Zoll M CCT User manual

Military Airworthiness-Certified
ZOLL M Series CCT Manual Insert
Part Number 9652-0096-30 Rev. A
Military Airworthiness-Certified ZOLL M Series CCT Manual Insert
This manual insert describes warnings, notices, and cautions that apply to the military airworthiness-certified ZOLL M Series CCT unit.
Warning:
When you operate the military M Series CCT unit from a 400Hz power source, please note that the patient auxiliary leakage current specifications are different than when operating from either 50Hz or 60Hz.
For the military M Series CCT unit, the patient auxiliary leakage current specification for ECG and MFC is
Type CF for 50Hz and 60Hz power
Type BF for 400Hz power
Do not operate the military M Series CCT unit from a 400Hz power source when using intracardiac monitoring catheters or internal paddles. To use intracardiac monitoring catheters or internal paddles, you must operate the M Series CCT unit from either a 50Hz or 60Hz power source or from battery power. The use of intracardiac monitoring catheters or internal paddles when operating the military M Series CCT unit from a 400Hz power source can result in significant risk to the patient.
Notice:
The product is tested for airworthiness without the VGA output capability. VGA output is not functional in the military M Series CCT unit.
Notice:
Input power to the airworthiness-certified military M Series CCT unit is restricted to 120 VAC (at 50Hz, 60Hz, or 400Hz) and battery power.
Caution:
The modem and ECG out ports have not been tested for airworthiness, however the modem and ECG out ports are functional in a non-airworthiness environment. The use of the military CCT unit’s modem and ECG out ports in an airworthiness environment has the potential to cause electronic interference, which can compromise the airworthiness environment and risk the safety of the aircraft.
Caution:
The external cables used on military M Series CCT units are not the same as those used on standard M Series CCT units. The use of standard CCT cables with the military CCT unit invalidates the military unit’s airworthiness certification.
9652-0096-30 1 Rev. A
The following table lists the cables and accessories that you can use with the ZOLL military M Series CCT unit:

Part Description ZOLL Part Number

(Military Use)
Defib Gel (12-Tubes) 8000-0053
Thermal Paper (1-Pack) 8000-0300
Thermal Paper (10-Pack) 8000-0302
Thermal Paper (20-Pack) 8000-0301
Military 12 lead ECG Cable 8000-0527-30
Military SpO2 cable 8000-0246-30
Commercial SpO2 Adult Sensor 8000-0245
Military EtCO2 Cable 8000-0410
Military AC Power Cord w/Ferrite 8000-0304
Commercial Paddles 8000-1010-01
Military MFC Cable 8000-0308-30
NIBP Hose 8000-0662
NSN Number
Adult NIBP Cuff 8000-0653
Military Temperature Cable 8000-0409
Military IBP Cable 8000-0408
XL Smart Ready Battery 8000-0500-01
Military softcase w/Flap 8000-0785
Xtreme Pack II 8000-0501-01
4 year shelf life DEP Pack 8900-4017-30
Commercial 1x1 Battery Charger 8050-0022-01
Military 4x4 AutoTest Battery Charger 8050-0002-30
Military M Series CCT Unit 43521911100124300
stat padz HVP Multi-Function Elec-1 pair 8900-4004 6515-01-443-0565
stat padz HVP Multi-Function Elec-1 case 8900-4003 6515-01-428-7546
Pro padz Pacing Only Adult Elec-1 case 8900-1055 6515-01-330-0175
9652-0096-30 2 Rev. A
Part Description ZOLL Part Number
(Military Use)
3 lead ECG Electrodes (4 electrodes) 1 case 8900-0014
12 lead ECG Electrodes (6 electrodes) 1 case 8900-0016
Single Use Airway Adaptor 1 box 8000-0260-01
NSN Number
9652-0096-30 3 Rev. A