Brookfield RVDV-II+Pro, LVDV-II+Pro, HBDV-II+Pro, HADV-II+Pro Operating Instructions Manual

SPECIALISTS IN THE
MEAS UREMENT AN D
CONTROL OF VISCOSITY
TEL 508-946-6200 FAX 508-946-6262
NTERNET
http://www.brookeldengineering.com
BROOKFIELD ENGINEERING LABORATORIES, INC.
11 Commerce Boulevard, M iddleboro, M A 02346 USA
with oces in :
Boston • Chicago • London • Stuttgart • Guangzhou
BROOKFIELD DV-II+Pro
Viscometer
Operating Instructions
Manual No. M/03-165 D0410
Brookeld Engineering Labs., Inc. Page 1 Manual No. M/03-165-D0410
TABLE OF CONTENTS
I. INTRODUCTION ......................................................................................................................................... 4
I.1 Components ................................................................................................................................................................................... 5
I.2 Utilities ..............................................................................................................................................................................................7
I.3 Specications .................................................................................................................................................................................7
I.4 Installation .......................................................................................................................................................................................8
I.5 Safety Symbols and Precautions ............................................................................................................................................. 9
I.6 Key Functions ...............................................................................................................................................................................10
I.7 Cleaning .........................................................................................................................................................................................11
II. GETTING STARTED .................................................................................................................................. 12
II.1 Autozero .........................................................................................................................................................................................12
II.2 Spindle Selection .......................................................................................................................................................................13
II.3 Speed Selection, Setting, Running ......................................................................................................................................14
II.4 Display Selection ........................................................................................................................................................................15
II.5 Autorange ......................................................................................................................................................................................16
II.6 Out of Range .................................................................................................................................................................................
II.7 Temperature Display ..................................................................................................................................................................18
II.8 Printing ..........................................................................................................................................................................................19
II.9 External Control Mode ..............................................................................................................................................................
II.10 Making Viscosity Measurements ..........................................................................................................................................20
II.11 Time Modes for Viscosity Measurement ............................................................................................................................21
17
20
III. OPTIONS .................................................................................................................................................... 22
III.1 Introduction to OPTIONS .........................................................................................................................................................22
III.2 Setup ................................................................................................................................................................................................
III.2.1 Temperature Display ....................................................................................................................................................25
III.2.2 Units of Measurement ..................................................................................................................................................26
III.2.3 Motor Speed Set Selection .........................................................................................................................................
III.2.4 Printer Output Port ........................................................................................................................................................28
III.2.5 Data Averaging ...............................................................................................................................................................28
III.3 Time Modes....................................................................................................................................................................................29
III.3.1 Time to Stop .....................................................................................................................................................................29
III.3.2 Time to Torque .................................................................................................................................................................32
III.3.3 Print Time Interval ..........................................................................................................................................................34
III.3.4 PC Program (On/O ) .....................................................................................................................................................35
III.3.5 Download a Program ....................................................................................................................................................36
III.3.6 Run a Program .................................................................................................................................................................37
24
26
IV. DVLOADER SOFTWARE ........................................................................................................................ 39
IV.1 B.E.V.I.S. Overview .......................................................................................................................................................................39
IV.2 Description of B.E.V.I.S. Commands ......................................................................................................................................40
IV.3 Creating a B.E.V.I.S. Program ...................................................................................................................................................41
IV.4 Downloading a B.E.V.I.S. Program .........................................................................................................................................43
IV.5 Example Programs ......................................................................................................................................................................44
V. AUTOMATED DATA GATHERING & ANALYSIS .............................................................................. 46
V.1 WINGATHER32..............................................................................................................................................................................46
V.2 RHEOCALC32 ................................................................................................................................................................................48
Brookeld Engineering Labs., Inc. Page 2 Manual No. M/03-165-D0410
Appendix A - Cone/Plate Viscometer Set-Up .....................................................................................................................50
Appendix B - Viscosity Ranges ...............................................................................................................................................55
Appendix C - Variables in Viscosity Measurements .........................................................................................................61
Appendix D - Spindle and Model Codes ..............................................................................................................................63
Appendix E - Calibration Procedures ....................................................................................................................................66
Appendix F - The Brookeld Guardleg .................................................................................................................................73
Appendix G - Speed Sets ...........................................................................................................................................................75
Appendix H - Communications ...............................................................................................................................................76
Appendix I - Model S Laboratory Stand .............................................................................................................................79
Appendix J - DVE-50A Probe Clip ..........................................................................................................................................81
Appendix K - Fault Diagnosis and Troubleshooting ........................................................................................................82
Appendix L - Instrument Dimensions .................................................................................................................................86
Appendix M - Warranty Repair and Service .........................................................................................................................87
Brookeld Engineering Labs., Inc. Page 3 Manual No. M/03-165-D0410
I. INTRODUCTION
The Brookeld DV-II+Pro Viscometer measures uid viscosity at given shear rates. Viscosity is a measure of a uid’s resistance to ow. You will nd a detailed description of the science of viscosity in the Brookeld publication “More Solutions to Sticky Problems” a copy of which was included with your DV-II+Pro.
The DV-II+Pro offers exceptional versatility in modes of control allowing for traditional standalone operation, automatic operation through programs downloaded from the PC or with complete control by PC using Brookeld Rheocalc32 Software.
The DV-II+Pro can be used as a traditional Brookeld viscometer for collection of single speed viscosity data through the easy to use keypad; just select the spindle and speed and read the value from the display. [see Section II, Getting Started]
The Brookeld DVLoader Software can be used to program the DV-II+Pro to control all aspects of the test and data collection without the need for the operator to monitor the instrument; just start the program and return to the printed test data (printer is optional).
[see Section IV, DVLoader Software]
The Brookeld Rheocalc32 Software will perform all control and data collection func tions of the DV-II+Pro from the PC while also providing a platform for advanced data collection and analysis. [see Section II.9, External Control]
In any of these modes of control, the DV-II+Pro will provide the best in viscosity measurement and control.
-
The principal of operation of the DV-II+Pro is to drive a spindle (which is immersed in the test uid) through a calibrated spring. The viscous drag of the uid against the spindle is measured by the spring deection. Spring deection is measured with a rotary transducer. The measurement range of a DV-II+Pro (in centipoise or milliPascal seconds) is determined by the rotational speed of the spindle, the size and shape of the spindle, the container the spindle is rotating in, and the full scale torque of the calibrated spring.
There are four basic spring torque series offered by Brookeld:
Spring Torque
Model dyne/cm milli Newton/m
LVDV-II+Pro 673.7 0.0673 RVDV-II+Pro 7,187.0 0.7187 HADV-II+Pro 14,374.0 1.4374 HBDV-II+Pro 57,496.0 5.7496
The higher the torque calibration, the higher the measurement range. The measurement range for each torque calibration may be found in Appendix B.
All units of measurement are displayed according to either the CGS system or the SI system.
1. Viscosity appears in units of centipoise (shown as “cP”) or milliPascal-seconds (shown as “mPa•s”) on the DV-II+Pro Viscometer display.
2. Shear Stress appears in units of dynes/square centimeter (“D/cm meter (“N/m
2
”).
2
”) or Newtons/square
3. Shear Rate appears in units of reciprocal seconds (“1/SEC”).
4. Torque appears in units of dyne-centimeters or Newton-meters (shown as percent “%” in both cases) on the DV-II+Pro Viscometer display.
Note: To change CGS to SI units on the display - see Section III.2.2.
Brookeld Engineering Labs., Inc. Page 4 Manual No. M/03-165-D0410
The equivalent units of measurement in the SI system are calculated using the following conversions:
SI CGS
Viscosity: 1 mPa•s = 1 cP
2
Shear Stress: 1 Newton/m
= 10 dyne/cm
Torque: 1 Newton/m = 10
7
dyne/cm
2
References to viscosity throughout this manual are done in CGS units. The DV-II+Pro Viscometer provides equivalent information in SI units.
I.1 Components
Please check to be sure that you have received all components, and that there is no damage. If you are missing any parts, please notify Brookeld Engineering or your local Brookeld agent immediately. Any shipping damage must be reported to the carrier.
Component Part Number Quantity
DV-II+Pro Viscometer varies 1
Model S Laboratory Stand MODEL S 1
Spindle Set with Case varies 1 LVDV-II+Pro set of four spindles SSL or RVDV-II+Pro set of six spindles (#2 - #7) SSR or HA/HBDV-II+Pro set of six spindles (#2 - #7) SSH
For Cone/Plate versions: a spindle wrench, one cone spindle and sample cup,
Part No. CPE-44Y replace the spindle set.
Power Cord 1 DVP-65 for 115 or DVP-65 DVP-66 for 230 DVP-66
RTD Temperature Probe DVP-94Y 1
Guard Leg: 1 LVDV-II+Pro B-20Y RVDV-II+Pro B-21Y
Carrying Case DVE-7Y 1
DVLOADER CD ROM DVLOADER 1
Cable (DV-II+Pro to computer) DVP-80 1
Operating Manual M/03-165 1
Brookeld Engineering Labs., Inc. Page 5 Manual No. M/03-165-D0410
COMPONENT DIAGRAM
Cone/Plate Optio
n
Temperature Probe Optio
n
Spindle Set
Cone Spindle
Wrench
Sample Cup
Temperature Probe Clip
Temperature Probe
Shipping Cap
DV-II Pro Viscometer
Model S Laboratory Stand
Guard Leg
LV Spindle Set shown above
Brookeld Engineering Labs., Inc. Page 6 Manual No. M/03-165-D0410
I.2 Utilities
Input Voltage: 115 VAC or 230 VAC Input Frequency: 50/60 Hz Power Consumption: 30 VA
Power Cord Color Code:
United States Outside United States
Hot (live) Black Brown
Neutral White Blue
Ground (earth) Green Green/Yellow
Main supply voltage uctuations are not to exceed ±10% of the nominal supply voltage.
I.3 Specications
Speeds: Choice of 3 options.
Instrument has “Interleaved” speeds when manufactured.
Standalone Interleaved: LV/RV (18 speeds) 8 LV speeds followed by
Sequential: LV/RV (18 speeds) 8 LV speeds and 10 RV
Custom: 54 speeds, user selectable
10 RV speeds
speeds arranged in sequential order from lowest 0.3 rpm to highest 100 rpm.
External (PC Control) .01 - 200 rpm
0.01 rpm increments from 0.01 to 0.99 rpm
0.1 rpm increments from 1.0 to 200 rpm
Note: Refer to Appendix G for detailed list of all speeds.
Weight: Gross Weight 23 lbs. 10.5 kg.
Net Weight 20 lbs. 9 kg. Carton Volume 1.65 cu. ft. 0.05 m
Temperature Sensing Range: -100˚C to 300˚C (-148˚F to 572˚F)
Analog Torque Output: 0 - 1 Volt DC (0 - 100% Torque)
Analog Temperature Output: 0 - 3.75 Volts DC (-100°C to +275°C)
RS232 Compatible Serial Port for use with an attached printer or PC.
Centronics Compatible Parallel Port for use with an attached printer.
Viscosity Accuracy: ±1.0% of full scale range
3
The use of accessory items will have an effect on the measurement accuracy. See Appendix B.
Viscosity Repeatability: ±0.2%
Brookeld Engineering Labs., Inc. Page 7 Manual No. M/03-165-D0410
Temperature Accuracy: ±1°C | -100°C to +149°C
±2°C | +150°C to +300°C
Operating Environment: 0°C to 40°C temperature range (32°F to 104°F)
20% - 80%R.H.: non-condensing atmosphere
Ball Bearing Option:
If you ordered the ball bearing suspension system with your new instrument please note the following:
1) The ball bearing suspension in your Brookeld instrument is noted on the serial tag on the back of the head by the letter “B” after the mode.
2) When attaching and detaching the spindle, it is not necessary to lift the coupling where the spindle connects to the instrument.
3) The Oscillation Check explained in the Appendix under Fault Diagnosis and Troubleshooting does not pertain to this instrument.
Electrical Certications:
Conforms to CE Standards:
BSEN 61326: Electrical equipment for measurement, control and laboratory use -
EMC requirements
BSEN 61010-1: Safety requirements for electrical equipment, for measurement, control
and laboratory use
Notice to customers:
This symbol indicates that this product is to be recycled at an appropriate collection center.
Users within the European Union:
Please contact your dealer or the local authorities in charge of waste management on how to dispose of this product properly. All Brookeld ofces and our network of representatives and dealers can be found on our website: www.brookeldengineering.com
Users outside of the European Union:
Please dispose of this product according to your local laws.
I.4 Installation
Note: “IQ, OQ, PQ”, a guideline document for installation, operation and performance validation
for your DV-II+Pro digital viscometer can be downloaded from our web site www. brookeldengineering.com.
1) Assemble the Model S Laboratory Stand (refer to assembly instructions in Appendix I).
Brookeld Engineering Labs., Inc. Page 8 Manual No. M/03-165-D0410
2) Put the viscometer on the stand.
3) Connect the RTD probe to the socket on the rear panel of the DV-II
+Pro.
4) The Viscometer must be leveled. The level is adjusted using the two leveling screws on the base. Adjust so that the bubble level on top of the DV-II+Pro is centered within the
circle.
Note: Check level periodically during use.
5) Remove the white shipping cap which secures lower coupling nut on Viscometer to pivot
cup.
6) Make sure that the AC power switch at the rear of the DV-II
+Pro is in the OFF position.
Connect the power cord to the socket on the back panel of the instrument and plug it into the appropriate AC line.
The AC input voltage and frequency must be within the appropriate range as shown on the nameplate of the viscometer.
Note: The DV-II+Pro must be earth grounded to ensure against electronic failure!!
7) Turn the power switch to the ON position and allow to warm up for 10 minutes before performing autozero.
8) For Cone/Plate models, refer to Appendix A.
9) If appropriate, connect interconnecting cable (DVP-80) to serial port for connection of DV-II+Pro to PC or printer.
10) If appropriate, connect interconnecting cable to parallel port for connection of DV-II+Pro
to printer.
11) If appropriate, connect interconnecting cable (DVP-96Y) to analog (serial) port for con nection of DV-II+Pro to chart recorder.
I.5 Safety Symbols and Precautions
Safety Symbols
The following explains safety symbols which may be found in this operating manual.
Indicates hazardous voltages may be present.
Refer to the manual for specic warning or caution information to avoid personal injury or damage to the instrument.
Precautions
If this instrument is used in a manner not specied by the manufacturer, the protection provided by the instrument may be impaired.
-
This instrument is not intended for use in a potentially hazardous environment.
Brookeld Engineering Labs., Inc. Page 9 Manual No. M/03-165-D0410
In case of emergency, turn off the instrument and then disconnect the electrical
DV- II+ P ro
VI SC OME TE R
B ROOKFIELD
PRINT
SELEC T
SPIND LE
SELEC T
DISPL AY
SET
SPEED
ENTER
AUTO
RANGE
MOTOR ON/OF F
ESCAP E
£
§
OPTIO NS
TAB
ô
ö
cord from the wall outlet.
The user should ensure that the substances placed under test do not release poisonous, toxic or ammable gases at the temperatures which they are subjected to during the testing.
I.6 Key Functions
Figure I-1 shows the control keys on the face of the DV-II+Pro Viscometer. The following describes the function of each key.
£
This key is used to scroll UP (in an increasing
UP ARROW
value direction) through the available speed, spindle and Option menu tables.
§
This key is used to scroll DOWN (in a
DOWN ARROW
decreasing value direction) through the available speed, spindle and option menu tables.
MOTOR ON/OFF
MOTOR ON/OFF/ESCAPE
ESCAPE
Turns the motor ON or OFF. ESCAPE exits the
Options menu.
SET
SPEED
Causes the DV-II+Pro to begin running at the
SET SPEED
currently selected speed. This function works only when the motor is ON. Also used to select custom speeds when in the Custom Speed
option.
Figure I-1
Selects the data parameter to be displayed:
SELECT
SELECT DISPLAY
DISPLAY
cP Viscosity (cP or mPa•s)
SS Shear Stress (dynes/cm2 or Newtons/m2)
ENTER
AUTO
ENTER/AUTO RANGE
RANGE
SR Shear Rate (1/sec)
ENTER: Used to execute the currently ashing option. AUTO RANGE: Presents the maximum (100% torque) viscosity attainable using the
Brookeld Engineering Labs., Inc. Page 10 Manual No. M/03-165-D0410
selected spindle at the current viscometer spindle speed.
SELECT
SPINDLE
SELECT SPINDLE
Initiates spindle selection on the rst press and then selects the currently scrolled-to
spindle when pressed a second time.
Selects printing and non-printing modes when a printer is attached.
OPTIONS: Presents the Options menu, ashing the last escaped option.
PRINT
OPTION
TAB
PRINT
OPTIONS/TAB
F
TAB: Toggles between selectable items when indicated, as shown in
Figure I-2.
Note: Symbol indicating
the OPTIONS/TAB key
f
°F(FAHRENHEIT) $
g f
CGS UNITS #
g
Figure I-2
Note: Inverted text (black background with white lettering) indicates that the
information is ashing on the viscometer display.
I.7 Cleaning
Be sure to remove the spindle from the instrument prior to cleaning. Severe instrument damage may result if the spindle is cleaned in place.
Instrument and Keypad: Clean with a dry, non-abrasive cloth. Do not use solvents
or cleaners.
Immersed Components (spindles): Spindles are made of stainless steel. Clean with a non-
abrasive cloth and solvent appropriate for sample mate-
rial.
Note: When cleaning, take care not to apply excessive force - it may bend the
spindles.
Brookeld Engineering Labs., Inc. Page 11 Manual No. M/03-165-D0410
II. GETTING STARTED
II.1 Autozero
Before readings may be taken, the Viscometer must be Autozeroed. This action is performed each time the power switch is turned on. (Note: If cable DVP-80 is connected for printer or computer communication see section II.9). The display window on the Viscometer will guide you through the procedure as follows:
Turn the power switch (located on the rear panel) to the ON position. This will result in the screen display shown in Figure II-1 indicating that the DV-II+Pro viscometer is in the standalone mode (is not connected to a computer).
BROOKFIELD DV-2+ PRO VISCOMETER
Figure II-1
After a few seconds, the following screen appears indicating the version of the operating rmware (the built in program which controls the instrument) and a two-digit alphanumeric code which indicates the Model number (see Table D2 in Appendix D; the code tells the spring torque rating or the viscosity measurement range of your viscometer). For most DV-II+Pro Viscometers, this information will be either “LV”, “RV” or “HB”:
BROOKFIELD DV-2+ RV V6.3
Figure II-2
No key press is required at this point. After a short time, the display will clear and the following will be displayed:
REMOVE SPINDLE PRESS ANY KEY
Figure II-3
After removing the spindle and pressing any key, the DV-II+Pro begins its Autozero. The screen will ash “Autozeroing.”
After approximately 15 seconds, the display shows the screen in Figure II-4:
REPLACE SPINDLE PRESS ANY KEY
Brookeld Engineering Labs., Inc. Page 12 Manual No. M/03-165-D0410
Figure II-4
Pressing any key at this point results in the display of the DV-II+Pro default screen:
CP 0.0 20.1C OFFRPM % 0.0
Figure II-5
The display will vary depending upon the selection of temperature (°F or °C) and units of viscosity
(cP or mPa•s).
II.2
SELECT
Spindle Selection
SPINDLE
LVDV-II+Pro Viscometers are provided with a set of four spindles and a narrow guardleg; RVDV-II+Pro Viscometers come with a set of six spindles and a wider guardleg; HADV­II+Pro and HBDV-II+Pro Viscometers come with a set of six spindles and no guardleg. (See Appendix F for more information on the guardleg.)
The spindles are attached to the viscometer by screwing them onto the lower shaft. Note that the spindles have a left-hand thread. The lower shaft should be secured and slightly lifted with one hand while screwing the spindle to the left. The face of the spindle nut and the matching surface on the lower shaft should be smooth and clean to prevent eccentric rotation of the spindle. Spindles can be identied by the number on the side of the spindle coupling nut.
The DV-II+Pro must have a Spindle Entry Code number to calculate Viscosity, Shear Rate and Shear Stress values. The DV-II+Pro memory contains parameters for all standard Brookeld spindles including custom spindles and the two digit entry code for each spindle (the complete list of entry codes may be found in Appendix D).
Note: The DV-II+Pro will remember the Spindle Entry Code which was in use when the
power was turned off.
Pressing the SELECT SPINDLE key will temporarily display the current selected spindle code in place of temperature and cause the character S to begin to blink . It will blink for about three seconds. If the UP or DOWN ARROW keys are pressed (while S is blinking) the two character spindle value to the right of the S character will begin to change (in either an increasing or decreasing direction depending upon which ARROW key is pressed) for each press of the key. If the ARROW key is pressed and held, the display will scroll through the spindle codes for as long as the ARROW key is depressed. When it reaches the last item in the list (either at the top or bottom of the list) the spindle code displayed will “roll-over” to either the rst or last spindle code and the scroll action will continue.
When the desired spindle code is displayed, release the ARROW key to halt further scrolling. Press the SELECT SPINDLE key once again. This will cause the S character to cease blinking and the new spindle code will be accepted for use in viscometer calculations. After 3 seconds the current spindle code will be replaced by the temperature display.
Note: You have approximately three seconds in which to press the SELECT SPINDLE
key before the blinking stops. If you fail to press the SELECT SPINDLE key before the blinking stops you will have to repeat the above steps and re-select the desired spindle.
Brookeld Engineering Labs., Inc. Page 13 Manual No. M/03-165-D0410
The DV-II+Pro will begin to calculate using the new spindle parameters as soon as the SELECT
SPINDLE key is pressed the second time.
Note: The number 99 spindle is for use with special spindles when using Brookeld’s
RHEOCALC32 computer program. Refer to the RHEOCALC32 operator manual for further information on using “99” spindles.
The DV-II+Pro may also be programmed at Brookeld Engineering for “special” user spindles. These “special” spindles will appear on the spindle scroll list starting with designation “AA” and continuing through “AZ”. Contact Brookeld Engineering regarding your needs for special spindles.
II.3
SET
Speed Selection, Setting, Running
SPEED
There are 54 speeds programmed into the DV-II+Pro. These speeds correspond to the standard LVT, RVT, HAT and HBT dial models (18 possible speeds altogether) plus 36 additional speeds.
The DV-II+Pro comes with the Sequential Speed Set already selected (see Appendix G). The speed set will start at speed 0.0. It will then scroll up through the LV speeds, pass through speed 0.0 again, and then scroll up through the RV speeds, pass through speed 0.0 again and then repeat the above sequence.
The DV-II+Pro can also be congured by the operator to interleave the LV and RV speeds. See Section III.2.3 on Setup for a description of how to install the Interleave Speed Set.
A complete list of speed sets and custom speeds is included in Appendix G. The DV-II+Pro can be programmed to select up to 19 of the 54 speeds for use at any one time. Speed 0.0 is the 20th speed and is automatically included. See Section III. 2.3.2 on Setup for a description of how to install a Custom Speed Set.
To select a Viscometer speed rst press either the UP or DOWN arrow keys which will cause the area to the right of RPM to display the currently selected speed. Figure II-6 shows the DV­II+Pro is operating at 6.0 RPM, and the current selected speed is 6.0 RPM.
If the ARROW key is pressed just once and then released, the characters “RPM” will blink for three seconds, then will cease blinking resulting in no change to the speed entry.
Note: The speed selection process remembers the last value of scrolled-to speed so that the
next time you initiate a speed change (by pressing an ARROW key), the DV-II+Pro will begin its scroll display from the last entered value.
The last-scrolled-to speed does not necessarily have to be the same as the speed at which the DV-II+Pro is currently running. The user may operate at a given speed and pre-set the DV­II+Pro to the next desired speed before that speed will be used. For example, if the DV-II+Pro is currently running at 6.0 RPM and was previously scrolled to 12 RPM, a single press of either
Brookeld Engineering Labs., Inc. Page 14 Manual No. M/03-165-D0410
CP 123.4 20.1C
6.0RPM6.0 % 15.6
Figure II-6
ARROW key would result in the Figure II-7 screen display:
cP 123.4 20.1C
6.0RPM12 % 15.6
Figure II-7
Pressing the SET SPEED key would cause the DV-II+Pro to begin running at 12 RPM.
If the user did not press the SET SPEED key, the DV-II+Pro would continue to run at its current speed of 6 RPM. In fact, you may scroll to a new speed (12 RPM in this example) and press the SET SPEED key at any future time (without further pressing an ARROW key) to immediately cause the DV-II+Pro to run at the new speed. Pressing the ARROW key at any time reminds the operator of what was selected for the next speed.
If an ARROW key is pressed and held the DV-II+Pro will scroll up (or down) through the speed table. When it reaches the last speed in the list (either at the top or bottom of the list) the speed displayed will “roll-over” to either the rst or last speed in the table and the scroll action will continue.
When the required speed is displayed, release the ARROW key to halt further scrolling. You have approximately two seconds (before the blinking RPM stops) in which to press the SET SPEED key to immediately begin rotation at the new speed.
Pressing the MOTOR ON/OFF/ESCAPE key stops the Viscometer spindle rotation. Pressing this key sets the DV-II+Pro to 0.0 RPM and causes the screen display to change as shown in Figure II-8:
cP 0.0 20.1C OFFRPM % 0.0
Figure II-8
Pressing the MOTOR ON/OFF/ESCAPE key again immediately starts the DV-II+Pro running at the last scrolled-to-speed. If you had been running at 12 RPM, pressed MOTOR ON/OFF/ ESCAPE and then re-started the DV-II+Pro by pressing MOTOR ON/OFF/ESCAPE once again, you would again be running at 12 RPM. However, if while the motor was off you had scrolled to a new speed of 0.5 RPM, pressing the MOTOR ON/OFF/ESCAPE key would start the DV-II+Pro running at 0.5 RPM.
Note: During both spindle or speed selection and scrolling operations, the DV-II+Pro
will continue to calculate and display Viscometer data as selected.
II.4
SELECT
Display Selection
DISPLAY
Viscosity (displayed in units of cP or mPa•s), Shear Stress and Shear Rate are displayed on the left
side of the top line. You may “step” through the three display options by pressing the SELECT DISPLAY key. For example, the DV-II+Pro is currently displaying Viscosity in Figure II-9:
Brookeld Engineering Labs., Inc. Page 15 Manual No. M/03-165-D0410
cP 123e3 20.1C
6.0RPM % 15.6
Figure II-9
If the viscosity value exceeds 99,999 scientic notation is used. In Figure II-9, the viscosity value is 123,000 cP.
The rst press of the SELECT DISPLAY key would display Shear Stress (SS) in Dynes/cm2 (or Newtons/m2), see Figure II-10:
SS 29.0 20.1C
6.0RPM % 15.6
Figure II-10
If the shear stress value exceeds 99,999, scientic notation is used.
The next press of the SELECT DISPLAY key would display Shear Rate (SR) in 1/Sec (Figure II-11).
SR 40.0 20.1C
6.0RPM % 15.6
Figure II-11
One more press of the SELECT DISPLAY key would result in a return to the viscosity screen, as shown in Figure II-9.
Notes:
1. You may step through the display at any time. This will not interrupt any Viscometer
calculations that are in progress.
2. Display of shear rate and shear stress requires selection of appropriate spindles. Otherwise, values displayed will be zero (0).
Units of Measurement
The DV-II+Pro Viscometer can be congured using the SETUP option (Section III.2.2) to
display/print in either the CGS or SI system of units.
ENTER
AUTO
RANGE
Autorange
II.5
The ENTER/AUTO RANGE key functions as auto range and allows you to determine the maximum calculated viscosity (full scale reading) possible with the current spindle/speed setting only when in the default screen. Pressing the key at any time will cause the current viscosity display to change and show that maximum viscosity. The screen area displaying % (torque) will now display a ashing “%100.0” to indicate this special condition. This maximum viscosity and ashing
Brookeld Engineering Labs., Inc. Page 16 Manual No. M/03-165-D0410
%100.0 value will be displayed for as long as the ENTER/AUTO RANGE key is depressed. Figure II-12 shows the AUTO RANGE function for the situation where the No. 1 LV spindle is rotating at 60 rpm. The full scale range is 100.0 cP (or 100.0 mPa
.
s).
cP 100.0 S61 60 RPM % 100
Figure II-12
Notes:
1. If the RPM is 0.0, the maximum viscosity displayed will be 0.0 cP (or 0.0 mPa.s).
2. While the Viscometer is in the Auto Range mode, any data sent to an attached printer or computer reects the displayed values (i.e. Auto Range values).
3. This function is only available when in the default screen.
II.6 Out of Range
The DV-II+Pro gives indications for out-of-range operation. When % (Torque) readings exceed
100% (over-range), the display changes to that shown in Figure II-13; EEEE will also appear in the display for % and viscosity or shear stress:
cP EEEE 20.1C 10 RPM % EEEE
Figure II-13
You must change either speed or spindle to correct this condition. If you operate at spindle speeds that produce % (Torque) below 10.0 %, the DV-II+Pro ashes the % (Torque), cP (Viscosity), SS (Shear Stress) and SR (Shear Rate) as shown in Figure II-14:
cP 12.4 20.1C 10 RPM % 8.2
Figure II-14
Negative % (Torque) will be displayed as shown in Figure II-15:
cP ---- 20.1C 10 RPM % -2.2
Figure II-15
Brookeld Engineering Labs., Inc. Page 17 Manual No. M/03-165-D0410
Figure II-16 is an example of the printed output of each of the above conditions.
Normal Operation:
RPM=50 M=RV S=29 %=51.4 cP=10280 D/CM2=1285 1/SEC=12.3 T=20.1C Z00:30
Over-Range Operation (>100% torque) (see Fig. 15):
RPM=50 M=RV S=29 %=EEEE cP=EEEE D/CM2=EEEE 1/SEC=12.3 T=20.1C Z00:30
Under-Range Operation (<10% torque) (see Fig. 16):
?RPM=50 M=RV S=29 %=5.2 cP=1040 D/CM2=130 1/SEC=12.3 T=20.1C Z00:30
Negative Torque Operation (see Fig. 17):
RPM=50 M=RV S=29 %=-0.1 cP=---- D/CM2=---- 1/SEC=12.3 T=20.1C Z00:30
M = Torque Range T = Temperature Z = Time
Figure II-16
II.7 Temperature Display
The DV-II+Pro displays the temperature measured by its RTD temperature probe. Temperature
may be displayed in either ˚C (Centigrade) or ˚F (Fahrenheit) units, depending upon selection from the Options menu. As received, the default temperature display will be in ˚C (Centigrade) units as shown in the Figure II-17:
CP 123.4 20.1C 10 RPM % 19.7
Figure II-17
If you turn on the DV-II+Pro with the temperature probe disconnected, or remove the temperature
probe at any point after power-up, the display will indicate “- - - -C”. The four “dashes” indicate the absence of the probe. If you were displaying temperature in Fahrenheit units the C would be replaced by an F. Accuracy of temperature measurement for the DV-II+Pro is shown in Table 1.
Table 1
Temperature Accuracies for
DV-II+Pro Viscometer
Temperature Range Temperature Accuracy
-100°C to +149°C ±1.0°C
Brookeld Engineering Labs., Inc. Page 18 Manual No. M/03-165-D0410
+150°C to +300°C ±2.0°C
II.8
For the case of CGS units with non-exponential results:
and CGS units with exponential results.
Similarly, for SI units with non-exponential results.
and SI units with exponential results.
M = Torque T = Temperature Z = Time
1 10 20 30 40 50 60 70 80
RPM=XXX M=XXXXX S=XX %=XXX.X cP=XXXXX D/CM2=XXXXX 1/SEC=XXXXX T=XX.XC ZXX=XX
RPM=XXX M=XXXXX S=XX %=XXX.X cP=XXXeX D/CM2=XXXeX 1/SEC=XXXXX T=XX.XC ZXX=XX
1 10 20 30 40 50 60 70 80
RPM=XXX M=XXXXX S=XX %=XXX.X mPas=XXXXX N/M2=XXXXX 1/SEC=XXXXX T=XX.XC ZXX=XX
1 10 20 30 40 50 60 70 80
1 10 20 30 40 50 60 70 80
RPM=XXX M=XXXXX S=XX %=XXX.X mPas=XXXeX N/M2=XXXeX 1/SEC=XXXXX T=XX.XC ZXX=XX
PRINT
Printing
The DV-II
+Pro will print data to an attached Serial (RS232) or Parallel (centronics) printer. The
printer must be attached to the appropriate rear panel output connector. See Appendix G for conguration and connection requirements.
Data may be printed in two ways:
1. Pressing the
PRINT key once (for less than three (3) seconds) will result in the printing of one
standard print line.
2. If the
PRINT key is pressed and held for more than three (3) seconds, the DV-II+Pro will then
begin continuous printer output at a print rate interval selected via the Options menu (see Section III.4). The display will show a ashing P in front of the % sign. See Figure II-18.
cP 123.4 20.1C 10 RPM P% 19.7
Figure II-18
To stop continuous printing, press the PRINT key for one (1) second. The ashing P will
disappear on the viscometer display.
Figure II-19 is an example of the print strings for CGS and SI units.
When printing via the parallel port, please note that if a printer is not attached to the viscometer,
the following screen appears:
Brookeld Engineering Labs., Inc. Page 19 Manual No. M/03-165-D0410
Figure II-19
PRINTERERROR
CHECKCONNECTION
Figure II-20
II.9 External Control Mode
The DV-II+Pro Viscometer can be used in conjunction with Brookeld software, Rheocalc32
(V2.4 or higher). Through Rheocalc32, all viscometer functions are controlled by the computer. The DV-II+Pro must be set to the external control mode to allow for proper communication with Rheocalc32. To congure the external control mode, connect cable DVP-80 to the serial port on the DV-II+Pro before turning on the DV-II+Pro. With the DVP-80 cable in place, the DV-II+Pro will present the screen shown in Figure II-21 when it is turned on. If external control is selected, the DV-II+Pro will display Figure II-22 and only accept control commands from Rheocalc32 software.
hEXTERNAL MODE iSTANDALONE MODE
Figure II-21
V6.3 RV EXTERNAL MODE
Figure II-22
The DV-II+Pro may be set to stand alone mode by turning it OFF and ON again and selecting
“Stand Alone” or by removing the DVP-80 cable prior to turning the DV-II+Pro on.
Note: The DV-II+Pro cannot communicate with DVLOADER software in the external
control mode. Choose “Stand Alone” when presented with Figure II-21 if you want to use DVLOADER.
For information on controlling the DV-II+Pro from Rheocalc32 software, check the HELP menu within Rheocalc32.
II.10 Making Viscosity Measurements
The following general procedure is used for making viscosity measurements. Brookeld
recommends the use of a 600 ml Low Form Grifn beaker (Brookeld Part No. BKR-600ml) when using LV/RV/HA/HB spindles.
1. Mount the guardleg on the DV-II+Pro
Viscometer (LV and RV series) and insert into the
container.
2. Insert and center spindle in the test material until the uid’s level is at the immersion groove on the spindle’s shaft. With a disc-type spindle, it is necessary to tilt the spindle slightly while immersing to avoid trapping air bubbles on its surface. Attach the spindle to the lower shaft of the viscometer. Lift the shaft slightly, holding it rmly with one hand while screwing the spindle on with the other (note left-hand thread). Avoid putting side thrust on the shaft. Verify the proper spindle immersion depth and that the viscometer is level.
Brookeld Engineering Labs., Inc. Page 20 Manual No. M/03-165-D0410
3. The process of selecting a spindle and speed for an unknown uid is normally trial and error. An
appropriate selection will result in measurements made between 10-100 on the instrument % torque scale.
Two general rules will help in the trial and error process.
1) Viscosity range is inversely proportional to the size of the spindle.
2) Viscosity range is inversely proportional to the rotational speed.
To measure high viscosity, choose a small spindle and/or a slow speed. If the chosen spindle/
speed results in a reading above 100%, then reduce the speed or choose a smaller spindle.
Experimentation may reveal that several spindle/speed combinations will produce satisfac
tory results between 10-100%. When this circumstance occurs, any of the spindles may be selected.
Non-Newtonian uid behavior can result in the measured viscosity changing if the spindle
and/or speed is changed. See our publication, “More Solutions to Sticky Problems,” for more detail.
Turn on motor.
Allow time for the indicated reading to stabilize. The time required for stabilization will depend
on the speed at which the Viscometer is running and the characteristics of the sample uid. For maximum accuracy, readings below 10% should be avoided.
Record values.
4. Press the
MOTOR ON/OFF/ESCAPE key and turn the motor “OFF” when changing a spindle or
changing samples. Remove spindle before cleaning.
5. Interpretation of results and the instrument’s use with non-Newtonian and thixotropic materials is discussed in the booklet, “More Solutions to Sticky Problems”, and in Appendix C, Variables in Viscosity Measurements.
II.11 Time Modes for Viscosity Measurement
-
The Time Modes
allow the viscometer user to implement the unattended Time to Stop and Time to Torque capabilities of the DV-II+Pro Viscometer. These features will allow the user to set up the viscometer (i.e. select spindle and speed) and then record readings for a xed period of time (Time to Stop) or until a set torque value is attained (Time to Torque). When timing begins, a message will be displayed showing time remaining (or time elapsed) and the appropriate display item (viscosity or torque) will be updated continuously during the event. Upon completion, the viscometer will stop and display a screen stating that the test is complete and will also display the nal recorded value for the viscosity in the rst case, or the time in minutes and seconds to reach the torque limit in the second case. Pressing the UP or DOWN ARROW keys will allow additional viscometer data to be examined. Pressing any other key (except the PRINT or ENTER/AUTORANGE key) will bring the user back to the default (normal) viscometer display with the motor OFF. Refer to the Time Modes in Section III.3.
Brookeld Engineering Labs., Inc. Page 21 Manual No. M/03-165-D0410
III. OPTIONS
III.1
OPTION
TAB
Introduction to OPTIONS
F
The OPTIONS/TAB key provides access to the conguration (Setup) of the DV-II+Pro Viscometer as
well as special functions that can enhance the user’s ability to make viscosity measurements.
The Options menu, shown in Table 1, gives a complete picture of the various conguration
choices and special functions.
Quick References to Options
Table 2
Options Menu
SETUP:
Temperature °F or °C Units CGS or SI * Speed Sets Sequential, Interleave, Custom Printer Port Serial (RS232) or Parallel Data Averaging — Display Only
*
TIME TO STOP
* TIME TO TORQUE
SET PRINT TIME:
Set the Printing Time
PC PROG (ON/OFF):
Enables/Disables Communication of Serial (RS232) Port
* DOWNLOAD A PROGRAM:
Link to PC to Receive a B.E.V.I.S. Program
(B.E.V.I.S. = Brookeld Engineering Viscometer Instruction Set)
* RUN A PROGRAM:
Execute a B.E.V.I.S. Program
*Not available when motor is ON
Pressing OPTIONS/TAB places you into the Options menu at the last option selected. The
following keys are active and perform as follows:
£
UP ARROW - Scrolls up through menu or selects new value from
list.
§
DOWN ARROW - Scrolls down through menu or selects new value
from list.
Brookeld Engineering Labs., Inc. Page 22 Manual No. M/03-165-D0410
OPTION
TAB
F
OPTIONS/TAB - Toggles between options.
ENTER
AUTO
MOTOR ON/OFF
ESCAPE
The Options menu screens will appear as shown in Figure III-1 if you cycle through the possible options using the UP/DOWN arrows.
ENTER/AUTORANGE
RANGE
MOTOR ON/OFF/ESCAPE - Cancels current operation and backs user out
- Accepts the currently ashing option and moves user to next level (if applicable) of the selected option.
one menu level. Repeated pressing will back the user out to the default screen. While in
the Options menu, the MOTOR ON/OFF/ESCAPE key does not cause the viscometer motor to
turn on or off!
$
SETUP # DTIME TO TORQUE $
OR
SETUP # DTIME TO STOP $
2
DTIME TO TORQUE # SET PRINT TIME $
OR
DTIME TO STOP # SET PRINT TIME $
£
§
2
SET PRINT TIME # DPC PROG OFF $
2
DPC PROG OFF # DOWNLOAD A PROG $
2
DOWNLOAD A PROG # RUN A PROG $
#
Figure III-1
OR
OR
OR
SET PRINT TIME # DPC PROG ON $
DPC PROG ON # DOWNLOAD A PROG $
Brookeld Engineering Labs., Inc. Page 23 Manual No. M/03-165-D0410
On entry to the Options menu, the following rules regarding current viscometer operation are in force:
1. Printer output will be suppressed when in the Custom Speed option, the Time to Torque and Time to Stop options, the Download A Program and Run A Program options. It will be continued when any other option is selected.
2. If the motor is ON when the user enters the Options menu, choices will be limited to: CGS/SI units (under SETUP), °F/°C units (under SETUP), PRINTING SELECTIONS and PC PROG.
3. The
last selected menu option will be ashing.
Selecting an Option
The following is a quick reference for entering and using the OPTIONS menu:
Press
OPTION
TAB
To enter Options Menu
F
Press
£
or
§
To scroll to a specic option
For Options:
Press
Press
OPTION
TAB
ENTER
AUTO
RANGE
To toggle between the choices available for a specic option when indicated
F
To select the ashing option
III.2 Setup
From the main Options screen, the user scrolls up or down until the following screen is displayed:
SETUP è DTIME TO TORQUE ê
Figure III-2
A press of the ENTER/AUTORANGE key takes you into the Setup sub-menu (Figure III-3). As in the main Options menu, you can scroll up or down through the various Setup options. In order to access all options, the motor must be turned off.
Brookeld Engineering Labs., Inc. Page 24 Manual No. M/03-165-D0410
$
£
§
D°F(FAHRENHEIT) # DCGS UNITS $
2
D°F(FAHRENHEIT) # DCGSUNITS $
2
DCGS UNITS # DSEQUENTIAL $
2
DSEQUENTIAL # DPRINTSERIAL $
2
1
1
1
1
D°C(CENTIGRADE) # DCGS UNITS $
D°F(FAHRENHEIT) # DSIUNITS $
DCGS UNITS # DINTERLEAVE $
DSEQUENTIAL # DPRINTPARALLEL $
DCGS UNITS #
1
DCUSTOMSPEEDS$
DPRINT SERIAL # DDATAAVERAGING $
OPTION TAB
F
#
Figure III-3
III.2.1 Temperature Display
The DV-II+Pro viscometer can display temperature in either degrees Centigrade or degrees Fahrenheit. On entry (assuming the viscometer is currently displaying °F) you will be presented with:
D°F(FAHRENHEIT) è DCGS UNITS ê
Brookeld Engineering Labs., Inc. Page 25 Manual No. M/03-165-D0410
Figure III-4
A press of the OPTIONS/TAB key at this point will “toggle” between the two available temperature scale options as shown in Figure III-5:
D°F(FAHRENHEIT) è DCGS UNITS ê
To select the temperature display mode, press the ENTER/AUTORANGE key. You automatically exit the Setup menu with the viscometer displaying temperature in the selected scale. You must press the ENTER/AUTORANGE key to select the ashing option.
III.2.2 Units of Measurement
Selecting units of measurement is identical to that for temperature described above. The screen display shows:
D°F(FAHRENHEIT) è DCGSUNITS ê
A press of the OPTIONS/TAB key at this point allows the user to “toggle” between the two available data display units as shown in Figure III-7:
"
Figure III-5
Figure III-6
D°C(CENTIGRADE) è DCGS UNITS ê
D°F(FAHRENHEIT) è DCGSUNITS ê
Pressing the ENTER/AUTORANGE key selects the display units, which are ashing, followed by an exit of the Setup menu. You must press the ENTER/AUTORANGE key to select the ashing option.
III.2.3 Motor Speed Set Selection
This selection must be done with the motor off. Scrolling in the Setup options menu to the speed set selection option yields the following screen display:
DCGS UNITS è DSEQUENTIAL ê
"
É
Figure III-7
Figure III-8
DF(FAHRENHEIT) è DSIUNITS ê
Brookeld Engineering Labs., Inc. Page 26 Manual No. M/03-165-D0410
The last selected speed set option is displayed, in this case, Sequential. For each press of the
OPTIONS/TAB key, the display shows selectable options (Figure III-9). You must press the ENTER/ AUTORANGE key to select the ashing option.
DCGS UNITS è DSEQUENTIAL ê
The speeds available in each of the above options are listed in Appendix G. The DV-II+Pro is initially set up with the Sequential Speed Set at Brookeld prior to shipment.
III.2.3.1 LV/RV Speeds
In the case of Sequential or Interleave, a press of the ENTER/AUTORANGE key immediately selects that option and exits the SETUP option menu.
III.2.3.2 Custom Speeds
Pressing the ENTER/AUTORANGE key when Custom Speeds is displayed results in the following screen display:
DCGS UNITS è
1
DINTERLEAVE ê
É
Figure III-9
.01 h °0.0 i
DCGS UNITS è
1
DCUSTOMSPEEDS ê
Figure III-10
Fifty-four (54) speeds are available for custom speed set selection (see Appendix G). You are allowed to select up to nineteen (19) of these (54) available speeds. Speed 0.0 is automatically included as one of the nineteen (19) speeds. Selecting a speed (or deleting a speed) is accomplished by pressing the SET SPEED key while the desired speed is blinking. This will cause an asterisk to appear (or to disappear if the speed is being cleared) to the left of the speed. Trying to select more than nineteen (19) speeds will result in a “beep” for each press of the SET SPEED key when over this limit. You may scroll up or down through the speed set in selecting speeds. When done, a press of the ENTER/AUTORANGE key will take you back to the default screen with the asterisked speeds now comprising the custom speed set. Regardless of order chosen, speeds will appear in ascending order for run selection, beginning with speed 0.0 RPM.
Brookeld Engineering Labs., Inc. Page 27 Manual No. M/03-165-D0410