Metrohm 899 User Manual

899 Coulometer
Manual
8.899.8001EN
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
899 Coulometer
8.899.8001EN
Manual
04.2012 ebe
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://products.metrohm.com under Literature/Technical documenta-
tion.
■■■■■■■■■■■■■■■■■■■■■■

Table of contents

1 Introduction 1
1.1 Instrument description ......................................................... 1
1.1.1 Connectors .............................................................................. 1
1.1.2 Intended use ........................................................................... 2
1.2 About the documentation ................................................... 2
1.2.1 Symbols and conventions ........................................................ 2
1.3 Safety instructions ................................................................ 3
1.3.1 General notes on safety ........................................................... 3
1.3.2 Electrical safety ........................................................................ 3
1.3.3 Tubing and capillary connections ............................................. 4
1.3.4 Flammable solvents and chemicals ........................................... 5
1.3.5 Recycling and disposal ............................................................. 5
2 Overview of the instrument 6
Table of contents
2.1 Front ...................................................................................... 6
2.2 Rear ........................................................................................ 7
3 Installation 8
3.1 Setting up the instrument .................................................... 8
3.1.1 Packaging ................................................................................ 8
3.1.2 Checks .................................................................................... 8
3.1.3 Location .................................................................................. 8
3.2 Setting up the coulometer cell ............................................ 9
3.2.1 Inserting the coulometer cell .................................................... 9
3.2.2 Preparing the coulometer cell .................................................. 9
3.2.3 Mounting the addition and aspiration tube (utilization with Ti
Stand) ................................................................................... 13
3.2.4 Using the coulometer cell with a Karl Fischer oven ................. 14
3.2.5 Using the coulometer cell with a sample changer .................. 14
3.3 Connecting the coulometer to the power supply ............ 14
3.3.1 Connecting the power supply unit ......................................... 14
3.3.2 Connecting the Power Box .................................................... 15
3.4 Connecting sensors ............................................................ 17
3.4.1 Connecting a generator electrode .......................................... 17
3.4.2 Connecting an indicator electrode ......................................... 18
3.4.3 Connecting a temperature sensor .......................................... 20
899 Coulometer
3.5 Connecting an additional stirrer ....................................... 21
3.6 Connecting a balance ......................................................... 22
3.7 Connecting a keyboard, printer and other USB devi-
ces ........................................................................................ 24
■■■■■■■■
III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
3.8 Connecting a sample changer to the remote connec-
tor ........................................................................................ 27
4 Coulometric titration 28
4.1 Principle of coulometry according to Karl Fischer .......... 28
4.2 Working with water standards ......................................... 29
4.2.1 Certified water standards ....................................................... 29
4.2.2 Practical recommendations .................................................... 29
4.3 Sample addition .................................................................. 31
4.3.1 Size of the sample size ........................................................... 31
4.3.2 Working with liquid samples .................................................. 32
4.3.3 Working with solid samples ................................................... 32
4.4 Optimum working conditions ........................................... 33
4.4.1 General ................................................................................. 33
4.4.2 Drift ...................................................................................... 33
4.4.3 Reagent replacement ............................................................. 34
4.4.4 Indicator electrode ................................................................ 34
5 Operation 35
5.1 Switching the instrument on and off ............................... 35
5.2 Fundamentals of operation ............................................... 36
5.2.1 The keypad ............................................................................ 36
5.2.2 Structure of the dialog windows ............................................ 36
5.2.3 Navigating in the dialog ......................................................... 37
5.2.4 Entering text and numbers ..................................................... 37
5.2.5 Selecting from a selection list ................................................. 38
5.3 Formula editor .................................................................... 39
5.4 Methods .............................................................................. 40
5.4.1 Method templates ................................................................. 40
5.4.2 Loading a method template ................................................... 40
5.4.3 Saving a method ................................................................... 41
5.4.4 Exporting a method ............................................................... 42
5.5 Control ................................................................................. 43
5.6 Sample data ........................................................................ 44
5.6.1 Entering sample data in the main dialog ................................ 44
5.6.2 Requesting sample data at the start of the determination ...... 45
5.7 Sample table ....................................................................... 45
5.7.1 General ................................................................................. 45
5.7.2 Editing the sample data ......................................................... 47
5.7.3 Sending the sample size from a balance ................................. 49
■■■■■■■■
IV
5.8 Carrying out a determination ............................................ 49
5.9 Live modifications .............................................................. 52
5.9.1 Editing the sample data of the running determination ............ 52
5.9.2 Editing the sample table while a determination is running ...... 53
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
6 System settings 62
Table of contents
5.9.3 Editing the live parameters ..................................................... 55
5.10 Results ................................................................................. 56
5.11 Statistics .............................................................................. 57
5.12 Printing a report manually ................................................. 59
5.13 Manual control ................................................................... 60
5.13.1 Stirring .................................................................................. 60
6.1 Basic settings ...................................................................... 62
6.2 Managing common variables ............................................ 65
6.2.1 General ................................................................................. 65
6.2.2 Editing common variables ...................................................... 65
6.3 File management ................................................................ 66
6.4 Configuring external devices ............................................. 68
6.5 Instrument diagnosis .......................................................... 71
6.5.1 Loading program versions and language files ......................... 71
6.5.2 Diagnosis functions ............................................................... 72
7 Parameters 73
7.1 Coulometric Karl Fischer titrations (KFC) ......................... 73
7.1.1 Conditioning ......................................................................... 73
7.1.2 Start conditions ..................................................................... 75
7.1.3 Control parameters ................................................................ 76
7.1.4 Titration parameters .............................................................. 79
7.1.5 Stop conditions ..................................................................... 81
7.1.6 Calculation ............................................................................ 81
7.1.7 Statistics ................................................................................ 85
7.1.8 Reports .................................................................................. 86
8 Operation and maintenance 88
8.1 Quality Management and validation with Metrohm ....... 88
9 Troubleshooting 89
9.1 Karl Fischer titration .......................................................... 89
9.2 Miscellaneous ..................................................................... 91
10 Appendix 92
10.1 Stirring rate ......................................................................... 92
899 Coulometer
10.2 Balance ................................................................................ 92
10.3 USB devices ......................................................................... 93
10.3.1 6.2147.000 numerical USB keypad ........................................ 93
10.3.2 Key assignment of a USB keyboard ........................................ 94
10.3.3 PC mouse .............................................................................. 95
■■■■■■■■
V
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
10.3.4 Printer ................................................................................... 95
10.4 Control parameters and polarization current .................. 95
10.5 Method examples for working with the Karl Fischer
oven ..................................................................................... 97
10.6 System initialization ........................................................... 98
10.7 Remote interface ................................................................ 99
10.7.1 Pin assignment of the remote interface .................................. 99
10.7.2 Status diagram of the remote interface ................................ 100
10.8 Remote control via an RS-232 connection ..................... 101
10.8.1 Commands and variables ..................................................... 102
10.9 Arithmetic algorithms in the 899 Coulometer ............... 103
11 Technical specifications 106
11.1 Measuring inputs .............................................................. 106
11.1.1 Indicator electrode .............................................................. 106
11.1.2 Temperature ........................................................................ 106
11.2 Generator connector ........................................................ 106
11.2.1 Generator electrode ............................................................ 106
11.3 Internal stirrer ................................................................... 107
11.4 Interfaces and connectors ............................................... 107
11.5 Power supply .................................................................... 107
11.6 Safety specifications ........................................................ 108
11.7 Electromagnetic compatibility (EMC) ............................. 108
11.8 Ambient temperature ...................................................... 109
11.9 Reference conditions ........................................................ 109
11.10 Dimensions ........................................................................ 109
12 Warranty (Guarantee) 110
13 Accessories 112
13.1 Scope of delivery .............................................................. 112
13.1.1 2.899.0010 ......................................................................... 112
13.1.2 2.899.0110 ......................................................................... 119
13.1.3 2.899.1010 ......................................................................... 125
13.1.4 2.899.1110 ......................................................................... 132
13.1.5 2.899.2110 ......................................................................... 139
13.1.6 2.899.3110 ......................................................................... 152
■■■■■■■■
VI
13.2 Optional accessories ........................................................ 164
Index 167
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■

Table of figures

Figure 1 Front 899 Coulometer ....................................................................... 6
Figure 2 Rear 899 Coulometer ........................................................................ 7
Figure 3 Inserting the coulometer cell ............................................................. 9
Figure 4 Filling the adsorber tube .................................................................. 10
Figure 5 Equipping the coulometer cell ......................................................... 11
Figure 6 Mounting the addition and aspiration tube ...................................... 13
Figure 7 Connecting the power supply unit ................................................... 15
Figure 8 Connecting the Power Box .............................................................. 16
Figure 9 Unscrewing the cover from the generator electrode ........................ 17
Figure 10 Screwing the electrode cable to the general electrode ..................... 18
Figure 11 Connecting a generator electrode ................................................... 18
Figure 12 Unscrew the cover from the indicator electrode .............................. 19
Figure 13 Screwing on the electrode cable to the indicator electrode .............. 19
Figure 14 Connecting an indicator electrode ................................................... 19
Figure 15 Connecting a temperature sensor .................................................... 20
Figure 16 Connecting a stirrer ......................................................................... 21
Figure 17 Connecting the balance via RS-232/USB Box .................................... 22
Figure 18 Connecting USB devices .................................................................. 24
Figure 19 Connecting the USB flash drive ........................................................ 26
Figure 20 Connecting the 6.2147.000 USB keyboard with USB flash drive and
printer ............................................................................................. 26
Figure 21 Connecting the USB hub with USB flash drive, printer and 6.2148.030
RS-232/USB Box .............................................................................. 27
Figure 22 Connecting a remote cable ............................................................. 27
Figure 23 Keypad 899 Coulometer .................................................................. 36
Figure 24 Directory structure on the USB flash drive ........................................ 67
Figure 25 Rotational speed depending on stirring rate .................................... 92
Figure 26 Control parameters and polarization current .................................... 96
Figure 27 Pin assignment of remote socket and remote plug .......................... 99
Figure 28 Remote status diagram .................................................................. 100
Figure 29 Connecting the RS-232/USB Box to the PC .................................... 101
Table of figures
899 Coulometer
■■■■■■■■
VII
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction

1.1 Instrument description

The 899 Coulometer is a titrator used for coulometric water content determination according to Karl Fischer. There are method templates available which are already configured except for a few parameters. The methods can be modified and stored under a new name. The methods can be exported to a connected USB flash drive. This function makes it possible for you to copy methods quickly and easily from one instrument to another.
The 899 Coulometer has an integrated magnetic stirrer that is visible on the top side of the housing. The coulometer cell can easily be attached to the magnetic stirrer. Thanks to its compact construction, you can use the device in a small space as a stand-alone titrator. Furthermore, the remote interface permits you to connect a sample changer with oven module and to perform automated determinations.

1 Introduction

As an alternative to the power supply with the power supply unit provi­ded, the 899 Coulometer can also be operated by means of an accumula­tor (a so-called Power Box). The Power Box is available as optional equip­ment. It is particularly appropriate for use in environments where power supplies are either unstable or absent.

1.1.1 Connectors

The instrument is equipped with the following connectors:
Electrical connection
MSB connector (Metrohm Serial Bus)
USB (OTG) connector
Sensor connectors
Grounding socket
For connecting to the mains supply with the aid of the power supply
unit provided or for connection to the 6.2164.500 Power Box.
For connecting an additional stirrer.
The 6.2151.100 adapter can be used to connect, for example, a
printer, a USB flash drive or a USB keyboard.
Three connectors for the following sensor types:
– Double Pt electrode – Generator electrode – Temperature sensor (Pt1000)
For grounding the Coulometer.
899 Coulometer
■■■■■■■■
1

1.2 About the documentation

Remote connector
For connecting instruments with a remote interface (e.g. 885 Compact Oven SC).

1.1.2 Intended use

The 899 Coulometer is designed for usage as a titrator in analytical labora­tories. Its application field is coulometric water content determination according to Karl Fischer.
This instrument is suitable for processing chemicals and flammable sam­ples. The usage of the 899 Coulometer therefore requires that the user have basic knowledge and experience in the handling of toxic and caustic substances. Knowledge with respect to the application of the fire preven­tion measures prescribed for laboratories is also mandatory.
1.2 About the documentation
Caution
■■■■■■■■■■■■■■■■■■■■■■
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe opera­tion of the instrument.

1.2.1 Symbols and conventions

The following symbols and formatting may appear in this documentation:
Method Dialog text, parameter in the software
File New Menu or menu item
[Next] Button or key
Cross-reference to figure legend
The first number refers to the figure number, the sec­ond to the instrument part in the figure.
Instruction step
Carry out these steps in the sequence shown.
Warning
■■■■■■■■
2
This symbol draws attention to a possible life hazard or risk of injury.
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
Warning
This symbol draws attention to a possible hazard due to electrical current.
Warning
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
Warning
This symbol draws attention to a possible biological hazard.
Caution
This symbol draws attention to a possible damage of instruments or instrument parts.
Note
This symbol marks additional information and tips.

1.3 Safety instructions

1.3.1 General notes on safety

Warning
This instrument may only be operated in accordance with the specifica­tions in this documentation.
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.

1.3.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
Warning
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
899 Coulometer
■■■■■■■■
3
1.3 Safety instructions
■■■■■■■■■■■■■■■■■■■■■■
Warning
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
Mains voltage
Warning
An incorrect mains voltage can damage the instrument.
Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.

1.3.3 Tubing and capillary connections

Caution
Leaks in tubing and capillary connections are a safety risk. Tighten all connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections.
Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are manda­tory.
■■■■■■■■
4
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■

1.3.4 Flammable solvents and chemicals

Warning
All relevant safety measures are to be observed when working with flammable solvents and chemicals.
Set up the instrument in a well-ventilated location (e.g. laboratory
flue).
Keep all sources of flame far from the workplace. Clean up spilled fluids and solids immediately. Follow the safety instructions of the chemical manufacturer.

1.3.5 Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
1 Introduction
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
899 Coulometer
■■■■■■■■
5

2.1 Front

1
2
3

2 Overview of the instrument

2.1 Front
■■■■■■■■■■■■■■■■■■■■■■
Figure 1 Front 899 Coulometer
Magnetic stirrer
1
For attaching the titration vessel holder.
Keypad
3
2
Display
■■■■■■■■
6
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
1 2
3
4
5 6
7
8 9

2.2 Rear

2 Overview of the instrument
Figure 2 Rear 899 Coulometer
Type plate
1
Contains the serial number.
Electrode connector (Ind.)
3
For connecting a metal electrode (double Pt wire electrode). Socket F.
MSB connector (MSB 1)
5
Metrohm Serial Bus. For connecting an external stirrer. Mini DIN, 8-pin.
Mains connection socket (Power)
7
For connecting the external power supply unit or a Power Box (6.2164.500).
Grounding socket
9
For grounding the Coulometer. Socket B, 4 mm.
Remote connector
2
For connecting instruments with a remote interface. D-Sub, 9-pin.
Temperature sensor connector (Temp.)
4
For connecting a temperature sensor of the type Pt1000. Two B sockets, 2 mm.
USB (OTG) connector
6
For connecting printers, USB flash drives, USB hubs, etc.
Electrode connector (Gen.)
8
For connecting a generator electrode.
899 Coulometer
■■■■■■■■
7

3.1 Setting up the instrument

3 Installation

3.1 Setting up the instrument

3.1.1 Packaging

The instrument is supplied in highly protective special packaging together with the separately packed accessories. Keep this packaging, as only this ensures safe transportation of the instrument.

3.1.2 Checks

Immediately after receipt, check whether the shipment has arrived com­plete and without damage by comparing it with the delivery note.

3.1.3 Location

The instrument has been developed for operation indoors and may not be used in explosive environments.
■■■■■■■■■■■■■■■■■■■■■■
Place the instrument in a location of the laboratory which is suitable for operation, free of vibrations, protected from corrosive atmosphere, and contamination by chemicals.
The instrument should be protected against excessive temperature fluctu­ations and direct sunlight.
■■■■■■■■
8
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
6.2047.030
6.1464.320
1
2

3.2 Setting up the coulometer cell

3.2.1 Inserting the coulometer cell

3 Installation
Figure 3 Inserting the coulometer cell
Set up the coulometer cell on the magnetic stirrer as follows:
Attach the 6.2047.030 titration vessel holder to the magnetic stirrer.
1
Insert the 6.1464.320 coulometer cell into the titration vessel holder.
2

3.2.2 Preparing the coulometer cell

Filling the adsorber tube
Before setting up the coulometer cell the 6.1403.030 adsorber tube has to be filled with 6.2811.000 molecular sieve. Proceed as follows:
899 Coulometer
■■■■■■■■
9
3.2 Setting up the coulometer cell
1
2
3
4
6.1403.030
6.2811.000
■■■■■■■■■■■■■■■■■■■■■■
Figure 4 Filling the adsorber tube
Insert a small cotton plug into the bottom of the adsorber tube. Do
1
not pack the cotton too tightly.
Fill the molecular sieve up to the ¾ level.
2
Place a small cotton plug on the molecular sieve. Do not pack the
3
cotton too tightly.
Seal the adsorber tube with the appropriate cover.
4
Note
Note that the molecular sieve must be replaced at regular intervals. Each time you refill the adsorber tube with molecular sieve, you can, for example, write the date directly on the adsorber tube.
■■■■■■■■
10
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
6.1903.030
6.1464.320
6.2713.010
6.1437.000
6.0345.100
6.2713.020
6.1403.030
6.1448.020
6.2701.040
6.2713.000
6.0341.100
2
1
2
5
2
3
4
6
6.0344.100
3 Installation
Equipping the coulometer cell
899 Coulometer
Figure 5 Equipping the coulometer cell
Equip the coulometer cell as follows:
Place the 6.1903.030 stirring bar in the coulometer cell.
1
Cut the 6.2713.0x0 ground-joint sleeves to the correct length and
2
attach them to the ground joints of the inserts (electrodes, adsorber tube, etc.).
Take care to ensure that the edges of the ground-joint sleeves are cut to size cleanly and that there are no fringes. The ground-joint sleeves are not permitted to protrude at the lower edge of the ground-joint opening.
Insert the 6.1403.030 adsorber tube into the generator electrode.
3
■■■■■■■■
11
3.2 Setting up the coulometer cell
■■■■■■■■■■■■■■■■■■■■■■
Insert the 6.0345.100 generator electrode without diaphragm or the
4
6.0344.100 generator electrode with diaphragm together with the adsorber tube into the large ground-joint opening at the rear.
Insert the 6.0341.100 indicator electrode into the left ground-joint
5
opening.
Place the 6.1448.020 septum on the front opening of the coulome-
6
ter cell and screw it shut with the 6.2701.040 screw cap.
Tighten the screw cap only enough so that it seals. The septum is not permitted to bend.
Filling the coulometer cell (generator electrode with dia­phragm)
Proceed as follows when using a generator electrode with a diaphragm:
Fill approximately 5 mL of catholyte into the generator electrode.
1
Fill approximately 100 mL of anolyte into the coulometer cell with
2
the aid of the 6.2738.000 funnel. The level of the anolyte should be roughly 1-2 mm above the level of the catholyte.
Close the remaining ground-joint opening on the right with the
3
6.1437.000 ground-joint stopper (with ground-joint sleeve attached).
■■■■■■■■
12
Filling the coulometer cell (generator electrode without diaphragm)
Proceed as follows when using a generator electrode without a dia­phragm:
Fill approximately 100 mL of reagent into the coulometer cell with
1
the aid of the 6.2738.000 funnel.
Close the remaining ground-joint opening on the right with the
2
6.1437.000 ground-joint stopper (with ground-joint sleeve attached).
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
6.1437.000
6.2713.000
6.2713.000
6.1439.010
1
2
4
5 6
3
3 Installation

3.2.3 Mounting the addition and aspiration tube (utilization with Ti Stand)

Figure 6 Mounting the addition and aspiration tube
Insert the addition and aspiration tube as follows into the coulometer cell:
Attach the 6.2713.000 ground-joint sleeve that has been cut to size
1
to the ground joint of the 6.1437.000 stopper.
Insert the stopper into the 6.1439.010 addition and aspiration tube.
2
Attach the 6.2713.000 ground-joint sleeve that has been cut to size
3
to the ground joint of the addition and aspiration tube.
Insert everything together into the ground-joint opening.
4
Connect the tubing for the reagent addition at the upper connector
5
of the addition and aspiration tube (5).
899 Coulometer
■■■■■■■■
13

3.3 Connecting the coulometer to the power supply

Connect the tubing for the aspiration of the coulometer cell at the
6
■■■■■■■■■■■■■■■■■■■■■■
lower connector of the addition and aspiration tube (6).
Details regarding how to connect the addition tubing and the aspiration tubing can be found in the manual for the 803 Ti Stand.

3.2.4 Using the coulometer cell with a Karl Fischer oven

When samples release their water only slowly or only at higher tempera­tures, the oven method is used. The sample is heated in a KF oven (e.g. 860 KF Thermoprep) and the water that is released is guided with a carrier gas into the coulometer cell. You will find recommended parameter set­tings for determinations with a Karl Fischer oven in chapter 10.5, page
97.
A detailed description of setting up the coulometer cell with the KF oven can be found in the respective manual.

3.2.5 Using the coulometer cell with a sample changer

When there is a large number of samples, the determination of the water content can be automated with the aid of a sample changer with oven module (e.g. 885 Compact Oven SC). You will find recommended para­meter settings for determinations with a Karl Fischer oven in chapter 10.5, page 97.
A detailed description of setting up the coulometer cell with the sample changer can be found in the respective manual.
3.3 Connecting the coulometer to the power supply
You can supply the 899 Coulometer with electricity two different ways:
Connect the coulometer directly to the mains supply with the aid of
the power supply unit provided.
Connect the coulometer to the 6.2164.500 Power Box if no stable
mains supply is available.

3.3.1 Connecting the power supply unit

The 899 Coulometer has an external power supply unit for a 24 V power supply (DC). This is connected to the mains connection of the Coulometer.
Warning
An incorrect mains voltage can damage the instrument.
Operate the instrument only with the mains voltage specified for it. Use the supplied power supply unit exclusively.
■■■■■■■■
14
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
Figure 7 Connecting the power supply unit
Proceed as follows:
Connect the plug of the external power supply unit with the mains
1
connection (2-7) of the Coulometer.
3 Installation
Note
The plug of the power supply unit is protected against accidental disconnection of the cable by means of a pull-out protection fea­ture. If you wish to pull out the plug, you will first need to pull back the outer plug sleeve.
Connect the mains cable with the external power supply unit of the
2
Coulometer and with the mains supply.
Caution
Switch off the Coulometer correctly by pressing the red [STOP] key before you interrupt the electricity supply. If this is not done, then there is a danger of data loss.

3.3.2 Connecting the Power Box

As an alternative to the power supply from the mains supply, the option exists of providing the 899 Coulometer with electricity through the
6.2164.500 Power Box. This means that you can also use the instrument in environments in which no stable mains supply is available. Details regarding the Power Box can be found in the respective manual.
899 Coulometer
■■■■■■■■
15
3.3 Connecting the coulometer to the power supply
Figure 8 Connecting the Power Box
■■■■■■■■■■■■■■■■■■■■■■
Proceed as follows:
Connect the plug of the Power Box with the mains connection (2-7)
1
of the Coulometer.
Note
The Power Box plug is protected against accidental disconnection of the cable by means of a "pull-out protection" feature. If you wish to pull out the plug, you will first need to pull back the outer plug sleeve.
Caution
Switch off the Coulometer correctly by pressing the red [STOP] key before you interrupt the connection with the Power Box. If this is not done, then there is a danger of data loss.
■■■■■■■■
16
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
1

3.4 Connecting sensors

The measuring interface contains the following measuring inputs:
Gen. for a generator electrode Ind. for a double Pt electrode Temp. for a temperature sensor of the Pt1000 type
Caution
Under all circumstances, avoid mixing up the electrode cable from the indicator electrode with the one from the generator electrode. Attach corresponding markings on the screw heads of the cables.

3.4.1 Connecting a generator electrode

Screwing the electrode cable to the general electrode
3 Installation
Unscrew the cover of the generator electrode.
1
Figure 9 Unscrewing the cover from the generator electrode
Screw the 6.2104.120 electrode cable tightly onto the generator
2
electrode.
899 Coulometer
■■■■■■■■
17
3.4 Connecting sensors
2
6.2104.120
■■■■■■■■■■■■■■■■■■■■■■
Figure 10 Screwing the electrode cable to the general electrode
Connecting the electrode cable to the coulometer
Plug the electrode plug into the Gen. socket of the Coulometer.
1
Figure 11 Connecting a generator electrode
Note
The electrode cable is protected against accidental disconnection of the cable by means of a pull-out protection. If you wish to pull out the plug again, you will first need to pull back the outer plug sleeve.

3.4.2 Connecting an indicator electrode

Screwing on the electrode cable to the indicator electrode
Unscrew the cover of the indicator electrode.
1
■■■■■■■■
18
899 Coulometer
■■■■■■■■■■■■■■■■■■■■■■
1
2
6.2104.020
3 Installation
Figure 12 Unscrew the cover from the indicator electrode
Screw the 6.2104.020 electrode cable tightly onto the indicator elec-
2
trode.
Figure 13 Screwing on the electrode cable to the indicator elec-
trode
Connecting the electrode cable to the coulometer
Plug the electrode plug into the Ind. socket of the Coulometer.
1
Figure 14 Connecting an indicator electrode
899 Coulometer
■■■■■■■■
19
3.4 Connecting sensors
Note
The electrode cable is protected against accidental disconnection of the cable by means of a pull-out protection. If you wish to pull out the plug again, you will first need to pull back the outer plug sleeve.

3.4.3 Connecting a temperature sensor

A temperature sensor of the Pt1000 type can be connected to the Temp. connector.
Connect the temperature sensor as follows:
Insert the plugs of the temperature sensor into the Temp. sockets of
1
the Coulometer.
■■■■■■■■■■■■■■■■■■■■■■
Figure 15 Connecting a temperature sensor
Note
Always insert the red plug into the red socket. It is only this way that the shielding against electrical interferences is ensured.
■■■■■■■■
20
899 Coulometer
Loading...
+ 148 hidden pages