The issue date for the M Series Service Manual (REF 9650-0450-01 Rev. R) is March, 2013.
If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if additional product information updates
are available.
Masimo is a trademark or registered trademark of Masimo Corporation in the United States
and/or other countries.
9650-0450-01 Rev R
M Series Service Manual
TABLEOF CONTENTS
PREFACE ....................................................................................................................................................................................... V
Additional Reference Material ........................................................................................................................................................... vi
Service Policy Warranty .................................................................................................................................................................... vii
Technical Service ..............................................................................................................................................................................vii
Technical Service for International Customers ................................................................................................................................viii
Before You Begin the Maintenance Tests ............................................................................................................................................2
Equipment that You Need to Perform the M Series Maintenance Tests ..............................................................................................3
Equipment You Need for the M Series Options Maintenance Tests ...................................................................................................4
1.0 Physical Inspection of the Unit...................................................................................................................................................... 5
2.0 Front Panel Button Test................................................................................................................................................................. 7
3.0 3, 5, and 12 Leads Test................................................................................................................................................................ 10
4.0 Power Supply Test (Optional)..................................................................................................................................................... 11
5.0 Leakage Current Test................................................................................................................................................................... 15
6.0 Paddles Test (If applicable)......................................................................................................................................................... 16
8.0 Calibrating Pulses on Strip Chart Test ........................................................................................................................................ 19
13.0 Shock Test ................................................................................................................................................................................. 26
15.0 Advisory Message Test ............................................................................................................................................................. 30
16.0 Pacer Test .................................................................................................................................................................................. 31
i
M Series Service Manual
17.0 SpO2 Monitor Test for SpO2 Option........................................................................................................................................ 33
18.0 EtCO2 Monitor Test (for EtCO2 Option).................................................................................................................................. 36
19.0 Temperature Test........................................................................................................................................................................ 38
23.0 NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer............................................................................................. 49
24.0 Bluetooth Test............................................................................................................................................................................ 51
Zoll M Series Error Messages ............................................................................................................................................................58
Parts That May Need Replacing After Disassembly ..........................................................................................................................76
Overview of Modules .........................................................................................................................................................................78
1. Removing the ZIF Keeper ............................................................................................................................................81
2. Removing the Front Panel ............................................................................................................................................82
2A.Removing the Display .................................................................................................................................................83
2B.Removing the Control Board ......................................................................................................................................84
3. Removing the Upper Housing Assembly .....................................................................................................................86
4. Removing the System Board Assembly .......................................................................................................................87
5. Removing the Battery Interconnect Board Assembly ..................................................................................................90
6. Removing the High Voltage/Charger Assembly .........................................................................................................91
7. Removing the High Voltage Module Assembly ..........................................................................................................94
8. Removing the High Voltage Capacitor Assembly .......................................................................................................96
9. Removing the System Interconnect Board ...................................................................................................................98
10.Removing the Printer/Recorder Motor ......................................................................................................................100
11.Removing the Lower Housing Assembly ..................................................................................................................102
12.Removing the Print Head Assembly ..........................................................................................................................103
13.Removing the PCMCIA Card Slot Assembly ...........................................................................................................104
ii
M Series Service Manual
14.Removing the Paddle Release Latch .........................................................................................................................105
CHAPTER 4 REPLACEMENT PARTS .........................................................................................................................................107
Replacement Parts ............................................................................................................................................................................108
Field Replacement Parts ...................................................................................................................................................................112
Main System Board ..........................................................................................................................................................................116
Main System Board Functions .........................................................................................................................................................118
Power Supply ................................................................................... .................................................................................................120
ECG Front End .................................................................................................................................................................................121
Multifunction Electrode (MFE)/PADS (System Board and High Voltage Module) .......................................................................121
CPU and EPU ...................................................................................................................................................................................121
High Voltage Module .......................................................................................................................................................................122
Defibrillator Charging and Discharging ...........................................................................................................................................123
High Voltage Capacitor Monitor ......................................................................................................................................................124
Pacer/Defibrillator Control Signals ..................................................................................................................................................125
System Interconnect Module ............................................................................................................................................................127
Front Panel and Controls PWBA ......................................................................................................................................................128
M Series Options ..............................................................................................................................................................................128
Isolated Power Supply Module .........................................................................................................................................................129
12 Lead Option .................................................................................................................................................................................129
Interconnect Diagram for the M Series Monophasic Unit ...............................................................................................................136
Interconnect Diagram for the M Series Biphasic Unit .....................................................................................................................137
Interconnect Diagram for the M Series CCT Biphasic Unit ............................................................................................................138
iii
M Series Service Manual
INDEX ..........................................................................................................................................................................................139
iv
M Series Service Manual
Overview
ZOLL Medical Corporation’s M Series Service Manual is intended for the service technician whose responsibility
is to identify malfunctions and/or make repairs at the subassembly level. The Zoll M Series Service Manual has five
main sections and one appendix.
Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to review this section
thoroughly before attempting to use or service the M Series unit.
Chapter 1—Maintenance T ests explains how to check the defibrillator’s performance using a series of recommended
checkout procedures to be conducted every six months.
Chapter 2—Troubleshooting provides a listing of the procedures and error messages to help the service technician
detect faults and repair them.
Chapter 3—Disassembly Procedures describes step-by-step procedures for removing subassemblies from the
M Series unit.
Chapter 4—Replacement Parts List displays a complete list of ZOLL part numbers for field replaceable parts
available for the M Series unit, allowing the service person to identify and order replacement parts from ZOLL.
Chapter 5—Functional Description provides technical descriptions for the M Series major subassembly modules.
Appendix A—M Series Operator’s Manual.
Preface
Safety Considerations
The following section describes general warnings and safety considerations for operators and patients. Service
technicians should review the safety considerations prior to servicing any equipment and read the manual carefully
before attempting to disassemble the unit. Only qualified personnel should service the M Series unit.
Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.
Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug Administration (FDA)
under section 510(K) of the Medical Device Act to obtain approval to market is based upon the use of ZOLL
accessories such as disposable electrodes, patient cables and batteries. The use of external pacing/defibrillation
v
M Series Service Manual
electrodes and adapter units from sources other than ZOLL is not recommended. ZOLL makes no representations or
warranties regarding the performance or effectiveness of its products when used in conjunction with pacing/
defibrillation electrodes and adapter units from other sources. If unit failure is attributable to pacing/defibrillation
electrodes or adapter units not manufactured by ZOLL, this may void ZOLL's warranty.
Only qualified personnel should disassemble the M Series unit.
WARNING! This unit can generate up to 4500 volts with sufficient current to cause lethal shocks.
All persons near the equipment must be warned to “STAND CLEAR” prior to discharging the defibrillator.
Do not discharge the unit’s internal energy more than three times in one minute or damage to the unit may result.
Do not discharge a battery pack except in a Base PowerCharger
Do not use the M Series in the presence of flammable agents (such as gasoline), oxygen-rich atmospheres, or
flammable anesthetics. Using the unit near the site of a gasoline spill may cause an explosion.
Do not use the unit near or within puddles of water.
NOTEThe M Series is protected against interference from radio frequency emissions typical of two-way radios and cellular
phones (digital and analog) used in emergency service/public safety activities. Users of the M Series should assess the
unit’s performance in their typical environment of use for the possibility of radio frequency interference from
high-power sources. Radio Frequency Interference (RFI) may be observed as shifts in monitor baseline, trace
compression, or transient spikes on the display.
4x4
or compatible ZOLL Battery Charging/Testing unit.
Additional Reference Material
In addition to this guide, there are several other components to the Zoll M Series documentation. They include:
• Operator’ s Guide - A comprehensive reference work that describes all the user tasks needed to operate the M Series.
• Configuration Guide - Describes the M Series features and functions whose operation can be customized by
authorized users.
vi
M Series Service Manual
Conventions
WARNING!Warning statements describe conditions or actions that can result in personal injury or death.
CAUTIONCaution statements describe conditions or actions that can result in damage to the unit.
NOTENotes contain additional informationon using the defibrillator.
Service Policy Warranty
In North America: Consult your purchasing agreement for terms and conditions associated with your warranty. Outside
of North America, consult ZOLL authorized representative.
In order to maintain this warranty , the instructions and procedures contained in this manual must be strictly followed.
For additional information, please call the ZOLL Technical Service Department 1-800-348-9011 in North America.
Technical Service
If the ZOLL M Series unit requires service, contact the ZOLL Technical Service Department:
Telephone: 1-978-421-9655; 1-800-348-9011
Fax 1-978-421-0010
Have the following information available for the Technical Service representative:
• Unit serial number.
• Description of the problem.
• Department where equipment is used.
• Sample chart recorder strips documenting the problem, if applicable.
• Purchase Order to allow tracking of loan equipment.
• Purchase Order for a unit with an expired warranty.
vii
M Series Service Manual
If the unit needs to be sent to ZOLL Medical Corporation, obtain a service order request number from the Technical
Service representative. Return the unit in its original container to:
Technical Service for International Customers
International customers should return the unit in its original container to the nearest authorized ZOLL Medical
Corporation Service Center. To locate an authorized service center, cont ac t the International Sales Department at
ZOLL Medical at the above address.
Units are available on loan while your unit is being repaired.
ZOLL Medical Corporation
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Attn: Technical Service Department
Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010
viii
M Series Service Manual
Overview
The M Series has two checkout procedures: the operator’s shift checklist and the extensive six-month maintenance
tests checkout procedures.
Because the M Series units must be maintained ready for immediate use, it is important for users to conduct the
Operator’s Shift Checklist procedure at the beginning of every shift. This procedure can be completed in a few minutes
and requires no additional test equipment. (See the ZOLL M Series Operator’s Guide for the Operator’s Shift
Checklist.)
A qualified biomedical technician must perform a more thorough maintenance test checkout every six months to
ensure that the functions of the M Series unit work properly. This chapter describes the step by step procedures for
performing the six month maintenance test checkout. Use the checklist at the back of this document (ZOLL M Series
Maintenance Tests Checklist) to record your results of the maintenance tests.
Chapter 1
Maintenance Tests
This chapter describes the following M Series maintenance tests:
1. Physical Inspection of the Unit
2. Front Panel Button
3. 3, 5, and 12 Leads
4. Power Supply
5. Leakage Current
6. Paddles
7. Heart Rate Display
8. Calibrating Pulses on Strip Chart
9. Notch Filter
10. Heart Rate Alarm
1
M Series Service Manual
11. Defibrillator Self Test
12. Synchronized Cardioversion
13. Shock
14. Summary Report
15. Advisory Message
16. Pacer
17. SpO
18. EtCO
19. Temperature Test
20. IBP Monitoring Test
21. NIBP Transducer Calibration Test
22. NIBP Monitor Test
23. NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer
24. Bluetooth Test
Monitor
2
Monitor
2
Before You Begin the Maintenance Tests
• Assemble the tools or specialized testing equipment listed in the “Equipment You Need to Perform the Maintenance
Tests” section shown below.
• Keep an extra fully charged ZOLL M Series compatible battery available.
• Schedule an hour to conduct the entire maintenance test.
• Photocopy the checklist at the back of this document and use the copy to record your results. As you con duct each
step of a procedure, mark the
Pass/Fail/NA check boxes on your checklist and then save it for your maintenance file.
• Perform the tests in the order presented.
• Perform all the steps of each test procedure.
• Complete all the steps of the procedure before evaluating the test results.
2
M Series Service Manual
Equipment that You Need to Perform the M Series Maintenance
This section lists equipment that we use to perform the maintenance tests that we describe in this chapter. You can
substitute an equivalent device for a listed device; however, not all simulators and analyzers will produce the same results. Be sure to follow the manufacturer’s recommendations for conducting the maintenance tests.
We recommend the use of the following equipment when performing M Series Maintenance Tests:
• ZOLL Medical Electrode Adapter from Fluke Biomedical (ZOLL Part Number 3010-0378).
• QED 6 Defibrillator Analyzer
• MFC Test Port Connector, 1004-0053-99
• Fluke Biomedical
• ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201).
• Stop watch.
• Standard series II PC flash memory cards.
• 1 red miniature alligator to miniature alligator test lead.
• 1 black miniature alligator to miniature alligator test lead.
• DC power supply (15 amp minimum).
Tests
601 Pro Series International Safety Analyzer.
• 0.1 ohm 1% resistor (1/4 watt or greater).
• 1000 ohm 1% resistor (1/4 watt or greater).
• Fluke 75 Multimeter or equivalent.
•Printer Paper.
• Battery.
• AC line cord.
• 3 lead, 5 lead and 12 lead ECG cables. (12 lead cable needed if 12 lead option is installed.
3
M Series Service Manual
Equipment You Need for the M Series Options Maintenance
• Fluke Biomedical Index 2 SpO2 Simulator or equivalent. (For SpO2 units only.)
•SpO
•EtCO
• Novametrix Capnostat Simulator Yb 1265/7100
• Paddles (if used).
•)
• Fluke Biomedical BP Pump 2 NIBP Monitor Analyzer (For NIBP units only) with NIBP cable and cuff (if NIBP
• Fluke Biomedical Cufflink Analyzer (if NIBP option is installed)
• IBP Temperature Simulator 9100-0402-TF
• DNI Nevada 214B Patient SImulator, with M Series interface cable
Tests
cable and sensor (if option is installed).
2
cable, and CAPNOST AT 3 Mainstream cable with airway adapter, or CAPNOSTAT 3 Sidestream cable with
2
cannula (if option is installed).
option is installed), or
NOTE The Fluke Biomedical BP Pump NIBP Monitor Analyzer and the Fluke Biomedical BP Pump NIBP Analyzer
use different technologies for testing NIBP monitors and therefore, the manual provides two different procedures for
performing the NIBP Volume Leak test with each of these types of test equipment.
4
M Series Service Manual
1.0 Physical Inspection of the Unit
Tools Needed Battery.
Test Setup None.
Observe this...Pass/Fail/NA
1.1
1.2
1.3
1.4
1.5
1.6
1.7
1.8
1.9
1.10
1.11
Housing
Is the unit clean and undamaged?
Does the unit show signs of excessive wear?
Does the handle work properly?
Does the recorder door open and close properly?
Are input connectors clean and undamaged?
Are there any cracks in the housing?
Do the front panel or selector switches have any damage or cracks?
Are there any loose housing parts?
Cables
Are all cables free of cracks, cuts, exposed or broken wires?
Are all cable bend/strain reliefs undamaged and free of excessive cable
wear?
Paddles
Do the adult and pedi plates have major scratches or show signs of damage?
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo
oo o
1.12
1.13
Do the adult shoes slide on and off easily to expose the covered pedi plates?
Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)
5
oo o
oo o
M Series Service Manual
Observe this...Pass/Fail/NA
1.14
1.15
Record your results on the Maintenance Test Checklist.
Battery
Place battery in battery well.
Is the battery seated in the battery well correctly?
oo
oo
6
M Series Service Manual
2.0 Front Panel Button Test
Tools Needed QED 6 Defibrillator Analyzer
Test Setup• Install strip chart paper into the recorder tray.
• Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electri cal outlet.
• Connect the universal cable and ECG cable (3 lead, 5 lead, or 12 lead) to the QED 6 Analyzer (or equivalent).
2.1
2.2
2.3
2.4
2.5
Do this...Verify that...Pass/Fail/
Turn the selector switch to
MONITOR.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Set the QED 6 analyzer to NSR
of 120 BPM. T o check the size of
the ECG waveform, press the
SIZE button.
Press the ALARM SUSPEND
button.
Press the RECORDER button.The strip chart paper moves out of the unit from the
Open the paper compartment
door, then press the
RECORDER button.
The unit sounds 4 beep tones. PADS and MONITOR
display on the monitor.
Note: PADS is a factory default setting.
As you press the SIZE button five times (0.5, 1.0, 1.5,
2.0, 3.0), the size of the ECG waveform appropriately
changes on the display.
Alarm symbol changes from disabled to enabled. If the
alarm sounds, press the ALARM SUSPEND button to
turn it off. The alarm will only be suspended for 90
seconds at this point. Press and hold the ALARM SUSPEND button for 3 seconds to disable alarms.
paper tray. Check that the correct time, date, ECG
lead annotation and waveform are recorded on the
paper. (Set Time and Date, if necessary.)
The CHECK RECORDER message appears on the
monitor.
NA
oo
oo
ooo
ooo
ooo
2.6
Close the paper compartment
door, then press the
RECORDER button.
The strip chart paper flows out of paper tray and the
monitor no longer displays the CHECK RECORDER
message.
7
ooo
M Series Service Manual
2.7
2.8
2.9
2.10
2.11
2.12
Do this...Verify that...Pass/Fail/
Press the RECORDER button.The strip chart paper stops flowing out of the paper
tray.
Press the VOLUME softkey.The volume bar graph displays.
NA
ooo
oo
Note: The QRS tone is on or off. There is no gradual
change in volume. If equipped, voice prompts
are gradual. Note: The voice volume has 5
settings. Setting 3 is in the mid-range.
Press the INC softkey.The bar graph appears on the display, indicating an
increase in volume.
oo
Note: This action does not increase the volume of the
unit’s audio prompts.
Press the DEC softkey.The bar graph appears on the display, indicating a
decrease in volume.
Verify that pressing the DEC softkey until only the last
bar appears silences the unit’s audio prompts,
otherwise the volume of the audio prompts does not
decrease.
Press the CONTRAST button. The monitor screen displays the Contrast Menu.
Press the INC softkey. The screen contrast and bar graph increases on the
display.
oo
oo
ooo
2.13
2.14
2.15
2.16
Press the DEC softkey.The screen contrast and bar graph decreases on the
display.
Press the SUMMARY button.The monitor screen displays the Summary menu,
showing the summary report options.
Press the CODEMARKER
button.
Connect to AC or DC current
and install the battery. Turn the
unit off.
The monitor screen displays the Code marker menu.
The CHARGER ON indicator lights (either the Green
or Amber indicator lights).
8
oo
ooo
ooo
ooo
M Series Service Manual
2.17
2.18
2.19
2.20
2.21
2.22
2.23
Do this...Verify that...Pass/Fail/
Remove the battery.The charger light alternately flashes green and amber.
Replace the battery and the turn
unit on.
Press the ANALYZE button.The SELECT DEFIB MODE message appears on the
Connect the universal cable to
the QED 6. Move the selector
switch to DEFIB. Select 2J.
Press the CHARGE button.
Press and hold the ENERGY SELECT down arrow.
Press and release the ENERGY SELECT up arrow 19 times.
Press the CHARGE button.The display shows the unit charged up to 200J (360J
The amber charger light illuminates.
monitor (for manual devices).
The display shows that the unit is charging. The
SHOCK button lights when the unit is charged, and
the Ready tone for DEFIB sounds.
The unit discharges internally and selected energy
decrements to 1J.
The following selections appear: Verify the following
energy amounts display incrementally:
Record your results on the Maintenance Test Checklist.
Press the SHOCK button.The unit discharges and the SHOCK button is no
longer illuminated.
A 15 second strip chart automatically prints, displaying
the number of joules delivered (if equipped with
recorder and configured to print event).
9
ooo
M Series Service Manual
3.0 3, 5, and 12 Leads Test
Tools Needed: 3 lead, 5 lead, and 12 lead cables. Test each cable separately. QED 6 Defibrillator Analyzer.
Test Setup: The M Series unit must be configured to display ECG LEAD OFF message.
Connect the lead wires appropriate to each test to the QED 6 Defibrillator Analyzer.
Do this...Observe this...Pass/Fail/
3.1
3.2
3.3
3.4
Turn the selector switch to
MONITOR. Select leads.
Disconnect one lead from the
simulator.
Reconnect the lead. Repeat
step 3.2 with the remaining
leads.
Repeat 3.2 and 3.3 for 5 lead
and 12 lead cables.
The monitor displays the NO ECG LEADS OFF
message.
The monitor displays the ECG LEAD OFF message
within 3 seconds (if configured).
The ECG LEAD OFF message appears when the
lead is disconnected and clears the lead is
reconnected.
Note: If heart rate alarm sounds, press and hold the
ALARM SUSPEND button for 4 seconds to
disable the alarms.
Note: When testing the 12 lead cable, the ECG
LEAD OFF message displays when you pull
off a limb lead. When you pull off a V lead, the
ECG VX LEAD OFF message displays, where
“X” is the number between 1 and 6.
Record your results on the Maintenance Tests Checklist.
NA
ooo
ooo
ooo
ooo
10
M Series Service Manual
4.0 Power Supply Test (Optional)
Tools Needed:• 1 red miniature alligator to miniature alligator test lead.
• 1 black miniature alligator to miniature alligator test lead.
• DC power supp ly (15 Amp min imu m).
• 0.1 ohm
• 1000 ohm 1% resistor (
• Fluke 75 multimeter or equivalent.
Test Setup:Make sure the unit and power supply are turned off.
Connect one end of the black lead to the “-” terminal in the battery well.
Connect the other end of the black lead to the “-” terminal of the power supply.
Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it.
Connect the other end of the red lead to the “+” terminal of the power supply.
Set the power supply voltage to 7V.
CAUTIONBe sure to connect the power supply properly to the M Series battery well terminals or damage to the unit may result. Do NOT raise
the power supply voltage above 12V.
resistor (¼W or greater).
¼W or greater).
11
M Series Service Manual
?
Battery Well
15 Amp
Supply
?
Red
Black
Do this...Verify that...Pass/
Turn the selector switch to
4.1
MONITOR.
(For AED units, turn the selector
switch to ON and select Manual
mode.)
Turn the selector switch to the
4.2
OFF position.
Adjust the power supply voltage to
4.3
10.3V and turn the selector switch
to MONITOR (for AED units, turn
the selector switch to the ON
position).
Low Battery Test
4.4
Set voltage to 9.8V.
Set voltage to 9.4V .The unit displays the LOW BATTER Y message within 30
4.5
Shut Down Voltage Test
4.6
Set voltage to 8.5V.
Record your results on the Maintenance Tests Checklist.
The unit does not turn on.
The unit turns on.
The unit does not display the LOW BATTERY message.
seconds.
The unit shuts off within 30 seconds.
Fall
oo
oo
oo
oo
oo
12
M Series Service Manual
Test Setup
Remove red lead from power supply and connect to 0.1ohm resistor.
Connect other end of resistor to “+” terminal of power supply using a second red label.
Connect multimeter across the resistor.
Set voltage scale (if DVM is not autoranging) to 220mV.
Red
Black
+
_
Battery Well
_
+
+
_
DMM
15 Amp
Supply
4.7
Do this....Observe this...Pass/
Fail/
System Current Test
Set power supply to 10.3V.
13
M Series Service Manual
Do this....Observe this...Pass/
4.8
4.9
Record your results on the Maintenance Test Checklist
Turn the selector switch to
MONITOR. (For AED units, turn
the selector switch to ON and
select Manual mode.
Turn unit off.
Voltage across resistor should be 80 mV or less (<800
mA of ON current). Note: Without optional parameters.
a) With green screen or LCD and no options <80mV.
b) With yellow screen and no options <81mV.
c) With yellow screen and SpO2 <104mV.
d) With yellow screen and voice recording <91mV.
e) With yellow screen, voice recording, and SpO2
<114mV.
f) All devices with EtCO2 <121mV.
Off Current Test
Fail/
oo
oo
Test SetupRemove 0.1 ohm resistor and replace with 1K
Connect DMM across resistor.
Set voltage scale to DCV.
Measure voltage across resistor.
Measure across resistor with unit
4.10
turned off.
Record your results on the Maintenance Test Checklist
Voltage should be less than 450mV (<450uA of current).
14
oo
M Series Service Manual
5.0 Leakage Current Test
Tools Needed: See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Test Setup: See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with
accessories: MFC, external paddles, and anterior/posterior paddles.
Table 1:
Maximum Leakage Acceptance Limits
Normal ConditionSingle Fault Condition*
ECG1050
Universal100100
Earth5001000
*Single fault considered AC mains on applied part.
15
M Series Service Manual
6.0 Paddles Test (If applicable)
Tools Needed: QED 6 Defibrillator Analyzer .
Test Setup: If applicable, connect the universal cable to the paddles and place the paddles in paddle wells.
6.1
6.2
6.3
6.4
6.5
6.6
Do this...Verify that...Pass/Fail/
Turn the selector switch to
DEFIB (For AED units,turn the
selector switch to ON and
select Manual mode.) Press
and hold the ENERGY DOWN
button on the sternum paddle.
Press and release the ENERGY UP button on the sternum paddle
for each setting.
Press and release the
RECORDER button on the
sternum paddle.
Press and release the
RECORDER button again.
Select 30J using the paddle
ENERGY button. Press the
CHARGE button on the Apex
paddle.
Press and release the APEX SHOCK button.
The energy selection decreases to 1J.
The joules selection increases as follows: Biphasic 110, 15, 20, 30, 50, 75, 100, 120, 150, 200J.