
121 Battery pre-charge timeout failure
Other error codes of
power board
143 Battery communication error
144 Power voltage error
145 Battery charging failure
146 Power-on self-test error
147 Main control failure
6-17
Page 72
Type Error Code Description
Other error codes of
power board
148 Disposable battery communication error
151 Low battery
152 Ultra-low battery
153 Battery failure
154 Battery aged
155 RTC self-test failed
156 Power-on/off button adhesion
161 VBUS failure
162 5VD failure
163 10V failure
164 DVDD failure
165 18V failure
166 AVSS failure
167 3VB failure

6.8.3 Main Control Error Codes

Error Code Description
401 No speaker
402 No storage card
403 Power board communication error
404 Therapy board communication error
405 Main control module power-on self-test error
406 Real-time clock error
407 Storage card read/write error
408 Keypad self-check error
409 Machine type recognition error
410 Recorder communication error
411 Front housing keypad self-test error
412 Speaker self-test error
413 Inaccurate real-time clock
414 Button adhesion
415 Program CRC check error
416 Poles expired
417 AT battery discharge error
419 UT battery discharge error
421 An error occurred when an SPI obtains the version
422 Wi-Fi module failure
423 Failed to load user configuration
424 Maintenance indicator user self-test error
425 M3 self-test error
426 Power board self-test error
6-18
Page 73

6.8.4 M51C Error Codes

Error Code Description
After the NIBP self-test fails or the equipment is powered on, an error occurs in the air
506
pump, A/D sampling, or pressure sensor, or a pointer error occurs during software
operation.
6-19
Page 74
FOR YOUR NOTES
6-20
Page 75

7 Disassembly and Repair

7.1 Tools Required

To disassemble and replace the parts and components, the following tools may be required:
Phillips screwdrivers
Tweezers
Sharp nose pliers
7# socket wrench
Adjustable spanner
Defibrillator/monitor high-voltage discharge fixture (0651-TF11)

7.2 Preparations for Disassembly

Before disassembling the equipment, finish the following preparations:
Stop patient monitoring and therapy, turn off the equipment and disconnect all the accessories and
peripheral devices.
Disconnect the AC power source and remove the battery.
To avoid high voltage hazard, strictly follow the procedure as defined in section 7.3.4 Discharging
the Capacitor for disassembling.
WARNING
Before disassembling the equipment, be sure to eliminate the static charges first. When
disassembling the parts labeled with static-sensitive symbols, make sure you are wearing
electrostatic discharge protection such as antistatic wristband or gloves to avoid damaging the
equipment.
Properly connect and route the cables and wires when reassembling the equipment to avoid
short circuit.
Select appropriate screws to assemble the equipment. If unfit screws are tightened by force,
the equipment may be damaged and the screws or part may fall off during use, causing
unpredictable equipment damage or human injury.
Follow correct sequence to disassembly the equipment. Otherwise, the equipment may be
damaged permanently.
Disconnect all the cables before disassembling any parts. Be careful not to damage any cables
or connectors.
7-1
Page 76
Place removed screws and disassembled parts properly, preventing them from being lost or
contaminated.
Place the screws and parts from the same module together to facilitate reassembling.
To reassemble the equipment, first assemble the assemblies, and then the main unit. Carefully
route the cables.
Make sure that the waterproof material is properly applied during reassembling.

7.3 Disassembling the Main Unit

WARNING
To disassemble the equipment, first remove the external assemblies, such as the hook mount
(if configured), paddle tray assembly, and front housing assembly in turn, and then the
internal assemblies and parts.
The power supply assembly and recorder can be removed without removing any other
assemblies.
To disassemble the equipment, place the equipment on a work surface free from foreign
material, avoiding damaging the antiglare screen, LCD and the knob. Be careful not to break
the two cotters on the front ends of rear housing.
All the operations should be performed by qualified service personnel only. Make sure to put
on the insulating gloves during service operations.
Before remove the therapy board, you must use the dicharge fixture to discharge the capacitor
first. If you do not have a discharge fixture, disconnect AC mains and remove batteries, wait for
at least 2 hours before removing the capacitor.
7-2
Page 77

7.3.1 Removing Hook Mount (if configured)

1. Stand the equipment on the work surface with the back of the equipment facing to you. Loose and
remove the two M3×16 Philips screws; take off the φ3 flat washers and spring washers.
2. Pull out the hooks.

7.3.2 Removing Paddle Tray

1. Stand the equipment on the work surface with the back of the equipment facing to you. Tweeze the
five plastic plugs filling the screw holes.
7-3
Page 78
P
assembly
2. Loose and unscrew the five M3×8 Philips screws. Remove the paddle tray.
addle tray
3. Unscrew the two PT3×8 Philips tapping screws and two M3×6 Philips combined screws. Disconnect
and remove the test load cable.
`
7-4
Page 79
NOTE

7.3.3 Separating the Housing

1. Lay the equipment on a padded work surface with the display facing down and the bottom of the
equipment nearest to you. Be careful not to damage the LCD and controls. Loose and remove the
seven M3×10 Philips screws.
2. Stand the equipment on the work surface. Carefully separate the front housing assembly and the rear
housing assembly. Remove the rocket latch , and then disconnect the cable between the main board
and keypad board to remove the front housing.
Socket latch
When reassemble the equipment, be sure to check if front housing water proof strip is correctly
placed.
7-5
Page 80

7.3.4 Discharging the Capacitor

1. Use the high-voltage discharge fixture (0651-TF11) to discharge the capacitor by hooking the
high-voltage ground end (TP1) with the black probe of the fixture, and hooking the high-voltage
socket (TP3) with the fixture’s red probe. Wait till all the indicating lamps on the fixture turns off. The
capacitor is not completely discharged if the indicator remains on.
2. Set the multimeter to DC 1000V. Measure the discharge resistance and check if the reading of the
multimeter is lower than 30V. If yes, you can safely disassemble the equipment now.
Use the black
probe to hook
the cooling fin

7.3.5 Disassembling the MPM Module Assembly

Latch of the
parameter
panel
Use the red
probe to touch
the copper pillar
1. Lay down the equipment flat, remove the latch of the parameter connector panel upwards, and then
remove the MPM module laterally.
2. Remove the recorder wire connected between the MPM module and CPU and power board and the
wire between the MPM module and the main control board from the main control board.
7-6
Page 81
WARNING

7.3.6 Removing the Parameter Panel Assembly

1. Remove the two M3×8 cross recessed pan head screws with washers from the parameter panel to
separate the MPM module from the parameter panel.

7.3.7 Removing the Therapy Module

Before removing the therapy board, you should discharge the capacitor first, refer to 7.3.4
Discharging the Capacitor for the instructions.
1. Remove the six M4×12 cross recessed combination screws from the therapy module, and then pull
out the therapy module upwards.
7-7
Page 82
2. Remove the in-position detection switch cable and two defibrillation port cables from the therapy
module.

7.3.8 Disassembling the Power Base Assembly

1. Lay the equipment on a padded work surface with the display side facing down and the bottom of
the equipment nearest to you. Loose and remove the three M3×8 Philips screws and pull out the
power base assembly.
7-8
Page 83
power board
head combination screw
3. Disconnect the cable between the power management board and AC/DC power supply module.
Remove the power base assembly.
Cable between the
power management
board and the AC/DC

7.3.9 Disassembling the AC/DC Power Supply Board

1. Use the screwdriver to loosen and remove the three M3×8 cross recessed pan head screws with
washers and one M4×8 cross recessed pan head combination screws, remove the AC socket cable,
and remove the defibrillation AC switching power board and shielding cover.
M4×8 cross recessed pan
AC socket cable
7-9
Page 84
2. Use the screwdriver to loosen and remove the four M3×8 cross recessed pan head screws with
washers, and separate and take out the power board from the shielding metal sheet.

7.3.10 Disconnecting the Therapy Port Cable

1. Remove the magnetic ring on the therapy port cable.
2. Pull the therapy port latch upward, and then disconnect the therapy port cable.
Magnetic ring on
therapy port cable
Therapy port latch
7-10
Page 85

7.3.11 Removing the Main Board Assembly

1. Remove the socket latch, and then remove the data cable between the main control board and the
keypad board from the power management board.
Data cable between the
main control board and
keypad board
Socket latch
2. Use the screwdriver to remove the six M3×8 cross recessed pan head screws with washers and use
the 7# socket wrench to remove the two M4×20 stud screws.
M4×20 stud screw
7-11
Page 86
Waterproof strip for
battery
3. Pull the cable between the power management board and the AC/DC power board out of the rear
housing gap and take out the main control board assembly.
Cable between the power
management board and
the AC/DC power board

7.3.12 Checking Waterproof Strips before Reassembling

Before reassembling the equipment, make sure that the waterproof material on the rear housing assembly
and power base assembly is stuck to the proper places.
1. Check that the waterproof strip is properly stuck on battery socket.
socket
7-12
Page 87
2. Check that the white waterproof strip is adhered to the proper place.
The joint of the waterproof
strip should be located at the
bottom and the gap should be
3. Check that the waterproof strip on the power base is adhered to the proper place.
Waterproof strip
7-13
Page 88
NOTE

7.4 Disassembling the Front Housing Assembly

To disassemble the equipment, place the equipment on a work surface free from foreign
material, avoiding damaging the antiglare screen, LCD and the knobs.
Make sure the speaker is not damaged after repairing any other front housing assembly. Verify
that the speaker works properly by powering on the equipment and testing the speaker.
Clear the LCD before reassembling it.

7.4.1 Removing the Keypad Board

1. Disconnect all the cables from the mode select knob cable, encoder cable, speaker cable, LCD cable,
assistant keypad board cable and the alarm lamp cable.
Mode select
knob cable
LCD cable
Encoder cable
Alarm lamp cable
Assistant
keypad cable
Speaker cable
7-14
Page 89
2. Remove the two M3×8 Philips screws securing the grounding plate, and the six M3×8 Philips screws
from the keypad board. Take out the kayped board.

7.4.2 Removing Display Assembly

1. Remove the four M3×8 Philips screws securing the keypad board. If the grounding plate is not
removed, you need to unscrew the two M3×8 Philips screws securing the grounding plate. Remove
the LCD bracket before removing the display assembly.
7-15
Page 90
applied on the LCD cable
2. Remove the hot-melt adhesive applied on the LCD cable, and then disconnect the LCD cable. Take out
the LCD from the silicone jacket.
Remove the hot-melt adhesive

7.4.3 Removing the Alarm Lamp Board and Assistant Keypad Board

To remove the alarm lamp board and assistant keypad board, remove the LCD bracket first, see 7.4.2
Removing Display Assembly.
1. Disconnect the alarm lamp cable.
2. Remove the two M3×8 Philips screws, and then take out the alarm lamp board and assitant keypad
board.
7-16
Page 91
with the knob

7.4.4 Removing the Speaker

1. Remove the two M3×8 Philips screws. Take off the speaker bracket, and then remove the speaker.
Speaker bracket
After repairing any part of the front housing assembly, verify that the speaker is not damaged by powering
on the equipment and testing the speaker.

7.4.5 Removing the Mode Select Knob

Pull the switch off its shaft. Loosen and remove the nut and washer using a socket wrench or sharp nose
pliers. Disconnect the cable from the knob.
Mode select
knob cable
When assemble the Mode Select knob, check that it is aligned with the labelling. Adjust the knob using
sharp nose pliers if necessary.
Washer and nut
Mode select
knob
7-17
Page 92

7.4.6 Removing the Encoder

1. Remove the navigation knob.
2. Disconnect the encoder cable.
3. Pull the encoder off its shaft. Loosen and remove the nut and washer using a socket wrench.
Disconnect the encoder cable.
navigation knob

7.4.7 Checking Waterproof Material on the Front Housing

Before reassembling the equipment, make sure that the waterproof material on the front housing are
placed to the proper places.
Check that the white waterproof strip is properly adhered to the slot around the edge of the front housing.
The joint of the waterproof
strip located at the bottom,
the gap less then 1mm
7-18
Page 93
board
Cut this cable if it is there

7.5 Removing the Recorder

1. Remove the two M3×6 Philips screws. Pull the recorder out of the recorder well.
2. Disconnect the cable from the recorder, and then remove the recorder.

7.6 Disassembling the Recorder

1. Loosen the two snaps and remove the recorder drive board.
Snaps
Recorder drive
7-19
Page 94
2. Loosen and remove the two PT2×6 crosshead tapping screws. Disconnect the flexible cable and the
connection cable between the recorder drive board and recorder keypad board. Remove the thermal
print head and recorder drive board.
Thermal
print head
Flexible cable
PT2×6 crosshead
tapping screw
Cable between the
recorder drive
board and recorder
keypad board
3. Remove the two PT2×6 crosshead tapping screws, and then remove the keypad board.
Keypad board
PT2×6 crosshead
tapping screw
7-20
Page 95

8 Parts

8.1 Introduction

This chapter contains the equipment’s exploded views and parts lists. It helps the engineer to identify the
parts during disassembling the equipment and replacing the parts. This manual is based on the maximum
configuration. Your equipment may not have some parts and the quantity of the screws, stacking sleeves,
and etc may be different with those included in the parts lists.
The figure below shows the hardware architecture of the equipment’s main unit.
Front housing assembly Rear housing assembly Paddle tray assembly Other assemblies
Front housing Rear housing Paddle tray External paddle set
Anti-glare screen
LCD Therapy module Paddle sheet metal Hook mount
Therapy mode select knob Defibrillator capacitor Paddle holding reed
Silicone therapy buttons MPM module Self test resistance
Monitor mode select knob Parameter panel assembly Silicon pad
Silicone monitor buttons Parameter interface FPC
Keyboard Recorder
Alarm lamp board Therapy port assembly
Waveform keyboard Capacitor fixing sheet
Status indicator shade
Alarm lamp shade Power base assembly
CPU and power
management board
On-position detection
switch kit
Main Unit
Handle kit Lithium battery
Conductive gel
container
Speaker Hook hole cover
Microphone Discharge board spring
8-1
Page 96

8.2 Main Unit

8.2.1 Exploded View

8.2.2 Parts List

No. Order Number Part Description Qty Remark
1 / Front Housing Assembly 1 /
2 / Rear Housing Assembly 1 /
3 / Paddle Tray Assembly 1 /
4 115-007858-00 Li-ion Bat Pack (14.8V3000mAh LI24I001A) 1 SN: EL-/ET-
115-062369-00 Li-ion Battery (5600mAh LI24I005A) SN: EZ-/EI-
5 115-007587-00 Pothook kit 1 /
8-2
Page 97

8.3 Front Housing Assembly

8.3.1 Exploded View

8.3.2 Parts List

No. Order Number Part Description Qty Remark
1 043-001114-00 knob-monitor 1 /
2 049-000216-00 Monitor Key(D3) 1 Without NIBP
047-011909-00 silica gel key (D3/NIBP/silk screen) With NIBP
3 047-003791-00 monitor key overlay(D3/English) 1 Without NIBP
047-011897-00 key label(D3/NIBP/english/silk screen) With NIBP
4 049-000217-00 Screen Key 1(D3) 1 /
5 049-000218-00 Screen Key 2(D3) 1 /
6 115-049435-00 Encoder assembly 1 Include 1,7
7 051-002778-00 Defibrillation Coder Board PCBA 1 /
8 0651-20-76734-51 mode knob 1 /
9 049-000215-00 Defib. Key(D3) 1 /
10 051-002638-00 0652 Front Keyboard PCBA 1 /
11 051-002851-00 0652 auxiliary keyboard PCBA 1 /
12 009-001102-00 assistant keyboard connecting line 1 /
13 115-049468-00 Truly Screen assembly 1 Include LCD & Cable
115-049469-00 Tianma Screen assembly Include LCD & Cable
14 115-049465-00 D3 front cover assembly 1 Include 20
115-049466-00 D2 front cover assembly Include 20
15 049-000170-00 lens-status 1 /
8-3
Page 98
No. Order Number Part Description Qty Remark
16 043-000986-00 lens-alarm 1 /
051-000491-00 0652 alarm light PCBA 1 /
17 801-0651-00126-00 encoder assembly 1
801-0651-00043-00 encoder assembly
18 0651-21-76884 mode switch cable 1 /
19 047-003789-00 mode overlay(D3/English) 1 /
047-005443-00 Mode label (D3/No pacer/English) /
20 047-003792-00 overlay of product maintenance led 1 /
21 115-041367-00 Speaker FRU(D3/D6) 1 /

8.4 Rear Housing Assembly

8.4.1 Exploded View

Include 8,18
Include 8,18
8-4
Page 99

8.4.2 Parts List

No. Order Number Part Description Qty Remark
Rear Housing Part
1 115-049467-00 D3 rear cover assembly 1 /
2 115-049429-00 Examine the switch assembly 1 /
3 043-001116-00 blot-defibrillation 1 /
Main Control Part
4 115-056282-00 0652 Main Control and Power Manage Board
FRU
5 115-050525-00 Wifi upgrade kit(5G/ NO Mainboard) 1 /
Therapy Module Part
6 047-018369-00 therapy module insulating trip 1 /
7 051-002615-00-00 Therapy Board FRU(with SW) 1 /
8 009-008106-00 Dry type Cable to treatment module 1 /
9 051-002616-00 Pace Board 1 /
10 006-000239-00
11 009-007382-00 Treatment of Plate load test cable (HVR) 1 /
MPM Assembly Part
12 043-007617-00 lock-recorder-D3 1 /
13 115-027548-00 AION Rhodium CO2 Analyzer 1 /
14 082-000862-00 pump. 12VDC with 120 wire and connector 1 /
21 082-000864-00 valve.CJV13-A12B2 1 /
15 051-002353-00 M51C-9008V3.0 SPO2 PCBA 1 /
100-000340-00 MSX2040 masimo SPO2 board
101-000469-00 Nellcor SPO2 PCBA /
16 051-002766-00 M51C 5L,MR/NC-SPO2,AO(D3) 1 /
051-002767-00 M51C 5L,3-SPO2,NIBP,AO(D3) /
17 115-036134-00 DRYLINE PRIME Receptacle with no panel 1 /
22 082-000098-00 630F Reducer 1 /
18 047-019272-00 Parameter label of 0652 ( ECG+SPO2 ) 1 /
047-019273-00 parameter label of 0652 ( ECG+SPO2+CO2 ) /
047-019274-00 parameter label of 0652 ( ECG+NIBP ) /
047-019275-00 parameter label of 0652 (ECG+SPO2+NIBP) /
047-018634-00 0652 Parameter Label(ECG+SPO2+NIBP+CO2) /
20 115-012155-00 Parameter Panel assembly 1 ECG
115-012152-00 Parameter Panel assembly ECG+Mindray
115-012153-00 Parameter Panel assembly ECG+Nellcor
115-012154-00 Parameter Panel assembly ECG+Masimo
115-021528-00 Parameter Panel assembly ECG+NIBP
115-021529-00 Parameter Panel assembly ECG+Mindray+NIBP
IND 1.5mH 5.6Ω
forNon-Mahwah
1 /
1 /
/
8-5
Page 100
No. Order Number Part Description Qty Remark
115-021530-00 Parameter Panel assembly ECG+Masimo+NIBP
115-021531-00 Parameter Panel assembly ECG+Nellcor+NIBP
19 801-6800-00080-00 TR6F Recorder 1 /
0000-10-11079 thermal printer head 50mm
Power Base Assembly Part
23 115-065041-00 DELTA Power Supply Board(18V 100W) FRU 1 /
24 115-049470-00 D3 Power base assembly 1 /
25 0651-20-76990 AC plug locked hook 1 /

8.5 Paddle Tray Assembly

8.5.1 Exploded View

8.5.2 Parts List

No. Order Number Part Description Qty Remark
1 801-0652-00014-00 Electrode base assembly 1 /
2 115-067390-00 bracket-electrode FRU 1 Include 2 pcs
3 801-0651-00124-00 test load cable 1 /
8-6
Loading...