B&K Precision 2120B User Manual

5 (1)
B&K Precision 2120B User Manual

Data Sheet

30 MHz Dual Trace Analog Oscilloscope With Probes

Model 2120B

B&K Precision's model 2120B is a dual trace oscilloscope that offers high performance at a low price. Most competitor's entry level oscilloscopes have a 20 MHz bandwidth, while B&K Precision's models 2120B has a bandwidth of 30 MHz. This oscilloscope is built by and backed by B&K Precision, a company that has been selling reliable, durable, value priced test instruments for over 50 years.

Dual or single trace operation

5 mV/div sensitivity

AUTO/NORM triggered sweep operation with AC, TVH, TVV and line coupling

Compact low profile design

cUL certified

Specifications

2120B

VERTICAL AMPLIFIERS (CH 1 and

CH 2)

 

 

Sensitivity

5 mV/div to 5 V/div, 1 mV/div to 1 V/div at X5

Attenuator

10 steps in 1-2-5 sequence. Vernier control provides full adjustment between steps

Accuracy

±3%, ±5% at X5

Input Resistance

1 MΩ ±2%

Input Capacitance

25 pF ±10pF

Frequency Response

5 mV to 5 V/div: DC to 30 MHz (-3dB). X5: DC to 10 MHz (-3dB)

Rise Time

12 ns (Overshoot ≤5%)

Operating Modes

CH 1: CH 1, single trace

CH 2

CH 2, single trace

ALT

dual trace, alternating

CHOP

dual trace, chopped

ADD

agebraic sum of CH 1 + CH 2

Polarity Reversal

CH 2 only

Maximum Input Voltage

400 V (DC + AC peak)

SWEEP SYSTEM

 

Sweep Speed

0.1 µs/div to 2 s/div in 1-2-5 sequence, 23 steps,

Vernier control provides fully adjustable sweep time between steps.

 

Accuracy

±3%

Sweep Magnification

10x

TRIGGERING

 

 

 

Triggering Modes

AUTO (free run) or NORM, TV-V, TV-H

Trigger Source

CH 1, CH 2, ALT, EXT, LINE

Maximum External Trigger Voltage

300 V (DC + AC peak)

Trigger Coupling

AC 30 Hz to 30 MHz

TV H

Used for triggering from horizontal sync pulses

TV V

Used for triggering from vertical sync pulses

TRIGGER SENSITIVITY

 

Auto

Bandwidth:100 Hz-30 MHz, Internal: 1.5 div, External: 100 mV

Norm

Bandwidth: DC to 30 MHz, Internal: 1.5 div, External: 100 mV

TV V

Bandwidth: 20 Hz-1 kHz, Internal: .5 div, External: 100 mV

TV H

Bandwidth:1 kHz-100 kHz, Internal: .5 div, External: 100 mV

HORIZONTAL AMPLIFIER (Input

through channel 2 input)

 

 

X-Y Mode

Switch selectable using X-Y switch. CH 1: X axis, CH 2: Y axis

Sensitivity

Same as vertical channel 1

Input Impedance

Same as vertical channel 1

Frequency Response

DC to 1 MHz typical (-3 dB)

X-Y Phase Difference

Approximately 3˚ at 50 kHz

Maximum Input Voltage

Same as vertical channel 1

CRT

 

 

 

Type

Rectangular with internal graticule

Display Area

8 x 10 div (1 div = 1 cm)

Accelerating Voltage

2 kV

Phosphor

P31

Trace Rotation

Electrical, front panel adjustable

Calibrating Voltage

1 kHz (±10%) Positive Square Wave, 2 V p-p (±3%)

GENERAL

 

 

 

 

Within Specified Accuracy: 50˚ to 95˚F (10˚ to 35˚C), ≤85% RH

Temperature

Full Operation: 32˚ to 104˚F (0˚ to 40˚C), ≤85% RH

 

Storage: -4˚ to 158˚F ( -20˚ to +70˚C

Power Requirements

100/120/220/240 VAC ±10%, 50/60 Hz, approximately 40 W.

Dimensions (WxHxD)

7 x 14.5 x 17.25" (180 x 370 x 440 mm)

Weight

Approximately 17.2 lbs (7.8 kg)

 

Two Year Warranty

Supplies Accessories

Instruction Manual, Two PR-33A x1/x10 Probes or equivalent,

AC Power Cord, Spare Fuse

 

Optional Accessories

PR-32A Demodulator Probe, PR-37A x1/x10/REF. Probe, PR-100A x100 Probe,

PR-55 High Voltage x1000 Probe, LC-210A Carrying Case

 

Technical data subject to change

www.bkprecision.com

© B&K Precision Corp. 2012

v013012

Tel.: 714.921.9095

Loading...