Getinge Sterilizer Operation Manual

TROUBLE SHOOTING
EXIT MANUAL
GUIDE
Getinge Sterilizers
Click here to
Click here to
Getinge/Castle, Inc.
1100 Towbin Avenue
Lakewood, NJ 08701
732 370-8800
Door System Trouble Shooting
Automatic Door Sterilizer
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
EXIT MANUAL
Door will not close. A. No air pressure or pres-
sure to low.
B. Door safety activated.
C. Defective pneumatic pilot operators. D. Broken air line.
Door closes slowly. A. Air pressure too low.
B. Air flow to door to low.
C. Door gasket not retracted.
A. Check air supply and verify correct pressure set­ting. B. Check for obstruction to door . C. Replace solenoid.
D. Inspect and replace air line.
A. Re-adjust air pressure to correct setting. B. Make adjustment to flow control valve or to flow control orifice (as equipped on unit). NOTE: To be adjusted only by qualified maintenance personnel. C. Remove gasket, lubri­cate or replace if necessary.
Manual Door Sterilizer
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Manual door will not close or is difficult to close.
REV C 1
A. Cable is worn or broken. B. Door guides.
C. Door gasket not retracted. D. Spring tension
A. Replace cable. B. Lubricate (grease) or readjust guides and eccen­trics. C. Remove gasket, lubri­cate or replace if necessary. D. Adjust spring tension NOTE: To be adjusted only by qualified maintenance personnel.
System Trouble Shooting
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
EXIT MANUAL
Jacket will not heat up whe n power is turned on.
Cycle does not begin when door is closed and cycle start button is pressed.
Door not sealed. A. Door gasket.
A. No steam. B. Steam to jacket valve not operating.
C. Trap for jacket not oper­ating.
A. Door limit switch not closing. B. Door limit switch defec­tive. C. Pneumatic door limit switch defective. D. Jacket not heating to
3oC below setpoint. E. Check input/output con­nector
B. Door seal supply pres­sure (steam or air).
A. Check steam supply. B. Check pneumatic pilot solenoid and valve, replace as needed. C. Clean or replace trap.
A. Check that door in com­pletely closed. B. Replace switch.
C. Replace Switch.
D. Refer to “Jacket will not heat up” problem above. E. Check for secure fit on pins and connectors.
A. Inspect gasket, replace if necessary. B. Check supply and regu­lator setting if equipped.
C. Door seal pressure switch. D. Steam to door seal valve not operating.
Unit not entering steam into cham­ber.
REV C 2
A. Steam supply.
B. Steam strainer.
C. Steam-to-chamber valve.
C. Replace Switch.
D. Check pneumatic pilot solenoid and replace if nec­essary.
A. Check steam supply to chamber and regulator set­ting if equipped. B. Turn off steam supply and check supply strainer. Clean or replace screen as necessary. C. Check pneumatic pilot solenoid and valve, replace if necessary.
EXIT MANUAL
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Unit not reaching sufficient vac­uum level (non-liquids cycle).
A. Water supply.
B. Worn or defective gas­ket. C. Power supply.
D. Vacuum pump.
E. Vacuum Leak. F. Load.
G. Drain line check valve.
H. Defective or worn door gasket
I. Needle valves.
A. Check water supply is in accordance with arrange­ment drawing. (Supply pressure and temperature effect performance). B. Inspect gasket, replace if necessary . C. Check all power to the unit. D. Pump tripped, reset heater. E. Locate leak and repair. F. Load placed over drain. Leave drain clear. G. Clean and replace valve if necessary. H. Check for faulty door gasket, replace if neces­sary. I. Check settings on needle valves (air to v ac pump and water to vac pump).
Chamber temperature not reach­ing programmed exposure temper­ature.
A. Steam pressure.
B. Chamber strainer.
C. Slow exhaust clogged.
D. Load.
A. No steam or low steam pressure to unit. Check that steam supply is in accor­dance with arrangement drawing. B. Check both strainer in chamber and drain strainer. Clean or replace screen as necessary. C. Clean slow exhaust nee­dle valve or orifice, check setting. D. Load is placed over drain in the chamber. Lea ve drain clear.
REV C 3
EXIT MANUAL
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Unit attains exposure temperature but then drops 2oC below setpoint
A. Defective steam to chamber valve.
for 2 minutes.
B. Slow exhaust clogged.
C. Steam quality.
D. Steam pressure.
E. Slow exhaust valve inoperative.
F. High condensate in drain.
Unit runs over temperature. A. Defective or worn door
gasket. B. Defective steam to chamber valve.
C. Calibration
D. Overdrive parameter.
A. Replace valve or pneu­matic pilot solenoid as nec­essary. B. Clean slow exhaust nee­dle valve or orifice. C. Check for “wet” steam being supplied to sterilizer. D. T emporary loss of steam during cycle. E. Replace valve or pneu­matic pilot solenoid as nec­essary. F. Check for high water or condensate around cham­ber drain probe.
A. Replace gasket.
B. Replace valve or pneu­matic pilot solenoid as nec­essary. C. Re-calibrate all temper­ature probes. D. Check setting of any adjustable overdrive parameters.
REV C 4
EXIT MANUAL
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Chamber exhausts slowly or not at all.
A. Clogged chamber or drain strainer.
B. Inoperative fast exhaust valve.
C. Water pressure.
D. Water strainer clogged.
E. Load placed on drain. F. Check Hi-Voltage power.
G. Slow exhaust clogged.
A. Check strainer in cham­ber and drain. Clean and replace screen as neces­sary. B. Check pneumatic pilot solenoid and valve, replace as necessary. C. Check that water pres­sure is within the require­ments specified on the arrangement drawing. D. Clean strainer on incoming water line. Replace screen as neces­sary. E. Re-adjust load. F. Check power, reset heater on starter for vac­uum pump. G. Clean slow exhaust needle valve or orifice.
Dry time will not count. A. Check “Chamber
exhausts slowly” problem above. B. Vacuum leak.
Dry time starts counting, but then stops.
Air not being admitted into cham­ber after dry time.
A. Vacuum leak B. Load.
A. Filter assembly clogged. B. Check valve clogged.
C. Defective air admit valve.
A. Check “Chamber exhausts slowly” actions above. B. Locate leak and repair.
A. Locate leak and repair. B. Load incorrectly placed in chamber, i.e. containers upright in chamber.
A. Replace filter. B. Clean or replace check valve. C. Replace valve or pneu­matic pilot solenoid as nec­essary.
REV C 5
EXIT MANUAL
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Doors will not open at Cycle End. A. Worn or defective door
gasket. B. Defective door seal valve.
C. Check “Chamber exhausts slowly” above
D. Check valve on steam trap for door seal.
Temperature display is false or dis­plays 999.9.
A. Faulty temperature device (RTD). B. Cable for temperature device is loose.
Pressure is high or pressure trans­ducer is faulty.
A. Check “Unit runs over temperature” problem above. B. Faulty pressure trans­ducer. C. Calibration
D. Loose cable for pressure transducer (PT).
A. Relubricate or replace door gasket. B. Check valve and pneu­matic pilot solenoid, replace if necessary. C. Check “Chamber exhausts slowly” actions above. D. Replace check valve.
A. Replace device.
B. Re-insert cable for tem­perature sensor.
A. Check “Unit runs over temperature” actions above. B. Replace transducer.
C. Re-calibrate the pres­sure transducer. D. Reinsert cable, check connection.
Steam vapor being admitted while door is open.
Jacket constantly over tempera­ture.
Water in chamber after unit is turned off for an extended period of time. (ie. overnight)
Water on chamber floor at cycle end.
A. Faulty steam to chamber valve.
A. Defective steam to jacket valve. B. Setpoint
A. Defective steam to chamber valve.
A. Chamber leveling.
B. Check “Chamber exhausts slowly” problem above. C. Jacket heating.
A. Replace valve.
A. Replace valve.
B. Check setpoint value in the program parameters.
A. Replace valve.
A. Check that chamber is lev el and make adjustments accordingly . B. See “Chamber exhausts slowly” actions above.
C. Check jacket heating problem above
REV C 6
EXIT MANUAL
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Steam coming from the door dur­ing cycle.
Excessive steam from drain in ser­vice area.
Goods are wet after textile or uten­sil cycle.
A. Worn or defective door gasket.
A. Cooling water valve faulty.
B. Safety relief valve.
C. Water supply strainer.
D. Drain quench.
A. Dry time too short for load. B. Water line clogged.
C. Vacuum level.
D. Steam quality.
A. Replace door gasket.
A. Check valve seat and pneumatic pilot solenoid, replace as necessary. B. Check safety relief v alve and replace if necessary. C. Clean or replace screen as necessary. D. Check drain quench valve setting, verify it is working properly.
A. Increase dry time.
B. Clean strainer screen or replace as necessary. C. Check vacuum pump performance. D. Check for “wet” steam.
Valves not operating. A. Air supply.
B. Defective pneumatic pilot solenoid.
A. Check that air pressure is within the requirements specified on the arrange­ment drawing. B. Check pneumatic pilot solenoid, replace as neces­sary.
REV C 7
Electrical System Trouble Shooting
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
EXIT MANUAL
Unit does not power on. A. No electrical power.
B. Tripped circuit breaker.
C. Defective power switch.
Start light not illuminated. A. No steam.
B. Doors not closed or sealed. C. Active Alarms
Power failure for more than 10 seconds during cycle.
Printer not operating. A. No electrical power to
Printer does not operate, red light on printer illuminated or flashing.
A. No electrical power. B. Tripped circuit breaker.
printer. B. Loose power cord.
A. No paper in unit. B. Paper not threaded prop­erly . C. Paper feed is jammed.
A. Check electrical source. B. Locate short and reset circuit breaker. C. Replace switch.
A. Check steam supply. B. Check doors and door seal. C. Check control display for active alarms, correct as necessary .
A. Check electrical source. B. Locate short and reset circuit breaker.
A. Check electrical source.
B. Check power cord is securely plugged in.
A. Replace paper. B. Re-thread paper.
C. Remove paper, clean printer feed.
REV C 8
Loading...