PPC2+ BG0002
Operation and Maintenance Manual
(Use with PPC2+ Operation
and Maintenance Manual)
©2001 DH Instruments, Inc.
PPC2+ BG0002 Operation and Maintenance Manual
High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly and with extreme force. High pressure systems should be assembled and operated only by personnel who have been instructed in proper safety practices.
© 2001 DH Instruments, Inc. All rights reserved.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, Inc. 4765 East Beautiful Lane Phoenix AZ 85044-5318 USA.
DH Instruments makes sincere efforts to ensure accuracy and quality of its’ published materials; however, no warranty, expressed or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual or products described in it. Mention of any product or brand does not constitute an endorsement by DH Instruments of that product or brand. This manual was originally composed in English and was subsequently translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version and other language versions, the English version predominates.
DH Instruments, DH, DHI, PPC2+ and CalTool are trademarks, registered and otherwise, of DH Instruments, Inc.
Swagelok is a registered trademark of the Swagelok Company
Document No. 550124 20010403
Printed in the USA.
©2001 DH Instruments, Inc.
PPC2+ BG0002 Operation and Maintenance Manual
TABLE OF CONTENTS ................................................................................ |
i |
|||
TABLES |
................................................................................................... |
|
|
iii |
FIGURES ................................................................................................. |
|
|
iii |
|
ABOUT THIS MANUAL .............................................................................. |
iv |
|||
1. INTRODUCTION |
.................................................................................... |
1 |
||
1.1 |
PRODUCT OVERVIEW............................................................................................................................. |
1 |
||
1.2 |
SPECIFICATIONS..................................................................................................................................... |
2 |
||
|
1.2.1 |
GENERAL SPECIFICATIONS........................................................................................................................ |
2 |
|
|
1.2.2 |
PRESSURE MEASUREMENT SPECIFICATIONS......................................................................................... |
2 |
|
|
1.2.3 |
PRESSURE ...................................................................................................CONTROL SPECIFICATIONS |
3 |
|
2. INSTALLATION ..................................................................................... |
|
5 |
||
2.1 |
UNPACKING AND ..............................................................................................................INSPECTION |
5 |
||
|
2.1.1 |
REMOVING ..................................................................................................................FROM PACKAGING |
5 |
|
|
2.1.2 |
INSPECTING .............................................................................................................................CONTENTS |
5 |
|
2.2 |
SITE REQUIREMENTS............................................................................................................................. |
6 |
||
2.3 |
INITIAL SETUP ......................................................................................................................................... |
|
6 |
|
|
2.3.1 |
PREPARING ....................................................................................................................FOR OPERATION |
6 |
|
|
2.3.2 |
POWER CONNECTION ................................................................................................................................. |
7 |
|
|
2.3.3 CONNECTING ............................................................................A PRESSURE SUPPLY (SUPPLY PORT) |
7 |
||
|
2.3.4 CONNECTING ..................................................................................A VACUUM PUMP (EXHAUST PORT) |
7 |
||
|
2.3.5 CONNECTING ........................................TO THE DEVICE UNDER TEST (TEST(+) AND TEST(-) PORTS) |
8 |
||
|
|
2.3.5.1 ................................................................................................................ |
INSTALLING THE SPLT |
9 |
|
2.3.6 THE VENT ......................................................................................................CONNECTION (VENT PORT) |
9 |
||
|
2.3.7 THE ATM .........................................................................................REF PASS THROUGH (UNLABELED) |
9 |
||
3. OPERATION........................................................................................ |
|
11 |
||
3.1 |
VENT ....................................................................................................................................................... |
|
|
11 |
3.2 |
CUSTOM CONTROL............................................................................................................................... |
11 |
||
3.3 |
UPPER LIMIT.......................................................................................................................................... |
|
11 |
|
|
3.3.1 |
OVERPRESSURE ....................................................................................................................FUNCTION |
12 |
|
3.4 |
SETUP MENU KEY................................................................................................................................. |
12 |
||
|
3.4.1 |
CONTROLREF............................................................................................................................................. |
12 |
|
3.5 |
SPECIAL MENU .............................................................................................................................KEY |
13 |
||
|
3.5.1 |
INTERNAL, ......................................................................................................................................RESET |
13 |
|
|
3.5.2 |
VALVE ......................................................................................................................................................... |
|
13 |
|
|
3.5.2.1 ....................................................................................................................................... |
BYPASS |
14 |
|
|
3.5.2.2 ..............................................................................................................................TEST(-) VENT |
14 |
|
|
3.5.3 |
INTERNAL, ..........................................................................................................................................ATM |
15 |
|
4. REMOTE OPERATION .......................................................................... |
17 |
|||
4.1 |
OVERVIEW ............................................................................................................................................. |
|
17 |
Page i |
©2001 DH Instruments, Inc. |
PPC2+ BG0002 Operation and Maintenance Manual
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION ............................. |
19 |
||
5.1 |
CALIBRATION OF REFERENCE PRESSURE TRANSDUCERS .......................................................... |
19 |
|
|
5.1.1 |
PRINCIPLE .................................................................................................................................................. |
19 |
|
5.1.2 |
EQUIPMENT REQUIRED ............................................................................................................................ |
20 |
|
5.1.3 |
SET-UP AND PREPARATION ..................................................................................................................... |
20 |
5.2 |
ADJUSTMENT OF ON-BOARD BAROMETER...................................................................................... |
21 |
|
5.3 |
PNEUMATIC CONTROL MODULE CONFIGURATION ......................................................................... |
21 |
|
5.4 |
SUBASSEMBLY DESCRIPTION AND LOCATION................................................................................ |
23 |
6. TROUBLESHOOTING ........................................................................... |
25 |
6.1 OVERVIEW ............................................................................................................................................. |
25 |
©2001 DH Instruments, Inc. |
Page ii |
PPC2+ BG0002 Operation and Maintenance Manual
|
TABLES |
|
Table 1. |
Default Pressure Control Parameters |
........................................................................................3 |
Table 2. |
PPC2+ Packing List ................................................................................................................... |
5 |
Table 3. |
System Default Control Limits.................................................................................................. |
11 |
Table 4. |
Troubleshooting Guide............................................................................................................. |
25 |
Figure 1. Pneumatic Module (Schematic) ............................................................................................... |
23 |
Page iii |
©2001 DH Instruments, Inc. |
PPC2+ BG0002 Operation and Maintenance Manual
This manual is designed to be used in conjunction with the PPC2+ Operation and Maintenance Manual (Document Number 550091). It contains information specific to the BG0002 model of PPC2+ Pressure Controller/Calibrator.
PPC2+ BG0002 is identical in conception and most features to other PPC2+ models. PPC2+ BG0002 is a specialized PPC2+ model designed to optimize very low gauge pressure performance. This manual covers the aspects of PPC2+ BG0002 that differ from other PPC2+ models.
To operate a PPC2+ BG0002, use the PPC2+ Operation and Maintenance Manual (Document Number 550091) and follow the instructions and information. Refer to this manual for details on BG0002 specifications, installation and set up, pneumatic schematic differences, and [SPECIAL], <5Valves> operation. PPC2+ BG0002 is normally used in conjunction with accessory package PK-PPC-BG-DVU (P/N 401699). This package is delivered with its own instruction sheet.
Manual Conventions
(CAUTION) is used in throughout the manual to identify user warnings and cautions.
(NOTE) is used throughout the manual to identify operating and applications advice and additional explanations.
[ ] indicates direct function keys (e.g., [RANGE]).
< > indicates PPC2+ BG0002 screen displays (e.g., <1yes>).
©2001 DH Instruments, Inc. |
Page iv |
PPC2+ BG0002 Operation and Maintenance Manual
PPC2+ BG0002 is a special model of the PPC2+ line of pressure controller/calibrators. PPC2+ BG0002 is identical to other PPC2+ models except for:
1)Use of a special low pressure “bi-directional gauge” reference pressure transducer (RPT).
2)This model has only one RPT. It does not support dual RPTs.
3)Inclusion and control of two additional internal control valves (a TEST(+)/TEST(-) bypass and a TEST(-) vent) with local and remote control capabilities for both.
4)Special recalibration recommendations for the very low pressure, bi-directional gauge RPT.
PPC2+ BG0002 is normally used in combination with accessory package PK-PPC-BG-DVU (P/N 401699) which includes two thermally isolated volumes installed in the test line to improve control stability. This package is delivered with its own instruction sheet.
Once a PPC2+ BG0002 has been put into service, its operation is identical to other PPC2+ models.
This manual is intended to be used as a supplement to the PPC2+ Operation and Maintenance Manual (Document Number 550091). To operate a PPC2+ BG0002, use the PPC2+ Operation and Maintenance Manual and refer to this manual for aspects specific to the BG0002 model. These are primarily the operating specifications, installation and use recommendations, recalibration recommendations.
Page 1 |
©2001 DH Instruments, Inc. |
PPC2+ BG0002 Operation and Maintenance Manual
1.2.1GENERAL SPECIFICATIONS
Specifications below are specific to PPC2+ BG0002. All other specifications are identical to PPC2+ specifications found in the PPC2+ Operation and Maintenance Manual (Document Number 550091).
Weight: 12.8 kg (28.2 lb)
Pressure Ranges: PPC2+ BG0002 has three pressure ranges: H3: -15 to +15 kPa (-2.2 to +2.2 psi)
H2: -10 to +10 kPa (-1.5 to +1.5 psi) H1: -5 to +5 kPa (-0.8 to +0.8 psi)
Pressure Limits: Line Pressure (min/max low side): 80 to 110 kPaa (11.6 to 16 psia) Maximum/Minimum Working Test Pressure: +15/-15 kPa (+2.2/-2.2 psi)
Maximum/Minimum Test Pressure without Damage: +16.5/-16.5 kPa (+2.4/-2.4 psi)
Required Supply Pressure: SUPPLY port: 70 kPa gauge (10 psi)
EXHAUST port: Vacuum of 70 kPaa (10 psia) or less with displacement of at least 90 lm (3 cfm)
1.2.2PRESSURE MEASUREMENT SPECIFICATIONS
All values are ± span of the active range unless otherwise indicated.
Warm Up Time: |
30 minutes recommended for RPT temperature stabilization |
|
Resolution: |
To 1 ppm, user settable by individual range |
|
Temperature Effect: |
Fully compensated with active independent temperature measurement |
|
|
from 10 to 40 °C |
|
|
± 0.005 % maximum temperature effect in normal ambient 15 to 35 °C |
|
|
operating range |
|
Acceleration Affect: |
± 0.02 %/g maximum, worst axis |
|
|
Allows operation at ± 20° from reference plane without significant effect |
|
Precision1: |
± 0.008 % |
|
Stability (w/ Autozero): |
90 day |
1 year |
|
0.003 % |
0.009 % |
Measurement Uncertainty2: |
90 day |
1 year |
|
0.010 % |
0.015 % |
1Measurement Precision: Combined linearity, hysteresis, repeatability of measurements made by the reference pressure transducer.
2Measurement Uncertainty: Maximum deviation of the reference pressure transducer indication from the true value of measured pressure including precision, stability, temperature effect and calibration standard accuracy of ± 0.003 % of reading.
©2001 DH Instruments, Inc. |
Page 2 |
PPC2+ BG0002 Operation and Maintenance Manual
1.2.3PRESSURE CONTROL SPECIFICATIONS
All values are ± span of the active range unless otherwise indicated. Control specifications assume that a PK-PPC-BG-DVU dual volume unit is installed on the TEST lines and that a vacuum is applied to the PPC2+ BG0002 EXHAUST port.
Modes and Ready Static: |
Sets pressure within hold limit and stops active control until hold |
Indication: |
limit is exceeded |
|
Pressure is Ready when within hold limit and stability test is met |
Dynamic: |
Sets pressure within hold limit and continuously adjusts as close as |
|
possible to target |
|
Pressure is Ready when within hold limit |
Control Hold limit, stability limit (optimum values set by default, can be customized Parameters: independently for each measurement range)
Table 1. Default Pressure Control Parameters
|
HOLD LIMIT |
STABILITY LIMIT |
|
|
|
STATIC |
± 0.5 % of the active |
± 0.0025 % FS/sec of the |
MODE |
pressure span |
active pressure span |
|
|
|
DYNAMIC |
± 0.0025 % FS of the |
± 0.0025 % FS of the |
MODE |
active pressure span |
active pressure span |
|
|
|
Maximum Control Precision1:
Normal Test Volume (in addition to system volume):
Control Speed:
Minimum Control Pressure:
Delivered Pressure Uncertainty2:
± 0.001 %
0 to 1 000 cc (250 cc optimal)
System volume includes PPC2+ internal volume and the PK-PPC-BG-DVU in-line volume of 840 cc
Slew Rate (0 to Controller FS) in
Optimal Volume with No Control: 10 to 15 seconds Pressure Setting (Typical Time To
Ready Indication in Dynamic Mode): 30 seconds
None |
|
90 day |
1 year |
0.010 |
0.015 % |
1Maximum Control Precision: Minimum useable hold limit in dynamic control mode.
2Delivered Pressure Uncertainty: Maximum deviation from the true value of pressure applied to the device under test in dynamic control mode with default control limits and assuming measurement uncertainty as in Section 1.2.2.
Page 3 |
©2001 DH Instruments, Inc. |
PPC2+ BG0002 Operation and Maintenance Manual
NOTES
©2001 DH Instruments, Inc. |
Page 4 |