Fluke QED 6 User Manual

4.5 (2)
PN 2204510 September 2007
© 2007 Fluke Corporation, All rights reserved. Printed in USA All product names are trademarks of their respective companies.
QED 6
Defibrillator Analyzer
Users Guide
Warranty and Product Support
Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase. During the war­ranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or conse­quential damages, this limitation of liability may not apply to you. If any pro­vision of this warranty is held invalid or unenforceable by a court or other de­cision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
07/07
Notices
All Rights Reserved
© Copyright 2007, Fluke Biomedical. No part of this publication may be reproduced, transmit­ted, transcribed, stored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physi­cal damage such as bent or broken parts, dents, or scratches.
Technical Support
For application support or answers to technical questions, either email
techservices@flukebiomedical.com
1-425-446-6945.
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.
or call 1-800- 648-7952 or
Standard Terms and Conditions
Refunds and Credits
Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for return or refund. Only products returned within 90 days from the date of original purchase are eligible for refund/credit. In order to receive a partial refund/credit of a product purchase price on a serialized product, the product must not have been damaged by the customer or by the carrier chosen by the customer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and re­salable condition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condition, are not eligible for credit return and will be returned to the cus­tomer. The Return Procedure (see below) must be followed to assure prompt refund/credit.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or missing parts and acces­sories will be applied to all returns.
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we rec­ommend the following guide for repackaging:
Use a double-walled carton of sufficient strength for the weight being shipped. Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive
material around all projecting parts.
Use at least four inches of tightly packed, industry-approved, shock-absorbent
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authoriza­tion (RMA) number, obtained from our Order Entry Group at 1-800-648-7952 or 1-425-446-
6945.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service
In the U.S.A.: Cleveland Calibration Lab Tel: 1-800-850-4606 Email: globalcal@flukebiomedical.com
Everett Calibration Lab Tel: 1-888-993-5853 Email: service.status@fluke.com
In Europe, Middle East, and Africa: Eindhoven Calibration Lab Tel: +31-402-675300 Email: ServiceDesk@fluke.com
In Asia: Everett Calibration Lab Tel: +425-446-6945 Email: service.international@fluke.com
material around the instrument.
, or
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards us­ing accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be re­sponsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorpo­rated in new editions of the publication. No responsibility is assumed by Fluke Biomedi­cal for the use or reliability of software or equipment that is not supplied by Fluke Bio­medical, or by its affiliated dealers.
Manufacturing Location
The QED 6 Defibrillator Analyzer is manufactured in Everett, Washington by Fluke Bio­medical, 6920 Seaway Blvd., Everett, WA, U.S.A.

Table of Contents

Chapter Title Page
1 Introduction and Specifications.............................................. 1-1
Description............................................................................................ 1-3
Unpacking and Inspection..................................................................... 1-4
General Safety Considerations.............................................................. 1-4
Symbols ............................................................................................ 1-4
Warnings and Cautions..................................................................... 1-5
Instrument Familiarization.................................................................... 1-6
Front Panel........................................................................................ 1-8
Back Panel ........................................................................................ 1-9
Upgrading the Analyzer........................................................................ 1-9
Specifications........................................................................................ 1-10
Accessories ........................................................................................... 1-14
2 Operation .................................................................................. 2-1
Introduction .......................................................................................... 2-3
Powering Up......................................................................................... 2-4
Adjusting Display Contrast................................................................... 2-9
Measuring Defibrillator Energy............................................................ 2-9
Evaluating Ability to Fire ..................................................................... 2-12
Defibrillator Pulse Playback ................................................................. 2-14
Viewing Oscilloscope Output............................................................... 2-14
Preparation for Viewing.................................................................... 2-14
Viewing a Test.................................................................................. 2-15
Measuring Synchronization .................................................................. 2-16
Generating Test Waveforms ................................................................. 2-18
Testing High Level Out ........................................................................ 2-20
Measuring Peak Voltage, Current, and Overshoot................................ 2-20
Measuring Charge Time (Models M and H)......................................... 2-21
Pacemaker (Non-Invasive) Testing....................................................... 2-22
Pacemaker Refractory Period Testing................................................... 2-24
Programming an Automatic Test Sequence.......................................... 2-26
Program Selection............................................................................. 2-27
Defib Setting..................................................................................... 2-28
Energy Limits ................................................................................... 2-28
Charge Time ..................................................................................... 2-28
i
QED 6
Users Guide
Sync Time ........................................................................................ 2-29
Peak.................................................................................................. 2-29
Overshoot ......................................................................................... 2-29
ECG Performance............................................................................. 2-30
Pacer................................................................................................. 2-30
Running an Automatic Test Sequence.................................................. 2-31
Printing the Analyzer Report Header ................................................... 2-32
Resetting the Analyzer to Factory Defaults.......................................... 2-35
Remote Operation ................................................................................ 2-36
Preparing for Serial Communications .............................................. 2-38
Ansur Software Control.................................................................... 2-39
3 Maintenance, Service, and Calibration .................................. 3-1
Maintenance ......................................................................................... 3-3
Avoiding Damage............................................................................. 3-3
Cleaning ........................................................................................... 3-3
Troubleshooting ................................................................................... 3-4
Service and Calibration ........................................................................ 3-5
ii

List of Tables

Table Title Page
1-1. Symbols ................................................................................................ 1-4
1-2. Front Panel Elements............................................................................ 1-8
1-3. Available QED 6 Models ...................................................................... 1-9
1-4. Standard Accessories ............................................................................ 1-14
1-5. Optional Accessories ............................................................................ 1-14
2-1. Available Waveforms ........................................................................... 2-10
2-2. Test Waveforms.................................................................................... 2-18
2-3. Autosequencing Defaults (Model H) .................................................... 2-26
2-4. Serial Port Wiring Configuration.......................................................... 2-37
2-5. Serial Cable Wiring Configuration ....................................................... 2-37
iii
QED 6
Users Guide
iv

List of Figures

Figure Title Page
1-1. Analyzer Isometric View ...................................................................... 1-6
1-2. Analyzer Front Panel Layout ................................................................ 1-7
2-1. Analyzer Front Panel Display ............................................................... 2-3
2-2. Main Menu 1 Functions ........................................................................ 2-6
2-3. Main Menu 2 Functions ........................................................................ 2-7
2-4. Autosequence Menu Structure (Main Menu 2)..................................... 2-8
2-5. Defibrillator Energy Testing ................................................................. 2-11
2-6. QEDR Performance Tag....................................................................... 2-12
2-7. ECG Lead Configuration ...................................................................... 2-13
2-8. Sync Time Measurements..................................................................... 2-17
2-9. Connecting Pacemaker Output to Analyzer.......................................... 2-23
2-10. Manual Output with Header.................................................................. 2-33
2-11. Automatic Sequence Output with Header............................................. 2-34
v
QED 6
Users Guide
vi
Chapter 1

Introduction and Specifications

Contents Page
Description................................................................................... 1-3
Unpacking and Inspection ........................................................... 1-4
General Safety Considerations..................................................... 1-4
Symbols.................................................................................... 1-4
Warnings and Cautions ............................................................ 1-5
Instrument Familiarization........................................................... 1-6
Front Panel ............................................................................... 1-8
Back Panel................................................................................ 1-9
Upgrading the Analyzer............................................................... 1-9
Specifications............................................................................... 1-10
Accessories .................................................................................. 1-14
1-1
QED 6
Users Guide
1-2
Introduction and Specifications Description
1

Description

The Fluke Biomedical QED 6 Defibrillator Analyzer, hereafter referred to as the Analyzer, is a highly versatile and portable instrument. Regular testing of defibrillators and pacemakers is critical to ensure safe and effective operation. The Analyzer accurately verifies the output characteristics of all defibrillators and tests the parameters of non-invasive pacemakers. The Analyzer is battery operated and completely portable. Simple-to-use menu softkeys allow quick access to tests.
The Analyzer measures:
The delivered energy in joules (watt-seconds) from a defibrillator by
simulating the human body’s resistance
The flow of current through that simulated resistance. The standard resistance used by the Analyzer is 50 Ω. Defibrillator energy is measured in one of two ranges: 0-100 joules, or 0-1000 joules.
Note
The defibrillator pulse waveform can be replayed via the ECG jacks or paddle plates for viewing on a recorder, or on an oscilloscope for greater detail.
Synchronization time in milliseconds. This parameter is measured by timing the firing delay from either the Q-wave (base) or R-wave (peak) simulated by the Analyzer. The simulated waveform is present at both the ECG jacks and the paddle plates.
The peak voltage and peak current (amps) of the defibrillator pulse. Overshoot voltage and current measurements of the defibrillator pulse are calculated and displayed.
The defibrillator’s charge time (the time it takes for a defibrillator to reach its maximum charge setting).
Waveforms, including ECG, arrhythmias, and performance, help verify monitor and recorder accuracy, and also test the automatic defibrillator’s ability to recognize the maximum charge and fire.
All waveforms are present at the ECG jacks, the paddle plates and scope output. Utilities allow the setting of Serial RS232 communication parameters to download results to a printer or computer. Display contrast can be adjusted to obtain the best view of the LCD display.
1-3
QED 6
Users Guide

Unpacking and Inspection

Use the following checklist when unpacking the Analyzer to check for damage during shipment. If the Analyzer has been damaged, call your Fluke Biomedical representative immediately. If you must return the Analyzer to Fluke for service, follow the procedure given under Packing Instructions.
Perform a visual inspection to ensure the front panel and case are intact.
Check the LCD display to ensure that it is unbroken.
Place the Analyzer on a level surface and power up the instrument. If the
message, WARNING - LOW BATTERY!! appears on the display, replace the battery.

General Safety Considerations

Read the Users Manual before operating the Analyzer.

Symbols

Table 1-1 describes the symbols associated with the Analyzer.
Table 1-1. Symbols
Symbol Description
W Important information; refer to manual.
~
Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
; Conforms to relevant Australian EMC requirements
X Hazardous voltage
P Conforms to European Union directives
IEC Measurement Category I – CAT I equipment designed to
CAT I
1-4
protect against transients in equipment on circuits not directly connected to MAINS. Under no circumstances should the terminals of the Analyzer be connected to any MAINS voltage.
Introduction and Specifications General Safety Considerations

Warnings and Cautions

A Warning identifies hazardous conditions and actions that could cause bodily harm or death.
A Caution identifies conditions and actions that could damage the Analyzer, the equipment under test, or cause permanent loss of data.
XW Warning
To avoid possible electrical shock or personal injury, follow these guidelines:
Use this Analyzer only in the manner specified by the
manufacturer or the protection provided may be impaired.
Do not use the product if it operates abnormally.
Remove all test leads and disconnect the battery
eliminator before replacing the battery.
Do not use the product around explosive gases or in
wet or dusty environments.
Inspect the defibrillator daily. Examine the paddles,
lead wires, and power cord for cracks and frays.
1
If the defibrillator is line powered, be sure that it is
plugged into a grounded receptacle. Do not touch the electrical contact surfaces of the defibrillator paddles.
Grip one paddle handle firmly in each hand. Apply to
the Analyzer plates. Keep the paddles firmly depressed to prevent arcing that can cause injury to the operator and/or damage to the Analyzer or defibrillator.
Do not touch the contact plates on the Analyzer when
the defibrillator paddles are being pressed onto the plates. Do not use any electrical paste or pads when testing a defibrillator with the Analyzer.
1-5
QED 6
Users Guide
W Caution
To avoid damage to the Analyzer or adverse affects on its performance, follow these guidelines:
Do not expose the system to temperature extremes.
Ambient temperatures should remain between 0 °C and 40 °C, with a relative humidity less than 90 %. System performance may be adversely affected if temperatures fluctuate above or below this range.
Clean the Analyzer only by wiping it down with a clean,
lint-free cloth dampened with a mild detergent solution. Do not spray liquid directly on or immerse the unit.

Instrument Familiarization

Figure 1-1 is an isometric illustration of the Analyzer, and the Front Panel Layout is shown in Figure 1-2.
1-6
DEFIB
BATTERY / 9VDC
RA
ECG
RS-232 / 10101
V
C
LL
F
LA
L
RL
N
R
SOFTKEYS
QED-6
DEFIBRILLATOR ANALYZER
STERNUM (-)
PACER
1000 j
100 j
Figure 1-1. Analyzer Isometric View
APEX (+)
fcf013.eps
Introduction and Specifications Instrument Familiarization
2 3 4
1
BATTERY / 9VDC RS-232 / 10101
6
LL
LA
RL
RA
R
L
N
QED-6
DEFIBRILLATOR ANALYZER
F
SOFTKEYS
V C
ECG
DEFIB
PACER
1000 j
100 j
1
5
7
8
STERNUM (-)
APEX (+)
9
fcf017.eps
Figure 1-2. Analyzer Front Panel Layout
1-7
QED 6
Users Guide

Front Panel

The front panel of the Analyzer includes the elements described in Table 1-2.
Table 1-2. Front Panel Elements
Number Element Function
A
B
C
D
E Range switch
F Power switch Enables the Analyzer (I = ON, O = OFF).
G
H Five softkeys
Universal ECG jacks
High Level ECG Banana jack
Common Banana jack
Defib Scope Out Banana jack
LCD display: 24 characters x 2 lines.
Utilize AHA and International color coding, allowing for waveform output to monitor/recorder.
Provides 1 volt peak output of the selected waveform.
Provides ground for the “High Level ECG” and “Defib Scope Out” jacks.
Provides pulse output to an oscilloscope.
Allows for defibrillator settings from 0 to 1000 joules (high), for power below 0-100 joules (low) for increased accuracy, and a PACER range setting for pacer output measurements.
The upper line of the LCD display provides messages and test results, while the bottom line displays menu choices.
Used to select the desired function highlighted on the lower line of the display.
1-8
I
Two nickel­plated Defibrillator Paddle Plates
Available for defibrillator paddle contact. All waveforms are present at the paddle plates simultaneously with the ECG jacks.
Introduction and Specifications Upgrading the Analyzer
1

Back Panel

The Back Panel includes a battery holder that houses a 9-volt alkaline battery, and a dc battery eliminator jack. An RS232 D-9-pin serial port allows communications to a computer, serial printer, or other Fluke test equipment.

Upgrading the Analyzer

A number of pre-configured Analyzer models are available. In addition, older models may be upgraded by contacting the Fluke Biomedical Service Center. Available Analyzer models are listed in Table 1-3.
Table 1-3. Available Analyzer Models
Model Characteristics
QED 6
QED 6M
QED 6H
Base unit. Features output energy, synchronization time, peak measurements, bi-directional RS232.
Features output energy, sync time, peak measurements, overshoot, bi-directional RS232, waveforms, charge time measurements, 28 programmable autosequences.
Output energy, sync time, peak measurements, overshoot, bi­directional RS232, waveforms, charge time measurements, 28 programmable autosequences, pacer output measurements and pacer refractory period measurements.
1-9
QED 6
Users Guide

Specifications

The following are specifications for the Analyzer. Please contact your Fluke Biomedical service representative for more information regarding the device specifications.
General
Display.....................................................................2-line x 24-character LCD
Power ......................................................................One 9 V Alkaline (Duracell MN1604 or
Weight ....................................................................4.54 lb
Dimensions .............................................................26.67 x 24.13 x 10.16 cm Environmental Operating Specs
Storage Temperature ..........................................-25 to 50 °C
Operating Temperature .......................................0 to 40 °C
Maximum Humidity..............................................90 % Relative Humidity
Output Power Measurement
Load Resistance......................................................50 Ω ±1 % non-inductance
Range
1000 J..................................................................0-1000.0 J
100 J....................................................................0-100.0 J
Resolution................................................................0.1 J Max. Vage
1000 J..................................................................5500 V
100 J....................................................................1750 V Max. Current
1000 J..................................................................110 A
100 J....................................................................35 A
Measurement...........................................................1000 J: 66 ±5 V
Trip Levels...............................................................100 J: 20 ±5 V
Pulse Width .............................................................1-50 ms Accuracy
1000 J Range ......................................................±2 % of reading
100-1000 J .......................................................... ±2 Js
100 J Range ........................................................±2 % of reading
supertwist alphanumeric
equivalent); 12 hours continuous operation; low battery indication; 9 V battery eliminator input.
(< 10 µH), 160 W
±0.1 J
1-10
Loading...
+ 50 hidden pages