Mindray BC-3000 Plus Hematology Analyzer User manual [gb]

BC-3000 Plus
Auto Hematology Analyzer
Operation Manual
Copyright
© 2003-2005 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved.
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this manual. This
manual may refer to information protected by copyrights or patents and does not convey any
license under the patent rights of Mindray, nor the rights of others. Mindray does not assume
any liability arising out of any infringements of patents or other rights of third parties.
Mindray intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner whatsoever without the written
permission of Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rent, adaption and translation of this manual
in any manner whatsoever without the written permission of Mindray is strictly forbidden.
, are the registered trademarks or trademarks owned by Mindray in China
and other countries. All other trademarks that appear in this manual are used only for
editorial purposes without the intention of improperly using them. They are the property of
their respective owners.
Responsibility on the Manufacturer Party
Contents of this manual are subject to changes without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for safety, reliability and performance of this product only in the
condition that:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and
local requirements;
I
the product is used in accordance with the instructions for use.
z This equipment must be operated by skilled/trained medical professionals.
z It is important for the hospital or organization that employs this equipment
to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
II
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
any Mindray product which has been subjected to misuse, negligence or accident;
any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
any product of any other manufacturer.
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to Mindray,
the following procedure should be followed:
1. Obtain return authorization: Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on the
outside of the shipping container. Returned shipments will not be accepted if the Mindray
number is not clearly visible. Please provide the model number, serial number, and a brief
description of the reason for return;
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges);
3. Return address: Please send the part(s) or equipment to the address offered by
Customer Service department.
III
Company Contact
Manufacture: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address:
Phone:
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, Shenzhen, P.R.China,518057
+86 755 26582479 26582888
Fax:
EC-Representative: Shanghai International Holding Corp. GmbH(Europe)
Address:
Phone:
Fax:
+86 755 26582500 26582501
Eiffestraβe 80, 20537 Hamburg Germany
0049-40-2513175
0049-40-255726
IV
Table of Contents
1 Using This Manual...................................................................................1-1
1.1 Introduction ............................................................................................ 1-1
1.2 Who Should Read This Manual ............................................................. 1-2
1.3 How to Find Information......................................................................... 1-3
1.4 Conventions Used in This Manual ......................................................... 1-4
1.5 Special Terms Used in This Manual....................................................... 1-5
1.6 Symbols ................................................................................................. 1-6
2 Understanding Your Analyzer.................................................................2-1
2.1 Introduction ............................................................................................ 2-1
2.2 Intended Use.......................................................................................... 2-2
2.3 User Interfaces....................................................................................... 2-3
2.4 Instrument Software............................................................................. 2-11
2.5 Reagents, Controls and Calibrators..................................................... 2-15
3 Understanding the System Principles................................................... 3-1
3.1 Introduction ............................................................................................ 3-1
3.2 Aspiration ............................................................................................... 3-2
3.3 Dilution ................................................................................................... 3-3
3.4 WBC/HGB Measurement....................................................................... 3-5
3.5 RBC/PLT Measurement ......................................................................... 3-9
3.6 Wash .................................................................................................... 3-13
4 Installing Your Analyzer .......................................................................... 4-1
4.1 Introduction ............................................................................................ 4-1
4.2 Installation Requirements....................................................................... 4-2
4.3 Unpacking .............................................................................................. 4-4
4.4 Installation Procedure ............................................................................ 4-6
4.5 Starting the Analyzer ............................................................................ 4-16
5 Customizing the Analyzer Software....................................................... 5-1
5.1 Introduction ............................................................................................ 5-1
5.2 Print........................................................................................................ 5-2
5.3 Count time.............................................................................................. 5-6
5.4 Password ............................................................................................... 5-8
5.5 Transmission........................................................................................ 5-14
5.6 Setting system time (Date & Time) ...................................................... 5-19
5.7 Gain...................................................................................................... 5-21
1
Table of Contents
5.8 Auto Clean Time .................................................................................. 5-26
5.9 Reagent Exp. Date............................................................................... 5-27
5.10 Report Title external keyboard required)....................................... 5-29
5.11 Parameter Units ................................................................................... 5-31
5.12 Other .................................................................................................... 5-34
6 Operating Y our Analyzer.........................................................................6-1
6.1 Introduction ............................................................................................ 6-1
6.2 Initial Checks.......................................................................................... 6-2
6.3 Power-on................................................................................................ 6-3
6.4 Daily Quality Control .............................................................................. 6-4
6.5 Sample Collection and Handling............................................................ 6-5
6.6 Running Whole Blood Samples ............................................................. 6-8
6.7 Running Prediluted Samples ............................................................... 6-17
6.8 Shutdown ............................................................................................. 6-26
7 Reviewing Sample Results..................................................................... 7-1
7.1 Introduction ............................................................................................ 7-1
7.2 Browsing All Sample Results ................................................................. 7-2
7.3 Searching for Interested Sample Results ............................................ 7-17
8 Using the QC Programs.......................................................................... 8-1
8.1 Introduction ............................................................................................ 8-1
8.2 “L-J Analysis” Program........................................................................... 8-2
8.3 “X Analysis ” Program ........................................................................ 8-15
8.4 “X-R Analysis” program ....................................................................... 8-28
8.5 “X-B Analysis” Program........................................................................ 8-39
9 Using the Calibration Programs.............................................................9-1
9.1 Introduction ............................................................................................ 9-1
9.2 When to calibrate ................................................................................... 9-2
9.3 How to Calibrate..................................................................................... 9-3
10 Maintaining Your Analyzer....................................................................10-1
10.1 Introduction .......................................................................................... 10-1
10.2 General Guidelines .............................................................................. 10-2
10.3 Using the “Maintenance” Program....................................................... 10-4
10.4 Using the “System Status” Program................................................... 10-24
10.5 Using the “Valve Test” Program ......................................................... 10-26
10.6 Using the “System Test” Program ...................................................... 10-27
10.7 Using the “Prepare to ship” Program ................................................. 10-28
10.8 Using the “Error Message” Program.................................................. 10-31
10.9 Calibrating Sample Probe Position .................................................... 10-33
10.10 Replacing the Probe Wipe. ................................................................ 10-37
2
Table of Contents
10.11 Replacing the Filter of the Vacuum Chamber.................................... 10-38
10.12 Using the [Flush] key ......................................................................... 10-39
10.13 Using the [Startup] key....................................................................... 10-40
11 Troubleshooting Your Analyzer............................................................ 11-1
11.1 Introduction .......................................................................................... 11-1
11.2 Errors without available error messages ............................................. 11-2
11.3 Errors indicated by error messages..................................................... 11-3
12 Appendices .............................................................................................. A-1
A Index ......................................................................................................A-1
B Specifications ......................................................................................... B-1
C Precautions, Limitations and Hazards ...................................................C-1
D Communication ......................................................................................D-1
3
1 Using This Manual
1.1 Introduction
This chapter explains how to use your BC-3000 Plus operation manual, which is shipped with
your BC-3000 Plus hematology analyzer and contains reference information about the
BC-3000 Plus and procedures for operating, troubleshooting and maintaining the analyzer.
Read this manual carefully before operating your analyzer and operate your analyzer strictly
as instructed in this manual.
z Be sure to operate your analyzer strictly as instructed in this manual.
1-1
Using This Manual
1.2 Who Should Read This Manual
This manual contains information written for clinical laboratory professionals to:
learn about the BC-3000 Plus hardware and software;
customize system settings;
perform daily operating tasks;
perform system maintenance and troubleshooting.
1-2
Using This Manual
1.3 How to Find Information
This operation manual comprises 11 chapters and 4 appendices. Refer to the table below to
find the information you need.
If you want to … See …
learn about the intended use and parameters of the BC-3000
Plus
learn about the hardware and software of the BC-3000 Plus
learn about how the BC-3000 Plus works
learn about how to install the BC-3000 Plus
learn about how to define/adjust system settings
learn about how to use the BC-3000 Plus to perform your daily
operating tasks
learn about how to review the saved analysis results
learn about how to use the quality control programs
Chapter 2 Understanding Y our Analyzer Chapter 2 Understanding Y our Analyzer Chapter 3 Understanding the System Principles Chapter 4 Installing Your Analyzer Chapter 5 Customizing the Analyzer Software Chapter 6 Operating Your Analyzer Chapter 7 Reviewing Sample Results Chapter 8 Using the QC
learn about how to calibrate the BC-3000 Plus
learn about how to maintain/service the BC-3000 Plus
learn about the meanings of the error messages and how to
correct the problems
learn about the technical specifications of the BC-3000 Plus
see the summary of all safety messages included in this
manual
learn about the communication protocol of the BC-3000 Plus
1-3
Programs Chapter 9 Using the Calibration Programs Chapter 10 Maintaining Your Analyzer Chapter 11 Troubleshooting Y our Analyzer Appendix B Specifications Appendix C Precautions, Limitations and Hazards Appendix D Communication
Using This Manual
1.4 Conventions Used in This Manual
This manual uses certain typographical conventions to clarify meaning in the text:
All capital letters enclosed in [ ] indicate a key name (either on the built-in keypad or the
external keyboard), such as [ENTER].
All capital, bold and italic letters indicate a special operation defined in the following
section, such as SELECT.
Bold letters included in “ ” indicate text you can find on the screen, such as “Prepare to ship”.
Bold letters indicate defined screen areas/fields, such as System Status area, or chapter titles, such as Chapter 1 Using This Manual.
All illustrations in this manual are provided as examples only. They may not necessarily
reflect your analyzer setup or data displayed.
1-4
Using This Manual
1.5 Special Terms Used in This Manual
When you read … It means …
CLICK
ENTER
DELETE
to press the arrow keys ([←][→] [][]) as needed to move the
cursor to a certain software button on screen and press
[ENTER].
to press the arrow keys ([←][] [][]) as needed to move
cursor to the desired edit box and use the built-in keypad
or the external keyboard to enter the desired characters
or digits. Note that besides the numeric keys you may
also use the [PgUp] or [PgDn] keys to enter digits; or
to scan the number in using the bar-code scanner.
to press the arrow keys ([←][→] [][]) as needed to move the
cursor to the character or digit to the left of the one you want to
delete and press [DEL]; or to press the arrow keys
([][][][]) as needed to move the cursor to the character
or digit to the right of the one you want to delete and press
[BackSpace] on the external keyboard.
to move the cursor to the character or digit you want to change
MODIFY
SELECT from “ ** ”
pull-down list
SELECT
and re-enter the desired one using either the built-in keypad or
the external keyboard.
to press the arrow keys ([←][→] [][]) as needed to move the
cursor to the desired edit box and press [ENTER] to display
the pull-down list and press [] or [] to move the cursor to the
desired item and press [ENTER] to select it.
to press the arrow keys ([←][→] [][]) as needed to the
desired item and press [ENTER].
z This analyzer adopts a fixed decimal point. You can enter the digits without
bothering to look for the [.] on the external keyboard.
1-5
Using This Manual
1.6 Symbols
You will find the following symbols in this manual.
When you see… Then…
read the statement below the symbol. The statement is
You may find the following symbols on the analyzer or the reagents.
When you see… It means…
alerting you to an operating hazard that can cause
personnel injury.
read the statement below the symbol. The statement is
alerting you to a possibility of analyzer damage or unreliable
analysis results.
read the statement below the symbol. The statement is
alerting you to information that requires your attention.
read the statement below the symbol . The statement is
alerting you to a potentially biohazardous condition.
EQUIPOTENTIALITY
CAUTION, CONSULT ACCOMPANYING
DOCUMENTS.
BIOLOGICAL RISK
HIGH VOLTAGE
ALTERNATING CURRENT
FOR IN VITRO DIAGNOSTIC USE
1-6
Using This Manual
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
TEMPERATURE LIMITATION
CONSULT INSTRUCTIONS FOR USE
THE DEVICE IS FULLY CONFORMANCE
WITH THE COUNCIL DIRECTIVE
CONCERNING IN VITRO DIAGNOSTIC
MEDICAL DEVICES 98/79/EC.
MANUFACTURER
AUTHORISED REPRESENTATIVE IN THE
EUROPEAN COMMUNITY
IRRITATING SUBSTANCE
THE FOLLOWING DEFINITION OF THE
WEEE LABEL APPLIES TO EU MEMBER
STATES ONLY: THE USE OF THIS SYMBOL
INDICATES THAT THIS PRODUCT SHOULD
NOT BE TREATED AS HOUSEHOLD
WASTE. BY ENSURING THAT THIS
PRODUCT IS DISPOSED OF CORRECTLY,
YOU WILL HELP PREVENT BRINGING
POTENTIAL NEGATIVE CONSEQUENCES
1-7
Using This Manual
TO THE ENVIRONMENT AND HUMAN
HEALTH. FOR MORE DETAILED
INFORMATION WITH REGARD TO
RETURNING AND RECYCLING THIS
PRODUCT, PLEASE CONSULT THE
DISTRIBUTOR FROM WHOM YOU
PURCHASED THE PRODUCT.
1-8
Using This Manual
5
6
(4)
(1)
(2)
(3)
Figure1-1 Back of the Analyzer
1
Equipotentiality.
2
Connect only to a properly earth grounded outlet;
To avoid electric shock, disconnect power cord prior to removing or replacing fuse;
Replace fuse only with the type and rating specified.
3
Biological risk.
4
The following definition of the WEEE label applies to EU member states only: The use of this symbol indicates that this product should not be treated as household waste. By ensuring
that this product is disposed of correctly, you will help prevent bringing potential negative
1-9
Using This Manual
consequences to the environment and human health. For more detailed information with
regard to returning and recycling this product, please consult the distributor from whom you
purchased the product.
1-10
Using This Manual
(5)
Figure1-2 Inside right of the analyzer
5
To avoid being injured, do not put hand under the motor when the machine is running.
1-11
2 Understanding Your Analyzer
2.1 Introduction
The BC-3000 Plus auto hematology analyzer is a quantitative, automated hematology
analyzer and leukocyte differential counter for In Vitro Diagnostic Use in clinical laboratories.
2-1
Understanding Your Analyzer
2.2 Intended Use
z The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that require additional studies.
The analyzer is used for the quantitative determination of the following 19 parameters and 3
histograms of blood samples.
White Blood Cell or leukocyte WBC
Lymphocyte Lymph#
Mid-sized cell Mid#
Granulocyte Gran#
Lymphocyte percentage Lymph%
Mid-sized cell percentage Mid%
Granulocyte percentage Gran%
Red Blood Cell or erythrocyte RBC
Hemoglobin Concentration HGB
Mean Corpuscular (erythrocyte) Volume MCV
Mean Cell (erythrocyte) Hemoglobin MCH
Mean Cell (erythrocyte) Hemoglobin Concentration
MCHC
Red Blood Cell (erythrocyte) Distribution Width
Coefficient of Variation
Red Blood Cell (erythrocyte) Distribution Width
Standard Deviation
Hematocrit HCT
Platelet PLT
Mean Platelet Volume MPV
Platelet Distribution Width PDW*
Plateletcrit PCT*
White Blood Cell Histogram WBC Histogram
Red Blood Cell Histogram RBC Histogram
Platelet Histogram PLT Histogram
2-2
RDW-CV
RDW-SD
Understanding Your Analyzer
2.3 User Interfaces
Figure2-1 Front view
1 ---- LCD 2 ---- Keypad
3 ---- Recorder 4 ---- Power indicator
5 ---- Aspirate key 6 ---- Sample probe
2-3
Understanding Your Analyzer
Figure2-2 Back view
1 --- RS-232 Port1 2 --- Parallel Port 3 --- RS-232 Port2 4 --- Keyboard interface 5 --- Power Interface of Floppy Disk Drive 6 --- Safety labeling 7 --- Diluent inlet 8 --- Diluent sensor connector 9 --- Rinse sensor connector 10 --- Waste outlet 11 --- Rinse inlet 12--- Power switch 13--- Equipotentiality
14--- WEEE labeling
2-4
Understanding Your Analyzer
1
2
3
4
5
Figure2-3 Inside front of the analyzer
1 --- Elevator motor 2 --- Sample probe 3 --- Probe wipe 4 --- WBC shielding box 5 --- RBC shielding box 6 --- Aspirate key
6
2-5
Understanding Your Analyzer
2
1
2
3
4
5
0
1
8
1
7
1
6
1
5
1
4
1
6
7
8
9
1
0
1
1
3
1
2
Figure2-4 Inside right of the analyzer
1 --- Valve8 2 --- Volumetric metering unit 3 --- Vacuum chamber 4 --- Valve13 5 --- Valve14 6 --- Valve12 7 --- Valve11 8 --- Valve10 9 --- Valve2 10 --- Valve9 11 --- 50ul and 2.5ml motor 12 --- 10ml motor 13 --- 2.5ml syringe 14 --- 50ul syringe 15 --- 10ml syringe 16 --- Valve6 17 --- Valve4 18 --- Valve3 19 --- Valve1 20 --- Valve5 21 --- Valve15 22 --- Valve16 23 --- Valve17 24 --- Valve7 25 --- Valve18
2-6
Understanding Your Analyzer
Figure2-5 Inside left of the analyzer
1 --- Fluid pump 2 --- Gas pump 3 --- Pressure chamber
2-7
Understanding Your Analyzer
2.3.1 LCD
The LCD is located on the front panel of the analyzer. It displays all alphanumeric and graphic
data.
2.3.2 Input Devices
The input devices include the aspirate key, built-in keypad and PS/2 keyboard.
Aspirate key
The aspirate key is located behind the sample probe, as Figure2-6 shows. You can press the
key to start the selected run cycle or dispense diluent.
Figure2-6 Sample probe and aspirate key
1.Sample Probe 2. Aspirate Key
2-8
Understanding Your Analyzer
Built-in keypad
The 23-key keypad is located below the LCD, as Figure2-7 shows.
Figure2-7 Built-in keypad
PS/2 keyboard
The analyzer can also be controlled by an external PS/2 keyboard that should be connected
to the analyzer’s keyboard interface. See the table below for the correspondence between
the keypad keys and the keyboard keys and for their functions.
Keypad PS/2 keyboard Function
[MENU] [Esc] To enter/exit the system menu.
[PRINT] [P]or [p] To print out data by the recorder or printer.
[DEL] [Delete] or [Del] To delete data and characters.
[ENTER] [Enter] To confirm or execute an operation.
[][][][] [][][][] To move the cursor.
[0]...[9] [0]...[9] To enter digits.
[PgUp][PgDn] [PageUp] [PageDown] To scroll screen.
[Flush] / To unclog the apertures.
[FEED] / To advance the recorder paper.
[MAIN] /
To go back to the “Count” screen.
[DILUENT] [D]or [d] To enter the dispensing diluent state.
[STARTUP] / To execute the startup procedure (flushing the
fluidic lines and checking background).
[ID] [ I ]or [i] To call out the screen to enter patient information.
/ Alphanumeric keys and
other function keys.
To enter alphanumeric data or initiate a function.
2-9
Understanding Your Analyzer
2.3.3 Recorder
The thermal recorder is located on the front panel. You can use it to print out analysis reports
and other interested information.
2.3.4 Keyboard Interface
A PS/2 keyboard can be connected here.
2.3.5 Serial Ports
The analyzer provides two RS-232 ports, one for connecting the scanner and the other for
connecting a computer (host).
2.3.6 Parallel Port
The analyzer provides a parallel port to connect a printer or a floppy disk drive (a floppy disk
drive is needed to upgrade the system software; the drive can only be connected by a
Mindray-supplied cable).
2.3.7 Power Supply for the Floppy Disk Drive
It supplies power to the connected floppy disk drive. Only the drive power cable supplied by
Mindray can be used.
2.3.8 Power Indicator
The power indicator tells you whether the analyzer is on, off or in the screen saver mode.
2.3.9 PrinterOptional
An external printer can be connected to the parallel port at the back of the analyzer. You can
use it to print out a detailed report and other interested information.
2.3.10 ScannerOptional
A bar-code scanner can be connected to the RS-232 port 1 of the analyzer. You can use it to
scan the bar-coded sample IDs into the analyzer.
z Be sure to use the printer and/or scanner of the specified model.
2-10
Understanding Your Analyzer
2.4 Instrument Software
2.4.1 Main Screen
After finishing the startup procedure, the analyzer enters the “Count” screen, which is the
screen to be used most frequently, hence the name “Main screen”. The main screen is shown
in Figure2-8.
Error message area
Title area
Menu area
Error Message Area
Status area
Analysis area
Help area
Figure2-8 “Count” screen
The Error Message area displays error messages one by one, alternating every two
seconds.
Title Area
The Title area displays the title of the current screen, which, in case of Figure2-8, is “Count.
Status Area
System Status Area
The System Status area displays whether this analyzer is ready for the next analysis. When it displays “Ready”, it means this analyzer is ready and you can proceed to analyze the next sample. When it displays “Waiting”, it means the analyzer is not ready for the next run yet. When it displays “Running”, it means this analyzer is analyzing a sample.
2-11
Understanding Your Analyzer
Count Mode Area
The Count Mode area displays in which analysis (count) mode, whole blood or prediluted,
the next sample is to be analyzed.
Transmission status
A live animation is displayed in this area when the transmission is in process.
System Time Area
The System Time area displays the system time (in the 24-hour format).
Analysis Result Area
The Analysis Result area displays the analysis result, including sample ID, analysis time of
the current sample.
Menu Area
When you press [MENU], this area displays the system menu.
Help Area
The Help area reminds you how to proceed to the next step.
2.4.2 System Menu
Press the [MENU] button and the system menu, shown in the figure below, will pop up.
Figure2-9 System menu
The system menu contains 9 programs. The programs followed by “
sub-menus. See the figure below for the fully expanded menu.
2-12
”s have further
Count
Sample Mode
Understanding Your Analyzer
Review
Quality Control
Setup
Sample Review
Search Review
Search Histogram Review
L-J Analysis
X Analysis
X-R Analysis
X-B Analysis Limit
Print
Count Time
Password
Ref. range Transmission Date & Time
L-J/X/X-R Edit L-J/X/X-R Count L-J/X/X-R Graph
L-J/X/X-R Table
Samples/Batch
Start/Stop
X-B Graph
X-B Table
General
Man
Woman
Sample Table Review
Sample Histogram ReviewSearch Table Review
File 1 File 2 File 3 File 4 File 5 File 6 File 7 File 8 File 9
File 1 File 2 File 3 File 4 File 5 File 6 File 7 File 8 File 9
File 1 File 2 File 3 File 4 File 5 File 6 File 7 File 8 File 9
File 1 File 2 File 3 File 4 File 5 File 6 File 7 File 8 File 9
Service
Calibration
Help
Shutdown
Gain
Auto Clean Time
Reagent Exp. Date
Report Title
Parameter Units
Other Setting
Maintenance
System Status
Valve Test
System Test
Prepare to Ship
Error Message
Manual Calibration
Auto Calibration
Fresh Blood
Child
Neonate
Figure 2-10 Fully expanded system menu
2-13
Understanding Your Analyzer
You can select the desired program as instructed below.
If you want to… Select…
analyze samples
select an appropriate analysis mode
review sample results
run the QC program
customize system software
maintain/service the analyzer
calibrate the analyzer
look for help
shut down the analyzer
Count Sample Mode Review Quality Control Setup Service Calibration Help Shutdown
2-14
Understanding Your Analyzer
2.5 Reagents, Controls and Calibrators
Because the analyzer, reagents (diluent, rinse, lyse, probe cleanser and E-Z cleanser),
controls, and calibrators are components of a system, performance of the system depends on
the combined integrity of all components. You should only use the Mindray-specified reagents
(see Appendix B Specifications), which are formulated specifically for the fluidic system of
your analyzer in order to provide optimal system performance. Do not use the analyzer with
reagents from multiple suppliers. In such use, the analyzer may not meet the performance
specified in this manual and may provide unreliable results. All references related to reagents
in this manual refer to the reagents specifically formulated for this analyzer.
Each reagent package must be examined before use. Inspect the package for signs of
leakage or moisture. Product integrity may be compromised in packages that have been
damaged. If there is evidence of leakage or improper handling, do not use the reagent.
z Store and use the reagents as instructed by instructions for use of the
reagents.
z When you have changed the diluent, rinse or lyse, run a background to see
if the results meet the requirement.
z Pay attention to the expiration dates and open-container stability days of all
the reagents. Be sure not to use expired reagents.
z After installing a new container of reagents, keep the reagents still for a
while before using them.
2.5.1 Diluent
The diluent is formulated to meet the following requirements:
To dilute the blood samples;
To provide the blood cells with an environment similar to the blood plasma;
To maintain the cell volume of each red blood cell and platelet during the count and
sizing portion of the measurement cycle;
To provide a conductive medium for impedance counting of white and red blood cells
and platelets.
2-15
Understanding Your Analyzer
2.5.2 Lyse
The lyse is formulated to meet the following requirements:
To rapidly break down red blood cell walls, release the hemoglobin from the cell, and
reduce the size of cellular debris to a level that does not interfere with white blood cell
counting.
To convert hemoglobin to a complex whose absorbance is determined by the
hemoglobin concentration.
2.5.3 Rinse
The rinse is formulated to rinse the baths and metering tubes and to provide proper meniscus
formation in the metering tubes and maintain it during each measurement cycle.
2.5.4 E-Z Cleanser
The E-Z (enzymatic) cleaner is an enzyme-based isotonic, cleaning solution and wetting
agent formulated to clean the fluidic lines and baths.
2.5.5 Probe Cleanser
The probe cleanser is an alkaline cleaning solution formulated to clean the fluidic lines,
apertures and baths.
2.5.6 Controls and Calibrators
The controls and calibrators are used to verify accurate operation of and calibrate the
analyzer.
The controls are commercially prepared whole-blood products used to verify that the analyzer
is functioning properly. They are available in low, normal, and high levels. Daily use of all
levels verifies the operation of the analyzer and ensures reliable results are obtained. The
calibrators are commercially prepared whole-blood products used to calibrate the analyzer.
Read and follow the instructions for use to use the controls and calibrators. All references
related to controls and calibrators in this manual refer to the controls and calibrators reagents
specifically formulated for this analyzer. You should buy those controls and calibrators from
Mindrayor Mindray-authorized distributors.
2-16
3 Understanding the System
Principles
3.1 Introduction
The two independent measurement methods used in this analyzer are:
the Coulter method for determining the WBC, RBC, and PLT data;
the colorimetric method for determining the HGB.
During each analysis cycle, the sample is aspirated, diluted and mixed before the
determination for each parameter is performed.
3-1
Understanding the System Principles
3.2 Aspiration
This analyzer can process two types of blood samples – whole blood samples and prediluted
blood samples.
If you are to analyze a whole blood sample, you can simply present the sample to the sample
probe and press the aspirate key to aspirate 13µL of the sample into the analyzer.
If you are to analyze a capillary blood sample, you should first manually dilute the sample (20 μL of capillary sample needs to be diluted by 0.7mL of diluent) and then present the
pre-diluted sample to the sample probe and press the aspirate key to aspirate 0.3ml of the
sample into the analyzer.
3-2
Understanding the System Principles
3.3 Dilution
Usually in blood samples, the cells are too close to each other to be identified or counted. For
this reason, the diluent is used to separate the cells so that they are drawn through the
aperture one at a time as well as to create a conductive environment for cell counting.
Moreover, red blood cells usually outnumber white blood cells by 1,000 times. For this reason,
lyse needs to be added to the sample to eliminate the red cells before the WBC counting.
When analyzing a whole blood sample, this analyzer aspirates 13µL of the sample and
follows the procedure presented in Figure3-1 to dilute it before proceeding to the actual
analysis.
13uL of Whole blood sample
3.5mL of diluent
About 1:269 dilution
0.5mL of lyse
About 1:308 dilution for
the WBC/HGB analysis
15.6uL
About 2.6mL of diluent
About 1:44872 dilution for
the RBC/PLT analysis
Figure3-1 How a whole blood sample is diluted
3-3
Understanding the System Principles
t
When analyzing a prediluted sample, you should first collect 20μL of capillary sample and
dispense 0.7mL of diluent from this analyzer to predilute it. Then the analyzer aspirates 0.3ml
of the pre-diluted sample for further dilution, as Figure3-2 shows.
20uL of capillary blood sample
0.7mL of diluent
1:36 dilution
0.3mL
2.9 mL of diluent
About 1:384 dilution
24.8uL
0.36 mL of lyse
About 2.8mL of diluent
About 1:428 dilution for
WBC/HGB measurement
About 1:43355 dilution
for RBC/PLT
measuremen
Figure3-2 How a capillary sample is diluted
3-4
Understanding the System Principles
3.4 WBC/HGB Measurement
3.4.1 Volumetric Metering
An accurate cell count cannot be obtained unless the precise volume of diluted sample that
passes through the aperture during the count portion of the analysis cycle (the count cycle) is
known. This analyzer uses a volumetric metering unit to control the count cycle and to ensure
that a precise volume of sample is analyzed.
The metering unit controlling the WBC count cycle consists of a metering tube with two
optical sensors mounted on it. This tube ensures that a precise amount of diluted sample is
measured during each count cycle. The exact amount is determined by the distance between
the two optical sensors. The rinse is used to create a meniscus in the metering tube. The
count cycle starts when the meniscus reaches the upper sensor and stops when the
meniscus reaches the lower sensor. The amount of time required for the meniscus to travel
from the upper sensor to the lower sensor is called the WBC Count Time and is measured in
seconds. At the end of the count cycle, the measured count time is compared to the
pre-defined reference count time (see Chapter 5.3 for details). If the former is less than or
greater than the latter by 2 seconds or more, the analyzer will report a WBC bubble or WBC
Clog error. Seeing the error message, you can refer to Chapter 11 Troubleshooting Your Analyzer for solutions.
Upper sensor
Lower sensor
1 Empty when start 2 Liquid surface fall down through the
volumeritc tube
Upper sensor
Lower sensor
3 Counting start when the liquid surface pass the upper sensor.
4 Counting end when the liquid surface pass the lower sensor
Upper sensor
Lower sensor
Upper sensor
Lower sensor
Figure3-3 Volumetric metering process
3-5
Understanding the System Principles
3.4.2 Measurement Principles
WBC measurement
WBCs are counted and sized by the Coulter method. This method is based on the
measurement of changes in electrical resistance produced by a particle, which in this case is
a blood cell, suspended in a conductive diluent as it passes through an aperture of known
dimensions. An electrode is submerged in the liquid on both sides of the aperture to create an
electrical pathway. As each particle passes through the aperture, a transitory change in the
resistance between the electrodes is produced. This change produces a measurable
electrical pulse. The number of pulses generated signals the number of particles that passed
through the aperture. The amplitude of each pulse is proportional to the volume of each
particle. Each pulse is amplified and compared to the internal reference voltage channels,
which only accepts the pulses of a certain amplitude. If the pulse generated is above the
WBC threshold, it is counted as a WBC.
Figure3-4
HGB measurement
HGB is determined by the colorimetric method. The WBC/HGB dilution is delivered to the
WBC bath where it is bubble mixed with a certain amount of lyse, which converts hemoglobin
to a hemoglobin complex that is measurable at 525 nm. An LED is mounted on one side of
the bath and emits a beam of light, which passes through the sample and a 525nm filter, and
then is measured by a photo-sensor that is mounted on the opposite side. The signal is then
amplified and the voltage is measured and compared to the blank reference reading
3-6
Understanding the System Principles
(readings taken when there is only diluent in the bath). The HGB is calculated per the
following equation and expressed in g/L.
HGB(g/L) = Constant×Log
(Blank Photocurrent/Sample Photocurrent)
10
3.4.3 Derivation of WBC-Related Parameters
WBC
9
WBC (10
passing through the aperture. Note that when you observe NRBCs (nucleated red blood
cells), which do not react with the lyse and can be mistaken by the analyzer for white cells, in
the microscope, be sure to correct the system-generated result by the following formula,
/ L) is the number of leukocytes measured directly by counting the white blood cells
WBC'WBC
×
where WBC represents the system-generated white cell number, NRBC the number of
NRBCs counted in 100 white cells and WBC′ the corrected white cell number.
WBC differentia
With the help of the diluent and lyse, this analyzer can size the white cells into three
sub-populations - lymphocytes, mid-sized cells (including monocytes, basophils and
eosinophils ) and granulocytes. Based on the WBC histogram, this analyzer calculates Lymph, Mid and Gran as follows and express the results in percents.
Lymph% ×
Mid% ×
Gran% ×
=
=
=
PL
PM
++
PG
100
NRBC100
100
PGPMPL
++
100
PGPMPL
100
PGPMPL
++
9
L/10
where PLT = particles in the lymphocyte region(
9
L/10
9
3-7
)
L/10
).
PM = particles in the mid size region(
PG = particles in the granulocyte region(
Having achieved the three parameters above, this analyzer proceeds to calculate the
Lymph# , Mid# and Gran# per the following equations and express them in
)
9
L/10
.
Understanding the System Principles
×
×
×
=
WBC histogram
Besides the parameters mentioned above, this analyzer also presents a WBC histogram, whose x-coordinate represents the cell volume(fL)and y-coordinate represents the number
of the cells. The histogram is presented in the Analysis area of the “Count” screen when the
analysis is done. You can also review the histograms of the stored patient results (see
Chapter 7 Reviewing Sample Results).
The first three discriminators of the WBC histogram can be adjusted in case you are not
satisfied with the result. Note that you cannot adjust them if the WBC result is less than 0.5
or out of the operating range.
#Lymph
=
#Mid
100
=
#Gran
100
WBC%Lymph
100
WBC%Mid
WBC%Gran
3.4.4 HGB
Using the colorimetric method, this analyzer calculates hemoglobin concentration (g/L) as
follows.
HGB(g/L)=Constant×Log
(Blank Photocurrent/Sample Photocurrent)
10
3-8
Understanding the System Principles
3.5 RBC/PLT Measurement
3.5.1 Volumetric Metering
An accurate cell count cannot be obtained unless the precise volume of diluted sample that
passes through the aperture during the count cycle is known. This analyzer uses a volumetric
metering unit to control the count cycle and to ensure that a precise volume of sample is
analyzed for the measurement.
The metering unit controlling the RBC/PLT count cycle consists of a metering tube with two
optical sensors mounted on it. This tube ensures that a precise amount of diluted sample is
measured during each count cycle. The exact amount is determined by the distance between
the two optical sensors. The rinse is used to create a meniscus in the metering tube. The
count cycle starts when the meniscus reaches the upper sensor and stops when the
meniscus reaches the lower sensor. The amount of time required for the meniscus to travel
from the upper sensor to the lower sensor is called the RBC Count Time and is measured in
seconds. At the end of the count cycle, the measured count time is compared to the
pre-defined reference count time (see Chapter 5.3 for details). If the former is less than or
greater than the latter by 2 seconds or more, the analyzer will report a RBC bubble or RBC
Clog error. Seeing the error message, refer to Chapter 11 Troubleshooting for solutions.
Upper sensor
Lower sensor
1 Empty when start 2 Liquid surface fall down through the
volumeritc tube
Upper sensor
Lower sensor
3 Counting start when the liquid surface pass the upper sensor.
4 Counting end when the liquid surface pass the lower sensor
Upper sensor
Lower sensor
Upper sensor
Lower sensor
Figure3-5 Volumetric metering process
3-9
Understanding the System Principles
3.5.2 Measurement Principles
RBC/PLT measurement
RBCs/PLTs are counted and sized by the Coulter method. This method is based on the
measurement of changes in electrical resistance produced by a particle, which in this case is
a blood cell, suspended in a conductive diluent as it passes through an aperture of known
dimensions. An electrode is submerged in the liquid on both sides of the aperture to create an
electrical pathway. As each particle passes through the aperture, a transitory change in the
resistance between the electrodes is produced. This change produces a measurable
electrical pulse. The number of pulses generated signals the number of particles that passed
through the aperture. The amplitude of each pulse is proportional to the volume of each
particle. Each pulse is amplified and compared to the internal reference voltage channels,
which only accepts the pulses of a certain amplitude. If the pulse generated is above the
RBC/PLT lower threshold, it is counted as a RBC/PLT.
Figure3-6
3.5.3 Derivation of RBC-Related Parameters
RBC
12
RBC (10 passing through the aperture.
/L) is the number of erythrocytes measured directly by counting the erythrocytes
3-10
Understanding the System Principles
×
MCV
Based on the RBC histogram, this analyzer calculates the mean cell volume (MCV) and expresses the result in fL .
This analyzer calculates the HCT (%), MCH (pg) and MCHC (g/L) as follows:
This analyzer calculates the HCT (%), MCH (pg) and MCHC(g/L) as follows:
=
Where the RBC is expressed in 10
HCT
MCH =
MCHC ×=
HGB RBC
MCVRBC
10
HGB
100
HCT
12
/LMCV in fL and HGB in g/L.
RDW-CV
Based on the RBC histogram, this analyzer calculates the CV (Coefficient of Variation) of the erythrocyte distribution width.
RDW-SD
RDW-SD (RBC Distribution Width – Standard Deviation, fL) is set on the 20% frequency level with the peak taken as 100%, Figure3-7 shows.
Figure3-7
RBC Histogram
Besides the parameters mentioned above, this analyzer also presents a RBC histogram, whose x-coordinate represents the cell volume(fL)and y-coordinate represents the number
of the cells. The histogram is presented in the Analysis area of the “Count” screen when the
analysis is done. You can also review the histograms of the stored patient results (see
Chapter 7 Reviewing Sample Results).
The two discriminators of the RBC histogram can be adjusted in case you are not satisfied
with the result. Note that you cannot adjust them if the RBC result is less than 0.2 or out of the
operating range.
3-11
Understanding the System Principles
3.5.4 Derivation of PLT-Related Parameters
PLT
PLT (10
MPV
Based on the PLT histogram, this analyzer calculates the mean platelet volume (MPV, fL).
PDW
Platelet distribution width (PDW) is the geometric standard deviation (GSD) of the platelet
size distribution. Each PDW result is derived from the platelet histogram data and is reported
as 10(GSD).
9
/L) is measured directly by counting the platelets passing through the aperture.
PCT
This analyzer calculates the PCT as follows and express it in %.
9
Where the PLT is expressed in 10
PCT
=
/L and the MPV in fL.
MPVPLT
×
10000
PLT Histogram
Besides the parameters mentioned above, this analyzer also presents a PLT histogram, whose x-coordinate represents the cell volume(fL)and y-coordinate represents the number
of the cells. The histogram is presented in the Analysis area of the “Count” screen when the
analysis is done. You can also review the histograms of the stored patient results (see
Chapter 7 Reviewing Sample Results).
The two discriminators of the PLT histogram can be adjusted in case you are not satisfied
with the result. Note that you cannot adjust them if the PLT result is less than 10 or out of the
operating range.
3-12
Understanding the System Principles
3.6 Wash
After each analysis cycle, each element of the analyzer is washed.
The sample probe is washed internally and externally with diluent;
The WBC bath is washed with diluent and rinse;
The RBC/PLT bath is washed with diluent and rinse;
The metering tube is washed with rinse.
3-13
4 Installing Your Analyzer
4.1 Introduction
This chapter introduces the installation procedure of the BC-3000 Plus. To ensure all system
components are functioning correctly and to verify system performance, a
Mindray-authorized representative will handle the installation and initial software setup.
z Installation by personnel not authorized or trained by Mindray may damage
your analyzer. Do not install your analyzer without the presence of Mindray
-authorized personnel.
4-1
Installing Your Analyzer
4.2 Installation Requirements
Before installation, you should ensure that the following space, power and environmental
requirements are met.
4.2.1 Space Requirements
Check the site for proper space allocation. In addition to the space required for the analyzer
itself, arrange for
at least 28 cm on each side, which is the preferred access to perform service
procedures;
at least 10 cm behind for cabling and ventilation;
enough room on or below the countertop to accommodate the diluent, rinse and waste
containers.
4.2.2 Power Requirements
Check the availability of a power outlet that is
a female receptacle;
single phase with ground (confirmed third-wire earth ground);
100 VAC - 240 VAC;
50/60 Hz.
z Make sure the analyzer is properly grounded. z If a power outlet with confirmed third-wire earth ground is not available, be
sure to connect the equipotentiality pole at the back of the analyzer to the ground.
z Only install a 250V T4A fuse on the analyzer. z Before turning on the analyzer , make sure the input voltage meets the above
requirements.
4.2.3 General Environment
 Operating temperature: 15 ℃ - 35 ℃;  Optimal operating temperature: 15 ℃ - 30 ℃;
4-2
Installing Your Analyzer
z The specified temperature range is necessary to obtain reliable analysis
results.
Relative humidity: 30% -85%;
Atmospheric pressure: 70 kPa -106 kPa;
The environment should be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference;
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close;
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
z Do not place the analyzer in a flammable or explosive environment.
4-3
Installing Your Analyzer
4.3 Unpacking
4.3.1 Unpacking and Inspecting the Analyzer
Your analyzer is tested before it is shipped from the factory. International symbols and special
handling instructions tell the carrier how to treat this electronic instrument. When you receive
your analyzer, carefully inspect the carton. If you see any signs of mishandling or damage,
contact Mindray customer service department or your local distributor immediately. When you
are sure the carton is fine, follow the steps below to unpack the analyzer:
1. Place the carton on the floor upright with the arrows on the side upwards;
2. Remove the tape and take out the accessory box. Check the accessories against the
packing list. Notify the Mindray customer service department or your local distributor
immediately if you find anything missing;
3. Open the main box and check the items inside against the packing list. Notify the Mindray
customer service department or your local distributor immediately if you find anything
missing;
4. Remove the top protective foam. Firmly grip the two cardboard handles and lift the
analyzer out of the box and place it on the floor. Carry the analyzer away from the foam
and set it on the countertop.
z Be sure to retain the shipping carton and all the packing materials, as they
can be used for packaging if analyzer must be reshipped.
4.3.2 How to move the analyzer
If the carton is intact, you may use a plate and fork-lifter to move the analyzer for a short
distance;
If your analyzer has been used for a while, do the”Empty Tubing” procedure and shut it
down before moving it;
4-4
Installing Your Analyzer
z Never move the analyzer without draining the fluidic lines.
For a short - distance moving on a smooth ground, you may use a trolley to facilitate the
transportation;
During the moving process, be sure to protect the LCD and the sample probe from
excessive force and from contact with other objects;
Be sure to keep the analyzer upright during the moving process. Do not tilt or incline it;
Do your best to minimize the mechanical shock when moving the analyzer. After a
long-distance moving, check and tune the analyzer before using it.
4-5
Installing Your Analyzer
4.4 Installation Procedure
4.4.1 Releasing sample probe
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
Before the analyzer is shipped out, the sample probe is fixed by a plastic cable tie to avoid
damage during shipment. After unpacking the analyzers, you need to release the sample
probe as follows:
1. Push the right door latch in the direction indicated in Figure 4-1 to open the right door;
Figure 4-1
4-6
Installing Your Analyzer
2. Lift up the front panel latch as indicated in Figure 4-2 and open the front panel;
Figure 4-2
3. Cut the plastic cable tie to release the probe, as Figure 4-3 shows;
Figure 4-3
4-7
Installing Your Analyzer
4. The released sample probe is shown in Figure 4-4;
Figure 4-4
5. Lift the front panel latch and close the front panel and then release the latch to lock it.
Finally close the right door.
4.4.2 Connecting Reagent Containers
z Consider all materials (samples, reagents, controls, calibrators, or
components that contain or have contacted human blood) as being potentially infectious. Wear proper laboratory attire (including rubber gloves, a laboratory coat, and eye protection) and follow safe laboratory procedures when handling any material in the laboratory.
z If the reagents accidentally spill on your skin, wash them off with plenty of
water and if necessary, go see a doctor; if the reagents accidentally spill into your eyes, wash them off with plenty of water and immediately go see a doctor.
4-8
Installing Your Analyzer
z Be sure to use the manufacturer-specified reagents. z Be sure to keep the reagents still for a while before using them. z Be sure not to use expired reagents. z To prevent contamination, be sure to tighten the container caps when the
installation is finished.
Connecting the lyse container
1. Push the left door latch in the direction indicated in Figure 4-5 to open the left door;
Figure 4-5
2. Locate the black and orange fittings as shown in Figure 4-6;
Figure 4-6
4-9
Installing Your Analyzer
3. Take out the lyse pickup tube with an orange connector (see Figure4-7) from the
accessory box. Take out the lyse container, in which there should be enough lyse.
Remove the container cap and insert the double-pronged end of the tube into the
container and turn (clockwise) the tube’s cap onto the lyse container until properly
secured;
Figure 4-7
4. Place the lyse container onto the shelf and connect the black connector on the cap to the
black fitting and the orange connector to the orange fitting, as Figure 4-8 shows.
Figure4-8
4-10
Installing Your Analyzer
Connecting diluent container
1. Take out the diluent pickup tube with a green connector (Figure 4-9) from the accessory
box;
2. Take out the diluent container, in which there should be enough diluent, and place it on or
below the countertop;
3. Remove the container cap and insert the double-pronged end of the tube into the diluent
container and turn the tube’s cap onto the diluent container until properly secured;
Figure4-9
4. Locate the green fitting, marked DILUENT, in the lower right corner of the back of the
analyzer. Plug the green connector of the tube into the fitting and turn it clockwise until
properly secured;
5. Locate the transducer fitting, marked DILUENT, in the lower right corner of the back of
the analyzer. Connect the wire by pushing in and turning until properly secured.
Connecting rinse container
1. Take out the diluent pickup tube with a blue connector (see Figure 4-10) from the
accessory box;
2. Take out the rinse container, in which there should be enough diluent, and place it on or
below the countertop;
3. Remove the container cap and insert the double-pronged end of the tube into the rinse
container and turn clockwise the tube’s cap onto the diluent container until properly
secured;
4-11
Installing Your Analyzer
Figure 4-10
4. Locate the green fitting, marked RINSE, in the lower right corner of the back of the
analyzer. Plug the blue connector of the tube into the fitting and turn it clockwise until
properly secured;
5. Locate the transducer fitting, marked RINSE, in the lower right corner of the back of the
analyzer. Connect the wire by pushing in and turning until properly secured.
Connect waste container
1. Take out the waste tube with a red connector from the accessory box;
2. Locate the red fitting, marked WASTE, in the lower right corner of the back of the
analyzer. Plug the red connector of the tube into the fitting and turn it clockwise until
properly secured;
3. Prepare a container to receive the waste and place it on or below the countertop;
4. Insert the waste tube into the waste container.
4-12
Installing Your Analyzer
4.4.3 Installing Recorder Paper
Follow the procedure below to install the recorder paper.
z Improper installation of recorder paper may jam the paper and/or result in
blank printouts.
1. Locate the projecting part in the upper right corner of the recorder and press it in the
direction shown in Figure 4-11 to open it;
Figure 4-11
2. Flip the paper tension lever on the left side upwards. Keep the printing side face-down.
Insert the pointed end of the paper into the slot below the paper rod and push the paper
until it comes out from above the rod. Pull the paper out. Keep the paper centered and
place the paper into the paper holder. See Figure 4-12;
4-13
Installing Your Analyzer
Figure 4-12
z The recorder paper is treated on one side for printing. To determine which
side is the printing side, gently scratch both sides with nails and the one with visible nail trace left is the printing side.
3. Flip the paper tension level downwards to lock the paper in place, as Figure 4-13 shows.
Figure 4-13
4-14
Installing Your Analyzer
4. Close the recorder door, as Figure 4-14 shows;
Figure 4-14
4.4.4 Connecting the Keyboard
Take out the keyboard from the accessory kit and connect it to the keyboard interface marked
“KB” at the back of the analyzer.
4.4.5 Connecting the Printer (Optional)
Follow the printer’s instructions for use to connect the printer to the parallel port at the back of
the analyzer.
4.4.6 Connecting the Bar-Code Scanner (Optional)
Follow the scanner’s instructions for use to connect the scanner to the serial port1 at the back
of the analyzer.
z Be sure to use the printer and/or scanner of the specified model.
4-15
Installing Your Analyzer
4.5 Starting the Analyzer
Take out the power cord from the accessory box. Plug the non-pronged end into the AC input
at the back of the analyzer and the pronged end into an electrical outlet. Place the power
switch at the back of the analyzer in the ON position (1) to turn on the analyzer. The power
indicator light will be illuminated and the screen will display “Initializing…“. The analyzer will
sequentially initialize the file, hardware and fluidic systems and the whole initializing process
lasts about 3 - 4 minutes. When the initialization is finished, the analyzer will automatically
enter the “Count” screen.
4-16
5 Customizing the Analyzer Software
5.1 Introduction
The BC-3000 Plus is a flexible laboratory instrument that can be tailed to your work
environment. You can use the “Setup”program to customize the software options as
introduced in chapters 5.2 - 5.13.
5-1
Customizing the Analyzer Software
5.2 Print
The “Print” screen is where you set printing options.
5.2.1 Entering the “Print” screen
Press [MENU] to enter the system menu.
Figure 5-1System menu
SELECT “Setup Print” (Figure 5-1) to enter the “Print” screen(Figure 5-2.
Figure 5-2 “Print” screen
5-2
Customizing the Analyzer Software
5.2.2 Selecting printing device
You can select either the built-in recorder or an external printer (if available) as the printing
device, as Figure 5-3 shows.
Figure 5-3 Selecting printing device
 If you prefer the recorder, SELECT “Recorder” from the Device pull-down list.  If you prefer the printer, SELECT “Printer” from the Device pull-down list.
5.2.3 Selecting printing format
If you have selected the printer, you can choose either of the following printing formats.
One page with histogram;
One page without histogram.
To choose the desired format, SELECT the desired format from the Print Format
pull-down list, as Figure 5-4 shows.
5-3
Customizing the Analyzer Software
Figure 5-4 Selecting printing format for the printer
If you have selected the recorder, you can choose any of the following 4 printing formats.
Format1 - parameter values + histograms;
Format2 – parameter values only;
Format3 - parameter values + histograms;
Format4 - parameter values only.
To choose the desired format, SELECT the desired format from the Print Format
pull-down list, as Figure 5-5 shows.
Figure 5-5 Selecting printing format for the recorder
5-4
Customizing the Analyzer Software
5.2.4 Enabling/disabling Auto print
If the “Auto Print ” function is on, the analysis result will be automatically printed out once the analysis is finished. To enable (or disable) this function, SELECT “ON”(or “OFF”) from the Auto Print pull-down list, as Figure 5-6 shows.
Figure 5-6 Enabling/disabling auto print
5.2.5 Exiting the “Print” screen
Press [MENU] to exit to the system menu or [MAIN] to exit to the “Count” screen, and the
changes will be saved automatically.
5-5
Customizing the Analyzer Software
5.3 Count time
The “Count Time” screen is where you view and/or set (if you have the administrator
password) the reference time for the WBC and/or RBC count portion of the measurement
cycle. If the actual WBC or RBC count time (see Chapter 3.4.1 and 3.5.1) deviates from the
reference time by 2 seconds or more, the analyzer will alarm you for clogging or bubbles and
invalidate the results of all related parameters.
5.3.1 Entering the “Count Time” screen and viewing the settings
Press [MENU] to enter the system menu.
Figure 5-7 System menu
SELECT “Setup Count Time” (Figure 5-7)to enter the “Count Time” screen (Figure 5-8).
5-6
Customizing the Analyzer Software
Figure 5-8 “Count Time” screen
5.3.2 Setting Count Time
1. Enter the administrator password as instructed by Chapter 5.4.1.
2. Enter the “Count Time” screen.
3. ENTER the desired number into the ”WBC Count Time” box or ”RBC Count Time” box
to set the reference WBC or RBC count time.
5.3.3 Exiting Count Timescreen
Press [MENU] to exit to the system menu or [MAIN] to exit to the “Count” screen, and the
changes will be saved automatically.
5-7
Customizing the Analyzer Software
5.4 Password
The BC-3000 Plus classifies users into two categories: common users (default) and
administrators. You need to enter the administrator password to adjust certain options such
as WBC/RBC Count Time, Gain, etc.
5.4.1 Entering the administrator password
Press [MENU] to enter the system menu.
Figure 5-9 System menu
SELECTSetup Password” (Figure 5-9) to enter the ”Passwordscreen (Figure 5-10).
Figure 5-10 “Password” screen
5-8
Customizing the Analyzer Software
ENTER “3000” and a message box will pop up to remind you of the current password level,
as Figure 5-11 shows.
Figure 5-11 A message box to confirm the user level
CLICK “Enter to confirm the password and exit to the system menu.
5.4.2 Resuming the common user password
Enter the “Password” screen and the default password is the common user password.
Press [MENU] again and a message box will pop up to remind you of the current password
level, as Figure 5-12 shows.
Figure 5-12 A message box to confirm the user level
CLICK “Enter to confirm the password and exit to the system menu.
5.4.3 Ref. Range
The ”Ref. Range” screen is where you view and/or set (if you have the administrator
password) the high and low limits for your patients. The analyzer flags any parameter value
above (H) or below (L) these limits.
This analyzer divides patients into 5 demographic groups, as Table 5-1shows.
5-9
Customizing the Analyzer Software
Table 5-1 Demographic groups
Group Gender Age
Not specified, male or female. Not specified. General
Not specified.
Man Male
Woman Female
Child Male or Female
Neonate Male or Female
12 years
12 years
12 years
28 days and 12 years
28 days
5.4.4 Viewing the limits (e.g. “General”)
Press [MENU] to enter the system menu.
Figure 5-13 System menu
SELECT SetupRef. RangeGeneral”(Figure 5-13) to view the limits.
5-10
Customizing the Analyzer Software
Figure 5-14 General screen
z Manufacturer-recommended ranges of all 19 parameters are available for
the General, Man, Woman and Child groups. As for the Neonate group, manufacturer-recommended ranges are available for WBC, Lymph#, RBC, HGB and PL T only and the rest, if deemed neces sary, can be set by the users themselves.
5.4.5 Setting the limits (e.g. the “General” group)
1. Enter the administrator password as introduced in Chapter 5.4.1;
2. Enter the ”General” screen, as Figure 5-15 shows;
3. ENTER the new limits as desired;
5-11
Customizing the Analyzer Software
Figure 5-15 “General”screen
5.4.6 Exiting the “Ref.Range” screen (e.g. the “General” group)
Press [MENU] to exit to the system menu; press [MAIN] to return to the “Count” screen. If
you have made any changes, a message box will pop up to ask you to save the changes, as
Figure 5-16 shows. CLICK “Enter” to save the changes and exit to the system menu or the main screen; CLICK “Cancel” to abort the changes and exit to the system menu or the “Count” screen.
Figure 5-16 A message box to confirm the changes
5-12
Customizing the Analyzer Software
z At the “General” screen (or the screen of any other group), you can press
[PRINT] to print out the displayed limits.
z At the “General” screen (or the screen of any other group), you can press
[DEL] to resume the manufacturer-recommended settings.
5-13
Customizing the Analyzer Software
5.5 Transmission
The “Transmission” screen is where you set communication parameters.
5.5.1 Entering the “Transmission” screen
Press [MENU] to enter the system menu.
Figure 5-17 System menu
SELECT “SetupTransmission” (Figure 5-17) to enter the “Transmission” screen (Figure 5-18).
Figure 5-18 “Transmission” screen
5-14
Customizing the Analyzer Software
5.5.2 Selecting baud rate
Five baud rate options are available: “19200”, “9600”(default), “4800”, “2400” and “1200”. To select the desired baud rate, SELECT the desired rate from the Baud Ratepull-down list,
as Figure 5-19 shows.
Figure 5-19 Selecting baud rate
5.5.3 Selecting parity
Three parity options are available: “Odd”, “Even” and “None” (default). To select the desired option, SELECT the desired rate from theParitypull-down list, as Figure 5-20 shows.
5-15
Customizing the Analyzer Software
Figure 5-20 Selecting parity check
5.5.4 Enabling/disabling handshake
If the “Handshake”function is enabled, the analyzer will send a handshake signal to the host
and wait for the response. If the host does not respond in 8 seconds, the analyzer will abort
the transmission and alarm you for a transmission error. If the “Handshake” function is
disabled, the analyzer will transmit the data without confirming the communication link. To
enable (or disable) the “Handshake” function, SELECT ”Yes” (or ”No”) from the Handshakepull-down list, as Figure 5-21 shows. The “Handshake” function is disabled
by default.
5-16
Customizing the Analyzer Software
Figure 5-21 Enabling/disabling handshake
5.5.5 Enabling/disabling auto transmission
When the auto transmission function is enabled, the analyzer will automatically transmit the
analysis result to the host once the analysis is finished. To enable (or disable) the auto
transmission function, SELECT ”ON” (or ”OFF”) from the Auto Trans.” pull-down list, as
Figure 5-22 shows.
Figure 5-22 Enabling/disabling auto transmission
5-17
Customizing the Analyzer Software
5.5.6 Exiting the “Transmission” screen
Press [MENU] to exit to the system menu or [MAIN] to exit to the “Count” screen, and the
changes will be saved automatically.
5-18
Customizing the Analyzer Software
5.6 Setting system time (Date & Time)
The “Date & Time” screen is where you set the system date and time.
5.6.1 Entering “Date & Time” screen
Press [MENU] to enter the system menu.
Figure 5-23 System menu
SELECTSetupDat e & Time” (Figure 5-23) to enter theDate & Time” screen (Figure
5-24).
Figure 5-24 “Date & Time” screen
5-19
Customizing the Analyzer Software
5.6.2 Selecting date format
Three date formats are available: “YYYY-MM-DD”, “MM-DD-YYYY” and “DD-MM-YYYY”. To select the desired format, SELECT the desired format from theFormat”pull-down list, as
Figure 5-25 shows.
Figure 5-25 Selecting date format
5.6.3 Setting system time
Respectively ENTER desired numbers into the ” Year”, “Month”, “Day”, “Hour”, “Minute” and “Second” boxes.
5.6.4 Exiting the “Date & Time” screen
Press [MENU] to exit to the system menu or [MAIN] to exit to the “Count” screen, and the
changes will be saved automatically.
5-20
Customizing the Analyzer Software
5.7 Gain
The ”Gain” screen is where you view and/or set (if you have the administrator password) gains of the “WBC (Whole Blood)”, “WBC (Predilute)”, “RBC” and “HGB” gains.
5.7.1 Entering the “Gain” screen
Press [MENU] to enter the system menu.
Figure 5-26 System menu
SELECT “SetupGain”(Figure 5-26) shows to enter the ”Gain” screen (Figure 5-27).
Figure 5-27 “Gain” screen
5-21
Customizing the Analyzer Software
5.7.2 Setting WBC channel gain
You can adjust the shape of the WBC histogram by adjusting the gain of the WBC channel.
When WBC histograms of most samples are similar to Figure 5-28, it implies too small a
WBC gain and you need to increase the gain appropriately.
Figure 5-28 WBC gain too small
When WBC histograms of most samples are similar to Figure 5-29, it implies too large a
WBC gain and you need to decrease the gain appropriately.
Figure 5-29 WBC gain too large
To increase (or decrease) the gain
1. Enter the administrator password as introduced in Chapter 5.4.1.
2. Enter the ”Gain” screen and ENTER the desired gain into the “WBC (Whole) ”, as Figure 5-30 shows, or “WBC (Predilute)”, as Figure 5-31 shows.
5-22
Customizing the Analyzer Software
Figure 5-30 Setting WBCWholegain
Figure 5-31 Setting WBCPredilutegain
5.7.3 Setting the RBC gain
If the MCV results of most calibration or QC runs deviate from the expected result by 6%, you
need to follow the rule below to change the RBC gain to adjust the MCV results.
Assume the expected MCV result is 90.0fL and the obtained MCV result is 82.0fL.
Then
5-23
Customizing the Analyzer Software
VExpectedMC
%100
ActualMCV
1. Enter the administrator password as introduced in Chapter 5.4.1.
2. Enter the “Gain” screen and ENTER a number into the ”RBC” box, as Figure 5-32 shows, so that RBC “Factor” is as close to 109.8% as possible.
0.90
=××
%8.109%100
0.82
Figure 5-32 Setting RBC gain
5.7.4 Setting HGB channel gain
You can adjust the HGB blank voltage by adjusting the HGB gain. Normally the HGB blank
voltage should be within 3.4 - 4.8 V (4.5V is recommended). To set the HGB channel gain,
1. Enter the administrator password as introduced in Chapter 5.4.1.
2. Enter the “Gain” screen and ENTER the desired gain into the ”HGB” box so that the HGB
voltage falls between 3.4 - 4.8 V, as Figure 5-33 shows.
5-24
Customizing the Analyzer Software
Figure 5-33 Setting HGB gain
5.7.5 Exiting the “Gain” screen
Press [MENU] or [MAIN] to exit the ”Gain” screen and a message box will pop up to ask you
save the changes, as Figure 5-34 shows.
Figure 5-34 A message box to confirm the changes
CLICK “Enter” to save the changes and exit to the system menu; CLICK “Cancel” to exit to the system menu or the “Count” screen without saving the changes.
5-25
Customizing the Analyzer Software
5.8 Auto Clean Time
The “Auto Clean Time” screen is where you set the interval for auto cleaning of the fluidic
lines. The valid interval is 2 - 24 hours and the default interval is 4 hours. To set the interval,
Press [MENU] to enter the system menu.
Figure 5-35 System menu
SELECTSetupAuto Clean Time” (Figure 5-35) to enter the “Auto Clean Time” screen
(Figure 5-36).
Figure 5-36 Auto Clean Time screen
ENTER the desired interval. Press [MENU] to exit to the system menu or [MAIN] to exit to the
Count” screen, and the changes will be saved automatically.
5-26
Customizing the Analyzer Software
5.9 Reagent Exp. Date
The ”Reagent Exp. Date” screen is where you set expiration dates for the diluent, rinse and
lyse. The analyzer will alarm you for expired reagents when the system time exceeds any of
the three expiration dates.
5.9.1 Entering the “Reagent Exp. Date” screen
Press [MENU] to enter the system menu.
Figure 5-37 System menu
SELECTSetup Reagent Exp. Date” (Figure 5-37) to enter the ”Reagent Exp. Date
screen (Figure 5-38).
Figure 5-38 “Reagent Exp. Date” screen
5-27
Customizing the Analyzer Software
5.9.2 Setting the expiration date
ENTER the desired expiration dates into the ” Diluent”, “Rinse” and ” Lyse” boxes.
z For any reagent, the entered expiration date should be either the expiration
date printed on the labeling or the open-container expiration d ate, whichever is earlier.
z The open-container expiration date is calculated as follows: the date that
container is opened + the open-container stability days.
5.9.3 Exiting the “ReagentExp. Date” screen
Press [MENU] or [MAIN] to exit the ” ReagentExp. Date” screen and a message box will pop
up to ask you save the changes, as Figure 5-39 shows.
Figure 5-39 A message box to confirm the changes
CLICKEnter” to save the changes and exit to the system menu or the “Count” screen;  CLICKCancel” to abort the changes and exit to the system menu or the “Count”
screen.
5-28
Customizing the Analyzer Software
5.10 Report Title external keyboard required
The “Report Title” screen is where you set the title of the report to be printed. To set the
report title,
Press [MENU] to enter the system menu.
SELECT Setup
5-41).
Figure 5-40 System menu
Report Title”(Figure 5-40) to enter the ”Report Title” screen (Figure
Figure 5-41 Report Title screen
5-29
Customizing the Analyzer Software
ENTER the desired title in the ”Report Title (by recorder) ” or ” Report Title (by printer)
box, depending on the printing device you choose to print out the report. Press [MENU] or
[MAIN] to save the changes and exit to the system menu or the system menu or the “Count” screen.
z To correct any erroneous entry, DELETE the wrong character.
5-30
Customizing the Analyzer Software
5.11 Parameter Units
The “Parameter Units” screen is where you view and/or set (if you have the administrator
password) the reporting units of the parameters.
See Table 5-2 for the available units for every parameter groups. Note that if you choose g/L
or g/dL for the HGB/MCHC group, the MCH unit will automatically change to pg and its
reporting format will be ***.*; if you choose mmol/L for the HGB/MCHC group, the MCH unit
will automatically change to fmol and its reporting format will be **.**.
Table 5-2 Reporting unit
Parameter group Reporting format Reporting unit Remarks
WBC
Lymph#
Mid#
***.*
***.*
****
9
10
/L Default
3
10
/uL
2
10
/uL
Gran#
Lymph%
Mid%
Gran%
HGB
MCHC
RBC
HCT
MCV
RDW-SD
***.*
**.*
***
**.*
**.*
*.**
*.**
***
*.**
**.*
.***
***.*
***.*
/nL
Default
%
g/L Default
g/dL
mmol/L
12
10
/L Default
6
10
/uL
4
10
/uL
/pg
% Default
L/L
fL Default
3
um
RDW-CV **.*
PLT
****
****
***.*
****
MPV **.*
% Default
10
10
10
/nL
fL Default
5-31
9
/L Default
3
/uL
4
/uL
Customizing the Analyzer Software
**.*
PDW **.*
PCT
.***
*.**
3
um
None Default
% Default
mL/L
5.11.1 Entering the “Parameter Units” screen and viewing the
settings
Press [MENU] to enter the system menu.
Figure 5-42 System menu
SELECTSetup Parameter Units” (Figure 5-42), to enter the “Parameter Units” screen
(Figure 5-43 ).
Figure 5-43 Parameter Units screen
5-32
Loading...