Metrohm 772 User Manual

772 Pump Unit
Manual
8.772.8001EN
Metrohm AG CH-9100 Herisau Switzerland Phone +41 71 353 85 85 Fax +41 71 353 89 01 info@metrohm.com www.metrohm.com
772 Pump Unit
8.772.8001EN
Manual
07.2012 fpe
Teachware Metrohm AG CH-9100 Herisau teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been checked with great care, errors cannot be entirely excluded. Should you notice any mistakes please send us your comments using the address given above.
Documentation in additional languages can be found on
http://documents.metrohm.com.
■■■■■■■■■■■■■■■■■■■■■■

Table of contents

1 Introduction 1
1.1 Instrument description ......................................................... 1
1.1.1 Model versions ........................................................................ 1
1.1.2 Connectors .............................................................................. 1
1.1.3 Intended use ........................................................................... 1
1.2 About the documentation ................................................... 2
1.2.1 Symbols and conventions ........................................................ 2
1.3 Safety instructions ................................................................ 3
1.3.1 General notes on safety ........................................................... 3
1.3.2 Electrical safety ........................................................................ 3
1.3.3 Tubing and capillary connections ............................................. 4
1.3.4 Flammable solvents and chemicals ........................................... 4
1.3.5 Recycling and disposal ............................................................. 4
2 Overview of the instrument 6
Table of contents
3 Installation 7
3.1 Setting up the instrument .................................................... 7
3.1.1 Packaging ................................................................................ 7
3.1.2 Checks .................................................................................... 7
3.1.3 Location .................................................................................. 7
3.2 Mounting the tubings .......................................................... 7
3.2.1 Mounting the pump tubing ..................................................... 7
3.2.2 Installing the inlet tubing and outlet tubing .............................. 9
3.3 Connecting the peristaltic pump ....................................... 10
3.3.1 Connection to a Sample Processor ......................................... 11
3.3.2 Connection to the 731 Relay Box .......................................... 11
4 Operation 14
5 Operation and maintenance 15
5.1 General notes ...................................................................... 15
5.1.1 Care ...................................................................................... 15
5.1.2 Maintenance by Metrohm Service .......................................... 15
5.2 Pump tubings ...................................................................... 16
5.2.1 Selecting the pump tubing ..................................................... 16
5.2.2 Suitability test for tubing materials ......................................... 17
5.2.3 Tubing dimensions ................................................................. 17
5.2.4 Tubing lifetime ...................................................................... 17
5.2.5 PharMed® pump tubing data sheet ....................................... 18
5.2.6 Viton® pump tubing data sheet ............................................ 19
772 Pump Unit
5.3 Tubings ................................................................................ 20
■■■■■■■■
III
Table of contents
■■■■■■■■■■■■■■■■■■■■■■
5.4 Quality Management and validation with Metrohm ....... 20
6 Troubleshooting 21
6.1 Problems ............................................................................. 21
7 Technical specifications 22
7.1 Peristaltic pump .................................................................. 22
7.2 Power supply ...................................................................... 22
7.3 Interfaces and connectors ................................................. 22
7.4 Safety specifications ........................................................... 22
7.5 Electromagnetic compatibility (EMC) ................................ 23
7.6 Ambient temperature ......................................................... 23
7.7 Reference conditions .......................................................... 23
7.8 Dimensions/material ........................................................... 24
8 Conformity and warranty 25
8.1 Quality Management Principles ........................................ 25
8.2 Warranty (Guarantee) ........................................................ 26
9 Accessories 28
9.1 Scope of delivery per model version ................................. 28
9.1.1 772 Pump Unit ...................................................................... 28
9.1.2 772 Pump Unit with aspiration equipment ............................. 29
9.1.3 772 Pump Unit with rinsing equipment .................................. 32
9.2 Optional accessories ........................................................... 35
9.2.1 772 Pump Unit ...................................................................... 35
9.2.2 772 Pump Unit with aspiration equipment ............................. 36
9.2.3 772 Pump Unit with rinsing equipment .................................. 36
Index 38
■■■■■■■■
IV
772 Pump Unit
■■■■■■■■■■■■■■■■■■■■■■

Table of figures

Figure 1 Front 772 Pump Unit ......................................................................... 6
Figure 2 Inserting the tubing olives ................................................................. 7
Figure 3 Installing the pump tubing ................................................................ 8
Figure 4 Fastening the pressure clamp in place ................................................ 8
Figure 5 Connecting the tubings ..................................................................... 9
Figure 6 Screwing the union nuts tightly ....................................................... 10
Figure 7 772 Pump Unit directly on the Sample Processor ............................. 11
Figure 8 772 Pump Unit via the Remote Box and 731 Relay Box .................... 12
Figure 9 Direction of rotation ........................................................................ 13
Figure 10 Calculation formula for the permeability (C) ..................................... 19
Table of figures
772 Pump Unit
■■■■■■■■
V
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction

1.1 Instrument description

The 772 Pump Unit is a versatile peristaltic pump for conveying or evacu­ating liquid media. It was designed specifically for the production opera­tion and laboratory environment and enables a wide range of applica­tions. Unlike membrane pumps, pumping organic solvents or liquids with solids content is no problem for the 772 Pump Unit.
Therefore, the 772 Pump Unit is equally well suited for use in titration, e.g. for evacuating titration samples with silver halide precipitation and as a delivery pump in larger automation systems. By using the respective suit­able tubing material, a wide variety of tasks can be managed.
The flow rate of up to 600 mL/min opens up a wide range of applications.

1.1.1 Model versions

The 772 Pump Unit is available in the following three versions:

1 Introduction

2.772.0110
2.772.0120 772 Pump Unit with aspiration equip-
2.772.0130 772 Pump Unit with rinsing equipment
Each version includes different accessories according to its use (see Chap- ter 9, page 28).

1.1.2 Connectors

The 772 Pump Unit needs 16 - 24 V DC voltage.
Power is supplied to the pump via the direct connection to a Sample Pro­cessor (e.g. 814 or 815) or to the 731 Relay Box (connection via adapter cable 6.2160.010).

1.1.3 Intended use

The 772 Pump Unit is designed for usage in automated systems in analyti­cal laboratories.
This instrument is suitable for pumping chemicals and flammable samples. Use of the 772 Pump Unit therefore requires the user to have basic knowl­edge and experience in handling toxic and caustic substances. Knowledge with respect to the application of the fire prevention measures prescribed for laboratories is also mandatory.
772 Pump Unit
ment
772 Pump Unit
■■■■■■■■
1

1.2 About the documentation

1.2 About the documentation
Caution
Please read through this documentation carefully before putting the instrument into operation. The documentation contains information and warnings which the user must follow in order to ensure safe opera­tion of the instrument.

1.2.1 Symbols and conventions

The following symbols and formatting may appear in this documentation:
Cross-reference to figure legend
The first number refers to the figure number, the sec­ond to the instrument part in the figure.
■■■■■■■■■■■■■■■■■■■■■■
Instruction step
Carry out these steps in the sequence shown.
Method Dialog text, parameter in the software
File New Menu or menu item
[Next] Button or key
Warning
This symbol draws attention to a possible life hazard or risk of injury.
Warning
This symbol draws attention to a possible hazard due to electrical current.
Warning
This symbol draws attention to a possible hazard due to heat or hot instrument parts.
Warning
This symbol draws attention to a possible biological hazard.
■■■■■■■■
2
Caution
This symbol draws attention to a possible damage of instruments or instrument parts.
772 Pump Unit
■■■■■■■■■■■■■■■■■■■■■■

1.3 Safety instructions

1.3.1 General notes on safety

Warning
This instrument may only be operated in accordance with the specifica­tions in this documentation.
This instrument has left the factory in a flawless state in terms of technical safety. To maintain this state and ensure non-hazardous operation of the instrument, the following instructions must be observed carefully.

1.3.2 Electrical safety

The electrical safety when working with the instrument is ensured as part of the international standard IEC 61010.
1 Introduction
Note
This symbol marks additional information and tips.
Warning
Only personnel qualified by Metrohm are authorized to carry out service work on electronic components.
Warning
Never open the housing of the instrument. The instrument could be damaged by this. There is also a risk of serious injury if live components are touched.
There are no parts inside the housing which can be serviced or replaced by the user.
Mains voltage
Warning
An incorrect mains voltage can damage the instrument.
772 Pump Unit
Only operate this instrument with a mains voltage specified for it (see rear panel of the instrument).
■■■■■■■■
3
1.3 Safety instructions
Protection against electrostatic charges
Warning
Electronic components are sensitive to electrostatic charges and can be destroyed by discharges.
Do not fail to pull the mains cable out of the mains connection socket before you set up or disconnect electrical plug connections at the rear of the instrument.

1.3.3 Tubing and capillary connections

Caution
Leaks in tubing and capillary connections are a safety risk. Tighten all connections well by hand. Avoid applying excessive force to tubing connections. Damaged tubing ends lead to leakage. Appropriate tools can be used to loosen connections.
■■■■■■■■■■■■■■■■■■■■■■
Check the connections regularly for leakage. If the instrument is used mainly in unattended operation, then weekly inspections are manda­tory.

1.3.4 Flammable solvents and chemicals

Warning
All relevant safety measures are to be observed when working with flammable solvents and chemicals.
Set up the instrument in a well-ventilated location (e.g. fume cup-
board).
Keep all sources of flame far from the workplace. Clean up spilled liquids and solids immediately. Follow the safety instructions of the chemical manufacturer.

1.3.5 Recycling and disposal

This product is covered by European Directive 2002/96/EC, WEEE – Waste from Electrical and Electronic Equipment.
■■■■■■■■
4
The correct disposal of your old equipment will help to prevent negative effects on the environment and public health.
772 Pump Unit
■■■■■■■■■■■■■■■■■■■■■■
1 Introduction
More details about the disposal of your old equipment can be obtained from your local authorities, from waste disposal companies or from your local dealer.
772 Pump Unit
■■■■■■■■
5

2 Overview of the instrument

1
2
3
4
5
6
7
■■■■■■■■■■■■■■■■■■■■■■
Figure 1 Front 772 Pump Unit
Status LED
1
For displaying the operating state.
Pressure clamp
3
For the counterpressure to the pump tubing.
Tubing clamp (outlet)
5
Connection cable
7
With M8 plug for connection to a Sample Processor.
Locking lever with clamping screw
2
Rotor
4
Tubing clamp (inlet)
6
■■■■■■■■
6
772 Pump Unit
Loading...
+ 32 hidden pages