Warnings, Cautions and Notes ........................................................................................................................vii
Notes ............................................................................................................................................................ ix
Theory of Operation .....................................................................................................1 - 1
Front View .......................................................................................................................................... 1 - 3
Side View........................................................................................................................................... 1 - 5
The Ventilator Unit ............................................................................................................................... 1 - 9
Ventilator Control and Drive....................................................................................................................... 1 - 12
BDU (Basic Digital Unit)...................................................................................................................... 1 - 12
Power Supply.................................................................................................................................... 1 - 28
Anesthesia System Components .................................................................................................................. 1 - 29
Gas Box Assembly............................................................................................................................. 1 - 36
Tube color coding.............................................................................................................................. 1 - 36
The Breathing System ................................................................................................................................ 1 - 37
Breathing System Parts List.......................................................................................................... 4 - 11
Electric Box....................................................................................................................................... 4 - 12
Electric Box Parts List.................................................................................................................. 4 - 12
Gas Circuit Box................................................................................................................................. 4 - 13
Gas Circuit Box Parts List............................................................................................................ 4 - 13
User Interface.................................................................................................................................... 4 - 14
User Interface Parts List............................................................................................................... 4 - 15
General Guidelines..................................................................................................................................... 5 - 4
Test Equipment and Special Tools Required.................................................................................................... 5 - 4
This Service Manual is intended as a guide for technically qualified personnel performing
repair and calibration procedures.
Warnings, Cautions and Notes
Please read and adhere to all warnings, cautions and notes listed here and in the
appropriate areas throughout this manual.
A WARNING is provided to alert the user to potential serious outcomes (death, injury, or
serious adverse events) to the patient or the user.
A CAUTION is provided to alert the user to use special care necessary for the safe and
effective use of the device. They may include actions to be taken to avoid effects on patients
or users that may not be potentially life threatening or result in serious injury, but about which
the user should be aware. Cautions are also provided to alert the user to adverse effects on
this device of use or misuse and the care necessary to avoid such effects.
A NOTE is provided when additional general information is applicable.
Warnings
WARNING: Whenever using anesthetic gases, nitrous oxide, oxygen, or
WARNING: For continued protection against fire hazard, replace all
WARNING: In order to prevent an electric shock, the machine
WARNING: Remove all accessory equipment from the shelf before
WARNING: Possible explosion hazard. Do not operate machine near
WARNING: The use of anti-static or electrically conductive respiration
any hospital gas always follow the appropriate agent
evacuation/collection procedures. Use the hospital gas
evacuation system.
fuses with the specified type and rating.
(protection class I) may only be connected to a correctly
grounded mains connection (socket outlet with grounding
contact).
moving the anesthesia machine over bumps or on any
inclined surface. Heavy top loading can cause the machine
to tip over causing injury.
flammable anesthetic agents or other flammable
substances. Do not use flammable anesthetic agents (i.e.,
ether or cyclopropane.)
tubes, when utilizing high frequency electric surgery
equipment, may cause burns and is therefore not
recommended in any application of this machine.
WARNING: Possible electric shock hazard. The machine may only be
opened by authorized service personnel.
AS3000™ Service Manual0070-10-0683vii
IntroductionCautions
WARNING: Compressed gasses are considered Dangerous Goods/
WARNING: Avoid exposure to respiratory gases by always directing
Hazardous Materials per I.A.T.A. and D.O.T. regulations. It is
a violation of federal and international law to offer any
package or over pack of dangerous goods for
transportation without the package being appropriately
identified, packed, marked, classified, labeled and
documented according to D.O.T. and I.A.T.A. regulations.
Please refer to the applicable I.A.T.A. Dangerous Goods
Regulations and /or the Code of Federal Regulations 49
(Transportation, Parts 171-180) for further information.
the fresh gas flow from the fresh gas outlet to the waste gas
scavenger.
Cautions
CAUTION:This device uses high pressure compressed gas. When
attaching or disconnecting backup gas cylinders, always
turn the cylinder valves slowly. Use the AS3000 flow meters
to bleed down the pressure, watching the cylinder gauge
indicate the depleting cylinder pressure, before
disconnecting the cylinder from the yoke. Always open and
close cylinder valves fully.
CAUTION:This device operates using compressed gas at high
pressures from the hospital central supply. When connecting
gas supply lines attach the hose connection to the machine
before connecting the quick disconnect fitting to the hospital
source. Disconnect the supply hose from the hospital source
connection prior to disconnecting it from the AS3000 gas
connection fittings.
CAUTION:Refer to the “Periodic Maintenance Schedule of Service
Activities” on page 6-2, in the Periodic Maintenance section
for assistance when performing scheduled periodic
maintenance.
CAUTION:Do not leave gas cylinder valves open if the pipeline supply
is in use and the system master switch is turned to 'ON'. If
used simultaneously, cylinder supplies could be depleted,
leaving an insufficient reserve supply in the event of
pipeline failure.
CAUTION:Use cleaning agent sparingly. Excess fluid could enter the
machine, causing damage.
CAUTION:This machine must only be operated by trained, skilled
medical staff.
CAUTION:Perform the electrical safety inspection as the last step after
completing a repair or after routine maintenance. Perform
this inspection with all covers, panels, and screws installed.
CAUTION:After changing the CO
CAUTION:Only Selectatec
CAUTION:After each exchange of a vaporizer, carry out a system Leak
test.
™
System may be used with the AS3000 unit.
test.
absorbent, carry out a system leak
2
compatible vaporizers with Interlock-
viii0070-10-0683AS3000™ Service Manual
NotesIntroduction
CAUTION:The bellows dome cannot be autoclaved.
CAUTION:Do not clean the machine while it is on and/or plugged in.
CAUTION:Pressing “Quit” at any time during the procedure will cancel
the session's settings and reload the previously-stored
calibration coefficients.
CAUTION:Depleted soda lime changes color. Replace the soda lime if
approximately 2/3 of the absorber content is discolored.
absorbent can be safely changed without stopping
CO
2
mechanical ventilation.
Notes
NOTE:Unauthorized servicing may void the remainder of the
warranty. Check with the factory or with a local authorized
distributor to determine the warranty status of a particular
instrument.
AS3000™ Service Manual0070-10-0683ix
IntroductionNotes
This page intentionally left blank.
x0070-10-0683AS3000™ Service Manual
1.0
Theory of Operation
1.1Introduction
The AS3000 is a continuous flow anesthesia system which offers manual or automatic
ventilation, easily adjustable fresh gas delivery, anesthetic agent delivery, ventilation
monitoring, convenient ergonomics, and state-of-the-art safety systems. The components of the
AS3000 Anesthesia System are described this chapter.
AS3000™ Service Manual0070-10-06831 - 1
Microprocessor-controlled VentilatorTheory of Operation
Gas Inlet
Block
N20
AIR
O2
Breathing
System
Ventilator
Pressure
Switch
Pressure
Regulator
Pressure
Regulator
Pressure
Regulator
Ratio•
Controller
N20•
Cuttoff
Flow
Tubes
Cylinder
Cylinder
Cylinder
Inlet Filter
Inlet Filter
Inlet Filter
O2
Flow
Meter
O2
Outlet
AIR•
Flow
Meter
Patient
Vaporizer
O2
Flush
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APLValve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp.Valve
Absorber
Canister
Exp.Valve
Breathing Bag
OOO
P
P
P
P
P
P
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Fresh Gas
Flow Sensor
V
Bellows
Vaporizer
1.2Microprocessor-controlled Ventilator
The Microprocessor-controlled ventilator, with its dedicated Breathing System, allows timecontrolled, pressure limited, constant volume ventilation for all patient groups within a tidal
volume range of 40 mL (4 kg infant) to 1400 mL (large adult).
Time-controlled, pressure limited, and compliance compensated constant volume ventilation
is provided through the Controlled Mandatory Ventilation (CMV) mode. The CMV mode
delivers a viable ventilation method for complicated lung conditions. The ventilator also
provides time-controlled, volume dependant ventilation, targeting a set (adjustable) target
pressure provided through the Pressure Controlled Ventilation (PCV) mode. Automatic and
comprehensive system startup tests and alarm management systems ensure controlled
ventilation conditions in every mode of operation.
The durable and ergonomically designed user interface and Navigator
™
Knob enables easy
operation. The display provides the selected ventilation modes (CMV, PCV, PS and SIMV)
and the following values: Tidal Volume, Peak Pressure, Mean Pressure, FiO
, Breath Rate, I:E
2
Ratio, PEEP, Plateau Pressure, Alarm limits, and real-time Airway Pressure and flow
waveform.
1 - 20070-10-0683AS3000™ Service Manual
FIGURE 1-1 The AS3000 Pneumatic System
Theory of OperationComponents
4
5
6
7
8
9
2
1
10
11
3
1.3Components
1.3.1Front View
FIGURE 1-2 AS3000 Front View
1. Monitor Arm
The Monitor Arm provides support for a bedside monitor. It can easily rotate for more
convenient viewing.
2. User Interface
The display of the user Interface provides waveforms, numeric data and menu tabs. The keys
and Navigator Knob enable the user to power up the system, silence alarms, access menu
AS3000™ Service Manual0070-10-06831 - 3
tabs, and switch between manual and mechanic ventilation.
ComponentsTheory of Operation
3. UI Arm
The UI Arm provides support for the user interface assembly. It can easily rotate for more
convenient viewing.
4. Oxygen Sensor
The Oxygen Sensor monitors the oxygen concentration of the inspired gas of the Breathing
System.
5. Breathing System
The Breathing System’s main function is to store anesthetic gas, oxygen, and air; vent exhaust
gas; and absorb carbon dioxide. It connects directly to the respiratory passage to help
complete the breathing process.
6. CGO (Common Gas Outlet) Subassembly
Mixed gas composed of O
, AIR, N2O, and anesthetic agent connects to the patient's
2
Breathing System via a flexible tube from the CGO Subassembly.
7. O2 Flush Valve
The O2 Flush Valve is located on the front of the AS3000. The supplied gas does not pass
the flowmeter and vaporizer. It is directly sent to the fresh gas outlet. Press this button to
supply gas (35 - 50 L/min). Release this switch to automatically close the gas supply.
8. AGSS (Anesthetic Gas Scavenging System)
The AGSS (Anesthetic Gas Scavenging System) reclaims exhausted gas generated during
anesthesia.
9. Drawer Subassembly
The AS3000 has three drawers for storage, which can be locked and fixed through the
uppermost drawer lock.
10. Vaporizer Mounting Manifold
The Vaporizer Mounting Manifold provides support for up to two Selectatec
®
compatible
anesthetic vaporizers.
11. Flowmeter
The Flowmeter displays gas flow values for N2O, O2, AIR, Auxiliary O2, and Auxiliary AIR.
It consists of coarse and fine flow tubes used to accurately measure gas flow with additional
flow tubes to measure auxiliary O
The Upper Mainframe Subassembly provides support for the pressure gauges, the Flowmeter,
the gas box assembly, the mains supply assembly, and the YOKE assembly.
13. User Interface Cable
The User Interface Cable provides signal transmission between the User Interface and the
main unit.
AS3000™ Service Manual0070-10-06831 - 5
ComponentsTheory of Operation
14. Auxiliary Gas Outlet Assembly
The Auxiliary Gas Outlet Assembly provides the patient with auxiliary oxygen and air mixtures
(of different concentrations). It also contains an O
connection for use with other equipment.
2
15. Breathing System Interface
The Breathing System Interface provides a connection interface for gas signal acquisition
between the Breathing System and gas circuit.
16. Breathing System Pneumatic Hose
The Breathing System Pneumatic Hose provides multiple connections including: PEEP control,
Auto/Manual control, and four pressure sampling connections between the Breathing System
and mainframe.
17. O
Sensor Cable
2
The O2 Sensor Cable provides a connection between the oxygen sensor component in the Breathing
System and the oxygen concentration signal acquisition port on outlet module of the mainframe.
18. Breathing System Interface
The Breathing System Interface provides the main unit with an interface for connection with
drive gas, heating system and the oxygen concentration sensor of the Breathing System.
19. The Breathing System Support Arm
The Breathing System Support Arm provides connection between the Breathing System and
the mainframe, in order to support the Breathing System.
20. Heater Wire
The Heater Wire provides connection to the heating system of the Breathing System. By
heating the Breathing System during anesthesia, accumulation of water in the Breathing
System is minimized. It also provides comfortable gas to the patient.
21. AGSS Transfer Hose
The AGSS Transfer Hose provides pipeline connection between the exhaust gas outlet of the
Breathing System and the AGSS evacuation system.
22. Lower Mainframe Assembly
The Lower Mainframe Assembly provides support for the drawer subassembly, outlet module,
CGO subassembly, O
flush valve and the electric box. Together with the Upper Mainframe
2
Subassembly and Base Assembly, forms the mainframe of the machine.
23. Base Assembly
The Base Assembly provides support to the whole machine. The four casters provide
movement for the machine in any direction. The front casters have a locking function.
1 - 60070-10-0683AS3000™ Service Manual
Theory of OperationComponents
24
25
26
27
28
29
30
31
32
33
1.3.3Rear View
FIGURE 1-4 AS3000 Rear View
24. Top Shelf Assembly
The Top shelf Assembly includes the work light and its switch. It also serves as a storage
area.
25. Gas Input Assembly
The Gas Input Assembly allows gas in the pipeline and backup gas cylinder to enter the
AS3000 for regulation. The Pressure regulator reduces pressure to 36 psi (250 kPa). The
O
operated valve can open allowing N2O to enter the flowmeter only when O2 pressure is
2
above 7.3 psi (50 kPa). Otherwise, N2O cannot enter the gas circuit. The pressure switch
generates a signal when the input pressure is below 29 psi (200 kPa), and the system will
provide an audible alarm.
AS3000™ Service Manual0070-10-06831 - 7
ComponentsTheory of Operation
26. Vaporizer Storage Mount
The Vaporizer Storage Mount provides the AS3000 with auxiliary placement for a
vaporizer.
27. Mains Supply Assembly
The Mains Supply Assembly attaches the AS3000 to an external AC wall outlet. It also
supplies power throughout the unit.
28. Electronic Flowmeter Assembly
The Electronic Flowmeter Assembly provides fresh gas flow measurement for the AS3000.
29. Gas Box Assembly
The Gas Box Assembly provides drive gas to the Breathing System and includes a safety
valve to insure that pressure stays below 85 cmH
O. It also contains the PEEP valve, and the
2
differential pressure sensors.
30. YOKE Assembly
The YOKE Assembly provides connection for three backup gas cylinders. The pressure
regulator for the backup gas cylinder reduces the pressure of a high pressure gas cylinder
down to 43.5 psi (300 kPa).
31. AMP Cables
The AMP Cables provide signal transmission between the electric box and gas circuit box.
32. Electric Box Assembly
The Electric Box Assembly provides, and distributes power to the AS3000.
33. Rear Panel Assembly
The Rear Panel Assembly protects, and provides heat dissipation for the electric box and
auxiliary support for the backup gas cylinder.
1.3.4Other components (not identified in graphics)
34. Inspiratory and Expiratory Valves
The Inspiratory and Expiratory Valves are unidirectional valves that allow air to flow in only
one direction.
35. Absorber Canister
The AS3000 uses two Absorber Canisters which can contain 1500 mL of soda lime each.
They can be used for 6 - 8 hours each if full. Water generated from the reaction with CO
drained from a valve on the lower side of the canister.
36. APL (Airway Pressure Limiting) Valve
The APL Valve is used for limiting maximum airway pressure during manual ventilation. The
adjustable range of the APL is 0 - 70 cmH
O.
2
is
2
1 - 80070-10-0683AS3000™ Service Manual
Theory of OperationComponents
37. Bellows Assembly
The AS3000 employs ascending ventilation bellows, in which mixed gas is stored. Drive
gas supplied by the ventilator forces the bellows to descend sending mixed gas into the
inspiratory passage of patient's airway. If a patient's airway suffers from gas leakage, the
bellows will collapse, informing the operator of a possible problem. A tidal volume scale is
provided on the transparent dome, through which a patient's tidal volume can be estimated.
38. Pressure-relief Valve
The Pressure-relief Valve is located at the base of the bellows. When end-expiration airway
pressure reaches 1 - 3 cmH
expelled.
39. Exhaust Gas Outlet
The Exhaust Gas Outlet is located on the lower part of the Breathing System. It is connected
to the AGSS or via the AGSS transfer tube.
40. Absorber Heating System
The Breathing System is heated to body temperature to avoid humidified gases condensing
within the Breathing System thus improving airway climatization for the patient's re-breathing
of respiratory gases.
O, the pressure-relief valve opens and redundant gas is
2
1.3.5Breathing System
The Breathing System is integrated into a compact aluminum block. This block is heated to
body temperature to prevent condensing of humidified gases within the Breathing System,
thus improving airway climatization for the patient's re-breathing of respiratory gases. The
heated Breathing System contains: an inspiratory valve with O
measurement, expiratory valve, APL Valve, breathing bag connection, and internal
inspiratory and expiratory flow sensors.
1.3.6The Ventilator Unit
The AS3000 ventilator offers multiple ventilation modes: Controlled Mandatory Ventilation
with volume control (CMV), Pressure Control Ventilation (PCV), Synchronized Intermittent
Mandatory Ventilation (SIMV), and Pressure Support (PS) ventilation. Electronic PEEP is
available in all ventilation modes. User control over inspiratory flow (SLOPE) is possible in
PCV, SIMV, and PS modes. Automatic fresh gas compensation limits the effect of user
changes in fresh gas flow rate on the patient. The traditional bellows system is driven by
oxygen and makes patient disconnections clearly visible.
adapter for FiO2
2
AS3000™ Service Manual0070-10-06831 - 9
ComponentsTheory of Operation
1.3.7Adjustable Alarms
Minimum and maximum alarm limits can be set for Peak Pressure, Mean Pressure and FiO2.
Minimum alarm limits can be set for Tidal Volume and Minute Volume. Exceeding the peak
pressure alarm limit automatically halts the inspiratory phase preventing airway pressure
from exceeding the high alarm setting. In the CMV mode, when reaching this pressure limit,
a “High Airway Pressure” alarm is displayed, and inspiration is discontinued. The next
inspiration occurs at the regular time interval, preventing increase of the respiratory rate. The
result is a decreased tidal volume (T Vol.) and minute volume (M Vol.). During pressure
limitation, the ventilator displays the alarm message until the condition is corrected.
1.3.8Compliance Compensation
Compliance compensation automatically corrects for the expansion of the circuit in CMV
ventilation mode. System compliance is measured by the ventilator to maintain the set tidal
volume (±15%). The compliance test may be bypassed at machine power up. When
bypassing the compliance test, the default settings are used.
1 - 100070-10-0683AS3000™ Service Manual
Theory of OperationElectrical Supply
Ventilator UI
Ventilator Control and Drive
Anesthesia
10.4” TFT-LCD
PC104
Backlight Inverter
Amplifier for
audio alarm
Communication
and Interface
Keypad
Knob
BDU
Amplifier
PWR MNG
DC/DC
BAT
PS
AC/DC
SW
Work L ight
FM Back Light
AC Heater
Ctrl &
Protect
Aux Mains
Outlet
Inlet
Filter
Power Supply
Pwr
DA
Cable
Switch
Switch
Internal Signals
Val. Exp. PEEP
Flow Insp.
FM Flow Sensor
Cal. Valve
Manual/Mech Val.
P. V a l. In sp.
Gas Source Pressure
FiO2 Sensor
Airway Pressure
Probe Flow
1.4Electrical Supply
1.4.1Electrical components
FIGURE 1-5 Electrical Components Overview
AS3000™ Service Manual0070-10-06831 - 11
Ventilator Control and DriveTheory of Operation
1.5Ventilator Control and Drive
1.5.1BDU (Basic Digital Unit)
The BDU serves as the active ventilator control. The BDU controls the actions of the PEEP,
calibration, and proportional valves, and reads the signal flow, and pressure sensors and
valves
FIGURE 1-6 BDU Control Board, Top View
1 - 120070-10-0683AS3000™ Service Manual
Theory of OperationVentilator Control and Drive
FIGURE 1-7 BDU Control Board, Bottom View
AS3000™ Service Manual0070-10-06831 - 13
Ventilator Control and DriveTheory of Operation
1.5.1.1Amplifier Board
FIGURE 1-8 Amplifier Board, Top View
FIGURE 1-9 Amplifier Board, Bottom View
1 - 140070-10-0683AS3000™ Service Manual
Theory of OperationVentilator Control and Drive
O2 Sensor Input, S4
PIN NAME FUNCTION
1. O
2. O
3. O
4. O
-O
2
-O
2
+O
2
+O
2
Sensor Input -
2
Sensor Input -
2
Sensor Input +
2
Sensor Input +
2
DC Power Input, J7
PIN NAME FUNCTION
1. GND Power Ground
2. +5V Controller Logic Power
3. P12V Power 12V
4. PWBUS Power Bus 24V
5. GND Analog Ground
6. A12V Analog 12V
7. 5V/4.8V Power 5V/4.8V
Power Control, J20
PIN NAME FUNCTION
1. GND Ground
2. /PWON Power On Enable, Low active
3.PWEN Power On Enable, High active
4. 5VAUX 5V Auxiliary
5. DC_M AC/DC Monitor signal
6. PWBUS Power bus
7. BAT_M Battery Monitor signal
8. APE Auxiliary Power Enable
9. CHAR_SIGNAL Charge Status, No connection
10. /MUTE Mute, No connection
Signals and Power to Keyboard, J5
PIN NAME FUNCTION
1. ENPW Power Enable, High active
2. /PWON /Power on Enable, Low active
3. 5VAUX 5V Auxiliary
4. APE Auxiliary Power Enable
5. GND Ground
6. TTLRX TTL Receive
7. 232ARX RS232A Receive
8. 232BRX RS232B Receive
9. GND Ground
AS3000™ Service Manual0070-10-06831 - 15
Ventilator Control and DriveTheory of Operation
Signals and Power to Keyboard, J5 (Continued)
PIN NAME FUNCTION
10. PWBUS PowerBus
11.PWBUS PowerBus
12. PWBUS PowerBus
13. GND Ground
14. GND Ground
15. DC24V DC24V monitor
16. BATV+ Battery monitor
17. /MUTE No connection
18. GND Ground
19. TTLTX TTL Transmit
20. 232ATX RS232A Transmit
21. 232BTX RS232B Transmit
22. GND Ground
23. PWBUS PowerBus
24. PWBUS PowerBus
25. GND Ground
Signals and Power to Sensor Board, J30
PIN NAME FUNCTION
1. A+12V Analogue +12V
2. FL_INS Import flow
3. FL_EXP Export flow
4. FM_FL Flowmeter flow
5. Pair Pressure of air
6. Paw Pressure inside air way
7. Pgas Pressure gas supply status
8. GND Ground
9. AGND Analog Ground
10. AGND Analog Ground
11. P24V Power 24V
12. P24V Power 24V
13. P24V Power 24V
14.P12V Power 12V
15. P12V Power 12
16. P5V Power 5V
17. P5V Power 5V
18. Inhale Inspire Valve
19. Man Manual/AUTO Valve
20. Exhale Expire Valve
21. Exhale Expire Valve
22. AGND Analog Ground
1 - 160070-10-0683AS3000™ Service Manual
Theory of OperationVentilator Control and Drive
Signals and Power to Sensor Board, J30 (Continued)
PIN NAME FUNCTION
23. AGND Analog Ground
24. AGND Analog Ground
25. AGND Analog Ground
Bus to BDU, J96
PIN NAME FUNCTION
1 - 3 5V 5V Power
4 - 6 GND Ground
Test Point Definition
DESIGNATORNAMEFUNCTIONRANGE
T1 FIO
T2 Pgas Pressure switch of gas supply,
T3 FL_INSP Inspire Flow signal0.20-0.30V (0 flow)
T4 Pair Absolute Pressure of air way1.7-2.1V (atmospheric
T5 FL_EXP Expire flow signal0.20-0.30V (0 flow)
T6 Fm_fl Flow of flowmeter0.25-0.51V (flowmeter off)
T7AD-ExhalePEEP valve, current feedbackLevel depends on PEEP
FIGURE 1-14 Breathing System Heater Board, Top View
FIGURE 1-15 Breathing System Heater Board, Bottom View
J2
PIN NAME FUNCTION
2. GND Power Ground
3. 5V Controller Power, Output Enable
4. 24V Heater Power Supply
5. NC No Connection
AS3000™ Service Manual0070-10-06831 - 21
Ventilator Control and DriveTheory of Operation
J3
PIN NAME FUNCTION
1. RET Heater Power Ground
2. DQ Sensor signal
3. 24V Heater Output
Test point definition
DESIGNATOR NAME FUNCTION
T10 GND Power Ground
T1 5V Controller Power
T7 24V Heater Power input
T9 POUT Heater Output
Indicating lamp definition
DESIGNATOR STATUS FUNCTION
LT2 RED Heater Output available
LT3 GREEN Heater Ready
1 - 220070-10-0683AS3000™ Service Manual
Theory of OperationVentilator Control and Drive
1.5.1.5Sensor Board
FIGURE 1-16 Sensor Board, Top View
FIGURE 1-17 Sensor Board, Bottom View
AS3000™ Service Manual0070-10-06831 - 23
Ventilator Control and DriveTheory of Operation
Sensor Board
Pair, J1
PIN NAME FUNCTION
1. AGND Analog Ground
2. Pair Pressure of air
3. NC No connection
4. AR10V 10V
5. NC No connection
PAW, J2
PIN NAME FUNCTION
1. AGND Analog Ground
2. Paw Pressure inside airway
3. AR5V 5V
4. NC No connection
5. NC No connection
Work Light Power, J11
PIN NAME FUNCTION
1. 24V Power 24V
2. GND Power Ground
Backlight Power, J15
PIN NAME FUNCTION
1. 24V Power 24V
2. GND Power Ground
Switch of Gas Supply, J6
PIN NAMEFUNCTION
1. GND Ground
2.NC No connection
3. Pgas Switch input
Expire Valve, J14
PIN NAME FUNCTION
1. 5V 5V
2. Valexp Ex p ire Valv e Drive
1 - 240070-10-0683AS3000™ Service Manual
Theory of OperationVentilator Control and Drive
Inspire Valve, J12
PIN NAME FUNCTION
1 - 2 24V 24V
3 - 4 Valexp inspire Valve Drive
Man/Auto Valve, J14
PIN NAME FUNCTION
1. 24V 24V
2. Valman Manual/Auto Valve Drive
Signals and Power of Sensor Board, J13
PIN NAME FUNCTION
1. A+12V Analogue +12V
2. FL_INS Import flow
3. FL_EXP Export flow
4. FM_FL Flowmeter flow
5. Pair Pressure of air
6. Paw Pressure inside air way
7. Pgas Pressure gas supply status
8. GND Ground
9. AGND Analog Ground
10. AGND Analog Ground
11. P24V Power 24V
12. P24V Power 24V
13. P24V Power 24V
14. P12V Power 12V
15. P12V Power 12
16. P5V Power 5V
17. P5V Power 5V
18. Inhale Inspire Valve
19. Man MANUAL/AUTO Valve
20. Exhale Expire Valve
21. Exhale Expire Valve
22. AGND Analog Ground
23. AGND Analog Ground
24. AGND Analog Ground
25. AGND Analog Ground
AS3000™ Service Manual0070-10-06831 - 25
Ventilator Control and DriveTheory of Operation
1.5.2Power Management
Power management is located behind the Rear Panel Assembly. This module serves as the
voltage supply for the ventilator control and drive BDU, the flowmeter backlight, the work
light, the Breathing System Heater, and the charging/discharging control for the battery.
FIGURE 1-18 Power Board, Top View
FIGURE 1-19 Power Board, Bottom View
1 - 260070-10-0683AS3000™ Service Manual
Theory of OperationVentilator Control and Drive
Power Board
Battery input, J4
PIN NAME FUNCTION
1. BAT Battery Input +
2. 5V Power Ground
3. BATTEST Battery test input
DC24V Power input, J3
PIN NAME FUNCTION
4. GND Power Ground
5. GNDPower Ground
6. 24V Power input
7. 24V Power input
DC Power output, J2
PIN NAME FUNCTION
1. GND Power Ground
2. +5V Controller Logic Power
3. P12V Power 12V
4. PWBUS Power Bus 24V
5. GND Analog Ground
6. A12V Analog 12V
7. 5V/4.8V Power 5V/4.8V
Heater Power, J1
PIN NAME FUNCTION
6. GND Power Ground
7. 5V Controller Power, Output Enable
8. 24V Heater Power Supply
9. PBUS Power Bus, No Connection
Power Control, J2
PIN NAME FUNCTION
1. GND Ground
2. /PWON Power On Enable, Low active
3. PWEN Power On Enable, High active
4. 5VAUX 5V Auxiliary
5. DC_M AC/DC Monitor signal
6. PWBUS Power bus
7. BAT_M Battery Monitor signal
AS3000™ Service Manual0070-10-06831 - 27
Ventilator Control and DriveTheory of Operation
Power Control, J2 (Continued)
PIN NAME FUNCTION
8. APE Auxiliary Power Enable
9. CHAR_SIGNAL Charge Status, No connection
10. /MUTE Mute, No connection
Power Switch, J5
PIN NAME FUNCTION
1. GND Ground
2. NC No Connection
3. PWON Power On Enable, Low active
Test Point definitions
DESIGNATOR NAME FUNCTION
T1 24V 24V DC Input
T2 PWBUS Power Bus
T3 A12V Analogue 12V
T4 P12V Power 12V
T5 +5V Logic Power 5V
T6 5V/4.8V Power 5V/4.8V
T7 BAT+ Battery Input +
T8 30V 30V for charger
T9 PWAUX Power Auxiliary
T10 CHARGE Charge Status
T11 5VAUX 5V Auxiliary
T12 VRs Charger Feedback
Led Designation
DESIGNATOR LED COLORSTATUS FUNCTION
DS1 GREEN 24V DC available
DS2 GREEN Power Bus On
DS3 GREEN Charge
1.5.3Battery
The sealed lead-acid battery is maintenance free, and has a maximum recharge time of 8
hours when fully discharged. The run time is at least 45 minutes with a fully charged battery.
To prevent unintended loss of battery operation, the recommended replacement period is
every 3 years.
1.5.4Power Supply
The power supply provides power to the machine and relevant controls.
1 - 280070-10-0683AS3000™ Service Manual
Theory of OperationAnesthesia System Components
1.6Anesthesia System Components
1.6.1Auxiliary Outlets
The AS3000 has four auxiliary outlets (120 VAC, 60 Hz, 2A maximum each). There are
two 2A fuses for each outlet.
The AS3000 has 10 fuses on the back panel of the unit. There are two 2A fuses for each of
the four outlets and two 10A fuses for the line cord.
AS3000™ Service Manual0070-10-06831 - 35
Ventilator Pneumatic - O2 Drive GasTheory of Operation
1.8Ventilator Pneumatic - O2 Drive Gas
1.8.1Ventilator pneumatic drive
Oxygen is the driving gas for the ventilator. In addition to the flowmeter block, a high
pressure regulator reduces the supply pressure to 25.4 psi (175 kPa). This pressure
represents the drive gas for the ventilator.
The drive pressure regulator is placed ahead of the proportional valve that generates the
driving gas flow during the inspiratory phase. This flow fills the bellows dome that surrounds
the bellows.
1.8.2Drive Pressure-High pressure regulator
The drive pressure regulator stabilizes the supply pressure provided to the proportional valve.
The flow generated by the proportional valve is therefore independent of pressure variations
at the supply.
Setting the drive pressure regulator at 25.4 psi (175 kPa) allows for a maximum inspiratory
flow of 70 L/min at the ventilator.
1.8.3Gas Box Assembly
The driving module consists of the proportional valve and a solenoid valve for the vent mode
switch. The proportional valve pressure regulator, and generates a driving gas flow of 0 - 70
L/min in relation to the control voltage of the proportional valve of 0 - 5VDC.
The control voltage of the proportional valve, required for the pre-selected parameter
settings, is generated by the BDU board. The driving gas flow Q
relationship with the tidal volume:
Q
with: Q
V
T
T = time
= VT/T
drive gas
= driving gas flow
drive gas
= generated tidal volume
1.8.4Tube color coding
All the pneumatic tubes used in the AS3000 are color coded for use in the United States
only.
GASUS STANDARD
O
2
OBlue
N
2
AIRYellow
Green
is in the following
drive gas
1 - 360070-10-0683AS3000™ Service Manual
Theory of OperationThe Breathing System
Breathing
System
Ventilator
Patient
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Bellows
Breathing
System
Ventilator
Patient
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Bellows
1.9The Breathing System
1.9.1CMV mode, inspiration
Tidal volume (T Vol.) compensates for variations in gas flow. This is to ensure that the set tidal
volume is delivered to the patient.
FIGURE 1-28 Breathing System Pneumatics, CMV Mode, Inspiration
1.9.2CMV mode, expiration
As the patient exhales tidal volume into the expiratory limb, fresh gas enters the bellows.
Fresh gas mixes with exhaled gas after the Absorber removes CO
through the exhaust valve to the AGSS
. Excess fresh gas passes
2
AS3000™ Service Manual0070-10-06831 - 37
FIGURE 1-29 Breathing System Pneumatics, CMV Mode, Expiration
The Breathing SystemTheory of Operation
Breathing
System
Ventilator
Patient
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Bellows
2 cmH2O
Breathing
System
Ventilator
Patient
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Bellows
2 cm H2O
1.9.3Manual mode, inspiration
As the breathing bag is compressed, the gas is directed to the patient. Pressures exceeding
the set value of the APL Valve will pass through the APL Valve the AGSS.
FIGURE 1-30 Breathing System Pneumatics, Manual Mode, Inspiration
1.9.4Manual mode, expiration
As the patient exhales tidal volume into the expiratory limb, fresh gas enters the Breathing
System. Fresh gas mixes with exhaled gas after the Absorber removes CO2. Excess fresh gas
passes through the exhaust valve to the AGSS
FIGURE 1-31 Breathing System Pneumatics, Manual Mode, Expiration
1 - 380070-10-0683AS3000™ Service Manual
Theory of OperationThe Breathing System
1.9.5Pneumatic PEEP
The PEEP valve regulates the pressure at which the exhaust valve opens, therefore the exhaust
valve opens only when the pressure exceeds the set PEEP pressure.
1.9.6Ventilator in Standby
When the AS3000 is in the standby mode, monitoring will be inactive, The patient should
not be ventilated when the system is in standby mode.
1.9.7Breathing System Components
1.9.7.1Ventilation Bellows System
The ventilator's driving system is a flow generator. Driving gas fills the bellows dome to
compress the bellows. The breathing gas is pressed out of the bellows into the patient
breathing circuit. The bellows is refilled with fresh gas and the expired gas from the patient.
1.9.7.2Manual Breathing Bag
In manual mode, this device acts as a normal breathing bag, enabling the user to ventilate
the patient manually. In mechanical ventilation mode, this bag is cut off from the breathing
circuit by the reversing valve.
1.9.7.3CO2 Absorber
The soda lime inside the absorber retains the carbon dioxide from the exhaled gas. The
AS3000 accommodates standard sized Pre-Paks or loose-fill CO2 absorbent.
1.9.7.4Inspiratory and Expiratory Valves
To ensure correct gas flow direction to and from the patient, one-way-valves are integrated in
the inspiratory and expiratory limb of the Breathing System.
1.9.7.5APL (Airway Pressure Limiting) valve
In manual mode, the APL Valve acts as a normal spring loaded pressure relief valve, limiting
the maximum pressure in the Breathing System.
AS3000™ Service Manual0070-10-06831 - 39
The Breathing SystemTheory of Operation
This page intentionally left blank.
1 - 400070-10-0683AS3000™ Service Manual
2.0
Installation Guide
2.1Delivery of The New Anesthesia Machine
The following customer supplied material must be present prior to installation. Missing
equipment can result in delays, incomplete installations and/or extra visits.
• Compatible emergency O2, N2O, and AIR cylinders
• Agent vaporizers and key fillers, if not purchased with the AS3000
• Liquid agent medication
•CO
absorbent Pre-Paks or loose fill
2
•Active O2, N2O, and AIR, lines at 50 psi
• Drop down hoses for ceiling mounted medical gas utilities, compatible with quick-
disconnect hoses if not purchased with the AS3000
• Activated medical gases (O
, N2O, AIR, VAC, and EVAC)
2
AS3000™ Service Manual0070-10-06832 - 1
AssemblyInstallation Guide
1
2
4
5
3
12
11
9810
7
6
2.2Assembly
NOTE:The AS3000 Breathing System Block is matched to the
AS3000 it is attached to by calibration and installation.
After removing the Breathing System Block from the
Mounting Arm of the AS3000, assure that the Breathing
System Block is returned to the same AS3000 that it came
from originally.
2.2.1Unpacking
1. Inspect the two boxes for any damage, check the tip watches, and report any damage
or sign of tipping to the dispatcher.
2. Ensure the serial numbers on both carton labels match prior to assembling the system.
3. Remove the shipping straps and caps from the 2 boxes.
4. Remove the foam from inside the top of the box.
5. Remove the Corrugated box from around AS3000.
6. Remove the bag and shrink wrap from around AS3000.
7. Remove the foam from AS3000 casters.
2.2.2Breathing System and Breathing System Accessories
FIGURE 2-1 The AS3000 and accessories
1. AGSS2. Breathing System
3. User Interface Mounting Arm4. User Interface
5. Patient Monitor Mounting Arm6. O
Cable
2
7. Pneumatic Hose Assembly8. Heater Cable
9. Drive Gas10. Breathing System Mounting Arm
11. AGSS Transfer Hose12. CGO Port
2 - 20070-10-0683AS3000™ Service Manual
Installation GuideAssembly
2.2.2.1Attaching the Mounting Arms and User Interface
(see FIGURE 2-1)
1. Remove the mounting arms from small box and install them onto the AS3000 using a 3
mm allen wrench.
2. Remove the keys from the parts box, and unlock the AS3000 drawers and put parts
box into top drawer.
3. Remove the AGSS and Breathing System mount from the foam packaging, and install
them onto the AS3000.
4. Remove the User Interface and slide it onto the User Interface mounting arm.
5. Apply Loctite 243 to the User Interface cable jackscrews and jackposts. Then, connect
the User Interface cable to the User Interface. Ensure the jackscrews are securely
tightened.
2.2.2.2Breathing System Connections
(see FIGURE 2-1)
1. Remove the foam and bag from around the Breathing System.
2. Ensure that the serial number tag attached to the bag arm matches the serial number
found on the package carton.
3. Install the Breathing System onto the arm on the left side of the AS3000 by attaching it
to the installed bracket.
4. Attach the Drive Gas Hose to the bottom port of the Breathing System
5. Attach the Heater cable into the port located on the bottom of the Breathing System
6. Install the O
7. Attach the O2 cell to the O2 cable and insert the cell into its designated port on the
Breathing System.
8. Attach the Pneumatic Hose Assembly to the designated port on the rear of the Breathing
System.
9. Attach the AGSS Transfer Hose to its designated port on the bottom of the Breathing
System.
Cell to the O2 Interface
2
2.2.2.3Anesthesia System Connections
(see FIGURE 2-1)
1. Attach the Fresh Gas Hose of the Breathing System to the CGO port on the system.
AS3000™ Service Manual0070-10-06832 - 3
AssemblyInstallation Guide
2. Attach the Drive Gas Hose to the designated port on the Breathing System Interface.
3. Attach the Heater Cable to the designated port on the Breathing System Interface.
4. Attach the O
5. Attach the Pneumatic Hose Assembly to the designated connector labeled Breathing
System Pneumatics.
6. Attach the AGSS Transfer Hose.
Cable to the designated port on the Breathing System Interface.
2
2.2.3Tank Wrench and Pre-operation Checklist
1. Mount the tank wrench on the rear of the AS3000 so that it can be used to open or
close each cylinder, without disconnecting it from the machine.
2. Attach the pre-operation checklist to a location on the AS3000 where the operator can
access it.
2.2.4Patient Suction Regulator and Arm
1. Remove the Patient Suction Regulator and mounting arm from its packaging.
2. Attach the mounting arm to the customers desired channel. It mounts to either side
channel.
3. Tighten the locking screw.
4. Thread the patient suction regulator into its mating port on the mounting arm.
5. Attach a barbed fitting adapter to the suction regulator's VAC fitting as necessary to
attach a hospital grade suction hose from the regulator output to the suction canister.
2.2.5Suction Canister Bracket
1. Remove the suction canister bracket from its packaging.
2. Attach the bracket to the right lower rail and secure.
2.2.6Utility Tray, Monitor Mounting Arm with Utility Hook(s)
1. Remove the items from their packing material.
2. Slide the utility tray into the desired channel.
3. Tighten the locking screw.
4. Remove the plastic cap from under the mounting arm's swivel head.
5. Mount the utility hook(s) under the swivel head of the arm using the screws and tool
provided with the hook.
6. Slide the mounting arm into the desired channel. Allow enough room for mounting
monitors at the desired height.
7. Tighten the locking screw.
2.2.7Vaporizers
1. Prior to mounting, set the vaporizer to the T position. (Does not apply for all vaporizers.)
2 - 40070-10-0683AS3000™ Service Manual
Installation GuideAssembly
2. Discard the washers that came with the vaporizer. Use only approved O-rings that come
mounted on the vaporizer mounting ports of the AS3000.
3. Mount a mechanical vaporizer to either side of the AS3000 vaporizer mount by
placing the vaporizer on the ports and locking down the lever.
4. Mount an electronic vaporizer in the same manner. Follow any installation instructions
that come with the specific vaporizer.
2.2.8High Pressure Hoses
1. Remove the hoses from their packing material.
2. Attach each hose to its mating connector by hand. Do not use any tools.
3. Use extreme care while attaching the EVAC hose to the waste gas scavenger (WGS); as
it is extremely delicate.
4. Attach an EVAC hose to a VAC source connection only if an EVAC source connection is
not available.
2.2.9Emergency Cylinder(s)
1. Remove the cover from a new O2, N2O, and AIR cylinder.
2. Mount one at a time onto the rear of the anesthesia machine.
3. Discard the cylinder's tank washer. Always use the approved tank washer provided with
the AS3000.
4. Open the bail of each yoke and mount the cylinder over the tank washer.
5. Ensure the O
yoke. Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke
O
2
port.
6. Open and close the cylinder valve and observe that the cylinder gauge on the
anesthesia machine rises to the colored range.
7. Ensure that the N2O cylinder mates to the N2O PISS connection on the N2O yoke.
Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke port.
8. Open and close the cylinder valve and observe that the cylinder gauge on the
anesthesia machine rises to the colored range.
9. Ensure that the AIR cylinder mates to the AIR PISS connection on the AIR yoke. Close the
yoke bail and use the hand-screw to tighten the cylinder to the yoke port.
10. Open and close the cylinder valve and observe the cylinder gauge on the anesthesia
machine rise to the colored range.
cylinder mates to the O2 Pin Index Safety System (PISS) connection on the
2
2.2.10Breathing Circuit, CO2 Absorbent, and Liquid Vaporizer Agent
1. Attach a breathing circuit to the inspiratory and expiratory ports as detailed in the
directions for use. Attach the breathing bag and any other respiratory accessories as
described.
2. Insert the Pre-Paks one on top of the other with the wider side facing up on both Pre-
Paks. (alternatively the absorber canisters may be filled with loose fill absorbent.)
3. Install the absorber canister with a quarter turn of the lever at the bottom the absorber
assembly, this ensures a tight seal.
AS3000™ Service Manual0070-10-06832 - 5
AssemblyInstallation Guide
2.2.11Monitoring Products - Mounting and Electrical Connection
1. Any monitoring system compatible with the GCX mounting system's swivel head may be
mounted to the AS3000’s arm.
NOTE:Use of other monitors and mounting hardware is the
responsibility of the installer.
2. Always make full use of all mounting fasteners and strap capturing devices when
mounting monitors to the AS3000.
3. After mounting a monitor to the AS3000, connect it to one of the AC outlets located on
the rear of the AS3000.
a. Turn on each monitor one at a time and ensure that the circuit breaker holds without
tripping.
b. Dress each line cord neatly along the side of the anesthesia machine so that it can
not be easily pulled or extend far from the main chassis.
2.2.12Agent Monitor Waste Gas Scavenging
1. Respiratory gas monitoring products have an exhaust port from which waste gas expels.
The exhaust port on the gas monitor must be connected to the open barbed fitting on the
waste gas scavenger.
2. Ensure that a tight connecting fitting is attached to the rear of the gas monitor. Ensure
that the other end of the same tube has a tight fitting connection attached to the waste
gas scavenger’s (WGS) barbed fitting.
3. Dress the exhaust tubing neatly along the side of the anesthesia machine so that it can
not be easily pulled or extend far from the main chassis.
2.2.13Oxygen Sensor Calibration
NOTE:See ‘‘Periodic Maintenance Schedule of Service Activities’’
on page 6-2 for when to calibrate the oxygen sensor.
1. Preparing the unit
a. Allow the breathing system to warm up and reach thermal equilibrium
(approximately 30-60 minutes).
b. In Standby mode, press the MENU button. The menu screen will appear.
c. Select Service, then input the password 2010 to enter the Service screen.
d. Select Calibration to enter the Calibration screen.
e. Select Oxygen Sensor to enter the Oxygen Sensor Calibration screen.
2. Calibration
NOTE:Do not shake the O2 sensor during calibration.
NOTE:During calibration, keep the O
connector side up, and bottom side exposed to room air;
keep the O
temperature difference from within the heated block.
sensor near the heated block to minimize the
2
sensor in a vertical position,
2
2 - 60070-10-0683AS3000™ Service Manual
Installation GuideAssembly
NOTE:If the system is going to be used during the calibration,
insert the O
sensor was removed using a push and turn motion.
cell plug into the port from which the oxygen
2
a. Select 21% to enter the 21% oxygen concentration calibration screen.
b. Remove the oxygen sensor from the Breathing System and expose it to room air for
at least 3 minutes.
c. Flush the O2 sensor with air from the auxiliary output for 5-10 seconds to ensure that
bubbles are trapped in the sensor.
no O
2
d. Select Next to start 21% oxygen concentration calibration.
NOTE:The O2 sensor voltage is displayed during the calibration.
This is the amplified O
the oxygen sensor. The O
for UI versions 2.24 and lower.
cell voltage at the A/D converter for
2
sensor voltage is not displayed
2
e. When calibration is successfully completed, install the oxygen sensor into the
Breathing System.
NOTE:The oxygen sensor must be installed in the Breathing
System for 10 minutes prior to 100% oxygen concentration
calibration to adjust to the temperature of the system.
f. Press the O2 Flush button for 5-10 seconds to clear out any non-O2 gases from the
system.
g. Expose the oxygen sensor to 100% pure oxygen (5 L/min) for at least 3 minutes.
h. Select 100% to enter the 100% oxygen concentration calibration screen.
i. Select Next to start 100% oxygen concentration calibration.
AS3000™ Service Manual0070-10-06832 - 7
Installation Checkout ProcedureInstallation Guide
2.3Installation Checkout Procedure
Complete each step to verify the functionality of the AS3000 prior to clinical use.
Also, perform this checkout after installation, reinstallation, servicing or after any periodic
maintenance activity. This checklist does not replace periodic maintenance actions that must
be performed to maintain peak performance.
1. Verify that all components are present, and inspect for physical damage.
a. Ensure that the Operating Instructions is present.
b. Verify unit is free from cosmetic defects. Plastics, labeling and display window are
free from inclusions, pitting, bulges, sink marks, nicks, scratches, gouges dents,
discolorations, etc. Verify all major hardware is fastened properly.
c. Verify that there are Tank Washers for each of the external cylinder yokes.
d. Verify that each vaporizer's locking spring is intact on the manifold by looking
through the opening and checking for the visible wire spring.
e. Verify that each connector of the vaporizer mount has an O-ring.
f. Verify that breathing circuit consumables and CO
g. Inspect the O
h. Turn the flow control knobs for O2, AIR and N2O and verify that the floats rise and
spin as the flow is increased and fall freely as the flow is decreased.
i. inspect the AC line cord for fraying and damages.
, N2O, AIR, VAC, and EVAC supply hoses for damages.
2
absorbent are present.
2
2. Verify that the laminated Preoperative Checkout card is attached.
3. Verify that the tank wrench is attached.
4. Verify that the waste scavenger flow rate is set and that the hose is
intact, connected and dry.
a. Attach the waste gas hose from the waste gas port on the Breathing System to one of
the waste gas ports on the AGSS.
b. Attach the EVAC hose from the waste gas assembly DISS (Diameter Index Safety
System) fitting to a source of vacuum.
c. Adjust the position of the float to be between the Min and Max lines by turning its
NOTE:If a specific gas source is not available, skip the
corresponding leak test.
, AIR and N2O hoses to the gas line inlets.
2
c. Veri f y O2, AIR and N2O Line Pressure gauges read the corresponding line pressure.
d. Set the AIR flow controller to minimum.
e. Pinch to occlude the O2, AIR and N2O hoses.
f. Disconnect each hose from its line pressure source.
g. Verify that the pressures do not drop more than 2 psi over a time period of 20
seconds.
6. Verify that all available line and cylinder gauges are operational.
a. Connect the O
supply hose and remove the AIR and N2O hoses.
2
b. Verify the O2 Line Pressure gauge shows the reading of the line pressure.
c. Connect the AIR supply hose and remove the O
and N2O hoses.
2
d. Verify that the AIR Line Pressure gauges show the reading of the line pressure.
e. Connect the O2 and N2O hoses and remove the AIR hose.
f. Verify that the N
O and O2 Line Pressure gauge shows the reading of the line
2
pressure.
g. Attach the cylinders to the AS3000's yokes (O2 and N2O or AIR). In the case
where multiple cylinders of the same gas are required, attach one at a time.
h. Open all attached cylinders including the O2 cylinder. If a cylinders pressure is less
than 500 psi, replace the respective cylinder.
i. Verify that the cylinder pressure gauges register a pressure for each type of gas
attached.
7. Test O
a. Connect the O
Flush operation with only O2 connected.
2
supply hose and remove the AIR and N2O hoses.
2
b. Attach a calibrated flow meter to the Common Gas Outlet.
c. Push the O2 Flush button and verify a reading between 35 L/min and 50 L/min at
the flow meter.
NOTE:O2 flow may be greater than 50 L/min at higher altitudes.
d. Reconnect the O2, AIR and N2O hoses.
8. Test the O2:N2O ratio system (applicable to -01 units only).
a. Set all flow control knobs to minimum.
b. Increase the flow of N
c. Verify that the N
O until the O2 reaches 1 L/min
2
O flow is not higher than 3.7 L/min.
2
d. Return flow control knobs to minimum.
AS3000™ Service Manual0070-10-06832 - 9
Installation Checkout ProcedureInstallation Guide
9. Test the O2:N2O ratio system (applicable to -02 units only).
a. Set all flow control knobs to minimum.
b. Set the flow of O
c. Open the N
to 1 L/min.
2
O flow knob and verify that the N2O flow will not increase higher than
2
3.7 L/min.
d. Return flow control knobs to minimum.
10. Perform System Test.
a. Power ON the AS3000.
b. Wait until the System Self Test is complete
FIGURE 2-2 Startup Self Test
c. Select Continue to prompt to the Safety Valve test
2 - 100070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
11. Perform Leak/Safety Valve Test
a. Follow the on-screen instructions.
FIGURE 2-3 Leak Test Setup
b. Select Continue to start the test.
FIGURE 2-4 Safety Valve Test Setup
AS3000™ Service Manual0070-10-06832 - 11
Installation Checkout ProcedureInstallation Guide
FIGURE 2-5 Safety Valve Test in Progress
FIGURE 2-6 Safety Valve Test Passed Message
c. After 3 seconds, the screen will prompt to the Leak Test.
12. Perform Leak Test
2 - 120070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
a. Follow the on-screen instructions.
b. Select Continue to start the test.
FIGURE 2-7 Leak Test Setup
AS3000™ Service Manual0070-10-06832 - 13
Installation Checkout ProcedureInstallation Guide
FIGURE 2-8 Leak Test in Progress
c. Select Continue to prompt to the Compliance Test.
FIGURE 2-9 Leak Test Results
2 - 140070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
13. Perform Compliance Valve Test
a. Follow the on-screen instructions.
b. Select Continue to start the test.
FIGURE 2-10 Compliance Test Setup
FIGURE 2-11 Compliance Test in Progress
AS3000™ Service Manual0070-10-06832 - 15
Installation Checkout ProcedureInstallation Guide
c. Select Continue to prompt to the Normal Screen.
FIGURE 2-12 Compliance Test Results
14. Manual Leak Test
NOTE:The Manual Leak Test detects smaller leaks than can be
detected in the Automatic Leak Test.
a. Ensure that the gas pressure for O2, N2O, and AIR are at 50 ±10 psi.
b. Power ON the AS3000.
c. Attach a reusable-silicone-rubber breathing circuit to the Breathing System.
NOTE:For testing purposes always use a reusable breathing
circuit.
d. Tightly connect the Y-fitting on the breathing circuit to the test port.
e. Attach a breathing bag to the bag arm.
f. Set the APL Valve to the fully closed position (70 cmH
g. Rotate the O
Flow Control Valve until 50 cmH2O pressure is observed on the
2
2
O).
Airway Pressure Gauge.
h. Verify that the flow required to stabilize the pressure is less than 300 mL/min.
2 - 160070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
15. Oxygen Sensor Calibration
NOTE:Oxygen Sensor Calibration can be performed in all
NOTE:See ‘‘Periodic Maintenance Schedule of Service Activities’’
ventilation modes.
on page 6-2 for when to calibrate the oxygen sensor.
a. Allow the breathing system to warm up and reach thermal equilibrium
(approximately 30-60 minutes).
b. Press the MENU key and then use the Navigator™ Knob to scroll to the
Calibrate menu tab (see FIGURE 2-13). Select the Start Calibration button.
FIGURE 2-13 Calibrate Menu Tab
c. After the Start Calibration button has been selected, the screen shown in
FIGURE 2-14 or FIGURE 2-16will be displayed, instructing the user to remove the
oxygen sensor from the Breathing System and expose it to room air for at least 3
minutes before proceeding. The O
calibration. This is the amplified O
sensor voltage is displayed during the
2
cell voltage at the A/D converter for the oxygen
2
sensor.
NOTE:The O2 sensor voltage is not displayed for UI versions 2.24
and lower.
d. Flush the O2 sensor with air from the auxiliary output for 5-10 seconds to ensure that
bubbles are trapped in the sensor.
no O
2
NOTE:Do not shake the O2 sensor during calibration.
NOTE:Keep the O
during calibration.
NOTE:Place the O
calibration to minimize the temperature difference from
within the heated block.
AS3000™ Service Manual0070-10-06832 - 17
sensor in a vertical position, connector side up,
2
sensor on top of the heated block during
2
Installation Checkout ProcedureInstallation Guide
NOTE:If the system is going to be used during the calibration,
insert the O
sensor was removed using a push and turn motion.
cell plug into the port from which the oxygen
2
e. After at least 3 minutes have passed, select the Next button to initiate the
calibration process. The progress bar shown in FIGURE 2-15 or FIGURE 2-17will be
displayed.
FIGURE 2-14 Oxygen Sensor
Calibration Instructions
FIGURE 2-16 Oxygen Sensor
Calibration Instructions (UI
versions 2.24 and lower)
FIGURE 2-15 Oxygen Sensor Calibration
Progress Bar
FIGURE 2-17 Oxygen Sensor Calibration
Progress Bar (UI versions
2.24 and lower)
2 - 180070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
16. Proceed based on one of the following two conditions:
• If the calibration is successful, the screen shown in FIGURE 2-18 will be
displayed, instructing the user to reinstall the oxygen sensor into the Breathing System.
Select the Done button to complete the process.
• If the calibration fails, the screen shown in FIGURE 2-19 will be displayed,
instructing the user to either repeat the calibration (by selecting the Repeat Cal
button) or to replace the oxygen sensor. If the oxygen sensor must be replaced, select
the Exit button, replace the oxygen sensor using a push and turn motion, and then
repeat the calibration.
FIGURE 2-18 Oxygen Sensor
Calibration Successful
FIGURE 2-19 Oxygen Sensor
Calibration Failed
17. Verify that the APNEA alarm activates in Manual mode
a. Press the MANUAL/AUTO key.
b. After 60 seconds, verify that:
• an alarm tone sounds
• the alarm message APNEA is displayed in red text
18. Verify that CMV ADULT ventilator mode operates and that the tidal
volume display reads within 15% of the CMV set value.
a. Attach a breathing circuit and breathing bag.
NOTE:For testing purposes always use a reusable breathing
circuit.
b. Attach an adult test lung to the Y-fitting of the breathing circuit.
c. Attach a Vent Tester between the EXP port and the expiratory hose.
d. Set the O
flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
2
AS3000™ Service Manual0070-10-06832 - 19
Installation Checkout ProcedureInstallation Guide
e. Set the ventilator controls to:
VENTILATOR CONTROLSVENTILATOR SETTINGS
Patient TypeAdult
Ventilation ModeCMV
Tidal Volume - V
Breath Rate - freq8
I:E Ratio - I:E1:2
Plateau - T
PEEP - PEEP Off
T
P
600
10
f. Select CMV again to begin ventilation.
g. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp.
rate and minute volume values appear on the screen.
h. Verify the Tidal Volume display is within 15% of the delivered volume measured with
the Vent Tester within approximately 1 minute from the start of ventilation.
i. Verify the delivered volume at the Vent Tester is within 15% of the Tidal Volume set
value within approximately 1 minute from the start of ventilation.
j. Verify the measured O
concentration is at least 97% after 5 minutes.
2
k. Set the AIR flow to 3 L/min and set the N2O and O2 flow rates to minimum flow.
l. Verify the measured O2 concentration is 21% ±3% vol. % after 5 minutes.
19. Verify that CMV CHILD ventilator mode operates and that the tidal
volume display reads within 10% of the CMV set value.
a. Attach a breathing circuit and breathing bag.
NOTE:For testing purposes always use a reusable breathing
circuit.
b. Attach an adult test lung to the Y-fitting of the breathing circuit.
NOTE:Limit the volume in the test lung to provide sufficient airway
pressure to satisfy the Low Peak Pressure alarm. Or reduce
the Peak Pressure alarm limit to a lower valve to prevent
the alarm when using an adult test lung.
2 - 200070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
c. Attach a Vent Tester between the EXP port and the expiratory hose.
d. Set the O
flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
2
e. Set the ventilator controls to:
VENTILATOR CONTROLSVENTILATOR SETTINGS
Patient TypeChild
Ventilation ModeCMV
Tidal Volume - V
Breath Rate - freq20
I:E Ratio - I:E1:2
Plateau - T
PEEP - PEEP Off
T
P
120
10
f. Select CMV to begin ventilation.
g. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp.
rate and minute volume values appear on the screen.
h. Verify the Tidal Volume display is within 25 ml or ±15% of the delivered volume
(whichever is greater) measured within approximately 1 minute from the start of
ventilation.
i. Verify the delivered volume as measured by a Vent Tester at the expiratory port, is
within 17% of the Tidal Volume set value within approximately 1 minute from the
start of ventilation.
j. Verify that the PEEP reading is 4 ±2 within approximately 1 minute from the start of
ventilation.
20. Verify that the PCV ADULT ventilator mode operates.
a. Attach a breathing circuit and breathing bag.
NOTE:For testing purposes always use a reusable breathing
circuit.
b. Attach an adult test lung to the Y-fitting of the breathing circuit.
c. Attach a Vent Tester between the EXP port and the expiratory hose.
d. Set the O
flow to 3 L/min and set the N2O and AIR flow rates to minimum flow.
2
AS3000™ Service Manual0070-10-06832 - 21
Installation Checkout ProcedureInstallation Guide
e. Set the ventilator controls to:
VENTILATOR CONTROLSVENTILATOR SETTINGS
Patient TypeAdult
Ventilation ModePCV
Target Pressure - P
Breath Rate - freq8
I:E Ratio - I:E1:2
PEEP - PEEP Off
Inspiratory Slope - T
TARGET
slope
20
0.5
f. Select PCV to begin ventilation.
g. Verify the Peak Pressure reading of the display is ±20%, at least 2 cmH
O of the
2
Peak Pressure measured with the Vent Tester.
h. Verify that the pressure waveform, Tidal Volume, Resp. Rate and minute volume
values appear on the screen.
i. Verify that the PEAK Value reaches 20 ±4 cmH2O within five breaths from the start
of ventilation.
21. Verify that the Low Airway Pressure alarm activates.
a. While the ventilator is running disconnect the Inspiratory hose.
b. Verify that the “Low Airway Pressure” and the “APNEA” message appear on the
screen.
c. Reconnect the hose to the Inspiratory port and verify that the alarm is cleared.
22. Verify that the O2 concentration alarm signals activate.
a. While the ventilator is running, press the ALARM LIMITS key.
b. Set the O
Minimum Alarm Limit to a value higher than the current O2 reading.
2
c. Verify that the FiO2 alarm activates indicated by an audio tone is present and that
the alarm message “Low FiO
” appears on the screen within three ventilation cycles.
2
d. Set the O2 Minimum Alarm Limit back to a value lower than the current O2 reading.
e. Verify that the FiO
alarm resets.
2
2 - 220070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
23. Verify that the drive gas pressure loss alarm signals activate and N2O
stops and AIR continues to flow.
a. Set the O2 flow for 1 L/min.
b. Set the N
O flow to 1 L/min.
2
c. Set the AIR flow to 1 L/min.
d. Disconnect the O2 line supply to the AS3000, and Close the O2 cylinder.
e. Verify that the flow of N2O stops when O2 is lost.
f. Verify that AIR continues to flow.
g. Ver i fy the “ O
Supply Failure” message appears on screen, a steady audio tone
2
should sound.
h. Reconnect O2 line pressure source to the AS3000.
i. Verify that the alarms are reset.
j. Ver i fy N
O flow has resumed once O2 pressure returns.
2
24. Verify that the battery operation and charging icons appears.
a. Set the vent mode to Standby.
b. Disconnect AC line cord.
c. Verify that the battery icon appears on the screen and indicates the charge level of
the battery.
d. Verify that the line cord icon is flashing.
e. Verify that the flow meter backlight is illuminated during battery operation.
f. When the battery icon is in a not full state, reconnect the AC line cord.
g. Verify that the battery charge icon appears on the screen.
h. Verify that the flow meter backlight is illuminated during AC voltage operation.
25. Verify operation of the work light.
a. Turn on the work light located on the bottom side of the top panel.
b. Verify that it lights in both on positions.
26. Verify that the Auxiliary O
and AIR Flowmeters operates.
2
a. Verify an AIR flow of 15 L/min can be obtained by connecting the auxiliary AIR hose
to the pressure source and opening the flow meter.
b. Ve rif y a n O
flow of 15 L/min can be obtained by connecting the auxiliary O2 hose
2
to the pressure source and opening the flow meter.
27. Verify that the Suction Regulator operates in LINE and REGULATE
positions.
a. Set the suction regulator's selection dial to LINE.
b. Verify maximum suction vacuum on the regulator's gauge.
c. Set the suction regulator's selection dial to REGULATE.
d. Verify the suction vacuum is adjustable on the regulator's gauge.
e. Set the suction regulator's selection dial to OFF.
AS3000™ Service Manual0070-10-06832 - 23
Installation Checkout ProcedureInstallation Guide
28. Verify that the Breathing System heats to body temperature.
a. Operate the AS3000 on AC operation for approximately 40 minutes.
b. Verify the Breathing System has heated up to body temperature
29. Complete electrical safety inspection.
NOTE:Perform the electrical safety inspection as the last step after
completing a repair or after routine maintenance. Perform
this inspection with all covers, panels, and screws installed.
a. Withdraw the Power cable(s) from the convenience receptacles at the rear of the
AS3000.
b. Plug the AS3000 into a Safety Analyzer.
c. Connect the case ground lead of the analyzer to the U-blade ground of one of the
convenience receptacles. Perform the following tests with the case grounded:
• Normal polarity
• Normal polarity with open neutral
d. Perform the following tests with the case ungrounded:
• Normal polarity
• Normal polarity with open neutral
• Reverse polarity
e. Verify that the maximum leakage current does not exceed 300 µA (0.3 mA).
NOTE:Ground Resistance (between the U-blade ground on any
convenience outlet to the U-blade ground on the AC line
cord).
f. Plug the AS3000 into the safety analyzer.
g. Attach the resistance-measuring probe on the analyzer to the AS3000 U-blade
ground on any convenience outlet.
h. Invoke the resistance function on the safety analyzer, following the instructions for the
analyzer.
i. Verify that resistance to ground is less than 0.2 Ohms (200 mOhms).
30. Check the vaporizer interlock.
a. Attach two vaporizers to the Vaporizer Mounting Manifold and lock them in place.
b. Rotate either of the vaporizer’s dial to 3% agent.
c. Verify that the other vaporizer dial cannot be rotated to a setting.
d. Set both vaporizer’s dials to 0.
e. Rotate the other vaporizer dial to 3%.
f. Verify that the first vaporizer dial cannot be rotated.
g. Rotate both vaporizer dials to T and remove both vaporizers.
h. Verify that the locking spring is intact.
i. Reconnect both vaporizers to the Vaporizer Mounting Manifold
2 - 240070-10-0683AS3000™ Service Manual
Installation GuideInstallation Checkout Procedure
31. Check each vaporizer's agent concentration output and accuracy.
a. Insert the agent measuring device sampling tube inside the common gas outlet port.
b. Fill the Vaporizer with anesthetic agent.
NOTE:Do not overfill by filling past the indicator line on the
vaporizer.
c. Test the vaporizer accuracy per the Vapor 2000 instructions or see the appropriate
vaporizer manual for testing details.
d. Test each vaporizer in turn.
e. Test any vaporizer on the Vaporizer Storage Mount.
f. Remove the measuring device from the common gas outlet port.
g. Connect the fresh gas tube to the common gas outlet port.
m. If the value displayed on the monitor is in % partial pressure, no correction is
required. If in vol.% convert to partial pressure:
• Concentration [% partial pressure] = measured value [vol.%] x atmospheric
pressure [cmH
O] / 1013 cmH2O
2
AS3000™ Service Manual0070-10-06832 - 25
Installation Checkout ProcedureInstallation Guide
n. For setting 0 and T there should be no output of anesthetic agent.
• At 1 vol.%; 0.8 - 1.2 vol.% *
• At 2 vol.%; 1.8 - 2.2 vol.% *
• At 3 vol.%; 2.8 - 3.2 vol.% *
• At 4 vol.%; 3.8 - 4.2 vol.% *
• At 5 vol.%; 4.8 - 5.2 vol.% *
• At 6 vol.%; 5.7 - 6.3 vol.% *
• At 7 vol.%; 6.7 - 7.3 vol.% *
• At 8 vol.%; 7.7 - 8.3 vol.% *
* = Correct for temperature and carrier gas if necessary.
o. Switch off the vaporizer until 0 engages.
p. Switch off the AIR or O
flow.
2
2 - 260070-10-0683AS3000™ Service Manual
3.0
Repair Information
3.1Introduction
This chapter of the Service Manual provides the necessary technical information to perform
repairs to the system. The most important prerequisites for effective troubleshooting are a
thorough understanding of the system’s functions, as well as understanding its principles of
operation.
AS3000™ Service Manual0070-10-06833 - 1
Warnings and CautionsRepair Information
3.2Warnings and Cautions
In the event the instrument covers are removed, observe the following warnings and cautions:
3.2.1Warnings
WARNING: Whenever using anesthetic gases, nitrous oxide, oxygen, or
any hospital gas always follow the appropriate agent
evacuation/collection procedures. Use the hospital gas
evacuation system.
3.2.2Cautions
CAUTION:This device uses high pressure compressed gas. When
attaching or disconnecting backup gas cylinders, always
turn the cylinder valves slowly. Use the AS3000 flow meters
to bleed down the pressure, watching the cylinder gauge
indicate the depleting cylinder pressure, before
disconnecting the cylinder from the yoke. Always open and
close cylinder valves fully.
CAUTION:This device operates using compressed gas at high
pressures from the hospital central supply. When connecting
gas supply lines attach the hose connection to the machine
before connecting the quick disconnect fitting to the hospital
source. Disconnect the supply hose from the hospital source
connection prior to disconnecting it from the AS3000 gas
connection fittings.
3.3Troubleshooting Guidelines
1. Identify the problem— Due to the wide variety of potential symptoms, certain
problems may be more subtle than others. Following the guidelines of the tests will help
determine the problem, if one exists.
2. Avoid shorting component leads together — During repair procedures, it can
be tempting to make a quick series of measurements. Always turn the power off before
connecting and disconnecting the test leads and probes. The accidental shorting of
leads can easily stress the components and cause a second failure (aside from the safety
risk).
3. Use the proper equipment — The equipment listed in “Special Tools Required” on
page 3-3, is suggested to fulfill a wide range of troubleshooting requirements. It is
imperative to use the designated equipment in order to ensure proper results of any and
all test procedures.
4. Clean up the repair area— After any repair, clean off the repair area.
3 - 20070-10-0683AS3000™ Service Manual
Repair InformationSpecial Tools Required
3.4Special Tools Required
PART NUMBER DESCRIPTION SPECIFICATION
10070-00-0683AS3000 Service ManualNA
2not applicableVaporizer Instruction ManualNA
3not applicableSafety AnalyzerDempsey 430 or equivalent
10not applicableHand tools, Allen wrench setMetric
110367-00-0080 Tank WrenchNA
120103-00-0508 Y-Fitting15 mm connection
130004-00-0076 Respiration Tube (2 required)0.6 meter silicone, 15 mm
140992-00-0139 Breathing Bag2.3 L silicone
150510-00-0020 Krytox LubricantNA
16not applicableVent TesterCertifier-FA PLUS Respiratory
170997-00-0641AS3000 Calibration hoseNA
180103-00-0691AS3000 test plugNA
190040-00-0448Plug KitNA
200453-00-1216Regulator Calibration HoseNA
O) Analyzer±0.3 V/V% + 5% of reading
2
NMC Digital Pressure Meter
, N2O, AIRMinimum of 35 psi, DISS
2
O, AIRFull PISS yoke connections.
2
connections.
Vent.Tester 4080
AS3000™ Service Manual0070-10-06833 - 3
3 - 40070-10-0683AS3000™ Service Manual
Troubleshooting ChartRepair Information
3.5Troubleshooting Chart
The following table shows common symptoms and corrective actions for problems when troubleshooting the AS3000. The information given
indicates failures during startup and runtime.
3.5.1Common Symptoms and Corrective Actions for Field Service Technicians
MESSAGE/ISSUEOCCURRENCECAUSESOLUTION
“BDU Communication . . .
FAI L”
“Software Version . . . FAIL”Startup test1Software versions of GUI, BDU and
“BDU EEPROM Data . . .
FAI L”
“EEPROM IC . . . FAIL”Startup test1The BDU control board is defective or its
“BDU WDT . . . FAIL”Startup test1The Watch Dog Timer in the digital signal
“AD/DA . . . FAIL”Startup test1The AD or 4052 in the digital signal
Startup test1The power supply switch is turned to ON,
immediately after turned to OFF, which
makes the time too short for the BDU control
board power to reboot.
2The data wire is disconnected or damaged.
3The BDU control board is defective.
Keyboard are incompatible.
2The power supply switch was turned ON,
and OFF quickly.
3The data wire is disconnected or damaged.
4The BDU control board is defective.
5The keyboard board is defective.
Startup test1Data stored in EEPROM on BDU control
board is lost (checksum error)
2The BDU is defective
EEPROM is damaged.
processor on the BDU control board is
damaged.
processor on BDU control board is
damaged.
1Turn the power supply switch to OFF and
power on again after waiting for a while (1 - 2
seconds).
2Check whether the data wire (signal wire at
backside of display) is disconnected or is
damaged.
3Replace the BDU control board and perform
each item of calibration and startup test again.
1Update software version.
2Turn the power supply switch to OFF, wait 1 -2
seconds, then power on again.
3Check whether the data wire (signal wire at
backside of display) is disconnected or is
damaged.
4Replace the BDU control board and perform
each item of calibration and startup test again.
5Replace the user interface assembly.
1Startup. Then perform each item of the
calibration and startup test again.
2Replace the BDU board
1Check that the EEPROM (BDU U12) is
connected correctly, or Replace the EEPROM
and BDU control board, and perform each
item of the calibration and startup test again.
1Replace the BDU control board, and perform
each item of calibration and startup test again.
1Replace the BDU control board, and perform
each item of the calibration and startup test
again.
AS3000™ Service Manual0070-10-06833 - 5
MESSAGE/ISSUEOCCURRENCECAUSESOLUTION
“Vent/Manual Valve Failure”Startup test1The wire that connects to the vent/manual
valve is disconnected or is damaged.
1Check whether the connection wire of the
vent/manual valve is disconnected or is
damaged.
2The vent/manual solenoid valve is
2Replace the vent/manual solenoid valve.
damaged.
3The amplifier board or connection cable
defective.
3Check the connection cable and replace the
amplifier board. Then perform each item of the
calibration and startup tests again.
“Expiration Sensor Failure”Startup test1Fresh gas flow is present at startup.
1Turn off all fresh gas flow and reboot the unit.
Repair InformationTroubleshooting Chart
2The sensor board is defective.
“Keyboard Communication .
. . FAIL”
Startup test1The data wire is disconnected or is
damaged.
2The keyboard board is defective.
“Pressure Sensor . . . FAIL”Startup test1The connection wire of the PAW board is
2The oxygen sensor is out of range.
3The main unit-oxygen sensor cable is
disconnected or damaged.
4The amplifier board is defective.
“PEEP Valve . . . FAIL”Startup test1The connection wire of the PEEP valve is
disconnected or damaged.
2The PEEP valve is defective.
3The Amplifier board is defective.
“Inspiration Valve . . . FAIL”Startup test1The connection wire of the inspiration valve
is disconnected or damaged.
2The inspiration valve is defective.
3The amplifier board is defective.
2Replace the sensor board, and then perform
each item of the calibration and startup test
again.
1Replace the user interface assembly.
2Replace the user interface assembly.
1Check whether the connection wire of the PAW
board is disconnected or damaged.
2Check the power supply and connection of the
sensor board. Replace the Paw board and
sensor board.
3Check the tubing for kinks and remove any
occlusions.
1Increase O
concentration.
2
2Replace the oxygen sensor.
3Check whether the main unit-oxygen sensor
cable is disconnected or damaged.
4Replace the amplifier board, and then perform
each item of the calibration and startup test
again.
1Check whether the PEEP valve is disconnected
or damaged.
2Replace the PEEP valve.
3Replace the amplifier board and then perform
each item of the calibration and startup test
again.
1Check whether the inspiration valve is
disconnected or damaged.
2Replace the inspiration valve.
3Replace the amplifier board and then perform
each item of the calibration and startup test
again.
3 - 60070-10-0683AS3000™ Service Manual
Troubleshooting ChartRepair Information
MESSAGE/ISSUEOCCURRENCECAUSESOLUTION
“Inspiration Sensor . . . FAIL”
or
“Inspiration Sensor . . . FAIL
Try re-starting the system with
all gas flows OFF” (UI
version 2.25 and higher)
Startup test1Fresh gas is flowing during the startup
procedure.
2The tubing kinked or occluded.
3The sensor board is defective.
1Verify that no fresh gas is flowing during the
startup procedure.
2Check the tubing for kinks and remove any
occlusions.
3Replace the sensor board and then perform
each item of the calibration and startup test
again.
Supply . . . FAIL”Startup test
“O
2
Runtime
1The gas source is not connected.
2The gas supply pressure switch or its cable
is damaged.
3The amplifier board is defective.
1Check gas source.
2Check and replace the switch or cable of the
gas source.
3Replace the amplifier board, and then perform
each item of the calibration and startup test
again.
“AC Power Failure”Startup test
Runtime
“Low Battery”Runtime1Battery capacity is low.
“Alarm Speaker . . . FAIL”Startup test1The connection wire to the speaker is
“Software Mismatch”
Startup test1Power is recycled to the unit too quickly, not
or
1AC power is not connected.
2A fuse is tripped
2The battery is defective.
3The connection wire or power board is
defective.
disconnected or damaged.
2The speaker is damaged.
3The keyboard board is defective.
allowing the memory to fully clear.
1Check the connection of AC power.
2Replace the fuse.
1Charge the battery by connecting AC power.
2Replace the battery.
3Check the connection wire or replace the
power board.
1Check whether the connection wire of the
speaker is disconnected or damaged.
2Replace the user interface assembly.
3Replace the user interface assembly.
1Shut down the unit and wait until the backlight
is fully extinguished before restoring power.
“Software Mismatch or
Failure to Shut Down
Completely” (UI version 2.25
and higher)
Charging state is always
displayed.
During startup leak test, the
bellows moves when it
Runtime1Battery capacity is low.
2The connection wire to the battery is
disconnected or damaged.
3The battery is defective.
1Charge the battery by connecting AC power.
2Check whether the connection wire to the
battery is disconnected or damaged.
3Replace the battery.
Startup test1A solenoid failure or wiring error.1Correct the wiring or replace the solenoid
valve.
shouldn’t
In leakage detection, The APL
Valve detection pressure
displayed on the pressure
gauge does not stay within
28 - 32 cmH
O.
2
Startup test1The Breathing System pneumatic hose leaks
gas when in manual mode.
2Pressure gauge failure.
3The APL Valve is out of spec
1Replace the Breathing System pneumatic hose
2Replace the pressure gauge
3Replace the APL Valve
AS3000™ Service Manual0070-10-06833 - 7
MESSAGE/ISSUEOCCURRENCECAUSESOLUTION
In leakage detection, Safety
valve detection failed.
In leakage detection, the
leakage volume is 501 999mL.
In leakage detection, the
leakage volume is larger than
1L.
In compliance test,
compliance test failed.
The power supply switch on
the main unit is defective.
All or some keys are non
responsive.
The encoder is non
responsive.
All or some of the work lights
are defective.
The switch on the work light
is defective.
The flow meter light is non
operational.
Startup test1The proportional valve on the airway
module failed.
2The safety valve in the airway module
failed.
Startup Test1The tubing connection is not tight.
2Incorrect installation of the absorber
canister.
3The pressure gauge or oxygen sensor on the
circuit is not installed tightly.
4There is a leakage in the fresh gas supply.
Startup Test1The tubing connection is not tight.
2Incorrect installation of the absorber
canister.
3The pressure gauge or oxygen sensor on the
circuit is not installed tightly.
4There is a leakage in the fresh gas supply.
Startup Test1The Y-fitting is not sealed or the Breathing
System has leakage.
2The respiration tubes are too long.
3Calibration of the pressure sensor is
inaccurate.
4The flow meter is set incorrectly
5The calibration of the flow meter is
inaccurate
All1The power supply switch on the main unit or
the cable connection failed.
2Communication board is defective.
All1The keyboard is defective.
2The communication board is defective.
All1The encoder failed.
2The communication board is defective.
All1The work light circuit board is defective.
2The cable is defective
All1The switch on the work light is defective.1Replace the work light switch.
All1The power supply to the light is defective
2The foil of light is defective.
3The inverter is defective.
1Calibrate the inspiration valve again.
2Replace the airway module.
1Check the connection on the tubing.
2Check the installation of the absorber canister.
3Check the installation of the pressure gauge or
oxygen sensor on circuit.
4Check the fresh gas supply for leakage
1Check the connection on the tubing.
2Check the installation of the absorber canister.
3Check the installation of the pressure gauge or
oxygen sensor on circuit.
4Check the fresh gas supply for leakage
1Check the connection and leakage of
respiration tubes.
2The respiration tubes cannot be too long.
3Calibrate the pressure oxygen again.
4Set the flow meter to the correct value
5Recalibrate the flow meter
1Check the connection cable of the power
supply switch of the main unit, or replace the
power supply switch of main unit.
2Replace the communication board.
1Replace the user interface assembly.
2Replace the user interface assembly.
1Replace the user interface assembly.
2Check and replace the user interface assembly.
1Replace the work light circuit board.
2Check the connection to the power supply.
1Replace the power supply to the light.
2Replace the foil of light.
3Replace the inverter
Repair InformationTroubleshooting Chart
3 - 80070-10-0683AS3000™ Service Manual
Troubleshooting ChartRepair Information
MESSAGE/ISSUEOCCURRENCECAUSESOLUTION
Heating of circuit is non
functional.
AC Outlets are non
Runtime1Line power is not connected.
2The circuit cable is defective.
3Heating control board is defective.
1Check the line power connection.
2Replace the cable connection to the circuit.
3Replace the heating control board.
All1A fuse is tripped1Replace the fuse
functional
reading is out of
The O
2
tolerance
Runtime1The calibration of the oxygen sensor is
inaccurate, or the oxygen sensor is
1Calibrate the oxygen sensor again, or replace
the oxygen sensor.
defective.
The display screen has a
hotspot or uneven brightness
The display screen is a solid
color
All1The display screen is defective.
2PC104 is defective.
Startup Test
Runtime
1The GUI program has crashed.
2The signal cable is disconnected.
3PC104 is defective.
4The communication board or power supply
1Replace the user interface assembly.
2Replace the user interface assembly.
1Reboot the system
2Check the signal cable.
3Replace the user interface assembly.
4Replace the user interface assembly.
for the light is defective.
The pressure reading is out of
tolerance
The PEEP reading is out of
tolerance
the tidal volume reading is
out of tolerance
5The display screen is damaged.
Runtime1Calibration of the pressure sensor is out of
spec.
Runtime1Calibration of expiration valve (PEEP) is
inaccurate.
2Calibration of pressure sensor is inaccurate.
Runtime1The compliance test after startup is
inaccurate.
2Calibration of the flow sensor is inaccurate.
3Calibration of the inspiration valve is
5Replace the user interface assembly.
1Calibrate the pressure sensor again.
1Calibrate the expiration valve (PEEP).
2Calibrate the pressure sensor.
1Perform the compliance test after startup again.
2Calibrate the flow sensor again.
3Calibrate the inspiration valve again.
inaccurate.
4Calibration of the pressure sensor is
4Calibrate the pressure sensor again.
inaccurate.
“System Self Test”Startup Test1Self-test during each startup.1No action needed.
“NON-FUNCTIONAL Call
Service Representative”
Startup Test1A failure has been found (Displayed on the
screen), which leads to the abnormal
1Handle it according to specific information on
the failure.
operation of anesthesia machine.
After the startup test, only the
MANUAL/AUTO key is
displayed, with the alarm:
(except O
supply fail)
2
After the startup test, there
are Bypass and Continue
keys, but with an alarm
Startup Test1A failure has been found (Displayed on the
screen), which leads to the abnormal
operation of anesthesia machine.
Startup Test1A failure has been found (Displayed on the
screen), but it does not affect the normal
operation of each ventilation mode.
1Handle it according to specific information on
the failure.
1Handle it according to specific information on
the failure.
displayed
AS3000™ Service Manual0070-10-06833 - 9
MESSAGE/ISSUEOCCURRENCECAUSESOLUTION
“Ventilator setting is not
possible!”
Runtime1A parameter has been set to a limit that can
be set.
1The parameter set cannot exceed the
adjustable range. If the limit is not reached,
another parameter may be limiting the range
and needs to be adjusted accordingly
“Displayed parameter values
are for the pending mode”
Runtime1The AS3000 is in the pre-selection mode,
but it is still working according to the
original operating mode.
1Set parameters 15 seconds in advance; or
directly confirm and enter ventilation mode; or
press the NORMAL SCREEN button to return
to the original ventilation mode interface; or do
not perform any operation and it will
automatically return to the original ventilation
mode interface after 15 seconds.
“Pressure, Volume and
Apnea Alarms are OFF!”
“Automatic Ventilation not
available!”
Runtime1The functional alarm in Manual mode is
disabled.
Startup Test
Runtime
1A failure has been found (Displayed on the
screen); the AS3000 can work only in
1Enable the functional alarm in Manual mode.
1Handle it according to specific information on
the failure.
Standby/Manual mode. This information is
displayed after entering the Main Screen.
“BDU Communication
Failure!”
Sensor failure”Runtime1The main unit-oxygen sensor cable is
“O
2
Startup Test
Runtime
1The communication cable is disconnected.
2The BDU software has crashed or reset.
1Check the communication cable.
2Power on again.
1Check the unit-oxygen sensor cable.
disconnected.
Supply failure”Runtime1The gas source is closed or disconnected.1Check the gas source.
“O
2
Continuous PressureRuntime1Airway pressure is too high.
2The expiration valve (PEEP) is defective.
3The airway is occluded.
1Relieve the airway pressure
2Repair/replace the expiration valve
3Clear the occlusion
High Airway PressureRuntime1Airway pressure is high.1Adjust the upper limit set for the airway
pressure alarm.
Low Airway PressureRuntime1Airway pressure is low.1Adjust the lower limit set for the airway
pressure alarm.
Negative PressureRuntime1Airway pressure is continuously negative.1Check the airway and fresh gas setting
Low FiO
2
Runtime1Oxygen concentration is lower than the
lower limit set for Oxygen Concentration
alarm.
2Oxygen calibration is inaccurate.
1Adjust the lower limit set for the Oxygen
Concentration alarm or the actual oxygen
concentration.
2Calibrate the oxygen concentration sensor
again.
Repair InformationTroubleshooting Chart
3 - 100070-10-0683AS3000™ Service Manual
Troubleshooting ChartRepair Information
MESSAGE/ISSUEOCCURRENCECAUSESOLUTION
High FiO
2
Runtime1Oxygen concentration is higher than the
upper limit set for the Oxygen Concentration
alarm.
2Oxygen calibration is inaccurate.
1Adjust the upper limit set for the Oxygen
Concentration alarm or the actual oxygen
concentration.
2Calibrate the oxygen concentration sensor
again.
APNEARuntime1Breathing hoses are disconnected.1Check the breathing hoses.
Low BatteryRuntime1Battery capacity is low.1Connect the unit to AC power.
High Minute VolumeRuntime1Minute ventilation is higher than the upper
limit set for the Oxygen Concentration
1Adjust the upper limit set for the Minute
Ventilation alarm.
alarm.
Low Minute VolumeRuntime1Minute ventilation is lower than the lower
limit set for the Minute Volume alarm.
APNEA BackupRuntime1PS mode is not triggered and APNEA
1Adjust the lower limit set for the Minute Volume
alarm.
1No action needed.
backup ventilation is given according to the
minimum breathing frequency that has been
set.
High PEEPRuntime1The expiration time is too short or expiration
limb is occluded
1Check the time setting and expiration pipeline.
2Check the expiration valve (PEEP).
2The expiration valve (PEEP) is defective.
Keyboard communication
Failure
Startup Test
Runtime
1The signal wire is disconnected or
damaged.
1Check the signal wire.
2Power on again.
2The keyboard board program crashed.
AC Power FailureAll1AC power is not connected or a fuse is
1Connect AC and check the fuses.
tripped.
Cal DueStartup Test
O
2
Runtime
Tidal volume displayed by
Runtime1This caused by the compliance and
the AS3000 is smaller than
the value displayed on
1The oxygen sensor has not been calibrated
for more than 3 days.
compensation of resisting force between the
airway and patient.
1Calibrate the oxygen sensor.
1No action needed.
bellow.
Bellows is deflated and at the
bottom of its travel during
runtime, after initial power-up
and Leak Test.
Runtime1This is caused by a disconnected breathing
circuit, a removed absorber canister, a
removed O
check valve dome rings, or no fresh gas
sensor or sensor adapter,
2
flow while ventilator is running.
1Install a breathing circuit and connect the Y-
piece to the Test port. Select CMV mode and
confirm the selection. Immediately push the O
Flush button until the bellows is completely
filled. Return to STANDBY mode and remove Y-
2
piece from Test port.
Repair InformationLeak Troubleshooting
PASSLeak Test
Manual
Leak Test
PASS
No Leaks
Detected
FAILFAIL
Manual
Leak Test
Seal the
scavenger port
and repeat the
test
PASSPASS
Check APL
Valve
FAILFAIL
Check the
Manual bag
and bag arm
Check the
Exhaust valve
Check the
Bellows Dome
Check the
Proportional
Valve
Check the PEEP
valve
Pinch the CGO
Hose and
repeat the Leak
test
PASS
Remove the
Vaporizers
and repeat the
Leak test
PASS
Check the
vaporizers and
the vaporizer
mount
FAILFAIL
Check the
Flowmeter
module /
CGO port
Remove
sampling
line
Check and if
necessary
exchange the
breathing hoses
Check
the O2
cell
o-ring
Check
the
Absorber
assembly
Verify the I/E Rings
are tight and not
cracked (do not
over tighten!)
Check the
PAW
gauge and
its o-rings
3.6Leak Troubleshooting
FIGURE 3-1 Leak Troubleshooting Flow Chart
AS3000™ Service Manual0070-10-06833 - 11
Gas Inlet
Block
N20
AIR
O2
Breathing
System
Ventilator
Pressure
Switch
Pressure
Regulator
Pressure
Regulator
Pressure
Regulator
Ratio•
Controller
N20•
Cuttoff
Flow
Tubes
Cylinder
Cylinder
Cylinder
Inlet Filter
Inlet Filter
Inlet Filter
O2
Flow
Meter
O2
Outlet
AIR•
Flow
Meter
Patient
Vaporizer
O2
Flush
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
OOO
P
P
P
P
P
P
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Fresh Gas
Flow Sensor
V
Bellows
Vaporizer
3 - 120070-10-0683AS3000™ Service Manual
Test P n e um a ti c sRepair Information
3.7Test Pneumatics
The following section shows the pneumatic components involved while performing the AS3000 startup tests.
3.7.1Leak Test - Manual Ventilation Test
FIGURE 3-2 Manual Ventilation Test Pneumatics
Gas Inlet
Block
N20
AIR
O2
Breathing
System
Ventilator
Pressure
Switch
Pressure
Regulator
Pressure
Regulator
Pressure
Regulator
Ratio•
Controller
N20•
Cuttoff
Flow
Tubes
Cylinder
Cylinder
Cylinder
Inlet Filter
Inlet Filter
Inlet Filter
O2
Flow
Meter
O2
Outlet
AIR•
Flow
Meter
Patient
Vaporizer
O2
Flush
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
OOO
P
P
P
P
P
P
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Fresh Gas
Flow Sensor
V
Bellows
Vaporizer
AS3000™ Service Manual0070-10-06833 - 13
3.7.2Safety Valve Test
Repair InformationTest P n eu m at ic s
FIGURE 3-3 Safety Valve Test Pneumatics
Gas Inlet
Block
N20
AIR
O2
Breathing
System
Ventilator
Pressure
Switch
Pressure
Regulator
Pressure
Regulator
Pressure
Regulator
Ratio•
Controller
N20•
Cuttoff
Flow
Tubes
Cylinder
Cylinder
Cylinder
Inlet Filter
Inlet Filter
Inlet Filter
O2
Flow
Meter
O2
Outlet
AIR•
Flow
Meter
Patient
Vaporizer
O2
Flush
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
OOO
P
P
P
P
P
P
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Fresh Gas
Flow Sensor
V
Bellows
Vaporizer
3 - 140070-10-0683AS3000™ Service Manual
Test P n e um a ti c sRepair Information
3.7.3Leak Test - Automatic Ventilation Test
FIGURE 3-4 Automatic Ventilation Test Pneumatics
Gas Inlet
Block
N20
AIR
O2
Breathing
System
Ventilator
Pressure
Switch
Pressure
Regulator
Pressure
Regulator
Pressure
Regulator
Ratio•
Controller
N20•
Cuttoff
Flow
Tubes
Cylinder
Cylinder
Cylinder
Inlet Filter
Inlet Filter
Inlet Filter
O2
Flow
Meter
O2
Outlet
AIR•
Flow
Meter
Patient
Vaporizer
O2
Flush
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
OOO
P
P
P
P
P
P
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Fresh Gas
Flow Sensor
V
Bellows
Vaporizer
AS3000™ Service Manual0070-10-06833 - 15
3.7.4Compliance Test
Repair InformationTest P n eu m at ic s
FIGURE 3-5 Compliance Test Pneumatics
1
2
3
4
5
6
7
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
36
37
38
39
40
41
42
8•
9
10
11
Gas Inlet
Block
N20
AIR
O2
Breathing
System
Ventilator
Pressure
Switch
Pressure
Regulator
Pressure
Regulator
Pressure
Regulator
Ratio•
Controller
N20•
Cuttoff
Flow
Tubes
Cylinder
Cylinder
Cylinder
Inlet Filter
Inlet Filter
Inlet Filter
O2
Flow
Meter
O2
Outlet
AIR•
Flow
Meter
Patient
Vaporizer
O2
Flush
Pressure
Regulator
Proportional
Valve
Solenoid
Valve
Fresh Gas
PEEP
Valve
Exhaust
Valve
APL Valve
AGSS
Electronic
Control
Reversing
Valve
O2 Sensor
Insp. Valve
Absorber
Canister
Exp. Valve
Breathing Bag
OOO
P
P
P
P
P
P
P
Exp. Flow
Sensor
Insp. Flow
Sensor
Fresh Gas
Flow Sensor
V
Bellows
Vaporizer
3 - 160070-10-0683AS3000™ Service Manual
Pneumatic Hose and Wiring DiagramsRepair Information
3.8Pneumatic Hose and Wiring Diagrams
3.8.1Pneumatic Hose Labeling
FIGURE 3-6 Pneumatic Hose Labeling Diagram
AS3000™ Service Manual0070-10-06833 - 17
10 Insp. Low
8 Exp. High
7 PEEP Control
12 Manual/Auto
Switch
11 Insp. High
9 Exp. Low
10 Insp. Low
8 Exp. High
7 PEEP Control
12 Manual/Auto
Switch
11 Insp. High
9 Exp. Low
3.8.2Sampling Pipeline Module Interface Labeling
Repair InformationPneumatic Hose and Wiring Diagrams
Pneumatic Hose and Wiring DiagramsRepair Information
4.0
Replacement Parts and Accessories
4.1Introduction
This section of the manual provides information that is necessary to identify replacement
parts and accessories.
4.2Available Replacement Parts and Sub-Assemblies
The following part lists are divided into two sections. The Isometric Drawings and the
accompanying part lists identify the available chassis mounted components.
4.2.1Exchange Program
Mindray DS USA, Inc. offers an exchange policy for many of the printed circuit board
assemblies. This program may provide the most expedient method of servicing the
equipment. A standard charge is associated with this service. Contact the Service
Department for details concerning this exchange program.
Many circuit boards make extensive use of multi-layer and surface mount technology.
Individual component replacement is not recommended on these boards. Board exchange or
replacement is the most efficient method of repair for these types of assemblies. Component
level repair is not recommended.
Circuit boards, returned as parts of the exchange program, that show evidence of improper
repair techniques are not considered for exchange. Damaged boards will be invoiced at full
value and no exchange credit will be applied.
4.2.2Replacement Parts Pricing Information
Current parts prices and exchange charges can be determined by contacting the Customer
Support Department.
AS3000™ Service Manual0070-10-06834 - 1
Available Replacement Parts and Sub-AssembliesReplacement Parts and Accessories
4.2.3Ordering Information
Replacement parts and assemblies are available. Please follow these guidelines when
ordering replacement items for the product:
1. Include the Model and Serial Number of the product.
2. Include the Part Number exactly as it appears in the Parts List under the column, “Part
Number.”
3. Include a description of the item.
EXAMPLE ORDERS: (1) ea. P/N 0334-00-2611-01
Label, Fuse Replacement, Serial No. XXXX
(2) ea. P/N 0213-07-0404
Screw, Self Tap, #4 x 0.25", Serial No. XXXX
NOTE:Mindray DS USA, Inc. maintains a policy of continuous
development for product improvement and reserves the
right to change materials, specifications, and prices without
notice.
4 - 20070-10-0683AS3000™ Service Manual
Replacement Parts and AccessoriesIsometric Drawings
8
17
16
15
14
13
12
11
10
9
7
1
2
3
4
6
5
4.3Isometric Drawings
4.3.1Chassis
FIGURE 4-1 AS3000 Frontview
AS3000™ Service Manual0070-10-06834 - 3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.