Getinge Sterilizer Operation Manual

TROUBLE SHOOTING
EXIT MANUAL
GUIDE
Getinge Sterilizers
Click here to
Click here to
Getinge/Castle, Inc.
1100 Towbin Avenue
Lakewood, NJ 08701
732 370-8800
Door System Trouble Shooting
Automatic Door Sterilizer
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
EXIT MANUAL
Door will not close. A. No air pressure or pres-
sure to low.
B. Door safety activated.
C. Defective pneumatic pilot operators. D. Broken air line.
Door closes slowly. A. Air pressure too low.
B. Air flow to door to low.
C. Door gasket not retracted.
A. Check air supply and verify correct pressure set­ting. B. Check for obstruction to door . C. Replace solenoid.
D. Inspect and replace air line.
A. Re-adjust air pressure to correct setting. B. Make adjustment to flow control valve or to flow control orifice (as equipped on unit). NOTE: To be adjusted only by qualified maintenance personnel. C. Remove gasket, lubri­cate or replace if necessary.
Manual Door Sterilizer
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Manual door will not close or is difficult to close.
REV C 1
A. Cable is worn or broken. B. Door guides.
C. Door gasket not retracted. D. Spring tension
A. Replace cable. B. Lubricate (grease) or readjust guides and eccen­trics. C. Remove gasket, lubri­cate or replace if necessary. D. Adjust spring tension NOTE: To be adjusted only by qualified maintenance personnel.
System Trouble Shooting
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
EXIT MANUAL
Jacket will not heat up whe n power is turned on.
Cycle does not begin when door is closed and cycle start button is pressed.
Door not sealed. A. Door gasket.
A. No steam. B. Steam to jacket valve not operating.
C. Trap for jacket not oper­ating.
A. Door limit switch not closing. B. Door limit switch defec­tive. C. Pneumatic door limit switch defective. D. Jacket not heating to
3oC below setpoint. E. Check input/output con­nector
B. Door seal supply pres­sure (steam or air).
A. Check steam supply. B. Check pneumatic pilot solenoid and valve, replace as needed. C. Clean or replace trap.
A. Check that door in com­pletely closed. B. Replace switch.
C. Replace Switch.
D. Refer to “Jacket will not heat up” problem above. E. Check for secure fit on pins and connectors.
A. Inspect gasket, replace if necessary. B. Check supply and regu­lator setting if equipped.
C. Door seal pressure switch. D. Steam to door seal valve not operating.
Unit not entering steam into cham­ber.
REV C 2
A. Steam supply.
B. Steam strainer.
C. Steam-to-chamber valve.
C. Replace Switch.
D. Check pneumatic pilot solenoid and replace if nec­essary.
A. Check steam supply to chamber and regulator set­ting if equipped. B. Turn off steam supply and check supply strainer. Clean or replace screen as necessary. C. Check pneumatic pilot solenoid and valve, replace if necessary.
Loading...
+ 6 hidden pages